(157 days)
The Labcorp Fentanyl Urine Visual Test is a lateral flow competitive immunoassay for the rapid qualitative detection of norfentanyl (fentanyl metabolite) in human urine at a cutoff of 5 ng/mL. It is intended for prescription use. For in vitro diagnostic use only.
This test provides only a preliminary result. Clinical consideration and professional judgment must be applied to any drug test result, particularly in evaluating a preliminary positive result. To confirm preliminary positive results, a more specific analytical method must be used. Gas Chromatography-Mass Spectrometry (GC-MS), Liquid Chromatography Tandem Mass Spectrometry (LC-MS/MS), and High Performance Liquid Chromatography (HPLC) are the preferred confirmatory methods.
The Labcorp Fentanyl Urine Visual Test (the "DEVICE") is a lateral flow competitive immunoassay for the rapid qualitative detection of Norfentanyl (nFEN), the primary urinary metabolite of Fentanyl, in human urine at concentrations above 5 ng/mL. It is intended for prescription use.
The single use, in vitro diagnostic DEVICE is available in a cassette format with a disposable dropper provided for sample transfer.
The DEVICE contains a test strip that gives a qualitative result for presence of Norfentanyl in human urine. The DEVICE is read visually and has labeling with instructions for interpreting test results.
N/A
U.S. Food & Drug Administration 510(k) Clearance Letter
Page 1
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
Doc ID # 04017.08.02
January 29, 2026
MEDTOX Diagnostics, Inc.
Elizabeth Brock
Regulatory Affairs Manager
1238 Anthony Road
Burlington, North Carolina 27215
Re: K252684
Trade/Device Name: Labcorp Fentanyl Urine Visual Test
Regulation Number: 21 CFR 862.3650
Regulation Name: Opiate Test System
Regulatory Class: Class II
Product Code: DJG
Dated: August 22, 2025
Received: August 25, 2025
Dear Elizabeth Brock:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
Page 2
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801 and Part 809); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-
Page 3
assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
JOSEPH A. KOTAREK -S
Digitally signed by JOSEPH A. KOTAREK -S
Date: 2026.01.29 16:21:43 -05'00'
Joseph Kotarek
Toxicology Branch Chief
Division of Chemistry and Toxicology Devices
OHT7: Office of In Vitro Diagnostics
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
Page 4
DEPARTMENT OF HEALTH AND HUMAN SERVICES
Food and Drug Administration
Indications for Use
Form Approved: OMB No. 0910-0120
Expiration Date: 07/31/2026
See PRA Statement below.
510(k) Number (if known): K252684
Device Name: Labcorp Fentanyl Urine Visual Test
Indications for Use (Describe)
The Labcorp Fentanyl Urine Visual Test is a lateral flow competitive immunoassay for the rapid qualitative detection of norfentanyl (fentanyl metabolite) in human urine at a cutoff of 5 ng/mL. It is intended for prescription use. For in vitro diagnostic use only.
This test provides only a preliminary result. Clinical consideration and professional judgment must be applied to any drug test result, particularly in evaluating a preliminary positive result. To confirm preliminary positive results, a more specific analytical method must be used. Gas Chromatography-Mass Spectrometry (GC-MS), Liquid Chromatography Tandem Mass Spectrometry (LC-MS/MS), and High Performance Liquid Chromatography (HPLC) are the preferred confirmatory methods.
Type of Use (Select one or both, as applicable)
☑ Prescription Use (Part 21 CFR 801 Subpart D)
☐ Over-The-Counter Use (21 CFR 801 Subpart C)
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services
Food and Drug Administration
Office of Chief Information Officer
Paperwork Reduction Act (PRA) Staff
PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
FORM FDA 3881 (8/23) Page 1 of 1 PSC Publishing Services (301) 443-6740 EF
Page 5
K252684 510(k) SUMMARY
[In accordance with 21 CFR 807.92]
Submitter
MEDTOX Diagnostics, Inc.
