K Number
K223124

Validate with FDA (Live)

Manufacturer
Date Cleared
2022-11-18

(46 days)

Product Code
Regulation Number
892.1680
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The EXPD 86P, EXPD 86PG, EXPD 129PG Digital X-ray detector is indicated for digital imaging solution designed for providing general radiographic diagnosis of human anatomy. This device is intended to replace film or screen based radiographic systems in all general purpose diagnostic procedures. This device is not intended for mammography applications.

Device Description

The EXPD 86P, EXPD 86PG, EXPD 129P, EXPD 129PG is a flat-panel type digital X-ray detector that captures projection radiographic images in digital format within seconds, eliminating the need for an entire x-ray film or an image plate as an image capture medium. EXPD 86P, EXPD 129P, EXPD 129PG differs from traditional X-ray systems in that, instead of exposing a film and chemically processing it to create a hard copy image, a device called a Detector is used to capture the image in electronic form.

The EXPD 86P, EXPD 86PG, EXPD 129P, EXPD 129PG Detector is an indirect conversion device in the form of a square plate in which converts the incoming X-rays into visible light. This visible light is then collected by an optical sensor, which generates an electric charges representation of the spatial distribution of the incoming X-ray quanta.

The charges are converted to a modulated electrical signal thin film transistors. The amplified signal is converted to a voltage signal and is then converted from an analog to digital signal which can be transmitted to a viewed image print out, transmitted to remote viewing or stored as an electronic data file for later viewing.

AI/ML Overview

The provided text describes the 510(k) summary for the EXPD 86P, EXPD 86PG, EXPD 129P, EXPD 129PG digital X-ray detectors. This document focuses on demonstrating substantial equivalence to a predicate device (EVS 4343A/AG and EVS 3643A/AG, K192400) rather than presenting a detailed clinical study for novel AI/ML-driven diagnostic aids.

Therefore, many of the requested criteria related to clinical studies, ground truth establishment, expert adjudication, and MRMC studies are not applicable to this type of submission. The performance assessment here is primarily based on technical benchmarks and non-clinical data demonstrating that the new device performs at least as well as the predicate device in fundamental radiographic characteristics.

Here's a breakdown of the requested information based on the provided text:

1. Table of acceptance criteria and the reported device performance

The acceptance criteria are implicitly defined by demonstrating that the subject devices (EXPD 86P/PG and EXPD 129P/PG) perform "equal or worth" (meaning equal to or better) than the predicate devices (EVS 4343A/AG and EVS 3643A/AG) across key performance metrics. The document states: "According to bench test report, it is proved that the DQE and MTF of predicated device and subject device are basically equal or worth than the predicate device. As a result, subject devices performance is equal or worth than the predicate device."

ParameterAcceptance Criteria (relative to predicate)Reported Device Performance (Subject Device Values)Predicate Device Values (for comparison)
DQE (Detective Quantum Efficiency)Equal to or better than predicateEXPD 129P: 50.0 % at 0.5lp/mm EXPD 129PG: 25.0 % at 0.5lp/mm EXPD 86P: 50.0 % at 0.5lp/mm EXPD 86PG: 25.0 % at 0.5lp/mmEVS 4343A: 52.9% at 1.0 lp/mm EVS 4343AG: 27.2 % at 1.0 lp/mm EVS 3643A: 50.5 % at 1.0 lp/mm EVS 3643AG: 26.3 % at 1.0 lp/mm
MTF (Modulation Transfer Function)Equal to or better than predicateEXPD 129P: 45.0 % at 2.0 lp/mm EXPD 129PG: 45.0 % at 2.0 lp/mm EXPD 86P: 45.0 % at 2.0 lp/mm EXPD 86PG: 45.0 % at 2.0 lp/mmEVS 4343A: 44.1 % at 2.0 lp/mm EVS 4343AG: 49.2 % at 2.0 lp/mm EVS 3643A: 44.5% at 2.0 lp/mm EVS 3643AG: 46.3 % at 2.0 lp/mm
Resolution3.5 lp/mm3.5 lp/mm3.5 lp/mm

Note: The DQE and MTF values are reported at different spatial frequencies for the subject and predicate devices (0.5 lp/mm vs 1.0 lp/mm for DQE, and both at 2.0 lp/mm for MTF). The document states the "bench test report... proved that the DQE and MTF... are basically equal or worth than the predicate device," suggesting that despite the different measurement points for DQE, the overall performance comparison satisfied the criteria.

