Search Results
Found 4 results
510(k) Data Aggregation
(135 days)
The ENNOVATE Spinal System is intended for anterior/anterolateral and posterior, non-cervical pedicle and non-pedicle fixation. Fixation is limited to skeletally mature patients and is intended to be used as an adjunct to fusion using autograft or allograft. The ENNOVATE System can be used in both an Open and Minimally Invasive Surgery (MIS). The device is indicated for treatment of the following acute and chronic instabilities or deformities.
- Degenerative disc disease (defined as discogenic back pain with degeneration of the disc confirmed by history and radiographic studies).
- Spondylolisthesis,
- Trauma (i.e., fracture or dislocation)
- Spinal Stenosis.
- Deformities or Curvatures (i.e., scoliosis, kyphosis, and/or lordosis),
- Tumor.
- Pseudoarthrosis, and
- Failed previous fusion
The ENNOVATE Spinal System is an implant system used to correct spinal deformity and facilitate the biological process of spinal fusion. This system is intended for posterior use in the thoracic, lumbar and sacral areas of the spine. This system includes polyaxial screws of varying diameters and lengths, rod-to-rod and cross connectors of various styles and lengths. All implant components are top loading and top tightening. The implants in this system are manufactured from titanium alloy (Ti-6Al-4V), conforming to ISO 5832-3.
The ENNOVATE Spinal System is a spinal rod and screw system. This system's polyaxial screws can be rigidly locked into a wide range of configurations, therefore allowing each construct to be formed to the needs of an individual patient. Rods of this system may be shaped intraoperatively to correct or maintain proper spinal curvature.
This document describes a spinal system, not a device that would typically have acceptance criteria and performance metrics in the way an AI/ML or diagnostic device would. This is a 510(k) premarket notification for a medical device called the "ENNOVATE Spinal System," which is a metallic implant system for spinal fusion.
The document focuses on demonstrating substantial equivalence to existing predicate devices, rather than establishing novelty or specific performance criteria like sensitivity, specificity, or accuracy for a diagnostic tool.
Therefore, many of the requested categories are not applicable to this type of device and regulatory submission. I will explain why each category is not applicable and what information is provided in the document.
1. A table of acceptance criteria and the reported device performance
- Not Applicable. For a spinal implant system like the ENNOVATE Spinal System, "acceptance criteria" are typically related to mechanical and material performance standards, not diagnostic accuracy or clinical outcomes in the same way as an AI/ML device. The "performance" reported is compliance with these standards through mechanical testing.
Acceptance Criteria (Mechanical / Material Standards) | Reported Device Performance |
---|---|
Mechanical Testing: | |
Dynamic/Static compression (ASTM F1717-15) | Testing completed. |
Static torsion (ASTM F1717-15) | Testing completed. |
Dynamic compression/tension (ASTM F2193-14) | Testing completed. |
Dynamic/Static flexion bending (ASTM F1798-13) | Testing completed. |
Static rod grip (ASTM F1798-13) | Testing completed. |
Static rod/cross rod torsion (ASTM F1798-13) | Testing completed. |
Axial compression (ASTM F543-13) | Testing completed. |
Pull out strength (ASTM F543-13) | Testing completed. |
Driving torque (ASTM F543-13) | Testing completed. |
Material Composition: | |
Titanium alloy (Ti-6Al-4V) conforming to ISO 5832-3 | Meets standard. |
Medical grade silicone, stainless steel, titanium alloy, and PEEK (for instruments) | Same materials as predicate devices. |
Note: The document states that "All required testing per 'Draft Guidance for the Preparation of Premarket Notifications (510(k)s) Applications for Orthopedic Devices-The Basic Elements' were done where applicable. In addition, testing per the guidance 'Spinal System 510(k)s May 3, 2004' was completed where applicable." This indicates that the device met the performance requirements outlined in these guidance documents, primarily through mechanical testing to demonstrate substantial equivalence.
