K Number
K242261

Validate with FDA (Live)

Manufacturer
Date Cleared
2024-11-06

(98 days)

Product Code
Regulation Number
872.3640
Panel
Dental
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

Accelx Implant System implants are indicated for use in partially or fully edentulous patients to support maxillary and mandibular single-unit, multiple-unit, and overdenture dental restorations. Accelx Implant System implants are indicated for immediate loading when good primary stability is achieved and with appropriate occlusal loading. Implant diameters larger than 6.0 mm are dedicated for the molar region and are indicated for delayed loading.

Accelx Implant System implants are compatible with Accelx Implant System abutments.

Accelx Implant System implants are compatible with MegaGen AnyRidge Internal Implant System titanium abutments as listed below.

Implant System CompatibilityPlatform Diameter, mmMaximum Angulation
MegaGen AnyRidge Internal Implant System3.5, 4.0, 4.25, 5.025°

Accelx Implant System cover screws are compatible with MegaGen AnyRidge Internal Implant System components as listed below.

Implant System CompatibilityImplant Body Diameter, mmImplant Platform, mm
MegaGen AnyRidge Internal Implant System4.03.5
4.43.5
4.93.5, 4.0
5.43.5, 4.25
5.93.5, 4.25
6.45.0
6.95.0
7.45.0
7.95.0
8.45.0
Device Description

The Accelx Implant System comprises Accelx dental implants, Accelx abutments previously cleared in K203240. The Accelx Implant System is compatible with the MegaGen AnyRidge Internal Implant System, regardless of the implant platform; this compatibility is due to the internal taper connection.

Accelx dental implants have an internal taper abutment interface connection and an internal hexagonal feature for abutment anti-rotation and instrument. The internal threaded section is for mating to the corresponding Accelx cover screws, abutments and screws previously cleared in K203240, and MegaGen AnyRidge Internal Implant System restorative devices.

Accelx dental implants are provided in ten body diameters ranging from 4.0 mm to 8.4 mm, and in total lengths ranging from 7.7 mm to 14.2 mm. Implants with a body diameter of 4.0 mm and 4.4 mm have a platform diameter of 3.5 mm. Implants with a body diameter of 4.9 mm have a platform diameter of 3.5 mm or 4.0 mm. Implants with a body diameter of 5.4 mm to 5.9 mm have a platform diameter of 3.5 mm. Implants with a body diameter of 6.4 mm to 8.4 mm have a platform diameter of 5.0 mm. The implants are manufactured from titanium alloy conforming to ASTM F136. The endosseous surface of the implant is aluminum oxide-blasted arom the implant collar to the apex. All Accelx dental implants are intended to be placed 0.5-1.0 mm sub crestal.

Accelx cover screws are provided with a coronal diameter of 3.5 mm in overall lengths of 5.7 mm, 6.5 mm and gingival heights (cuff heights) of 0.8 mm, 1.6 mm, respectively. The cover screws are manufactured from titanium alloy confirming to ASTM F136 and anodized. The cover screws are single use only. The smallest cover screw is provided sterile within the Subject device implant packaging. They are also provided individually as replacement or alternative parts in the same non-sterile packaging configuration as cleared in K203240.

AI/ML Overview

The provided document is a 510(k) premarket notification for the Accelx Implant System. It primarily focuses on demonstrating substantial equivalence to predicate devices based on non-clinical performance data and a comparison of technological characteristics. It does not contain information about clinical studies, AI/algorithm performance, multi-reader multi-case (MRMC) studies, or any form of human-in-the-loop performance evaluation.

Therefore, I cannot provide a table of acceptance criteria or details about a study proving the device meets those criteria in the context of an AI/ML-driven device, as the submission is for a medical device (dental implant system) that does not incorporate AI/ML.

However, I can extract the non-clinical acceptance criteria and reported device performance from the document.

