K Number
K252520

Validate with FDA (Live)

Date Cleared
2025-09-11

(31 days)

Product Code
Regulation Number
862.3650
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Immunalysis SEFRIA™ Hydrocodone Oral Fluid Enzyme Immunoassay is an enzyme immunoassay with a cutoff of 30 ng/mL in neat oral fluid collected by Quantisal™ or Quantisal™ II Oral Fluid Collection Device. The assay is intended for the qualitative and semi-quantitative analysis of hydrocodone in human oral fluid to be used with clinical analyzers. This assay is calibrated against hydrocodone.

The semi-quantitative mode is for purposes of enabling laboratories to determine an appropriate dilution of the specimen for confirmation by a confirmatory method such as Gas Chromatography/Mass Spectrometry (GC-MS) or Liquid Chromatography/Tandem Mass Spectrometry (LC-MS/MS) or permitting laboratories to establish quality control procedures.

The Immunalysis SEFRIA™ Hydrocodone Oral Fluid Enzyme Immunoassay provides only a preliminary analytical test result. A more specific alternate chemical method must be used in order to obtain a confirmed analytical result. GC-MS or LC-MS/MS are the preferred confirmatory methods. Clinical consideration and professional judgment should be applied to any test result, particularly when preliminary positive results are used.

Device Description

The Immunalysis SEFRIA™ Hydrocodone Oral Fluid Enzyme Immunoassay is a homogenous enzyme immunoassay with a cutoff of 30 ng/mL in neat oral fluid. The assay is intended for use in laboratories with clinical chemistry analyzers for the qualitative and semi-quantitative analysis of hydrocodone in human oral fluid collected with Quantisal™ Oral Fluid Collection Device or Quantisal™ II Oral Fluid collection device.

AI/ML Overview

N/A

FDA 510(k) Clearance Letter - SEFRIA™ Hydrocodone Oral Fluid

Page 1

U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov

Doc ID # 04017.08.01

September 11, 2025

Immunalysis Corporation
Shubhajit Mitra
Senior Manager, Regulatory Affairs
829 Towne Center Drive
Pomona, California 91767

Re: K252520
Trade/Device Name: SEFRIA™ Hydrocodone Oral Fluid
Regulation Number: 21 CFR 862.3650
Regulation Name: Opiate test system
Regulatory Class: Class II
Product Code: DJG
Dated: August 8, 2025
Received: August 11, 2025

Dear Shubhajit Mitra:

We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

Page 2

K252520 - Shubhajit Mitra Page 2

Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).

Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801 and Part 809); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.

All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-

Page 3

K252520 - Shubhajit Mitra Page 3

assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

JOSEPH A. KOTAREK -S
Digitally signed by JOSEPH A. KOTAREK -S
Date: 2025.09.11 07:46:56 -04'00'

Joseph Kotarek
Branch Chief for Toxicology
Division of Chemistry and Toxicology Devices
OHT7: Office of In Vitro Diagnostics
Office of Product Evaluation and Quality
Center for Devices and Radiological Health

Enclosure

Page 4

FORM FDA 3881 (8/23) Page 1 of 1

DEPARTMENT OF HEALTH AND HUMAN SERVICES
Food and Drug Administration

Indications for Use

Form Approved: OMB No. 0910-0120
Expiration Date: 07/31/2026
See PRA Statement below.

510(k) Number (if known): K252520

Device Name: SEFRIA™ Hydrocodone Oral Fluid

Indications for Use (Describe)

The Immunalysis SEFRIA™ Hydrocodone Oral Fluid Enzyme Immunoassay is an enzyme immunoassay with a cutoff of 30 ng/mL in neat oral fluid collected by Quantisal™ or Quantisal™ II Oral Fluid Collection Device. The assay is intended for the qualitative and semi-quantitative analysis of hydrocodone in human oral fluid to be used with clinical analyzers. This assay is calibrated against hydrocodone.

