Search Results
Found 416 results
510(k) Data Aggregation
(268 days)
FMF
Monoject Syringes with Luer tip are intended for single use by health care professionals for general purpose fluid aspiration and and/or injection.
Not Found
The provided FDA 510(k) clearance letter pertains to Monoject™ Syringes, which are physical medical devices (piston syringes) and not an AI/ML software device. Therefore, the concepts of acceptance criteria for algorithm performance, study methodology for AI models, sample sizes for training/test sets, ground truth establishment, expert adjudication, or MRMC studies are not applicable to this document.
The letter indicates that the device has received substantial equivalence to legally marketed predicate devices, meaning it meets the safety and effectiveness requirements for its intended use, which is general purpose fluid aspiration and/or injection by healthcare professionals.
To answer your request, if this were an AI/ML device, the information would typically be found in the 510(k) submission summary or a separate clinical study report, neither of which is present in the provided FDA clearance letter. The letter itself is a notice of clearance, not a detailed technical report of the studies performed.
Therefore, I cannot extract the requested information from the provided document as it does not contain details about an AI/ML study or its performance criteria.
Ask a specific question about this device
(90 days)
FMF
The BD Plastipak™ Syringe is intended for use by health care professionals for general purpose fluid aspiration/injection.
The BD Plastipak™ Syringe is a three-piece, single use, sterile, hypodermic syringe with a 6% (Luer) male connector in 20 mL and 50 mL eccentric luer slip tip configurations. The BD Plastipak™ Syringe is intended for use by health care professionals for general purpose fluid aspiration/injection. The syringe assembly consists of a lubricated polypropylene barrel with a graduated scale, a lubricated synthetic rubber stopper and a polypropylene plunger rod. The plunger rod is pulled back to aspirate fluids or depressed to inject or expel fluids. The syringe barrel incorporates a male 6% (Luer) connector which is connectable to a compatible female 6% (Luer) connector. The BD Plastipak™ Syringe is provided sterile by Ethylene Oxide Gas (ETO) sterilization method.
The provided text is a 510(k) Clearance Letter for a medical device (BD Plastipak™ Syringe). It details the device's characteristics, intended use, and comparison to a predicate device. However, it does not describe an AI/ML-driven medical device or a study involving human readers or expert consensus for ground truth establishment.
The document discusses bench performance testing and biocompatibility tests for a physical device (syringe), not a software or AI-based diagnostic tool. Therefore, many of the requested criteria (like sample size for test/training set, number of experts, adjudication method, MRMC study, standalone performance, ground truth types) are not applicable to this specific submission.
Despite the irrelevance of some questions to the provided document, I will structure the answer based on the questions asked, indicating "Not Applicable" or providing the information that is present in the document.
Here's an analysis of the provided 510(k) clearance letter in the context of the requested information about acceptance criteria and study data:
This 510(k) clearance letter pertains to a physical medical device, the BD Plastipak™ Syringe, not an AI/ML-driven diagnostic or image analysis tool. As such, many of the typical acceptance criteria and study methodologies applicable to AI models (e.g., ground truth established by experts, MRMC studies, training/test set sizes for algorithms, human reader improvement with AI assistance) are not relevant or described in this document.
The "study" referenced in the document primarily consists of non-clinical performance and biocompatibility testing to demonstrate the substantial equivalence of the new syringe (with a changed barrel resin) to a previously cleared predicate syringe.
1. Table of Acceptance Criteria and Reported Device Performance
The document states, "The subject device met all the predetermined acceptance criteria for the above listed performance and biocompatibility tests." However, it does not explicitly list the quantitative acceptance criteria or the specific numerical performance results for each test. It only lists the tests performed and the standards they adhere to.
Test Category | Test | Purpose (as described) | Acceptance Criteria (Not Explicitly Stated Quantitatively) | Reported Performance (Not Explicitly Stated Quantitatively) |
---|---|---|---|---|
Performance Testing | Breakout Force | ISO 7886-1:2017- Performance evaluation of force to operate the piston. | Met predetermined criteria (implied by "met all") | Subject device performed equivalent to predicate (implied) |
Sustaining Force | ISO 7886-1:2017- Evaluation of force to operate the piston. | Met predetermined criteria (implied by "met all") | Subject device performed equivalent to predicate (implied) | |
Leakage Past Stopper | ISO 7886-1:2017- Evaluation of Freedom from air and liquid leakage past plunger stopper. | Met predetermined criteria (implied by "met all") | Subject device performed equivalent to predicate (implied) | |
Volumetric Accuracy | ISO 7886-1:2017- Evaluation of Volumetric Accuracy | Met predetermined criteria (implied by "met all") | Subject device performed equivalent to predicate (implied) | |
Dead Space | ISO 7886-1:2017- Evaluation of residual volume | Met predetermined criteria (implied by "met all") | Subject device performed equivalent to predicate (implied) | |
Luer Leakage | ISO 80369-7:2021 - Evaluation of the luer fittings for leakage | Met predetermined criteria (implied by "met all") | Subject device performed equivalent to predicate (implied) | |
Stress Cracking | ISO 80369-7:2021- Evaluation of the luer fittings for stress cracking | Met predetermined criteria (implied by "met all") | Subject device performed equivalent to predicate (implied) | |
Resistance to separation from axial load | ISO 80369-7:2021- Evaluation of the luer fitting for separation when subjected to axial force. | Met predetermined criteria (implied by "met all") | Subject device performed equivalent to predicate (implied) | |
Biocompatibility Testing | Cytotoxicity | ISO 10993-5:2009- Biological evaluation of medical devices - Part 5: Tests for in vitro cytotoxicity | Met predetermined criteria (implied by "met all") | Subject device performed equivalent to predicate (implied) |
Sensitization | ISO 10993-10:2021- Biological Evaluation of Medical Devices - Part 10: Tests for Skin Sensitization | Met predetermined criteria (implied by "met all") | Subject device performed equivalent to predicate (implied) | |
Intracutaneous Reactivity | ISO 10993-23:2021- Biological evaluation of medical devices- Test for Irritation | Met predetermined criteria (implied by "met all") | Subject device performed equivalent to predicate (implied) | |
Acute Systemic Toxicity | ISO 10993-11:2017- Biological evaluation of medical devices — Part 11: Tests for systemic toxicity | Met predetermined criteria (implied by "met all") | Subject device performed equivalent to predicate (implied) | |
Material Mediated Pyrogenicity | ISO 10993-11:2017- Biological evaluation of medical devices — Part 11: Tests for systemic toxicity USP43-NF38 Pyrogen Test (USP Rabbit Test) | Met predetermined criteria (implied by "met all") | Subject device performed equivalent to predicate (implied) | |
Hemocompatibility | ISO 10993-4:2017- Biological evaluation of medical devices Part 4: Selection of tests for interactions with blood ASTM F756-17- Standard Practice for Hemolytic Properties of Materials | Met predetermined criteria (implied by "met all") | Subject device performed equivalent to predicate (implied) |
2. Sample Size Used for the Test Set and Data Provenance
- Sample Size for Test Set: Not specified in the provided document. The tests are "bench performance testing" on various syringe units.
