Search Filters

Search Results

Found 4 results

510(k) Data Aggregation

    K Number
    K220716

    Validate with FDA (Live)

    Device Name
    Spartan SC 069
    Date Cleared
    2022-05-05

    (55 days)

    Product Code
    Regulation Number
    870.1250
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    Why did this record match?
    Reference Devices :

    K150107

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Spartan SC 069TM is indicated for the introduction of interventional devices into the peripheral and neuro vasculature.

    Device Description

    The Spartan SC 069™ is a support catheter designed for the introduction of interventional devices into the peripheral and neuro vasculature. The SC 069™ is a single lumen, flexible, variable stiffness composite catheter with a Nitinol structure. A radiopaque marker band on the distal tip of the device is used for visualization under fluoroscopy. The distal section of the catheter is coated with a hydrophilic coating to reduce the overall frictional force during intravascular use. The device is supplied sterile and intended for single use only.

    AI/ML Overview

    The provided text describes the performance data for the Spartan SC 069™ support catheter. This device appears to be a medical device, and the information presented is from a 510(k) premarket notification to the FDA, which focuses on demonstrating substantial equivalence to a predicate device, rather than a clinical study evaluating the device's diagnostic performance. Therefore, many of the requested fields regarding diagnostic acceptance criteria, multi-reader multi-case studies, and human-in-the-loop performance are not applicable.

    Here's the information extracted from the document:

    1. A table of acceptance criteria and the reported device performance

    The document lists various non-clinical bench tests and biocompatibility tests, with each test summary implying a pass/fail acceptance criterion. The results consistently state that the Spartan SC 069™ "met the acceptance criteria" or showed "acceptable" performance, or similar to the predicate device.

    TestTest Method Summary (Implied Criteria)Reported Device Performance
    Coating Lubricity (Friction Force)Acceptable friction force under simulated use conditions similar to predicate.The Spartan SC 069™ was found to have acceptable friction force under simulated use conditions similar to the predicate device.
    Hub Functional & DimensionalMeet ISO 594-1:1986-06-15 and ISO 594-2:1998-09-01 requirements.The Spartan SC 069™ met the acceptance criteria for hub functional and dimensional requirements.
    Torque StrengthAcceptable torsional strength similar to predicate in simulated path model.The Spartan SC 069™ exhibited acceptable torsional strength similar to the predicate device.
    TensileMeet ISO 10555-1:2013 Annex B.The Spartan SC 069™ met the acceptance criteria for tensile strength.
    Air AspirationMeet ISO 10555-1:2013(E) Annex D for air leakage.The Spartan SC 069™ met the acceptance criteria for air aspiration.
    Liquid LeakMeet ISO 10555-1:2013(E) Annex C.The Spartan SC 069™ met the acceptance criteria for liquid leakage.
    Particulate and Coating IntegrityAcceptable particulate generation and coating integrity, substantially equivalent to predicate.The Spartan SC 069™ met the acceptance criteria for particulate generation and coating integrity, and was found substantially equivalent to the predicate.
    StiffnessMeet acceptance criteria for catheter stiffness, substantially equivalent to predicate.The Spartan SC 069™ met the acceptance criteria for catheter stiffness and found substantially equivalent to the predicate.
    Kink ResistanceMeet acceptance criteria for kink resistance, substantially equivalent to predicate.The Spartan SC 069™ met the acceptance criteria for kink resistance and was found substantially equivalent to the predicate.
    RadiopacityMeet acceptance criteria for radiopacity, substantially equivalent to predicate.The Spartan SC 069™ met the acceptance criteria for radiopacity and was found substantially equivalent to the predicate.
    CorrosionMeet ISO 10555-1:2013(E) Annex A.The Spartan SC 069™ met the acceptance criteria for corrosion resistance.
    Static Burst PressureMeet ISO 10555-1:2013(E) Annex F for burst pressure.The Spartan SC 069™ met the acceptance criteria for static burst pressure.
    BucklingMeet acceptance criteria for tip deflection and buckling, substantially equivalent to predicate.The Spartan SC 069™ met the acceptance criteria for tip deflection and buckling and was found substantially equivalent to the predicate.
    Design ValidationMeet requirements for interventional device introduction, trackability, stability, guidewire interaction, and overall device integrity in a challenging use setting.The Spartan SC 069™ met the requirements for which it was designed and tested.
    Dimensional TestingMeet device specifications.The Spartan SC 069™ met the dimensional requirements.
    Visual InspectionMeet acceptance criteria.The Spartan SC 069™ met the visual inspection requirements.
    Dead Space VolumeMeasured and reported in labeling.The Spartan SC 069™ dead space volume was measured. The dead space volume is reported in the labeling.
    Pressure Flow Testing with Contrast MediaEvaluated flow rates and pressures of various solutions; reported in labeling.The Spartan SC 069™ pressure-flow performance was evaluated. Flow rates and pressures of various saline and contrast media solutions are reported in the labeling.
    Biocompatibility Tests
    Sensitization (Guinea Pig)No reaction at challenge; classified as non-sensitizer per ISO 10993-10.Non-sensitizer.
    Genotoxicity (Mouse Lymphoma)IMF of Test Article < GEF; considered non-mutagenic per ISO 10993-3.Non-mutagenic.
    Genotoxicity (Salmonella Typhimurium & Escherichia Coli)No statistically significant increase in revertant colonies compared to negative controls per ISO 10993-3.Non-genotoxic.
    Cytotoxicity (L929 MEM Elution)No cell lysis or reduction of cell growth at 24/48 hours; considered non-cytotoxic per ISO 10993-5.Non-cytotoxic.
    Irritation/Intracutaneous ReactivityNo significantly greater biological reaction than control per ISO 10993-10.Non-irritant.
    Acute Systemic ToxicityNo significantly greater biological reaction than control extracts per ISO 10993-11.No systemic toxicity.
    Pyrogenicity (Rabbit Pyrogen Test)Temperature increases did not exceed test limit; considered non-pyrogenic per ISO 10993-11.Non-pyrogenic.
    Hemocompatibility (Rabbit Blood Hemolysis)Hemolysis < 5% (direct and indirect contact) per ISO 10993-4.Non-hemolytic.
    Hemocompatibility (Partial Thromboplastin Time)Test article clotting time not significantly different from comparison article per ASTM International F2382-18 and ISO 10993-4.Hemocompatible.
    Hemocompatibility (SC5B-9 Complement Activation)No statistically significant increase in SC5b-9 concentrations compared to controls per ISO 10993-4.Hemocompatible.
    Hemocompatibility (Platelet and Leukocyte Binding)No effect on platelet and leukocyte concentrations per ISO 10993-4.Hemocompatible.
    Sterilization
    Sterilization ValidationAchieved a Sterility Assurance Level of 10-6 using overkill (half cycle) approach per ISO 11135:2014.Successful validation.
    Shelf Life
    Accelerated AgingMaintained safety and effectiveness after 2-year equivalent accelerated aging (55 °C, 75 days) and simulated shipping.Confirmed 2-year shelf life.