1238 Anthony Road
Burlington, North Carolina 27215
Phone: 336-226-6311
Facsimile: 336-229-4471
Contact Person: Elizabeth Brock
Regulatory Affairs Manager
Phone: 336-482-2862
Date Prepared: December 19, 2025
Trade or Proprietary Name of Device: Labcorp Fentanyl Urine Visual Test
Common or Usual Name: Drugs of Abuse Screen Urine Test
Regulatory Class: Class II
Regulatory information:
| Product Code | Classification | Regulation Section | Panel |
|---|---|---|---|
| DJG | Class II | 862.3650, Opiate Test System (Fentanyl) | Clinical Toxicology |
Predicate Devices
Chemtron Biotech, Inc., Chemtrue® Drug Screen Fentanyl Test (K232736)
Reference Method: Fentanyl Liquid Chromatography / Mass Spectroscopy / Mass Spectroscopy (LC-MS/MS)
Page 1 of 12
Page 6
K252684 510(k) SUMMARY
Device Description
The Labcorp Fentanyl Urine Visual Test (the "DEVICE") is a lateral flow competitive immunoassay for the rapid qualitative detection of Norfentanyl (nFEN), the primary urinary metabolite of Fentanyl, in human urine at concentrations above 5 ng/mL. It is intended for prescription use.
The single use, in vitro diagnostic DEVICE is available in a cassette format with a disposable dropper provided for sample transfer.
The DEVICE contains a test strip that gives a qualitative result for presence of Norfentanyl in human urine. The DEVICE is read visually and has labeling with instructions for interpreting test results.
Intended Use / Indications for Use
The Labcorp Fentanyl Urine Visual Test is a lateral flow competitive immunoassay for the rapid qualitative detection of norfentanyl (fentanyl metabolite) in human urine at a cutoff of 5 ng/mL. It is intended for prescription use. For in vitro diagnostic use only.
This test provides only a preliminary result. Clinical consideration and professional judgment must be applied to any drug test result, particularly in evaluating a preliminary positive result. To confirm preliminary positive results, a more specific analytical method must be used. Gas Chromatography-Mass Spectrometry (GC-MS), Liquid Chromatography Tandem Mass Spectrometry (LC-MS/MS), and High Performance Liquid Chromatography (HPLC) are the preferred confirmatory methods.
Summary of Technological Characteristics – comparison with predicate
The Labcorp Fentanyl Urine Visual Test (the "DEVICE") has the same technological characteristics as the predicate device. Both the subject DEVICE and the predicate device are in vitro diagnostic devices utilizing colloidal gold based lateral flow immunoassay technology to detect the presence of norfentanyl in human urine. Both devices have the same test procedure; a urine sample is added to the test device and allowed to react for a specified period of time, after which the user interprets the test results visually. Both devices are single-use devices.
The subject DEVICE differs from the predicate device in terms of test device format, and aspects such as packaging or labeling. The subject DEVICE is a cassette design with a dropper for sample transfer, whereas the predicate device comes in a dip card or cup format. These differences do not raise new questions of safety or effectiveness.
Overall characteristics of the subject DEVICE and the predicate device are summarized in Table 1 below:
Table 1. Comparison of Similarities/Differences for the DEVICE and Predicate Device.
| Item | Subject Device | Predicate K232736 |
|---|---|---|
| Device Trade Name | Labcorp Fentanyl Urine Visual Test | Chemtrue® Drug Screen Fentanyl Cup/Dip Card Tests |
| Intended Use/ Indications for Use | Same | Qualitative detection of drugs of abuse in human urine |
Page 2 of 12
Page 7
K252684 510(k) SUMMARY
| Item | Subject Device | Predicate K232736 |
|---|---|---|
| Device Type | Same | In vitro diagnostic |
| Analyte | Same | Norfentanyl |
| Cutoff | Same | 5 ng/mL |
| Specimen Type | Same | Human Urine |
| Method | Same | Competitive lateral flow immunoassay |
| Test Device Format | Cassette | Cup/Dip Card |
| Test Procedure | Sample is added to a single use test cassette by disposable dropper. The test is then timed and interpreted visually. | Sample is added to a single use test cup or dip card. The test is then timed and interpreted visually. |
| Conditions for Use | Prescription use | OTC – Over The Counter |
Principle of Operation
The DEVICE uses lateral flow competitive immunoassay technology in which a modified drug (drug-protein conjugate) competes with drug that may be present in the urine. The colorless drug-protein conjugate is applied to the membrane of the test strip at the test line position. The colored drug-specific antibody-colloidal gold is applied to the sample pad of the test strip. When urine is transferred onto the sample pad it dissolves and migrates the colored antibody-colloidal gold across the strip membrane. If drug is not present or is below the cutoff, the drug-specific antibodies will bind to the conjugate line on the membrane and the colloidal gold will form a visible line, which indicates a negative result. If drug is present in the urine, the drug-specific antibodies will bind to the drug in solution and bypass the conjugate on the membrane and no line will form, which indicates a preliminary positive result.