2. Sample size used for the test set and the data provenance

Not applicable in the context of a clinical test set for AI/ML performance. This submission is for a digital X-ray detector, and the performance assessment is based on non-clinical, technical measurements (e.g., DQE, MTF) rather than a patient image test set for diagnostic accuracy. The testing performed would be lab-based measurements on the physical detectors.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

Not applicable. Ground truth in this context refers to physical characteristics measured on the device itself, not clinical diagnoses or interpretations of images by experts.

4. Adjudication method for the test set

Not applicable. No adjudication process is described as the evaluation is based on technical specifications rather than human interpretation.

5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

No. This is not an AI-assisted diagnostic device. It is a digital X-ray detector, which is hardware for acquiring images. No MRMC study was conducted or is relevant for this type of device submission.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

Not applicable. This is a hardware device (X-ray detector), not a standalone algorithm. Its performance is measured directly through physical parameters.

7. The type of ground truth used

The "ground truth" for this device's performance evaluation is based on physical and engineering measurements of objective image quality parameters such as DQE (Detective Quantum Efficiency), MTF (Modulation Transfer Function), and spatial resolution (lp/mm). These are standard metrics for evaluating X-ray detectors.

8. The sample size for the training set

Not applicable. This is a hardware device; there is no "training set" in the machine learning sense. Its design and performance are based on engineering principles and fabrication, not learned from data.

9. How the ground truth for the training set was established

Not applicable. As there is no training set for a machine learning model, there is no corresponding ground truth to establish for it.

{0}------------------------------------------------

November 18, 2022

Image /page/0/Picture/1 description: The image shows the logo for the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo includes the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG" in blue, and "ADMINISTRATION" in a smaller font size below.

DRTECH Corporation % Ella An Regulatory Affairs Suite No. 1, 2 Floor/Suite No. 2, 3 Floor, 29, Dunchon-daero 541beon-gil Jungwon-gu, Seongnam-si, Gyeonggi-do 13216 KOREA

Re: K223124

Trade/Device Name: EXPD 86P, EXPD 86PG, EXPD 129P, EXPD 129PG Regulation Number: 21 CFR 892.1680 Regulation Name: Stationary X-Ray System Regulatory Class: Class II Product Code: MOB Dated: September 29, 2022 Received: October 3, 2022

Dear Ella An:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal

{1}------------------------------------------------

statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Zhkke

Laurel Burk, Ph.D. Assistant Director Diagnostic X-Ray Systems Team DHT8B: Division of Radiological Imaging Devices and Electronic Products OHT8: Office of Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K223124

Device Name EXPD 86P, EXPD 86PG, EXPD 129P, EXPD 129PG

Indications for Use (Describe)

The EXPD 86P, EXPD 86PG, EXPD 129PG Digital X-ray detector is indicated for digital imaging solution designed for providing general radiographic diagnosis of human anatomy. This device is intended to replace film or screen based radiographic systems in all general purpose diagnostic procedures. This device is not intended for mammography applications.