    2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    The document relates to non-clinical bench testing and biocompatibility assessments of a medical device (catheter). It does not involve a "test set" in the context of diagnostic accuracy, human subjects, or data provenance like country of origin. The tests were performed in a laboratory setting for regulatory submission.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

    This information is not applicable as the study involves bench testing and biocompatibility of a medical device, not diagnostic performance requiring expert ground truth.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    This information is not applicable as the study involves bench testing and biocompatibility of a medical device, not diagnostic performance requiring adjudication.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    This information is not applicable. This is a 510(k) submission for a physical medical device (catheter), not an AI-assisted diagnostic tool. No MRMC study was conducted.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    This information is not applicable. This is a 510(k) submission for a physical medical device (catheter), not an algorithm or AI.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

    The "ground truth" for these performance studies is established by the specifications and standards (e.g., ISO standards, internal design requirements) that the device must meet, as well as comparison to the predicate device. For biocompatibility, it's about not eliciting adverse biological responses.

    8. The sample size for the training set

    This information is not applicable. There is no "training set" as this is not an AI/machine learning study.

    9. How the ground truth for the training set was established

    This information is not applicable. There is no "training set" as this is not an AI/machine learning study.

    Ask a Question

    Ask a specific question about this device

    K Number
    K180705

    Validate with FDA (Live)

    Date Cleared
    2018-07-04

    (107 days)

    Product Code
    Regulation Number
    870.1250
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    Why did this record match?
    Reference Devices :

    K172448, K150107

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Riptide™ Aspiration System is intended for use in the revascularization of patients with acute iscoke secondary to intracranial large vessel occlusive disease (within the internal carotid, middle cerebral - M1 and M2 segments, basilar, and vertebral arteries) within 8 hours of symptom onset. Patients who are ineligible for intravenous tissue plasminogen activator (IV t-PA) or who fail IV t-PA therapy are candidates for treatment.

    Device Description

    The Riptide™ Aspiration System React™ 68 Catheter is a single lumen, flexible, variable stiffness composite catheter with a nitinol structure that is jacketed with a durable polymer outer layer. A lubricous, polytetrafluoroethylene liner is used to create a structure that has both proximal stiffness and distal flexibility, and an encapsulated radiopaque distal platinum-iridium markerband which is used for visualization under fluoroscopy. The Riptide™ Aspiration System React™ 68 Catheter is introduced into the vasculature through the split y-introducer sheath.

