Search Results
Found 1 results
510(k) Data Aggregation
(137 days)
The product is intended for use in male, female, and pediatric patients requiring bladder drainage as determined by their physician. This device is indicated for those individuals unable to promote a natural urine flow or for those individuals who have a significant volume of residual urine following a natural bladder-voiding episode.
SureCath Set is a sterile, single-use hydrophilic coated intermittent catheter with integrated urine bag. An ampoule of saline solution is provided inside the urine bag. This saline solution from the ampoule is poured over the catheter to activate the coating before use.
Here's an analysis of the provided text regarding acceptance criteria and the study:
The document (K242049) is a Traditional 510(k) Summary for the Coloplast SureCath Set, a sterile, single-use hydrophilic coated intermittent catheter with an integrated urine bag. It demonstrates substantial equivalence to predicate devices. The information provided primarily focuses on non-clinical performance testing rather than human clinical studies involving AI assistance.
Based on the provided text, the device SureCath Set is a medical device, not an AI/ML powered device. Therefore, a multi-reader multi-case (MRMC) comparative effectiveness study, effect size of human readers with/without AI, or standalone algorithm performance are not applicable.
Here's a breakdown of the requested information based on the provided document:
1. A table of acceptance criteria and the reported device performance
The document broadly states that "Bench performance testing and usability testing were conducted to verify the proposed subject devices met the pre-determined acceptance criteria per specified requirements." It also mentions various ISO and ASTM standards to which the device was tested. However, it does not explicitly list specific quantitative acceptance criteria or detailed reported performance values for each test. It only states that the device "met the pre-determined acceptance criteria."
Implicit Acceptance Criteria (based on standards and test descriptions):
Acceptance Criteria Category | Reported Device Performance |
---|---|
Biocompatibility | Met acceptance criteria per ISO 10993 series for cytotoxicity, irritation/intracutaneous reactivity, sensitization, material mediated pyrogenicity, acute systemic toxicity, and subacute toxicity. |
Catheter Performance (ISO 20696, ASTM F623-19, Coloplast Test Methods) | Met acceptance criteria for: |
- Strength of connection between catheter and urine bag.
- Aseptic insertion (non-touch approach) of catheter through urine bag.
- Water amount of 20 ml in ampoule must activate coating on catheter.
- Friction of the coated catheter after 30 seconds activation time.
- Tearing and emptying of the urine bag.
- Volume of the urine bag.
- Strength of the urine bag. |
| Usability | Met acceptance criteria per EN/IEC 62366-1. |
| Packaging Integrity | Met acceptance criteria per ISO 11607-1, DS EN ISO 20696, ASTM D4169-22 for maintenance of sterile barrier and impact of transportation. |
| Shelf Life/Aging Stability | Met acceptance criteria after accelerated aging per ASTM F1980-21, demonstrating stability for the defined shelf life (24 months at time of submission, planned 36 months). |
2. Sample size used for the test set and the data provenance
The document does not specify the exact sample sizes used for each non-clinical test. It states "Testing was performed on final, finished, and sterilized devices." The provenance of the data is from Coloplast A/S, Denmark. The tests are non-clinical, bench performance tests, so the concepts of "retrospective or prospective" human data don't directly apply here.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts
This information is not applicable to this type of non-clinical device testing. There are no "experts" establishing a "ground truth" for a test set in the way it would be for an AI/ML clinical study (e.g., radiologists interpreting images). The "ground truth" for non-clinical tests is based on objective measurements against established engineering and biological standards.
4. Adjudication method (e.g. 2+1, 3+1, none) for the test set
This information is not applicable. Adjudication methods are typically used in clinical studies where expert consensus is needed to establish ground truth for ambiguous cases. Non-clinical bench tests rely on objective measurements against defined standards.
5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
No, an MRMC comparative effectiveness study was not conducted. The device is a physical medical device (intermittent catheter), not an AI-powered diagnostic or assistive tool. Therefore, there is no "AI assistance" for human readers to improve upon.
6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done
No, a standalone algorithm-only performance study was not done. The device is not an algorithm or software requiring human-in-the-loop performance.
7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)
The "ground truth" for the non-clinical tests is based on:
- Compliance with recognized international standards (ISO, ASTM, EN/IEC).
- Pre-determined acceptance criteria defined by Coloplast for specific performance aspects (e.g., strength, friction, volume, activation).
- Objective measurements of physical and chemical properties.
8. The sample size for the training set
This information is not applicable. "Training set" is a concept for AI/ML models. This document describes testing for a physical medical device.
9. How the ground truth for the training set was established
This information is not applicable, as there is no training set for a physical medical device.
Ask a specific question about this device
Page 1 of 1