510(k) Data Aggregation

    K Number
    K252384

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2025-10-23

    (84 days)

    Product Code
    Regulation Number
    872.3630
    Panel
    Dental
    Age Range
    All
    Predicate For
    N/A
    Why did this record match?
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    DESS® Dental Smart Solutions abutments are intended to be used in conjunction with endosseous dental implants in the maxillary or mandibular arch to provide support for prosthetic restorations.

    All digitally designed custom abutments for use with DESS® Pre-Milled Blank abutments are to be sent to a Terrats Medical validated milling center for manufacture.

    Device Description

    The purpose of this submission is to add abutments to the DESS Dental Smart Solutions system, which includes abutments cleared previously in K242340 and K240208. The subject device abutment designs include Pre-Milled Blank Abutments, Multi-Unit Abutments (straight and angled), and abutment screws.

    Pre-Milled Blank Abutments are designed for custom abutment fabrication by a CAD-CAM process. All patient-specific custom abutment fabrication is by prescription on the order of the clinician. The Pre-Milled Blank Abutments have a maximum (before milling) diameter of 10 mm or 14 mm and a solid cylindrical design with an engaging or non-engaging implant connection. They are available in versions with a straight screw channel and with an angled screw channel for esthetic considerations. The Pre-Milled Blank Abutments are manufactured from titanium alloy (Ti 6Al-4V).

    The design parameters for the CAD-CAM fabrication of a custom abutment from the Pre-Milled Blank Abutment are:

    • Minimum wall thickness – 0.45 mm
    • Minimum cementable post height for single-unit restoration – 4.0 mm (minimum cementable post height for single-unit restoration is defined as the height above the restorative margin)
    • Minimum gingival height – 0.5 mm
    • Maximum gingival height – 6.0 mm
    • Pre-Milled Blanks are for straight abutments only

    Multi-Unit Abutments: Straight and Angled are designed for attachment of multi-unit screw-retained restorations and are provided in three (3) designs, straight (0°), angled 17°, and angled 30°. The designs of the subject Multi-Unit Abutments are similar to the designs of Multi-Unit Abutments cleared in K242340 and K240208. All Multi-Unit Abutments are manufactured from titanium alloy (Ti-6Al-4V).

    The subject device straight Multi-Unit Abutments have a non-engaging, threaded design that attaches directly to the implant. The subject device straight (0°) Multi-Unit Abutments are included in this submission are similar in design to straight Multi-Unit Abutments cleared in K242340, K230143, and K222288. Subject device straight Multi-Unit Abutments are provided with a prosthetic platform diameter of 4.8 mm, and with a gingival height ranging from 1 mm to 5.5 mm.

    The subject device angled Multi-Unit Abutments are provided only in an engaging design that requires an abutment screw. The subject device Multi-Unit Abutments angled 17° and 30° are provided with a prosthetic platform diameter of 4.8 mm, and with a gingival height ranging from 2.5 to 4.5 mm.

    Select straight and angled Multi-Unit Abutments have an added coating of zirconium nitride (ZrN). This coating is identical to the ZrN coating used on abutments with ZrN cleared in K242340.

    DESS® Dental Smart Solutions Screws are designed to attach the abutment to the implant or the prosthesis to the abutment. There are a total of seventeen (17) subject device screws compatible with the subject device components or previously cleared components. The new screws have designs that are similar to those of screws cleared in K242340 and K240208. Screws are made of titanium alloy (Ti-6Al-4V). Select subject device screws are available with DLC (Diamond-like Carbon) coating and are identical to that used on screws cleared in K240208.

    AI/ML Overview

    N/A

    Ask a Question

    Ask a specific question about this device

    K Number
    K230143

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2023-04-13

    (85 days)

    Product Code
    Regulation Number
    872.3630
    Panel
    Dental
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    Why did this record match?
    Reference Devices :

    K061319, K192221, K130572, K201553

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    DESS Multi-Unit Abutments are intended to be used in conjunction with endosseous dental implants in the maxillary or mandibular arch to provide support for prosthetic restorations.

    Device Description

    The purpose of this submission is to add components to the DESS Dental Smart Solutions system, which includes abutments cleared previously in K170588, K173908, K191986, K212577, K212628, K222269, and K22288. These previously cleared abutments are compatible with a variety of original equipment manufacturers (OEM) of dental implants. This submission adds abutments for implant lines from Implant Direct Sybron Manufacturing LLC (hereinafter, Implant Direct). The subject device abutments are compatible with various Implant Direct internal hex and internal conical implant bodies. The subject device abutments include Multi-Unit Abutments in straight, 17° angled, and 30° angled designs. All abutments are provided with the appropriate abutment screw (ff applicable) for attachment to the corresponding implant. All abutments and screws are provided non-sterile. The subject device is only intended for multi-unit restorations such as bridges and bars.

