Search Filters

Search Results

Found 5 results

510(k) Data Aggregation

    Why did this record match?
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The EXSYS DEXi is a diagnostic X-ray system intended for use in generating radiographic images of human anatomy for general purpose. The system obtains necessary information of patient's anatomical structure by an image processing (workstation) after process of examination using radiation exposure with DR. This system is not intended for mammography applications.

    Device Description

    The EXSYS DEXi is a diagnostic X-ray system intended for use in generating radiographic images of human anatomy for general purpose. The system obtains necessary information of patient's anatomical structure by an image processing (workstation) after process of examination using radiation exposure with DR. This system is not intended for mammography applications.

    The EXSYS DEXi composed of a x-ray generator, tube, collimator, tube stand, bucky stand, patient table, flat panel detector and console.

    AI/ML Overview

    This FDA 510(k) summary does not contain information about acceptance criteria and device performance as it pertains to AI/ML or image analysis aspects. The document focuses on the substantial equivalence of the modified EXSYS DEXi diagnostic X-ray system to a previously cleared predicate device (K233530) based on hardware and software updates, and compliance with general safety and performance standards for X-ray systems.

    Specifically, the document refers to non-clinical data and verification and validation testing demonstrating compliance with various international and FDA-recognized consensus standards (e.g., IEC 60601 series for medical electrical equipment, ISO 14971 for risk management, IEC 62304 for medical device software, UL ANSI 2900-1 and IEC 81001-5-1 for cybersecurity). It states, "The test results support that all the specifications have met the acceptance criteria. Verification and validation testing were found acceptable to support the claim of substantial equivalence." However, it does not provide a table of acceptance criteria with reported device performance metrics in an AI/ML context, nor does it describe specific studies that would typically prove such performance (e.g., standalone performance studies, MRMC studies, details on ground truth establishment for a diagnostic algorithm, sample sizes for test/training sets relevant to AI performance).

    The "technological characteristics" table (Table 1) compares design parameters of the subject device (new models of EXSYS DEXi) with the predicate device, highlighting additions like new collimators, mechanical parts, detectors, and software (EConsole2). The discussion column for these additions generally states, "The system has been tested and there is 'No negative impact on safety or efficacy' and there are no new potential or increased safety risks concerning this difference." This refers to overall system safety and performance in line with a general X-ray system, not specific AI/ML diagnostic performance.

    Therefore, based on the provided document, the following information cannot be extracted:

    1. A table of acceptance criteria and the reported device performance (for AI-specific functions): Not provided. The document states general compliance with standards and "test results support that all the specifications have met the acceptance criteria," but does not detail these criteria or performance metrics specific to an AI component's diagnostic accuracy, sensitivity, specificity, etc.
    2. Sample size used for the test set and the data provenance: Not provided for AI/ML performance evaluation.
    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts: Not provided for AI/ML performance evaluation.
    4. Adjudication method for the test set: Not provided for AI/ML performance evaluation.
    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, and the effect size: Not mentioned.
    6. If a standalone (algorithm only without human-in-the-loop performance) was done: Not mentioned.
    7. The type of ground truth used: Not specified for AI/ML performance evaluation.
    8. The sample size for the training set: Not provided for AI/ML performance evaluation.
    9. How the ground truth for the training set was established: Not provided for AI/ML performance evaluation.

    The document states, "Clinical studies are unnecessary to validate the safety and effectiveness of the Stationary x-ray system, EXSYS DEXi, the subject of this 510(k) notification," further indicating that specific performance data from clinical trials or detailed AI algorithm validation studies (which typically involve such criteria) are not included in this submission summary. The software updates mentioned (EConsole2) were previously cleared via K240243, suggesting that any specific performance data for that software might be found in its own 510(k) submission, but not in this document.

    Ask a Question

    Ask a specific question about this device

    K Number
    K210919

    Validate with FDA (Live)

    Device Name
    AcuityDRe
    Manufacturer
    Date Cleared
    2021-04-30

    (32 days)

    Product Code
    Regulation Number
    892.1680
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    Why did this record match?
    Reference Devices :

    K192400, K193017, K171137, K162552, K162555

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Intended for use by a qualified/trained physician or technician on both adult and pediatric subjects for taking diagnostic x-rays. Not for mammography.