Type of Use (Select one or both, as applicable)

Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

510(k) Number: K223124

DRTECH Corporation Suite No.1, 2 Floor / Suite No. 2, 3 Floor, 29, Dunchon-Daero 541beon-gil, Jungwon-gu, Seongnam-si, Gyeonggi-do, South Korea

510(k) Summary

[As required by 21 CFR 807.92]

This 510(k) summary of safety and effectiveness information is prepared in accordance with 21 CFR 807.92

1. Date Prepared [21 CFR 807.92(a) (1)]

September/29/2022

2. Submitter's Information [21 CFR 807.92(a) (1)]

• Name of Sponsor:DRTECH Corporation
• Address:Suite No.1, 2 Floor / Suite No. 2, 3 Floor, 29, Dunchon-daero541 beon-gilJungwon-gu, Seongnam-si, Gyeonggi-do, South Korea
  • Contact Name: Ella An ●
  • Telephone No.: + 82-31-779-7787
  • Fax No .: + 82-31-779-7790
  • Email Address : ● ysan96@drtech.co.kr
  • Registration Number: ● 3005172103
  • Name of Manufacturer: Same as Sponsor ●

3. Trade Name, Common Name, Classification [21 CFR 807.92(a) (2)]

• Trade Name:EXPD 86P, EXPD 86PG, EXPD 129P, EXPD 129PG
-----------------------------------------------------------
  • . Common Name: Digital Flat Panel X-ray Detector
  • Classification Name: Stationary X-ray System ●
  • Classification Panel: Radiology
  • Classification Regulation: 21 CFR 892.1680 ●
  • Product Code: MOB
  • Device Class: II ●

4. Identification of Predicate Device(s) [21 CFR 807.92(a) (3)]

  • 510(k) Number: ● K192400
  • Applicant: DRTECH Corporation
  • Trade Name: EVS 4343A / EVS 4343AG / EVS 3643A / EVS 3643AG ●
  • Classification Name: Stationary X-ray System ●
  • Classification Panel: Radiology ●
  • Classification Regulation: 21 CFR 892.1680
  • Product Code: MQB ●
  • Device Class: II ●

{4}------------------------------------------------

5. Description of the Modified Device [21 CFR 807.92(a) (4)]

Addition of EXPD 86P, EXPD 86PG, EXPD 129P, EXPD 129PG: The differences between the subject devices and the predicate devices are Dimension and Configuration of Components(Optional). The Dimension of Subject devices is longer than predicate devices.

6. Indication for Use [21 CFR 807.92(a)(5)]

The EXPD 86P, EXPD 86PG, EXPD 129P, EXPD 129PG Digital X-ray detector is indicated for digital imaging solution designed for providing general radiographic diagnosis of human anatomy. This device is intended to replace film or screen based radiographic systems in all general purpose diagnostic procedures. This device is not intended for mammography applications.

7. Technological Characteristics [21 CFR 807.92(a)(6)]

The EXPD 86P, EXPD 86PG, EXPD 129P, EXPD 129PG is a flat-panel type digital X-ray detector that captures projection radiographic images in digital format within seconds, eliminating the need for an entire x-ray film or an image plate as an image capture medium. EXPD 86P, EXPD 129P, EXPD 129PG differs from traditional X-ray systems in that, instead of exposing a film and chemically processing it to create a hard copy image, a device called a Detector is used to capture the image in electronic form.

The EXPD 86P, EXPD 86PG, EXPD 129P, EXPD 129PG Detector is an indirect conversion device in the form of a square plate in which converts the incoming X-rays into visible light. This visible light is then collected by an optical sensor, which generates an electric charges representation of the spatial distribution of the incoming X-ray quanta.

The charges are converted to a modulated electrical signal thin film transistors. The amplified signal is converted to a voltage signal and is then converted from an analog to digital signal which can be transmitted to a viewed image print out, transmitted to remote viewing or stored as an electronic data file for later viewing.

Comparisons with the predicate, devices show the technological characteristics of the EXPD 86P, EXPD 86PG, EXPD 129P, EXPD 129PG to be same to the predicate devices. EXPD 86P, EXPD 86PG, EXPD 129P, EXPD 129PG is functionally identical to the predicate devices.

{5}------------------------------------------------

DRTECH

8. Intergrated Configuration for X-Ray system

The EXPD 86P, EXPD 86PG, EXPD 129P, EXPD 129PG is connected with General diagnostic X-ray system.

