510(k) Data Aggregation

    K Number
    K201491

    Validate with FDA (Live)

    Date Cleared
    2020-12-22

    (201 days)

    Product Code
    Regulation Number
    872.3640
    Panel
    Dental
    Age Range
    All
    Reference & Predicate Devices
    Why did this record match?
    Reference Devices :

    K150367, K132881, K153624, K180477, K182620, K192893

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Indications for Use for Zirconia Implants, Cover Screw and Healing Abutment:
    The Neodent Implant System is intended to be surgically placed in the bone of the upper or lower jaw to provide support for prosthetic devices, such as artificial teeth, to restore chewing function. It may be used with singlestage or two-stage surgical procedures, for single or multiple unit restorations, and may be loaded immediately when good primary stability is achieved and with physiological occlusal loading. Multiple teeth applications can be rigidly splinted.

    Indications for Use for Zirconia Bases:
    The Zirconia Base is an abutment placed over Neodent Zirconia Implants in order to provide support for patient-specific prosthetic restorations, such as copings or crowns. It may be used for cement- or screw retained single unit restorations. All digitally designed copings and/or crowns to be used with the Neodent Zirconia Base Abutment System are intended to be sent to Straumann for manufacture at a validated milling center.

    Device Description
    • Intended for single use;
    • Zirconia implants provided sterile via Ethylene Oxide; Zirconia Base abutments are provided sterile but require end user sterilization via moist heat following cementation of the restoration and before use
    • All devices of Zirconia Implant System are manufactured of Y-TZP (Yttrium-stabilized zirconium dioxide), except for the Zirconia Implant Cover Screw and the Basal Screw, which is manufactured of titanium alloy according to ASTM F136 standard;
    • ZiLock prosthetic interface with internal indexer;
    • Apically tapered implant with trapezoidal thread profile;
    • Implant provided with sand-blasted, acid etched surface finish to facilitate osseointegration.
    • The final finished Zirconia Base abutments are two-piece abutments composed of a zirconia base bottom-half bonded to a CAD-CAM zirconia top-half.
    AI/ML Overview

    The provided text describes a medical device submission (K201491) for the Neodent Implant System - Zirconia Implant System to the FDA. This document focuses on demonstrating substantial equivalence to predicate devices, rather than presenting a standalone study with specific acceptance criteria and performance results for a new AI/software device.

    Therefore, the requested information about acceptance criteria, detailed device performance, sample sizes for test/training sets, ground truth establishment, expert adjudication, and MRMC studies cannot be extracted from this document as it pertains to a different type of medical device submission (dental implants and their components) and regulatory pathway (510(k) for substantial equivalence).

    The document details the comparison of the subject device (Neodent Implant System - Zirconia Implant System) to predicate and reference devices in terms of indications for use, technological characteristics, and performance data from various tests. However, these are evaluations against established standards and comparisons to existing devices, not metrics of an AI's performance.

    Here's what can be extracted, though it doesn't directly answer the detailed questions about AI performance criteria:

    1. Table of Acceptance Criteria and Reported Device Performance:

    The document doesn't provide a table of acceptance criteria for software performance in the typical sense (e.g., sensitivity, specificity, AUC). Instead, it presents a comparison of technological characteristics and states that the "tested subject devices exhibit a level of performance equivalent to that reviewed for the predicate devices" in mechanical testing.

    Here's a summary of the performance data presented, which implies meeting an "equivalent" performance to predicates rather than specific numerical acceptance criteria for a new software.

    Performance AreaReported Device Performance
    BiocompatibilityRepresentative samples of each device subjected to tests per ISO 10993-1, ISO 10993-5, ISO 10993-12, and ISO 10993-18. (Implied acceptance is successful completion according to these standards).
    Mechanical TestingDynamic Fatigue Test (per ISO 14801): "exhibit a level of performance equivalent to that reviewed for the predicate devices."
    Wear Assessment (Titanium/Ceramic contacting pieces): "concluded of comparable behavior of the subject device to the reference devices in terms of wear on the implant-to-abutment connection."
    Torsion Test: Performed to evaluate static torsional loading. (Implied acceptance is meeting design specifications and being comparable to predicates).
    Insertion Test: Performed to evaluate insertion torque in sawbones materials (bone type I, II, III, IV). (Implied acceptance is meeting design specifications and being comparable to predicates).
    SterilizationAchieved a "Sterility Assurance Level of 1x10⁻⁶" via Ethylene Oxide per ISO 11135-1. Residuals according to ISO 10993-7. Not represented as "pyrogen free."
    MRI TestingDone according to documents presented in K182620 (a previously cleared device for MRI compatibility). (Implied acceptance is being MRI compatible as per the reference device).
    Shelf Life5 years. Packaging assessment per ISO 11607-1, with maintenance of sterile barrier and sealing integrity confirmed by Dye Penetration, Sealing Strength, and Bubble Test after Accelerated Aging.

    No information is available regarding points 2-9 as they are specific to AI/software device studies to establish performance against a ground truth. This document is a 510(k) summary for a physical dental implant system and components, establishing substantial equivalence to existing devices.

