K Number
K240698

Validate with FDA (Live)

Date Cleared
2024-05-10

(57 days)

Regulation Number
862.3650
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

Dochek® Multi-Drug Urine Test Dipcard Rx is an immunoassay for the qualitative determination of single or multiple drugs in human urine at the following cutoff concentrations:
Amphetamine (AMP) 1000 or 500 ng/mL
Secobarbital (BAR) 300 ng/mL
Buprenorphine (BUP) 10 ng/mL
Oxazepam (BZO) 300 ng/mL
Cocaine (COC) 300 or 150 ng/mL
2-ethylidene-1,5-dimethyl-3,3-diphenylpyrrolidine (EDDP) 300 ng/mL
Methylenedioxymethamphetamine (MDMA) 500 ng/mL
Methamphetamine (MET) 1000 or 500 ng/mL
Morphine (MOP/OPI) 300 or 2000 ng/mL
Methadone (MTD) 300 ng/mL
Oxycodone (OXY) 100 ng/mL
Phencyclidine (PCP) 25 ng/mL
Propoxyphene (PPX) 300 ng/mL
Nortriptyline (TCA) 1000 ng/mL
Cannabinoids (THC) 50 ng/mL
6-Monoacetylmorphine (6-MAM) 10 ng/mL
Dochek® Multi-Drug Urine Test Dipcard Rx offers any combinations from 1 to 16 drugs but only one cutoff concentration under same drug condition will be included per device. It is intended for prescription use. For in vitro diagnostic use only.
The test provides only preliminary results. Clinical consideration and professional judgment should be applied to any drug of abuse test result, particularly in evaluating a preliminary positive result. To obtain a confirmed analytical result, a more specific alternate chemical method is needed. GC/MS is the recommended confirmatory method.

Dochek® Multi-Drug Urine Test Dipcard is an immunoassay for the qualitative determination of single or multiple drugs in human urine at the following cutoff concentrations:
Amphetamine (AMP) 1000 or 500 ng/mL
Secobarbital (BAR) 300 ng/mL
Buprenorphine (BUP) 10 ng/mL
Oxazepam (BZO) 300 ng/mL
Cocaine (COC) 300 or 150 ng/mL
2-ethylidene-1,5-dimethyl-3,3-diphenylpyrrolidine (EDDP) 300 ng/mL
Methylenedioxymethamphetamine (MDMA) 500 ng/mL
Methamphetamine (MET) 1000 or 500 ng/mL
Morphine (MOP/OPI) 300 or 2000 ng/mL
Methadone (MTD) 300 ng/mL
Oxycodone (OXY) 100 ng/mL
Phencyclidine (PCP) 25 ng/mL
Propoxyphene (PPX) 300 ng/mL
Nortriptyline (TCA) 1000 ng/mL
Cannabinoids (THC) 50 ng/mL
6-Monoacetylmorphine (6-MAM) 10 ng/mL
Dochek® Multi-Drug Urine Test Dipcard offers any combinations from 1 to 16 drugs but only one cutoff concentration under same drug condition will be included per device. It is intended for over-the-counter (OTC) use. For in vitro diagnostic use only.
The test provides only preliminary results. Clinical consideration and professional judgment should be applied to any drug of abuse test result, particularly in evaluating a preliminary positive result. To obtain a confirmed analytical result, a more specific alternate chemical method is needed. GC/MS or LC/MS is the recommended confirmatory method.

Device Description

Dochek® Multi-Drug Urine Test Dipcard Rx and Dochek® Multi-Drug Urine Test Dipcard are immunochromatographic assays that use a lateral flow system for the qualitative detection of single or multiple drugs in human urine at or above the cut-off levels as indicated. The products are singleuse in vitro diagnostic devices.
This device is a dipcard format in which the test strips are integrated into the plastic dipcard. After removing the cap of the dipcard, the absorbent end of the test strips is exposed and can be in direct contact with the urine sample. The device is in a ready-to-use format and no longer requires assembly before use.

AI/ML Overview

Here's a breakdown of the acceptance criteria and study details for the Dochek® Multi-Drug Urine Test Dipcard, based on the provided FDA 510(k) summary:

This device is an in-vitro diagnostic test, not an AI/ML device, so many of the requested criteria (e.g., number of experts, adjudication, MRMC study, training set) typically apply to AI/ML software. Therefore, I will adapt the response to describe the data provided for this type of diagnostic device.


Acceptance Criteria and Device Performance for Dochek® Multi-Drug Urine Test Dipcard

This device is a qualitative immunoassay for the detection of various drugs in human urine. The acceptance criteria and performance are primarily based on its analytical accuracy (precision, specificity, and comparison to a gold standard) and usability.

1. Table of Acceptance Criteria and Reported Device Performance

For this type of qualitative immunoassay, the primary acceptance criteria revolve around the device's ability to correctly classify samples as positive or negative at and around the specified cutoff concentrations. The "performance" is demonstrated through the observed concordance with confirmatory methods (LC-MS/MS) and its precision.

Acceptance Criteria CategorySpecific Criteria (Implicitly from Study Design)Reported Device Performance (Summary)
Analytical PrecisionConsistent positive results for samples ≥ 25% above cutoff. Consistent negative results for samples ≤ 25% below cutoff. Acceptable percentage of positive/negative results at cutoff.Observed Precision (Examples from various drugs and lots at cutoff): - AMP (1000 ng/mL): Lot I: 13-/37+, Lot II: 14-/36+, Lot III: 13-/37+ (out of 50 tests) - BAR (300 ng/mL): Lot I: 14-/36+, Lot II: 12-/38+, Lot III: 14-/36+ - COC (300 ng/mL): Lot II: 13-/37+, Lot III: 12-/38+ - MDMA (500 ng/mL): Lot I: 10-/40+, Lot II: 10-/40+, Lot III: 11-/39+ All samples ≥ +25% cutoff consistently showed "0-/50+" (0 negative, 50 positive). All samples ≤ -25% cutoff consistently showed "50-/0+" (50 negative, 0 positive).
Analytical SpecificityMinimal or no cross-reactivity with structurally unrelated compounds at specified concentrations. Acceptable cross-reactivity with structurally related compounds.Extensive cross-reactivity tables provided (pages 13-22 in original document) for all 16 drugs. Many structurally related compounds showed cross-reactivity, requiring careful interpretation of results (e.g., Amphetamine panel detects other amphetamine-like substances). Many non-structurally related compounds showed no interference at 100 µg/mL.
InterferenceUrinary pH (4-9) and Specific Gravity (1.000-1.035) should not affect results. Substances commonly found in urine should not interfere.pH levels of 4-9 and specific gravity levels of 1.000-1.035 did not affect assay results. Over 100 non-structurally related compounds (e.g., acetaminophen, ibuprofen, glucose) showed no interference at 100 µg/mL. (Pages 22-23)
Method Comparison (Accuracy)High concordance with a confirmed analytical method (LC-MS/MS) for both negative and positive samples, especially around the cutoff concentrations.High concordance with LC-MS/MS. For each drug, 80 clinical samples (40 negative, 40 positive) were tested. The detailed tables (pages 24-27) show the breakdown of positive/negative calls by the dipcard vs. LC-MS/MS concentrations (drug-free, low negative, near cutoff negative, near cutoff positive, high positive). While some discordant results (false positives/negatives near cutoff) were observed, they are typical for qualitative immunoassay screening tests designed to be highly sensitive around the cutoff. The device generally performed as expected for a qualitative screening test.
Lay Person UsabilityLay users should be able to correctly interpret the device results and understand the instructions for use.High (≥90%) correct result interpretation by lay users across various drug concentrations for the majority of tests. The "Percentage of correct results" for samples +/- 25% of cutoff varied (e.g., BAR 95%, BUP 90% for -25% cutoff; BUP 95% for +25% cutoff), indicating some variability by lay users near the cutoff, which is expected for visual interpretation of a qualitative test. All participants found the instructions easy to understand and follow, with a Flesch-Kincaid Grade Level of 7. (Pages 28-33)
StabilityThe device should demonstrate stability over its claimed shelf life.Demonstrated stability at 2-30°C for 36 months based on real-time stability study.

