Search Filters

Search Results

Found 2 results

510(k) Data Aggregation

    K Number
    K243418
    Date Cleared
    2024-12-02

    (28 days)

    Product Code
    Regulation Number
    878.4780
    Reference & Predicate Devices
    Why did this record match?
    Reference Devices :

    K172448

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Riptide™ Aspiration Pump is intended as a vacuum source for Medtronic aspiration devices or systems for use in hospitals or clinics. Refer to the Instructions for Use included with the accessory aspiration device for indications related to the procedure for which this device will be used.

    Device Description

    The Riptide™ Aspiration Pump is an externally powered electromechanical device capable of generating a vacuum designed for drawing fluids and thrombus into the Riptide™ Collection Canister during interventional procedures. It is intended for general suction use in hospitals or clinics and is not intended for transport or field applications. The Riptide™ Aspiration Pump includes a receptacle that holds the Riptide™ Collection Canister with Intermediate Tubing in place. The intermediate tubing is connected to the vacuum inlet port.

    The Riptide™ Aspiration Pump package includes the following:

    • Riptide™ Aspiration Pump (LMT-RAP)
    • Region specific power cord for connection to earthed receptacle.
    • Riptide™ Aspiration Pump User's Manual

    The following item is required for use with Riptide™ Aspiration Pump (packaged separately):

    • Riptide™ Collection Canister with Intermediate Tubing (LMT-RCT)

    The Riptide™ Collection Canister is provided non-sterile and is pre-assembled with the Intermediate Tubing. The Riptide™ Collection Canister with Intermediate Tubing is single use. The Riptide™ Collection Canister is the repository for aspirated material. The Riptide™ Canister is placed into the receptacle of the Riptide™ Aspiration Pump as shown in Figure 3. The Intermediate Tubing is then connected to the vacuum inlet port.

    AI/ML Overview

    Here's an analysis of the Riptide™ Aspiration Pump's acceptance criteria and the study proving it, based on the provided document:

    This document is a 510(k) premarket notification summary, which focuses on demonstrating substantial equivalence to a predicate device rather than comprehensive de novo clinical trial data. Therefore, the device's acceptance criteria are framed in terms of meeting established standards and matching the performance of a legally marketed predicate device.


    1. Table of Acceptance Criteria and Reported Device Performance

    The acceptance criteria for the Riptide™ Aspiration Pump are primarily based on compliance with international standards for medical electrical equipment and medical suction equipment, as well as demonstrating equivalent performance to the predicate device.

    Acceptance Criteria (Standard / Characteristic)Reported Device Performance
    Electrical Safety (IEC 60601-1, Edition 3.2 2020-08)Pass
    Electromagnetic Compatibility (IEC 60601-1-2, Edition 4.1 2020-09)Pass
    User Manual Design Validation (FDA Guidance & ISO 14971, ANSI/AAMI HE75)Pass (evaluated per FDA guidance "Applying Human Factors and Usability Engineering to Medical Devices" and recognized consensus standards ISO 14971 and ANSI/AAMI HE75)
    Medical Suction Equipment (ISO 10079-1, Fourth Edition 2022-03)Pass
    Medical Suction Equipment (ISO 10079-4, First Edition 2021-08)Pass
    Vacuum Range0-29 inHg (Same as predicate device Penumbra Pump MAX™)
    Flow Rate (60 Hz US)0-0.8 SCFM (0-23 LPM) (Same as predicate device Penumbra Pump MAX™)
    Voltage110-115 Vac (Comparable to predicate device 100-115 Vac)
    Frequency60 Hz (US) (Same as predicate device Penumbra Pump MAX™)
    Duty CycleNon-continuous: 97% (58.2 minutes on, 1.8 minutes off) (Compared to predicate device 97.8% (45 minutes on, 1 minute off) - changed to match reference device Riptide™ Aspiration System)
    Applied Part ClassificationType CF (Same as predicate device Penumbra Pump MAX™; change from Reference Device Type BF to match predicate)
    Noise
    Ask a Question

    Ask a specific question about this device

    K Number
    K180705
    Date Cleared
    2018-07-04

    (107 days)

    Product Code
    Regulation Number
    870.1250
    Reference & Predicate Devices
    Why did this record match?
    Reference Devices :

    K172448, K150107

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Riptide™ Aspiration System is intended for use in the revascularization of patients with acute iscoke secondary to intracranial large vessel occlusive disease (within the internal carotid, middle cerebral - M1 and M2 segments, basilar, and vertebral arteries) within 8 hours of symptom onset. Patients who are ineligible for intravenous tissue plasminogen activator (IV t-PA) or who fail IV t-PA therapy are candidates for treatment.