Non-Clinical Acceptance Criteria and Reported Device Performance

Acceptance Criteria CategorySpecific Criterion / Test Method (as recommended in FDA guidance)Reported Device Performance
Modified SurfacesEvaluation by scanning electron microscope (SEM) and energy dispersive X-ray spectroscopy (EDS) following media blasting and cleaning."Implant surfaces were evaluated after media blasting and cleaning by scanning electron microscope (SEM) and energy dispersive X-ray spectroscopy (EDS) characterization meeting acceptance criteria."
BiocompatibilityCytotoxicity testing according to ISO 10993-5. Bacterial endotoxin testing to ANSI/AAMI ST72, USP <161> and USP <85>.Material mediated pyrogenicity per USP <151> Pyrogen Test."Cytotoxicity testing according to ISO 10993-5, bacterial endotoxin to ANSI/AAMI ST72, USP <161> and USP <85> and material mediated pyrogenicity per USP <151> Pyrogen Test was performed on representative devices." (Implies successful completion as per general context of 510(k) submission for substantial equivalence).
SterilizationGamma sterilization validation for implants according to ISO 13004.Moist heat (steam) sterilization for non-sterile cover screws according to ISO 17665-2."Gamma sterilization validation was performed according to ISO 13004 was performed for implants; moist heat (steam) sterilization according to ISO 17665-2 for non-sterile cover screws." (Implies successful validation as per general context of 510(k) submission for substantial equivalence).
Magnetic Resonance Imaging (MRI) SafetyNon-clinical worst-case MRI review performed using scientific rationale and published literature (e.g., Woods, Terry O., Jana G. Delfino, and Sunder Rajan, "Assessment of Magnetically Induced Displacement Force and Torque on Metal Alloys Used in Medical Devices." Journal of Testing and Evaluation 49.2 (2019): 783-795).Rationale addressed parameters per the FDA guidance "Testing and Labeling Medical Devices for Safety in the Magnetic Resonance (MR) Environment," including magnetically induced displacement force and torque."Non-clinical worst-case MRI review was performed to evaluate the subject device components in the MRI environment using scientific rationale and published literature... rationale addressed parameters per the FDA guidance Testing and Labeling Medical Devices for Safety in the Magnetic Resonance (MR) Environment, including magnetically induced displacement force and torque." (Implies favorable assessment supporting MRI safety, likely "MR Conditional" labeling although not explicitly stated here for all components). The document also states the results demonstrate the device is suitable for intended use.
Reverse EngineeringTo demonstrate physical compatibility with MegaGen AnyRidge Internal Implant connection. Testing performed on OEM implant bodies, OEM abutments, and OEM abutment screws."Reverse engineering of the MegaGen AnyRidge Internal Implant connection was performed to demonstrate physical compatibility. This testing was performed on OEM implant bodies, OEM abutments, and OEM abutment screws." (Implies successful demonstration of physical compatibility).
Mechanical PerformanceMechanical performance testing according to ISO 14801:2016, "Dentistry – Implants – Dynamic loading test for Endosseous Dental Implants."Worst-case scenario chosen based on FDA guidance document "Root Form Endosseous Dental Implants and Endosseous Dental Implant Abutments.""The fatigue limit data demonstrated that constructs of the previously cleared Accelx abutments (K203240) and previously cleared MegaGen AnyRidge Internal Implant System OEM abutments in combination with Accelx implants have sufficient strength for their intended use."

Since the device described is not an AI/ML device, the following points are not applicable and thus, the document does not contain information on them:

  1. Sample sizes used for the test set and data provenance (e.g., country of origin of the data, retrospective or prospective): No clinical test set data from patients is referenced. The "test set" here refers to the physical units tested in non-clinical lab settings.
  2. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g., radiologist with 10 years of experience): Not applicable, as no ground truth for image interpretation or diagnosis is relevant here. The ground truth for mechanical testing is established by engineering standards and measurements.
  3. Adjudication method (e.g., 2+1, 3+1, none) for the test set: Not applicable.
  4. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance: Not applicable.
  5. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done: Not applicable.
  6. The type of ground truth used (expert consensus, pathology, outcomes data, etc): Ground truth is based on engineering specifications, material science standards (ASTM, ISO), and validated test methods for physical properties.
  7. The sample size for the training set: Not applicable, as there is no AI/ML model for training.
  8. How the ground truth for the training set was established: Not applicable.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logos of the Department of Health & Human Services and the Food and Drug Administration (FDA). The Department of Health & Human Services logo is on the left, and the FDA logo is on the right. The FDA logo includes the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.

November 6, 2024

Biotech Dental LLC % Chris Brown Manager Aclivi. LLC 3250 Brackley Drive Ann Arbor, Michigan 48105

Re: K242261

Trade/Device Name: Accelx Implant System Regulation Number: 21 CFR 872.3640 Regulation Name: Endosseous Dental Implant Regulatory Class: Class II Product Code: DZE, NHA Dated: August 1, 2024 Received: November 5, 2024

Dear Chris Brown:

We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{1}------------------------------------------------

Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).

Your device is also subject to, among other requirements, the Quality System (OS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.

