Search Results
Found 6 results
510(k) Data Aggregation
(49 days)
The Stryker Global Modular Replacement System, Modular Replacement System and Modular Rotating Hinge Knee Component are sterile, single-use devices intended for use in situations where there is a need for replacement of bone due to radical bone loss. This loss can be related to oncology, trauma or failed previous prosthesis. Specific Indications for Use are listed below.
Indications for the MRS Stems and Intercalary Stems (presented in K952970):
- This device is intended for use in patients requiring extensive reconstruction of the femur . and/or proximal tibia, including the hip or knee joint, resulting from extensive bone loss. Tumor resection for skeletal lesions (Oncology patients where radical bone resection and replacement may be required), revision surgery for failed arthroplasty, and acute trauma are the primary causes of extensive bone loss. These prostheses are intended for use with bone cement.
- The Intercalary System is intended for use in situations arising from femoral mid-shaft tumor resection, or for prosthetic knee fusion.
Indications for the Global Modular Replacement System (presented in K023087):
- Femoral and/or proximal tibial replacement and total femoral replacement in Oncology cases where radical resection and replacement of bone is required, and in limb salvage procedures where radical resection and replacement of the bone is required. Limb salvage procedures would include surgical intervention for severe trauma, failed previous prosthesis, and/or Oncology indications.
Indications for the GMRS Anteverted Proximal Femoral Component (presented in K032581):
- Femoral replacement in Oncology cases where radical resection and replacement of bone is required, and in limb salvage procedures where radical resection and replacement of bone is required. Limb salvage procedures would include surgical intervention for severe trauma, failed previous prosthesis, and/or Oncology indications. This component may also be used with the Distal Femoral segment components of the GMRS in total femoral replacement.
Indications for the Global Modular Replacement System Press Fit Stems with HA Coating (presented in K022403, and K031217):
Proximal femoral reconstruction secondary to:
o Trauma
o Failed previous prosthesis
o Tumor resection
Indications for the Modular Replacement System Cemented Stems (presented in K040749):
Femoral and/or proximal tibial replacement due to:
o Trauma
o Failed previous prosthesis
o Tumor resection
Indications for the Modular Rotating Hinge Knee System (presented in K002552):
- The Modular Rotating Hinge Knee System is intended for use with bone cement in cases where there is destruction of the joint surfaces, with or without significant bone deformity; the cruciate and/or collateral ligaments do not stabilize the knee joint; the ligaments are inadequate and/or the musculature is weak; and revision of a failed previous prosthesis where there is instability, with or without bone loss or inadequate soft tissue. Expanded indications include for use with bone cement in situations where there is extensive bone loss, including limb salvage procedures for severe trauma, failed previous prosthesis, and/or Oncology indications.
The Global Modular Replacement System (GMRS) is comprised of a number of components that are intended to be used in conjunction with each other, or in conjunction with components of the Modular Replacement System (MRS) or the Modular Rotating Hinge Knee System (MRH). The Modular Rotating Hinge Knee System is a tricompartmental knee system. The system consists of a stemmed femoral component and a stemmed tibial bearing component connected by a set of bushings and an axle. The MRH Tibial Rotating component is a part of the MRH system and mates with the GMRS Standard Proximal Tibia. These devices are intended to be used in clinical situations where there is radical bone loss of the femur and/or proximal tibia. This radical bone loss can be related to oncology (tumor), trauma or failed previous prosthesis.
The provided FDA 510(k) clearance letter and summary for the Stryker Orthopaedics Global Modular Replacement System, Modular Replacement System, and Modular Rotating Hinge Knee Component does not contain information regarding acceptance criteria or a study proving the device meets specific performance criteria.
This document is a clearance letter indicating a determination of "substantial equivalence" to previously marketed predicate devices. It does not present novel performance data or studies against pre-defined acceptance criteria.
The key statements from the document that confirm this are:
- "Non-Clinical Testing: Non-Clinical testing was not required as a basis for substantial equivalence." (Page 8)
- "Clinical Testing: Clinical testing was not required as a basis for substantial equivalence." (Page 8)
The submission focuses on adding a contraindication to the labeling, updating MRI Safety information, and making clerical updates to the Instructions for Use (IFUs). These changes do not necessitate new performance studies for substantial equivalence.
