(115 days)
The LZI Buprenorphine II Enzyme Immunoassay is intended for the qualitative and semi-quantitative determination of norbuprenorphine in human urine at the cutoff value of 5 ng/mL when calibrated against norburprenorphine. The assay is designed for prescription use with a number of automated clinical chemistry analyzers.
The semi-quantitative mode is for purposes of (1) enabling laboratories to determine an appropriate dilution of the specimen for confirmation by a confirmatory method such as gas or liquid chromatography/mass spectrometry (GC/MS or LC/MS) or (2) permitting laboratories to establish quality control procedures
The assay provides only a preliminary analytical result. A more specific alternative chemical method (e.g., gas or liquid chromatography and mass spectrometry) must be used in order to obtain a confirmed analytical result. Clinical consideration and professional judgment should be exercised with any drug of abuse test result, particularly when the preliminary test result is positive.
The LZI Buprenorphine II Enzyme Immunoassay is a homogeneous enzyme immunoassay ready-to-use liquid reagent. The assay is based on competition between drug in the sample and drug labeled with the enzyme glucose-6-phosphate dehydrogenase (G6PDH) for a fixed amount of antibody in the reagent. Enzyme activity decreases upon binding to the antibody, and the drug concentration in the sample is measured in terms of enzyme activity. In the absence of drug in the sample, norbuprenorphine-labeled G6PDH conjugate is bound to antibody, and the enzyme activity is inhibited. On the other hand, when drug is present in the sample, antibody would bind to free drug; the unbound norbuprenorphine-labeled G6PDH then exhibits its maximal enzyme activity. Active enzyme converts nicotinamide adenine dinucleotide (NAD) to NADH, resulting in an absorbance change that can be measured spectrophotometrically at 340 nm.
The LZI Buprenorphine II Enzyme Immunoassay is a kit comprised of two reagents, R₁ and R₂, which are bottled separately but sold together within the kit.
The R₁ solution contains mouse monoclonal anti-norbuprenorphine antibody, glucose-6-phosphate (G6P), nicotinamide adenine dinucleotide (NAD), stabilizers, and sodium azide (0.09%) as a preservative. The R₂ solution contains glucose-6-phosphate dehydrogenase (G6PDH) labeled with norbuprenorphine in buffer with sodium azide (0.09%) as a preservative.
N/A
FDA 510(k) Clearance Letter - LZI Buprenorphine II Enzyme Immunoassay
Page 1
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
Doc ID # 04017.08.02
Jan 16, 2026
Lin-Zhi International, Inc.
Thet Wai
Manager, TDM Immunoassay Development
2945 Oakmead Village Ct.
Santa Clara, California 95051
Re: K253082
Trade/Device Name: LZI Buprenorphine II Enzyme Immunoassay
Regulation Number: 21 CFR 862.3650
Regulation Name: Opiate test system
Regulatory Class: Class II
Product Code: DJG
Dated: December 5, 2025
Received: December 8, 2025
Dear Thet Wai:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Page 2
K253082 - Thet Wai Page 2
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801 and Part 809); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-
Page 3
K253082 - Thet Wai Page 3
assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
JOSEPH A. KOTAREK -S
Digitally signed by JOSEPH A. KOTAREK -S
Date: 2026.01.16 15:05:50 -05'00'
Joseph Kotarek
Branch Chief
Division of Chemistry and Toxicology Devices
OHT7: Office of In Vitro Diagnostics
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
Page 4
FORM FDA 3881 (8/23) Page 1 of 1
DEPARTMENT OF HEALTH AND HUMAN SERVICES
Food and Drug Administration
Indications for Use
Form Approved: OMB No. 0910-0120
Expiration Date: 07/31/2026
See PRA Statement below.
510(k) Number (if known): K253082
Device Name: LZI Buprenorphine II Enzyme Immunoassay
Indications for Use (Describe)
The LZI Buprenorphine II Enzyme Immunoassay is intended for the qualitative and semi-quantitative determination of norbuprenorphine in human urine at the cutoff value of 5 ng/mL when calibrated against norburprenorphine. The assay is designed for prescription use with a number of automated clinical chemistry analyzers.
The semi-quantitative mode is for purposes of (1) enabling laboratories to determine an appropriate dilution of the specimen for confirmation by a confirmatory method such as gas or liquid chromatography/mass spectrometry (GC/MS or LC/MS) or (2) permitting laboratories to establish quality control procedures
The assay provides only a preliminary analytical result. A more specific alternative chemical method (e.g., gas or liquid chromatography and mass spectrometry) must be used in order to obtain a confirmed analytical result. Clinical consideration and professional judgment should be exercised with any drug of abuse test result, particularly when the preliminary test result is positive.
Type of Use (Select one or both, as applicable)
☒ Prescription Use (Part 21 CFR 801 Subpart D) ☐ Over-The-Counter Use (21 CFR 801 Subpart C)
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services
Food and Drug Administration
Office of Chief Information Officer
Paperwork Reduction Act (PRA) Staff
PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
Page 5
Lin-Zhi International, Inc.
510(k) Summary
510(k) Number
Prepared On
January 14, 2026
Submitter Name and Contact Person:
Thet Wai
Manager, TDM Assay Development
Phone: (408) 970-8811
Fax: (408) 970-9030
E-mail: submissions@lin-zhi.com
Company Address:
Lin-Zhi International, Inc.
2945 Oakmead Village Court
Santa Clara, CA 95051
Page 6
Lin-Zhi International, Inc.