    Device Description

    AcuityDRe components into a complete digital x-ray system upgrade kit, including software and digital radiography panels. The customer selects one of the following digital x-ray receptor panels: AcuityDRe 1417w, AcuityDRe 1717w, AcuityDRe 1717t. The "w" indicates wireless wi-fi while the "t" indicates tethered. The indications for use remains unchanged: Intended for use by a qualified/trained physician or technician on both adult and pediatric subjects for taking diagnostic x-rays. Not for mammography. So the only difference between this submission and the predicate submission is the generator/tubestand combination. Each system consists of the following items: Customer supplies: Diagnostic x-ray generator (HF) Class I Code IZO. + Tubehead: Class I Code ITY + Tube Mount: Class I Code IYB + Attached Collimator, Manual (IZX) Class II 510(k) Exempt We supply: Digital X-Ray Receptor Panel 892.1680 Class II Code MQB. Digital X-ray Software 892.2050 Class II Code LLZ. The software offered for sale with this system has received previous 510(k) clearance in K201058.

    AI/ML Overview

    The provided text describes the 510(k) premarket notification for the AcuityDRe device, which is an upgrade kit for digital x-ray systems. It primarily focuses on demonstrating substantial equivalence to a predicate device (Accuvue+, K201058), rather than detailing a clinical study with specific acceptance criteria and performance against those criteria in a typical sense of an AI/ML medical device.

    The "study" cited here is non-clinical testing, specifically clinical image evaluation performed by a Board Certified Radiologist, and bench testing for physical performance characteristics. It's not a comparative effectiveness study involving human readers with and without AI assistance, nor a standalone AI performance study.

    Here's an attempt to extract and present the information based on the provided text, acknowledging that many requested fields regarding AI/ML device studies are not applicable or detailed in this 510(k) summary for an x-ray hardware upgrade kit.


    Device: AcuityDRe (Digital X-ray System Upgrade Kit)

    Description of Testing (Non-Clinical and Image Evaluation)

    The primary goal of the testing was to demonstrate substantial equivalence of the AcuityDRe system (comprising new digital x-ray receptor panels, software, and generator compatibility) to a legally marketed predicate device (Accuvue+, K201058). This was achieved through:

    • Bench testing for physical performance characteristics of the new digital x-ray panels.
    • Clinical image evaluation by a Board Certified Radiologist to assess image quality.
    • Confirmation of compliance with relevant medical device standards and FDA guidance documents.

    1. Table of Acceptance Criteria and Reported Device Performance

    Given that this 510(k) is for an X-ray system upgrade kit rather than an AI/ML algorithm, the "acceptance criteria" are not framed as typical AI performance metrics (e.g., sensitivity, specificity, AUC). Instead, the key "acceptance criteria" are based on demonstrating substantial equivalence to the predicate device in terms of:

    • Same Indications for Use
    • Similar Technological Characteristics (especially panel performance: DQE, MTF, and panel sizes)
    • Compliance with safety and performance standards.
    • "Excellent quality" of clinical images as assessed by an expert.
    Feature / Criteria (Implied for Substantial Equivalence)Acceptance Criteria (Compared to Predicate / Standards)Reported Device Performance (AcuityDRe)Comparison Result
    Indications for UseSame as predicate."Intended for use by a qualified/trained physician or technician on both adult and pediatric subjects for taking diagnostic x-rays. Not for mammography." - SAME as predicate.SAME
    X-ray Generator CompatibilityCompatible with CPI or Sedecal generators.Compatible CPI generators: CMP 200 Series. Compatible Sedecal generators: SFHR and SHF Series. - SAME as predicate. Software control of technique factor possible with certain compatible generators.SAME
    Digital X-Ray DetectorsNew models, but comparable performance and size.AcuityDRe 1417w, AcuityDRe 1717w, AcuityDRe 1717t. New models had not received previous FDA clearance, hence testing was performed for them.SAME (as new versions)
    Panel Performance (DQE @ 1.0 lp/mm)Comparable to predicate's DQE.AcuityDRe 1417w: 35% (Predicate AcuityDR 1417: 34.6%) AcuityDRe 1717w/1717t: 42% (Predicate AcuityDR 1717: 23.6%) (Note: The AcuityDRe 1717w/1717t shows improved DQE compared to the predicate's AcuityDR 1717, which is a positive attribute for image quality).Almost IDENTICAL / Improved for some models
    Panel Performance (MTF @ 2.0 lp/mm)Comparable to predicate's MTF.AcuityDRe 1417w: 31% (Predicate AcuityDR 1417: 34%) AcuityDRe 1717w/1717t: 38% (Predicate AcuityDR 1717: 34%) (Note: Ranges are close, with slight variations).Almost IDENTICAL
    Panel SizesComparable to predicate's sizes.AcuityDRe 1417w: 148 µm AcuityDRe 1717w: 140 µm AcuityDRe 1717t: 140 µm (Predicate AcuityDR 1417: 140µm; AcuityDR 1717: 140μm)Almost IDENTICAL
    Operator ConsoleWindows PC using Windows 10-IoT.SAMESAME
    Acquisition SoftwareAccuVue / AccuVue+.SAMESAME
    Power SourceAC Line or rechargeable batteries.SAMESAME
    Compliance with StandardsDemonstrated compliance with listed standards.IEC 60601-1:2005/(R)2012 And A1:2012 (Medical Electrical Equipment Safety) IEC 60601-1-2:2014 (Electromagnetic Disturbances) NEMA PS 3.1 - 3.20 (2011) (DICOM Set) - SAME as predicate.SAME
    Clinical Image Quality"Of excellent quality" (qualitative assessment)."The images were found to be of excellent quality." (Qualitative assessment by expert)Achieved

    2. Sample Size Used for the Test Set and Data Provenance

    • Test Set Sample Size: Not explicitly stated numerically for the clinical image evaluation. The text only mentions "Clinical image evaluation was performed on the proposed new panels."
    • Data Provenance: Not specified. It's likely retrospective, as it's an evaluation of images generated by the new panels. Country of origin not mentioned; presumed to be images from typical clinical settings where such panels would be used.

    3. Number of Experts Used to Establish Ground Truth and Qualifications

    • Number of Experts: One.
    • Qualifications: A "Board Certified Radiologist." Further details like years of experience are not provided.

    4. Adjudication Method for the Test Set

    • Adjudication Method: Not applicable/None, as only a single expert was used for the qualitative assessment of image quality.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

    • MRMC Study: No. This type of study (AI vs. human-assisted AI) was not performed as the device is an x-ray system upgrade, not an AI diagnostic algorithm.
    • Effect Size of Human Reader Improvement: Not applicable.

    6. Standalone (Algorithm Only Without Human-in-the-Loop Performance) Study

    • Standalone Study: No, as the device itself is a component of an x-ray system, not a standalone AI algorithm. The image quality assessment was of the images produced by the hardware, which would then be interpreted by humans.

    7. Type of Ground Truth Used

    • Ground Truth Type: Expert Consensus / Qualitative Expert Assessment. The "ground truth" for the image quality was the subjective assessment by the Board Certified Radiologist that the images were "of excellent quality." This is distinct from, for example, pathology-confirmed diagnoses or patient outcomes data.

    8. Sample Size for the Training Set

    • Training Set Sample Size: Not applicable. This device is a hardware component (digital x-ray panels) and associated software, not an AI/ML algorithm that requires a "training set" in the context of machine learning model development. The software mentioned (AccuVue/AccuVue+) has received previous clearance for control of generators.

    9. How the Ground Truth for the Training Set was Established

    • Ground Truth Establishment for Training Set: Not applicable, as there is no "training set" in the context of an AI/ML algorithm for this device.
    Ask a Question

    Ask a specific question about this device

    K Number
    K203703

    Validate with FDA (Live)

    Device Name
    I-Q View
    Manufacturer
    Date Cleared
    2021-03-24

    (96 days)

    Product Code
    Regulation Number
    892.1680
    Age Range
    All
    Predicate For
    N/A
    Why did this record match?
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    This software is intended to generate digital radiographic images of the skull, spinal column, extremities, and other body parts in patients of all ages. Applications can be performed with the patient sitting, or lying in the prone or supine position and is intended for use in all routine radiography exams. The product is not intended for mammographic applications.

