Search Filters

Search Results

Found 3 results

510(k) Data Aggregation

    K Number
    K211952

    Validate with FDA (Live)

    Date Cleared
    2022-07-21

    (393 days)

    Product Code
    Regulation Number
    872.3640
    Panel
    Dental
    Age Range
    All
    Predicate For
    N/A
    Why did this record match?
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The BTI Dental Implant System UnicCa® for oral implant surgery is to be used for the partial or total replacement of teeth in edentulate patients. Once attached to the bone, the implants act as an anchor for various fixed or removable prosthetic solutions that can be used to improve or restore a patient's mastication function.

    In the case of 5.5 - 6.5mm long UnicCa® implants should be used in a two-stage surgical procedure. These implants are indicated for delayed loading. These implants are indicated only for straight abutments and to support permanently fixed restorations.

    In the case of Tiny® 3.0 UnicCa® implants: These implants shall be used only to replace maxillary lateral incisors and mandibular lateral and central incisors. Immediate loading is recommended when there is good primary stability and an appropriate occlusal load.

    Device Description

    BTI Dental Implant System UnicCa® is a self-tapping, threaded, root form dental titanium implant provided in two types of connections: external (i.e., Externa®) and interna®), in a variety of platforms and range of diameters (3.0 – 6.0 mm) and lengths (5.5 – 18.0 mm). BTI Dental Implant System UnicCa® features an implant surface treatment that improves the hydrophilicity of the implant.

    The purpose of this 510(k) is to expand the product offering and include a new dental implant platform (Interna Narrow), and compatible abutments, together with additional implants for the already existing Universal Plus platform. Interna Narrow implants will be offered with implant diameters ranging from 3.3 to 4.75 mm and lengths of 5.5 to 15 mm. Additionally, new implants for the existing Universal Plus platform of diameter 6.0 mm, and lengths of 7.5 to 11.5 mm are included.

    AI/ML Overview

    This document describes the BTI Interna Narrow/Plus Dental Implant System UnicCa®, which is a dental implant system. The information provided is for a 510(k) premarket notification to the FDA, asserting substantial equivalence to previously cleared predicate devices.

    1. Table of Acceptance Criteria and Reported Device Performance:

    The document does not explicitly state "acceptance criteria" in the typical sense of numerical thresholds for outcomes. Instead, it demonstrates substantial equivalence to predicate devices through various non-clinical tests and comparisons. The "performance" is largely reported as demonstrating "substantial equivalence" or being "identical" or "similar" to the predicate devices.

