Search Results
Found 240 results
510(k) Data Aggregation
(91 days)
General Total Knee Arthroplasty (TKR) Indications:
- Painful, disabling joint disease of the knee resulting from: noninflammatory degenerative joint disease (including osteoarthritis, traumatic arthritis, or avascular necrosis), rheumatoid arthritis or post-traumatic arthritis.
- Post-traumatic loss of knee joint configuration and function.
- Moderate varus, valgus, or flexion deformity in which the ligamentous structures can be returned to adequate function and stability.
- Revision of previous unsuccessful knee replacement or other procedure.
- Fracture of the distal femur and/or proximal tibia that cannot be stabilized by standard fracture-management techniques.
- Ligamentous instability requiring implant bearing surface geometries with increased constraint.
- Absent or non-functioning posterior cruciate ligament.
The subject device is intended for primary or revision total knee arthroplasty.
The subject device is a sterile, single use device that is intended for use in primary or revision total knee arthroplasty with previously cleared Triathlon® femoral and tibial baseplate components. The subject device adds to the Triathlon® Total Knee System a tibial bearing insert option with an asymmetric articulating surface which is designed to increase medial conformity with the compatible femoral components. The design is intended to provide increased sagittal stability and allow for more natural rotational kinematics. This tibial bearing insert also introduces a material change to the locking wire subcomponent located on the anterior portion of the distal baseplate mating profile. The locking wire component is intended to assemble the insert with a compatible tibial baseplate. The locking wire provided with the Triathlon® X3® Medial Stabilized Tibial Bearing Insert is made from titanium alloy (Ti-6Al-4V).
N/A
Ask a specific question about this device
(95 days)
• Painful, disabling joint disease of the knee resulting from: degenerative arthritis, rheumatoid arthritis or post-traumatic arthritis.
• Post-traumatic loss of knee joint configuration and function.
• Moderate varus, valgus, or flexion deformity in which the ligamentous structures can be returned to adequate function and stability.
• Revision of previous unsuccessful knee replacement or other procedure.
Triathlon® Hinge Knee System is intended to be implanted with bone cement for the following condition(s):
• There is destruction of the joint surfaces, with or without significant bone deformity.
• The cruciate and/or collateral ligaments do not stabilize the knee joint.
• The ligaments are inadequate and/or the musculature is weak. And/or
• Revision is required of a failed prosthesis where there has been gross instability, with or without bone loss or inadequate soft tissue.
When used with MRH femur and/or MRH tibial baseplate replacement indicated in revision of an existing prosthesis:
• Revision is required of a failed prosthesis where there has been gross instability, with or without bone loss or inadequate soft tissue.
When used with compatible GMRS components:
• Where segmental resection and/or replacement of femur and/or proximal tibia is required.
• Painful, disabling joint disease of the knee resulting from: noninflammatory degenerative joint disease (including osteoarthritis, traumatic arthritis, or avascular necrosis), rheumatoid arthritis or post-traumatic arthritis.
• Post-traumatic loss of knee joint configuration and function.
• Moderate varus, valgus, or flexion deformity in which the ligamentous structures can be returned to adequate function and stability.
• Revision of previous unsuccessful knee replacement or other procedure.
• Fracture of the distal femur and/or proximal tibia that cannot be stabilized by standard fracture management techniques.
• Painful, disabling joint disease of the knee resulting from: noninflammatory degenerative joint disease (including osteoarthritis, traumatic arthritis, or avascular necrosis), rheumatoid arthritis or post-traumatic arthritis.
• Post-traumatic loss of knee joint configuration and function.
• Moderate varus, valgus, or flexion deformity in which the ligamentous structures can be returned to adequate function and stability.
• Revision of previous unsuccessful knee replacement or other procedure.
• Fracture of the distal femur and/or proximal tibia that cannot be stabilized by standard fracture management techniques.
The Triathlon Tritanium Tibial Baseplate and Tritanium Metal-Backed Patella components are indicated for both uncemented and cemented use.
The devices covered by this Traditional 510(k) Premarket Notification are Stryker UHMWPE devices (including knee patella devices and hinged knee bumper) that are packaged in a N2Vac environment. The packaging for these devices consists of an impermeable foil pouch in a Polyethylene Terephthalate Glycol (PETG) blister sealed with a Tyvek lid. All subject devices are commercially available and have been found to be substantially equivalent in previous 510(k)s.
N/A
Ask a specific question about this device
(116 days)
General Total Knee Arthroplasty (TKR) Indications:
- Painful, disabling joint disease of the knee resulting from: noninflammatory degenerative joint disease (including osteoarthritis, traumatic arthritis, or avascular necrosis), rheumatoid arthritis or post-traumatic arthritis.
- Post-traumatic loss of knee joint configuration and function.
- Moderate varus, valgus, or flexion deformity in which the ligamentous structures can be returned to adequate function and stability.
- Revision of previous unsuccessful knee replacement or other procedure.
- Fracture of the distal femur and/or proximal tibia that cannot be stabilized by standard fracture-management techniques
The Triathlon® Gold Femoral Components are indicated for both uncemented and cemented use.
The Triathlon® Total Knee System is a total knee prosthesis designed to resurface the articulating surface of the femoral and tibial bones. In a total knee replacement, Triathlon® Gold Femoral Components are intended to replace the articulating surface of the distal femur and articulate against an articular surface of the tibial component as well as an optional patellar component. Triathlon® Gold Femoral Components are additively manufactured from Ti-6Al-4V powder, using Electron Beam Melting (EBM) technology, and then fully coated with titanium nitride (TiN), using Physical Vapor Deposition (PVD). In addition to TiN coating, the bone cut facing surfaces of these femoral components contain the Tritanium® porous structure. Indicated for both uncemented and cemented use, the subject femoral components are provided in right and left configurations and eight proportional sizes (sizes 1-8) for a total of 16 unique components to accommodate differences in patient anatomy. The subject femoral components are sterile, single use devices.
N/A
Ask a specific question about this device
(56 days)
Triathlon® Hinge Knee System is intended to be implanted with bone cement for the following condition(s):
- There is destruction of the joint surfaces, with or without significant bone deformity.
- The cruciate and/or collateral ligaments do not stabilize the knee joint.
- The ligaments are inadequate and/or the musculature is weak. And/or
- Revision is required of a failed prosthesis where there has been gross instability, with or without bone loss or inadequate soft tissue.
