510(k) Data Aggregation
(270 days)
Solidex® Ti-Links and Screws is intended to be used in conjunction with endosseous dental implants in the maxillary or mandibular arch to provide support for prosthetic restorations. All digitally designed custom abutments for use with Solidex® Ti-Links is to be sent to a CREODENT HUDSON VALLEY validated milling center for manufacture, or to be designed and manufactured according to the digital dentistry workflow. The digital dentistry workflow integrates multiple components: scan files from intra-oral and lab (desktop) scanners, CAD software, CAM software, ceramic material, milling machine, and associated tooling and accessories.
Solidex® Ti-Links & Screws consists of a two-piece abutment, where the titanium base will be used to support a CAD/CAM designed superstructure (the second part of the two-piece abutment) that composes the final abutment. The system also includes a Solidex® Screw for fixation to the implant body. Solidex® Ti-Link are made of titanium alloy conforming to ASTM F136 standard specification for Wrought Titanium-6Aluminum-4Vanadium ELI (Extra Low Interstitial) Alloy for Surgical Implant Applications and are provided in various prosthetic platform diameters. Solidex® Screw is composed of titanium alloy per ASTM F136.
Solidex® Ti-Links & Screws is designed for fabrication of a patient-specific CAD/CAM zirconia superstructure on which a crown may be placed. There are two-piece abutments for which the second part (or top half) is the zirconia superstructure. They also may be used for support of a crown directly on the Ti-Link. CAD/CAM zirconia superstructure fabrication for Ti-Link is by prescription on the order of the clinician.
Solidex® Ti-Links & Screws all zirconia superstructures for use with the subject device Ti-Link will be manufactured using a digital dentistry workflow, either at a Validated Milling Center (VMC) or at the Point of Care (POC), by using scan files from intra-oral and lab (desktop) scanners, CAD/CAM software, ceramic material, and associated tooling and accessories. The digital dentistry workflow uses scan files from intra-oral and lab (desktop) scanners, CAD/CAM software, ceramic material, and associated tooling and accessories. The digital workflow includes the following products:
-
Ceramic material: Sagemax NexxZr T (K130991)
-
Cement: Panavia SA cement Universal (K183537)
-
Desktop scanner: 3Shape E series Lab Scanner (510(k) exempt under 21 CFR 872.3661)
-
Design software: 3Shape Abutment Designer Software (K151455)
-
Milling machine: PrograMill PM7 (510(k) exempt under 21 CFR 872.3661)
-
Milling software: PrograMill CAM (510(k) exempt under 21 CFR 872.3661)
-
Digital Library / Templates (3 Shape): The Solidex® Ti-Link CAD/CAM System shall only be used with the official "Solidex® Ti-Link Digital Library" and templates identified. The library is imported into the 3Shape software, and the design restriction parameters (minimum wall thickness, post height, maximum gingival height, etc.) are automatically applied. The official "Solidex® Ti-Link Digital Library" can be obtained from CREODENT HUDSON VALLEY by contacting support email (inquiry@creodental.com) or from CREODENT HUDSON VALLEY web-site.
When imported into the 3Shape software, the design restriction parameters (minimum wall thickness, post height, maximum gingival height, etc.) and built-in protection features ("stops" and restriction shells) are automatically applied.
The design parameters for the CAD/CAM zirconia superstructure to be used Solidex® Ti-Links & Screws:
- Minimum wall thickness: 0.5 mm
- Minimum abutment post height: 6.0 mm
(Note)
: The abutment post height is the dimension above the gingival height.
: Feature maximum abutment post height are based on patient anatomy and clinician discretion. - Minimum gingival height: 0.5 mm
- Maximum gingival height: 6.0 mm
- Minimum Angulation: 0°
- Maximum Angulation: 20°
The zirconia superstructure must be cemented to the Solidex® Ti-Links using Panavia SA cement Universal in a dental laboratory. Solidex® Ti-Links & Screws is provided non-sterile therefore must be sterilized after the cementation of the zirconia superstructure on Solidex® Ti-Links. Solidex® Screws are designed to attach the abutment to the implant.
N/A
Ask a specific question about this device
(80 days)
DESS Dental Smart Solutions abutments are intended to be used in conjunction with endosseous dental implants in the maxillary or mandibular arch to provide support for prosthetic restorations.