Introduction
This submission is provided in accordance with 21 CFR 807.92 as a 510(k) for the LZI Buprenorphine II Enzyme Immunoassay. The LZI Buprenorphine II Enzyme Immunoassay is similar to its legally marketed predicate device, LZI Buprenorphine Enzyme Immunoassay (K090844), in terms of intended use, method principle, device components, and clinical performance.
Verification and validation activities were conducted using well-established methods consistent with those performed for the predicate device. These included method comparison with LC/MS, precision studies, cross-reactivity evaluation, and interference testing. Collectively, the data confirm that the modified device supports its intended use, safety, and effectiveness, and performs substantially equivalent to the predicate.
This submission includes a summary of design control activities and performance characteristics, which together provide a complete basis for a determination of substantial equivalence.
Device Name and Classification
Classification Name: Enzyme Immunoassay, Opiates
Regulation Number: 21 CFR 862.3650
Product Code: Class II, DJG
Common Name: Homogeneous Buprenorphine Enzyme Immunoassay
Proprietary Name: LZI Buprenorphine II Enzyme Immunoassay
Submission Type: 510(k)
510(k) Number: K253082
Legally Marketed Predicate Device
The subject device is compared to the predicate:
- Predicate Device: LZI Buprenorphine Enzyme Immunoassay
- 510(k) Number: K090844
The LZI Buprenorphine II Enzyme Immunoassay is substantially equivalent to the LZI Buprenorphine Enzyme Immunoassay (K090844) manufactured by LZI in terms of intended use, method principle, device components, and clinical performance.
Page 7
Lin-Zhi International, Inc.
Device Description
The LZI Buprenorphine II Enzyme Immunoassay is a homogeneous enzyme immunoassay ready-to-use liquid reagent. The assay is based on competition between drug in the sample and drug labeled with the enzyme glucose-6-phosphate dehydrogenase (G6PDH) for a fixed amount of antibody in the reagent. Enzyme activity decreases upon binding to the antibody, and the drug concentration in the sample is measured in terms of enzyme activity. In the absence of drug in the sample, norbuprenorphine-labeled G6PDH conjugate is bound to antibody, and the enzyme activity is inhibited. On the other hand, when drug is present in the sample, antibody would bind to free drug; the unbound norbuprenorphine-labeled G6PDH then exhibits its maximal enzyme activity. Active enzyme converts nicotinamide adenine dinucleotide (NAD) to NADH, resulting in an absorbance change that can be measured spectrophotometrically at 340 nm.
The LZI Buprenorphine II Enzyme Immunoassay is a kit comprised of two reagents, R₁ and R₂, which are bottled separately but sold together within the kit.
The R₁ solution contains mouse monoclonal anti-norbuprenorphine antibody, glucose-6-phosphate (G6P), nicotinamide adenine dinucleotide (NAD), stabilizers, and sodium azide (0.09%) as a preservative. The R₂ solution contains glucose-6-phosphate dehydrogenase (G6PDH) labeled with norbuprenorphine in buffer with sodium azide (0.09%) as a preservative.
Intended Use
The LZI Buprenorphine II Enzyme Immunoassay is intended for the qualitative and semi-quantitative determination of norbuprenorphine in human urine at the cutoff value of 5 ng/mL when calibrated against norbuprenorphine. The assay is designed for prescription use with a number of automated clinical chemistry analyzers.
The semi-quantitative mode is for purposes of (1) enabling laboratories to determine an appropriate dilution of the specimen for confirmation by a confirmatory method such as gas or liquid chromatography/mass spectrometry (GC/MS or LC/MS) or (2) permitting laboratories to establish quality control procedures
The assay provides only a preliminary analytical result. A more specific alternative chemical method (e.g., gas or liquid chromatography and mass spectrometry) must be used in order to obtain a confirmed analytical result. Clinical consideration and professional judgment should be exercised with any drug of abuse test result, particularly when the preliminary test result is positive.
Substantial Equivalence Comparison to Predicate Device
The LZI Buprenorphine II Enzyme Immunoassay is substantially equivalent to the LZI Buprenorphine Enzyme Immunoassay which was cleared by the FDA under the premarket notification K090844 for its stated intended use.
Page 8
Lin-Zhi International, Inc.
The following table compares the LZI Buprenorphine II Enzyme Immunoassay with the predicate device.