All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-device-advicecomprehensive-regulatory-assistance/unique-device-identification-system-udi-system.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

{2}------------------------------------------------

3

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely.

Andrew I. Steen -S

Andrew I. Steen Assistant Director DHT1B: Division of Dental and ENT Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT, and Dental Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{3}------------------------------------------------

Indications for Use

510(k) Number (if known) K242261

Device Name Accelx Implant System

Indications for Use (Describe)

Accelx Implant System implants are indicated for use in partially or fully edentulous patients to support maxillary and mandibular single-unit, multiple-unit, and overdentions. Accelx Implant System implants are indicated for immediate loading when good primary stability is achieved and with appropriate occlusal loading. Implant diameters larger than 6.0 mm are dedicated for the molar region and are indicated for delayed loading.

Accelx Implant System implants are compatible with Accelx Implant System abutments.

Accelx Implant System implants are compatible with MegaGen AnyRidge Internal Implant System titanium abutments as listed below.

Implant System CompatibilityPlatform Diameter, mmMaximum Angulation
MegaGen AnyRidge Internal Implant System3.5, 4.0, 4.25, 5.025°

Accelx Implant System cover screws are compatible with MegaGen AnyRidge Internal Implant components as listed below.

Implant System CompatibilityImplant Body Diameter, mmImplant Platform, mm
MegaGen AnyRidge Internal Implant System4.03.5
4.43.5
4.93.5, 4.0
5.43.5, 4.25
5.93.5, 4.25
6.45.0
6.95.0
7.45.0
7.95.0
8.45.0

Type of Use (Select one or both, as applicable)

X Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

{4}------------------------------------------------

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{5}------------------------------------------------

510(k) Summary K242261 Accelx Implant System November 5, 2024

ADMINISTRATIVE INFORMATION

Manufacturer NameBiotech Dental LLC145 Cedar Ln, Suite 205Englewood, NJ 07631Telephone: +1 (201) 676-2456Fax: n/a
Official ContactDavid Singh, CEO
Email:david.singh@biotechdentalusa.com

DEVICE NAME AND CLASSIFICATION

Trade/Proprietary Name:Accelx Implant System
Common Name:Implant, Endosseous, Root-Form
Regulation Name:Endosseous dental implant
Regulation Number:21 CFR 872.3640
Device Class:Class II
Product Code:DZE, NHA
Review Panel:Dental Products Panel
Reviewing Branch:Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices (OHT1)Dental Devices (DHT1B)

PREDICATE DEVICE INFORMATION

The devices within this submission are substantially equivalent in indications, intended use and technological characteristics to the following Predicate device shares technological characteristics with the following Reference devices.

510(k)Predicate Device NameCompany Name
K123870Xpeed AnyRidge Internal Implant SystemMegaGen Implant Co., Ltd.
510(k)Reference Device NameCompany Name
K203240Accelx AbutmentsBiotech Dental LLC
K233231Dental Implants and AbutmentsDitron Precision Ltd.
K110955AnyRidge Internal Implant SystemMegaGen Implant Co., Ltd.

INDICATIONS FOR USE

Accelx Implant System implants are in partially or fully edentulous patients to support maxillary and mandibular single-unit, multiple-unit, and overdenture dental restorations. Accelx Implants are indicated for immediate loading when good primary stability is achieved and with appropriate occlusal loading. Implant diameters larger than 6.0 mm are dedicated for the molar region and are indicated for delayed loading.

Accelx Implant System implants are compatible with Accelx Implant System abutments.

Accelx Implant System implants are compatible with MegaGen AnyRidge Internal Implant System titanium abutments as listed below.

Implant System CompatibilityPlatform Diameter, mmMaximum Angulation
MegaGen AnyRidge Internal Implant System3.5, 4.0, 4.25, 5.025°

{6}------------------------------------------------

Accelx Implant System cover screws are compatible with MegaGen AnyRidge Internal Implant components as listed below.

Implant System CompatibilityImplant Body Diameter, mmImplant Platform, mm
MegaGen AnyRidge Internal Implant System4.03.5
MegaGen AnyRidge Internal Implant System4.43.5
MegaGen AnyRidge Internal Implant System4.93.5, 4.0
MegaGen AnyRidge Internal Implant System5.43.5, 4.25
MegaGen AnyRidge Internal Implant System5.93.5, 4.25
MegaGen AnyRidge Internal Implant System6.45.0
MegaGen AnyRidge Internal Implant System6.95.0
MegaGen AnyRidge Internal Implant System7.45.0
MegaGen AnyRidge Internal Implant System7.95.0
MegaGen AnyRidge Internal Implant System8.45.0

DEVICE DESCRIPTION

The Accelx Implant System comprises Accelx dental implants, Accelx abutments previously cleared in K203240. The Accelx Implant System is compatible with the MegaGen AnyRidge Internal Implant System, regardless of the implant platform; this compatibility is due to the internal taper connection.