2. Sample Size and Data Provenance:

  • Test Set Sample Sizes:
    • Precision/Reproducibility: For each drug and each of the three lots, 50 tests were performed at each of the 9 concentration levels around the cutoff (total of 450 tests per lot per drug for analytical precision). This was repeated for multiple lots.
    • Method Comparison: 80 clinical urine samples were used for each drug. These 80 samples consisted of 40 negative and 40 positive samples, further categorized by their concentration relative to the cutoff (drug-free, low negative, near cutoff negative, near cutoff positive, high positive).
    • Lay Person Study: 140 participants (79 male, 61 female for Configuration 1; 73 male, 67 female for Configuration 2). For each drug and concentration level, 20 samples were tested by lay users.
  • Data Provenance: The document generally indicates "in-house" studies for method comparison and precision. The clinical samples for method comparison were "unaltered urine clinical samples." The lay person study used pooled urine specimens spiked with drugs, confirmed by LC-MS/MS. The country of origin for the data is implied to be China, given the manufacturer's address (Guangzhou, China). The studies appear to be prospective in nature as they involved preparing samples and testing them using the device under controlled conditions to evaluate performance.

3. Number of Experts and Qualifications for Ground Truth:

  • This device is a qualitative immunoassay, not an AI/ML algorithm requiring expert interpretation of images. Therefore, the concept of "experts" to establish ground truth in the same way as for imaging AI is not directly applicable.
  • The ground truth for the analytical studies was established by LC-MS/MS (Liquid Chromatography-Mass Spectrometry/Mass Spectrometry). This is a highly accurate chemical method considered the gold standard for quantitative drug identification and concentration determination in urine.
  • For the method comparison study, "three operators" performed the tests. Their qualifications are not explicitly stated, but for a laboratory-based study of an immunoassay, these would typically be trained laboratory personnel or technicians.

4. Adjudication Method for the Test Set:

  • Not applicable in the context of an immunoassay device. The results are physical readouts (presence/absence of a line).
  • For the Method Comparison study, the device result (positive/negative) was compared directly to the quantitative LC-MS/MS result. Discordant results are individually listed (pages 27-28), showing which operators, if any, had differing readings from the LC-MS/MS ground truth. Since three operators independently read the same samples for the method comparison, consistency among them can be inferred, but formal "adjudication" is not described as they were simply recording what they observed.

5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study:

  • No, an MRMC comparative effectiveness study was not conducted for this device. This type of study is specifically designed for AI-assisted diagnostic tools (often in imaging) to compare human performance with and without AI assistance.
  • This device is a standalone diagnostic test; it does not "assist" human readers in the interpretation of complex data (like medical images).

6. Standalone (Algorithm Only) Performance:

  • Yes, in a sense. The "Performance Characteristics" section details the device's inherent analytical performance (precision, specificity, and method comparison against LC-MS/MS). These studies evaluate the device's ability to detect drugs independently of human interpretation nuances (though human "operators" did conduct the tests, the output is a visual line).
  • The "Lay Person Study" then evaluates the human interpretation of the device's output.

7. Type of Ground Truth Used:

  • The primary ground truth used throughout the studies (precision, method comparison) was GC/MS or LC/MS (Gas Chromatography/Mass Spectrometry or Liquid Chromatography/Mass Spectrometry) results. These are highly accurate, quantitative analytical chemistry methods considered definitive for drug presence and concentration.
  • For the lay person study, the "ground truth" for the samples was also established by LC-MS/MS confirmation of spiked drug concentrations.

8. Sample Size for the Training Set:

  • This device is a chemical immunoassay, not a machine learning model. Therefore, there is no "training set" in the context of algorithm development. The device's performance is inherent to its biochemical design.

9. How Ground Truth for Training Set was Established:

  • As stated above, no training set for an algorithm exists for this type of device. The scientific and engineering principles of immunoassay design and manufacturing determine its performance, which is then validated through the analytical and clinical studies described.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. Underneath the square are the words "U.S. FOOD & DRUG ADMINISTRATION" in blue.

Guangzhou Decheng Biotechnology Co., Ltd. Mango Huang Regulatory Affairs Manager Floor 3/4/5/7, Building A1, No.12, Nanyun 1st Road Science City, Huangpu District Guangzhou, 510663 China

Re: K240698

Trade/Device Name: Dochek® Multi-Drug Urine Test Dipcard Rx; Dochek® Multi-Drug Urine Test Dipcard Regulation Number: 21 CFR 862.3650 Regulation Name: Opiate Test System Regulatory Class: Class II Product Code: DJG, NFT, DKZ, PTH, DIS, NGL, NFV, JXM, NFY, DIO, PTG, DJR, NGG, DJC, NGM, LCM, QBF, JXN, QAW, LFG, NFW, LDJ Dated: March 14, 2024 Received: March 14, 2024

Dear Mango Huang:

We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{1}------------------------------------------------

Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).

Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30. Design controls; 21 CFR 820.90. Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801 and Part 809); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safetyreporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Joseph A. Digitally signed by Joseph A. Kotarek - Kotarek - S Date: 2024.05.09 S 14:56:49 -04'00'

Joseph Kotarek Branch Chief

{2}------------------------------------------------

Division of Chemistry and Toxicology Devices OHT7: Office of In Vitro Diagnostics Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{3}------------------------------------------------

Indications for Use

510(k) Number (if known) K240698

Device Name

Dochek® Multi-Drug Urine Test Dipcard Rx

Indications for Use (Describe)

Dochek® Multi-Drug Urine Test Dipcard Rx is an immunoassay for the qualitative determination of single or multiple drugs in human urine at the following cutoff concentrations:

Drug (Identifier)
Amphetamine (AMP)
Secobarbital (BAR)
Buprenorphine (BUP)
Oxazepam (BZO)
Cocaine (COC)
2-ethylidene-1,5-dimethyl-3,3-diphenylpyrrolidine (EDDP)
Methylenedioxymethamphetamine (MDMA)
Methamphetamine (MET)
Morphine (MOP/OPI)
Methadone (MTD)
Oxycodone (OXY)
Phencyclidine (PCP)
Propoxyphene (PPX)
Nortriptyline (TCA)
Cannabinoids (THC)
6-Monoacetylmorphine (6-MAM)

Cut-off level 1000 or 500 ng/mL 300 ng/mL 10 ng/mL 300 ng/mL 300 or 150 ng/mL 300 ng/mL 500 ng/mL 1000 or 500 ng/mL 300 or 2000 ng/mL 300 ng/mL 100 ng/mL 25 ng/mL 300 ng/mL 1000 ng/mL 50 ng/mL 10 ng/mL

Dochek® Multi-Drug Urine Test Dipcard Rx offers any combinations from 1 to 16 drugs but only one cutoff concentration under same drug condition will be included per device. It is intended for prescription use. For in vitro diagnostic use only.

The test provides only preliminary results. Clinical consideration and professional judgment should be applied to any drug of abuse test result, particularly in evaluating a preliminary positive result. To obtain a confirmed analytical result, a more specific alternate chemical method is needed. GC/MS is the recommended confirmatory method.

