K Number
K220046

Validate with FDA (Live)

Date Cleared
2022-12-13

(342 days)

Product Code
Regulation Number
862.3650
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Superbio Fentanyl Urine Detection Kit is a fluorescence immunoassay intended for the qualitative detection of fentanyl in human urine at a cutoff concentration of 1.0 ng/mL. The assay is intended for use with Superbio Immunofluorescence Analyzer EASY-11. This in vitro diagnostic device is for prescription use only.

The test provides only preliminary test results. A more specific alternative chemical method must be used in order to obtain a confirmed analytical result. GC/MS or LC/MS is the preferred confirmatory method.

Device Description

This test uses a lateral flow design with location-dependent lines and zones. The Immunofluorescence Analyzer EASY-11 scans the test strip and displays results. The sample is added to the sample well of the test card, and the sample is drawn by capillary action into and through the fluorescent labeled pad, through the nitrocellulose strip and into the adsorption pad.

Within the fluorescent labeled pad, the specimen comes into contact with antibodies conjugated with fluorescent microspheres. During this interaction, if the amount of fentanyl antigen in the sample is greater than or equal to the detection limit, the antigen in the sample and the fluorescence-labeled antibody bind to the FTY antigen-antibody complex when the sample passes through a pad of fluorescence-microbead-labeled antibody conjugate. As the sample flows and reaches the FTY antigen coated by the T-line of nitrocellulose membrane, the FTY antigen coated by the T-line and the antigen in the sample competitively bind the FTY antibody labeled with fluorescence, then the T-line captures no fluorescence signal. When the samples do not contain fentanyl antigen or levels below the detection limit, as the sample flow, fluorescent microsphere labeled antibody to nitrocellulose membrane T line captures fluorescent signal. Whether or not FTY antigen was present in the sample, the rabbit IgG fluorescent microsphere conjugate not bound to the test line continued to flow with the rest of the sample and soon encountered a control line composed of sheep antirabbit IgG. The position of C-line will accumulate fluorescence signal. The C-line control area was scanned to confirm that adequate sample flow had occurred. High resolution, narrow band SMD LED was used as light source in the Immunofluorescence Analyzer. The central wavelength of the excitation spectrum is 365nm. The central response wavelength is 610nm.

AI/ML Overview

Here's a summary of the acceptance criteria and the study proving the device meets them, based on the provided text.

Acceptance Criteria and Device Performance for Superbio Fentanyl Urine Detection Kit

1. Table of Acceptance Criteria and Reported Device Performance

The document does not explicitly state formal "acceptance criteria" with numerical thresholds for performance metrics. Instead, the performance characteristics are reported as results of various analytical studies.

Table 1: Analytical Performance of the Superbio Fentanyl Urine Detection Kit

Performance CharacteristicAcceptance Criteria (Implied/Standard for IVD)Reported Device Performance
PrecisionConsistent results around the cutoff.Lot 1: -100% cut off: 60-/0+ -75% cut off: 60-/0+ -50% cut off: 60-/0+ -25% cut off: 56-/4+ cut off: 27-/33+ +25% cut off: 56+/4- +50% cut off: 60+/0- +75% cut off: 60+/0- +100% cut off: 60+/0- Lot 2 & 3: Similar results, showing expected distribution of positive/negative results around the cutoff.
StabilityStable for a specified duration at certain conditions.Stable at 4-30 °C for 12 months (based on accelerated stability at 45 °C). Real-time study ongoing.
InterferenceNo significant interference from common substances/conditions at specified concentrations.No interference observed from 60+ listed compounds/substances at specified concentrations (e.g., Acetaminophen, Ibuprofen, Albumin (100mg/dL), Glucose (3000mg/dL), pH 4-9, Specific Gravity 1.000-1.035). All samples at and above +50% Cut-Off were positive; all at and below -50% Cut-Off were negative.
Specificity (Cross-Reactivity)Minimal or no cross-reactivity with structurally similar but distinct compounds or common opioids not meant to be detected at the specified cutoff.Nor-fentanyl: 0.01% cross-reactivity (10000 ng/mL required for positive). Acetyl fentanyl / Acrylfentanyl: 83.33% cross-reactivity (1.20 ng/mL required for positive). Other fentanyl analogs: Varied cross-reactivity (e.g., Isobutyryl fentanyl 66.67%, Furanyl fentanyl 57.14%). No cross-reactivity observed with a long list of other opioid compounds (e.g., Morphine, Codeine, Hydrocodone) at 100 ug/mL (or specified lower concentrations for some).
Method Comparison (Clinical Concordance)High agreement with a validated confirmatory method (LC/MS) across negative, positive, and near-cutoff samples.Site 1: - 7/7 low negative samples correctly identified. - 17/17 very low negative samples correctly identified. - 13/16 near-cutoff negative samples correctly identified. - 21/24 near-cutoff positive samples correctly identified. - 16/16 high positive samples correctly identified. Site 2 & 3: Very similar results with minimal variations in near-cutoff performance. Total 80 samples per site.

