Search Results
Found 765 results
510(k) Data Aggregation
(126 days)
870.1130 | Class II | DXN | System, measurement, blood pressure, non-invasive |
| General Hospital 21 CFR 880.2910
The Perin Health Platform is a wireless remote monitoring system intended for use by healthcare professionals for spot check collection of physiological data in healthcare and home settings for long-term monitoring. The Perin Health Patch can monitor auscultation data of heart and lung sounds, photoplethysmography waveforms (PPG), oxygen saturation (%SpO2), heart rate, electrocardiography (ECG), heart rate variability, R-R interval, respiratory rate, skin temperature, activity detection (including step count), and posture (body position relative to gravity including fall).
The Perin Health System is intended for spot-checking and tracking changes of adult patients in hospitals, clinics, long-term care, and at home. In home-use environments, the Perin Health Platform is able to integrate with optional third-party devices for blood pressure, and weight data collection via the mobile application. The mobile application transmits data from the Health Patch and third party devices to the cloud and web-based portal for storage, analysis, and review by healthcare professionals. The Perin Health Platform can include the ability to notify healthcare professionals when physiological data falls outside set limits or manual trigger by the patient.
The device is intended to provide physiological information for non-critical, adult population.
The Perin Health System is a wireless remote patient monitoring platform that enables healthcare professionals to perform spot-checking and retrospective monitoring of physiological data from adult patients. The Perin Health System is designed for use in hospitals, clinics, long-term care facilities, physician offices, and home environments.
The Perin Health System comprises the following components:
- Perin Health Patch wearable device
- Perin Health Patient Mobile Application
- Perin Health Cloud
- Perin Health Provider Portal
- Perin Health Inpatient Application
1. The Perin Health Patch
The Perin Health Patch is a chest-worn wearable device that performs scheduled spot-check measurements of multiple physiological parameters. Unlike continuous monitoring systems, the Perin Health Patch captures measurements at predetermined intervals configured by healthcare providers based on clinical need.
The device integrates six primary sensing modalities:
- Auscultation (heart and lung sounds)
- Electrocardiography (1-channel ECG)
- Pulse oximetry via photoplethysmography (PPG)
- Bioimpedance (BioZ) for respiratory monitoring
- Temperature sensing (skin)
- Motion and orientation detection via accelerometer
The combination of these modalities in a small, low-power wearable form allows for the spot-checking of primary vital signs:
- Heart rate and R-R intervals
- Heart rate variability (HRV) parameters
- ECG waveform data
- Auscultation sound data (heart and lung sounds)
- Respiratory rate
- Pulse (PPG) waveform
- Oxygen saturation (SpO2%)
- Skin temperature
- Fall detection events
- Body posture
- Activity level and step count
The device adheres to the patient's upper left chest at the second intercostal space with a medical-grade long-term wear adhesive. The adhesive is placed on the patient-facing side of the wearable, with cutouts for the sensors to make direct contact with the skin. The wearable device is lightweight and semi-flexible, allowing for the device to conform to the natural curvature of the chest. It is water resistant, allowing for bathing and normal activities while the patient is wearing the system.
The wearable communicates to the receiving unit (mobile phone) via an encrypted Bluetooth Low Energy connection. Measurements, all notifications and control commands, and software updates are transmitted over the BLE connection. The wearable uses Near Field Communication (NFC) to facilitate the Bluetooth pairing process with the mobile phone by simply having to tap their phone to the device to initiate a Bluetooth connection. The wearable device also contains on-board memory that can store over two weeks of spot-check data. When measurements are taken and no receiving unit is present, the wearable can store recordings in the onboard memory. Recordings are stored in a stack, such that at the next connection possibility between the wearable and the receiving unit, the most recent data will be transmitted first followed by other measurements in reverse chronological order.
Other key features of the wearable include:
- Customizable recording schedule set by the healthcare provider in their care program
- Replaceable battery
- Patient-triggered recordings via double-tap
- Signal quality indicators for measurement validation and identification of noisy measurements
2. The Patient Mobile Application
The Patient Mobile Application, available on iOS or Android platforms, is intended exclusively for use in home environments by patients under healthcare provider supervision. The application serves as a data relay and display interface, allowing the patient to complete key tasks, including onboarding, device setup, device communication, and patient-reported data.
The application serves as the primary interface between the Perin Health Patch and the cloud infrastructure, receiving spot-check measurements from the device and uploading them for provider review. The application establishes and manages secured BLE communication with the Health Patch. Given that the Health Patch operates on provider-configured recording schedules, the application manages data transfer in the background with minimal patient interaction required. When internet connectivity is unavailable, the application stores measurements locally until transmission becomes possible. The system also manages firmware updates for the Perin Health Patch.