    Ask a Question

    Ask a specific question about this device

    K Number
    K193234

    Validate with FDA (Live)

    Date Cleared
    2020-06-18

    (206 days)

    Product Code
    Regulation Number
    872.3640
    Panel
    Dental
    Age Range
    All
    Predicate For
    Why did this record match?
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Indications for Use for NUVO IF Implants and conventional abutments:

    The NUVO IF Implant System is intended to be surgically placed in the maxilla or mandible to provide support for prosthetic devices such as artificial teeth in order to restore chewing function. It may be used with single-stage or twostage procedures, for single or multiple unit restorations, and may be loaded immediately when good primary stability is achieved and with appropriate occlusal loading. Multiple tooth applications may be rigidly splinted.

    The Titanium Temporary Abutment is indicated to provide temporary support for prosthesis structure for up to 6 months.

    The Attachment Equator and Attachment Removable Prosthesis are indicated for the attachment of full or partial dentures to NUVO implants.

    Indications for Use for Rotational and Anti-rotational Titanium Bases abutments:

    Titanium Base Abutment is a titanium base placed onto dental implants to provide support for customized prosthetic restorations. It is used with a coping and crown, or crown alone, and is indicated for cement or screw-retained single or multi-unit restorations or screw-retained single restorations.

    All digitally designed copings and/or crowns to be used with the Titanium Base Abutment System are intended to be sent to Straumann for manufacture at a validated milling center.

    The Rotational Titanium Base is indicated for cement or screw-retained multi-unit restorations.

    Indications for Use for CARES® Abutment IF:

    The CARES® Abutment is a customized prosthetic abutment, manufactured in titanium alloy, placed onto dental implants to provide support for customized prosthetic restorations (copings or crowns). It is indicated for cementretained single restorations. All digitally designed abutments for use with the CARES® Abutment are intended to be sent to Straumann for manufacturing at a validated milling center.

    Device Description

    The NUVO IF Implant System consists of implants, cover screws, and abutments.

    Implants: Intended for single use, provided sterile via gamma irradiation, manufactured of commercially pure titanium (Grade 4), bone level design, apically tapered thread-form with apical cutting flutes in various diameters and lengths, and provided with an internal hexagonal implant-to-abutment interface.

    Cover Screws: Intended for single use, used to protect the internal geometry of the implants during healing, placed out of occlusion, provided sterile via gamma irradiation or ethylene oxide gas, manufactured of titanium alloy (Ti6Al4V-ELI), and provided with an implant-to-abutment interface compatible with the internal hexagonal geometry of the implants.

    Abutments: Intended for single use, provided sterile via ethylene oxide gas or non-sterile, manufactured of titanium alloy (Ti6Al4V-ELI) or a combination of POM polymer and cobalt-chromium alloy, conical format available in different diameters, height of cementable area and gingival height, screw-retained to the implant, provided with an implant-to-abutment interface compatible with the internal hexagonal geometry of the implants, provided with coronal geometries in rotational (non-indexed) and anti-rotational (indexed) versions, and Titanium Base abutments are two-piece abutments composed of a titanium base and a patient-specific CAD/CAM top-half.

    AI/ML Overview

    This looks like a 510(k) summary for a dental implant system (NUVO IF Implant System), not a study describing acceptance criteria and performance data for an AI/ML medical device. Therefore, I cannot extract the requested information regarding acceptance criteria and performance studies for an AI device.

    The document discusses:

    • Trade/Device Name: NUVO IF Implant System
    • Regulation Number/Name: 21 CFR 872.3640, Endosseous Dental Implant
    • Regulatory Class: Class II
    • Product Code: DZE, NHA
    • Indications for Use: Surgical placement in maxilla or mandible for prosthetic device support, including immediate loading, temporary support, and attachment of dentures.
    • Device Description: Details on implants (material, design, size, sterilization), cover screws (material, use, sterilization), and various abutment types (materials, design, sterilization, use).
    • Predicate Devices: Several predicate devices are listed and compared for substantial equivalence.
    • Performance Data: This section details non-AI/ML performance testing, including:
      • Dynamic fatigue testing per ISO 14801.
      • Sterilization validation (gamma irradiation per ISO 11137-1/2, ethylene oxide per ISO 11135, moist heat per ISO 17665-2).
      • Ethylene oxide residuals testing per ISO 10993-7.
      • Biological Safety Assessment per ISO 10993-1 (cytotoxicity per ISO 10993-5, chemical characterization per ISO 10993-18).
      • Package transport integrity per ISTA 2A.
      • Product and package stability per ASTM F1980.
      • Product endotoxin testing (LAL) per ANSI/AAMI ST72.

    There is no mention of an AI/human-in-the-loop study, ground truth establishment, sample sizes for training/test sets for AI, or expert adjudication.

    Ask a Question

    Ask a specific question about this device

    K Number
    K182620

    Validate with FDA (Live)

    Date Cleared
    2019-01-18

    (116 days)

    Product Code
    Regulation Number
    872.3640
    Panel
    Dental
    Age Range
    All
    Predicate For
    Why did this record match?
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use
    • HE (External Hex) implants, stock abutments and stock copings (originally cleared per K101207)
    • Titamax CM (Cone Morse) implants, stock abutments and stock copings (originally cleared per K101945 ● and extended per K123022, K133696 and K150199)
    • Titamax CM (Cone Morse) EX implants, stock abutments and stock copings (originally cleared per K101945 and extended per K123022, K133696 and K150199)
    • Alvim CM (Cone Morse) implants, stock abutments and stock copings (originally cleared per K101945 and ● extended per K123022, K133696 and K150199)
    • CM (Cone Morse) Drive implants, stock abutments and stock copings (originally cleared per K123022 and extended per K133696, K150182 and K150199)
    • GM (Grand Morse) Titamax implants, stock abutments and stock copings (originally cleared per K163194)
    • GM (Grand Morse) Drive implants, stock abutments and stock copings (originally cleared per K163194) ●
    • . GM (Grand Morse) Helix implants, stock abutments and stock copings (originally cleared per K163194)

