Search Results
Found 3 results
510(k) Data Aggregation
(96 days)
This software is intended to generate digital radiographic images of the skull, spinal column, extremities, and other body parts in patients of all ages. Applications can be performed with the patient sitting, or lying in the prone or supine position and is intended for use in all routine radiography exams. The product is not intended for mammographic applications.
This software is not meant for mammography, fluoroscopy, or angiography.
The I-Q View is a software package to be used with FDA cleared solid-state imaging receptors. It functions as a diagnostic x-ray image acquisition platform and allows these images to be transferred to hard copy, softcopy, and archive devices via DICOM protocol. The flat panel detector is not part of this submission. In the I-Q View software, the Digital Radiography Operator Console (DROC) software allows the following functions:
-
- Add new patients to the system; enter information about the patient and physician that will be associated with the digital radiographic images.
-
- Edit existing patient information.
-
- Emergency registration and edit Emergency settings.
-
- Pick from a selection of procedures, which defines the series of images to be acquired.
-
- Adiust technique settings before capturing the x-ray image.
-
- Preview the image, accept or reject the image entering comments or rejection reasons to the image. Accepted images will be sent to the selected output destinations.
-
- Save an incomplete procedure, for which the rest of the exposures will be made at a later time.
-
- Close a procedure when all images have been captured.
-
- Review History images, resend and reprint images.
-
- Re-exam a completed patient.
-
- Protect patient records from being deleted by the system.
-
- Delete an examined Study with all images being captured.
-
- Edit User accounts.
-
- Check statistical information.
-
- Image QC.
-
- Image stitching.
-
- Provides electronic transfer of medical image data between medical devices.
The provided document is a 510(k) summary for the I-Q View software. It focuses on demonstrating substantial equivalence to a predicate device through bench testing and comparison of technical characteristics. It explicitly states that clinical testing was not required or performed.
Therefore, I cannot provide details on clinical acceptance criteria or a study proving the device meets them, as such a study was not conducted for this submission. The document relies on bench testing and comparison to a predicate device to establish substantial equivalence.
Here's a breakdown of what can be extracted from the provided text regarding acceptance criteria and the "study" (bench testing) that supports the device:
1. Table of Acceptance Criteria and Reported Device Performance
Since no clinical acceptance criteria or performance metrics are provided, this table will reflect the general statements made about the device performing to specifications.
Acceptance Criteria (Implied) | Reported Device Performance |
---|---|
Device functions as intended for image acquisition. | Demonstrated intended functions. |
Device performs to specification. | Performed to specification. |
Integration with compatible solid-state detectors performs within specification. | Verified integration performance within specification. |
Software is as safe and functionally effective as the predicate. | Bench testing confirmed as safe and functionally effective as predicate. |
2. Sample size used for the test set and the data provenance
- Test Set Sample Size: Not applicable/not reported. The document describes bench testing, not a test set of patient data.
- Data Provenance: Not applicable. Bench testing generally involves internal testing environments rather than patient data from specific countries or retrospective/prospective studies.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts
- Not applicable. As no clinical test set was used, no experts were needed to establish ground truth for patient data. Bench testing typically relies on engineering specifications and verification.
4. Adjudication method for the test set
- Not applicable. No clinical test set or human interpretation was involved.
5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
- No, an MRMC comparative effectiveness study was not done. The document explicitly states: "Clinical Testing: The bench testing is significant enough to demonstrate that the I-Q View software is as good as the predicate software. All features and functionality have been tested and all specifications have been met. Therefore, it is our conclusion that clinical testing is not required to show substantial equivalence." The device is software for image acquisition, not an AI-assisted diagnostic tool.
6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done
- Yes, in a sense. The "study" described is bench testing of the software's functionality and its integration with solid-state detectors. This is an evaluation of the algorithm/software itself.
7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)
- For bench testing, the "ground truth" would be the engineering specifications and expected functional behavior of the software and its interaction with hardware components. It's about verifying that the software performs according to its design requirements.
8. The sample size for the training set
- Not applicable. The I-Q View is described as an image acquisition and processing software, not an AI/machine learning model that typically requires a training set of data.
9. How the ground truth for the training set was established
- Not applicable, as there is no mention of a training set or AI/machine learning component.
Summary of the "Study" (Bench Testing) for K203703:
The "study" conducted for the I-Q View software was bench testing. This involved:
- Verification and validation of the software.
- Demonstrating the intended functions and relative performance of the software.
- Integration testing to verify that compatible solid-state detectors performed within specification as intended when used with the I-Q View software.
The conclusion drawn from this bench testing was that the software performs to specification and is "as safe and as functionally effective as the predicate software." This was deemed sufficient to demonstrate substantial equivalence, and clinical testing was explicitly stated as not required.
