Search Filters

Search Results

Found 3 results

510(k) Data Aggregation

    K Number
    K220331
    Manufacturer
    Date Cleared
    2022-07-29

    (175 days)

    Product Code
    Regulation Number
    870.1250
    Reference & Predicate Devices
    Why did this record match?
    Reference Devices :

    K112404

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The 091 Balloon Guide Catheter is indicated for use in facilitating the insertion and guidance of intravascular catheters into a selected blood vessel in the peripheral and neuro vascular systems. The balloon provides temporary vascular occlusion during these and other angiographic procedures. The Balloon Guide Catheter is also indicated for use as a conduit for Retrieval devices.

    Device Description

    The 091 Balloon Guide Catheter is a coaxial-lumen, coil-reinforced, variable stiffness catheter with a compliant balloon at the distal end to provide temporary vascular occlusion during angiographic procedures. A radiopaque marker is included at the distal end of the distal tip of the catheter) for fluoroscopic visualization. A bifurcated luer hub on the proximal end of the catheter allows attachments for flushing, and aspiration. This catheter is designed for use in facilitating the insertion and guidance of an intravascular catheter into a selected blood vessel in the peripheral and neuro vascular systems. The dimensions and maximum recommended balloon inflation volume for the 091 Balloon Guide Catheter are indicated on the product label. A Peel-Away Introducer accessory is included within the packaging. It is supplied sterile, non-pyrogenic, and is intended for single use only.

    AI/ML Overview

    This document describes the validation of the 091 Balloon Guide Catheter. Since the provided text focuses on the device's substantial equivalence to a predicate device for regulatory approval, the "acceptance criteria" and "device performance" primarily refer to engineering performance specifications and bench testing results, rather than clinical efficacy metrics for an AI/ML device. There is no mention of an AI/ML component in this premarket notification.

    Therefore, the requested information specifically related to AI/ML device validation (e.g., sample sizes for training/test sets, expert ground truth adjudication, MRMC studies, standalone algorithm performance) is not applicable to this document. The document details the testing performed to ensure the device meets safety and performance standards for a medical device.

    Table of Acceptance Criteria and Reported Device Performance

    The acceptance criteria are generally implied by the "Pass" result for each test. The reported device performance is that it met all predetermined acceptance criteria for each test.

    Acceptance Criteria (Test Performed)Reported Device Performance (Results)
    Design Verification Testing
    Tensile Strength (per ISO 10555-1, Section 4.6 and Annex B)Pass
    PTFE delamination (visual inspection)Pass
    Torque Strength (withstand one turn of the hub)Pass
    Catheter Burst (per ISO10555-1, Section 4.10 and Annex F)Pass
    Balloon Burst (not burst below specified volume)Pass
    Visual Inspection (under 2.5X magnification)Pass
    Particulates (per USP )Pass
    Liquid Leak Test (per ISO 10555-1, Section 4.7 and Annex C)Pass
    Air Leak Test (per ISO 10555-1 section 4.7.2 and Annex D)Pass
    Balloon Leak Test (per ISO 10555-4, section 4.4.2 and Annex B)Pass
    Dimensional Verification (catheter and introducer meet specs)Pass
    Chemical Compatibility (withstand exposure to saline, dextrose, heparin, contrast)Pass
    Hub Compatibility (per ISO 594-1:1986 and ISO 594-2:1998)Pass
    Kink Resistance (test to determine resistance to kinking)Pass
    Balloon Air Purge Test (acceptable level of air removed)Pass
    Balloon Fatigue Test (withstand specified inflation/deflation cycles)Pass
    Balloon Compliance Test (not exceed specified dimensions for given volume)Pass
    Flow Arrest (minimum occlusion time confirmed in flow model)Pass
    Balloon Deflation Time (time to restore flow measured)Pass
    Packaging Testing
    Pouch Leak Test (per ASTM F-1929)Pass
    Pouch Peel Test (per ASTM F88/F88M)Pass
    Packaging - Visual Inspection (perforations, nicks, cuts, punctures, seal damage)Pass
    Packaging - Seal Width (meet specified width)Pass
    Design Validation Testing
    In-vitro Simulated Use Study – Benchtop (performance verification by physicians)Pass
    Usability Testing
    In-vitro Simulated Use Study – Usability (IFU and labeling review, simulated use, worst-case models)Pass
    Biocompatibility Testing
    Cytotoxicity (MEM elution, 48 hr. inc., triplicate L929, 24 hr. ext.)Pass
    Sensitization (Magnusson-Kligman Method, 2 extracts)Pass
    Irritation (Intracutaneous Toxicity (ISO), 2 extracts)Pass
    Material mediated pyrogenicity (Material Mediated Pyrogen)Pass
    Acute Systemic Toxicity (Systemic Injection (ISO), 2 extracts)Pass
    Hemolysis (ASTM Method, indirect and direct)Pass
    Complement Activation (SC5b-9)Pass
    Dog ThrombogenicityPass
    Sterilization Validation
    Confirmatory sterilization study (sterility, EO residual, LAL, bioburden)Pass

