510(k) Data Aggregation

    K Number
    K183024

    Validate with FDA (Live)

    Date Cleared
    2020-02-13

    (469 days)

    Product Code
    Regulation Number
    872.3640
    Panel
    Dental
    Age Range
    All
    Predicate For
    Why did this record match?
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Implacil Implant System is intended for placement in the maxillary or mandibular arch to provide support for single-unit and/or multi-unit restorations. When a one-stage surgical is applied, the Implacil Implant System is intended for immediate loading when good primary stability is achieved and with appropriate occlusal loading.

    Device Description

    Implacil Implant System is composed of three implant lines that are divided according to the implant-toabutment interface: External Hex (HE), Internal Hex (HI) and Morse Taper AR Due Cone (CM AR). HE and HI lines are composed of tissue-level implants while CM AR line of bone-level implants. Each implant line is composed of implants and related prosthetic components available in multiple designs (temporary, screwed, cementable, angled, straight, UCLA, ball).

    HE implant line implants are available in two root-forms designs: conical (tapered) and cylindrical. Conical implants are available in three diameters and platforms (3.5, 4.0 and 5.0 mm) and five lengths (7.0, 9.0, 11.0, 13.0 and 15.0 mm). Cylindrical implants are provided in four diameters (3.3, 3.75, 4.0 and 4.75 mm), three platforms (3.5, 4.0 and 5.0 mm) and five lengths (8.0, 10.0, 11.5, 13.0 and 15.0 mm). HE cylindrical implants of diameters 3.75 and 4.0 share the same platform of 4.0 mm.

    HI implant line implants are available in two root-forms designs: conical (tapered) and cylindrical. Conical implants are available in three diameters and platforms (3.5, 4.0 and 5.0 mm). For diameters 3.5 and 4.0 are available in the lengths 8.0, 9.0, 11.0, 13.0 and 15.0 mm, and for diameters 5.0 are available in the lengths 7.0, 9.0, 11.0, 13.0 and 15.0 mm. Cylindrical implants are provided in four diameters (3.3, 3.75, 4.3 and 4.75 mm) and three platforms (3.5, 4.0 and 5.0 mm). For diameter 3.3 are available in the lengths 8.0, 9.0, 11.0, 13.0 and 15.0 mm, and for diameters 3.75 and 4.3 are available in the lengths 7.0, 9.0, 11.0, 13.0 and 15.0 mm. HI cylindrical implants of diameters 3.75 and 4.3 share the same platform of 4.0 mm.

    CM AR implant line implants are available in conical root-form design only, in four diameters (3.5, 4.0, 4.5 and 5.0 mm) and five lengths (7.0, 9.0, 11.0, 13.0 and 15.0 mm).

    lmplacil implants are made of commercially pure titanium (ASTM F67). Implacil prosthetic components are made of commercially pure titanium (ASTM F67) or titanium alloy (ASTM F136). Implant System screws (abutment screw, UCLA screws and coping screws) are made of titanium alloy (ASTM F136).

    The subject device abutments components mate exclusively with the subject implants of the same line (HJ, HE, CM AR).

    AI/ML Overview

    The provided document is a 510(k) Summary for a dental implant system. It outlines the device description, intended use, and a comparison to predicate and reference devices to demonstrate substantial equivalence. Crucially, this document does not contain information about acceptance criteria or a study proving the device meets specific acceptance criteria in the context of AI/ML-based medical devices.

    The "Performance Data" section explicitly states: "No clinical data were included in this submission." Instead, it lists non-clinical data for physical and material properties of the dental implants, such as sterilization validation, shelf life testing, biocompatibility, and mechanical performance (fatigue and torsional loading).

    Therefore, I cannot provide the requested information regarding acceptance criteria, study details, sample sizes, expert involvement, ground truth establishment, or MRMC comparative effectiveness studies, as these types of studies are not described in this 510(k) submission for a traditional medical device (dental implants), which are not a software device or AI/ML-based device.

    If you are looking for information on acceptance criteria and study data for AI/ML-based medical devices, you would need to consult a different type of FDA submission, specifically for software as a medical device (SaMD) or AI-enabled medical devices, where such performance data would be critical for demonstrating safety and effectiveness.