When used with MRH femur and/or MRH tibial baseplate replacement indicated in revision of an existing prosthesis:
- Revision is required of a failed prosthesis where there has been gross instability, with or without bone loss or inadequate soft tissue.
When used with compatible GMRS components:
- Where segmental resection and/or replacement of femur and/or proximal tibia is required
The subject THK components are a line extension of the previously cleared Triathlon® Hinge Knee System (K223528, K230416). The THK System (K223528, K230416) is a tricompartmental knee system consisting of a stemmed femoral component and a stemmed tibial bearing component connected by a set of previously cleared MRH or THK Bushings and an MRH or THK Axle (K994207, K002552, K223528, K230416). A bumper locks this assembly. This assembly provides motion through the MRH or THK Axle/Bushings combination in the flexion/extension plane. The articulation between the bearing surfaces on the underside of a tibial bearing component and a tibial insert provides motion in the rotating plane. A hinge tibial insert is assembled to a Triathlon® Hinge Revision Tibial Baseplate that incorporates a longitudinal bore to accept a previously cleared MRH Tibial Sleeve (K994207, K002552) or Triathlon Tibial Sleeve (K223528, K230416). Optional distal femoral and tibial augments are available to fill bone defects.
The subject Triathlon® Hinge Small Proximal Tibial Bearing Component and Triathlon® End Cap are sterile, single use devices intended for cemented use only and are being added to the previously cleared THK system (K223528, K230416) as an extension. They can be used with previously cleared MRH Knee components (K994207, K002552), GMRS (K023087, K222056), and Triathlon® Knee System components (K172634, K172326, K190991, K143393, K141056, K132624, K070095, K061521, K053514, K052917, K051948, K051146, K040267).
This FDA 510(k) clearance letter is for a mechanical medical device (a knee implant system), not an AI/software as a medical device. Therefore, the request for "acceptance criteria and the study that proves the device meets the acceptance criteria" using metrics relevant to AI/software (like sensitivity, specificity, MRMC studies, ground truth for training/test sets, etc.) is not applicable here.
The document provided details about the substantial equivalence of the "Triathlon® Hinge Knee System" to previously cleared predicate devices. The "study" proving the device meets acceptance criteria for a mechanical implant typically involves non-clinical (benchtop) testing and engineering analyses to demonstrate performance, material compatibility, and safety, rather than clinical trials or AI performance evaluations with ground truth.
Here's how the provided information relates to the typical evaluation of such a device, addressing the closest analogues to your requested points:
Acceptance Criteria and Device Performance for a Mechanical Implant:
For mechanical implants like the Triathlon® Hinge Knee System, "acceptance criteria" are generally tied to the successful completion of various non-clinical (benchtop) tests and analyses. These tests simulate the physical stresses and conditions the implant will experience in the human body. The "study" proving the device meets these criteria is the execution and successful outcome of these non-clinical tests.
No direct table of "acceptance criteria" vs. "reported device performance" in the AI sense is provided in this 510(k) summary. Instead, the summary lists the types of non-clinical tests performed to demonstrate performance and substantial equivalence to predicate devices. The implied acceptance is that the device passed these tests, meaning its performance in these simulated scenarios was deemed acceptable and comparable to existing, legally marketed devices.
Table of Non-Clinical Tests (Analogous to Performance Metrics):
| Acceptance Criteria (Test Type) | Reported Device Performance (Implied "Pass") | Notes |
|---|---|---|
| Triathlon Hinge Bearing Component Varus/Valgus Fatigue | Met performance standards | Tests the component's ability to withstand repeated bending stresses in side-to-side directions, simulating loading during gait. |
| Triathlon Hinge Bearing Component Chair Rise Testing | Met performance standards | Simulates stresses experienced during a common activity (standing up from a chair). |
| Triathlon Hinge Full Construct Fatigue | Met performance standards | Evaluates the fatigue life of the entire assembled knee system under cyclic loading. |
| Wear Test Rationale for New Constructs | Acceptable wear characteristics demonstrated | Rationale provided to show that wear performance is acceptable, likely through comparison to predicate devices or established standards. Direct wear testing is implied by "wear test rationale". |
| Analysis of Contact Area/Contact Stress & Constraint Analysis | Acceptable contact mechanics & constraint | Computer modeling and/or benchtop tests to evaluate how the components interact under load, ensuring appropriate stress distribution and constrained movement. |
| Range of Motion and Rotational Freedom Analysis (ASTM F1223-20) | Met specified ranges and freedoms | Tested for flexion/extension, internal/external rotation, varus/valgus, and translation (medial/lateral, proximal/distal, anterior/posterior) to ensure physiological movement. |
| Total Femur Construct Compatibility & Triathlon Hinge Bushing/Axle Compatibility | Compatible with legacy components | Ensured new components integrate properly with existing cleared femoral and tibial components. |
| Triathlon Revision Baseplate-End Cap Locking Strength Analysis | Adequate locking strength | Evaluated the mechanical strength of the connection points. |
| Triathlon End Cap Tightening Analysis | Optimal tightening confirmed | Ensured secure fixation of the end cap. |
| Triathlon End Cap Load Carrying Capacity Rationale | Acceptable load capacity | Analysis provided to demonstrate the end cap can withstand expected loads. |
| Triathlon End Cap Stability Analysis | Stable under normal use | Evaluated the end cap's ability to remain in place and function without loosening. |
| MRI Testing (Safety for imaging) - Displacement, Torque, Artifacts, RF Heating | All MRI safety criteria met | Performed to confirm the device is safe for patients undergoing MRI procedures (e.g., no excessive magnetic pull, acceptable image distortion, no dangerous heating). Standards: ASTM F2052-15, ASTM F2213-17, ASTM F2119-07 (reapproved 2013), ASTM F2182-191ae. |
| Biocompatibility (ISO 10993-1:2020) | Biocompatible with human tissue | Testing to ensure the materials used in the implant do not cause adverse biological reactions in the body. |
| Shelf-life Validation (ISO 11607-1:2019, ISO 11607-2:2019, ASTM F1980-21) | Validated shelf-life | Ensures the sterility and integrity of the packaging system and the device itself are maintained over the specified shelf-life. Methods: ASTM F1886/F1886M-16, ASTM F88/88M-21, ASTM F2096-11(2019). |
| Bacterial Endotoxin Testing (ANSI/AAMI ST72:2019) | Endotoxin limit < 20EU/Device | Confirms the device is free from pyrogens (substances that can cause fever). |
Since this is for a mechanical knee implant system (not an AI/software device), most of the requested points related to AI are not applicable. Here's a breakdown based on the provided document:
-
A table of acceptance criteria and the reported device performance:
- As shown above, the acceptance criteria are implicitly defined by the successful completion of the listed non-clinical tests and analyses. The performance is "Met performance standards" or "Acceptable," as detailed in the Notes column. The document doesn't provide specific numerical results of these tests (e.g., how many cycles endured, specific force resistance values) but states they were performed to determine substantial equivalence.