The application integrates with FDA-cleared third-party blood pressure cuff and scale using BLE and transfers the data to the Cloud System. Healthcare providers determine which patients require the additional third-party device monitoring as part of their individualized care programs. The system also allows users to optionally enter manual data for blood pressure and weight if no third-party device is connected.
Patients are able to review their historical measurement data taken throughout their monitoring program and their goals and thresholds set by their providers. The patient can view metrics assigned within their care program:
- Heart Rate and Heart Rate Variability
- Respiratory Rate
- Oxygen Saturation
- Step Count
- Temperature
- Blood Pressure
- Weight
Patients can also select audio segments captured by the device for playback (no visualization).
The application provides comprehensive patient engagement features. Patients can complete customized questionnaires with up to 20 questions in various formats, review educational content delivered through their care programs, and submit non-critical medical reports to their care team. The reporting feature includes anatomical body mapping for location-specific symptoms, severity scaling, and photo attachment capabilities. The application supports secure messaging with care providers, virtual appointment attendance with waiting room functionality, and comprehensive offline operation with automatic synchronization upon connectivity restoration.
3. The Perin Health Cloud
The Perin Health Cloud infrastructure serves as the central hub for data management and processing. The cloud system receives encrypted spot-check data from relay systems and manages raw data processing (for Health Patch data only), storage, and retrieval of physiological measurements for retrospective clinical review. Algorithms are run in the cloud to process measurements from the Health Patch and generate Signal Quality Index, Heart Rate, Heart Rate Variability, Respiratory Rate, Oxygen Saturation, and Posture.
The alert and notification system enables healthcare professionals to configure multi-level alerts based on clinical parameters, technical issues, or manual patient triggers. Clinical alerts are based on provider-configured thresholds that are set in during the enrollment of a patient in a care program. The system supports complex notification rules including threshold exceedances, percentage changes, trending patterns, and consecutive violations. Alerts are displayed to providers for the purpose of highlighting data during their retrospective review and are not intended to support real-time patient monitoring or urgent care provider action.
The cloud infrastructure includes comprehensive audit logging of all user actions, data access, and system events. The system provides API access for integration with electronic health records with HL7 v2.x, HL7 FHIR R4, and other standard protocols, enabling bidirectional data exchange with major EHR systems.
4. The web-based Provider Portal
The web-based Provider Portal enables healthcare professionals to access and manage patient data and alert statuses remotely through any compatible web browser. Through the portal, providers can review spot-check measurements and historical trends, playback audio recordings of auscultation sounds captured by the Patch, configure individualized care programs, set measurement schedules and alert thresholds, and communicate with patients through various modalities.
Through the portal, providers can review spot-check measurements with customizable vital sign charts displaying trends over days, weeks, or months. Advanced visualization includes waveform analysis for ECG and PPG signals, audio playback for auscultation recordings, and comprehensive annotation tools. The portal displays signal quality indicators and out-of-range values with appropriate visual highlighting based on configured thresholds. The portal also displays patient severity levels (Low/Medium/High) based on the NEWS2 scoring methodology. Additional clinical measures, such blood pressure and weight, that are manually input into the EHR can be read into the Perin Health System and viewed in the Provider Portal using the EHR interface.
The system employs a structured care program architecture that ensures appropriate clinical oversight throughout the monitoring process. Healthcare organizations create standardized care program templates for common conditions. Individual providers can then select from these approved templates and customize them for specific patient needs, prescribing the specific devices needed, measurement frequencies appropriate to the condition, and recording schedules tailored to clinical requirements.
The portal includes comprehensive communication capabilities supporting both patient and care team interactions. Providers can conduct virtual appointments with integrated video calling, AI-powered real-time transcription using AWS HealthScribe, and automated clinical note generation structured into standard sections. The messaging system supports secure text communication with file attachments, while the task management system enables care coordination across team members. Providers can create and deploy customized questionnaires with various response types and scoring algorithms, manage educational content delivery, and review patient-submitted reports with collaborative response capabilities.
Additional portal features include appointment scheduling with EHR integration, comprehensive alert management with acknowledgment workflows, administrative functions for user and device management, and organization hierarchy configuration. The portal provides detailed audit trails, performance analytics, and compliance reporting to support quality improvement initiatives.
5. The Perin Health Inpatient Module
The Perin Health Inpatient Module provides a monitoring dashboard for monitoring capabilities in healthcare facility environments. The modules leverage the existing architectures for the Mobile Application and Provider Portal but offer unique interfaces for inpatient spot-check measurements.
The web-based monitoring dashboard, a page accessible through the Provider Portal, displays vital signs for up to 50 concurrent patients in a grid layout. Each patient card shows the latest values for heart rate, respiratory rate, oxygen saturation, temperature, and device status, with automatic sorting by alert priority and visual indicators for threshold violations. The dashboard refreshes every second, updating as new spot-check recordings are captured from patients across the unit.