Type of Use (Select one or both, as applicable)

X Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

{4}------------------------------------------------

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{5}------------------------------------------------

Indications for Use

510(k) Number (if known) K240698

Device Name

Dochek® Multi-Drug Urine Test Dipcard

Indications for Use (Describe)

Dochek® Multi-Drug Urine Test Dipcard is an immunoassay for the qualitative determination of single or multiple drugs in human urine at the following cutoff concentrations:

Cut-off level 1000 or 500 ng/mL

300 ng/mL 10 ng/mL 300 ng/mL

300 ng/mL 100 ng/mL 25 ng/mL 300 ng/mL 1000 ng/mL 50 ng/mL 10 ng/mL

300 or 150 ng/mL 300 ng/mL 500 ng/mL

1000 or 500 ng/mL 300 or 2000 ng/mL

Drug (Identifier)
Amphetamine (AMP)
Secobarbital (BAR)
Buprenorphine (BUP)
Oxazepam (BZO)
Cocaine (COC)
2-ethylidene-1,5-dimethyl-3,3-diphenylpyrrolidine (EDDP)
Methylenedioxymethamphetamine (MDMA)
Methamphetamine (MET)
Morphine (MOP/OPI)
Methadone (MTD)
Oxycodone (OXY)
Phencyclidine (PCP)
Propoxyphene (PPX)
Nortriptyline (TCA)
Cannabinoids (THC)
6-Monoacetylmorphine (6-MAM)

Dochek® Multi-Drug Urine Test Dipcard offers any combinations from 1 to 16 drugs but only one cutoff concentration under same drug condition will be included per device. It is intended for over-the-counter (OTC) use. For in vitro diagnostic use only.

The test provides only preliminary results. Clinical consideration and professional judgment should be applied to any drug of abuse test result, particularly in evaluating a preliminary positive result. To obtain a confirmed analytical result, a more specific alternate chemical method is needed. GC/MS or LC/MS is the recommended confirmatory method.

Type of Use (Select one or both, as applicable)

Prescription Use (Part 21 CFR 801 Subpart D)

X Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

{6}------------------------------------------------

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{7}------------------------------------------------

510(k) SUMMARY

510(k) number: K240698

  • Date: May 9, 2024 1. Submitter: Submitter: Guangzhou Decheng Biotechnology Co., Ltd. 2. Address: Floor 3/4/5/7, Building A1, No.12, Nanyun 1st Road, Science City, Huangpu District, Guangzhou, Guangdong, 510663, P.R. China Contact Person: Mango Huang Contact Email Address: mango.huang(@)dochekbio.com Telephone: +86-020-82557192 Correspondent: Correspondent: Guangzhou Decheng Biotechnology Co., Ltd. 3. Address: Floor 3/4/5/7, Building A1, No.12, Nanyun 1st Road, Science City, Huangpu District, Guangzhou, Guangdong, 510663, P.R. China Contact Person: Mango Huang Contact Email Address: mango.huang(@dochekbio.com Telephone: (888) 695-5248 4. Device Name: Dochek® Multi-Drug Urine Test Dipcard Rx Dochek® Multi-Drug Urine Test Dipcard

5. Classification: Class II

Product CodeRegulation SectionPanel
Target Drug
DKZ, NFTAmphetamine (AMP)862.3100, Amphetamine Test SystemToxicology
DIS, PTHSecobarbital (BAR)862.3150, Barbiturate Test SystemToxicology
DJG, NGLBuprenorphine (BUP)Morphine (MOP/OPI)Oxycodone (OXY)6-Monoacetylmorphine (6-MAM)862.3650, Opiate Test SystemToxicology
JXM, NFVOxazepam (BZO)862.3170, Benzodiazepine Test SystemToxicology
DIO, NFYCocaine (COC)862.3250, Cocaine and cocainemetabolite test systemToxicology
DJR, PTG2-ethylidene-1,5-dimethyl-3,3-diphenylpyrrolidine (EDDP)Methadone (MTD)862.3620, Methadone Test SystemToxicology

{8}------------------------------------------------

DJC, NGGMethylenedioxymethamphetamine(MDMA)Methamphetamine (MET)862.3610,Methamphetamine Test SystemToxicology
LCM, NGMPhencyclidine (PCP)UnclassifiedToxicology
JXN, QBFPropoxyphene (PPX)862.3700 Propoxyphene test system.Toxicology
LFG, QAWNortriptyline (TCA)862.3910 Tricyclic antidepressantdrugs test systemToxicology
LDJ, NFWCannabinoids (THC)862.3870, Cannabinoids Test SystemToxicology

Predicate Devices 6.

K232659Dochek® Multi-Drug Urine Test Cup Rx
K232659Dochek® Multi-Drug Urine Test Cup

7. Intended Use

Dochek® Multi-Drug Urine Test Dipcard Rx is an immunoassay for the qualitative determination of single or multiple drugs in human urine at the following cutoff concentrations:

Drug (Identifier)Cut-off level
Amphetamine (AMP)1000 or 500 ng/mL
Secobarbital (BAR)300 ng/mL
Buprenorphine (BUP)10 ng/mL
Oxazepam (BZO)300 ng/mL
Cocaine (COC)300 or 150 ng/mL
2-ethylidene-1,5-dimethyl-3,3-diphenylpyrrolidine (EDDP)300 ng/mL
Methylenedioxymethamphetamine (MDMA)500 ng/mL
Methamphetamine (MET)1000 or 500 ng/mL
Morphine (MOP/OPI)300 or 2000 ng/mL
Methadone (MTD)300 ng/mL
Oxycodone (OXY)100 ng/mL
Phencyclidine (PCP)25 ng/mL
Propoxyphene (PPX)300 ng/mL
Nortriptyline (TCA)1000 ng/mL
Cannabinoids (THC)50 ng/mL
6-Monoacetylmorphine (6-MAM)10 ng/mL

Dochek® Multi-Drug Urine Test Dipcard Rx offers any combinations from 1 to 16 drugs but only one cutoff concentration under same drug condition will be included per device. It is intended for prescription use. For in vitro diagnostic use only.

{9}------------------------------------------------

The test provides only preliminary results. Clinical consideration and professional judgment should be applied to any drug of abuse test result, particularly in evaluating a preliminary positive result. To obtain a confirmed analytical result, a more specific alternate chemical method is needed. GC/MS or LC/MS is the recommended confirmatory method.

Dochek® Multi-Drug Urine Test Dipcard is an immunoassay for the qualitative determination of single or multiple drugs in human urine at the following cutoff concentrations:

Drug (Identifier)Cut-off level
Amphetamine (AMP)1000 or 500 ng/mL
Secobarbital (BAR)300 ng/mL
Buprenorphine (BUP)10 ng/mL
Oxazepam (BZO)300 ng/mL
Cocaine (COC)300 or 150 ng/mL
2-ethylidene-1,5-dimethyl-3,3-diphenylpyrrolidine (EDDP)300 ng/mL
Methylenedioxymethamphetamine (MDMA)500 ng/mL
Methamphetamine (MET)1000 or 500 ng/mL
Morphine (MOP/OPI)300 or 2000 ng/mL
Methadone (MTD)300 ng/mL
Oxycodone (OXY)100 ng/mL
Phencyclidine (PCP)25 ng/mL
Propoxyphene (PPX)300 ng/mL
Nortriptyline (TCA)1000 ng/mL
Cannabinoids (THC)50 ng/mL
6-Monoacetylmorphine (6-MAM)10 ng/mL

Dochek® Multi-Drug Urine Test Dipcard offers any combinations from 1 to 16 drugs but only one cutoff concentration under same drug condition will be included per device.

It is intended for over-the-counter (OTC) use. For in vitro diagnostic use only.

The test provides only preliminary results. Clinical consideration and professional judgment should be applied to any drug of abuse test result, particularly in evaluating a preliminary positive result. To obtain a confirmed analytical result, a more specific alternate chemical method is needed. GC/MS or LC/MS is the recommended confirmatory method.

Device Description 8.

Dochek® Multi-Drug Urine Test Dipcard Rx and Dochek® Multi-Drug Urine Test Dipcard are immunochromatographic assays that use a lateral flow system for the qualitative detection of single or multiple drugs in human urine at or above the cut-off levels as indicated. The products are singleuse in vitro diagnostic devices.

This device is a dipcard format in which the test strips are integrated into the plastic dipcard. After removing the cap of the dipcard, the absorbent end of the test strips is exposed and can be in direct

{10}------------------------------------------------

contact with the urine sample. The device is in a ready-to-use format and no longer requires assembly before use.

