Search Filters

Search Results

Found 126 results

510(k) Data Aggregation

    K Number
    K253565

    Validate with FDA (Live)

    Date Cleared
    2026-03-04

    (107 days)

    Product Code
    Regulation Number
    872.5470
    Panel
    Dental
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The SureSmile Software is indicated for use as a front-end device for management of orthodontic models, systematic inspection, detailed analysis, treatment planning and simulation and the design of custom appliance systems, which may include sequential aligners, retainers, IDB trays, brackets, or wires, and dental casts. These applications are based on 3D models of the patient's dentition before the start of the orthodontic treatment. It can also be applied during the treatment to inspect and analyze the progress of the treatment. It can be used at the end of the treatment to evaluate if the outcome is consistent with the planned/desired treatment objectives.

    The use of the SureSmile Software requires the user to have the necessary training and domain knowledge in the practice of orthodontics, as well as to have received a dedicated training in the use of the software.

    Device Description

    The SureSmile® Software is suitable for use for patients requiring orthodontic treatment; it allows the user to plan therapy and then order patient matched orthodontic appliances. The software includes many features to support diagnosis, planning, appliance design and treatment tracking. Additionally, SureSmile Software acts as the order management system for all SureSmile Digital Lab services and appliance production. SureSmile Software contains features to support ordering and shipment tracking. The software does not have any contact with the body.

    AI/ML Overview

    N/A

    Ask a Question

    Ask a specific question about this device

    K Number
    K250203

    Validate with FDA (Live)

    Device Name
    SureAx-Guide™
    Date Cleared
    2025-10-17

    (266 days)

    Product Code
    Regulation Number
    870.1330
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The SureAx-Guide guidewire is indicated for use in facilitating delivery of percutaneous catheters into the peripheral vasculature. The SureAx-Guide guidewire is not intended for use in the coronary or neuro vasculature.

    Device Description

    The SureAx-Guide™ guidewire is comprised of a stainless-steel wire with a PTFE coating and distal coils and soldered tip. The distal end has a urethane sheath and PVP-based hydrophilic coating. The maximum outside diameter is 0.0142 inches and it will be available in 135cm and 190cm lengths. It will be supplied sterile and intended for single use.

    AI/ML Overview

    N/A

    Ask a Question

    Ask a specific question about this device

    K Number
    K250510

    Validate with FDA (Live)

    Date Cleared
    2025-10-15

    (236 days)

    Product Code
    Regulation Number
    880.5860
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Sure-Fine Insulin Syringes are hypodermic insulin syringes for subcutaneous injection of U100 insulin.

    Device Description

    Sure-Fine Insulin syringes are designed for the subcutaneous injection of a desired dose of insulin. The syringe has a graduated barrel(U-100), a plunger rod, needle cap, protective end cap and needle permanently affixed to the tip of the syringe with epoxy.

    The syringes are available in the following size and cap color.

    CategoryInsulin SyringeNeedle gaugeNeedle lengthCap Color
    Needle Cap
    U-1001 cc27 Gauge1/2″Orange
    1 cc27 Gauge5/16″
    1/2cc and 1 cc28 Gauge1/2″
    1/2cc and 1 cc28 Gauge5/16″
    3/10cc, 1/2cc and 1 cc29 Gauge1/2″
    3/10cc, 1/2cc and 1 cc29 Gauge5/16″
    3/10cc, 1/2cc and 1 cc30 Gauge1/2″
    3/10cc, 1/2cc and 1 cc30 Gauge5/16″
    3/10cc and 1/2cc30 Gauge1/4″
    3/10cc, 1/2cc and 1 cc31 Gauge5/16″
    3/10cc and 1/2cc31 Gauge1/4″
    AI/ML Overview

    N/A

    Ask a Question

    Ask a specific question about this device

    K Number
    K251104

    Validate with FDA (Live)

    Date Cleared
    2025-09-29

    (171 days)

    Product Code
    Regulation Number
    876.4300
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    This device is used for cannulation of the ductal system and for sphincterotomy. If preloaded, also aids in bridging difficult strictures during ERCP.

    Device Description

    The SureTome™ SW Sphincterotome with DomeTip ® are sterile, single use devices compatible with the accessory channel of endoscope. The device consists of a long, thin plastic tube (cannula) with a wire running the length of its interior. A small portion of that wire is exposed at its distal end. The roof of the papilla is opened by passing high frequency current through the wire, exposing the biliary or pancreatic orifices for selective cannulation.

