Search Filters

Search Results

Found 496 results

510(k) Data Aggregation

    Why did this record match?
    510k Summary Text (Full-text Search) :

    Re: K251181
    Trade/Device Name: Splashwire Hydrophilic Guide Wire
    Regulation Number: 21 CFR 870.1330
    code:** | DQX |
    | Classification Name: | Wire, Guide, Catheter |
    | Regulation Number: | 21 CFR 870.1330
    code:** | DQX |
    | Classification Name: | Wire, Guide, Catheter |
    | Regulation Number: | 21 CFR 870.1330

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Merit Hydrophilic Guide Wire is intended to be used in the peripheral vascular system to facilitate the placement of devices during diagnostic and interventional procedures. These guide wires are not intended for Percutaneous Transluminal Coronary Angioplasty use or for use in the neurovasculature.

    Device Description

    The Splashwire Hydrophilic guide wire consists of a nitinol core wire with a ground tapered distal section. A polyurethane jacket which contains tungsten for radiopacity is applied over the core wire and a hydrophilic coating is applied over the polyurethane jacket. The surface of the jacket is uniform with both the distal and proximal ends fully coated. The wire is placed inside a multiple loop flush dispenser, also referred to as a hoop. The dispenser has a flush port which facilitates solution flushing through the hoop to hydrate the guide wire. The wire is placed so that the distal end of the wire comes out of the outer portion of the hoop. A J-straightener is placed on the other end of the hoop to introduce the wire into the catheter. The wire is sold sterile and is a single use device. The subject Splashwire Hydrophilic Guide Wire Line Extensions and the predicate Splashwire Hydrophilic Guide Wire share the device characteristics described above, with the subject guide wire incorporating the following additional modifications:

    J-Tip Guide Wires:
    The J Tip devices incorporate a distal tip formed into a J shape. These guide wires are provided with a 1.5 mm J Tip.

    Long Taper Guide Wires:
    The Long Taper Guide Wires incorporate an extended core wire taper length providing enhanced flexibility.

    AI/ML Overview

    The provided FDA 510(k) clearance letter and summary for the Splashwire Hydrophilic Guide Wire (K251181) outlines the regulatory approval process and demonstrates substantial equivalence to a predicate device (K201595). However, it does not contain the specific information required to address your request regarding acceptance criteria and the detailed study that proves the device meets those criteria, particularly for an AI-enabled medical device.

    The document describes a traditional medical device (a guide wire) and the non-clinical bench testing performed to ensure its safety and performance. This typically involves physical and mechanical property tests. It does not refer to AI/ML software, algorithms, or any studies involving human readers or ground truth expert consensus in the way you've outlined.

    Therefore, I cannot provide a table of acceptance criteria and reported device performance for an AI-enabled device, nor details on sample sizes, expert ground truth establishment, or multi-reader multi-case studies, as this information is not present in the provided text.

    Based on the provided document, the device is a physical medical device (a guide wire), not an AI-enabled one. The acceptance criteria and "study" described are standard bench tests for physical properties, not performance metrics for an AI algorithm.

    Here's a breakdown of what can be inferred from the document regarding the device's performance, but it does not align with the parameters of an AI/ML study:

    1. Table of Acceptance Criteria and Reported Device Performance:

    The document states: "Testing was conducted on the subject Splashwire Hydrophilic Guide Wire in accordance with protocols based on requirements outlined in guidance's and industry standards and these were shown to meet the acceptance criteria that were determined to demonstrate substantial equivalence." And "All test results were comparable to the predicate Splashwire Hydrophilic Guide Wire and the subject Splashwire Hydrophilic Guide Wire met the predetermined acceptance criteria."

    While specific numerical acceptance criteria and performance data are not provided in this summary, the document indicates that the following categories of non-clinical bench tests were successfully completed and met their predetermined acceptance criteria:

    Acceptance Criteria Category (Test Type)Reported Device Performance (Implied)
    Dimensional VerificationMet predetermined specifications (comparable to predicate)
    Finished Wire SurfaceMet predetermined specifications (comparable to predicate)
    Tip Tensile StrengthMet predetermined specifications (comparable to predicate)
    Tip StiffnessMet predetermined specifications (comparable to predicate)
    Tip FlexibilityMet predetermined specifications (comparable to predicate)
    Prolapse TestMet predetermined specifications (comparable to predicate)
    Torque StrengthMet predetermined specifications (comparable to predicate)
    TorqueabilityMet predetermined specifications (comparable to predicate)
    Coating IntegrityMet predetermined specifications (comparable to predicate)
    Catheter CompatibilityMet predetermined specifications (comparable to predicate)
    Fracture TestMet predetermined specifications (comparable to predicate)
    Flexing TestMet predetermined specifications (comparable to predicate)
    Simulated Use TestingMet predetermined specifications (comparable to predicate)
    Kink ResistanceMet predetermined specifications (comparable to predicate)
    Ancillary Device CompatibilityMet predetermined specifications (comparable to predicate)

    2. Sample Size Used for the Test Set and Data Provenance:

    This information is not available in the provided document. For a physical medical device, "test set" would refer to the number of units tested for each characteristic. The provenance of the data is not applicable in the context of clinical images or patient data; it would refer to the samples of the manufactured guide wire used for the bench tests. These are likely internal lab tests.

    3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications:

    This information is not applicable to this type of device and study. "Ground truth" in this context would be derived from the physical properties of the device measured by laboratory instruments, not expert human interpretation of data.

    4. Adjudication Method for the Test Set:

    This information is not applicable as there are no human interpretations or adjudications involved in the physical bench tests.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done:

    This information is not applicable. MRMC studies are relevant for software that aids human interpretation (e.g., AI for radiology). This device is a physical guide wire.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was done:

    This information is not applicable. This describes the performance of a physical device, not an algorithm.

    7. The Type of Ground Truth Used:

    The "ground truth" for this device is based on measurements of its physical and mechanical properties according to established industry standards and FDA guidance for guide wires (e.g., ISO standards, FDA guidance on guidewire performance). It's not expert consensus, pathology, or outcomes data.

    8. The Sample Size for the Training Set:

    This information is not applicable. This device is not an AI/ML algorithm that requires a training set. The "design controls" mentioned refer to the engineering and manufacturing process, not data training.

    9. How the Ground Truth for the Training Set Was Established:

    This information is not applicable as there is no training set for a physical device.

    In summary, the provided document details the 510(k) clearance for a physical medical device (Premarket Notification: K251181) and its substantial equivalence based on standard bench testing. It does not provide information relevant to the assessment of an AI-enabled medical device.

    Ask a Question

    Ask a specific question about this device

    K Number
    K244061
    Device Name
    X-Wire Guidewire
    Date Cleared
    2025-08-26

    (238 days)

    Product Code
    Regulation Number
    870.1330
    Reference & Predicate Devices
    Why did this record match?
    510k Summary Text (Full-text Search) :

    California 95008

    Re: K244061
    Trade/Device Name: X-Wire Guidewire
    Regulation Number: 21 CFR 870.1330
    Classification Name: Catheter guide wire
    Product Codes: MOF, DQX
    Regulation: 21 CFR 870.1330
    Classification Name:* Catheter guide wire
    Product Codes: MOF, DQX
    Regulation: 21 CFR. 870.1330
    Subject Device X-Wire Guidewire (K244061) |
    |---|---|---|---|
    | Regulation Number | 21 CFR 870.1330

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The X-Wire Guidewire is indicated for general intravascular use within the peripheral and neuro vasculature to introduce and position catheters and other interventional devices. The guidewire is not intended for use in the coronary vasculature.

