510(k) Data Aggregation

    K Number
    K252585

    Validate with FDA (Live)

    Date Cleared
    2026-01-08

    (146 days)

    Product Code
    Regulation Number
    872.3640
    Panel
    Dental
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    Why did this record match?
    Reference Devices :

    K140091, K181138, K203240, K210826, K231557

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    ZENEX Implant System R-System is indicated for use in partially or fully edentulous mandibles and maxillae, in support of single or multiple unit restorations including; cemented retained, screw retained, or overdenture restorations, and final or temporary abutment support for fixed bridgework. It is intended for delayed loading. Wide Fixture System is intended to be used in the molar region.

    Device Description

    ZENEX Implant System_R-System is a thread type implant made of pure titanium according to ASTM F 67 and supplied sterile, which will be placed in the alveolar bone in order to support or maintain the prosthetic tooth or denture when a patient's teeth are partially or totally lost.

    The fixture's surface is treated with SLA (Sandblasted with Large-grit and Acid-etching).

    There are 2 types of fixtures, and the dimensions are as following:

    NameFixture TypeDiameter (mm)Length (mm)Material
    ZENEX MULTI Fixture∅ 3.85/4.25/4.6/5.05/5.4/5.9/6.757mm/8.5mm/10mm/11.5mm/13mm/15mmTi CP4 (ASTM F67)
    ZENEX PLUS Fixture∅ 3.85/4.25/4.6/5.05/5.47mm/8.5mm/10mm/11.5mm/13mm/15mm

    Tolerance of dimension shall be within ± 1% range.

    The Abutments are below:

    Abutment NameDiameter (∅)AngulationLength or Cuff(mm)Material + Surface Treatment
    Cover Screw∅3.5P/H: 0.4/1.4/2.0Ti-6Al-4V ELI
    Healing Abutment∅4.2/5.2/6.2/7.2P/H: 3/4/5/6/7/9Ti-6Al-4V ELI
    ∅8.0P/H: 3/4/5/6/7
    ∅9.0P/H: 3/4/5
    Cemented Abutment∅4.0/5.0/6.0/7.0G/H: 1.8/2.8/3.8/4.8 P/H: 5.5/7Ti-6Al-4V ELI + TiN Coating
    Angled Abutment∅4.0/5.0/6.0/7.015˚/25˚G/H: 1.8/2.8/3.8/4.8 P/H: 7Ti-6Al-4V ELI + TiN Coating
    Temporary Abutment∅4.0G/H: 0.8/2.8 P/H: 10Ti-6Al-4V ELI
    Ball Abutment∅3.5G/H: 0.8/1.8/2.8/3.8/4.8/5.8Ti-6Al-4V ELI
    Multi Abutment∅4.80˚/17˚/30˚G/H: 1.3/2.3/3.3/4.3Ti-6Al-4V ELI
    CCM Cast Abutment∅4.0G/H: 0.8/2.8 P/H: 10Co-Cr-Mo Alloy
    FreeMilling Abutment∅4.0G/H: 1.8/2.8/3.8 P/H: 9Ti-6Al-4V ELI + TiN Coating
    ∅5.0/6.0/7.0G/H: 1.8/2.8/3.8 P/H: 8
    Abutment Screw∅2.05/2.16.75/7.7Ti-6Al-4V ELI

    Fixture, Cover Screw, Healing Abutment is provided sterile, and valid for 5 years.
    Other abutments are provided non-sterile and sterilized by end users.

    AI/ML Overview

    N/A

    Ask a Question

    Ask a specific question about this device

    K Number
    K242261

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2024-11-06

    (98 days)

    Product Code
    Regulation Number
    872.3640
    Panel
    Dental
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    Why did this record match?
    Reference Devices :

    K203240, K233231, K110955

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Accelx Implant System implants are indicated for use in partially or fully edentulous patients to support maxillary and mandibular single-unit, multiple-unit, and overdenture dental restorations. Accelx Implant System implants are indicated for immediate loading when good primary stability is achieved and with appropriate occlusal loading. Implant diameters larger than 6.0 mm are dedicated for the molar region and are indicated for delayed loading.