A visible control line should form in the control line position regardless of whether drug is present to indicate that the test has been performed properly.
Performance Characteristics
Performance data for the DEVICE are summarized and presented below.
A. Analytical Performance:
1. Precision/Reproducibility:
Precision/reproducibility studies were carried out for the DEVICE using urine samples containing Norfentanyl spiked into a drug-free urine pool with LC-MS/MS confirmed concentrations of 0%, 25%, 50%, 75%, 100%, 125%, 150%, and 200% of the test cutoff. These samples were prepared in a blind randomized panel in triplicate and evaluated by three (3) in-house operators over five (5) days using three (3) different lots. The results are summarized in Table 2 below.
Page 3 of 12
Page 8
K252684 510(k) SUMMARY
Table 2. Sensitivity/Precision/Distribution of Random Error
| % of Cutoff | 1st Lot Neg/Pos | 2nd Lot Neg/Pos | 3rd Lot Neg/Pos | Total Neg/Pos |
|---|---|---|---|---|
| Neg | 45/0 | 45/0 | 45/0 | 135/0 |
| 25% | 45/0 | 45/0 | 45/0 | 135/0 |
| 50% | 45/0 | 45/0 | 45/0 | 135/0 |
| 75% | 40/5 | 35/10 | 31/14 | 106/29 |
| 100% | 20/25 | 22/23 | 18/27 | 60/75 |
| 125% | 3/42 | 8/37 | 3/42 | 14/121 |
| 150% | 0/45 | 0/45 | 0/45 | 0/135 |
| 200% | 0/45 | 0/45 | 0/45 | 0/135 |
2. Linearity:
Not applicable. These devices are intended for qualitative use only.
3. Analytical Specificity/Interference:
Analytical specificity (cross reactivity and interference) data are summarized below.
Related Compounds and Cross-Reactants
Structurally related and reacting compounds for Norfentanyl were evaluated on the DEVICE for cross reactivity by diluting reference standards in drug-free urine pool.
Results for related and reacting compounds shown in Table 3 below are expressed as follows:
- Reactive compound: approximate minimum concentration required to produce a positive result, listed first in the table
- Non-reactive compound: concentration where no interference was observed, listed last in the table (results negative at all concentrations)
- The "% Cross-Reactive" values were calculated from the cutoff level for the DEVICE divided by the minimum positive concentration as percentage.
Table 3. Related Compounds and Cross-reactants
| Compound | Result | % Cross-Reactive |
|---|---|---|
| Norfentanyl (nFEN) | (Norfentanyl, 5 ng/mL) | |
| Parent drug: Fentanyl | Positive at 7.5 ng/mL | 67% |
| Fentanyl/Norfentanyl analog cross-reactants | ||
| 3-Methylfentanyl | Positive at 150 ng/mL | 3% |
| 4-Fluoro-Isobutyryl Fentanyl | Positive at 15 ng/mL | 33% |
| 4'-Methylacetyl Fentanyl (para-Methacetyl Fentanyl) | Positive at 750 ng/mL | 1% |
| Acetyl Fentanyl | Positive at 5 ng/mL | 100% |
| Acetyl Norfentanyl | Positive at 50 ng/mL | 10% |
| Acryl Fentanyl | Positive at 10 ng/mL | 50% |
Page 4 of 12
Page 9
K252684 510(k) SUMMARY
| Compound | Result | % Cross-Reactive |
|---|---|---|
| Alfentanil | Positive at 2,500 ng/mL | < 1% |
| Butyryl Fentanyl | Positive at 10 ng/mL | 50% |
| Cis-DL-3-Methylfentanyl | Positive at 100 ng/mL | 5% |
| Cyclopropyl Fentanyl | Positive at 10 ng/mL | 50% |
| Cyclopropyl Norfentanyl | Positive at 7.5 ng/mL | 67% |
| Fluorocyclopropyl Fentanyl (para-Fluorocyclopropyl Fentanyl) | Positive at 15 ng/mL | 33% |