    Device Description

    The Imperative Care's X-Wire Guidewires are guidewires with shapeable tips to aid in accessing the peripheral and neuro vasculature. The guidewires are available in 200cm - 300cm lengths, standard (S) and support (T) stiffness profiles, and 0.014", 0.018" and 0.024" diameters. The distal portion of the guidewire tip includes a radiopaque marker to facilitate fluoroscopic visualization. A hydrophilic coating on the distal segment and PTFE coating on the proximal segment serve to reduce friction during manipulation in vessels. The X-Wire Guidewire is supplied with a shaping mandrel, introducer, and torque device.

    AI/ML Overview

    This document is a 510(k) clearance letter for a medical device and therefore does not contain acceptance criteria and study data in the format typically found in clinical efficacy studies for AI/software devices. The acceptance criteria and "study" described herein relate to the safety and performance of the physical guidewire device (X-Wire Guidewire) based on bench testing, biocompatibility testing, sterilization validation, and shelf-life testing, rather than a clinical study evaluating the performance of an AI algorithm.

    Here's a breakdown of the requested information based on the provided document:

    1. Table of Acceptance Criteria and Reported Device Performance

    The document summarizes the performance testing done in Table 2: Summary of Bench Tests and Performance Specifications and Table 3: Biocompatibility Test Summary – X-Wire Guidewire.

    Table: Acceptance Criteria and Reported Device Performance (Synthesized from document)

    Test AttributeAcceptance Criteria (Specification)Reported Device Performance (Results)
    Dimensional VerificationAll defined guidewire dimensions are within the specified tolerances.Pass
    Visual InspectionThe guidewire shall be free of visual defects when removed from packaging.Pass
    TorqueabilityThe guidewire shall transmit rotation from the proximal end to the distal tip to allow users to select branches of the vasculature and reach the target location.Pass
    Torque StrengthDevice shall not fail under expected torsional input when distal end is unable to move/rotate.Pass
    Tip FlexibilityTip of guidewire shall not cause vessel damage; defined by ability to buckle to reduce contact pressure and prevent perforation.Pass
    Tensile Strength and Tip PullGuidewire shall withstand tensile forces expected in clinical use without breaking.Pass
    Kink ResistanceGuidewire shall be able to traverse through clinically relevant bends without kinking.Pass
    Coating IntegrityCoating shall remain intact during clinical use.Pass
    Coating LubricityCoating shall be lubricious to reduce frictional forces for navigation.Pass
    Particulate EvaluationGuidewire shall not generate particles at a level greater than the range generated by the predicate.Pass
    Corrosion ResistanceGuidewire shall not corrode from time of manufacture through its shelf life.Pass
    RadiopacityOpacity to x-ray shall allow physicians to visualize guidewire under fluoroscopy.Pass
    Simulated UseGuidewire must be able to reach anatomical locations and deliver catheters and other interventional devices used in common neurovascular procedures.Pass
    CytotoxicityTest article extracts must yield grade 2 or lower.Pass, Noncytotoxic
    SensitizationOverall pattern, intensity, duration, character of reactions compared to control conditions.Pass, Nonsensitizer
    Irritation or Intracutaneous ReactivityDifference between test extract mean score and corresponding control mean score ≤1.Pass, Nonirritating
    Acute Systemic ToxicityNone of the animals treated with test article extracts must show significantly greater biological reactivity than control-treated animals.Pass, Non-toxic (acute systemic)
    Material Mediated PyrogenicityTest article extract must yield
    Ask a Question

    Ask a specific question about this device

    K Number
    K250552
    Manufacturer
    Date Cleared
    2025-07-25

    (150 days)

    Product Code
    Regulation Number
    870.1330
    Reference & Predicate Devices
    Why did this record match?
    510k Summary Text (Full-text Search) :

    Hi-Torque Command 14 ST Guide Wire and Hi-Torque Command 14 MT Guide Wire
    Regulation Number: 21 CFR 870.1330
    Product Classification** | Class II |
    | Product Code | DQX |
    | Product Regulation Number | 21 CFR 870.1330

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Hi-Torque Command™ 14 ST Guide Wire and Hi-Torque Command™ 14 MT Guide Wire are indicated to facilitate the placement of balloon dilatation catheters during percutaneous transluminal angioplasty (PTA), in arteries such as the femoral, popliteal and infra-popliteal arteries. The guide wires may also be used with compatible stent devices during therapeutic procedures.

    The guide wires may also be used to reach and cross a target lesion, provide a pathway within the vessel structure, facilitate the substitution of one diagnostic or interventional device for another, and to distinguish the vasculature.

    Device Description

    The Hi-Torque Command™ 14 ST Guide Wire (guide wire with short taper) and Hi-Torque Command™ 14 MT Guide Wire (guide wire with medium taper) have a maximum diameter of 0.0144" (0.366 mm) and are provided in 210 cm and 300 cm lengths. The distal tip of the guide wire is available as a straight tip or an angled tip, both of which are shapeable.

    AI/ML Overview

    The provided FDA 510(k) clearance letter pertains to guide wires, which are physical medical devices, not AI/software systems. Therefore, the questions related to AI/software performance (e.g., ground truth establishment, MRMC studies, training/test sets, expert adjudication) are not applicable to this document.

    For a physical device like a guide wire, the "acceptance criteria" and "study that proves the device meets the acceptance criteria" refer to non-clinical performance testing to demonstrate that the device performs as intended and is substantially equivalent to a predicate device, ensuring its safety and effectiveness.

    Here's an analysis based on the provided document for the guide wires:

    Analysis of Acceptance Criteria and Performance for Abbott Medical Guide Wires (K250552)

    The substance of this 510(k) clearance is that the subject devices are identical to predicate devices, and the submission is primarily for a labeling change. This means extensive new performance testing demonstrating equivalence to a new set of criteria beyond what was already established for the predicate is not explicitly detailed as part of "new" acceptance criteria in this document. Instead, the testing listed serves to confirm that the labeling changes do not negatively impact the device's safe and effective use, and that the subject device remains substantially equivalent to its predicate.

    Given that the device design, materials, and technological characteristics are identical between the subject and predicate devices, the "acceptance criteria" for this specific 510(k) are implicitly tied to confirming that the existing performance characteristics (established for the predicate) are maintained and that the labeling changes introduce no new risks or compromises to performance.

    1. Table of Acceptance Criteria and Reported Device Performance

    Since this 510(k) is for a device identical to its predicate and addressing labeling changes, the "acceptance criteria" are implied to be meeting established performance standards for guide wires and demonstrating no negative impact from labeling updates. Actual numerical performance data is not typically a part of the publicly available 510(k) summary, but rather part of the internal design verification and validation report submitted to the FDA.

    Acceptance Criteria Category (Implied)Reported Device Performance (as stated in 510(k) Summary)
    Mechanical/Physical PerformanceSubstantial equivalence demonstrated through:
    Friction performanceTesting completed to support labeling changes.
    Kink resistanceData support conclusion that labeling updates do not negatively impact safe and effective use.
    Bending durabilitySubject and predicate devices are substantially equivalent.
    Fracture resistance
    Torsional wire strength
    Tip tensile strength
    Particulate generation
    Rotational accuracy
    Coating integrity
    Simulated use
    Safety and Effectiveness"The subject and predicate devices have the identical intended use, indications for use, device design, material composition, and technological characteristics. Testing data to evaluate the labeling changes in scope of the 510(k) submission demonstrate that the subject and predicate devices are substantially equivalent in safety and effectiveness."