    The proximal end of the Riptide™ Aspiration System React™ 68 Catheter is designed with a thermoplastic elastomer strain relief and a clear hub. The Riptide™ Aspiration System React™ 68 Catheter is designed with a hydrophilic coating. The Riptide™ Aspiration System React™ 68 Catheter is navigated to the intended treatment site and positioned proximal to the site of occlusion.

    AI/ML Overview

    The document describes the acceptance criteria and supporting studies for the Riptide Aspiration System (React 68 Catheter). However, it's important to note that this document is a 510(k) summary, which focuses on demonstrating substantial equivalence to a predicate device rather than providing a detailed clinical study report with specific performance metrics against pre-defined acceptance criteria for a new AI/software device. The device in question is a medical device (catheter) for stroke revascularization, not an AI/software device that typically has performance metrics like sensitivity, specificity, or F1-score.

    The acceptance criteria provided in the document are primarily for bench testing and biocompatibility, and the reported "performance" is whether the device "met the acceptance criteria" or specific qualitative observations, rather than a quantitative measurement against a numerical threshold.

    Here's the information extracted from the provided text, structured to best fit your request, with an understanding that the device is a physical catheter and not an AI/software product in the context of typical AI acceptance criteria.


    1. Table of Acceptance Criteria and Reported Device Performance

    For the Riptide Aspiration System (React 68 Catheter), acceptance criteria are primarily related to bench testing, biocompatibility, and animal testing. The reported "performance" is whether the device met these criteria.

    Test DescriptionAcceptance Criteria (Implied / Explicit)Reported Device Performance
    Biocompatibility
    Cytotoxicity (Elution Method)No evidence of causing cell lysis or toxicity (grade of zero reactivity)Test article extract showed no evidence of causing cell lysis or toxicity and had a grade of zero (no reactivity). Considered non-cytotoxic.
    Sensitization (Guinea Pig Max. Test)Grade less than two (mild reactivity)Test article extract met requirements (grade less than two). Does not elicit a sensitization response.
    Irritation (Intracutaneous Reactivity)Difference between test article and control extract mean scores = 0.0Difference between test article extract overall mean score and corresponding control extract mean score was 0.0. Considered a non-irritant.
    Acute Systemic Toxicity (Systemic Toxicity)No mortality or evidence of systemic toxicity; clinically normal animalsNo mortality or evidence of systemic toxicity. All animals clinically normal. Does not indicate signs of toxicity.
    Hemocompatibility (Hemolysis)Hemolytic index for direct contact blood < 5% (typically); no hemolysisHemolytic index for direct contact with blood was 0.9%, for extract was 0.0%. Considered non-hemolytic.
    Hemocompatibility (Complement Activation)Not statistically higher concentration of SC5b-9 than controlsConcentration of SC5b-9 in test article sample was not statistically higher than controls. Not considered potential activators of complement system.
    Hemocompatibility (Thrombogenicity)Minimal thrombus formationDetermined to have minimal thrombus formation after four (4) hours. Demonstrates minimal thrombus formation.
    Pyrogenicity (Material Mediated)No animal showed a temperature rise of 0.5°C or more above baselineNot a single animal showed a temperature rise of 0.5°C or more. Total rise was 0.0°C. Considered non-pyrogenic.
    Bench Performance Data
    Ethylene Oxide ResidualMet acceptance criteria per ISO 10993-7Met the acceptance criteria for ethylene oxide residual.
    Ethylene Chlorohydrin ResidualMet acceptance criteria per ISO 10993-7Met the acceptance criteria for ethylene chlorohydrin residual.
    BioburdenMet acceptance criteria per ISO 11737-1Met the acceptance criteria for bioburden.
    Bacterial EndotoxinMet acceptance criteria per ANSI/AAMI ST72 and USP <161>Met the acceptance criteria for bacterial endotoxin.
    Visual InspectionMet acceptance criteria for visual inspectionMet the acceptance criteria for visual inspection.
    Dimensional MeasurementsMet acceptance criteria for dimensional measurementsMet the acceptance criteria for dimensional measurements.
    Tip BucklingMet acceptance criteria for maximum compressive forceMet the acceptance criteria for tip buckling.
    Kink ResistanceMet acceptance criteria for maximum kink diameterMet the acceptance criteria for kink resistance.
    ParticulateMet acceptance criteria per USP <788>Met the acceptance criteria for particulate evaluation.
    Coating LubricityMet acceptance criteria for average friction forcesMet the acceptance criteria for coating lubricity.
    Tensile StrengthMet acceptance criteria per ISO 10555-1. Annex BMet the acceptance criteria for tensile strength at the hub and shaft.
    Liquid LeakMet acceptance criteria per ISO 10555-1. Annex CMet the acceptance criteria for liquid leak.
    Corrosion ResistanceMet acceptance criteria per ISO 10555-1. Annex AMet the acceptance criteria for corrosion resistance.
    Hub Aspiration ResistanceMet acceptance criteria per ISO 10555-1. Annex DMet the acceptance criteria for hub air aspiration.
    RadiopacityMet acceptance criteria for markerband length, wall thickness, and fluoroscopy confirmationMet the acceptance criteria for radiopacity.
    Luer StandardsMet acceptance criteria per ISO 594-1 and ISO 80369-7Met the acceptance criteria for luer standards.
    Torque StrengthAble to withstand torsional forces typical of clinical useWas able to withstand torsional forces that are typical of clinical use.
    Dynamic PressureAble to withstand pressures typical of clinical useWas able to withstand pressures that are typical of clinical use.
    Coating IntegrityCoating coverage and lubricity maintainedRemained coated and lubricous.
    RecanalizationMet acceptance criteria for recanalizationMet the acceptance criteria for recanalization.
    Vacuum PressureMet acceptance criteria for vacuum pressureMet the acceptance criteria for vacuum pressure.
    Suction Flow RateMet acceptance criteria for suction flow rateMet the acceptance criteria for suction flow rate.
    UsabilityMet acceptance criteria for maneuverability, flexibility, aspiration, and clot retrievalMet the acceptance criteria for usability.
    Animal Testing
    Safety, Efficacy, UsabilityDemonstrated comparable safety, efficacy, and usability to predicate deviceEvaluated in a porcine model at acute and chronic time points, in presence/absence of simulated clot. This testing supports substantial equivalence.