Accelx dental implants have an internal taper abutment interface connection and an internal hexagonal feature for abutment anti-rotation and instrument. The internal threaded section is for mating to the corresponding Accelx cover screws, abutments and screws previously cleared in K203240, and MegaGen AnyRidge Internal Implant System restorative devices.

Accelx dental implants are provided in ten body diameters ranging from 4.0 mm to 8.4 mm, and in total lengths ranging from 7.7 mm to 14.2 mm. Implants with a body diameter of 4.0 mm and 4.4 mm have a platform diameter of 3.5 mm. Implants with a body diameter of 4.9 mm have a platform diameter of 3.5 mm or 4.0 mm. Implants with a body diameter of 5.4 mm to 5.9 mm have a platform diameter of 3.5 mm. Implants with a body diameter of 6.4 mm to 8.4 mm have a platform diameter of 5.0 mm. The implants are manufactured from titanium alloy conforming to ASTM F136. The endosseous surface of the implant is aluminum oxide-blasted arom the implant collar to the apex. All Accelx dental implants are intended to be placed 0.5-1.0 mm sub crestal.

Accelx cover screws are provided with a coronal diameter of 3.5 mm in overall lengths of 5.7 mm, 6.5 mm and gingival heights (cuff heights) of 0.8 mm, 1.6 mm, respectively. The cover screws are manufactured from titanium alloy confirming to ASTM F136 and anodized. The cover screws are single use only. The smallest cover screw is provided sterile within the Subject device implant packaging. They are also provided individually as replacement or alternative parts in the same non-sterile packaging configuration as cleared in K203240.

Non-clinical Performance Data

The Acclex implants and cover screws were evaluated and tested as recommended in the FDA guidance document, Root Form Endosseous Dental Implants and Endosseous Dental Implant Abutments (issued May 12, 2004).

Non-clinical data submitted to demonstrate substantial equivalence included: evaluation of the modified surface of the dental implant, biocompatibility testing according to ISO 1099-1, bacterial endotoxin testing, sterilizations,

{7}------------------------------------------------

shelf-life testing, a magnetic resonance information (MRI) safety assessment, reverse engineering of OEM implants to confirm OEM compatibility, and static and compression fatigue testing according to ISO 14801.

Modified Surfaces

lmplant surfaces were evaluated after media blasting and cleaning by scanning electron microscope (SEM) and energy dispersive X-ray spectroscopy (EDS) characterization meeting acceptance criteria.

Biocompatibility

Cytotoxicity testing according to ISO 10993-5, bacterial endotoxin to ANSI/AAMI ST72, USP <161> and USP <85> and material mediated pyrogenicity per USP <151> Pyrogen Test was performed on representative devices.

Sterilization

Gamma sterilization validation was performed according to ISO 13004 was performed for implants; moist heat (steam) sterilization according to ISO 17665-2 for non-sterile cover screws.

Magnetic Resonance Imaging Safety

Non-clinical worst-case MRI review was performed to evaluate the subject device components in the MRI environment using scientific rationale and published literature (e.g., Woods, Terry O., Jana G. Delfino, and Sunder Rajan, "Assessment of Magnetically Induced Displacement Force and Torque on Metal Alloys Used in Medical Devices." Journal of Testing and Evaluation 49.2 (2019): 783-795), based on the entire system including all compatible implant bodies, abutments, and fixation screws) and material composition. rationale addressed parameters per the FDA guidance Testing and Labeling Medical Devices for Safety in the Magnetic Resonance (MR) Environment, including magnetically induced displacement force and torque.

Reverse Engineering

Reverse engineering of the MegaGen AnyRidge Internal Implant connection was performed to demonstrate physical compatibility. This testing was performed on OEM implant bodies, OEM abutments, and OEM abutment screws.

Mechanical Performance

Mechanical performance testing of the Accelx Implant System was performance to ISO 14801:2016, Dentistry – Implants – Dynamic loading test for Endosseous Dental Implants. The worst-case scenario was chosen based on FDA guidance document Root Form Endosseous Dental Implants and Endosseous Dental Implant Abutments. The fatigue limit data demonstrated that constructs of the previously cleared Accelx abutments (K203240) and previously cleared MegaGen AnyRidge Internal Implant System OEM abutments in combination with Accelx implants have sufficient strength for their intended use.

