The Labcorp Fentanyl Urine Visual Test is a lateral flow competitive immunoassay for the rapid qualitative detection of norfentanyl (fentanyl metabolite) in human urine at a cutoff of 5 ng/mL. It is intended for prescription use. For in vitro diagnostic use only.
This test provides only a preliminary result. Clinical consideration and professional judgment must be applied to any drug test result, particularly in evaluating a preliminary positive result. To confirm preliminary positive results, a more specific analytical method must be used. Gas Chromatography-Mass Spectrometry (GC-MS), Liquid Chromatography Tandem Mass Spectrometry (LC-MS/MS), and High Performance Liquid Chromatography (HPLC) are the preferred confirmatory methods.
The Labcorp Fentanyl Urine Visual Test (the "DEVICE") is a lateral flow competitive immunoassay for the rapid qualitative detection of Norfentanyl (nFEN), the primary urinary metabolite of Fentanyl, in human urine at concentrations above 5 ng/mL. It is intended for prescription use.
The single use, in vitro diagnostic DEVICE is available in a cassette format with a disposable dropper provided for sample transfer.
The DEVICE contains a test strip that gives a qualitative result for presence of Norfentanyl in human urine. The DEVICE is read visually and has labeling with instructions for interpreting test results.
N/A
Ask a specific question about this device
(318 days)
The Evidence MultiSTAT DOA Urine MultiPlex is intended for use with the Evidence MultiSTAT. The Evidence MultiSTAT is an analyser intended for the qualitative determination of parent drug molecule and metabolites of drugs in human urine at the associated cut offs.
The Evidence MultiSTAT DOA Urine MultiPlex detects the following drugs at the following cut offs:
| Analyte | Analyte in Cut Off Material | Cut Off |
|---|---|---|
| Methamphetamine | S-(+)-Methamphetamine | 500ng/ml |
| Noroxycodone | Noroxycodone | 100ng/ml |
| Benzodiazepines 1 | Oxazepam | 200ng/ml |
| Methadone | (+)-Methadone | 300ng/ml |
| Phenobarbital | Phenobarbital | 200ng/ml |
| Tramadol | Tramadol | 200ng/ml |
| Phencyclidine | Phencyclidine | 25ng/ml |
| Buprenorphine | Norbuprenorphine | 5ng/ml |
| 6-Acetylmorphine | 6-Acetylmorphine | 10ng/ml |
| Fentanyl | Fentanyl | 1ng/ml |
| Benzodiazepines 2 | Lorazepam | 200ng/ml |
| Opiates | Morphine | 300ng/ml |
| Benzoylecgonine/Cocaine | Benzoylecgonine | 150ng/ml |
| Cannabinoids (THC) | (-)-11-nor-9-Carboxy-Δ9-THC | 50ng/ml |
| Amphetamine | S-(+)-Amphetamine | 500ng/ml |
The Evidence MultiSTAT DOA Urine Multiplex provides only a preliminary analytical test result. A more specific alternative chemical method must be used to obtain a confirmed analytical result. Gas Chromatography/Mass Spectrometry (GC/MS) and/or Liquid Chromatography/Tandem Mass Spectrometry (LC/MS/MS) and the preferred confirmatory methods. Other chemical confirmation methods are available. Clinical consideration and professional judgement should be applied to any drug of abuse test result, particularly when preliminary positive results are obtained.
For In Vitro Diagnostic use only.
FOR PRESCRIPTION USE ONLY
The Evidence MultiSTAT analyzer is a benchtop fully automated Biochip Array System. It performs simultaneous detection of multiple analytes from a single sample. The core technology is the Randox Biochip, a solid-state device containing an array of discrete test regions containing immobilized antibodies specific to different Drugs of Abuse (DOA) compound classes. A competitive chemiluminescent immunoassay is used for the DOA assays with the drug in the specimen and drug labelled with horseradish peroxidase (HRP) being in direct competition for the antibody binding sites. Increased levels of drug in a specimen will lead to reduced binding of drug labelled with HRP and thus a reduction in chemiluminescence being emitted. The light signal generated from each of the test regions on the biochip is detected by a Charge Coupled Device (CCD) camera in the Evidence MultiSTAT system which, together with the analyzer software, is used to quantify the light output and produce meaningful results.
The immunoassay processes are performed automatically in a self-contained and sealed biochip cartridge, which holds the biochips, the reagents, wash buffer and other fluids required for the test to be conducted.
Evidence MultiSTAT assays employ a qualitative reporting method. Each test sample is assayed against the provided Cut Off material of known concentration, which is used to determine the classification of the samples based on the comparison of the signal output.
The Evidence MultiSTAT DOA Urine MultiPlex (EV4393) will be supplied as a test kit comprising:
- 12 x Urine Test Cartridges
- 6 x 1 ml Urine Cut Off Material (lyophilized)
- 4 x 1 ml Urine Positive Control Material (lyophilized)
- 2 x 10 ml Reconstitution Buffer
- 1 x Batch Barcodes
Each kit is supplied with the Evidence MultiSTAT Accessory kit (EV4116) which contains:
- 12 x MultiSTAT Tip Cartridges
- 1 x Tip/Waste Cartridge
- 6 x 1000 µl Pipette Tip
- 1 x Liquid Absorber
Reagent Composition
MultiSTAT DOA Urine MultiPlex Assay Diluent
20 mM phosphate buffer, pH 7.0 containing protein, detergents, and preservatives. This is contained within the cartridge.
MultiSTAT DOA Urine MultiPlex Conjugate
20 mM Tris based buffer, pH 7.0 containing protein, preservatives, and horseradish peroxidase - labelled drug derivatives. This is contained within the cartridge.
MultiSTAT DOA Urine MultiPlex Biochip
Solid substrate containing immobilized antibody discrete test regions. This is contained within the cartridge.
MultiSTAT DOA Urine MultiPlex Wash Buffer
20 mM Tris buffered saline, pH 7.4, containing surfactant and preservatives. This is contained within the cartridge.
LUM-EV934/PX
Luminol-EV934 and Peroxide are contained within the cartridge and are mixed in a ratio of 1:1 by the analyser to give the working signal reagent
MultiSTAT DOA Urine MultiPlex Cut Off
Lyophilised, 20 mM phosphate buffer, pH 7.2 containing stabilizers, preservatives and drug concentrations at the assay cut off values (detailed in Table 3 above).
MultiSTAT DOA Urine MultiPlex Positive Control
Lyophilised, 20 mM phosphate buffer, pH 7.2 containing stabilizers, preservatives, and drug concentrations.
MultiSTAT Reconstitution Buffer
A solution at a neutral pH containing preservatives.
N/A
Ask a specific question about this device
(115 days)
The LZI Buprenorphine II Enzyme Immunoassay is intended for the qualitative and semi-quantitative determination of norbuprenorphine in human urine at the cutoff value of 5 ng/mL when calibrated against norburprenorphine. The assay is designed for prescription use with a number of automated clinical chemistry analyzers.
The semi-quantitative mode is for purposes of (1) enabling laboratories to determine an appropriate dilution of the specimen for confirmation by a confirmatory method such as gas or liquid chromatography/mass spectrometry (GC/MS or LC/MS) or (2) permitting laboratories to establish quality control procedures
The assay provides only a preliminary analytical result. A more specific alternative chemical method (e.g., gas or liquid chromatography and mass spectrometry) must be used in order to obtain a confirmed analytical result. Clinical consideration and professional judgment should be exercised with any drug of abuse test result, particularly when the preliminary test result is positive.
The LZI Buprenorphine II Enzyme Immunoassay is a homogeneous enzyme immunoassay ready-to-use liquid reagent. The assay is based on competition between drug in the sample and drug labeled with the enzyme glucose-6-phosphate dehydrogenase (G6PDH) for a fixed amount of antibody in the reagent. Enzyme activity decreases upon binding to the antibody, and the drug concentration in the sample is measured in terms of enzyme activity. In the absence of drug in the sample, norbuprenorphine-labeled G6PDH conjugate is bound to antibody, and the enzyme activity is inhibited. On the other hand, when drug is present in the sample, antibody would bind to free drug; the unbound norbuprenorphine-labeled G6PDH then exhibits its maximal enzyme activity. Active enzyme converts nicotinamide adenine dinucleotide (NAD) to NADH, resulting in an absorbance change that can be measured spectrophotometrically at 340 nm.
The LZI Buprenorphine II Enzyme Immunoassay is a kit comprised of two reagents, R₁ and R₂, which are bottled separately but sold together within the kit.
The R₁ solution contains mouse monoclonal anti-norbuprenorphine antibody, glucose-6-phosphate (G6P), nicotinamide adenine dinucleotide (NAD), stabilizers, and sodium azide (0.09%) as a preservative. The R₂ solution contains glucose-6-phosphate dehydrogenase (G6PDH) labeled with norbuprenorphine in buffer with sodium azide (0.09%) as a preservative.
N/A
Ask a specific question about this device
(31 days)
The Immunalysis SEFRIA™ Hydrocodone Oral Fluid Enzyme Immunoassay is an enzyme immunoassay with a cutoff of 30 ng/mL in neat oral fluid collected by Quantisal™ or Quantisal™ II Oral Fluid Collection Device. The assay is intended for the qualitative and semi-quantitative analysis of hydrocodone in human oral fluid to be used with clinical analyzers. This assay is calibrated against hydrocodone.
