    DESS Dental Smart Solutions abutments are intended to be used in conjunction with endosseous dental implants in the maxillary or mandibular arch to provide support for prosthetic restorations.

    All digitally designed custom abutments for use with DESS Bases or Pre-milled Blanks are to be sent to a Terrats Medical validated milling center for manufacture, or to be designed and manufactured according to the digital dentistry workflow. The digital dentistry workflow integrates multiple components: scan files from intra-oral and lab (desktop) scanners, CAD software, CAM software, ceramic material, milling machine, and associated tooling and accessories.

    Device Description

    The purpose of this submission is to expand the DESS Dental Smart Solutions abutment system cleared under K221301 and K240982 to allow additional options of zirconia material, scanners, CAM software, and milling machines to the digital dentistry workflow. The subject devices are to be sent to Terrats Medical validated milling centers for manufacture, or to be designed and manufactured via a digital dentistry workflow. The digital dentistry workflow integrates multiple components: scan files from intra-oral and lab (desktop) scanners, CAD software, CAM software, titanium and ceramic material, milling machine, and associated tooling and accessories. There are no changes to the abutment design, implant compatibilities, or design parameters. All part numbers have been cleared for manufacturing via a validated milling center and digital dentistry workflows (also referred to as point of care) under K221301 and K240982.

    The subject device DESS Dental Smart Solutions abutments provide a range of prosthetic solutions for dental implant restoration. DESS abutments are offered in a variety of connection types to enable compatibility with currently marketed dental implants. All abutments are provided non-sterile, and each abutment is supplied with the appropriate abutment screw (if applicable) for attachment to the corresponding implant.

    Subject device Base Abutments are designed for fabrication of a patient-specific CAD/CAM zirconia superstructure on which a crown may be placed. They are two-piece abutments for which the second part (or top half) is the ceramic superstructure. They also may be used for support of a crown directly on the abutment.

    All patient-specific custom abutment fabrication for Base Abutments and Pre-milled (Blank) Abutments is by prescription on the order of the clinician. The subject device Pre-milled (Blank) Abutments and all zirconia superstructures for use with the subject device Ti Base Interface, DESS Aurum Base, ELLIPTIBase, and DESS C-Base will be manufactured using a validated milling center or a digital dentistry workflow. A validated milling center will be under FDA quality system regulations. The digital dentistry workflow scans files from intra-oral and lab (desktop) scanners, CAD software, CAM software, titanium and ceramic material, milling machine and associated tooling and accessories.

    The digital dentistry workflow uses scan files from intra-oral and lab (desktop) scanners, CAD software, CAM software, ceramic material, milling machine and associated tooling and accessories.

    AI/ML Overview

    The provided 510(k) summary for DESS Dental Smart Solutions focuses on demonstrating substantial equivalence to predicate devices for dental implant abutments. It primarily addresses the expansion of compatible materials, scanners, CAM software, and milling machines within an existing digital dentistry workflow. The document does not describe an AI/ML-based device that would typically have acceptance criteria related to diagnostic performance.

    Therefore, many of the requested items related to AI/ML device performance (like acceptance criteria for diagnostic metrics, sample size for test sets, data provenance, expert qualifications, adjudication methods, MRMC studies, standalone performance, and training set details) are not applicable to this submission.

    The acceptance criteria and supporting "study" (non-clinical data) for this device are related to its mechanical performance, biocompatibility, and integration within the digital workflow, demonstrating that the expanded components maintain the safety and effectiveness of the previously cleared predicate devices.

    Here's a breakdown based on the information provided and the non-applicability of AI/ML-specific questions:

    1. A table of acceptance criteria and the reported device performance

    Since this is not an AI/ML diagnostic device, the acceptance criteria are not in terms of traditional diagnostic metrics (sensitivity, specificity, AUC). Instead, they are related to material properties, mechanical integrity, and the digital workflow's accuracy.

    Acceptance Criteria CategoryReported Device Performance (Summary from Submission)
    Mechanical Integrity / Compatibility- Fatigue testing of OEM implant bodies with patient-specific abutments made at worst-case angled conditions. (Implies successful completion to similar or better standards than predicate.)
    Biocompatibility- Biocompatibility testing according to ISO 10993-5 and ISO 10993-12. (Implies successful demonstration of biocompatibility.)
    Sterilization- Sterilization validation according to ISO 17665-1, ISO 17665-2, and ISO 14937. (Implies successful validation for sterilization.)
    CAD Design Restrictions- Software verification included testing of restrictions that prevent design of components outside of the stated design parameters. - Abutment design library validated to demonstrate established design limitations are locked and cannot be modified by the user. (Implies successful implementation and verification of design constraints.)
    CAM Restriction Zones / Manufacturing Accuracy- Validation testing of CAM restriction zones conducted, including verification to show avoidance of damage or modifications of the connection geometry, and locking of restriction zones from user editing in CAM software. (Implies successful validation to ensure manufacturing precision and prevent damage.)
    Material Conformance- Zirconia materials conform to ISO 6872. - Titanium alloy conforms to ASTM F136. - Co-Cr-Mo alloy conforms to ASTM F15337. (Implies materials meet standards.)
    Physical Dimensions- Device encompasses the same range of physical dimensions as the predicate device. (Implies dimensional equivalence.)