The bedside Inpatient Application is built on top of the Android architecture of the Patient Mobile app and operates in kiosk mode. The Beside app only interfaces with the Perin Health Patch and relays information to the Cloud to provide clinicians with access to recent measurements in the Provider Portal Inpatient view. The application also maintains local data storage for backup operation and automatically synchronizes with the cloud upon connectivity restoration. Providers are unable to manually input clinical data (e.g., blood pressure measurements) directly into the bedside Inpatient Application but manual data input into the EHR can be read into and visualized in the Provider Portal over the EHR interface.
The Perin Health System supports monitoring in hospitals and out-of-hospital patient care settings where care is administered by healthcare professionals. Visual alarm indicators highlight parameter exceedances according to configured thresholds. High-priority alerts display prominently with appropriate color coding, though all clinical responses and acknowledgments must be performed through the Provider Portal to maintain proper documentation and workflow management.
The Perin Health System facilitates comprehensive spot-checking and retrospective monitoring across the continuum of care. Data flows from the wearable patch and third-party devices through the patient mobile application to the central cloud infrastructure, where processing algorithms derive clinical insights. Healthcare providers access this information through the web portal or inpatient displays for clinical review and analysis, enabling healthcare providers to track patient progress, adjust treatment plans based on measurements, and identify patients requiring intervention based on retrospective data trends.
Here's a breakdown of the acceptance criteria and the study details for the Perin Health System (PHD80060-2), based on the provided FDA 510(k) clearance documentation:
Acceptance Criteria and Device Performance Study (Perin Health System PHD80060-2)
1. Acceptance Criteria and Reported Device Performance
The acceptance criteria and reported device performance for key physiological parameters are summarized below:
| Parameter | Acceptance Criteria | Reported Device Performance |
|---|---|---|
| Heart Rate | 20-200 bpm ± 3 BPM or 5%, whichever is greater (based on primary predicate UbiqVue) | High levels of agreement between the Perin Health Patch and the reference Holter monitor across all evaluated parameters for ECG, HR, and HRV for 243 participants. |
| Respiratory Rate | Bench Testing: 5-30 Breaths per Minute ± 1 Breaths per Minute (Accuracy Root Mean Square (Arms)).Clinical Study: ± 3 Breaths per Minute (Accuracy Root Mean Square (Arms)) derived from Trans-thoracic Impedance (TTI) and ECG Derived Respiration (EDR) based on RS Amplitude. (Predicate UbiqVue had ≤ 1 Breath per minute MAE for simulation, ≤ 3 Breaths per minute MAE for clinical study) | Clinical Validation: Arms of 1.7 breaths per minute for 259 points. Subgroups exhibited Arms between 0.5 and 2.8. Clinical Validation: Mean Absolute Error (MAE) of 0.8 breaths per minute for 259 points. Subgroups exhibited MAE between 0.4 and 1.3. |
| Skin Temperature | 15 C - 50°C ± 0.3°C Resolution: 0.008°C Time response: 30 minutes Measurement mode: Direct ISO 80601-2-56 (Matching primary predicate UbiqVue) | Verified by using bench testing as per ISO 80601-2-56:2017(E). (Specific accuracy values beyond "verified" are not explicitly stated for the Perin Health System in this summary, but implied to meet the criteria) |
| SpO2% | 70% - 100% ± 3 % (Predicate UbiqVue 0 to 100% ± 3 % (100 to 70%), Less than 70% unspecified) | Clinical Validation: Overall measured Arms in the range of 70 to 100% SpO2 was 3.3%. Arms of 3.5% for 67% to <80%, 3.1% for 80% to <90%, and 3.3% for 90% to 100%. (This implicitly meets the ± 3% criterion for the 70-100% range, with Arms values slightly above 3% for the lower range. The predicate allows unspecified below 70%). |
| Posture | Prone, supine, left lateral recumbent, right lateral recumbent, Fowler's, Trendelenburg, upright, leaning forward (> 80% average sensitivity and specificity, compared to visual) | Verified by using bench testing as per the acceptance criteria. (Specific sensitivity and specificity values are not explicitly stated beyond "verified" but implied to meet the criteria.) |
| Body Motion | Active or sedentary (> 90% sensitivity and specificity) | Verified by using bench testing as per the acceptance criteria. (Specific sensitivity and specificity values are not explicitly stated beyond "verified" but implied to meet the criteria.) |
| Fall Detection | Fall or no fall (> 80% sensitivity and specificity) | Verified by using bench testing as per the acceptance criteria. (Specific sensitivity and specificity values are not explicitly stated beyond "verified" but implied to meet the criteria.) |
| Step Count | < 5% Absolute Error Compared to Manual Count for speeds of at least 2 miles per hour Compliance: ANSI/CTA-2056-A | Verified via bench testing as per ANSI/CTA-2056-A. (Specific absolute error is not explicitly stated beyond "verified" but implied to meet the criteria.) |
| Auscultation Data | Verified by using bench testing in accordance with acceptance criteria. (No specific numerical accuracy given) | Verified by using bench testing in accordance with acceptance criteria. (Specific performance metrics are not detailed beyond meeting acceptance criteria). |
| ECG, R-R Interval, HRV | Performance testing in compliance with ANSI/AAMI/IEC 60601-2-27:2011, ANSI/AAMI/IEC 60601-2-47:2012 (No specific numerical accuracy given for these parameters directly here) | Clinical Validation: Demonstrated high levels of agreement between the Perin Health Patch and the reference Holter monitor across all evaluated parameters (timing intervals, SNR, morphological features) and for all demographic and clinical subgroups for 243 participants. |
| Wear-life | Sustained adhesion to the body for 360 hours. | Demonstrated stable performance across all evaluated parameters (timing intervals, SNR, morphological features) and for all demographic and clinical subgroups over 360 hours. |
Note: For several parameters (Skin Temperature, Posture, Body Motion, Fall Detection, Step Count, Auscultation data), the document states they were "verified by using bench testing as per the acceptance criteria" or "in accordance with acceptance criteria," implying they met the specified thresholds without explicitly re-stating the achieved performance metrics.