The semi-quantitative mode is for purposes of enabling laboratories to determine an appropriate dilution of the specimen for confirmation by a confirmatory method such as Gas Chromatography/Mass Spectrometry (GC-MS) or Liquid Chromatography/Tandem Mass Spectrometry (LC-MS/MS) or permitting laboratories to establish quality control procedures.
The Immunalysis SEFRIA™ Hydrocodone Oral Fluid Enzyme Immunoassay provides only a preliminary analytical test result. A more specific alternate chemical method must be used in order to obtain a confirmed analytical result. GC-MS or LC-MS/MS are the preferred confirmatory methods. Clinical consideration and professional judgment should be applied to any test result, particularly when preliminary positive results are used.
The Immunalysis SEFRIA™ Hydrocodone Oral Fluid Enzyme Immunoassay is a homogenous enzyme immunoassay with a cutoff of 30 ng/mL in neat oral fluid. The assay is intended for use in laboratories with clinical chemistry analyzers for the qualitative and semi-quantitative analysis of hydrocodone in human oral fluid collected with Quantisal™ Oral Fluid Collection Device or Quantisal™ II Oral Fluid collection device.
N/A
Ask a specific question about this device
(50 days)
The Healgen® AccuFluor Fentanyl Fluorescence Immunoassay (FIA) Test Kit-Qualitative is a fluorescence immunoassay intended for the qualitative detection of fentanyl in human urine at a cutoff concentration of 1.0 ng/mL. The assay is intended for use with Healgen® Immunofluorescence analyzer OG-H180. This in vitro diagnostic device is for prescription use only.
This assay provides only a preliminary analytical test result. A more specific alternate chemical method must be used to obtain a confirmed analytical result. Gas Chromatography-Mass Spectrometry (GC-MS) and Liquid Chromatography-Mass Spectrometry (LC-MS) are the preferred confirmatory methods. Clinical consideration and professional judgment should be applied to a Fentanyl test result, particularly when a preliminary positive result is obtained.
The Healgen® Immunofluorescence analyzer OG-H180 is a portable fluorescence instrument for in vitro diagnostic use only. The analyzer is designed to detect test results from in vitro diagnostic tests on clinical specimens. This analyzer can be used in a laboratory or point-of-care setting.
The AccuFluor Fentanyl FIA Test Kit-Qualitative is a rapid fluorescence immunoassay based on the principle of competitive binding, which uses fluorescent microspheres-labeled antibody as the indicator marker to qualitatively detect fentanyl in human urine. Drugs which may be present in the urine specimen compete against the drug conjugate for binding sites on the antibody.
During testing, a urine specimen migrates upward by capillary action. Fentanyl, if present in the urine specimen below 1.0 ng/mL, will not saturate the binding sites of antibody-coated particles in the test device. The antibody coated fluorescence particles will then be captured by immobilized Fentanyl conjugate, and the signal will be detected in the test line (T) region to show a negative result. The signal will not be detected in the test line (T) region if the Fentanyl level exceeds 1.0 ng/mL because all the binding sites for the anti-Fentanyl antibodies will be saturated and the result will show as positive. To serve as a procedural control, a signal will be detected at the control line (C) region indicating the proper volume of specimen has been added and membrane wicking has occurred. The test is interpreted by the Healgen® Immunofluorescence analyzer OG-H180 and the result will be interpreted by the analyzer.
The provided FDA 510(k) clearance letter pertains to the Healgen® AccuFluor Fentanyl Fluorescence Immunoassay (FIA) Test Kit - Qualitative and the Healgen® Immunofluorescence Analyzer (OG-H180). This document outlines the general regulatory approval and provides some performance characteristics, but it is not a comprehensive study report detailing all aspects of the acceptance criteria and the full study that proves the device meets those criteria.
Specifically, the document does not explicitly state "acceptance criteria" as a defined set of metrics and thresholds prior to presenting performance data. Instead, it presents results from various analytical performance studies which are implicitly used to demonstrate equivalence to a predicate device. Similarly, it does not describe "human expert ground truth establishment," "adjudication methods," or "MRMC comparative effectiveness studies" because these are typically relevant for AI/ML-based diagnostic devices utilizing image interpretation or complex decision support, which is not the primary function described for this immunoassay and analyzer.
This device is an in vitro diagnostic (IVD) test for qualitative detection of fentanyl in urine, which relies on a chemical reaction read by an analyzer. Therefore, the "study" described is a series of analytical performance tests, rather than a clinical study with human readers and ground truth established by medical experts in the way that would be done for an AI radiology device, for example.
Despite these limitations in the provided text for certain categories, I will extract and infer information where possible based on the provided document and common IVD device clearance practices.
Acceptance Criteria and Device Performance for Healgen® AccuFluor Fentanyl FIA Test Kit
1. Table of Acceptance Criteria and Reported Device Performance
As noted, the document does not explicitly list pre-defined "acceptance criteria" with specific numerical thresholds for all metrics. However, based on the provided performance data, here's an interpretation of the implied criteria and the reported performance. The "acceptance criteria" inferred here are based on what constitutes successful demonstration of performance for an IVD device of this type, often aiming for high accuracy, precision, and lack of interference, especially around the cutoff concentration.
| Performance Characteristic | Implied Acceptance Criteria (Inferred) | Reported Device Performance |
|---|---|---|
| Analytical Precision (Around Cutoff) | High agreement (low false negatives/positives) at concentrations near the 1.0 ng/mL cutoff. | At -25% Cutoff (0.75 ng/mL): 56-57 out of 60 negative readings across 3 lots (93.3% - 95% negative agreement). At Cutoff (1.0 ng/mL): 22-24 negative and 36-38 positive readings out of 60 total across 3 lots. This demonstrates the expected transition around the cutoff. At +25% Cutoff (1.25 ng/mL): 60 out of 60 positive readings across 3 lots (100% positive agreement). At -100%, -75%, -50% Cutoff: 60 out of 60 negative readings. At +50%, +75%, +100% Cutoff: 60 out of 60 positive readings. |
| Analytical Specificity (Cross-Reactivity) | Minimal to no cross-reactivity with common related compounds (e.g., other opioids, metabolites) or other substances found in urine, beyond expected concentrations where some cross-reactivity is acceptable and quantified. | Norfentanyl: 0.003% cross-reactivity at 30,000 ng/mL. Carfentanil: 0.013% cross-reactivity at 8,000 ng/mL. Cyclopropyl fentanyl, Para-fluoro fentanyl, Acetyl fentanyl: 100% cross-reactivity at 1 ng/mL. Many other fentanyl analogs: various quantified cross-reactivity percentages. Numerous unrelated compounds/metabolites: No cross-reactivity at 100 µg/mL. |
| Interference | No significant interference from common substances (physiological or exogenous) found in human urine at specified concentrations. | Many compounds (e.g., Acetaminophen, Ethanol, Glucose, Albumin, Hemoglobin) showed no interference at high concentrations (e.g., 100µg/mL or 1% for Ethanol) for both negative and ±50% Cut-Off fentanyl spiked samples. |
| Stability | Device maintains stated performance characteristics over its shelf life under specified storage conditions. | Stable at 2-30°C for 27 months based on real-time stability study. |
| Effect of Urine Specific Gravity & pH | Test performance (positive/negative call) should remain consistent across a physiological range of urine specific gravity and pH. | Results were consistent (all positive for samples at/above +50% Cut-Off, all negative for samples at/below -50% Cut-Off) across specific gravity 1.000-1.035 and pH 4-9. |
| Method Comparison (Clinical Samples) | High agreement (concordance) with a confirmed analytical method (LC-MS/MS) for clinical samples, especially for samples near the cutoff. | Overall Concordance: Across 3 sites, for 80 clinical samples (40 negative, 40 positive) compared to LC-MS/MS: True Negative Ranges: 7 negative, 19 low negative (less than -50% cutoff) correctly identified as negative. Near Cutoff Negative: 11-12 samples (between -50% and cutoff) correctly identified as negative, with 2-3 false positives. Discordant Results: 6 samples near cutoff showed discordance. For example, 3 samples (0.802, 0.841, 0.916 ng/mL) were LC-MS/MS negative but device positive. 3 samples (1.013, 1.092, 1.113 ng/mL) were LC-MS/MS positive but device negative. This indicates typical variability at the qualitative cutoff. Near Cutoff Positive: 20-21 samples (between cutoff and +50%) correctly identified as positive, with 2-3 false negatives. High Positive: 17 samples (greater than +50%) correctly identified as positive. |
2. Sample Size Used for the Test Set and Data Provenance
- Analytical Precision: 60 replicates per concentration (6 replicates/day for 10 days) per lot, across 9 concentration levels, for 3 device lots. Total: 60 * 9 * 3 = 1620 individual tests.
- Interference: Samples with various interfering substances were tested, each at both drug-free and ±50% Cut-Off spiked fentanyl concentrations, using three batches of device. (Exact number of tests not specified, but implies a comprehensive set).
- Specificity: Various drug metabolites and other compounds tested, each using three batches of device. (Exact number of tests not specified).
- Effect of Urine Specific Gravity and pH: Samples across the specified ranges were tested at -50% and +50% Cut-Off levels by three different operators using three device lots. (Exact number of tests not specified).
- Method Comparison (Clinical Samples): 80 unaltered clinical samples (40 negative, 40 positive). These samples were run at three different testing sites.
- Data Provenance: The document does not explicitly state the country of origin for the clinical samples. It does state they were "unaltered clinical samples," implying they were retrospective real-world samples collected from patients. It does not indicate if they were prospective.