    The implants, abutments and copings of the Neodent Implant System are intended to be surgically placed in the bone of the upper or lower jaw to provide support for prosthetic devices such as artificial teeth, to restore chewing function. It may be used with single-stage or two-stage procedures, for single or multiple unit restorations, and may be loaded immediately when good primary stability is achieved and with appropriate occlusal loading. Multiple tooth applications may be rigidly splinted.

    • Titamax WS implants, abutments and copings (originally cleared per K123022)
      The implants, abutments and copings of the Neodent Implant System are intended to be surgically placed in the bone of the upper or lower jaw to provide support for prosthetic devices such as artificial teeth to restore chewing function. The Neodent Implant System may be used for single or multiple unit restorations. Multiple tooth applications may be rigidly splinted.

    The Titamax WS implants are indicated for a delayed loading protocol.

    • Facility implants, stock abutments and stock copings (originally cleared per K123022 and extended per K133696 and K150199)
      The implants, abutments and copings of the Neodent Implant System are intended to be surgically placed in the bone of the upper or lower jaw to provide support for prosthetic devices such as artificial teeth to restore chewing function. The Neodent Implant System may be used for single or multiple unit restorations. Multiple tooth applications may be rigidly splinted.

    The Facility implants of the Neodent Implant System are indicated for replacement of maxillary lateral incisors, mandibular incisors or retention of overdentures.

    • Zygomatic implants, stock abutments and stock copings (originally cleared per K141777)
      The Zygomatic implants, abutments and copings of the Neodent Implant System are intended to be surgically placed in the bone of the upper or lower jaw to provide support for prosthetic devices such as artificial teeth, to restore chewing function. It may be used with single-stage or two-stage procedures, for single or multiple unit restorations, and may be loaded immediately when good primary stability is achieved and with appropriate occlusal loading. Multiple tooth applications may be rigidly splinted.

    The Zygomatic implants of the Neodent Implant System are indicated for surgical installation in the zygoma region in cases of severe jaw resorption in order to restore patient esthetics and chewing function. Zygomatic implants are recommended for the posterior (pre-molar/ molar) region, one implant on each side, with at least two standard dental implants in the anterior region to support a fixed restoration.

    • CM (Cone Morse) Titanium Base Abutment component of two-piece patient-specific abutment (originally cleared per K150367 and extended per K153624)
    • o GM (Grand Morse) Exact Titanium Base Abutment component of two-piece patient-specific abutment (originally cleared per K163194)

    The implants, abutments and copings of the Neodent Implant System are intended to be surgically placed in the bone of the upper or lower jaw to provide support for prosthetic devices such as artificial teeth, to restore chewing function. It may be used with single-stage or two-stage procedures, for single or multiple unit restorations, and may be loaded immediately when good primary stability is achieved and with appropriate occlusal loading. Multiple tooth applications may be rigidly splinted.

    The Titanium Base Abutment is a titanium base placed onto Neodent dental implants to provide support for customized prosthetic restorations. The Titanium Base is used with a coping and crown, or crown alone, and is indicated for cement-retained single or multi-unit restorations, or screw-retained single restorations.

    All digitally designed copings and/or crowns for use with the Neodent Titanium Base Abutment are intended to be sent to Straumann for manufacture at a validated milling center.

    • CM (Cone Morse) Titanium Base for CEREC Abutment component of two-piece patient-specific abutment (originally cleared per K160964)
      The implants, abutments and copings of the Neodent Implant System are intended to be surgically placed in the bone of the upper or lower jaw to provide support for prosthetic devices such as artificial teeth, to restore chewing function. It may be used with single-stage or two-stage procedures, for single or multiple unit restorations, and may be loaded immediately when good primary stability is achieved and with appropriate occlusal loading. Multiple tooth applications may be rigidly splinted.

    The Titanium Base for CEREC Abutment is a titanium base placed onto Neodent dental implants to provide support for customized prosthetic restorations. They are used with a coping and crown, or crown alone, and is indicated for cement-retained single or multi-unit restorations, or screw-retained single restorations.

    All digitally designed copings and/or crowns for use with the Neodent Titanium Base for CEREC Abutment are to be designed using Sirona inLab software or Sirona CEREC Software and manufactured using a Sirona CEREC or inLab MC X or MC XL milling unit.

    • CM Preface Abutment component of patient-specific abutment (originally cleared per K150367)
      The implants, abutments and copings of the Neodent Implant System are intended to be surgically placed in the bone of the upper or lower jaw to provide support for prosthetic devices such as artificial teeth, to restore chewing function. It may be used with single-stage or two-stage procedures, for single or multiple unit restorations, and may be loaded immediately when good primary stability is achieved and with appropriate occlusal loading. Multiple tooth applications may be rigidly splinted.

    The Preface Abutment is a titanium abutment to be used in fabricating a full custom abutment placed onto Neodent dental implants to provide support for customized prosthetic restorations. The Preface Abutments is indicated for screw-retained single restorations or cement-retained single or multi-unit restorations.