    Accelx Implant System implants are compatible with Accelx Implant System abutments.

    Accelx Implant System implants are compatible with MegaGen AnyRidge Internal Implant System titanium abutments as listed below.

    Implant System CompatibilityPlatform Diameter, mmMaximum Angulation
    MegaGen AnyRidge Internal Implant System3.5, 4.0, 4.25, 5.025°

    Accelx Implant System cover screws are compatible with MegaGen AnyRidge Internal Implant System components as listed below.

    Implant System CompatibilityImplant Body Diameter, mmImplant Platform, mm
    MegaGen AnyRidge Internal Implant System4.03.5
    4.43.5
    4.93.5, 4.0
    5.43.5, 4.25
    5.93.5, 4.25
    6.45.0
    6.95.0
    7.45.0
    7.95.0
    8.45.0
    Device Description

    The Accelx Implant System comprises Accelx dental implants, Accelx abutments previously cleared in K203240. The Accelx Implant System is compatible with the MegaGen AnyRidge Internal Implant System, regardless of the implant platform; this compatibility is due to the internal taper connection.

    Accelx dental implants have an internal taper abutment interface connection and an internal hexagonal feature for abutment anti-rotation and instrument. The internal threaded section is for mating to the corresponding Accelx cover screws, abutments and screws previously cleared in K203240, and MegaGen AnyRidge Internal Implant System restorative devices.

    Accelx dental implants are provided in ten body diameters ranging from 4.0 mm to 8.4 mm, and in total lengths ranging from 7.7 mm to 14.2 mm. Implants with a body diameter of 4.0 mm and 4.4 mm have a platform diameter of 3.5 mm. Implants with a body diameter of 4.9 mm have a platform diameter of 3.5 mm or 4.0 mm. Implants with a body diameter of 5.4 mm to 5.9 mm have a platform diameter of 3.5 mm. Implants with a body diameter of 6.4 mm to 8.4 mm have a platform diameter of 5.0 mm. The implants are manufactured from titanium alloy conforming to ASTM F136. The endosseous surface of the implant is aluminum oxide-blasted arom the implant collar to the apex. All Accelx dental implants are intended to be placed 0.5-1.0 mm sub crestal.

    Accelx cover screws are provided with a coronal diameter of 3.5 mm in overall lengths of 5.7 mm, 6.5 mm and gingival heights (cuff heights) of 0.8 mm, 1.6 mm, respectively. The cover screws are manufactured from titanium alloy confirming to ASTM F136 and anodized. The cover screws are single use only. The smallest cover screw is provided sterile within the Subject device implant packaging. They are also provided individually as replacement or alternative parts in the same non-sterile packaging configuration as cleared in K203240.

    AI/ML Overview

    The provided document is a 510(k) premarket notification for the Accelx Implant System. It primarily focuses on demonstrating substantial equivalence to predicate devices based on non-clinical performance data and a comparison of technological characteristics. It does not contain information about clinical studies, AI/algorithm performance, multi-reader multi-case (MRMC) studies, or any form of human-in-the-loop performance evaluation.

    Therefore, I cannot provide a table of acceptance criteria or details about a study proving the device meets those criteria in the context of an AI/ML-driven device, as the submission is for a medical device (dental implant system) that does not incorporate AI/ML.

    However, I can extract the non-clinical acceptance criteria and reported device performance from the document.