3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications
- Not Applicable in the traditional sense for this device. For this IVD device, the primary ground truth for its performance studies (precision, specificity, method comparison) is established by analytical gold standards, specifically:
- LC/MS-MS (Liquid Chromatography-Mass Spectrometry/Mass Spectrometry) for confirming fentanyl concentrations in precision studies and as the comparator method in the method comparison study.
- This is a highly accurate and precise laboratory method for quantifying drug concentrations, and its results are considered the "ground truth" for chemical concentration data.
- There were "three different operators" for the specific gravity/pH study, but these are not "experts" in the sense of medical professionals establishing a clinical diagnosis ground truth. They are laboratory personnel performing the test.
4. Adjudication Method for the Test Set
- Not Applicable in the traditional sense. Given that the ground truth is established by LC-MS/MS, there is no human "adjudication" process like consensus reading by multiple radiologists for image interpretation. The LC-MS/MS results serve as the definitive reference.
5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done, and Effect Size
- No, an MRMC comparative effectiveness study was not done. This type of study (comparing human readers with and without AI assistance on complex interpretation tasks) is not applicable to a qualitative immunoassay and analyzer like the Healgen AccuFluor Fentanyl FIA Test Kit, which determines the presence or absence of a substance based on a fluorescent signal. The device performance is assessed on its analytical accuracy against a gold standard method.
6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done
- Yes, the performance presented is primarily standalone. The Healgen® Immunofluorescence Analyzer (OG-H180) automatically interprets the fluorescent signal from the test kit. The performance data (precision, specificity, interference, method comparison) directly reflects the analytical capability of the device and test kit combination, without any human interpretation or intervention in the final "positive" or "negative" determination. A human loads the sample and the device performs the analysis and provides the result.
7. The Type of Ground Truth Used
- Analytical Gold Standard (LC-MS/MS): This is the primary method used to establish the true concentration of fentanyl in samples for precision studies and as the comparative reference for clinical samples.
- Spiked Samples: For analytical performance studies (precision, interference, specificity), known concentrations of fentanyl or interfering substances were added to negative urine samples, establishing a controlled ground truth.
8. The Sample Size for the Training Set
- Not explicitly stated in the document, and likely not applicable in the typical AI/ML sense. This device is an immunoassay, not an AI/ML diagnostic algorithm that undergoes a "training" phase with a large dataset. Immunoassays are based on biochemical principles and do not "learn" from data in the same way. Performance is optimized during development and validated analytically.
9. How the Ground Truth for the Training Set Was Established
- Not Applicable / Not Described. As it's not an AI/ML device relying on a training set, the concept of establishing ground truth for training does not apply here. The analytical performance is characterized through rigorous testing under controlled conditions and comparison to established reference methods (like LC-MS/MS).
Ask a specific question about this device
(20 days)
The LZI Fentanyl III Enzyme Immunoassay is intended for the qualitative determination of fentanyl in human urine at the cutoff value of 1 ng/mL when calibrated against fentanyl. The assay is designed for prescription use with a number of automated clinical chemistry analyzers.
The assay provides only a preliminary analytical result. A more specific alternative chemical method (e.g., gas or liquid chromatography and mass spectrometry) must be used in order to obtain a confirmed analytical result. Clinical consideration and professional judgment should be exercised with any drug of abuse test result, particularly when the preliminary test result is positive.
The LZI Fentanyl III Enzyme Immunoassay is a homogeneous enzyme immunoassay with ready-to-use liquid reagents. The assay is based on competition between the drug in the sample and the drug labeled with the enzyme glucose-6-phosphate dehydrogenase (G6PDH) for a fixed amount of antibody in the reagent. The drug-labeled G6PDH conjugate is traceable to a commercially available fentanyl standard and referred to as fentanyl-labeled G6PDH conjugate. Enzyme activity decreases upon binding to the antibody, and the drug concentration in the sample is measured in terms of enzyme activity. In the absence of a drug in the sample, fentanyl-labeled G6PDH conjugate is bound to the antibody, and the enzyme activity is inhibited. On the other hand, when the free drug is present in the sample, the antibody would bind to the free drug; the unbound fentanyl-labeled G6PDH then exhibits its maximal enzyme activity. Active enzyme converts nicotinamide adenine dinucleotide (NAD) to NADH, resulting in an absorbance change that can be measured spectrophotometrically at 340 nm.
The LZI Fentanyl III Enzyme Immunoassay is a kit comprised of two reagents, R1 and R2, which are bottled separately but sold together within the kit.
The R1 solution contains mouse monoclonal anti-fentanyl antibody, glucose-6-phosphate (G6P), nicotinamide adenine dinucleotide (NAD), stabilizers, and sodium azide (0.09%) as a preservative. The R2 solution contains glucose-6-phosphate dehydrogenase (G6PDH) labeled with fentanyl in buffer with sodium azide (0.09%) as a preservative.
The provided FDA 510(k) Clearance Letter for the LZI Fentanyl III Enzyme Immunoassay details the device's acceptance criteria (in terms of performance characteristics) and the study results that demonstrate the device meets these criteria.
Here's a breakdown of the requested information:
1. Acceptance Criteria and Reported Device Performance
The acceptance criteria are implicitly defined by the satisfactory performance demonstrated in the various studies, aiming to prove substantial equivalence to the predicate device. The performance characteristics sections show the device meets the expected qualitative and quantitative accuracy, precision, and specificity.
| Acceptance Criteria Category | Specific Criteria (Implicitly Derived from Study Design/Outcomes) | Reported Device Performance |
|---|---|---|
| Intended Use Equivalence | Device is substantially equivalent to the predicate (LZI Fentanyl II Enzyme Immunoassay) in terms of intended use, method principle, device components, and clinical performance. | The submission states, "The LZI Fentanyl III Enzyme Immunoassay is substantially equivalent to the LZI Fentanyl II Enzyme Immunoassay (K201938) manufactured by LZI in terms of intended use, method principle, device components, and clinical performance." The changes are clearly identified (target analyte from norfentanyl to fentanyl, updated cutoff, and assay application parameters), and subsequent studies address the impact of these changes. |
| Qualitative Accuracy | High concordance with LC/MS confirmation for true positives and true negatives, especially at and around the cutoff concentration. | Method Comparison - Clinical Samples (1 ng/mL Cutoff):- Negative by LC/MS analysis: 35 negatives, 0 positives by EIA.- < 50% of cutoff by LC/MS: 0 negatives, 20 positives by EIA (discrepant, attributed to norfentanyl cross-reactivity).- Near Cutoff Negative (50% below to cutoff) by LC/MS: 0 negatives, 12 positives by EIA (discrepant, attributed to norfentanyl cross-reactivity).- Near Cutoff Positive (cutoff to 50% above) by LC/MS: 0 negatives, 22 positives by EIA.- High Positive (> 50% above cutoff) by LC/MS: 0 negatives, 61 positives by EIA.Overall, 35/35 true negatives, 83/83 true positives (excluding near-cutoff positives and positives above cutoff). The remaining 32 discrepant samples (20 + 12) were positive by EIA but negative by LC/MS for fentanyl, indicating cross-reactivity with norfentanyl as a contributing factor. |
| Precision | Consistent and reliable qualitative results (positive/negative) across multiple runs and days, especially around the cutoff concentration. | Precision: 1 ng/mL Cutoff (N=88 total replicates over 22 days):- 0 ng/mL, 0.25 ng/mL, 0.5 ng/mL, 0.75 ng/mL: 88/88 Negative (100% agreement).- 1 ng/mL (Cutoff): 4 Neg / 84 Pos (95.5% Positive, 4.5% Negative - demonstrating near-cutoff variability, which is expected for qualitative assays at the cutoff).- 1.25 ng/mL, 1.5 ng/mL, 1.75 ng/mL, 2 ng/mL: 88/88 Positive (100% agreement). |
| Cross-Reactivity (Specificity) | Minimal or no false positives from structurally unrelated compounds. Acceptable cross-reactivity with fentanyl metabolites and structurally related compounds as defined by the assay's intended specificity. | Fentanyl and Metabolites: Fentanyl (100%), Norfentanyl (40%).Structurally Related Compounds: Varying levels of cross-reactivity (e.g., Acetyl fentanyl 25%, Acryl fentanyl 100%, Thiofentanyl 200%). Many substances showed "ND" (Not Detected) at high concentrations (e.g., 100,000 ng/mL).Structurally Unrelated Pharmacological Compounds: No interference (all Neg/Neg/Pos results for 0 ng/mL Fentanyl, -50% Cutoff, +50% Cutoff respectively) at 100,000 ng/mL for almost all listed compounds, except Dextromethorphan, which showed interference at 20,000 ng/mL (Neg/Pos/Pos) but was acceptable at 15,000 ng/mL (Neg/Neg/Pos). This demonstrates good specificity against common drugs. |
| Interference | Test performance (positive/negative results) should not be significantly affected by common endogenous substances or preservatives found in urine, or by varying specific gravity and pH within physiological ranges. | Endogenous and Preservative Compound Interference: No interference observed for most compounds (e.g., Acetone, Ascorbic acid, Bilirubin, Glucose, Hemoglobin) at high concentrations (e.g., 100,000 mg/dL), showing Neg/Neg/Pos results for 0 ng/mL, -50% Cutoff, +50% Cutoff respectively. Boric acid showed interference at 1,000 mg/dL (Neg/Neg/Neg), indicating a potential issue if present at high concentrations. This is a known interference for certain urine assays.Specific Gravity Interference: No interference observed across the range of 1.000 to 1.030 (all Neg/Neg/Pos).pH Interference: No major interference observed between pH 3 to pH 11 (all Neg/Neg/Pos). |
2. Sample Size and Data Provenance
- Test Set Sample Size:
- Precision Study: 88 replicates for each concentration level (0 ng/mL to 2 ng/mL fentanyl).