    AI/ML Overview

    N/A

    Ask a Question

    Ask a specific question about this device

    K Number
    K250658

    Validate with FDA (Live)

    Date Cleared
    2025-04-03

    (29 days)

    Product Code
    Regulation Number
    880.5570
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    SureFine Pen Needle is intended for use with a pen injector for the subcutaneous injection of insulin.

    Device Description

    SureFine Pen Needle is designed for use with a pen injector for the subcutaneous injection of insulin. The pen needle consists of a needle, hub, and shield assembly. Blister paper covers primary container. The primary container maintains sterility of the needle because primary container covers the hub and needle cap with blister paper sealed on the opening hole of primary container.

    The needle hub can be connected screwed onto the pen. The needle shield is intended to provide physical protection to the needle tube.

    SureFine Pen Needle is offered in various gauges sizes (29G, 30G, 31G, 32G) and Lengths (4mm, 6mm, 8mm, 12.7mm). These are sterile (EO gas sterilization), non-toxic, and non-pyrogenic. The pen needles are disposable, single use devices.

    AI/ML Overview

    The SureFine Pen Needle device, K250658, is a Class II hypodermic single lumen needle intended for use with a pen injector for the subcutaneous injection of insulin. The information provided is from an FDA 510(k) Clearance Letter.

    1. Acceptance Criteria and Reported Device Performance

    The acceptance criteria for the SureFine Pen Needle are based on compliance with international standards ISO 9626, ISO 11608-2, and ISO 10993-1, 10993-4, 10993-5, 10993-7, 10993-10, and 10993-11 for biocompatibility and sterilization validation according to ISO 11135.

    Test ItemAcceptance Criteria (from referenced ISO standards)Reported Device Performance
    ISO 9626 (Needle Characteristics)All criteria specified in ISO 9626 for hypodermic needlesPass
    MaterialsConformance to specified material propertiesPass
    CleanlinessConformance to cleanliness standardsPass
    Limits for acidity and alkalinityConformance to pH limitsPass
    Size designationAccurate representation of gauge and lengthPass
    DimensionConformance to specified dimensional tolerancesPass
    StiffnessConformance to stiffness requirementsPass
    Resistance to breakageResistance to breakage under specified conditionsPass
    Resistance to corrosionResistance to corrosion under specified conditionsPass
    ISO 11608-2 (Needle-Based Injection Systems)All criteria specified in ISO 11608-2 for pen needlesPass
    DimensionConformance to specified dimensional tolerancesPass
    Flow rateConformance to prescribed flow ratePass
    Bond between hub and needle tubeSecure bond between componentsPass
    Needle point freedom from defects lubricationAbsence of defects and proper lubricationPass
    Dislocation of measuring point at patientMinimal dislocation to ensure accurate deliveryPass
    Compatibility of needles and injector systemFunctional compatibility with pen injectorsPass
    Biocompatibility (ISO 10993 Series)All evaluation acceptance criteriaMet
    Cytotoxicity testSatisfy requirements of ISO 10993-5 (in vitro cytotoxicity)Pass
    Hemolysis testSatisfy requirements of ISO 10993-4 (hemolytic properties)Pass
    Intracutaneous reactivity testSatisfy requirements of ISO 10993-10 (intracutaneous reactivity)Pass
    Skin sensitization testSatisfy requirements of ISO 10993-10 (skin sensitization)Pass
    Acute systemic toxicity testSatisfy requirements of ISO 10993-11 (acute systemic toxicity)Pass
    Pyrogen TestSatisfy requirements of ISO 10993-11 (material-mediated pyrogens)Pass
    Sterilization & Shelf-lifeMax levels of EO/ECH within ISO 10993-7 limits; adequate mitigation of bacterial endotoxins; 5-year shelf-life stability.Conforms to ISO 11135 validated half-cycle method. EOx/ECH levels meet ISO 10993-7. LAL testing performed. Supports 5-year shelf-life.

