K Number
K231827

Validate with FDA (Live)

Date Cleared
2023-09-15

(86 days)

Product Code
Regulation Number
872.3630
Panel
Dental
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The BTI Interna Dental Implant System UnicCa® - Prosthetic Components are intended to function in the mandible or maxilla to support single and multiple-unit temporary or definitive restorations on the BTI Interna® Dental Implant System implants.

All digitally designed zirconia components for use with Square Aesthetic Abutments and Aesthetic Interfaces for Transepithelials are to be sent to a BTI validated milling center for manufacture.

Device Description

The purpose of this submission is to obtain marketing clearance for the subject device abutments, Square Aesthetic Abutments and Aesthetic Interfaces for Transepithelials, as part of a two-piece abutment with a zirconia superstructure fabricated using CAD-CAM technology.

The subject Square Aesthetic abutments are stock manufactured abutments, to be attached directly to BTI dental implants with the Interna implant connection (Narrow, Universal Plus and Wide platforms), to hold single or multiple restorations. The Square Aesthetic abutments are compatible with BTI implants previously cleared in K211952, K202825, K173257, and K151391. The subject Square Aesthetic abutments are provided in a variety of gingival heights ranging from 0.5 mm, with a prosthetic platform diameter ranging from 4.0 mm to 5.9 mm, and a prosthetic post height of 3.5 mm. All versions include anti-rotation indexes to prevent crown or bridge rotation. Engaging and non-engaging designs are available for single and multiple restorations, respectively. The abutments are used with cemented and screw-retained restorations.

The design parameters for the CAD-CAM fabrication of the patient-specific superstructures for use with the Square Aesthetic Abutments are:

minimum wall thickness - 0.45 mm minimum post height for single-unit restoration - 4.2 mm maximum gingival height - 6.0 mm minimum gingival height - 0 mm in the zirconia superstructure maximum angle - 0°, straight only.

The subject Aesthetic Interfaces for Transepithelials are stock manufactured abutments, used to support single or multi-unit definitive prosthetic restorations over BTI Transepthelial Abutments (Unit and Multi-im, respectively). The compatible Transepithelial Abutments were previously cleared in K213106 and the Transepithelial Abutments are compatible with the BTI Interna connection implants (Narrow, Universal, Universal Plus and Wide platforms), previously cleared in K211952, K202825, K173257, and K151391. The previously cleared Transepithelial Abutments have a minimum gingival height of 1.0 mm and the Aesthetic Interfaces for Transepithelials have a minimum gingival height of 0.3 mm; when used together (as required) the minimum gingival height is 1.3 mm. The Aesthetic Interfaces for Transepithelials are provided in a range of prosthetic diameters from 3.5 mm, with a range of post heights from 3.5 mm. The Aesthetic Interfaces for Transepithelials are provided in two configurations, Straight and Expanded configurations have a larger prosthetic platform diameter than the Straight configurations with the same platform diameter. Overall, the prosthetic platform diameters range from 3.5 mm to 6.5 mm.

The design parameters for the CAD-CAM fabrication of the patient-specific superstructures for use with the Aesthetic Interfaces for Transepithelials vary for the minimum wall thickness and prostheight depending on the design of the Aesthetic Interface and the corresponding Transepithelial Abutments. The overall design parameters for the CAD-CAM fabrication of the patient-specific superstructures for use with the Aesthetic Interfaces for Transepithelials are:

minimum wall thickness - 0.4 mm minimum post height for single-unit restoration - 4 mm maximum gingival height - 6.0 mm minimum gingival height - 0 mm in the zirconia superstructure maximum angle - 0°, straight only.

All zirconia superstructures for use with the subject device will be made at a B.T.I. Biotechnology Institute, S.L. validated milling center under FDA quality system regulations, and the material will conform to ISO 13356. All superstructures will be bonded to the abutment using Multilink Hybrid Abutment Cement (Ivoclar AG, cleared in K130436).

