Search Filters

Search Results

Found 46 results

510(k) Data Aggregation

    K Number
    K173352

    Validate with FDA (Live)

    Date Cleared
    2018-02-15

    (113 days)

    Product Code
    Regulation Number
    882.5600
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use
    1. The Trevo Retriever is indicated for use to restore blood flow in the neurovasculature by removing thrombus for the treatment of acute ischemic stroke to reduce disability in patients with a persistent, proximal anterior circulation, large vessel occlusion, and smaller core infarcts who have first received intravenous tissue plasminogen activator (IV t-PA). Endovascular therapy with the device should start within 6 hours of symptom onset.

    2. The Trevo Retriever is intended to restore blood flow in the neurovasculature by removing thrombus in patients experiencing ischemic stroke within 8 hours of symptom onset. Patients who are ineligible for intravenous tissue plasminogen activator (IV t-PA) or who fail IV t-PA therapy are candidates for treatment.

    3. The Trevo Retriever is indicated for use to restore blood flow in the neurovasculature by removing thrombus for the treatment of acute ischemic stroke to reduce disability in patients with a persistent, proximal anterior circulation, large vessel occlusion of the internal carotid artery (ICA) or middle cerebral artery (MCA)-M1 segments with smaller core infarcts (0-50 cc for age < 80 years, 0-20 cc for age ≥ 80 years). Endovascular therapy with the device should start within 6-24 hours of time last seen well in patients who are ineligible for intravenous tissue plasminogen activator (IV t-PA) or who fail IV t-PA therapy.

    Device Description

    The Trevo Retriever family, including the Trevo ProVue and Trevo XP ProVue Retriever consists of a flexible, tapered core wire with a shaped section at the distal end. Platinum markers at the distal end allow fluoroscopic visualization. In addition, the shaped section is also radiopaque. Retriever dimensions are indicated on product label. The Retriever has a hydrophilic coating to reduce friction during use. The Retriever has a shaft marker to indicate proximity of Retriever tip relative to Microcatheter tip. A torque device is provided with the Retriever to facilitate manipulation. The torque device is used to lock the core wire to the microcatheter during the procedure. Locking of the torque device to the wire allows the microcatheter and Retriever to be retracted as a system during clot retrieval. An insertion tool is provided to introduce the Retriever into a Microcatheter. The Insertion Tool is a sheath in which the Retriever comes preloaded. Once half the retriever's length is inserted into the microcatheter, the insertion tool is removed. Retrievers have a modified proximal end that permits attachment of the Abbott Vascular DOC Guide Wire Extension (REF 22260). Joining Guide Wire Extension to Retriever facilitates removal or exchange of a catheter while maintaining Retriever position in anatomy. After exchange has been completed, the extension can be detached.

    AI/ML Overview

    Here's a breakdown of the acceptance criteria and study information for the Trevo Retriever, based on the provided document:

    1. Table of Acceptance Criteria and Reported Device Performance:

    The document doesn't explicitly state "acceptance criteria" in a bulleted or numbered list with specific thresholds. Instead, the study aims to demonstrate superior clinical outcomes for the treatment arm compared to the control arm across several endpoints. The success thresholds are defined within the study design.

    Performance Metric (Acceptance Criteria)Reported Device Performance (Treatment Arm vs. Control Arm)
    Primary Effectiveness:
    90-day Utility Weighted Modified Rankin Scale (UW-mRS): Posterior probability the Trevo thrombectomy increases expected scores relative to medical management alone, with a threshold for success of at least 0.986.Mean UW-mRS: 5.5 (Treatment) vs. 3.4 (Control) Posterior mean treatment benefit, core-adjusted: 2.0 (95% credible interval 1.1 to 3.0) Probability of superiority: > 0.9999
    90-day Dichotomized mRS (0-2 vs. 3-6) (Functional Independence): Success based on the same model and hierarchical fashion as UW-mRS.Functional Independence (mRS 0-2): 48.6% (Treatment) vs. 13.1% (Control) Posterior treatment benefit, core-adjusted: 32.8% (95% credible interval 21.1% to 44.1%) Probability of superiority: > 0.9999
    Primary Safety:
    Incidence of stroke-related mortality at 90 days: No significant difference between study arms.Treatment Arm: 15.9% (17/107) vs. Control Arm: 18.2% (18/99) Difference: -2.3% [-12.6%, 8.0%] p-value = 0.7126 (No significant difference)
    Secondary Effectiveness:
    Early Neurological Response (Day 5-7/Discharge): Defined as NIHSS drop of ≥ 10 points or NIHSS score of 0 or 1.Treatment Arm: 47.7% vs. Control Arm: 19.2% Mean Absolute Difference: 28.5% (16.2%, 40.7%) Risk Ratio: 2.5 (1.6, 3.9) P-Value: <0.001 (Superiority for Treatment Arm)
    Revascularization Rates at 24 Hours: Presence of partial or complete recanalization.Treatment Arm: 76.6% (82/107) vs. Control Arm: 38.4% (38/99) Difference: 40.2% [27.1%, 51.5%] p-value: < 0.0001 (Superiority for Treatment Arm)
    Secondary Safety:
    Incidence of sICH (by ECASS III definition) within 24 hours post-randomization: No statistical difference between arms.Treatment Arm: 5.6% (6/107) vs. Control Arm: 3.0% (3/99) Difference: 2.6% [-2.9%, 8.1%] p-value = 0.5011 (No statistical difference)
    Incidence of Neurological Deterioration (≥ 4-point increase in NIHSS from baseline through Day 5-7/Discharge):Treatment Arm: 14.0% (15/107) vs. Control Arm: 26.3% (26/99) Difference: -12.2% [-23.1%, -1.4%] p-value = 0.0358 (Less frequently in Treatment Arm)
    All-Cause Mortality at 90 days: No significant difference between study arms.Treatment Arm: 18.7% (20/107) vs. Control Arm: 18.2% (18/99) Difference: 0.5% [-10.1%, 11.1%] p-value = 1.0000 (No statistical difference)
    Serious Adverse Events (Procedure-related and Device-related): No new safety risks identified, comparable across treatment groups.Vascular perforation: 0.9% (Treatment) Intramural arterial dissection: 1.9% (Treatment) Embolization to a new territory: 3.7% (Treatment) Full list in Table 11 & 12 No new safety risks identified, comparable risk profile to predicate.

    2. Sample Size Used for the Test Set and Data Provenance:

    • Test Set Sample Size: 206 subjects (107 in the Treatment Arm, 99 in the Control Arm).
    • Data Provenance: The DAWN™ study (IDE G130223) was a "multi-center, ... conducted in multiple countries." The document doesn't specify the exact countries, but it indicates an international scope. The study was prospective, randomized, open, masked endpoint trial.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Their Qualifications:

    The document mentions "masked endpoint trial" and "CEC adjudication" (Clinical Events Committee). While it confirms the involvement of a committee for adjudication, it does not specify the exact number of experts or their specific qualifications (e.g., number of years of experience, subspecialty) for establishing the ground truth.

    4. Adjudication Method for the Test Set:

    The exact adjudication method (e.g., 2+1, 3+1) is not explicitly stated. However, it refers to "CEC adjudication," indicating that a Clinical Events Committee independently reviewed and determined certain outcomes (e.g., sICH, neurological deterioration, all-cause mortality, serious adverse events). This typically implies a consensus-based approach among committee members.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study:

    A multi-reader multi-case (MRMC) comparative effectiveness study was not performed as described in the document. The study was a randomized controlled trial comparing the Trevo thrombectomy device plus medical management to medical management alone. This is a direct patient outcome study, not a study of human readers' performance with or without AI assistance. The Trevo Retriever is a medical device for mechanical thrombectomy, not an AI diagnostic tool.

    6. Standalone (Algorithm Only Without Human-in-the-Loop Performance) Study:

    This question is not applicable as the Trevo Retriever is a physical medical device (neurovascular mechanical thrombectomy device), not a standalone algorithm or AI. Its performance is directly measured in terms of patient outcomes and procedural metrics.

    7. Type of Ground Truth Used:

    The ground truth used in the DAWN study was based on patient outcomes data, specifically:

    • Clinical Outcomes: Modified Rankin Scale (mRS) at 90 days (utility-weighted and dichotomized).
    • Neurological Assessment: NIHSS scores (for early neurological response and deterioration).
    • Imaging Data: Recanalization rates (CTA or MRA at 24 hours), core infarct volume (for inclusion criteria and statistical adjustment).
    • Safety Endpoints: Stroke-related mortality, sICH (ECASS III definition), all-cause mortality, and adverse events (adjudicated by CEC).

    8. Sample Size for the Training Set:

    The document describes a clinical trial (DAWN study) to evaluate the device in general, not specifically an AI model with a separate training and test set. Therefore, there is no "training set" for an AI algorithm mentioned in the context of this device's premarket notification. The entire DAWN study cohort (206 subjects) served as the dataset for evaluating the device's effectiveness and safety.

    9. How the Ground Truth for the Training Set Was Established:

    As there is no "training set" for an AI algorithm, this question is not applicable. The clinical trial data was gathered through the rigorous protocol of the DAWN study, including masked endpoint assessment and CEC adjudication for relevant clinical and safety outcomes, which served as the "ground truth" for evaluating the device's performance in comparison to standard medical management.

