K Number
K250658

Validate with FDA (Live)

Date Cleared
2025-04-03

(29 days)

Product Code
Regulation Number
880.5570
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

SureFine Pen Needle is intended for use with a pen injector for the subcutaneous injection of insulin.

Device Description

SureFine Pen Needle is designed for use with a pen injector for the subcutaneous injection of insulin. The pen needle consists of a needle, hub, and shield assembly. Blister paper covers primary container. The primary container maintains sterility of the needle because primary container covers the hub and needle cap with blister paper sealed on the opening hole of primary container.

The needle hub can be connected screwed onto the pen. The needle shield is intended to provide physical protection to the needle tube.

SureFine Pen Needle is offered in various gauges sizes (29G, 30G, 31G, 32G) and Lengths (4mm, 6mm, 8mm, 12.7mm). These are sterile (EO gas sterilization), non-toxic, and non-pyrogenic. The pen needles are disposable, single use devices.

AI/ML Overview

The SureFine Pen Needle device, K250658, is a Class II hypodermic single lumen needle intended for use with a pen injector for the subcutaneous injection of insulin. The information provided is from an FDA 510(k) Clearance Letter.

1. Acceptance Criteria and Reported Device Performance

The acceptance criteria for the SureFine Pen Needle are based on compliance with international standards ISO 9626, ISO 11608-2, and ISO 10993-1, 10993-4, 10993-5, 10993-7, 10993-10, and 10993-11 for biocompatibility and sterilization validation according to ISO 11135.

Test ItemAcceptance Criteria (from referenced ISO standards)Reported Device Performance
ISO 9626 (Needle Characteristics)All criteria specified in ISO 9626 for hypodermic needlesPass
MaterialsConformance to specified material propertiesPass
CleanlinessConformance to cleanliness standardsPass
Limits for acidity and alkalinityConformance to pH limitsPass
Size designationAccurate representation of gauge and lengthPass
DimensionConformance to specified dimensional tolerancesPass
StiffnessConformance to stiffness requirementsPass
Resistance to breakageResistance to breakage under specified conditionsPass
Resistance to corrosionResistance to corrosion under specified conditionsPass
ISO 11608-2 (Needle-Based Injection Systems)All criteria specified in ISO 11608-2 for pen needlesPass
DimensionConformance to specified dimensional tolerancesPass
Flow rateConformance to prescribed flow ratePass
Bond between hub and needle tubeSecure bond between componentsPass
Needle point freedom from defects lubricationAbsence of defects and proper lubricationPass
Dislocation of measuring point at patientMinimal dislocation to ensure accurate deliveryPass
Compatibility of needles and injector systemFunctional compatibility with pen injectorsPass
Biocompatibility (ISO 10993 Series)All evaluation acceptance criteriaMet
Cytotoxicity testSatisfy requirements of ISO 10993-5 (in vitro cytotoxicity)Pass
Hemolysis testSatisfy requirements of ISO 10993-4 (hemolytic properties)Pass
Intracutaneous reactivity testSatisfy requirements of ISO 10993-10 (intracutaneous reactivity)Pass
Skin sensitization testSatisfy requirements of ISO 10993-10 (skin sensitization)Pass
Acute systemic toxicity testSatisfy requirements of ISO 10993-11 (acute systemic toxicity)Pass
Pyrogen TestSatisfy requirements of ISO 10993-11 (material-mediated pyrogens)Pass
Sterilization & Shelf-lifeMax levels of EO/ECH within ISO 10993-7 limits; adequate mitigation of bacterial endotoxins; 5-year shelf-life stability.Conforms to ISO 11135 validated half-cycle method. EOx/ECH levels meet ISO 10993-7. LAL testing performed. Supports 5-year shelf-life.

2. Sample Size Used for the Test Set and Data Provenance

The document does not explicitly state the specific sample sizes for each test mentioned (ISO 9626, ISO 11608-2, Biocompatibility). However, it indicates that "Bench tests for the device's performance were conducted" and "Biocompatibility of the SureFine Pen Needle was evaluated". The testing was conducted by Shina Med Corporation, a company based in the Republic of Korea. The nature of these tests (bench testing, lab-based biocompatibility) suggests they are prospective tests performed on manufactured samples of the device. The data provenance is primarily from internal testing and validation by the manufacturer, following international standards.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

The document does not specify the number or qualifications of experts used to establish ground truth for the "test set" in the context of device performance. The testing performed is against established international standards (ISO 9626, ISO 11608-2, ISO 10993 series), which themselves represent a form of "ground truth" for device safety and performance criteria. The evaluation results are reported as "Pass" or "Met," implying adherence to these predefined standards rather than an expert panel establishing a de novo ground truth for each specific test.

4. Adjudication Method for the Test Set

No information is provided regarding an adjudication method (such as 2+1, 3+1, none). The evaluation is based on objective measurements and established criteria set forth by the ISO standards. For instance, a flow rate test would have a pass/fail threshold, not require expert adjudication in the traditional sense.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done

No. The document describes bench tests and biocompatibility evaluations for a medical device (pen needle), not a diagnostic or AI-assisted system that would typically undergo an MRMC comparative effectiveness study involving human readers. Therefore, there is no mention of an effect size of human readers improving with or without AI assistance.