-
Sample sized used for the test set and the data provenance:
- This concept (test set, data provenance) is typically for AI/software. For mechanical device testing, the "sample size" refers to the number of physical units tested according to each standard. These details are generally found in the full test reports, which are not part of the public 510(k) summary. The "data provenance" would refer to the lab where tests were conducted (e.g., an accredited testing lab).
- Not explicitly stated in the 510(k) summary.
-
Number of experts used to establish the ground truth for the test set and the qualifications of those experts:
- This is an AI/software concept. For mechanical device testing, "ground truth" is defined by established engineering and medical device standards (e.g., ASTM, ISO guidelines) and the physical properties and performance characteristics of the materials and design. Expert engineers and quality assurance specialists verify the test results against these standards.
- Not applicable for a mechanical device.
-
Adjudication method (e.g. 2+1, 3+1, none) for the test set:
- This is an AI/software concept for resolving discordant expert opinions. For mechanical testing, test results are typically objective and quantitative, allowing for direct comparison against predefined acceptance criteria from standards.
- Not applicable for a mechanical device.
-
If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:
- This is an AI/software concept measuring clinical impact with human-AI interaction. For this mechanical device, the comparison is to legally marketed predicate devices through substantial equivalence, primarily based on design, materials, and non-clinical performance, rather than an MRMC study.
- No MRMC study was done, as it's not applicable to a mechanical implant.
-
If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:
- This refers to AI performance. For a mechanical device, its "standalone" performance is assessed through the non-clinical benchtop tests listed.
- The non-clinical tests listed in the summary represent the "standalone" performance assessment of the device.
-
The type of ground truth used (expert consensus, pathology, outcomes data, etc.):
- For a mechanical implant, "ground truth" for performance is established by engineering principles, material science, and performance standards (e.g., ASTM, ISO). For biocompatibility, the ground truth is established per ISO 10993-1. For MRI safety, it's established by ASTM standards.
- Ground truth established by relevant engineering and medical device standards.
-
The sample size for the training set:
- This refers to AI algorithm development. For a mechanical device, there isn't a "training set" in this sense. The
trainingfor the design of the device comes from established engineering principles, prior device designs, and material science research. - Not applicable for a mechanical device.
- This refers to AI algorithm development. For a mechanical device, there isn't a "training set" in this sense. The
-
How the ground truth for the training set was established:
- Again, this is an AI concept.
- Not applicable for a mechanical device.
In conclusion, the Triathlon® Hinge Knee System is a physical implant, and its clearance is based on demonstrating substantial equivalence to existing legally marketed devices through a series of rigorous non-clinical (benchtop) tests and engineering analyses that confirm its material properties, mechanical performance, biocompatibility, and safety (e.g., MRI compatibility, sterility). Clinical testing was explicitly stated as not required for this 510(k) submission, a common occurrence for devices demonstrating substantial equivalence through non-clinical data.
Ask a specific question about this device
Indications for Restoration Modular Hip System
- · Noninflammatory degenerative joint disease, including osteoarthritis and avascular necrosis:
- · Rheumatoid arthritis;
- · Correction of functional deformity;
- · Revision procedures where other treatments or devices have failed; and,
· Nonunions, femoral neck fractures, and trochanteric fractures of the proximal femur with head involvement that are unmanageable using other techniques.
· The RESTORATION Modular Hip System is intended for primary and revision total hip arthroplasty as well as in the presence of severe proximal bone loss. These femoral stems are designed to be proximal femur.
Indications for Accolade C, Accolade HFx, SYSTEM 12 CROSSFIRE, ACCOLADE DISTAL SPACER, CITATION TMZF, PCA Duration Insert, PCA Femoral Head
- · Noninflammatory degenerative joint disease, including osteoarthritis and avascular necrosis;
- · Rheumatoid arthritis;
- · Correction of functional deformity;
- · Revision procedures where other treatments or devices have failed; and,
· Nonunions, femoral neck fractures, and trochanteric fractures of the proximal femur with head involvement that are unmanageable using other techniques
Indications for Accolade II Stems
The indications for use of the total hip replacement prostheses include:
- · noninflammatory degenerative joint disease, including osteoarthritis and avascular necrosis;
- rheumatoid arthritis;
- · correction of functional deformity;
- · revision procedures where other treatments or devices have failed; and.
· nonunions, femoral neck and trochanteric fractures of the proximal femur with head involvement that are unmanageable using other techniques.
Additional indication specific to use of ACCOLADE II Femoral Stems with compatible Howmedica Osteonics Constrained Liners
· When the stem is to be used with compatible Howmedica Osteonics Constrained Liners, the device is intended for use in primary or revision patients at high risk of hip dislocation due to a history of prior dislocation, bone loss, soft tissue laxity, neuromuscular disease, or intraoperative instability.
ACCOLADE II Femoral Stems are intended for cementless use only and are intended for total and hemiarthroplasty procedures.
Indications for OMNIFIT SERIES Acetbular Inserts, Omnifit Crossfire 10 Deg Inserts, Crossfire Inserts, Omnifit Crossfire Inserts, Trident Crossfire Inserts, Trident X3 Inserts, Trident X3 Eccentric Inserts
· Painful, disabling joint disease of the hip from: degenerative arthritis, theumatoid arthritis, posttraumatic arthritis or late stage avascular necrosis.
· Revision of previous unsuccessful femoral head replacement, cup arthroplasty or other procedure.
· Clinical management problems where arthrodesis or alternative techniques are less likely to achieve satisfactory results.
· Where bone stock is of poor quality or inadequate for other reconstructive techniques as indicated by deficiencies of the acetabulum.
Indications for ALL POLY CONSTRAINED INSERT, TRIDENT 0 DEG CONSTRAINED INSERT, TRIDENT CONSTRAINED INSERT
A Constrained Acetabular Insert is indicated for use as a component of a total hip prosthesis in primary and revision patients at high risk of hip dislocation due to a history of prior dislocation, bone loss, joint or soft tissue laxity, neuromuscular disease, or intraoperative instability.
