K Number
K180332

Validate with FDA (Live)

Device Name
17HK700G-W
Manufacturer
Date Cleared
2018-06-08

(122 days)

Product Code
Regulation Number
892.1680
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Flat Panel Digital X-ray Detector 17HK700G-W is indicated for digital imaging solution designed for general radiographic system for human anatomy. It is intended to replace film or screen based radiographic systems in all general purpose diagnostic procedures. Not to be used for mammography.

Device Description

The 17HK700G-W is the solid state x-ray imager, which can generate radiographic images of any part of the body. These devices intercept x-ray photons and the scintillator (CsI:TI) emits visible spectrum photons that illuminate an array of photo-detectors that create an electrical signals. After the electrical signals are generated, it is converted to digital value, and the images are displayed on monitors. The digital value can be communicated to the operator console via wiring connection.

The 17HK700G-W consists of the following components: Flat Panel Detector, Control Box, Calibration Software and power cord and cables. The 17HK700G-W can be used for general X-ray system excluding fluoroscopic, angiographic, and mammographic applications.

AI/ML Overview

The provided text describes a 510(k) premarket notification for a Flat Panel Digital X-ray Detector (17HK700G-W) and its substantial equivalence to a predicate device (S4343-W of GC85A). However, it does not contain the detailed information necessary to fully answer all aspects of your request regarding acceptance criteria and the comprehensive study that proves the device meets those criteria.

Specifically, the document primarily focuses on demonstrating substantial equivalence to an existing predicate device rather than outright proving performance against specific acceptance criteria for a novel AI/Algorithm-driven diagnostic device.

Here's what can be extracted and what information is missing:

Missing Information (Crucial for a typical AI/Algorithm device study):

  • A formal table of acceptance criteria for specific diagnostic performance metrics (e.g., sensitivity, specificity, AUC) for the device's intended use.
  • Detailed results of a clinical study demonstrating the device's diagnostic performance against these acceptance criteria. The document states "Clinical data has been provided... but provided further evidence in addition to the laboratory performance data to show that the device works as intended," which suggests a supportive rather than a primary diagnostic efficacy study.
  • Sample size used for a test set in a diagnostic performance study.
  • Provenance of clinical data (country, retrospective/prospective).
  • Number of experts, their qualifications, and adjudication methods for establishing ground truth for a diagnostic test set.
  • Whether a Multi-Reader Multi-Case (MRMC) comparative effectiveness study was done, and if so, the effect size.
  • Information on a standalone (algorithm only) performance study against diagnostic ground truth.
  • The type of ground truth used for diagnostic performance (e.g., pathology, outcomes data).
  • Sample size for the training set and how its ground truth was established.

Information Extracted from the Document (related to device performance and testing, but not a full diagnostic study):

1. Table of Acceptance Criteria (as reported or inferred from performance tests):

The document presents technical performance characteristics and non-clinical test summaries rather than diagnostic acceptance criteria. The device is a digital X-ray detector, and its performance is compared to a predicate device based on imaging quality metrics.

Criteria CategorySpecific MetricPredicate Device PerformanceProposed Device PerformanceAcceptance Criteria (Implicit for Substantial Equivalence)
Technical PerformanceHigh Contrast Limiting Resolution (LP/mm)3.573.6Comparable to predicate (demonstrated as similar or better)
DQE @0.1lp/mmTyp. 73%Typ. 72%Comparable to predicate (demonstrated as similar or better)
MTF @0.5lp/mmTyp. 84%Typ. 89%Comparable to predicate (demonstrated as similar or better)
Resolution3.57lp3.6lpComparable to predicate (demonstrated as similar or better)
Safety & CompatibilityElectrical Safety Standard (AAMI ES60601-1)N/ACompliantCompliance Required (Demonstrated)
Electromagnetic Compatibility (IEC 60601-1-2)N/ACompliantCompliance Required (Demonstrated)
Software ValidationSoftware Level of Concern: MODERATEN/AValidatedCompliance with FDA guidance (Demonstrated)
Imaging PerformanceAccording to IEC 62220-1 (Detective Quantum Efficiency - DQE)N/AConductedConducted and results support substantial equivalence

2. Sample Size and Data Provenance:

  • Test Set Sample Size: Not explicitly stated for a diagnostic efficacy test. The "clinical data" mentioned is described as "further evidence" rather than a primary diagnostic study with a defined test set.
  • Data Provenance: Not specified (e.g., country of origin, retrospective or prospective).

