Search Filters

Search Results

Found 3 results

510(k) Data Aggregation

    K Number
    K162134

    Validate with FDA (Live)

    Date Cleared
    2016-12-14

    (135 days)

    Product Code
    Regulation Number
    888.3070
    Age Range
    All
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The ENNOVATE Spinal System is intended for anterior/anterolateral and posterior, non-cervical pedicle and non-pedicle fixation. Fixation is limited to skeletally mature patients and is intended to be used as an adjunct to fusion using autograft or allograft. The ENNOVATE System can be used in both an Open and Minimally Invasive Surgery (MIS). The device is indicated for treatment of the following acute and chronic instabilities or deformities.

    1. Degenerative disc disease (defined as discogenic back pain with degeneration of the disc confirmed by history and radiographic studies).
    2. Spondylolisthesis,
    3. Trauma (i.e., fracture or dislocation)
    4. Spinal Stenosis.
    5. Deformities or Curvatures (i.e., scoliosis, kyphosis, and/or lordosis),
    6. Tumor.
    7. Pseudoarthrosis, and
    8. Failed previous fusion
    Device Description

    The ENNOVATE Spinal System is an implant system used to correct spinal deformity and facilitate the biological process of spinal fusion. This system is intended for posterior use in the thoracic, lumbar and sacral areas of the spine. This system includes polyaxial screws of varying diameters and lengths, rod-to-rod and cross connectors of various styles and lengths. All implant components are top loading and top tightening. The implants in this system are manufactured from titanium alloy (Ti-6Al-4V), conforming to ISO 5832-3.
    The ENNOVATE Spinal System is a spinal rod and screw system. This system's polyaxial screws can be rigidly locked into a wide range of configurations, therefore allowing each construct to be formed to the needs of an individual patient. Rods of this system may be shaped intraoperatively to correct or maintain proper spinal curvature.

    AI/ML Overview

    This document describes a spinal system, not a device that would typically have acceptance criteria and performance metrics in the way an AI/ML or diagnostic device would. This is a 510(k) premarket notification for a medical device called the "ENNOVATE Spinal System," which is a metallic implant system for spinal fusion.

    The document focuses on demonstrating substantial equivalence to existing predicate devices, rather than establishing novelty or specific performance criteria like sensitivity, specificity, or accuracy for a diagnostic tool.

    Therefore, many of the requested categories are not applicable to this type of device and regulatory submission. I will explain why each category is not applicable and what information is provided in the document.


    1. A table of acceptance criteria and the reported device performance

    • Not Applicable. For a spinal implant system like the ENNOVATE Spinal System, "acceptance criteria" are typically related to mechanical and material performance standards, not diagnostic accuracy or clinical outcomes in the same way as an AI/ML device. The "performance" reported is compliance with these standards through mechanical testing.
    Acceptance Criteria (Mechanical / Material Standards)Reported Device Performance
    Mechanical Testing:
    Dynamic/Static compression (ASTM F1717-15)Testing completed.
    Static torsion (ASTM F1717-15)Testing completed.
    Dynamic compression/tension (ASTM F2193-14)Testing completed.
    Dynamic/Static flexion bending (ASTM F1798-13)Testing completed.
    Static rod grip (ASTM F1798-13)Testing completed.
    Static rod/cross rod torsion (ASTM F1798-13)Testing completed.
    Axial compression (ASTM F543-13)Testing completed.
    Pull out strength (ASTM F543-13)Testing completed.
    Driving torque (ASTM F543-13)Testing completed.
    Material Composition:
    Titanium alloy (Ti-6Al-4V) conforming to ISO 5832-3Meets standard.
    Medical grade silicone, stainless steel, titanium alloy, and PEEK (for instruments)Same materials as predicate devices.

    Note: The document states that "All required testing per 'Draft Guidance for the Preparation of Premarket Notifications (510(k)s) Applications for Orthopedic Devices-The Basic Elements' were done where applicable. In addition, testing per the guidance 'Spinal System 510(k)s May 3, 2004' was completed where applicable." This indicates that the device met the performance requirements outlined in these guidance documents, primarily through mechanical testing to demonstrate substantial equivalence.


    The following sections are NOT APPLICABLE because this document describes a physical spinal implant system, not a diagnostic AI/ML device requiring clinical performance evaluation with ground truth.