    Feature / Test CategoryAcceptance Criteria (Implied by Predicate Equivalence)Reported Device Performance
    Product ClassificationIdentical product codes (DZE, NHA) and regulation number (21 CFR 872.3640), Class II.Identical.
    Indications for UseIdentical to primary predicate device (K151391) for general use in partial/total replacement of teeth, specific uses for 5.5-6.5mm implants (two-stage, delayed loading, straight abutments, fixed restorations), and Tiny® 3.0 implants (maxillary lateral, mandibular lateral/central incisors, immediate loading with good primary stability/appropriate occlusal load).Identical to primary predicate device. Includes specific indications for 5.5-6.5mm long UnicCa® implants and Tiny® 3.0 UnicCa® implants that match the predicate.
    Implant DesignSimilar designs, sterilization, and packaging as identified predicate devices (K022258, K053355, K061383) for use as dental implants.Similar designs, sterilization and packaging are included in subject and the identified predicate devices, intended for similar use.
    Platform Diameter (mm)Range of implant platform diameters covered by identified predicate devices (e.g., Interna: 4.1, 5.5; Externa: 3.5, 4.1, 5.5).Similar. A new platform diameter for the Interna connection (Interna Narrow, 3.5 mm) is included, but the range is covered by predicate devices. Interna Universal Plus is identical to predicate.
    Implant Length (mm)5.5 - 18 mm, matching predicate K151391.Identical (Interna Narrow: 5.5 – 15 mm; Interna Universal Plus: 7.5 – 11.5 mm, which falls within the predicate's 5.5-18 mm range).
    Implant Body Diameter (mm)3.0 - 6.0 mm, matching predicates K151391, K022258, K053355.Identical (Interna Narrow: 3.3 – 4.75 mm; Interna Universal Plus: 6.0 mm, which falls within the predicate's 3.0-6.0 mm range).
    Implant MaterialCommercially pure titanium (Grade 4).Identical.
    Implant Surface TreatmentCalcium surface treatment.Identical to primary predicate (K151391).
    RestorationSingle and multiple restorations.Identical.
    Connection DesignEngaging, non-engaging.Identical.
    Abutment Prosthetic Diameter (mm)Range of diameters covered by identified predicate devices (K022258, K053355, K061383), e.g., 3.5 - 6.5 mm.Similar. The subject device's range (3.5 - 5.1 mm) is covered.
    Abutment Gingival Height (mm)Range of gingival heights covered by identified predicate and reference devices (K022258, K053355, K061383, K130808), e.g., 1 - 4 mm.Similar. The subject device's range (0.5 - 4 mm) is covered.
    Abutment AngulationStraight (0°) with divergence allowance supported by bench testing.Identical. Divergence allowance in subject device is supported by bench testing provided.
    Abutment MaterialCommercially pure titanium (Grade 4).Identical to identified predicate devices.
    Prosthetic Screws MaterialTi6Al4V, or other materials in predicate (e.g., gold/palladium alloy in K022258).Identical to K053355 (Ti6Al4V).
    Packaging, Sterilization, Shelf-LifeSimilar packaging and sterilization processes, sterile by gamma irradiation or non-sterile for end-user sterilization.Similar packaging as the identified predicate devices. Healing abutments/Transepithelial abutments: sterile by gamma irradiation, thermoform tray with peel top lid. Temporary Titanium abutments, Titanium abutments, Healing Caps, Aesthetic Interfaces and Screws: non-sterile, to be sterilized by end user (moist heat), thermosealed bag.
    BiocompatibilityConformance with ISO 10993-1, with satisfactory results for all biological endpoints relative to a long-term implant.All biological endpoints relative to a long-term implant as identified in the FDA guidance have been addressed with satisfactory results.
    Corrosion, Surface Hydrophilicity, TOF-SIMSLeveraged from predicate K151391. Performance should be comparable.Leveraged from K151391.
    Fatigue TestingConformance with ISO 14801 and FDA guidance document, showing substantial equivalence to predicate for intended use.Test results show that the proposed device is substantially equivalent to the predicate device for its intended use.
    Comparative Area Analyses (Bone Resorption, Bone-to-Implant Contact)Subject implants and reference device should show substantial equivalence in these aspects.Results showed that subject device is substantially equivalent.
    Sterilization ValidationSterility assurance level (SAL) of 10-6, according to ISO 11137-1, ISO 11137-2 and ISO 17665-1.Validation performed to a SAL of 10-6, according to specified ISO standards.
    Packaging/Sterile Barrier/Shelf-Life ValidationConformance with ISO 11607-1, ISO 11607-2, ASTM F88, ASTM F1886, ASTM 1929 and ASTM F2096.Validation performed according to specified ISO and ASTM standards.
    Transport ValidationConformance with ISTA 3A.Validation performed according to ISTA 3A.
    Endotoxin TestingConformance with ANSI/AAMI ST72.Testing performed according to ANSI/AAMI ST72.

    2. Sample size used for the test set and the data provenance:

    • Sample Size for Test Set: The document mentions "Comparative area analyses for full external area and for bone resorption of up to 3 mm have been performed for subject implants and reference device." However, specific numerical sample sizes for this or other non-clinical tests (e.g., fatigue testing, biocompatibility) are not provided.
    • Data Provenance: The document does not specify the country of origin for the data or whether it is retrospective or prospective. It refers to manufacturing in Spain (B.T.I. Biotechnology Institute, SL Jose Rivero Qualified Person Leonardo Da Vinci 14, Parque Tecnologico de Alava Minano, Alava 01510 SPAIN). The tests are non-clinical, implying laboratory or bench testing, not human patient data.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

    This information is not applicable and not provided as the submission relies on non-clinical, bench testing for substantial equivalence, not clinical study data requiring expert-established ground truth.