Therefore, I cannot populate the requested table or answer the specific questions about acceptance criteria and studies because the provided document explicitly states that such testing was not required for this particular 510(k) submission.
Ask a specific question about this device
(70 days)
Indications for the MRS Stems and Intercalary Stems (presented in K952970):
· This device is intended for use in patients requiring extensive reconstruction of the femur and/or proximal tibia, including the hip or knee joint, resulting from extensive bone loss. Tumor resection for skeletal lesions (Oncology patients where radical bone resection and replacement may be required), revision surgery for failed arthroplasty, and acute trauma are the primary causes of extensive bone loss. These prostheses are intended for use with bone cement.
· The Intercalary System is intended for use in situations arising from in femoral mid-shaft tumor resection. or for prosthetic knee fusion.
Indications for the Global Modular Replacement System (presented in K023087):
· Femoral and/or proximal tibial replacement and total femoral replacement in Oncology cases where radical resection and replacement of bone is required, and in limb salvage procedures where radical resection and replacement of the bone is required. Limb salvage procedures would include surgical intervention for severe trauma, failed previous prosthesis, and/or Oncology indications.
Indications for the GMRS Anteverted Proximal Femoral Component (presented in K032581)
· Femoral replacement in Oncology cases where radical resection and replacement of bone is required, and in limb salvage procedures where radical resection and replacement of bone is required. Limb salvage procedures would include surgical intervention for severe trauma, failed previous prosthesis, and/or Oncology indications. This component may also be used with the Distal Femoral segment components of the GMRS in total femoral replacement.
Indications for the Modular Replacement System Cemented Stems (cleared in K040749):
· Femoral and/or proximal tibial replacement due to:
- Trauma
- Failed previous prosthesis
- Tumor resection
The Global Modular Replacement System is comprised of a number of components that are intended to be used in conjunction with each other, or in conjunction with components of the Modular Replacement System or the Modular Rotating Hinge Knee System. These devices are intended to be used in clinical situations where there is radical bone loss of the femur and/or proximal tibia. This radical bone loss can be related to oncology (tumor), trauma or failed previous prosthesis. The subject Global Modular Replacement System components are unchanged since the last premarket notification for the devices in K222056.
This document is entirely focused on a 510(k) premarket notification for a medical device called the "Global Modular Replacement System." It is a request for approval to market the device, and the content details its intended use, a comparison to legally marketed predicate devices, and a summary of its technological characteristics.
Crucially, this document does not contain any information about acceptance criteria, device performance testing results, ground truth establishment, sample sizes for training or test sets, expert qualifications, or MRMC studies for an AI/software device. The request is specifically for a physical medical device (hip and knee prostheses) and its purpose is to update labeling, including adding a contraindication and updating MRI safety information.
The sections you've requested (acceptance criteria, study details, sample sizes, expert qualifications, ground truth, MRMC studies) are typically relevant for AI/Software as a Medical Device (SaMD) submissions where software algorithms are being evaluated for diagnostic or prognostic purposes.
Therefore, I cannot fulfill your request as the provided text does not contain the information you are asking for. It explicitly states:
"No additional testing was conducted for this submission, as the only changes being made are to modify the labeling to add a contraindication and an update to the MRI safety information to align with an FDA guidance document. Testing performed in the previously cleared premarket notifications is applicable to this submission." (Page 10)
This confirms that no new performance studies, as would be required for an AI/Software component, were performed or are detailed in this 510(k) submission.
Ask a specific question about this device
(198 days)
The Stryker Global Modular Replacement System components are sterile, single-use devices intended for use in situations where there is a need for replacement of bone due to radical bone loss. This loss can be related to oncology, trauma or failed previous prosthesis.
Specific Indications for Use are listed below.
