Found 7 results
510(k) Data Aggregation
(32 days)
Intended for use by a qualified/trained physician or technician on both adult and pediatric subjects for taking diagnostic x-rays. Not for mammography.
AcuityDRe components into a complete digital x-ray system upgrade kit, including software and digital radiography panels. The customer selects one of the following digital x-ray receptor panels: AcuityDRe 1417w, AcuityDRe 1717w, AcuityDRe 1717t. The "w" indicates wireless wi-fi while the "t" indicates tethered. The indications for use remains unchanged: Intended for use by a qualified/trained physician or technician on both adult and pediatric subjects for taking diagnostic x-rays. Not for mammography. So the only difference between this submission and the predicate submission is the generator/tubestand combination. Each system consists of the following items: Customer supplies: Diagnostic x-ray generator (HF) Class I Code IZO. + Tubehead: Class I Code ITY + Tube Mount: Class I Code IYB + Attached Collimator, Manual (IZX) Class II 510(k) Exempt We supply: Digital X-Ray Receptor Panel 892.1680 Class II Code MQB. Digital X-ray Software 892.2050 Class II Code LLZ. The software offered for sale with this system has received previous 510(k) clearance in K201058.
The provided text describes the 510(k) premarket notification for the AcuityDRe device, which is an upgrade kit for digital x-ray systems. It primarily focuses on demonstrating substantial equivalence to a predicate device (Accuvue+, K201058), rather than detailing a clinical study with specific acceptance criteria and performance against those criteria in a typical sense of an AI/ML medical device.
The "study" cited here is non-clinical testing, specifically clinical image evaluation performed by a Board Certified Radiologist, and bench testing for physical performance characteristics. It's not a comparative effectiveness study involving human readers with and without AI assistance, nor a standalone AI performance study.
Here's an attempt to extract and present the information based on the provided text, acknowledging that many requested fields regarding AI/ML device studies are not applicable or detailed in this 510(k) summary for an x-ray hardware upgrade kit.
Device: AcuityDRe (Digital X-ray System Upgrade Kit)
Description of Testing (Non-Clinical and Image Evaluation)
The primary goal of the testing was to demonstrate substantial equivalence of the AcuityDRe system (comprising new digital x-ray receptor panels, software, and generator compatibility) to a legally marketed predicate device (Accuvue+, K201058). This was achieved through:
- Bench testing for physical performance characteristics of the new digital x-ray panels.
- Clinical image evaluation by a Board Certified Radiologist to assess image quality.
- Confirmation of compliance with relevant medical device standards and FDA guidance documents.
1. Table of Acceptance Criteria and Reported Device Performance
Given that this 510(k) is for an X-ray system upgrade kit rather than an AI/ML algorithm, the "acceptance criteria" are not framed as typical AI performance metrics (e.g., sensitivity, specificity, AUC). Instead, the key "acceptance criteria" are based on demonstrating substantial equivalence to the predicate device in terms of:
- Same Indications for Use
- Similar Technological Characteristics (especially panel performance: DQE, MTF, and panel sizes)
- Compliance with safety and performance standards.
- "Excellent quality" of clinical images as assessed by an expert.