Image /page/5/Figure/4 description: This image shows a diagram of an X-ray system setup. The diagram includes components such as an X-ray System Generator, EXPD129P, Console PC, and Hospital Network. Cables labeled as LLD POE Cable, Ethernet Cable, and LLD SYNC Cable connect these components, with annotations indicating voltage, Gigabit Ethernet, and bidirectional TTL signal specifications.

No.CableCommunication
LLD POE Cable1) DC 24V voltage2) It is supplied from Adaptor to Detector.
LLD POE Cable1) Gigabit Ethernet Communications2) It is transmitted from the detector to the CONSOLE PC.
LLD SYNC Cable1) two-way TTL communication2) It is communicated between the detector and the X-ray System Generator.
LAN Cable1) 1) DICOM communication2) It is transmitted from the CONSOLE PC to the hospital network.

{6}------------------------------------------------

DRTECH

For successful integration of detector with X-ray system, we designed the hardware of EXPD 86P, EXPD 86PG, EXPD 129PG as below function. The Energy of generator is controlled by Radiography imaging software(Econsole1, K152172), not Detector.

ModeDescription
AED/AWC ModeThe detector detects actual amount of X-rays without any connection to the X-raygenerator.No signal used (No need of connector interface cable)
Sync. Trigger ModeThe detector receives EXP_REQ signal that X-ray generator is prepared togenerate X-rays.The detector prepares image acquiring and then responds EXP_OK signal to theX-ray generator.The X-ray generator confirms EXP_OK signal and generates X-ray, then thedetector performs image acquiring, according to image acquisition time andtransmits the image data.EXP_REQ (Generator→ Detector), EXP_OK (Detector →Generator)

9. Substantial Equivalence [21 CFR 807.92(b)]

ParameterSubject DevicePredicate Device
510(K) NumberUnknownK192400
ManufacturerDRTECH CorporationDRTECH Corporation
Model NameEXPD 129P/ EXPD 129PG/EXPD 86P/ EXPD 86PGEVS 4343A / EVS 4343AGEVS 3643A / EVS 3643AG
Classification NameStationary X-ray System
Classification PanelRadiology
ClassificationRegulation21 CFR 892.1680
Product CodeMQB
Device ClassClass II
Intended UseThe EXPD 86P, EXPD 86PG, EXPD129P, EXPD 129PG Digital X-raydetector is indicated for digital imagingsolution designed for providing generalradiographic diagnosis of humananatomy. This device is intended toreplace film or screen basedradiographic systems in all generalpurpose diagnostic procedures. Thisdevice is not intended formammography applicationsThe EVS 4343A / EVS 4343AG / EVS3643A / EVS 3643AG Digital X-raydetector is indicated for digital imagingsolution designed for providing generalradiographic diagnosis of humananatomy. This device is intended toreplace film or screen basedradiographic systems in all generalpurpose diagnostic procedures. Thisdevice is not intended formammography applications.

