K Number
K180705

Validate with FDA (Live)

Date Cleared
2018-07-04

(107 days)

Product Code
Regulation Number
870.1250
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Riptide™ Aspiration System is intended for use in the revascularization of patients with acute iscoke secondary to intracranial large vessel occlusive disease (within the internal carotid, middle cerebral - M1 and M2 segments, basilar, and vertebral arteries) within 8 hours of symptom onset. Patients who are ineligible for intravenous tissue plasminogen activator (IV t-PA) or who fail IV t-PA therapy are candidates for treatment.

Device Description

The Riptide™ Aspiration System React™ 68 Catheter is a single lumen, flexible, variable stiffness composite catheter with a nitinol structure that is jacketed with a durable polymer outer layer. A lubricous, polytetrafluoroethylene liner is used to create a structure that has both proximal stiffness and distal flexibility, and an encapsulated radiopaque distal platinum-iridium markerband which is used for visualization under fluoroscopy. The Riptide™ Aspiration System React™ 68 Catheter is introduced into the vasculature through the split y-introducer sheath.

The proximal end of the Riptide™ Aspiration System React™ 68 Catheter is designed with a thermoplastic elastomer strain relief and a clear hub. The Riptide™ Aspiration System React™ 68 Catheter is designed with a hydrophilic coating. The Riptide™ Aspiration System React™ 68 Catheter is navigated to the intended treatment site and positioned proximal to the site of occlusion.

AI/ML Overview

The document describes the acceptance criteria and supporting studies for the Riptide Aspiration System (React 68 Catheter). However, it's important to note that this document is a 510(k) summary, which focuses on demonstrating substantial equivalence to a predicate device rather than providing a detailed clinical study report with specific performance metrics against pre-defined acceptance criteria for a new AI/software device. The device in question is a medical device (catheter) for stroke revascularization, not an AI/software device that typically has performance metrics like sensitivity, specificity, or F1-score.

The acceptance criteria provided in the document are primarily for bench testing and biocompatibility, and the reported "performance" is whether the device "met the acceptance criteria" or specific qualitative observations, rather than a quantitative measurement against a numerical threshold.

Here's the information extracted from the provided text, structured to best fit your request, with an understanding that the device is a physical catheter and not an AI/software product in the context of typical AI acceptance criteria.


1. Table of Acceptance Criteria and Reported Device Performance

For the Riptide Aspiration System (React 68 Catheter), acceptance criteria are primarily related to bench testing, biocompatibility, and animal testing. The reported "performance" is whether the device met these criteria.