Feature / Criteria (Implied for Substantial Equivalence) | Acceptance Criteria (Compared to Predicate / Standards) | Reported Device Performance (AcuityDRe) | Comparison Result |
---|---|---|---|
Indications for Use | Same as predicate. | "Intended for use by a qualified/trained physician or technician on both adult and pediatric subjects for taking diagnostic x-rays. Not for mammography." - SAME as predicate. | SAME |
X-ray Generator Compatibility | Compatible with CPI or Sedecal generators. | Compatible CPI generators: CMP 200 Series. Compatible Sedecal generators: SFHR and SHF Series. - SAME as predicate. Software control of technique factor possible with certain compatible generators. | SAME |
Digital X-Ray Detectors | New models, but comparable performance and size. | AcuityDRe 1417w, AcuityDRe 1717w, AcuityDRe 1717t. New models had not received previous FDA clearance, hence testing was performed for them. | SAME (as new versions) |
Panel Performance (DQE @ 1.0 lp/mm) | Comparable to predicate's DQE. | AcuityDRe 1417w: 35% (Predicate AcuityDR 1417: 34.6%) AcuityDRe 1717w/1717t: 42% (Predicate AcuityDR 1717: 23.6%) (Note: The AcuityDRe 1717w/1717t shows improved DQE compared to the predicate's AcuityDR 1717, which is a positive attribute for image quality). | Almost IDENTICAL / Improved for some models |
Panel Performance (MTF @ 2.0 lp/mm) | Comparable to predicate's MTF. | AcuityDRe 1417w: 31% (Predicate AcuityDR 1417: 34%) AcuityDRe 1717w/1717t: 38% (Predicate AcuityDR 1717: 34%) (Note: Ranges are close, with slight variations). | Almost IDENTICAL |
Panel Sizes | Comparable to predicate's sizes. | AcuityDRe 1417w: 148 µm AcuityDRe 1717w: 140 µm AcuityDRe 1717t: 140 µm (Predicate AcuityDR 1417: 140µm; AcuityDR 1717: 140μm) | Almost IDENTICAL |
Operator Console | Windows PC using Windows 10-IoT. | SAME | SAME |
Acquisition Software | AccuVue / AccuVue+. | SAME | SAME |
Power Source | AC Line or rechargeable batteries. | SAME | SAME |
Compliance with Standards | Demonstrated compliance with listed standards. | IEC 60601-1:2005/(R)2012 And A1:2012 (Medical Electrical Equipment Safety) IEC 60601-1-2:2014 (Electromagnetic Disturbances) NEMA PS 3.1 - 3.20 (2011) (DICOM Set) - SAME as predicate. | SAME |
Clinical Image Quality | "Of excellent quality" (qualitative assessment). | "The images were found to be of excellent quality." (Qualitative assessment by expert) | Achieved |
2. Sample Size Used for the Test Set and Data Provenance
- Test Set Sample Size: Not explicitly stated numerically for the clinical image evaluation. The text only mentions "Clinical image evaluation was performed on the proposed new panels."
- Data Provenance: Not specified. It's likely retrospective, as it's an evaluation of images generated by the new panels. Country of origin not mentioned; presumed to be images from typical clinical settings where such panels would be used.
3. Number of Experts Used to Establish Ground Truth and Qualifications
- Number of Experts: One.
- Qualifications: A "Board Certified Radiologist." Further details like years of experience are not provided.
4. Adjudication Method for the Test Set
- Adjudication Method: Not applicable/None, as only a single expert was used for the qualitative assessment of image quality.
5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study
- MRMC Study: No. This type of study (AI vs. human-assisted AI) was not performed as the device is an x-ray system upgrade, not an AI diagnostic algorithm.
- Effect Size of Human Reader Improvement: Not applicable.
6. Standalone (Algorithm Only Without Human-in-the-Loop Performance) Study
- Standalone Study: No, as the device itself is a component of an x-ray system, not a standalone AI algorithm. The image quality assessment was of the images produced by the hardware, which would then be interpreted by humans.
7. Type of Ground Truth Used
- Ground Truth Type: Expert Consensus / Qualitative Expert Assessment. The "ground truth" for the image quality was the subjective assessment by the Board Certified Radiologist that the images were "of excellent quality." This is distinct from, for example, pathology-confirmed diagnoses or patient outcomes data.
8. Sample Size for the Training Set
- Training Set Sample Size: Not applicable. This device is a hardware component (digital x-ray panels) and associated software, not an AI/ML algorithm that requires a "training set" in the context of machine learning model development. The software mentioned (AccuVue/AccuVue+) has received previous clearance for control of generators.
9. How the Ground Truth for the Training Set was Established
- Ground Truth Establishment for Training Set: Not applicable, as there is no "training set" in the context of an AI/ML algorithm for this device.
Ask a specific question about this device
(102 days)
Intended for use by a qualified/trained doctor or technician on both adult and pediatric subjects for taking diagnostic radiographic exposures of the skull, spinal column, extremities, and other body parts. Applications can be performed with the patient sitting, standing, or lying in the prone or supine position. Not for mammography.