Similarities
ItemNew DevicePredicate Device(K232659)
Intended useQualitative detection of drugs of abuse in urine.For prescription use or over-the-counter useSame
MethodologyCompetitive binding, lateral flowimmunochromatographic assay based on antigen-antibody reactionSame
Type of TestQualitativeSame
Specimen TypeHuman urineSame
Target Drugand CutoffValueTarget DrugCutoff(ng/mL)
Amphetamine (AMP)1000 or 500
Secobarbital (BAR)300
Buprenorphine (BUP)10
Oxazepam (BZO)300
Cocaine (COC)300 or 150
2-ethylidene-1,5-dimethyl-3,3-diphenylpyrrolidine (EDDP)300
Methylenedioxymethamphetamine(MDMA)500
Methamphetamine (MET)1000 or 500
Morphine (MOP300/OPI2000)2000 or 300
Methadone (MTD)300
Oxycodone (OXY)100
Phencyclidine (PCP)25
Propoxyphene (PPX)300
Nortriptyline (TCA)1000
Cannabinoids (THC)50
6-Monoacetylmorphine (6-MAM)10
Differences
ConfigurationsTest dipcardTest cup

Substantial Equivalence Information 9.

10. Standard/Guidance Document Reference (if applicable)

None referenced.

11. Test Principle

Dochek® Multi-Drug Urine Test Dipcard Rx or Dochek® Multi-Drug Urine Test Dipcard is a

{11}------------------------------------------------

competitive immunoassay that is used to screen for the presence of various drugs and drug metabolites in urine. It is chromatographic absorbent device in which, drugs within a urine sample, competitively combined to a limited number of drug monoclonal antibody (mouse) conjugate binding sites.

When the test is activated, the urine is absorbed into each test strip by capillary action, mixes with the respective drug monoclonal antibody conjugate, and flows across a pre-coated membrane. When drug in the urine sample is below the detection level of the test, respective drug monoclonal antibody conjugate binds to the respective drug-protein conjugate immobilized in the Test Region (T) of the test strip. This produces a colored Test line in the Test Region (T) of the strip, which, regardless of its intensity, indicates a negative test result.

When drug level is at or above the detection level of the tree drug in the sample binds to the respective drug monoclonal antibody conjugate, preventing the respective drug monoclonal antibody conjugate from binding to the respective drug-protein conjugate immobilized in the Test Region (T) of the device. This prevents the development of a distinct colored band in the test region, indicating a preliminary positive result.

To serve as a procedure control, a colored line will appear at the Control Region (C) of each strip, if the test has been performed properly.

12. Performance Characteristics

A. Analytical Performance

a. Precision/Reproducibility:

Precision studies were carried out for samples with concentrations of +100% cutoff, +50% cutoff, +25% cutoff, cutoff, -25% cutoff, -50% cutoff, -75% cut off and -100% cutoff. Other samples were prepared by spiked target drug in drug-free urine samples. Each drug concentration was confirmed by LC-MS/MS. For each concentration, tests were performed two runs per day for 25 days using three lots of test dipcard. The results obtained are summarized in the following tables:

DrugLotNumber+100%cutoff+75%cutoff+50%cutoff+25%cutoffCutoff-25%cutoff-50%cutoff-75%cutoff-100%cut-off
AMP1000Lot I0-/50+0-/50+0-/50+0-/50+13-/37+50-/0+50-/0+50-/0+50-/0+
1000Lot II0-/50+0-/50+0-/50+0-/50+14-/36+50-/0+50-/0+50-/0+50-/0+
Lot III0-/50+0-/50+0-/50+0-/50+13-/37+50-/0+50-/0+50-/0+50-/0+
BAR300Lot I0-/50+0-/50+0-/50+0-/50+14-/36+50-/0+50-/0+50-/0+50-/0+
300Lot II0-/50+0-/50+0-/50+0-/50+12-/38+50-/0+50-/0+50-/0+50-/0+
Lot III0-/50+0-/50+0-/50+0-/50+14-/36+50-/0+50-/0+50-/0+50-/0+
BUP10Lot I0-/50+0-/50+0-/50+0-/50+16-/34+50-/0+50-/0+50-/0+50-/0+
10Lot II0-/50+0-/50+0-/50+0-/50+15-/35+50-/0+50-/0+50-/0+50-/0+
Lot III0-/50+0-/50+0-/50+0-/50+15-/35+50-/0+50-/0+50-/0+50-/0+
BZO300Lot I0-/50+0-/50+0-/50+0-/50+11-/39+50-/0+50-/0+50-/0+50-/0+
300Lot II0-/50+0-/50+0-/50+0-/50+13-/37+50-/0+50-/0+50-/0+50-/0+
Lot III0-/50+0-/50+0-/50+0-/50+12-/38+50-/0+50-/0+50-/0+50-/0+
COCLot II0-/50+0-/50+0-/50+0-/50+13-/37+50-/0+50-/0+50-/0+50-/0+
300Lot III0-/50+0-/50+0-/50+0-/50+12-/38+50-/0+50-/0+50-/0+50-/0+
EDDPLot I0-/50+0-/50+0-/50+0-/50+13-/37+50-/0+50-/0+50-/0+50-/0+
300Lot II0-/50+0-/50+0-/50+0-/50+12-/38+50-/0+50-/0+50-/0+50-/0+
MDM A 500Lot I0-/50+0-/50+0-/50+0-/50+10-/40+50-/0+50-/0+50-/0+50-/0+
Lot II0-/50+0-/50+0-/50+0-/50+10-/40+50-/0+50-/0+50-/0+50-/0+
Lot III0-/50+0-/50+0-/50+0-/50+11-/39+50-/0+50-/0+50-/0+50-/0+
MET 1000Lot I0-/50+0-/50+0-/50+0-/50+11-/39+50-/0+50-/0+50-/0+50-/0+
Lot II0-/50+0-/50+0-/50+0-/50+10-/40+50-/0+50-/0+50-/0+50-/0+
Lot III0-/50+0-/50+0-/50+0-/50+11-/39+50-/0+50-/0+50-/0+50-/0+
OPI 2000Lot I0-/50+0-/50+0-/50+0-/50+10-/40+50-/0+50-/0+50-/0+50-/0+
Lot II0-/50+0-/50+0-/50+0-/50+9-/41+50-/0+50-/0+50-/0+50-/0+
Lot III0-/50+0-/50+0-/50+0-/50+10-/40+50-/0+50-/0+50-/0+50-/0+
MTD 300Lot I0-/50+0-/50+0-/50+0-/50+14-/36+50-/0+50-/0+50-/0+50-/0+
Lot II0-/50+0-/50+0-/50+0-/50+15-/35+50-/0+50-/0+50-/0+50-/0+
Lot III0-/50+0-/50+0-/50+0-/50+14-/36+50-/0+50-/0+50-/0+50-/0+
OXY 100Lot I0-/50+0-/50+0-/50+0-/50+13-/37+50-/0+50-/0+50-/0+50-/0+
Lot II0-/50+0-/50+0-/50+0-/50+15-/35+50-/0+50-/0+50-/0+50-/0+
Lot III0-/50+0-/50+0-/50+0-/50+15-/35+50-/0+50-/0+50-/0+50-/0+
PCP 25Lot I0-/50+0-/50+0-/50+0-/50+15-/35+50-/0+50-/0+50-/0+50-/0+
Lot II0-/50+0-/50+0-/50+0-/50+16-/34+50-/0+50-/0+50-/0+50-/0+
Lot III0-/50+0-/50+0-/50+0-/50+14-/36+50-/0+50-/0+50-/0+50-/0+
PPX 300Lot I0-/50+0-/50+0-/50+0-/50+12-/38+50-/0+50-/0+50-/0+50-/0+
Lot II0-/50+0-/50+0-/50+0-/50+12-/38+50-/0+50-/0+50-/0+50-/0+
Lot III0-/50+0-/50+0-/50+0-/50+12-/38+50-/0+50-/0+50-/0+50-/0+
TCA 1000Lot I0-/50+0-/50+0-/50+0-/50+12-/38+50-/0+50-/0+50-/0+50-/0+
Lot II0-/50+0-/50+0-/50+0-/50+11-/39+50-/0+50-/0+50-/0+50-/0+
Lot III0-/50+0-/50+0-/50+0-/50+11-/39+50-/0+50-/0+50-/0+50-/0+
THC 50Lot I0-/50+0-/50+0-/50+0-/50+14-/36+50-/0+50-/0+50-/0+50-/0+
Lot II0-/50+0-/50+0-/50+0-/50+13-/37+50-/0+50-/0+50-/0+50-/0+
Lot III0-/50+0-/50+0-/50+0-/50+14-/36+50-/0+50-/0+50-/0+50-/0+
6-MAMLot I0-/50+0-/50+0-/50+0-/50+14-/36+50-/0+50-/0+50-/0+50-/0+
Lot II0-/50+0-/50+0-/50+0-/50+15-/35+50-/0+50-/0+50-/0+50-/0+
Lot III0-/50+0-/50+0-/50+0-/50+15-/35+50-/0+50-/0+50-/0+50-/0+
AMP 500Lot IV0-/50+0-/50+0-/50+0-/50+12-/38+50-/0+50-/0+50-/0+50-/0+
Lot V0-/50+0-/50+0-/50+0-/50+12-/38+50-/0+50-/0+50-/0+50-/0+
Lot VI0-/50+0-/50+0-/50+0-/50+12-/38+50-/0+50-/0+50-/0+50-/0+
COC 150Lot IV0-/50+0-/50+0-/50+0-/50+15-/35+50-/0+50-/0+50-/0+50-/0+
Lot V0-/50+0-/50+0-/50+0-/50+14-/36+50-/0+50-/0+50-/0+50-/0+
Lot VI0-/50+0-/50+0-/50+0-/50+14-/36+50-/0+50-/0+50-/0+50-/0+
METLot IV0-/50+0-/50+0-/50+0-/50+12-/38+50-/0+50-/0+50-/0+50-/0+
500Lot V0-/50+0-/50+0-/50+0-/50+13-/37+50-/0+50-/0+50-/0+50-/0+
Lot VI0-/50+0-/50+0-/50+0-/50+12-/38+50-/0+50-/0+50-/0+50-/0+
MOPLot IV0-/50+0-/50+0-/50+0-/50+14-/36+50-/0+50-/0+50-/0+50-/0+
300Lot V0-/50+0-/50+0-/50+0-/50+15-/35+50-/0+50-/0+50-/0+50-/0+
Lot VI0-/50+0-/50+0-/50+0-/50+14-/36+50-/0+50-/0+50-/0+50-/0+