| Furanyl Fentanyl | Positive at 25 ng/mL | 20% |
| Furanyl Norfentanyl | Positive at 7.5 ng/mL | 67% |
| (±) β-Hydroxythiofentanyl | Positive at 10 ng/mL | 50% |
| Isobutyryl Fentanyl | Positive at 7.5 ng/mL | 67% |
| Isobutyryl Norfentanyl | Positive at 5 ng/mL | 100% |
| Methoxyacetylfentanyl (MAF) | Positive at 15 ng/mL | 33% |
| N-Benzylfuranyl Norfentanyl | Positive at 50 ng/mL | 10% |
| N-Benzyl-Para-Fluoro Norfentanyl | Positive at 25 ng/mL | 20% |
| Octfentanil | Positive at 500 ng/mL | 1% |
| Para-Fluoro-Butyryl Fentanyl (p-FBF) | Positive at 75 ng/mL | 7% |
| Para-Fluorofentanyl | Positive at 10 ng/mL | 50% |
| Phenylacetyl Fentanyl | Positive at 7.5 ng/mL | 67% |
| Remifentanil | Positive at 100 ng/mL | 5% |
| THF Fentanyl (THFF) | Positive at 35 ng/mL | 14% |
| Thienyl Fentanyl | Positive at 15 ng/mL | 33% |
| Trans-DL-3-Methylfentanyl | Positive at 10 ng/mL | 50% |
| Valeryl Fentanyl | Positive at 50 ng/mL | 10% |
| Non-Fentanyl cross-reactants | ||
| Fluphenazine | Positive at 50,000 ng/mL | < 1% |
| Perphenazine | Positive at 75,000 ng/mL | < 1% |
| Quinidine | Positive at 25,000 ng/mL | < 1% |
| Quinine | Positive at 50,000 ng/mL | < 1% |
| Risperidone | Positive at 2,500 ng/mL | < 1% |
| Risperidone Metabolite: 9-Hydroxyrisperidone (Paliperidone) | Positive at 2,500 ng/mL | < 1% |
| Non-reactive Fentanyl/Norfentanyl analogs | ||
| Carfentanil | Negative at 500 ng/mL | Non-reactive* |
| Despropionylfentanyl (4-ANPP) | Negative at 1,000 ng/mL | Non-reactive |
| MT-45 | Negative at 3,500 ng/mL | Non-reactive* |
| 4-Methoxybutyryl Fentanyl | Negative at 5,000 ng/mL | Non-reactive* |
| Norcarfentanil | Negative at 1,000 ng/mL | Non-reactive |
| Sufentanil | Negative at 1,000 ng/mL | Non-reactive |
| U-47700 | Negative at 5,000 ng/mL | Non-reactive* |
*These Non-reactive compounds produced negative interference with the test at higher concentrations, however this interference is not expected to impact test performance.
Page 5 of 12
Page 10
K252684 510(k) SUMMARY
Interference Data
The DEVICE was evaluated for cross-reactivity by assaying each of the below compounds or conditions in a drug-free urine pool, and for interference by assaying each compound or condition in drug-free urine pool fortified with Norfentanyl at 50% of cutoff or 150% of cutoff separately. Per the Precision/Reproducibility data Norfentanyl at 50% of cutoff is the concentration where results for all lots were 95% negative or greater (C5), and Norfentanyl at 150% of cutoff is the concentration where results for all lots were 95% positive or greater (C95). Each compound or condition was assayed one at a time using five (5) replicates for each concentration of Norfentanyl and each result was visually interpreted. The compound concentrations and conditions below produced consistently negative results with the negative urine pool, and the 50% / 150% of cutoff urines gave the expected negative / positive results.
Non-cross-reactive Endogenous Compounds:
The DEVICE was evaluated for cross-reactivity and interference with endogenous compounds. None of these compounds demonstrated cross-reactivity or interference at the indicated concentration levels tested with the DEVICE. The endogenous compounds and their non-reactive concentration are summarized in Table 4 below.