The semi-quantitative mode is for purposes of enabling laboratories to determine an appropriate dilution of the specimen for confirmation by a confirmatory method such as Gas Chromatography/Mass Spectrometry (GC-MS) or Liquid Chromatography/Tandem Mass Spectrometry (LC-MS/MS) or permitting laboratories to establish quality control procedures.

The Immunalysis SEFRIA™ Hydrocodone Oral Fluid Enzyme Immunoassay provides only a preliminary analytical test result. A more specific alternate chemical method must be used in order to obtain a confirmed analytical result. GC-MS or LC-MS/MS are the preferred confirmatory methods. Clinical consideration and professional judgment should be applied to any test result, particularly when preliminary positive results are used.

Type of Use (Select one or both, as applicable)
☒ Prescription Use (Part 21 CFR 801 Subpart D)
☐ Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services
Food and Drug Administration
Office of Chief Information Officer
Paperwork Reduction Act (PRA) Staff
PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

Page 5

510(k) Summary Page 1 of 19

SEFRIA™ Hydrocodone Oral Fluid Enzyme Immunoassay Premarket Notification

510(k) Summary

This 510(k) summary of safety and effectiveness information is being submitted in accordance with 21 CFR Section 807.92.

Submitter

510(k) Number: K252520

Applicant Name: Immunalysis Corporation
829 Towne Center Drive
Pomona, CA 91767

FDA Establishment #: 2020952

Primary Correspondent: Shubhajit Mitra
Senior Manager, Regulatory Affairs

Primary Phone: 508-330-4796

Primary Email: shubhajit.mitra@abbott.com

Alternate Correspondent: Iris Saliba
Director, Regulatory Affairs

Secondary Phone: 619-540-3931

Alternate Email: iris.saliba@abbott.com

Date Prepared: Sep 08, 2025

Page 6

510(k) Summary Page 2 of 19

1. Device Information

Trade or Proprietary Names: SEFRIA™ Hydrocodone Oral Fluid

Common Name: Immunalysis SEFRIA™ Hydrocodone Oral Fluid Enzyme Immunoassay

Device Classification Name: Enzyme Immunoassay, Opiates

Product Codes: DJG

Regulatory Class: Class II

Classification Regulation: 21 CFR 862.3650

Panel: Toxicology (91)

2. Predicate Information

Company: Immunalysis Corporation

Device: SEFRIA™ Oxycodone Oral Fluid Enzyme Immunoassay (K203564)

3. Device Description

The Immunalysis SEFRIA™ Hydrocodone Oral Fluid Enzyme Immunoassay is a homogenous enzyme immunoassay with a cutoff of 30 ng/mL in neat oral fluid. The assay is intended for use in laboratories with clinical chemistry analyzers for the qualitative and semi-quantitative analysis of hydrocodone in human oral fluid collected with Quantisal™ Oral Fluid Collection Device or Quantisal™ II Oral Fluid collection device.

4. Indication for Use

The Immunalysis SEFRIA™ Hydrocodone Oral Fluid Enzyme Immunoassay is an enzyme immunoassay with a cutoff of 30 ng/mL in neat oral fluid collected by Quantisal™ or Quantisal™ II Oral Fluid Collection Device. The assay is intended for the qualitative and semi-quantitative analysis of hydrocodone in human oral fluid to be used with clinical analyzers. This assay is calibrated against hydrocodone.

Page 7

510(k) Summary Page 3 of 19

The semi-quantitative mode is for purposes of enabling laboratories to determine an appropriate dilution of the specimen for confirmation by a confirmatory method such as Gas Chromatography/Mass Spectrometry (GC-MS) or Liquid Chromatography/Tandem Mass Spectrometry (LC-MS/MS) or permitting laboratories to establish quality control procedures.

The Immunalysis SEFRIA™ Hydrocodone Oral Fluid Enzyme Immunoassay provides only a preliminary analytical test result. A more specific alternate chemical method must be used in order to obtain a confirmed analytical result. Gas Chromatography/ Mass Spectrometry (GC-MS) or Liquid Chromatography- Tandem Mass Spectrometry (LC-MS/MS) is the preferred confirmatory method. Professional judgment should be applied to any test result, particularly when preliminary positive results are used.