The following sections are NOT APPLICABLE because this document describes a physical spinal implant system, not a diagnostic AI/ML device requiring clinical performance evaluation with ground truth.
2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)
- Not Applicable. This is a mechanical device, not an AI/ML diagnostic. No 'test set' of clinical data for diagnostic performance is described. The 'test set' here refers to physical devices tested in a lab.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)
- Not Applicable. Ground truth, in the context of expert review, is for diagnostic devices or AI, not for mechanical spinal implants.
4. Adjudication method (e.g. 2+1, 3+1, none) for the test set
- Not Applicable. Adjudication methods are relevant for clinical endpoints or diagnostic interpretations, not for mechanical testing of an implant.
5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
- Not Applicable. This is specific to AI/ML diagnostic tools, not an orthopedic implant.
6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done
- Not Applicable. This is specific to AI/ML diagnostic tools.
7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)
- Not Applicable. Ground truth refers to the definitive determination of a condition for diagnostic purposes. For a spinal implant, the "truth" is whether it meets engineering specifications and functions mechanically as intended.
8. The sample size for the training set
- Not Applicable. This is for AI/ML models. No training set is involved for this mechanical device.
9. How the ground truth for the training set was established
- Not Applicable. As above, no training set or ground truth in this context.
Ask a specific question about this device
(120 days)
The Polaris Spinal System is a non-cervical spinal fixation device intended for immobilization as an adjunct to fusion as a pedicle screw fixation system, a posterior hook and sacral/iliac screw fixation system, or as an anterior or anterolateral fixation system for use with autograft. The Polaris Spinal System is indicated for the following conditions: degenerative disc disease (defined as discogenic back pain with degeneration of the disc confirmed by history and radiographic studies), spondylolisthesis, tracture or dislocation), deformity or curvature (i.e., scoliosis, kyphosis, Scheuerman's disease, and/or lordosis,), tumor, stenosis, pseudarthrosis, or failed previous fusion.
The Ballista and Cypher MIS instruments are intended to be used with Ballista/Cypher MIS/Polaris 5.5mm implants. Cannulated screws and percutaneous rods may be used with the Ballista/Cypher MIS instruments to provide the surgeon with a percutaneous approach for posterior spinal surgery for the above indications.
For pediatric patients, the Polaris Spinal System may be used for posterior, non-cervical pedicle screw fixation as an adjunct to fusion to treat adolescent idiopathic scoliosis and is also indicated for treatment of the following conditions: spondylolisthesis/spondylolysis and fractures caused by tumor and/or trauma. Pedicle screw fixation is limited to a posterior approach.
The Polaris Spinal System may be used with the instruments in the AccuVision Minimally Invasive Spinal Exposure System to provide the surgeon with a minimally invasive approach for posterior spinal surgery.
The dominoes in the Polaris Spinal System can be used to connect the Polaris Spinal System to the Altius Spinal System, Lineum OCT Spine System, the Array Spinal System, the Biomet Omega21 Spinal System, or the Synergy Spinal System to achieve additional levels of fixation. Please refer to the individual system's Package Insert for a list of the indications for use for each system.
The Polaris Spinal System is a non-cervical spinal fixation device made from titanium alloy (Ti-6Al-4V) per ASTM F136, unalloyed titanium per ASTM F67, stainless steel per ASTM F138 or ASTM F1314 and Cobalt Chrome Alloy (Co-28Cr-6Mo) per ASTM F1537. The system includes screws, various types and sizes of rods, locking nuts, hooks, lateral connectors, plugs, fixation washers, rod connectors/dominos, various cross connectors and accessories. Various instruments are also available for use by the surgeon to facilitate implantation of the device. The purpose of this Premarket Notification is to add modified Ø4.75mm multi-axial screws.
The provided document describes the clearance of the Biomet Spine Polaris Spinal System (K151974) by the FDA. It is a 510(k) premarket notification, which means the device is seeking clearance based on substantial equivalence to a legally marketed predicate device, rather than proving safety and effectiveness through clinical trials. Therefore, the information typically found in a study proving acceptance criteria for a novel device (like specific performance metrics, sample sizes for test sets, expert consensus for ground truth, etc.) is generally not present in this type of submission.