All digitally designed custom abutments for use with DESS Bases or Pre-milled Blanks are to be sent to a Terrats Medical validated milling center for manufacture, or to be designed and manufactured according to the digital dentistry workflow. The digital dentistry workflow integrates multiple components: scan files from intra-oral and lab (desktop) scanners, CAD software, CAM software, ceramic material, milling machine, and associated tooling and accessories.
The purpose of this submission is to expand the DESS Dental Smart Solutions abutment system cleared under K221301 and K240982 to allow additional options of zirconia material, scanners, CAM software, and milling machines to the digital dentistry workflow. The subject devices are to be sent to Terrats Medical validated milling centers for manufacture, or to be designed and manufactured via a digital dentistry workflow. The digital dentistry workflow integrates multiple components: scan files from intra-oral and lab (desktop) scanners, CAD software, CAM software, titanium and ceramic material, milling machine, and associated tooling and accessories. There are no changes to the abutment design, implant compatibilities, or design parameters. All part numbers have been cleared for manufacturing via a validated milling center and digital dentistry workflows (also referred to as point of care) under K221301 and K240982.
The subject device DESS Dental Smart Solutions abutments provide a range of prosthetic solutions for dental implant restoration. DESS abutments are offered in a variety of connection types to enable compatibility with currently marketed dental implants. All abutments are provided non-sterile, and each abutment is supplied with the appropriate abutment screw (if applicable) for attachment to the corresponding implant.
Subject device Base Abutments are designed for fabrication of a patient-specific CAD/CAM zirconia superstructure on which a crown may be placed. They are two-piece abutments for which the second part (or top half) is the ceramic superstructure. They also may be used for support of a crown directly on the abutment.
All patient-specific custom abutment fabrication for Base Abutments and Pre-milled (Blank) Abutments is by prescription on the order of the clinician. The subject device Pre-milled (Blank) Abutments and all zirconia superstructures for use with the subject device Ti Base Interface, DESS Aurum Base, ELLIPTIBase, and DESS C-Base will be manufactured using a validated milling center or a digital dentistry workflow. A validated milling center will be under FDA quality system regulations. The digital dentistry workflow scans files from intra-oral and lab (desktop) scanners, CAD software, CAM software, titanium and ceramic material, milling machine and associated tooling and accessories.
The digital dentistry workflow uses scan files from intra-oral and lab (desktop) scanners, CAD software, CAM software, ceramic material, milling machine and associated tooling and accessories.
The provided 510(k) summary for DESS Dental Smart Solutions focuses on demonstrating substantial equivalence to predicate devices for dental implant abutments. It primarily addresses the expansion of compatible materials, scanners, CAM software, and milling machines within an existing digital dentistry workflow. The document does not describe an AI/ML-based device that would typically have acceptance criteria related to diagnostic performance.
Therefore, many of the requested items related to AI/ML device performance (like acceptance criteria for diagnostic metrics, sample size for test sets, data provenance, expert qualifications, adjudication methods, MRMC studies, standalone performance, and training set details) are not applicable to this submission.
The acceptance criteria and supporting "study" (non-clinical data) for this device are related to its mechanical performance, biocompatibility, and integration within the digital workflow, demonstrating that the expanded components maintain the safety and effectiveness of the previously cleared predicate devices.
Here's a breakdown based on the information provided and the non-applicability of AI/ML-specific questions:
1. A table of acceptance criteria and the reported device performance
Since this is not an AI/ML diagnostic device, the acceptance criteria are not in terms of traditional diagnostic metrics (sensitivity, specificity, AUC). Instead, they are related to material properties, mechanical integrity, and the digital workflow's accuracy.