    2. Sample Size Used for the Test Set and Data Provenance

    The document does not explicitly state the specific sample sizes for each test mentioned (ISO 9626, ISO 11608-2, Biocompatibility). However, it indicates that "Bench tests for the device's performance were conducted" and "Biocompatibility of the SureFine Pen Needle was evaluated". The testing was conducted by Shina Med Corporation, a company based in the Republic of Korea. The nature of these tests (bench testing, lab-based biocompatibility) suggests they are prospective tests performed on manufactured samples of the device. The data provenance is primarily from internal testing and validation by the manufacturer, following international standards.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

    The document does not specify the number or qualifications of experts used to establish ground truth for the "test set" in the context of device performance. The testing performed is against established international standards (ISO 9626, ISO 11608-2, ISO 10993 series), which themselves represent a form of "ground truth" for device safety and performance criteria. The evaluation results are reported as "Pass" or "Met," implying adherence to these predefined standards rather than an expert panel establishing a de novo ground truth for each specific test.

    4. Adjudication Method for the Test Set

    No information is provided regarding an adjudication method (such as 2+1, 3+1, none). The evaluation is based on objective measurements and established criteria set forth by the ISO standards. For instance, a flow rate test would have a pass/fail threshold, not require expert adjudication in the traditional sense.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done

    No. The document describes bench tests and biocompatibility evaluations for a medical device (pen needle), not a diagnostic or AI-assisted system that would typically undergo an MRMC comparative effectiveness study involving human readers. Therefore, there is no mention of an effect size of human readers improving with or without AI assistance.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done

    No. This device is a physical medical instrument (a pen needle), not a software algorithm or AI system. Therefore, the concept of "standalone performance" in the context of algorithms is not applicable here.

    7. The Type of Ground Truth Used

    The ground truth used for demonstrating the device's performance is adherence to established international standards and regulatory requirements:

    • ISO 9626: Specifies requirements for stainless steel needle tubing for the manufacture of hypodermic needles.
    • ISO 11608-2: Specifies requirements and test methods for single-use needles used with pen-injectors.
    • ISO 10993 series: Biological evaluation of medical devices, covering various aspects of biocompatibility (cytotoxicity, sensitization, irritation, systemic toxicity, hemocompatibility, pyrogenicity).
    • ISO 11135: Specifies requirements for the development, validation, and routine control of an ethylene oxide sterilization process for medical devices.
    • ISO 10993-7: EtO residual limits.
    • LAL testing: To ensure absence of bacterial endotoxins.

    8. The Sample Size for the Training Set

    Not applicable. This is a physical medical device, not a machine learning or AI algorithm, so there is no "training set."

    9. How the Ground Truth for the Training Set Was Established

    Not applicable, as there is no training set for a physical medical device.

    Ask a Question

    Ask a specific question about this device

    K Number
    K242006

    Validate with FDA (Live)

    Date Cleared
    2025-02-12

    (218 days)

    Product Code
    Regulation Number
    892.1000
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The SureWave Elastography device is intended for use with Siemens 1.5T and 3.0T MRI systems to generate shear wave vibrations in the body of adult patients during an MRI exam that are translated into images representing tissue stiffness. The images may be used for diagnostic purposes when interpreted by a trained physician.

    Device Description

    The SureWave Elastography device is an accessory to the MRI system comprised of both hardware and software components. The hardware induces shear wave vibrations in the body through a transducer which is driven by a mobile tower and two flexible rotating axes. The transducer is fastened to the patient's body and contains a rotatable eccentric mass which induces vibrations in the body during an MRI scan. The SureWave Elastography reconstruction software uses the acquired image data from the MRI system to create images that show tissue stiffness.

    AI/ML Overview

    The provided text describes the non-clinical and clinical tests performed to demonstrate the substantial equivalence of the SureWave Elastography device to its predicate devices. However, it does not explicitly define acceptance criteria in a table format with specific thresholds. It rather focuses on demonstrating similarity and accuracy to known values and predicate devices.

    Based on the information provided, here's a structured response, with some sections noting what information is not present in the document.

    1. A table of acceptance criteria and the reported device performance

    The document does not explicitly state acceptance criteria in a quantitative table format with predefined thresholds. However, the study aims to demonstrate accuracy and repeatability comparable to, or better than, the predicate device and known phantom values. The reported performance is summarized below:

    Performance MetricImplicit Acceptance Criteria (based on study design)Reported Device Performance (SureWave 2D)Reported Device Performance (SureWave 3D)Notes/Comparison to Predicate
    Accuracy (Phantom, kPa)Values fall between predicate and expected values, and are comparable to expected values.See table below.See table below.- SureWave Elastography 2D and 3D values fall between the predicate and expected stiffness values. - Computed stiffness values were "at least as accurate if not more accurate than the predicate method."
    Specific Phantom Values:
    Expected: 0.82 kPaClose to 0.82 kPa0.76 ± 0.01 kPa (95% CI)0.73 ± 0.02 kPa (95% CI)
    Expected: 2.02 kPaClose to 2.02 kPa2.53 ± 0.94 kPa (95% CI)2.01 ± 0.07 kPa (95% CI)
    Expected: 2.77 kPaClose to 2.77 kPa3.83 ± 0.44 kPa (95% CI)3.27 ± 0.32 kPa (95% CI)
    Expected: 5.80 kPaClose to 5.80 kPa5.80 ± 0.67 kPa (95% CI)5.20 ± 0.09 kPa (95% CI)
    RepeatabilityWithin 10% of the mean of the respective method, and comparable/better than predicate.Within 10% of the mean.Within 10% of the mean."Slightly better than that of the predicate."
    Equivalency to Predicate (SureWave 2D)Strong agreement with predicate, low bias.Linear regression slope: 1.02 (R² = 0.99). Bland-Altman bias: 4%.Not applicable."Excellent agreement," "no significant difference observed," "equivalent."
    Equivalency to Predicate (SureWave 3D)Strong linear relationship. Expected bias due to volumetric analysis.Not applicable.Linear regression R² = 0.99. Bland-Altman bias: approximately 20%."Strong linear relationship." Bias expected and potentially indicates more accuracy.

    2. Sample sizes used for the test set and the data provenance

    • Accuracy Test (Phantom):
      • Sample Size: Not explicitly stated as a "sample size" for a test set, but referred to as "MRE acquisitions from SureWave Elastography 2D, SureWave Elastography 3D, and the predicate MRE methods." The number of acquisitions per method for the general QA phantom repeatability testing was 240 measurements. It is implied similar extensive measurements were taken for the multi-component QA phantom.
      • Data Provenance: Not explicitly stated (e.g., country of origin, retrospective/prospective clinical data). The phantom study is a bench test, not human data.
    • Equivalency Test (Volunteer):
      • Sample Size: 22 healthy adult volunteers.
      • Data Provenance: Not explicitly stated (e.g., country of origin). The study involved "healthy adult volunteers," implying a prospective human study.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    • Phantom Studies: Ground truth for the phantom was established by the phantom manufacturer ("expected stiffness values established by the phantom manufacturer"). No human experts were involved in establishing this ground truth.
    • Volunteer Studies: For the healthy volunteer equivalency study, the "ground truth" was effectively established by the predicate device's measurements. The document states that the SureWave Elastography measurements were plotted against the predicate device to assess their relationship and bias. No independent expert consensus or interpretation was described as establishing a separate ground truth for this human volunteer data beyond the predicate device as the comparison standard.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    • No adjudication method (e.g., expert consensus or multi-reader review beyond the device's output) is described for either the phantom or volunteer test sets. The focus appears to be on direct numerical comparison of the device's output against known phantom values or predicate device measurements.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    • No MRMC comparative effectiveness study involving human readers and AI assistance is described. The study focuses on the device's performance in terms of stiffness measurement accuracy and repeatability, and its equivalence to a predicate device, not on diagnostic improvement for human readers. The device outputs images that are "interpreted by a trained physician," but the study does not evaluate this physician interpretation process.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    • Yes, the conducted tests, precisely the phantom and volunteer studies, evaluate the standalone performance of the SureWave Elastography device in generating stiffness measurements. The device produces quantitative stiffness values (kPa) and qualitative stiffness maps, confidence overlays, and wave images. The accuracy and repeatability of these outputs were directly assessed. While a trained physician interprets the images, the study's focus is on the device's ability to accurately produce those measurements, independent of a specific physician's interpretation improvement.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

    • Phantom Studies: The ground truth was known physical properties of the phantom components, as provided by the phantom manufacturer.
    • Volunteer Studies: The ground truth for comparison was the measurements obtained from the legally marketed predicate device.

    8. The sample size for the training set

    • The document does not provide information about a training set size. The described tests relate to the validation or verification of the device's performance, implying it has already been developed. As this is a 510(k) submission for a device that generates physical properties (stiffness) rather than AI-based diagnostic interpretations, it's possible a separate "training set" in the machine learning sense was not explicitly required or detailed in this summary.