3. Number of Experts and Qualifications for Ground Truth:

  • Not applicable/mentioned in the context of a diagnostic performance study. The ground truth for this device's validation primarily relies on physics-based imaging performance metrics and engineering standards adherence, as well as comparison to a predicate device.

4. Adjudication Method:

  • Not applicable/mentioned for a diagnostic performance study.

5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study:

  • Not performed based on the provided text. This is a detector, not an AI-driven image interpretation algorithm that assists human readers.

6. Standalone Performance (Algorithm Only):

  • Not applicable in the typical sense of an AI algorithm. The device itself (the detector) is the "standalone" component being evaluated for its image acquisition performance, which is measured through metrics like DQE, MTF, and resolution.

7. Type of Ground Truth Used:

  • For the non-clinical performance tests, the "ground truth" refers to established physical standards and measurement techniques for X-ray detector performance (e.g., IEC 62220-1 for DQE).
  • For the "clinical data" mentioned, the specific type of ground truth is not detailed, but it's presented as supportive evidence for the device working as intended, not as a primary diagnostic accuracy study against a clinical ground truth (like pathology or clinical outcomes).

8. Sample Size for the Training Set:

  • Not applicable/mentioned. This document is about a hardware device (X-ray detector), not an AI algorithm that requires a training set.

9. How Ground Truth for Training Set was Established:

  • Not applicable.

Summary based on the document:

This 510(k) submission for the LG 17HK700G-W Flat Panel Digital X-ray Detector is primarily focused on demonstrating substantial equivalence to an existing predicate device (Samsung S4343-W). The "acceptance criteria" can be inferred as successfully demonstrating that the new device's technical specifications and safety/performance (via non-clinical tests and supportive clinical data) are comparable to or better than the predicate, and that it meets general regulatory standards for X-ray equipment. The document does not describe a clinical diagnostic performance study with specific diagnostic outcomes or reader studies often seen for AI/Algorithm-based devices.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logos of the Department of Health & Human Services and the U.S. Food & Drug Administration (FDA). The Department of Health & Human Services logo is on the left, and the FDA logo is on the right. The FDA logo is a blue square with the letters "FDA" in white, followed by the words "U.S. Food & Drug Administration" in blue.

LG Electronics % Jinhwan Jun Chief Research Engineer 222, Lg-ro, Cheongho-ri Jinwi-myeon Pyeongtaek-si, Gyeonggi-do 17709 SOUTH KOREA

June 8th, 2018

Re: K180332

Trade/Device Name: 17HK700G-W Regulation Number: 21 CFR 892.1680 Regulation Name: Stationary x-ray system Regulatory Class: II Product Code: MOB Dated: May 11, 2018 Received: May 16, 2018

Dear Jinhwan Jun:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

{1}------------------------------------------------

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Jeff Bodge

for Robert Ochs, Ph.D. Director Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known)

K180332

Device Name 17HK700G-W

Indications for Use (Describe)

The Flat Panel Digital X-ray Detector 17HK700G-W is indicated for digital imaging solution designed for general radiographic system for human anatomy. It is intended to replace film or screen based radiographic systems in all general purpose diagnostic procedures. Not to be used for mammography.

Type of Use (Select one or both, as applicable)

X Prescription Use (Part 21 CFR 801 Subpart D)

| Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Image /page/3/Picture/0 description: The image shows the logo for LG, a South Korean multinational electronics company. The logo consists of two parts: a stylized image of a face within a circle and the text "LG Life's Good" to the right. The face is formed by the letters "L" and "G", and the circle is a deep red color. The text "LG Life's Good" is in a dark gray color.