| Acceptance Criteria Category | Reported Device Performance (Summary from Submission) |
|---|---|
| Mechanical Integrity / Compatibility | - Fatigue testing of OEM implant bodies with patient-specific abutments made at worst-case angled conditions. (Implies successful completion to similar or better standards than predicate.) |
| Biocompatibility | - Biocompatibility testing according to ISO 10993-5 and ISO 10993-12. (Implies successful demonstration of biocompatibility.) |
| Sterilization | - Sterilization validation according to ISO 17665-1, ISO 17665-2, and ISO 14937. (Implies successful validation for sterilization.) |
| CAD Design Restrictions | - Software verification included testing of restrictions that prevent design of components outside of the stated design parameters. - Abutment design library validated to demonstrate established design limitations are locked and cannot be modified by the user. (Implies successful implementation and verification of design constraints.) |
| CAM Restriction Zones / Manufacturing Accuracy | - Validation testing of CAM restriction zones conducted, including verification to show avoidance of damage or modifications of the connection geometry, and locking of restriction zones from user editing in CAM software. (Implies successful validation to ensure manufacturing precision and prevent damage.) |
| Material Conformance | - Zirconia materials conform to ISO 6872. - Titanium alloy conforms to ASTM F136. - Co-Cr-Mo alloy conforms to ASTM F15337. (Implies materials meet standards.) |
| Physical Dimensions | - Device encompasses the same range of physical dimensions as the predicate device. (Implies dimensional equivalence.) |
2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)
- Sample Size: Not explicitly stated in terms of a "test set" for diagnostic performance. The validation involved physical testing of components (e.g., fatigue testing) and software verification. The specific number of abutments or digital design instances used for these non-clinical tests is not detailed in this summary.
- Data Provenance: Not applicable in the context of patient data for an AI/ML device. The "data" here refers to engineering and material testing results, likely conducted in controlled lab environments (implied to be in accordance with international standards like ISO and ASTM). The manufacturer is Terrats Medical SL, in Spain, so testing would likely originate from their facilities or contracted labs.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)
Not applicable. This is not a diagnostic device requiring expert interpretation for ground truth. The "ground truth" for this device relates to engineering specifications and material science.
4. Adjudication method (e.g. 2+1, 3+1, none) for the test set
Not applicable. This is not a diagnostic device involving expert review adjudication.
5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
Not applicable. This device is not an AI-assisted diagnostic tool.
6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done
Not applicable. This is not an AI/ML algorithm.
7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)
For this device, the "ground truth" is based on:
- Engineering Specifications: Defined design parameters (e.g., minimum wall thickness, post height, angulation limits).
- Material Standards: Conformance to international standards such as ASTM F136, ISO 6872.
- Benchmarking/Predicate Equivalence: Performance is assessed against established performance of the predicate devices and OEM implant systems.
- Software Validation Logic: Verification that software correctly enforces design rules and CAD/CAM restrictions.
8. The sample size for the training set
Not applicable. This device does not involve a machine learning training set.
9. How the ground truth for the training set was established
Not applicable. This device does not involve a machine learning training set.
Ask a specific question about this device
(138 days)
DESS Dental Smart Solutions abutments are intended to be used in conjunction with endosseous dental implants in the maxillary or mandibular arch to provide support for prosthetic restorations.
The purpose of this submission is to expand the DESS Dental Smart Solutions abutment system cleared under K221301 to add the ability for the subject device Base Abutments and Pre-milled (Blank) Abutments to be designed using AbutmentCAD software in the digital dentistry workflow, as well as add angulation to some of the Pre-Milled (Blank) Abutments. The subject devices are to Terrats Medical validated milling centers for manufacture, or to be designed and manufactured via a digital dentistry workflow. The digital dentistry workflow integrates multiple components: scan files from intra-oral and lab (desktop) scanners, CAD software, CAM software, titanium and ceramic material, milling machine, and associated tooling and accessories. The proposed change is to allow the subject devices to be designed using AbutmentCAD by exocad GmbH, the current workflow allows only the use of 3Shape by 3Shape A/S for the design software. Another purpose of this submission is to expand the design parameters to allow angulation (up to 30°) on Pre-milled (Blank) Abutments that are compatible with Neodent Grand Morse, Nobel Active/Nobel Parallel Conical 3.0 mm, and Straumann BLX implants. There are no changes to the abutment design or implant compatibilities. All part numbers have been cleared for manufacturing via a validated milling center and digital dentistry workflows in under K221301.
The subject device DESS Dental Solutions abutments provide a range of prosthetic solutions for dental implant restoration. DESS abutments are offered in a variety of connection types to enable compatibility with currently marketed dental implants. All abutments are provided non-sterile, and each abutment is supplied with the appropriate abutment screw (if applicable) for attachment to the corresponding implant.