| Device Characteristics | Subject Device LZI Buprenorphine II Enzyme Immunoassay | Predicate Device (K090844) LZI Buprenorphine Enzyme Immunoassay |
|---|---|---|
| Intended Use | The LZI Buprenorphine II Enzyme Immunoassay is intended for the qualitative and semi-quantitative determination of norbuprenorphine in human urine at the cutoff value of 5 ng/mL when calibrated against norbuprenorphine. The assay is designed for prescription use with a number of automated clinical chemistry analyzers. | The Lin-Zhi International (LZI) Buprenorphine Enzyme Immunoassay is intended for the qualitative and semiquantitative determination of norbuprenorphine (buprenorphine metabolite) in human urine, at cutoff values of 5 ng/mL and 10 ng/mL. The assay is designed for prescription use with a number of automated clinical chemistry analyzers. |
| The semi-quantitative mode is for purposes of (1) enabling laboratories to determine an appropriate dilution of the specimen for confirmation by a confirmatory method such as gas or liquid chromatography/mass spectrometry (GC/MS or LC/MS) or (2) permitting laboratories to establish quality control procedures | ||
| The assay provides only a preliminary analytical result. A more specific alternative chemical method (e.g., gas or liquid chromatography and mass spectrometry) must be used in order to obtain a confirmed analytical result. Clinical consideration and professional judgment should be exercised with any drug of abuse test result, particularly when the preliminary test result is positive. | ||
| Analyte | norbuprenorphine | norbuprenorphine |
| Cutoff | 5 ng/mL | 5 ng/mL and 10 ng/mL |
| Matrix | Urine | Urine |
| Calibrator Level | 0, 2.5, 5, 10, and 20 ng/mL | 0, 5, 10, 20, 40, and 75 ng/mL |
| Controls Level | 5 ng/mL Cutoff: 2 Levels 3.75 and 6.25 ng/mL | 5 ng/mL Cutoff: 2 Levels 3 and 7 ng/mL |
| 10 ng/mL Cutoff: 2 Levels 7 and 13 ng/mL | ||
| Storage | 2-8 °C until expiration date | 2-8 °C until expiration date |
| Detection | Absorbance change measured spectrophotometrically at 340 nm. | same |
| User Environment | Clinical laboratories; Prescription use only | same |
| Mass Spectrometry Confirmation | Required to confirm preliminary positive analytical results | same |
| Platform Required | Automated clinical chemistry analyzer | same |
Page 9
Lin-Zhi International, Inc.
| Device Characteristics | Subject Device LZI Buprenorphine II Enzyme Immunoassay | Predicate Device (K090844) LZI Buprenorphine Enzyme Immunoassay |
|---|---|---|
| Reagents Form | Liquid – ready-to-use | same |
| Reagent Materials | Two (2) reagent system: Antibody/substrate reagent (R₁) and enzyme labeled conjugates (R₂) with sodium azide preservative | same |
Performance Characteristics Summary:
All 510(k) studies below were conducted on the Beckman Coulter AU480 Analyzer
Precision: 5 ng/mL Cutoff
The assay was tested in qualitative (AOD, mAU) mode using a modified NCCLS-EP5 protocol. Norbuprenorphine sample concentrations were prepared by spiking a norbuprenorphine standard into a pool of negative human urine at the cutoff concentration and ±25%, ±50%, ±75%, and ±100% of the cutoff concentration.
Results shown below were obtained by testing all norbuprenorphine samples in replicate of two, two runs a day (one in the morning and one in the afternoon) for 22 days (within-run precision) on one Beckman Coulter AU480 automated clinical analyzer for a total of 88 runs (total precision). One single lot of reagents and calibrators and controls were used and stored at 2-8°C when not in use.
Precision: 5 ng/mL Cutoff
Semi-Quantitative Positive/Negative Results:
| 5 ng/mL Cutoff Result: | Within Run (N=22) | Total Precision (N=88) |
|---|---|---|
| Norbuprenorphine Concentration | % of Cutoff | Number of Determination |
| 0 ng/mL | 0% | 22 |
| 1.25 ng/mL | 25% | 22 |
| 2.5 ng/mL | 50% | 22 |
| 3.75 ng/mL | 75% | 22 |
| 5 ng/mL | 100% | 22 |
| 6.25 ng/mL | 125% | 22 |
| 7.5 ng/mL | 150% | 22 |
| 8.75 ng/mL | 175% | 22 |
| 10 ng/mL | 200% | 22 |
Page 10
Lin-Zhi International, Inc.
Performance Characteristics Summary, continued:
Beckman Coulter® AU480 Analyzer
Semi-Quantitative Precision Analysis Summary (ng/mL):
| NBUP Concentration | Within Run (N=22) | Total Precision (N=88) |
|---|---|---|
| SD | % CV | |
| 0 ng/mL | 0.15 | N/A |
| 1.25 ng/mL | 0.17 | 12.0% |
| 2.5 ng/mL | 0.17 | 6.4% |
| 3.75 ng/mL | 0.19 | 4.7% |
| 5 ng/mL | 0.19 | 3.5% |
| 6.25 ng/mL | 0.19 | 2.8% |
| 7.5 ng/mL | 0.20 | 2.5% |
| 8.75 ng/mL | 0.21 | 2.2% |
| 10 ng/mL | 0.30 | 2.8% |
Qualitative Positive/Negative Results:
| 5 ng/mL Cutoff Result: | 5 ng/mL Cutoff Result: | Total Precision (N=88) |
|---|---|---|
| Norbuprenorphine Concentration | % of Cutoff | Number of Determination |
| 0 ng/mL | 0% | 22 |
| 1.25 ng/mL | 25% | 22 |
| 2.5 ng/mL | 50% | 22 |
| 3.75 ng/mL | 75% | 22 |
| 5 ng/mL | 100% | 22 |
| 6.25 ng/mL | 125% | 22 |
| 7.5 ng/mL | 150% | 22 |
| 8.75 ng/mL | 175% | 22 |
| 10 ng/mL | 200% | 22 |
Limit of Detection: 5 ng/mL Cutoff
The lowest concentration that can be differentiated from the negative urine with 95% confidence is determined as 1.0 ng/mL.
Page 11
Lin-Zhi International, Inc.
Performance Characteristics Summary, continued:
Beckman Coulter® AU480 Analyzer
Linearity: 5 ng/mL Cutoff
To demonstrate recovery of the entire assay range, a drug-free urine pool spiked with norbuprenorphine at 20 ng/mL was serially diluted. Each sample was run in 10 replicates and the average was used to determine percent recovery compared to the expected target value.