All subject device components are manufactured from unalloyed titanium conforming to ASTM F67 and ISO 5832-2 and are provided with a Titanium Nitride (TiN) coating to enhance of the device. All subject device components are used with abutment and prosthetic screws previously cleared in K213106, and K053355.

AI/ML Overview

The provided text is a 510(k) summary for the BTI Dental Implant System UnicCa® - Aesthetic Post Abutments. It primarily focuses on demonstrating substantial equivalence to predicate devices based on design, materials, manufacturing, and intended use, rather than presenting a performance study with acceptance criteria for an AI/ML powered device.

Therefore, the requested information regarding acceptance criteria, study details (sample sizes, data provenance, expert qualifications, adjudication, MRMC studies, standalone performance), and ground truth establishment for a device performance study cannot be extracted from this document, as it describes a dental implant abutment, not a device whose performance is measured through such clinical or AI-centric metrics.

The document states: "No clinical data were included in this submission." This explicitly indicates that the type of study typically associated with performance criteria, and the establishment of ground truth for AI/ML models, was not part of this 510(k) submission.

The "Performance Data" section solely discusses non-clinical data related to:

  • MRI review for components.
  • Moist heat sterilization for non-sterile devices.
  • Biocompatibility testing (cytotoxicity) for the final abutment.

These are standard engineering and material science performance tests for medical devices, not an evaluation of a diagnostic or therapeutic AI/ML device's accuracy or effectiveness in a clinical context.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

9-15-23

B.T.I. Biotechnology Institute, S.L. % Kevin Thomas Vice President & Director of Regulatory Affairs PaxMed International, LLC 12264 El Camino Real, Suite 400 San Diego, California 92130

Re: K231827

Trade/Device Name: BTI Dental Implant System UnicCa® - Aesthetic Post Abutments Regulation Number: 21 CFR 872.3630 Regulation Name: Endosseous Dental Implant Abutment Regulatory Class: Class II Product Code: NHA Dated: June 21, 2023 Received: June 21, 2023

Dear Kevin Thomas:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{1}------------------------------------------------

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely.

Sherrill Lathrop Blitzer

for Andrew Steen Assistant Director DHT1B: Division of Dental and ENT Devices OHT1: Office of Ophthalmic, Anesthesia. Respiratory, ENT and Dental Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K231827

Device Name

BTI Dental Implant System UnicCa® - Aesthetic Post Abutments

Indications for Use (Describe)

The BTI Interna Dental Implant System UnicCa® - Prosthetic Components are intended to function in the mandible or maxilla to support single and multiple-unit temporary or definitive restorations on the BTI Interna® Dental Implant System implants.

All digitally designed zirconia components for use with Square Aesthetic Abutments and Aesthetic Interfaces for Transepithelials are to be sent to a BTI validated milling center for manufacture.

Type of Use (Select one or both, as applicable)

Prescription Use (Part 21 CFR 801 Subpart D) Over-The-Counter Use (21 CFR 801 Subpart C)
---------------------------------------------------------------------------------------------------------------------------------------------------

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

510(k) Summary K231827 B.T.I. Biotechnology Institute, S.L. BTI Dental Implant System UnicCa® - Aesthetic Post Abutments September 11, 2023

ADMINISTRATIVE INFORMATION

Manufacturer NameB.T.I. Biotechnology Institute, S.L.
Leonardo Da Vinci 14
Miñano, Álava, 01510 Spain
Telephone+34 945 297 030
Fax+34 945 297 031
Official ContactJosé Ramón Rivero, Technical Manager
Representative/ConsultantKevin A. Thomas, PhD
Floyd G. Larson, MS, MBA
PaxMed International, LLC
12264 El Camino Real, Suite 400
San Diego, CA 92130
Telephone+1 858-792-1235
Fax+1 858-792-1236
Emailkthomas@paxmed.com
flarson@paxmed.com

DEVICE NAME AND CLASSIFICATION

Trade/Proprietary NameBTI Dental Implant System UnicCa® – Aesthetic Post Abutments
Common NameDental implant abutment
Regulation Number21 CFR 872.3630
Regulation NameEndosseous dental implant abutment
Regulatory ClassClass II
Product CodeNHA
Classification PanelDental
Reviewing OfficeOffice of Health Technology 1 (Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices)
Reviewing DivisionDivision of Dental and ENT Devices

PREDICATE DEVICE INFORMATION

Primary Predicate Device K213106, BTI Interna Dental Implant System UnicCa® - Prosthetic Components, B.T.I. Biotechnology Institute, S.L.