- Method Comparison (Clinical Samples): 150 unaltered clinical samples.
- Cross-Reactivity: Duplicates for each compound and concentration tested.
- Interference (Endogenous/Preservative): Duplicates for each compound and concentration tested.
- Specific Gravity/pH Interference: Duplicates for each specific gravity/pH level and fentanyl concentration.
- Data Provenance:
- The clinical samples for the Method Comparison study were obtained by LZI and "through collaboration with various clinical laboratories across the United States and Canada." This indicates a prospective and/or retrospective collection of real-world clinical samples from diverse geographical regions. The nature (retrospective/prospective) of the collection for these specific 150 samples isn't explicitly stated but "clinical samples" usually implies samples collected from patients.
3. Number of Experts and Qualifications for Ground Truth
- The document does not specify the number of experts used to establish ground truth.
- Qualifications of Experts: Not explicitly stated. However, for LC/MS confirmation, it implicitly relies on the expertise of the laboratory personnel performing and interpreting the LC/MS (Liquid Chromatography-Mass Spectrometry) analysis, which is a gold standard for chemical identification and quantification. These would typically be trained analytical chemists or lab technicians with experience in mass spectrometry.
4. Adjudication Method for the Test Set
- The document does not describe an adjudication method for the test set in the typical sense of multiple human readers or reviewers resolving discrepancies.
- For the Method Comparison study, the "ground truth" was established independently by LC/MS. Discrepancies between the EIA result and the LC/MS result were reported and attributed to norfentanyl cross-reactivity. This is a comparison against a reference method, not a subjective adjudication by experts.
5. Multi Reader Multi Case (MRMC) Comparative Effectiveness Study
- No, an MRMC comparative effectiveness study was not done. This type of study is more common for diagnostic imaging AI algorithms where human interpretation is a primary component and AI aims to augment or replace it. This document describes an in vitro diagnostic (IVD) assay, an enzyme immunoassay, which is an automated lab test. Its performance is evaluated against a reference standard (LC/MS), not against human reader performance.
6. Standalone Performance
- Yes, standalone performance was done. The entire submission details the performance of the LZI Fentanyl III Enzyme Immunoassay as a standalone device. The qualitative accuracy, precision, cross-reactivity, and interference studies all evaluate the algorithm/device's performance independently in generating a preliminary analytical result from a urine sample. It does not involve human-in-the-loop performance; rather, it provides a result that a clinician then interprets in conjunction with other clinical data.
7. Type of Ground Truth Used
- The primary type of ground truth used for the quantitative confirmation of fentanyl concentration in the clinical samples (Method Comparison study) was LC/MS (Liquid Chromatography-Mass Spectrometry) analysis. This is considered a highly specific and sensitive reference method for drug quantification.
8. Sample Size for the Training Set
- The document does not explicitly state the sample size for the training set. The described studies are verification and validation activities for a modified assay (Special 510(k) for an existing predicate device, LZI Fentanyl II). For IVD assays, "training" often refers to the internal development and optimization of the assay performed by the manufacturer, which precedes the formal V&V studies presented for regulatory submission. The reported studies are the test set performance.
9. How the Ground Truth for the Training Set Was Established
- Since the training set size is not explicitly stated, the method for establishing ground truth for it is also not explicitly described. However, it is highly probable that internal development and optimization of the assay would have utilized similar rigorous analytical methods, likely LC/MS or other established analytical techniques, to calibrate and refine the assay's performance against known concentrations of fentanyl and its metabolites. This would be part of the manufacturer's design control and development process.
Ask a specific question about this device
(150 days)
BioSieve™ Fentanyl FIA Test Kit is a fluorescence immunoassay (FIA) for the qualitative determination of fentanyl in human urine at a cutoff concentration of 1.0 ng/mL. The assay is intended for use with BioSieve™ ToxiSmart FIA Reader.
It is for in vitro diagnostic use only. It is intended for prescription use.
The tests provide only preliminary results. A more specific alternative chemical must be used in order to obtain a confirmed analytical result. Gas Chromatography/Mass Spectrometry (GC/MS) or Liquid Chromatography/Tandem Mass Spectrometry (LC/MS-MS) is the preferred confirmatory method.
Clinical consideration and professional judgment should be exercised with any drug test result, particularly when the preliminary test result is positive.
BioSieve™ ToxiSmart FIA Reader is a portable fluorescence instrument for in vitro diagnostic use only. The Reader is designed to perform in vitro diagnostic tests on urine specimens. This Reader can be used in a laboratory or in a point-of-care setting.
This test uses a lateral flow design with location-dependent lines and zones. BioSieve™ ToxiSmart FIA Reader scans the test strip and displays results. The sample is added to the sample well of the test card, and the sample is drawn by capillary action into and through the fluorescent labeled pad, through the nitrocellulose strip and into the adsorption pad. Within the fluorescent labeled pad, the specimen comes into contact with antibodies conjugated with fluorescent microspheres. During this interaction, if the amount of fentanyl antigen in the sample is greater than or equal to the cutoff value, the antigen in the sample and the fluorescence-labeled antibody bind to the FTY antigenantibody complex when the sample passes through a pad of fluorescence-microbead-labeled antibody conjugate. As the sample flows and reaches the FTY antigen coated by the T-line of nitrocellulose membrane, the FTY antigen coated by the T-line antigen in the sample competitively bind the FTY antibody labeled with fluorescence, then the T-line captures fluorescence signal is weaker than the cutoff fluorescence signal. When the samples do not contain fentanyl antigen or levels below the cutoff value, as the sample flow, fluorescent microsphere labeled antibody to nitrocellulose membrane T line captures fluorescent signal is stronger than the cutoff fluorescence signal. Whether or not FTY antigen was present in the sample, the rabbit IgG fluorescent microsphere conjugate not bound to the test line continued to flow with the rest of the sample and soon encountered a control line composed of goatanti-rabbit IgG. The position of C-line will accumulate fluorescence signal. The C-line control area was scanned to confirm that adequate sample flow had occurred. High resolution, narrow band SMD LED was used as light source in the Immunofluorescence Analyzer. The central wavelength of the excitation spectrum is 365nm. The central response wavelength is 610nm.
The provided text describes the acceptance criteria and the study that proves the device meets those criteria for the BioSieve™ Fentanyl FIA Test Kit and BioSieve™ ToxiSmart FIA Reader.
Here's an analysis of the requested information:
1. Table of Acceptance Criteria and Reported Device Performance
The acceptance criteria are not explicitly stated as distinct pass/fail thresholds in a table format. However, the performance characteristics tested and summarized imply the criteria for acceptable performance. The study aims to demonstrate that the device performs as expected for a qualitative immunoassay for fentanyl in urine, particularly around the 1.0 ng/mL cutoff.
The reported device performance is presented in several sections:
| Performance Metric | Implied Acceptance Criteria (Based on typical immunoassay requirements) | Reported Device Performance |
|---|---|---|
| Precision | High agreement for samples well below and well above cutoff; reasonable agreement near cutoff. | For -100%, -75%, -50% cutoff, all 60 tests for each lot were negative (60-/0+). For +25%, +50%, +75%, +100% cutoff, all 60 tests for each lot were positive (60+/0-). For -25% cutoff, 58-59 negative and 1-2 positive out of 60. For cutoff (1.0 ng/mL), 28-29 negative and 31-32 positive out of 60. |
| Stability | Device remains effective under specified storage conditions for its shelf life. | Stable at 2-30°C for 24 months based on accelerated stability study; real-time ongoing. |
| Interference | No significant interference from common substances found in urine. | Numerous compounds (e.g., Acetaminophen, Ethanol, Glucose) at specific concentrations showed no interference. |
| Specificity (Cross-Reactivity) | Low or no cross-reactivity with other related compounds or metabolites not intended to be detected, especially at clinically relevant concentrations. | Lists various fentanyl-related compounds and their cross-reactivity percentages (e.g., Acetyl fentanyl 83.33%, Carfentanil 0.20%). Many opiate compounds showed no cross-reactivity at 100 µg/mL. |
| Effect of Urine Specific Gravity and pH | Performance should be robust across a physiological range of urine specific gravity and pH. | Samples spiked at -50% and +50% Cut-Off levels across specific gravity 1.000-1.035 and pH 4-9 showed expected negative and positive results respectively. |
| Method Comparison (Clinical Samples) | High concordance with a reference method (LC-MS/MS), especially for samples far from the cutoff. Acceptable levels of discordance near the cutoff. | Site 1: 0 False Positives (FP) from negative, 0 FP from low negative, 4 FP from near cutoff negative. 1 False Negative (FN) from near cutoff positive, 0 FN from high positive. Site 2: 0 FP from negative, 0 FP from low negative, 3 FP from near cutoff negative. 2 FN from near cutoff positive, 0 FN from high positive. Site 3: 0 FP from negative, 0 FP from low negative, 2 FP from near cutoff negative. 1 FN from near cutoff positive, 0 FN from high positive. Discordant Results: Specific sample numbers and their LC-MS/MS vs. BioSieve results are listed, showing cases where the device gave a positive result for a true negative near the cutoff, and a negative result for a true positive near the cutoff. |
2. Sample Size Used for the Test Set and Data Provenance
- Precision Study Test Set: For each of 9 concentrations (e.g., -100% cutoff, cutoff, +100% cutoff), 6 runs per day for 10 days per device lot were performed. With 3 lots, this suggests: 9 concentrations * 6 tests/run * 10 days * 3 lots = 1620 individual tests (results reported as sums across these).