    This software is not meant for mammography, fluoroscopy, or angiography.

    Device Description

    The I-Q View is a software package to be used with FDA cleared solid-state imaging receptors. It functions as a diagnostic x-ray image acquisition platform and allows these images to be transferred to hard copy, softcopy, and archive devices via DICOM protocol. The flat panel detector is not part of this submission. In the I-Q View software, the Digital Radiography Operator Console (DROC) software allows the following functions:

      1. Add new patients to the system; enter information about the patient and physician that will be associated with the digital radiographic images.
      1. Edit existing patient information.
      1. Emergency registration and edit Emergency settings.
      1. Pick from a selection of procedures, which defines the series of images to be acquired.
      1. Adiust technique settings before capturing the x-ray image.
      1. Preview the image, accept or reject the image entering comments or rejection reasons to the image. Accepted images will be sent to the selected output destinations.
      1. Save an incomplete procedure, for which the rest of the exposures will be made at a later time.
      1. Close a procedure when all images have been captured.
      1. Review History images, resend and reprint images.
      1. Re-exam a completed patient.
      1. Protect patient records from being deleted by the system.
      1. Delete an examined Study with all images being captured.
      1. Edit User accounts.
      1. Check statistical information.
      1. Image QC.
      1. Image stitching.
      1. Provides electronic transfer of medical image data between medical devices.
    AI/ML Overview

    The provided document is a 510(k) summary for the I-Q View software. It focuses on demonstrating substantial equivalence to a predicate device through bench testing and comparison of technical characteristics. It explicitly states that clinical testing was not required or performed.

    Therefore, I cannot provide details on clinical acceptance criteria or a study proving the device meets them, as such a study was not conducted for this submission. The document relies on bench testing and comparison to a predicate device to establish substantial equivalence.

    Here's a breakdown of what can be extracted from the provided text regarding acceptance criteria and the "study" (bench testing) that supports the device:

    1. Table of Acceptance Criteria and Reported Device Performance

    Since no clinical acceptance criteria or performance metrics are provided, this table will reflect the general statements made about the device performing to specifications.

    Acceptance Criteria (Implied)Reported Device Performance
    Device functions as intended for image acquisition.Demonstrated intended functions.
    Device performs to specification.Performed to specification.
    Integration with compatible solid-state detectors performs within specification.Verified integration performance within specification.
    Software is as safe and functionally effective as the predicate.Bench testing confirmed as safe and functionally effective as predicate.

    2. Sample size used for the test set and the data provenance

    • Test Set Sample Size: Not applicable/not reported. The document describes bench testing, not a test set of patient data.
    • Data Provenance: Not applicable. Bench testing generally involves internal testing environments rather than patient data from specific countries or retrospective/prospective studies.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    • Not applicable. As no clinical test set was used, no experts were needed to establish ground truth for patient data. Bench testing typically relies on engineering specifications and verification.

    4. Adjudication method for the test set

    • Not applicable. No clinical test set or human interpretation was involved.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    • No, an MRMC comparative effectiveness study was not done. The document explicitly states: "Clinical Testing: The bench testing is significant enough to demonstrate that the I-Q View software is as good as the predicate software. All features and functionality have been tested and all specifications have been met. Therefore, it is our conclusion that clinical testing is not required to show substantial equivalence." The device is software for image acquisition, not an AI-assisted diagnostic tool.

    6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done

    • Yes, in a sense. The "study" described is bench testing of the software's functionality and its integration with solid-state detectors. This is an evaluation of the algorithm/software itself.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

    • For bench testing, the "ground truth" would be the engineering specifications and expected functional behavior of the software and its interaction with hardware components. It's about verifying that the software performs according to its design requirements.

    8. The sample size for the training set

    • Not applicable. The I-Q View is described as an image acquisition and processing software, not an AI/machine learning model that typically requires a training set of data.