{12}------------------------------------------------

{13}------------------------------------------------

b. Linearity/assay reportable range:

Not applicable. This device is intended for qualitative use only.

c. Stability:

The device can be stable at 2-30°C for 36 months based on real time stability study.

d. Detection limit:

Not applicable.

e. Analytical specificity/Interference:

Cross-Reactivity:

Analytical specificity for this device was determined through adding the potential interfering substances to drug-free urine samples. The relative cross-reactivity represents the minimum concentration necessary to yield a result similar to the cutoff level of the respective assay. Percent cross-reactivity, provided in the below table, was calculated as the concentration of analyte tested that yielded a positive result, divided by the cutoff concentration, multiplied by 100. If no cross reactivity was observed, the highest concentration tested is shown.

Drug/CutoffCompoundConcentration(ng/mL)% Cross-Reactivity
AMP 1000d-Amphetamine1,000100%
d/l-Amphetamine3,00033.3%
l-Amphetamine50,0002%
(+/-) 3,4-methylenedioxyamphetamine(MDA)5,00020%
Phentermine3,00033.3%
Hydroxyamphetamine10,00010%
d-Methamphetamine100,000(Negative)Not detected
l-Methamphetamine100,000(Negative)Not detected
(+/-)3,4-Methylenedioxyethylamphetamine(MDEA)100,000(Negative)Not detected
(+/-)3,4-Methylenedioxymethamphetamine(MDMA)100,000(Negative)Not detected
(1R,2S)-(-)-Ephedrine100,000(Negative)Not detected
β-Phenylethylamine100,0001%
Tyramine100,0001%
p-Hydroxynorephedrine100,0001%
Phenylpropanolamine100,000(Negative)Not detected
(±)Phenylpropanolamine100,000(Negative)Not detected
p-Hydroxyamphetamine100,0001%
d/l-Norephedrine100,0001%
Benzphetamine100,000(Negative)Not detected
1-Epinephrine100,000(Negative)Not detected
d/l-Epinephrine100,000(Negative)Not detected
Secobarbital300100%
Amobarbital1,00030%
Alphenal75400%
Aprobarbital250120%
BAR 300Butabarbital100300%
Butalbital5,0006%
Butethal50060%
Cyclopentobarbital50060%
Pentobarbital200150%
Phenobarbital300100%
Buprenorphine10100%
Norbuprenorphine5020%
Buprenorphine 3-D-glucuronide10100%
BUP 10Norbuprenorphine 3-D-glucuronide10100%
Morphine100000 (Negative)Not Detected
Oxymorphone100000 (Negative)Not Detected
Hydromorphone100000 (Negative)Not Detected
Oxazepam300100%
Alprazolam150200%
α-Hydroxyalprazolam1,50020%
Bromazepam100300%
Chlordiazepoxide50060%
Clobazam75040%
Clonazepam1,50020%
Clorazepate dipotassium100300%
Diazepam50060%
Estazolam50060%
Flunitrazepam2,50012%
BZO 300Midazolam2,00015%
Nitrazepam2,00015%
Nordiazepam50060%
Temazepam250120%
Triazolam1,00030%
Desalkylflurazepam50060%
Lorazepam5,0006%
Norchlordiazepoxide50060%
Nordazepam1,00030%
Delorazepam2,00015%
Demoxepam5,0006%
Flurazepam50060%
Benzoylecgonine300100%
Cocaine HCl75040%
Cocaethylene12,5002.4%
COC 300Ecgonine30,0001%
Ecgonine methyl ester100,000(negative)Not detected
Norcocaine100,000(negative)Not detected
2-ethylidene-1,5-dimethyl-3,3-300100%
diphenylpyrrolidine
EDDP 300Methadone100,000(negative)Not detected
EMDP100,000(negative)Not detected
Doxylamine100,000(negative)Not detected
Levacetylmethadol (LAAM)100,000(negative)Not detected
Disopyramide100,000(negative)Not detected
Alpha Methadol100,000(negative)Not detected
MDMA 500(+/-)3,4-MethylenedioxymethamphetamineHCI(MDMA)500100%
(+/-)3,4-Methylenedioxyamphetamine HCl(MDA)3,00017%
(+/-)3,4-Methylenedioxyethylamphetamine(MDEA)300167%
d-Methamphetamine100,000(Negative)Not detected
d-Amphetamine100,000(Negative)Not detected
1-Methamphetamine100,000(Negative)Not detected
1-Amphetamine100,000(Negative)Not detected
d-Methamphetamine1,000100%
d-Amphetamine50,0002%
Chloroquine50,0002%
(1R,2S)-(-)-Ephedrine50,0002%
(-)-Methamphetamine25,0004%
MET 1000(+/-)3,4-methylenedioxumethamphetamine(MDMA)4,00025%
β-Phenylethylamine50,0002%
Trimethobenzamide10,00010%
1-Amphetamine75,0001.3%
(+/-)3,4-Methylenedioxyethylamphetamine(MDEA)30,0003.3%
Mephentermine50,0002%
Methoxyphenamine50,0002%
Fenfluramine75,0001.3%
Procaine100,000(Negative)Not detected
d/l-Amphetamine100,000(Negative)Not detected
p-Hydroxymethamphetamine30,0003.3%
l-Phenylephrine100,000(Negative)Not detected
d/l-Methamphetamine1,000100%
(+/-) 3,4-Methylenedioxyamphetamine(MDA)100,000(Negative)Not detected
Morphine2,000100%
Codeine2,000100%
Hydrocodone12,50016%
Hydromorphone5,00040%
6-Monoacetylmorphine1,500133%
Morphine 3-β-D-glucuronide2,000100%
Ethylmorphine1,500133%
Diacetylmorphine (heroin)2,000100%
OPI 2000Levorphanol75,0002.7%
Norcodeine12,50016%
Oxycodone100,000(Negative)Not detected
Thebaine5,00040%
Normorphine50,0004%
Oxymorphone100,000(Negative)Not detected
Procaine100,000(Negative)Not detected
Codeine-6-β-D-glucuronide3,00067%
d-Norpropoxyphene hydrochloride5,00040%
Methadone300100%
MTD 300EDDP100,000(Negative)Not detected
Doxylamine50,0000.6%
Levacetylmethadol (LAAM)100,000(negative)Not detected
EMDP100,000(negative)Not detected
Alpha Methadol100,000(negative)Not detected
Oxycodone100100%
OXY 100Hydrocodone5,0002%
Hydromorphone50,0000.2%
Oxymorphone1,00010%
Codeine>100,000Not detected
Ethylmorphine>100,000Not detected
Dihydrocodeine20,0000.5%
Oxymorphone-3β-D- glucuronide5,0002%