Indications for the MRS Stems and Intercalary Stems presented in K952970:
- This device is intended for use in patients requiring extensive reconstruction of the femur and/or proximal tibia, including the hip or knee joint, resulting from extensive bone loss. Tumor resection for skeletal lesions (Oncology patients where radical bone resection and replacement may be required), revision surgery for failed arthroplasty, and acute trauma are the primary causes of extensive bone loss. These prostheses are intended for use with bone cement.
- . The Intercalary System is intended for use in situations arising from femoral mid-shaft turnor resection, or for prosthetic knee fusion.
Indications for the Global Modular Replacement System presented in K023087:
- . Femoral and/or proximal tibial replacement and total femoral replacement in Oncology cases where radical resection and replacement of bone is required, and in limb salvage procedures where radical resection and replacement of the bone is required. Limb salvage procedures would include surgical intervention for severe trauma, failed previous prosthesis, and/or Oncology indications.
Indications for the Global Modular Replacement System Press Fit Stems with HA Coating (presented in K022403, and K031217):
- Proximal femoral reconstruction secondary to: 0
- Trauma o
- Failed previous prosthesis о
- Tumor resection O
Indications for the GMRS Anteverted Proximal Femoral Component (presented in K032581)
-
Femoral replacement in Oncology cases where radical resection and replacement of bone is . required, and in limb salvage procedures where radical resection and replacement of bone is required. Limb salvage procedures would include surgical intervention for severe trauma, failed previous prosthesis, and/or Oncology indications. This component may also be used with the Distal Femoral segment components of the GMRS in total femoral replacement.
Indications for the Modular Replacement System Cemented Stems (cleared in K040749): -
Femoral and/or proximal tibial replacement due to: ●
- Trauma O
- Failed previous prosthesis O
- Tumor resection o
Indications for the Modular Rotating Hinge Knee System (cleared in K002552)
The Modular Rotating Hinge Knee System is intended for use with bone cement in cases where there is destruction of the joint surfaces, with or without significant bone deformity; the cruciate and/or collateral ligaments do not stabilize the knee joint; the ligaments are inadequate and/or the musculature is weak; and revision of a failed previous prosthesis where there is instability, with or without bone loss or inadequate soft tissue.
Expanded indications include for use with bone cement in situations where there is extensive bone loss, including limb salvage procedures for severe trauma, failed previous prosthesis, and/or Oncology indications.
The Global Modular Replacement System is comprised of a number of components that are intended to be used in conjunction with each other, or in conjunction with components of the Modular Replacement System or the Modular Rotating Hinge Knee System. These devices are intended to be used in clinical situations where there is radical bone loss of the femur and/or proximal tibia. This radical bone loss can be related to oncology (tumor), trauma or failed previous prosthesis.
This document is a 510(k) premarket notification from Howmedica Osteonics Corp., dba Stryker Orthopaedics, for their Global Modular Replacement System, Modular Replacement System, and Modular Rotating Hinge Knee. The submission, K222056, aims to modify the labeling of these devices to include MR Conditional language and to add an additional contraindication to the GMRS Press Fit Stems with Hydroxyapatite.
Here's an analysis of the provided text regarding acceptance criteria and supporting studies:
-
Table of Acceptance Criteria and Reported Device Performance:
The document does not explicitly present a table of acceptance criteria with corresponding device performance metrics for clinical outcomes. Instead, it focuses on non-clinical testing for MR compatibility and states that these tests were performed according to recognized ASTM standards. The "acceptance criteria" can be inferred as meeting the requirements of these standards.
Acceptance Criteria (Inferred) Reported Device Performance Compliance with ASTM F2052-15 for Magnetically Induced Displacement Force Testing performed as outlined in FDA guidance. Compliance with ASTM F2213-17 for Magnetically Induced Torque Testing performed as outlined in FDA guidance. Compliance with ASTM F2119-07 (2013) for Image Artifacts Testing performed as outlined in FDA guidance. Compliance with ASTM F2182-19e1 for Heating by RF Fields (including additional analyses for tissue heating) Testing performed as outlined in FDA guidance. -
Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective):
The document describes non-clinical testing performed on the devices themselves, not on human subjects. Therefore, the concept of a "test set" in the context of clinical data (e.g., patient cases) does not apply here. The "samples" would be the physical devices or components tested. The document does not specify the number of devices tested for each non-clinical evaluation (e.g., how many different GMRS components were subjected to MR compatibility testing). The data provenance is non-clinical laboratory testing, not human-derived data, and thus country of origin or retrospective/prospective status is not applicable in this context.