    4. Adjudication method for the test set:

    This information is not applicable and not provided as the submission relies on non-clinical, bench testing.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, if so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    This is not applicable. The device is a physical dental implant system, not an AI-assisted diagnostic or therapeutic tool for which an MRMC study would be relevant.

    6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done:

    This is not applicable as the device is a physical dental implant system, not a software algorithm.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):

    The "ground truth" for this submission is based on established engineering and biological performance standards (e.g., ISO, ASTM standards, FDA guidance documents) and comparative data to already cleared predicate physical devices. For example:

    • Biocompatibility: Conformance with ISO 10993-1.
    • Fatigue Testing: Conformance with ISO 14801 and FDA guidance.
    • Sterilization Validation: Conformance with ISO 11137 series.
    • Physical measurements and material specifications are compared to those of the predicate devices.

    8. The sample size for the training set:

    This information is not applicable as there is no mention of a "training set" in the context of this physical device submission, which does not involve machine learning or AI models.

    9. How the ground truth for the training set was established:

    This information is not applicable for the reasons stated in point 8.

    Ask a Question

    Ask a specific question about this device

    K Number
    K170421

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2018-01-11

    (332 days)

    Product Code
    Regulation Number
    872.3630
    Panel
    Dental
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    Why did this record match?
    Reference Devices :

    K132219, K123627

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    ET Hybrid Abutment is a customized abutment intended for use with HIOSSEN ET dental implant in the edentulous or partially edentulous maxilla or mandible to provide support for prosthetic restorations such as crowns and bridges. All digitally designed copings for use with the ET Hybrid Abutment for CAD/CAM are intended to be sent to a HIOSSEN Inc. manufacturing facility for manufacture.

    Device Description

    ET Hybrid Abutment is intended to be placed onto Hiossen dental implants to provide support for customized prosthetic restorations such as crowns and bridges. The proposed devices are composed of ET Link Abutment and Coping. The ET Link Abutment is premanufactured abutment by Hiossen Inc. only and the coping would be manufactured and bonded into the ET Link Abutment by Hiossen only with design input using CAD software from both by dental laboratories and by Hiossen Inc. The final device of ET Hybrid Abutment is under controlled by Hiossen's Quality System and provided to the customer. ET Hybrid Abutment is indicated for screw-retained single tooth or cement-retained single tooth and bridge restorations. Titanium alloy is the raw material for the Link Abutment and Zirconia material is for the Coping. The ET Hybrid Abutment gives the flexibility to design customized prosthetics.

    AI/ML Overview

    This document is a 510(k) summary for the ET Hybrid Abutment by HIOSSEN Inc., intended for use with HIOSSEN ET dental implants to support prosthetic restorations. The submission relies on the substantial equivalence to a primary predicate device, ET Hybrid Abutment (K162390), and two reference predicates: Straumann Variobase Abutments (K132219) and ET SmartFit Abutment (K123627). No clinical studies were submitted. Therefore, the information provided below will be based on the non-clinical testing and comparison to predicate devices, rather than a clinical study establishing the device's meeting of acceptance criteria from human data.

    1. A table of acceptance criteria and the reported device performance

    The document does not explicitly state "acceptance criteria" in a quantitative format for the entire device. Instead, it relies on demonstrating substantial equivalence to predicate devices through material similarity, manufacturing processes, intended use, and non-clinical testing (biocompatibility and fatigue). The performance is demonstrated by conformance to relevant ISO standards for dental implants.