- Interference Study Test Set: Numerous interfering substances were tested with 3 batches of each device for drug-free urine and target fentanyl urine at -50% and +50% Cut-Off levels. The exact number of tests per substance is not specified, but it implies a substantial number.
- Specificity (Cross-Reactivity) Test Set: Various drug metabolites and opioid compounds were tested using three batches of device. The number of individual tests per compound is not explicitly stated.
- Effect of Urine Specific Gravity and pH Test Set: Urine samples across the specified ranges were spiked at -50% and +50% Cut-Off. Tested by three different operators per lot of device, with a total of three lots. The exact number of individual tests is not specified.
- Method Comparison Test Set: 80 unaltered clinical samples (40 negative and 40 positive based on an internal classification) were used per site across three different testing sites. This totals 240 clinical samples.
- Data Provenance: The document does not explicitly state the country of origin for the samples. It mentions "unaltered clinical samples" for the method comparison study, implying they were retrospective if they were already collected clinical samples. The precision study samples were "prepared by spiking fentanyl in negative samples," indicating these were artificially contrived samples, not naturally occurring clinical specimens, and were conducted in a controlled lab setting.
3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts
- For the Precision Study, "Each fentanyl concentration was confirmed by LC/MS-MS." LC/MS-MS (Liquid Chromatography/Mass Spectrometry) is a highly accurate analytical chemistry method, typically considered a gold standard for quantifying substances in biological samples. This method itself provides the "ground truth" and does not typically involve human "experts" in the sense of clinicians or radiologists reading results.
- For the Method Comparison Study, "The samples were blind labeled and compared to LC-MS/MS results." Again, LC-MS/MS is the ground truth. No human experts (e.g., radiologists) were used to establish the ground truth; it was established by an analytical instrument method.
- No information is provided about the qualifications of the operators who performed the tests on the BioSieve™ device at the three sites, nor is there mention of a panel of experts for subjective adjudication.
4. Adjudication Method for the Test Set
- None in the context of human expert adjudication. The ground truth for quantitative accuracy and comparative performance was established by LC-MS/MS, a definitive analytical method, not by human consensus or adjudication. The BioSieve™ device provides a qualitative result (positive/negative), which is then compared against the quantitative LC-MS/MS value relative to the 1.0 ng/mL cutoff.
5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study
- No, an MRMC study was not done. This device is a quantitative/qualitative immunoassay read by an instrument (BioSieve™ ToxiSmart FIA Reader), not interpreted by human readers in the same way an imaging AI would be. The "operators" in the method comparison study are likely technicians performing the test, not interpreting complex outputs. Therefore, a study on human reader improvement with AI assistance is not applicable.
6. Standalone (Algorithm Only Without Human-in-the-Loop Performance) Study
- Yes, in essence. The BioSieve™ ToxiSmart FIA Reader is an automated instrument that reads the test strip and displays results. The results presented for precision, interference, specificity, and comparison studies demonstrate the standalone performance of the BioSieve™ Fentanyl FIA Test Kit and the BioSieve™ ToxiSmart FIA Reader system. The device produces a qualitative result directly from the sample, without human interpretation of the signal beyond observing the final positive/negative display.
7. Type of Ground Truth Used
- LC-MS/MS (Liquid Chromatography/Mass Spectrometry): This highly accurate analytical method was used to confirm fentanyl concentrations in precision samples and as the reference method for clinical samples in the method comparison study. It provides objective, quantitative biochemical data, which is then translated to the qualitative positive/negative ground truth based on the 1.0 ng/mL cutoff.
8. Sample Size for the Training Set
- Not specified. The document describes performance validation studies for regulatory submission (510(k)). It does not provide details about a specific "training set" for an AI algorithm because the BioSieve™ ToxiSmart FIA Reader is not described as an AI-powered device in the typical sense of a deep learning model that requires a labeled training dataset. It's a fluorescence immunoassay reader. The "development modes" refer to how the test is run (Standard vs. Quick), not algorithmic training.
9. How the Ground Truth for the Training Set Was Established
- Not applicable. As noted above, the device is an immunoassay system and reader, not explicitly an AI/machine learning algorithm requiring a training set with established ground truth in the context of typical AI development. The document describes analytical validation, not algorithmic training.
Ask a specific question about this device
(57 days)
Dochek® Multi-Drug Urine Test Dipcard Rx is an immunoassay for the qualitative determination of single or multiple drugs in human urine at the following cutoff concentrations:
Amphetamine (AMP) 1000 or 500 ng/mL
Secobarbital (BAR) 300 ng/mL
Buprenorphine (BUP) 10 ng/mL
Oxazepam (BZO) 300 ng/mL
Cocaine (COC) 300 or 150 ng/mL
2-ethylidene-1,5-dimethyl-3,3-diphenylpyrrolidine (EDDP) 300 ng/mL
Methylenedioxymethamphetamine (MDMA) 500 ng/mL
Methamphetamine (MET) 1000 or 500 ng/mL
Morphine (MOP/OPI) 300 or 2000 ng/mL
Methadone (MTD) 300 ng/mL
Oxycodone (OXY) 100 ng/mL
Phencyclidine (PCP) 25 ng/mL
Propoxyphene (PPX) 300 ng/mL
Nortriptyline (TCA) 1000 ng/mL
Cannabinoids (THC) 50 ng/mL
6-Monoacetylmorphine (6-MAM) 10 ng/mL
Dochek® Multi-Drug Urine Test Dipcard Rx offers any combinations from 1 to 16 drugs but only one cutoff concentration under same drug condition will be included per device. It is intended for prescription use. For in vitro diagnostic use only.
The test provides only preliminary results. Clinical consideration and professional judgment should be applied to any drug of abuse test result, particularly in evaluating a preliminary positive result. To obtain a confirmed analytical result, a more specific alternate chemical method is needed. GC/MS is the recommended confirmatory method.
Dochek® Multi-Drug Urine Test Dipcard is an immunoassay for the qualitative determination of single or multiple drugs in human urine at the following cutoff concentrations:
Amphetamine (AMP) 1000 or 500 ng/mL
Secobarbital (BAR) 300 ng/mL
Buprenorphine (BUP) 10 ng/mL
Oxazepam (BZO) 300 ng/mL
Cocaine (COC) 300 or 150 ng/mL
2-ethylidene-1,5-dimethyl-3,3-diphenylpyrrolidine (EDDP) 300 ng/mL
Methylenedioxymethamphetamine (MDMA) 500 ng/mL
Methamphetamine (MET) 1000 or 500 ng/mL
Morphine (MOP/OPI) 300 or 2000 ng/mL
Methadone (MTD) 300 ng/mL
Oxycodone (OXY) 100 ng/mL
Phencyclidine (PCP) 25 ng/mL
Propoxyphene (PPX) 300 ng/mL
Nortriptyline (TCA) 1000 ng/mL
Cannabinoids (THC) 50 ng/mL
6-Monoacetylmorphine (6-MAM) 10 ng/mL
Dochek® Multi-Drug Urine Test Dipcard offers any combinations from 1 to 16 drugs but only one cutoff concentration under same drug condition will be included per device. It is intended for over-the-counter (OTC) use. For in vitro diagnostic use only.
The test provides only preliminary results. Clinical consideration and professional judgment should be applied to any drug of abuse test result, particularly in evaluating a preliminary positive result. To obtain a confirmed analytical result, a more specific alternate chemical method is needed. GC/MS or LC/MS is the recommended confirmatory method.
Dochek® Multi-Drug Urine Test Dipcard Rx and Dochek® Multi-Drug Urine Test Dipcard are immunochromatographic assays that use a lateral flow system for the qualitative detection of single or multiple drugs in human urine at or above the cut-off levels as indicated. The products are singleuse in vitro diagnostic devices.
This device is a dipcard format in which the test strips are integrated into the plastic dipcard. After removing the cap of the dipcard, the absorbent end of the test strips is exposed and can be in direct contact with the urine sample. The device is in a ready-to-use format and no longer requires assembly before use.
Here's a breakdown of the acceptance criteria and study details for the Dochek® Multi-Drug Urine Test Dipcard, based on the provided FDA 510(k) summary:
This device is an in-vitro diagnostic test, not an AI/ML device, so many of the requested criteria (e.g., number of experts, adjudication, MRMC study, training set) typically apply to AI/ML software. Therefore, I will adapt the response to describe the data provided for this type of diagnostic device.
Acceptance Criteria and Device Performance for Dochek® Multi-Drug Urine Test Dipcard
This device is a qualitative immunoassay for the detection of various drugs in human urine. The acceptance criteria and performance are primarily based on its analytical accuracy (precision, specificity, and comparison to a gold standard) and usability.