-
Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience):
This information is not applicable. Ground truth, in the sense of clinical diagnoses or outcomes, is not established for this type of non-clinical device testing. The "ground truth" for MR compatibility is determined by the physical properties of the materials and design, evaluated against established engineering standards like ASTM.
-
Adjudication method (e.g. 2+1, 3+1, none) for the test set:
This information is not applicable. Adjudication methods are used in clinical studies or expert reviews to resolve discrepancies in diagnoses or assessments from multiple reviewers. For non-clinical engineering tests, the outcome is determined by measurements against predefined tolerances within the ASTM standards.
-
If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:
No MRMC comparative effectiveness study was done. The submission is for a medical device (orthopedic implants), not an AI-powered diagnostic or assistive tool.
-
If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:
Not applicable. The device is a physical orthopedic implant, not an algorithm.
-
The type of ground truth used (expert consensus, pathology, outcomes data, etc.):
The "ground truth" for the non-clinical testing is based on established engineering principles and the requirements defined in the ASTM standards (e.g., limits for magnetically induced displacement, torque, image artifact levels, and localized heating).
-
The sample size for the training set:
Not applicable. This is not an AI/machine learning device; therefore, there is no "training set."
-
How the ground truth for the training set was established:
Not applicable. There is no training set for this type of device.
Ask a specific question about this device
(225 days)
Indications:
- Correction of revision of unsuccessful osteotomy, arthrodesis, or previous joint replacement.
- Tumor resections.
- Revision of previously failed total joint arthroplasty
- Trauma
The Compress Segmental Femoral Replacement System components are intended for uncemented use only.
The ComPreSs® Segmental Femoral Replacement System consists of three main components, the anchor plug, the spindle assembly, and the intercalary segment. The Anchor Plug is embedded within the medullary canal, the Spindle Assembly attaches to the Anchor Plug and is compressed against the bone / implant interface at the osteotomy site. The Intercalary Segment attaches to the Spindle Assembly, and completes the Femoral Stem Component.
This system is intended for use in both the proximal and distal femur.
This document does not contain information about acceptance criteria or a study that proves the device meets specific performance criteria. Instead, it is a 510(k) summary for a medical device (ComPreSs® Segmental Femoral Replacement System) seeking substantial equivalence to legally marketed predicate devices.
The key points from the provided text are:
- Device: ComPreSs® Segmental Femoral Replacement System
- Applicant: Biomet Manufacturing Corp.
- Date: August 5, 2005 (with a correction letter dated April 23, 2009)
- Purpose of Submission: To demonstrate substantial equivalence to predicate devices, not to prove specific performance against acceptance criteria in a new study.
- Basis for Equivalence: The submission states that "All components of this system are identical in material, processing, function, and design to their respective predicate components."
- Non-clinical Testing: "The engineering rationale with the attached literature and mechanical test reports support the claim for substantial equivalence to the predicate devices." (No specific acceptance criteria or performance numbers are given here).
- Clinical Testing: "Clinical testing in support of the predicate Compress® Distal Femoral Replacement System is sufficient to ensure safety and efficacy. No further clinical testing is necessary to support the claim for substantial equivalence to the predicate devices." This explicitly states that no new clinical study was performed for this specific device to establish novel acceptance criteria.
Therefore, the requested information regarding acceptance criteria, device performance, sample sizes, ground truth establishment, expert involvement, adjudication methods, or MRMC studies is not present in the provided text. The document focuses on demonstrating equivalence through comparison to existing, approved devices and their past testing, rather than presenting new performance data for the subject device against defined acceptance metrics.