2. Sample size used for the test set and data provenance

  • Test Set Sample Size:
    • Precision Study: For each of 9 concentration levels, 6 runs per day for 10 days per device lot were performed. With 3 lots, this implies 9 concentrations * 6 runs * 10 days * 3 lots = 1620 individual tests (though not explicitly stated as unique samples for each test). Samples were prepared by spiking fentanyl in negative samples.
    • Interference Study: Not explicitly stated, but "three batches of each device" were used, and samples were drug-free urine and target drug fentanyl urine spiked at specific concentrations.
    • Specificity (Cross-Reactivity) Study: "Three batches of device" were used with various fentanyl analogs and other opioid compounds.
    • Effect of Urine Specific Gravity and Urine pH: "Three lots of device" were used.
    • Method Comparison Study: 80 unaltered clinical samples were used per site across three different testing sites, totaling 240 clinical samples.
  • Data Provenance: The document does not explicitly state the country of origin for the clinical samples used in the method comparison study. The samples were described as "unaltered clinical samples." It does not specify if these were retrospective or prospective, but the context of "clinical samples" for method comparison suggests they were collected for diagnostic purposes, which could be either.

3. Number of experts used to establish the ground truth for the test set and qualifications of those experts

  • Ground Truth for Test Set: The ground truth for the clinical samples in the method comparison study was established by LC/MS (Liquid Chromatography/Mass Spectrometry). This is a highly accurate and commonly accepted confirmatory method for drug testing, often considered the gold standard.
  • Number and Qualifications of Experts: The document does not mention the use of human experts to establish the ground truth for the test set. LC/MS results are objective chemical measurements, not subjective expert interpretations.

4. Adjudication method for the test set

  • Not applicable as the ground truth was established by LC/MS, which is an objective analytical method, not human interpretation requiring adjudication.

5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

  • No MRMC comparative effectiveness study was done or reported. This device is an in vitro diagnostic (IVD) kit that is read by an automated immunofluorescence analyzer (Superbio Immunofluorescence Analyzer EASY-11), not by human readers interpreting images. Therefore, the concept of human readers improving with AI assistance is not relevant in this context.

6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done

  • Yes, the performance studies described are essentially standalone algorithm performance (though the "algorithm" is embedded in the immunoassay and analyzer's interpretation). The Superbio Fentanyl Urine Detection Kit, combined with the Superbio Immunofluorescence Analyzer EASY-11, functions as a system that provides a qualitative result (positive/negative) based on the fluorescent signal. The method comparison study directly compares the device's output to the LC/MS ground truth, representing its standalone performance. The operator's role is to run the test, not interpret the result.

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

  • LC/MS (Liquid Chromatography/Mass Spectrometry) was used as the ground truth for the clinical samples in the method comparison study. For the analytical studies (precision, interference, specificity), ground truth was established by preparing samples with known concentrations (e.g., spiking fentanyl into negative urine) and confirming these concentrations with LC/MS where applicable.