{7}------------------------------------------------

ParameterSubject DevicePredicate Device
DetectorSizeEXPD 86P/ EXPD 86PG: Rectangular PanelEVS 3643A: Rectangular PanelEVS 3643AG: Rectangular Panel
EXPD 129P/ EXPD 129PG: 17" X 51"EVS 4343A: 17" X 17"EVS 4343AG: 17" X 17"
EXPD 86P/ EXPD 86PG: 17" X 34"EVS 3643A: 13" X 17"EVS 3643AG: 13" X 17"
DimensionsEXPD 129P/EXPD 129PG: 460mm (W) × 1324mm (L) × 20mm (H) [±0.5mm]EVS 4343A:460(W) x 483(L) x 15.5(H)EVS 4343AG:460(W) x 483(L) x 15.5(H)
PD 86P/EXPD 86PG: 460mm (W) × 894.14mm (L) × 15.5mm (H) [±0.5mm]EVS 3643A:460(W) × 409(L) × 15.5(H)EVS 3643AG:460(W) × 409(L) × 15.5(H)
Pixel Pitch140μm140μm
Image SizeEXPD 129P/EXPD 129PG: 3,072 x 9,216EVS 4343A: 3,072 x 3,072EVS 4343AG: 3,072 x 3,072
EXPD 86P/EXPD 86PG: 3,072 x 6,144EVS 3643A: 2,560 x 3,072EVS 3643AG: 2,560 x 3,072
Scintillator & TFTMaterialsTFT -amorphous SiliconTFT -amorphous Silicon
EXPD 129P/ EXPD 86P: CsIEVS 4343A/EVS 3643A: CsI
EXPD 129PG/ EXPD 86PG: GadoxEVS 4343AG/ EVS 3643AG: GOS
PerformanceDQEEXPD 129P: 50.0 % at 0.5lp/mmEVS 4343A: 52.9% at 1.0 lp/mm
EXPD 129PG: 25.0 % at 0.5lp/mmEVS 4343AG: 27.2 % at 1.0 lp/mm
EXPD 86P: 50.0 % at 0.5lp/mmEXPD 86PG: 25.0 % at 0.5lp/mmEVS 3643A: 50.5 % at 1.0 lp/mmEVS 3643AG: 26.3 % at 1.0 lp/mm
MTFEXPD 129P: 45.0 % at 2.0 lp/mmEXPD 129PG: 45.0 % at 2.0 lp/mmEXPD 86P: 45.0 % at 2.0 lp/mmEXPD 86PG: 45.0 % at 2.0 lp/mmEVS 4343A: 44.1 % at 2.0 lp/mmEVS 4343AG: 49.2 % at 2.0 lp/mmEVS 3643A: 44.5% at 2.0 lp/mmEVS 3643AG: 46.3 % at 2.0 lp/mm
Resolution3.5 lp/mm3.5 lp/mm
Anatomical SitesGeneral RadiographyGeneral Radiography
Power Supply100240V, 50/60 Hz100240V, 50/60 Hz
ComponentAdaptor & AC Power CordLLD POE CableLLD Sync CablePoE AdaptorLan cables(2 ea)AC Power CableEthernet Connector Bracket
CommunicationWired: EthernetWired: Ethernet
ParameterModified DevicePredicate DeviceRemark
DOEEXPD 129P: 50.0 % at 0.5lp/mmEVS 4343A: 52.9 % at 1.0 lp/mm
EXPD 129PG: 25.0 % at 0.5lp/mmEVS 4343AG: 27.2 % at 1.0 lp/mm
EXPD 86P: 50.0 % at 0.5lp/mmEVS 3643A: 50.5 % at 1.0 lp/mm
EXPD 86PG: 25.0 % at 0.5lp/mmEVS 3643AG: 26.3 % at 1.0 lp/mm
MTFEXPD 129P: 45.0 % at 2.0 lp/mmEVS 4343A: 44.1 % at 2.0 lp/mm
EXPD 129PG: 45.0 % at 2.0 lp/mmEVS 4343AG: 49.2 % at 2.0 lp/mm
EXPD 86P: 45.0 % at 2.0 lp/mmEVS 3643A: 44.5 % at 2.0 lp/mm
EXPD 86PG: 45.0 % at 2.0 lp/mmEVS 3643AG: 46.3 % at 2.0 lp/mm

DRTECH

{8}------------------------------------------------

When compared to the predicate devices (K192400), the EXPD 86P, EXPD 86PG, EXPD 129P, EXPD 129PG presented in this submission have similar:

  • . Intended Use
  • . Technological characteristics
  • . Operating principle
  • Performance (Resolution)
  • Communication Method
  • . Performance (DQE and MTF)
  • . Panel Material
  • . Software(Firmware)

A few differences are as follows

  • . Size
  • Configuration of Components(Optional)

There are no significant differences between the EXPD 86PG, EXPD 129P, EXPD 129PG and the predicate device(s) that would adversely affect the use of the product. It is substantially equivalent to these devices in design, function, materials, operational principles and intended use.