Subject device Base Abutments are designed for fabrication of a patient-specific CAD/CAM zirconia superstructure on which a crown may be placed. They are two-piece abutments for which the second part (or top half) is the ceramic superstructure. They also may be used for support of a crown directly on the abutment.
All patient-specific custom abutment fabrication for Base Abutments and Pre-milled (Blank) Abutments is by prescription on the order of the clinician. The subject device Pre-milled (Blank) Abutments and all zirconia superstructures for use with the subject device Ti Base Interface, DESS Aurum Base, ELLIPTIBase, and DESS C-Base will be manufactured using a validated milling center or a digital dentistry workflow. A validated milling center will be under FDA quality system regulations. The digital dentistry workflow scans files from intra-oral and lab (desktop) scanners, CAD software, titanium and ceramic material, milling machine and associated tooling and accessories.
The digital dentistry workflow uses scan files from intra-oral and lab (desktop) scanners, CAD software, CAM software, ceramic material, milling machine and associated tooling and accessories. The digital workflow includes the following products (not subject devices of this submission):
- Ceramic material: VITA YZ ST and VITA YZ XT (K180703)
- . Cement: Ivoclar Vivadent Multilink Hybrid Abutment Cement (K130436)
- . Intraoral Scanner: 3Shape TRIOS A/S Series Intraoral Scanner (510(k) exempt under 21 CFR 872.3661)
- Desktop scanner: 3Shape D900 Dental Lab Scanner (510(k) exempt under 21 CFR 872.3661)
- Abutment design software: 3Shape Abutment Designer Software (K151455) and AbutmentCAD ● (K193352)
- . Milling machine: VHF R5 by vhf camfacture AG with DentalCAM and DentalCNC 7 software
The provided text describes a 510(k) premarket notification for DESS Dental Smart Solutions, which are dental implant abutments. This type of submission focuses on demonstrating substantial equivalence to a legally marketed predicate device, rather than proving a device's effectiveness through clinical performance studies with specific statistical acceptance criteria for accuracy, sensitivity, or specificity.
Therefore, the document does not contain the information requested regarding:
- A table of acceptance criteria and reported device performance (in terms of clinical metrics like accuracy, sensitivity, specificity).
- Sample size used for the test set or its provenance.
- Number of experts used to establish ground truth or their qualifications.
- Adjudication method for the test set.
- Multi-Reader Multi-Case (MRMC) comparative effectiveness study or its effect size.
- Standalone (algorithm only) performance.
- Type of ground truth used (expert consensus, pathology, outcomes data).
- Sample size for the training set.
- How the ground truth for the training set was established.
The study described in this document focuses on non-clinical performance data to demonstrate substantial equivalence, specifically:
- Sterilization validation: According to ISO 17665-1, ISO 17665-2, and ISO 14937.
- Biocompatibility testing: According to ISO 10993-5 and ISO 10993-12.
- Fatigue testing and reverse engineering analysis: Of OEM implant bodies, OEM abutments, and OEM abutment screws to confirm compatibility. This includes fatigue testing of OEM implant bodies with patient-specific abutments made at worst-case angled conditions.
- MR Conditional labeling.
- Validation testing of CAM restriction zones: Including verification to show avoidance of damage or modifications of the connection geometry, and locking of restriction zones from user editing in the CAM software.
- Software verification: Included testing of restrictions that prevent design of components outside of the stated design parameters. The abutment design library was validated to demonstrate that established design limitations and specifications are locked and cannot be modified by the user.
The acceptance criteria and reported "performance" for this submission are based on these engineering and design validations, ensuring the device meets safety and performance standards equivalent to the predicate device, K221301. The key "performance" metrics are about maintaining physical and material integrity and compatibility.
The core of the submission is to expand the DESS Dental Smart Solutions abutment system to:
- Allow design using AbutmentCAD software (in addition to 3Shape software).
- Add angulation (up to 30°) to some Pre-milled (Blank) Abutments for specific implant systems.