Test DescriptionAcceptance Criteria (Implied / Explicit)Reported Device Performance
Biocompatibility
Cytotoxicity (Elution Method)No evidence of causing cell lysis or toxicity (grade of zero reactivity)Test article extract showed no evidence of causing cell lysis or toxicity and had a grade of zero (no reactivity). Considered non-cytotoxic.
Sensitization (Guinea Pig Max. Test)Grade less than two (mild reactivity)Test article extract met requirements (grade less than two). Does not elicit a sensitization response.
Irritation (Intracutaneous Reactivity)Difference between test article and control extract mean scores = 0.0Difference between test article extract overall mean score and corresponding control extract mean score was 0.0. Considered a non-irritant.
Acute Systemic Toxicity (Systemic Toxicity)No mortality or evidence of systemic toxicity; clinically normal animalsNo mortality or evidence of systemic toxicity. All animals clinically normal. Does not indicate signs of toxicity.
Hemocompatibility (Hemolysis)Hemolytic index for direct contact blood < 5% (typically); no hemolysisHemolytic index for direct contact with blood was 0.9%, for extract was 0.0%. Considered non-hemolytic.
Hemocompatibility (Complement Activation)Not statistically higher concentration of SC5b-9 than controlsConcentration of SC5b-9 in test article sample was not statistically higher than controls. Not considered potential activators of complement system.
Hemocompatibility (Thrombogenicity)Minimal thrombus formationDetermined to have minimal thrombus formation after four (4) hours. Demonstrates minimal thrombus formation.
Pyrogenicity (Material Mediated)No animal showed a temperature rise of 0.5°C or more above baselineNot a single animal showed a temperature rise of 0.5°C or more. Total rise was 0.0°C. Considered non-pyrogenic.
Bench Performance Data
Ethylene Oxide ResidualMet acceptance criteria per ISO 10993-7Met the acceptance criteria for ethylene oxide residual.
Ethylene Chlorohydrin ResidualMet acceptance criteria per ISO 10993-7Met the acceptance criteria for ethylene chlorohydrin residual.
BioburdenMet acceptance criteria per ISO 11737-1Met the acceptance criteria for bioburden.
Bacterial EndotoxinMet acceptance criteria per ANSI/AAMI ST72 and USP <161>Met the acceptance criteria for bacterial endotoxin.
Visual InspectionMet acceptance criteria for visual inspectionMet the acceptance criteria for visual inspection.
Dimensional MeasurementsMet acceptance criteria for dimensional measurementsMet the acceptance criteria for dimensional measurements.
Tip BucklingMet acceptance criteria for maximum compressive forceMet the acceptance criteria for tip buckling.
Kink ResistanceMet acceptance criteria for maximum kink diameterMet the acceptance criteria for kink resistance.
ParticulateMet acceptance criteria per USP <788>Met the acceptance criteria for particulate evaluation.
Coating LubricityMet acceptance criteria for average friction forcesMet the acceptance criteria for coating lubricity.
Tensile StrengthMet acceptance criteria per ISO 10555-1. Annex BMet the acceptance criteria for tensile strength at the hub and shaft.
Liquid LeakMet acceptance criteria per ISO 10555-1. Annex CMet the acceptance criteria for liquid leak.
Corrosion ResistanceMet acceptance criteria per ISO 10555-1. Annex AMet the acceptance criteria for corrosion resistance.
Hub Aspiration ResistanceMet acceptance criteria per ISO 10555-1. Annex DMet the acceptance criteria for hub air aspiration.
RadiopacityMet acceptance criteria for markerband length, wall thickness, and fluoroscopy confirmationMet the acceptance criteria for radiopacity.
Luer StandardsMet acceptance criteria per ISO 594-1 and ISO 80369-7Met the acceptance criteria for luer standards.
Torque StrengthAble to withstand torsional forces typical of clinical useWas able to withstand torsional forces that are typical of clinical use.
Dynamic PressureAble to withstand pressures typical of clinical useWas able to withstand pressures that are typical of clinical use.
Coating IntegrityCoating coverage and lubricity maintainedRemained coated and lubricous.
RecanalizationMet acceptance criteria for recanalizationMet the acceptance criteria for recanalization.
Vacuum PressureMet acceptance criteria for vacuum pressureMet the acceptance criteria for vacuum pressure.
Suction Flow RateMet acceptance criteria for suction flow rateMet the acceptance criteria for suction flow rate.
UsabilityMet acceptance criteria for maneuverability, flexibility, aspiration, and clot retrievalMet the acceptance criteria for usability.
Animal Testing
Safety, Efficacy, UsabilityDemonstrated comparable safety, efficacy, and usability to predicate deviceEvaluated in a porcine model at acute and chronic time points, in presence/absence of simulated clot. This testing supports substantial equivalence.

2. Sample Sizes and Data Provenance (Not applicable for AI/software test set, but for testing performed)

  • Biocompatibility and Bench Testing: The sample sizes are not explicitly stated for individual tests but are implied to be sufficient for standard regulatory testing (e.g., "The test article extract showed no evidence...", "Not a single animal showed..."). These are laboratory tests on physical devices/extracts.
  • Animal Testing: Conducted in a porcine model. Data provenance is from non-clinical laboratory studies performed in accordance with 21 CFR Part 58 for Good Laboratory Practice (GLP). This is prospective animal testing.

3. Number of Experts and Qualifications (Not applicable to this type of device submission for establishing ground truth)

This information is typically relevant for AI/software devices where human experts establish ground truth for image or data interpretation. For a physical catheter, ground truth is established through physical measurements, chemical analysis, and physiological responses as per standardized test methods. The document does not mention experts for ground truth establishment in this context.

4. Adjudication Method (Not applicable)

Adjudication methods (like 2+1, 3+1) are used to resolve disagreements among human labelers/experts when establishing ground truth for AI/software performance evaluation. Not relevant for this physical device submission.

5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study (Not done/applicable)

MRMC studies are typically performed for AI/software devices to assess the impact of AI assistance on human reader performance. This device is a physical catheter, not an AI/software system, so such a study was not performed. The submission is based on demonstrating substantial equivalence to a predicate device through bench and animal testing.