    Non-Clinical Acceptance Criteria and Reported Device Performance

    Acceptance Criteria CategorySpecific Criterion / Test Method (as recommended in FDA guidance)Reported Device Performance
    Modified SurfacesEvaluation by scanning electron microscope (SEM) and energy dispersive X-ray spectroscopy (EDS) following media blasting and cleaning."Implant surfaces were evaluated after media blasting and cleaning by scanning electron microscope (SEM) and energy dispersive X-ray spectroscopy (EDS) characterization meeting acceptance criteria."
    BiocompatibilityCytotoxicity testing according to ISO 10993-5. Bacterial endotoxin testing to ANSI/AAMI ST72, USP <161> and USP <85>.Material mediated pyrogenicity per USP <151> Pyrogen Test."Cytotoxicity testing according to ISO 10993-5, bacterial endotoxin to ANSI/AAMI ST72, USP <161> and USP <85> and material mediated pyrogenicity per USP <151> Pyrogen Test was performed on representative devices." (Implies successful completion as per general context of 510(k) submission for substantial equivalence).
    SterilizationGamma sterilization validation for implants according to ISO 13004.Moist heat (steam) sterilization for non-sterile cover screws according to ISO 17665-2."Gamma sterilization validation was performed according to ISO 13004 was performed for implants; moist heat (steam) sterilization according to ISO 17665-2 for non-sterile cover screws." (Implies successful validation as per general context of 510(k) submission for substantial equivalence).
    Magnetic Resonance Imaging (MRI) SafetyNon-clinical worst-case MRI review performed using scientific rationale and published literature (e.g., Woods, Terry O., Jana G. Delfino, and Sunder Rajan, "Assessment of Magnetically Induced Displacement Force and Torque on Metal Alloys Used in Medical Devices." Journal of Testing and Evaluation 49.2 (2019): 783-795).Rationale addressed parameters per the FDA guidance "Testing and Labeling Medical Devices for Safety in the Magnetic Resonance (MR) Environment," including magnetically induced displacement force and torque."Non-clinical worst-case MRI review was performed to evaluate the subject device components in the MRI environment using scientific rationale and published literature... rationale addressed parameters per the FDA guidance Testing and Labeling Medical Devices for Safety in the Magnetic Resonance (MR) Environment, including magnetically induced displacement force and torque." (Implies favorable assessment supporting MRI safety, likely "MR Conditional" labeling although not explicitly stated here for all components). The document also states the results demonstrate the device is suitable for intended use.
    Reverse EngineeringTo demonstrate physical compatibility with MegaGen AnyRidge Internal Implant connection. Testing performed on OEM implant bodies, OEM abutments, and OEM abutment screws."Reverse engineering of the MegaGen AnyRidge Internal Implant connection was performed to demonstrate physical compatibility. This testing was performed on OEM implant bodies, OEM abutments, and OEM abutment screws." (Implies successful demonstration of physical compatibility).
    Mechanical PerformanceMechanical performance testing according to ISO 14801:2016, "Dentistry – Implants – Dynamic loading test for Endosseous Dental Implants."Worst-case scenario chosen based on FDA guidance document "Root Form Endosseous Dental Implants and Endosseous Dental Implant Abutments.""The fatigue limit data demonstrated that constructs of the previously cleared Accelx abutments (K203240) and previously cleared MegaGen AnyRidge Internal Implant System OEM abutments in combination with Accelx implants have sufficient strength for their intended use."

    Since the device described is not an AI/ML device, the following points are not applicable and thus, the document does not contain information on them:

    1. Sample sizes used for the test set and data provenance (e.g., country of origin of the data, retrospective or prospective): No clinical test set data from patients is referenced. The "test set" here refers to the physical units tested in non-clinical lab settings.
    2. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g., radiologist with 10 years of experience): Not applicable, as no ground truth for image interpretation or diagnosis is relevant here. The ground truth for mechanical testing is established by engineering standards and measurements.
    3. Adjudication method (e.g., 2+1, 3+1, none) for the test set: Not applicable.
    4. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance: Not applicable.
    5. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done: Not applicable.
    6. The type of ground truth used (expert consensus, pathology, outcomes data, etc): Ground truth is based on engineering specifications, material science standards (ASTM, ISO), and validated test methods for physical properties.
    7. The sample size for the training set: Not applicable, as there is no AI/ML model for training.
    8. How the ground truth for the training set was established: Not applicable.
    Ask a Question