Indications for UHR Bipolar
- · Femoral head/neck fractures or non-unions.
- · Aseptic necrosis of the femoral head.
· Osteo-, rheumatoid, and post-traumatic arthritis of the hip with minimal acetabular involvement or distortion.
Other Considerations:
· Pathological conditions or age considerations which indicate a more conservative acetabular procedure and an avoidance of the use of bone cement in the acetabulum.
· Salvage of failed total hip arthroplasty
Indications for Artisan Bone Plug
These bone plugs are intended to be placed in the femoral canal prior to the introduction of bone cemented hip procedure.
The plug is placed distally to the femoral stem to help allow cement pressurization and to help prevent cement migration further down the femoral canal.
Indications for C-Taper Alumina Ceramic Heads, V40 Taper Alumina Ceramic V40™ Femoral Head
• Painful, disabling joint disease of the hip from: degenerative arthritis, theumatoid arthritis, posttraumatic arthritis or late stage avascular necrosis.
· Revision of previous unsuccessful femoral head replacement, cup arthroplasty or other procedure.
· Clinical management problems where arthrodesis or alternative techniques are less likely to achieve satisfactory results.
• Where bone stock is of poor quality or is inadequate for other reconstructive techniques as indicated by deficiencies of the acetabulum
Indications for BIOLOX Delta Ceramic Heads, C-Taper to Universal Taper Adapter Sleeve, Universal V40™ Taper Adapter Sleeve
For Use as a Total Hip Replacement:
· Painful disabling joint disease of the hip resulting from: degenerative arthritis, rheumatic arthritis or late stage avascular necrosis.
• Revision of previous cup arthroplasty or other procedures
· Clinical management problems where arthrodesis or alternative techniques are less likely to achieve satisfactory results.
· Where bone stock is of poor quality or is inadequate for other reconstructive techniques as indicated by deficiencies in the acetabulum.
For Use as a Bipolar Hip Replacement
- · Femoral head/neck fractures or non-unions.
· Aseptic necrosis of the femoral head.
• Osteo-, rheumatoid, and post traumatic arthritis of the hip with minimal acetabular involvement or distortion.
· Pathological considerations or age considerations which indicate a more conservative acetabular procedure and an avoidance of the use of bone cement in the acetabulum.
· Salvage of failed total hip arthroplasty
Indications for ADM/MDM X3 Inserts, MDM Acetabular Inserts, MDM Acetabular Liners
The indications for use for total hip arthroplasty include:
- Noninflammatory degenerative joint disease including osteoarthritis and avascular necrosis;
- Rheumatoid arthritis;
- Correction of functional deformity;
- Revision procedures where other treatments or devices have failed; and,
- Treatment of nonunion, femoral neck and trochanteric fractures of the proximal femur with head involvement that are unmanageable using other techniques.
- Dislocation risks
MDM Liners are intended for cementless use only.
The devices included in this submission are femoral heads, acetabular inserts, distal spacers, and bone plugs, used in hip arthroplasty procedures. All devices have been previously deemed substantially equivalent in prior premarket submissions and are commercially available.
This document, K243784, is a 510(k) premarket notification from Stryker Orthopaedics for a "Hip Devices Labeling Update." It states that the purpose of the submission is to add a contraindication to the labeling of existing Stryker Orthopaedics Hip Devices.
Based on the provided document, the device in question is a physical medical device (hip implants and components) and the 510(k) submission is for a labeling update, specifically adding a contraindication. This is not an AI/ML-enabled medical device. Therefore, the questions regarding acceptance criteria and studies proving the device meets those criteria, particularly those related to AI/ML device performance (e.g., sample sizes for training/test sets, expert adjudication, MRMC studies, standalone algorithm performance, ground truth establishment), are not applicable to this submission.
The document explicitly states under "Non-Clinical Testing" and "Clinical Testing":
- "Non-Clinical testing was not required as a basis for substantial equivalence."
- "Clinical testing was not required as a basis for substantial equivalence."
This indicates that no performance studies (clinical or non-clinical in the context of device performance metrics) were conducted or needed to be submitted for this specific 510(k) approval, as the change is solely a labeling update concerning a contraindication, and the device itself (the physical hip implant) has previously been deemed substantially equivalent.
Therefore, I cannot populate the table or answer the specific questions related to acceptance criteria and performance studies for an AI/ML device, as this context does not apply to the provided document.
Ask a specific question about this device
(56 days)
The Stryker Scorpio Total Knee system components are intended for the replacement of the bearing and/or articulating surfaces of the distal femur and proximal tibia to remove pain, instability, and restriction of motion due to degenerative bone disease, including osteoarthritis, rheumatoid arthritis, failure of other devices or trauma. The devices are intended for single-use only, and are intended for cemented fixation in patients indicated for total knee arthroplasty.
The devices included in this submission are tibial inserts, and all-polyethylene patellar components used in total knee arthroplasty procedures. All devices have been previously deemed substantially equivalent in prior premarket submissions and are commercially available.
This document does not contain information about acceptance criteria or a study proving that a device meets acceptance criteria.
The document is a 510(k) premarket notification letter from the FDA regarding a change in labeling for orthopedic knee implants (Scorpio Universal Dome Patella, Scorpio Total Stabilizer Insert, Scorpio-Flex Posterior Stabilized Tibial Insert, Scorpio-Flex Cruciate Retaining Tibial Insert, Scorpio NRG Tibial Bearing Insert – Cruciate Retaining Insert, Scorpio NRG Tibial Bearing Insert – Posteriorly Stabilized Insert).
The key takeaway from the document regarding the submission is:
- Reason for 510(k) Submission: "The purpose of this "Change Being Effected" bundled submission is to add a contraindication to the labeling of the subject Scorpio Total Knee System." (Page 7)
- Non-Clinical and Clinical Testing: "Non-Clinical testing was not required as a basis for substantial equivalence." and "Clinical testing was not required as a basis for substantial equivalence." (Page 10)
This means the submission is primarily administrative (a labeling change) and does not involve new performance testing or studies. Therefore, the requested information about acceptance criteria, study details, sample sizes, expert involvement, and ground truth establishment is not present in this document.
Ask a specific question about this device
General Total Knee Arthroplasty (TKR) Indications:
• Painful, disabling joint disease of the knee resulting from: noninflammatory degenerative joint disease (including osteoarthritis, traumatic arthritis, or avascular necrosis), rheumatoid arthritis or post-traumatic arthritis.
