K Number
K120961

Validate with FDA (Live)

Device Name
TREVO RETRIEVER
Date Cleared
2012-08-03

(126 days)

Product Code
Regulation Number
870.1250
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Trevo Retriever is intended to restore blood flow in the neurovasculature by removing thrombus in patients experiencing ischemic stroke within 8 hours of symptom onset. Patients who are ineligible for intravenous tissue plasminogen activator (IV t-PA) or who fail IV t-PA therapy are candidates for treatment.

Device Description

Like the predicate device, the Trevo Retriever consists of a flexible, tapered core wire with a shaped section at the distal end and is designed to restore blood flow in the neurovasculature by removing thrombus in patients experiencing ischemic stroke. A radiopaque coil at the distal end allows fluoroscopic visualization. Retriever dimensions are indicated on the product label. The Retriever has a hydrophilic coating to reduce friction during use. A torque device and an insertion tool are provided with the Retriever. The proximal end of the device is compatible with the Abbott guide wire extension to facilitate removal or exchange of a catheter while maintaining the Retriever position in the vessel.

AI/ML Overview

Here's a breakdown of the acceptance criteria and the study proving the device meets them, based on the provided text:

Acceptance Criteria and Device Performance

The study aimed to demonstrate the non-inferiority of the Trevo Retriever compared to the predicate Merci Retriever.

Table 1: Acceptance Criteria and Reported Device Performance

CriterionAcceptance (Non-Inferiority Margin)Trevo Retriever Performance (vs. Merci)Met?
Primary Efficacy Endpoint: Post-Device Revascularization Success (TICI ≥2a)Non-inferiority margin of 10%Trevo: 87.3% (69/79) Merci: 58.0% (47/81) Difference: 29.3% (95% CI: 15.0%, 42.4%) p-value < 0.0001 (non-inferiority & superiority)Yes
Primary Safety Endpoint: Incidence of procedure-related serious adverse events through 24 hours post-procedure (Composite Events)Not explicitly defined as a statistical hypothesis for non-inferiority. The goal was to show safety comparable to or better than the predicate.Trevo: 13.9% (11/79) Merci: 23.5% (19/81) Difference: -9.5% (95% CI: -22.1%, 2.8%) p-value = 0.1567 (Fisher's exact test of difference). The rate was numerically lower for Trevo.Yes

Note: While a specific numerical acceptance criterion for the primary safety endpoint wasn't stated in terms of non-inferiority, the clinical trial demonstrated a numerically lower (and thus favorable) rate of adverse events for the Trevo device.


Study Details:

2. Sample size used for the test set and the data provenance

  • Test Set Sample Size:
    • Total randomized patients: 178
    • Patients meeting angiographic entry criteria and included in primary analysis: 160 (79 for Trevo, 81 for Merci)
  • Data Provenance: The data was obtained from a multi-center, prospectively controlled IDE clinical trial (TREVO 2). The document does not explicitly state the country of origin, but "IDE clinical trial" implies it was conducted under FDA regulations, typically within the United States, although international sites are possible. The study design is prospective.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

The document mentions that "Primary Efficacy was assessed by the independent Core Lab." While it specifies a "Core Lab," it doesn't state the exact number of experts involved or their specific qualifications (e.g., "Radiologist with 10 years of experience").

4. Adjudication method for the test set

The primary efficacy endpoint (revascularization success) was assessed by an "independent Core Lab." The primary safety endpoint (procedure-related serious adverse events) was adjudicated by an independent clinical events committee per protocol. The specific method of adjudication (e.g., 2+1, 3+1) for either endpoint is not detailed.

5. If a multi-reader, multi-case (MRMC) comparative effectiveness study was done, if so, what was the effect size of how much human readers improve with AI vs without AI assistance

  • No, this was not an MRMC study. This clinical trial evaluated the efficacy and safety of a medical device (thrombus retriever) for ischemic stroke treatment, comparing it directly to a predicate device. It did not involve AI assistance or human reader performance evaluation in diagnosing or interpreting medical images.

6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done

  • No, this evaluates a physical medical device (thrombus retriever), not an algorithm or AI. Therefore, a standalone algorithm performance study is not applicable.

7. The type of ground truth used

The ground truth for efficacy was angiographic data interpreted by an independent Core Lab using the Thrombolysis in Cerebral Infarction (TICI) score. For safety, the ground truth was occurrence of procedure-related serious adverse events adjudicated by an independent clinical events committee.