    Ask a Question

    Ask a specific question about this device

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Trevo Retrievers are indicated for use to restore blood flow in the neurovasculature by removing thrombus for the treatment of acute ischemic stroke to reduce disability in patients with a persistent, proximal anterior circulation, large vessel occlusion, and smaller core infarcts who have first received intravenous tissue plasminogen activator (IV t-PA). Endovascular therapy with the device should start within 6 hours of symptom onset.

    Device Description

    The Trevo ProVue and XP ProVue Retrievers consist of a flexible, tapered core wire with a shaped self-expanding section at the distal end for clot capture and removal. Radiopaque platinum wires in the shaped section and radiopaque markers on the distal end allow fluoroscopic visualization. The Trevo Retrievers have a hydrophilic coating to reduce friction during use. A torque device and an insertion tool are provided with the Retrievers. The Trevo Retrievers are delivered to the site of occlusion in the neurovasculature through a microcatheter. The torque device may be used to lock the core wire of the Trevo Retriever to the microcatheter during the procedure, allowing the Trevo Retriever and microcatheter to be retracted as a system through the guide catheter and removed from the body with captured clot.

    AI/ML Overview

    The acceptance criteria for the Trevo ProVue and XP ProVue Retrievers, a neurovascular mechanical thrombectomy device, are derived from the clinical study, bench testing, animal studies, and various regulatory requirements. The study supporting these criteria is primarily the MR CLEAN (Multicenter Randomized Clinical Trial of Endovascular Treatment for Acute Ischemic Stroke in the Netherlands) trial, with additional non-clinical and animal studies.

    Here's a breakdown of the acceptance criteria and reported device performance:

    1. Table of Acceptance Criteria and Reported Device Performance

    Note: The document does not explicitly state "acceptance criteria" for each performance metric with specific thresholds (e.g., "must achieve X%"). Instead, the assessment is based on demonstrating safety and effectiveness as compared to medical management alone, or meeting benchmarks set by non-clinical testing. Where specific targets are not given, the trial's statistically significant improvement or demonstration of comparable safety/effectiveness serves as the "meeting" of the criteria.

    CategoryAcceptance Criteria (Implied / Explicit)Reported Device Performance
    EffectivenessPrimary Effectiveness Endpoint: Clinically meaningful functional independence (mRS ≤ 2) at 90 days, demonstrating superiority over Medical Management (MM) alone.29.2% (28/96) of patients in the Trevo Retriever group achieved mRS ≤ 2 at 90 days, compared to 19.3% (48/249) in the MM group. Odds Ratio [95% CI]: 1.88 [1.07, 3.29], p-value = 0.014. (Demonstrated statistically significant superiority).
    Secondary Effectiveness Endpoint 1: Percentage of patients with no intra-cranial occlusion after 24 hours.77.5% (62/80) in the Trevo group achieved no intra-cranial occlusion, compared to 33.51% (65/194) in the MM group. (Demonstrated significantly higher rates of recanalization).
    Secondary Effectiveness Endpoint 2: Percentage of Trevo patients with recanalization (TICI ≥ 2a).81.3% (78/96) of Trevo patients achieved TICI ≥ 2a.
    Secondary Effectiveness Endpoint 3: Neurological outcomes assessed by NIHSS at 24 hours and 5-7 days post-randomization, demonstrating improvement compared to MM alone.NIHSS (24 Hours): Trevo Mean ± SD: 14.23 ± 9.23 (N=96) vs. MM Mean ± SD: 16.19 ± 7.77 (N=240). NIHSS (5-7 Days): Trevo Mean ± SD: 13.02 ± 12.16 (N=94) vs. MM Mean ± SD: 15.76 ± 11.49 (N=237). (Directionally favorable for Trevo, though statistical significance for these endpoints isn't explicitly stated).
    SafetyPrimary Safety Endpoint: All-cause mortality rate at 90 days, demonstrating non-inferiority or reduced mortality compared to MM alone.Trevo FDA Cohort: 13.3% (16/120) mortality. MM: 22.89% (57/249) mortality. (Lower mortality rate in Trevo group).
    Secondary Safety Endpoint 1: Proportion of patients with symptomatic intracranial hemorrhage (sICH) within 24 hours.Trevo FDA Cohort: 6.67% (8/120) sICH. MM: 4.42% (11/249) sICH. (Slightly higher incidence in Trevo group, but likely considered acceptable within the benefit-risk profile).
    Secondary Safety Endpoint 2: Proportion of patients with neurological deterioration within 5-7 days or discharge.Trevo FDA Cohort: 9.17% (10/109) . MM: 9.57% (22/230). (Comparable rates).
    Other Adverse Events: Monitor rates of various SAEs and adverse events to ensure an acceptable safety profile.Comprehensive reporting of SAEs (Table 10) and MedDRA coded AEs (Table 11) indicates generally comparable or acceptable rates across various categories, with some expected differences due to the invasive nature of the procedure (e.g., higher haemorrhage intracranial in Trevo, but also higher "other infection" and "other complication"). The overall benefit-risk determination concluded that benefits outweigh risks.
    BiocompatibilityDevice materials demonstrated to be biocompatible (ISO 10993-1:2009/AC:2010 compliance).Leveraged from prior submissions (K120961). Tests included hemocompatibility/coagulation, hemolysis, cytotoxicity, maximization test for delayed hypersensitivity, intracutaneous reactivity, acute systemic toxicity, and material-mediated rabbit pyrogenicity. (Met standard).
    SterilitySterility assurance level (SAL) of 10^-6 (ISO 11135-1:2007). Sterilant residuals per ISO 10993-7:2008. Endotoxin testing <2.15 EU/device (USP <161>).Achieved SAL of 10^-6. Sterilant residuals tested. Endotoxin testing met criterion. (Met standards).
    Shelf-Life24-month shelf-life supported by accelerated and real-time shelf-life testing, including packaging integrity and component functional performance.Supported by accelerated testing (Trevo XP ProVue) and real-time testing (Trevo ProVue), including: dimensional verification, tensile strength, radial force, tip flexibility, torque/tensile durability, retriever platinum wire and joint durability, radiopacity, deliverability, simulated use, coating integrity, and particulate testing. Packaging integrity tested to ASTM D4169, F1980, F1929, F2906, F88/M. (Met standard).
    Performance (Bench)Device performs as intended under anticipated conditions of use (Table 2 tests: Dimensional Verification, Tensile Strength, Radial Force, Tip Flexibility, Torque/Tensile Durability, Platinum Wire Joint Strength/Durability, Radiopacity, Deliverability, Particulate Evaluation, Coating Integrity Evaluation, Simulated Use).All listed bench tests were conducted and passed: Verified dimensions; recorded peak tensile force; constrained and released shaped section to record radial force; flexed distal tip to record compression/flex force; applied rotations and tensile cycles for durability; recorded peak tensile force for platinum wires; performed wrap/unwrapped and deploy/reload cycles; assessed radiopacity visually; measured force to push device through tortuous model; measured particulates; performed visual assessment of coating integrity before/after simulated use; simulated use in silicone neurovascular model with thrombus. (All tests passed, indicating performance as intended).
    Performance (Animal)Safety (vessel response) demonstrated in animal studies, including no arterial dissection, perforation, or thrombosis.Acute and chronic swine studies (up to 6 passes). Acute study: no evidence of vessel dissection, perforation, or thrombosis at Day 0. Chronic study: no evidence of dissection, perforation, or thrombosis at Day 0, and no stenosis, vessel irregularity, intimal flap or pseudoaneurysm at Day 30. Histopathology showed arterial healing consistent with routine catheterization. (Demonstrated safety in animal models).

    2. Sample Sizes Used for the Test Set and Data Provenance

    The primary clinical study supporting the device's expanded indication for the Trevo Retrievers was the MR CLEAN trial.

    • Test Set Sample Size:

      • TREVO FDA Cohort: 120 patients (from the IAT arm, met specific criteria) (N=120 for primary safety; N=96 for primary effectiveness after exclusions).
      • Medical Management (MM) Control Group: 249 patients (after exclusions for IV t-PA timing).
      • Total Patients in MR CLEAN (originally randomized): 500 patients (233 IAT, 267 control).
    • Data Provenance:

      • Country of Origin: The Netherlands (MR CLEAN was conducted in every endovascular hospital center in the Netherlands).
      • Retrospective or Prospective: Prospective, randomized, open-label, controlled, multicenter trial.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Their Qualifications

    The document mentions expert review for several endpoints:

    • Recanalization and Occlusion Assessment (Table 4 & 5):
      • Number of Experts: "Consensus review by up to three readers" for occlusion assessment (Table 4). "Consensus review by at least three readers" for TICI score recanalization (Table 5).
      • Qualifications: The document does not explicitly state the specific qualifications (e.g., years of experience, subspecialty) of these readers/experts. However, given the nature of the study (neurovascular stroke treatment), it is highly probable they were experienced neurologists, neuroradiologists, or stroke specialists involved in the study's clinical centers.