This is a multifunctional fixed X-ray equipment consisting of Examination Table, Bucky Stand, Tube/Bucky Stand, X-ray Source Assembly (Tube/Collimator), High Voltage Generator plus Operating Panel, Conventional Image Receivers and, in specific version, Image System with Flat panel detector and Workstation for image acquisition, processing and visualization. The equipment was developed to perform radiographic examinations of patients in reclining, standing or sitting positions. Six possible tube head configurations are available: FLOOR TO CELLING TUBE STAND, FLOOR MOUNTED TUBE STAND, ROTARY U-ARM mount with integrated film/digital cassette mount, MANUAL CEILING-MOUNTED TELESCOPIC TUBE STAND, MOTORIZED CEILING-MOUNTED TELESCOPIC TUBE STAND, ROTARY STRAIGHT ARM mount with integrated film/digital cassette mount. The Apolo provides a complete x-ray system with generator, tube head, and collimator. The generator is made by us, VMI, whereas the tube head, collimator, and digital x-ray receptor panels/software are made by other manufacturers. Tube heads are typically Toshiba and collimators are typically Ralco. The Apolo D comes with digital x-ray receptor panels (see the comparison table below) while the Apolo S comes without digital x-ray receptor panels. This system employs without modification software Econsole cleared in K152172. This software has a Moderate level of concern. The reference detectors made by DRTECH can be used with anti-scatter grids.
This document is a 510(k) premarket notification for the Apolo D / Apolo S stationary x-ray system. It does not contain information about an AI/ML-based device or a study involving acceptance criteria for such a device. The device described is a traditional x-ray system, and the clearance is based on substantial equivalence to a predicate device, not on AI-driven performance metrics.
Therefore, I cannot extract the requested information (acceptance criteria, details of a study proving device meets acceptance criteria, sample sizes, expert involvement, adjudication, MRMC, standalone performance, ground truth, training set specifics) from the provided text because these elements are related to the evaluation of AI/ML devices, which is not applicable here.
The document explicitly states regarding clinical testing:
"Clinical testing was not required to establish substantial equivalence because all digital x-ray receptor panels have had previous FDA clearance." This further confirms that no clinical performance study, particularly one measuring AI-driven diagnostic accuracy, was conducted for this submission.
Ask a specific question about this device
(30 days)
Intended for use by a qualified/trained doctor or technician on both adult and pediatric subjects for taking diagnostic radiographic exposures of the skull, spinal column, extremities, and other body parts. Applications can be performed with the patient sitting, standing, or lying in the prone or supine position. Not for mammography.
Aquila 320-D Series: Small, light, easy handling. High Performance, High Power 320 mA, Leading Technology. Capacitive discharge technology, wireless image capture and transmission technology, smartphone image access. Wireless Detector and Workstation on touchscreen notebook or tablet. High frequency generator with micro-processor controls: Power 35.2 KW (The Aquila 320-S comes without the digital panels and workstation)
The provided text is a 510(k) summary for the AQUILA 320 D / AQUILA 320 S mobile x-ray system. It details the device, its intended use, and its substantial equivalence to a legally marketed predicate device. However, it does not describe an AI-powered device or a study where an AI device meets acceptance criteria.
The document explicitly states regarding clinical testing:
"Clinical testing was not required to establish substantial equivalence because all digital x-ray receptor panels have had previous FDA clearance."
Therefore, I cannot provide the requested information about acceptance criteria and a study proving an AI device meets them based on the provided text, as this document pertains to a traditional medical imaging device (mobile x-ray system) and not an AI-powered one, and no clinical study was conducted or described.
Ask a specific question about this device
(15 days)
Intended for use by a qualified/trained physician or technician on both adult and pediatric subjects for taking diagnostic x-rays. Not for mammography.
Accuvue+ components into a complete digital x-ray system upgrade kit, including software and digital radiography panels. The customer selects one (or more) of the following digital x-ray receptor panels: DRTech 4343A, (K192400); DRTech 4343W, (K193017); AcuityDR (K171137); AcuityDR 1417 (K162552 EVS 3643G) or AcuityDR 1717 (K162555 EVS 4343, EVS 4343G). We wish to offer for sale a subset of our recently cleared submission K200726. This in essence would be an upgrade kit. The indications for use remains unchanged: Intended for use by a qualified/trained physician or technician on both adult and pediatric subjects for taking diagnostic x-rays. Not for mammography. So the only difference between this submission and the predicate submission is the generator/tubestand combination.
