The S4 Spinal System, when used as a pedicle screw fixation system, is indicated for use in patients: a) having severe spondylolisthesis (Grade 3 and 4) at the L5-S1 ioin: b) who are receiving fusion using autogenous graft only; c) who are having the device fixed or attached to the lumbar or sacral spine; and d) who are having the device removed after the development of a solid fusion mass may not go above the L5-S1 joint, the levels of pedicle screw fixation may span from L3 to the sacrum.
The S4 Spinal System is a pedicle screw system intended to provide immobilization and stabilization of spinal segments in skeletally mature patients as an adjunct to fusion in the treatment of the following acute and chronic instabilities or deformities of the thoracic, lumbar, and sacral spin: degenerative spondyloisthesis with objective evidence of neurological impairment, fracture, dislocation, scoliosis, spinal tumor, and failed previous fusion (pseudarthrosis).
When used as a hook and sacral screw system (other than pedicle screw fixation system for high grade spondylolisthesis), the S4 Spinal System is intended for use in the treatment of degenerative disc disease (as defined by chronic back pain of discogenic origin with the degeneration of the disc confirmed by a history and radiographic studies), idiopathic scoliosis, spondylolisthesis, kyphotic or lordotic deformity of the spine, loss of stability due to tumors, spinal stenosis, vertebral fracture of dislocation, pseudoarthrosis, and previous failed spinal surgery/fusion. When used for this indication, screws of the S4 Spinal System are intended for sacral/iliac attachment only. Hooks and transverse connectors of the system are intended for posterior throracic and/or lumbar use only. The levels of use for hook and sacral screw fixation of this system are T1 to the sacrum.
The S4 Spinal System consists of polyaxial screws and monoaxial screws of varying diameters and lengths, various hook styles, rods of varying lengths, and fixed and adjustable rod to rod connectors. All implant components are top loading and top tightening. The S4 Spinal System is manufactured from Titanium and Titanium alloy in accordance with ISO 5832/3 and ISO 5832/2.
Here's an analysis of the provided text regarding the S4 Spinal System, focusing on information typically found in acceptance criteria and study designs for medical devices.
Important Note: The provided document is a 510(k) summary and an FDA clearance letter for a line extension of an existing spinal system. It primarily focuses on demonstrating substantial equivalence to predicate devices based on technological characteristics and mechanical testing. It does not describe a study involving human or animal subjects to assess clinical performance or comparative effectiveness with human readers using AI. Therefore, many of the requested fields related to clinical studies, human readers, and ground truth establishment will be marked as "Not Applicable" or "Not Provided" based on the content.
Acceptance Criteria and Device Performance
The document does not explicitly state quantitative "acceptance criteria" in the format of specific performance metrics that the device had to achieve for clinical efficacy (e.g., success rates, complication rates). Instead, the acceptance is based on demonstrating substantial equivalence to existing, legally marketed predicate devices through technological similarities and mechanical performance testing.
Table of Acceptance Criteria and Reported Device Performance
| Acceptance Criteria Category | Specific Criteria (Implicit from Submission) | Reported Device Performance (Summary from Submission) |
|---|---|---|
| Material Equivalence | Material composition must be same as predicate or an accepted biocompatible material. | Manufactured from Titanium and Titanium alloy in accordance with ISO 5832/3 and ISO 5832/2, same as predicate devices. |
| Design/Dimensional Equivalence | Similar shapes and sizes to predicate devices. | New components are offered in similar shapes and sizes as the predicate devices. All implant components are top loading and top tightening. |
| Mechanical Performance | Must meet relevant standards for spinal implants (fatigue, static strength, etc.). | "All required testing per 'Draft Guidance for the Preparation of Premarket Notifications (510(k)s) Applications for Orthopedic Devices-The Basic Elements' were done where applicable. In addition, testing per the 'Spinal System 510(k)s' was completed where applicable." |
| Intended Use Equivalence | Indications for use must be same or a subset of predicate device indications. | Indications for use are consistent with the predicate devices, covering various spinal conditions requiring fusion and stabilization. |
Study Details (Based on Provided Text)
Given the nature of a 510(k) submission for a line extension of an orthopedic implant, the "study" referred to is almost entirely mechanical and material testing, not a clinical study involving human patients or AI algorithms.
-
Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective):
- Sample Size: Not specified for individual tests. The "test set" refers to the physical device components subjected to mechanical tests. The number of components tested for each specific mechanical evaluation (e.g., fatigue, static bending) is not provided in this summary.
- Data Provenance: Not applicable in the context of geographical origin for mechanical testing. The testing was conducted internally or by accredited labs as part of the manufacturer's due diligence for regulatory submission. It is inherently prospective in the sense that the new components were manufactured and then tested to demonstrate performance.