6. Standalone Performance (Algorithm only without human-in-the-loop performance) (Not applicable)

This concept applies to AI/software algorithms. The Riptide Aspiration System is a physical catheter intended for use by a clinician. Its "performance" is inherently linked to its physical properties and interaction within a biological system, not a standalone algorithmic output.

7. Type of Ground Truth Used

The "ground truth" for this device's evaluation is primarily established by:

  • Standardized Test Methods and Criteria: For biocompatibility, microbial, and performance bench tests, "ground truth" is defined by adherence to specific ISO standards (e.g., ISO 10993, ISO 10555, ISO 11737), USP standards, and ANSI/AAMI standards. These standards define the acceptable range or qualitative outcome.
  • Physiological Observations and Measurements: In animal studies, ground truth for safety, efficacy, and usability comes from direct observation of animal health, physiological responses, and procedural outcomes.

8. Sample Size for the Training Set (Not applicable)

This concept is for AI/machine learning models. The Riptide Aspiration System is a physical medical device. It does not have a "training set" in the context of AI. Design and manufacturing improvements would be based on engineering principles and iterative testing.

9. How the Ground Truth for the Training Set was Established (Not applicable)

Again, this is an AI/ML concept. The "ground truth" for the development of catheter designs comes from engineering specifications, material science, and prior knowledge of predicate devices and clinical needs, rather than a labeled training set.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which consists of a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

July 4, 2018

Micro Therapeutics, Inc. d/b/a ev3 Neurovascular Ryan Kenney Senior Regulatory Affairs Specialist 9775 Toledo Wav Irvine, California 92618

Re: K180705

Trade/Device Name: Riptide Aspiration System (React 68 Catheter) Regulation Number: 21 CFR 870.1250 Regulation Name: Percutaneous Catheter Regulatory Class: Class II Product Code: NRY Dated: June 4, 2018 Received: June 6, 2018

Dear Ryan Kenney:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820);

{1}------------------------------------------------

and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Xiaolin Zheng -S

for Carlos L. Peña, PhD, MS Director Division of Neurological and Physical Medicine Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K180705

Device Name Riptide Aspiration System (React 68 Catheter)

Indications for Use (Describe)

The Riptide™ Aspiration System is intended for use in the revascularization of patients with acute iscoke secondary to intracranial large vessel occlusive disease (within the internal carotid, middle cerebral - M1 and M2 segments, basilar, and vertebral arteries) within 8 hours of symptom onset. Patients who are ineligible for intravenous tissue plasminogen activator (IV t-PA) or who fail IV t-PA therapy are candidates for treatment.

Type of Use (Select one or both, as applicable)
☑ Prescription Use (Part 21 CFR 801 Subpart D)☐ Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

510(k) Summary: K180705

510(k) Owner:Micro Therapeutics, Inc. d/b/a ev3 Neurovascular
9775 Toledo Way
Irvine, CA 92618
Establishment Registration No.: 2029214
Contact Person:Ryan Kenney
Senior Regulatory Affairs Specialist
Telephone: (949) 297-5489
Email: ryan.j.kenney@medtronic.com
Date SummaryPrepared:June 25, 2018
Trade Name ofDevice:RiptideTM Aspiration System (React 68 Catheter)
Common Name ofDevice:Catheter, Thrombus Retriever
Review Panel:Neurology
Product Code:NRY
Regulation Number:21 CFR 870.1250
Regulation Name:Percutaneous Catheter
Device Classification:Class II
Predicate Device(s):Penumbra System® ACE 68 Reperfusion Catheter510(k)#: K161064
Reference Device(s):RiptideTM Aspiration System510(k)#: K172448ArcTM Intracranial Support Catheter510(k)#: K150107

Device Description:

The Riptide™ Aspiration System React™ 68 Catheter is a single lumen, flexible, variable stiffness composite catheter with a nitinol structure that is jacketed with a durable polymer outer layer. A lubricous, polytetrafluoroethylene liner is used to create a structure that has both proximal stiffness and distal flexibility, and an encapsulated radiopaque distal platinum-iridium markerband which is used for visualization under fluoroscopy. The Riptide™ Aspiration System React™ 68 Catheter is introduced into the vasculature through the split y-introducer sheath.