5. Comparison to Predicate Device

The subject device has the same design and functionality as the predicate device.

Table 1: Device Comparison

Device CharacteristicsPredicate Device SEFRIA™ Oxycodone Oral Fluid Enzyme Immunoassay (K203564)Candidate device SEFRIA™ Hydrocodone Oral Fluid Enzyme Immunoassay
Similarities
Assay MaterialsReady to use reagents: Antibody reagent, drug conjugate reagentSame
Test PrincipleHomogenous, competitiveSame
Measurement modeQualitative and semi-quantitativeSame
Cutoff Level30 ng/mL in neat oral fluidSame
Sample matrixOral fluidSame
Reagent Storage2 - 8°C until expiration dateSame
InstrumentationAutomated clinical chemistry analyzerSame

Page 8

510(k) Summary Page 4 of 19

Device CharacteristicsPredicate Device SEFRIA™ Oxycodone Oral Fluid Enzyme Immunoassay (K203564)Candidate device SEFRIA™ Hydrocodone Oral Fluid Enzyme Immunoassay
Mass Spectrometry ConfirmationRequired for preliminary positive resultsSame
Associated oral fluid collection deviceQuantisal™ or Quantisal™ II Oral Fluid Collection DeviceSame
Differences
Intended UseQualitative and semi-quantitative analysis of oxycodone in human oral fluid collected by Quantisal™ or Quantisal™ II Oral Fluid Collection DeviceQualitative and semi-quantitative analysis of hydrocodone in human oral fluid collected by Quantisal™ or Quantisal™ II Oral Fluid Collection Device
Target analyteOxycodoneHydrocodone

6. Performance Characteristics

The following laboratory performance studies were performed to determine substantial equivalence of the SEFRIA™ Hydrocodone Oral Fluid Enzyme Immunoassay to the predicate device. Assay performance was established using the Beckman Coulter AU5800 chemistry analyzer.

6.1 Precision

A precision study was performed over 15 days, 2 runs per day with 2 collection devices per run (N=60), one replicate per collection device on 1 lot of reagent and 3 lots of Quantisal™ Oral Fluid Collection Device. Drug free negative oral fluid was spiked to concentrations of assay cutoff and ±25%, ±50%, ±75%, ±100% of the cutoff and was collected using the collection devices. The spiked concentrations were confirmed by mass spectrometry (LC-MS/MS) before collection. The study established the repeatability of the testing system, including assay and oral fluid collection device. Test results in qualitative and semi-quantitative modes are presented in Table 2 and Table 3.

Page 9

510(k) Summary Page 5 of 19

An additional 20-day study was performed on 3 lots of assay reagent to demonstrate the repeatability across multiple reagent lots.

Table 2 – 15 Days Precision -Qualitative- Quantisal™ – All 3 lots

Concentration (ng/mL)% of cutoff# of determinationsResults
Quantisal™ Lot 1
0.0-100%60Negative/100%
7.5-75%60Negative/100%
15-50%60Negative/100%
22.5-25%60Negative/100%
30Cutoff6027 Positive/33 Negative
37.5+25%60Positive/100%
45+50%60Positive/100%
52.5+75%60Positive/100%
60+100%60Positive/100%
Concentration (ng/mL)% of cutoff# of determinationsResults
Quantisal™ Lot 2
0.0-100%60Negative/100%
7.5-75%60Negative/100%
15-50%60Negative/100%
22.5-25%60Negative/100%
30Cutoff6033 positive/27 Negative
37.5+25%60Positive/100%
45+50%60Positive/100%
52.5+75%60Positive/100%
60+100%60Positive/100%
Concentration (ng/mL)% of cutoff# of determinationsResults
Quantisal™ Lot 3

Page 10

510(k) Summary Page 6 of 19

Concentration (ng/mL)% of cutoff# of determinationsResults
0.0-100%60Negative/100%
7.5-75%60Negative/100%
15-50%60Negative/100%
22.5-25%60Negative/100%
30Cutoff6027 Positive/33 Negative
37.5+25%60Positive/100%
45+50%60Positive/100%
52.5+75%60Positive/100%
60+100%60Positive/100%