    2. Sample Sizes and Data Provenance (Not applicable for AI/software test set, but for testing performed)

    • Biocompatibility and Bench Testing: The sample sizes are not explicitly stated for individual tests but are implied to be sufficient for standard regulatory testing (e.g., "The test article extract showed no evidence...", "Not a single animal showed..."). These are laboratory tests on physical devices/extracts.
    • Animal Testing: Conducted in a porcine model. Data provenance is from non-clinical laboratory studies performed in accordance with 21 CFR Part 58 for Good Laboratory Practice (GLP). This is prospective animal testing.

    3. Number of Experts and Qualifications (Not applicable to this type of device submission for establishing ground truth)

    This information is typically relevant for AI/software devices where human experts establish ground truth for image or data interpretation. For a physical catheter, ground truth is established through physical measurements, chemical analysis, and physiological responses as per standardized test methods. The document does not mention experts for ground truth establishment in this context.

    4. Adjudication Method (Not applicable)

    Adjudication methods (like 2+1, 3+1) are used to resolve disagreements among human labelers/experts when establishing ground truth for AI/software performance evaluation. Not relevant for this physical device submission.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study (Not done/applicable)

    MRMC studies are typically performed for AI/software devices to assess the impact of AI assistance on human reader performance. This device is a physical catheter, not an AI/software system, so such a study was not performed. The submission is based on demonstrating substantial equivalence to a predicate device through bench and animal testing.

    6. Standalone Performance (Algorithm only without human-in-the-loop performance) (Not applicable)

    This concept applies to AI/software algorithms. The Riptide Aspiration System is a physical catheter intended for use by a clinician. Its "performance" is inherently linked to its physical properties and interaction within a biological system, not a standalone algorithmic output.

    7. Type of Ground Truth Used

    The "ground truth" for this device's evaluation is primarily established by:

    • Standardized Test Methods and Criteria: For biocompatibility, microbial, and performance bench tests, "ground truth" is defined by adherence to specific ISO standards (e.g., ISO 10993, ISO 10555, ISO 11737), USP standards, and ANSI/AAMI standards. These standards define the acceptable range or qualitative outcome.
    • Physiological Observations and Measurements: In animal studies, ground truth for safety, efficacy, and usability comes from direct observation of animal health, physiological responses, and procedural outcomes.

    8. Sample Size for the Training Set (Not applicable)

    This concept is for AI/machine learning models. The Riptide Aspiration System is a physical medical device. It does not have a "training set" in the context of AI. Design and manufacturing improvements would be based on engineering principles and iterative testing.

    9. How the Ground Truth for the Training Set was Established (Not applicable)

    Again, this is an AI/ML concept. The "ground truth" for the development of catheter designs comes from engineering specifications, material science, and prior knowledge of predicate devices and clinical needs, rather than a labeled training set.