    2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    • Not Applicable. This is a mechanical device, not an AI/ML diagnostic. No 'test set' of clinical data for diagnostic performance is described. The 'test set' here refers to physical devices tested in a lab.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

    • Not Applicable. Ground truth, in the context of expert review, is for diagnostic devices or AI, not for mechanical spinal implants.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    • Not Applicable. Adjudication methods are relevant for clinical endpoints or diagnostic interpretations, not for mechanical testing of an implant.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    • Not Applicable. This is specific to AI/ML diagnostic tools, not an orthopedic implant.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    • Not Applicable. This is specific to AI/ML diagnostic tools.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

    • Not Applicable. Ground truth refers to the definitive determination of a condition for diagnostic purposes. For a spinal implant, the "truth" is whether it meets engineering specifications and functions mechanically as intended.

    8. The sample size for the training set

    • Not Applicable. This is for AI/ML models. No training set is involved for this mechanical device.

    9. How the ground truth for the training set was established

    • Not Applicable. As above, no training set or ground truth in this context.
    Ask a Question

    Ask a specific question about this device

    K Number
    K123549

    Validate with FDA (Live)

    Date Cleared
    2013-02-25

    (98 days)

    Product Code
    Regulation Number
    888.3070
    Age Range
    All
    Why did this record match?
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Polaris Spinal System is a non-cervical spinal fixation device intended for immobilization and stabilization as an adjunct to fusion as a pedicle screw fixation system, a posterior hook and sacral/iliac screw fixation system, or as an anterior or anterolateral fixation system. Pedicle screw fixation is limited to skeletally mature patients and for use with autograft and/or allograft. The device is indicated for all the following indications: degenerative disc disease (defined as discogenic back pain with degeneration of the disc confirmed by history and radiographic studies). spondylolisthesis, trauma, (i.e., fracture or dislocation), deformity or curvature (i.e., scoliosis, kyphosis, and/or lordosis), tumor, stenosis, pseudoarthrosis, or failed previous fusion.

    The Ballista/Ballista II instruments are intended to be used with Ballista II/Polaris 5.5mm implants. Cannulated screws and percutaneous rods may be used with the Ballista/Ballista II instruments to provide the surgeon with a percutaneous approach for posterior spinal surgery for the following indications: degenerative disc disease (defined as discogenic back pain with degeneration of the disc confirmed by history and radiographic studies), spondylolisthesis, trauma, (i.e., fracture or dislocation), deformity or curvature (i.e., scoliosis, and/or lordosis), tumor, stenosis, pseudoarthrosis, or failed previous fusion that warrant the use of a non-cervical spinal fixation device intended for use as a pedicle screw fixation system or sacral/iliac screw fixation system. Pedicle screw fixation is limited to skeletally mature patients and for use with autograft and/or allograft.

    The Polaris Spinal System may be used with the instruments in the AccuVision Minimally Invasive Spinal Exposure System to provide the surgeon with a minimally invasive approach for posterior spinal surgery.

    The dominos in the Polaris Spinal System can be used to connect the Polaris Spinal System to the Altius Spinal System, the Array Spinal System, the Biomet Omega21 Spinal System, or the Synergy Spinal System to achieve additional levels of fixation. Please refer to the individual system's Package Insert for a list of the indications for use for each system.

    Device Description

    The Polaris Spinal System is a non-cervical spinal fixation device. The system includes screws, various types and sizes of rods, locking nuts, hooks, lateral connectors, plugs, fixation washers, rod connectors/dominos and various cross connectors. Various instruments are also available for use by the surgeon to facilitate implantation of the device. This submission is a line extension to Polaris Spinal System to add other styles of multiaxial screw, including a screw with translating feature, and to provide the Polaris Spinal System sterile packed.

    AI/ML Overview

    The provided document describes a 510(k) summary for the Biomet Spine Polaris Spinal System - Translation Screw, which is a non-cervical spinal fixation device.