    2. Sample Size Used for the Test Set and Data Provenance

    • Sample Size: The document does not specify the sample sizes used for each of the non-clinical tests (Friction, Kink resistance, etc.). These details are typically found in the full test reports submitted to the FDA, not in the 510(k) summary.
    • Data Provenance: The document does not specify the country of origin of the data. For physical device performance testing, this often occurs in internal labs or contract research organizations. It is inherently prospective for the specific tests performed to support the 510(k) submission, as new data is generated to demonstrate compliance.

    3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications

    This question is not applicable. For physical device testing (mechanical, durability, etc.), "ground truth" is established by direct measurement against engineering specifications and industry standards, not by expert human interpretation like in medical imaging AI.

    4. Adjudication Method for the Test Set

    This question is not applicable. Adjudication methods (like 2+1, 3+1) are relevant for subjective human interpretations of data, such as medical image readings, typically in AI/software performance studies. They are not used for objective physical device performance testing.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done

    No, an MRMC study was not done. MRMC studies are specific to evaluating the diagnostic performance of software or imaging systems, often comparing human readers with and without AI assistance. This is a physical guide wire, so such a study is irrelevant.

    6. If a Standalone (i.e., Algorithm Only Without Human-in-the-Loop Performance) Was Done

    This question is not applicable. "Standalone performance" refers to the performance of an algorithm or software system without human interaction. This document describes a physical medical device.

    7. The Type of Ground Truth Used

    For physical device testing, the "ground truth" is based on:

    • Engineering Specifications: The design parameters and tolerances the device is intended to meet.
    • Industry Standards: Recognized standards for guide wire performance (e.g., ISO, ASTM).
    • Predicate Device Performance: The established, acceptable performance of the legally marketed predicate device.
      Data from the listed tests (Friction, Kink resistance, etc.) are measured and compared against these objective criteria.

    8. The Sample Size for the Training Set

    This question is not applicable. "Training set" refers to data used to train AI models. This document describes a physical medical device.

    9. How the Ground Truth for the Training Set Was Established

    This question is not applicable for the same reason as above.

    In summary, the provided 510(k) clearance letter is for a physical medical device (guide wires) that is stated to be identical in design, materials, and characteristics to a legally marketed predicate device, with the clearance primarily relating to labeling changes. Therefore, the "acceptance criteria" and "proof" focus on demonstrating that these labeling changes do not adversely affect the device's established safety and performance, and that the device remains substantially equivalent to its predicate through standard non-clinical performance testing. The concepts of AI/software-specific studies, ground truth establishment by experts, and training/test sets are not relevant to this type of device and clearance.

    Ask a Question

    Ask a specific question about this device

    K Number
    K243756
    Date Cleared
    2025-07-17

    (223 days)

    Product Code
    Regulation Number
    870.1330
    Reference & Predicate Devices
    Why did this record match?
    510k Summary Text (Full-text Search) :

    California 94965

    Re: K243756
    Trade/Device Name: 0.014" Willow Guidewire
    Regulation Number: 21 CFR 870.1330
    Guidewire
    Common Name: Wire, Guide, Catheter
    Classification Name: Wire, Guide, Catheter, 21 CFR 870.1330
    0.014" Willow Guidewire (K243756) |
    |---|---|---|
    | Classification Name | Wire, Guide, Catheter, 21 CFR 870.1330

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The 0.014" Willow Guidewire is intended for general intravascular use, including neurovascular and peripheral vasculatures. It can be used to selectively introduce and position catheters and other interventional devices within the peripheral and neuro vasculature. This device should be used only by physicians trained in percutaneous, intravascular techniques and procedures.

    Device Description

    Like the predicate device, the 0.014" Willow Guidewire is a single-use product with a shapeable tip available in straight and pre-shaped configurations, used to gain intravascular access to facilitate the positioning and exchange of interventional devices in small diameter, tortuous vasculature for neuro and peripheral diagnostic and interventional procedures. The wire can be torqued to facilitate navigation through the vasculature. The 0.014" Willow Guidewire comes in three stiffness profiles: Soft, Standard and Support.

    AI/ML Overview

    The provided FDA 510(k) clearance letter and summary for the 0.014" Willow Guidewire do not describe a study involving an AI/Machine Learning device.

    Instead, this document describes a traditional medical device (a guidewire) and its clearance process through substantial equivalence to a legally marketed predicate device. The "acceptance criteria" and "study that proves the device meets the acceptance criteria" in this context refer to engineering and biocompatibility bench testing, not clinical performance studies for an AI algorithm.

    Therefore, many of the questions related to AI/ML device performance (like sample size for test/training sets, expert ground truth, adjudication methods, MRMC studies, standalone performance, effect size of AI assistance) are not applicable to this document.

    However, I can extract the relevant information regarding the acceptance criteria and the engineering study (bench performance testing) for the 0.014" Willow Guidewire.


    Acceptance Criteria and Study for the 0.014" Willow Guidewire

    The 0.014" Willow Guidewire is a traditional medical device, not an AI/ML device. Therefore, the "acceptance criteria" and "study" refer to engineering specifications and bench performance testing, along with biocompatibility testing, to demonstrate substantial equivalence to a predicate device.

    1. Table of Acceptance Criteria and Reported Device Performance

    The acceptance criteria are generally implied by "Acceptance criteria were met" or "considered acceptable" for each test. The reported device performance is that these criteria were successfully achieved.