· Post-traumatic loss of knee joint configuration and function.
· Moderate varus, valgus, or flexion deformity in which the ligamentous structures can be returned to adequate function and stability.
· Revision of previous unsuccessful knee replacement or other procedure.
- Fracture of the distal femur and/or proximal tibia that cannot be standard fracture -management techniques.
The Triathlon® Tritanium® Total Knee System components are indicated for both uncemented use. The Triathlon® Total Knee System beaded with Peri-Apatite components are intended for uncemented use only.
The Triathlon® All Polyethylene tibial components are indicated for cemented use only.
Additional Indications for Posterior Stabilized (PS) and Total Stabilizer (TS) Components: - · Ligamentous instability requiring implant bearing surface geometries with increased constraint.
- · Absent or non-functioning posterior cruciate ligament.
· Severe anteroposterior instability of the knee joint.
Additional Indications for Total Stabilizer (TS) Components:
· Severe instability of the knee secondary to compromised collateral ligament integrity or function.
Indications for Bone Augments:
· Painful, disabling joint disease of the knee secondary to: degenerative arthritis, rheumatic arthritis, complicated by the presence of bone loss.
· Salvage of previous unsuccessful total knee replacement or other surgical procedure, accompanied by bone loss.
Additional Indications for Cone Augments: - · Severe degeneration or trauma requiring extensive resection and replacement
- · Femoral and Tibial bone voids
- Metaphyseal reconstruction
The Triathlon TS Cone Augment components are intended for cemented or cementless use.
Triathlon Pro Posterior Stabilized Femoral Components Indications for Use:
General Total Knee Arthroplasty (TKR) Indications:
· Painful, disabling joint disease of the knee resulting from: noninflammatory degenerative joint disease (including osteoarthritis, traumatic arthritis, or avascular necrosis), rheumatoid arthritis or post-traumatic arthritis.
- · Post-traumatic loss of knee joint configuration and function.
· Moderate varus, valgus, or flexion deformity in which the ligamentous structures can be returned to adequate function and stability.
· Revision of previous unsuccessful knee replacement or other procedure.
· Fracture of the distal femur and/or proximal tibia that cannot be standard fracture -management techniques. Additional Indications for Posterior Stabilized (PS) components: - · Ligamentous instability requiring implant bearing surface geometries with increased constraint.
- · Absent or non-functioning posterior cruciate ligament.
- · Severe anteroposterior instability of the knee joint.
The Triathlon® Pro PS Femoral Components are intended for cemented use only.
Triathlon Tritanium Tibial Baseplate Indications for Use:
General Total Knee Arthroplasty (TKR) Indications:
· Painful, disabling joint disease of the knee resulting from: non-inflammatory degenerative joint disease (including osteoarthritis, traumatic arthritis or avascular necrosis) or rheumatoid arthritis
- · Post-traumatic loss of knee joint configuration and function
- · Moderate varus, valgus, or flexion deformity in which the ligamentous structures can be returned to adequate function and stability
- · Revision of previous unsuccessful knee replacement or other procedure
- · Fracture of the distal femur and/or proximal tibia that cannot be standard fracture management techniques
Additional General Total Knee Arthroplasty (TKR) Indications specific to the PS implant: - · Ligamentous instability requiring implant bearing surface geometries with increased constraint
- Absent or non-functioning posterior cruciate ligament
- · Severe anteroposterior instability of the knee joint
The Triathlon Tritanium Tibial Baseplates are indicated for both cemented and uncemented use.
Triathlon Low Profile Tibial Tray Indications for Use:
The Triathlon Low Profile Tibial Tray is intended to be used with commercially available Triathlon® femoral components and associated patellar components, and tibial bearing inserts in primary cemented total knee arthroplasty. The indications for the Triathlon® Low Profile Tibial Tray are outlined below:
Indications for Use:
· Painful, disabling joint disease of the knee resulting from: degenerative arthritis, rheumatoid arthritis or post-traumatic arthritis.
· Post-traumatic loss of knee joint configuration and function.
· Moderate varus, valgus, or flexion deformity in which the ligamentous structures can be returned to adequate function and stability.
Triathlon Metal Backed Patella Indications for Use:
- · Noninflammatory degenerative joint disease including osteoarthritis or avascular necrosis;
- Rheumatoid arthritis;
- · Correction of functional deformity;
- · Revision procedures where other treatments or devices have failed;
- · Post traumatic loss of joint anatomy, particularly when there is patello-femoral erosion, dysfunction or prior patellectomy; and,
- · Irreparable fracture of the knee.
These products are intended to achieve fixation without the use of bone cement.
Triathlon Partial Knee System Indications for Use:
Moderately disabling joint disease of the knee resulting from painful osteo- or post traumatic arthritis
· Revision of previous unsuccessful surgical procedures, either involving, or not involving, previous use of a unicompartmental knee prosthesis
· As an alternative to tibial osteotomy in patients with unicompartmental osteoarthritis
· Where bone stock is of poor quality or inadequate for other reconstructive techniques as indicated by deficiencies of the femoral condyle/tibial plateau.
These components are intended for implantation with bone cement.
Avon Patello-femoral Joint Prosthesis Indications for Use:
The Avon Patello-femoral Joint Prosthesis is intended to be used in cemented patellofemoral arthroplasty in patients with degenerative arthritis in the distal femur and patients with a history of patellar dislocation or patella fracture, or patients with failed previous surgery (arthroscopy, tibial tubercle elevation, lateral release) where pain, deformity, or dysfunction persists. These components are single use only and are intended for implantation with bone cement.
Restoris Multi-Compartmental Knee System Indications for Use:
Restoris MCK is indicated for single or multi-compartmental knee replacement used in conjunction with RIO, the Robotic Arm Interactive Orthopedic System, in individuals with osteoarthritis of the tibiofemoral and/or patellofemoral articular surfaces.
The specific knee replacement configurations include:
- Medial unicondylar
- Lateral unicondylar
- Patellofemoral
- · Medial bi-compartmental (medial unicondylar and patellofemoral)
Restoris Multi Compartmental Knee is for single use only and is intended for implantation with bone cement.
All of the subject devices have been found substantially equivalent in previous 510(k)s. All the subject devices have been cleared for MR conditional labeling in previous 510(k)s. The purpose of this submission is to modify the MR conditional information in the instructions for use to update the parameters in which a patient who has the device can be safely scanned, per testing conducted accordance to updated FDA guidance. There have been no changes made to the devices requiring 510(k) clearance - only the MR conditional information in the instruction for use is being modified.