1. Table of Acceptance Criteria and Reported Device Performance
For this type of qualitative immunoassay, the primary acceptance criteria revolve around the device's ability to correctly classify samples as positive or negative at and around the specified cutoff concentrations. The "performance" is demonstrated through the observed concordance with confirmatory methods (LC-MS/MS) and its precision.
| Acceptance Criteria Category | Specific Criteria (Implicitly from Study Design) | Reported Device Performance (Summary) |
|---|---|---|
| Analytical Precision | Consistent positive results for samples ≥ 25% above cutoff. Consistent negative results for samples ≤ 25% below cutoff. Acceptable percentage of positive/negative results at cutoff. | Observed Precision (Examples from various drugs and lots at cutoff): - AMP (1000 ng/mL): Lot I: 13-/37+, Lot II: 14-/36+, Lot III: 13-/37+ (out of 50 tests) - BAR (300 ng/mL): Lot I: 14-/36+, Lot II: 12-/38+, Lot III: 14-/36+ - COC (300 ng/mL): Lot II: 13-/37+, Lot III: 12-/38+ - MDMA (500 ng/mL): Lot I: 10-/40+, Lot II: 10-/40+, Lot III: 11-/39+ All samples ≥ +25% cutoff consistently showed "0-/50+" (0 negative, 50 positive). All samples ≤ -25% cutoff consistently showed "50-/0+" (50 negative, 0 positive). |
| Analytical Specificity | Minimal or no cross-reactivity with structurally unrelated compounds at specified concentrations. Acceptable cross-reactivity with structurally related compounds. | Extensive cross-reactivity tables provided (pages 13-22 in original document) for all 16 drugs. Many structurally related compounds showed cross-reactivity, requiring careful interpretation of results (e.g., Amphetamine panel detects other amphetamine-like substances). Many non-structurally related compounds showed no interference at 100 µg/mL. |
| Interference | Urinary pH (4-9) and Specific Gravity (1.000-1.035) should not affect results. Substances commonly found in urine should not interfere. | pH levels of 4-9 and specific gravity levels of 1.000-1.035 did not affect assay results. Over 100 non-structurally related compounds (e.g., acetaminophen, ibuprofen, glucose) showed no interference at 100 µg/mL. (Pages 22-23) |
| Method Comparison (Accuracy) | High concordance with a confirmed analytical method (LC-MS/MS) for both negative and positive samples, especially around the cutoff concentrations. | High concordance with LC-MS/MS. For each drug, 80 clinical samples (40 negative, 40 positive) were tested. The detailed tables (pages 24-27) show the breakdown of positive/negative calls by the dipcard vs. LC-MS/MS concentrations (drug-free, low negative, near cutoff negative, near cutoff positive, high positive). While some discordant results (false positives/negatives near cutoff) were observed, they are typical for qualitative immunoassay screening tests designed to be highly sensitive around the cutoff. The device generally performed as expected for a qualitative screening test. |
| Lay Person Usability | Lay users should be able to correctly interpret the device results and understand the instructions for use. | High (≥90%) correct result interpretation by lay users across various drug concentrations for the majority of tests. The "Percentage of correct results" for samples +/- 25% of cutoff varied (e.g., BAR 95%, BUP 90% for -25% cutoff; BUP 95% for +25% cutoff), indicating some variability by lay users near the cutoff, which is expected for visual interpretation of a qualitative test. All participants found the instructions easy to understand and follow, with a Flesch-Kincaid Grade Level of 7. (Pages 28-33) |
| Stability | The device should demonstrate stability over its claimed shelf life. | Demonstrated stability at 2-30°C for 36 months based on real-time stability study. |
2. Sample Size and Data Provenance:
- Test Set Sample Sizes:
- Precision/Reproducibility: For each drug and each of the three lots, 50 tests were performed at each of the 9 concentration levels around the cutoff (total of 450 tests per lot per drug for analytical precision). This was repeated for multiple lots.
- Method Comparison: 80 clinical urine samples were used for each drug. These 80 samples consisted of 40 negative and 40 positive samples, further categorized by their concentration relative to the cutoff (drug-free, low negative, near cutoff negative, near cutoff positive, high positive).
- Lay Person Study: 140 participants (79 male, 61 female for Configuration 1; 73 male, 67 female for Configuration 2). For each drug and concentration level, 20 samples were tested by lay users.
- Data Provenance: The document generally indicates "in-house" studies for method comparison and precision. The clinical samples for method comparison were "unaltered urine clinical samples." The lay person study used pooled urine specimens spiked with drugs, confirmed by LC-MS/MS. The country of origin for the data is implied to be China, given the manufacturer's address (Guangzhou, China). The studies appear to be prospective in nature as they involved preparing samples and testing them using the device under controlled conditions to evaluate performance.
3. Number of Experts and Qualifications for Ground Truth:
- This device is a qualitative immunoassay, not an AI/ML algorithm requiring expert interpretation of images. Therefore, the concept of "experts" to establish ground truth in the same way as for imaging AI is not directly applicable.
- The ground truth for the analytical studies was established by LC-MS/MS (Liquid Chromatography-Mass Spectrometry/Mass Spectrometry). This is a highly accurate chemical method considered the gold standard for quantitative drug identification and concentration determination in urine.
- For the method comparison study, "three operators" performed the tests. Their qualifications are not explicitly stated, but for a laboratory-based study of an immunoassay, these would typically be trained laboratory personnel or technicians.
4. Adjudication Method for the Test Set:
- Not applicable in the context of an immunoassay device. The results are physical readouts (presence/absence of a line).
- For the Method Comparison study, the device result (positive/negative) was compared directly to the quantitative LC-MS/MS result. Discordant results are individually listed (pages 27-28), showing which operators, if any, had differing readings from the LC-MS/MS ground truth. Since three operators independently read the same samples for the method comparison, consistency among them can be inferred, but formal "adjudication" is not described as they were simply recording what they observed.
5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study:
- No, an MRMC comparative effectiveness study was not conducted for this device. This type of study is specifically designed for AI-assisted diagnostic tools (often in imaging) to compare human performance with and without AI assistance.
- This device is a standalone diagnostic test; it does not "assist" human readers in the interpretation of complex data (like medical images).
6. Standalone (Algorithm Only) Performance:
- Yes, in a sense. The "Performance Characteristics" section details the device's inherent analytical performance (precision, specificity, and method comparison against LC-MS/MS). These studies evaluate the device's ability to detect drugs independently of human interpretation nuances (though human "operators" did conduct the tests, the output is a visual line).
- The "Lay Person Study" then evaluates the human interpretation of the device's output.
7. Type of Ground Truth Used:
- The primary ground truth used throughout the studies (precision, method comparison) was GC/MS or LC/MS (Gas Chromatography/Mass Spectrometry or Liquid Chromatography/Mass Spectrometry) results. These are highly accurate, quantitative analytical chemistry methods considered definitive for drug presence and concentration.
- For the lay person study, the "ground truth" for the samples was also established by LC-MS/MS confirmation of spiked drug concentrations.
8. Sample Size for the Training Set:
- This device is a chemical immunoassay, not a machine learning model. Therefore, there is no "training set" in the context of algorithm development. The device's performance is inherent to its biochemical design.
9. How Ground Truth for Training Set was Established:
- As stated above, no training set for an algorithm exists for this type of device. The scientific and engineering principles of immunoassay design and manufacturing determine its performance, which is then validated through the analytical and clinical studies described.
Ask a specific question about this device
(254 days)
The Rapid Fentanyl (FYL) Test Strip is a rapid, screening test for the qualitative detection of Fentanyl (FYL) in human urine at the cut-off concentration of 1 ng/mL.
The tests is intended for in vitro diagnostics use. It is intended for prescription use including point of care sites. This assay provides only a preliminary analytical test result. To obtain a confirmed analytical result, a more specific alternative chemical method must be used. Gas Chromatography/ Mass spectrometry (GC/MS) or Liquid chromatography/Mass spectrometry (LC/MS) is the preferred confirmatory method. Clinical consideration and professional judgment should be applied to any drug of abuse test result, particularly when preliminary positive results are indicated.
The Rapid Fentanyl (FYL) Test Dipcard is a rapid, screening test for the qualitative detection of Fentanyl (FYL) in human urine at the cut-off concentration of 1 ng/mL.
The tests is intended for in vitro diagnostics use. It is intended for prescription use including point of care sites. This assay provides only a preliminary analytical test result. To obtain a confirmed analytical result, a more specific alternative chemical method must be used. Gas Chromatography/ Mass spectrometry (GC/MS) or Liquid chromatography/Mass spectrometry (LC/MS) is the preferred confirmatory method. Clinical consideration and professional judgment should be applied to any drug of abuse test result, particularly when preliminary positive results are indicated.
Rapid Fentanyl (FYL) Test Strip and Rapid Fentanyl (FYL) Test Dipcard are competitive binding, lateral flow immunochromatographic assays for qualitatively the detection of fentanyl at or above the cut-off concentration of 1 ng/mL. The tests can be performed without the use of an instrument. Test Strip and Test Dipcard use identical test strips made with same chemical formulation and manufacturing procedures.
The provided document describes the Co-Innovation Biotech Co.,Ltd. Rapid Fentanyl (FYL) Test Strip and Rapid Fentanyl (FYL) Test Dipcard, which are rapid screening tests for the qualitative detection of Fentanyl (FYL) in human urine at a cut-off concentration of 1 ng/mL. The document includes performance data to demonstrate substantial equivalence to a predicate device (K220046 Superbio Fentanyl Urine Detection Kit).