    4. Adjudication Method for the Test Set

    • Adjudication Method: "Consensus review by up to three readers" or "at least three readers" was used for evaluating arterial occlusion and recanalization (TICI scores). This implies a method where multiple experts independently (or collectively) assess the imaging and arrive at a joint decision. The specific process (e.g., if there were disagreements, how they were resolved beyond just "consensus") is not detailed beyond "consensus review."

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done, and the Effect Size of Human Reader Improvement

    • No, a typical MRMC comparative effectiveness study, as often seen for diagnostic AI, was not performed in this context.
      • The MR CLEAN trial directly compared patient outcomes with and without the device (with medical management), rather than evaluating how human readers' diagnostic accuracy or treatment decisions improved with AI assistance.
      • The Trevo Retrievers are a mechanical device for thrombectomy, not a diagnostic AI algorithm intended to assist human readers in interpreting images. Therefore, the concept of "how much human readers improve with AI vs. without AI assistance" does not directly apply to this device's clinical evaluation.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Study Was Done

    • Yes, in essence, the "device" (Trevo Retrievers) was evaluated in a standalone manner as a treatment in the context of the clinical trial.
      • The MR CLEAN trial evaluated the performance of the neurovascular mechanical thrombectomy device (Trevo Retrievers) in conjunction with best medical management against best medical management alone. This is effectively a "standalone" evaluation of the device's therapeutic effect, rather than an "algorithm only" evaluation, as it's a physical medical device.
      • The device's performance is measured by its direct impact on patient outcomes when used by clinicians, not as an interpretive AI tool.

    7. The Type of Ground Truth Used

    • Clinical Outcomes/Expert Assessment (Blinded):
      • Primary Effectiveness (mRS): The 90-day mRS assessment was performed by an assessor who was blinded to the subject's treatment allocation. This is a high-quality method for establishing ground truth for functional independence.
      • Imaging-based Endpoints (Occlusion/Recanalization): Assessed by a consensus review by up to three/at least three readers (presumably domain experts) after 24 hours using CTA or MRA and the AOL/TICI scales.
      • Safety Endpoints: Based on collected adverse events, neurological assessments (NIHSS), and mortality data, which are objective clinical measures or physician-reported events.
      • Animal Studies: Safety ground truth established by intra-procedural angiography and histopathology assessments performed by experts.

    8. The Sample Size for the Training Set

    • Not Applicable in the traditional sense for an AI/algorithm: The information provided describes the regulatory approval of a physical medical device (mechanical thrombectomy retriever), not an AI algorithm. Therefore, there is no "training set" for an AI model.
    • The clinical data from the MR CLEAN trial serves as the pivotal clinical evidence supporting the device's effectiveness and safety, analogous to a validation set for an AI, but it's not a "training set" for an algorithm.

    9. How the Ground Truth for the Training Set Was Established

    • Not Applicable: As explained above, this is a physical medical device, not an AI algorithm, so there is no "training set" or ground truth establishment process for an AI model. The "ground truth" for the device's performance is established through rigorous clinical trials and non-clinical testing as described in the previous sections.
    Ask a Question

    Ask a specific question about this device

    K Number
    K153729

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2016-01-27

    (30 days)

    Product Code
    Regulation Number
    870.1250
    Age Range
    All
    Reference & Predicate Devices
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Flow Gate Balloon Guide Catheters are indicated for use in facilitating the insertion and guidance of an intravascular catheter into a selected blood vessel in the peripheral and neuro vascular systems. The balloon provides temporary vascular occlusion during these and other angiographic procedures. The Balloon Guide Catheter is also indicated for use as a conduit for Retrieval devices.

    Device Description

    Like the predicate device, the 8F FlowGate2™ Balloon Guide Catheter is a coaxial- lumen, braidreinforced, variable stiffness catheter designed for use in facilitating the insertion and guidance of an intravascular catheter into a selected blood vessel in the peripheral and neuro vascular systems. A radiopaque marker is included on the distal end for angiographic visualization. A compliant balloon is mounted on the distal end to provide temporary vascular occlusion during angiographic procedures. A bifurcated luer hub on the proximal end allows attachments for flushing, inflation and aspiration. Modified FlowGate™ Balloon Guide Catheter dimensions and maximum recommended balloon inflation volume are indicated on product label.

    AI/ML Overview

    The provided text describes a 510(k) premarket notification for the 8F FlowGate2™ Balloon Guide Catheter. It outlines the acceptance criteria and testing performed to demonstrate its substantial equivalence to predicate devices.

    1. Table of Acceptance Criteria and Reported Device Performance:

    The document broadly states that all performed tests "Met established acceptance criteria." However, it does not provide specific quantitative acceptance criteria or detailed numerical results for each test. Instead, it offers a list of tests and a blanket statement of compliance.

    TestReported Device Performance
    Modifications to catheter shaft
    Catheter Simulated Use (Retriever/Microcatheter Compatibility)Met established acceptance criteria.
    Catheter Kink Resistance, Variables (Test Kink Resistance)Met established acceptance criteria.
    Catheter Tip Flexibility (Catheter/Vessel Interaction)Met established acceptance criteria.
    Torque TransmissionMet established acceptance criteria.
    High Pressure LeakMet established acceptance criteria.
    Distal Shaft TrackabilityMet established acceptance criteria.
    Lumen CompatibilityMet established acceptance criteria.
    6F Dilator Accessory
    Dimensional VerificationMet established acceptance criteria.
    Dilator Shape RetentionMet established acceptance criteria.
    Air Leak ResistanceMet established acceptance criteria.
    Liquid Leak ResistanceMet established acceptance criteria.
    Tensile StrengthMet established acceptance criteria.
    High Pressure LeakMet established acceptance criteria.
    Catheter/Vessel Interaction (Catheter Tip Flexibility)Met established acceptance criteria.
    Kink ResistanceMet established acceptance criteria.
    Tuohy Borst Valve with Sideport Accessory
    Hub GaugingMet established acceptance criteria.
    Liquid LeakageMet established acceptance criteria.
    Air LeakageMet established acceptance criteria.
    Separation ForceMet established acceptance criteria.
    Unscrewing TorqueMet established acceptance criteria.
    Ease of AssemblyMet established acceptance criteria.
    Resistance to OverridingMet established acceptance criteria.
    Stress CrackingMet established acceptance criteria.
    Catheter Simulated Use (Tuohy Borst Valve with Sideport seal over devices and wires)Met established acceptance criteria.
    BiocompatibilityMet biological safety requirements per EN ISO 10993-1: 2009 for externally communicating medical devices with circulating blood contact for less than 24 hours.
    SterilizationAdopted into a validated sterilization process in accordance with ISO 11135:2007 and AAMI TIR 28:2009.

    2. Sample Size Used for the Test Set and Data Provenance:

    The document does not specify the sample sizes (number of devices or cases) used for the verification and validation (test set) activities. It also does not explicitly state the data provenance (e.g., country of origin, retrospective/prospective). The testing appears to be primarily bench testing performed on manufactured devices.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts:

    This information is not provided in the document. The study described is a technical verification and validation of a medical device's physical and functional properties, not a clinical study involving human assessment of data requiring expert ground truth establishment in the typical sense (e.g., for diagnostic accuracy).

    4. Adjudication Method for the Test Set:

    Not applicable. The tests performed are objective technical assessments (e.g., measuring leak resistance, tensile strength, kink resistance), not subjective assessments requiring expert adjudication.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done, and the Effect Size:

    No, an MRMC comparative effectiveness study was not done. This submission is for a percutaneous catheter, not a diagnostic or AI-assisted interpretation device that would typically involve human readers. The study focuses on demonstrating the substantial equivalence of the new device to existing predicate devices through material, design, and performance testing.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done:

    Not applicable. This device is not an algorithm or AI system. It is a physical medical device (catheter). The testing focuses on the physical and functional performance of the device itself.

    7. The Type of Ground Truth Used:

    The "ground truth" for the tests performed is based on established engineering principles, industry standards (e.g., ISO 11135, EN ISO 10993-1), and internal performance specifications determined through design and risk assessment processes. For instance, for "High Pressure Leak," the ground truth is a pass/fail criterion based on whether the device exhibits leakage under specific pressure. These are objective measures rather than subjective expert consensus, pathology, or outcomes data in the context of diagnostic performance.

    8. The Sample Size for the Training Set:

    Not applicable. This is not an AI or machine learning device that requires a training set.

    9. How the Ground Truth for the Training Set Was Established:

    Not applicable, as there is no training set for this device.

    Ask a Question

    Ask a specific question about this device

    K Number
    K150616

    Validate with FDA (Live)

    Date Cleared
    2015-06-12

    (94 days)

    Product Code
    Regulation Number
    870.1250
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Trevo XP ProVue Retriever (4x30mm) is intended to restore blood flow in the neurovasculature by removing thrombus in patients experiencing ischemic stroke within 8 hours of symptom onset. Patients who are ineligible for intravenous tissue plasminogen activator (IV t-PA) or who fail IV t-PA therapy are candidates for treatment.