An additional contraindication is being added to the components of the Triathlon Total Knee System. This contraindication regarding material sensitivity to implant materials is being added.
This document is a 510(k) premarket notification for knee arthroplasty devices, specifically for the Triathlon Knee System and related components. It is important to note that this document does NOT describe a study evaluating the performance of a device driven by an AI algorithm or software.
Instead, this submission is for physical medical devices (knee implants) and their associated labeling. The core purpose of this 510(k) submission is to modify the MR conditional information in the instructions for use for these existing, previously cleared devices due to updated FDA guidance on MR safety testing. The document explicitly states:
- "There have been no changes made to the devices requiring 510(k) clearance - only the MR conditional information in the instruction for use is being modified." (Page 16)
- "There have been no changes requiring 510(k) clearance to the technological characteristics of the Stryker Knee systems as a result of the revision to the labeling." (Page 19)
- "Clinical testing was not required as a basis for substantial equivalence." (Page 20)
Therefore, the requested information regarding "acceptance criteria and the study that proves the device meets the acceptance criteria" in the context of an AI-driven device's performance (e.g., accuracy, sensitivity, specificity, human-in-the-loop studies) cannot be extracted from this document, as it pertains to entirely different types of evaluation (mechanical performance, biocompatibility, and in this specific case, MR safety of physical implants, not AI algorithm performance).
To explicitly answer your questions based on the provided document, even if they don't fully apply to the nature of this submission:
-
A table of acceptance criteria and the reported device performance:
- Acceptance Criteria (Implicit for this submission): The new MR safety testing results must demonstrate that the device is safe for use in an MR environment under the updated parameters to justify the labeling changes, and that the device remains substantially equivalent to its predicates.
- Reported Device Performance: The document states: "New testing was performed to comprehensively assess the RF-related heating effects induced by the subject devices when implanted into bone, following the FDA guidance document, "Testing and Labeling Medical Devices for Safety in the Magnetic Resonance (MR) Environment," dated May 20, 2021." (Page 19). While the specific quantitative results of this testing are not provided in this summary document, the FDA's clearance (the letter at the beginning) implies the acceptance criteria for MR safety were met.
-
Sample sized used for the test set and the data provenance:
- This document describes non-clinical (laboratory) testing for MR safety. It does not involve "test sets" in the sense of patient data for AI evaluation. The "sample size" would refer to the number of physical implants tested for MR compatibility. This specific number is not provided in the summary.
- Data Provenance: N/A for clinical data; the testing was performed per FDA guidance, implying controlled laboratory conditions.
-
Number of experts used to establish the ground truth for the test set and the qualifications of those experts:
- N/A. This is not a study requiring expert readers or ground truth establishment in the context of AI performance. The "ground truth" for MR safety is established by quantitative measurements in a laboratory setting based on physics principles and regulatory standards.
-
Adjudication method (e.g. 2+1, 3+1, none) for the test set:
- N/A. Not a clinical study requiring human adjudication.
-
If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:
- No. This document does not pertain to AI or human reader performance.
-
If a standalone (i.e. algorithm only, without human-in-the-loop performance) was done:
- No. This document does not describe an algorithm.
-
The type of ground truth used (expert concensus, pathology, outcomes data, etc):
- N/A for artificial intelligence context. The ground truth for this submission relates to physical properties and safety in an MR environment, established through standardized physical testing and engineering measurements.
-
The sample size for the training set:
- N/A. There is no training set mentioned or implied for an AI algorithm in this submission.
-
How the ground truth for the training set was established:
- N/A. There is no training set mentioned or implied.
In summary, this 510(k) notification is for knee replacement components and deals with updating their MR conditional labeling based on physical testing, not with the performance evaluation of an AI-powered device.
Ask a specific question about this device
(49 days)
The Stryker Global Modular Replacement System, Modular Replacement System and Modular Rotating Hinge Knee Component are sterile, single-use devices intended for use in situations where there is a need for replacement of bone due to radical bone loss. This loss can be related to oncology, trauma or failed previous prosthesis. Specific Indications for Use are listed below.
Indications for the MRS Stems and Intercalary Stems (presented in K952970):
- This device is intended for use in patients requiring extensive reconstruction of the femur . and/or proximal tibia, including the hip or knee joint, resulting from extensive bone loss. Tumor resection for skeletal lesions (Oncology patients where radical bone resection and replacement may be required), revision surgery for failed arthroplasty, and acute trauma are the primary causes of extensive bone loss. These prostheses are intended for use with bone cement.
- The Intercalary System is intended for use in situations arising from femoral mid-shaft tumor resection, or for prosthetic knee fusion.
Indications for the Global Modular Replacement System (presented in K023087):
- Femoral and/or proximal tibial replacement and total femoral replacement in Oncology cases where radical resection and replacement of bone is required, and in limb salvage procedures where radical resection and replacement of the bone is required. Limb salvage procedures would include surgical intervention for severe trauma, failed previous prosthesis, and/or Oncology indications.
Indications for the GMRS Anteverted Proximal Femoral Component (presented in K032581):
- Femoral replacement in Oncology cases where radical resection and replacement of bone is required, and in limb salvage procedures where radical resection and replacement of bone is required. Limb salvage procedures would include surgical intervention for severe trauma, failed previous prosthesis, and/or Oncology indications. This component may also be used with the Distal Femoral segment components of the GMRS in total femoral replacement.
Indications for the Global Modular Replacement System Press Fit Stems with HA Coating (presented in K022403, and K031217):
Proximal femoral reconstruction secondary to:
o Trauma
o Failed previous prosthesis
o Tumor resection
Indications for the Modular Replacement System Cemented Stems (presented in K040749):
Femoral and/or proximal tibial replacement due to:
o Trauma
o Failed previous prosthesis
o Tumor resection
Indications for the Modular Rotating Hinge Knee System (presented in K002552):
- The Modular Rotating Hinge Knee System is intended for use with bone cement in cases where there is destruction of the joint surfaces, with or without significant bone deformity; the cruciate and/or collateral ligaments do not stabilize the knee joint; the ligaments are inadequate and/or the musculature is weak; and revision of a failed previous prosthesis where there is instability, with or without bone loss or inadequate soft tissue. Expanded indications include for use with bone cement in situations where there is extensive bone loss, including limb salvage procedures for severe trauma, failed previous prosthesis, and/or Oncology indications.