- Data Provenance: Not specified, but generally, bench testing for physical devices is conducted in a controlled lab environment by the manufacturer. It is non-clinical.
- Retrospective or Prospective: Not applicable for this type of physical device testing.
3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts
- Not Applicable. Ground truth for this physical device testing is established through standardized laboratory test methods and measurements against international or internal specifications, not by human experts interpreting clinical data.
4. Adjudication Method for the Test Set
- Not Applicable. (See point 3)
5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
- No. This is not an AI/ML device. Therefore, no MRMC study or assessment of human reader improvement with AI assistance was conducted or is relevant.
6. If a Standalone (i.e. algorithm only without human-in-the loop performance) was done
- No. This is not an AI/ML device. Therefore, no standalone algorithm performance was assessed. The performance tests are for the physical syringe itself.
7. The Type of Ground Truth Used
- The "ground truth" for this device's performance is based on measurements against established engineering specifications and international standards (e.g., ISO, ASTM, USP) for physical and material properties (e.g., force, leakage, volume accuracy, biocompatibility reactions). It is not based on expert consensus, pathology, or outcomes data in the clinical sense.
8. The Sample Size for the Training Set
- Not Applicable. This is not an AI/ML device. There is no concept of a "training set" in the context of the reported non-clinical bench testing for a physical syringe.
9. How the Ground Truth for the Training Set was Established
- Not Applicable. (See point 8)
Ask a specific question about this device
(270 days)
FMF
FORCYTE Autograft Harvest Kit is intended to harvest cancellous bone and bone marrow for combination with commercially available allograft or bone void fillers for filling bony voids or gaps that are not intrinsic to the stability of the bony structure.
The Subject device is a kit comprised of a bone and marrow harvesting needle, collection jar, suction tubing, drill-tip guide pin, and bone void filler hydration chamber. The FORCYTE system harvests autologous bone and marrow for use as a bone void filler.
Based on the provided FDA 510(k) clearance letter for the Forcyte Autograft Harvest Kit (K243407), the device is a Gastroenterology-Urology Biopsy Instrument. The letter explicitly states that "No clinical evaluations were conducted" for this device. This means that the 510(k) clearance relies solely on non-clinical (bench) testing, and therefore, it is not possible to provide acceptance criteria or a study description related to AI algorithm performance or diagnostic accuracy, human reader performance, or multi-reader multi-case (MRMC) comparative effectiveness studies from this document.
The information requested in the prompt (AI algorithm performance data, expert ground truth, MRMC studies, etc.) is typically associated with software as a medical device (SaMD) or AI/ML-enabled devices that perform diagnostic or prognostic functions based on image analysis, pattern recognition, etc. The Forcyte Autograft Harvest Kit, however, is a physical instrument for harvesting bone and bone marrow.
Therefore, many of the requested items are not applicable to the information provided in this 510(k) clearance letter. I will address the relevant points based on the document's content and explicitly state when information is not available.
Analysis of the Forcyte Autograft Harvest Kit (K243407) based on the Provided 510(k) Clearance Letter:
The Forcyte Autograft Harvest Kit (K243407) is a physical medical device, not an AI/ML-enabled device or software medical device. The 510(k) clearance for this device was based on non-clinical (bench) testing and substantial equivalence to a predicate device, not on clinical performance studies involving human readers or AI algorithms.
Therefore, most of the requested information regarding AI acceptance criteria, clinical study design (sample size, expert ground truth, MRMC), and training data is not applicable to this specific device and the provided document.
Here's an attempt to address each point based on the provided text, clarifying what information is present and what is absent:
1. A table of acceptance criteria and the reported device performance
Based on the document, the acceptance criteria and performance data are related to the physical characteristics and functionality of the harvesting kit, not diagnostic accuracy.
Acceptance Criteria and Reported Device Performance (Non-Clinical)
Performance Metric (Type of Test) | Acceptance Criteria (Implicit) | Reported Device Performance (Implicit) |
---|---|---|
Simulated Use Testing | Satisfactory Operation | All testing had satisfactory results. |
Tensile Testing | Within Engineering Specifications | All testing had satisfactory results. |
Compression Testing | Within Engineering Specifications | All testing had satisfactory results. |
Torque Testing | Within Engineering Specifications | All testing had satisfactory results. |
Flexural Testing | Within Engineering Specifications | All testing had satisfactory results. |
Vacuum Testing | Achieve/Maintain Required Vacuum | All testing had satisfactory results. |
Biocompatibility | Meet ISO 10993 Standards | Indicating safety and effectiveness. |
Sterilization/Shelf Life | Maintain Sterility & Integrity | All testing had satisfactory results. |
Note: The document states "All testing had satisfactory results" but does not provide specific numerical acceptance criteria or detailed numerical results for each test. The acceptance criteria are implicitly that the device performs as intended and meets safety requirements.