    All digitally designed restorations for use with the Neodent Preface Abutment are intended to be sent to Straumann for manufacture at a validated milling center.

    • GM (Grand Morse) ProPEEK Abutments (originally cleared per K163194) ●
    • CM (Cone Morse) ProPEEK Abutments (originally cleared per K170080)

    The implants, abutments and copings of the Neodent Implant System are intended to be surgically placed in the bone of the upper or lower jaw to provide support for prosthetic devices such as artificial teeth, to restore chewing function. It may be used with single-stage or two-stage procedures, for single or multiple unit restorations, and may be loaded immediately when good primary stability is achieved and with appropriate occlusal loading. Multiple tooth applications may be rigidly splinted.

    The Pro PEEK Abutment is indicated to be used on Neodent implants to provide temporary support for prosthesis structure for up to 6 months.

    • Neodent Implant for Orthodontic Anchor (originally cleared per K102769) ●
      The Neodent Implant for Orthodontic Anchor is a surgical device in the form of a temporary screw used as an aid in orthodontic movement procedures.

    • . The Neodent Graft Screw (originally cleared per K103084)
      The Neodent Graft Screw is an implantable device used for fixation of bone blocks for the regeneration of bone in the oral cavity. The product is intended for temporary use only.

    Device Description

    Not Found

    AI/ML Overview

    The provided text is a 510(k) Premarket Notification from the FDA regarding the MRI compatibility of various Neodent Implant System devices. It primarily focuses on defining the indications for use for different components of the implant system and does not contain information about acceptance criteria, device performance studies, sample sizes, expert qualifications, or ground truth establishment for an AI/Medical Device.

    The document is a regulatory approval letter and descriptive indications for use, not a study report or technical performance assessment. Therefore, I cannot extract the requested information from this text.

    Ask a Question

    Ask a specific question about this device

    K Number
    K173961

    Validate with FDA (Live)

    Date Cleared
    2018-06-05

    (159 days)

    Product Code
    Regulation Number
    872.3640
    Panel
    Dental
    Age Range
    All
    Reference & Predicate Devices
    Why did this record match?
    Reference Devices :

    Temporary Abutments, SC Variobase Abutments, SC CARES Abutment., K101545 Genesis Implant System ●, K153624

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Straumann® BLX Implants are suitable for endosteal implantation in the upper and lower jaw and for the functional and esthetic oral rehabilitation of edentulous and partially edentulous patients. BLX Implants can be placed with immediate function on single-tooth applications when good primary stability is achieved and with appropriate occlusal loading to restore chewing function. The prosthetic restorations are connected to the implants through the corresponding abutment components.

    Straumann® Closure Caps and Healing Abutments are indicated to be placed in the patient's mouth at the end of the implant placement to protect the inner configuration of the implant and to shape, maintain and stabilize the soft tissue during the healing process. Closure caps and healing abutments should be used only with suitable implant connections. Straumann Closure Caps and Healing Abutments have a maximum duration of usage of 6 months.

    Prosthetic components directly or indirectly connected to the endosseous dental implant are intended for use as an aid in prosthetic rehabilitations. Temporary components can be used prior to the insertion of the final components to maintain, stabilize and shape the soft tissue during the healing phase; they may not be placed into occlusion. Final abutments may be placed into occlusion when the implant is fully osseointegrated. BLX Temporary Abutments have a maximum duration of usage of 180 days.

    Straumann® Variobase® prosthetic components directly or indirectly connected to the endosseous dental implant are intended for use as an aid in prosthetic rehabilitations. The prosthetic restoration (crowns) can be cemented onto the Straumann® Variobase® prosthetic components. A temporary restoration can be used prior to the insertion of the final components to maintain, stabilize and shape the soft tissue during the healing phase; they must be placed out of occlusion. Final abutments and restorations may be placed into occlusion when the implant is fully osseointegrated. All digitally designed copings and/or crowns for use with the Straumann® Variobase® Abutment system are intended to be sent to Straumann for manufacture at a validated milling center.

    Device Description

    The Straumann BLX Implants are fully tapered implants manufactured utilizing the Roxolid material and are finished with SLActive® surface. The connection is identified as conical fitting with Torx style engaging feature. The prosthetic platforms are identified as RB (Regular Base) and WB (Wide Base). The implants with a RB platform have a "small top/head", and implants with WB platform have a "large top/head", whereas the internal connection is identical for both platform and all the implant diameters and lengths.

    The closure cap and healing abutments are manufactured from Titanium Grade 4, per ISO 5832-2 and ISO 5832-11, and are anodized signal violet for the parts compatible with the RB platform and brown for the parts compatible with the WB platform for identification purposes. Closure caps are screwed into the implant to protect the inner configuration and shoulder of the implant during the healing phase in cases of submerged (submucosal) healing. Healing abutments are screwed into the implant to protect the inner configuration in cases of transmucosal healing and are placed out of occlusion.

    The temporary abutments are manufactured from TAN and are anodized signal violet (RB platform) and brown (WB platform) for identification purposes. The temporary abutments are fixed in the implant with a basal screw which is also manufactured from TAN. The basal screw will be delivered with the temporary abutment.