    9. How the ground truth for the training set was established

    • As no training set is described, information on how its ground truth was established is also not present in the document.
    Ask a Question

    Ask a specific question about this device

    K Number
    K242049

    Validate with FDA (Live)

    Device Name
    SureCath Set
    Manufacturer
    Date Cleared
    2024-11-26

    (137 days)

    Product Code
    Regulation Number
    876.5130
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The product is intended for use in male, female, and pediatric patients requiring bladder drainage as determined by their physician. This device is indicated for those individuals unable to promote a natural urine flow or for those individuals who have a significant volume of residual urine following a natural bladder-voiding episode.

    Device Description

    SureCath Set is a sterile, single-use hydrophilic coated intermittent catheter with integrated urine bag. An ampoule of saline solution is provided inside the urine bag. This saline solution from the ampoule is poured over the catheter to activate the coating before use.

    AI/ML Overview

    Here's an analysis of the provided text regarding acceptance criteria and the study:

    The document (K242049) is a Traditional 510(k) Summary for the Coloplast SureCath Set, a sterile, single-use hydrophilic coated intermittent catheter with an integrated urine bag. It demonstrates substantial equivalence to predicate devices. The information provided primarily focuses on non-clinical performance testing rather than human clinical studies involving AI assistance.

    Based on the provided text, the device SureCath Set is a medical device, not an AI/ML powered device. Therefore, a multi-reader multi-case (MRMC) comparative effectiveness study, effect size of human readers with/without AI, or standalone algorithm performance are not applicable.

    Here's a breakdown of the requested information based on the provided document:

    1. A table of acceptance criteria and the reported device performance

    The document broadly states that "Bench performance testing and usability testing were conducted to verify the proposed subject devices met the pre-determined acceptance criteria per specified requirements." It also mentions various ISO and ASTM standards to which the device was tested. However, it does not explicitly list specific quantitative acceptance criteria or detailed reported performance values for each test. It only states that the device "met the pre-determined acceptance criteria."

    Implicit Acceptance Criteria (based on standards and test descriptions):

    Acceptance Criteria CategoryReported Device Performance
    BiocompatibilityMet acceptance criteria per ISO 10993 series for cytotoxicity, irritation/intracutaneous reactivity, sensitization, material mediated pyrogenicity, acute systemic toxicity, and subacute toxicity.
    Catheter Performance (ISO 20696, ASTM F623-19, Coloplast Test Methods)Met acceptance criteria for: - Strength of connection between catheter and urine bag. - Aseptic insertion (non-touch approach) of catheter through urine bag. - Water amount of 20 ml in ampoule must activate coating on catheter. - Friction of the coated catheter after 30 seconds activation time. - Tearing and emptying of the urine bag. - Volume of the urine bag. - Strength of the urine bag.
    UsabilityMet acceptance criteria per EN/IEC 62366-1.
    Packaging IntegrityMet acceptance criteria per ISO 11607-1, DS EN ISO 20696, ASTM D4169-22 for maintenance of sterile barrier and impact of transportation.
    Shelf Life/Aging StabilityMet acceptance criteria after accelerated aging per ASTM F1980-21, demonstrating stability for the defined shelf life (24 months at time of submission, planned 36 months).

    2. Sample size used for the test set and the data provenance

    The document does not specify the exact sample sizes used for each non-clinical test. It states "Testing was performed on final, finished, and sterilized devices." The provenance of the data is from Coloplast A/S, Denmark. The tests are non-clinical, bench performance tests, so the concepts of "retrospective or prospective" human data don't directly apply here.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    This information is not applicable to this type of non-clinical device testing. There are no "experts" establishing a "ground truth" for a test set in the way it would be for an AI/ML clinical study (e.g., radiologists interpreting images). The "ground truth" for non-clinical tests is based on objective measurements against established engineering and biological standards.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    This information is not applicable. Adjudication methods are typically used in clinical studies where expert consensus is needed to establish ground truth for ambiguous cases. Non-clinical bench tests rely on objective measurements against defined standards.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    No, an MRMC comparative effectiveness study was not conducted. The device is a physical medical device (intermittent catheter), not an AI-powered diagnostic or assistive tool. Therefore, there is no "AI assistance" for human readers to improve upon.

    6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done

    No, a standalone algorithm-only performance study was not done. The device is not an algorithm or software requiring human-in-the-loop performance.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

    The "ground truth" for the non-clinical tests is based on:

    • Compliance with recognized international standards (ISO, ASTM, EN/IEC).
    • Pre-determined acceptance criteria defined by Coloplast for specific performance aspects (e.g., strength, friction, volume, activation).
    • Objective measurements of physical and chemical properties.