This document is a 510(k) summary for a medical device called Accuvue+. It describes the device and its intended use, and argues for its substantial equivalence to a predicate device. However, it does not contain information about acceptance criteria or a study proving that the device meets specific performance criteria through a clinical or non-clinical study.
The document explicitly states:
- "6. Clinical testing. Not required for a determination of substantial equivalence."
- "5. Non clinical testing: Testing was performed in previous submssions according to the following standards:" (It then lists general safety and communication standards, not performance criteria for diagnostic accuracy or image quality.)
Therefore, I cannot extract the requested information regarding acceptance criteria and a study proving the device meets those criteria from the provided text. The submission focuses on demonstrating substantial equivalence based on technological characteristics and adherence to general safety standards, rather than a performance study with acceptance criteria.
Ask a specific question about this device
(24 days)
Intended for use by a qualified/trained physician or technician on both adult and pediatric subjects for taking diagnostic x-rays. Not for mammography.
RadmediX AcuityPDR combines components into a complete mobile x-ray system, including software, a generator/collimator combination, and digital radiography panels. Radmedix combines components from various manufacturers into a complete mobile x-ray system. The customer selects one (or more) of the following digital x-ray receptor panels: DRTech 4343A, (K192400); DRTech 4343W, (K193017); AcuityDR (K171137); AcuityDR 1417 (K162552 EVS 3643, EVS 3643G) or AcuityDR 1717 (K162555 EVS 4343, EVS 4343G). In addition, the customer selects one of three software packages: Accuvue+ (K130883), AccuVueMED, (K152172) or AccuVue (K141440). The generator can be battery operated. A single battery charge produces 200 Exposures at max KV and MAS settings. The battery charger is UL Listed and the internal lithium ion battery is overcharge and overcurrent protected. A typical acquisition computer would be a Lenovo P53S or a Dell Precision 3541.
This document describes a medical device, AcuityPDR, a mobile X-ray system, and its 510(k) submission for FDA clearance. The submission relies on demonstrating substantial equivalence to a predicate device (MinXray CMDR 2CW).
Here's an analysis of the requested information based on the provided text:
1. A table of acceptance criteria and the reported device performance
The document does not explicitly state acceptance criteria in the form of quantitative performance metrics for the AcuityPDR device itself. Instead, it focuses on demonstrating "substantial equivalence" to a predicate device. The acceptance criteria for this type of submission are typically that the device performs as safely and effectively as the predicate and has the same indications for use.
The device performance is reported implicitly through the comparison table on page 4, showing that the AcuityPDR either matches or has comparable/equivalent functionality to the predicate in various technical parameters.
Acceptance Criteria (Implicit for Substantial Equivalence) | Reported Device Performance (AcuityPDR vs. Predicate CMDR 2CW) |
---|---|
Indications for Use (Identical) | Intended for use by a qualified/trained physician or technician on both adult and pediatric subjects for taking diagnostic x-rays. Not for mammography. (Identical) |
X-ray Tube Functionality (Equivalent/Comparable) | TOSHIBA D-125SB (Equivalent Functionality) |
Peak Generator Output (Same) | 2 kW (SAME) |
Tube Current Range (Comparable) | 40-60kV: 25mA, 61-100kV: 20mA (Comparable ranges) |
Tube Voltage Adjustable Range (More Flexible) | 40-100kV, step value 1kV (More flexible kV adjustment) |
mAs Range (Comparable Adjustability) | 0.4mAs ~ 50mAs (Comparable adjustability) |
Collimator (Equivalent Functionality) | Built in (Equivalent Functionality) |
X-ray Generator (Equivalent Functionality) | One model, up to 100 kVp (Equivalent Functionality) |
Digital X-Ray Detectors (Equivalent Functionality) | DRTech 4343A, DRTech 4343W, AcuityDR (various models) (Equivalent Functionality) |
Operator Console (Similar Functionality) | SAME (Touch Control or Touch Screen) (Similar Functionality) |
Acquisition Software (Equivalent Functionality) | AccuVueMED, AccuVue, or AccuVue+ (Equivalent Functionality, only cleared software supplied) |
Power Source (Same) | AC Line or rechargeable batteries (Generator only) (SAME) |
Compliance with Standards (Same) | 60601-1:2005; 60601-1-2:2014, etc. (SAME) |
2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)
The document explicitly states in section 6, "Clinical testing. Not required for a determination of substantial equivalence." This indicates that no clinical test set was used for this specific submission. The submission relies on non-clinical testing (bench testing, standards compliance) and comparison to the predicate device.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)
Since no clinical test set was required or used for this submission, there is no mention of experts being used to establish ground truth for a test set.