Determined concentration averages were obtained and all averages were ±15 % of the target concentrations. The determined average percent recovery (Determined Concentration Average divided by the Target Concentration) was considered acceptable between 85 – 115 %.
The recovery of norbuprenorphine spiked to various concentrations was evaluated and the average recovery for the linear range of the assay was found to range between 99.6% - 106.3%.
| Target Concentration (ng/mL) | Determined Concentration Range (ng/mL) | Determined Concentration Average (ng/mL) | Average % Recovery |
|---|---|---|---|
| 20 | 18.73 - 20.41 | 19.91 | 99.6% |
| 18 | 18.06 – 18.96 | 18.54 | 103.0% |
| 16 | 16.37 – 16.95 | 16.66 | 104.1% |
| 14 | 13.51 - 15.14 | 14.18 | 101.3% |
| 12 | 12.17 – 13.23 | 12.75 | 106.3% |
| 10 | 9.74 - 10.68 | 10.24 | 102.4% |
| 8 | 7.92 – 8.56 | 8.21 | 102.6% |
| 6 | 6.08 – 6.59 | 6.29 | 104.8% |
| 4 | 3.70 – 4.29 | 4.07 | 101.9% |
| 2 | 1.70 – 2.09 | 1.90 | 94.8% |
| 1 | 0.60 – 1.36 | 0.94 | 94.2% |
| 0 | -0.22 – 0.29 | 0.05 | N/A |
Page 12
Lin-Zhi International, Inc.
Performance Characteristics Summary, continued:
Beckman Coulter® AU480 Analyzer
Method Comparison - Clinical Samples: 5 cutoff
A total of ninety-four (94) unaltered clinical samples were tested with the LZI Buprenorphine II Enzyme Immunoassay on the Beckman Coulter AU480 automated clinical analyzer. All samples were tested in singlet. All samples were confirmed with LC/MS for both buprenorphine and norbuprenorphine concentrations. Samples were collected by:
- Lin-Zhi International, Inc. (LZI) at Santa Clara, CA, USA
- University of California, San Francisco (UCSF) at San Francisco, CA, USA
- Millenium Health (Millenium) at San Diego, CA, USA
- Boca Biolistics at Pompano Beach, FL, USA
- University of Leeds at Leeds, UK
- BirdRock Laboratories (BirdRock) at San Diego, CA, USA
- APC Health at Pearland, TX),
- Altius Diagnostics (Altius) at Bellevue, WA, USA
- Tricore Research Institute (Tricore) at Albuquerque, NM, USA
- Cozart at Oxfordshire, UK
Semi-Quantitative Accuracy Study:
| BUP II Results 5 ng/mL Cutoff | Negative by LC/MS analysis | < 50% of the cutoff concentration by LC/MS analysis | Near Cutoff Negative between 50% below the cutoff and the cutoff concentration by LC/MS analysis | Near Cutoff Positive between the cutoff and 50% above the cutoff concentration by LCMS analysis | High Positive greater than 50% above the cutoff concentration by LC/MS analysis |
|---|---|---|---|---|---|
| Positive at or above the cutoff by EIA analysis | 0 | 7* | 12** | 14 | 33 |
| Negative below the cutoff by EIA analysis | 20 | 6 | 2 | 0 | 0 |
*Discrepant <50% of the cutoff concentration (0.1 – 2.49 ng/mL)
**Discrepant between 50% of the cutoff to the cutoff concentration (2.5 – 4.99 ng/mL)
Page 13
Lin-Zhi International, Inc.
Performance Characteristics Summary, continued:
Beckman Coulter® AU480 Analyzer
Method Comparison 5 cutoff, continued:
Discrepant samples determined when comparing total buprenorphine and norbuprenorphine LC/MS results with EIA results on the Beckman Coulter AU480 automated clinical analyzer are shown in the table below.
| Sample # | BUP LC/MS (ng/mL) | NBUP LC/MS (ng/mL) | Total LC/MS (ng/mL) | Pos/Neg Result | AU480 EIA Semi-Quantitative Result (ng/mL) | Pos/Neg Result |
|---|---|---|---|---|---|---|
| 21* | 0.75 | 0.27 | 1.02 | - | 21.59 | + |
| 22* | 0.00 | 1.09 | 1.09 | - | 5.20 | + |
| 24* | 1.04 | 0.07 | 1.11 | - | 21.96 | + |
| 26* | 0.57 | 0.68 | 1.25 | - | 9.79 | + |
| 28* | 1.23 | 0.06 | 1.29 | - | 16.74 | + |
| 31* | 1.24 | 0.16 | 1.40 | - | 20.56 | + |
| 32* | 0.62 | 0.83 | 1.45 | - | 8.98 | + |
| 35** | 0.24 | 2.38 | 2.62 | - | 9.59 | + |
| 36** | 0.20 | 2.48 | 2.68 | - | 12.35 | + |
| 37** | 2.05 | 0.67 | 2.72 | - | 24.57 | + |
| 38** | 0.11 | 2.64 | 2.75 | - | 7.87 | + |
| 39** | 0.94 | 1.83 | 2.77 | - | 5.89 | + |
| 40** | 0.23 | 2.55 | 2.78 | - | 15.47 | + |
| 41** | 1.69 | 1.29 | 2.98 | - | 23.60 | + |
| 42** | 1.04 | 1.97 | 3.01 | - | 5.62 | + |
| 43** | 2.54 | 0.52 | 3.06 | - | 22.99 | + |
| 44** | 0.19 | 2.99 | 3.18 | - | 8.02 | + |
| 45** | 0.00 | 3.19 | 3.19 | - | 5.78 | + |
| 47** | 1.61 | 1.66 | 3.27 | - | 24.45 | + |
*Discrepant <50% of the cutoff concentration (0.1 – 2.49 ng/mL)
**Discrepant between 50% of the cutoff to the cutoff concentration (2.5 – 4.99 ng/mL)
*These discrepancies are due to known causes. LZI Buprenorphine II Enzyme Immunoassay cross-reacts with both buprenorphine glucuronide and norbuprenorphine glucuronide, which explains the elevated EIA values and false positives observed in discrepant samples #21, #22, #24, #26, #28, #31 and #32.