According to bench test report, it is proved that the DQE and MTF of predicated device and subject device are basically equal or worth than the predicate device. As a result, subject devices performance is equal or worth than the predicate device.

10. Summary of Non-Clinical Data [21 CFR 807.92(b)(1)]

The non-clinical performance testing constrains that the main physical values for comparison of Xray devices like DQE and MTF are basically equal or worth than the predicate device as following table:

The EXPD 86P / EXPD 86PG/ EXPD 129P/ EXPD 129PG detector complies with the following international and FDA-recognized consensus standards:

Recognition No.Standard No.Title of StandardRemark
----------------------------------------------------------

{9}------------------------------------------------

Recognition No.Standard No.Title of StandardRemark
19-46ANSI AAMI ES60601-1:2005/(R)2012 & A1:2012,C1:2009/(R)2012 &A2:2010/(R)2012 (Cons. Text)[Incl. AMD2:2021]Medical electrical equipment - Part 1:General requirements for basic safetyand essential performance (IEC 60601-1:2005, MOD) [Including Amendment2 (2021)]
12-289IEC 62220-1-1 Edition 1.0 2015-03Medical electrical equipment-Characteristics of digital X-ray imagingdevices Part 1-1: Determination of thedetective quantum efficiency Detectorsused in radiographic imaging
5-125ISO 14971 Third Edition 2019-12Medical devices - Application of riskmanagement to medical devices
5-132IEC 60601-1-6 Edition 3.2 2020-07CONSOLIDATED VERSIONMedical electrical equipment - Part 1-6:General requirements for basic safetyand essential performance - Collateralstandard: Usability
12-342PS 3.1 - 3.20 2021eDigital Imaging and Communications inMedicine (DICOM) Set
19-36IEC 60601-1-2 Edition 4.1 2020-09CONSOLIDATED VERSIONMedical electrical equipment - Part 1-2:General requirements for basic safetyand essential performance - CollateralStandard: Electromagnetic disturbances- Requirements and tests
13-79IEC 62304 Edition 1.1 2015-06CONSOLIDATED VERSIONMedical device software - Software lifecycle processes

DRTECH

11. Related FDA Guidance documents

We reviewed the below FDA guidance documents and it was reflected to development of the modified EXPD 86P/ EXPD 86PG/ EXPD 129P/ EXPD 129PG detector.

Title of GuidanceRemark
Content of Premarket Submissions for Management of Cybersecurity in Medical Devices
Pediatric Information for X-ray Imaging Device Premarket Notifications

{10}------------------------------------------------

13. Conclusion [21 CFR 807.92(b)(3)]

The modified EXPD 86P / EXPD 86PG/ EXPD 129P/ EXPD 129PG detector is substantially equivalent to the currently marketed and predicate device (EVS 4343A / EVS 4343AG / EVS 3643A / EVS 3643AG - K192400) in terms of fundamental scientific technology, indications for use, and safety and effectiveness.

Additionally, Substantial equivalence was demonstrated through the non-clinical performance, which complied with the requirements specified in the international and FDA-recognized consensus standards, ANSI AAMI ES60601-1:2005/AMD2:2021, IEC 60601-1-2:2020, IEC 60601-1-6 Edition 3.2, IEC 62304:2015 and IEC 6220-1-1 which complied with the requirements specified in the CDRH's Guidance for the Submission of 510(k)'s for Solid State X-ray Imaging Devices.

The results of these tests demonstrate that The EXPD 86P / EXPD 86PG/ EXPD 129P/ EXPD 129PG meets the acceptance criteria and is adequate for this intended use. The comparison of technological characteristics, non-clinical performance data and safety testing demonstrates that the device is as safe, as effective, and performs as well or better than the predicate devices.

N/A