Instead, the submission focuses on demonstrating that the new device (modified Ø4.75mm multi-axial screws for the Polaris Spinal System) is as safe and effective as its predicate device (Polaris Spinal System – K133746). This is primarily achieved through mechanical testing to verify that the modified components meet established standards for spinal implants.
Here's a breakdown of the requested information based on the provided document:
1. Table of Acceptance Criteria and Reported Device Performance
Acceptance Criteria (Implicit from Guidance and Standards): The device must meet the mechanical test requirements based on the worst-case construct testing according to "FDA's Guidance for Industry and FDA Staff - Spinal System 510(k)s" dated May 3, 2004, and ASTM standards. This implies demonstrating sufficient strength, durability, and resistance to wear for safe use as a spinal implant.
Reported Device Performance:
Test Type | Standard/Guidance | Reported Performance |
---|---|---|
Static Compression Bending | ASTM F1717 and FDA Guidance | Complied; Verified substantial equivalence to other spinal systems on the market and met all mechanical test requirements based on worst-case construct testing. |
Static Torsion | ASTM F1717 and FDA Guidance | Complied; Verified substantial equivalence to other spinal systems on the market and met all mechanical test requirements based on worst-case construct testing. |
Dynamic Compression Bending Fatigue | ASTM F1717 and FDA Guidance | Complied; Verified substantial equivalence to other spinal systems on the market and met all mechanical test requirements based on worst-case construct testing. |
Modified Flexion-Extension Testing | ASTM F1798 and FDA Guidance | Complied; Verified substantial equivalence to other spinal systems on the market and met all mechanical test requirements based on worst-case construct testing. |
Note: The document explicitly states: "The mechanical testing verifies that the subject components are substantially equivalent to other spinal systems currently on the market for its intended use and has met all mechanical test requirements based on the worst-case construct testing." Specific numerical acceptance values or pass/fail thresholds for each test are not detailed in this summary, as is common for 510(k) summaries where the focus is on meeting recognized standards.
2. Sample Size Used for the Test Set and Data Provenance
- Sample Size for Test Set: Not explicitly stated in the 510(k) summary. Mechanical testing is typically conducted on a representative sample of devices, but the exact number of units tested per configuration is not disclosed.
- Data Provenance: The mechanical testing was conducted by the applicant, Biomet Spine, to demonstrate compliance with FDA guidance and ASTM standards. This data is part of their submission to the FDA. It is not patient or clinical data, but rather engineering test data.
3. Number of Experts Used to Establish the Ground Truth for the Test Set and Their Qualifications
This question is not applicable to this type of submission. "Ground truth" in the context of mechanical testing refers to the physical properties and performance of the device under specific loads, as measured by standardized testing procedures. It does not involve human expert consensus in the way clinical diagnostic or image interpretation studies do. The "truth" is established by the test setup and measurement equipment's accuracy and precision, adhering to established engineering standards.
4. Adjudication Method for the Test Set
Not applicable. Adjudication methods (like 2+1, 3+1) are used in clinical trials or image interpretation studies where there might be disagreement among human evaluators. Mechanical testing results are objective measurements derived from physical tests following defined protocols, not subject to human subjective interpretation requiring adjudication.
5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
Not applicable. This device is a mechanical spinal implant, not an AI-powered diagnostic or assistive tool. Therefore, MRMC studies involving human readers or AI assistance are irrelevant to its evaluation.
6. If a Standalone (i.e. algorithm only without human-in-the-loop performance) was done
Not applicable. This device is a mechanical spinal implant, not an algorithm or software. Its performance is evaluated through physical mechanical testing.
7. The Type of Ground Truth Used
The "ground truth" for the mechanical testing is defined by the objective performance parameters as measured against the specified ASTM standards and FDA guidance. These standards dictate acceptable loads, deformation, and fatigue life for spinal implants. The "truth" is whether the device physically withstands these tests as required by the standards, not a diagnostic or pathological determination.