-
Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience):
- Not Applicable. The "ground truth" for mechanical testing is established by engineering standards and specifications (e.g., ISO standards, FDA guidance documents for spinal systems), not by human expert consensus or clinical interpretation.
-
Adjudication method (e.g. 2+1, 3+1, none) for the test set:
- None. Adjudication methods like 2+1 or 3+1 refer to resolving discrepancies among human expert reviewers in clinical studies. This is not relevant for mechanical testing. Mechanical test results are objective measurements against defined pass/fail criteria.
-
If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:
- No, an MRMC comparative effectiveness study was not done. This is a mechanical device, not an AI-assisted diagnostic tool.
-
If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:
- No, a standalone algorithm performance study was not done. This device does not involve any AI algorithms.
-
The type of ground truth used (expert consensus, pathology, outcomes data, etc):
- Engineering Standards and Specifications. The "ground truth" for the performance of this device in a 510(k) context consists of meeting predefined mechanical performance limits and material specifications outlined in national and international standards (e.g., ISO 5832 series for materials, and relevant FDA guidance documents for spinal implants concerning fatigue, static strength, and construct integrity).
-
The sample size for the training set:
- Not Applicable. There is no "training set" in the context of a mechanical device.
-
How the ground truth for the training set was established:
- Not Applicable. There is no "training set" for this type of device.
{0}------------------------------------------------
SEP 2 0 2007
Page 1 of 2
510(k) SUMMARY (as required by 21 CFR 807.92) B.
S4 Spinal System
July 13, 2007
COMPANY:
Aesculap Implant Systems, Inc. 3773 Corporate Parkway Center Valley, PA 18034 Establishment Registration Number: 3005673311
CONTACT:
Kathy A. Racosky 610-984-9291 (phone) 610-791-6882 (fax) kathy.racosky@aesculap.com (email)
TRADE NAME: ਟੈਪੇ
COMMON NAME: S4 Spinal System
DEVICE CLASS: Class III
PRODUCT CODE: NKB, MNI, MNH, and KWP
888.3070 -- Orthosis, Spinal Pedicle Fixation For REGULATION NUMBER: Degenerative Disc Disease 888.3070 - Orthosis, Spinal Pedicle Fixation 888.3070 - Orthosis, Spondyloisthesis Spinal Fixation 888.3050 - Appliance, Fixation, Spinal Fixation
REVIEW PANEL: Orthopedics
SUBSTANTIAL EQUIVALENCE
Aesculap Implant Systems, Inc. believes that the S4 Spinal System additions are substantially equivalent to the existing components of the S4 Spinal System ({Revolution Spinal Fixation System K032219} / K062065) and Stryker Spine Xia Spinal System (K043473).
DEVICE DESCRIPTION
The S4 Spinal System consists of polyaxial screws and monoaxial screws of varying diameters and lengths, various hook styles, rods of varying lengths, and fixed and adjustable rod to rod connectors. All implant components are top loading and top tightening. The S4 Spinal System is manufactured from Titanium and Titanium alloy in accordance with ISO 5832/3 and ISO 5832/2.
{1}------------------------------------------------
S4 Spinal System
Line Extension
Page 2 of 2
INDICATIONS FOR USE
The S4 Spinal System, when used as a pedicle screw fixation system, is indicated for use in patients: a) having severe spondylolisthesis (Grade 3 and 4) at the L5-S1 ioin: b) who are receiving fusion using autogenous graft only; c) who are having the device fixed or attached to the lumbar or sacral spine; and d) who are having the device removed after the development of a solid fusion mass may not go above the L5-S1 joint, the levels of pedicle screw fixation may span from L3 to the sacrum.
The S4 Spinal System is a pedicle screw system intended to provide immobilization and stabilization of spinal segments in skeletally mature patients as an adjunct to fusion in the treatment of the following acute and chronic instabilities or deformities of the thoracic, lumbar, and sacral spin: degenerative spondyloisthesis with objective evidence of neurological impairment, fracture, dislocation, scoliosis, spinal tumor, and failed previous fusion (pseudarthrosis).
When used as a hook and sacral screw system (other than pedicle screw fixation system for high grade spondylolisthesis), the S4 Spinal System is intended for use in the treatment of degenerative disc disease (as defined by chronic back pain of discogenic origin with the degeneration of the disc confirmed by a history and radiographic studies), idiopathic scoliosis, spondylolisthesis, kyphotic or lordotic deformity of the spine, loss of stability due to tumors, spinal stenosis, vertebral fracture of dislocation, pseudoarthrosis, and previous failed spinal surgery/fusion. When used for this indication, screws of the S4 Spinal System are intended for sacral/iliac attachment only. Hooks and transverse connectors of the system are intended for posterior throracic and/or lumbar use only. The levels of use for hook and sacral screw fixation of this system are T1 to the sacrum.