8. The sample size for the training set

  • The document does not describe a training set or machine learning model in the conventional sense. The Superbio Fentanyl Urine Detection Kit and Superbio Immunofluorescence Analyzer EASY-11 operate based on established immunochemical principles and fixed analytical protocols, not a trainable artificial intelligence algorithm. Therefore, there is no "training set" to report.

9. How the ground truth for the training set was established

  • Not applicable, as there is no training set for this type of IVD device.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo features the letters 'FDA' in a blue square, followed by the words 'U.S. FOOD & DRUG ADMINISTRATION' in blue text.

December 13, 2022

Shenzhen Superbio Technology Co., Ltd. % Joe Shia, Director LSI 504 E Diamond Ave., Suite I Gaithersburg, MD 20877

Re: K220046

Trade/Device Name: Superbio Fentany] Urine Detection Kit, Superbio Immunofluorescence Analyzer EASY-11 Regulation Number: 21 CFR 862.3650 Regulation Name: Opiate Test System Regulatory Class: Class II Product Code: DJG, KHO Dated: August 29, 2022 Received: August 29, 2022

Dear Joe Shia:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part

{1}------------------------------------------------

801 and Part 809); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Digitally signed by Paula Paula Caposino -S Caposino -S Date: 2022.12.13
17:19:25 -05'00'

Paula V. Caposino, Ph.D. Acting Deputy Director Division of Chemistry and Toxicology Devices OHT7: Office of In Vitro Diagnostics Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K220046

Device Name Superbio Fentanyl Urine Detection Kit Superbio Immunofluorescence Analyzer EASY-11

Indications for Use (Describe)

The Superbio Fentanyl Urine Detection Kit is a fluorescence immunoassay intended for the qualitative detection of fentanyl in human urine at a cutoff concentration of 1.0 ng/mL. The assay is intended for use with Superbio Immunofluorescence Analyzer EASY-11. This in vitro diagnostic device is for prescription use only.

The test provides only preliminary test results. A more specific alternative chemical method must be used in order to obtain a confirmed analytical result. GC/MS or LC/MS is the preferred confirmatory method.

The Superbio Immunofluorescence Analyzer EASY-11 is a portable fluorescence instrument for in vitro diagnostic use only. The analyzer is designed to perform in vitro diagnostic tests on clinical specimens. This analyzer can be used in a laboratory or in a point-of-care setting.

Type of Use (Select one or both, as applicable)
---------------------------------------------------

Prescription Use (Part 21 CFR 801 Subpart D)

| Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

510(k) SUMMARY K220046

    1. Date: December 2, 2022 2. Submitter: Shenzhen Superbio Technology Co., Ltd. Building B. Xinzheng, 71 Area, Xinan Sub-District, Baoan Shenzhen, China 518101 3. Contact person: Joe Shia LSI International Inc. 504E Diamond Ave., Suite I Gaithersburg, MD 20877 Telephone: 240-505-7880 Email: shiajl@yahoo.com 4. Device Names: Superbio Fentanyl Urine Detection Kit Superbio Immunofluorescence Analyzer EASY-11
Classification:Class 2
Product CodeClassificationRegulation SectionPanel
DJGII21 CFR § 862.3650Opiate Test SystemToxicology (91)
KHOI21 CFR § 862.2560Fluorometer for clinical useClinical Chemistry
    1. Predicate Devices:
      The ARK™ Fentanyl Assay K180427; TRIAGE STATMETER K973547
    1. Indications for Use
      The Superbio Fentanyl Urine Detection Kit is a fluorescence immunoassay intended for the qualitative detection of fentanyl in human urine at a cutoff concentration of 1.0 ng/mL. The assay is intended for use with Superbio Immunofluorescence Analyzer EASY-11. This in vitro diagnostic device is for prescription use only.