    TestTest Method SummaryReported Device Performance / Conclusions
    Bench Performance Testing
    Coating LubricityHydrophilic coating lubricity was assessed after multiple pull cycles through silicone pads.Acceptance criteria were met.
    Coating DurabilityCoating durability was assessed after repeating multiple pull cycles through silicone pads.Acceptance criteria were met.
    Coating IntegrityPer FDA guidance "Coronary, Peripheral, and Neurovascular Guidewires – Performance Tests and Recommended Labeling" (2019). Inspected pre- and post-simulated use tracking with comparison to predicate.Coating integrity was reported and considered acceptable.
    Corrosion ResistancePer FDA guidance "Coronary, Peripheral, and Neurovascular Guidewires – Performance Tests and Recommended Labeling" (2019).Acceptance criteria were met.
    Dimensional & Visual Insp.Per FDA guidance "Coronary, Peripheral, and Neurovascular Guidewires – Performance Tests and Recommended Labeling" (2019). Verified key dimensions.Acceptance criteria were met.
    Flex FatigueSubjected guidewire to multiple flexure cycles around cylindrical pins.Acceptance criteria were met.
    FractureSubjected guidewire to multiple wrappings around cylinder and visually inspected for signs of fracture.Acceptance criteria were met.
    Kink ResistancePer FDA guidance "Coronary, Peripheral, and Neurovascular Guidewires – Performance Tests and Recommended Labeling" (2019). Wrapped guidewire around test fixture with clinically relevant radii.Acceptance criteria were met.
    Particulate CharacterizationPer FDA guidance "Coronary, Peripheral, and Neurovascular Guidewires – Performance Tests and Recommended Labeling" (2019). Counted particulates after tracking through tortuous simulated use model, with comparison to predicate.Particulate counts were reported and considered acceptable.
    RadiopacityPer FDA guidance "Coronary, Peripheral, and Neurovascular Guidewires – Performance Tests and Recommended Labeling" (2019). Evaluated subject and predicate guidewires under fluoroscopy.All Willow Guidewires demonstrated acceptable radiopacity; comparable to predicate.
    Simulated UsePer FDA guidance "Coronary, Peripheral, and Neurovascular Guidewires – Performance Tests and Recommended Labeling" (2019). Tested with microcatheter, introducer, and torque device while navigating to target in a tortuous simulated use model.Acceptance criteria were met.
    Tensile StrengthPer FDA guidance "Coronary, Peripheral, and Neurovascular Guidewires – Performance Tests and Recommended Labeling" (2019). Measured force to break at each bond.Acceptance criteria were met.
    Tip FlexibilityPer FDA guidance "Coronary, Peripheral, and Neurovascular Guidewires – Performance Tests and Recommended Labeling" (2019). Measured force to deflect tip at 5 mm, 10 mm, and 20 mm gauge lengths, compared to predicate.Acceptance criteria were met.
    Tip ShapeabilityShaped the guidewire tip three times per labeling.Acceptance criteria were met.
    Tip Shape RetentionMeasured tip shape retention after tracking through a tortuous simulated use model.Acceptance criteria were met.
    TorqueabilityPer FDA guidance "Coronary, Peripheral, and Neurovascular Guidewires – Performance Tests and Recommended Labeling" (2019). Tracked through tortuous simulated use model; evaluated steerability and torque response.Acceptance criteria were met.
    Torque StrengthPer FDA guidance "Coronary, Peripheral, and Neurovascular Guidewires – Performance Tests and Recommended Labeling" (2019). Recorded rotations until failure.Acceptance criteria were met.
    Biocompatibility Testing
    Acute Systemic ToxicityPer ISO 10993-11.Pass: No evidence of acute systemic toxicity.
    Compliment ActivationPer ISO 10993-4.Pass: Non-activator.
    Cytotoxicity – MEM ElutionPer ISO 10993-5.Pass: Non-cytotoxic.
    HemolysisPer ISO 10993-4.Pass: Non-hemolytic.
    Intracutaneous ReactivityPer ISO 10993-10.Pass: Non-reactive.
    In Vivo ThrombogenicityPer ISO 10993-4.Pass: Non-Thrombogenic.
    PyrogenicityPer ISO 10993-11.Pass: Non-pyrogenic.
    SensitizationPer ISO 10993-10.Pass: Non-sensitizing.

    Information Not Applicable (NA) to this Device/Submission:

    The following questions are specifically targeted at AI/ML medical devices and are not relevant to the 0.014" Willow Guidewire, which is a traditional physical medical device.

    1. Sample size used for the test set and the data provenance: NA (This is a physical device, not an AI/ML algorithm requiring a data test set.)
    2. Number of experts used to establish the ground truth for the test set and the qualifications of those experts: NA (No "ground truth" in the AI/ML sense is established for this type of device.)
    3. Adjudication method (e.g. 2+1, 3+1, none) for the test set: NA (No adjudication method described for physical device testing.)
    4. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance: NA (No MRMC study or AI assistance involved.)
    5. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done: NA (Not an algorithm.)
    6. The type of ground truth used (expert consensus, pathology, outcomes data, etc.): NA (Ground truth is not a concept applied in this context; performance is measured against engineering specifications and predicate characteristics.)
    7. The sample size for the training set: NA (No training set for a physical device.)
    8. How the ground truth for the training set was established: NA (No training set or ground truth in this context.)

    Ask a Question

    Ask a specific question about this device

    K Number
    K250075
    Manufacturer
    Date Cleared
    2025-06-13

    (154 days)

    Product Code
    Regulation Number
    870.1330
    Reference & Predicate Devices
    Why did this record match?
    510k Summary Text (Full-text Search) :

    **
    Trade/Device Name: Medtronic Stedi Extra Support Guidewire
    Regulation Number: 21 CFR 870.1330
    **
    Trade/Device Name: Medtronic Stedi Extra Support Guidewire
    Regulation Number: 21 CFR 870.1330
    Classification Name:** Wire, Guide, Catheter
    Classification: Class II
    Regulation Number: 21 CFR 870.1330

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Stedi Guidewire is intended for use to introduce and position catheters during interventional procedures within the chambers of the heart, including transcatheter aortic valve replacement (TAVR).

    Device Description

    The Medtronic Stedi™ Extra Support Guidewire (herein after referred as Stedi Guidewire) is design for use to introduce and position catheters during interventional procedures within the chambers of the heart, including transcatheter aortic valve replacement (TAVR) procedures.

    The Stedi Guidewire has a 0.035" diameter and is 275cm in length and composed of two primary components: a core, and a coil. Both components are made of stainless steel. The core wire component is a piece of stainless-steel wire which is ground on the distal end to fit into the coil and provide flexibility. The coil and the ground core are joined in two locations: a proximal bond and a distal weld. The distal end of the Stedi Guidewire is comprised of a preformed 540° curved tip is available in 2 sizes (3cm and 4cm). The Stedi Guidewire has a polytetrafluoroethylene (PTFE) coating applied to the entire length of the device in order to aid in lubricity.

    The Stedi Guidewire is sterilized using ethylene oxide, nonpyrogenic, disposable, and for single use only.

    AI/ML Overview

    Based on the provided FDA 510(k) clearance letter for the Medtronic Stedi Extra Support Guidewire, here's a detailed breakdown of the acceptance criteria and the study information.

    It's important to note that the provided document is a 510(k) clearance letter, which focuses on demonstrating substantial equivalence to a predicate device. For medical devices like guidewires, the "studies" primarily consist of non-clinical (bench) performance testing to ensure the new device meets established safety and performance requirements, rather than clinical trials with human subjects in the way AI/software devices typically undergo. Therefore, many of the questions related to human readers, ground truth, and training sets are not applicable in this context.


    Acceptance Criteria and Reported Device Performance

    The acceptance criteria for the Medtronic Stedi Extra Support Guidewire are demonstrated through various non-clinical (bench) performance tests. The FDA guidance "Coronary, Peripheral and Neurovascular Guidewires Performance Tests and Recommended Labeling (October 10, 2019)" was utilized to establish these tests. The conclusion states that the device "met all design input requirements based on the intended use."