    Ask a Question

    Ask a specific question about this device

    K Number
    K172448

    Validate with FDA (Live)

    Date Cleared
    2017-12-21

    (129 days)

    Product Code
    Regulation Number
    870.1250
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    Why did this record match?
    Reference Devices :

    K150107, K133317, K122756, K090752

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Riptide™ Aspiration System is intended for use in the revascularization of patients with acute ischemic stroke secondary to intracranial large vessel occlusive disease (within the internal carotid, middle cerebral – M1 and M2 segments, basilar, and vertebral arteries) within 8 hours of symptom onset. Patients who are ineligible for intravenous tissue plasminogen activator (IV t-PA) or who fail IV t-PA therapy are candidates for treatment.

    Device Description

    The Riptide™ Aspiration System is designed to restore blood flow in patients with acute ischemic stroke secondary to intracranial large vessel occlusive disease. The Riptide™ Aspiration System is designed for use within the internal carotid, middle cerebral – M1 and M2 segments, basilar, and vertebral arteries. The Riptide™ Aspiration System is composed of the following components:

    • Arc™ Catheter .
    • Riptide™ Aspiration Tubing .
    • Riptide™ Aspiration Pump ●
    • Riptide™ Collection Canister with Intermediate Tubing .

    The Arc™ Catheter is introduced into the vasculature through the Split-Y Introducer Sheath. A lubricous, tapered liner is used to create a structure that has both proximal stiffness and distal flexibility. The Arc™ Catheter has a radiopaque marker band encapsulated at the distal tip for visualization under fluoroscopy. The Arc™ Catheter is navigated to the intended treatment site and positioned proximal to the site of occlusion. The Arc™ Catheter is the only component of the Riptide™ Aspiration System that is used intravascularly.

    The Riptide™ Aspiration Tubing serves as a conduit to supply vacuum from the Riptide™ Aspiration Pump to the distal tip of the Arc™ Catheter. The Riptide™ Aspiration Tubing provides a connection between the sterile and non-sterile environments. The proximal end of the Riptide™ Aspiration Tubing is connected to the Riptide™ Collection Canister (outside of the sterile environment) while the distal end of the Riptide™ Aspiration Tubing is connected to the Arc™ Catheter (inside the sterile environment). The Riptide™ Collection Canister is connected to the Riptide™ Aspiration Pump (also outside of the sterile environment) via the Intermediate Tubing.

    The Riptide™ Aspiration Pump is designed to generate vacuum for the Riptide™ Aspiration System. The vacuum pressure of the Riptide™ Aspiration Pump is set by turning the vacuum control valve until the vacuum gauge reads a minimum of 20inHg but not exceeding 25inHg. The Riptide™ Aspiration Pump is reusable, non-sterile, and intended to be utilized outside of the sterile environment.

    The Riptide™ Collection Canister is provided non-sterile and is pre-assembled with the Intermediate Tubing. The Riptide™ Collection Canister with Intermediate Tubing is single-use and the repository for aspirated material. The Riptide™ Collection Canister is placed into the receptacle of the Riptide™ Aspiration Pump while the Intermediate Tubing is connected to the vacuum inlet port.

    AI/ML Overview

    The provided text describes the acceptance criteria and the studies conducted to demonstrate the substantial equivalence of the Riptide™ Aspiration System to its predicate device, the Penumbra System® and Penumbra Pump MAX™.

    Here's the breakdown of the information requested:


    1. Table of Acceptance Criteria and Reported Device Performance

    The document lists numerous tests conducted on various components of the Riptide™ Aspiration System, with an explicit statement that the device met the acceptance criteria for each. The specific acceptance criteria values are generally not quantified in the "Results" column, but rather stated qualitatively (e.g., "met the acceptance criteria").

    Table 1: Acceptance Criteria and Reported Device Performance (Summary)