The test provides only preliminary test results. A more specific alternative chemical must be used in order to obtain a confirmed analytical result. GC/MS or LC/MS is the preferred confirmatory method.

The Superbio Immunofluorescence Analyzer EASY-11 is a portable fluorescence instrument for in vitro diagnostic use only. The analyzer is designed to perform in vitro diagnostic tests on clinical specimens. This analyzer can be used in a laboratory or in a point-of-care setting.

    1. Device Description
      This test uses a lateral flow design with location-dependent lines and zones. The Immunofluorescence Analyzer EASY-11 scans the test strip and displays results. The sample is added to the sample well of the test card, and the sample is drawn by capillary action into and through the fluorescent labeled pad, through the nitrocellulose strip and into the adsorption pad.

{4}------------------------------------------------

Within the fluorescent labeled pad, the specimen comes into contact with antibodies conjugated with fluorescent microspheres. During this interaction, if the amount of fentanyl antigen in the sample is greater than or equal to the detection limit, the antigen in the sample and the fluorescencelabeled antibody bind to the FTY antigen-antibody complex when the sample passes through a pad of fluorescence-microbead-labeled antibody conjugate. As the sample flows and reaches the FTY antigen coated by the T-line of nitrocellulose membrane, the FTY antigen coated by the T-line and the antigen in the sample competitively bind the FTY antibody labeled with fluorescence, then the T-line captures no fluorescence signal. When the samples do not contain fentanyl antigen or levels below the detection limit, as the sample flow, fluorescent microsphere labeled antibody to nitrocellulose membrane T line captures fluorescent signal. Whether or not FTY antigen was present in the sample, the rabbit IgG fluorescent microsphere conjugate not bound to the test line continued to flow with the rest of the sample and soon encountered a control line composed of sheep antirabbit IgG. The position of C-line will accumulate fluorescence signal. The C-line control area was scanned to confirm that adequate sample flow had occurred. High resolution, narrow band SMD LED was used as light source in the Immunofluorescence Analyzer. The central wavelength of the excitation spectrum is 365nm. The central response wavelength is 610nm.

    1. Substantial Equivalence Information
      A summary comparison of features of the Superbio Fentanyl Urine Detection Kit and the predicate devices is provided in following table.
ItemDevicePredicate - K180427
Indication(s)for UseFor the qualitative determination offentanyl in human urine.Same
Calibrator and Cut-OffValuesFentanyl (FTY)1 ng/mlSame
MethodologyCompetitive binding, lateral flowimmunochromatographic assays basedon the principle of antigen antibodyimmunochemistry.Homogenous enzymeimmunoassay(EIA)
Type of TestQualitativeSame
Specimen TypeHuman UrineSame
Intended UseFor prescription useSame
ConfigurationsDip Card and StripCup
Platform RequiredImmunofluorescence AnalyzerAutomated clinical chemistryanalyzer
Storage4-30°C2-8°C

Table 1: Features Comparison of Superbio Fentanyl Urine Detection Kit and the Predicate Devices

Table 2: Instrument Similarities and Differences

ItemPredicate - K973547Subject Device
Intended Use/Indication for UseImmunofluorescence analyzerdesigned to perform in vitrodiagnostic tests on clinical specimensincluding drug urine test.Same
Principles of AssayOperationSandwich immunofluorescenceimmunoassaySame
Calibration CheckRun two levels or POS and NEG externalliquid control samples as appropriate witheach new lot of reagents and once every30 days with continued use of the samereagent lot numberA Quality control card is suppliedwith Easy-11 and used for checkEasy-11 optics and calculation.systems.
Development Modesone test mode• Run TestTwo basic assay developmentmodes:• Standard test: In standard test, theuser immediately inserts TestCassette into Easy-11 and click"start test". Easy-11 automaticallycounts the time.• Quick test: Manually timing, theninsert the test card into Easy-11, andclick "Start Test". The instrumentwill read the results.
User interface4 inch LCD Screen display8 inch Color LCD touchscreendisplay
Barcode scanner(sample)External Barcode Scanner (optional)Same
Assay/instrumentinterfaceDrawerSame
Light SourceLaser DiodeLED Light
Power Supply6v DC at 1 amp – supplied via 4 AAbatteries or AC/DC ConverterAC100-240V
Dimensions22.5 cm x 19 cm x 7 cm24.5 cm x 27 cm x 16 cm
Weight~1.5 lbs~4.5 lbs