Table 3 - 15 Days Precision -Semi-Quantitative- Quantisal™ – All 3 lots

Concentration (ng/mL)% of cutoff# of determinationsResult
Quantisal™ Lot 1
0.0-100%60Negative/100%
7.5-75%60Negative/100%
15-50%60Negative/100%
22.5-25%60Negative/100%
30Cutoff6043 Positive/ 17 Negative
37.5+25%60Positive/100%
45+50%60Positive/100%
52.5+75%60Positive/100%
60+100%60Positive/100%
Concentration (ng/mL)% of cutoff# of determinationsResults
Quantisal™ Lot 2
0.0-100%60Negative/100%
7.5-75%60Negative/100%

Page 11

510(k) Summary Page 7 of 19

Concentration (ng/mL)% of cutoff# of determinationsResult
15-50%60Negative/100%
22.5-25%60Negative/100%
30Cutoff6050 positive/10 Negative
37.5+25%60Positive/100%
45+50%60Positive/100%
52.5+75%60Positive/100%
60+100%60Positive/100%
Concentration (ng/mL)% of cutoff# of determinationsResults
Quantisal™ Lot 3
0.0-100%60Negative/100%
7.5-75%60Negative/100%
15-50%60Negative/100%
22.5-25%60Negative/100%
30Cutoff6052 Positive/8 Negative
37.5+25%60Positive/100%
45+50%60Positive/100%
52.5+75%60Positive/100%
60+100%60Positive/100%

6.2 Specificity and Cross Reactivity

Structurally and functionally similar compounds were spiked into drug free pooled oral fluid at levels that will yield a result that is equivalent to the cutoff, if cross reacting. The study verified the cross reactivity of the hydrocodone assay to related drugs and drug metabolites, in both the qualitative and semi-quantitative modes. Cross-reactivity test results in qualitative mode and semi-quantitative mode are presented in Table 4 below.

Page 12

510(k) Summary Page 8 of 19

Table 4 – Cross Reactivity – Qualitative and Semi-Quantitative

CompoundAnalyte Conc. (ng/mL)Hydrocodone Conc. (ng/mL)Hydrocodone Semi-Quantitative Mean Value (ng/mL)Hydrocodone Qualitative ResultCross Reactivity (%)
6-Acetylcodeine40,0003015.9Negative<0.1%
6-Acetylmorphine40,0003018.3Negative<0.1%
Buprenorphine40,000304.2Negative<0.1%
Dextromethorphan40,0003010.7Negative<0.1%
Fentanyl40,000305.2Negative<0.1%
Heroin40,0003011.0Negative<0.1%
(+) Methadone40,000305.4Negative<0.1%
Morphine-6β-D-Glucuronide40,000304.6Negative<0.1%
Norbuprenorphine40,000305.6Negative<0.1%
Norcodeine40,000308.6Negative<0.1%
Normorphine40,000307.5Negative<0.1%
Tapentadol40,000304.8Negative<0.1%
Meperidine40,000305.4Negative<0.1%
Tramadol40,000305.3Negative<0.1%
Codeine30,0003030.4Positive0.1%
Levorphanol4,0003030.3Positive0.8%
Noroxycodone16,5003030.1Positive0.2%
Noroxymorphone19,0003030.8Positive0.2%
Oxycodone2,0003030.6Positive1.5%
Dihydrocodeine1,8003031.2Positive1.7%
Morphine32,0003030.9Positive0.1%
Desomorphine2253032.9Positive13.3%
Morphine-3β-D-Glucuronide19,0003031.0Positive0.2%
Nalorphine39,0003031.5Positive0.1%
Norhydrocodone1,9003031.3Positive1.6%
Oxymorphone2,0003030.8Positive1.5%
Hydromorphone343031.9Positive88.2%
Hydromorphone-3β-D-Glucuronide203031.4Positive150.0%
Naloxone2,0003030.4Positive1.5%