4. Adjudication method (e.g. 2+1, 3+1, none) for the test set
As no clinical test set was used, no adjudication method is mentioned.
5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
No MRMC study was performed as clinical testing was "Not required for a determination of substantial equivalence." The device is a mobile X-ray system, not an AI-powered diagnostic tool, so the concept of human readers improving with AI assistance is not applicable in this context.
6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done
This is not an AI algorithm; it's a mobile X-ray system. Therefore, standalone algorithm performance is not applicable. The performance is assessed through compliance with electrical and safety standards for X-ray devices and comparison of technical specifications to a predicate.
7. The type of ground truth used (expert concensus, pathology, outcomes data, etc)
For this type of 510(k) submission, "ground truth" is primarily established by:
- Compliance with recognized electrical, safety, and performance standards for X-ray generating equipment (e.g., IEC 60601 series, 21 CFR 1020).
- Demonstration that the device's technical specifications and intended use are substantially equivalent to a legally marketed predicate device.
- Component certifications: "All of the components subject to the CDRH performance standard are certified to comply with the standard by their respective manufacturers."
8. The sample size for the training set
The document does not describe a training set for an AI model, as this is not an AI device. The "training" for such a device would refer to its design and manufacturing processes adhering to established engineering principles and standards.
9. How the ground truth for the training set was established
Not applicable, as no AI model or training set is described. The "ground truth" for the device's design and manufacturing relies on established medical device regulations, design controls, and industry standards for X-ray imaging equipment.
Ask a specific question about this device
(29 days)
The purpose of Acuity is to acquire, store, communicate, display and process medical X-ray images. These radiographic systems are intended for use by a qualified/trained physician or technician on both adult and pediatric subjects for taking diagnostic x-rays. Not for mammography, angiography, interventional, or fluoroscopy use.
RadmediX Acuity combines components into a complete stationary x-ray system, including software, tube stands, tube heads, collimators, generators, tables, and digital radiography panels.
Here's a breakdown of the acceptance criteria and the study information based on the provided text:
Acceptance Criteria and Device Performance:
The document describes the Acuity SDR Standard, Acuity SDR Plus, and Acuity FDR Standard as stationary x-ray systems. The core claim for substantial equivalence is that the device performs the same functions using the same technological methods as the predicate device (Visaris Vision®) to produce diagnostic x-ray images.