    2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    • Sample Size: Not explicitly stated in terms of a "test set" for diagnostic performance. The validation involved physical testing of components (e.g., fatigue testing) and software verification. The specific number of abutments or digital design instances used for these non-clinical tests is not detailed in this summary.
    • Data Provenance: Not applicable in the context of patient data for an AI/ML device. The "data" here refers to engineering and material testing results, likely conducted in controlled lab environments (implied to be in accordance with international standards like ISO and ASTM). The manufacturer is Terrats Medical SL, in Spain, so testing would likely originate from their facilities or contracted labs.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

    Not applicable. This is not a diagnostic device requiring expert interpretation for ground truth. The "ground truth" for this device relates to engineering specifications and material science.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    Not applicable. This is not a diagnostic device involving expert review adjudication.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    Not applicable. This device is not an AI-assisted diagnostic tool.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    Not applicable. This is not an AI/ML algorithm.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

    For this device, the "ground truth" is based on:

    • Engineering Specifications: Defined design parameters (e.g., minimum wall thickness, post height, angulation limits).
    • Material Standards: Conformance to international standards such as ASTM F136, ISO 6872.
    • Benchmarking/Predicate Equivalence: Performance is assessed against established performance of the predicate devices and OEM implant systems.
    • Software Validation Logic: Verification that software correctly enforces design rules and CAD/CAM restrictions.

    8. The sample size for the training set

    Not applicable. This device does not involve a machine learning training set.

    9. How the ground truth for the training set was established

    Not applicable. This device does not involve a machine learning training set.

    Ask a Question

    Ask a specific question about this device

    K Number
    K190079

    Validate with FDA (Live)

    Device Name
    ArgenZ HT+
    Date Cleared
    2019-04-12

    (86 days)

    Product Code
    Regulation Number
    872.6660
    Panel
    Dental
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    Why did this record match?
    Reference Devices :

    K071410

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    ArgenZ HT+ (high translucent plus) zirconia can be used for the production of full contour and substructures restorations up to a full arch.

    Device Description

    ArgenZ HT+ is a pressed Yttria stabilized Zirconia with primary oxides of Hafnium and Aluminum with naturally occurring trace oxides used to produce various shades.

    AI/ML Overview
    1. Acceptance Criteria and Reported Device Performance
    Acceptance Criteria (from BS EN ISO 6872:2015 via predicate device or standard)Reported Device Performance (ArgenZ HT+)
    Flexural strength: ≥ 800 MPa1348 MPa (ArgenZ HT+ unshaded)
    1172 MPa (ArgenZ HT+ Shade A4)
    1143 MPa (ArgenZ HT+ Shade OM1)
    Chemical solubility: < 100 µg/cm²66.11 µg/cm² (ArgenZ HT+ unshaded)
    3.10 µg/cm² (ArgenZ HT+ Shade A4)
    3.10 µg/cm² (ArgenZ HT+ Shade OM1)
    Coefficient of Thermal Expansion: (Not explicitly stated as a pass/fail criterion, but a comparative value)10.3 µm/m-C° (25-500°C)
    Cytotoxicity: No evidence of cell lysis or toxicityMet the requirements of the test.
    1. Sample Size and Data Provenance

      • The document does not explicitly state the sample size used for the mechanical and chemical property tests. The results are presented as single values (e.g., "Test Result").
      • The data provenance is not specified in terms of country of origin or whether the studies were retrospective or prospective. It implies the performance testing was conducted by or for The Argen Corporation as part of their 510(k) submission.
    2. Number of Experts and Qualifications

      • This submission is for a dental material (zirconia), not an AI/software device that requires expert review for ground truth. Therefore, there were no experts establishing ground truth in the context of human interpretation of data. The ground truth for material properties is established by standardized laboratory testing methods.
    3. Adjudication Method

      • Not applicable for this type of device (dental material). Adjudication methods like 2+1 or 3+1 are typical for studies involving human interpretation or AI model output where consensus among experts is needed.
    4. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

      • No, an MRMC comparative effectiveness study was not conducted. This type of study is relevant for diagnostic devices where human readers interpret images, and the AI assists in this interpretation. This submission is for a dental material with physical and chemical properties.
    5. Standalone (Algorithm Only) Performance

      • Not applicable. This is not an algorithm or AI device. The performance data relates to the physical and chemical properties of the zirconia material itself.
    6. Type of Ground Truth Used

      • The ground truth for the device's performance relies on objective laboratory measurements based on established international standards (BS EN ISO 6872:2015 for mechanical/chemical properties and ISO 10993-5 for biocompatibility/cytotoxicity). This is akin to "outcomes data" in a broader sense of directly measured physical properties and biological responses, rather than expert consensus on subjective interpretations.
    7. Sample Size for the Training Set

      • Not applicable. This is a material, not a machine learning model, so there is no concept of a "training set" in this context.
    8. How Ground Truth for the Training Set Was Established

      • Not applicable, as there is no training set for a material. The "ground truth" for the material's properties is established through the standardized testing methodologies detailed above.
    Ask a Question

    Ask a specific question about this device

    K Number
    K182833

    Validate with FDA (Live)

    Date Cleared
    2019-02-08

    (122 days)

    Product Code
    Regulation Number
    872.6660
    Panel
    Dental
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    Why did this record match?
    Reference Devices :

    K071410, K150919

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    ArgenZ Liquids are intended to be used by trained dental technicians as an accessory for shading ArgenZ allzirconia, monolithic restorations for anterior and posterior dental prosthetics.

