(62 days)
The 5.5 Helical Flange Spinal System is a non-cervical spinal fixation device intended for use as a pedicle screw fixation system, a posterior hook and sacral/iliac screw fixation system, or as an anterolateral fixation system. Pedicle screw fixation is limited to skeletally mature patients. The device is indicated for all of the following indications regardless of the intended use; degenerative disc disease (defined as discogenic back pain with degeneration of the disc confirmed by history and radiographic studies), spondylolisthesis, trauma, (i.e., fracture or dislocation), deformity or curvature (i.e., scoliosis, kyphosis, and/or lordosis), tumor, stenosis, pseudoarthrosis, and failed previous fusion.
The 5.5 Helical Flange Spinal System Percutaneous Instruments, when used with the 5.5 Helical Flange Spinal System cannulated screws and percutaneous rods, are indicated to provide the surgeon with a percutaneous approach for posterior spinal surgery for the following indications, regardless of intended use: degenerative disc disease (defined as discogenic back pain with degeneration of the disc confirmed by history and radiographic studies), spondylolisthesis, trauma, (i.e., fracture or dislocation), deformity or curvature (i.e., kyphosis, and lordosis), stenosis, pseudoarthrosis, and failed previous fusion that warrant the use of a non-cervical spinal fixation device intended for use as a pedicle screw fixation system or sacral/iliac screw fixation system. Pedicle screw fixation is limited to skeletally mature patients.
The 5.5 Helical Flange Spinal System Mini-Open Instruments, when used with the 5.5 Helical Flange Spinal System implants are indicated to provide the surgeon with a minimally invasive approach for posterior spinal surgery for the following indications, regardless of intended use: degenerative disc disease (defined as discogenic back pain with degeneration of the disc confirmed by history and radiographic studies), spondylolisthesis, trauma, (i.e., fracture or dislocation), deformity or curvature (i.e., kyphosis, and lordosis), tumor, stenosis, pseudoarthrosis, and failed previous fusion that warrant the use of a non-cervical spinal fixation device intended for use as a pedicle screw fixation system or sacral/iliac screw fixation system. Pedicle screw fixation is limited to skeletally mature patients.
Not Found
The provided text describes a 510(k) premarket notification for the EBI® 5.5 Helical Flange Spinal System. This document focuses on the substantial equivalence of a medical device to legally marketed predicate devices, rather than on the performance of an AI/ML powered device. As such, the information typically required for acceptance criteria and studies proving the device meets these criteria for AI/ML devices (e.g., accuracy metrics, sample sizes for training/test sets, ground truth establishment, expert qualifications, MRMC studies) is not applicable or present in this document.
The document primarily discusses:
- Technological Characteristics: States that mechanical testing was conducted and demonstrates conformity to design specifications, which were based on predicate devices. It confirms the system "adequately meets the requirements established in design specifications for its mechanical performance."
- Substantial Equivalence: Asserts that the device is substantially equivalent to other legally marketed spinal fixation devices regarding intended use, indications, technological characteristics, and basic principles of operation.
Since this document pertains to a traditional spinal fixation system and not an AI/ML powered device, the requested information (performance metrics, study design, expert involvement, etc. for AI/ML) cannot be extracted from the provided text.
{0}------------------------------------------------
- lof 2
510(k) Summary
EBI, L.P.'s 5.5 Helical Flange Spinal System
JUL 25 2006
| SUBMITTER: | EBI. L.P. |
|---|---|
| ADDRESS: | 100 Interpace ParkwayParsippany, NJ 07054 |
| PHONE: | (973) 299-9300 |
| FAX: | (973) 257-0232 |
| CONTACT PERSON: | Jennifer P. Harakal |
| DATE PREPARED: | May 22, 2006 |
| TRADE NAME: | EBI® 5.5 Helical Flange Spinal System |
| COMMON NAME: | Spinal Fixation Device |
| CLASSIFICATION NAMES: | Spondylolisthesis Spinal Fixation Device SystemSpinal Intervertebral Body Fixation OrthosisSpinal Interlaminal Fixation Orthosis |
| REGULATION NUMBERS: | 21 CFR 888.3050, 21 CFR 888.3070, 21 CFR 888.3060 |
| PRODUCT CODES: | NKB, MNH, MNI, KWQ, KWP |
| PREDICATE DEVICES: | -EBI® Array Spinal Fixation System/EBI® SpineLink II SpinalFixation system with the VuePASS™ Portal Access SurgicalSystem-EBI® Array Spinal Fixation System-EBI® Webb Morley Spine System-Interpore Cross International Synergy™ Spinal System(Posterior)-Interpore Cross International Synergy Spinal System (Anterior)-Medtronic Sofamor Danek CD HORIZON® Spinal System |
INTENDED/INDICATIONS FOR USE:
The 5.5 Helical Flange Spinal System is a non-cervical spinal fixation device intended for use as a pedicle screw fixation system, a posterior hook and sacral/iliac screw fixation system, or as an anterolateral fixation system. Pedicle screw fixation is limited to skeletally mature patients. The device is indicated for all of the following indications regardless of the intended use; degenerative disc disease (defined as discogenic back pain with degeneration of the disc confirmed by history and radiographic studies), spondylolisthesis, trauma, (i.e., fracture or dislocation), deformity or curvature (i.e., scoliosis, kyphosis, and/or lordosis), tumor, stenosis, pseudoarthrosis, and failed previous fusion.
{1}------------------------------------------------
Image /page/1/Picture/0 description: The image shows the text "K061441 PG 2 of 2". The text appears to be handwritten. The top line contains the alphanumeric string "K061441", while the bottom line indicates page 2 of 2.