    Ask a Question

    Ask a specific question about this device

    K Number
    K182620

    Validate with FDA (Live)

    Date Cleared
    2019-01-18

    (116 days)

    Product Code
    Regulation Number
    872.3640
    Panel
    Dental
    Age Range
    All
    Predicate For
    Why did this record match?
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use
    • HE (External Hex) implants, stock abutments and stock copings (originally cleared per K101207)
    • Titamax CM (Cone Morse) implants, stock abutments and stock copings (originally cleared per K101945 ● and extended per K123022, K133696 and K150199)
    • Titamax CM (Cone Morse) EX implants, stock abutments and stock copings (originally cleared per K101945 and extended per K123022, K133696 and K150199)
    • Alvim CM (Cone Morse) implants, stock abutments and stock copings (originally cleared per K101945 and ● extended per K123022, K133696 and K150199)
    • CM (Cone Morse) Drive implants, stock abutments and stock copings (originally cleared per K123022 and extended per K133696, K150182 and K150199)
    • GM (Grand Morse) Titamax implants, stock abutments and stock copings (originally cleared per K163194)
    • GM (Grand Morse) Drive implants, stock abutments and stock copings (originally cleared per K163194) ●
    • . GM (Grand Morse) Helix implants, stock abutments and stock copings (originally cleared per K163194)

    The implants, abutments and copings of the Neodent Implant System are intended to be surgically placed in the bone of the upper or lower jaw to provide support for prosthetic devices such as artificial teeth, to restore chewing function. It may be used with single-stage or two-stage procedures, for single or multiple unit restorations, and may be loaded immediately when good primary stability is achieved and with appropriate occlusal loading. Multiple tooth applications may be rigidly splinted.

    • Titamax WS implants, abutments and copings (originally cleared per K123022)
      The implants, abutments and copings of the Neodent Implant System are intended to be surgically placed in the bone of the upper or lower jaw to provide support for prosthetic devices such as artificial teeth to restore chewing function. The Neodent Implant System may be used for single or multiple unit restorations. Multiple tooth applications may be rigidly splinted.

    The Titamax WS implants are indicated for a delayed loading protocol.

    • Facility implants, stock abutments and stock copings (originally cleared per K123022 and extended per K133696 and K150199)
      The implants, abutments and copings of the Neodent Implant System are intended to be surgically placed in the bone of the upper or lower jaw to provide support for prosthetic devices such as artificial teeth to restore chewing function. The Neodent Implant System may be used for single or multiple unit restorations. Multiple tooth applications may be rigidly splinted.

    The Facility implants of the Neodent Implant System are indicated for replacement of maxillary lateral incisors, mandibular incisors or retention of overdentures.

    • Zygomatic implants, stock abutments and stock copings (originally cleared per K141777)
      The Zygomatic implants, abutments and copings of the Neodent Implant System are intended to be surgically placed in the bone of the upper or lower jaw to provide support for prosthetic devices such as artificial teeth, to restore chewing function. It may be used with single-stage or two-stage procedures, for single or multiple unit restorations, and may be loaded immediately when good primary stability is achieved and with appropriate occlusal loading. Multiple tooth applications may be rigidly splinted.

    The Zygomatic implants of the Neodent Implant System are indicated for surgical installation in the zygoma region in cases of severe jaw resorption in order to restore patient esthetics and chewing function. Zygomatic implants are recommended for the posterior (pre-molar/ molar) region, one implant on each side, with at least two standard dental implants in the anterior region to support a fixed restoration.

    • CM (Cone Morse) Titanium Base Abutment component of two-piece patient-specific abutment (originally cleared per K150367 and extended per K153624)
    • o GM (Grand Morse) Exact Titanium Base Abutment component of two-piece patient-specific abutment (originally cleared per K163194)

    The implants, abutments and copings of the Neodent Implant System are intended to be surgically placed in the bone of the upper or lower jaw to provide support for prosthetic devices such as artificial teeth, to restore chewing function. It may be used with single-stage or two-stage procedures, for single or multiple unit restorations, and may be loaded immediately when good primary stability is achieved and with appropriate occlusal loading. Multiple tooth applications may be rigidly splinted.