    Device Description

    The Trevo XP ProVue Retriever (4x30mm) consists of a flexible, tapered core wire with a shaped section at the distal end. It is designed to restore blood flow in the neurovasculature by removing thrombus in patients experiencing ischemic stroke. Radiopaque platinum wires in the shaped section and radiopaque markers on the distal end allow fluoroscopic visualization. Retriever dimensions are indicated on the product label. The Retriever has a hydrophilic coating to reduce friction during use. A torque device is provided with the Retriever to facilitate manipulation. The torque device is used to lock the core wire to the microcatheter during the procedure. Locking of the torque device to the wire allows the microcatheter and Retriever to be retracted as a system during clot retrieval. An insertion tool is provided to introduce the Retriever into a microcatheter. The proximal end of the device is compatible with the Abbott guide wire extension to facilitate removal or exchange of a catheter while maintaining the Retriever position in the vessel.

    AI/ML Overview

    The provided text describes a 510(k) premarket notification for the Trevo XP ProVue Retriever (4x30mm). This document focuses on demonstrating substantial equivalence to a predicate device, rather than proving performance against specific acceptance criteria for a new clinical outcome. Therefore, many of the requested items (e.g., sample size for test set, number of experts for ground truth, MRMC study, training set details) are not explicitly present in the provided text as they are typically associated with de novo approvals or significant device modifications requiring new clinical evidence.

    However, I can extract information related to the acceptance criteria for bench and animal testing, and the study methodologies employed to demonstrate substantial equivalence.

    Here's the information derived from the provided text:

    1. Table of Acceptance Criteria and Reported Device Performance

    The acceptance criteria for each test were "meets acceptance criteria." The device performance is generally stated as having met these unspecified criteria. The document emphasizes substantial equivalence to the predicate device through bench and animal testing.

    TestAcceptance CriteriaReported Device Performance
    Dimensional VerificationMeets acceptance criteriaDimensional verification meets acceptance criteria.
    Retriever Mid Joint Tensile StrengthMeets acceptance criteriaRetriever Mid Joint Tensile Strength meets acceptance criteria.
    Retriever Tip Tensile StrengthMeets acceptance criteriaRetriever Tip Tensile Strength meets acceptance criteria.
    Retriever Shaped Section Radial ForceMeets acceptance criteriaRetriever Shaped Section Radial Force meets acceptance criteria.
    Retriever / Vessel Interaction (Tip Flexibility)Meets acceptance criteriaRetriever/Vessel Interaction (Tip Flexibility) meets acceptance criteria.
    Retriever Torque Tensile DurabilityMeets acceptance criteriaRetriever Torque Tensile Durability meets acceptance criteria.
    Retriever Platinum Wire Joint StrengthMeets acceptance criteriaRetriever Platinum Wire Joint Strength meets acceptance criteria.
    Retriever Platinum Wire and Joint DurabilityMeets acceptance criteriaRetriever Platinum Wire and Joint Durability meet acceptance criteria.
    RadiopacityMeets acceptance criteriaRadiopacity meets acceptance criteria. (Due to same platinum weave wires and configuration as predicate).
    Retriever / Microcatheter DeliverabilityMeets acceptance criteriaRetriever/Microcatheter Deliverability meets acceptance criteria.
    Particulate EvaluationAll samples meet acceptance criteriaAll samples meet acceptance criteria.
    Coating Integrity EvaluationAll samples meet acceptance criteriaAll samples meet acceptance criteria.
    Simulated UseMeets acceptance criteriaSimulated Use meets acceptance criteria.
    Animal Studies (Safety and Usability)Met acceptance criteriaThe safety and usability results from the animal studies met acceptance criteria.

    2. Sample Size Used for the Test Set and Data Provenance

    • Bench Testing: The sample sizes for individual bench tests are not explicitly stated in the summary document. The phrasing "All samples meet acceptance criteria" suggests that a sufficient number of devices were tested per manufacturing and quality control standards, but the exact N is not provided.
    • Animal Testing: The document mentions that a "larger diameter device (6x30mm) than the subject device (4x30mm)" was used for animal studies as a "worst case." The number of animals used is not specified.
    • Data Provenance: The bench and animal studies were conducted by Concentric Medical, Inc. (the submitter). These are prospective studies performed specifically for this 510(k) submission. No country of origin for the data is explicitly mentioned beyond the submitter's location in Mountain View, California, USA.

    3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications

    • Bench Testing: No external experts are mentioned for establishing ground truth for the bench tests. These tests are industrial performance evaluations against internal specifications.
    • Animal Testing: The animal studies were conducted in compliance with GLP regulation (21 CFR Part 58), which implies qualified personnel (e.g., veterinarians, pathologists, study directors) were involved in the conduct and evaluation, but their specific number or qualifications are not detailed.
    • Clinical Data: No clinical study was performed for this 510(k) submission (stated in the document as "No clinical study was performed as there is no change to the indications for use or the fundamental scientific technology for the subject device."). Therefore, no experts were used to establish ground truth from a clinical test set.

    4. Adjudication Method for the Test Set

    Not applicable, as no human-read test set or clinical study requiring adjudication is described. The evaluations were based on quantifiable bench tests and animal observations.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done

    No, an MRMC comparative effectiveness study was not done. The document explicitly states: "No clinical study was performed as there is no change to the indications for use or the fundamental scientific technology for the subject device."

    6. If a Standalone (i.e. algorithm only without human-in-the-loop performance) was done

    This refers to a medical device in the context of hardware, specifically a "thrombus retriever" catheter. The concept of "standalone algorithm performance" is not applicable here as it is not an AI/software device.

    7. The Type of Ground Truth Used

    • Bench Testing: Ground truth for bench tests is based on engineering specifications and physical measurements, which are inherently objective and quantitative (e.g., tensile strength, radial force, dimensions).
    • Animal Testing: Ground truth in animal studies would typically involve pathological examination and observations for adverse events (e.g., "evidence of arterial transmural dissection or perforation"). This is based on direct biological and histological assessment.

    8. The Sample Size for the Training Set

    Not applicable. This device is a physical medical instrument (thrombus retriever), not a machine learning algorithm requiring a training set.

    9. How the Ground Truth for the Training Set was Established

    Not applicable. This device is a physical medical instrument (thrombus retriever), not a machine learning algorithm.

    Ask a Question

    Ask a specific question about this device

    K Number
    K143077

    Validate with FDA (Live)

    Date Cleared
    2015-04-06

    (161 days)

    Product Code
    Regulation Number
    870.1250
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Trevo XP ProVue Retriever (6x25 mm) is intended to restore blood flow in the neurovasculature by removing thrombus in patients experiencing ischemic stroke within 8 hours of symptom onset. Patients who are ineligible for intravenous tissue plasminogen activator (IV t-PA) or who fail IV t-PA therapy are candidates for treatment.

    Device Description

    The Trevo XP ProVue Retriever (6x25mm) consists of a flexible, tapered core with a shaped section at the distal end. It is designed to restore blood flow in the neurovasculature by removing thrombus in patients experiencing ischemic stroke. Radiopaque platinum wires in the shaped section and radiopaque markers on the distal end allow fluoroscopic visualization. Retriever dimensions are indicated on the product label. The Retriever has a hydrophilic coating to reduce friction during use. A torque device and an insertion tool are provided with the Retriever. The proximal end of the device is compatible with the Abbott guide wire extension to facilitate removal or exchange of a catheter while maintaining the Retriever position in the vessel.

    AI/ML Overview

    The provided text describes the Trevo XP ProVue Retriever (6x25 mm) and summarizes the testing conducted to support its substantial equivalence to a predicate device (Trevo XP ProVue Retriever (4x20mm)). However, it does not contain specific acceptance criteria for each test or detailed performance metrics in the format of a table comparing criteria to reported performance. Instead, it generally states that tests "meet acceptance criteria" or provides qualitative conclusions.

    Therefore, for aspects requiring quantitative acceptance criteria and reported device performance (like a comparative table), a direct answer cannot be fully generated as that level of detail is not explicitly present. Similarly, information regarding sample size for test sets, data provenance, expert involvement for ground truth, adjudication methods, MRMC studies, standalone performance with specific metrics, or training set details are not provided in the given document.

    Based on the available information, here's what can be inferred and stated:

    1. A table of acceptance criteria and the reported device performance

    The document lists various tests and provides a "Conclusions" column for each, generally stating that the device "meets acceptance criteria." It does not provide the numerical acceptance criteria themselves or quantitative reported device performance in most cases.