Page 13

510(k) Summary Page 9 of 19

Table 4 – Cross Reactivity – Qualitative and Semi-Quantitative

CompoundAnalyte Conc. (ng/mL)Hydrocodone Conc. (ng/mL)Hydrocodone Semi-Quantitative Mean Value (ng/mL)Hydrocodone Qualitative ResultCross Reactivity (%)
6-Acetylcodeine40,0003015.9Negative<0.1%
6-Acetylmorphine40,0003018.3Negative<0.1%
Buprenorphine40,000304.2Negative<0.1%
Dextromethorphan40,0003010.7Negative<0.1%
Fentanyl40,000305.2Negative<0.1%
Heroin40,0003011.0Negative<0.1%
(+) Methadone40,000305.4Negative<0.1%
Morphine-6β-D-Glucuronide40,000304.6Negative<0.1%
Norbuprenorphine40,000305.6Negative<0.1%
Norcodeine40,000308.6Negative<0.1%
Normorphine40,000307.5Negative<0.1%
Tapentadol40,000304.8Negative<0.1%
Meperidine40,000305.4Negative<0.1%
Tramadol40,000305.3Negative<0.1%
Codeine30,0003030.4Positive0.1%
Levorphanol4,0003030.3Positive0.8%
Noroxycodone16,5003030.1Positive0.2%
Noroxymorphone19,0003030.8Positive0.2%
Oxycodone2,0003030.6Positive1.5%
Dihydrocodeine1,8003031.2Positive1.7%
Morphine32,0003030.9Positive0.1%
Desomorphine2253032.9Positive13.3%
Morphine-3β-D-Glucuronide19,0003031.0Positive0.2%
Nalorphine39,0003031.5Positive0.1%
Norhydrocodone1,9003031.3Positive1.6%
Oxymorphone2,0003030.8Positive1.5%
Hydromorphone343031.9Positive88.2%
Hydromorphone-3β-D-Glucuronide203031.4Positive150.0%
Naloxone2,0003030.4Positive1.5%
Naltrexone13,0003031.1Positive0.2%
Oxymorphone-3β-D-Glucuronide1,5003031.0Positive2.0%

6.3 Interference – Structurally Unrelated Compounds

Structurally unrelated compounds were evaluated in qualitative and semi-quantitative modes by spiking the potential interferent into drug free oral fluid containing hydrocodone at ±25% of the cutoff. At the levels tested, there was no interference with structurally unrelated compounds. The concentration levels of structurally unrelated compounds are presented in Table 5.

Table 5 - Non-Interfering Structurally Unrelated Compounds

CompoundConcentration Tested (ng/mL)
7-Aminoclonazepam5,000
7-Aminoflunitrazepam40,000
7-Aminonitrazepam40,000
α-Hydroxyalprazolam40,000
Alprazolam40,000
Amitriptyline40,000
Amobarbital40,000
(+)S-Amphetamine40,000
Atomoxetine40,000
Barbital40,000
Benzoylecgonine40,000
Benzylpiperazine40,000
Bromazepam40,000
Brompheniramine40,000
Buprenorphine40,000
Bupropion40,000
Butabarbital40,000

Page 14

510(k) Summary Page 10 of 19

CompoundConcentration Tested (ng/mL)
Butalbital40,000
Cannabidiol40,000
Cannabinol40,000
Carbamazepine40,000
Carisoprodol40,000
Cetirizine40,000
Chlordiazepoxide40,000
1-(3-Chlorophenyl)piperazine (mCPP)40,000
Chlorpromazine40,000
Chlorpheniramine40,000
Cimetidine40,000
cis-Tramadol40,000
Citalopram40,000
Clobazam40,000
Clomipramine40,000
Clonazepam40,000
Clozapine20,000
Cocaine40,000
Cyclobenzaprine40,000
Dehydronorketamine10,000
Demoxepam40,000
Desalkylflurazepam40,000
Desipramine40,000
Diazepam40,000
Digoxin40,000
(±)-10,11-Dihydro-10-Hydroxycarbamazepine40,000
Diphenhydramine40,000
Doxepin40,000
Doxylamine40,000
Duloxetine40,000