2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)
- Sample Size for Test Set: Not specified in terms of number of units or test runs for the bench tests. The document only states "bench testing including simulated use testing, tensile, compression, torque, flexural, and vacuum testing were conducted."
- Data Provenance (Country of Origin, Retrospective/Prospective): Not specified. As these are bench tests, the concept of "retrospective" or "prospective" data collection from patients is not applicable. The tests would have been performed in a laboratory setting. No geographical origin of the testing data is mentioned.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)
- Not Applicable. This device is a biopsy instrument, not an imaging or diagnostic AI device requiring expert interpretation for ground truth establishment. Bench testing does not typically involve "experts" establishing a "ground truth" in the diagnostic sense; rather, it involves engineers and quality control personnel assessing physical performance against specifications.
4. Adjudication method (e.g. 2+1, 3+1, none) for the test set
- Not Applicable. As no expert interpretation or diagnostic "ground truth" was established, there was no need for an adjudication method. Bench testing results are typically assessed against pre-defined engineering and material specifications.
5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
- No. The document explicitly states: "No clinical evaluations were conducted." Therefore, no MRMC study, human reader improvement, or AI assistance was involved in the clearance. This type of study is irrelevant for a physical bone harvesting instrument.
6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done
- Not Applicable. This is a physical medical device, not an algorithm or software. There is no "standalone" algorithm performance to evaluate.
7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)
- Not Applicable in the diagnostic sense. For a physical device, "ground truth" is established by manufacturing specifications, engineering design, and material properties. The device's performance is measured against these established physical and functional parameters (e.g., tensile strength, proper vacuum seal, successful harvesting of material, biocompatibility).
8. The sample size for the training set
- Not Applicable. This is not an AI/ML device that requires a "training set" of data.
9. How the ground truth for the training set was established
- Not Applicable. As there is no training set for an AI/ML algorithm, no ground truth was established for it.
Ask a specific question about this device
(25 days)
FMF
The ProfojectTM Disposable Syringe and the ProfojectTM Disposable Syringe with Needle are intended to be used for medical purposes to inject fluid into or withdraw fluid from body.
The ProfojectTM Disposable Syringe is available in a sterile state. It is a single-use device comprising a polypropylene barrel with a Luer slip or Luer lock nozzle, a polypropylene plunger, and a polyisoprene rubber plunger stopper. Syringes are available in various volume specifications, ranging from 1mL to 60mL. The syringe barrel is printed with a legible graduated scale in milliliters. The syringe is compatible with U.S. legally marketed devices featuring Luer slip or Luer lock connectors that comply with ISO 80369-7. The syringe is intended for manual use only. The plunger of the syringe is pulled back to aspirate fluids or depressed to inject or expel fluids.
The ProfojectTM Disposable Syringe with Needle is supplied in a sterile state. It is a single-use device comprising a Luer slip or Luer lock syringe and a hypodermic needle. The hypodermic needle consists of a stainless steel needle tube, a polypropylene needle hub, and a polypropylene needle cap. The hypodermic needle is available in variety of combinations of needle gauge (18G, 20G, 21G, 22G, 23G, 25G, 26G, 27G) and needle length (15mm, 20mm, 25mm, 30mm, 38mm). Needle hubs are color-coded to identify the gauge of the needle tube. The Disposable Syringe with Needle is available in various combinations of syringe volume and needle size. It is intended for manual use only. The plunger of the syringe is pulled back to aspirate fluids or depressed to inject or expel fluids.
The provided FDA 510(k) clearance letter and summary for the Profoject™ Disposable Syringe and Profoject™ Disposable Syringe with Needle focuses on establishing substantial equivalence to a predicate device (K211211) through non-clinical testing. It does not describe a study involving an AI component, human readers, or a ground truth established in the way typical for AI/ML medical devices.
Therefore, the requested information regarding acceptance criteria, device performance, sample size for test sets, data provenance, expert ground truth establishment, adjudication methods, MRMC studies, standalone performance, and training set details are not applicable as this is a traditional medical device submission without an AI/ML component.
Instead, the submission relies on a comparison of technological characteristics and performance testing against recognized international standards for syringes and needles to demonstrate substantial equivalence.
Here's a summary based on the provided document, addressing the relevant points and noting where information is not applicable due to the nature of the device:
Acceptance Criteria and Study Details for Profoject™ Disposable Syringe
1. Table of Acceptance Criteria and Reported Device Performance
The acceptance criteria for this device are based on compliance with a comprehensive set of international standards rather than specific performance metrics directly stated in a table for each criterion. The "reported device performance" is indicated by the statement that the tests met the requirements of these standards.
Acceptance Criterion (Compliance with Standard) | Reported Device Performance |
---|---|
Biocompatibility: | |
ISO 10993-1:2018 (Cytotoxicity, Skin sensitization, Intracutaneous reactivity, Acute systemic toxicity, Pyrogen test, Hemocompatibility) | Met the requirements in the standards. |
USP Particulate Matter in Injections | Met the USP acceptance criteria. |
Performance Testing: | |
ISO 7886-1:2017 (Sterile hypodermic syringes for single use - Part 1: Syringes for manual use) | Complied with the standard. |
ISO 7864:2016 (Sterile hypodermic needles for single use - Requirements and test methods) | Complied with the standard. |
ISO 9626:2016 (Stainless steel needle tubing for the manufacture of medical devices - Requirements and test methods) | Complied with the standard. |
ISO 80369-7:2021 (Small-bore connectors for liquids and gases in healthcare applications - Part 7: Connectors for intravascular or hypodermic applications) | Complied with the standard. |
ISO 6009:2016 (Hypodermic needles for single use — Colour coding for identification) | Complied with the standard. |
Sterilization, Package, and Shelf Life Testing: | |
ISO 11135:2014 (Ethylene Oxide Sterilization Validation) | Validated, establishing routine control and monitoring parameters. |
ISO 11737-1:2018+AMD1:2021 (Bioburden test) | Test results demonstrated compliance. |
ISO 11737-2:2019 (Sterility test) | Test results demonstrated compliance. |
ISO 10993-7:2008+AMD1:2019 (EO and ECH residue test) | Test results demonstrated compliance. |
USP (Bacterial Endotoxins Test) | Test results demonstrated compliance. |
ISTA 3A:2018 (Simulated distribution testing for packaging) | Test results demonstrated compliance. |
ASTM F88/F88M-23 (Seal strength) | Packaging found acceptable. |
ASTM F1929-23 (Dye penetration) | Packaging found acceptable. |
ASTM F1886/F1886M-16 (Visual inspection) | Packaging found acceptable. |
ASTM F1140/F1140M-13 (Reapproved 2020)e1 (Internal pressurization) | Packaging found acceptable. |
USP (Sterility test for packaging) | Packaging found acceptable. |
ASTM F1980-21 (Accelerated aging for shelf life) | Device shelf-life determined to be 5 years. |
Labeling Compliance | Met the requirements of 21 CFR Part 801. |
2. Sample Size Used for the Test Set and Data Provenance
The document does not specify exact sample sizes for each test. It states that "the following tests have been performed" and that "the testing results demonstrate that the proposed device complies with the applicable standards requirements." For biocompatibility, performance, sterilization, package, and shelf-life testing, samples of the device were used as required by the specific standards.