    TestAcceptance Criteria (Not explicitly stated quantitatively in the document for most tests)Reported Device Performance (Qualitative conclusions)
    Dimensional VerificationNot specified quantitativelyDimensional verification meets acceptance criteria.
    Retriever Mid Joint Tensile StrengthNot specified quantitativelyRetriever Mid Joint Tensile Strength meets acceptance criteria.
    Retriever Tip Tensile StrengthNot specified quantitativelyRetriever Tip Tensile Strength meets acceptance criteria.
    Retriever Shaped Section Radial ForceNot specified quantitativelyRetriever Shaped Section Radial Force meets acceptance criteria.
    Retriever/Vessel Interaction (Tip Flexibility)Not specified quantitativelyRetriever/Vessel Interaction (Tip Flexibility) meets acceptance criteria.
    Retriever Torque Tensile DurabilityNot specified quantitativelyRetriever Torque Tensile Durability meets acceptance criteria.
    Retriever Platinum Wire Joint StrengthNot specified quantitativelyRetriever Platinum Wire Joint Strength meets acceptance criteria.
    Retriever Platinum Wire and Joint DurabilityNot specified quantitativelyRetriever Platinum Wire and Joint Durability meet acceptance criteria.
    RadiopacityVisual assessment of device under fluoroscopy.Radiopacity meets acceptance criteria.
    Retriever / Microcatheter DeliverabilityNot specified quantitativelyRetriever/Microcatheter Deliverability meets acceptance criteria.
    Particulate EvaluationMaximum allowable number/size of particulates generated during simulated use.All samples meet acceptance criteria (i.e., total number and size of particulates generated were within limits for ≥10µm, ≥25µm, and ≥50µm size ranges).
    Coating Integrity EvaluationNo delamination at coating edges, minimal defects along coating length after tracking.Neither Baseline nor Post-tracking test articles exhibited delamination at coating edges. Baseline showed no defects; Post-tracking showed minimal defects. Coating functions as intended and exhibited appropriate integrity.
    Simulated UseSuccessful removal of simulated thrombus, no adverse events.Simulated Use meets acceptance criteria (i.e., device successfully retrieved simulated thrombus in a neurovascular model).
    Animal Studies (Acute)No arterial transmural dissection or perforation due to Stentriever use.Histopathology consistent with arterial healing after routine catheterization. Angiography showed no evidence of vessel dissection, perforation, or thrombosis.
    Animal Studies (Chronic)No arterial transmural dissection or perforation; no stenosis, vessel irregularity, or pseudoaneurysm at treatment sites.Histopathology consistent with arterial healing. No evidence of vessel dissection or perforation. Angiography showed no evidence of dissection, perforation, stenosis, vessel irregularity, intimal flap, or pseudoaneurysm at Day 0 and Day 30. Animal studies meet acceptance criteria and are comparable to the predicate device.

    2. Sample size used for the test set and the data provenance

    • Sample Size for Test Set: Not explicitly stated for most tests.
      • For Animal Studies: 12 test articles (devices) were deployed (4 per animal) across 6 treatment runs in both internal maxillary arteries (IMAs) of animals, resulting in the assessment of 6 arteries (2 per animal) for both acute and chronic studies.
    • Data Provenance:
      • Simulated Use: Conducted using a silicone neurovascular model cast from actual human neurovascular arteries.
      • Animal Studies: Conducted in swine, in compliance with GLP regulation (21 CFR Part 58). This is prospective animal data.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    This information is not provided in the document. The tests primarily involve engineering performance metrics, simulated use outcomes, and animal study observations (angiography, histopathology) conducted by unspecified personnel. There is no mention of external human experts establishing ground truth for the test set in the context of diagnostic interpretation or similar tasks.

    4. Adjudication method for the test set

    This information is not provided in the document. The tests described are primarily physical/mechanical performance tests, simulated use, and animal studies. Adjudication methods like "2+1" or "3+1" are typically used for establishing ground truth in clinical or image-based diagnostic studies involving human interpretation.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    There is no mention of an MRMC comparative effectiveness study, AI assistance, or human reader performance improvement in this document. The device is a mechanical thrombectomy retriever, not an AI-powered diagnostic or assistive tool.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    This question is not applicable as the device is a medical instrument (thrombectomy retriever) and not an algorithm or AI system. Its performance is inherent to the device itself (mechanical, physical, and biocompatibility characteristics), not an algorithm that operates standalone. The animal and simulated use studies demonstrate its "standalone" performance as a device.

    7. The type of ground truth used

    • For physical/mechanical tests (e.g., tensile strength, radial force, deliverability, particulate evaluation): The "ground truth" is defined by established engineering specifications and physical measurements, and compliance with these objective criteria.
    • For Radiopacity: Visual assessment under fluoroscopy.
    • For Coating Integrity: Visual assessment, comparison of test articles, and consideration of other correlated tests (lubricity, durability, particulate generation).
    • For Simulated Use: The ability to retrieve simulated thrombus in a physical model, guided by procedural instructions derived from the Instructions for Use.
    • For Animal Studies: Histopathology and angiography results are used to assess vessel response, dissection, perforation, thrombosis, stenosis, etc., against predefined safety criteria.

    8. The sample size for the training set

    This information is not provided and is not applicable to the type of device (a medical instrument). Training sets are typically associated with machine learning or AI models, which this device is not.

    9. How the ground truth for the training set was established

    This information is not provided and is not applicable for the same reasons as #8.

    Ask a Question

    Ask a specific question about this device

    K Number
    K133177

    Validate with FDA (Live)

    Date Cleared
    2014-02-25

    (131 days)

    Product Code
    Regulation Number
    870.1250
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Modified HD Guide Catheter is indicated for use in facilitating the insertion and guidance of an occlusion catheter, infusion catheter or other appropriate microcatheter into a selected blood vessel in the peripheral, coronary and neuro vascular systems. It may also be used as a diagnostic angiographic catheter and as a conduit for retrieval devices.

    Device Description

    The Modified HD Guide Catheters are single lumen, braided, variable stiffness shaft catheters designed for use in facilitating the insertion and guidance of an occlusion catheter, infusion catheter or other appropriate microcatheter into a selected blood vessel in the peripheral, coronary or neuro vascular system. The catheters include a radiopaque marker on the distal and for angiographic visualization and a luer hub on the proximal end allowing attachments for flushing and aspiration. The catheter shaft has a hydrophilic coating to reduce friction during use. A rotating hemostatic valve with side-arm adapter is provided with each catheter.

    AI/ML Overview

    Here's an analysis of the acceptance criteria and study details for the Modified HD Guide Catheter, based on the provided 510(k) summary:

    1. Table of Acceptance Criteria and Reported Device Performance

    TestAcceptance CriteriaReported Device Performance
    Tip Patency during AspirationDistal tip of catheter remains patent during vigorous aspiration (to avoid loss of suction).Pass. All samples met acceptance criteria for expanded indications for use. Device continues to meet same design requirements as predicate devices (K090335 and K110483).
    Lumen CompatibilityGuidewires and other devices pass through the inner shaft with no more than moderate resistance.Pass. All samples met acceptance criteria for expanded indications for use. Device continues to meet same design requirements as predicate devices (K090335 and K110483).
    Lumen and Retriever CompatibilityRetriever and microcatheter (as a system) can be withdrawn through the catheter and completely removed from the model without device fracture.Pass. All samples met acceptance criteria for expanded indications for use. Device continues to meet same design requirements as predicate devices (K090335 and K110483).
    Infusion and AspirationUser can aspirate and inject fluid through the Extension Tubing with a 60cc syringe. Catheter maintains mechanical integrity and function.Pass. All samples met acceptance criteria for expanded indications for use. Device continues to meet same design requirements as predicate devices (K090335 and K110483).
    Distal Tip StabilityDistal tip does not move (forward or back) during retraction of the Retriever and microcatheter.Pass. All samples met acceptance criteria for expanded indications for use. Device continues to meet same design requirements as predicate devices (K090335 and K110483).
    Mechanical and Functional IntegrityCatheter maintains mechanical integrity (e.g., no damage, kink, etc.) and function is not impaired after simulated use.Pass. All samples met acceptance criteria for expanded indications for use. Device continues to meet same design requirements as predicate devices (K090335 and K110483).
    Retriever FractureRetriever is free from fractures following simulated use testing.Pass. All samples met acceptance criteria for expanded indications for use. Device continues to meet same design requirements as predicate devices (K090335 and K110483).
    Liquid Leak ResistanceCatheter does not leak from the shaft or hub/shaft interface during use.Pass. All samples met acceptance criteria for expanded indications for use. Device continues to meet same design requirements as predicate devices (K090335 and K110483).
    Design Verification (various tests listed in document)Performance as designed, suitability for intended use, substantial equivalence to predicate devices.Successfully evaluated and met criteria for all tests (Tip Patency during Aspiration, Air Leak Resistance during Aspiration, Leak Testing, Dimensional Testing, Tensile Testing, Kink Resistance Testing, Flexural Fatigue Testing, Torque Testing, Tip Flexibility Testing, Coating Lubricity and Durability Testing, Flow Rate Testing, Luer Testing).

    2. Sample Size Used for the Test Set and Data Provenance

    The document categorizes testing into "Testing and Non-Clinical Performance Data" (bench testing) and "Clinical Performance Data" (review of existing clinical studies and post-market data).