The proximal end of the Riptide™ Aspiration System React™ 68 Catheter is designed with a thermoplastic elastomer strain relief and a clear hub. The Riptide™ Aspiration System React™ 68 Catheter is designed with a hydrophilic coating. The Riptide™ Aspiration System React™ 68 Catheter is navigated to the intended treatment site and positioned proximal to the site of occlusion.

{4}------------------------------------------------

Indications for Use:

The Riptide™ Aspiration System is intended for use in the revascularization of patients with acute ischemic stroke secondary to intracranial large vessel occlusive disease (within the internal carotid, middle cerebral – M1 and M2 segments, basilar, and vertebral arteries) within 8 hours of symptom onset. Patients who are ineligible for intravenous tissue plasminogen activator (IV t-PA) or who fail IV t-PA therapy are candidates for treatment.

Penumbra System® ACE 68Reperfusion CatheterReact™ 68 Catheter
Indication for Use (IFU)StatementThe Penumbra System® isintended for use in therevascularization of patients withacute ischemic stroke secondaryto intracranial large vesselocclusive disease (within theinternal carotid, middle cerebral –M1 and M2 segments, basilar,and vertebral arteries) within 8hours of symptom onset.The Riptide™ Aspiration Systemis intended for use in therevascularization of patients withacute ischemic stroke secondaryto intracranial large vesselocclusive disease (within theinternal carotid, middle cerebral –M1 and M2 segments, basilar,and vertebral arteries) within 8hours of symptom onset. Patientswho are ineligible for intravenoustissue plasminogen activator (IVt-PA) or who fail IV t-PA therapyare candidates for treatment.
Materials
Proximal HubGrilamid™Trogamid®
Strain Relief(Hub Sleeve)Grilamid™DynaFlex®
Strain Relief304 Stainless SteelN/A
Inner Diameter BandPolyolefin, PET Yellow(Black Ink)N/A
LinerPTFESame
Braid ReinforcementN/ANitinol
Coil Reinforcement304V Stainless SteelNickel (55%)/Titanium (45%)Nitinol
Proximal ExtrusionsVestamid®Pebax® 72DGrilamid™Pebax® 45DPebax® 55DPebax® 63DPebax® 72D
Distal ExtrusionsPebax® 35DPebax® 40DPebax® 55DPebax® 63DPellathane® 80ATecoflex® 80APebax® 25DPebax® 35D
Penumbra System® ACE 68Reperfusion CatheterReact™ 68 Catheter
Extrusion ColorantsClear/Natural or PurpleClear/Natural or Green
MarkerbandPlatinum (90%)/Iridium (10%)Same
CoatingSRDX HarmonySurmodics Serene®
Working Length132 cmSame
Proximal Inner Diameter(ID)0.068" (Min.)0.068" (Min.)
Proximal Outer Diameter(OD)0.084" (Max.)0.083" (Max.)
Distal Inner Diameter (ID)0.068" (Min.)0.068" (Min.)
Distal Outer Diameter(OD)0.084" (Max.)0.083" (Max.)
Coating Length30cm40cm
Tip ShapeStraightSame
Accessories
Peelable SheathPTFESame
Rotating Hemostasis ValvePolycarbonate Silicone O-RingN/A
Shaping Mandrel0.038" OD Stainless SteelN/A
Packaging
CartonSBS PaperboardSame
Packaging CardPolyethyleneSame
Packaging HoopPolyethyleneSame
Packaging TrayPolyethyleneTerephthalatePolystreneN/A
PouchPolyesterPolyethyleneTyvek®NylonTyvek®
Other
Shelf Life8 Months3 Months
SterilizationEthylene Oxide (EO)Same
UseSingle Use, DisposableSame

Device Comparison:

{5}------------------------------------------------

Biocompatibility:

Biocompatibility was conducted for the React™ 68 Catheter. The React™ 68 Catheter is categorized as a limited exposure (< 24 hours), external communicating device contacting circulating blood. The following biocompatibility was conducted for the React™ 68 Catheter:

Test DescriptionResultsConclusions
Cytotoxicity(Elution Method)The test article extract showedno evidence of causing celllysis or toxicity and had agrade of zero (no reactivity).The React TM 68 Catheter isconsidered non-cytotoxic.
Test DescriptionResultsConclusions
Sensitization(Guinea Pig Maximization Test)The test article extract met therequirements of the test sincethe grade was less than a gradetwo (mild reactivity).The ReactTM 68 Catheter doesnot elicit a sensitizationresponse.
Irritation(Intracutaneous Reactivity)The difference between eachtest article extract overallmean score and correspondingcontrol extract overall meanscore was 0.0 and 0.0 for thesodium chloride and sesameoil test article extracts,respectively.The ReactTM 68 Catheter isconsidered a non-irritant.
Acute Systemic Toxicity(Systemic Toxicity)There was no mortality orevidence of systemic toxicityfrom the extracts injected. Allanimals were clinically normalthroughout the study.The ReactTM 68 Catheter doesnot indicate signs of toxicity.
Hemocompatibility(Hemolysis)The hemolytic index for thetest article in direct contactwith blood was 0.9%, and thehemolytic index for the testarticle extract was 0.0%.The ReactTM 68 Catheter isconsidered non-hemolytic.
Hemocompatibility(Complement Activation)The concentration of SC5b-9in the sponsor providedcontrol sample was notstatistically higher than theactivated normal human serumcontrol or the negative control.The concentration of SC5b-9in the test article sample wasnot statistically higher than theactivated normal human serumcontrol, the negative control orthe sponsor provided control.The control and test articlesamples are not considered tobe potential activators of thecomplement system.
Hemocompatibility(Thrombogenicity)The thrombogenic potential ofthe control article wasevaluated in comparison to thetest article. Both the controland test articles wereThe ReactTM 68 Catheterdemonstrates minimalthrombus formation.
Test DescriptionResultsConclusions
determined to have minimalthrombus formation after four(4) hours (±30 minutes)without anticoagulation.
Pyrogenicity(Material Mediated)Not a single animal showed atemperature rise of 0.5°C ormore above its baselinetemperature. The total rise ofthe temperature during three(3) hours was 0.0°C.The React™ 68 Catheter isconsidered non-pyrogenic.

{6}------------------------------------------------

{7}------------------------------------------------

The React™ 68 Catheter has been evaluated to meet requirements specified in ISO 10993-1.

Performance Data – Bench:

Non-clinical bench testing was conducted to evaluate the performance of the React™ 68 Catheter. The following non-clinical bench testing was conducted for the React™ 68 Catheter:

TestTest Method SummaryResults
Microbial
Ethylene Oxide ResidualThe ReactTM 68 Catheter wasevaluated per ISO 10993-7.The ReactTM 68 Catheter metthe acceptance criteria forethylene oxide residual.
Ethylene ChlorohydrinResidualThe ReactTM 68 Catheter wasevaluated per ISO 10993-7.The ReactTM 68 Catheter metthe acceptance criteria forethylene chlorohydrin residual.
BioburdenThe ReactTM 68 Catheter wasevaluated per ISO 11737-1.The ReactTM 68 Catheter metthe acceptance criteria forbioburden.
Bacterial EndotoxinThe ReactTM 68 Catheter wasevaluated per ANSI/AAMIST72 and USP <161>.The ReactTM 68 Catheter metthe acceptance criteria forbacterial endotoxin.
Performance
Visual InspectionThe ReactTM 68 Catheter wasinspected under x2.5magnification.The ReactTM 68 Catheter metthe acceptance criteria forvisual inspection.
Dimensional MeasurementsThe proximal ID, distal ID,proximal OD, distal OD, usablelength, total length, coatinglength, and distal tip length ofthe ReactTM 68 Catheter weremeasured.The ReactTM 68 Catheter metthe acceptance criteria fordimensional measurements.
Tip BucklingThe ReactTM 68 Catheter wasevaluated for the maximumcompressive force.The ReactTM 68 Catheter metthe acceptance criteria for tipbuckling
TestTest Method SummaryResults
Kink ResistanceThe ReactTM 68 Catheter was evaluated for the maximum kink diameter.The ReactTM 68 Catheter met the acceptance criteria for kink resistance.
ParticulateThe ReactTM 68 Catheter was evaluated per USP <788>.The ReactTM 68 Catheter met the acceptance criteria for particulate evaluation.
Coating LubricityThe ReactTM 68 Catheter was evaluated for the average friction forces.The ReactTM 68 Catheter met the acceptance criteria for coating lubricity.
Tensile StrengthThe ReactTM 68 Catheter was evaluated per ISO 10555-1. Annex B.The ReactTM 68 Catheter met the acceptance criteria for tensile strength at the hub and shaft.
Liquid LeakThe ReactTM 68 Catheter was evaluated per ISO 10555-1. Annex C.The ReactTM 68 Catheter met the acceptance criteria for liquid leak.
Corrosion ResistanceThe ReactTM 68 Catheter was evaluated per ISO 10555-1. Annex A.The ReactTM 68 Catheter met the acceptance criteria for corrosion resistance.
Hub Aspiration ResistanceThe ReactTM 68 Catheter was evaluated per ISO 10555-1. Annex D.The ReactTM 68 Catheter met the acceptance criteria for hub air aspiration.
RadiopacityThe markerband length and wall thickness of the ReactTM 68 Catheter were measured. In addition, radiopacity was confirmed via fluoroscopy.The ReactTM 68 Catheter met the acceptance criteria for radiopacity.
Luer StandardsThe ReactTM 68 Catheter was evaluated per ISO 594-1 and ISO 80369-7.The ReactTM 68 Catheter met the acceptance criteria for luer standards.
Torque StrengthThe ReactTM 68 Catheter was evaluated for transmission of proximal torque to the distal tip.The ReactTM 68 Catheter was able to withstand torsional forces that are typical of clinical use.
Dynamic PressureThe ReactTM 68 Catheter was evaluated for the amount of pressure it can withstand.The ReactTM 68 Catheter was able to withstand pressures that are typical of clinical use.
Coating IntegrityThe ReactTM 68 Catheter was evaluated for coating coverage and lubricity.The ReactTM 68 Catheter remained coated and lubricous.
TestTest Method SummaryResults
RecanalizationThe Riptide™ AspirationSystem React™ 68 Catheterwas evaluated forrecanalization.The Riptide™ AspirationSystem React™ 68 Cathetermet the acceptance criteria forrecanalization.
Vacuum PressureThe Riptide™ AspirationSystem React™ 68 Catheterwas evaluated for vacuumpressure.The Riptide™ AspirationSystem React™ 68 Cathetermet the acceptance criteria forvacuum pressure.
Suction Flow RateThe Riptide™ AspirationSystem React™ 68 Catheterwas evaluated for suction flowrate.The Riptide™ AspirationSystem React™ 68 Cathetermet the acceptance criteria forsuction flow rate.
UsabilityThe Riptide™ AspirationSystem React™ 68 Catheterand the Penumbra System®ACE 68 Reperfusion Catheterwere evaluated formaneuverability, flexibility, theability to engage and applyaspiration, and the ability tosafely retrieve clot.The Riptide™ AspirationSystem React™ 68 Cathetermet the acceptance criteria forusability.

{8}------------------------------------------------

Note: Packaging for the React™ 68 Catheter was leveraged from the Riptide™ Aspiration Tubing.

In addition, the following non-clinical bench testing was conducted to evaluate the performance of the Riptide™ Aspiration System React™ 68 Catheter:

{9}------------------------------------------------

Performance Data - Animal:

Non-clinical animal testing was conducted to evaluate the safety, efficacy, and usability of the Riptide™ Aspiration System React™ 68 Catheter in comparison to the Penumbra System® ACE 68 Reperfusion Catheter at acute and chronic time points in a porcine model, both in the presence and absence of simulated clot. Non-clinical animal testing was conducted in accordance with 21 CFR Part 58 for Good Laboratory Practice (GLP) for Non-Clinical Laboratory Studies.

Performance Data - Clinical:

Not Applicable. A determination of substantial equivalence is based upon successful completion of non-clinical bench and animal testing.

Conclusion:

The design modifications incorporated do not alter the fundamental scientific technology or intended use.

Non-clinical bench and animal testing supports a determination that the subject Riptide™ Aspiration System React™ 68 Catheter is substantially equivalent to the predicate Penumbra System® ACE 68 Reperfusion Catheter.

§ 870.1250 Percutaneous catheter.

(a)
Identification. A percutaneous catheter is a device that is introduced into a vein or artery through the skin using a dilator and a sheath (introducer) or guide wire.(b)
Classification. Class II (performance standards).