    9. How the ground truth for the training set was established

    • Not applicable, as there is no mention of a training set or AI/machine learning component.

    Summary of the "Study" (Bench Testing) for K203703:

    The "study" conducted for the I-Q View software was bench testing. This involved:

    • Verification and validation of the software.
    • Demonstrating the intended functions and relative performance of the software.
    • Integration testing to verify that compatible solid-state detectors performed within specification as intended when used with the I-Q View software.

    The conclusion drawn from this bench testing was that the software performs to specification and is "as safe and as functionally effective as the predicate software." This was deemed sufficient to demonstrate substantial equivalence, and clinical testing was explicitly stated as not required.

    Ask a Question

    Ask a specific question about this device

    K Number
    K201058

    Validate with FDA (Live)

    Device Name
    Accuvue+
    Manufacturer
    Date Cleared
    2020-05-06

    (15 days)

    Product Code
    Regulation Number
    892.1680
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    Why did this record match?
    Reference Devices :

    K192400, K193017, K171137, K162552, K162555

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Intended for use by a qualified/trained physician or technician on both adult and pediatric subjects for taking diagnostic x-rays. Not for mammography.

    Device Description

    Accuvue+ components into a complete digital x-ray system upgrade kit, including software and digital radiography panels. The customer selects one (or more) of the following digital x-ray receptor panels: DRTech 4343A, (K192400); DRTech 4343W, (K193017); AcuityDR (K171137); AcuityDR 1417 (K162552 EVS 3643G) or AcuityDR 1717 (K162555 EVS 4343, EVS 4343G). We wish to offer for sale a subset of our recently cleared submission K200726. This in essence would be an upgrade kit. The indications for use remains unchanged: Intended for use by a qualified/trained physician or technician on both adult and pediatric subjects for taking diagnostic x-rays. Not for mammography. So the only difference between this submission and the predicate submission is the generator/tubestand combination.

    AI/ML Overview

    This document is a 510(k) summary for a medical device called Accuvue+. It describes the device and its intended use, and argues for its substantial equivalence to a predicate device. However, it does not contain information about acceptance criteria or a study proving that the device meets specific performance criteria through a clinical or non-clinical study.

    The document explicitly states:

    • "6. Clinical testing. Not required for a determination of substantial equivalence."
    • "5. Non clinical testing: Testing was performed in previous submssions according to the following standards:" (It then lists general safety and communication standards, not performance criteria for diagnostic accuracy or image quality.)

    Therefore, I cannot extract the requested information regarding acceptance criteria and a study proving the device meets those criteria from the provided text. The submission focuses on demonstrating substantial equivalence based on technological characteristics and adherence to general safety standards, rather than a performance study with acceptance criteria.

    Ask a Question

    Ask a specific question about this device

    K Number
    K200726

    Validate with FDA (Live)

    Device Name
    AcuityPDR
    Manufacturer
    Date Cleared
    2020-04-13

    (24 days)

    Product Code
    Regulation Number
    892.1720
    Age Range
    0 - 0.166
    Predicate For
    Why did this record match?
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Intended for use by a qualified/trained physician or technician on both adult and pediatric subjects for taking diagnostic x-rays. Not for mammography.

    Device Description

    RadmediX AcuityPDR combines components into a complete mobile x-ray system, including software, a generator/collimator combination, and digital radiography panels. Radmedix combines components from various manufacturers into a complete mobile x-ray system. The customer selects one (or more) of the following digital x-ray receptor panels: DRTech 4343A, (K192400); DRTech 4343W, (K193017); AcuityDR (K171137); AcuityDR 1417 (K162552 EVS 3643, EVS 3643G) or AcuityDR 1717 (K162555 EVS 4343, EVS 4343G). In addition, the customer selects one of three software packages: Accuvue+ (K130883), AccuVueMED, (K152172) or AccuVue (K141440). The generator can be battery operated. A single battery charge produces 200 Exposures at max KV and MAS settings. The battery charger is UL Listed and the internal lithium ion battery is overcharge and overcurrent protected. A typical acquisition computer would be a Lenovo P53S or a Dell Precision 3541.