- Sample Size: Not explicitly stated as a single number for a "test set." Samples were drawn and tested according to the requirements of each individual standard (e.g., ISO, ASTM, USP).
- Data Provenance: The tests were conducted by the manufacturer (CMT Health Pte. Ltd., based in Singapore) or their designated testing facilities to demonstrate compliance with international standards. This is prospective testing performed on manufactured devices.
3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications
N/A. This device is a traditional medical device (syringes) and not an AI/ML device requiring expert-established ground truth for a diagnostic test. The "ground truth" here is compliance with established engineering and biological safety standards.
4. Adjudication Method for the Test Set
N/A. Not applicable to engineering performance and biocompatibility testing of a physical device. Test results are compared directly against the quantitative or qualitative pass/fail criteria defined within the referenced international standards.
5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done
N/A. This is a traditional medical device (syringe), not an AI-assisted diagnostic or interpretative device. Therefore, MRMC studies are not relevant.
6. If a Standalone (i.e. algorithm only without human-in-the-loop performance) was done
N/A. This device does not involve an AI algorithm. Its performance is evaluated through bench testing against established standards.
7. The Type of Ground Truth Used
The "ground truth" for this device's acceptance is defined by the objective, measurable requirements and pass/fail criteria specified within the referenced international and national standards (ISO, ASTM, USP) for its physical, chemical, and biological properties. This includes:
- Engineering specifications (e.g., dimensions, force required for operation, leak integrity).
- Material safety (biocompatibility, leachable substances).
- Sterility assurance.
- Packaging integrity.
- Shelf-life stability.
8. The Sample Size for the Training Set
N/A. This is a traditional medical device and does not involve machine learning algorithms requiring a training set.
9. How the Ground Truth for the Training Set Was Established
N/A. No training set is applicable for this device.
Ask a specific question about this device
(114 days)
FMF
The 3DOSE 1ml Syringe is intended for use by health care professionals for general purpose fluid aspiration/ injection.
The 3DOSE 1ml Syringe is a sterile, single-use, disposable and non-reusable manual syringe which is intended for injection of fluids into the body. The syringe can be used for pulling up any mixture, and after operating the locking handle for accurate dosing according to the ml/unit scale on the plunger.
The 3DOSE 1ml Syringe is used like a classic syringe, but includes a feature that provides tactile and audible feedback to the user during dosing.
Models: 3DOSE 1ml Syringe (125 GREEN)
3DOSE 1ml Syringe (100 ORANGE)
The provided FDA 510(k) clearance letter for the 3DOSE 1ml Syringe does not detail any study involving human readers, AI, or diagnostic performance metrics typically associated with AI/ML-based medical devices. This clearance is for a piston syringe, a physical medical device, not an AI/ML software as a medical device (SaMD).
Therefore, I cannot provide information for the following points as they are not applicable to this device and its clearance documentation:
- A table of acceptance criteria and the reported device performance (in terms of diagnostic metrics like sensitivity, specificity, AUC)
- Sample sizes used for the test set and data provenance (for diagnostic performance of an algorithm)
- Number of experts used to establish the ground truth for the test set and their qualifications
- Adjudication method for the test set
- If a multi-reader multi-case (MRMC) comparative effectiveness study was done
- If a standalone (algorithm only) performance study was done
- The type of ground truth used (expert consensus, pathology, outcomes data, etc.)
- The sample size for the training set
- How the ground truth for the training set was established
However, I can extract information related to the device's technical specifications, performance standards, and the non-clinical testing performed to establish its safety and effectiveness relative to a predicate device.
Here's the summary of acceptance criteria and the study that proves the device meets them, based on the provided document, specifically for a physical medical device (syringe):
1. Acceptance Criteria and Reported Device Performance (Non-Diagnostic)
For a physical device like a syringe, acceptance criteria relate to its physical properties, functionality, and compliance with recognized standards. The "performance" reported is its compliance with these standards, indicating mechanical functionality, material safety, and sterility.
Category | Acceptance Criteria (Standard Compliance) | Reported Device Performance |
---|---|---|
Functional Performance | - ISO 7886-1:2017: Sterile Hypodermic syringes for single use - Part 1: Syringes for manual use |
- ISO 80369-7:2016: Small-bore connectors for liquids and gases in healthcare applications - Part 7: Connectors for intravascular or hypodermic applications
- ISO 80369-20:2015: Small-bore connectors for liquids and gases in healthcare applications - Part 20: Common test methods | - Complies with ISO 7886-1:2017
- Complies with ISO 80369-7:2016
- Complies with ISO 80369-20:2015
(Demonstrated through "Functional Performance Testing") |
| Sterility | - ISO 11135: Sterilization of health-care products - Ethylene oxide - ISO 11607-1: Packaging for terminally sterilized medical devices - Part 1: Requirements for materials, sterile barrier systems and packaging systems
- ISO 11607-2: Packaging for terminally sterilized medical devices - Part 2: Validation requirements for forming, sealing and assembly processes
- ISO 10993-7: Biological evaluation of medical devices - Part 7: Ethylene oxide sterilization residuals | - Device is sterile, single-use, disposable.