    AI/ML Overview

    This appears to be a 510(k) Summary for a medical device (dental abutments) rather than a study report with acceptance criteria for an AI/ML powered device. The document describes the device, its intended use, comparison to predicate devices, and non-clinical performance data (mechanical testing, biocompatibility, sterilization validation, MRI compatibility). It explicitly states: "No clinical data were included in this submission."

    Therefore, I cannot extract acceptance criteria or information about a study that proves the device meets those criteria in the context of an AI/ML device, as the provided text doesn't describe such a device or study.

    If the request was based on a misunderstanding of the document's content, please clarify. If you're looking for general information on what would be included in such a study for an AI/ML device, I can provide a theoretical outline.

    Ask a Question

    Ask a specific question about this device

    K Number
    K222211

    Validate with FDA (Live)

    Date Cleared
    2023-02-02

    (192 days)

    Product Code
    Regulation Number
    872.3640
    Panel
    Dental
    Age Range
    All
    Predicate For
    N/A
    Why did this record match?
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Simply Iconic™ dental implants are two-piece implants for one-stage surgical procedures. These implants are intended for use in partially and fully edentulous upper and lower jaws in support of single or multiple-unit restorations and termediate abutment support for fixed bridgework. Implants can be indicated for immediate loading when stability has been achieved and with appropriate occlusal loading.

    Legacy2, simplyLegacy2, Legacy3, simplyLegacy3, and Legacy4 dental implants are two-piece implants for one-stage or twostage surgical procedures. These implants are intended for use in partially and fully edentulous upper and lower jaws in support of single or multiple-unit restorations and terminal or intermediate abutment support for fixed bridgework. Implants can be indicated for immediate loading when good primary stability has been achieved and with appropriate occlusal loading. The Legacy 2, Legacy3, and Legacy4 fixture-mounts are intended for use with the corresponding dental implants (Legacy3, and Legacy4, respectively). The fixture-mounts can function as an abutment, fixture-mounts are intended for use with dental implants in the maxillary and/or mandbular arches to provide support for crowns or bridges for edentulous or partially edentulous patients.

    Custom Titanium Abutments are customizable devices intended for use with dental implants in the maxillary and or mandibular arches to provide support for crowns or bridges for edentulous or partially edentulous patients.

    The Legacy™ SMARTBase Abutment system is designed to be used in support of a dental implant(s) to provide support for prosthetic restorations in a partially edentulous patient. Legacy SMARTBase engaging abutments are intended for use in the mandible or maxilla in support of single unit restorations.

    InterActive/SwishActive Implant System consists of two-piece implants for one-stage surgical procedures. These implants are intended for use in partially and fully edentulous upper and lower jaws in support of single or multiple-unit restorations and terminal or internediate SMARTBase support for fixed bridgework.

    Implant Direct Custom Bars are patient specific devices indicated for attachment to dental implants in the treatment of partially or fully edentulous jaws for the purpose of restoring chewing function.

    GPS® Angled Abutments are designed to be used in support of a dental implant(s) to provide support for prosthetic restorations in a partially or completely edentulous patient. These abutments are designed to only receive a fabricated multi-unit bridge or overdenture. Abutments are intended for use in the mandible or maxilla. Implants can be indicated for immediate loading when good primary stability has been achieved and with appropriate occlusal loading.

    InterActive/SwishActive Implant System consists of two-piece implants for one-stage surgical procedures. These implants are intended for use in partially and fully edentulous upper and lower jaws in support of single or multipleunit restorations and terminal or intermediate abutment support for fixed bridgework. Implants can be indicated for immediate loading when good primary stability has been achieved and with appropriate occlusal loading.

    Legacy3 6mm Length consists of two-piece implants for one-stage or two-stage surgical procedures. There implants are intended for use in partially and fully edentulous upper and lower jaws in support of single or multiple-unit restorations and terminal or internediate abutment support for fixed bridgework. Implants can be indicated for immediate loading when good primary stability has been achieved and with appropriate occlusal loading.

    InterActive/SwishPlus2 Implant System consists of two-piece implants for one-stage or two-stage surgical procedures. These implants are intended for use in partially and fully edentulous upper and lower jaws in support of single or multiple-unit restorations and terminal or intermediate abutment support for fixed bridgework: Implants can be indicated for immediate loading when good primary stability has been achieved and with appropriate occlusal loading.