    The Titanium Base Abutment is a titanium base placed onto Neodent dental implants to provide support for customized prosthetic restorations. The Titanium Base is used with a coping and crown, or crown alone, and is indicated for cement-retained single or multi-unit restorations, or screw-retained single restorations.

    All digitally designed copings and/or crowns for use with the Neodent Titanium Base Abutment are intended to be sent to Straumann for manufacture at a validated milling center.

    • CM (Cone Morse) Titanium Base for CEREC Abutment component of two-piece patient-specific abutment (originally cleared per K160964)
      The implants, abutments and copings of the Neodent Implant System are intended to be surgically placed in the bone of the upper or lower jaw to provide support for prosthetic devices such as artificial teeth, to restore chewing function. It may be used with single-stage or two-stage procedures, for single or multiple unit restorations, and may be loaded immediately when good primary stability is achieved and with appropriate occlusal loading. Multiple tooth applications may be rigidly splinted.

    The Titanium Base for CEREC Abutment is a titanium base placed onto Neodent dental implants to provide support for customized prosthetic restorations. They are used with a coping and crown, or crown alone, and is indicated for cement-retained single or multi-unit restorations, or screw-retained single restorations.

    All digitally designed copings and/or crowns for use with the Neodent Titanium Base for CEREC Abutment are to be designed using Sirona inLab software or Sirona CEREC Software and manufactured using a Sirona CEREC or inLab MC X or MC XL milling unit.

    • CM Preface Abutment component of patient-specific abutment (originally cleared per K150367)
      The implants, abutments and copings of the Neodent Implant System are intended to be surgically placed in the bone of the upper or lower jaw to provide support for prosthetic devices such as artificial teeth, to restore chewing function. It may be used with single-stage or two-stage procedures, for single or multiple unit restorations, and may be loaded immediately when good primary stability is achieved and with appropriate occlusal loading. Multiple tooth applications may be rigidly splinted.

    The Preface Abutment is a titanium abutment to be used in fabricating a full custom abutment placed onto Neodent dental implants to provide support for customized prosthetic restorations. The Preface Abutments is indicated for screw-retained single restorations or cement-retained single or multi-unit restorations.

    All digitally designed restorations for use with the Neodent Preface Abutment are intended to be sent to Straumann for manufacture at a validated milling center.

    • GM (Grand Morse) ProPEEK Abutments (originally cleared per K163194) ●
    • CM (Cone Morse) ProPEEK Abutments (originally cleared per K170080)

    The implants, abutments and copings of the Neodent Implant System are intended to be surgically placed in the bone of the upper or lower jaw to provide support for prosthetic devices such as artificial teeth, to restore chewing function. It may be used with single-stage or two-stage procedures, for single or multiple unit restorations, and may be loaded immediately when good primary stability is achieved and with appropriate occlusal loading. Multiple tooth applications may be rigidly splinted.

    The Pro PEEK Abutment is indicated to be used on Neodent implants to provide temporary support for prosthesis structure for up to 6 months.

    • Neodent Implant for Orthodontic Anchor (originally cleared per K102769) ●
      The Neodent Implant for Orthodontic Anchor is a surgical device in the form of a temporary screw used as an aid in orthodontic movement procedures.

    • . The Neodent Graft Screw (originally cleared per K103084)
      The Neodent Graft Screw is an implantable device used for fixation of bone blocks for the regeneration of bone in the oral cavity. The product is intended for temporary use only.

    Device Description

    Not Found

    AI/ML Overview

    The provided text is a 510(k) Premarket Notification from the FDA regarding the MRI compatibility of various Neodent Implant System devices. It primarily focuses on defining the indications for use for different components of the implant system and does not contain information about acceptance criteria, device performance studies, sample sizes, expert qualifications, or ground truth establishment for an AI/Medical Device.

    The document is a regulatory approval letter and descriptive indications for use, not a study report or technical performance assessment. Therefore, I cannot extract the requested information from this text.