    ComponentTest CategoryIndividual TestAcceptance Criteria Statement (from document)
    Arc™ CatheterBiocompatibilityUSP PhysicochemicalMeets USP Physicochemical (Pass)
    ISO MEM Elution Using L-929 Mouse Fibroblast CellsNon-cytotoxic
    ISO Guinea Pig Maximization Sensitization TestNon-sensitizer
    ISO Intracutaneous Irritation TestNon-irritant
    ISO Acute Systemic Injection TestNon-cytotoxic
    ISO Materials Mediated Rabbit PyrogenNon-pyrogenic
    ASTM Hemolysis Assay – Direct Contact and Extract MethodNon-hemolytic
    Complement activation C3a and SC5b-9 AssayThe compliment activation of the C3a and SC5b assays were similar for test and comparison articles.
    Thromboresistance EvaluationThe test and control articles exhibited similar thromboresistant characteristics.
    ISO Bacterial Mutagenicity Test - Ames AssayNon-mutagenic
    ISO in vitro Mouse Lymphoma with Extended TreatmentNon-mutagenic (non-genotoxic and non-clastogenic)
    ISO in vivo Mouse Micronucleus AssayNon-mutagenic
    Partial Thromboplastin TimeNon-activator (of intrinsic coagulation pathway)
    in vitro Hemocompatibility AssayNo adverse effect on leukocyte or platelet counts.
    Arc™ CatheterBench - MicrobialEthylene Oxide (EO) ResidualMet the acceptance criteria for EO Residual.
    Ethylene Chlorohydrin (ECH)Met the acceptance criteria for ECH.
    Bacterial EndotoxinMet the acceptance criteria for Bacterial Endotoxin.
    Bench - PackagingVisual InspectionMet the acceptance criteria for Visual Inspection.
    Bubble LeakMet the acceptance criteria for Bubble Leak.
    Seal StrengthMet the acceptance criteria for Seal Strength.
    Bench - PerformanceLumen PatencyMet the acceptance criteria for Lumen Patency.
    Dimensional InspectionMet the acceptance criteria for Dimensional Inspection.
    Tip BucklingMet the acceptance criteria for Tip Buckling.
    Injection Flow RateMet the acceptance criteria for Injection Flow Rate.
    Suction Flow RateMet the acceptance criteria for Suction Flow Rate.
    Vacuum ResistanceMet the acceptance criteria for Vacuum Resistance.
    Air Aspiration LeakMet the acceptance criteria for Air Aspiration Leak.
    Hub/Shaft Peak Tensile ForceMet the acceptance criteria for Hub/Shaft Peak Tensile Force.
    Coating IntegrityCharacterization only.
    Coating Lubricity/DurabilityMet the acceptance criteria for Coating Lubricity/Durability.
    ParticulateMet the acceptance criteria for Particulate.
    Kink ResistanceMet the acceptance criteria for Kink Resistance.
    Liquid LeakMet the acceptance criteria for Liquid Leak.
    Static/Dynamic BurstMet the acceptance criteria for Static/Dynamic Burst.
    Corrosion ResistanceMet the acceptance criteria for Corrosion Resistance.
    Torque to FailureCharacterization only.
    Riptide™ Aspiration TubingBench - MicrobialEthylene Oxide (EO) ResidualMet the acceptance criteria for EO Residual.
    Ethylene Chlorohydrin (ECH)Met the acceptance criteria for ECH.
    Bioburden RecoveryMet the acceptance criteria for Bioburden Recovery.
    BioburdenMet the acceptance criteria for Bioburden.
    Bacterial EndotoxinMet the acceptance criteria for Bacterial Endotoxin.
    Bench - PackagingAseptic PresentationMet the acceptance criteria for Aseptic Presentation.
    Seal WidthMet the acceptance criteria for Seal Width.
    Dye LeakMet the acceptance criteria for Dye Leak.
    Visual InspectionMet the acceptance criteria for Visual Inspection.
    Bubble LeakMet the acceptance criteria for Bubble Leak.
    LegibilityMet the acceptance criteria for Legibility.
    Foreign MaterialMet the acceptance criteria for Foreign Material.
    Seal StrengthMet the acceptance criteria for Seal Strength.
    Bench - PerformanceDimensional InspectionMet the acceptance criteria for Dimensional Inspection.
    Identification of Heat Shrink BandsMet the acceptance criteria for Identification of the Heat Shrink Bands.
    Identification of Flow SwitchMet the acceptance criteria for Identification of the Flow Switch.
    Joint Tensile StrengthMet the acceptance criteria for Joint Tensile Strength.
    Leak (Vacuum and Pressure Decay)Met the acceptance criteria for Leak.
    Degree of CollapseMet the acceptance criteria for Degree of Collapse.
    Compatibility (Arc™ Catheter & Collection Canister)Met the acceptance criteria for Compatibility to the Arc™ Catheter and Riptide™ Collection Canister.
    Riptide™ Aspiration PumpBench - PerformanceDegrees of TiltMet the acceptance criteria for Degrees of Tilt.
    DurabilityMet the acceptance criteria for Durability.
    Maximum VacuumMet the acceptance criteria for Maximum Vacuum.
    Flow CapacityMet the acceptance criteria for Flow Capacity.
    Electrical SafetyMet the acceptance criteria for Electrical Safety.
    Electromagnetic CompatibilityMet the acceptance criteria for Electromagnetic Compatibility.
    Riptide™ Collection Canister with Intermediate TubingBench - PackagingPackaging IntegrityMet the acceptance criteria for Packaging Integrity.
    Packaging LabelingMet the acceptance criteria for Packaging Labeling.
    Bench - PerformanceVolume CapacityMet the acceptance criteria for Volume Capacity.
    Volume MarkingMet the acceptance criteria for Volume Marking.
    Canister StrengthMet the acceptance criteria for Canister Strength.
    Overfill ProtectionMet the acceptance criteria for Overfill Protection.
    Dimensional InspectionMet the acceptance criteria for Dimensional Inspection.
    Degree of CollapseMet the acceptance criteria for Degree of Collapse.
    Port IdentificationMet the acceptance criteria for Port Identification.
    Compatibility (Riptide™ Aspiration Pump)Met the acceptance criteria for Compatibility.
    Riptide™ Aspiration SystemBench - System PerformanceSimulated Clot RetrievalMet the acceptance criteria for Simulated Clot Retrieval.
    Vacuum PressureMet the acceptance criteria for Vacuum Pressure.
    Flow RateMet the acceptance criteria for Flow Rate.
    Lumen CollapseThe Arc™ Catheter is resistant to Lumen Collapse.
    UsabilityMet the acceptance criteria for Usability.