    Details of the Study:

    • 1. A table of acceptance criteria and the reported device performance: Provided above. All tests "Passed," indicating the device met all predetermined acceptance criteria.

    • 2. Sample size used for the test set and the data provenance: The document states that "Some of the tests were also conducted on the predicate device to help establish substantial equivalence." However, specific sample sizes for each test are not explicitly detailed in this summary. The data provenance is from bench testing and simulated use, conducted to support a regulatory submission in the United States. It is not patient data, so "retrospective or prospective" is not applicable.

    • 3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts: Not applicable in the context of an AI/ML device. For the "In-vitro Simulated Use Study – Benchtop," it states "A simulated interventional procedure was performed by physicians." For "Usability Testing," it mentions "Evaluators representative of the intended user population shall review... then attempt to use the catheters and accessories along with the expected compatible products in a simulated use environment, using worst-case neurovascular models." The specific number or qualifications of these physicians/evaluators are not detailed.

    • 4. Adjudication method (e.g. 2+1, 3+1, none) for the test set: Not applicable as this is not an AI/ML device with image interpretation or complex diagnostic outputs requiring expert adjudication. Performance was assessed against pre-defined engineering and functional criteria.

    • 5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance: Not applicable. This is a physical medical device (catheter), not an AI/ML diagnostic or assistive tool.

    • 6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done: Not applicable. This is a physical medical device (catheter), not an AI/ML algorithm.

    • 7. The type of ground truth used:

      • Engineering/Performance Specifications: For most tests (e.g., Tensile Strength, Torque Strength, Burst, Leak tests, Dimensional Verification), the "ground truth" is defined by established engineering and materials standards (e.g., ISO 10555-1, USP , ASTM F-1929, ISO 594-1:1986).
      • Simulated Use/Usability: For these tests, the "ground truth" is the successful and safe performance of the device by intended users in a simulated clinical environment, meeting pre-defined operational criteria.
      • Biocompatibility: The "ground truth" is compliance with ISO 10993-1 and specific biological test methodologies (e.g., MEM elution, Magnusson-Kligman Method, ASTM Method for Hemolysis) to ensure the device is not harmful to biological systems.
    • 8. The sample size for the training set: Not applicable. This is a physical medical device, not an AI/ML device trained on data.

    • 9. How the ground truth for the training set was established: Not applicable.

    Ask a Question

    Ask a specific question about this device

    K Number
    K172335
    Device Name
    Bleep DreamPort
    Manufacturer
    Date Cleared
    2018-03-30

    (240 days)

    Product Code
    Regulation Number
    868.5905
    Reference & Predicate Devices
    Why did this record match?
    Reference Devices :

    K112404, K102502

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Bleep™ DreamPort nasal adhesive mask is intended to provide an interface from a Continuous Positive Airway Pressure (CPAP) or bi-level system. The mask to be used by adult patients (>66 lbs. / 30 kg, for whom positive airway pressure has been prescribed. The mask is for single-patient reuse in the home, hospital/institutional environment.

    Device Description

    The Bleep™ DreamPort nasal adhesive mask is similar to standard nasal pillow CPAP mask for use with patients using CPAP and bi-level equipment for the treat of OSA. The Bleep™ DreamPort nasal adhesive mask is comprised of 2 sub-assemblies: Nasal Port assembly and Airflow management assembly. The device uses single use foam tape to create the seal at the nostrils. This allows the interface to be used without head gear. There is a reusable interface which connects the nasal adhesive to the circuit of the CPAP equipment. It includes integrated exhalation ports.