TECHNOLOGICAL CHARACTERISTICS(compared to Predicate(s))
The new components of the S4 Spinal System are offered in similar shapes and sizes as the predicate devices. All the components are manufactured from Titanium and Titanium Alloy, which is the same material as the predicate devices.
PERFORMANCE DATA
All required testing per "Draft Guidance for the Preparation of Premarket Notifications (510(k)s) Applications for Orthopedic Devices-The Basic Elements" were done where applicable. In addition, testing per the "Spinal System 510(k)s" was completed where applicable.
{2}------------------------------------------------
Image /page/2/Picture/1 description: The image shows the seal of the Department of Health & Human Services USA. The seal features a stylized eagle with its wings spread, facing right. The eagle is composed of three curved lines that suggest feathers. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" is arranged in a circular pattern around the eagle. The text is in all caps and is evenly spaced around the circle.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
SEP 2 0 2007
Aesculap Implant Systems, Incorporated % Ms. Kathy Racosky Regulatory Affairs Specialist 3773 Corporate Parkway Center Valley, Pennsylvania 18034
Re: K071945
Trade/Device Name: S4 Spinal System Regulation Number: 21 CFR 888.3070 Regulation Name: Pedicle screw spinal system Regulatory Class: Class III Product Code: NKB, MNI, MNH and K WP Dated: September 12, 2007 Received: September 13, 2007
Dear Ms. Racosky:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendmouts, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Trug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA), You may, therefore, market the device, subject to the general controls provisions of the Act The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{3}------------------------------------------------
Page 2 - Ms. Kathy Racosky
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Center for Devices and Radiological Health's (CDRH's) Office of Compliance at (240) 276-0120. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding postmarket surveillance, please contact CDRH's Office of Surveillance and Biometric's (OSB's) Division of Postmarket Surveillance at 240-276-3474. For questions regarding the reporting of device adverse events (Medical Device Reporting (MDR)), please contact the Division of Surveillance Systems at (240) 276-3464. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at toll-free number (800) 638-2041 or (240) 276-3150 or the Internet address http://www.fda.gov/cdrh/industry/support/index.html
Sincerely yours,
Mark N. Melkerson Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health
{4}------------------------------------------------
Line Extension
Page 1 of 1
INDICATIONS FOR USE STATEMENT A.
K071945 510(k) Number:
Device Name: S4 Spinal System
Indications for Use:
The S4 Spinal System, when used as a pedicle screw fixation system, is indicated for use in patients: a) having severe spondylolisthesis (Grade 3 and 4) at the L5-S1 join; b) who are receiving fusion using autogenous graft only; c) who are having the device fixed or attached to the lumbar or sacral spin; and d) who are having the device removed after the development of a solid fusion mass may not go above the L5-S1 joint, the levels of pedicle screw fixation may span from L3 to the sacrum.
The S4 Spinal System is a pedicle screw system intended to provide immobilization and stabilization of spinal segments in skeletally mature patients as an adjunct to fusion in the treatment of the following acute and chronic instabilities or deformities of the thoracic, lumbar, and sacral spin: degenerative spondyloisthesis with objective evidence of neurological impairment, fracture, dislocation, scoliosis, spinal tumor, and failed previous fusion (pseudarthrosis).
When used as a hook and sacral screw system (other than pedicle screw fixation system for high grade spondylolisthesis), the S4 Spinal System is intended for use in the treatment of degenerative disc disease (as defined by chronic back pain of discogenic origin with the degeneration of the disc confirmed by a history and radiographic studies), idiopathic scoliosis, spondylolisthesis, kyphotic or lordotic deformity of the spine loass of stability due to tumors, spinal stenosis, vertebral fracture of dislocation, pe euron, roos sis, and previous failed spinal surgery/fusion. When used for this indication, screws of the S4 Spinal System are intended for sacral/iliac attachment only. Hooks and transverse connectors of the system are intended for posterior throracic and/or lumbar use only. The levels of use for hook and sacral screw fixation of this system are T1 to the sacrum.
| Prescription Use | X | and/or Over-the-Counter Use | |
|---|---|---|---|
| (Part 21 CFR 801 Subpart D) | (21 CFR 801 Subpart C) | ||
| (PLEASE DO NOT WRITE BELOW THIS LINE . CONTINIA ON ANOTHER PAGE IF NEEDED) | |||
| Concurrence of CPN - Office Nige Nige Evaluation (ODE) | |||
| Division of General, Restorative, | |||
| and Neurological Devices | 002 | ||