    Here's a table summarizing the types of tests conducted, which imply the acceptance criteria were met by the device's performance in these areas:

    Acceptance Criterion (Type of Test)Reported Device Performance
    Mechanical Performance:
    Dimensional VerificationMet applicable design and performance requirements
    Visual InspectionMet applicable design and performance requirements
    Tensile Strength (Proximal & Distal Bond)Met applicable design and performance requirements
    Torque StrengthMet applicable design and performance requirements
    Lubricity/Pinch ForceMet applicable design and performance requirements
    Kink ResistanceMet applicable design and performance requirements
    Tip Flexibility/Spiral Tip CompressionMet applicable design and performance requirements
    Flex TestMet applicable design and performance requirements
    Fracture TestMet applicable design and performance requirements
    Three-Point Bend TestMet applicable design and performance requirements
    Material/Biocompatibility:
    Coating IntegrityMet applicable design and performance requirements
    Corrosion ResistanceMet applicable design and performance requirements
    Particulate Evaluation & Chemical CharacterizationMet applicable design and performance requirements
    Biocompatibility Testing (Cytotoxicity, Sensitization, Irritation, Acute Systemic Toxicity, Material Mediated Pyrogenicity, Hemolysis, Complement Activation, Thrombogenicity)Compliant with ISO 10993-1 requirements
    Sterility/Packaging:
    Sterilization ValidationCompliant with ISO 11135 requirements
    Packaging Design Verification TestingCompliant with ISO 11607 requirements
    Durability:
    Shelf Life TestingMet applicable design and performance requirements
    Simulated Use:
    Simulated Use/CompatibilityMet applicable design and performance requirements
    RadiopacityMet applicable design and performance requirements

    Study Details

    1. Sample size used for the test set and the data provenance:

      • Sample Size: The document does not specify exact sample sizes for each non-clinical test. However, it indicates "samples were analyzed according to predetermined acceptance criteria" for the various bench tests. In medical device bench testing, sample sizes are typically determined statistically to ensure sufficient power to detect differences or to demonstrate compliance with specifications.
      • Data Provenance: The data is generated from non-clinical bench testing performed by Medtronic Inc. This is not clinical data (i.e., no patient data is involved). It is prospective in the sense that the tests were designed and executed to evaluate the new device.
    2. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

      • N/A. For this type of medical device (guidewire), ground truth is established through engineering specifications, material science standards (e.g., ISO standards), and performance benchmarks derived from predicate devices and historical data. It does not involve human expert consensus in the diagnostic sense. The "experts" are the engineers, material scientists, and testers who design and conduct the tests and interpret the results against predetermined specifications.
    3. Adjudication method for the test set:

      • N/A. Adjudication methods like "2+1" or "3+1" are relevant for clinical studies or studies involving human interpretation of data (e.g., image analysis). For bench testing of a guidewire, results are quantitative or qualitative against predetermined engineering specifications, and "adjudication" typically refers to the pass/fail criteria established for each test.
    4. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

      • No. This is a physical medical device (guidewire), not an AI/software device that assists human readers/clinicians, so an MRMC study is not applicable.
    5. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

      • N/A. This is a physical medical device, not an algorithm. The "standalone" performance is the device's performance in the bench tests, independent of its use in a patient for the initial testing and FDA clearance.
    6. The type of ground truth used:

      • The "ground truth" for each test is based on pre-established engineering specifications, material standards (e.g., ISO), and performance characteristics derived from the predicate device and FDA guidance documents. For example, the "ground truth" for tensile strength is a minimum force value, for biocompatibility it's compliance with ISO 10993, and for dimensions it's adherence to specified tolerances.
    7. The sample size for the training set:

      • N/A. This is a physical medical device undergoing non-clinical testing, not a machine learning model, so there is no "training set."
    8. How the ground truth for the training set was established:

      • N/A. As there is no training set for a physical device, this question is not applicable.
    Ask a Question

    Ask a specific question about this device

    K Number
    K243383
    Device Name
    CHIKAI Nexus 014
    Date Cleared
    2025-04-21

    (172 days)

    Product Code
    Regulation Number
    870.1330
    Reference & Predicate Devices
    Why did this record match?
    510k Summary Text (Full-text Search) :

    California 92780

    Re: K243383
    Trade/Device Name: CHIKAI Nexus 014
    Regulation Number: 21 CFR 870.1330
    katsuhiko.fujimura@asahi-intecc.com

    TRADE NAME: CHIKAI Nexus 014

    DEVICE CLASSIFICATION: Class II, 21 CFR § 870.1330
    K163426 | K160659 | K191714 |
    | Manufacturer | ASAHI INTECC | | | | |
    | Regulation Number | 21 CFR 870.1330

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    This guide wire is intended to be used in the neuro vasculature to facilitate the placement and exchange of therapeutic devices such as cerebral catheters during intravascular therapy. This guide wire is intended for use only in the neuro vasculature.

    Device Description

    The CHIKAI Nexus 014 consists of a stainless-steel tapered core wire, with an inner coil and outer coil made of radiopaque Pt-Ni alloy and stainless-steel. Surrounding the inner coil and the distal core wire is a radiopaque Pt-Ni alloy and stainless-steel outer coil. The radiopaque distal tip enables the user to view the position of the tip under X-ray fluoroscopy. Outer and inner coils are soldered to the tapered core wire with Ag-Sn solder. A similar coil design is used with other ASAHI guide wires, such as the predicate CHIKAI black (K141751).

    Additionally, the CHIKAI Nexus 014 employs hydrophilic, polyurethane, PTFE, and silicone coatings which are used in the predicate and reference devices.

    The nominal outer diameter of the CHIKAI Nexus 014 is 0.36 mm (0.014 inch). The device is available in two lengths: 215 cm and 300 cm.

    Both sizes are available with straight, pre-shape, and angled designs.

    AI/ML Overview

    Acceptance Criteria and Study to Prove Device Meets Criteria for CHIKAI Nexus 014

    The provided FDA 510(k) clearance letter and summary for the CHIKAI Nexus 014 detail performance testing to demonstrate substantial equivalence to predicate devices, rather than a clinical study evaluating human reader performance with an AI device. The CHIKAI Nexus 014 is a catheter guide wire, a physical device, and not an AI/software as a medical device (SaMD) that would involve human readers or AI-assisted interpretation of images.

    Therefore, many of the requested categories (e.g., number of experts, adjudication methods, MRMC study, standalone performance for an algorithm, ground truth for training/test sets for an AI) are not applicable to the information provided for this specific physical medical device.

    The acceptance criteria and performance data provided relate to the physical and functional characteristics of the guide wire.


    1. Table of Acceptance Criteria and Reported Device Performance

    Device: CHIKAI Nexus 014 (Catheter Guide Wire)

    Test CategoryAcceptance Criteria (Implicit from "All samples met the acceptance criteria" or "Results/Conclusion")Reported Device Performance
    Dimensional VerificationMeasurements (overall length, outer diameters, coating lengths) within specified tolerances.All samples met the acceptance criteria.
    Simulated UseSuccessful advancement and removal through a clinically relevant model according to IFU; proper and safe operation.All samples met the acceptance criteria.
    Visual InspectionAbsence of foreign matter, damage, or excessive droplet-like residue of coating liquid.All samples met the acceptance criteria.
    Tensile StrengthWithstand specified tensile forces without failure.All samples met the acceptance criteria.
    Torque StrengthWithstand specified torque forces without failure when distal end is stationary.All samples met the acceptance criteria.
    TorqueabilityAchieve specified distal end rotation/response upon proximal end rotation through a rotational response model.All samples met the acceptance criteria.
    Coating IntegrityCoating adhesion and integrity maintained after winding around a test jig; no visual damage.All samples met the acceptance criteria.
    Coating Integrity/Particulate EvaluationComparable to predicate device in terms of particulate generation during simulated use in a glass vascular model.The results were comparable to the predicate device.
    LubricityMeet specified slipping resistance against a catheter.All samples met the acceptance criteria.
    Corrosion ResistanceNo signs of corrosion after immersion in sodium chloride solution.All samples met the acceptance criteria.
    Kink ResistanceNo damage (microscopic) after multiple bends of clinically relevant radii.All samples met the acceptance criteria.
    Tip FlexibilityBending loads at various points of the distal end meet specified criteria.All samples met the acceptance criteria.
    RadiopacityDistal end clearly visible under fluoroscopy.All samples met the acceptance criteria.