Here's an analysis of the acceptance criteria and the study that proves the device meets them, based on the provided text:
1. Table of Acceptance Criteria and Reported Device Performance:
The document doesn't explicitly state "acceptance criteria" with numerical targets for accuracy, sensitivity, or specificity. However, the precision study and accuracy study demonstrate performance relative to the cut-off concentration. The implied acceptance criteria are that the device should accurately detect fentanyl around the 1 ng/mL cut-off and exhibit minimal interference and cross-reactivity. The precision study illustrates the device's ability to classify samples correctly relative to the cut-off, and the accuracy study compares device results to LC/MS reference results.
| Performance Metric | Implied Acceptance Criteria (Based on typical drug screening test expectations) | Reported Device Performance (Rapid Fentanyl (FYL) Test Strip and Dipcard - data presented is identical for both) |
|---|---|---|
| Precision | Consistent results for samples below, at, and above the cut-off. | Rapid Fentanyl (FYL) Test Strip: - 0 ng/mL, 0.25 ng/mL, 0.5 ng/mL: 100% negative (60/60 for each lot). - 0.75 ng/mL (-25% cutoff): 4/60 (Lot 1), 4/60 (Lot 2), 2/60 (Lot 3) positive results (majority negative). - 1 ng/mL (cutoff): 34/60 (Lot 1), 34/60 (Lot 2), 36/60 (Lot 3) positive results (near 50% positive/negative as expected at cutoff). - 1.25 ng/mL (+25% cutoff): 56/60 (Lot 1), 58/60 (Lot 2), 56/60 (Lot 3) positive results (majority positive). - 1.5 ng/mL, 1.75 ng/mL, 2 ng/mL: 100% positive (60/60 for each lot).Rapid Fentanyl (FYL) Test Dipcard: - 0 ng/mL, 0.25 ng/mL, 0.5 ng/mL: 100% negative (60/60 for each lot). - 0.75 ng/mL (-25% cutoff): 4/60 (Lot 1), 2/60 (Lot 2), 6/60 (Lot 3) positive results (majority negative). - 1 ng/mL (cutoff): 36/60 (Lot 1), 34/60 (Lot 2), 32/60 (Lot 3) positive results (near 50% positive/negative as expected at cutoff). - 1.25 ng/mL (+25% cutoff): 54/60 (Lot 1), 56/60 (Lot 2), 56/60 (Lot 3) positive results (majority positive). - 1.5 ng/mL, 1.75 ng/mL, 2 ng/mL: 100% positive (60/60 for each lot). |
| Accuracy (Agreement with Reference Method) | High agreement with a confirmed analytical method (LC/MS). | Rapid Fentanyl (FYL) Test Strip & Dipcard: - Discordant results for both devices and all sites: One positive at 0.76 ng/mL (LC/MS result, below cutoff) and one negative at 1.04 ng/mL (LC/MS result, above cutoff). - Drug-free: 27/27 negative calls by device. - Less than half cutoff: 5/5 negative calls by device. - Near Cutoff Negative (0.5-1 ng/mL): 7/8 negative calls by device, 1/8 positive call (at 0.76ng/mL). - Near Cutoff Positive (1-1.5 ng/mL): 8/9 positive calls by device, 1/9 negative call (at 1.04ng/mL). - High Positive (>1.5 ng/mL): 31/31 positive calls by device. |
| Cross-reactivity | Minimal cross-reactivity with structurally similar compounds. | Fentanyl analogs showed varying degrees of cross-reactivity, some as low as 1.2 ng/mL (Acetyl fentanyl), while Norfentanyl and Acetyl norfentanyl showed very low (0.01%) cross-reactivity at 10,000 ng/mL. This indicates the device can detect some fentanyl analogs at relevant concentrations. |
| Interference | No interference from common opioids, medications, or endogenous substances. | None of the tested opioids, commonly ingested medications, or substances (at specified concentrations) were shown to interfere with the test. |
| Effect of urinary pH | Consistent results across a range of urinary pH. | Varying ranges of pH (3 to 9) did not interfere with performance. |
| Effect of urinary specific gravity | Consistent results across a range of urinary specific gravity. | Varying ranges of urinary specific gravity (1.000 to 1.040) did not affect the test result. |
2. Sample Size and Data Provenance for the Test Set:
- Precision Study (Test Set):
- Sample Size: 1620 observations in total (3 lots x 9 concentrations x 60 determinations/lot for each device type). Each concentration tested in 3 replicates daily for 10 non-consecutive days with 3 aliquots per location (2 operators per location, 3 locations).
- Data Provenance: Drug-free human urine spiked with target fentanyl at various concentrations. The document does not specify the country of origin but implies laboratory-controlled samples, not patient data. Retrospective, as urine samples were collected and then spiked.
- Accuracy Study (Test Set):
- Sample Size: 80 clinical urine specimens.
- Data Provenance: Clinical urine specimens. The country of origin is not specified. Retrospective.
3. Number of Experts (or Reference Method) and Qualifications to Establish Ground Truth for the Test Set:
- Precision Study: Ground truth was established by precise spiking of fentanyl into drug-free urine at known concentrations, confirmed by LC/MS. No human experts were involved in establishing the ground truth for these spiked samples, as the concentration was analytically determined.
- Accuracy Study: Ground truth was established by LC/MS (Gas Chromatography/ Mass spectrometry or Liquid chromatography/Mass spectrometry) analysis of the clinical urine specimens. LC/MS is the "preferred confirmatory method" and considered the gold standard for quantitative drug testing. The qualifications of the LC/MS operators or analysts are not specified.
4. Adjudication Method for the Test Set:
Not applicable in the conventional sense of expert adjudication.
- In the precision study, known concentrations (LC/MS confirmed) served as the reference.
- In the accuracy study, LC/MS served as the reference method. Discordant results were analyzed by comparing the device's reading to the objective LC/MS concentration around the cutoff.
5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study:
- No MRMC comparative effectiveness study was done in the context of human readers improving with AI vs. without AI assistance. This device is a rapid, screening test (immunochromatographic assay) and is not an AI-powered diagnostic imaging or interpretative device that augments human reader performance.
- However, the precision study did involve multiple operators (6 operators at 3 Point-of-Care sites) reading the results, although it wasn't designed as a comparative effectiveness study of human reader improvement. Each operator independently read the samples.
6. Standalone (Algorithm Only Without Human-in-the-Loop Performance):
- Yes, this device is a standalone test. The "Rapid Fentanyl (FYL) Test Strip and Rapid Fentanyl (FYL) Test Dipcard are competitive binding, lateral flow immunochromatographic assays for qualitatively the detection of fentanyl... The tests can be performed without the use of an instrument." The results are visually interpreted by a human operator, but the device itself does not involve an algorithm or AI. It functions as a direct chemical/immunological reaction.
7. Type of Ground Truth Used:
- Precision Study: Known concentrations of Fentanyl in urine, confirmed by LC/MS.
- Accuracy Study: Quantitative results from LC/MS analysis of clinical urine specimens.
8. Sample Size for the Training Set:
This information is typically not applicable or explicitly stated for rapid, immunoassay-based tests like the one described. These devices are developed through chemical and biological formulation and optimization rather than machine learning training on a "training set" of data. The document describes analytical validation studies, which are performance evaluations, not algorithm training.
9. How the Ground Truth for the Training Set Was Established:
Not applicable, as there is no "training set" in the context of an AI/ML algorithm for this type of immunochromatographic device. The development process would involve optimizing reagent concentrations, membrane properties, and other manufacturing parameters to achieve the desired sensitivity and specificity, typically guided by analytical testing against known standards.
Ask a specific question about this device
(211 days)
Accurature Multi-drug Urine Test Cup and Accurature Multi-drug Urine Test Dip Card are competitive binding, lateral flow immunochromatographic assays for qualitative and simultaneous detection of Amphetamine, Secobarbital, Buprenorphine, Oxazepam, Cocaine, Methylenedioxy-methamphetamine, Morphine, Methadone, Oxycodone, Phencyclidine, Marijuana in human urine at the cut off concentrations of:
| Drug | Abbreviation | Cut-off |
|---|---|---|
| Amphetamine | AMP | 1000ng/mL |
| Secobarbital | BAR | 300 ng/mL |
| Buprenorphine | BUP | 10 ng/mL |
| Opzaepam | BZO | 300 ng/mL |
| Cocaine | COC | 300 ng/mL |
| Methylenedioxy-methamphetamine | MDMA | 500 ng/mL |
| Methamphetamine | MET | 1000ng/mL |
| Morphine | MOR | 300 ng/mL |
| Methadone | MTD | 300 ng/mL |
| Oxycodone | OXY | 100 ng/mL |
| Phencyclidine | PCP | 25 ng/mL |
| Marijuana | THC | 50 ng/mL |
Configuration of the Accurature Multi-drug Urine Test Cup and Accurature Multi-drug UrineTest Dip Card can consist of any combination of the above listed drug analytes.
The test may yield positive results for the prescription drugs Buprenorphine, Oxazepam Secobarbital and Oxycodone when taken at or above prescribed doses. It is not intended to distinguish between prescription use or abuse of these drugs. Clinical consideration and professional judgment should be exercised with any drug of abuse test result, particularly when the preliminary result is positive. The test provides only preliminary test results. A more specific alternative chemical method must be used in order to obtain a confirmed analytical result. GC/MS is the preferred confirmatory method.
For over-the-counter use.