- Sterilized using Ethylene Oxide.
- Validated shelf-life of 5 years.
(Demonstrated through "Sterility" testing per listed ISO standards) |
| Biocompatibility | - ISO 10993-1: Biological Evaluation of Medical Devices - Part 1: Evaluation and Testing Within a Risk Management Process (including Cytotoxicity, Sensitization, Irritation, Acute Systemic Toxicity, Pyrogenicity, Hemocompatibility) - ISO 10993-4: Tests for interactions with blood
- ISO 10993-5: Tests for in vitro cytotoxicity
- ISO 10993-10: Tests for irritation and skin sensitization
- ISO 10993-11: Tests for systemic toxicity
- USP Particulate Matter In Injections | - Biocompatibility evaluation conducted in accordance with ISO 10993-1 and FDA recognized standards.
- Syringe of testing included Cytotoxicity, Sensitization, Irritation, Acute Systemic Toxicity, Pyrogenicity, Hemocompatibility.
- All a-forementioned tests passed, demonstrating the device is biocompatible. (Demonstrated through "Biocompatibility Testing") |
| Chemical Properties | - ISO 7886-1 (Section for chemical properties: Limits for acidity or alkalinity, Limits for extractable metals) | - Chemical performances inspection based on ISO 7886-1. - Results conform to ISO 7886-1 for limits for acidity or alkalinity and extractable metals. (Demonstrated through "Chemical properties" testing) |
| Material/Design Integrity | - Functionality of the "audible/tactile feature" (ratcheting dose mechanism) - No negative effect on overall device performance due to new features (locking arm, plastic part, audible/tactile feedback) | - In-process quality control tests performed to confirm the audio notification mechanism works correctly and does not have a negative effect on overall performance.
- Performance testing conducted for the ratcheting lock mechanism (locking arm, plastic part) to ensure it does not raise new questions of safety or effectiveness. (Demonstrated through implicit in-house testing/QC) |
2. Sample Sizes used for the Test Set and Data Provenance
The document does not specify exact sample sizes for each non-clinical test (e.g., how many syringes were tested for extractable metals or sterility). It generally states "The collective results of the nonclinical testing demonstrate..." This implies a sufficient number of samples were tested to meet the statistical requirements of the respective ISO standards.
- Data Provenance: The tests would have been performed in a laboratory setting, likely in Turkey (Bimed Teknik Aletler Sanayi & Ticaret A.S. is based in Istanbul, Turkey) or by accredited labs contracted by them. These are prospective tests performed specifically for this 510(k) submission.
3. Number of Experts used to establish the ground truth for the test set and the qualifications of those experts
This concept is not applicable here as there is no "ground truth" in the clinical diagnostic sense. The "truth" for device functionality, sterility, and biocompatibility is established by compliance with internationally recognized consensus standards (ISO, USP) and physical/chemical measurements by qualified laboratory personnel.
4. Adjudication method for the test set
Not applicable. This is not a study assessing human or AI diagnostic performance.
5. If a multi reader multi case (MRMC) comparative effectiveness study was done
No. This type of study is for evaluating observer performance, typically in medical imaging or diagnostics, and is not relevant to a physical medical device like a syringe.
6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done
No. This device is not an algorithm or software. It is a manually operated piston syringe.
7. The type of ground truth used
For this device, the "ground truth" is defined by the specifications and limits set by the referenced international standards (e.g., ISO 7886-1, ISO 10993 series). For example:
- Dimensional accuracy of volume: defined by ISO 7886-1 tolerances.
- Sterility: defined by ISO 11135 and ISO 11607 series (e.g., SAL of 10^-6).
- Biocompatibility: Defined by the absence of adverse biological reactions as per ISO 10993 series tests.
- Chemical properties: Defined by limits for acidity/alkalinity and extractable metals in ISO 7886-1.
8. The sample size for the training set
Not applicable. This is a physical device, not an AI/ML model that requires training data.
9. How the ground truth for the training set was established
Not applicable.
Ask a specific question about this device
(26 days)
FMF
GPS Advanced and GPS Advanced Cannula are intended for use to deliver FIBERGRAFT™ BG Putty GPS to a bone grafting site.
GPS Advanced is a sterile, single use dispenser for the delivery of FIBERGRAFT™ BG Putty GPS bone graft substitute. The dispenser has a ratcheted plunger that advances with each squeeze of its handle. The 5 cc GPS Advanced Cannula ("GPS Advanced Cannula") is available individually packaged and can be attached to the dispenser.
GPS Advanced is provided sterile via irradiation and GPS Advanced Cannula is provided sterile via ethylene oxide.
The GPS Advanced Cannula is provided empty and needs to be filled with FIBERGRAFT™ BG Putty GPS prior to use. FIBERGRAFT™ BG Putty GPS is a bioactive synthetic bone graft substitute in putty format made from 45S5 bioactive glass and has been previously cleared under K170306 (FIBERGRAFT™ BG Putty – Bone Graft Substitute) and K222276 (CONDUIT™ Cages and FIBERGRAFT™ BG Putty). FIBERGRAFT™ BG Putty GPS is not included in the GPS Advanced dispensing system.
Based on the provided FDA 510(k) clearance letter for "GPS Advanced and GPS Advanced Cannula," it's important to understand that this device is a piston syringe used for delivering a bone graft substitute. As such, the "acceptance criteria" and "study that proves the device meets acceptance criteria" are focused on the mechanical and functional performance of the syringe as a medical device, rather than the performance of an AI/ML algorithm or diagnostic tool.
Therefore, many of the typical questions asked about AI/ML device studies (like MRMC studies, standalone algorithm performance, AI training/test sets, ground truth establishment for AI) are not applicable to this type of device.