    Study Proving Device Meets Acceptance Criteria

    The study described is a series of non-clinical, bench-top performance tests, and biocompatibility assessments comparing the CHIKAI Nexus 014 to its predicate device, the ASAHI Neurovascular Guide Wire CHIKAI black (K141751), and other reference devices. The goal was to establish substantial equivalence for a physical medical device.

    2. Sample Size Used for the Test Set and Data Provenance:

    • Sample Size: The document repeatedly states "All samples met the acceptance criteria" or "The samples are visually inspected..." or "All samples are performed/tested..." indicating that multiple samples were used for each test, but the exact number of samples for each specific test is not explicitly quantified (e.g., "n=X").
    • Data Provenance: The data are from laboratory bench testing and biocompatibility testing. Given "ASAHI INTECC CO., LTD." is based in "Aichi, Japan," it's highly probable the testing was conducted prospectively at their R&D facilities or authorized contract labs, likely in Japan or globally recognized testing centers.

    3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications of those Experts:

    • N/A for this type of device. This device is a physical guidewire, not an AI or imaging device requiring expert interpretation for ground truth. "Ground truth" here refers to the measured physical and performance characteristics determined by established test methods and standards.

    4. Adjudication Method for the Test Set:

    • N/A. Adjudication methods like 2+1 or 3+1 are typical for subjective human assessments (e.g., in reading medical images) to establish a consensus ground truth. For objective physical device testing, results are typically quantitative measurements against predefined specifications.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done, and what was the effect size of how much human readers improve with AI vs without AI assistance:

    • N/A. This is a physical guidewire, not an AI-powered diagnostic or assistive tool for human readers. No human clinical performance study (like an MRMC) involving "human readers" or "AI assistance" was conducted as part of this 510(k) submission.

    6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done:

    • N/A. This applies to AI algorithms. The performance of this guide wire was evaluated through bench testing against established physical and mechanical properties.

    7. The Type of Ground Truth Used:

    • For performance testing: Objective engineering measurements and observations against predetermined specifications and widely accepted industry standards (e.g., "All samples met the acceptance criteria" implying specific numerical or qualitative thresholds were achieved). The FDA guidance document "Coronary, Peripheral, and Neurovascular Guidewires - Performance Tests and Recommended Labeling," October 2019, served as a reference for the test methods, which implicitly defines the "ground truth" of what constitutes acceptable performance for such a device.
    • For biocompatibility testing: Established laboratory test results compared against ISO 10993 series standards and control samples.

    8. The Sample Size for the Training Set:

    • N/A. This is not an AI/ML device that requires a training set. The "design" of the device is based on engineering principles and comparison to existing predicate devices, not data-driven machine learning.

    9. How the Ground Truth for the Training Set Was Established:

    • N/A. As above, no training set or ground truth in the context of an AI/ML model was established for this device.
    Ask a Question

    Ask a specific question about this device

    K Number
    K243733
    Device Name
    SION blue PLUS
    Date Cleared
    2025-04-02

    (119 days)

    Product Code
    Regulation Number
    870.1330
    Reference & Predicate Devices
    Why did this record match?
    510k Summary Text (Full-text Search) :

    , California 92780

    Re: K243733
    Trade/Device Name: SION blue PLUS
    Regulation Number: 21 CFR 870.1330
    , California 92780

    Re: K243733
    Trade/Device Name: SION blue PLUS
    Regulation Number: 21 CFR 870.1330
    | Manufacturer | ASAHI INTECC | ASAHI INTECC | ASAHI INTECC |
    | Regulation Number | 21 CFR§ 870.1330
    | 21 CFR§ 870.1330 | 21 CFR§ 870.1330 |
    | Regulation Name | Catheter, Guide Wire | Catheter, Guide
    | 21 CFR§ 870.1330 | 21 CFR§ 870.1330 |
    | Regulation Name | Catheter, Guide Wire | Catheter, Guide

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    PCI Guide Wires are intended to facilitate the placement of balloon dilatation catheters during percutaneous coronary intervention (PCI) and percutaneous transluminal angioplasty (PTA). The PCI Guide Wires are not to be used in the neurovasculature.

    Device Description

    SION blue PLUS is steerable guide wire with a maximum diameter of 0.36mm (0.014 inches) and available in 190cm and 300cm lengths. The guide wire is constructed from stainless-steel core wire with platinum-nickel and stainless-steel coils. The coils assembly consists of an inner coil and an outer coil, as well as a safety wire which is soldered to the inner and outer coils and the core wire. The distal end of the guide wire has a 3cm radiopaque tip to achieve visibility and is available in a straight, Pre-shape and J-shape to bend with the vessel curve. A silicone and hydrophilic coatings are applied to the distal portion of SION blue PLUS. The proximal sections of the SION blue PLUS are coated with PTFE. The extension wire is connected to the end of the guide wire outside of the body for 190cm model.

    AI/ML Overview

    This FDA 510(k) clearance letter pertains to a conventional medical device (a guide wire), not an AI/ML-enabled device. Therefore, the information requested about acceptance criteria, study data, ground truth establishment, expert adjudication, and MRMC studies, which are typical for AI/ML device clearances, is not present in this document.

    The document details non-clinical bench testing for the SION blue PLUS guide wire to demonstrate its substantial equivalence to a predicate device.

    Here's a breakdown of the available information based on your request, highlighting what is included and what is not applicable to this type of device submission:

    1. A table of acceptance criteria and the reported device performance

    The document states that "The in vitro bench tests demonstrated the SION blue PLUS met all acceptance criteria and performed similarly to the predicate/reference devices." However, it does not provide a specific table of acceptance criteria with corresponding performance data. It lists the types of tests performed:

    • Dimensional Verification
    • Visual Inspection
    • Simulated Use/Human body Phantom
    • Tensile Strength (including Tip Pull test)
    • Torque Strength
    • Torqueability
    • Coating Integrity
    • Particulate Evaluation
    • Lubricity/Catheter Compatibility
    • Corrosion Resistance
    • Kink Resistance
    • Tip Flexibility
    • Radiopacity

    2. Sample sizes used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    • Sample Sizes: Not specified. For physical device bench testing, sample sizes are typically determined by engineering standards and statistical power analysis for the specific tests (e.g., n=3, n=5, n=10 per test) but are not typically detailed in 510(k) summaries as they are for clinical or AI/ML studies.
    • Data Provenance: Not applicable in the sense of clinical study data provenance (country, retrospective/prospective). The data comes from in vitro bench testing conducted by the manufacturer, ASAHI INTECC CO., LTD. (Global Headquarters and R&D Center in Seto, Aichi, Japan).

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    • Not applicable. This information is relevant for AI/ML device evaluations where human experts establish ground truth for image interpretation or diagnosis. For a physical device like a guide wire, performance is measured against engineering specifications and industry standards, not against "expert ground truth" in a clinical diagnostic sense.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    • Not applicable. Adjudication methods are typically used in clinical or AI/ML studies where multiple human readers might disagree on a diagnosis or interpretation, requiring a consensus mechanism. This is not relevant for bench testing of a physical medical device.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    • Not applicable. An MRMC study is a type of clinical study used to evaluate the diagnostic performance of a device (often AI-enabled) by comparing multiple readers' interpretations across multiple cases, sometimes with and without AI assistance. This device is a physical catheter guide wire, not a diagnostic or AI-assisted device.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    • Not applicable. This refers to the evaluation of an AI algorithm's performance independent of human interaction. Since SION blue PLUS is a physical guide wire, this concept does not apply.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

    • Not applicable in the clinical sense. For this device, "ground truth" equates to the pre-defined engineering specifications, performance standards, and material properties that the device must meet during bench testing. For example, a tensile strength test would have an acceptance criterion (e.g., withstand X Newtons of force), and the device's measured performance against this criterion is the "truth."