    Ask a Question

    Ask a specific question about this device

    K Number
    K173902

    Validate with FDA (Live)

    Date Cleared
    2018-04-20

    (119 days)

    Product Code
    Regulation Number
    872.3630
    Panel
    Dental
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    Why did this record match?
    Reference Devices :

    K101207, K121843, K133696

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Neodent Implant System is intended to be surgically placed in the bone of the upper or lower jaw to provide support for prosthetic devices such as artificial teeth, to restore chewing function. It may be used with single-stage or two-stage procedures, for single or multiple unit restorations, and may be loaded immediately when good primary stability is achieved and with appropriate occlusal loading.

    Device Description

    The purpose of this submission is to expand the Neodent Implant System - GM Line components cleared under K163194, which included dental implants with a Morse taper abutment interface (called Grand Morse, or GM), mating abutment screws, and other associated components. This submission includes the GM Exact Co-Cr Abutment for Crown and the GM Equator Attachment. The subject device components mate exclusively with the GM implants cleared in K163194.

    The GM Exact Co-Cr Abutment for Crown is provided in three sizes to fit the prosthetic diameter sizes of the GM implants cleared in K163194. The subject device abutments consist of a Co-Cr alloy component that interfaces directly to the implant and a cvlinder of polyoxymethylene (POM) for fabrication of a cast prosthesis by a burn-out technique. The subject device abutments are provided in three (3) sizes to match the various platforms of the mating GM implants: 3.5/3.75 mm platform diameter, 4.0/4.3 mm platform diameter, and 5.0/6.0 mm platform diameter. The subject device abutments are for single-unit prosthetic restorations only. The final prosthetic restoration may be cement-retained or screw-retained.

    The GM Equator Attachment abutments are straight, ball-type abutments for the attachment of full or partial overdentures. GM Equator Attachments are provided in five gingival heights (1.5, 2.5, 3.5, 4.5, and 5.5 mm), and may be used with all GM implant diameter sizes cleared in K163194.

    The subject device GM Exact Co-Cr abutments are made of Co-Cr alloy conforming to ASTM F1537, with mating abutment screws made of titanium alloy conforming to ASTM F136. The GM Equator Attachments made of titanium alloy conforming to ASTM F136 with a TiN coating identical to the TiN coating cleared in K133696.

    AI/ML Overview

    This document describes the Neodent Implant System - GM Line, specifically introducing new abutment components. Based on the provided text, the acceptance criteria and study information are focused on demonstrating substantial equivalence to predicate devices rather than proving performance against specific acceptance criteria for a novel device. The document mainly uses non-clinical performance data (biocompatibility, sterilization validation, shelf-life, bacterial endotoxin testing) to support this claim.

    Here's a breakdown of the requested information based on the provided text:

    1. A table of acceptance criteria and the reported device performance

    The document does not explicitly state "acceptance criteria" in terms of specific performance metrics for the device (e.g., accuracy, sensitivity, specificity, or clinical outcomes). Instead, the performance data presented is aimed at demonstrating the device's characteristics are comparable to predicate devices. The "performance" is primarily shown through a comparison of technological characteristics and non-clinical testing results to existing standards and predicate devices to establish substantial equivalence.

    Acceptance CriterionReported Device Performance (as demonstrated by comparison/testing)
    BiocompatibilityDemonstrated through testing according to ISO 10993-5, ISO 10993-12, and ISO 10993-18 for the GM Exact Co-Cr Abutment for Crown. Biocompatibility of titanium alloy components referenced to K163194 and K133696.
    Sterilization ValidationPerformed according to ISO 17665-1, ISO TS 17665-2, ISO 11135, and ISO 10993-7.
    Sterile Barrier Shelf LifePerformed on real-time aged samples (referenced from K163194), including seal strength, helium leak detection, and sterility testing of package contents.
    Bacterial Endotoxin TestingPerformed using methods described in AAMI / ANSI ST72.
    Material CompositionGM Exact Co-Cr Abutment for Crown: Co-Cr alloy conforming to ASTM F1537; Mating abutment screws: Ti-6Al-4V alloy, ASTM F136. GM Equator Attachments: Ti-6Al-4V alloy, ASTM F136 with TiN coating. (Demonstrates equivalence or compliance to accepted material standards).
    Mechanical Design/InterfaceSubject device abutments have identical implant interface design to Neodent GM implants cleared in K163194. Morse taper connection with internal hex prosthetic indexing. Straight (0°) angles only.
    Intended UseFunctional and esthetic rehabilitation of the edentulous mandible or maxilla, consistent with predicate devices.