Table 4. Non-reactive Endogenous Compounds
| Endogenous Compound | Concentration (ng/mL) | Endogenous Compound | Concentration (ng/mL) |
|---|---|---|---|
| Acetone | 10,000,000 | Glucose | 30,000,000 |
| Albumin, Human | 5,000,000 | Hemoglobin, Human | 3,000,000 |
| Ascorbic Acid | 5,000,000 | Niacinamide | 100,000 |
| Atropine | 100,000 | Nicotinic Acid | 100,000 |
| Beta-Hydroxybutyric Acid | 100,000 | Octopamine | 150,000 |
| Bilirubin | 20,000 | Oxalic Acid | 100,000 |
| Biotin | 100,000 | Potassium Chloride | 10,000,000 |
| Calcium Chloride | 3,000,000 | Pyridoxine | 90,000 |
| Cholesterol | 100,000 | Quinolinic Acid | 100,000 |
| Creatine Hydrate | 100,000 | Riboflavin | 75,000 |
| Creatinine | 5,000,000 | Sodium Chloride | 10,000,000 |
| Deoxycorticosterone | 100,000 | Thiamine | 100,000 |
| Dopamine (3-Hydroxytyramine) | 100,000 | Tryptamine | 100,000 |
| Ethanol | 10,000,000 | Tyramine | 100,000 |
| Galactose | 100,000 | Urea | 60,000,000 |
| Gamma Globulin | 5,000,000 | Uric Acid | 100,000 |
pH and Specific Gravity:
The DEVICE was evaluated for pH interference. Each negative and fortified urine control was adjusted by either 6N NaOH or 6N HCl to pH values across the normal range for human urine – 4.0, 4.5, 5.0, 5.5, 6.0, 6.5, 7.0, 7.5, 8.0, 8.5, and 9.0. None of the pH conditions demonstrated cross-reactivity or interference with the DEVICE.
Page 6 of 12
Page 11
K252684 510(k) SUMMARY
The DEVICE was evaluated for specific gravity interference by assaying negative and fortified urine samples that had specific gravity values across the normal range for human urine – 1.001, 1.004, 1.010, 1.014, 1.020, 1.025, and 1.030. None of the specific gravity conditions demonstrated cross-reactivity or interference with the DEVICE.
Common Drugs:
The DEVICE was evaluated for cross-reactivity and interference with commonly prescribed drugs, common over-the-counter drugs, and common drugs of abuse. Concentrations were used that were at or above a medically relevant concentration when feasible (concentrations in urine known/calculatable and compound availability). None of the compounds demonstrated cross-reactivity or interference at the indicated concentration levels tested with the DEVICE. The common drugs and their non-reactive concentration are summarized in Table 5 below.
Table 5. Non-reactive Common Drugs
| Common Drug | Concentration (ng/mL) | Common Drug | Concentration (ng/mL) |
|---|---|---|---|
| 1-(3-Trifluoromethylphenyl)-piperazine | 100,000 | Isotonitazene | 10,000 |
| 11-Hydroxy-Δ9-THC | 10,000 | Ketamine | 100,000 |
| 11-Nor-9-carboxy-Δ9-THC | 10,000 | Labetalol | 1,800,000 |
| 4-Bromo-2,5, Dimethoxyphenethylamine | 100,000 | Lamotrigine | 100,000 |
| 6-Acetylmorphine (6-MAM) | 100,000 | Lansoprazole | 100,000 |
| 7-Aminoclonazepam | 100,000 | L-Cotinine | 100,000 |
| 7-Aminoflunitrazepam | 100,000 | L-Erythromycin | 100,000 |
| 7-Aminonitrazepam | 10,000 | Levonorgestrel | 100,000 |
| Acetaminophen | 2,900,000 | Levorphanol | 100,000 |
| Acetylsalicylic Acid | 100,000 | Levothyroxine (L-Thyroxine) | 100,000 |
| AH-7921 (Doxylam) | 10,000 | Lisinopril | 100,000 |
| AH-8529 | 100,000 | L-Methamphetamine | 100,000 |
| AH-8533 | 100,000 | Loperamide | 100,000 |
| Albuterol | 100,000 | Loratadine | 100,000 |
| Allopurinol | 100,000 | Lorazepam | 100,000 |
| Alprazolam | 100,000 | Losartan | 100,000 |