    BLX Variobase abutments is a two-piece abutment ultimately composed by three components: Variobase™ Abutment (Ti-base), Prosthetic restoration (coping and/or crown), and Basal Screw. The BLX Variobase Abutments are manufactured and are delivered with the corresponding basal screw. The prosthetic restoration (crowns) can be cemented onto the Variobase prosthetic components.

    Basal screws are used to seat the temporary abutments and the BLX Variobase Abutments to the dental implant, and can be also be used during lab procedures to fix lab prosthetic parts on implant analogs. There is one basal screw for the RB platform and one for the WB platform. They have identical designs and differ only in color-coding (signal violet and brown) to ease the handling. They are provided along the abutments but they are also provided as standalone screws. The BLX basal screws are manufactured from TAN.

    AI/ML Overview

    This document describes the Straumann® BLX Implant System, a dental implant system, and its various components. The information provided is for a 510(k) premarket notification to the FDA. The goal is to demonstrate substantial equivalence to legally marketed predicate devices.

    Here's an analysis of the acceptance criteria and study information based on the provided text:

    1. Table of Acceptance Criteria and Reported Device Performance

    The provided document is a 510(k) summary, which focuses on demonstrating substantial equivalence to predicate devices rather than setting and meeting specific, quantitative acceptance criteria for novel performance. Instead, the "acceptance criteria" are implied by successful comparative testing and adherence to existing standards and guidance documents. The "reported device performance" is the outcome of these tests.

    Test TypeImplied Acceptance Criteria (Demonstrated by Equivalence/Compliance)Reported Device Performance
    Sterilization ValidationAchieves a Sterility Assurance Level (SAL) of 10⁻⁶ via gamma irradiation. Complies with ISO 11137-1:2006 and ISO 11137-2:2013. For end-user sterilized components, validation according to ISO 17665-1, ISO 17665-2, and FDA guidance "Reprocessing Medical Devices in Health Care Settings."BLX implants, closure caps, and healing abutments are sterile via gamma irradiation (25 kGy minimum), validated to SAL 10⁻⁶ per ISO 11137-1:2006 and ISO 11137-2:2013 (overkill bioburden method). BLX Temporary Abutments and BLX Variobase Abutments are end-user sterilized by moist heat (steam), validated per ISO 17665-1, ISO 17665-2, and FDA reprocessing guidance.
    Shelf LifePackaging ensures maintaining sterility for a specified period and materials are not adversely affected.Shelf life for sterile devices remains 5 years, as packaging is equivalent to predicate/reference devices and materials are not adversely affected by time.
    PyrogenicityMeets pyrogen limit specifications based on LAL Endotoxin Analysis (e.g., < 20 EU/device).Pyrogenicity determined by LAL Endotoxin Analysis, with a testing limit of 20 EU/device for blood contacting and implanted devices.
    BiocompatibilityComplies with ISO 10993-1:2009 and FDA guidance "Use of International Standard ISO 10993-1." Equivalent nature of body contact, contact duration, material formulation, and sterilization methods compared to predicates.Biological assessment performed per ISO 10993-1:2009 and FDA guidance. Subject devices have equivalent body contact, duration, material formulation, and sterilization methods compared to predicates. Differences in manufacturing assessed via Chemical Characterization and Cytotoxicity tests.
    Dynamic Fatigue & Static StrengthFatigue load limits are equivalent to reference devices, especially for worst-case configurations.Tests performed according to FDA guidance "Class II Special Controls Guidance Document: Root-form Endosseous Dental Implants and Endosseous Dental Abutments." Both RB and WB platforms, for worst-case considerations, proved equivalent to reference devices.
    Insertion TorqueAllows reaching suitable implant insertion torques.Tests conducted on worst-case defined for each diameter (Ø4.5, Ø 5.5, Ø 6.5 mm) of the BLX portfolio (shortest, middle, longest implants). Demonstrated that BLX Implants and related cutting instruments allow reaching suitable implant insertion torques.
    Pullout Strength, Surface Area, B-I ContactEquivalent to reference devices.Performed for 6mm long BLX implants with Ø 4.5, 5.5, and 6.5 mm. Equivalency to reference devices was demonstrated.
    Software Verification & ValidationComplies with FDA guidance documents for CAD/CAM systems and general software validation. Software functions correctly within the digital workflow.Conducted and documented according to FDA guidance documents "Class II Special Controls Guidance Document: Optical Impression Systems for Computer Assisted Design and Manufacturing (CAD/CAM) of Dental Restorations" and "General Principles of Software Validation; Final Guidance for Industry and FDA Staff." Software considered "moderate" level of concern. Implementation verification shows Variobase devices and modeling components work together within the digital workflow for copings/crowns.

    2. Sample Size Used for the Test Set and Data Provenance

    The document does not explicitly state numerical sample sizes for most bench tests (e.g., fatigue, insertion torque, pullout strength). It refers to "worst case considerations" for selecting test configurations (e.g., shortest, middle, and longest implants for insertion torque; both RB and WB platforms for fatigue). This implies a representative selection but not a specific count.

    The data provenance is not explicitly stated as "country of origin." However, as this is an FDA submission for a device from "Institut Straumann AG" and "Straumann USA, LLC" (Andover, Massachusetts), the testing would typically be conducted in facilities compliant with US/international standards relevant to medical device approval, likely either in the US or Europe where Straumann has its main operations. The tests are retrospective, as they are being submitted to demonstrate the device meets prior requirements for market clearance.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

    This information is not applicable to this document. The document describes a medical device (dental implants and components) and its engineering and biological testing, not an AI/diagnostic software. Therefore, there is no "ground truth" to be established by experts in the context of diagnostic interpretation, and no test set in that specific sense.