510(k) Summary

[As Required by 21 CFR 807.92]

1. Date Prepared [21 CFR 807.92(a)(a)]

January 31, 2018

2. Submitter's Information [21 CFR 807.92(a)(1)]

Name of Sponsor:- Address:LG Electronics222, LG-ro, Cheongho-ri, Jinwi-myeon, Pyeongtaek-si,Gyeonggi-do, Republic of Korea
Contact Name:- Telephone No.:- Email Address:Jinhwan Jun / Chief Research Engineer+82-31-8066-5641jinhwan.jun@lge.com
Name of Manufacturer:- Address:Same as sponsorSame as sponsor

3. Trade Name, Common Name, Classification [21 CFR 807.92(a)(2)]

Trade Name17HK700G-W
Common NameFlat Panel Digital X-ray Detector
Device Classification NameSolid State X-Ray Imager
Regulation Number21 CFR 892.1680
Classification Product CodeMQB
Device ClassII
510k Review PanelRadiology

{4}------------------------------------------------

Image /page/4/Picture/0 description: The image shows the logo for LG, a South Korean multinational electronics company. The logo consists of two parts: a stylized image of a face inside a circle and the company's slogan, "Life's Good." The face is formed by the letters "L" and "G," with the "L" serving as the nose and the "G" forming the outline of the face. The color scheme is primarily red and gray.

4. Identification of Predicate Device(s) [21 CFR 807.92(a)(3)]

The identified predicate devices within this submission are shown as follow:

Predicate Device

510(k) Number:K150165
• Applicant:Samsung Electronics Co., Ltd.
• Trade Name:S4343-W (of GC85A)
• Common Name:Digital Diagnostic X-ray System
• Classification Name:Stationary System, X-ray
• Regulation Number:21 CFR 892.1680
• Classification Product Code:KPR
• Device Class:II
• 510(k) Review Panel:Radiology

5. Description of the Device [21 CFR 807.92(a)(4)]

The 17HK700G-W is the solid state x-ray imager, which can generate radiographic images of any part of the body. These devices intercept x-ray photons and the scintillator (CsI:TI) emits visible spectrum photons that illuminate an array of photo-detectors that create an electrical signals. After the electrical signals are generated, it is converted to digital value, and the images are displayed on monitors. The digital value can be communicated to the operator console via wiring connection.

The 17HK700G-W consists of the following components: Flat Panel Detector, Control Box, Calibration Software and power cord and cables. The 17HK700G-W can be used for general X-ray system excluding fluoroscopic, angiographic, and mammographic applications.

6. Indications for Use [21 CFR 807.92(a)(5)]

The Flat Panel Digital X-ray Detector 17HK700G-W is indicated for digital imaging solution designed for general radiographic system for human anatomy. It is intended to replace film or screen based radiographic systems in all general purpose diagnostic procedures. Not to be used for mammography.

{5}------------------------------------------------

Image /page/5/Picture/0 description: The image shows the logo for LG, a South Korean multinational electronics company. The logo consists of two parts: a stylized face formed by the letters 'L' and 'G' inside a circle, and the text 'LG Life's Good' to the right. The face is in a shade of red, while the letters 'LG' and the slogan 'Life's Good' are in a dark gray color. The logo is simple, modern, and recognizable, reflecting the company's brand identity.

7. Technological Characteristics (Equivalence to Predicate Device) [21 CFR 807.92(a)(6)]

There are no significant differences in the technological characteristics of these devices compared to the predicate devices which adversely affect safety or effectiveness. Provided below is a table summarizing and comparing the technological characteristics of the 17HK700G-W and the predicate devices:

Proposed DevicePredicate Device
K NumberNot knownK150165
ManufacturerLG ElectronicsSamsung Electronics Co., Ltd.
Trade Name17HK700G-WS4343-W (of GC85A)
Common NameFlat Panel Digital X-ray DetectorDigital Diagnostic X-ray System
Product CodeMQBKPR
Regulation Number21 CFR 892.168021 CFR 892.1680
510(k) ReviewPanelRadiologyRadiology
Indications for UseThe Flat Panel Digital X-rayDetector 17HK700G-W is indicatedfor digital imaging solution designedfor general radiographic system forhuman anatomy. It is intended toreplace film or screen basedradiographic systems in all generalpurpose diagnostic procedures. Notto be used for mammography.The GC85A Digital X-ray ImagingSystem is intended for use ingenerating radiographic images ofhuman anatomy by a qualified/traineddoctor or technician. This device is notintended for mammographicapplications.
Detector
ScintillatorCsICsI
Imaging Area17 x 17 inches17 x 17 inches
Pixel Matrix3,072 x 3,072 pixels3,036 x 3,040 pixels
Pixel Pitch140 um140 µm
High ContrastLimiting Resolution(LP/mm)3.63.57
CommunicationWiredWired/Wireless
DQETyp.72% @0.1lp/mmTyp.73% @0.1lp/mm
MTFTyp.89% @0.5lp/mmTyp.84% @0.5lp/mm
Resolution3.6lp3.57lp
Anatomical SitesGeneralGeneral
Exposure ModeManual, Auto(AED)Manual, Auto(AED)
Power Supply100-240V~, 50/60Hz100-240V~, 50/60Hz