    Here's an analysis of the acceptance criteria and study information:

    1. Table of Acceptance Criteria and Reported Device Performance

    Acceptance CriteriaReported Device Performance
    Mechanical Testing:
    Static compression bending per ASTM 1717-11aMet all mechanical test requirements based on worst-case construct testing.
    Static torsion per ASTM 1717-11aMet all mechanical test requirements based on worst-case construct testing.
    Dynamic compression bending fatigue per ASTM 1717-11aMet all mechanical test requirements based on worst-case construct testing.
    Substantial Equivalence:Reported as substantially equivalent to predicate devices regarding material, design, operational principles, intended use, indications, and fundamental technology.

    2. Sample Size Used for the Test Set and Data Provenance

    The document does not specify a "test set" in the context of clinical data for algorithmic performance. The testing performed was mechanical testing of the device hardware.

    • Test Set Sample Size: Not applicable in the context of clinical data or algorithm performance.
    • Data Provenance: Not applicable. The "data" comes from physical mechanical tests conducted on the device components.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

    Not applicable. Ground truth, in the context of algorithm performance, typically refers to expert-validated clinical diagnoses or outcomes. This submission focuses on mechanical performance data of a physical medical device.

    4. Adjudication Method for the Test Set

    Not applicable. Adjudication methods are relevant for resolving discrepancies in expert opinions for clinical ground truth. Mechanical testing has well-defined pass/fail criteria per ASTM standards.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done, If So, What Was the Effect Size of How Much Human Readers Improve with AI vs. Without AI Assistance

    No, an MRMC comparative effectiveness study was not done. This device is a physical spinal fixation system, not an AI or imaging diagnostic tool.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done

    No, a standalone algorithm performance study was not done. This submission is for a physical medical device.

    7. The Type of Ground Truth Used (Expert Consensus, Pathology, Outcomes Data, etc.)

    The "ground truth" for the mechanical testing was the established pass/fail criteria defined by ASTM 1717-11a standard test methods for static and fatigue testing of spinal implant constructs in a vertebrectomy model.

    8. The Sample Size for the Training Set

    Not applicable. This device is a physical implant, not an AI algorithm requiring a training set.

    9. How the Ground Truth for the Training Set Was Established

    Not applicable, as there is no training set for a physical device.

    Ask a Question

    Ask a specific question about this device

    K Number
    K120838

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2012-07-31

    (133 days)

    Product Code
    Regulation Number
    888.3070
    Age Range
    All
    Why did this record match?
    Reference Devices :

    Medtronic CD Horizon, K113529, EBI/Biomet Polaris, K111957, Aesculap S4, K112551, Synthes USS, K111358

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Synthes USS are non-cervical spinal fixation devices intended for posterior pedicle screw fixation (T1-S2/ilium), posterior hook fixation (T1-L5), or anterolateral fixation (T8-L5). Pedicle screw fixation is limited to skeletally mature patients with the exception of the Small Stature USS, which includes small stature and pediatric patients. These devices are indicated as an adjunct to fusion for all of the following indications: degenerative disc disease (defined as discogenic back pain with degeneration of the disc confirmed by history and radiographic studies), spondylolisthesis, trauma (i.e., fracture or dislocation), deformities or curvatures (i.e., scoliosis, kyphosis, and/or lordosis, Scheuermann's Disease), tumor, stenosis, and failed previous fusion (pseudoarthrosis).

    When treating patients with Degenerative Disc Disease (DDD), transverse bars are not cleared for use as part of the posterior pedicle screw construct.

    When used with the 3.5 mm/6.0mm parallel connectors, the Synthes USS 6.0 mm rod systems can be linked to the CerviFix 3.5mm Systems. In addition, when used with 3.5 mm/5.0mm parallel connectors, the Synthes Small Stature USS can be linked to the CerviFix 3.5mm Systems. When used with the 5.0 mm/6.0mm parallel connectors, the Synthes Small Stature USS can be linked to the Synthes USS 6.0 mm rod systems. When used with the 5.5 mm/6.0mm parallel or extension connectors, Synthes USS 5.5 mm rod systems can be linked to the Synthes USS 6.0 mm rod systems. 5.5 mm/5.5mm parallel or extension connectors can be used to link all Synthes USS 5.5 mm rod systems to one another. 6.0 mm/6.0mm parallel or extension connectors can be used to link all Synthes USS 6.0 mm rod systems to one another.

    Rod-to-rod connectors can be used to link all Synthes USS 5.5 mm rod systems to one another.