Multi-drug Urine Test Cup and Multi-drug Urine Test Dip Card are competitive binding, lateral flow immunochromatographic assays for qualitative and simultaneous detection of Amphetamine, Secobarbital, Buprenorphine, Oxazepam, Cocaine, Methylenedioxy-methamphetamine, Morphine, Methadone, Oxycodone, Phencyclidine, Marijuana in human urine at the cut off concentrations of:
| Drug | Abbreviation | Cut-off |
|---|---|---|
| Amphetamine | AMP | 1000ng/mL |
| Secobarbital | BAR | 300 ng/mL |
| Buprenorphine | BUP | 10 ng/mL |
| Opzaepam | BZO | 300 ng/mL |
| Cocaine | COC | 300 ng/mL |
| Methylenedioxy-methamphetamine | MDMA | 500 ng/mL |
| Methamphetamine | MET | 1000ng/mL |
| Morphine | MOR | 300 ng/mL |
| Methadone | MTD | 300 ng/mL |
| Oxycodone | OXY | 100 ng/mL |
| Phencyclidine | PCP | 25 ng/mL |
| Marijuana | THC | 50 ng/mL |
Configuration of the Multi-drug Urine Test Cup and Multi-drug Test Dip Card can consist of any combination of the above listed drug analytes.
The test may yield positive results for the prescription drugs Buprenorphine, Oxazepam Secobarbital and Oxycodone when taken at or above prescribed doses. It is not intended to distinguish between prescription use or abuse of these drugs. Clinical consideration and professional judgment should be exercised with any drug of abuse test result, particularly when the preliminary result is positive. The test provides only preliminary test results. A more specific alternative chemical method must be used in order to obtain a confirmed analytical result. GC/MS is the preferred confirmatory method.
For prescription use only.
The subject device will be sold in two formats: Multi-drug Urine Test Dip Card and Multi-drug Urine Test Cup. Multi-Drug Urine Test Cup and Multi-drug Urine Test Dip Card are competitive binding, lateral flow immunochromatographic assays for qualitative and simultaneous detection of Amphetamine, Marijuana, Cocaine, Methamphetamine, Morphine, Oxazepam, Oxycodone, Secobarbital, Buprenorphine, Methylenedioxy-methamphetamine, Phencyclidine and Methadone in human urine samples. Multi-Drug Urine Test devices detect each of analytes on different strips. A positive urine sample will not generate a colored-line for the specific drug tested in the designated region. A negative urine specimen or a urine sample containing Amphetamine, Marijuana, Cocaine, Methamphetamine, Morphine, Oxazepam, Oxycodone, Secobarbital, Buprenorphine, Methylenedioxy-methamphetamine, Phencyclidine and Methadone at the concentration below the designated cutoff levels will generate a colored line in the designated test region for the drug. To serve as a test control, a color line will always appear at the control region.
The provided document is a 510(k) Summary for the Shanghai Accurature Diagnostics Co., Ltd. Multi-drug Urine Test Cup and Multi-drug Urine Test Dip Card. It describes the device's indications for use, its substantial equivalence to a predicate device, and performance data from various studies.
Here's an analysis of the acceptance criteria and the study that proves the device meets them:
1. A table of acceptance criteria and the reported device performance:
The document doesn't explicitly state "acceptance criteria" for precision or agreement. However, it indicates a "qualified" result for percentage agreement among positives and negatives in the comparison study (Section 10.1). For precision (Section 10.7), it states a cutoff coincidence rate of 79% (which is >55%) and a ±25% cutoff coincidence rate of 88% (which is >84%), both deemed "qualified." For the lay user study, the percentages of agreement for specific concentrations are presented. Given that these devices are screening tools, the implicit acceptance criteria would be high agreement rates (ideally 100% or very close) for negative and strongly positive samples, and reasonable agreement around the cutoff concentrations.
Here's a summary of the reported performance from the comparison study, which appears to be the primary study for establishing initial performance characteristics:
| Drug | Device | Agreement among positives | Agreement among negatives | Reported Performance (Agreement) |
|---|---|---|---|---|
| Amphetamine | Cup | 98.3% | 100.0% | Qualified |
| Amphetamine | Card | 98.3% | 100.0% | Qualified |
| Secobarbital | Cup | 98.3% | 100.0% | Qualified |
| Secobarbital | Card | 100.0% | 100.0% | Qualified |
| Buprenorphine | Cup | 98.3% | 100.0% | Qualified |
| Buprenorphine | Card | 100.0% | 100.0% | Qualified |
| Oxazepam | Cup | 98.3% | 100.0% | Qualified |
| Oxazepam | Card | 98.3% | 100.0% | Qualified |
| Cocaine | Cup | 97.5% | 100.0% | Qualified |
| Cocaine | Card | 98.3% | 100.0% | Qualified |
| Methylenedioxy-methamphetamine | Cup | 95.8% | 100.0% | Qualified |
| Methylenedioxy-methamphetamine | Card | 98.3% | 100.0% | Qualified |
| Methamphetamine | Cup | 95.8% | 100.0% | Qualified |
| Methamphetamine | Card | 99.2% | 100.0% | Qualified |
| Morphine | Cup | 97.5% | 100.0% | Qualified |
| Morphine | Card | 99.2% | 100.0% | Qualified |
| Methadone | Cup | 97.5% | 100.0% | Qualified |
| Methadone | Card | 97.5% | 100.0% | Qualified |
| Oxycodone | Cup | 99.2% | 100.0% | Qualified |
| Oxycodone | Card | 96.7% | 100.0% | Qualified |
| Phencyclidine | Cup | 95.8% | 100.0% | Qualified |
| Phencyclidine | Card | 99.2% | 100.0% | Qualified |
| Marijuana | Cup | 99.2% | 100.0% | Qualified |
| Marijuana | Card | 99.2% | 100.0% | Qualified |
2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)
- Comparison Study (Section 10.1): The sample size is not explicitly stated as a single number. The tables show "agreement among positives" and "agreement among negatives" for each drug with percentages. While not directly stated, the implication is that a sufficient number of samples were tested to generate these percentages. The provenance of the data (country of origin, retrospective/prospective) is not mentioned.
- Precision Study (Section 10.7): For each drug, the study tested "drug free," "-75%Cut off," "-50%Cut off," "+100% cutoff," "+75%Cut off," and "+50%Cut off" samples. The number of samples per concentration is not explicitly given, but the results for "-75%Cut off," "-50%Cut off," "drug free" samples were all negative, and "+100% cutoff," "+75%Cut off," "+50%Cut off" samples were all positive. The specific number of samples at the cutoff and ±25% cutoff concentrations that led to the "79%(>55%)" and "88%(>84%)" coincidence rates, respectively, is not provided.
- Lay User Study (Section 10.8):
- Test Cup: 603 laypersons. Urine samples were prepared at negative, +/-75%, +/-50%, +/-25% of the cutoff levels. Each layperson was given 1 blind-labeled sample.
- Test Dip Card: 300 laypersons. Urine samples were prepared at negative, +/-75%, +/-50%, +/-25% of the cutoff levels. Each layperson was given 1 blind-labeled sample.
- Data Provenance: Not specified (e.g., country of origin or retrospective/prospective). However, the study involved "three intended user sites" and "diverse educational and professional backgrounds."
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)
- Lay User Study (Section 10.8): The ground truth for the lay user study was established by LC/MS (Liquid Chromatography-Mass Spectrometry), which is a highly reliable analytical chemical method. This is not dependent on human expert interpretation of the test results themselves, but rather on instrument analysis. The samples were "spiking drug free-pooled urine specimens" with known concentrations, and these concentrations were "confirmed by LC/MS." Therefore, the ground truth is analytical, not expert consensus on the device's outcome.
- For other studies like the Comparison Study and Precision, the method for establishing ground truth isn't explicitly detailed, but generally, for drug screening devices, analytical methods (like GC/MS or LC/MS) are considered the gold standard for ground truth.
4. Adjudication method (e.g. 2+1, 3+1, none) for the test set
- No information on adjudication methods such as 2+1 or 3+1 is provided. The studies appear to rely on objective measures (percentages of agreement with an assumed or analytically confirmed truth). For the lay user study, the ground truth was analytical (LC/MS), eliminating the need for reader adjudication.
5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
- This question is not applicable to the provided document. The device is a lateral flow immunochromatographic assay (a rapid drug test), not an AI-powered diagnostic system or an imaging modality that would involve human readers interpreting images. Therefore, a multi-reader multi-case (MRMC) comparative effectiveness study with or without AI assistance is irrelevant to this device.
6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done
- This question is not applicable. The device is a physical diagnostic kit (test cup/dip card) that provides a qualitative result (presence or absence of a line). It is not an algorithm, and its performance is inherently "standalone" in that it produces a result without further computational analysis. The "human-in-the-loop" aspect for this device involves a person performing the test and visually interpreting the lines, particularly in the over-the-counter (OTC) use case. The lay user study evaluates this "human-in-the-loop" performance.
7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)
- The primary type of ground truth used, particularly for the lay user study, was analytical chemical confirmation, specifically LC/MS (Liquid Chromatography-Mass Spectrometry). For the precision study, samples were prepared at specific concentrations relative to the cutoff, implying a known, analytically verified concentration as the ground truth. This is considered a highly objective and accurate method for determining the presence and concentration of substances.
8. The sample size for the training set
- The document describes performance studies, cross-reactivity, interference, and a lay user study. It does not mention a "training set" in the context of machine learning or algorithm development. The device is a lateral flow immunoassay, which does not involve a training set as would be found in AI/ML applications. The studies described are for validation of the physical test device.
9. How the ground truth for the training set was established
- As there is no "training set" for this type of device, this question is not applicable. The underlying principles of the immunoassay are based on chemical reactions, not on data learned from a training set.