Morphine100,000(negative)Not detected
6-Monoacetylmorphine100,000(negative)Not detected
Buprenorphine100,000(negative)Not detected
Thebaine100,000(negative)Not detected
Phencyclidine25100%
PCP 254-Hydroxy Phencyclidine1,5001.7%
d-Propoxyphene300100%
PPX 300d-Norpropoxyphene300100%
Nortriptyline1,000100%
Nordoxepin1,000100%
Trimipramine3,00033.3%
Promazine1,50066.7%
Desipramine200500%
Imipramine750133%
TCA 1000Clomipramine10,00010%
Doxepin1,25080%
Maprotiline2,00050%
Amitriptyline1,50066.7%
Promethazine25,0004%
Cyclobenzaprine1,000100%
Norclomipramine12,5008%
11-nor-Δ9-THC-9-COOH50100%
THC 5011-nor-Δ8-THC-9-COOH30167%
(±)-11-Hydroxy-Δ9-THC2,5002%
Δ8- Tetrahydrocannabinol2,0002.5%
Δ9- Tetrahydrocannabinol5,0001%
Cannabinol10,0000.5%
Cannabidiol(CBD)100,0000.05%
(±)-11-nor-9-carboxy-Δ9-THC10050%
11-nor-Δ9-THC-carboxyglucuronide
6-MAM6-Monoacetylmorphine10100%
Codeine100,000(Negative)Not detected
Ethylmorphine100,000(Negative)Not detected
Hydrocodone50,0000.02%
Hydromorphone10,0000.1%
Levorphanol100,000(Negative)Not detected
Morphine 3-β-D-glucuronide100,000(Negative)Not detected
Morphine100,0000.01%
Norcodeine100,000(Negative)Not detected
Normorphine100,000(Negative)Not detected
Oxycodone100,000(Negative)Not detected
Oxymorphone10,0000.1%
Procaine50,0000.02%
Thebaine10,0000.1%
Diacetylmorphine (heroin)2540%
10Acetylcodeine10,000(Negative)<0.1%
Buprenorphine10,000(Negative)<0.1%
Dihydrocodeine10,000(Negative)<0.1%
Nalorphine5,0000.2%
Dextromethorphan100,000(Negative)Not detected
Imipramine100,000(Negative)Not detected
Levacetylmethadol (LAAM)100,000(Negative)Not detected
Meperidine100,000(Negative)Not detected
Methadone100,000(Negative)Not detected
Mitragynine (kratom)20,000(Negative)<0.05%
Morphine 6-D-glucuronide100,000(Negative)Not detected
Naloxone100,000(Negative)Not detected
Naltrexone100,000(Negative)Not detected
Naproxen100,000(Negative)Not detected
Norbuprenorphine10,000(Negative)<0.1%
Norbuprenorphine glucuronide100,000(Negative)Not detected
Norhydrocodone100,000(Negative)Not detected
Noroxycodone100,000(Negative)Not detected
Noroxymorphone100,000(Negative)Not detected
Norpropoxyphene100,000(Negative)Not detected
Oxymorphone-3β-D- glucuronide100,000(Negative)Not detected
Tapentadol HCl100,000(Negative)Not detected
Tramadol100,000(Negative)Not detected
d-Amphetamine500100%
d/l-Amphetamine1,50033.3%
l-Amphetamine25,0002%
(+/-) 3,4-methylenedioxyamphetamine(MDA)2,50020%
Phentermine1,50033.3%
Hydroxyamphetamine5,00010%
d-Methamphetamine100,000(Negative)Not detected
l-Methamphetamine100,000(Negative)Not detected
AMP 500(+/-)3,4-Methylenedioxyethylamphetamine(MDEA)100,000(Negative)Not detected
(+/-)3,4-Methylenedioxymethamphetamine(MDMA)100,000(Negative)Not detected
(1R,2S)-(-)-Ephedrine100,000(Negative)Not detected
β-Phenylethylamine100,0000.5%
Tyramine100,0000.5%
p-Hydroxynorephedrine100,0000.5%
Phenylpropanolamine100,000(Negative)Not detected
(±)Phenylpropanolamine100,000(Negative)Not detected
p-Hydroxyamphetamine100,0000.5%
d/l-Norephedrine100,0000.5%
Benzphetamine100,000(Negative)Not detected
l-Epinephrine100,000(Negative)Not detected
d/l-Epinephrine100,000(Negative)Not detected
COC 150Benzoylecgonine150100%
Cocaine HCl50030%
Cocaethylene5,0003%
Ecgonine15,0001%
Ecgonine methyl ester100,000(negative)Not detected
Norcocaine100,000(negative)Not detected
d-Methamphetamine500100%
d-Amphetamine25,0002%
Chloroquine25,0002%
(1R,2S)-(-)-Ephedrine25,0002%
(-)-Methamphetamine12,5004%
(+/-)3,4-methylenedioxumethamphetamine(MDMA)2,00025%
β-Phenylethylamine25,0002%
Trimethobenzamide5,00010%
l-Amphetamine50,0001%
MET 500(+/-)3,4-Methylenedioxyethylamphetamine(MDEA)15,0003.3%
Mephentermine25,0002%
Methoxyphenamine25,0002%
Fenfluramine37,5001.3%
Procaine>100,000Not detected
d/l-Amphetamine75,0000.7%
p-Hydroxymethamphetamine15,0003.3%
l-Phenylephrine>100,000Not detected
d/l-Methamphetamine500100%
(+/-) 3,4-Methylenedioxyamphetamine(MDA)75,0000.7%
MOP 300Morphine300100%
Codeine300100%
Hydrocodone5,0006%
Hydromorphone1,00030%
6-Monoacetylmorphine150200%
Morphine 3-β-D-glucuronide1,00030%
Ethylmorphine100300%
Diacetylmorphine (heroin)300100%
Levorphanol10,0003%
Norcodeine5,0006%
Oxycodone75,0000.4%
Thebaine3,00010%
Normorphine3,00010%
Oxymorphone25,0001.2%
Procaine100,000(Negative)Not detected
Codeine-6-β-D-glucuronide50060%
d-Norpropoxyphene hydrochloride300100%

{14}------------------------------------------------

{15}------------------------------------------------

{16}------------------------------------------------

{17}------------------------------------------------

{18}------------------------------------------------

{19}------------------------------------------------

{20}------------------------------------------------

{21}------------------------------------------------

{22}------------------------------------------------

Interfering Substances:

To evaluate potential interference, non-structurally related compounds were added to drug-free urine and to urine samples containing the target drugs at 25% below and 25% above each corresponding cutoff. Compounds that show no interference at a concentration of 100μg/mL are summarized in the following table.