| Aminopyrine (4-Dimethylaminoantipyrine) | 100,000 | Lurasidone | 10,000 |
| Amlodipine Besylate | 100,000 | Lysergic Acid Diethylamide (LSD) | 10,000 |
| Amoxicillin | 2,600,000 | Maprotiline | 100,000 |
| Ampicillin | 100,000 | Meperidine | 100,000 |
| Apixaban | 100,000 | Meprobamate | 100,000 |
| Apomorphine | 100,000 | Metformin | 2,900,000 |
| Aspartame | 100,000 | Methadone | 100,000 |
| Atenolol | 200,000 | Methapyrilene | 100,000 |
| Atorvastatin | 100,000 | Methylphenidate | 100,000 |
| Baclofen | 100,000 | Metonitazene | 10,000 |
Page 7 of 12
Page 12
K252684 510(k) SUMMARY
| Common Drug | Concentration (ng/mL) | Common Drug | Concentration (ng/mL) |
|---|---|---|---|
| Benzocaine | 100,000 | Metoprolol | 550,000 |
| Benzoic Acid | 100,000 | Mitragynine | 100,000 |
| Benzoylecgonine | 100,000 | Morphine | 300,000 |
| Benzylpiperiazine | 10,000 | Morphine-3-glucuronide | 10,000 |
| Brompheniramine | 100,000 | Nalidixic Acid | 100,000 |
| Buprenorphine | 100,000 | Naloxegol | 100,000 |
| Bupropion | 100,000 | Naloxone | 100,000 |
| Caffeine | 1,200,000 | Naltrexone | 100,000 |
| Cannabidiol | 100,000 | Naproxen | 1,800,000 |
| Carbamazepine | 100,000 | N-Desmethyltapentadol | 10,000 |
| Carisoprodol | 100,000 | Nifedipine | 100,000 |
| Carvedilol | 100,000 | Norbuprenorphine | 100,000 |
| Cetirizine | 100,000 | Norcodeine | 100,000 |
| Chloramphenicol | 100,000 | Nordiazepam | 100,000 |
| Chlorcyclizine | 100,000 | Norethindrone | 100,000 |
| Chlordiazepoxide | 100,000 | Norketamine | 100,000 |
| Chlorpheniramine | 100,000 | Normeperidine | 100,000 |
| Chlorpromazine | 15,000 | Normorphine | 10,000 |
| Cimetidine | 1,800,000 | Noroxycodone | 100,000 |
| Citalopram | 100,000 | Norpropoxyphene | 100,000 |
| Clofibrate | 100,000 | Norpseudoephedrine (Cathine) | 100,000 |
| Clomipramine | 100,000 | Noscapine | 100,000 |
| Clonazepam | 100,000 | Omeprazole | 100,000 |
| Clonidine | 100,000 | Oxazepam | 100,000 |
| Clopidogrel | 200,000 | Oxazepam Glucuronide | 1,000 |
| Codeine | 120,000 | Oxycodone | 100,000 |
| Cortisone | 100,000 | Oxymetazoline | 100,000 |
| Cyclobenzaprine | 100,000 | Oxymorphone | 100,000 |
| Cyclodextrin-r | 100,000 | Pantoprazole | 100,000 |
| Cyproheptadine | 100,000 | Papaverine | 100,000 |
| DL-Isoproterenol | 100,000 | Penicillin-G (Benzylpenicillin) | 100,000 |
| D-Amphetamine | 350,000 | Pentazocine | 35,000 |
| Demoxepam | 100,000 | Perospirone | 10,000 |
| Dextromethorphan | 100,000 | Phenelzine | 100,000 |
| Diclofenac | 200,000 | Pheniramine | 100,000 |
| Difunisal | 100,000 | Phentermine | 100,000 |
| Dihydrocodeine | 100,000 | Phenylephrine | 100,000 |
| Diphenhydramine | 100,000 | Phenylethylamine | 100,000 |
| Diphenylhydantoin (Phenytoin) | 100,000 | Pravastatin | 100,000 |
| D-Methamphetamine | 100,000 | Prednisone | 100,000 |
| Doxylamine | 100,000 | Promazine | 100,000 |
Page 8 of 12
Page 13
K252684 510(k) SUMMARY
| Common Drug | Concentration (ng/mL) | Common Drug | Concentration (ng/mL) |
|---|---|---|---|
| D-Pseudoephedrine | 250,000 | Promethazine | 100,000 |
| Duloxetine | 150,000 | Propoxyphene | 10,000 |
| EDDP (Methadone metabolite) | 100,000 | Propranolol | 700,000 |
| Ephedrine | 100,000 | Propylhexedrine | 50,000 |
| Ergocalciferol | 100,000 | Pyrilamine | 150,000 |
| Escitalopram | 100,000 | Pyrogallol | 100,000 |
| Esomeprazole Mg hydrate | 250,000 | Ranitidine | 150,000 |