Page 14
Lin-Zhi International, Inc.
Performance Characteristics Summary, continued:
Beckman Coulter® AU480 Analyzer
Qualitative Accuracy Study:
| BUP II Results 5 ng/mL Cutoff | Negative by LC/MS analysis | < 50% of the cutoff concentration by LC/MS analysis | Near Cutoff Negative between 50% below the cutoff and the cutoff concentration by LC/MS analysis | Near Cutoff Positive between the cutoff and 50% above the cutoff concentration by LCMS analysis | High Positive greater than 50% above the cutoff concentration by LC/MS analysis |
|---|---|---|---|---|---|
| Positive at or above the cutoff by EIA analysis | 0 | 7* | 12** | 14 | 33 |
| Negative below the cutoff by EIA analysis | 20 | 6 | 2 | 0 | 0 |
*Discrepant <50% of the cutoff concentration (0.1 – 2.49 ng/mL)
**Discrepant between 50% of the cutoff to the cutoff concentration (2.5 – 4.99 ng/mL)
Page 15
Lin-Zhi International, Inc.
Performance Characteristics Summary, continued:
Beckman Coulter® AU480 Analyzer
Discrepant samples determined when comparing total buprenorphine and norbuprenorphine LC/MS results with EIA results on the Beckman Coulter AU480 automated clinical analyzer are shown in the table below.
| Sample # | BUP LC/MS (ng/mL) | NBUP LC/MS (ng/mL) | Total LC/MS (ng/mL) | Pos/Neg Result | AU480 EIA Qualitative Result (mAU) | Qualitative Cutoff Rate (mAU) | Pos/Neg Result |
|---|---|---|---|---|---|---|---|
| 21* | 0.75 | 0.27 | 1.02 | - | 359.3 | 88.6 | + |
| 22* | 0.00 | 1.09 | 1.09 | - | 99.8 | 87.8 | + |
| 24* | 1.04 | 0.07 | 1.11 | - | 372.5 | 88.6 | + |
| 26* | 0.57 | 0.68 | 1.25 | - | 185.0 | 88.6 | + |
| 28* | 1.23 | 0.06 | 1.29 | - | 296.1 | 88.6 | + |
| 31* | 1.24 | 0.16 | 1.40 | - | 358.2 | 88.6 | + |
| 32* | 0.62 | 0.83 | 1.45 | - | 171.6 | 88.6 | + |
| 35** | 0.24 | 2.38 | 2.62 | - | 181.9 | 88.6 | + |
| 36** | 0.20 | 2.48 | 2.68 | - | 233.1 | 88.6 | + |
| 37** | 2.05 | 0.67 | 2.72 | - | 410.0 | 88.6 | + |
| 38** | 0.11 | 2.64 | 2.75 | - | 155.9 | 88.6 | + |
| 39** | 0.94 | 1.83 | 2.77 | - | 105.7 | 88.6 | + |
| 40** | 0.23 | 2.55 | 2.78 | - | 275.8 | 88.6 | + |
| 41** | 1.69 | 1.29 | 2.98 | - | 404.6 | 88.6 | + |
| 42** | 1.04 | 1.97 | 3.01 | - | 107.2 | 88.6 | + |
| 43** | 2.54 | 0.52 | 3.06 | - | 389.8 | 88.6 | + |
| 44** | 0.19 | 2.99 | 3.18 | - | 153.0 | 88.6 | + |
| 45** | 0.00 | 3.19 | 3.19 | - | 105.6 | 88.6 | + |
| 47** | 1.61 | 1.66 | 3.27 | - | 412.3 | 88.6 | + |
*Discrepant <50% of the cutoff concentration (0.1 – 2.49 ng/mL)
**Discrepant between 50% of the cutoff to the cutoff concentration (2.5 – 4.99 ng/mL)
*These discrepancies are due to known causes. LZI Buprenorphine II Enzyme Immunoassay cross-reacts with both buprenorphine glucuronide and norbuprenorphine glucuronide, which explains the elevated EIA values and false positives observed in discrepant samples #21, #22, #24, #26, #28, #31 and #32.
Page 16
Lin-Zhi International, Inc.
Performance Characteristics Summary, continued:
Beckman Coulter® AU480 Analyzer
Cross-reactivity: 5 cutoff
The cross-reactivity of various potentially interfering drugs were tested by spiking various concentrations of each substance into a pool of negative human urine and then evaluated against the assay's calibration curve in qualitative mode. All samples were tested in duplicates.
The table below lists the concentration of each test compound that gave a response approximately equivalent to that of the cutoff calibrator (as positive) or the maximal concentration of the compound tested that gave a response below the response of the cutoff calibrator (as negative). Compounds tested at high concentration (100,000 ng/mL) with results below the cutoff value were listed as Not Detected (ND).