The results of the non-clinical testing demonstrate that the Subject device is suitable for the intended use and is the same or highly similar to the Predicate device.

No clinical or animal testing data is included in this premarket notification.

{8}------------------------------------------------

Equivalence to Marketed Devices

Overall, the Subject device is substantially equivalent in indications and design principles to the sponsor's Predicate and Reference devices listed above. Provided at the end of this summary are tables comparing the Indications for Use Statements and the technological characteristics of the Subject, Predicate device, and Reference devices.

The Subject device Accelx dental implants are compatible with the Accelx Abutments cleared in K203240 and together complete the Accelx Implant System.

Indications for Use Statement

The Subject device Indications for Use Statement (IFUS) is similar to that of the K123870 Predicate and the K110955 and K233231 Reference devices. Slight differences in the language of the IFUS do not affect the intended use as an endosseous dental implant for use with dental implant abutments and screws for support of a prosthesis to restore chewing function.

Minor differences between the IFUS for the Subject device and the K123870 Predicate include the following: the Subject device IFUS includes the term "single-unit, and overdenture dental restorations" and the Predicate IFUS uses alternative terminology "Crown, bridges, and overdentures"; the Subject device IFUS includes references to compatible OEM implant systems and the Predicate IFUS does not include OEM compatibility; the Subject device IFUS states "Implant diameters larger than 6.0 mm are dedicated for the molar region and are indicated for delayed loading" and the predicate IFUS states more generically "Larger implants are dedicated for the molar region and are indicated for delayed loading"; and differences in tradenames. These same differences apply to comparisons of the Subject device to the K110955 Reference device. These minor differences do not raise new questions of safety or effectiveness because both IFUS express equivalent intended use to provide support for prosthetic devices to restore chewing function, expressed equivalently using different specific wording.

Minor differences between the Subject device and the K233231 Reference device include the following: differences in tradename; the Subject device IFUS includes references to compatible OEM implant systems and the Reference IFUS does not include OEM compatibility; the Reference device IFUS distinguishes models for one stage restorations and the Subject device IFUS distinguishes implant diameters and lengths for immediate and delayed loading; and the Reference IFUS includes indications specific to the use of angled multi-unit abutments and the Subject device IFUS does not. These minor differences do not raise new questions of safety or effectiveness as both IFUS express equivalent intended use to provide support for prosthetic devices to restore chewing function, expressed equivalently using different specific wording.

The Subject device IFUS includes a reference to OEM prosthetic components the Subject device dental implants are compatible with. Unless an endosseous dental implant is a one-part implant including an inherent support structure for a definitive restoration, dental implants are used in conjunction with compatible abutments. Normally these abutments are not specifically listed in the IFUS. Since the Subject device implants are compatible with another OEM abutments, the critical design parameters of those OEM abutments are listed in the IFUS. This minor difference in IFUS language does not raise new questions of safety or effectiveness as the IFUSs express equivalent intended use to provide support for prosthetic devices to restore chewing function, expressed equivalently using different specific wording and is supported through non-clinical bench testing.

Technological Characteristics

Accelx Implants

Predicate Device K123870, Xpeed AnyRidge Internal Implant System

The Subject device Accelx dental implants are similar in designs and sizes to the Xpeed AnyRidge Internal Implant System implants cleared in the K123870 Predicate device implants also the same internal tapered abutment interface connection and internal heature for abutment anti-rotation and instrument attachment.

The Subject device and K123870 Predicate implants are provided in the same body diameters ranging from 4.0 mm to 8.4 mm. Both the Subject device implants and the K123870 Predicate with body diameters of 4.0 mm to 5.9 mm are available with a platform diameter of 3.5 mm. The Subject device and the K123870 Predicate device implants

{9}------------------------------------------------

with body diameters of 4.9 mm are also available with a platform diameter of 4.0 mm. The Subject device and the K123870 Predicate device implants with body diameters of 5.4 and 5.9 mm are also available with a platform diameter of 4.25 mm. Similarly, both the Subject device and K123870 Predicate device implants with body diameters of 6.4 mm to 8.4 mm have a platform diameter of 5.0 mm. The Subject device implant lengths range from 7.7 mm to 14.2 mm and are encompassed by the K123870 Predicate device implants provided in lengths ranging from 6.2 mm . The Subject device implants are fabricated from a similar titanium material and have a similar modified surface treatment as the K123870 Predicate device. The Subject device implants have the same sterilization method and single-patient, single-use usage as the K123870 Predicate device.