Ask a specific question about this device
(382 days)
The Howmedica® Modular Replacement System - Distal Femoral/Proximal Tibial Segments are presently cleared for use with bone cement in Oncology patients where radical resection and replacement of the distal femur is required (reference premarket notification K952970). Howmedica seeks clearance to expand this indication to include use with bone cement in limb salvage procedures where radical resection and replacement of the distal femur/proximal tibia is required. Limb salvage procedures would include surgical intervention for severe trauma, failed previous knee arthroplasties, and/or Oncology indications. All of these surgical interventions have in common the need for radical resection of the distal femur/proximal tibia, and the need for a prosthetic replacement. This device achieves fixation by the use of bone cement.
The Howmedica® Modular Replacement System - Distal Femoral/Proximal Tibial Segments are presently cleared for use with bone cement in Oncology patients where radical resection and replacement of the distal femur/proximal tibia is required (reference premarket notification K952970). Howmedica seeks clearance to expand this indication to include use with bone cement in limb salvage procedures where radical resection and replacement of the distal femur/proximal tibia is required. Limb salvage procedures would include surgical intervention for severe trauma, failed previous knee arthroplasties, and/or Oncology indications. All of these surgical interventions have in common the need for radical resection of the distal femur/proximal tibia, and the need for a prosthetic replacement. This device achieves fixation by the use of bone cement.
This document is a 510(k) summary for a medical device (Howmedica® Modular Replacement System - Distal Femoral/Proximal Tibial Segments) seeking to expand its indications for use. It does not contain information about acceptance criteria or a study proving the device meets acceptance criteria.
Instead, it focuses on demonstrating substantial equivalence to previously marketed devices (a process for regulatory clearance, not clinical performance evaluation). The only "testing" mentioned is for "modular tapers," but no details on acceptance criteria or results are provided.
Therefore, I cannot provide the requested information from the given text.
Here's why and what's missing:
- Acceptance Criteria & Device Performance: The document states "A comparison to other legally marketed products was made, and testing of the modular tapers was presented." This is incredibly vague. It doesn't specify what the acceptance criteria for this testing were, nor does it present any numeric or descriptive performance data.
- Sample Size & Data Provenance (Test Set): No information is provided about any test set used to evaluate clinical or technical performance.
- Number & Qualifications of Experts (Ground Truth): No experts are mentioned in the context of establishing ground truth for performance evaluation.
- Adjudication Method: Not applicable as no ground truth establishment is described.
- MRMC Comparative Effectiveness Study: No such study is mentioned or implied. The device is a physical implant, not an AI/software device that would typically undergo MRMC studies for human-in-the-loop improvement.
- Standalone Performance Study: As above, no such study is mentioned for clinical performance. The focus is on demonstrating substantial equivalence based on materials and design comparison, and some unspecified "testing of modular tapers."
- Type of Ground Truth: Not applicable, as no performance study is detailed.
- Sample Size for Training Set: Not applicable, as this is a physical device approval, not a machine learning model.
- Ground Truth for Training Set: Not applicable.
In summary, the provided text describes a regulatory submission for expanding the indications of a medical implant and does not contain the detailed performance study information requested.
Ask a specific question about this device
(86 days)
The Kinemax® Superstabilizer (K904208), Kinematic® II Modular Condylar and Stabilizer Inserts (K871349), PCA® Modular Inserts (K894403), Kinematio® Rotating Hinge Knee (K792089), Duracon® Unicompartmental Knee (K 926231), and the Modular Replacement Metal Encapsulated Components (K952970) are intended to be used with the PCA®, Duracon®, or Kinemax® family of femoral components, tibial baseplates, and patellar components in primary or revision cemented total knee arthroplasty.
This submission describes additional UHMWPE Knee components that can be sterilized by the Duration® Stabilized process previously cleared in submission K936292, as an alternate to standard air irradiated gamma sterilization. Corresponding wear claims, previously cleared in submission K 965173 and listed below, are also applicable to these additional components. This submission includes the Kinemax® Superstabilizer (K904208), the Kinematic® II Modular Condylar and Stabilizer Inserts (K871349), the PCA Modular Inserts(K894403), the Kinematic® Rotating Hinge Knee (K792089), the Duracon® Unicompartmental Knee (K926231), and the Modular Replacement Metal Encapsulated Components (K952970). These components are intended to be used with the PCA®, Duracon®, or Kinemax® family of femoral components, tibial baseplates, and patellar components in primary or revision cemented total knee arthroplasty.