    • Non-Clinical Performance Data (Bench Testing):

      • The document states "All samples met acceptance criteria." but does not specify the exact number of samples used for each bench test (e.g., Tip Patency during Aspiration, Lumen Compatibility, etc.). It only implies that multiple samples (denoted by "All samples") were tested for each criterion.
      • Data Provenance: The bench testing utilized a "neurovascular model cast from actual human neurovascular arteries." This is a controlled experimental setup, not data from live patients.
      • Specifically for the expanded indications, "Finished sterilized devices were used for the simulated use testing."
    • Clinical Performance Data:

      • TREVO 2 Study: 99 patients where the DAC (Modified HD Guide Catheter) was used.
        • Provenance: Retrospective analysis of an IDE trial ("TREVO 2 study... designed to support FDA clearance of the Trevo Retriever in the U.S."). Enrolled at 26 sites in the United States and one site in Spain.
      • TREVO Study: 30 patients where aspiration through DAC was applied (a subset of 34 patients where DAC was used).
        • Provenance: Retrospective analysis of a post-marketing prospective, multi-center, single-arm study performed at seven sites in Europe.
      • Post-Market Surveillance: A review of product complaints and Medical Device Reports (MDRs) from 2008 to July 31, 2013. The exact number of complaints reviewed is not given, but it is stated that 10 MDRs were filed.
        • Provenance: Real-world post-market data.
      • Literature Review: 15 published articles pertaining to DAC use.
        • Provenance: Published scientific literature.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

    • Non-Clinical (Bench) Testing: The document does not specify the number or qualifications of experts involved in observing and confirming the "Pass" results for the bench tests. It likely involved internal engineering or research staff.
    • Clinical Performance Data (TREVO and TREVO 2 Studies):
      • The original TREVO 2 and TREVO studies would have had clinical investigators and potentially core labs or adjudication committees, but their specific roles in establishing "ground truth" for catheter performance (as opposed to primary study endpoints for the Retriever device) are not detailed here.
      • The 510(k) summary states that "None of the procedure related or CEC adjudicated events were related to DAC," implying that a Clinical Events Committee (CEC) was used in the TREVO study to adjudicate events. Their specific qualifications are not provided in this 510(k) summary.

    4. Adjudication Method for the Test Set

    • Non-Clinical (Bench) Testing: No specific adjudication method is mentioned. Results were simply reported as "Pass" based on visual verification or adherence to specified methods.
    • Clinical Performance Data: The TREVO study mentioned events being "CEC adjudicated," indicating a Clinical Events Committee (CEC) was involved in reviewing adverse events. The specific method within the CEC (e.g., 2+1, 3+1) is not detailed. The TREVO 2 study also implies similar clinical oversight.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    There is no MRMC comparative effectiveness study mentioned, nor is there any AI assistance component to this device. This is a physical medical device (catheter), not an AI-powered diagnostic tool. Therefore, this section is not applicable.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was done

    Not applicable, as this device is a physical catheter and does not involve an algorithm.

    7. The Type of Ground Truth Used

    • Non-Clinical (Bench) Testing:
      • For most bench tests, the "ground truth" was based on pre-defined acceptance criteria and direct observation of the device's physical performance characteristics (e.g., patency, resistance, integrity, absence of fracture/leakage) in a simulated environment (neurovascular model).
    • Clinical Performance Data (TREVO 2 and TREVO Studies):
      • The "ground truth" in these studies relates to the clinical outcomes of the overall thrombectomy procedure, specifically revascularization rates and the absence of device-specific adverse events. While the primary ground truth would be patient outcomes as assessed by clinicians (e.g., angiography for revascularization, neurological exams for NIHSS), the review here focuses on whether the DAC was implicated in any negative outcomes.

    8. The Sample Size for the Training Set

    Not applicable. This is a physical device and not an AI/machine learning model that requires a training set. The existing clinical data reviewed served as real-world performance evidence, not a "training set."

    9. How the Ground Truth for the Training Set was Established

    Not applicable, as there is no training set for this device.

    Ask a Question

    Ask a specific question about this device

    K Number
    K133464

    Validate with FDA (Live)

    Date Cleared
    2014-01-17

    (66 days)

    Product Code
    Regulation Number
    870.1250
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Trevo Retriever is intended to restore blood flow in the neurovasculature by removing thrombus in patients experiencing ischemic stroke within 8 hours of symptom onset. Patients who are ineligible for intravenous tissue plasminogen activator (IV t-PA) or who fail IV t-PA therapy are candidates for treatment.

    Device Description

    The Modified Trevo ProVue Retriever consists of a flexible, tapered core with a shaped section at the distal end. It is designed to restore blood flow in the neurovasculature by removing thrombus in patients experiencing ischemic stroke. A radiopaque coil at the distal end allows fluoroscopic visualization. The shaped section is also radiopaque. Retriever dimensions are indicated on the product label. The Retriever has a hydrophilic coating to reduce friction during use. A torque device and an insertion tool are provided with the Retriever. The proximal end of the device is compatible with the Abbott guide wire extension to facilitate removal or exchange of a catheter while maintaining the Retriever position in the vessel.

    AI/ML Overview

    The provided document describes the Modified Trevo ProVue Retriever, a device intended to restore blood flow in the neurovasculature for ischemic stroke patients. The submission primarily focuses on demonstrating substantial equivalence to a predicate device (Cleared Trevo ProVue Retriever, K122478) based on design, materials, and function, with a specific modification to the platinum wire weave.

    Here's an analysis of the acceptance criteria and the study that proves the device meets them:

    1. A table of acceptance criteria and the reported device performance:

    Acceptance Criteria (Implicit from Testing)Reported Device Performance
    Maintain deliverabilityAll samples passed
    Maintain deployment capabilityAll samples passed
    Maintain reloadabilityAll samples passed
    No new failure modes introducedNo new failure modes detected
    No changes to existing failure modesNo changes to existing failure modes
    Perform as designedDevice performs as designed
    Suitable for intended useDevice is suitable for its intended use
    Substantially equivalent to predicate deviceDemonstrated as substantially equivalent

    2. Sample size used for the test set and the data provenance:

    • Sample Size: The document does not specify the exact number of Modified Trevo ProVue Retriever samples tested. It states "All samples passed the acceptance criteria."
    • Data Provenance: The testing was conducted as bench testing using a silicone neurovascular model cast from actual human neurovascular arteries. This is a retrospective approach in terms of model derivation, but the testing itself is prospective for the device being evaluated. The 'country of origin' of the data is not explicitly stated, but it's part of a submission to the US FDA by a US-based company (Concentric Medical, Inc. in Mountain View, CA).

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

    • This information is not provided in the document. The "ground truth" for this type of bench testing is typically the physical and functional characteristics of the device itself and its interaction with the simulated environment, rather than expert interpretation of images or clinical outcomes. The "success" or "failure" of deliverability, deployment, and reloadability would be directly observed and measured against pre-defined engineering specifications.

    4. Adjudication method for the test set:

    • An adjudication method is not applicable or mentioned. This was a bench test evaluating physical performance, not a study requiring human interpretation or consensus for outcomes.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, if so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    • No MRMC study was conducted. This device is a mechanical thrombectomy device, not an AI-assisted diagnostic or therapeutic tool for which an MRMC study would be relevant.

    6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done:

    • This question is not applicable as the device is a mechanical thrombectomy device, not an algorithm or AI system.

    7. The type of ground truth used:

    • The ground truth for this bench testing was based on engineering specifications and observable physical performance criteria (e.g., successful navigation, deployment, and retrieval within a simulated anatomical model without device failure). It involved assessing whether the device met its design requirements for deliverability, deployment, and reloadability.

    8. The sample size for the training set:

    • This question is not applicable. There is no training set mentioned, as this is bench testing of a medical device, not a machine learning model.

    9. How the ground truth for the training set was established:

    • This question is not applicable as there is no training set involved.
    Ask a Question

    Ask a specific question about this device

    K Number
    K132641

    Validate with FDA (Live)

    Date Cleared
    2014-01-13

    (140 days)

    Product Code
    Regulation Number
    870.1250
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Trevo Retriever is intended to restore blood flow in the neurovasculature by removing thrombus in patients experiencing ischemic stroke within 8 hours of symptom onset. Patients who are ineligible for intravenous tissue plasminogen activator (IV t-PA) or who fail IV t-PA therapy are candidates for treatment.

    Device Description

    The Modified Trevo ProVue Retriever consists of a flexible, tapered core with a shaped section at the distal end. It is designed to restore blood flow in the neurovasculature by removing thrombus in patients experiencing ischemic stroke. Radiopaque platinum wires in the shaped section and new radiopaque markers on the new truncated distal end allow fluoroscopic visualization. Retriever dimensions are indicated on the product label. The Retriever has a hydrophilic coating to reduce friction during use. A torque device and an insertion tool are provided with the Retriever. The proximal end of the device is compatible with the Abbott guide wire extension to facilitate removal or exchange of a catheter while maintaining the Retriever position in the vessel. The Modified Trevo ProVue Retriever is offered in two sizes.

    AI/ML Overview

    The provided text describes a 510(k) submission for a Modified Trevo ProVue Retriever, a medical device intended to remove thrombus in patients experiencing ischemic stroke. The submission focuses on demonstrating substantial equivalence to a predicate device (Concentric Trevo ProVue Retriever K122478) rather than an AI-powered diagnostic device. Therefore, many of the requested categories related to AI performance, ground truth, expert adjudication, and MRMC studies are not applicable or cannot be extracted from this document.