8. The sample size for the training set

  • Not applicable. This study is a clinical trial evaluating the performance of a physical medical device, not a machine learning model, so there is no "training set" in the context of AI/ML.

9. How the ground truth for the training set was established

  • Not applicable, as there is no training set for an AI/ML model.

{0}------------------------------------------------

K120961

3 2012 AUG

510(K) SUMMARY

Trade Name:Trevo Retriever
Common Name:Catheter, Thrombus Retriever
Classification:Catheter, Thrombus Retriever, 21CFR 870.1250 - Class II
Submitter:Concentric Medical, Inc.
301 E. Evelyn Avenue
Mountain View, CA 94041
Tel 650-938-2100
Fax 650-237-5230
Facility Registration #2954917
Contact:Kirsten ValleyVice President, Technology & Regulatory Affairs
Date Prepared:July 27, 2012
1 64 6 4 4 4 4 8 8 8 8 8 8 8 8 8 8 8 8 8 8 8 8 8 8 8 8 8 8 8 8 8 8 8 8 8 8 8 8 8 8 8 8 8 8 8 8 8 8 8 8 8 8 8 8 8 8 8 8 8 8 8 8 8 8 8 8 8 8 8 8 8 8 8 8 8 8 8 8 8 8 8 8 8 8 8 8Uly ( ( ) ( LV LL
Predicate Device:Merci® Retriever (K063774

Device Description

Like the predicate device, the Trevo Retriever consists of a flexible, tapered core wire with a shaped section at the distal end and is designed to restore blood flow in the neurovasculature by removing thrombus in patients experiencing ischemic stroke. A radiopaque coil at the distal end allows fluoroscopic visualization. Retriever dimensions are indicated on the product label. The Retriever has a hydrophilic coating to reduce friction during use. A torque device and an insertion tool are provided with the Retriever. The proximal end of the device is compatible with the Abbott guide wire extension to facilitate removal or exchange of a catheter while maintaining the Retriever position in the vessel.

Indications for Use

The Indications for Use are as follows:

The Trevo Retriever is intended to restore blood flow in the neurovasculature by removing thrombus in patients experiencing ischemic stroke within 8 hours of symptom onset. Patients who are ineligible for intravenous tissue plasminogen activator (IV t-PA) or who fail IV t-PA therapy are candidates for treatment. This is comparable to the predicate device's Indications for Use and does not impact the intended therapeutic use of the device in patients experiencing ischemic stroke or raise different issues of safety and effectiveness. As a result, the safety and effectiveness of the device are not impacted when used as indicated.

Technological Characteristics

There are technological differences between Trevo and the predicate device. The effect of these differences on safety and effectiveness has been addressed in a clinical study which supports that the clinical performance of the two devices is substantially equivalent. Table 1 provides a comparison between the Trevo device and the predicate Merci device.

{1}------------------------------------------------

AttributePredicate Merci Retriever Device (K063774)Subject Trevo Retriever
Device DescriptionThe Retriever consists of a flexible, tapered core wire with a shaped section at the distal end. A platinum coil allows fluoroscopic visualization. The Retriever has a hydrophilic coating to reduce friction. The Retriever has a shaft marker to indicate proximity of Retriever tip relative to Microcatheter tip. A torque device and insertion tool are provided with the Retriever.Same as the predicate device
Distal End (shaped section) ConfigurationImage: Distal End Configuration of Merci Retriever DeviceImage: Distal End Configuration of Trevo Retriever Device
Anatomical SitesNeurovasculatureSame as cleared device
Target PopulationPatients with symptoms of an ischemic strokeSame as cleared device
Single/Multiple UseSingle UseSame as cleared device
Materials
Core Wire MaterialNitinol (nickel titanium alloy)Same as cleared device
Distal Shaped Section MaterialNitinolSame as cleared device
Coil Material Distal to Distal Shaped SectionPlatinum/TungstenSame as cleared device
Coil Material Proximal to Shaped Section304 Stainless SteelSame as cleared device
SolderGold/TinSame as cleared device
Nominal Design Attributes
Proximal Core Wire Diameter0.0137"0.0180"
Shaped Section Diameter (nominal)Variable along length ranging from 3mm-1.5mm4 mm
Shaped Section Length (nominal)7 mm20 mm
Distal Tip Length (nominal)7 mm4 mm
Accessory DevicesProvided (not in directcontact with patient)Insertion tool and torque device providedin product packageSame as cleared device
Packaging
Materials andConfigurationPolyethylene Hoop, polycarbonatemounting card, Tyvek/Film Pouch, MDPETubing Clips, Chipboard cartonImage: packagingSame as cleared device
Sterilization Method100% EtOSame as cleared device
How SuppliedSterile/Single UseSame as cleared device