6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done

No. This device is a physical medical instrument (a pen needle), not a software algorithm or AI system. Therefore, the concept of "standalone performance" in the context of algorithms is not applicable here.

7. The Type of Ground Truth Used

The ground truth used for demonstrating the device's performance is adherence to established international standards and regulatory requirements:

  • ISO 9626: Specifies requirements for stainless steel needle tubing for the manufacture of hypodermic needles.
  • ISO 11608-2: Specifies requirements and test methods for single-use needles used with pen-injectors.
  • ISO 10993 series: Biological evaluation of medical devices, covering various aspects of biocompatibility (cytotoxicity, sensitization, irritation, systemic toxicity, hemocompatibility, pyrogenicity).
  • ISO 11135: Specifies requirements for the development, validation, and routine control of an ethylene oxide sterilization process for medical devices.
  • ISO 10993-7: EtO residual limits.
  • LAL testing: To ensure absence of bacterial endotoxins.

8. The Sample Size for the Training Set

Not applicable. This is a physical medical device, not a machine learning or AI algorithm, so there is no "training set."

9. How the Ground Truth for the Training Set Was Established

Not applicable, as there is no training set for a physical medical device.

FDA 510(k) Clearance Letter - SureFine Pen Needle

Page 1

U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov

Doc ID # 04017.07.05

April 3, 2025

Shina Med Corporation
℅ Peter Chung
President
Plus Global
300 Atwood Street
Pittsburgh, Pennsylvania 15213

Re: K250658
Trade/Device Name: SureFine Pen Needle
Regulation Number: 21 CFR 880.5570
Regulation Name: Hypodermic Single Lumen Needle
Regulatory Class: Class II
Product Code: FMI
Dated: March 5, 2025
Received: March 5, 2025

Dear Peter Chung:

We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).

Page 2

K250658 - Peter Chung Page 2

Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.

All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Page 3

K250658 - Peter Chung Page 3

Sincerely,

Shruti N. Mistry -S

Shruti Mistry
Assistant Director
DHT3C: Division of Drug Delivery and
General Hospital Devices, and
Human Factors
OHT3: Office of Gastrorenal, ObGyn,
General Hospital, and Urology Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health

Enclosure

Page 4

FORM FDA 3881 (8/23) Page 1 of 1

DEPARTMENT OF HEALTH AND HUMAN SERVICES
Food and Drug Administration

Indications for Use

Form Approved: OMB No. 0910-0120
Expiration Date: 07/31/2026
See PRA Statement below.

510(k) Number (if known): K250658

Device Name: SureFine Pen Needle

Indications for Use (Describe):
SureFine Pen Needle is intended for use with a pen injector for the subcutaneous injection of insulin.

Type of Use (Select one or both, as applicable):

  • ☒ Prescription Use (Part 21 CFR 801 Subpart D)
  • ☒ Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.
DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services
Food and Drug Administration
Office of Chief Information Officer
Paperwork Reduction Act (PRA) Staff
PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

Page 5

510(k) Summary

Page 1 / 3

1. Date of Preparation

March 27, 2025

2. Applicant

  1. Company: SHINA MED CORPORATION
  2. Address: 455-30, Bogaewonsam-ro Bogae-myun, Anseong-si, Gyeonggi-do, 17509, Republic of Korea
  3. Tel: +82-31-8057-2125
  4. Fax: +82-31-8057-2150
  5. Contact person: Peter Chung, 412-512-8802
  6. Contact person address: 300 Atwood Street, Pittsburgh, PA, 15213, USA
  7. Submission date: March 5, 2025
  8. Prior related submission: K152877

3. Subject Device Information

  1. Trade name: SureFine Pen Needle
  2. Common name: Insulin Pen Needle
  3. Classification name: Hypodermic single lumen needle
  4. Product code: FMI
  5. Regulation number: 880.5570
  6. Class of device: Class II
  7. Panel: General Hospital

4. Predicate Device

  1. Trade name (K152877): SureFine Pen Needle
  2. Manufacturer: SHINA MED CORPORATION

5. Reference Device

  1. Trade name (K210399): Unifine SafeControl
  2. Manufacturer: Owen Mumford Ltd

6. Device description

SureFine Pen Needle is designed for use with a pen injector for the subcutaneous injection of insulin. The pen needle consists of a needle, hub, and shield assembly. Blister paper covers primary container. The primary container maintains sterility of the needle because primary container covers the hub and needle cap with blister paper sealed on the opening hole of primary container.

The needle hub can be connected screwed onto the pen. The needle shield is intended to provide physical protection to the needle tube.

SureFine Pen Needle is offered in various gauges sizes (29G, 30G, 31G, 32G) and Lengths (4mm, 6mm, 8mm, 12.7mm). These are sterile (EO gas sterilization), non-toxic, and non-pyrogenic. The pen needles are disposable, single use devices.

7. Indications for Use

SureFine Pen Needle is intended for use with a pen injector for the subcutaneous injection of insulin.