    Here's the information that can be extracted or inferred from the provided text, with an explanation of why other requested information is not present:


    1. Table of Acceptance Criteria and Reported Device Performance

    TestAcceptance CriteriaReported Device Performance
    Dimensional VerificationMet"Dimensional verification meets acceptance criteria."
    Retriever Proximal Joint Tensile StrengthMet"Retriever Proximal Joint Tensile Strength meets acceptance criteria."
    Retriever Mid Joint Tensile StrengthMet"Retriever Mid Joint Tensile Strength meets acceptance criteria."
    Retriever Tip Tensile StrengthMet"Retriever Tip Tensile Strength meets acceptance criteria."
    Retriever Shaped Section Radial ForceMet"Retriever Shaped Section Radial Force meets acceptance criteria."
    Retriever/Vessel Interaction (Tip Flexibility)Met"Retriever/Vessel Interaction (Tip Flexibility) meets acceptance criteria."
    Retriever Torque Tensile DurabilityMet"Retriever Torque Tensile Durability meets acceptance criteria."
    Retriever Platinum Wire Joint StrengthMet"Retriever Platinum Wire Joint Strength meets acceptance criteria."
    Retriever Platinum Wire and Joint DurabilityMet"Retriever Platinum Wire and Joint Durability meets acceptance criteria."
    RadiopacityAssessed based on visual assessment of the device being used under fluoroscopy."Radiopacity meets acceptance criteria."
    Retriever/Microcatheter DeliverabilityMet"Retriever/Microcatheter Deliverability meets acceptance criteria."
    Proximal Curl ResistanceMet"Proximal Curl Resistance meets acceptance criteria."
    Simulated UseDemonstrated successful performance in a silicone neurovascular model with simulated thrombus."Simulated Use meets acceptance criteria."
    Biocompatibility (Cytotoxicity - ISO Elution Method)Non-cytotoxic (Dilutions: Undiluted (3 - Moderate reactivity - noted as consistent with historical data), 50%, 25%, 12.5%, 6.25%, 3.13% - 0 reactivity)Pass for all dilutions, with specific note on historical consistency for undiluted moderate reactivity.
    Biocompatibility (ASTM Hemolysis)Non-HemolyticPass. Hemolytic index: 0.0% (direct contact) and 0.4% (extract)
    Biocompatibility (Physicochemical Tests Plastics)No heavy metals or leachables (Non-volatile residue: 2 mg, Residue on ignition: ≤ 2 mg, Heavy metal: < 1 ppm, Buffering capacity: < 1.0 mL)Pass. Stated conclusion: "No heavy metals or leachables."

    Explanation: The acceptance criteria are implicitly stated by the "Conclusion" column in the "Testing Summary" and "Biocompatibility" tables, which consistently reports that the device "meets acceptance criteria." The specific quantitative acceptance criteria are not explicitly detailed (e.g., what is the exact tensile strength required), but the summary confirms they were met.


    2. Sample Size Used for the Test Set and the Data Provenance

    • Sample Size: The document does not specify the exact sample sizes (number of devices, number of tests, or number of simulated emboli) used for the various physical and simulated use tests.
    • Data Provenance:
      • Simulated Use: The test used a "silicone neurovascular model cast from actual human neurovascular arteries," replicating tortuosity, diameter, and location of arteries. This indicates a benchtop, in vitro model.
      • Other Tests: The other tests (dimensional, tensile strength, radial force, durability, radiopacity, deliverability, curl resistance) are laboratory benchtop tests performed on physical device samples.
      • All testing was conducted by Concentric Medical, Inc. (the submitter), implying an internal, controlled testing environment.
      • The tests are prospective in nature, as they were conducted specifically for this submission.
      • The country of origin of the data is not explicitly stated but is implicitly the country where Concentric Medical, Inc. conducts its R&D and testing, likely the USA given the submission to the FDA.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and the Qualifications of Those Experts

    This information is Not Applicable to this submission. The Modified Trevo ProVue Retriever is a physical medical device, not an AI diagnostic algorithm. Ground truth for the performance characteristics of the device is established through physical and simulated benchtop testing, not through expert review of clinical data or images. The "Radiopacity" test mentions "visual assessment," which likely implies assessment by trained personnel (e.g., engineers, quality control specialists) using fluoroscopy, but no details on their qualifications or number are provided, as it's a standard engineering/device performance check.


    4. Adjudication Method for the Test Set

    This information is Not Applicable to this submission, as it pertains to expert consensus for clinical ground truth in AI studies. The tests conducted are objective physical and performance measurements.


    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    This information is Not Applicable. This is a submission for a physical medical device. There is no AI component or human-in-the-loop performance being evaluated in this K132641 summary.


    6. If a Standalone (i.e. algorithm only without human-in-the-loop performance) was done

    This information is Not Applicable. This is a physical medical device, not a standalone algorithm.


    7. The Type of Ground Truth Used

    The "ground truth" for this device's performance is established through objective, quantifiable physical and mechanical tests, and simulated use in a validated in vitro model. This implicitly serves as the "ground truth" for its functional characteristics.

    • Physical/Mechanical Properties: Measurements directly obtained during testing (e.g., force, dimensions, durability cycles) compared against predetermined specifications.
    • Simulated Use: The ability of the device to successfully retrieve simulated thrombus in a anatomically correct (cast from human arteries) silicone neurovascular model.
    • Biocompatibility: Laboratory test results (cytotoxicity, hemolysis, physicochemical properties) compared against ISO 10993-1 standards and internal specifications.

    8. The Sample Size for the Training Set

    This information is Not Applicable. The device is a physical medical device; there is no AI training set involved.


    9. How the Ground Truth for the Training Set was Established

    This information is Not Applicable. There is no "training set" for this physical medical device.

    Ask a Question

    Ask a specific question about this device

    K Number
    K131437

    Validate with FDA (Live)

    Date Cleared
    2013-10-11

    (144 days)

    Product Code
    Regulation Number
    870.1200
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Modified Concentric Microcatheter is indicated for use in the selective placement of fluids and/or other devices or agents into the peripheral, coronary and neuro vasculature during diagnostic and/or therapeutic procedures.

    Device Description

    The Modified Concentric Microcatheter is a line extension to the existing Concentric Microcatheter, cleared under K113260. Like the predicate device, the Modified Concentric Microcather is a single-lumen, braided shaft, variable stiffness catheter designed for use in the selective placement of fluids and/or other devices or agents into the peripheral, coronary and neuro vasculature during diagnostic and/or therapeutic procedures. A luer hub on the proximal end of the shaft enables connection to the rotating hemostasis valve included in the package. The radiopaque shaft and distal marker facilitate fluoroscopic visualization. The catheter shaft is coated with a hydrophilic coating to reduce friction during use. The Modified Concentric Microcatheter differs from its predicate in that it has smaller shaft diameters and a longer effective shaft length. Device dimensions and configuration are shown on the product label. A rotating hemostasis valve with side-arm adapter is provided with each microcatheter.

    AI/ML Overview

    The provided submission describes a medical device, the "Modified Concentric Microcatheter," which is a line extension to an existing product. The submission focuses on demonstrating substantial equivalence to the predicate device rather than presenting a de novo study with acceptance criteria for a new AI/software device. Therefore, many of the requested categories related to AI/software performance evaluation (e.g., effect size with human readers, standalone performance, ground truth establishment for training) are not applicable to this document.

    Here's an analysis of the provided information concerning acceptance criteria and the supporting study, tailored to the context of a medical device submission demonstrating substantial equivalence:

    1. A table of acceptance criteria and the reported device performance

    The submission does not explicitly list quantitative acceptance criteria in a tabular format with corresponding performance metrics for the Modified Concentric Microcatheter. Instead, it describes various tests and states that the device successfully met the evaluation for each. The acceptance criteria generally seem to be "successful evaluation" or "meets its intended use" for each test. The performance is reported qualitatively as "successfully evaluated."

    Acceptance Criteria (Implied)Reported Device Performance
    Withstand kinking when flexedKink Resistance: successfully evaluated
    Withstand leak when subjected to both high pressure and vacuumLeak Resistance: successfully evaluated
    Navigate tight bendsFlexibility Testing: successfully evaluated
    Maintain mechanical integrity under tensile loadsTensile Testing: successfully evaluated
    Maintain mechanical integrity when subjected to torsionTorque Testing: successfully evaluated
    Durability and lubricity of the device coatingCoating Testing: successfully evaluated
    Satisfactory dimensions (as verified)Dimensional verification testing: performed
    Visibility under fluoroscopyRadiopacity Testing: successfully evaluated
    Meets intended use (in a simulated environment)Design Validation Simulated Use Testing: successfully evaluated
    Biocompatibility (relying on predicate device data)Biocompatibility: met pre-determined acceptance criteria
    Sterilization (relying on predicate device data)Sterilization: met pre-determined acceptance criteria

    2. Sample size used for the test set and the data provenance (e.g., country of origin of the data, retrospective or prospective)

    The document does not explicitly state the sample sizes (number of devices or tests performed) for each specific test. It simply mentions that tests were "performed" or "evaluated."

    For the Design Validation Simulated Use Testing, it states "testing was performed in a porcine model." This indicates the use of an ex vivo or in vivo animal model. Data provenance is implied to be from internal testing by Concentric Medical, Inc. The timing of testing is prospective to the submission.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g., radiologist with 10 years of experience)

    This section is not applicable as the submission is for a medical device (microcatheter) and not an AI/software device requiring expert interpretation for ground truth. The "ground truth" for this device relates to its physical and functional performance, which is assessed through engineering and simulated-use testing.