    8. The sample size for the training set

    This information is not applicable. "Training set" is a concept for AI/ML models. This document describes testing for a physical medical device.

    9. How the ground truth for the training set was established

    This information is not applicable, as there is no training set for a physical medical device.

    Ask a Question

    Ask a specific question about this device

    K Number
    K230891

    Validate with FDA (Live)

    Date Cleared
    2023-12-28

    (272 days)

    Product Code
    Regulation Number
    888.3040
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The SURESTITCH™ UHMWPE Suture PEEK Button Meniscus Repair Implant is indicated for percutaneous or endoscopic soft tissue procedures which requires usage of a suture retention device. For ex: Meniscus repair.

    Device Description

    The SURESTITCH™ UHMWPE Suture PEEK Button Meniscus Repair Implant is an all-inside meniscal repair device with two PEEK non-absorbable polymer implants, preloaded with UHMWPE sutures. SURESTITCH™ UHMWPE Suture PEEK Button Meniscus Repair Implant is preloaded with USP #3-0 UHMWPE suture whereas SURESTITCH™ UHMWPE Suture PEEK Button Meniscus Repair Implant XL is preloaded with USP #2-0 suture. The adjustable depth control knob is used to define the needle piercing length in the meniscus prior to deployment of the implant. These meniscal repair devices are available in various needle curvature.

    AI/ML Overview

    The provided document, K230891, describes the SURESTITCH™ UHMWPE Suture PEEK Button Meniscus Repair Implant and its substantial equivalence to a predicate device. However, the document does not contain acceptance criteria for the device's performance, nor does it detail a specific study proving the device meets said criteria with the requested level of detail for an AI/ML medical device.

    The provided text focuses on the 510(k) submission for a physical medical implant (a suture and PEEK button for meniscus repair) and its comparison to a predicate device, rather than an AI/ML-driven device. As such, information regarding test sets, data provenance, expert ground truth establishment, adjudication methods, MRMC studies, standalone performance, training sets, or how ground truth was established for a training set (all typical for AI/ML device evaluations) are absent.

    Here's an analysis based on the information that is available in the document:

    1. Table of acceptance criteria and the reported device performance

    The document does not explicitly state acceptance criteria in a quantitative manner (e.g., "pull-out strength must be X N"). Instead, it states the conclusion of the test:

    Acceptance Criteria (Implied)Reported Device Performance
    Device performs adequately regarding mechanical strengthSURESTITCH™ UHMWPE Suture PEEK Button Meniscus Repair Implant performs statistically equivalent to the predicate device.
    Device meets pyrogen limit specificationsBacterial endotoxin per USP <85> was conducted to demonstrate that the device meets pyrogen limit specifications.
    Packaging maintains sterile barrier and product integrityPackaging testing was conducted to demonstrate shelf-life and confirm the packaging design provides a protective sterile barrier and protects the integrity of the products post sterilization during shipping and handling.

    2. Sample size used for the test set and the data provenance

    • Sample Size: Not specified for the pull-out strength, cyclical loading, bacterial endotoxin, or packaging tests.
    • Data Provenance: Not specified (e.g., country of origin, retrospective or prospective). The tests are non-clinical, likely bench testing.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    Not applicable. The tests performed are non-clinical mechanical and biological compatibility tests, not evaluations that require expert interpretation of results to establish ground truth in the context of an AI/ML device.

    4. Adjudication method for the test set

    Not applicable. This concept is relevant for AI/ML performance ground truth establishment, not for the types of non-clinical tests described.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    Not applicable. This is not an AI/ML device. The study described compares the subject device's mechanical performance to a predicate device, not human reader performance with or without AI assistance.

    6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done

    Not applicable. This is not an AI/ML device.

    7. The type of ground truth used

    For the non-clinical tests described:

    • Mechanical Performance (Pull-out strength, Cyclical loading): The "ground truth" is measured physical properties and comparison to a predicate device's measured physical properties.
    • Biological Compatibility (Bacterial Endotoxin): The "ground truth" is adherence to established pyrogen limit specifications (USP <85>).
    • Packaging: The "ground truth" is the confirmation of sterile barrier integrity and product protection during simulated conditions.

    8. The sample size for the training set

    Not applicable. This is not an AI/ML device.

    9. How the ground truth for the training set was established

    Not applicable. This is not an AI/ML device.