2. Sample Size Used for the Test Set and Data Provenance
-
SpO2% (Induced Hypoxia Study):
- Sample Size: 12 healthy adults (5 female, 7 male)
- Data Provenance: Not explicitly stated (e.g., country of origin), but implied to be prospective clinical validation conducted for this submission.
-
Respiratory Rate (Clinical Validation):
- Sample Size: 35 participants (17 males, 18 females)
- Data Provenance: Not explicitly stated (e.g., country of origin), but implied to be prospective clinical validation conducted for this submission.
-
ECG, Heart Rate, R-R Interval, and Heart Rate Variability (Clinical Validation):
- Sample Size: 243 participants
- Data Provenance: Not explicitly stated (e.g., country of origin), but implied to be prospective clinical validation conducted for this submission.
-
Wear-life Performance (Internal Clinical Wear Life Evaluation):
- Sample Size: 26 participants
- Data Provenance: Across 3 clinical sites. Implied to be prospective clinical evaluation.
3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts
The document does not explicitly state the number or specific qualifications of experts used to establish ground truth for the clinical test sets. However, it references:
- SpO2%: "arterial blood samples analyzed by a laboratory co-oximeter" as the gold standard. This implies specialized laboratory personnel for analysis, but their number and specific qualifications are not detailed.
- Respiratory Rate: "manually counted end-tidal CO2" as the gold standard. This would typically be performed by trained clinical staff, but their number and qualifications are not specified.
- ECG, HR, HRV: "standard Holter monitor" as the reference for comparison. Interpretation of Holter data would involve cardiologists or trained technicians, but the document doesn't specify if this was used as "ground truth" to establish the Holter reference itself or if it refers to the Holter output as the reference measurement.
4. Adjudication Method for the Test Set
The document does not describe any specific adjudication method (e.g., 2+1, 3+1, none) for the test sets. The studies compare the device's measurements directly to a "gold standard" or "reference monitor" without mentioning a multi-reader adjudication process for discrepancies.
5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study
There is no indication of a Multi-Reader Multi-Case (MRMC) comparative effectiveness study being done to evaluate how much human readers improve with AI vs. without AI assistance. The document focuses on the standalone performance of the device's measurements against established standards.
6. Standalone (Algorithm Only Without Human-in-the-Loop Performance) Study
Yes, standalone performance was done for several key parameters. The clinical validation studies directly assess the Perin Health System's ability to measure physiological data (SpO2%, Respiratory Rate, ECG/HR/HRV) against a specified gold standard or reference device. These studies inherently evaluate the algorithm's performance without direct human interpretation influencing the measurement output. For example:
- SpO2% accuracy is measured against arterial blood samples.
- Respiratory rate accuracy is measured against manually counted end-tidal CO2.
- ECG, HR, HRV performance is validated against a standard Holter monitor.
7. Type of Ground Truth Used
The types of ground truth used for the clinical validation studies include:
- Laboratory Standard / Direct Measurement: For SpO2%, the ground truth was "arterial blood samples analyzed by a laboratory co-oximeter."
- Clinical Gold Standard: For Respiratory Rate, the ground truth was "manually counted end-tidal CO2."
- Reference Clinical Device: For ECG, Heart Rate, R-R Interval, and Heart Rate Variability, the ground truth/reference was a "standard Holter monitor."
8. Sample Size for the Training Set
The document does not provide any information regarding the sample size for the training set. This information is typically proprietary to the manufacturer and not usually disclosed in 510(k) summaries unless specifically relevant to a novel AI/ML algorithm requiring such details for FDA review.