Reference Devices K211952, BTI Interna Narrow/Plus Dental Implant System UnicCa®, B.T.I. Biotechnology Institute, S.L. K212628, DESS Dental Implants, Terrats Medical SL

{4}------------------------------------------------

510(k) Summary Page 2 of 6

INDICATIONS FOR USE STATEMENT

The BTI Interna Dental Implant System UnicCa® - Prosthetic Components are intended to function in the mandible or maxilla to support single and multiple-unit temporary or definitive restorations on the BTI Interna" Dental Implant System implants.

All digitally designed zirconia components for use with Square Aesthetic Abutments and Aesthetic Interfaces for Transepithelials are to be sent to a BTI validated milling center for manufacture.

SUBJECT DEVICE DESCRIPTION

The purpose of this submission is to obtain marketing clearance for the subject device abutments, Square Aesthetic Abutments and Aesthetic Interfaces for Transepithelials, as part of a two-piece abutment with a zirconia superstructure fabricated using CAD-CAM technology.

The subject Square Aesthetic abutments are stock manufactured abutments, to be attached directly to BTI dental implants with the Interna implant connection (Narrow, Universal Plus and Wide platforms), to hold single or multiple restorations. The Square Aesthetic abutments are compatible with BTI implants previously cleared in K211952, K202825, K173257, and K151391. The subject Square Aesthetic abutments are provided in a variety of gingival heights ranging from 0.5 mm, with a prosthetic platform diameter ranging from 4.0 mm to 5.9 mm, and a prosthetic post height of 3.5 mm. All versions include anti-rotation indexes to prevent crown or bridge rotation. Engaging and non-engaging designs are available for single and multiple restorations, respectively. The abutments are used with cemented and screw-retained restorations.

The design parameters for the CAD-CAM fabrication of the patient-specific superstructures for use with the Square Aesthetic Abutments are:

minimum wall thickness - 0.45 mm minimum post height for single-unit restoration - 4.2 mm maximum gingival height - 6.0 mm minimum gingival height - 0 mm in the zirconia superstructure maximum angle - 0°, straight only.

The subject Aesthetic Interfaces for Transepithelials are stock manufactured abutments, used to support single or multi-unit definitive prosthetic restorations over BTI Transepthelial Abutments (Unit and Multi-im, respectively). The compatible Transepithelial Abutments were previously cleared in K213106 and the Transepithelial Abutments are compatible with the BTI Interna connection implants (Narrow, Universal, Universal Plus and Wide platforms), previously cleared in K211952, K202825, K173257, and K151391. The previously cleared Transepithelial Abutments have a minimum gingival height of 1.0 mm and the Aesthetic Interfaces for Transepithelials have a minimum gingival height of 0.3 mm; when used together (as required) the minimum gingival height is 1.3 mm. The Aesthetic Interfaces for Transepithelials are provided in a range of prosthetic diameters from 3.5 mm, with a range of post heights from 3.5 mm. The Aesthetic Interfaces for Transepithelials are provided in two configurations, Straight and Expanded configurations have a larger prosthetic platform diameter than the Straight configurations with the same platform diameter. Overall, the prosthetic platform diameters range from 3.5 mm to 6.5 mm.

The design parameters for the CAD-CAM fabrication of the patient-specific superstructures for use with the Aesthetic Interfaces for Transepithelials vary for the minimum wall thickness and prostheight depending on the design of the Aesthetic Interface and the corresponding Transepithelial Abutments. The overall design parameters for the CAD-CAM fabrication of the patient-specific superstructures for use with the Aesthetic Interfaces for Transepithelials are:

minimum wall thickness - 0.4 mm minimum post height for single-unit restoration - 4 mm maximum gingival height - 6.0 mm minimum gingival height - 0 mm in the zirconia superstructure maximum angle - 0°, straight only.