3-HydroxytyramineDiflunisalOxalic Acid
AcetaminophenDigoxinOxolinic Acid
Acetylsalicylic AcidDiphenhydramineOxymetazoline
AcyclovirDopamine HClPaliperidone
Albumin (100 mg/dL)D-PseudoephedrinePapaverine
Albuterol sulfate (Proair HFA)DuloxetinePenicillin-G
AminophyllineErythromycinPenicillinV Potassium
AminopyrineEsomeprazole MagnesiumPhenacetin (Acetophenetidin)
AmoxicillinEthanol (1%)Phenelzine
AmpicillinFenoprofenPrednisone
ApomorphineFluoxetine HydrochloridePregablin
AripiprazoleFurosemideQuinine
AspartameGabapentinRanitidine
AtomoxetineGentisic AcidRifampicin
Atorvastatin CalciumGlucoseRisperidone
AtropineHemoglobinSalicylic Acid
AzithromycinHydralazineSerotonin
Benzilic acidHydrochlorothiazideSertraline Hydrochloride
BenzocaineHydrocortisoneSildenafil Citrate
Benzoic acidIbuprofenSimvastatin
BilirubinIsoxsuprineSulfamethazine
BupropionKetamineSulindac
CaptoprilKetoprofenTetrahydrozoline
CarbamazepineLabetalolTheophylline
CefradineLevofloxacin HydrochlorideThiamine
CephalexinLevonorgestrelThioridazine
Chloral HydrateLevothyroxine SodiumTramadol Hydrochloride
ChloramphenicolLidocaine HydrochlorideTrazodone Hydrochloride
ChlorothiazideLisinoprilTriamterene
chlorpheniramineLoperamideTrifluoperazine
CholesterolLoratadineTrimethoprim
Ciprofloxacin HydrochlorideMagnesiumUric Acid
CitalopramMeperidineVenlafaxine HCl
ClarithromycinMeprobamateVerapamil
ClonidineMetoprolol TartrateVitamin B2
ClozapineMifepristoneVitamin C (Ascorbic acid)
Conjugated EstrogensN-AcetylprocainamideZomepirac
CortisoneNalidixic Acidẞ-Estradiol
CotinineNaproxenChlorpromazine
CreatinineNiacinamidePerphenazine
D,L- IsoproterenolNicotineTetrahydrocortisone 3-(ß-D-glucuronide)
D,L-OctopamineNifedipineTetrahydrocortisone 3-acetate
D,L-PropranololNitroglycerinEcgonine methyl ester
D,L-TryptophanNorethindroneMethoxyphenamine (exceptMET test)
D,L-TyrosineNoscapineNaloxone
DeoxycorticosteroneO-Hydroxyhippuric AcidNaltrexone
DextromethorphanOmeprazoleTyramine (except AMP test)
Diclofenac

{23}------------------------------------------------

{24}------------------------------------------------

Effect of Urinary pH and Specific Gravity:

Interference by urinary pH and specific gravity were also evaluated using pooled urine specimens with concentrations of 0 (drug-free), at 25% below and 25% above each corresponding cutoff. The results demonstrated that pH levels of 4 to 9 and specific gravity levels of 1.000 to 1.035 do not affect the results of the assays.

B. Method Comparison Study

The method comparison studies for the device were performed in-house with three operators. Operators ran 80 (40 negative and 40 positive) unaltered urine clinical samples for each drug. The samples were blind labeled and compared to LC-MS/MS results are presented in the table below:

DrugtestTest DipcardResultDrug-FreeLowNegative byLC-MS/MS(less than-50%)Near CutoffNegative byLC-MS/MS(Between-50% andthe Cutoff)Near CutoffPositive byLC-MS/MS(Betweenthe cutoffand +50%)HighPositive byLC-MS/MS(greaterthan +50%)
AMP(AMP1000)Viewer A+000731
-1581720
Viewer B+000631
-1581730
Viewer C+000631
-1581730
AMP(AMP500)Viewer A+000830
-15121320
Viewer B+000830
-15121320
Viewer C+000830
-15121320
BARViewer A+0001622
-1518720
Viewer B+0001622
-1518720
Viewer C+0001522
-1518730
BUPViewer A+0022910
-15131010
Viewer B+0022910
-131010
Viewer C+0012910
-131110
BZOViewer A+001831
-111310
Viewer B+002731
-111220
Viewer C+001731
-111320
COCViewer A+0021127
(COC-131020
300)Viewer B+0011127
-131120
Viewer C+0021127
-131020
COCViewer A+0011325
(COC-121220
150)Viewer B+0011325
-121220
Viewer C+0011325
-121220
EDDPViewer A+0011129
-61800
Viewer B+0001129
-61900
Viewer C+0011129
-61800
MDMAViewer A+001930
-141010
Viewer B+000930
-141110
Viewer C+001930
-141010
METViewer A+000830
(MET-81720
1000)Viewer B+000830
-81720
Viewer C+000830
-81720
METViewer A+0001030
(MET-141100
500)Viewer B+000930
-141110
Viewer C+000930
-15141110
OPI(MOP2000)Viewer A+002929
-1591420
Viewer B+002929
-1591420
Viewer C+002929
-1591420
MOP(MOP300)Viewer A+0021821
-15121110
Viewer B+0021921
-15121100
Viewer C+0021821
-15121110
MTDViewer A+002930
-15121110
Viewer B+002830
-15121120
Viewer C+002930
-15121110
OXYViewer A+0021029
-15121110
Viewer B+002929
-15121120
Viewer C+0021029
-15121110
PCPViewer A+0002712
-15131210
Viewer B+0002812
-15131200
Viewer C+0002712
-15131210
PPXViewer A+0021128
-15131010
Viewer B+0011228
-15131100
Viewer C+0011028
-15131120
TCAViewer A+000632
-15131220
Viewer B+000632
-15131220
Viewer C+000632
-15131220
THCViewer A+002930
-15131010
Viewer B+001930
-15131110
Viewer C+001930
-15131110
6-MAMViewer A+0032810
-1513920
Viewer B+0032810
-1513920
Viewer C+0032810
-1513920

{25}------------------------------------------------

{26}------------------------------------------------

{27}------------------------------------------------

Discordant Results are summarized below.

DrugOperatorSample NumberLC/MS/MSResult (ng/mL)Dochek Result
6-MAM 10Viewer A, B, C10919.006+
Viewer A, B, C10819.465+
Viewer A, B, C10539.967+
Viewer A, B, C104810.237-
Viewer A, B, C105210.257-
AMP 1000Viewer A, B, C09551014.625-
Viewer A, B, C09361056.71-
Viewer B, C09031141.794-
BAR 300Viewer A, B, C0221302.963-
Viewer C0238304.207-
Viewer A, B, C0173312.828-
BUP 10Viewer A, B, C01368.76+
Viewer A, B01378.785+
Viewer A, B, C014110.761-
BZO 300Viewer B0405269.424+
Viewer A, B, C0428290.449+
Viewer B, C0422301.512-
Viewer A, B, C0409301.586-
COC 300Viewer A, B, C0632284.614+
Viewer A, C0602295.071+
Viewer A, B, C0620307.773-
Viewer A, B, C0630308.012-
EDDP 300Viewer A, C1591296.944+
MDMA 500Viewer A, C0762458.975+
Viewer A, B, C0750565.340-
MET 1000Viewer A, B, C09721002.105-
Viewer A, B, C10251026.596-
OPI 2000Viewer A, B, C08171872.771+
Viewer A, B, C08551996.034+
Viewer A, B, C08792022.484-
Viewer A, B, C08802137.854-
MTD 300Viewer A, B, C1159271.186+
Viewer A, B, C1185273.035+
Viewer A, B, C1129308.556-
Viewer B1123324.733-
OXY 100Viewer A, B, C037985.212+
Viewer A, B, C037587.649+
Viewer B0397106.706-
Viewer A, B, C0340108.7-
PCP 25Viewer A, C005325.901-
PPX 300Viewer A, B, C0504285.846+
Viewer A0487295.988+
Viewer A, C0536347.008-
Viewer C0551355.750-
TCA 1000Viewer A, B, C06881,051.49-
Viewer A, B, C06531,135.56-
THC 50Viewer A, C030041.179+
Viewer A, B031749.391+
Viewer A, B, C031255.644-
AMP 500Viewer A, B, C1345512.815-
Viewer A, B, C1297545.751-
COC 150Viewer A, B, C1446138.861+
Viewer A, B, C1512156.470-
Viewer A, B, C1505162.099-
MET 500Viewer B, C1403507.553-
MOP 300Viewer A, B, C1279248.796+
Viewer A, B, C1259249.557+
Viewer A, C1264308.896-