    Device Description

    ArqenZ Zirconia Coloring Liguids are dilute aqueous solutions of transition and lanthanoid metal salts used in combination with ArgenZ Zirconia material (K071410 and K150919) to properly shade teeth. Elements, which may include (but are not limited to) Co, Cr, Er, Fe, Mn, Pr, Ni, Bi, and Nd are combined with water to result in the finished product.

    AI/ML Overview

    The provided text describes Argen Corporation's ArgenZ Liquids, which are dental coloring liquids for zirconia restorations. The document is a 510(k) premarket notification for the device, seeking substantial equivalence to a predicate device.

    Here's an analysis of the acceptance criteria and study information:

    1. Table of Acceptance Criteria and Reported Device Performance

    Acceptance Criteria (Product Standard BS EN ISO 6872:2015)CriteriaReported Performance (ArgenZ Liquids)
    Flexural Strength
    ArgenZ HT+>= 800 MPa1230 MPa
    ArgenZ HT+ Shaded with ArgenZ HT Liquid (A4)>= 800 MPa1223 MPa
    ArgenZ Anterior>= 500 MPa868 MPa
    ArgenZ Anterior Shaded with ArgenZ ST Liquid (A3)>= 500 MPa765 MPa
    ArgenZ Anterior Shaded with ArgenZ ST Liquid (C4)>= 500 MPa963 MPa
    Chemical Solubility
    ArgenZ HT+< 100 ug/cm²4.16 ug/cm²
    ArgenZ HT+ Shaded with ArgenZ HT Liquid (A4)< 100 ug/cm²4.16 ug/cm²
    ArgenZ Anterior< 100 ug/cm²20.80 ug/cm²
    ArgenZ Anterior Shaded with ArgenZ ST Liquid (A3)< 100 ug/cm²10.32 ug/cm²
    ArgenZ Anterior Shaded with ArgenZ ST Liquid (C4)< 100 ug/cm²8.30 ug/cm²
    Cytotoxicity (ISO 10993-5)Reactivity grade <= 2 (mild reactivity)All test articles showed no evidence of cell lysis or toxicity (i.e. grade <= 2).

    2. Sample Size for Test Set and Data Provenance

    • Mechanical & Chemical Properties: The document does not specify the exact sample size for each test (flexural strength and chemical solubility). It mentions "test records of parent materials and representative accessories infused into the parent materials," and lists several variations (e.g., ArgenZ HT+, ArgenZ HT+ Shaded with ArgenZ HT Liquid (A4)). The provenance is not explicitly stated, but these are laboratory tests conducted by the manufacturer.
    • Cytotoxicity: The document states that "test articles for the cytotoxicity testing were selected as representative samples for each product family, and determined to be worst case scenarios for potential biocompatibility issues." Four specific liquid/zirconia combinations were selected as representative samples (ST on K150919 (A3), ST on K150919 (C4), HT on K071410 (A4), HT on K071410 (B4)). The provenance is not explicitly stated, but these are laboratory tests conducted by the manufacturer.

    3. Number of Experts and Qualifications

    This document describes a pre-market notification for a dental material, not a diagnostic device involving human interpretation. Therefore, there are no "experts" in the context of establishing ground truth for diagnostic accuracy, nor qualifications for such experts. The evaluation relies on standardized material testing.

    4. Adjudication Method

    Not applicable, as this is related to material properties testing, not human interpretation of diagnostic results.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

    Not applicable. This device is a dental coloring liquid used by trained dental technicians, not a diagnostic tool where human readers assess cases. The study focuses on material properties, not reader performance.

    6. Standalone (Algorithm Only) Performance

    Not applicable. This device is a material, not an algorithm.

    7. Type of Ground Truth Used

    The "ground truth" for the performance evaluation is based on product standards (BS EN ISO 6872:2015 for mechanical and chemical properties, and ISO 10993-5 for cytotoxicity). These standards define the acceptable limits for various material properties.

    8. Sample Size for Training Set

    Not applicable. This is a material, not a machine learning model, so there is no training set in the conventional sense. The "training" in manufacturing refers to established processes and quality control.

    9. How Ground Truth for Training Set was Established

    Not applicable, as there is no training set for an AI/algorithm. For materials, the "ground truth" for manufacturing involves adherence to established specifications, quality control procedures, and compliance with relevant industry standards and regulations.