    8. The sample size for the training set

    • Not applicable. This concept pertains to AI/ML devices where a "training set" is used to develop the algorithm. This device is a physical instrument, not an AI.

    9. How the ground truth for the training set was established

    • Not applicable. See point 8.

    In summary, the provided 510(k) clearance letter is for a physical medical device. The types of detailed information requested regarding AI/ML study design, ground truth, expert adjudication, and sample sizes for training/test sets are not relevant to this specific device submission and therefore are not present in the document. The document focuses on demonstrating substantial equivalence through non-clinical bench testing of the device's physical and material properties.

    Ask a Question

    Ask a specific question about this device

    Why did this record match?
    510k Summary Text (Full-text Search) :

    24 Guidewire; Zoom Wire 14 Guidewire; Aristotle Colossus Guidewire
    Regulation Number: 21 CFR 870.1330
    14 Guidewire |
    | Common Name: | Guidewire |
    | Classification Name | Catheter Guide Wire per 21 CFR 870.1330
    18 Guidewire |
    | Common Name: | Guidewire |
    | Classification Name | Catheter Guide Wire per 21 CFR 870.1330
    24 Guidewire |
    | Common Name: | Guidewire |
    | Classification Name | Catheter Guide Wire per 21 CFR 870.1330
    14 Guidewire |
    | Common Name: | Guidewire |
    | Classification Name | Catheter Guide Wire per 21 CFR 870.1330

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Aristotle 14 Guidewire is intended for general vascular use within the neuro and peripheral vasculatures to introduce and position catheters and other interventional devices. The guidewire is not intended for use in the coronary vasculature.

    The Aristotle 18 Guidewire is intended for general vascular use within the neuro and peripheral vasculatures to introduce and position catheters and other interventional devices. The guidewire is not intended for use in the coronary vasculature.

    The Aristotle 24 Guidewire is intended for general vascular use within the neuro and peripheral vasculatures to introduce and position catheters and other interventional devices. The guidewire is not intended for use in the coronary vasculature.

    The Zoom Wire 14 Guidewire is intended for general vascular use within the neuro and peripheral vasculatures to introduce and position catheters and other interventional devices. The guidewire is not intended for use in the coronary vasculature.

    The Aristotle Colossus Guidewire is intended for general vascular use within the neuro and peripheral vasculatures to introduce and position catheters and other interventional devices. The guidewire is not intended for use in the coronary vasculature.

    Device Description

    The Scientia Vascular's Aristotle 14, Aristotle 18, Aristotle 24, Zoom Wire 14, and Aristotle Colossus Guidewires ("the guidewires") are steerable guidewires with a shapeable tip to aid in accessing the neuro and peripheral vasculatures. The guidewires are supplied sterile, for single use only.

    The distal portion of each guidewire's tip includes a radiopaque platinum wire marker coil to facilitate fluoroscopic visualization. All guidewires have a hydrophilic polymer coating on the distal portion and a polytetrafluoroethylene (PTFE) coating on the proximal portion to reduce friction during manipulation in vessels.

    The guidewires are provided with an accessory kit consisting of an introducer (to aid with the insertion of the guidewire into a catheter hub and/or a hemostasis valve), a torque device (to attach to the proximal portion of the guidewire to facilitate gripping and manipulation of the guidewire during use), and a shaping mandrel (to aid in shaping the flexible tip of the guidewire). These accessory devices are included to facilitate use of the guidewires and are not intended to contact the patient's body.

    AI/ML Overview

    This document describes a 510(k) clearance for several guidewire devices, not an AI/ML medical device. Therefore, the information requested about acceptance criteria and studies related to AI/ML device performance (such as sample sizes for test/training sets, expert ground truth establishment, MRMC studies, etc.) is not applicable and cannot be extracted from this document.

    The document primarily focuses on establishing substantial equivalence for physical medical devices—guidewires—by comparing their technological characteristics and performing non-clinical performance tests (biocompatibility, sterilization, bench performance).

    Here's a breakdown of why this document does not contain the requested information about an AI/ML device:

    • Device Type: The clearance is for "Aristotle 14 Guidewire; Aristotle 18 Guidewire; Aristotle 24 Guidewire; Zoom Wire 14 Guidewire; Aristotle Colossus Guidewire" which are physical catheter guide wires. These are traditional medical devices, not software or AI/ML devices.
    • Study Types: The studies described are "Biocompatibility," "Sterilization," and "Bench Performance Testing" (Visual Inspection, Coating Integrity, Corrosion Resistance, Agent Compatibility, Simulated Use, Particulate). These are mechanical and biological tests relevant to physical devices, not algorithmic performance.
    • Missing Information: All points you requested regarding AI/ML performance are absent because they are not relevant to this type of device clearance:
      • No acceptance criteria for AI performance metrics (e.g., sensitivity, specificity, AUC).
      • No test or training set sample sizes for AI data.
      • No information about experts establishing ground truth for images or data used by an AI.
      • No adjudication methods.
      • No MRMC comparative effectiveness studies.
      • No standalone algorithm performance.
      • No ground truth type related to AI (e.g., imaging labels, pathology).

    The document explicitly states:

    • "No animal testing was deemed necessary to support the substantial equivalence of the subject devices."
    • "No clinical testing was deemed necessary to support the substantial equivalence of the subject devices."

    This indicates that the clearance relies on non-clinical data and comparison to predicate devices, which is common for guidewires, especially when the changes are minor (like a different PTFE coating).

    Ask a Question

    Ask a specific question about this device

    K Number
    K241962
    Device Name
    Crossloop
    Date Cleared
    2025-03-27

    (267 days)

    Product Code
    Regulation Number
    870.1330
    Reference & Predicate Devices
    Why did this record match?
    510k Summary Text (Full-text Search) :

    Suite 212 Tustin, California 92780

    Re: K241962

    Trade/Device Name: Crossloop Regulation Number: 21 CFR 870.1330
    | | | | | |
    | Device Classification: | Class 2 per 21 CFR § 870.1330
    |
    | Regulation
    Number | 21 CFR 870.1330

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    This product is intended to direct a catheter to the desired anatomical vasculature (excluding coronary and cerebral vessels).

    Device Description

    The CROSSLOOP is steerable guide wire with a maximum diameter of 0.018 inches (0.46 mm) and available in various lengths of 200 cm, 235 cm and 300 cm. The CROSSLOOP consists of a stainless-steel (SUS) core wire with a Pt-Ni alloy coil, soldered with Au-Sn solder. The distal tip is loop-structured. The coil is radiopaque to achieve visibility and can be made to bend easily to accommodate vessel tortuosity. A silicone and hydrophilic coating are applied to the distal portion of the guide wire. A hydrophobic coating (PTFE) is applied to proximal portion. The purpose of these surface coatings is to provide lubricity when the guide wire is passed through percutaneous catheters.

    AI/ML Overview

    The provided text describes a 510(k) premarket notification for a medical device called "CROSSLOOP," a catheter guide wire. This type of submission focuses on demonstrating substantial equivalence to a legally marketed predicate device, rather than proving a new device's absolute safety and effectiveness through extensive clinical trials as would be required for a PMA (Premarket Approval Application).