    Device Description

    The Riptide™ Aspiration System React™ 68 Catheter is a single lumen, flexible, variable stiffness composite catheter with a nitinol structure that is jacketed with a durable polymer outer layer. A lubricous, polytetrafluoroethylene liner is used to create a structure that has both proximal stiffness and distal flexibility, and an encapsulated radiopaque distal platinum-iridium markerband which is used for visualization under fluoroscopy. The Riptide™ Aspiration System React™ 68 Catheter is introduced into the vasculature through the split y-introducer sheath.

    The proximal end of the Riptide™ Aspiration System React™ 68 Catheter is designed with a thermoplastic elastomer strain relief and a clear hub. The Riptide™ Aspiration System React™ 68 Catheter is designed with a hydrophilic coating. The Riptide™ Aspiration System React™ 68 Catheter is navigated to the intended treatment site and positioned proximal to the site of occlusion.

    AI/ML Overview

    The document describes the acceptance criteria and supporting studies for the Riptide Aspiration System (React 68 Catheter). However, it's important to note that this document is a 510(k) summary, which focuses on demonstrating substantial equivalence to a predicate device rather than providing a detailed clinical study report with specific performance metrics against pre-defined acceptance criteria for a new AI/software device. The device in question is a medical device (catheter) for stroke revascularization, not an AI/software device that typically has performance metrics like sensitivity, specificity, or F1-score.

    The acceptance criteria provided in the document are primarily for bench testing and biocompatibility, and the reported "performance" is whether the device "met the acceptance criteria" or specific qualitative observations, rather than a quantitative measurement against a numerical threshold.

    Here's the information extracted from the provided text, structured to best fit your request, with an understanding that the device is a physical catheter and not an AI/software product in the context of typical AI acceptance criteria.


    1. Table of Acceptance Criteria and Reported Device Performance

    For the Riptide Aspiration System (React 68 Catheter), acceptance criteria are primarily related to bench testing, biocompatibility, and animal testing. The reported "performance" is whether the device met these criteria.

    Test DescriptionAcceptance Criteria (Implied / Explicit)Reported Device Performance
    Biocompatibility
    Cytotoxicity (Elution Method)No evidence of causing cell lysis or toxicity (grade of zero reactivity)Test article extract showed no evidence of causing cell lysis or toxicity and had a grade of zero (no reactivity). Considered non-cytotoxic.
    Sensitization (Guinea Pig Max. Test)Grade less than two (mild reactivity)Test article extract met requirements (grade less than two). Does not elicit a sensitization response.
    Irritation (Intracutaneous Reactivity)Difference between test article and control extract mean scores = 0.0Difference between test article extract overall mean score and corresponding control extract mean score was 0.0. Considered a non-irritant.
    Acute Systemic Toxicity (Systemic Toxicity)No mortality or evidence of systemic toxicity; clinically normal animalsNo mortality or evidence of systemic toxicity. All animals clinically normal. Does not indicate signs of toxicity.
    Hemocompatibility (Hemolysis)Hemolytic index for direct contact bloodMet the acceptance criteria for bacterial endotoxin.
    Visual InspectionMet acceptance criteria for visual inspectionMet the acceptance criteria for visual inspection.
    Dimensional MeasurementsMet acceptance criteria for dimensional measurementsMet the acceptance criteria for dimensional measurements.
    Tip BucklingMet acceptance criteria for maximum compressive forceMet the acceptance criteria for tip buckling.
    Kink ResistanceMet acceptance criteria for maximum kink diameterMet the acceptance criteria for kink resistance.
    ParticulateMet acceptance criteria per USPMet the acceptance criteria for particulate evaluation.
    Coating LubricityMet acceptance criteria for average friction forcesMet the acceptance criteria for coating lubricity.
    Tensile StrengthMet acceptance criteria per ISO 10555-1. Annex BMet the acceptance criteria for tensile strength at the hub and shaft.
    Liquid LeakMet acceptance criteria per ISO 10555-1. Annex CMet the acceptance criteria for liquid leak.
    Corrosion ResistanceMet acceptance criteria per ISO 10555-1. Annex AMet the acceptance criteria for corrosion resistance.
    Hub Aspiration ResistanceMet acceptance criteria per ISO 10555-1. Annex DMet the acceptance criteria for hub air aspiration.
    RadiopacityMet acceptance criteria for markerband length, wall thickness, and fluoroscopy confirmationMet the acceptance criteria for radiopacity.
    Luer StandardsMet acceptance criteria per ISO 594-1 and ISO 80369-7Met the acceptance criteria for luer standards.
    Torque StrengthAble to withstand torsional forces typical of clinical useWas able to withstand torsional forces that are typical of clinical use.
    Dynamic PressureAble to withstand pressures typical of clinical useWas able to withstand pressures that are typical of clinical use.
    Coating IntegrityCoating coverage and lubricity maintainedRemained coated and lubricous.
    RecanalizationMet acceptance criteria for recanalizationMet the acceptance criteria for recanalization.
    Vacuum PressureMet acceptance criteria for vacuum pressureMet the acceptance criteria for vacuum pressure.
    Suction Flow RateMet acceptance criteria for suction flow rateMet the acceptance criteria for suction flow rate.
    UsabilityMet acceptance criteria for maneuverability, flexibility, aspiration, and clot retrievalMet the acceptance criteria for usability.
    Animal Testing
    Safety, Efficacy, UsabilityDemonstrated comparable safety, efficacy, and usability to predicate deviceEvaluated in a porcine model at acute and chronic time points, in presence/absence of simulated clot. This testing supports substantial equivalence.