Reference Device K203240, Accelx Abutments

The K203240 Reference device is included to support the previously cleared Accelx abutments which are compatible with the Subject device implants.

Reference Device K233231, Dental Implants and Abutments

The Subject device implants are similar in body diameter and length to the K233231 Reference device implants.

The Subject device implants are fabricated from the same ASTM F136 titanium alloy and have the same alumina grit blasting and acid etching surface treatment, and the same sterilization process as the K233231 Reference device implants. The Subject device implants and cover screws have the same single-use usage as the K233231 Reference device. Additionally, both the Subject device and K233231 Reference device cover screws are color anodized.

Reference Device K110955, AnyRidge Internal Implant System

The Subject device cover screws have the same coronal diameter and gingival heights as the K110955 Reference device cover screws and are manufactured from the same ASTM F136 titanium ally, both the Subject device and K110955 Reference device cover screws are color anodized.

For Subject device cover screws provided with the implants, the Subject device cover screws are supplied sterile, the same as the K110955 Reference device. Unlike the K110955 Reference device cover screws supplied as replacements or alternate sizes are not delivered sterile. However, they are the cover screws are intended to be used sterile and the sponsor has provided a validated moist-heat sterilization process in device labeling.

Reverse engineering and compatibility analysis was performed to validate physical compatibility of the Subject device implants with the K123870 Predicate and K110955 Reference device Anyridge Internal Implant System implant/abutment connections. Additionally, fatigue testing according to ISO 14801, Dentistry - Implants - Dynamic loading test for endosseous dental implants was performed to validate compatibility.

Minor differences in technological characteristics the Subject and Reference devices do not affect substantial equivalence. These minor differences do not impact safety or effectiveness are mitigated by bench performance testing.

CONCLUSION

Overall, the Indications for Use statements are similar, differing in specific wording, but have the same intended use to provide prosthetic support for dental restorations. The Subject device additional language regarding compatibility is supported through bench performance testing.

Overall, the Technological Characteristics of the Subject device are highly similar or the same as the Predicate and Reference devices. Any minor differences in Technological Characteristics are supported through bench performance testing.

{10}------------------------------------------------

The Accelx implants and cover screws are substantially equivalent in indications and design principles to the Predicate device and Reference devices listed above. Provided at the end of this summary is a table comparing the Indications for Use Statements and Technological Characteristics of the Accelx Implant System, the Preference devices.