{2}------------------------------------------------

Testing Summary

Bench Testing

The results of verification and validation conducted on the Trevo Retriever demonstrate that it performs as designed, is suitable for its intended use and is substantially equivalent to the predicate device. Specifically, the following tests were performed on the proposed device:

  • Simulated Use Testing: the device's ability to be used in a neurovascular model per . procedural instructions outlined in the Instructions for Use was successfully evaluated.
  • Tensile Testing: the device's mechanical integrity under tensile loads was successfully . evaluated.
  • Radial Force Testing: the resulting radial force when the device is constrained radially 0 was successfully evaluated.
  • . Tip Deflection Force Testing: the force to deflect the distal tip of the device was successfully evaluated.
  • 0 Torque/Tensile Durability: the ability of the device to withstand torque and tensile load cycles without fracture was successfully evaluated.
  • Kink Resistance: the ability of the device shaft to resist kinking was successfully . evaluated.
  • Device safety (vessel response) was successfully evaluated in an animal model .

Clinical Testing

To demonstrate substantial equivalence, a randomized, multi-center, prospectively controlled IDE clinical trial (TREVO 2) was conducted comparing the efficacy of the subject Trevo Retriever to the predicate Merci Retriever for removing occlusive thrombi in patients experiencing an ischemic stroke.

{3}------------------------------------------------

A total of 178 patients were randomized, 88 to treatment with the Trevo device and 90 to the Merci device. On review by the Core Lab, 81 subjects randomized to Trevo and 82 subjects randomized to Merci met all angiographic entry criteria. An attempt to treat with the device was made in 79 Trevo and 81 Merci subjects. The population for analysis of the primary effectiveness and safety endpoints includes these 160 subjects were treated with the device to which they were randomized. There were no subjects with missing endpoint data. Statistical Analysis

Analysis of the primary efficacy endpoint (successful revascularization of the target occlusion, using the TICI score as the metric) was performed using a one-sided test per Blackwelder's formulation of the primary efficacy non-inferiority hypothesis at the 0.025 level of significance. The primary safety endpoint of the study was the incidence of procedure-related serious adverse events (PRSAEs) through 24 hours post procedure. These events were adjudicated by an independent clinical events committee per protocol. There was no statistical hypothesis related to the primary safety endpoint.

Study Procedures

Diagnostic angiographic data was obtained at baseline (prior to patient randomization), again at the conclusion of revascularization attempts with the assigned study device, and post procedure. 24-hour post-procedure follow-up included CT or MR imaging and NIHSS examination. Follow-up at day 7-10 (or discharge if earlier) and at day 90 (+/-14 days) included NIHSS exam and modified Rankin Scale (mRS).

Inclusion Criteria

  • Patient presenting with clinical signs and symptoms consistent with a diagnosis of Acute . Ischemic Stroke, and:
    • o a. Patient has failed IV t-PA therapy
      • Or
    • b. Patient is contraindicated for IV t-PA administration o
  • Age 18-85 (has had 18th birthday, but not yet had 86th birthday) ●
  • . NIHSS 8 8 < NIHSS < 29
  • Anticipated life expectancy of at least 6 months .
  • . No significant pre-stroke disability (mRS < 1)
  • . Written informed consent to participate given by patient or legal representative
  • . Angiographic confirmation of a persistent large vessel occlusion in the internal carotid, middle cerebral (M1 and/or M2 segments), basilar and/or vertebral arteries
  • . Treatable within 8 hours of symptom onset, defined as the first pass being made with the assigned study device

Exclusion Criteria

  • . Baseline glucose < 50 mg/dL (2.78 mmol) or > 400 mg / dL (22.20 mmol)
  • . Known hemorrhagic diathesis, coagulation factor deficiency, or oral anticoagulant therapy with INR > 3.0
  • Treated with Heparin within 48 hours with a PTT greater than 2 times the lab normal .
  • Baseline platelet count < 30,000
  • History of severe allergy (more than rash) to contrast medium or nitinol .
  • Severe, sustained hypertension (SBP>185 mm Hg or DBP>110 mm Hg) .