| Estrone | 100,000 | Rosuvastatin | 100,000 |
| Ethinyl Estradiol | 100,000 | Salicylic Acid | 1,000,000 |
| Famotidine | 1,000,000 | Sertindole | 10,000 |
| Fenfluramine | 100,000 | Sertraline | 30,000 |
| Fenofibrate | 100,000 | Simvastatin | 100,000 |
| Fexofenadine | 75,000 | Sodium Azide | 100,000 |
| Fluoxetine | 100,000 | Sulfamethazine | 100,000 |
| Fluticasone | 100,000 | Sulindac | 100,000 |
| Furosemide | 500,000 | Tamsulosin | 100,000 |
| Gabapentin | 4,500,000 | Temazepam | 100,000 |
| Gemfibrozil | 100,000 | Tetracycline | 100,000 |
| Gentisic Acid | 100,000 | Tetrahydrozoline | 100,000 |
| Glipizide | 100,000 | Thioridazine | 100,000 |
| Guaifenesin | 100,000 | Tianeptine | 100,000 |
| Heroin (Diacetylmorphine) | 100,000 | Tramadol | 450,000 |
| Hexobarbital | 10,000 | Trazodone | 10,000 |
| Hydralazine | 100,000 | Trifluoperazine | 100,000 |
| Hydrochlorothiazide | 150,000 | Venlafaxine | 100,000 |
| Hydrocodone | 100,000 | Xylazine | 100,000 |
| Hydrocortisone | 100,000 | Ziprasidone | 100,000 |
| Hydromorphone | 100,000 | Zolpidem | 100,000 |
| Ibuprofen | 2,700,000 | Zolpidem Tartrate | 100,000 |
| Insulin | 100,000 |
4. Assay Reportable Range:
Characterization of how the device performs around the claimed cutoff concentration appears in the precision study (section A.1 above). Higher concentrations perform equivalent to the highest concentration evaluated in the precision study. The DEVICE is a competitive test and there is no hook-effect at higher concentrations.
5. Traceability, Stability, Expected Values (Controls, Calibrators, or Methods):
Page 9 of 12
Page 14
K252684 510(k) SUMMARY
All drug calibrators of the device are traceable to available commercial reference materials.
6. Detection Limit:
Characterization of how the device performs at low analyte concentrations appears in the precision study (section A.1 above).
7. Assay Cut-Off:
Characterization of how the device performs around the claimed cutoff concentration appears in the precision study (section A.1 above).
8. Stability:
To establish and support the shelf life and expiration date, stability studies were conducted under accelerated temperature (at 45°C), and real time (2°C to 30°C) with three (3) lots of the DEVICE. The accelerated stability study results support two (2) years shelf life of the products stored at 2°C to 30°C. Real-time stability studies are still on going.
9. Flex Testing:
To evaluate robustness of the DEVICE, flex testing studies were conducted after the devices were exposed to various sample volumes, common types of contamination or damage, non-urine specimen types, preservatives, and humidity and temperature conditions. The results support the robustness of the DEVICE.
B. Comparison Studies:
1. Method Comparison with Predicate Device:
The accuracy of the DEVICE was evaluated by a method comparison study comparing the performance of the DEVICE to the LC-MS/MS reference method. Clinical urine samples were assayed in a randomized panel of blind coded urine samples that were interpreted visually. The eighty-seven (87) unaltered clinical urine samples contained either no target drug/metabolite or varying concentrations of Norfentanyl as determined by LC-MS/MS results. Three (3) operators each tested the 87 samples on a different test lot. The results are summarized in Table 6 below.