Here's a breakdown of the acceptance criteria and performance data as inferred from the document:
1. Table of Acceptance Criteria and Reported Device Performance
Acceptance Criteria Category | Specific Criteria (Inferred) | Reported Device Performance (Inferred) |
---|---|---|
Biological Safety | Biocompatibility and safety of materials in contact with body fluids/tissue. | Met through Biological Risk Assessment. |
Sterilization Efficacy | Device is sterile. | Met through Sterilization Validation (Irradiation for GPS Advanced, Ethylene Oxide for GPS Advanced Cannula). |
Packaging Integrity | Packaging maintains sterility and protects device integrity over shelf-life. | Met through Packaging and Shelf-Life Testing. |
Functional Performance | Device delivers the intended substance (FIBERGRAFT™ BG Putty GPS) effectively and reliably to the bone grafting site. | Met through functional evaluation (likely as part of Packaging and Shelf-Life Testing, and implicitly through similarity to predicate). The ratcheted plunger advances with each squeeze, allowing delivery. |
Shelf-Life Stability | Device maintains its functional and sterile properties for its specified shelf-life. | Met through Shelf-Life Testing. |
Design & Material Conformity | Design features, materials, and principle of operation are consistent with predicate devices and suitable for intended use. | "The technological characteristics, including design, materials, principle of operation and performance as well as intended use of GPS Advanced and GPS Advanced Cannula are consistent with those of the primary predicate and reference device." |
Explanation of Inferences:
- The document explicitly states that "The performance data for the subject devices consists of the following evaluations: Biological Risk Assessment, Sterilization Validation, Packaging and Shelf-Life Testing, including functional evaluation." These evaluations implicitly define the acceptance criteria for those aspects.
- The statement about "consistent with those of the primary predicate and reference device" implies that the new device must perform comparably to previously cleared devices in terms of its core technological characteristics and function.
2. Sample Size Used for the Test Set and Data Provenance
This information is not explicitly detailed in the provided 510(k) summary. For a mechanical device like a syringe, "sample size" typically refers to the number of units tested for various validations (e.g., how many syringes were subjected to sterilization, packaging, or functional tests). The document does not specify these numbers.
- Data Provenance: Not applicable in the context of clinical data for AI/ML. The "data" here refers to test results from various engineering and biological validations of the device itself.
3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications of Those Experts
This concept is not applicable to this device. "Ground truth" in this context is established through engineering specifications, validated test methods (e.g., sterility testing, material biocompatibility standards), and direct functional testing, not by expert consensus on clinical images or interpretations.
4. Adjudication Method for the Test Set
This concept is not applicable to this device, as there's no "interpretation" or "diagnosis" being made that requires human adjudication. Device performance is determined by objective measurements and validation against established standards.
5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
No, an MRMC comparative effectiveness study was not done. This type of study is relevant for AI-assisted diagnostic devices where human readers (e.g., radiologists) interpret images with and without AI assistance. The GPS Advanced and GPS Advanced Cannula is a delivery device, not an AI diagnostic tool.
6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was done
No, a standalone algorithm performance study was not done. This concept applies to AI/ML algorithms that perform a task independently (e.g., detecting a lesion on an image). The GPS Advanced and GPS Advanced Cannula is a physical medical device, not an algorithm.
7. The Type of Ground Truth Used
The "ground truth" for this device's performance is based on objective scientific and engineering standards and tests:
- Sterility Validation: Demonstration that the device meets established sterility assurance levels (e.g., SAL of 10^-6).
- Biocompatibility Standards: Compliance with ISO 10993 series for medical device biocompatibility.
- Functional Specifications: Ability to successfully deliver the specified volume of material, maintain pressure, resist leakage, and operate smoothly.
- Packaging Integrity Standards: Ability of the packaging to maintain sterility and physical integrity under specified conditions.
- Shelf-Life Parameters: Demonstration that the device retains its functional and sterile properties over its specified shelf-life.
8. The Sample Size for the Training Set
This concept is not applicable for this device. "Training set" refers to data used to train AI/ML algorithms. The GPS Advanced and GPS Advanced Cannula is a mechanical device, not an AI/ML system.
9. How the Ground Truth for the Training Set Was Established
This concept is not applicable for this device, as there is no "training set" in the AI/ML sense.
Ask a specific question about this device
(85 days)
FMF
GO-PEN® is a reusable insulin pen intended for the self-injection of NovoLog® (U-100, insulin aspart for injection) available in 10 ml vials.
GO-PEN® is an assembly of a durable pen-shaped dosing unit with a special disposable sterile reservoir that fits into the dosing unit. The GO-PEN® device can only be used with NovoLog® U-100 insulin aspart for injection. The user is intended to fill up the disposable reservoir with 1.3 mL of insulin from a 10 ml vial and mount it on the dosing unit. The GO-PEN® device allows the user to set the desired dose from 1 to 60 Units in 1 Unit increments using the dial feature. Once the GO-PEN® disposable reservoir is filled, it can be used over a period of up to 3 days at room temperature (max. 30°C/86°F) for multiple injections of precise doses of insulin. When the reservoir is empty or the time has been exceeded, the reservoir must be discarded, and a new sterile reservoir is used for the next filling. GO-PEN® is used with standard single-use hypodermic needles with a luer lock which are supplied separately by the user.
Based on the provided FDA 510(k) Clearance Letter for the GO-PEN®, here's an analysis of the acceptance criteria and study information:
The document focuses on non-clinical (bench) testing to demonstrate substantial equivalence, rather than clinical studies with human participants. Therefore, several of the requested categories related to human studies (e.g., sample size for test sets/training sets in a clinical context, expert ground truth, MRMC studies, standalone performance with humans) are not applicable or not provided in this regulatory submission.
1. Table of Acceptance Criteria and Reported Device Performance
The acceptance criteria are primarily defined by adherence to ISO 11608-1:2022 and specific design specifications. The reported performance indicates that the device passed these requirements.