Ask a specific question about this device
(34 days)
As a part of a radiographic system, the Philips Eleva Workspot with SkyFlow is intended to acquire, process, store, display, and export digital radiographic images. The Philips Eleva Workspot with SkyFlow is suitable for all routine radiography exams, including specialist areas like intensive care, trauma, or pediatric work, excluding fluoroscopy, angiography and mammography.
The Philips Eleva Workspot with SkyFlow is a workstation (computer keyboard, display, mouse), combined with a flat solid state X-ray detector. It is used by the operator to preset examination data and to generate process and handle digital X-ray images. The Philips Eleva Workspot with SkyFlow will be used as a common software platform in the following currently marketed Philips X-ray systems: Philips Digital Diagnost (K131483 – October 7, 2013), Philips MobileDiagnost (K111725 – July 19, 2011), Philips PCR Eleva (K093355– October 28, 2009), Philips EasyDiagnost Eleva (K031535 – September 6, 2006), and Philips BuckyDiagnost (K945278 – December 29, 1994). As a part of a radiographic system, the Philips Eleva Workspot with SkyFlow is intended to acquire, process, store, display, and export digital radiographic images. The Philips Eleva Workspot with SkyFlow is intended for clinical situations where practitioners deem necessary to remove the anti-scatter grid in critical care departments of hospitals (such as ICU and Emergency), where patients require portable radiographs. Whereas the Pre-Market Notification K140771 of the predicate device limited the indications for use of the SkyFlow option to bedside chest exams only, this Pre-Market Notification covers also other anatomical regions where scattered radiation might have an impact on image quality. There is a standalone version with minimal integration into the X-ray system. With the fully integrated version, the workstation screen also provides displays area and controls for X-ray generator control. The workstation computer can also host parts of the system control software.
The provided text does not contain acceptance criteria or a detailed study of device performance for the Philips Eleva Workspot with SkyFlow.
Instead, the document is a 510(k) summary for the Philips Eleva Workspot with SkyFlow, seeking substantial equivalence to a predicate device (Philips Eleva Workspot, K140771). This type of submission relies heavily on demonstrating that the new device is as safe and effective as an already cleared device, without necessarily requiring new, extensive clinical performance studies.
Here's what can be extracted and why the requested information is largely absent:
1. A table of acceptance criteria and the reported device performance
- Not present. The document states: "The image quality test results were equivalent or better with the modified SkyFlow function turned on than with this function turned off." This is a qualitative statement about an internal test, not quantitative acceptance criteria with reported performance metrics.
2. Sample size used for the test set and the data provenance
- Not present. The document mentions "non-clinical software verification and validation tests" and "image quality test results," but does not specify the sample size of images or the data provenance for these tests. The focus is on the device's technical specifications and adherence to standards rather than diagnostic performance on a patient dataset.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts
- Not present. Since there's no detailed diagnostic performance study described, there's no mention of experts or ground truth establishment in this context.
4. Adjudication method (e.g. 2+1, 3+1, none) for the test set
- Not present. As above, no diagnostic performance study details are provided.
5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
- No. The document explicitly states: "The Philips Eleva Workspot with SkyFlow did not require clinical studies since substantial equivalence to the currently marketed and predicate device was demonstrated with the following attributes: Design features; Indication for use; Fundamental scientific technology; Non-clinical performance testing including validation; and Safety and effectiveness." This indicates that a clinical study, including an MRMC study, was not performed or deemed necessary for this 510(k) clearance.
6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done
- Not explicitly detailed as a diagnostic performance study. The "image quality test results" are a form of standalone testing, but no specific metrics or comparison against a diagnostic ground truth are provided. The device itself is an "Eleva Workspot with SkyFlow," which is a workstation with an image processing function (SkyFlow). It's essentially a standalone image processing system, but its performance is described in terms of image quality rather than diagnostic accuracy.
7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)
- Not present. For a device focused on image acquisition and processing rather than diagnostic interpretation, the "ground truth" for image quality might involve physical phantoms and objective image metrics rather than clinical diagnoses. However, these specifics are not provided.
8. The sample size for the training set
- Not applicable. The document describes "non-clinical software verification and validation tests." There's no mention of an "AI" or machine learning component that would require a "training set" in the conventional sense. The "SkyFlow" is described as a "modified" function, implying a fixed algorithm rather than a learning one.
9. How the ground truth for the training set was established
- Not applicable. See point 8.
In summary: The provided document is a 510(k) summary focused on demonstrating substantial equivalence through technical specifications, compliance with standards, and non-clinical performance testing. It does not contain the details of a clinical performance study with acceptance criteria, human reader studies, or ground truth establishment that would typically be associated with AI-powered diagnostic devices. The "SkyFlow" functionality seems to be an image processing algorithm (likely for scatter correction) that improves image quality, rather than a diagnostic AI that requires a training set and extensive clinical validation against a diagnostic ground truth.