    Ask a specific question about this device

    K Number
    K232750

    Validate with FDA (Live)

    Date Cleared
    2023-12-08

    (91 days)

    Product Code
    Regulation Number
    872.3630
    Panel
    Dental
    Age Range
    All
    Predicate For
    N/A
    Why did this record match?
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The BTI Interna Prosthetic Components are intended to function in the mandible or maxilla to support single and multiple-unit temporary or definitive restorations on the BTI Interna® Dental Implant System.

    All digitally designed zirconia components for use with Aesthetic post abutments are to be sent to a BTI validated milling center for manufacture.

    Device Description

    The purpose of this 510(k) is to expand the BTI product offering and include a variety of abutments compatible with the already cleared BTI Dental Implants. Subject device abutments include 15° Angled abutments, Transepithelial abutments, Temporary titanium abutments, Healing caps, Aesthetic post abutments and Screws. Aesthetic post abutments include Square abutments and Aesthetic interfaces for Transepithelials, as the bottom part of a two-piece abutment. A zirconia superstructure fabricated through CAD-CAM technology is the upper part of the two-piece abutment.

    The subject 15º Angled Titanium abutments are designed to a BTI Interna Narrow implant, to hold single and multiple cement-retained restorations. The compatible BTI implants have been previously cleared in K211952. The abutments are offered with a 15º angulation and gingival heights of 3.0, 4.0, and 5.0 mm. They are manufactured in commercially pure titanium, with a TiN coating to enhance abutment aesthetic appearance.

    The subject Temporary abutments are designed for single-unit screw-retained restorations on Transepithelial abutments are offered in prosthetic diameters of 3.5, 5.0 and 5.5 mm, with no angulation. They are manufactured from commercially pure titanium.

    The subject Transepithelial abutments are designed for single-unit and multi-unit screw-retained restorations on previously cleared BTI implants. The compatible BTI implants have been previously cleared in K211952, K202825, K173257, and K151391. They are available in an engaging and a non-enqaging connection. They are offered in a prosthetic platform diameter of 3.5, 5.0 and 5.5 mm, and gingival heights ranging from 0.5 to 5.0 mm, with 0°, 17° and 30° angulation. They are manufactured from commercially pure titanium, with a TiN coating on abutment surface. The retention screw is provided with the abutment and is manufactured from titanium alloy with a DLC coating.

    The subject Healing caps are designed to be connected to a Transepithelial abutment implant during the healing period, between Transepithelial placement and final dental restoration placement. These abutments are provided in prosthetic diameters of 3.5 and 5.0 mm. They are manufactured from commercially pure titanium and anodized.

    The subject Aesthetic interfaces are two-piece abutments composed of a bottom half titanium component, and a patient-specific designed CAD-CAM zirconia superstructure as a top half component. The Aesthetic interfaces are to be attached to a Transepithelial abutment to hold single or multi-unit restorations. They are fabricated from commercially pure titanium, with a TiN coating to enhance abutment aesthetic appearance.

    The design parameters for this top-half abutment fabrication are as follows:

    • Minimum wall thickness: 0.4 mm -
    • Minimum post height for single-unit restorations: 4.0 mm -
    • Minimum gingival height: 0 mm in the zirconia superstructure -
    • -Maximum gingival height: 6.0 mm
    • -Maximum angulation: 0°

    The subject Square abutments are two-piece abutments, composed of a bottom half titanium component, and a patient-specific designed CAD-CAM zirconia superstructure as a top half component. The Square abutments are to be attached to a BTI implant to hold single or multi-unit restorations. The compatible BTI implants have been previously cleared in K211952, K202825, K173257, and K151391. They are fabricated from commercially pure titanium, with a TiN coating to enhance abutment aesthetic appearance. The subject Square abutments are straight abutments (0%, and are provided in a variety of gingival heights ranging from 0.5 mm to 3.0 mm.