{5}------------------------------------------------

    1. Test Principle
      Superbio Fentanyl Urine Detection Kit is a competitive and fluorescence immunochromatography assay, and uses fluorescence microspheres-labeled monoclonal antibody as the indicator marker to qualitatively detect fentanyl in human urine.

{6}------------------------------------------------

The test card contains fentanyl test strip. The nitrocellulose membrane test area (T) of the test strip is correspondingly coated with fentanyl-bovine serum albumin conjugate, and the quality control area (C) is coated with goat anti-rabbit IgG polyclonal antibody. Both Fentanyl monoclonal antibody and rabbit IgG polyclonal antibody labeled with fluorescent microspheres were embedded on the conjugate pad. When the concentration of fentany] in the sample is higher than or equal to the cut-off of the product, it will compete with the corresponding conjugate coated on the test area (T) to bind to the fluorescently-labeled monoclonal antibody, the fluorescence rendering of the test line is inhibited and the result is positive; while when the sample does not contain fentanyl or its concentration is lower than the cut-off of the product, the corresponding conjugate on the test line reacts with sufficient fluorescently-labeled monoclonal antibodies, the test line will have fluorescence and the result is negative. No matter whether the sample contains the corresponding analyte or not, the quality control area (C) will develop fluorescence, which is the criteria for judging whether the chromatography process is normal or not. The test card is tested by the Superbio Immunofluorescence Analyzer EASY-11 and the results are interpreted by the analyzer.

10. Performance Characteristics

    1. Analytical Performance
    • a. Precision

Precision studies were carried out for samples with concentrations of -100% cut off, -75% cut off, -50% cut off, -25% cut off, +25% cut off, +50% cut off, +75% cut off and +100% cut off. These samples were prepared by spiking fentanyl in negative samples. Each fentanyl concentration was confirmed by LC/MS. All sample aliquots were blindly labeled by the person who prepared the samples and didn't take part in the sample testing. For each concentration, tests were performed six runs per day for 10 days per device lot in a randomized order. Each device was read on one Superbio Immunofluorescence Analyzer EASY-11. The results obtained are summarized in the following tables.

LotNumber-100%cut off-75%cut off-50%cut off-25%cutoffcut off+25%cut off+50%cut off+75%cut off+100%cut off
Lot 160-/0+60-/0+60-/0+56-/4+27-/33+56+/4-60+/0-60+/0-60+/0-
Lot 260-/0+60-/0+60-/0+57-/3+32-/28+57+/3-60+/0-60+/0-60+/0-
Lot 360-/0+60-/0+60-/0+56-/4+27-/33+55+/5-60+/0-60+/0-60+/0-

c. Stability

The devices are stable at 4-30 ℃ for 12 months based on the accelerated stability study at 45 °C. The real time stability study is ongoing.

d. Interference

Potential interfering substances found in human urine of physiological or pathological conditions were added to drug-free urine and target drug fentanyl urine with concentrations at 50% below and 50% above Cut-Off levels. These urine samples were tested using three batches of each device. Compounds that showed no interference at a concentration of 100µg/mL or specified concentrations are summarized in the following tables.