Page 10 of 12
Page 15
K252684 510(k) SUMMARY
Table 6. Accuracy with Clinical Samples Compared to LC-MS/MS
| Fentanyl Test (nFEN) (5 ng/mL Cutoff) | Concentration by LC-MS/MS (ng/mL) | ||||
|---|---|---|---|---|---|
| (-) | (+) | ||||
| No Drug Present | Low Negative (<50% of cutoff) | Near Cutoff Negative (50% - 100% of cutoff) | Near Cutoff Positive (100% - 150% of cutoff) | True Positive (>150% of cutoff) | |
| 1st Lot | Positive (+) | 0 | 0 | 3* | 4 |
| Negative (-) | 40 | 3 | 1 | 0 | |
| 2nd Lot | Positive (+) | 0 | 0 | 3* | 4 |
| Negative (-) | 40 | 3 | 1 | 0 | |
| 3rd Lot | Positive (+) | 0 | 0 | 3* | 4 |
| Negative (-) | 40 | 3 | 1 | 0 |
*Summary of concentrations for discordant samples (50-100% of cutoff)
| Fentanyl Test (nFEN) Result | Norfentanyl by LC-MS/MS (ng/mL) |
|---|---|
| Positive (+) | 4.76 |
| Positive (+) | 3.86 |
| Positive (+) | 3.34 |
The same 3 clinical samples produced positive results with all 3 lots tested.
2. Matrix Comparison:
Not applicable. These devices are for use with urine samples only.
C. Clinical Studies:
1. Clinical Sensitivity:
Not applicable.
2. Clinical Specificity:
Not applicable.
D. Clinical Cut-Off:
Not applicable.
E. Expected Values/Reference Range:
Not applicable.
Page 11 of 12
Page 16
K252684 510(k) SUMMARY
Conclusions
The Labcorp Fentanyl Urine Visual Test (the "DEVICE") is as safe, as effective, and performs as well as or better than the legally marketed predicate device. The DEVICE has the same intended use, technological characteristics, and principles of operation as its predicate device. The DEVICE indications for use fall within the use population of the predicate device, and the DEVICE format is a cassette design with a dropper for sample transfer while the predicate device is a dip card or cup format. Comparison of similarities and differences between the DEVICE and the predicate device raises no new issues of safety or effectiveness. Thus, the DEVICE is substantially equivalent to its predicate device.
The submitted information in this premarket notification is complete and supports a substantial equivalence decision.
References:
-
Disposition of Toxic Drugs and Chemicals in Man, 10th ed., R.C. Baselt. Biomedical Publications, Seal Beach (USA) (2014), pp. 846–849, ISBN: 978-0-9626523-9-4.
-
Silverstein JH, Rieders MF, McMullin M, Schulman S, Zahl K. An analysis of the duration of fentanyl and its metabolites in urine and saliva. Anesth Analg. 1993 Mar;76(3):618-21. DOI: 10.1213/00000539-199303000-00030. PMID: 8452277.
Page 12 of 12
§ 862.3650 Opiate test system.
(a)
Identification. An opiate test system is a device intended to measure any of the addictive narcotic pain-relieving opiate drugs in blood, serum, urine, gastric contents, and saliva. An opiate is any natural or synthetic drug that has morphine-like pharmocological actions. The opiates include drugs such as morphine, morphine glucoronide, heroin, codeine, nalorphine, and meperedine. Measurements obtained by this device are used in the diagnosis and treatment of opiate use or overdose and in monitoring the levels of opiate administration to ensure appropriate therapy.(b)
Classification. Class II (special controls). An opiate test system is not exempt if it is intended for any use other than employment or insurance testing or is intended for Federal drug testing programs. The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 862.9, provided the test system is intended for employment and insurance testing and includes a statement in the labeling that the device is intended solely for use in employment and insurance testing, and does not include devices intended for Federal drug testing programs (e.g., programs run by the Substance Abuse and Mental Health Services Administration (SAMHSA), the Department of Transportation (DOT), and the U.S. military).