    The design parameters for this top-half abutment fabrication are as follows:

    • Minimum wall thickness: 0.4 mm -
    • -Minimum post height for single-unit restorations: 4.0 mm
    • Minimum qingival height: 0 mm in the zirconia superstructure -
    • Maximum gingival height: 6.0 mm -

    The zirconia superstructures for use with the subject Square abutments and Aesthetic Interfaces will be made at a BTI validated milling center and the material will conform to ISO 13356. The bonding cement recommended for the zirconia superstructure is Multilink Hybrid Abutment Cement (Ivoclar Vivadent AG), cleared in K130436.

    The subject Screws are designed to attach Square abutments and Aesthetic Interfaces to the compatible BTI dental implant or Transepithelial abutment, respectively. The Screws are manufactured from titanium allov, with a DLC coating on screw thread. One subject Screw is also anodized for ease identification.

    AI/ML Overview

    I am sorry, but the provided text does not contain the necessary information to describe the acceptance criteria and the study proving the device meets them, as it is a 510(k) summary for dental prosthetic components. This type of document focuses on demonstrating substantial equivalence to previously cleared devices rather than on performance studies with specific statistical acceptance criteria for a device involving AI or software.

    The document discusses:

    • Device Name: BTI Interna Prosthetic Components
    • Regulatory Information: Class II device, product code NHA, regulation number 21 CFR 872.3630 (Endosseous Dental Implant Abutment).
    • Indications for Use: To function in the mandible or maxilla to support single and multiple-unit temporary or definitive restorations on the BTI Interna® Dental Implant System. It also mentions that digitally designed zirconia components are to be sent to a BTI validated milling center for manufacture.
    • Predicate Devices: Lists several predicate and reference devices, primarily other BTI Interna components and systems, and components from other manufacturers like AccelX Abutments and Medentika GmbH.
    • Device Description: Detailed descriptions of various abutment types (15° Angled, Transepithelial, Temporary titanium, Healing caps, Aesthetic post abutments, and Screws), their materials (commercially pure titanium, zirconia), coatings (TiN, DLC), and design parameters (e.g., wall thickness, post height, gingival height, angulation).
    • Performance Data (Non-clinical):
      • Moist heat sterilization validation.
      • Biocompatibility testing (ISO 10993-5 and ISO 10993-12).
      • Mechanical testing (ISO 14801).
      • Coating characterization.
      • Non-clinical worst-case MRI analysis.
      • Crucially, it explicitly states: "No clinical data were included in this submission."
    • Technological Characteristics and Substantial Equivalence Discussion: A comparative analysis of the subject device with predicate and reference devices regarding indications for use, materials, surface treatment, restoration types, fixation methods, compatible implant platforms, dimensions (gingival height, angulation, prosthetic diameter), sterilization status, and packaging.

    Why the requested information cannot be found in the provided text:

    The request is for information typically found in the clinical study section of a submission for a device, especially those involving AI or software, where performance against a ground truth is evaluated. The provided document is a 510(k) summary for implant prosthetic components, which are mechanical devices, not AI/software. Therefore, concepts like "acceptance criteria" related to AI performance metrics (e.g., sensitivity, specificity), test set sample sizes, data provenance, expert ground truth establishment, adjudication methods, MRMC studies, standalone algorithm performance, or training set details are not applicable and not present in this document.

    The "Performance Data" section in this document refers to non-clinical bench testing (sterilization, biocompatibility, mechanical strength, material characterization, and MRI compatibility) to demonstrate the physical and material properties of the dental components, and it explicitly states that no clinical data was included.

    To answer your request, a different type of FDA submission document (e.g., one for an AI/ML-driven diagnostic or screening tool) would be required.