KIZOq

{4}------------------------------------------------

NOTE: If the blood pressure can be successfully reduced and maintained at the acceptable level using medication, the patient can be enrolled

  • . Pregnancy
  • Patient participating in another investigational drug or device study .
  • CT showing hypodensity or MR showing hyperintensity involving greater than 1/3 of the ● MCA territory. For non-MCA strokes, CT showing hypodensity or MR showing hyperintensity involving > 100cc of tissue.
  • . Baseline CT/MR evidence of significant mass effect with midline shift
  • . Baseline CT evidence of hemorrhage
  • . Baseline CT/MR evidence of intracranial tumor (except small meningioma)
  • . Angiographic evidence of vasculitis or arterial dissection
  • . Stenosis in a proximal vessel that requires treatment or that prevents access to the thrombus with the assigned study device
  • � Angiographic evidence of excessive arterial tortuosity that would preclude the assigned study device from reaching the thrombus
  • . Bilateral stroke
Reason for ExclusionNumber ofSubjects
Angiographic evidence of vasculitis or arterial dissection0
Angiographic evidence of excessive arterial tortuosity that wouldpreclude the assigned study device from reaching the thrombus1
Vessel too small1
No angiographic evidence of a persistent large vessel occlusion in theinternal carotid, middle cerebral (M1 and/or M2 segments), basilarand/or vertebral arteries26*
CT showing hypodensity or MR showing hyperintensity involvinggreater than 1/3 of the MCA territory (discovery concurrent withangiography)1
Stenosis in a proximal vessel that requires treatment or that preventsaccess to the thrombus with the assigned study device3*

Table 2: Summary of Reasons for Frelusion during Angiographic Screening

*1 subject had 2 exclusion criteria

Primary Efficacy Endpoint

Primary Efficacy was assessed by the independent Core Lab. Revascularization as measured by TICI was determined for each subject following the use of the assigned study device. Use of any IA lytic or treatment of a proximal carotid stenosis was automatically counted as a revascularization failure regardless of the subjects's revascularization status after use of the device (Table 3a). In an additional analysis, subjects were counted as revascularization failures if any adjunctive therapy was used at the site of an occlusion regardless of revascularization status after the use of the assigned study device (Table 3b). Subjects with a baseline TICI 2a by Core Lab and subjects in whom the device was never attempted were excluded from both analyses. The test for non-inferiority is highly significant and establishes that the Trevo Retriever is non-inferior and substantially equivalent to the Merci Retriever.

{5}------------------------------------------------

Lab), With Study Device Attempted, Any IA lytic Considered as Treatment Failure
Primary EffectivenessEndpointTrevo (N=79)% (n/N)[95% CI]aMerci (N=81)% (n/N)[95% CI]aDifference[95% CI]bp-value
Post-Device RevascularizationSuccess (TICI ≥2a)87.3% (69/79)[78.0%, 93.8%]58.0% (47/81)[46.5%, 68.9%]29.3%[15.0%, 42.4%]<0.0001c<0.0001d

Table 3a: Primary Effectiveness Endpoint for Subjects with Baseline TICI 0 or 1 (by Core Lab), with Study Device Attempted. Any IA Jytic Considered as Treatment Failur

a: Exact Clopper Pearson confidence intervals on individual proportions; b: Exact confidence intervals on differences in proportions computed with StatXact Version 8; c: Non-inferiority hypothesis using Blackwelder's method with noninferiority margin of 10%; d: One-sided Fisher's exact test of superiority

Table 3b: Primary Effectiveness Endpoint for Subjects with Baseline TICI 0 or 1 (by Core Lab), with Study Device Attempted, Any Adjunctive Treatment Considered as Treatment Failure

Primary EffectivenessEndpointTrevo (N=79)% (n/N)[95% CI]aMerci (N=81)% (n/N)[95% CI]aDifference[95% CI]bp-value
Post-Device RevascularizationSuccess (TICI ≥2a)79.7% (63/79)49.4% (40/81)30.4%<0.0001c
[69.2%, 88.0%][38.1%, 60.7%][15.0%,44.1%]<0.0001d

a: Exact Clopper Pearson confidence intervals on individual proportions; b: Exact confidence intervals on differences in proportions computed with StatXact Version 8; c: Non-inferiority hypothesis using Blackwelder's method with noninferiority margin of 10%; d: One-sided Fisher's exact test of superiority

{6}------------------------------------------------

Primary Safety Endpoint and Good Outcomes

The incidence of procedure-related serious adverse events through 24 hours post-procedure was reported for the Trevo and Merci arms (Table 4). The rate for this composite endpoint was numerically lower in the Trevo arm supporting the safety of the Trevo Retriever. Subjects with a baseline TICI 2a by Corelab and subjects in whom the device was never attempted were excluded from this analysis.