    When used with the 3.5 mm/6.0mm and 4.0 mm/6.0mm tapered rods, the Synthes USS 6.0 mm rod systems can be linked to the CerviFix 3.5 mm and 4.0 mm Systems, respectively. When used with the 3.5 mm/5.5mm and 4.0 mm/5.5 mm tapered rods, Synthes USS 5.5 mm rod systems can be linked to the CerviFix 3.5 mm and 4.0 mm Systems, respectively. When used with the 5.5 mm/6.0mm tapered rods, the Synthes USS 6.0 mm rod systems can be linked to Synthes USS 5.5 mm rod systems.

    In addition, Synthes USS 6.0 mm rod systems can be interchanged with all USS 6.0 mm rods and transconnectors except Synthes 6.0 mm cobalt-chromium-molybdenum alloy and titanium grade 3 rods, which can only be used with Pangea. Synthes USS 5.5 mm rod systems can be interchanged with all USS 5.5 mm rods and transconnectors.

    Device Description

    This is an addition to Synthes' existing non-cervical spinal fixation devices intended for are non-cervical spinal fixation devices intended for posterior pedicle screw fixation (T1-S2/ilium), posterior hook fixation (T1-L5), or anterolateral fixation (T8-L5). The current system is comprised of monoaxial and polyaxial screws, rods, locking caps, transverse bars and connectors. The additional locking cap in this submission is manufactured of cobalt chrome (Cobalt- 28Chromium - 6Molybdenum per ASTM F1537 - 08).

    AI/ML Overview

    The provided text describes a 510(k) Summary for a medical device called the "Synthes Matrix System," which is an addition to an existing line of non-cervical spinal fixation devices. The document explicitly states that clinical data and conclusions were NOT needed for this device. This means that the information requested regarding acceptance criteria, study design, sample sizes, expert ground truth, MRMC studies, standalone performance, and training sets is not available in this document because a clinical study was not performed or deemed necessary for this 510(k) submission.

    The acceptance criteria and proof of meeting them are based on non-clinical performance data (bench testing), which demonstrated equivalence or superiority to predicate devices.

    Here's a breakdown of the available information:

    1. A table of acceptance criteria and the reported device performance

    Acceptance Criteria (Implied)Reported Device Performance
    Equivalence or superiority to predicate devices in mechanical propertiesBench testing results demonstrate that Synthes Matrix performs equivalently or superiorly to the above-mentioned predicates in static compression bending, static torsion, and dynamic compression bending (in accordance with ASTM F1717-11a).
    Compliance with relevant ASTM standardsThe testing was conducted in accordance with ASTM F1717-11a (Standard Test Methods for Spinal Implant Constructs in a Vertebrectomy Model (for Posterior Spinal Implant Constructs)).

    2. Sample size used for the test set and the data provenance

    • Sample Size for Test Set: Not specified. (Refers to mechanical testing specimens, not human subjects or a data set for an AI model).
    • Data Provenance: Not applicable in the context of human data. The data comes from bench testing of the device components.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    • Experts: Not applicable. Ground truth for mechanical testing is established by engineering specifications and direct measurement, not expert consensus on medical images or diagnoses.

    4. Adjudication method for the test set

    • Adjudication Method: Not applicable. This is mechanical bench testing, not a clinical study requiring adjudication of diagnostic outcomes.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    • MRMC Study: No. This device is a spinal implant, not an AI or diagnostic imaging device. No human reader studies were performed or relevant.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    • Standalone Performance: Not applicable. This is a physical medical device (spinal implant), not an algorithm or AI.

    7. The type of ground truth used

    • Ground Truth: For mechanical testing, the ground truth is based on the physical properties and performance characteristics defined by engineering standards (e.g., ASTM F1717-11a) and compared against the performance of legally marketed predicate devices.

    8. The sample size for the training set

    • Training Set Sample Size: Not applicable. There is no AI model or learning algorithm mentioned or relevant to this device.

    9. How the ground truth for the training set was established

    • Ground Truth for Training Set: Not applicable. There is no AI model or learning algorithm mentioned or relevant to this device.

    In summary, this 510(k) submission relied on non-clinical (bench) testing to demonstrate substantial equivalence to predicate devices, and explicitly stated that clinical data was not required. Therefore, all questions pertaining to clinical study design, human data, expert review, and AI model specifics are not addressed by the provided document.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1