The Global Modular Replacement System (GMRS) is comprised of a number of components that are intended to be used in conjunction with each other, or in conjunction with components of the Modular Replacement System (MRS) or the Modular Rotating Hinge Knee System (MRH). The Modular Rotating Hinge Knee System is a tricompartmental knee system. The system consists of a stemmed femoral component and a stemmed tibial bearing component connected by a set of bushings and an axle. The MRH Tibial Rotating component is a part of the MRH system and mates with the GMRS Standard Proximal Tibia. These devices are intended to be used in clinical situations where there is radical bone loss of the femur and/or proximal tibia. This radical bone loss can be related to oncology (tumor), trauma or failed previous prosthesis.
The provided FDA 510(k) clearance letter and summary for the Stryker Orthopaedics Global Modular Replacement System, Modular Replacement System, and Modular Rotating Hinge Knee Component does not contain information regarding acceptance criteria or a study proving the device meets specific performance criteria.
This document is a clearance letter indicating a determination of "substantial equivalence" to previously marketed predicate devices. It does not present novel performance data or studies against pre-defined acceptance criteria.
The key statements from the document that confirm this are:
- "Non-Clinical Testing: Non-Clinical testing was not required as a basis for substantial equivalence." (Page 8)
- "Clinical Testing: Clinical testing was not required as a basis for substantial equivalence." (Page 8)
The submission focuses on adding a contraindication to the labeling, updating MRI Safety information, and making clerical updates to the Instructions for Use (IFUs). These changes do not necessitate new performance studies for substantial equivalence.
Therefore, I cannot populate the requested table or answer the specific questions about acceptance criteria and studies because the provided document explicitly states that such testing was not required for this particular 510(k) submission.
Ask a specific question about this device
(126 days)
-
Painful, disabling joint disease of the hip from: degenerative arthritis, rheumatoid arthritis, post traumatic arthritis or late-stage avascular necrosis.
-
Revision of previous unsuccessful femoral head replacement, cup arthroplasty or other procedure.
-
Clinical management problems where arthrodesis or alternative techniques are less likely to achieve satisfactory results.
-
Where bone stock is of poor quality or is inadequate for other reconstructive techniques as indicated of the acetabulum.
The subject 22.2mm Biolox® delta Ceramic V40™ Femoral Heads with neck length +0mm and +3mm are sterile, single-use devices that are manufactured with high-purity alumina matrix with zirconia reinforcement and are compatible with the 22.2mm ID sizes of compatible acetabular inserts and UHMWPE acetabular cups. The Biolox® delta Ceramic V40TM Femoral Heads may be used in conjunction with compatible V40™ taper femoral stems and acetabular shell components to achieve reconstructive replacement of the hip joint.
This document is a FDA 510(k) premarket notification letter for a medical device, specifically the 22.2mm Biolox® delta Ceramic V40™ Femoral Heads. This type of submission focuses on demonstrating substantial equivalence to a legally marketed predicate device, rather than detailed, de novo clinical or algorithmic performance studies typically seen for novel AI/ML or complex diagnostic devices.
Therefore, the request for "acceptance criteria and the study that proves the device meets the acceptance criteria" in the context of an AI/ML algorithm or a complex diagnostic cannot be fully answered by this document. This clearance is for a mechanical orthopedic implant.
However, I can extract the information relevant to device performance and substantiation as presented for this type of mechanical device, even if it doesn't align with the typical AI/ML study parameters.
Here's what can be extracted and how it relates to your request:
Understanding the Device and its Substantiation Approach:
The device is a femoral head for hip joint replacement, a mechanical implant. Its "performance" is assessed through engineering tests (mechanical strength, wear, range of motion, etc.) rather than clinical diagnostic accuracy or algorithmic output. The FDA's acceptance is based on demonstrating substantial equivalence to an existing, legally marketed device (the predicate). This means showing that the new device is as safe and effective as the predicate, typically through comparison of design, materials, intended use, and non-clinical (laboratory) performance data.
Addressing Your Specific Points (adjusted for a mechanical device 510(k) submission):
-
A table of acceptance criteria and the reported device performance:
- Acceptance Criteria (Mechanical/Engineering Standards): For this type of device, the "acceptance criteria" are the performance specified by recognized consensus standards (ISO, ASTM) for orthopedic implants. The document lists the standards followed, implying the device met the requirements of these standards. It doesn't present a table with specific pass/fail values or measured results against those values, as that level of detail is typically in the underlying test reports, not the 510(k) summary.
- Reported Device Performance:
- "The device comparisons and performance testing show that the 22.2mm Biolox® delta Ceramic V40™ Femoral Heads are substantially equivalent to the cited predicate V40™ Biolox® delta Ceramic Femoral Heads cleared via K052718, based on intended use, indications for use, design, material, technological characteristics, operational principles, and non-clinical performance data."
- Non-Clinical Testing Performed:
- Burst testing, Fatigue testing and post-fatigue burst testing (ISO 7206-10 and ASTM F2345-21)
- Axial Pull-off testing
- Wear evaluation
- Range of Motion
- Torsional resistance (ISO 7206-13)
- MRI Safety testing
- Coating Characterization Standards (for compatible Accolade II hip stems):
- ISO 13179-1
- ASTM F1160
- ASTM F1147
- ASTM F1044
- ASTM F1978
- ASTM F1580
- ASTM F1854
Note: The document only lists the tests and standards; it does not provide the quantitative results or the specific acceptance criteria thresholds for each test. This information would be in the detailed engineering reports submitted to the FDA, not typically in the public 510(k) summary. The "performance" is implicitly stated as meeting the "substantial equivalence" criteria by successfully passing these tests.
-
Sample sizes used for the test set and the data provenance:
- Sample Size: Not explicitly stated in this summary. For mechanical tests, sample sizes are typically determined by the specific ASTM/ISO standards followed (e.g., n=3 or n=6 for certain fatigue tests).
- Data Provenance: Not specified for the non-clinical testing. These are laboratory tests, not clinical data from patients.
-
Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience):
- This question is not applicable to this type of device and submission. "Ground truth" established by experts (like radiologists for AI/ML) is irrelevant for mechanical performance testing of an orthopedic implant. The "ground truth" here is objective physical measurement against established engineering standards.
-
Adjudication method (e.g. 2+1, 3+1, none) for the test set:
- This question is not applicable. Adjudication is for subjective assessment, often in clinical trials or for establishing ground truth in image interpretation. Mechanical tests are objective.