Acceptance Criteria Category | Specific Criteria (from ISO 11608-1:2022 unless otherwise specified) | Reported Device Performance |
---|---|---|
Dose Accuracy & Injection Force | After various preconditioning: |
- Standard Atmosphere
- Cool Atmosphere
- Warm Atmosphere
- Free Fall
- Dry Heat
- Cold Storage
- Damp Heat
- Cyclical (Temperature cycling)
- Vibration
- In-Use Lifetime Test (Life-cycle test for 5 years)
- Simulated Transport
- Aging to Expiration (Functional Stability) | Passed dose accuracy requirements after all specified preconditioning conditions. |
| Last Dose Accuracy | As defined in ISO 11608-1:2022 | Passed. |
| Fluid Leakage (into dosing mechanism) | As defined in ISO 11608-1:2022 | Passed. |
| Leakage (during injection) | As defined in ISO 11608-1:2022 | Passed. |
| Visible Particles in Reservoir | As defined in ISO 11608-1:2022 | Passed. |
| Break Loose and Glide Force | As defined in ISO 11608-1:2022 | Passed. |
| Biocompatibility (Durable Pen – intact skin contact) | Meets ISO 10993-1:2018; evaluated for in-vitro cytotoxicity. Materials excluded from Attachment G of FDA guidance also evaluated. | Exterior materials tested and evaluated accordingly. |
| Biocompatibility (Reservoir – externally communicating) | Meets ISO 10993-1:2018; evaluated for in-vitro cytotoxicity, skin irritation, material mediated pyrogenicity, and sensitization. | Materials of construction evaluated accordingly. |
| Usability/Human Factors | Critical tasks (filling/labeling reservoirs, injecting, handling) passed without mistakes; user interface did not adversely impact performance or usability. | Confirmed all critical tasks passed without mistakes. |
| Luer Testing | As defined in ISO 80369-7 and 80369-20 for: - Dimensions
- Fluid leakage
- Sub-atmospheric pressure leakage
- Stress cracking
- Resistance to separation from axial load
- Resistance to separation from unscrewing
- Resistance to overriding | Passed all luer lock testing requirements. |
| Insufficient Remaining Dose (IRD) Feature | Complies with ISO 11608-1:2022, §5.6(k) method (4). | Complies (pen displays un-injected dose after emptying reservoir). |
| Life Cycle (Durable Pen) | 3 years once use begins (bench tested for 5 years). | Tested for a 5-year life-cycle; stated use life is 3 years once use begins. |
Study Proving Device Meets Acceptance Criteria
The study described is a non-clinical bench testing program.
-
Sample Size used for the Test Set and the Data Provenance:
- Sample Size: The document does not specify the exact number of GO-PEN® devices or components used for each bench test. Regulatory submissions typically use statistically relevant sample sizes for each test to achieve confidence in the results, though specific numbers aren't listed here.
- Data Provenance: The testing was conducted by or on behalf of GO-Pen ApS, located in Lyngby, Denmark. The nature of these are prospective tests performed specifically to support this regulatory submission, based on established international standards.
-
Number of Experts Used to Establish the Ground Truth for the Test Set and the Qualifications of Those Experts:
- This is not applicable as the "ground truth" for bench testing involves adherence to objective physical and engineering standards (e.g., ISO, precise measurements) rather than expert human interpretation of data. The "experts" would be the engineers and technicians conducting the tests and interpreting results against the standard's pass/fail criteria.
-
Adjudication Method for the Test Set:
- Not applicable in the sense of expert consensus for clinical data. Adjudication is automated based on whether measured parameters fall within the acceptable ranges defined by the ISO standards.
-
If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:
- No. This is a medical device (insulin pen), not an AI/imaging diagnostic device. Therefore, MRMC studies and AI performance metrics are not applicable.
-
If a standalone (i.e. algorithm only without human-in-the loop performance) was done:
- This is not applicable as the GO-PEN® is a mechanical medical device, not an algorithm or AI. Its "standalone performance" is its mechanical function as tested in the bench studies.
-
The Type of Ground Truth Used:
- For bench testing, the ground truth is defined by international consensus standards (e.g., ISO 11608-1:2022, ISO 80369-7, ISO 80369-20, ISO 10993-1:2018) and the device's own pre-defined engineering specifications derived from these standards. Metrics like dose accuracy are objectively measured against these established limits.
-
The Sample Size for the Training Set:
- Not applicable. As a mechanical medical device undergoing non-clinical bench testing, there is no "training set" in the context of machine learning or AI.
-
How the Ground Truth for the Training Set was Established:
- Not applicable, as there is no "training set."
Ask a specific question about this device
(78 days)
FMF
The GraftGun Universal Graft Delivery System (GDS®) is intended to be used for the delivery of hydrated allograft, autograft, or synthetic bone graft material to an orthopedic surgical site, via direct visualization or minimally invasive.
The GraftGun Universal Graft Delivery System (GDS®) consists of a sterile, single-use, disposable device, and non-sterile reusable Adapter Tips intended for the delivery of prepared allograft, autograft, or synthetic bone graft material to an orthopedic surgical site. The GraftGun Universal Graft Delivery System (GDS®) includes the GraftGun Dispensing Unit, Loading Device, Graft Tubes, and Adapter Tips.
The provided text describes a medical device, the GraftGun Universal Graft Delivery System (GDS®), and its 510(k) clearance process. However, this document does not contain information about the acceptance criteria or a study proving the device meets those criteria in the context of an Artificial Intelligence (AI) or machine learning (ML) system.
The device described is a piston syringe used for delivering bone graft material. The performance testing mentioned (Compatibility Assessments, Sterilization Validations, Cleaning Validations, Biocompatibility Assessment, Package Performance, Shelf-Life Testing) are standard tests for mechanical medical devices to ensure their safety, functionality, and sterility. They are not related to AI/ML performance metrics such as sensitivity, specificity, AUC, or reader studies.
Therefore, I cannot fulfill your request for the specific details about AI/ML acceptance criteria and study data based on the provided text. The document is for a non-AI/ML device.
Ask a specific question about this device
(130 days)
FMF
The Insulin Syringes are intended for subcutaneous injection of U100 insulin or U40 insulin.