{5}------------------------------------------------

All zirconia superstructures for use with the subject device will be made at a B.T.I. Biotechnology Institute, S.L. validated milling center under FDA quality system regulations, and the material will conform to ISO 13356. All superstructures will be bonded to the abutment using Multilink Hybrid Abutment Cement (Ivoclar AG, cleared in K130436).

All subject device components are manufactured from unalloyed titanium conforming to ASTM F67 and ISO 5832-2 and are provided with a Titanium Nitride (TiN) coating to enhance of the device. All subject device components are used with abutment and prosthetic screws previously cleared in K213106, and K053355.

PERFORMANCE DATA

Non-clinical data submitted or referenced to demonstrate substantial equivalence included:

  • provided in this submission was non-clinical worst-case MRI review was performed to evaluate the subject device . components in the MR environment using scientific rationale and published literature (T.O. Woods, J.G. Delfino, and S. Raian. "Assessment of Magnetically Induced Displacement Force and Torque on Metal Alloys Used in Medical Devices," Journal of Testing and Evaluation Volume 49, No. 2 (March/April 2021): 783-795), based on the entire system including all variations (all compatible implant bodies, abutments, and fixation screws) and material composition. Rationale addressed parameters per the FDA guidance Testing and Labeling Medical Devices for Safety in the Magnetic Resonance (MR) Environment, including magnetically induced displacement force and torque.
  • . provided in this submission was moist heat sterilization for the subject devices provided non-sterile to the end user, to a sterility assurance level of 10° by the overkill method according to ANSI/AAMI/ISO 17665-1, ANSI/AAMI/ISO TIR 17665-2, and ANSI/AAMI/ISO 14937;
  • provided in this submission was biocompatibility testing according to ISO 10993-5 (cytotoxicity) for the final ● abutment (unalloyed titanium base and bonded zirconia superstructure).

No clinical data were included in this submission.

EQUIVALENCE TO MARKETED DEVICES

The subject device is substantially equivalent in indications and design principles to the primary predicate device and the reference devices listed above. Provided at the end of this summary is a table comparing the Indications for Use Statements and the technological characteristics of the primary predicate device, and the reference devices.

The primary predicate device K213106 is in support of substantial equivalence for the abutment designs, materials, manufacturing, and compatible implants. The reference device K211952 is for the Indications for Use, abutment designs, materials, manufacturing, compatible implants, and sterilization. The reference device K212628 is for the Indications for Use for CAD-CAM workflow, abutment designs, superstructure design and materials, and bonding cement. Minor differences in the IFUS for the subject device, the primary predicate device, and the reference devices do not affect substantial equivalence because all devices have intended use for functional and esthetic rehabilitation of the edentulous mandible or maxilla.

The subject device components are compatible with the same BTI Interna dental implants as the primary predicate K213106 and reference device K211952. The subject device Square Aesthetic Abutments and Aesthetic Interfaces for Transepithelials are provided in the same designs and materials as these devices in K213106 and K211952 (for use with a conventional workflow). The subject devices have the same designs for single-unit and multi-unit restorations (engaging and non-engaging connections), and the same ranges of prosthetic platform diameter and gingival heights.

The subject device Square Aesthetic Abutments and Aesthetic Interfaces for Transepithelials are the apical part of a two-piece abutment and requre a superstructure to complete the abutment. The design limits for the subject device

{6}------------------------------------------------

510(k) Summary Page 4 of 6

digitally designed superstructures are similar to those of the reference device K212628. The digitally designed superstructures for the subject device are to be manufactured at a validated milling center; this requirement is the same as that of the reference device K212628.

The subject device components are manufactured from the same material (unalloyed titanium conforming to ASTM F67 and ISO 5832-2). All subject device abutments are provided with a Titanium Nitride (TiN) coating the aesthetic appearance of the device. This material and TiN coating are identical to the material and coating for the abutments cleared in K213106 and K211952.