{28}------------------------------------------------

C. Lay Person Study

79 male and 61 female tested Dochek® Multi-Drug Urine Test Dipcard Configuration 1; 73 male and 67 female tested Dochek® Multi-Drug Urine Test Dipcard Configuration 2. They had diverse educational and occupational backgrounds and their age range from 21 to > 50. Urine samples were prepared at the following concentrations; -100%, +/-75%, +/-25% of the cutoff by spiking drug(s) into drug free-pooled urine specimens. The concentrations of the samples were confirmed by LC-MS/MS. Each sample was aliquoted into individual containers and blind-labeled. Each

{29}------------------------------------------------

participant was provided with the package insert, 1 blind labeled sample and a device. The results are summarized below.

DrugCutoff(ng/mL)ResultsDrug Concentration
-100%cutoff-75%cutoff-50%cutoff-25%cutoff+25%cutoff+50%cutoff+75%cutoff
AMP1000Negative20202020000
Positive0000202020
Total20202020202020
Percentage of correctresults (%)100%100%100%100%100%100%100%
BAR300Negative20202019000
Positive0001202020
Total20202020202020
Percentage of correctresults (%)100%100%100%95%100%100%100%
BUP10Negative20202018100
Positive0002192020
Total20202020202020
Percentage of correctresults (%)100%100%100%90%95%100%100%
BZO300Negative20202019000
Positive0001202020
Total20202020202020
Percentage of correctresults (%)100%100%100%95%100%100%100%
COC300Negative20202019000
Positive0001202020
Total20202020202020
Percentage of correctresults (%)100%100%100%95%100%100%100%
EDDP300Negative20202019000
Positive0001202020
Total20202020202020
Percentage of correctresults (%)100%100%100%95%100%100%100%
Negative20202020000
Positive0000202020
MDMA500Total20202020202020
Percentage of correctresults (%)100%100%100%100%100%100%100%
Negative20202019000
Positive0001202020
MET1000Total20202020202020
Percentage of correctresults (%)100%100%100%95%100%100%100%
Negative20202020000
OPI2000Positive0000202020
Total20202020202020
Percentage of correctresults (%)100%100%100%100%100%100%100%
300Negative20202019100
Positive0001192020
MTDTotal20202020202020
Percentage of correctresults (%)100%100%100%95%95%100%100%
Negative20202019000
Positive0001202020
OXY100Total20202020202020
Percentage of correctresults (%)100%100%100%95%100%100%100%
Negative20202018000
PCPPositive0002202020
25Total20202020202020
Percentage of correctresults (%)100%100%100%90%100%100%100%
PPX300Negative20202020000
Positive0000202020
Total20202020202020
Percentage of correctresults (%)100%100%100%100%100%100%100%
Negative20202020000
TCA1000Positive0000202020
Total20202020202020
Percentage of correctresults (%)100%100%100%100%100%100%100%
THC50Negative20202019100
Positive0001192020
Total20202020202020
Percentage of correctresults (%)100%100%100%95%95%100%100%
6-MAM10Negative20202020000
Positive0000202020
Total20202020202020
Percentage of correctresults (%)100%100%100%100%100%100%100%

Result of Dochek® Multi-Drug Urine Test Dipcard Configuration 1:

{30}------------------------------------------------

{31}------------------------------------------------

Result of Dochek® Multi-Drug Urine Test Dipcard Configuration 2:

DrugCutoff(ng/mL)ResultsDrug Concentration
-100%cutoff-75%cutoff-50%cutoff-25%cutoff+25%cutoff+50%cutoff+75%cutoff
AMP500Negative20202019000
Positive0001202020
Total20202020202020
Percentage ofcorrect results (%)100%100%100%95%100%100%100%
BAR300Negative20202019000
Positive0001202020
Total20202020202020
Percentage ofcorrect results (%)100%100%100%95%100%100%100%
BUP10Negative20202020000
Positive0000202020
Total20202020202020
Percentage ofcorrect results (%)100%100%100%100%100%100%100%
BZO300Negative20202019100
Positive0001192020
Total20202020202020
Percentage ofcorrect results (%)100%100%100%95%95%100%100%
COC150Negative20202018000
Positive0002202020
Total20202020202020
Percentage ofcorrect results (%)100%100%100%90%100%100%100%
EDDP300Negative20202019000
Positive0001202020
Total20202020202020
Percentage ofcorrect results (%)100%100%100%95%100%100%100%
MDMA500Negative20202019000
Positive0001202020
Total20202020202020
Percentage ofcorrect results (%)100%100%100%95%100%100%100%
MET500Negative20202020100
Positive0000192020
Total20202020202020
Percentage ofcorrect results (%)100%100%100%100%95%100%100%
MOP300Negative20202019000
Positive0001202020
Total20202020202020
Percentage ofcorrect results (%)100%100%100%95%100%100%100%
MTD300Negative20202020100
Positive0000192020
Total20202020202020
Percentage ofcorrect results (%)100%100%100%100%95%100%100%
Negative20202018000
OXY100Positive0002202020
Total20202020202020
Percentage ofcorrect results (%)100%100%100%90%100%100%100%
Negative20202019000
PCP25Positive0001202020
Total20202020202020
Percentage ofcorrect results (%)100%100%100%95%100%100%100%
PPX300Negative20202018100
Positive0002192020
Total20202020202020
Percentage ofcorrect results (%)100%100%100%90%95%100%100%
TCA1000Negative20202019000
Positive0001202020
Total20202020202020
Percentage ofcorrect results (%)100%100%100%95%100%100%100%
THC50Negative20202018000
Positive0002202020
Total20202020202020
Percentage ofcorrect results (%)100%100%100%90%100%100%100%
6-MAM10Negative20202019000
Positive0001202020
Total20202020202020
Percentage of correctresults (%)100%100%100%95%100%100%100%

{32}------------------------------------------------

{33}------------------------------------------------

Participants were given surveys on the ease of understanding the instruction for use. All participants indicated that the device instruction is easy to understand and follow. A Flesch-Kincaid reading analysis was performed on each package insert and the scores revealed a reading Grade Level of 7.

{34}------------------------------------------------

D. Clinical Studies

Not applicable.

13. Conclusion

Based on the test principle and performance characteristics of the device including precision, cut-off, interference, specificity, method comparison and lay-user studies of the devices, it's concluded that Dochek® Multi-Drug Urine Test Dipcard Rx and Dochek® Multi-Drug Urine Test Dipcard are substantially equivalent to the predicate devices.

§ 862.3650 Opiate test system.

(a)
Identification. An opiate test system is a device intended to measure any of the addictive narcotic pain-relieving opiate drugs in blood, serum, urine, gastric contents, and saliva. An opiate is any natural or synthetic drug that has morphine-like pharmocological actions. The opiates include drugs such as morphine, morphine glucoronide, heroin, codeine, nalorphine, and meperedine. Measurements obtained by this device are used in the diagnosis and treatment of opiate use or overdose and in monitoring the levels of opiate administration to ensure appropriate therapy.(b)
Classification. Class II (special controls). An opiate test system is not exempt if it is intended for any use other than employment or insurance testing or is intended for Federal drug testing programs. The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 862.9, provided the test system is intended for employment and insurance testing and includes a statement in the labeling that the device is intended solely for use in employment and insurance testing, and does not include devices intended for Federal drug testing programs (e.g., programs run by the Substance Abuse and Mental Health Services Administration (SAMHSA), the Department of Transportation (DOT), and the U.S. military).