Feature / Criterion | Acceptance Criteria (Implied by Predicate Equivalence) | Reported Device Performance (Acuity Systems) |
---|---|---|
Indications for Use | Acquisition, storage, communication, display, and processing of medical X-ray images for diagnostic X-rays on adult and pediatric subjects. Not for mammography, angiography, interventional, or fluoroscopy use. | Identical to predicate. |
Wall Stand Functionality | Motorized vertical movable wall stand, tiltable tray. | "SAME" – Equivalent Functionality. |
Table Functionality | Free-floating and height-adjustable, maximum patient weight 660 lbs., working table height 20-5/16 inch to 37-5/8 inch. | "SAME" – Equivalent Functionality. |
X-ray Tube Specifications | 150 kVp 0.6/1.2mm focal spots (Toshiba). | "SAME" – Equivalent Functionality. |
Collimator Certification | All CFR Certified. (510(k) exempt) | Collimaire collimators, all CFR Certified. (510(k) exempt) – Equivalent Functionality. |
X-ray Generator Types and Certification | Various HF Models (Claymount up to 63 kW, CPI 32kW to 100kW, EMD 45 kW to 80 kW, POSKOM 32 kW to 50 kW, Sedecal 40, 50, 65, or 80 kW), all CFR Certified. | Various HF Models (Summit 40 to 50 kW, CPI up to 80 kW), all CFR Certified. (510(k) exempt) – Equivalent Functionality. |
Wireless Detector Capability | Uses FDA cleared 14" x 17" detectors and software (e.g., Pixium 3543 EZ C). | Optional enclosure for 14" x 17" fixed and removable FDA cleared detectors and software. – Equivalent Functionality. |
Fixed Detector Capability | Uses FDA cleared 17"x17" detectors and software (e.g., Pixium 4343RC). | Uses FDA cleared 17"x17" detectors and software. – Equivalent Functionality. -------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------- |
Compatibility with Conventional Film/CR Cassettes | Comes with FDA cleared digital x-ray panels, but conventional film and CR cassettes can still be used. | "SAME" – Similar Functionality. -------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------- |
Operator Console Interface | GUI-based. | "SAME" – Similar Functionality. ---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------- |
Acquisition Software | FDA cleared digital x-ray acquisition software (e.g., Visaris Avanse, K150725). | Customer selects AccuVueMED (K152172) or AccuVue (K141440). Only cleared software is supplied. – Equivalent Functionality. ---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------- |
Power Source | AC LINE, single or three phase depending on the generator. | "SAME" – SAME. -------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------- |
Compliance with Standards | IEC 60601-1:2005 (Basic Safety and Essential Performance), IEC 60601-1-2:2014 (Electromagnetic Disturbances), IEC 60601-2-54 Ed 1.0 (X-Ray Equipment for Radiography and Radioscopy), NEMA PS 3.1 - 3.20 (2011) DICOM Set, 21CFR1020 (Electronic Products; Performance Standard for Diagnostic X-Ray Systems). Component certification by manufacturers. | "SAME" – Generators are UL Listed. Components certified by manufacturers. Software used unmodified from previous FDA clearances. – SAME. ---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------- |
Image Quality (Clinical Acceptability) | Produce images of high quality and contrast, clinically acceptable for diagnostic x-rays across various anatomical regions. | A constructed and field-tested complete system (AcuitySDR with specific components) acquired DICOM images from all major body structures, all images were of high quality and contrast and clinically acceptable. Each new configuration undergoes a digital image integration protocol and a total system validation protocol prior to ensure the configuration works as intended. "The flat-panel detectors used with the x-ray system Acuity present comparable or better performance than the predicate detectors." |
Cybersecurity | Managed as per FDA guidance "Content of Premarket Submissions for Management of Cybersecurity in Medical Devices Guidance for Industry and Food and Drug Administration Staff October 2014". | Updated internal standard operating procedures and added cybersecurity precautions to software users' manuals. ---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------- |
Pediatric Considerations in User Manuals | Included in user manuals. | Generator User Manuals have been updated to add Pediatric Considerations. ---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------- |
Study Information:
-
Sample Size and Data Provenance:
- Test Set Sample Size: Not explicitly stated as a numerical sample size of clinical cases for a formal clinical trial. Instead, the non-clinical testing involved acquiring DICOM images from "all major body structures" using one AcuitySDR system configuration.
- Data Provenance: The document does not specify the country of origin for the image acquisition or whether it was retrospective or prospective. It was a "field tested" system, implying real-world or simulated real-world scenarios.
-
Number of Experts and Qualifications (for Ground Truth):
- Number of Experts: Not specified.
- Qualifications of Experts: Not specified, beyond the statement that images were generated for "diagnostic x-rays" and assessed as "clinically acceptable." This implies evaluation by trained personnel (e.g., radiologic technologists, radiologists), but their specific qualifications or number are not detailed.
-
Adjudication Method:
- No formal adjudication method is described. The assessment of image quality was based on whether images were "of high quality and contrast and clinically acceptable."