All superstructures for use with the subject abutments will be manufactured from ziroonia conforming to ISO 13356. All superstructures will be bonded to the abutment using Multilink Hybrid Abutment (Ivoclar Vivadent AG, cleared in K130436). This zirconia material and bonding cement are the same as those required for the reference device K212628.

The subject device components are provided non-sterile. After fabrication of the zirconia superstructure by the validated milling center and bonding of the superstructure to the final device is to be sterilized by moist heat by the end user. This requirement is the same as that of the reference device K212628.

CONCLUSION

The subject device, the primary predicate device, and the reference devices have the same intended use, have similar technological characteristics, and are materials. The subject device, the primary predicate device, and the reference devices encompass the same of physical dimensions, are packaged in similar materials, and are to be sterilized using similar methods. The data included in this submission demonstrate to the predicate devices listed above.

{7}------------------------------------------------

Table of Substantial Equivalence

Subject DevicePrimary Predicate DeviceReference DeviceReference Device
K231827BTI Dental Implant System UnicCa® – AestheticPost AbutmentsK213106BTI Interna Dental Implant System UnicCa® - ProstheticComponentsK211952BTI Interna Narrow/Plus Dental Implant System UnicCa®K212628DESS Dental Smart Solutions
B.T.I. Biotechnology Institute, S.L.B.T.I. Biotechnology Institute, S.L.B.T.I. Biotechnology Institute, S.L.Terrats Medical SL
Indications for Use StatementThe BTI Interna Dental Implant System UnicCa® - ProstheticComponents are intended to function in the mandible or maxilla tosupport single and multiple-unit temporary or definitive restorations onthe BTI Interna® Dental Implant System implants.All digitally designed zirconia components for use with SquareAesthetic Abutments and Aesthetic Interfaces for Transepithelials are tobe sent to a BTI validated milling center for manufacture.The BTI Interna Dental Implant System UnicCa® - ProstheticComponents are intended to function in the mandible or maxilla tosupport single and multiple-unit temporary or definitive restorations onthe BTI Interna® Dental Implant System.The BTI Dental Implant System UnicCa® for oral implant surgery is tobe used for the partial or total replacement of teeth in edentulatepatients. Once attached to the bone, the implants act as an anchor forvarious fixed or removable prosthetic solutions that can be used toimprove or restore a patient's mastication function.In the case of 5.5 – 6.5mm long UnicCa® implants should be used in atwo-stage surgical procedure. These implants are indicated for delayedloading. These implants are indicated only for straight abutments and tosupport permanently fixed restorations.In the case of Tiny® 3.0 UnicCa® implants: These implants shall beused only to replace maxillary lateral incisors and mandibular lateraland central incisors. Immediate loading is recommended when there isgood primary stability and an appropriate occlusal load.DESS Dental Smart Solutions abutments are intended to be used inconjunction with endosseous dental implants in the maxillary ormandibular arch to provide support for prosthetic restorations.All digitally designed custom abutments for use with DESS Bases orBlanks are to be sent to a Terrats Medical validated milling center formanufacture.For the complete Indications for Use Statement including the list ofcompatible implant systems, see the 510(k) Summary for K212628.
Reason for Predicate DeviceNot applicableIFUS; compatible implants; abutment designs; materials;manufacturingIFUS; compatible implants; abutment designs; materials;manufacturing; sterilizationIFUS for CAD-CAM workflow; abutment designs; bondingcement