Buprenorphine and Major Metabolites:
| Compound | Test Concentration (ng/mL) | Qualitative Result (mAU) | Semi-Quantitative Result (ng/mL) | % Cross-reactivity |
|---|---|---|---|---|
| Norbuprenorphine | 5.0 | 66.9 | 5.3 | 100.0% |
| Buprenorphine | 5.0 | 71.0 | 5.6 | 100.0% |
| Buprenorphine Glucuronide | 32.0 | 83.4 | 6.1 | 15.6% |
| Norbuprenorphine Glucuronide | 210.0 | 66.4 | 5.0 | 2.4% |
Structurally related or unrelated Opiate compounds:
| Compound | Test Concentration (ng/mL) | Qualitative Result (mAU) | Semi-Quantitative Result (ng/mL) | % Cross-reactivity |
|---|---|---|---|---|
| Codeine | 100,000 | -17.2 | -0.98 | ND |
| Codeine 6β D-Glucuronide | 100,000 | 0.0 | 0.18 | ND |
| Dextromethorphan | 100,000 | 14.3 | 1.43 | ND |
| Dextrophan Tartrate | 100,000 | -7.3 | -0.15 | ND |
| Dihydrocodeine | 100,000 | 3.5 | 0.51 | ND |
| Dihydromorphine | 100,000 | -7.4 | -0.23 | ND |
| EDDP | 100,000 | -6.0 | -0.08 | ND |
| EMDP | 100,000 | 12.1 | 1.28 | ND |
| Ethylmorphine | 100,000 | -8.0 | -0.15 | ND |
| Fentanyl | 100,000 | -22.4 | -0.66 | ND |
| Heroin | 100,000 | -8.8 | -0.09 | ND |
| Hydrocodone | 100,000 | -18.4 | -0.63 | ND |
| Hydromorphone | 100,000 | -20.8 | -0.81 | ND |
| Hydromorphone 3β D-Glucuronide | 10,000 | -5.0 | -0.02 | ND |
| LAAM | 100,000 | 12.3 | 1.33 | ND |
| Levorphanol | 100,000 | -5.8 | -0.05 | ND |
| Meperidine | 100,000 | 12.6 | 1.38 | ND |
| Meperidine | 100,000 | -15.6 | -0.68 | ND |
Page 17
Lin-Zhi International, Inc.
Performance Characteristics Summary, continued:
Beckman Coulter® AU480 Analyzer
Cross-reactivity: 5 cutoff, continued
| Compound | Test Concentration (ng/mL) | Qualitative Result (mAU) | Semi-Quantitative Result (ng/mL) | % Cross-reactivity |
|---|---|---|---|---|
| Methadone | 100,000 | -18.7 | -0.73 | ND |
| Morphine | 100,000 | -11.8 | -0.73 | ND |
| Morphine 3β D-Glucuronide | 100,000 | -6.0 | -0.32 | ND |
| Morphine 6 D-Glucuronide | 100,000 | -3.1 | 0.24 | ND |
| N-Desmethyl-cis-Tramadol | 100,000 | -7.5 | -0.10 | ND |
| Nalbuphine | 100,000 | -5.8 | -0.21 | ND |
| Nalorphine | 100,000 | -2.1 | 0.12 | ND |
| Naloxegol | 100,000 | 36.9 | 1.86 | ND |
| Naloxone | 100,000 | 1.0 | -0.20 | ND |
| Naltrexone | 100,000 | 15.2 | 1.26 | ND |
| Norcodeine | 100,000 | -8.9 | -0.35 | ND |
| Norhydrocodone | 100,000 | -7.4 | -0.09 | ND |
| Normorphine | 100,000 | -8.6 | -0.27 | ND |
| Noroxycodone | 100,000 | -4.6 | -0.05 | ND |
| Noroxymorphone | 100,000 | -4.5 | 0.29 | ND |
| Norpropoxyphene | 100,000 | 12.0 | 1.40 | ND |
| O-Desmethyl-cis-Tramadol | 100,000 | -8.5 | -0.36 | ND |
| Oxycodone | 100,000 | -13.5 | -0.57 | ND |
| Oxymorphone | 100,000 | -15.6 | -0.80 | ND |
| Oxymorphone 3β D-Glucuronide | 10,000 | -3.9 | 0.14 | ND |
| Pentazocine | 100,000 | -8.2 | -0.32 | ND |
| Tapentadol | 100,000 | -6.8 | -0.38 | ND |
| Thebaine | 100,000 | 11.7 | 1.37 | ND |
| Tilidine | 100,000 | -7.0 | -0.17 | ND |
| Tramadol | 100,000 | -5.2 | -0.33 | ND |
Page 18
Lin-Zhi International, Inc.
Performance Characteristics Summary, continued:
Beckman Coulter® AU480 Analyzer
Cross-reactivity: 5 cutoff, continued
Structurally unrelated compounds were additionally spiked into pooled negative human urine to desired concentrations (as described above). These solutions were then split into three portions; one without norbuprenorphine, and the remaining two that were further spiked with norbuprenorphine standards to a final norbuprenorphine concentration of 3.75 ng/mL or 6.25 ng/mL (as negative or positive controls, ±25% of the cutoff concentration, respectively). Samples were then evaluated against the assay's calibration curve in qualitative mode. All samples were tested in duplicates. If discrepant results were observed, the lowest tested concentration at which discrepancies occurred is presented in the following table.