TrevoN= 79 ptsMerciN= 81 ptsDifference[95% CI]ap-Value
Primary Safety Endpoint
Composite Events[95% Conf. Interval]c13.9% (11/79)[7.2%, 23.6%]23.5% (19/81)[14.8%, 34.2%]-9.5%[-22.1%, 2.8%]0.1567b
Vessel Perforation0.0% (0/79)9.9% (8/81)-9.9%[-18.5%, -3.9%]
Intramural Arterial Dissection0.0% (0/79)1.2% (1/81)-1.2%[-6.7%, 3.5%]
Symptomatic ICH5.1% (4/79)9.9% (8/81)-4.8%[-14.1%, 3.8%]
Embolization to PreviouslyUninvolved Territory7.6% (6/79)4.9% (4/81)2.7%[-5.6%, 11.4%]
Access Site ComplicationRequiring Surgical Repair orBlood Transfusion1.3% (1/79)0.0% (0/81)1.3%[-3.3%, 6.9%]
Mortality within 24 hrs1.3% (1/79)0.0% (0/81)1.3%[-3.3%, 6.9%]
in vivo Device Failure0.0% (0/79)0.0% (0/81)0.0%[-4.6%, 4.6%]
Other PR-SAE0.0% (0/79)0.0% (0/81)0.0%[-4.6%, 4.6%]
Secondary Endpoint
Good Outcome at 90 days (mRS ≤2)38.2% (29/76)17.9% (14/78)20.2%[6.4%, 34.1%]

Table 4: Summary of Primary Safety Endpoint and 90-day Good Outcomes

a. Exact computations with StatXact Version 8; b. Fisher's exact test; c. Exact Clopper Pearson confidence intervals on individual proportions

{7}------------------------------------------------

Concentric Medical ClassificationTermTrevo PatientsN=86 ptsMerci PatientsN=89 pts
Units% (Number of pts) [Number of AEs]% (Number of pts) [Number of AEs]
Total Adverse Events (AE)96.5% (83) [388]96.6% (86) [493]
Mortality at 90 days33.7% (29/86) [29]23.8% (21/88) [21]
Neurologic73.3% (63) [136]83.1% (74) [178]
Cerebral Edema15.1% (13) [14]18.0% (16) [16]
Headache10.5% (9) [10]9.0% (8) [8]
Dysphagia (Difficulty Swallowing)16.3% (14) [14]27.0% (24) [24]
IVH5.8% (5) [5]6.7% (6) [6]
SAH12.8% (11) [11]23.6% (21) [21]
ICH - HI -114.0% (12) [12]21.3% (19) [19]
ICH – HI -28.1% (7) [7]6.7% (6) [6]
ICH - PH114.0% (12) [12]21.3% (19) [19]
ICH – PH28.1% (7) [7]5.6% (5) [5]
Neurologic Decline11.6% (10) [10]24.7% (22) [22]
Late ICH4.7% (4) [4]6.7% (6) [6]
Depression4.7% (4) [4]5.6% (5) [5]
Progression of index Stroke9.3% (8) [8]6.7% (6) [6]
Cardiac37.2% (32) [39]30.3% (27) [43]
Arrhythmia - Tachycardia7.0% (6) [6]3.4% (3) [3]
Atrial Fibrillation10.5% (9) [9]7.9% (7) [7]
Hypotension - Sustained - Tx7.0% (6) [6]5.6% (5) [6]
Dermatologic9.3% (8) [11]7.9% (7) [7]
Gastrointestinal17.4% (15) [16]16.9% (15) [19]
Constipation2.3% (2) [2]5.6% (5) [5]
Nausea and Vomiting (Non-Neuro)9.3% (8) [8]3.4% (3) [3]
Hematologic18.6% (16) [16]27.0% (24) [28]
Anemia16.3% (14) [14]19.1% (17) [17]
Metabolic26.7% (23) [34]33.7% (30) [45]
Hyperglycemia9.3% (8) [8]10.1% (9) [9]
Electrolyte Imbalance18.6% (16) [23]27.0% (24) [32]
Musculoskeletal14.0% (12) [13]13.5% (12) [19]
Joint/Extremity Pain7.0% (6) [6]11.2% (10) [13]
Concentric Medical ClassificationTermTrevo PatientsN=86 ptsMerci PatientsN=89 pts
Procedural12.8% (11) [11]18.0% (16) [17]
Access Site Complication5.8% (5) [5]1.1% (1) [1]
Embolization to PreviouslyUninvolved Territory5.8% (5) [5]4.5% (4) [4]
Vessel Perforation1.2% (1) [1]10.1% (9) [9]
Pulmonary38.4% (33) [53]50.6% (45) [71]
Pneumonia10.5% (9) [11]23.6% (21) [23]
Pulmonary Edema4.7% (4) [4]5.6% (5) [5]
Respiratory Distress9.3% (8) [9]3.4% (3) [3]
Respiratory Failure - Acute5.8% (5) [5]21.3% (19) [19]
Vascular12.8% (11) [13]15.7% (14) [14]
Deep Vein Thrombosis (DVT)11.6% (10) [12]14.6% (13) [13]
Constitutional15.1% (13) [15]16.9% (15) [18]
Fever4.7% (4) [4]6.7% (6) [6]
Positive Cultures4.7% (4) [4]7.9% (7) [7]
Urogenital27.9% (24) [30]29.2% (26) [32]
Urinary Tract Infection16.3% (14) [15]21.3% (19) [20]
Hematuria7.0% (6) [6]1.1% (1) [2]
Urinary Retention3.5% (3) [3]5.6% (5) [5]