Substantial Equivalence - Indications for Use Statement

Subject DevicePredicate DeviceReference DeviceReference Device
Accelx Implant SystemXpeed AnyRidge Internal Implant SystemDental Implants and AbutmentsAnyRidge Internal Implant System
Biotech Dental LLCMegaGen Implant Co., Ltd.Ditron Precision Ltd.MegaGen Implant Co., Ltd.
K123870K233231K110955
Accelx Implant System implants are indicated for use inpartially or fully edentulous patients to supportmaxillary and mandibular single-unit, multiple-unit, andoverdenture dental restorations. Accelx Implant Systemimplants are indicated for immediate loading whengood primary stability is achieved and with appropriateocclusal loading. Implant diameters larger than 6.0 mmare dedicated for the molar region and are indicated fordelayed loading.The Xpeed AnyRidge Internal ImplantSystem is intended to be surgicallyplaced in the maxillary or mandibularmolar areas for the purpose providingprosthetic support for dentalrestorations (Crown, bridges, andoverdentures) in partially or fullyedentulous individuals. It is used torestore a patient's chewing function.Smaller implants (less than Ø6.0 mm)are dedicated for immediate loadingwhen good primary stability isachieved and with appropriateocclusal loading. Larger implants arededicated for the molar region andare indicated for delayed loading.Ditron's Dental Implants andAbutments are indicated for use insurgical and restorative applicationsfor placement in the bone of the upperor lower jaw to provide support forprosthetic devices, such as artificialteeth, in order to restore the patient'schewing function.• Two stage: MPI, ULT, API and CPImodels• One stage: OPI and TPI modelsThe 3.3 and 3.0mm diameter modelsfor One stage OPI and TPI, Two stageMPI, Two stage and API implants areintended only for the incisors andcuspids of the maxilla and mandible.They are also indicated for denturestabilization using multiple implants.Two stage and One stage implants fortemporary or long-term use: MPI, ULT,API, CPI, OPI and TPI are self-tappingTitanium threaded screws indicatedfor long term intra bony applications.They permit immediate splint stabilityand long-term fixation of new orexisting crown, bridge and prosthesisand protection of graft sites.MPI, ULT, API, CPI, OPI and TPI designsare indicated for immediate loading(except for MPI and API in 6.0mmlength) when good primary stability isachieved and with appropriateocclusal loading.MPI, ULT, API, CPI, OPI and TPI areindicated for immediate loading(except for MPI and API in 6.0mmlength) in single tooth restorationswhen good primary stability isachieved with appropriate occlusalloading.The 30° multi-unit abutments must beused within 45 degrees of parallelismfor a splinted restoration.The 17° multi-unit abutments must beused within 32 degrees of parallelismfor a splinted restorationThe AnyRidge Internal Implant Systemis intended to be surgically placed inthe maxillary or mandibular molarareas for the purpose providingprosthetic support for dentalrestorations (Crown, bridges, andoverdentures) in partially or fullyedentulous individuals. It is used torestore a patient's chewing function.Smaller implants (less than Ø6.0 mm)are dedicated for immediate loadingwhen good primary stability isachieved and with appropriateocclusal loading. Larger implants arededicated for the molar region andare indicated for delayed loading.
Accelx Implant System implants are compatible withAccelx Implant System abutments.Accelx Implant System implants are compatible withMegaGen AnyRidge Internal Implant System titaniumabutments as listed below.
Implant SystemCompatibilityPlatformDiameter, mmMaximumAngulation
MegaGen AnyRidgeInternal ImplantSystem3.5, 4.0, 4.25, 5.025°
Accelx Implant System cover screws are compatiblewith MegaGen AnyRidge Internal Implant Systemcomponents as listed below.
Implant SystemCompatibilityImplant BodyDiameter, mmImplantPlatform, mm
MegaGen AnyRidgeInternal ImplantSystem4.03.5
4.43.5
4.93.5, 4.0
5.43.5, 4.25
5.93.5, 4.25
6.45.0
6.95.0
7.45.0
7.95.0
8.45.0