Structurally Unrelated Pharmacological Compounds:
| Compound | Test Concentration Norbuprenorphine (ng/mL) | 0 ng/mL % Cross | -25% Norbuprenorphine Cutoff (3.75 ng/mL) Result | +25% Norbuprenorphine Cutoff (6.25 ng/mL) Result |
|---|---|---|---|---|
| Acetaminophen | 100,000 | Neg | Neg | Pos |
| 6-Acetylmorphine | 100,000 | Neg | Neg | Pos |
| Acetylsalicylic Acid | 100,000 | Neg | Neg | Pos |
| Amitriptyline | 100,000 | Neg | Neg | Pos |
| Amlodipine Besylate | 100,000 | Neg | Neg | Pos |
| Amoxicillin | 100,000 | Neg | Neg | Pos |
| d-Amphetamine | 100,000 | Neg | Neg | Pos |
| Atorvastatin | 100,000 | Neg | Neg | Pos |
| Benzoylecgonine | 100,000 | Neg | Neg | Pos |
| Bupropion | 100,000 | Neg | Neg | Pos |
| Caffeine | 100,000 | Neg | Neg | Pos |
| Carbamazepine | 100,000 | Neg | Neg | Pos |
| Cetirizine | 100,000 | Neg | Neg | Pos |
| Chlorpheniramine | 100,000 | Neg | Neg | Pos |
| Chlorpromazine | 100,000 | Neg | Neg | Pos |
| Clomipramine | 100,000 | Neg | Neg | Pos |
| Desipramine | 100,000 | Neg | Neg | Pos |
| Diphenhydramine | 100,000 | Neg | Neg | Pos |
| Duloxetine | 100,000 | Neg | Neg | Pos |
| Fluoxetine | 100,000 | Neg | Neg | Pos |
| Fluphenazine | 100,000 | Neg | Neg | Pos |
| Gabapentin | 100,000 | Neg | Neg | Pos |
| Ibuprofen | 100,000 | Neg | Neg | Pos |
| Imipramine | 100,000 | Neg | Neg | Pos |
| Lisinopril | 100,000 | Neg | Neg | Pos |
| Losartan | 100,000 | Neg | Neg | Pos |
| Loratidine | 100,000 | Neg | Neg | Pos |
| MDA (3,4-methylenedioxyamphetamine) | 100,000 | Neg | Neg | Pos |
Page 19
Lin-Zhi International, Inc.
Performance Characteristics Summary, continued:
Beckman Coulter AU480 Analyzer
Cross-reactivity, continued:
Structurally Unrelated Pharmacological Compounds, continued:
| Compound | Test Concentration Norbuprenorphine (ng/mL) | 0 ng/mL % Cross | -25% Norbuprenorphine Cutoff (3.75 ng/mL) Result | +25% Norbuprenorphine Cutoff (6.25 ng/mL) Result |
|---|---|---|---|---|
| MDEA | 100,000 | Neg | Neg | Pos |
| MDMA (3,4-methylenedioxymethamphetamine) | 100,000 | Neg | Neg | Pos |
| Metformin | 100,000 | Neg | Neg | Pos |
| Metoprolol | 100,000 | Neg | Neg | Pos |
| d-Methamphetamine | 100,000 | Neg | Neg | Pos |
| Nalmefene | 100,000 | Neg | Neg | Pos |
| Nicotine | 100,000 | Neg | Neg | Pos |
| Nortriptyline | 100,000 | Neg | Neg | Pos |
| Omeprazole | 100,000 | Neg | Neg | Pos |
| Oxazepam | 100,000 | Neg | Neg | Pos |
| Phenobarbital | 100,000 | Neg | Neg | Pos |
| (1S,2S)-(+)Pseudoephedrine | 100,000 | Neg | Neg | Pos |
| Quetiapine | 100,000 | Neg | Neg | Pos |
| Ranitidine | 100,000 | Neg | Neg | Pos |
| Salbutamol (Albuterol) | 100,000 | Neg | Neg | Pos |
| Sertraline | 100,000 | Neg | Neg | Pos |
| THC-COOH (11-Nor-Delta-9-THC-9-carboxylic acid) | 100,000 | Neg | Neg | Pos |
| L-Thyroxine | 100,000 | Neg | Neg | Pos |
| Zolpidem | 100,000 | Neg | Neg | Pos |
Page 20
Lin-Zhi International, Inc.
Performance Characteristics Summary, continued:
Beckman Coulter AU480 Analyzer
Endogenous and Preservative Compound Interference 5 cutoff:
Endogenous and Preservative compounds were spiked into pooled negative human urine to desired concentrations. These solutions were then split into three portions; one without norbuprenorphine, and the remaining two that were further spiked with norbuprenorphine standards to a final norbuprenorphine concentration of 3.75 ng/mL or 6.25 ng/mL (as negative or positive controls, ±25% of the cutoff concentration, respectively). Samples were then evaluated against the assay's calibration curve in qualitative mode. All samples were tested in duplicates.