Page 15

510(k) Summary Page 11 of 19

CompoundConcentration Tested (ng/mL)
Ecgonine40,000
Ecgonine Methyl Ester40,000
EDDP40,000
EMDP40,000
1R,2S(-)-Ephedrine40,000
1S,2R(+)-Ephedrine40,000
Ethotoin40,000
Ethylmorphine5,000
Ethyl-β-D-Glucuronide40,000
Fenfluramine40,000
Flunitrazepam40,000
Fluoxetine40,000
Flurazepam40,000
Haloperidol40,000
Hexobarbital40,000
Gabapentin40,000
Imipramine40,000
Ketamine40,000
Lamotrigine40,000
Labetalol40,000
Lidocaine40,000
Lorazepam40,000
Lorazepam Glucuronide15,000
Loratadine40,000
Lormetazepam40,000
LSD40,000
Maprotiline40,000
(±)MDA40,000
(±)MDEA40,000
(±)MDMA40,000

Page 16

510(k) Summary Page 12 of 19

CompoundConcentration Tested (ng/mL)
Meprobamate40,000
(+)S-Methamphetamine40,000
Methaqualone40,000
Methoxetamine40,000
Methylone40,000
Methylphenidate40,000
α-Methyl-α-phenylsuccinimide40,000
Mirtazapine40,000
Naproxen40,000
n-desmethylcitalopram40,000
n-desmethyltapentadol40,000
Nitrazepam40,000
Nordiazepam40,000
(±)Norketamine40,000
Normesuximide40,000
(±)Norpropoxyphene40,000
Nortriptyline40,000
Olanzapine40,000
Oxazepam40,000
PCP40,000
Pentazocine40,000
Pentobarbital40,000
Phenazepam40,000
Phenobarbital40,000
Phentermine40,000
(±)Phenylpropanolamine (PPA)40,000
Phenytoin40,000
PMA40,000
PMMA40,000
Prazepam40,000

Page 17

510(k) Summary Page 13 of 19

CompoundConcentration Tested (ng/mL)
Procaine40,000
(±)Propoxyphene35,000
Propranolol40,000
Protriptyline40,000
R(-)-Phenylephrine40,000
R,R(-)-Pseudoephedrine40,000
Risperidone40,000
Ritalinic Acid40,000
S,S(+)-Pseudoephedrine40,000
Salicylic Acid40,000
Secobarbital40,000
Sertraline40,000
Sufentanil40,000
Temazepam40,000
(-)11-nor-9-carboxy-Δ9-THC40,000
(-)-Δ9-THC40,000
(±)-11-hydroxy-Δ9-THC40,000
Theophylline40,000
Thioridazine10,000
Trazodone40,000
Triazolam40,000
3-Trifluoromethylphenylpiperazine40,000
Trimipramine40,000
Tyramine40,000
Venlafaxine40,000
Verapamil40,000
Zolpidem40,000
2-CB(4-bromo-2,5-dimethoxyphenyl-benzeneethamine)40,000
Midazolam40,000
Oxazepam Glucuronide35,000

Page 18

510(k) Summary Page 14 of 19

CompoundConcentration Tested (ng/mL)
Norpseudoephedrine7500
O-desmethylvenlafaxine40,000
N-desmethylvenlafaxine40,000
N-desmethyltramadol40,000
O-desmethylcistramadol40,000

6.4 Interference – Endogenous Compounds and Exogenous Compounds

Endogenous and commonly ingested exogenous compounds were assessed using qualitative and semi-quantitative methods by spiking potential interferents into drug-free oral fluid containing hydrocodone at ±25% of the cutoff concentration. Additionally, oral products were evaluated by collecting oral fluid samples from volunteers using the Quantisal™ Oral Fluid Collection Device after product use. No interference was observed from endogenous substances, exogenous compounds, or orally administered products at the tested levels. Endogenous compounds and exogenous compounds are presented in Table 6, Table 7 and Table 8.