Table A. Substantial Equivalence - Indications for Use Statement

{11}------------------------------------------------

Comparison of Technological Characteristics

Subject DevicePredicate DeviceReference DeviceReference DeviceReference DeviceComparison
Accelx Implant SystemBiotech Dental LLCXpeed AnyRidge Internal Implant SystemMegaGen Implant Co., Ltd.K123870Accelx AbutmentsBiotech Dental LLCK203240Dental Implants and AbutmentsDitron Precision Ltd.K233231AnyRidge Internal Implant SystemMegaGen Implant Co., Ltd.K110955
Product CodeDZE, NHADZE, NHANHADZE, NHADZE, NHASame
Regulation872.3640872.3640872.3630872.3640872.3640Same
Intended UseFunctional and esthetic rehabilitation of partiallyor fully edentulous maxilla and mandible torestore chewing function.Functional and esthetic rehabilitation of partially or fullyedentulous maxilla and mandible to restore chewingfunction.Functional and esthetic rehabilitation ofpartially or fully edentulous maxilla andmandible to restore chewing function.Functional and esthetic rehabilitation ofpartially or fully edentulous maxilla andmandible to restore chewing function.Functional and esthetic rehabilitation ofpartially or fully edentulous maxilla andmandible to restore chewing function.Same
Reason forPredicate/ReferenceNot ApplicableOEM Abutment Compatibility; Implant DesignsPreviously cleared abutments compatible withSubject device implantsImplant material and surface treatment,anodized cover screwsOEM Compatibility, Cover Screw Designsn/a
Implants
Prosthetic InterfaceConnectionInternal HexInternal Hexn/aInternal Hexn/aSame
Body/PlatformDiameter, mmImplant Body DiameterImplant Body Diametern/aImplant Body Diametern/aSame – PredicateHighly Similar - Reference
4.0, 4.4, 4.9, 5.4, 5.94.0, 4.4, 4.9, 5.4, 5.93.5, 3.75
4.94.94.2, 5.0
6.4, 6.9, 7.4, 7.9, 8.46.4, 6.9, 7.4, 7.9, 8.46.4, 6.9, 7.4, 7.9, 8.4
Body diameter andLengths, mmImplant Body DiameterImplant Body Diametern/aImplant Diametern/aHighly SimilarSubject device dimensionsencompassed by Predicateand Reference Devices
4.04.03.5, 3.75
4.44.44.2, 5.0
4.94.9
5.45.4
6.46.4
6.96.9
7.47.4
7.97.9
8.48.4
LengthsLengthsLengths
7.7, 9.2, 10.7, 12.2,14.26.2, 7.7, 9.2, 10.7,12.2, 14.28.0, 10, 11, 13, 16
6.4, 7.9, 9.4, 10.9, 12.46.4, 7.9, 9.4, 10.9,12.4, 14.48.0, 10, 11, 13, 16
Implant MaterialTitanium Alloy (ASTM F136)CP Ti Grade 4 (ASTM F67)n/aTitanium Alloy (ASTM F136)n/aSimilar - PredicateSame – K233231Reference device
Implant SurfaceTreatmentAlumina Grit blasting and Acid EtchingSand-blasted, Large grit, Acid-etched (SLA)n/aAlumina Grit blasting and Acid EtchingColor anodization of Tissue-Level Neckn/aHighly Similar - PredicateSame – K233231Reference device (bodybelow the neck)
SterilityGamma IrradiationGamma Irradiationn/aGamma Irradiationn/aSame
UsageSingle Patient, Single-UseSingle Patient, Single-Usen/aSingle Patient, Single-Usen/aSame
Cover Screws
MaterialAnodized Titanium Alloy (ASTM F136)n/aAnodized Titanium Alloy (ASTM F136)Anodized Titanium Alloy (ASTM F136)Anodized Titanium Alloy (ASTM F136)Same as K110955Reference Device
Coronal Diameter, mm3.5n/a3.54.83.5Same as K110955Reference Device
Cuff/Gingival Height, mm0.8, 1.6, 2.6n/a0.8, 1.6, 2.63.5, 4.5, 5.50.8, 1.6, 2.6Same as K110955Reference Device
SterilitySupplied with implant - SterileSupplied as Replacement/Alternate - Non-Sterilen/aSupplied with implant - SterileSupplied as Replacement/Alternate - Non-SterileSterileSterileHighly Similar
Sterilization MethodSupplied with implant - Gamma IrradiationSupplied as Replacement/Alternate - Moist Heatn/aSupplied with implant - Gamma IrradiationSupplied as Replacement/Alternate - Moist HeatGamma IrradiationGamma IrradiationSimilar
UsageSingle patient, single usen/aSingle patient, single useSingle Patient, Single-UseSingle patient, single useSame as ReferenceDevices

Table B. Substantial Equivalence – Technological Characteristics

§ 872.3640 Endosseous dental implant.

(a)
Identification. An endosseous dental implant is a prescription device made of a material such as titanium or titanium alloy that is intended to be surgically placed in the bone of the upper or lower jaw arches to provide support for prosthetic devices, such as artificial teeth, in order to restore a patient's chewing function.(b)
Classification. (1) Class II (special controls). The device is classified as class II if it is a root-form endosseous dental implant. The root-form endosseous dental implant is characterized by four geometrically distinct types: Basket, screw, solid cylinder, and hollow cylinder. The guidance document entitled “Class II Special Controls Guidance Document: Root-Form Endosseous Dental Implants and Endosseous Dental Implant Abutments” will serve as the special control. (See § 872.1(e) for the availability of this guidance document.)(2)
Classification. Class II (special controls). The device is classified as class II if it is a blade-form endosseous dental implant. The special controls for this device are:(i) The design characteristics of the device must ensure that the geometry and material composition are consistent with the intended use;
(ii) Mechanical performance (fatigue) testing under simulated physiological conditions to demonstrate maximum load (endurance limit) when the device is subjected to compressive and shear loads;
(iii) Corrosion testing under simulated physiological conditions to demonstrate corrosion potential of each metal or alloy, couple potential for an assembled dissimilar metal implant system, and corrosion rate for an assembled dissimilar metal implant system;
(iv) The device must be demonstrated to be biocompatible;
(v) Sterility testing must demonstrate the sterility of the device;
(vi) Performance testing to evaluate the compatibility of the device in a magnetic resonance (MR) environment;
(vii) Labeling must include a clear description of the technological features, how the device should be used in patients, detailed surgical protocol and restoration procedures, relevant precautions and warnings based on the clinical use of the device, and qualifications and training requirements for device users including technicians and clinicians;
(viii) Patient labeling must contain a description of how the device works, how the device is placed, how the patient needs to care for the implant, possible adverse events and how to report any complications; and
(ix) Documented clinical experience must demonstrate safe and effective use and capture any adverse events observed during clinical use.