AcetaminophenDoxepin (50ug/ml)Nortriptyline (25ug/ml)
Acetone (1000mg/dL)Ecgonine methyl esterNoscapine
AcetophenetidinEphedrineO-Hydroxyhippuric acid

{7}------------------------------------------------

Acetylsalicylic acidErythromycinOctopamine
Albumin (100mg/dL)Ethanol (1%)Oxalic acid (100 mg/dL)
AlbuterolFenoprofenOxazepam
AminopyrineFluphenazineOxolinic acid
Amitriptyline (35ug/ml)FurosemideOxymetazoline
AmobarbitalGalactose (10mg/dL)Papaverine
AmoxicillinGamma Globulin (500mg/dL)Penicillin G
AmpicillinGentisic acidPerphenazine
ApomorphineGlucose (3000mg/dL)Phencyclidine
Ascorbic acidHemoglobinPhenelzine
AspartameHydralazinePhenobarbital
AtropineHydrochlorothiazidePrednisone
Benzilic acidHydrocortisonePropoxyphene (50ug/ml)
Benzoic acidHydroxytyraminePropranolol
BenzoylecgonineIbuprofenPseudoephedrine
BilirubinImipramine (30ug/ml)Quinine
Boric Acid (1%)IsoproterenolRanitidine
Bupropion (50ug/ml)IsoxsuprineRiboflavin (7.5mg/dL)
CaffeineKetamineSalicylic acid
CarbamazepineKetoprofenSecobarbital
Chloral hydrateLabetalolSerotonin (5-Hydroxytyramine)
ChloramphenicolLidocaine (50ug/ml)Sulfamethazine
ChlorothiazideLoperamideSulindac
ChlorpromazineMaprotiline (50ug/ml)Tetrahydrocortisone 3-(β- Dglucuronide)
CholesterolMeperidineTetrahydrocortisone 3-acetate
Clomipramine (50ug/ml)MeprobamateTetrahydrozoline
ClonidineMethapyrilene (10ug/ml)Thiamine
CortisoneMethaqualone (50ug/ml)Thioridazine
CotinineMethoxyphenamineTriamterene
CreatinineMetronidazole (300ug/ml)Trifluoperazine
Cyclobenzaprine(10ug/ml)N-AcetylprocainamideTrimethoprim
DeoxycorticosteroneNaCl (4000mg/dL)Tyramine
Desipramine (50ug/ml)Nalidixic acidUrea (2000mg/dL)
DextromethorphanNaloxoneUric acid
DiclofenacNaltrexoneValproic acid (250ug/ml)
DiflunisalNaproxenVenlafaxine
DigoxinNiacinamideVerapamil
DiphenhydramineNicotine (10ug/ml)Zomepirac
DL-TryptophanNifedipineβ-Estradiol

e.Specificity

To test specificity, drug metabolites and other components that are likely to interfere in urine samples were tested using three batches of device. The lowest concentration that caused a positive result for each compound are listed below.

Fentanyl (Cutoff=1ng/mL)Minimumconcentration requiredto obtain a positiveresult (ng/mL)% Cross-Reactivity
Norfentanyl100000.01
Acetyl fentanyl1.2083.33

{8}------------------------------------------------

Acrylfentanyl1.2083.33
ω-1-Hydroxyfentanyl200000.005
Isobutyryl fentanyl1.5066.67
Ocfentanil1.5066.67
Butyryl fentanyl1.6062.50
Furanyl fentanyl1.7557.14
Valeryl fentanyl2.5040.00
(±) β-hydroxythiofentanyl2.8035.71
4-Fluoro-isobutyrylfentanyl3.0033.33
Para-fluorobutyrylfentanyl (p-FBF)3.0033.33
Para-fluoro fentanyl3.0033.33
(±)-3-cis-methylfentanyl5.0020.00
Carfentanil5000.20
Sufentanil6250.16
Acetyl norfentanyl10,0000.01

The following opioids compounds were tested at a concentration of 100ug/mL (despropionyl fentanyl (4-ANPP) was tested at 50 ug/mL, norcarfentanil was tested at 5 ug/mL and remifentanil was tested at 10 ug/mL). Negative results were obtained for all these compounds. There is no cross-reactivity for these compounds using the Superbio Fentanyl Urine Detection Kit