    4. Adjudication Method for the Test Set

    This information is not applicable for the same reason as point 3. There is no diagnostic "test set" requiring adjudication by multiple readers or experts.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    This information is not applicable. This document describes a physical medical device (dental implants), not an AI-assisted diagnostic or clinical decision support tool.

    6. If a Standalone (i.e. algorithm only without human-in-the-loop performance) was done

    This information is not applicable. This document describes a physical medical device and some associated software for design/milling, not a standalone AI algorithm with performance metrics.

    7. The Type of Ground Truth Used (expert consensus, pathology, outcomes data, etc.)

    For the engineering tests (fatigue, strength, torque, pullout), the "ground truth" is defined by established engineering and material science principles, industry standards (e.g., ISO, ASTM, FDA guidance), and the performance of legally marketed predicate devices. The "truth" is whether the device meets the performance requirements set forth by these standards and demonstrates equivalence to existing, safe, and effective devices. For biocompatibility, the ground truth is compliance with ISO 10993 standards and a lack of adverse biological reactions.

    8. The Sample Size for the Training Set

    This information is not applicable. The device is a physical dental implant system, not an AI model that requires a "training set." The software mentioned is for CAD/CAM and is validated, not "trained" in the machine learning sense.

    9. How the Ground Truth for the Training Set was Established

    This information is not applicable for the same reason as point 8.

    Ask a Question

    Ask a specific question about this device

    K Number
    K163194

    Validate with FDA (Live)

    Date Cleared
    2017-07-14

    (241 days)

    Product Code
    Regulation Number
    872.3640
    Panel
    Dental
    Age Range
    12 - 150
    Reference & Predicate Devices
    Why did this record match?
    Reference Devices :

    K150182, K153624, K093027, K151455, K130808, K061804

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Indications for Use for GM implants and conventional abutments:

    The Neodent Implant System is intended to be surgically placed in the bone of the upper or lower jaw to provide support for prosthetic devices such as artificial teeth, to restore chewing function. It may be used with single-stage or two-stage procedures, for single or multiple unit restorations, and may be loaded immediately when good primary stability is achieved and with appropriate occlusal loading.

    Indications for Use for GM Titanium Base abutments:

    Titanium Base Abutment is a titanium base placed onto Neodent dental implants to provide support for customized prosthetic restorations. It is used with a coping and crown, or crown alone, and is indicated for cement-retained single or multi-unit restorations, or screw-retained single restorations.

    All digitally designed copings and/or crowns for use with the Neodent Titanium Base Abutment System are intended to be sent to Straumann for manufacture at a validated milling center.

    Indications for Use for GM Pro Peek Abutments:

    The Pro PEEK Abutments are indicated to be used on Neodent implants to provide temporary support for prosthesis structure for up to 6 months. They can be used in one or two stage procedures and also immediate load when there is good primary stability.

    Device Description

    The GM Line of the Neodent Implant System (GM Line) consists in an expansion of the Neodent Implant System previously cleared under K101945, K150182 and K153624, presenting a new design of Morse taper implant-to-abutment interface trade named GM line.

    The GM line comprises a range of endosseous dental implants as well as conventional and CAD/CAM abutments described as follows:

    The GM Line dental implants are threaded, self-tapping, root form with a Morse taper implantto-abutment interface with an internal hexagonal index exclusive to the GM line. They are made of titanium grade 4 conforming to ASTM F67. They are available in two types of surface treatment:

    NeoPoros: rough surface created using an abrasive particle jet concept with controlled grain oxides, followed by acid etching creating uniform cavities in the implant surface.

    Acqua: hydrophilic surface created from Neoporos rough surface that undergoes additional processing that renders hydrophilic surface.

    GM Titamax implants: cylindrical shape with double threads of pyramidal profile, rounded apex.

    GM Helix implants: conical shape, double threads with trapezoidal profile, a rounded apex.

    GM Drive implants: conical shape with double and progressive threads, cutting chambers in the counterclockwise direction, blade-shaped apical thread, rounded apex.

    The GM Line dental abutments are made of titanium alloy conforming to ASTM F136 and have a machined surface. The abutments are intended to be placed directly onto implants and are recommended according to the available interocclusal space, gingival height, and threedimensional position of the implant. They can be used in immediate or conventional rehabilitation procedures on the maxilla or mandible.

    GM Cover Screws: devices for temporary use during the implant healing phase.

    GM Healings: devices for temporary use for maintenance of the soft tissue during the osseointegration phase.

    GM Micro Abutments / GM Mini Conical Abutment / GM Exact Mini Conical Abutment: intermediary prosthetic components to be installed onto GM implants to support the final prosthesis.

    GM Exact Abutments: intermediary prosthetic components to be installed onto GM implants to support the final prosthesis.

    GM Exact Click Universal Abutments (straight and angled): intermediary prosthetic components to be installed onto GM implants to support the final prosthesis.

    GM Exact Titanium Base abutments: intermediary prosthetic components to be installed onto GM implants to support the final prosthesis; the Titanium Base two-piece abutments used as a base when fabricating a CAD/CAM customized restoration.

    GM Pro PEEK Abutment: temporary intermediary prosthetic components to be installed onto the Implants to support the provisional prosthesis up to 6 months.