This 510(k) summary describes the Duraction® Stabilized UHMWPE Knee Components and their wear claims. The submission details three in vitro wear tests comparing the Duraction® Stabilized UHMWPE with conventionally gamma sterilized UHMWPE.
1. Table of Acceptance Criteria and Reported Device Performance
Acceptance Criteria (Implied) | Reported Device Performance (Reduction in Volumetric Wear) | Study Type | Articulating Counterface | Lubricant | Cycles (Duration) |
---|---|---|---|---|---|
Significant Reduction in Volumetric Wear (vs. conventionally sterilized) | 30% reduction | Ring-on-Block | CoCr circular disk | Bovine calf serum | Over 5 million cycles |
Significant Reduction in Volumetric Wear (vs. conventionally sterilized) | 68% reduction | Pin-on-Disk | CoCr cylindrical pin (1" spherical end) | Bovine calf serum | Over 4 million cycles |
Significant Reduction in Volumetric Wear after aging (vs. aged conventionally sterilized) | 91% reduction | Pin-on-Disk (Aged) | CoCr cylindrical pin (1" spherical end) | Bovine calf serum | Over 2.5 million cycles (after 23 days simulated aging) |
Note: The document does not explicitly state "acceptance criteria" but rather presents the results as "wear claims" based on the performed in vitro tests. The implied acceptance criterion for each test is a demonstrated reduction in volumetric wear compared to the conventionally gamma sterilized UHMWPE.
2. Sample Size Used for the Test Set and Data Provenance
The document describes material-level testing rather than patient or imaging data.
- Sample Size:
- Ring-on-Block Test: "A block of Howmedica's Duration® Stabilized UHMWPE" and "the same block of Howmedica's conventionally gamma sterilized UHMWPE." (Implies n=1 for each material type for the block test, though likely multiple replicates were used within the block).
- Pin-on-Disk Tests (both non-aged and aged): "A 9mm thick circular disk of Howmedica's Duration® Stabilized UHMWPE" and "the same circular disk of Howmedica's conventionally gamma sterilized UHMWPE." (Implies n=1 for each material type for the disk test, though likely multiple replicates were used).
- Data Provenance: The tests are described as "in vitro tests" and "independent laboratory test(s)." The country of origin of the data is not specified, but the manufacturer is Howmedica Inc. in the USA. The data is prospective for these specific tests as they were conducted to support the submission.
3. Number of Experts Used to Establish the Ground Truth for the Test Set and the Qualifications of Those Experts
Not applicable. This is a materials science engineering test, not a medical device performance assessment based on expert interpretation (e.g., radiology reads). The "ground truth" is measured volumetric wear.
4. Adjudication Method for the Test Set
Not applicable. This is a direct measurement of material wear, not a consensus-based assessment.
5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
Not applicable. This is an in vitro materials test, not a clinical study involving human readers or AI.
6. If a Standalone (i.e. algorithm only without human-in-the loop performance) was done
Not applicable. This refers to the performance of the UHMWPE material itself, not an algorithm.
7. The Type of Ground Truth Used
The ground truth used for these tests is measured volumetric wear of the material samples under controlled in vitro laboratory conditions.
8. The Sample Size for the Training Set
Not applicable. This is not a machine learning or AI device that requires a training set. The "training" in this context would refer to the development and optimization of the UHMWPE material and sterilization process, which is not detailed in terms of sample sizes for individual material property tests.
9. How the Ground Truth for the Training Set Was Established
Not applicable. (See point 8). The "ground truth" for material development would involve various material characterization techniques and performance testing during the R&D phase, but this specific 510(k) focuses on the final wear claims derived from the described in vitro tests.
Ask a specific question about this device
Page 1 of 1