The document explicitly states: "No clinical data were included in this submission." and "The subject device, the predicate device, and reference devices have the same intended use, technological characteristics, and are materials. The subject device, the predicate device, and reference devices encompass the same range of physical dimensions, manufactured by similar methods, are packaged in similar materials, and are to be sterilized using similar methods. The data included in this submission demonstrate substantial equivalence to the predicate devices listed above."
Therefore, this FDA submission is for a physical medical device (dental implant abutment) and its manufacturing/design software modifications, not an AI or diagnostic device that would involve clinical performance metrics like sensitivity or specificity.
Ask a specific question about this device
(308 days)
DESS Dental Smart Solutions abutments are intended to be used in conjunction with endosseous dental implants in the maxillary or mandibular arch to provide support for prosthetic restorations.
All digitally designed custom abutments for use with DESS Bases or Pre-milled Blanks are to be sent to a Terrats Medical validated milling center for manufacture, or to be designed and manufactured according to the digital dentistry workflow integrates multiple components: scan files from intra-oral and lab (desktop) scanners, CAM software, ceramic material, milling machine, and associated tooling and accessories.
The purpose of this submission is to expand the DESS Dental Smart Solutions abutment system cleared under K22288 to add the ability for the subject device Base Abutments and Pre-milled (Blank) Abutments to be manufactured via a digital dentistry workflow by using scan files from intra-oral and lab (desktop) scanners, CAD software, CAM software, ceramic material, milling machine and associated tooling and accessories. There are no changes to the abutment design or implant compatibilities, however, there is a new material for the zircomia superstructure. All part numbers have been cleared within previous submissions for manufacturing at a validated milling center. The purpose of this submission is to allow manufacturing via digital dentistry workflow and to add a new zirconia material for the superstructures of the two-piece abutment.
The subject device DESS Dental Smart Solutions abutments provide a range of prosthetic solutions for dental implant restoration. DESS abutments are offered in a variety of connection types to enable compatibility with currently marketed dental implants. All abutments are provided non-sterile, and each abutment is supplied with the appropriate abutment screw (if applicable) for attachment to the corresponding implant.
Subject device Base Abutments are designed for fabrication of a patient-specific CAD/CAM zirconia superstructure on which a crown may be placed. They are two-piece abutments for which the second part (or top half) is the ceramic superstructure. They also may be used for support of a crown directly on the abutment.
All patient-specific custom abutment fabrication for Base Abutments and Pre-milled (Blank) Abutments is by prescription on the order of the clinician. The subject device Pre-milled (Blank) Abutments and all zirconia superstructures for use with the subject device Ti Base Interface, DESS Aurum Base, ELLIPTIBase, and DESS C-Base will be manufactured using a digital dentistry workflow by using scan files from intra-oral and lab (desktop) scanners, CAD software, CAM software, ceramic material, milling machine and associated tooling and accessories.
The digital dentistry workflow uses scan files from intra-oral and lab (desktop) scanners, CAD software, CAM software, ceramic material, milling machine and associated tooling and accessories.
The provided text is a 510(k) summary for a dental device, specifically DESS Dental Smart Solutions abutments. It focuses on demonstrating substantial equivalence to predicate devices rather than proving performance against specific acceptance criteria for an AI/ML-based device.
Therefore, I cannot provide the requested information as the document does not contain details about:
- Acceptance criteria for an AI/ML device: This document is about dental implant abutments, not a medical device driven by AI/ML.
- Study proving device meets acceptance criteria: There is no performance study described for an AI/ML component. The "Performance Data" section refers to non-clinical data for biocompatibility and fatigue testing of the physical dental abutments, and software verification to ensure design parameters are met and restriction zones are locked, not AI/ML performance.
- Sample size for test set and data provenance: Not applicable to a non-AI/ML device.
- Number of experts and their qualifications for ground truth: Not applicable.
- Adjudication method for the test set: Not applicable.
- Multi-reader multi-case (MRMC) comparative effectiveness study: Not applicable.
- Standalone (algorithm only) performance: Not applicable.
- Type of ground truth used: Not applicable.
- Sample size for training set: Not applicable.
- How ground truth for training set was established: Not applicable.
The document describes the physical characteristics, materials, manufacturing processes (including digital dentistry workflows), and compatibility of the dental abutments, and asserts their substantial equivalence to previously cleared devices. It does not mention any AI/ML components or their performance evaluation.