    Protection Cylinders: prosthetic components with a tapered shape intended to be placed on abutments to protect them during the fabrication of the prosthesis and/or healing of the peri-implant tissues.

    Coping Screws: prosthetic devices intended for coping fixation on the corresponding GM (screw-retained) abutment.

    AI/ML Overview

    This document is a 510(k) Premarket Notification from the FDA regarding the "Neodent Implant System - GM Line." It primarily focuses on demonstrating substantial equivalence to previously cleared predicate devices, rather than an independent study proving the device meets specific acceptance criteria in a clinical or performance study context.

    Therefore, the information requested in your prompt (Acceptance Criteria, Study Design for proving performance, Sample Sizes, Expert consensus, MRMC studies, etc.) is not present in this document. This document details the comparison of the new device with existing ones based on:

    1. Indications for Use: The new device's indications align with or are subsumed by those of the predicates.
    2. Technological Characteristics: Materials, design features, sterilization methods, and compatibility are compared.
    3. Performance Data (Mechanical Testing and Biocompatibility): This section confirms that the materials meet established standards (ASTM, ISO) and that the device components (implants, abutments) have undergone mechanical fatigue testing according to ISO 14801 and FDA guidance. Biocompatibility has been assessed for the materials used. However, it does not provide specific acceptance criteria or detailed results of these tests, simply stating that the performance is "equivalent" to predicate devices.

    Summary of missing information based on your prompt:

    • Table of acceptance criteria and reported device performance: Not provided in a quantifiable manner for clinical performance. Only mentions "equivalent" mechanical performance based on ISO standards.
    • Sample sizes for test set and data provenance: No clinical test set involving human subjects is described. Mechanical tests likely involved specific numbers of devices, but these are not disclosed.
    • Number of experts and qualifications for ground truth: Not applicable as no clinical ground truth assessment is detailed.
    • Adjudication method for test set: Not applicable.
    • MRMC comparative effectiveness study: Not conducted or reported. This submission relies on substantial equivalence to existing devices, not a comparative clinical trial with human readers/outcomes.
    • Standalone performance (algorithm only): Not applicable as this is a physical medical device (dental implant system), not an AI algorithm.
    • Type of ground truth used: For regulatory purposes, the "ground truth" for this submission is established through comparison to legally marketed predicate devices and adherence to relevant industry standards (ISO, ASTM) and FDA guidance for mechanical and biocompatibility testing. No clinical "ground truth" derived from patient outcomes or expert consensus on images is mentioned regarding the device's function.
    • Sample size for training set: Not applicable as this is not an AI/ML device.
    • How ground truth for training set was established: Not applicable.

    In conclusion, this document demonstrates regulatory compliance and substantial equivalence through a comparison of device characteristics and a statement of meeting mechanical and biocompatibility standards, rather than presenting a detailed study proving the device meets specific, quantifiable clinical acceptance criteria.

    Ask a Question

    Ask a specific question about this device

    K Number
    K160964

    Validate with FDA (Live)

    Date Cleared
    2016-09-16

    (163 days)

    Product Code
    Regulation Number
    872.3630
    Panel
    Dental
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    Why did this record match?
    Reference Devices :

    K153624, K111421

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Titanium Base for CEREC is a titanium component that is placed over Neodent Implants to provide support for custom prosthetic restorations, such as copings or crowns. It is indicated for single-tooth screw-retained restorations.

    All digitally designed copings and/or crowns for use with the Neodent Titanium Base for CEREC are to be designed using Sirona inLab software or Sirona CEREC Software and manufactured using a Sirona CEREC or inLab MC X or MC XL milling unit.

    Device Description

    The Titanium Base for CEREC consists of a two-piece abutment, where the titanium base is a pre-manufactured abutment that will be used to support a CAD/CAM designed superstructure (the second part of the two-piece abutment) that composes the final abutment.

    The Titanium Base for CEREC abutments are made of titanium alloy conforming to ASTM F136 Standard Specification for Wrought Titanium-6Aluminum-4Vanadium ELI (Extra Low Interstitial) Alloy for Surqical Implant Applications and are provided in a 4.65 mm prosthetic platform diameter and seven gingival heights (0.8, 1.5, 2.5, 3.5, 4.5, 5.5 and 6.5 mm), all having a 4.7 mm prosthetic height. They also feature:

    • cylindrical shape
    • hexagonal indexing at the apical end of the Morse taper connection
    • indexing guide in the cementable portion for coping fitting

    The CAD/CAM customized superstructure that composes the final abutment must be designed and milled through the Sirona Dental CAD/CAM System, according to the prosthetic planning and patient clinical situation.

    The Titanium Base for CEREC is provided sterile by Ethylene Oxide and steam sterilization is recommended after the cementation of the customized superstructure on the Titanium Base for CEREC.

    The Titanium Base for CEREC is compatible with the following devices:

    Dental implants
    All Neodent dental implants having Morse taper implant-to-abutment interface (CM line) cleared under K101945, K123022, K133592, K150182 and K150199.

    Raw material blanks

    • -InCoris Zi (ZrO2) by Sirona Dental Systems GmbH, L size blanks, cleared under K123664;
    • -IPS e.max CAD Abutments Solutions (LiSi₂) by Ivoclar Vivadent AG, L size blanks, cleared under K132209.