    2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    The document focuses on non-clinical testing and refers to standards. It does not specify "sample sizes" in the context of clinical test sets or patient data.

    • Sample size for testing: Not explicitly stated for each non-clinical test (e.g., number of abutments tested for biocompatibility or number of packages for shelf-life). However, regulatory standards (like ISO 10993, ISO 17665-1) inherently require specific sample sizes for valid testing.
    • Data Provenance: The manufacturer is JJGC Indústria e Comércio de Materiais Dentários S.A. based in Curitiba, Paraná, Brazil. The non-clinical testing would presumably have been conducted by or commissioned by this company, possibly in Brazil or by certified labs internationally. The data is non-clinical, not from human subjects, hence "retrospective or prospective" does not directly apply in the clinical sense.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

    This information is not applicable. The provided document details the submission for new components of an existing dental implant system, demonstrating substantial equivalence through non-clinical testing (biocompatibility, sterilization, material specifications). It does not involve diagnostic imaging or other subjective assessments that would require expert consensus for ground truth.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    This information is not applicable, as there is no clinical test set requiring adjudication in the context of diagnostic interpretation or outcomes.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    This information is not applicable. The device is a dental implant system (hardware), not a diagnostic AI device. Clinical data was explicitly not submitted: "Clinical data were not submitted in this premarket notification." Therefore, no MRMC study or AI-related comparative effectiveness was performed or is relevant to this submission.

    6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done

    This information is not applicable. The device is a dental implant system (hardware), not an algorithm or AI device.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

    For the non-clinical performance data, the "ground truth" or reference for evaluating compliance consists of:

    • International Standards: ISO 10993 series (biocompatibility), ISO 17665-1, ISO TS 17665-2, ISO 11135, ISO 10993-7 (sterilization), AAMI / ANSI ST72 (bacterial endotoxin testing).
    • Material Standards: ASTM F1537 (Co-Cr alloy), ASTM F136 (Ti-6Al-4V alloy).
    • Predicate Device Characteristics: Comparison of physical dimensions, design principles, intended use, and materials to established predicate devices (K163194, K101207, K121843, K133696) to demonstrate substantial equivalence.

    There is no "ground truth" in the sense of clinical expert consensus, pathology, or outcomes data, as clinical data was not submitted.

    8. The sample size for the training set

    This information is not applicable. This is a submission for a physical medical device (dental implant components), not a machine learning or AI algorithm that requires a "training set."

    9. How the ground truth for the training set was established

    This information is not applicable, as there is no training set for an algorithm.

    Ask a Question

    Ask a specific question about this device

    K Number
    K133510

    Validate with FDA (Live)

    Date Cleared
    2014-02-24

    (102 days)

    Product Code
    Regulation Number
    872.3640
    Panel
    Dental
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    Why did this record match?
    Reference Devices :

    K101207

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Neodent Implant System is intended to be surgically placed in the bone of the upper or lower jaw to provide support for prosthetic devices such as artificial teeth, to restore chewing function. It may be used with single-stage or two-stage procedures, for single or multiple unit restorations, and may be loaded immediately when good primary stability is achieved and with appropriate occlusal loading. Multiple tooth applications may be rigidly splinted.

    Device Description

    The purpose of this submission is to expand the Neodent Implant System components cleared under K101207, which included dental implants with an external hex abutment interface. The external hex additions include Titamax Smart EX, and Drive Smart implants that are provided with a new implant mount that can be used as a temporary abutment. Additional straight, angled and ball-type abutments with external hex interfaces are also included in this submission.