8. Performance data

  1. Bench tests for the device's performance were conducted.
Test itemsResult
ISO 9626
MaterialsPass
CleanlinessPass
Limits for acidity and alkalinityPass
Size designationPass
DimensionPass
StiffnessPass
Resistance to breakagePass
Resistance to corrosionPass

Page 6

Page 2 / 3

ISO 11608-2
DimensionPass
Flow ratePass
Bond between hub and needle tubePass
Needle point freedom from defects lubricationPass
Dislocation of measuring point at patientPass
Compatibility of needles and injector systemPass
  1. Biocompatibility

Biocompatibility of the Surefine Pen Needle was evaluated in accordance with ISO 10993-1:2018. The following tests were performed: Cytotoxicity; Skin sensitization; Hemocompatibility; Intracutaneous reactivity; Acute systemic toxicity; Pyrogenicity. All evaluation acceptance criteria were met.

Test itemTest method / Test criteriaTest result
Cytotoxicity testWhen it was tested according to ISO 10993-5, tests for in vitro cytotoxicity-Test on extracts method, it should satisfy the requirements.Pass
Hemolysis testWhen it was tested according to ISO 10993-4, Selection of tests for interactions with blood-evaluation of hemolytic properties of medical devices and medical device materials, it should satisfy the requirements.Pass
Intracutaneous reactivity testWhen it was tested according to ISO 10993-10, Tests for irritation and skin sensitization-Animal intracutaneous (Intradermal) reactivity test, it should satisfy the requirements.Pass
Skin sensitization testwhen it was tested according to ISO 10993-10, Tests for irritation and skin sensitization-Guinea pig maximization test (GPMT), it should satisfy the requirements.Pass
Acute systemic toxicity testWhen it was tested according to ISO 10993-11, Tests for systemic toxicity-Acute systemic toxicity, it should satisfy the requirements.Pass
Pyrogen TestWhen it was tested according to ISO 10993-11, Tests for systemic toxicity-Information on material-mediated pyrogens, it should satisfy the requirements.Pass

9. Sterilization and Shelf-life Testing

Sterilization of the SureFine Pen Needle has been validated using the half-cycle method as outlined in ISO 11135. Testing demonstrated maximum levels of residues of ethylene oxide and ethylene chlorohydrins do not exceed the limits presented in ISO 10993-7. LAL testing was performed to demonstrate that bacterial endotoxins are adequately mitigated. Shelf-life testing supports a shelf-life of 5-years after sterilization.

10. Substantially Equivalent (SE) Comparison

Proposed devicePredicate deviceRemark
ManufacturerSHINA MED CORPORATIONSHINA MED CORPORATIONSame
510(K) No.K250658K152877N/A
Trade NameSureFine Pen NeedleSureFine Pen NeedleSame
Indications for useSureFine Pen Needle is intended for use with a pen injector for the subcutaneous injection of insulin.SureFine Pen Needle is intended for use with a pen injector for the subcutaneous injection of insulin.Same
Type of usePrescription use and Over-The-Counter usePrescription useNote #1
ComponentsPrimary container, Paper seal, Needle shield, Cannula, Needle hubPrimary container, Paper seal, Needle shield, Cannula, Needle hubSame
Needle length4mm, 6mm, 8mm, 12.7mm4mm, 6mm, 8mm, 12.7mmSame
Needle gauge29G, 30G, 31G, 32G29G, 30G, 31G, 32GSame

Page 7

Page 3 / 3

Proposed devicePredicate deviceRemark
BiocompatibilityConform ISO10993-1: Cytotoxicity, Sensitization, Irritation, Systemic Toxicity, Hemocompatibility, PyrogenicityConform ISO10993-1: Cytotoxicity, Sensitization, Irritation, Systemic Toxicity, Hemocompatibility, PyrogenicitySame
Needle performance requirementsISO 11608-2 : 2012 Needle-based injection systems for medical use — Requirements and test methodsISO 11608-2 : 2012 Needle-based injection systems for medical use — Requirements and test methodsSame
Shelf life5 years3 yearsNote #2

Note 1 – Type of use
Over-The-Counter use is added to the type of use of for the subject device; however, this change does not alter the indications for use, nor does it affect the safety and effectiveness of the subject device when compared to the predicate device. Thus, the change from Rx only to Rx and OTC does not raise any additional questions of safety or effectiveness when compared to the predicate device.

Note 2 – Shelf-life
The Shelf-life has changed to 5 years from 3 years in comparison with the predicate device. To validate this claim, shelf-life testing for the pen needle was conducted. Therefore, the different shelf-life does not raise new questions of safety or effectiveness compared to the predicate device.

11. Substantially Equivalent (SE) Conclusion

Based on the information provided, the SureFine Pen Needle is substantially equivalent to the predicate device (K152877).

§ 880.5570 Hypodermic single lumen needle.

(a)
Identification. A hypodermic single lumen needle is a device intended to inject fluids into, or withdraw fluids from, parts of the body below the surface of the skin. The device consists of a metal tube that is sharpened at one end and at the other end joined to a female connector (hub) designed to mate with a male connector (nozzle) of a piston syringe or an intravascular administration set.(b)
Classification. Class II (performance standards).