    AI/ML Overview

    This document describes a medical device, AcuityPDR, a mobile X-ray system, and its 510(k) submission for FDA clearance. The submission relies on demonstrating substantial equivalence to a predicate device (MinXray CMDR 2CW).

    Here's an analysis of the requested information based on the provided text:

    1. A table of acceptance criteria and the reported device performance

    The document does not explicitly state acceptance criteria in the form of quantitative performance metrics for the AcuityPDR device itself. Instead, it focuses on demonstrating "substantial equivalence" to a predicate device. The acceptance criteria for this type of submission are typically that the device performs as safely and effectively as the predicate and has the same indications for use.

    The device performance is reported implicitly through the comparison table on page 4, showing that the AcuityPDR either matches or has comparable/equivalent functionality to the predicate in various technical parameters.

    Acceptance Criteria (Implicit for Substantial Equivalence)Reported Device Performance (AcuityPDR vs. Predicate CMDR 2CW)
    Indications for Use (Identical)Intended for use by a qualified/trained physician or technician on both adult and pediatric subjects for taking diagnostic x-rays. Not for mammography. (Identical)
    X-ray Tube Functionality (Equivalent/Comparable)TOSHIBA D-125SB (Equivalent Functionality)
    Peak Generator Output (Same)2 kW (SAME)
    Tube Current Range (Comparable)40-60kV: 25mA, 61-100kV: 20mA (Comparable ranges)
    Tube Voltage Adjustable Range (More Flexible)40-100kV, step value 1kV (More flexible kV adjustment)
    mAs Range (Comparable Adjustability)0.4mAs ~ 50mAs (Comparable adjustability)
    Collimator (Equivalent Functionality)Built in (Equivalent Functionality)
    X-ray Generator (Equivalent Functionality)One model, up to 100 kVp (Equivalent Functionality)
    Digital X-Ray Detectors (Equivalent Functionality)DRTech 4343A, DRTech 4343W, AcuityDR (various models) (Equivalent Functionality)
    Operator Console (Similar Functionality)SAME (Touch Control or Touch Screen) (Similar Functionality)
    Acquisition Software (Equivalent Functionality)AccuVueMED, AccuVue, or AccuVue+ (Equivalent Functionality, only cleared software supplied)
    Power Source (Same)AC Line or rechargeable batteries (Generator only) (SAME)
    Compliance with Standards (Same)60601-1:2005; 60601-1-2:2014, etc. (SAME)

    2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    The document explicitly states in section 6, "Clinical testing. Not required for a determination of substantial equivalence." This indicates that no clinical test set was used for this specific submission. The submission relies on non-clinical testing (bench testing, standards compliance) and comparison to the predicate device.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

    Since no clinical test set was required or used for this submission, there is no mention of experts being used to establish ground truth for a test set.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    As no clinical test set was used, no adjudication method is mentioned.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    No MRMC study was performed as clinical testing was "Not required for a determination of substantial equivalence." The device is a mobile X-ray system, not an AI-powered diagnostic tool, so the concept of human readers improving with AI assistance is not applicable in this context.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    This is not an AI algorithm; it's a mobile X-ray system. Therefore, standalone algorithm performance is not applicable. The performance is assessed through compliance with electrical and safety standards for X-ray devices and comparison of technical specifications to a predicate.

    7. The type of ground truth used (expert concensus, pathology, outcomes data, etc)

    For this type of 510(k) submission, "ground truth" is primarily established by:

    • Compliance with recognized electrical, safety, and performance standards for X-ray generating equipment (e.g., IEC 60601 series, 21 CFR 1020).
    • Demonstration that the device's technical specifications and intended use are substantially equivalent to a legally marketed predicate device.
    • Component certifications: "All of the components subject to the CDRH performance standard are certified to comply with the standard by their respective manufacturers."

    8. The sample size for the training set

    The document does not describe a training set for an AI model, as this is not an AI device. The "training" for such a device would refer to its design and manufacturing processes adhering to established engineering principles and standards.

    9. How the ground truth for the training set was established

    Not applicable, as no AI model or training set is described. The "ground truth" for the device's design and manufacturing relies on established medical device regulations, design controls, and industry standards for X-ray imaging equipment.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1