8. The Sample Size for the Training Set
Not applicable. This is not a machine learning or AI-based device, so there is no concept of a "training set" in the context of device development or evaluation.
9. How the Ground Truth for the Training Set Was Established
Not applicable, as there is no training set for a mechanical spinal implant.
Ask a specific question about this device
(98 days)
The Polaris Spinal System is a non-cervical spinal fixation device intended for immobilization and stabilization as an adjunct to fusion as a pedicle screw fixation system, a posterior hook and sacral/iliac screw fixation system, or as an anterior or anterolateral fixation system. Pedicle screw fixation is limited to skeletally mature patients and for use with autograft and/or allograft. The device is indicated for all the following indications: degenerative disc disease (defined as discogenic back pain with degeneration of the disc confirmed by history and radiographic studies). spondylolisthesis, trauma, (i.e., fracture or dislocation), deformity or curvature (i.e., scoliosis, kyphosis, and/or lordosis), tumor, stenosis, pseudoarthrosis, or failed previous fusion.
The Ballista/Ballista II instruments are intended to be used with Ballista II/Polaris 5.5mm implants. Cannulated screws and percutaneous rods may be used with the Ballista/Ballista II instruments to provide the surgeon with a percutaneous approach for posterior spinal surgery for the following indications: degenerative disc disease (defined as discogenic back pain with degeneration of the disc confirmed by history and radiographic studies), spondylolisthesis, trauma, (i.e., fracture or dislocation), deformity or curvature (i.e., scoliosis, and/or lordosis), tumor, stenosis, pseudoarthrosis, or failed previous fusion that warrant the use of a non-cervical spinal fixation device intended for use as a pedicle screw fixation system or sacral/iliac screw fixation system. Pedicle screw fixation is limited to skeletally mature patients and for use with autograft and/or allograft.
The Polaris Spinal System may be used with the instruments in the AccuVision Minimally Invasive Spinal Exposure System to provide the surgeon with a minimally invasive approach for posterior spinal surgery.
The dominos in the Polaris Spinal System can be used to connect the Polaris Spinal System to the Altius Spinal System, the Array Spinal System, the Biomet Omega21 Spinal System, or the Synergy Spinal System to achieve additional levels of fixation. Please refer to the individual system's Package Insert for a list of the indications for use for each system.
The Polaris Spinal System is a non-cervical spinal fixation device. The system includes screws, various types and sizes of rods, locking nuts, hooks, lateral connectors, plugs, fixation washers, rod connectors/dominos and various cross connectors. Various instruments are also available for use by the surgeon to facilitate implantation of the device. This submission is a line extension to Polaris Spinal System to add other styles of multiaxial screw, including a screw with translating feature, and to provide the Polaris Spinal System sterile packed.
The provided document describes a 510(k) summary for the Biomet Spine Polaris Spinal System - Translation Screw, which is a non-cervical spinal fixation device.
Here's an analysis of the acceptance criteria and study information:
1. Table of Acceptance Criteria and Reported Device Performance
Acceptance Criteria | Reported Device Performance |
---|---|
Mechanical Testing: | |
Static compression bending per ASTM 1717-11a | Met all mechanical test requirements based on worst-case construct testing. |
Static torsion per ASTM 1717-11a | Met all mechanical test requirements based on worst-case construct testing. |
Dynamic compression bending fatigue per ASTM 1717-11a | Met all mechanical test requirements based on worst-case construct testing. |
Substantial Equivalence: | Reported as substantially equivalent to predicate devices regarding material, design, operational principles, intended use, indications, and fundamental technology. |
2. Sample Size Used for the Test Set and Data Provenance
The document does not specify a "test set" in the context of clinical data for algorithmic performance. The testing performed was mechanical testing of the device hardware.
- Test Set Sample Size: Not applicable in the context of clinical data or algorithm performance.
- Data Provenance: Not applicable. The "data" comes from physical mechanical tests conducted on the device components.