    Acceptance Criteria (Implied / Indirect)Reported Device Performance (Demonstrated vs. Predicate)
    Biocompatibility: Meet requirements for implantable devices.Demonstrated by relying on the predicate K162390. The subject device is manufactured using identical materials (Titanium Alloy Ti-6AL 4V, Zirconia Oxide) and manufacturing processes. Chemical composition, body contact (BONE / TISSUE, permanent > 30 days), and sterilization methods are the same as predicate devices. No additional biocompatibility testing was deemed necessary.
    Fatigue Strength: Withstand static and dynamic forces in oral environment.Successfully demonstrated through additional non-clinical testing: static and dynamic fatigue tests according to ISO 14801 in saline at 2Hz on the final finished sterilized abutment. The details of the performance (e.g., specific load cycles endured) are not provided in this summary but are implicit in the conclusion of substantial equivalence.
    Sterilization Efficacy: Ensure sterility after processing.Demonstrated by relying on the predicate device's sterilization validation. The subject device uses the same materials (Titanium & Zirconia, Cement) and the predicate's validation for steam sterilization was conducted according to ISO 17665-1, ISO/TS 17665-2 for gravity displacement on wrapped conditions. The worst-case sample (longest height, largest diameter) was used.
    Dimensional Specifications: Adhere to design limitations.The device specifications (Angulation, Gingival Diameter, Total Height, Contact Diameter, Wall Thickness, Gingival Height, Post Diameter, Post Height) are explicitly listed (e.g., Angulation 030°, Gingival Diameter 4.215 mm), indicating controlled design parameters. These are compared to the predicate, with the only difference noted as the coping part angle.
    Material Composition: Use specified materials.Uses Titanium Alloy Ti-6AL 4V and Zirconia Oxide, identical to the primary predicate device.

    2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    • Sample Size for Test Set: Specific sample sizes for the fatigue tests (ISO 14801) are not explicitly stated in this 510(k) summary. ISO standards typically specify minimum sample sizes for such tests.
    • Data Provenance: The origin of the testing data is not explicitly stated in terms of country, but it would have been conducted by or for HIOSSEN Inc., which is based in Fairless Hills, Pennsylvania, USA. The testing is prospective as it was conducted to support the
      510(k) submission.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

    Not applicable. This is a medical device (dental abutment) undergoing non-clinical mechanical and biological testing for substantial equivalence, not an AI or diagnostic device requiring expert consensus for a ground truth in image interpretation or patient diagnosis.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    Not applicable. As this is not a diagnostic device or a study involving human interpretation of data, there is no adjudication method used. The testing relies on objective, standardized physical and material property measurements.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    Not applicable. This is not an AI-assisted diagnostic device, and therefore, no MRMC study or AI-related effectiveness study was conducted.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    Not applicable. This is a physical medical device (dental abutment), not an algorithm or software.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

    The "ground truth" in this context refers to the established standards and specifications for material properties, mechanical performance, and biocompatibility of dental implant abutments. These are based on:

    • International Standards: Primarily ISO 14801 for fatigue testing, and ISO 17665-1, ISO/TS 17665-2 for sterilization validation.
    • Predicate Device Data: The performance and safety profile of the legally marketed predicate devices (K162390, K132219, K123627) serve as the benchmark for demonstrating substantial equivalence.
    • Material Specifications: Established specifications for Titanium Alloy Ti-6AL 4V and Zirconia Oxide.

    8. The sample size for the training set

    Not applicable. There is no training set as this is not a machine learning or AI device.

    9. How the ground truth for the training set was established

    Not applicable. No training set was used.

    Ask a Question

    Ask a specific question about this device

    Why did this record match?
    Reference Devices :

    K132219, K120822, K080239, K990342, K140737

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Straumann® Variobase™ prosthetic components directly connected to the endosseous dental implants are indicated for use as an aid in prosthetic rehabilitations. The prosthetic restoration (bridge or over-denture) can be cemented on the Straumann® Variobase™ prosthetic components. A temporary restoration can be used prior to the insertion of the final components to maintain, stabilize and form the soft tissue during the healing phase. They may not be placed into occlusion. Final abutments and restorations may be placed into occlusion when the implant is fully osseointegrated.

    Straumann® Variobase® abutments are intended to be placed into Straumann dental implants to provide support for prosthetic reconstructions such as bridges and over-dentures.

    Device Description

    The Straumann Variobase Abutments for Bars & Bridges are non-engaging abutments made from Ti-6Al-7Nb (TAN) that support a bridge or bar reconstruction (framework or full contour) on two or more dental implants. The corresponding basal screw is delivered with the abutment for connecting the abutment to the implant. A dental laboratory technician designs and manufactures the bridge/bar reconstruction via their preferred workflow using traditional or CAD/CAM methods.

    AI/ML Overview

    The provided text is a 510(k) Summary for the Straumann® Variobase® Abutment for Bridges/Bars. It outlines the characteristics of the device and its substantial equivalence to predicate devices, but it does not contain information about acceptance criteria or a specific study proving the device meets those criteria in the typical sense of a diagnostic or clinical performance study.