    2. Sample Sizes and Data Provenance (Not applicable for AI/software test set, but for testing performed)

    • Biocompatibility and Bench Testing: The sample sizes are not explicitly stated for individual tests but are implied to be sufficient for standard regulatory testing (e.g., "The test article extract showed no evidence...", "Not a single animal showed..."). These are laboratory tests on physical devices/extracts.
    • Animal Testing: Conducted in a porcine model. Data provenance is from non-clinical laboratory studies performed in accordance with 21 CFR Part 58 for Good Laboratory Practice (GLP). This is prospective animal testing.

    3. Number of Experts and Qualifications (Not applicable to this type of device submission for establishing ground truth)

    This information is typically relevant for AI/software devices where human experts establish ground truth for image or data interpretation. For a physical catheter, ground truth is established through physical measurements, chemical analysis, and physiological responses as per standardized test methods. The document does not mention experts for ground truth establishment in this context.

    4. Adjudication Method (Not applicable)

    Adjudication methods (like 2+1, 3+1) are used to resolve disagreements among human labelers/experts when establishing ground truth for AI/software performance evaluation. Not relevant for this physical device submission.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study (Not done/applicable)

    MRMC studies are typically performed for AI/software devices to assess the impact of AI assistance on human reader performance. This device is a physical catheter, not an AI/software system, so such a study was not performed. The submission is based on demonstrating substantial equivalence to a predicate device through bench and animal testing.

    6. Standalone Performance (Algorithm only without human-in-the-loop performance) (Not applicable)

    This concept applies to AI/software algorithms. The Riptide Aspiration System is a physical catheter intended for use by a clinician. Its "performance" is inherently linked to its physical properties and interaction within a biological system, not a standalone algorithmic output.

    7. Type of Ground Truth Used

    The "ground truth" for this device's evaluation is primarily established by:

    • Standardized Test Methods and Criteria: For biocompatibility, microbial, and performance bench tests, "ground truth" is defined by adherence to specific ISO standards (e.g., ISO 10993, ISO 10555, ISO 11737), USP standards, and ANSI/AAMI standards. These standards define the acceptable range or qualitative outcome.
    • Physiological Observations and Measurements: In animal studies, ground truth for safety, efficacy, and usability comes from direct observation of animal health, physiological responses, and procedural outcomes.

    8. Sample Size for the Training Set (Not applicable)

    This concept is for AI/machine learning models. The Riptide Aspiration System is a physical medical device. It does not have a "training set" in the context of AI. Design and manufacturing improvements would be based on engineering principles and iterative testing.

    9. How the Ground Truth for the Training Set was Established (Not applicable)

    Again, this is an AI/ML concept. The "ground truth" for the development of catheter designs comes from engineering specifications, material science, and prior knowledge of predicate devices and clinical needs, rather than a labeled training set.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1