    AI/ML Overview

    The provided text describes the Bleep™ DreamPort nasal adhesive mask and its acceptance criteria for FDA clearance, comparing it to a predicate device (Resmed Swift LT). The study primarily focuses on non-clinical performance testing and a limited user study to demonstrate substantial equivalence, rather than a typical multi-reader multi-case (MRMC) or standalone AI performance study. Therefore, some of the requested information, particularly regarding AI-specific criteria, human reader improvement, and extensive expert ground truth establishment, is not applicable to this device's clearance process as described.

    Here's an analysis based on the provided document:

    Acceptance Criteria and Reported Device Performance

    The acceptance criteria are generally framed around demonstrating substantial equivalence to the predicate device (Resmed Swift LT, K073638) and compliance with international standards like ISO 17510-2:2015 and ISO 10993-1.

    Acceptance Criteria CategorySpecific Acceptance CriteriaReported Device Performance (Bleep™ DreamPort)Assessment
    Indications for UseIntended to provide an interface from a Continuous Positive Airway Pressure (CPAP) or bi-level system for adult patients (≥66 lbs. / 30 kg) for whom positive airway pressure has been prescribed. For single-patient reuse in home, hospital/institutional environment.The Bleep™ DreamPort nasal adhesive mask is intended to provide an interface from a Continuous Positive Airway Pressure (CPAP) or bi-level system. The mask to be used by adult patients (>66 lbs. / 30 kg.) for whom positive airway pressure has been prescribed. The mask is for single-patient reuse in the home, hospital/institutional environment.Met: Identical Indications for Use to predicate.
    Patient PopulationAdults > 66 lbs. / > 30 kg.Adults > 66 lbs. / > 30 kg.Met: Identical Patient Population to predicate.
    Environment of UseSingle-patient reuse in the home, hospital/institutional environment.Single-patient reuse in the home, hospital/institutional environment.Met: Identical Environment of Use to predicate.
    Therapy Pressures & CompatibilityUsed with equipment providing therapy pressures within a defined range.4 to 15 cm H2O. Used with equipment which provide pressures within the defined therapy pressure range.Met: Compatible within a defined range. While the predicate is 4-20 cm H2O, the Bleep still operates within a common CPAP range.
    CO2 Washout Profile (ISO 17510-2)1 for leachable/extractable components for permanent contact with intact skin.Non-cytotoxic, non-irritating, non-sensitizing. MOS > 1 for leachable/extractable. Risk-based assessment.Met: Passed all required biocompatibility tests.
    User Study (Clinical Testing)Demonstrate effectiveness of seal under different CPAP pressure settings and compare leak rates and mean AHI (Apnea-Hypopnea Index).Found to be substantially equivalent in performance (seal effectiveness, leak rates, mean AHI) over 2 nights.Met: "Substantially equivalent in performance" is the stated finding.

    Study Information:

    1. Sample size used for the test set and the data provenance:

      • User study (Clinical Testing): The document states a "user study was performed," but does not explicitly mention the sample size (number of participants/patients).
      • Non-clinical performance testing (Bench Tests): This refers to the physical device being tested against standards (e.g., CO2 washout, exhaust flow). No "test set" in the sense of patient data is applicable here; tests are performed on the device itself.
      • Data Provenance: Not explicitly stated for the user study (e.g., country of origin, specific clinics). The user study was conducted over "2 nights" for each participant. It's implicitly prospective as it's a "user study" where performance was compared.
    2. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

      • User Study (Clinical Testing): The document does not specify the number or qualifications of experts involved in establishing ground truth (e.g., reading sleep study data). It mentions "mean AHI during use" was compared, implying that existing clinical methods for AHI calculation would be the "ground truth." This is not an AI-specific ground truth establishment (like image annotation by multiple radiologists).
      • Non-clinical performance testing: Ground truth here are the engineering standards and measurements, not expert consensus.
    3. Adjudication method (e.g. 2+1, 3+1, none) for the test set:

      • User Study (Clinical Testing): The document does not specify any adjudication method for the clinical/user study results (e.g., for AHI measurements or leak rates). The comparison was seemingly direct or averaged.
      • Non-clinical performance testing: Not applicable.
    4. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

      • No, an MRMC comparative effectiveness study was not done. This device is a medical accessory (mask) and not an AI-powered diagnostic tool, so such a study is not relevant to its clearance.
    5. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

      • No, not applicable. This is a physical medical device, not an algorithm. Bench testing is the "standalone" performance for the device itself.
    6. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):