9. How the Ground Truth for the Training Set Was Established
Since no information about the training set or its sample size is provided, there is no information available on how the ground truth for the training set was established.
Ask a specific question about this device
(268 days)
W0001ES, W0028ES, W0101FS-A, W0101FS-P, W0099LS, W0099FS, W0099PS)
Regulation Number: 21 CFR 880.2910
W0099PS)
Classification name: Clinical Electronic Thermometer
Regulation number: 21 CFR 880.2910
Monitoring Probe | Temperature Monitoring Probe | DeRoyal Industries, Inc. | Class II, FLL | 21 CFR 880.2910
The sterile temperature probe is used with Mindray BeneVision N17 patient monitor to monitor body or skin surface temperature. The device is for use by qualified healthcare personnel.
The probe is offered in the following three configurations:
- Body cavity temperature probe for monitoring of the core temperature in adult and pediatric patients by insertion into the esophageal; and for monitoring of the body temperature in adult and pediatric patients by insertion into the rectal or nasopharyngeal cavity.
- Skin contact temperature probe for monitoring of skin temperature to an adult and pediatric patient’s skin surface.
- Ear cavity temperature probe for monitoring of body temperature by insertion of the foam into pediatric or adult’s auditory canal.
For pediatric population, the subpopulations are children and adolescents.
The sterile temperature probe is a compatible sensor for use with Mindary BeneVision N17 patient monitor, as an accessory of the legally marketed patient monitor on the US market, the sterile temperature probe is indicated for continuous monitoring of body temperature in hospital settings. It mainly consists of plug, cable, and probe at patient side. When in use, the probe should be connected to the compatible device, and the temperature change of the patient's measured part is sensed through the built-in NTC thermistor at the probe end and the patient's body temperature is measured. The subject devices are used with Mindray BeneVision N17 patient monitor, which was cleared under K182075.
The temperature probes are packed individually into a paper plastic pouch in sterile condition and are single use only.
N/A
Ask a specific question about this device
(128 days)
plethysmograph | 21 CFR 870.2770 |
| | Thermometer, Electronic, Clinical | 21 CFR 880.2910
plethysmograph | 21 CFR 870.2770 |
| | Thermometer, Electronic, Clinical | 21 CFR 880.2910
HemoSphere Alta™ Advanced Monitoring Platform with Swan-Ganz™ Technology
The HemoSphere Alta™ Advanced Monitor when used with the HemoSphere Alta Swan-Ganz™ Patient Cable and Swan-Ganz™ Catheters is indicated for use in adult and pediatric critical care patients requiring monitoring of cardiac output (continuous [CO] and intermittent [iCO]) and derived hemodynamic parameters in a hospital environment. Pulmonary artery blood temperature monitoring is used to compute continuous and intermittent CO with thermodilution technologies. It may be used for monitoring hemodynamic parameters in conjunction with a perioperative goal directed therapy protocol in a hospital environment. Refer to the Swan-Ganz™ Catheter and Swan-Ganz Jr™ Catheter indications for use statement for information on target patient population specific to the catheter being used.
The Global Hypoperfusion Index (GHI) algorithm provides the clinician with physiological insight into a patient's likelihood of future hemodynamic instability. The GHI algorithm is intended for use in surgical or non-surgical patients receiving advanced hemodynamic monitoring with the Swan-Ganz™ Catheter. The GHI algorithm is considered to provide additional information regarding the patient's predicted future risk for clinical deterioration, as well as identifying patients at low risk for deterioration. The product predictions are for reference only and no therapeutic decisions should be made based solely on the GHI algorithm predictions.
When used in combination with a Swan-Ganz™ Catheter connected to a pressure cable and pressure transducer, the Smart Wedge™ Algorithm measures and provides pulmonary artery occlusion pressure and assesses the quality of the pulmonary artery occlusion pressure measurement. The Smart Wedge™ Algorithm is indicated for use in critical care patients over 18 years of age receiving advanced hemodynamic monitoring. The Smart Wedge™ Algorithm is considered to be additional quantitative information regarding the patient's physiological condition for reference only and no therapeutic decisions should be made based solely on the Smart Wedge™ Algorithm parameters.
HemoSphere Alta™ Advanced Monitoring Platform with HemoSphere™ Oximetry Cable
The HemoSphere Alta™ Advanced Monitor when used with the HemoSphere™ Oximetry Cable and oximetry catheters is indicated for use in adult and pediatric critical care patients requiring monitoring of venous oxygen saturation (SvO2 and ScvO2) and derived hemodynamic parameters in a hospital environment. Refer to the oximetry catheter indications for use statement for information on target patient population specific to the catheter being used.