-
Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study:
- No MRMC comparative effectiveness study was performed or described. The submission focuses on demonstrating substantial equivalence through technical characteristics and non-clinical image quality assessment, not on human reader performance improvement with or without AI assistance.
-
Standalone Performance Study (Algorithm Only):
- Yes, a standalone study (non-clinical testing) was performed. The algorithm (software) alongside the hardware components was tested to acquire and process images. The output (DICOM images) was then evaluated for inherent quality and acceptability. The assessment "all images were of high quality and contrast and clinically acceptable" refers to the output of the integrated system.
-
Type of Ground Truth Used:
- Expert Consensus / Clinical Acceptability: The ground truth for image quality was established through subjective expert assessment of "high quality and contrast and clinically acceptable." It does not mention pathology or outcomes data.
-
Training Set Sample Size:
- Not applicable. This device is an X-ray system, not an AI/ML algorithm that requires a training set in the typical sense for image interpretation or diagnosis. The "software" components mentioned (AccuVueMED, AccuVue) are image acquisition and processing software, which are generally deterministic and not "trained" on data in the way a diagnostic AI would be. They are "previously cleared software supplied with the system."
-
How Ground Truth for Training Set Was Established:
Ask a specific question about this device
(35 days)
ProRad Series Stationary Radiographic System is intended for use by a qualified, trained doctor or technician on both adult and paediatric subjects for taking diagnostic radiographic exposures of the skull, spinal column, chest, abdomen, extremities, and other body parts. Applications can be performed with the patient sitting, standing, or lying in the prone or supine position. Not for mammography.
The ProRad series Stationary Radiographic System is a diagnostic x-ray system intended for general purpose radiographic imaging of the human body. There are two types of configurations (2FC and 3NC) for ProRad; the difference is in the mounting of the X-ray tube. For X-ray tube mounting the configuration is either the floor mounted (2FC) or ceiling suspension (3NC) assembly.
The devices are a new combination of a previously cleared solid state digital x-ray acquisition panel and software with the diagnostic x-ray components (including Xray tube, high frequency X-ray generator, a tilting vertical bucky, X-ray table and collimator) required to make a complete system. The purchaser may select any of the digital panels and software based on the user's requirements. The other components are also available in different configurations to meet specific customer needs. The X-ray panel and imaging software have been previously cleared by the FDA, and most of the other components are used in previously cleared 510(k) devices.
Here's an analysis of the acceptance criteria and study information for the ProRad 2FC and ProRad 3NC Digital Stationary Radiographic Systems, based on the provided text:
Acceptance Criteria and Device Performance Table:
The document primarily focuses on demonstrating substantial equivalence to predicate devices rather than setting and reporting specific performance metrics with acceptance criteria in a comparative table for novel device features. The "Comparable Properties" table (pages 6-8) is used to show alignment with predicate devices.
Acceptance Criteria Category | Specific Criteria (Implicit from Equivalence) | Reported Device Performance (as stated in the document) |
---|---|---|
Intended Use | Equivalent to predicate devices (K150816, K153119) for diagnostic radiographic exposures of various body parts in adults and pediatrics, excluding mammography. | "Equivalent" to predicate devices. |
Configuration of Digital Panels | Battery or AC operated wireless IEEE 802.11n or Wired Ethernet. | "Same" as predicate devices. |
Digital Panel Models and their Clearance Numbers | Utilize previously cleared FDA digital flat panel detectors (specific K numbers listed in Note 1). | "Similar functionality" to predicate devices (Note 1 states the subject device uses different flat panel detectors that are previously FDA cleared, and testing demonstrates no increased safety concern or effect on effectiveness). |
Image Acquisition Panel Specifications | Comparable pixel resolution and size range (e.g., 3,320 x 3,408 Pixels 125 µm, 3,072 x 3,072 Pixels 139 µm, etc.). | "Similar functionality" to predicate devices (Note 1). |
DICOM Compliance | DICOM 3 compliant. | "Same" as predicate devices. |
WiFi Wireless (if applicable) | Functionality similar to predicate wireless detectors. | "Similar functionality" to predicate devices (specific Canon, Varex, PerkinElmer, and DRTECH wireless detectors listed). |