      • User Study: The "ground truth" for the user study appears to be clinical measurements such as leak rates and AHI, which are standard metrics in sleep apnea therapy. This would typically be derived from objective CPAP machine data and sleep study results, often interpreted by sleep physicians. It is not described as involving extensive expert consensus or pathology data.
      • Non-clinical performance testing: The ground truth is engineering measurements against established ISO standards.
    7. The sample size for the training set:

      • Not applicable. This device is not an AI/machine learning product and therefore does not have a "training set" in that sense. It's a manufactured product whose performance is demonstrated through physical and limited user testing.
    8. How the ground truth for the training set was established:

      • Not applicable. See point 7.
    Ask a Question

    Ask a specific question about this device

    K Number
    K133177
    Date Cleared
    2014-02-25

    (131 days)

    Product Code
    Regulation Number
    870.1250
    Reference & Predicate Devices
    Why did this record match?
    Reference Devices :

    K112404

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Modified HD Guide Catheter is indicated for use in facilitating the insertion and guidance of an occlusion catheter, infusion catheter or other appropriate microcatheter into a selected blood vessel in the peripheral, coronary and neuro vascular systems. It may also be used as a diagnostic angiographic catheter and as a conduit for retrieval devices.

    Device Description

    The Modified HD Guide Catheters are single lumen, braided, variable stiffness shaft catheters designed for use in facilitating the insertion and guidance of an occlusion catheter, infusion catheter or other appropriate microcatheter into a selected blood vessel in the peripheral, coronary or neuro vascular system. The catheters include a radiopaque marker on the distal and for angiographic visualization and a luer hub on the proximal end allowing attachments for flushing and aspiration. The catheter shaft has a hydrophilic coating to reduce friction during use. A rotating hemostatic valve with side-arm adapter is provided with each catheter.

    AI/ML Overview

    Here's an analysis of the acceptance criteria and study details for the Modified HD Guide Catheter, based on the provided 510(k) summary:

    1. Table of Acceptance Criteria and Reported Device Performance

    TestAcceptance CriteriaReported Device Performance
    Tip Patency during AspirationDistal tip of catheter remains patent during vigorous aspiration (to avoid loss of suction).Pass. All samples met acceptance criteria for expanded indications for use. Device continues to meet same design requirements as predicate devices (K090335 and K110483).
    Lumen CompatibilityGuidewires and other devices pass through the inner shaft with no more than moderate resistance.Pass. All samples met acceptance criteria for expanded indications for use. Device continues to meet same design requirements as predicate devices (K090335 and K110483).
    Lumen and Retriever CompatibilityRetriever and microcatheter (as a system) can be withdrawn through the catheter and completely removed from the model without device fracture.Pass. All samples met acceptance criteria for expanded indications for use. Device continues to meet same design requirements as predicate devices (K090335 and K110483).
    Infusion and AspirationUser can aspirate and inject fluid through the Extension Tubing with a 60cc syringe. Catheter maintains mechanical integrity and function.Pass. All samples met acceptance criteria for expanded indications for use. Device continues to meet same design requirements as predicate devices (K090335 and K110483).
    Distal Tip StabilityDistal tip does not move (forward or back) during retraction of the Retriever and microcatheter.Pass. All samples met acceptance criteria for expanded indications for use. Device continues to meet same design requirements as predicate devices (K090335 and K110483).
    Mechanical and Functional IntegrityCatheter maintains mechanical integrity (e.g., no damage, kink, etc.) and function is not impaired after simulated use.Pass. All samples met acceptance criteria for expanded indications for use. Device continues to meet same design requirements as predicate devices (K090335 and K110483).
    Retriever FractureRetriever is free from fractures following simulated use testing.Pass. All samples met acceptance criteria for expanded indications for use. Device continues to meet same design requirements as predicate devices (K090335 and K110483).
    Liquid Leak ResistanceCatheter does not leak from the shaft or hub/shaft interface during use.Pass. All samples met acceptance criteria for expanded indications for use. Device continues to meet same design requirements as predicate devices (K090335 and K110483).
    Design Verification (various tests listed in document)Performance as designed, suitability for intended use, substantial equivalence to predicate devices.Successfully evaluated and met criteria for all tests (Tip Patency during Aspiration, Air Leak Resistance during Aspiration, Leak Testing, Dimensional Testing, Tensile Testing, Kink Resistance Testing, Flexural Fatigue Testing, Torque Testing, Tip Flexibility Testing, Coating Lubricity and Durability Testing, Flow Rate Testing, Luer Testing).