Adverse Events and Mortality in Subjects in whom the Assigned Study Device was Used Table 5: Summary of Adverse Events

{8}------------------------------------------------

Conclusion

The data collected on both the primary efficacy endpoint and the primary safety in the TREVO 2 clinical trial demonstrate that the Trevo Retriever is substantially equivalent to the predicate Merci Retriever.

Summary of Substantial Equivalence

The Trevo Retriever is comparable to the predicate device with regard to device design, materials, intended use, and patient population. The conclusions drawn from the verification and validation testing, animal studies and clinical trial conducted using the Trevo Retriever demonstrate that it performs as designed, is suitable for its intended use and is substantially equivalent to the legally marketed predicate device.

{9}------------------------------------------------

DEPARTMENT OF HEALTH & HUMAN SERVICES

Image /page/9/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a stylized eagle with three stripes forming its wing and body. The eagle is enclosed in a circle of text that reads "DEPARTMENT OF HEALTH & HUMAN SERVICES · USA".

Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002

AUG 3 2012

Concentric Medical, Inc. c/o Ms. Kirsten Valley Vice President, Technology & Regulatory Affairs 301 East Evelyn Avenue Mountain View, CA 94041

Re: K120961

Trade/Device Name: Trevo Retriever Regulation Number: 21 CFR 870.1250 Regulation Name: Percutaneous Catheter Regulatory Class: Class II Product Code: NRY Dated: July 30, 2012 Received: July 31, 2012

Dear Ms. Valley:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{10}------------------------------------------------

Page 2 - Ms. Kirsten Valley

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.

Sincerely vours.

-Malvina B. Evdelman, M.D. Director Division of Ophthalmic, Neurological, and Ear, Nose and Throat Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{11}------------------------------------------------

Indications for Use

510(k) Number (if known):

Device Name: Trevo Retriever

Indications for Use:

The Trevo Retriever is intended to restore blood flow in the neurovasculature by removing thrombus in patients experiencing ischemic stroke within 8 hours of symptom onset. Patients who are ineligible for intravenous tissue plasminogen activator (IV-tPA) or who fail IV-tPA therapy are candidates for treatment.

Prescription Use X (Part 21 CFR 801 Subpart D) AND/OR

Over-The-Counter Use (21 CFR 807 Subpart C)

(PLEASE DO NOT WRITE BELOW LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Page 1 of 1

JOE HVTTER

(Division Sign-Off) Division of Ophthalmic, Neurological and Ear, Nose and Throat Devices

KI20961

510(k) Number.

§ 870.1250 Percutaneous catheter.

(a)
Identification. A percutaneous catheter is a device that is introduced into a vein or artery through the skin using a dilator and a sheath (introducer) or guide wire.(b)
Classification. Class II (performance standards).