The Insulin Syringes are 0.3 mL, 0.5 mL, or 1.0 mL synnges designed for subcutaneous injection of a desired dose of insulin. The Insulin Syringes consist of a graduated barrel, plunger rod and needle/hub assembly. It is available in various needle gauge sizes (25G, 26G, 27G, 28G, 29G, 30G and 31G) and various needle lengths (6mm). The Insulin Syringes are sterile, single use, and non-toxic. These devices operate on the principles of a piston syringe.
The provided document is an FDA 510(k) premarket notification for "Insulin Syringes," which are medical devices, not AI/ML-driven software with specific acceptance criteria related to a test set, expert ground truth, or MRMC studies. The document describes a traditional medical device (insulin syringes) and its substantial equivalence to previously cleared predicate devices.
Therefore, the requested information regarding acceptance criteria, study design (test set, ground truth, expert adjudication, MRMC, standalone performance, training set details) for an AI/ML device is not available in this document. The document primarily focuses on:
- Device Description: Physical characteristics (volume, needle gauge/length, materials), intended use (subcutaneous injection of U100/U40 insulin).
- Modifications: Updates to packaging options and syringe sizes compared to the predicate.
- Substantial Equivalence Justification: Comparison of indications for use, technological characteristics (design, materials, sterilization) with predicate devices.
- Performance Testing: References to ISO standards for sterile single-use syringes, hypodermic syringes, stainless steel needle tubing, hypodermic needles, and sterilization validation (e.g., ISO 8537, ISO 7886-1, ISO 9626, ISO 7864, ISO 11135, ASTM F1980-16). These are standard tests for physical and functional properties of syringes, not AI/ML performance metrics.
The "Performance Testing" section refers to engineering and manufacturing standards for physical device performance, not for the performance of an algorithm or software. There are no mentions of AI/ML, image data, diagnostic accuracy, or human reader studies.
Ask a specific question about this device
(179 days)
FMF
Single use Needle is intended for use with syringes and injection devices for general purpose fluid injection/aspiration.
Single use Syringe with or without Needle is intended to be used for medical purposes to inject fluid into or withdraw from body.
Membrane Filter Syringe is intended to be used for medical purposes to inject fluid into or withdraw from body. The filter operates when injecting the human body to remove foreign substances from the drug solutions.
Single use Syringe with or without Needle consist of Syringe, Cap, Needle and Hub and Blister paper. Blister paper function to sustain sterilization of the product. The Cap function to protect the needle. This device is single use. This device is provided EO Sterilization, and the syringes of Luer Slip and Luer Lock.
These devices are injecting the medicine with syringes. Biocompatibility testing and performance testing has been completed to support the substantial equivalence of the device.
This document describes a Special 510(k) submission for modifications to the "Sofjec Single use Syringe with or without Needle." The submission primarily addresses the addition of a new 60mL syringe volume and the inclusion of "without needle" options for existing syringe models.
1. Table of Acceptance Criteria and Reported Device Performance
The acceptance criteria are based on compliance with harmonized international standards for sterile hypodermic syringes and needles. The document does not provide a specific table of numerical acceptance criteria with corresponding performance data. Instead, it states that the device "Complied with ISO 7886-1, ISO 7864, ISO 9626."
- ISO 7886-1: Sterile hypodermic syringes for single use — Part 1: Syringes for manual use. This standard details requirements for material, design, performance, and packaging. Key performance aspects typically include:
- Absence of air and liquid leakage past the plunger during aspiration and injection.
- Dimensional accuracy (e.g., barrel capacity, graduations).
- Smoothness of plunger movement.
- Resistance to breakage.
- Sterility.
- ISO 7864: Sterile hypodermic needles for single use. This standard specifies requirements for sterile hypodermic needles. Key performance aspects typically include:
- Needle dimensions (e.g., gauge, length).
- Integrity of the bond between needle and hub.
- Resistance to breakage and corrosion.
- Patency.
- Sterility.
- ISO 9626: Stainless steel needle tubing for the manufacture of medical devices — Requirements and test methods. This standard specifies requirements for the tubing used to make hypodermic needles.
The reported device performance, in summary, is that the modified devices complied with these standards. This implies that the devices met all specified requirements within these standards.
2. Sample Size Used for the Test Set and Data Provenance
The document does not specify the exact sample size used for performance testing. It states that "Successful performance testing has been completed using FDA recognized standards." The data provenance is implied to be from the manufacturer, HLB Life Science Co., Ltd., located in the Republic of Korea. The testing conducted is for a "Special 510(k) Premarket Notification" for device modification, which typically involves prospective testing of the modified device to ensure it meets established safety and performance criteria.
3. Number of Experts Used to Establish the Ground Truth for the Test Set and Their Qualifications
Not applicable. This is a submission for medical devices (syringes and needles), not an AI/Software as a Medical Device (SaMD). The "ground truth" here is defined by compliance with established international standards for physical device performance and sterility. The evaluation is done through laboratory testing, not expert interpretation of data like in image analysis.
4. Adjudication Method for the Test Set
Not applicable. As this is not an AI/SaMD submission, there is no expert adjudication process for ground truth establishment. Performance is determined by objective laboratory tests against pre-defined criteria in the ISO standards.
5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study
Not applicable. This is not an AI/SaMD product.
6. Standalone (Algorithm Only Without Human-in-the-Loop Performance) Study
Not applicable. This is not an AI/SaMD product.
7. Type of Ground Truth Used
The "ground truth" for this device's performance is established by compliance with recognized international consensus standards (ISO 7886-1, ISO 7864, ISO 9626). This means the device's physical and functional properties were tested against the objective specifications outlined in these standards.
8. Sample Size for the Training Set
Not applicable. This is not an AI/SaMD product; therefore, there is no "training set" in the context of machine learning.
9. How the Ground Truth for the Training Set was Established
Not applicable. As there is no training set, the concept of establishing ground truth for it does not apply.
Ask a specific question about this device
Page 1 of 42