    Therefore, the information you're asking for, such as "acceptance criteria and the study that proves the device meets the acceptance criteria," "sample size for the test set," "number of experts used to establish ground truth," "adjudication method," "MRMC comparative effectiveness study," "standalone performance," "type of ground truth," "sample size for the training set," and "how the ground truth for the training set was established," are not directly applicable or available in this document.

    This 510(k) summary focuses on non-clinical testing/performance data and biocompatibility to demonstrate that the device is substantially equivalent to existing cleared devices, not that it independently proves clinical efficacy or diagnostic accuracy (as would be the case for an AI-powered diagnostic device, for example).

    Here's how to address your points based on the provided text:

    1. A table of acceptance criteria and the reported device performance

    The document states: "The in vitro bench tests demonstrated the CROSSLOOP met all acceptance criteria and performed similarly to the predicate and reference devices." However, it does not provide a table specifying the precise acceptance criteria or quantitative performance results for each test. It only lists the types of non-clinical tests performed:

    Acceptance Criteria (Implied)Reported Device Performance
    Tensile StrengthMet all acceptance criteria and performed similarly to predicate/reference devices.
    Torque StrengthMet all acceptance criteria and performed similarly to predicate/reference devices.
    TorqueabilityMet all acceptance criteria and performed similarly to predicate/reference devices.
    Tip FlexibilityMet all acceptance criteria and performed similarly to predicate/reference devices.
    Coating IntegrityMet all acceptance criteria and performed similarly to predicate/reference devices.
    Catheter CompatibilityMet all acceptance criteria and performed similarly to predicate/reference devices.
    Visual InspectionMet all acceptance criteria and performed similarly to predicate/reference devices.
    Corrosion ResistanceMet all acceptance criteria and performed similarly to predicate/reference devices.
    Kink ResistanceMet all acceptance criteria and performed similarly to predicate/reference devices.
    Radio DetectabilityMet all acceptance criteria and performed similarly to predicate/reference devices.
    Dimensional VerificationMet all acceptance criteria and performed similarly to predicate/reference devices.
    Coating Integrity / Acute Particulate CharacterizationMet all acceptance criteria and performed similarly to predicate/reference devices.
    Biocompatibility (Cytotoxicity, Sensitization, Intracutaneous Irritation, Systemic Toxicity, USP Rabbit Pyrogen, Hemolysis, Partial Thromboplastin Time, In Vivo Thrombogenicity, Sc5b-9 Complement Activation)Performed to assess biocompatibility. (Implied to have met criteria, as stated in the overall conclusion).

    2. Sample sized used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    • Sample Size: Not specified for any of the non-clinical tests.
    • Data Provenance: The tests are "in vitro bench tests" and "biocompatibility" tests. The manufacturer is ASAHI INTECC CO., LTD. whose global headquarters and R&D center are in Japan. The specific country where the testing was performed is not explicitly stated, but it's likely linked to the manufacturer's R&D facilities. These are laboratory tests, not clinical data, so terms like "retrospective" or "prospective" don't apply in the common clinical sense.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

    This information is not applicable to this type of device and submission. "Ground truth" established by experts (like radiologists for imaging devices) is relevant for AI/ML-powered diagnostic tools or clinical studies, neither of which are described here. The "ground truth" for the non-clinical tests would be the established engineering and materials science standards and specifications.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    This is not applicable. Adjudication methods are typically used in clinical trials involving human observation or interpretation, especially with imaging reads. For bench testing, results are typically quantitative measurements compared against predefined engineering specifications.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    This is not applicable. An MRMC study is designed for AI-assisted diagnostic devices to assess the impact of AI on human reader performance. The "CROSSLOOP" is a physical medical device (catheter guide wire), not an AI/ML software tool.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    This is not applicable. This question refers to the performance of an AI algorithm in isolation. The CROSSLOOP is a physical medical device.

    7. The type of ground truth used (expert concensus, pathology, outcomes data, etc)

    For a physical device like a guide wire, the "ground truth" for non-clinical testing refers to pre-defined engineering specifications, material properties, and performance standards (e.g., tensile strength required to withstand certain forces, flexibility within a specified range, biocompatibility per ISO standards). This is not derived from expert consensus on clinical cases, pathology, or outcomes data, but rather from material science and mechanical engineering principles.

    8. The sample size for the training set

    This is not applicable. "Training set" refers to data used to train AI/ML models. This device is not an AI/ML product.

    9. How the ground truth for the training set was established

    This is not applicable for the same reason as point 8.

    Ask a Question

    Ask a specific question about this device

    K Number
    K242597
    Date Cleared
    2025-03-05

    (187 days)

    Product Code
    Regulation Number
    870.1330
    Reference & Predicate Devices
    Why did this record match?
    510k Summary Text (Full-text Search) :

    Tustin, California 92780

    Re: K242597

    Trade/Device Name: CROSSLEAD 0.018inch Regulation Number: 21 CFR 870.1330
    | |
    | Device Classification: | Class 2 per 21 CFR § 870.1330

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    This product is intended to direct a catheter to the desired anatomical vasculature (excluding coronary and cerebral vessels) during diagnostic or interventional procedures.

    Device Description

    The CROSSLEAD 0.018inch is steerable guide wire with a maximum diameter of 0.46mm (0.018inches) and available in various lengths of 100 cm, 200 cm, 235 cm and 300 cm. The quide wire consists of a hybrid nitinol (Ni-Ti) and stainlesssteel (SUS) core wire with a stainless steel inner coil, a stainless steel and Platinum-Nickel (Pt-Ni) outer coil. The coil is radiopaque to achieve visibility and can be made to bend easily to accommodate vessel tortuosity. A hydrophilic and polyurethane coating are applied to the distal portion of the guide wire. A hydrophobic coating (PTFE) is applied to proximal portion. The purpose of these surface coatings is to provide lubricity when the guide wire is passed through percutaneous catheters.

    AI/ML Overview

    The provided text is a 510(k) Premarket Notification from the FDA regarding the "CROSSLEAD 0.018inch" guide wire. This document primarily focuses on establishing substantial equivalence to predicate devices through non-clinical bench testing and biocompatibility assessments. It does not contain information about a study involving human-in-the-loop performance, expert ground truth establishment for a test set, or a multi-reader multi-case (MRMC) comparative effectiveness study, which are typically associated with the evaluation of artificial intelligence (AI) or software-based medical devices.

    Therefore, I cannot fulfill your request for information related to:

    • A table of acceptance criteria and reported device performance in the context of an AI/software study. The document lists non-clinical tests and states they met acceptance criteria, but these are for the physical device, not an AI.
    • Sample size used for a test set (in the context of an AI study).
    • Number of experts and their qualifications for establishing ground truth.
    • Adjudication method for a test set.
    • MRMC comparative effectiveness study or related effect sizes.
    • Standalone (algorithm only) performance.
    • Type of ground truth used (expert consensus, pathology, outcomes data).
    • Sample size for the training set.
    • How ground truth for the training set was established.

    The document discusses the physical characteristics and performance of a medical guide wire, not a software or AI device that would require such studies. The "non-clinical testing/performance data" section refers to bench tests for the physical guide wire's properties (e.g., tensile strength, torqueability, coating integrity, biocompatibility), not an algorithm's performance on a dataset.

    In summary, the provided text does not describe the kind of study you are asking about, which typically pertains to the evaluation of AI/software in medical devices.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 50