HemoSphere Alta™ Advanced Monitoring Platform with HemoSphere™ Pressure Cable or HemoSphere Alta™ Monitor - Pressure Cable
The HemoSphere Alta™ Advanced Monitor when used with the HemoSphere™ Pressure Cable or HemoSphere Alta™ Monitor – Pressure Cable is indicated for use in adult and pediatric critical care patients in which the balance between cardiac function, fluid status, vascular resistance and pressure needs continuous assessment. It may be used for monitoring of hemodynamic parameters in conjunction with a perioperative goal directed therapy protocol in a hospital environment. Refer to the FloTrac™ Sensor, FloTrac Jr™ Sensor, Acumen IQ™ Sensor, and TruWave™ Disposable Pressure Transducer indications for use statements for information on target patient populations specific to the sensor/transducer being used.
The Acumen Hypotension Prediction Index™ Software Feature (HPI™ Parameter) provides the clinician with physiological insight into a patient's likelihood of future hypotensive events and the associated hemodynamics. The Acumen HPI™ Feature is intended for use in surgical or non-surgical patients receiving advanced hemodynamic monitoring. The Acumen HPI™ Feature is considered to be additional quantitative information regarding the patient's physiological condition for reference only and no therapeutic decisions should be made based solely on the Acumen Hypotension Prediction Index™ Parameter.
When used in combination with the HemoSphere™ Pressure Cable or HemoSphere Alta™ Monitor – Pressure Cable connected to a compatible Swan-Ganz™ Catheter, the Right Ventricular Pressure (RVP) algorithm provides the clinician with physiological insight into the hemodynamic status of the right ventricle of the heart. The RVP algorithm is indicated for critically ill patients over 18 years of age receiving advanced hemodynamic monitoring in the operating room (OR) and intensive care unit (ICU). The RVP algorithm is considered to be additional quantitative information regarding the patient's physiological condition for reference only and no therapeutic decisions should be made based solely on the Right Ventricular Pressure (RVP) parameters.
When used in combination with the HemoSphere™ Pressure Cable or HemoSphere Alta™ Monitor – Pressure Cable connected to a compatible Swan-Ganz™ Catheter, the Right Ventricular Cardiac Output (RVCO) feature provides the clinician with physiological insight into the hemodynamic status of the right ventricle of the heart. The RVCO algorithm is intended for use in surgical or non-surgical patients over 18 years of age that require advanced hemodynamic monitoring. The Right Ventricular Cardiac Output provides a continuous cardiac output and derived parameters.
The Cerebral Autoregulation Index (CAI) algorithm is an informational index intended to represent a surrogate measurement of whether cerebral autoregulation is likely intact or is likely impaired as expressed by the level of coherence or lack thereof between Mean Arterial Pressure (MAP) and the Absolute Levels of Blood Oxygenation Saturation (StO2) in patient's cerebral tissue. MAP is acquired by the HemoSphere™ Pressure Cable and StO2 is acquired by the ForeSight™ Oximeter Cable. CAI is intended for use in patients over 18 years of age receiving advanced hemodynamic monitoring. CAI is not indicated to be used for treatment of any disease or condition and no therapeutic decisions should be made based solely on the Cerebral Autoregulation Index (CAI) algorithm.
HemoSphere Alta Advanced Monitoring Platform with ForeSight™ Oximeter Cable
The non-invasive ForeSight™ Oximeter Cable is intended for use as an adjunct monitor of absolute regional hemoglobin oxygen saturation of blood under the sensors in individuals at risk for reduced flow or no-flow ischemic states. The ForeSight™ Oximeter Cable is also intended to monitor relative changes of total hemoglobin of blood under the sensors. The ForeSight™ Oximeter Cable is intended to allow for the display of StO2 and relative change in total hemoglobin on the HemoSphere Alta™ Advanced Monitoring Platform.
• When used with large sensors, the ForeSight™ Oximeter Cable is indicated for use on adults and transitional adolescents ≥40 kg.
• When used with medium sensors, the ForeSight™ Oximeter Cable is indicated for use on pediatric subjects ≥3 kg.
• When used with small sensors, the ForeSight™ Oximeter Cable is indicated for cerebral use on pediatric subjects <8 kg and non-cerebral use on pediatric subjects <5kg.
The algorithm for measurement of blood hemoglobin is indicated for continuously monitoring changes to hemoglobin concentration in the circulating blood of adults ≥40 kg receiving advanced hemodynamic monitoring using HemoSphere ForeSight™ Oximeter Cable and noninvasive ForeSight IQ™ Sensors in cerebral locations.
HemoSphere Alta™ Advanced Monitoring Platform with Non-invasive technology
The HemoSphere Alta™ Monitor when used with the pressure controller and a compatible finger cuff are indicated for adult and pediatric patients in which the balance between cardiac function, fluid status and vascular resistance needs continuous assessment. It may be used for monitoring hemodynamic parameters in conjunction with a perioperative goal directed therapy protocol in a hospital environment. In addition, the non-invasive system is indicated for use in patients with co-morbidities for which hemodynamic optimization is desired and invasive measurements are difficult. The HemoSphere Alta™ Advanced Monitor and compatible finger cuffs non-invasively measures blood pressure and associated hemodynamic parameters. Refer to the non-invasive finger cuff indications for use statements for information on target patient population specific to the finger cuff being used.