3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts
Not applicable. Ground truth, in the context of algorithm performance, typically refers to expert-validated clinical diagnoses or outcomes. This submission focuses on mechanical performance data of a physical medical device.
4. Adjudication Method for the Test Set
Not applicable. Adjudication methods are relevant for resolving discrepancies in expert opinions for clinical ground truth. Mechanical testing has well-defined pass/fail criteria per ASTM standards.
5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done, If So, What Was the Effect Size of How Much Human Readers Improve with AI vs. Without AI Assistance
No, an MRMC comparative effectiveness study was not done. This device is a physical spinal fixation system, not an AI or imaging diagnostic tool.
6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done
No, a standalone algorithm performance study was not done. This submission is for a physical medical device.
7. The Type of Ground Truth Used (Expert Consensus, Pathology, Outcomes Data, etc.)
The "ground truth" for the mechanical testing was the established pass/fail criteria defined by ASTM 1717-11a standard test methods for static and fatigue testing of spinal implant constructs in a vertebrectomy model.
8. The Sample Size for the Training Set
Not applicable. This device is a physical implant, not an AI algorithm requiring a training set.
9. How the Ground Truth for the Training Set Was Established
Not applicable, as there is no training set for a physical device.
Ask a specific question about this device
(30 days)
K0902037
The LifeBed Network Patient Vigilance System is intended for use with adult patients by health care professionals in the continuous measurement of heart rate, respiratory rate, and as an integral part of fall prevention protocols.
The LifeBed Network Patient Vigilance System ("LifeBed Network") is composed of a bedside unit (either a LifeBed Display or Vigilance Display, generically referred to as a Display), a sensor array packaged in the form of a mattress coverlet, and the LifeBed Controller. The LifeBed Displays are optionally equipped with a wireless adapter. The LifeBed Network Patient Vigilance System collects, stores, and reports patient medical and status information. Each Display is a bedside unit that collects and processes data from a sensor array on the patients' bed. It reports patient heart rate, respiratory rate, whether the patient is in bed, and any alerts. The configuration of a Display may be changed on the Display or remotely from the LifeBed Network. The Displays are connected to the LifeBed Controller via a TCP/IP network which is provided by the customer. This network may be wired (using Ethernet), or wireless (using WiFi) as required by the facility. The hub of the LifeBed Network is the LifeBed Controller'), which serves one unit (e.g., a hospital ward). From the Controller a user has access to the collected information and the ability to view and modify the configuration of connected Displays. The LifeBed Network provides an interface which may be used to integrate with a Clinical Information System (CIS). When used in this way an auxiliary adapter isolates and translates the protocols between the LifeBed Network and the CIS. This interface allows the collected data and stored configuration for the Displays. The LifeBed Network centralizes information access and configuration for a group of Displays. It DOES NOT change the manner in which alerts are communicated by the Display to the Nurse Call system of the hospital. All user access to the LifeBed Network is constrained by a security system. Each user is authenticated by a user name and password and their ability to access data is controlled by their assigned role.
This document is a 510(k) premarket notification for an update to the LifeBed Network Patient Vigilance System. The 510(k) summary explicitly states that no clinical trials were performed in support of this submission. Therefore, it does not contain information about acceptance criteria, device performance from clinical studies, sample sizes, ground truth establishment, or expert involvement as requested.
The submission focuses on software changes to the LifeBed Controller; it relies on previous testing for substantial equivalence. The document clearly states:
- "No changes were made to the LifeBed Patient Vigilance System (KO82366) which contains the algorithm and control of the patient interface. The LifeBed Network only adds the networking and data collection capabilities. No clinical trials were performed in support of this submission." (Section labeled "Clinical Testing")
- "All verification and validation testing conducted demonstrate that this release of the LifeBed Network Patient Vigilance System is substantially equivalent to the prior release (K092037)." (Conclusion)
Therefore, I cannot provide the requested information from the provided text.
Ask a specific question about this device
Page 1 of 1