    Spectra-System Dental Implants 2008 are comprised of dental implant fixtures and prosthetic devices that compose a two-piece implant system. The implants are intended for use in the mandible and maxilla, in support of single unit or multiple unit cement or screw-receiving restorations and for the retention and support of overdentures. The implants are intended for immediate placement and function for the support of single-tooth restorations, recognizing bone stability and appropriate occlusal load requirements.

    The SwissPlant Dental Implant system consists of two-piece implants for one or two-stage surgical procedures that are intended for use in partially or fully edentulous maxillae, in support of single or multiple-unit restorations including: cement retained, screw retained or overdenture restorations and in terminal or immediate abutment support for fixed bridgework. The SwissPlant dental implants are intended for immediate placement and function on single tooth and/or multiple tooth applications recognizing initial implant stability and appropriate occlusal loading, to restore normal masticatory function.

    Spectra-System Abutments are designed to be used in support of a dental implant(s) to provide support for prosthetic restorations such as crowns, bridges, overdentures or custom prosthetic fabrications in a partially or completely edentulous patient. Spectra-System Abutments are intended for use in the mandble or maxilla. Prostheses can be screw or cement retained to the abutment.

    The ReActive Dental Implant System are dental implant fixtures that are a part of a two-piece implant system. The ReActive implants are intended for use in the mandible and maxilla, in support of single or multiple-unit cement or screw receiving fixed restorations and for retention and support of overdentures. The implants are intended for immediate placement and function for support of single tooth and/or multiple tooth restorations, recognizing bone stability and appropriate occlusal load requirements.

    ScrewIndirect Narrow Dental Implants for single-stage surgical procedures intended for use in partially or fully edentulous mandibles and maxillae, in support of complete or partial denture prostheses or as a terminal or intermediary attachment for fixed or removable bridgework via interface with copings. These implants are intended for for support of splinted multiple tooth restorations, provided initial implant stability and appropriate occlusal load requirements are met.

    The RePlus Dental Implants with HA coating are dental implant fixtures that are a part of a two-piece implant system. The RePlus implants are intended for use in the mandible and maxilla, in support of single or multiple-unit cement or screw receiving fixed restorations and for retention and support of overdentures. The immediate loading for support of single tooth and/or multiple tooth restorations, provided initial implant stability and appropriate occlusal load requirements are met.

    The ScrewPlus Dental Implant with HA coating is a dental implant fixture that is a part of a two-piece implant system that is to be used for single-stage or two-stage surgical procedures. The ScrewPlus implants are intended for use in the mandible and maxilla, in support of single or multiple-unit cement or screw receiving fixed restoration and support of overdentures. The implants are intended for immediate placement and function for multiple tooth and/or multiple tooth restorations, recognizing bone stability and appropriate occlusal load requirements.

    The Legacy Dental Implant with HA coating is a dental implant fixture that is a part of a two-piece implant system. The Legacy implants are intended for use in the mandible and maxilla, in support of single or multiple-unit cement or screw receiving fixed restorations and for retention and support of overdentures. The implants are intended for immediate placement and function for support of single tooth and/or multiple tooth restorations, recognizing bone stability and appropriate occlusal load requirements.

    The Spectra Dental Implant System consists of one-piece implants for single-stage or two- stage surgical procedures that are intended for use in partially or fully edentulous mandillae, in support of single or multiple-unit restorations and terminal or intermediate abutment support for fixed bridgework. They may be placed in immediate function if initial implant stability can be established.

    The Legacy Abutment System is intended for use with dental implants in the maxillary and/or mandbular arches to provide support for crowns, bridges, or overdentures for edentulous or partially edentulous patients.

    Device Description

    Not Found

    AI/ML Overview

    The provided text is a compilation of Indications for Use statements for various dental implant systems and components, all under the umbrella of "Implant Direct Dental Implant Systems Portfolio - MR Conditional." Each section describes the intended use for a specific product or family of products, often referencing previous 510(k) clearances.

    Crucially, this document does not contain any information about acceptance criteria or studies that prove the device meets these criteria. It is a regulatory filing for the Indications for Use for a medical device which has already been cleared by the FDA to be "substantially equivalent" to predicate devices.

    Therefore, I cannot fulfill the request to describe acceptance criteria and the study that proves the device meets the acceptance criteria based solely on the provided text. The document is for describing what the device is for, not how its performance was tested or validated.

    To provide the requested information, I would need a different type of document, such as a summary of safety and effectiveness, a clinical study report, or a validation plan for a specific performance characteristic.

    Ask a Question

    Ask a specific question about this device