6-Acetyl morphineNaltrexone
AmphetamineNorbuprenorphine
BuprenorphineNorcodeine
BuprenorphineglucuronideNorketamine
CodeineNormeperidine
DextromethorphanNormorphine
DihydrocodeineNoroxycodone
EDDPOxycodone
EMDPOxymorphone
FluoxetinePentazocine (Talwin)
HeroinPipamperone
HydrocodoneRisperidone
HydromorphoneTapentadol
KetamineThioridazine
LevorphanolTilidine
MeperidineTramadol
MethadoneTramadol-O- Desmethyl
MorphineTramadol-N- Desmethyl
Morphine-3-glucuronideTrazodone
NaloxoneAlfentanil
Despropionyl fentanyl (4-ANPP) (50 ug/mL)Norcarfentanil (5 ug/mL)
Norcarfentanil (10 ug/mL)

f. Effect of Urine Specific Gravity and Urine pH

To investigate the effect of urine specific gravity and urine pH, urine samples, with 1.000 to 1.035 specific gravity or urine samples with pH 4 to 9 were spiked with target fentanyl at 50% below and 50% above Cut-Off levels. These samples were tested using three lots

{9}------------------------------------------------

of device. Results were all positive for samples at and above +50% Cut-Off and all negative for samples at and below -50% Cut-Off.

2. Comparison Studies

Method comparison studies for the Superbio Fentanyl Urine Detection Kit were performed at three different testing sites. Operators ran 80 (40 negative and 40 positive) unaltered clinical samples. The samples were blind labeled and compared to LC/MS results are presented in the tables below.

NegativeLow Negative byLC/MS(less than-50%)Near CutoffNegative byLC/MS(Between-50% andcutoff)Near CutoffPositive byLC/MS(Between thecutoff and+50%)High Positiveby LC/MS(greater than+50%)
Site1Positive0032116
Negative7171330
Site2Positive0032116
Negative7171330
Site3Positive0032216
Negative7171320

Discordant Results

OperatorSample NumberLC/MS ResultSUPERBIOResults
Site 1FM0150.984Positive
Site 1FM0260.866Positive
Site 1FM0580.921Positive
Site 1FM0311.05Negative
Site 1FM0511.18Negative
Site 1FM0651.07Negative
Site 2FM0120.833Positive
Site 2FM0150.984Positive
Site 2FM0260.866Positive
Site 2FM0131.09Negative
Site 2FM0741.1Negative
Site 2FM0981.14Negative
Site 3FM0120.833Positive
Site 3FM0160.953Positive
Site 3FM0320.815Positive
Site 3FM0651.07Negative
Site 3FM0911.12Negative
    1. Clinical Studies
      Not applicable.
    1. Conclusion

{10}------------------------------------------------

Based on the test principle and acceptable performance characteristics including precision, cut-off, interference, specificity, and method comparison studies of the devices, it's concluded a substantial equivalence decision.

§ 862.3650 Opiate test system.

(a)
Identification. An opiate test system is a device intended to measure any of the addictive narcotic pain-relieving opiate drugs in blood, serum, urine, gastric contents, and saliva. An opiate is any natural or synthetic drug that has morphine-like pharmocological actions. The opiates include drugs such as morphine, morphine glucoronide, heroin, codeine, nalorphine, and meperedine. Measurements obtained by this device are used in the diagnosis and treatment of opiate use or overdose and in monitoring the levels of opiate administration to ensure appropriate therapy.(b)
Classification. Class II (special controls). An opiate test system is not exempt if it is intended for any use other than employment or insurance testing or is intended for Federal drug testing programs. The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 862.9, provided the test system is intended for employment and insurance testing and includes a statement in the labeling that the device is intended solely for use in employment and insurance testing, and does not include devices intended for Federal drug testing programs (e.g., programs run by the Substance Abuse and Mental Health Services Administration (SAMHSA), the Department of Transportation (DOT), and the U.S. military).