    Study Information:

    The provided document describes non-clinical bench testing and non-clinical animal testing. It explicitly states "Not Applicable" for clinical performance data. Therefore, questions related to human reader studies, ground truth establishment for a test set of clinical images, and training set details are not directly addressed in this document.

    2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    • Test Set Sample Size:
      • Bench Testing: The document does not specify the exact sample sizes (number of units tested) for each individual bench test. The nature of these tests often involves a batch of devices or components.
      • Animal Testing: The document mentions "a porcine model" but does not specify the number of animals used for the animal testing.
    • Data Provenance:
      • Bench Testing: Conducted by the manufacturer, Micro Therapeutics, Inc. d/b/a ev3 Neurovascular, as part of their R&D and regulatory submission process. Location not specified, but the company is based in Irvine, California, USA.
      • Animal Testing: Conducted "in a porcine model." Location not specified.
      • Retrospective/Prospective: These are non-clinical studies (bench and animal), so the terms retrospective/prospective in the context of human data acquisition do not directly apply. They are inherently prospective in the sense that the tests were designed and executed to evaluate this specific device for its regulatory submission.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

    • Not Applicable. The studies described are non-clinical (bench and animal). There is no mention of human expert-established ground truth for a test set of clinical images or data. The "ground truth" for these tests comes from objective measurements against defined standards (e.g., ISO, ASTM, USP standards, or internal specifications) and observations in animal models.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    • Not Applicable. As no human expert evaluation of clinical data for ground truth establishment is described, adjudication methods are not relevant to this submission.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    • No. The document explicitly states "Performance Data - Clinical: Not Applicable." This means no human reader studies (MRMC or otherwise) were conducted or submitted as part of this 510(k). The device is not an AI-powered system that assists human readers.

    6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done

    • No. This device is a medical instrument (aspiration system for stroke), not an AI algorithm. Therefore, a standalone algorithm performance study is not applicable.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

    • Bench Testing: Ground truth is established by objective measurements against pre-defined engineering specifications, international standards (ISO, ASTM, USP), and validated test methods. For example, "The Arc™ Catheter met the acceptance criteria for EO Residual" means the measured residual levels were below the established safe limits per ISO 10993-7.
    • Animal Testing: Ground truth is established through direct observation and measurement in a controlled porcine model, in accordance with "21 CFR Part 58 for Good Laboratory Practice (GLP) for Non-Clinical Laboratory Studies". The goal was to evaluate "safety, efficacy, and usability" and compare it to the predicate device.

    8. The sample size for the training set

    • Not Applicable. No AI/machine learning component is described for this device, so there is no concept of a "training set" in the context of algorithm development.

    9. How the ground truth for the training set was established

    • Not Applicable. As there is no training set for an AI algorithm mentioned, this question is not relevant.
    Ask a Question

    Ask a specific question about this device

    K Number
    K171268

    Validate with FDA (Live)

    Date Cleared
    2017-05-31

    (30 days)

    Product Code
    Regulation Number
    870.1250
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    Why did this record match?
    Reference Devices :

    K161152, K150107

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Reverse™M Micro Catheter is intended for use in neuro, peripheral, and coronary vasculature for the infusion of diagnostic agents, such as contrast media, and therapeutic agents, such as occlusion coils.