-
If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:
- This question is not applicable. MRMC studies are for evaluating diagnostic performance of AI or human readers, typically with medical imaging. This is a mechanical implant.
-
If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:
- This question is not applicable. This is not an algorithm. "Standalone performance" for this device would refer to its mechanical integrity and function. The non-clinical tests serve this purpose.
-
The type of ground truth used (expert consensus, pathology, outcomes data, etc.):
- Ground Truth: For this mechanical device, the "ground truth" is defined by the objective performance requirements outlined in established international and national engineering standards (ISO and ASTM). Results of the physical tests are measured and compared against the pass/fail criteria defined in these standards.
-
The sample size for the training set:
- This question is not applicable. This is not an AI/ML device, so there is no "training set."
-
How the ground truth for the training set was established:
- This question is not applicable. No training set exists.
Summary for this specific document:
This FDA 510(k) clearance document is for a mechanical orthopedic implant, not an AI/ML or diagnostic device. As such, the typical criteria and study types asked about (AI performance, human readers, ground truth consensus, training/test sets for algorithms) are not relevant or applicable to this submission.
The "acceptance criteria" for this device are meeting the requirements of recognized industry standards for mechanical testing (e.g., ISO, ASTM) to demonstrate substantial equivalence in terms of safety and performance to a predicate device. The "study" proving this involves non-clinical (laboratory) testing, the results of which (though not detailed quantitatively in this summary) are presented as having successfully met the engineering specifications for substantial equivalence.
Ask a specific question about this device
(59 days)
DALL-MILES Cable System
The DALL-Miles System is indicated for reattachment of the trochanter in any hip procedure using the trochanteric osteotomy (total or partial) approach.
The DALL-MILES Mini Cleat is indicated for vertical reattachment or reinforcement of the trochanter in any situation where the surgeon feels that the trochanter is at risk for detachment. The Mini Cleat is intended for use with the DALL-MILES System for trochanteric reattachment only.
The DALL-MILES Cables and Cable Sleeves are indicated for trochanteric reattachment and trauma surgery of the hip; to stabilize bone graft material; and for supplementary cerclage fixation with plates and screws for fracture fixation.
The DALL-MILES Trochanteric Grips and Grip Plates are indicated for use in the fixation of the greater trochanter due to trochanteric fracture or osteotomy with intramedullary fixation as the primary device.
The DALL-MILES Trochanteric Grip Plate is additionally indicated for use in the fixation of the greater trochanter due to extended trochanteric osteotomies.
Femoral Heads
The indications for use for total hip and hemi hip arthroplasty include:
- Noninflammatory degenerative joint disease including osteoarthritis and avascular necrosis;
- Rheumatoid arthritis;
- Correction of functional deformity;
- Revision procedures where other treatments or devices have failed; and,
- Treatment of nonunion, femoral neck and trochanteric fractures of the proximal femur with head involvement that are unmanageable using other techniques.
Femoral Mesh
The indications for use for total hip and hemi hip arthroplasty include:
- Noninflammatory degenerative joint disease including osteoarthritis and avascular necrosis;
- Rheumatoid arthritis;
- Correction of functional deformity;
- Revision procedures where other treatments or devices have failed; and,
- Treatment of nonunion, femoral neck and trochanteric fractures of the proximal femur with head involvement that are unmanageable using other techniques.
Surgical Mesh. Surgical Mesh is intended to be used to reinforce bone or tissue in any situation where additional strengthening and support is required due to poor bone/tissue quality.
Intramedullary Plug, Centralizer
The indications for use of total hip replacement prostheses include:
· noninflammatory degenerative joint disease including osteoarthritis and avascular necrosis;
· rheumatoid arthritis;
· correction of functional deformity;
· revision procedures where other treatments or devices have failed; and,
· treatment of nonunion, femoral neck and trochanteric fractures of the proximal femur with head involvement that are unmanageable using other techniques.
The EXETER Centralizer is intended to be used to centralize the femoral stem within the intramedullary canal and is intended to be used with bone cement.
The EXETER 2.5mm Intramedullary Bone Plug is intended to be used to restrict the migration of bone cement down the femoral canal and permit cement pressurization during total hip arthroplasty and is intended to be used with bone cement.
Exeter X3 RimFit Cups
The indications for use for total hip arthroplasty include:
- Painful, disabling joint disease of the hip resulting from: degenerative arthritis, rheumatoid arthritis, post-traumatic arthritis or late stage avascular necrosis.
- Revision of previous unsuccessful femoral head replacement, cup arthroplasty or other procedure.
- Clinical management problems where arthrodesis or alternative reconstructive techniques are less likely to achieve satisfactory results.
- Where bone stock is of poor quality or inadequate for other reconstructive techniques, such as cementless fixation, as indicated by deficiencies of the acetabulum.
The Exeter X3 RimFit Cup is intended for cemented use only.
Femoral Stems
The indications for use for total hip and hemi hip arthroplasty include:
- noninflammatory degenerative joint disease including osteoarthritis and avascular necrosis;
- rheumatoid arthritis;
- correction of functional deformity;
- revision procedures where other treatments or devices have failed; and,
- treatment of nonunion, femoral neck and trochanteric fractures of the proximal femur with head involvement that are unmanageable using other techniques.
The Exeter V40 Femoral Stem is intended for use in total or hemi hip replacement. It is intended for cemented use only.
The devices covered by this bundled submission are Stryker Total Hip Systems which include Dall-Miles cable system components, femoral heads, femoral mesh, Intramedullary Plug, Centralizer, acetabular cups, and femoral stems. All devices are commercially available and have been cleared in prior 510(k) submissions.
This FDA 510(k) letter and summary (K240418) is for a "Stryker Orthopaedics Hip Systems Labeling Update." This submission is not for a new medical device or an AI/software device that requires performance evaluation. Instead, it is a "Change Being Effected" bundled submission to add a contraindication to the labeling of existing, already cleared Stryker Orthopaedics Hip Systems.
Therefore, the provided document does not contain information about acceptance criteria or a study proving the device meets acceptance criteria. It explicitly states under "Non-Clinical Testing" and "Clinical Testing" that these were not required as a basis for substantial equivalence because the technological characteristics of the subject devices are identical to those of the predicate devices.
In summary, the request for information regarding acceptance criteria, device performance, sample sizes, ground truth, expert qualifications, adjudication methods, MRMC studies, or standalone performance is not applicable to this document, as it pertains to a labeling update for previously cleared devices, not a new device or a device requiring new performance data.
Ask a specific question about this device
Page 1 of 24