Image Acquisition Software | Utilize previously cleared FDA image processing software (specific K numbers listed in Note 2). | "Similar functionality" to predicate devices (Note 2 states Prognosys uses previously cleared FDA software and does not modify it). |
Power Source | AC Line, various voltages available. | "Same" as predicate devices. |
X-ray Generator | Range of power ratings (e.g., 32 kW to 80 kW, 125 kV/150 kV). | "Same or similar functionality" to predicate devices (Note 3 states the subject device uses similar X-ray generators with different power ratings and that these generators have been used in previously cleared 510(k) devices). |
X-ray Tubes | Range of kV, focal spot sizes, and heat units (e.g., Toshiba E7239FX, Varex RAD14, etc.). | "Similar functionality" to predicate devices (Note 4 states both configurations provide similar imaging resolution and these tubes have been used in previously cleared 510(k) devices). |
Collimator | Adequate inherent filtration. | "Similar functionality" to predicate devices (Note 5 states the inherent filtration is different but does not affect safety and effectiveness). |
Performance Standard Compliance | Compliance with FDA 21 CFR 1020.30-31 (Performance Standards for Diagnostic X-Ray Systems and Their Major Components). | "Same" as predicate devices. |
Electrical Safety | Compliance with IEC 60601-1. | "Same" as predicate devices. |
Electromagnetic Compatibility (EMC) | Compliance with IEC 60601-1-2. | Device complies with this standard. |
Radiation Protection | Compliance with IEC 60601-1-3. | Device complies with this standard. |
Specific Safety/Performance for X-ray Equipment | Compliance with IEC 60601-2-54. | Device complies with this standard. |
Usability | Compliance with IEC 60601-1-6. | Device complies with this standard. |
Risk Management | Compliance with ISO 14971. | Device complies with this standard. |
Cybersecurity | Adherence to "Content of Premarket Submissions for Management of Cybersecurity in Medical Devices, 2014" guidance for device labeling. | Cybersecurity concerns were addressed based on the US FDA Guidance document. |
1. Sample sized used for the test set and the data provenance:
- Sample Size: Not explicitly stated as a separate "test set" for a dedicated algorithm performance study. Instead, compliance is demonstrated through testing of the integrated system and reliance on previous FDA clearances for individual components.
- Data Provenance: Not applicable in the context of an algorithm performance test set. The clinical images reviewed by a radiologist were "acquired by the device," but the origin (e.g., country, prospective/retrospective status) is not specified.
2. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:
- Number of Experts: One radiologist.
- Qualifications of Experts: Only "a radiologist" is mentioned. Specific qualifications (e.g., years of experience, board certification) are not detailed.
3. Adjudication method for the test set:
- Adjudication method: Not applicable. The radiologist's review was a single assessment, not a consensus or adjudication process among multiple readers.
4. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:
- MRMC Study: No. The document explicitly states: "Since the digital x-ray panels and software have previously received FDA clearance, a clinical study was not required as per the FDA guidance document." The clinical image review was supplementary.
- Effect size of human reader improvement with AI: Not applicable, as no MRMC study or AI assistance evaluation was conducted. The device is an imaging system, not an AI diagnostic tool.
5. If a standalone (i.e. algorithm only without human-in-the loop performance) was done:
- Standalone Performance Study: No. This device is a diagnostic X-ray system, which intrinsically requires a human (a qualified doctor or technician) in the loop for operation and interpretation. The performance of individual cleared components (digital panels, software) was relied upon.
6. The type of ground truth used (expert consensus, pathology, outcomes data, etc):
- Type of Ground Truth: For the supplementary review, the "ground truth" was the radiologist's assessment that the images were "acceptable and allowed the radiologist to make an accurate diagnosis." This is a form of expert opinion on image quality and diagnostic utility, rather than an objective "truth" like pathology or outcomes.
7. The sample size for the training set:
- Sample Size for Training Set: Not applicable. This document does not describe a machine learning algorithm that requires a training set. The device is a conventional X-ray system composed of cleared components.
8. How the ground truth for the training set was established:
- Ground Truth for Training Set: Not applicable, as there is no machine learning algorithm described.
Ask a specific question about this device