Table 6 – Non-Interfering Endogenous Compounds

CompoundConcentration Tested
Ascorbic Acid3 mg/mL
Bilirubin0.15 mg/mL
Cholesterol0.45 mg/mL
γ-globulin0.8 mg/mL
Hemoglobin3 mg/mL
Human Serum Albumin (HSA)15 mg/mL
IgA1 mg/mL
IgG1 mg/mL
IgM0.5 mg/mL
Salivary-α-amylase1000 U/mL

Page 19

510(k) Summary Page 15 of 19

Table 7 – Non-Interfering Exogenous Compounds (Commonly Ingested)

CompoundConcentration Tested
Acetaminophen0.05 mg/mL
Acetylsalicylic Acid0.1 mg/mL
Baking Soda0.6% v/v
Cotinine0.03 mg/mL
Denture Adhesive0.6% w/v
Ibuprofen0.05 mg/mL
Alcohol (Ethanol)6% v/v
Caffeine0.05 mg/mL
Coffee6% v/v
Cranberry Juice6% v/v
Milk2% v/v
Mouthwash6% v/v
Orange Juice6% v/v
Soft Drink (Pepsi)6% v/v
Sodium Chloride18 mg/mL
Tea6% v/v
Toothpaste6% w/v
Sugar30 mg/mL

Page 20

510(k) Summary Page 16 of 19

Table 8 – Non-Interfering Orally Used Exogenous Products

ProductQuantity Ingested
Teeth Whitener2 strips
Cigarette1 cigarette
Hard Candy1 piece
Chewing Gum1 piece
Sugar2 teaspoon
Cough Syrup2 teaspoon
Milk100 mL
Orange Juice100 mL
Hydrogen Peroxide (3% OTC)10 mL
Ibuprofen200 mg
Acetaminophen1000 mg

6.5 Interference – pH

To evaluate potential interference from the effect of oral fluid pH, device performance in the qualitative and semi-quantitative modes was tested using a range of oral fluid pH values (3.0, 4.0, 5.0, 6.0, 7.0, 8.0, 9.0, 10.0 and 11.0). All test samples were prepared in drug free oral fluid containing hydrocodone at ±25% of the cutoff. At the pH levels tested, there was no interference observed for each test mode.

6.6 Linearity/Recovery

Linearity of the SEFRIA™ Hydrocodone Oral Fluid Immunoassay in semi-quantitative mode was assessed by spiking pooled drug-free oral fluid with a high concentration of hydrocodone. Serial dilutions were performed using drug-free oral fluid to generate specimens with target concentrations ranging from 10 ng/mL to 110 ng/mL. A 0 ng/mL sample was prepared using unspiked drug-free oral fluid. Each concentration level was collected using Quantisal™ Oral Fluid Collection Devices and tested in five replicates. Mean values from replicate testing were used to

Page 21

510(k) Summary Page 17 of 19

calculate recovery. The assay demonstrated acceptable linearity and recovery across the tested concentration range. Results are summarized in Table 9. The linear range of the assay was confirmed to be 10 – 100 ng/mL using Quantisal™ Oral Fluid Collection Device, and the drug recovery was within the ±15% of the expected value.

Table 9 - Linearity/Recovery – Quantisal

Expected Concentration (ng/mL)Mean Concentration (ng/mL)Recovery (%)
04.6N/A
1011.3113.2
2019.798.6
3027.591.5
4034.486.1
5045.390.7
6052.988.2
7068.798.1
8077.797.1
9099.6110.6
10090.990.9
110103.694.2

6.7 Hydrocodone Stability in Oral Fluid

The stability of hydrocodone in the oral fluid specimens collected with the Quantisal™ and Quantisal™ II Oral Fluid Collection Device was evaluated with low positive samples (+50% cutoff) at room temperature (8-25°C) and refrigerated (2-8°C). Sample stability testing was performed using LCMS/MS or GC/MS at multiple timepoints post collection at 8°C - 25°C and at 2°C - 8°C. The stability study results demonstrate that oral fluid samples containing hydrocodone are stable at 2°C - 8°C for up to 12 months and 10 days at ambient temperature 8°C - 25°C when stored in

Page 22

510(k) Summary Page 18 of 19

Quantisal™ or Quantisal™ II Oral Fluid Collection Device. The data to support hydrocodone sample stability was submitted and cleared in K223781 (Quantisal™ II) and K232898 (Quantisal™).