    Ask a Question

    Ask a specific question about this device

    K Number
    K231227

    Validate with FDA (Live)

    Date Cleared
    2023-12-20

    (236 days)

    Product Code
    Regulation Number
    870.1340
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The sheath is indicated for introducing various cardiovascular catheters to the heart, including the left side of the heart through the interatrial septum.

    The sheath curve can be visualized when used with impedance-based electroanatomical mapping systems.

    Device Description

    The proposed SureFlex HD Steerable Sheath and VersaCross HD Steerable Sheath subject devices are non-pyrogenic, single use devices that are supplied sterile to the UD Steerable Sheath families are comprised of the following components that are packaged together:

    • . One Steerable Sheath
    • . One Transseptal Dilator
    • . One J-tip Mechanical Guidewire

    The SureFlex and VersaCross HD Steerable Sheaths are intended for introducing various cardiovascular catheters to the heart, including the left side of the heart through the interatrial septum. The sheath curve can be visualized when used with separately cleared impedancebased electroanatomical mapping (EAM) systems. The steerable sheath component used in conjunction with a compatible transseptal dilator can facilitate transseptal puncture using separately cleared compatible needle or wire-based puncture devices. The steerable sheath component is connected to the impedance-based EAM system via a compatible legally marketed accessory cable. The subject device is used in a sterile environment where facilities are appropriately equipped and staffed to perform transseptal procedures.

    The steerable sheath component is 8.5 Fr with a 63 cm or 73 cm useable length and is available in three curve options (small, medium, large). It is flexible and bi-directionally deflectable at the distal end (180 degrees clockwise and counterclockwise). The radiopaque tip (containing a platinum/iridium marker band) maximizes visualization of the sheath during manipulation. The sheath contains 4 electrodes equi-spaced along the shaft to enable shaft visualization. The sheath has two distal sideports and the proximal end includes a hemostasis valve and a 3-way stopcock to facilitate the injection or fluids. A silicone lubricant is applied to the outer surface of the shaft for smoother device manipulation. The sheath handle allows the user to control the deflection of the sheath tip through rotation of the knob in the direction of desired distal deflection.

    The compatible dilator component is 8.5F with a 95cm usable length and is comprised of HDPE material with a tungsten radiopaque marker coil at the tip for visualization. The dilator has a tapered distal tip that provides support for the sheath. The compatible mechanical guidewire (0.032″ or 0.035″) component is J-tipped, 180 cm in length and is comprised of a stainless-steel core mandrel with a PTFE coating.

    AI/ML Overview

    The provided document is a 510(k) Premarket Notification for medical devices, specifically steerable sheaths (SureFlex HD Steerable Sheath and VersaCross HD Steerable Sheath). It details the substantial equivalence determination made by the FDA.

    Crucially, this document does NOT describe the acceptance criteria and study proving a device meets acceptance criteria for an AI/Software as a Medical Device (SaMD) product. The medical devices in question are physical catheters/sheaths, and the performance testing summarized refers to mechanical, electrical, biocompatibility, sterilization, pyrogen, and packaging testing, as well as a simulated-use animal study. These are typical validation activities for a physical medical device.

    The only mention of "visualization" or "mapping systems" is in the context of the sheath's curve being visualized when used with impedance-based electroanatomical mapping systems, and the predicate device's compatibility with CARTO® 3 EP Navigation Systems. This refers to the ability of the physical sheath to be seen on separate, compatible systems, not an AI or SaMD performing a diagnostic or therapeutic function itself.

    Therefore, I cannot provide the requested information regarding acceptance criteria and study proving a device meets acceptance criteria for an AI/Software as a Medical Device (SaMD) product based on the provided text, as this information is not present. The document focuses on the substantial equivalence of a physical medical device.

    Ask a Question

    Ask a specific question about this device

    K Number
    K230646

    Validate with FDA (Live)

    Date Cleared
    2023-11-29

    (266 days)

    Product Code
    Regulation Number
    876.1075
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The UR0-1 SUREcore Prime Device is a reusable system for histological core biopsies. It is used to obtain tissue from transrectal or transperineal biopsy of the prostate. It has a throw (advancement) of 25 mm and is used in conjunction with a single use SUREcore biopsy needle.