    Device Description

    The Reverse™ Micro Catheter is a single lumen catheter designed to be introduced over a steerable guide wire to access small, tortuous vasculature. The semi-rigid proximal section transitions to a flexible distal tip to facilitate vessel navigation. The micro catheter incorporates a PTFE liner to facilitate movement or introduction of devices passed through its lumen. Dual radiopaque markers at the distal end facilitate fluoroscopic visualization. The outer surface of the catheter body also employs a hydrophilic coating to enhance lubricity and promote deliverability. The Reverse™ Micro Catheter incorporates a standard luer fitting to facilitate the attachment of ancillary devices. The catheter and included accessories (steam shaping mandrel and peel away introducer sheath) are provided sterile, nonpyrogenic, and are intended for single use only. To accommodate physician preference and anatomical variations, the Reverse™ Micro Catheter is available in two models, Reverse 021, with a 0.021" inner diameter (ID) and, Reverse 027, with a 0.027" ID.

    AI/ML Overview

    The document you provided describes a medical device, the Reverse™ Micro Catheter, and its non-clinical performance evaluation for substantial equivalence to a predicate device. It does not describe a study involving an AI model or a specific algorithm. Therefore, I cannot provide information about acceptance criteria, device performance, sample sizes, ground truth, expert involvement, or comparative effectiveness studies in the context of AI.

    The document focuses on demonstrating that the modified Reverse™ Micro Catheter is substantially equivalent to a previously cleared device (Reverse™ Medical Microcatheter, K130858) based on non-clinical performance data.

    Here's a breakdown of what the document does provide, structured to address the spirit of your request where possible, but re-contextualized for a medical device rather than an AI study:

    1. A table of acceptance criteria and the reported device performance

    The document states that "All testing passed the acceptance criteria." However, it does not explicitly list the quantitative acceptance criteria for each test. The performance results are reported as successful completion of tests without new safety or effectiveness issues.

    Test CategorySpecific Tests EvaluatedReported PerformanceAcceptance Criteria (Not explicitly quantified in document)
    Bench Testing- Surface inspection- Dimensional requirements- Lubricity/friction- Particulates- Coating characterization- Simulated use device compatibility- Simulated use navigation and delivery- Shelf-life verification (1 year accelerated aging)"All testing passed the acceptance criteria.""Comprehensive verification and validation activities were successfully completed; raising no new issues of safety or effectiveness."Implicitly, meeting specifications and predicate equivalence.
    Biological Safety- MEM Elution Cytotoxicity- Guinea Pig Maximization Sensitization- Intracutaneous Reactivity- Acute System Injection- Materials Mediated Rabbit Pyrogen- Hemolysis Direct Contact- Hemolysis Indirect Extract Method- Complement Activation- In vivo Thrombogenicity- Genotoxicity (Ames Assay, in vitro Mouse Lymphoma, in vivo Mouse Micronucleus)- USP Physicochemical"Testing demonstrated that the Reverse™ Micro Catheter finished device, accessories, and packaging materials have no residual risk of biological hazards; and are therefore considered biocompatible for their intended use."Compliance with ISO 10993-1 and related standards.
    Microbial Assessments- Bioburden- Endotoxin- Sterility assurance"Sterilization has been validated and is controlled... to provide a minimum SAL of 10-6."Minimum SAL of 10-6; compliance with ISO 11135-1.
    Packaging Validation- Visual Inspection- Bubble Leak- Seal Strength"To evaluate integrity of the packaging configuration." (Implied successful evaluation)Implicitly, maintaining package integrity after sterilization, environmental conditioning, and simulated shipping.

    2. Sample sized used for the test set and the data provenance

    The document does not specify sample sizes for individual bench tests, biological safety tests, or packaging validation. It refers broadly to "comprehensive verification and validation activities." The data provenance is internal testing by the manufacturer, Medtronic Neurovascular, for a new medical device. This is prospective testing of the device being submitted for clearance.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    This concept is not applicable to this type of device submission. There isn't "ground truth" established by experts in the context of an AI algorithm's performance. Instead, the "truth" is determined by established engineering specifications, biocompatibility standards, and functional performance requirements for a physical medical device.

    4. Adjudication method for the test set

    Not applicable. There's no human-in-the-loop performance or diagnostic adjudication involved for this physical device's non-clinical testing.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    Not applicable, as this is not an AI device.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    Not applicable, as this is not an AI device. The testing described is "standalone" in the sense that it's the device itself being tested, not an algorithm.

    7. The type of ground truth used

    For this medical device, the "ground truth" for the non-clinical tests would be:

    • Engineering Specifications: Conformance to pre-defined dimensional, material, and performance requirements (e.g., inner diameter, lubricity, navigation).
    • International Standards: Adherence to established standards for biocompatibility (ISO 10993-1), sterilization (ISO 11135-1), and packaging integrity.
    • Predicate Device Performance: Demonstrating that the subject device performs "at least as safe and effective as the predicate device."

    8. The sample size for the training set

    Not applicable, as this is not an AI device.

    9. How the ground truth for the training set was established

    Not applicable, as this is not an AI device.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1