6.8 Calibration Duration

To evaluate calibration stability, drug-free oral fluid samples were spiked with hydrocodone at ±25% of the cutoff concentration and tested in both qualitative and semi-quantitative modes over a 15-day period. At the initial time point, a two-point calibration curve was established in qualitative mode and five-point calibration curve was established in semi-quantitative mode. These calibrations were used through the duration of the study. The test results met acceptance criteria at each time point up to 15 days, providing a validated margin beyond the claimed duration without extrapolation. The calibration duration stability study confirmed that the SEFRIA™ Hydrocodone Oral Fluid Immunoassay supports a calibration interval of up to 14 days.

6.9 Method Comparison

A method comparison study was conducted to evaluate the analytical performance of the SEFRIA™ Hydrocodone Oral Fluid Enzyme Immunoassay using the Beckman Coulter AU5800 analyzer, compared to the reference method, Liquid Chromatography-Tandem Mass Spectrometry (LC-MS/MS). A total of eighty (80) deidentified, unaltered clinical oral fluid samples (40 positive, 40 negative) were tested. Qualitative concordance was 100% for positive samples and 92.5% for negative samples. Semi-quantitative concordance was 97.5% for positives and 95% for negatives. Discordant results were limited and primarily attributed to cross-reactivity or concentrations near the assay cutoff.

Table 10 – Method Comparison

SEFRIA™ Hydrocodone Oral Fluid Enzyme ImmunoassayLC-MS/MS Hydrocodone ConcentrationAgreement (%)
< 15 ng/mL (Less than -50% cutoff)15 to <30 ng/mL (Between -50% cutoff and cutoff)

Page 23

510(k) Summary Page 19 of 19

-50% cutoff)(Between -50% cutoff and cutoff)(Between cutoff and +50% cutoff)than +50% cutoff)
Qualitative
Positive21436100.0
Negative3430092.5
Semi-Quantitative
Positive2033697.5
Negative3441095.0

Table 11 – SEFRIA™ Hydrocodone Method Comparison Discordant Results

Discordant Result Summary at 30 ng/mL Cutoff

Sample IDConcentration (LCMS)LCMS Qualitative ResultSEFRIA™ Qualitative Result (POS/NEG)SEFRIA™ Semi-Quantitative Result (POS/NEG)
HYC-031910.3962NEGPOSPOS
HYC-03240NEGPOSPOS
HYC-032721.996NEGPOSNEG
HYC-035530.4135NEGPOSNEG

7. Conclusion

The information provided in this pre-market notification demonstrates that the SEFRIA™ Hydrocodone Oral Fluid Enzyme Immunoassay is substantially equivalent to the legally marketed predicate device.

§ 862.3650 Opiate test system.

(a)
Identification. An opiate test system is a device intended to measure any of the addictive narcotic pain-relieving opiate drugs in blood, serum, urine, gastric contents, and saliva. An opiate is any natural or synthetic drug that has morphine-like pharmocological actions. The opiates include drugs such as morphine, morphine glucoronide, heroin, codeine, nalorphine, and meperedine. Measurements obtained by this device are used in the diagnosis and treatment of opiate use or overdose and in monitoring the levels of opiate administration to ensure appropriate therapy.(b)
Classification. Class II (special controls). An opiate test system is not exempt if it is intended for any use other than employment or insurance testing or is intended for Federal drug testing programs. The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 862.9, provided the test system is intended for employment and insurance testing and includes a statement in the labeling that the device is intended solely for use in employment and insurance testing, and does not include devices intended for Federal drug testing programs (e.g., programs run by the Substance Abuse and Mental Health Services Administration (SAMHSA), the Department of Transportation (DOT), and the U.S. military).