The Acumen Hypotension Prediction Index™ Software Feature (HPI™ Parameter) provides the clinician with physiological insight into a patient's likelihood of future hypotensive events and the associated hemodynamics. The Acumen HPI™ Feature is intended for use in surgical or non-surgical patients receiving advanced hemodynamic monitoring. The Acumen HPI™ Feature is considered to be additional quantitative information regarding the patient's physiological condition for reference only and no therapeutic decisions should be made based solely on the Acumen Hypotension Prediction Index™ Parameter.
HemoSphere Alta Advanced Monitoring Platform with Acumen Assisted Fluid Management Feature and Acumen IQ Sensor
The Acumen assisted fluid management (AFM) software feature provides the clinician with physiological insight into a patient's estimated response to fluid therapy and the associated hemodynamics. The Acumen AFM software feature is intended for use in surgical patients ≥18 years of age, that require advanced hemodynamic monitoring. The Acumen AFM software feature offers suggestions regarding the patient's physiological condition and estimated response to fluid therapy. Acumen AFM fluid administration suggestions are offered to the clinician; the decision to administer a fluid bolus is made by the clinician, based upon review of the patient's hemodynamics. No therapeutic decisions should be made based solely on the assisted fluid management suggestions.
The HemoSphere Alta Advanced Monitoring Platform is the next-generation platform that provides a means to interact with and visualize hemodynamic and volumetric data on a screen. It incorporates a comprehensive view of patient hemodynamic parameters with an intuitive and easy user interface. The HemoSphere Alta Advanced Monitoring Platform is designed to provide monitoring of cardiac flow with various core technologies coupled with other technologies-based features such as Algorithms and Interactions. It integrates existing hemodynamic monitoring technologies into a unified platform.
The HemoSphere Alta Advanced Monitoring Platform's FDA 510(k) clearance letter and associated 510(k) summary (K252533) primarily focus on software modifications and the integration of previously cleared hardware components to an existing platform (K242451). The document states that no new clinical testing was performed in support of the subject 510(k). Therefore, the information provided mainly pertains to performance verification studies rather than standalone clinical performance studies involving ground truth establishment by experts for a novel algorithm.
However, based on the provided text, we can infer the acceptance criteria and study information as follows:
1. Table of Acceptance Criteria and Reported Device Performance
The document describes several verification activities without providing specific numerical acceptance criteria for each, except implicitly stating "All tests passed" or "All acceptance criteria were met."
| Acceptance Criteria Category | Reported Device Performance |
|---|---|
| Usability | Demonstrated that intended users can perform primary operating functions and critical tasks without usability issues that may lead to patient or user harm. All acceptance criteria were met for human factors validation. |
| System Verification (Non-Clinical Performance) | Met predetermined design and performance specifications. Differences in design and materials did not adversely affect safety and effectiveness. All tests passed. |
| Electrical Safety and EMC | Complies with IEC 60601-1, IEC 60601-1-2, IEC 60601-1-6, IEC 60601-1-8, IEC 62304, IEC 62366-1, IEC 60601-2-34, IEC 60601-2-57, IEC 60601-2-49, IEC 60529-1, and IEC 80601-2-49. All tests passed. |
| Software Verification | Extensive software verification testing ensured safety for use, demonstrating substantial equivalence to predicate devices. All tests passed. |
2. Sample Size Used for the Test Set and Data Provenance
- For Usability Study: The document mentions "the intended users." It does not specify the numerical sample size of users or the provenance of the data (e.g., retrospective or prospective, country of origin).
- For Non-Clinical Performance (Bench Simulation): "Measured and derived parameters were tested using a bench simulation." No sample size in terms of patient data or data provenance is applicable here, as it's a bench test.
- For Software Verification: "Extensive software verification testing was conducted." No specific sample size of test cases or data provenance is provided.
- For Clinical Performance: "No new clinical testing was performed in support of the subject 510(k)." This indicates no patient-level test set data was used for this specific submission. The algorithms within the device (e.g., GHI, Smart Wedge, HPI, CAI, RVP, RVCO, AFM) likely had clinical performance studies for their initial clearances, but those details are not provided in this 510(k) for the HemoSphere Alta platform updates.
3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications
- As "no new clinical testing was performed" for this 510(k) submission, there is no mention of experts establishing ground truth for a new clinical test set.
- For the Usability Study, "intended users" participated, implying clinical professionals, but their specific qualifications or their role in establishing "ground truth" (beyond identifying usability issues) are not detailed.