There are no significant differences between the 17HK700G-W and the predicate device that would adversely affect the use of the product. It is substantially equivalent to these devices in design, function, materials, operational principles and intended use. The proposed device, 17HK700G-W has been tested about electrical safety, EMC and performance, and the software has been validated. In addition, the clinical data has been provided to support the substantial equivalence to the predicate device.

{6}------------------------------------------------

Image /page/6/Picture/0 description: The image shows the logo for LG. The logo consists of two parts: a stylized face inside a circle on the left and the letters "LG" in gray on the right. Below the letters "LG" are the words "Life's Good" in a smaller font.

The predicate device is the X-ray system but the proposed device provides only a detector. The proposed device is substantially equivalent to the previously cleared detector which is provided as part of the complete imaging system, K150165.

8. Integration Specifications/Requirements for 17HK700G-W

No.ItemSpecification
1Power frequency50Hz ~ 60Hz
2KV40kVp ~ 150kVp
3mA Range10 to 500mA
4Exposure Time0.001 to 10sec
5mAs Range0.1 to 500mAs
6AccuracykV < ±(1%+1kV), mA < ±(3%+1mA)

The proposed device is provided with the recommended generator specification as follows:

9. Non-Clinical Test Summary [21 CFR 807.92(b)(1)]

1) Electrical Safety and Electromagnetic Compatibility
StandardsNo.StandardsOrganizationStandard TitleVersionPublicationYear
ES60601-1AAMIMedical Electrical Equipment - Part 1: General Requirements For Basic Safety And Essential Performance (IEC 60601-1:2005, MOD)ES60601-1:2005(R)2012 andA1:20122014
60601-1-2IECMedical Electrical Equipment - Part 1-2: General Requirements for Safety – Collateral Standard: Electromagnetic Compatibility - Requirements and Tests60601-1-2Edition 4.02014-022016

2) Software Validation

The 17HK700G-W contains MODERATE level of concern software was designed and developed according to a software development process and was verified and validated.

The software information is provided in accordance with FDA quidance: The content of premarket submissions for software contained in medical devices, on May 11, 2005.

3) Performance Test

Imaging performance test has been conducted according to IEC 62220-1. Medical Electrical Equipment - Characteristics of Digital X-ray Imaging Devices - Part 1-1: Determination of the Detective Quantum Efficiency - Detectors Used in Radiographic Imaging.

{7}------------------------------------------------

Image /page/7/Picture/0 description: The image shows the logo for LG, a South Korean multinational electronics company. The logo consists of two parts: a stylized image of a face inside a circle and the letters "LG" in a bold, sans-serif font. Below the letters "LG" is the slogan "Life's Good" in a smaller, italicized font. The face and circle are in a shade of red, while the letters and slogan are in gray.

10. Clinical Test Summary [ 21 CFR 807.92(b)(2)]

Clinical data has been provided according to FDA quidance document "Guidance for the Submission of 510(k)s for Solid Sate X-ray Imaging Devices". The data was not necessary to establish substantial equivalence based on the modifications to the device but provided further evidence in addition to the laboratory performance data to show that the device works as intended.

11. Substantial Equivalence [21 CFR 807.92(b)(1) and 807.92]

There are no significant differences between 17HK700G-W and the predicate devices, K150165 that would adversely affect the use of the product. It is substantially equivalent to these devices in indications for use and technology characteristics.

12. Conclusion [21 CFR 807.92(b)(3)]

In according with the Federal Food & Drug and cosmetic Act, 21 CFR Part 807, and based on the information provided in this premarket notification LG Electronics, concludes that the 17HK700G-W is substantially equivalent in safety and effectiveness to the predicate devices as described herein.

§ 892.1680 Stationary x-ray system.

(a)
Identification. A stationary x-ray system is a permanently installed diagnostic system intended to generate and control x-rays for examination of various anatomical regions. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.(b)
Classification. Class II (special controls). A radiographic contrast tray or radiology diagnostic kit intended for use with a stationary x-ray system only is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.