    All of the subject device external implants are threaded, self-tapping, root form, endosseous dental implants with a grit blasted and acid etched endosseous surface. All implants have a smooth machined collar on the transgingival surface. The Titamax Smart implants are provided in five endosseous thread diameters (3.3, 3.75, 4.0, 4.5, 5.0 mm), three platform diameters (3.3, 4.1, 5.0 mm), and five overall lengths (9, 11, 13, 15, 17 mm). The Titamax Smart EX implants are provided in two endosseous thread diameters (3.75, 4.0 mm), one platform diameter (4.1 mm), and six overall lengths (9, 11, 13, 15, 17, 19 mm). The Drive Smart implants are provided in three endosseous thread diameters (3.5, 4.3, 5.0 mm), three platform diameters (3.3, 4.3, 5.0 mm), and five overall lengths (8, 10, 11.5, 13, 16 mm). All implants are made from commercially pure titanium conforming to ASTM F67.

    This submission includes the following abutments: the Smart Mount in three platform diameters (3.3, 4.1/4.3, 5.0 mm); Conical Abutments (platform diameter 3.3 mm); Angled Mini Conical abutments angled 17° or 30°, in four platform diameters (3.3, 4.1, 4.3, 5.0 mm), and gingival heights from 2 mm to 5 mm; SF Universal Post abutments (platform diameter 4.5 mm); Angled Post abutments angled 17° or 30° in four platform diameters (3.3. 4.1. 4.3, 5.0 mm), and gingival heights from 2 mm to 5 mm; and Equator Attachments for overdentures or partial dentures in three platform diameters (4.1, 4.3, 5.0 mm) and gingival heights from 2 mm to 5 mm. All abutments are made from titanium alloy conforming to ASTM F136.

    AI/ML Overview

    The Neodent Implant System is a dental implant system designed to support prosthetic devices like artificial teeth to restore chewing function.

    1. A table of acceptance criteria and the reported device performance

    CriteriaAcceptance Criteria (Predicate Device K101207, K101945, K123022)Reported Device Performance (Neodent Implant System)
    BiocompatibilityDemonstratedDemonstrated
    Engineering AnalysisDemonstratedDemonstrated
    Dimensional AnalysisDemonstratedDemonstrated
    Static and Dynamic Compression-Bending Testing (ISO 14801)Demonstrated equivalence or superiority to predicate devicesDemonstrated equivalence to predicate devices
    Similar Design and DimensionsSame range of physical dimensions (diameter, length of implants; diameter, height, angle of abutments)Similar design and dimensions, same range of physical dimensions
    Same Materials UsedImplants: Commercially pure titanium (ASTM F67); Abutments: Titanium alloy (ASTM F136)Implants: Commercially pure titanium (ASTM F67); Abutments: Titanium alloy (ASTM F136)
    Same SurfaceGrit blasted and acid etched endosseous surface; smooth machined collarGrit blasted and acid etched endosseous surface; smooth machined collar
    Similar PackagingSimilar to predicate devicesSimilar to predicate devices
    Similar Sterilization MethodsSame materials and processes as predicate devicesSame materials and processes as predicate devices

    2. Sample size used for the test set and the data provenance

    • Sample Size: Not explicitly stated in the provided text. The submission refers to "non-clinical data" for testing.
    • Data Provenance: The document does not specify the country of origin for the non-clinical data. It is a retrospective comparison to previously cleared predicate devices.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    This information is not applicable to this submission. The "ground truth" for demonstrating substantial equivalence relied on engineering and material characteristic comparisons, not expert interpretation of clinical data or images.

    4. Adjudication method for the test set

    Not applicable. This was a non-clinical equivalence study, not a clinical trial requiring adjudication of patient outcomes or expert diagnoses.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    Not applicable. This submission is for a physical medical device (dental implants and abutments), not an AI-powered diagnostic or assistive tool.

    6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done

    Not applicable. This is not an algorithm or AI-based device.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

    The ground truth used for demonstrating substantial equivalence was based on engineering standards, material specifications, and physical performance testing (e.g., fatigue testing according to ISO 14801), comparing the subject device's attributes to those of legally marketed predicate devices.

    8. The sample size for the training set

    Not applicable. This is a medical device, not a machine learning model, so there is no concept of a "training set" in this context.

    9. How the ground truth for the training set was established

    Not applicable. There is no training set for this device.