Interference was observed with Boric Acid at 1% w/v. No other significant cross-reactivity was observed.
| Interfering Substance | Concentration of Compound (mg/dL) | 0 ng/mL Norbuprenorphine | -25% Norbuprenorphine Cutoff (3.75 ng/mL) | +25% Norbuprenorphine Cutoff (6.25 ng/mL) |
|---|---|---|---|---|
| Acetone | 1,000 | Neg | Neg | Pos |
| Ascorbic acid | 500 | Neg | Neg | Pos |
| Bilirubin | 2 | Neg | Neg | Pos |
| Biotin | 2 | Neg | Neg | Pos |
| Boric acid | 1,000 | Neg | Neg | Neg |
| Calcium chloride | 300 | Neg | Neg | Pos |
| Citric acid | 200 | Neg | Neg | Pos |
| Creatinine | 500 | Neg | Neg | Pos |
| Ethanol | 1,000 | Neg | Neg | Pos |
| Galactose | 10 | Neg | Neg | Pos |
| γ-Globulin | 500 | Neg | Neg | Pos |
| Glucose | 3,000 | Neg | Neg | Pos |
| Hemoglobin | 300 | Neg | Neg | Pos |
| Human urine (pooled) | N/A | Neg | Neg | Pos |
| Human serum albumin | 500 | Neg | Neg | Pos |
| β-Hydroxybutyric acid | 100 | Neg | Neg | Pos |
| Oxalic acid | 100 | Neg | Neg | Pos |
| Potassium chloride | 1,000 | Neg | Neg | Pos |
| Riboflavin | 7.5 | Neg | Neg | Pos |
| Sodium azide | 1,000 | Neg | Neg | Pos |
| Sodium chloride | 1,000 | Neg | Neg | Pos |
| Sodium fluoride | 1,000 | Neg | Neg | Pos |
| Sodium phosphate | 300 | Neg | Neg | Pos |
| Urea | 6,000 | Neg | Neg | Pos |
| Uric acid | 10 | Neg | Neg | Pos |
| Urine-based calibrator buffer | N/A | Neg | Neg | Pos |
Page 21
Lin-Zhi International, Inc.
Performance Characteristics Summary, continued:
Beckman Coulter AU480 Analyzer
Specific Gravity Interference 5 cutoff:
Samples ranging in specific gravity from 1.000 to 1.030 were split into three portions each and either left un-spiked or further spiked to a final norbuprenorphine concentration of either 3.75 ng/mL or 6.25 ng/mL (as negative or positive controls, ±25% of the cutoff concentration, respectively). These samples were then evaluated in qualitative and semi-quantitative modes. No interference was observed.
| Specific Gravity Value | 0 ng/mL Norbuprenorphine | -25% Norbuprenorphine Cutoff (3.75 ng/mL) | +25% Norbuprenorphine Cutoff (6.25 ng/mL) |
|---|---|---|---|
| 1.000 | Neg | Neg | Pos |
| 1.005 | Neg | Neg | Pos |
| 1.007 | Neg | Neg | Pos |
| 1.010 | Neg | Neg | Pos |
| 1.013 | Neg | Neg | Pos |
| 1.015 | Neg | Neg | Pos |
| 1.017 | Neg | Neg | Pos |
| 1.020 | Neg | Neg | Pos |
| 1.030 | Neg | Neg | Pos |
Page 22
Lin-Zhi International, Inc.
Performance Characteristics Summary, continued:
Beckman Coulter AU480 Analyzer
pH Interference 5 cutoff:
Negative urine and urine spiked with norbuprenorphine to the final norbuprenorphine concentration of either 3.75 ng/mL or 6.25 ng/mL (as negative or positive controls, ±25% of the cutoff concentration, respectively) were adjusted to the following pH levels and tested by the assay. The pH adjusted solutions were evaluated in qualitative mode.
No major interference was observed between pH 3 to pH 11. Results are summarized in the following table:
| Interfering Substance | 0 ng/mL Norbuprenorphine | -25% Norbuprenorphine Cutoff (3.75 ng/mL) | +25% Norbuprenorphine Cutoff (6.25 ng/mL) |
|---|---|---|---|
| pH 3 | Neg | Neg | Pos |
| pH 4 | Neg | Neg | Pos |
| pH 5 | Neg | Neg | Pos |
| pH 6 | Neg | Neg | Pos |
| pH 7 | Neg | Neg | Pos |
| pH 8 | Neg | Neg | Pos |
| pH 9 | Neg | Neg | Pos |
| pH 10 | Neg | Neg | Pos |
| pH 11 | Neg | Neg | Pos |
Page 23
Lin-Zhi International, Inc.
Conclusion:
The information provided in this pre-market notification demonstrates that the LZI Buprenorphine II Enzyme Immunoassay is substantially equivalent to the legally marketed predicate device for its general intended use. Substantial equivalence was demonstrated through comparison of intended use and physical properties to the commercially available predicate device as confirmed by chromatography/mass spectrometry (LC/MS), an independent analytical method. The information supplied in this pre-market notification provides reasonable assurance that the LZI Buprenorphine II Enzyme Immunoassay is safe and effective for its stated intended use.
§ 862.3650 Opiate test system.
(a)
Identification. An opiate test system is a device intended to measure any of the addictive narcotic pain-relieving opiate drugs in blood, serum, urine, gastric contents, and saliva. An opiate is any natural or synthetic drug that has morphine-like pharmocological actions. The opiates include drugs such as morphine, morphine glucoronide, heroin, codeine, nalorphine, and meperedine. Measurements obtained by this device are used in the diagnosis and treatment of opiate use or overdose and in monitoring the levels of opiate administration to ensure appropriate therapy.(b)
Classification. Class II (special controls). An opiate test system is not exempt if it is intended for any use other than employment or insurance testing or is intended for Federal drug testing programs. The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 862.9, provided the test system is intended for employment and insurance testing and includes a statement in the labeling that the device is intended solely for use in employment and insurance testing, and does not include devices intended for Federal drug testing programs (e.g., programs run by the Substance Abuse and Mental Health Services Administration (SAMHSA), the Department of Transportation (DOT), and the U.S. military).