Ask a specific question about this device
(29 days)
As a part of a radiographic system, the Philips Eleva Workspot is intended to acquire, process, store, display, and export digital radiographic images. The Philips Eleva Workspot is suitable for all routine radiography exams, including specialist areas like intensive care, trauma, or pediatric work, excluding fluoroscopy, angiography and mammography.
The Philips Eleva Workspot is a workstation (computer, keyboard, display, mouse), combined with a flat solid state X-ray detector. It is designed to be used with the following set of flat solid state X-ray detectors: Philips Pixium 4600, Philips Wireless Portable Detector FD-W17, Philips Pixium 4343RC. It is used by the operator to preset examination data and to generate process and handle digital X-ray images. The Philips Eleva Workspot will be used as a common software platform in the following currently marketed Philips X-ray systems: Philips Digital Diagnost (K131483 October 7, 2013), MobileDiagnost (K111725 July 19, 2011), Philips PCR Eleva (K093355- October 28, 2009), Philips EasyDiagnost Eleva (K031535 September 6, 2006), and Philips BuckyDiagnost (K945278 December 29, 1994). As a part of a radiographic system, the Philips Eleva Workspot is intended to acquire, process, store, display, and export digital radiographic images. The Philips Eleva Workspot is also intended for clinical situations where physicians decide not to use an anti-scatter grid in situations where patients require bedside chest AP digital radiographs. There is a standalone version with minimal integration into the X-ray system. This standalone version does not connect to a solid state X-ray detector. Instead, it is intended to connect to a Philips PCR x-ray cassette reader. With the fully integrated version, the workstation screen also provides displays area and controls for X-ray generator control. The workstation computer can also host parts of the system control software. The device modification employs an additional software algorithm (referred to as "SkyFlow" in this premarket notification) to post-process digital radiographs that are generated in clinical situations where physicians decide not to use an anti-scatter grid in critical care departments of hospitals such as ICU and Emergency, where patients require bedside chest AP digital radiographs. The software modifications enhance image contrast, producing images that have similar detail contrast as images acquired with an anti-scatter grid. Image quality and detail detectability improvements depend on the clinical task, patient size, anatomical location, and clinical practice. The additional SkyFlow software feature is an optional and reversible image processing option that is not required by the Philips Eleva Workspot to reach its intended use.
The Philips Eleva Workspot is a workstation that acquires, processes, stores, displays, and exports digital radiographic images. The device was cleared under K140771. The 510(k) Summary does not contain a specific section outlining a clinical study to prove the device met acceptance criteria, nor does it present device performance metrics against defined acceptance criteria in a table.
Instead, the submission states that no clinical studies were required. The claim of substantial equivalence to the predicate device (Philips XD-S Direct Radiography Workstation/Package, K063781) was demonstrated through non-clinical performance testing (verification and validation) and compliance with international and FDA-recognized consensus standards (IEC 62304, IEC 62366, and ISO 14971).
The relevant sections from the 510(k) Summary state:
- "The Philips Eleva Workspot did not require clinical studies since Summary of substantial equivalence to the currently marketed and predicate device Clinical Data: was demonstrated with the following attributes: Design features; Indication for use; Fundamental scientific technology; Nonclinical performance testing including validation; and Safety and effectiveness."
- "The results of these tests demonstrate that Philips Eleva Workspot met the acceptance criteria and is adequate for this intended use."
Without a conducted clinical study providing specific performance metrics, the following requested information cannot be directly extracted from the provided text:
- A table of acceptance criteria and the reported device performance: Not available. The document states non-clinical tests "met the acceptance criteria" but does not detail these criteria or performance specifics.
- Sample size used for the test set and the data provenance: Not applicable, as no clinical test set was described.
- Number of experts used to establish the ground truth for the test set and the qualifications of those experts: Not applicable.
- Adjudication method for the test set: Not applicable.
- If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance: Not applicable.
- If a standalone (i.e. algorithm only without human-in-the-loop performance) was done: The document describes a "standalone version with minimal integration into the X-ray system" but does not provide performance data for it. The primary evaluation focused on software verification and validation, implying an algorithm-only assessment of functionality without human interaction metrics.
- The type of ground truth used: Not applicable, as no clinical ground truth was established for performance evaluation.
- The sample size for the training set: Not applicable, as no details about machine learning model training or a training set are provided. The "SkyFlow" algorithm is mentioned as an additional software algorithm for post-processing, but no specifics on its development or training data are given.
- How the ground truth for the training set was established: Not applicable.
Ask a specific question about this device
Page 1 of 1