    4. Adjudication method (e.g., 2+1, 3+1, none) for the test set

    This is not applicable as the testing described pertains to the physical and mechanical properties of a microcatheter, not to the interpretation of images or data that would necessitate an adjudication method among experts.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    This is not applicable. The device is an intravascular catheter, not an AI-powered diagnostic tool. Therefore, an MRMC study related to human reader performance with AI assistance is irrelevant to this submission.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    This is not applicable as the device is a physical medical instrument and does not involve an algorithm.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

    The "ground truth" for this device's performance is established through:

    • Engineering specifications and standards: For physical and mechanical properties (kink resistance, leak resistance, flexibility, tensile strength, torque, coating).
    • Dimensional measurements: For shaft diameters and length.
    • Fluoroscopic visibility standards: For radiopacity.
    • Functional assessment in a simulated environment: In the porcine model to confirm intended use.

    There is no "expert consensus" or "pathology" in the diagnostic sense, nor is there "outcomes data" presented in this 510(k) summary, as it's a pre-market submission focused on substantial equivalence.

    8. The sample size for the training set

    This is not applicable as the device is a physical medical instrument and does not involve AI or machine learning algorithms that require a training set.

    9. How the ground truth for the training set was established

    This is not applicable for the same reason as point 8.

    Ask a Question

    Ask a specific question about this device

    K Number
    K131492

    Validate with FDA (Live)

    Date Cleared
    2013-10-03

    (133 days)

    Product Code
    Regulation Number
    870.1250
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Concentric Balloon Guide Catheter is indicated for use in facilitating the insertion and guidance of an intravascular catheter into a selected blood vessel in the peripheral and neuro vascular systems. The balloon provides temporary vascular occlusion during these and other angiographic procedures. The Balloon Guide Catheter is also indicated for use as a conduit for Retrieval devices.

    Device Description

    Like the predicate device, the Modified FlowGate™ Balloon Guide Catheter is a coaxiallumen, braid-reinforced, variable stiffness catheter designed for use in facilitating the insertion and guidance of an intravascular catheter into a selected blood vessel in the peripheral and neuro vascular systems. A radiopaque marker is included on the distal end for angiographic visualization. A compliant balloon is mounted on the distal end to provide temporary vascular occlusion during angiographic procedures. A bifurcated luer hub on the proximal end allows attachments for flushing, inflation and aspiration. Modified FlowGate™ Balloon Guide Catheter dimensions and maximum recommended balloon inflation volume are indicated on product label.

    AI/ML Overview

    The provided text is a 510(k) Summary for a medical device (Modified Balloon Guide Catheter) and does not describe a study involving an AI/ML device. Therefore, it does not contain information on acceptance criteria, reported device performance in the context of AI/ML, sample sizes for test or training sets, data provenance, expert ground truth establishment, adjudication methods, MRMC studies, standalone algorithm performance, or the type of ground truth used for AI/ML validation.

    Instead, the document focuses on demonstrating substantial equivalence of a modified medical device to its predicate devices, as required for 510(k) clearance by the FDA. The "acceptance criteria" and "study" mentioned in this document refer to the verification and validation testing performed to ensure the modified device meets its design specifications and performs as safely and effectively as the predicate devices.

    Here's an interpretation of the requested information based on the provided text, adapted for a medical device rather than an AI/ML system:


    1. Table of Acceptance Criteria and Reported Device Performance:

    The document doesn't present a single table of acceptance criteria and reported full device performance in a summary format. Instead, it lists various tests performed and states that the results were successful or met acceptance criteria. For biocompatibility, specific results and conclusions are provided.

    Here's a synthesized table based on the information available:

    Acceptance Criteria Category/TestReported Device Performance Summary (Modified Balloon Guide Catheter)
    Mechanical & Performance Tests
    Dimensional VerificationDevice met predetermined dimensional requirements.
    Simulated Use TestingDevice and accessories were successfully used per procedural instructions in a neurovascular model.
    Balloon Testing (Symmetry, Compliance, Inflation/Deflation, Fatigue, Burst, Leakage)Tests performed successfully; balloon demonstrated appropriate symmetry, compliance, inflation/deflation rates, fatigue strength, constrained burst, and leakage.
    Tensile Testing (Mechanical Integrity)Device demonstrated mechanical integrity under tensile loads.
    Distal Shaft Flexibility TestingForce to flex the distal shaft was within acceptable limits.
    Soft Tip Deflection TestingForce to deflect the distal edge of the soft tip was within acceptable limits.
    Tip Patency TestingDevice distal tip maintained patency under aspiration.
    Kink ResistanceDevice shaft demonstrated resistance to kinking.
    Hub Luer Fitting TestingExtension Tubing accessory demonstrated compatibility with the hub luer fitting.
    Extension Tubing AspirationAbility to aspirate using the Extension Tubing Accessory was confirmed.
    Packaging VerificationPackaging successfully protected the device following climatic conditioning and distribution simulation.
    Shelf Life Testing (Product and Packaging)Demonstrated acceptable shelf life for both product and packaging.
    Biocompatibility Tests (Predicate Device K122581)
    Sensitization (ISO 10993-10)Non-sensitizer (no evidence of delayed dermal contact sensitization).
    Cytotoxicity (ISO 10993-5)Non-cytotoxic (no biological activity with L929 cells).
    Intracutaneous Reactivity (ISO 10993-10)Non-irritant (no significantly greater biological reaction than controls).
    Systemic Toxicity (Acute) (ISO 10993-11)No acute systemic toxicity (no significantly greater biological reaction than controls).
    Systemic Toxicity (Rabbit Pyrogen Test) (ISO 10993-11, USP 34 <151>)Non-pyrogenic (no individual temperature rise of 0.5°C or more).
    Hemocompatibility (Hemolysis) (ASTM F756)Non-hemolytic (0.17% direct, 0.12% indirect hemolysis above negative control).
    Hemocompatibility (Complement Activation) (ISO 10993-4)Minimal complement activation potential (no statistically significant increase in C3a; significant increase in SC5b-9).
    Hemocompatibility (Thrombogenicity) (ISO 10993-4)No significant thrombosis (Grade 0-1 observed).
    Biocompatibility Screening Tests (Subject Device K131492)
    Cytotoxicity (ISO 10993-5)Non-cytotoxic (no cytotoxicity or cell lysis).
    Hemocompatibility (Hemolysis) (ISO 10993-4, ASTM Direct/Extract)Non-hemolytic (hemolytic index 0.0% direct, 0.4% extract).
    Fourier Transform Infrared (FTIR) Scan (ISO 10993-18)Materials for test article and control sample are equivalent (97.90% match for distal tip).
    Physicochemical Tests (USP <661>)Met USP <661> requirements (Non-volatile residue: 1 mg; Residue on ignition: < 1 mg; Heavy metal: < 1 ppm; Buffering capacity: < 1.0 ml).

    2. Sample Size Used for the Test Set and Data Provenance:

    • The document does not specify general "sample sizes" for the test sets (e.g., number of catheters tested for mechanical properties) beyond implying that tests were conducted on "the subject device and accessories."
    • For biocompatibility tests, it does mention "Swiss Albino mice" for systemic toxicity and "rabbit blood" for hemolysis, and "dogs" for the thrombogenicity study.
    • Data Provenance: Not explicitly stated, but assumed to be internal laboratory testing by Concentric Medical, Inc. or its contractors. No information on country of origin or whether it's retrospective or prospective is provided. It represents prospective testing conducted specifically for this submission.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts:

    • This concept is not applicable in this context. For medical devices undergoing 510(k) clearance based on substantial equivalence, "ground truth" is established through engineering and biocompatibility testing against defined specifications and recognized standards (e.g., ISO, ASTM, USP), not expert consensus on medical images or clinical outcomes. The "experts" would be the engineers, toxicologists, and other scientists performing and evaluating the tests.

    4. Adjudication Method for the Test Set:

    • Not applicable. Adjudication methods (like 2+1, 3+1) are typically used in clinical or imaging studies involving human interpretation or subjective assessments. Here, the "adjudication" is based on objective measurements and compliance with predetermined pass/fail criteria from recognized standards.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done:

    • No. An MRMC comparative effectiveness study is performed for AI/ML systems often involving diagnostic imaging to assess the impact of an AI tool on human reader performance. This document is for a physical medical device (catheter) and such a study is not relevant.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done:

    • No. This is relevant to AI/ML devices, not a physical medical catheter. The device's performance is inherently standalone in that it is evaluated on its physical and material properties, not an algorithm.

    7. The Type of Ground Truth Used (Expert Consensus, Pathology, Outcomes Data, etc.):

    • For the engineering and mechanical tests, the "ground truth" is defined by the engineering specifications and performance requirements derived from the intended use, risk assessment, and comparison to predicate devices, and adherence to relevant industry standards (e.g., for tensile strength, flexibility, balloon characteristics).
    • For biocompatibility, the "ground truth" is established by internationally recognized standards (e.g., ISO 10993 series, ASTM F756, USP <661>, USP 34 <151>) which define acceptable biological responses and material properties. Test results are compared directly against the pass/fail criteria outlined in these standards.

    8. The Sample Size for the Training Set:

    • Not applicable. This document describes a physical medical device, not an AI/ML system, so there is no "training set." The development of the device would involve engineering design, prototyping, and iterative testing, but not in the sense of a data-driven "training set" for an algorithm.

    9. How the Ground Truth for the Training Set Was Established:

    • Not applicable, as there is no training set for this type of device.
    Ask a Question

    Ask a specific question about this device

    Page 1 of 5