    Instead, the "study" referred to is a bench study focused on mechanical performance and material characteristics to demonstrate substantial equivalence to legally marketed predicate devices, rather than a clinical study evaluating diagnostic accuracy or effectiveness.

    Here's a breakdown of the requested information based on the provided text, highlighting what is available and what is not:

    1. A table of acceptance criteria and the reported device performance

    The document doesn't present explicit "acceptance criteria" in a quantitative, pass/fail table for clinical performance. Instead, it discusses "equivalence" based on various features and a dynamic fatigue test.

    Feature/CharacteristicAcceptance Criteria (Implied by Equivalence)Reported Device Performance
    Indications for UseWithin the scope of the predicate devices."The basic indication of providing support for prostheses is identical. Where the predicate is indicated for both single-unit and multi-unit restorations, the subject devices are only suitable for multi-unit restorations. This limited indication is within the scope of the indication of the predicate devices." (Equivalent)
    Abutment MaterialIdentical to predicate device material (Titanium-Aluminum-Niobium alloy (Ti-6Al-7Nb))."Titanium-Aluminum-Niobium alloy (Ti-6Al-7Nb)" (Identical)
    Abutment DiameterWithin the range of predicate device diameters."Subject device diameters are within the range of diameters for the predicate device." (Subject: 4.2 - 7.0 mm, Predicate: 3.8 - 7.0 mm) (Equivalent)
    Abutment HeightSame minimum height as predicate; taller variants' performance addressed via bench testing."The subject devices have the same minimum height as the Variobase predicates. Performance of the taller subject device is addressed via bench testing." (Subject: 3.5 – 5.2 mm, Predicate: 3.5 – 4.5 mm) (Equivalent)
    Abutment Apical DesignMorse taper portion identical to predicate; non-engaging feature for multi-unit restorations."The Morse taper portion of both the subject and Variobase predicate devices are identical. Non-engaging abutments are necessary so as not to introduce mechanical constraints in the framework that would affect fit." (Equivalent)
    Abutment Coronal DesignConic taper within the range established by reference predicates."The conic taper of the subject devices is within the range established by the synOcta and Multibase predicates." (Equivalent)
    Restoration MaterialWithin the scope of materials indicated for the predicate devices."The range of materials cleared for use with the predicate devices is within the scope of materials indicated for the subject devices." (Equivalent)
    Design/Manufacturing WorkflowIdentical to predicate."Identical"
    Mode of AttachmentIdentical to predicate (Screw-retained or cement retained)."Identical"
    ReusableIdentical to predicate (No)."Identical"
    Dynamic FatigueSatisfactorily addressed via bench studies and consistent with FDA guidance and ISO 14801. (Implied: Device must meet or exceed performance of predicate or relevant standards.)"Dynamic fatigue test data consistent with FDA guidance and ISO 14801 have been referenced in support of this submission." This indicates the device passed these structural integrity tests relevant to dental implants and abutments.

    2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    The document primarily relies on bench testing for dynamic fatigue. It doesn't specify a sample size for this testing other than "data have been referenced." No information is provided about data provenance ( країни of origin or retrospective/prospective nature) as it's not a clinical data study.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

    This information is not applicable here because the "test set" refers to physical samples for mechanical testing (dynamic fatigue), not a dataset for a diagnostic algorithm where expert ground truth would be established.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    This information is not applicable. Adjudication methods are typically used for establishing ground truth in clinical or image-based studies, not for mechanical bench testing.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    This is not applicable. The device is an endosseous dental implant abutment, a physical component, not an AI or diagnostic tool. Therefore, MRMC studies and AI assistance comparisons are irrelevant.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    This is not applicable as the device is a physical medical device, not a software algorithm.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

    The "ground truth" in this context is the mechanical performance standards (e.g., ISO 14801 for dynamic fatigue) and the material properties of the device, which are evaluated through bench testing. There is no biological or diagnostic "ground truth" from human subjects involved in this 510(k) submission for mechanical equivalence.

    8. The sample size for the training set

    This is not applicable. There is no "training set" as this is not a machine learning or AI-based device.

    9. How the ground truth for the training set was established

    This is not applicable for the same reason as point 8.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1