    Device Description

    The SUREcore Prime Biopsy Instrument facilitates collection of tissue for analysis by pathology in order to assist with a diagnosis of a disease condition in a patient. The SUREcore Prime device consists of two elements: (1) a handle and (2) a needle set. The handle contains springs that energize the biopsy needles and when used will cause rapid advance of the needles into target tissue. This action causes tissue to be held within a component of the needle (the core collector) intact from the body. The tissue samples are removed from the needle set and prepared for transfer to a pathologist or lab. The user can reenergize the springs by pulling back on the cocking slide so that multiple tissue samples can be collected from different locations within the target tissue.

    AI/ML Overview

    This document is a 510(k) summary for the URO-1 SUREcore Prime Biopsy Instrument. It focuses on demonstrating substantial equivalence to a predicate device, the Cook BX Biopsy Device, rather than presenting a study proving that the device meets specific acceptance criteria based on diagnostic performance metrics (like sensitivity, specificity, or AUC) as one might expect for an AI/software as a medical device (SaMD) product.

    Therefore, many of the requested elements regarding acceptance criteria and study design for diagnostic performance (such as sample size, expert ground truth, MRMC studies, standalone performance, etc.) are not applicable to this submission, as it is for a mechanical biopsy instrument.

    However, the document does describe performance testing relevant to the function of a mechanical biopsy device.

    Here's a breakdown based on the provided text, addressing the points where information is available or indicating when a point is not applicable to this type of device submission:


    1. A table of acceptance criteria and the reported device performance

    The document doesn't explicitly list "acceptance criteria" in a quantitative diagnostic sense with target thresholds (e.g., "sensitivity > 90%"). Instead, the performance testing described is functional and safety-related for a mechanical device. The "performance" is implicitly demonstrated by passing these tests and showing similarity to the predicate.

    Acceptance Criteria (Implied)Reported Device Performance
    Biocompatibility: No cytotoxicity, sensitization, irritation, or systemic toxicity from patient-contacting materials (needle set). Handle material should also not show cytotoxicity.Needle set passed all tests: Cytotoxicity, Sensitization, Irritation, and Systemic toxicity (as noted in K220611). Handle assessed for cytotoxicity and found to pass.
    Mechanical Function: Capacity to collect tissue from a prostate model.Successfully collected tissue in a prostate model.
    Mechanical Strength: Withstand deformation.Force necessary to deform the needle set (16 and 18 gauge) was tested. (Specific force values not provided, but implies it met internal specifications).
    Electrical Safety/EMC: (Not applicable for non-electrical device)"Not required" as the device has no electronic components.
    Software V&V: (Not applicable for non-software device)"Neither the subject nor predicate devices contain software."

    2. Sample size used for the test set and the data provenance

    • Sample Size for Test Set: Not specified in terms of a "test set" of patient cases for diagnostic performance, as this is a mechanical device. For mechanical testing, the number of units tested (e.g., prostate models, individual needles) is not detailed.
    • Data Provenance: Not applicable in the context of clinical data for diagnostic performance. The mechanical testing was described as "bench testing."

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    • Not applicable. This submission does not involve diagnostic interpretation studies requiring expert ground truth for imaging or similar data. The "ground truth" for mechanical testing would be the physical properties and functionality of the device itself (e.g., did it puncture, did it collect tissue).

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    • Not applicable. No expert adjudication was needed for this type of mechanical device testing.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    • No. This is a mechanical biopsy instrument, not an AI/SaMD product. Therefore, MRMC studies and human reader improvement with AI assistance are not relevant.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    • No. This is a mechanical device, not an algorithm.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

    • For biocompatibility: Laboratory testing against ISO standards (Cytotoxicity, Sensitization, Irritation, Systemic toxicity).
    • For mechanical function: Physical observation in a prostate model and force measurement for deformation. There isn't a "ground truth" in the diagnostic sense; it's about meeting pre-defined mechanical specifications.

    8. The sample size for the training set

    • Not applicable. This is a mechanical device and does not involve AI or machine learning algorithms that require training data.

    9. How the ground truth for the training set was established

    • Not applicable. As above, no training set or ground truth in this context is relevant.

    In summary, this FDA submission for the URO-1 SUREcore Prime Biopsy Instrument focuses on demonstrating "substantial equivalence" to a predicate mechanical device through design comparisons and basic functional/safety bench testing, rather than through complex clinical studies evaluating diagnostic performance or AI algorithms. The "acceptance criteria" are implicitly met by passing the described biocompatibility and mechanical tests, and by showing equivalence to a legally marketed predicate.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 13