The 5.5 Helical Flange Spinal System Percutaneous Instruments, when used with the 5.5 Helical Flange Spinal System cannulated screws and percutaneous rods, are indicated to provide the surgeon with a percutaneous approach for posterior spinal surgery for the following indications, regardless of intended use: degenerative disc disease (defined as discogenic back pain with degeneration of the disc confirmed by history and radiographic studies), spondylolisthesis, trauma, (i.e., fracture or dislocation), deformity or curvature (i.e., kyphosis, and lordosis), stenosis, pseudoarthrosis, and failed previous fusion that warrant the use of a non-cervical spinal fixation device intended for use as a pedicle screw fixation system or sacral/iliac screw fixation system. Pedicle screw fixation is limited to skeletally mature patients.
The 5.5 Helical Flange Spinal System Mini-Open Instruments, when used with the 5.5 Helical Flange Spinal System implants are indicated to provide the surgeon with a minimally invasive approach for posterior spinal surgery for the following indications, regardless of intended use: degenerative disc disease (defined as discogenic back pain with degeneration of the disc confirmed by history and radiographic studies), spondylolisthesis, trauma, (i.e., fracture or dislocation), deformity or curvature (i.e., kyphosis, and lordosis), tumor, stenosis, pseudoarthrosis, and failed previous fusion that warrant the use of a non-cervical spinal fixation device intended for use as a pedicle screw fixation system or sacral/iliac screw fixation system. Pedicle screw fixation is limited to skeletally mature patients.
TECHNOLOGICAL CHARACTERISTICS:
Performance Testing
Mechanical testing of the EBI® 5.5 Helical Flange Spinal System was conducted and demonstrates that the proposed system conforms to its design specifications. The design requirements were established based on those of the previously cleared predicate devices. The results of testing conducted demonstrate that the proposed system adequately meets the requirements established in design specifications for its mechanical performance.
Substantial Equivalence
The EB1º 5.5 Helical Flange Spinal System is substantially equivalent to other legally marketed spinal fixation devices with respect to intended use and indications, technological characteristics, and basic principles of operation.
{2}------------------------------------------------
Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a stylized eagle with three lines representing its body and wings. The logo is surrounded by the text "DEPARTMENT OF HEALTH AND HUMAN SERVICES. USA" in a circular arrangement.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
JUL 25 2006
EBI. LP % Ms. Jennifer P. Harakal Regulatory Affairs Specialist 100 Interspace Parkway Parsippany, New Jersey 07054
Re: K061441
Trade/Device Name: EBI® 5.5 Helical Flange Spinal System Regulation Number: 21 CFR 888.3070 Regulation Name: Orthosis, Spinal pedicle fixation, for degenerative disc disease Regulatory Class: III Product Code: NKB, MNH, MNI, KWQ, KWP Dated: May 22, 2006 Received: May 24, 2006
Dear Ms. Harakal:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act s requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable. the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{3}------------------------------------------------
Page 2 - Ms. Jennifer P. Harakal
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0120. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely yours,
Sincerely yours,
Barbara Muchut
Mark N. Mè Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
Indications for Use Statement
510(k) Number (if known):
Device Name: EBI® 5.5 Helical Flange Spinal System
Indications for Use:
The 5.5 Helical Flange Spinal System is a non-cervical spinal fixation device intended for use as a pedicle screw fixation system, a posterior hook and sacral/iliac screw fixation system, or as an anterolateral fixation system. Pedicle screw fixation is limited to skeletally mature patients. The device is indicated for all of the following indications regardless of the intended use; degenerative disc disease (defined as discogenic back pain with degeneration of the disc confirmed by history and radiographic studies). sporidylolisthesis, trauma, (i.e., fracture or dislocation), deformity or curvature (i.e., scoliosis, kyphosis, and/or lordosis), tumor, stenosis, pseudoarthrosis, and failed previous fusion.
The 5.5 Helical Flange Spinal System Percutaneous Instruments, when used with the 5.5 Helical Flange Spinal System cannulated screws and percutaneous rods, are indicated to provide the surgeon with a percutaneous approach for posterior spinal surgery for the following indications, regardless of intended use: degenerative disc disease (defined as discogenic back pain with degeneration of the disc confirmed by history and radiographic studies), spondylolisthesis, trauma, (i.e., fracture or dislocation), deformity or curvature (i.e., yohosis, and lordosis), stenosis, pseudoarthrosis, and failed previous fusion that warrant the use of a non-cervical spinal fixation device intended for use as a pedicle screw fixation system or sacral/iliac screw fixation system. Pedicle screw fixation is limited to skeletally mature patients.
The 5.5 Helical Flange Spinal System Mini-Open Instruments, when used with the 5.5 Helical Flange Spinal System implants are indicated to provide the surgeon with a minimally invasive approach for posterior spinal surgery for the following indications, regardless of intended use: degenerative disc disease (defined as discogenic back pain with degeneration of the disc confirmed by history and radiographic studies), spondylolisthesis, trauma, (i.e., fracture or dislocation), deformity or curvature (j.e., kyphosis, and lordosis), turnor, stenosis, pseudoarthrosis, and failed previous fusion that warrant the use of a non-cervical spinal fixation device intended for use as a pedicle screw fixation system or sacral/iliac screw fixation system. Pedicle screw fixation is limited to skeletally mature patients.
Prescription Use X (Part 21 C.F.R. 801 Subpart D) AND/OR
Over-The-Counter Use (21 C.F.R. 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE -- CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
| Barbare Bondi for | |
|---|---|
| (Division Sign-Off) |
Division of General, Restorative and Neurological Devices
| 510(k) Number | K061441 |
|---|---|
| --------------- | --------- |
Page 1 of 1
N/A