Product CodesNHANHADZE, NHANHA
Intended UseFunctional and esthetic rehabilitation of the edentulousmandible or maxillaFunctional and esthetic rehabilitation of the edentulousmandible or maxillaFunctional and esthetic rehabilitation of the edentulousmandible or maxillaFunctional and esthetic rehabilitation of the edentulousmandible or maxilla
Abutment Designs
Square Aesthetic AbutmentsAesthetic Interfaces for TransepithelialsVarious abutment designs includingSquare Aesthetic AbutmentsAesthetic Interfaces for TransepithelialsVarious abutment designs includingSquare Aesthetic AbutmentsAesthetic Interfaces for TransepithelialsVarious abutment designs including multiple Ti Basetype abutment designs
Restoration typesSingle-unit and Multi-unitSingle-unit and Multi-unitSingle-unit and Multi-unitSingle-unit and Multi-unit
Compatible Implant ConnectionsInternalInternalInternalInternal, External
Compatible Implants andImplant platform diameterInterna Narrow, 3.5 mmInterna Universal and Interna Universal Plus, 4.1 mmInterna Wide, 5.5 mmInterna Universal and Interna Universal Plus, 4.1 mmInterna Wide, 5.5 mmInterna Narrow, 3.5 mmInterna Universal Plus, 4.1 mmVarious implants3.0 mm - 6.0 mm (for Ti Base type abutments)
Prosthetic Platform diameterSquare Aesthetic Abutments, 4 mm - 5.9 mmAesthetic Interfaces for Transepithelials, 3.5 mm - 6.5 mm4.1 mm - 6.5 mm3.5 mm - 5.1 mm4.5 mm - 6.5 mm
Gingival Height in AbutmentSquare Aesthetic Abutments, 0.5 mm - 3.5 mmAesthetic Interfaces for Transepithelials, 0.3 mm – 1.1 mmUp to 4 mm maximum0.5 mm - 4.0 mmNot stated in 510(k) Summary
Superstructure design limits
Minimum wall thickness0.4 mm0.4 mm - 0.45 mm
Minimum post height for single-unit restoration4.0 mm – 4.2 mm4.0 mm - 4.2 mm
Maximum gingival height6.0 mm6.0 mm
Minimum gingival height0 mm - in the zirconia superstructure0.5 mm
Maximum angle0°, straight only0°, straight only
Subject DevicePrimary Predicate DeviceReference DeviceReference Device
K231827BTI Dental Implant System UnicCa® – AestheticPost AbutmentsB.T.I. Biotechnology Institute, S.L.K213106BTI Interna Dental Implant System UnicCa® - ProstheticComponentsB.T.I. Biotechnology Institute, S.L.K211952BTI Interna Narrow/Plus Dental Implant System UnicCa®B.T.I. Biotechnology Institute, S.L.K212628DESS Dental Smart SolutionsTerrats Medical SL
Abutment Materials
Abutment BaseUnalloyed titanium, ASTM F67 / ISO 5832-2Unalloyed titanium, ASTM F67 / ISO 5832-2Unalloyed titanium, ASTM F67 / ISO 5832-2Ti-6Al-4V alloy, ASTM F136Co-Cr-Mo alloy, ASTM F1537
Abutment Base coatingTitanium nitride (TiN)Titanium nitride (TiN)Titanium nitride (TiN)None
Superstructure, CAD-CAM workflowZirconia, ISO 13356Zirconia, ISO 13356
Cement, CAD-CAM workflowMultilink Hybrid Abutment Cement(Ivoclar Vivadent AG, cleared in K130436)Multilink Hybrid Abutment Cement(Ivoclar Vivadent AG, cleared in K130436)
How Provided
AbutmentsNon-sterile, requires moist heat sterilizationSterile by gamma irradiationSterile by gamma irradiation, andNon-sterile, requires moist heat sterilizationNon-sterile, requires moist heat sterilization
Usage - All ComponentsSingle patient, single useSingle patient, single useSingle patient, single useSingle patient, single use

{8}------------------------------------------------

Table of Substantial Equivalence

§ 872.3630 Endosseous dental implant abutment.

(a)
Identification. An endosseous dental implant abutment is a premanufactured prosthetic component directly connected to the endosseous dental implant and is intended for use as an aid in prosthetic rehabilitation.(b)
Classification. Class II (special controls). The guidance document entitled “Class II Special Controls Guidance Document: Root-Form Endosseous Dental Implants and Endosseous Dental Implant Abutments” will serve as the special control. (See § 872.1(e) for the availability of this guidance document.)