    2. Sample Size Used for the Test Set and Data Provenance

    The document categorizes testing into "Testing and Non-Clinical Performance Data" (bench testing) and "Clinical Performance Data" (review of existing clinical studies and post-market data).

    • Non-Clinical Performance Data (Bench Testing):

      • The document states "All samples met acceptance criteria." but does not specify the exact number of samples used for each bench test (e.g., Tip Patency during Aspiration, Lumen Compatibility, etc.). It only implies that multiple samples (denoted by "All samples") were tested for each criterion.
      • Data Provenance: The bench testing utilized a "neurovascular model cast from actual human neurovascular arteries." This is a controlled experimental setup, not data from live patients.
      • Specifically for the expanded indications, "Finished sterilized devices were used for the simulated use testing."
    • Clinical Performance Data:

      • TREVO 2 Study: 99 patients where the DAC (Modified HD Guide Catheter) was used.
        • Provenance: Retrospective analysis of an IDE trial ("TREVO 2 study... designed to support FDA clearance of the Trevo Retriever in the U.S."). Enrolled at 26 sites in the United States and one site in Spain.
      • TREVO Study: 30 patients where aspiration through DAC was applied (a subset of 34 patients where DAC was used).
        • Provenance: Retrospective analysis of a post-marketing prospective, multi-center, single-arm study performed at seven sites in Europe.
      • Post-Market Surveillance: A review of product complaints and Medical Device Reports (MDRs) from 2008 to July 31, 2013. The exact number of complaints reviewed is not given, but it is stated that 10 MDRs were filed.
        • Provenance: Real-world post-market data.
      • Literature Review: 15 published articles pertaining to DAC use.
        • Provenance: Published scientific literature.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

    • Non-Clinical (Bench) Testing: The document does not specify the number or qualifications of experts involved in observing and confirming the "Pass" results for the bench tests. It likely involved internal engineering or research staff.
    • Clinical Performance Data (TREVO and TREVO 2 Studies):
      • The original TREVO 2 and TREVO studies would have had clinical investigators and potentially core labs or adjudication committees, but their specific roles in establishing "ground truth" for catheter performance (as opposed to primary study endpoints for the Retriever device) are not detailed here.
      • The 510(k) summary states that "None of the procedure related or CEC adjudicated events were related to DAC," implying that a Clinical Events Committee (CEC) was used in the TREVO study to adjudicate events. Their specific qualifications are not provided in this 510(k) summary.

    4. Adjudication Method for the Test Set

    • Non-Clinical (Bench) Testing: No specific adjudication method is mentioned. Results were simply reported as "Pass" based on visual verification or adherence to specified methods.
    • Clinical Performance Data: The TREVO study mentioned events being "CEC adjudicated," indicating a Clinical Events Committee (CEC) was involved in reviewing adverse events. The specific method within the CEC (e.g., 2+1, 3+1) is not detailed. The TREVO 2 study also implies similar clinical oversight.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    There is no MRMC comparative effectiveness study mentioned, nor is there any AI assistance component to this device. This is a physical medical device (catheter), not an AI-powered diagnostic tool. Therefore, this section is not applicable.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was done

    Not applicable, as this device is a physical catheter and does not involve an algorithm.

    7. The Type of Ground Truth Used

    • Non-Clinical (Bench) Testing:
      • For most bench tests, the "ground truth" was based on pre-defined acceptance criteria and direct observation of the device's physical performance characteristics (e.g., patency, resistance, integrity, absence of fracture/leakage) in a simulated environment (neurovascular model).
    • Clinical Performance Data (TREVO 2 and TREVO Studies):
      • The "ground truth" in these studies relates to the clinical outcomes of the overall thrombectomy procedure, specifically revascularization rates and the absence of device-specific adverse events. While the primary ground truth would be patient outcomes as assessed by clinicians (e.g., angiography for revascularization, neurological exams for NIHSS), the review here focuses on whether the DAC was implicated in any negative outcomes.

    8. The Sample Size for the Training Set

    Not applicable. This is a physical device and not an AI/machine learning model that requires a training set. The existing clinical data reviewed served as real-world performance evidence, not a "training set."

    9. How the Ground Truth for the Training Set was Established

    Not applicable, as there is no training set for this device.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1