    Software
    Sirona Dental CAD/CAM System, by Sirona Dental Systems GmbH, cleared under K111421.

    Instruments

    • -Neodent screwdrivers (class I - exempt devices)
    • Sirona scanbodies (class I exempt devices)
    AI/ML Overview

    The provided text is a 510(k) premarket notification for the "Neodent Implant System - Titanium Bases For Cerec" and primarily focuses on demonstrating substantial equivalence to predicate devices, rather than establishing specific acceptance criteria and proving device performance through a standalone study with a test set, expert ground truth, or MRMC studies.

    Therefore, much of the requested information regarding acceptance criteria, specific study design elements (sample sizes, expert qualifications, adjudication methods, MRMC studies, standalone performance), and details on training set ground truth cannot be directly extracted from this document, as it outlines a comparison and testing against existing standards and predicate devices.

    However, I can extract information about the types of performance data used to support substantial equivalence, which acts as a proxy for meeting certain "acceptance criteria" related to safety and effectiveness by demonstrating similarity to already cleared devices.

    Here's a breakdown of what can be inferred:

    1. A table of acceptance criteria and the reported device performance

    The document does not present explicit acceptance criteria with numerical targets in the traditional sense, as it relies on demonstrating substantial equivalence to predicate devices and adherence to recognized standards. The "reported device performance" is framed as meeting these standards or being comparable to predicate devices.

    Acceptance Criteria (Inferred from testing performed)Reported Device Performance (as stated in document)
    Biocompatibility: Device material is biocompatible.Made of titanium alloy conforming to ASTM F136; same material as previously cleared Neodent Implant System components (K150367). No new biocompatibility testing performed, as material is commonly used and cleared.
    Mechanical Strength: Device demonstrates sufficient strength and fatigue resistance.Strength demonstrated through fatigue testing according to ISO 14801 and FDA Guidance for Industry. No new fatigue tests performed as the device does not present a new worst case compared to reference device (K153624).
    Software Verification & Validation: Software functions correctly and does not pose undue risk.Considered "moderate" level of concern. Verified and validated with respect to functionality and design using the Sirona validated workflow (same as primary predicate).
    Sterilization Efficacy (Manufacturer): Sterilization process achieves appropriate Sterility Assurance Level (SAL).Validated according to ISO 11135 (Ethylene Oxide) to achieve a SAL of 10⁻⁶.
    Sterilization Efficacy (End-user): Recommended end-user sterilization method is effective.Moist heat (steam) sterilization validated according to ISO 17665-1 and ISO/TR 17665-2 to achieve a SAL of 10⁻⁶.
    Substantial Equivalence: Features and performance are comparable to identified predicate devices.Detailed comparison table (Table 2) provided demonstrating equivalence in indications, dimensions, connections, materials, manufacturing workflow, and sterilization.

    2. Sample sized used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    • Sample Size for Test Set: Not explicitly stated as a separate "test set" for performance evaluation in the typical sense.

      • Biocompatibility: No new testing was performed; reliance on prior clearance of the material.
      • Mechanical Testing: Not specified for this submission, but it refers to tests presented in K153624 (a reference predicate). For fatigue testing per ISO 14801, a specific number of samples (typically 5, 8, or 10, depending on the method) would be tested. The number isn't provided here.
      • Software Verification & Validation: The document states the "Titanium Base for CEREC was verified and validated with respect to its functionality and design using the Sirona validated workflow." This implies testing against the workflow, but specific sample sizes (e.g., number of test cases, scenarios) are not provided.
      • Sterilization Validation: Sterilization validation studies typically involve a defined number of biological indicators (BIs) or process challenge devices (PCDs) per run, over multiple runs. Specific numbers are not provided, only the standards used (ISO 11135, ISO 17665-1).
    • Data Provenance: Not explicitly stated. Given Neodent's location in Brazil (JJGC Indústria e Comércio de Materiais Dentários SA, Curtiba, Parana, BRAZIL), it's reasonable to infer that any original data generation would have been conducted in either Brazil or relevant testing laboratories supporting their operations, or by the manufacturers of the predicate devices. The study type (retrospective/prospective) is not detailed.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

    Not applicable. This device is a dental implant component, not an AI or diagnostic imaging device that requires expert adjudication for ground truth. Its performance is evaluated through material properties, mechanical testing, and software validation.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    Not applicable, as this is not an AI or diagnostic imaging device for which such adjudication methods would be relevant.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    Not applicable. This is not an AI-assisted diagnostic device.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    Not applicable. This is a physical medical device. Software verification was performed, but not as a "standalone algorithm" in the typical sense of AI performance evaluation.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

    The "ground truth" for this device's performance is established by:

    • Material Specifications: Adherence to ASTM F136 for titanium alloy.
    • Mechanical Performance: Demonstrating strength and fatigue resistance according to ISO 14801.
    • Sterilization Efficacy: Achieving performance per ISO 11135 and ISO 17665-1 standards.
    • Software Functionality: Verification and validation against predefined functional requirements within the Sirona validated workflow.
    • Predicate Device Equivalence: The ultimate "ground truth" for this 510(k) is that the device is as safe and effective as its legally marketed predicate devices.

    8. The sample size for the training set

    Not applicable. This device is not an AI algorithm trained on data sets. The "training set" concept is not relevant here.

    9. How the ground truth for the training set was established

    Not applicable, for the same reason as point 8.