4. Adjudication Method for the Test Set
- Since no new clinical test set data with expert adjudication is described in this submission, no adjudication method is mentioned.
5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study
- The document does not mention any MRMC comparative effectiveness study comparing human readers with and without AI assistance for this 510(k) submission.
6. Standalone Performance (Algorithm Only Without Human-in-the-Loop)
- While the device contains various algorithms (e.g., GHI, HPI, Smart Wedge, CAI, RVP, RVCO, AFM), this 510(k) primarily addresses software updates and hardware integration to an existing platform. It doesn't detail standalone performance studies for these specific algorithms within this document. The description of these algorithms (e.g., "additional information regarding the patient's physiological condition for reference only and no therapeutic decisions should be made based solely on the GHI algorithm predictions") implies a non-standalone, assistive role, but explicit standalone performance studies are not part of this submission's provided information.
7. Type of Ground Truth Used
- "No new clinical testing was performed." Therefore, for this specific 510(k) submission, no new patient-level ground truth (expert consensus, pathology, outcomes data, etc.) was established for performance evaluation of new algorithms or features. The verification activities relied on bench simulations and usability testing, not clinical ground truth.
8. Sample Size for the Training Set
- The document pertains to the clearance of a device (HemoSphere Alta Advanced Monitoring Platform) with software modifications and hardware integration, not the development or training of new AI algorithms. Therefore, no information on the sample size of a training set is provided. The algorithms included in the HemoSphere Alta system (e.g., GHI, HPI, CAI) would have been developed and trained using data sets prior to their initial clearance. This current 510(k) does not detail those previous training sets.
9. How the Ground Truth for the Training Set Was Established
- Similar to the training set sample size, this information is not provided in this 510(k) document, as it focuses on software updates and hardware integration to an already cleared platform, not the initial development and training of novel algorithms.
Ask a specific question about this device
(271 days)
. |
| General Hospital | §880.2910, II | FLL | Thermometer, Electronic, Clinical |
| Anesthesiology |
The N10, N12, N15, N10MPro, N12MPro, NM15Pro Multi-parameter Patient Monitors are intended for monitoring, displaying, reviewing, storing, alarming and transferring of multiple physiological parameters including ECG (3-lead, 5-lead, 12-lead selectable), Arrhythmia Analysis, ST Segment Analysis, QT Analysis, Heart Rate (HR) and Heart-Rate-Variability(HRV)), interpretations of resting 12-lead ECG, Respiration rate(Resp), Temperature(Temp), Pulse Oxygen Saturation (SpO2), Pulse Rate (PR), Non-invasive Blood Pressure (NIBP), Invasive Blood Pressure (IBP), Pulmonary Artery Wedge Pressure (PAWP), Cardiac Output (C.O.), Carbon Dioxide (CO2). The N10MPro, N12MPro, NM15Pro Multi-parameter Patient Monitors are also intended for monitoring, displaying, reviewing, storing, alarming and transferring of physiological parameters including Masimo Rainbow SpO2, Anesthesia gas (AG), oxygen (O2) respiratory gas monitoring, Bispectral Index (BIS), Respiration Mechanics (RM) and Neuromuscular Transmission Monitoring (NMT). All the parameters can be monitored on single adult, pediatric, and neonatal patient except for the following:
- Arrhythmia analysis is intended to use on adult patients only and is not intended and shall not be used on pediatric and neonatal population.
- NIBP measurement continual mode is not applicable to neonates.
- When using COMEN SpO2, the monitor is intended to be used on adult patient only.
- PAWP is intended for adult and pediatric patients only.
- C.O. measurement is intended for adult patients only.
- BIS monitoring is intended for adult patients only.
- RM is intended for adult and pediatric patients only.
- NMT monitoring is intended for adult and pediatric patients only.
The monitors are to be used in healthcare facilities by healthcare professionals or under their guidance.
The Multi-parameter Patient monitors are not intended for emergency and transport use, aircraft environment or home use.
The monitors are not intend for use as apnea monitors.
The monitors are not intended for use in MRI or CT environments.
The monitors are not used on patients who have a demonstrated need for cardiac monitoring known arrhythmias of VT, Accelerated Idioventricular rhythm and Torsades de Pointes.
There are six (6) models under evaluation, namely N10, N12, N15, N10MPro, N12MPro, N15MPro. All models share the same intended condition of use, the same intended patient population and operator profile, biological safety characteristic and principle of operation. All these models are the same on electric and electrical circuit and components, mechanical construction, software and alarm system. The only difference lies on the screen and configuration of with/without plug-in module slot and the number of battery packs.
N/A
Ask a specific question about this device
(96 days)
Transmitters and Receivers, Telephone
Regulation Number: 21 CFR 870.2340, 21 CFR 870.1875, 21 CFR 880.2910
510(k)** | K252474 | K130921 | K160544 | K181612 | K170874 | N/A |
| Regulation Number | 21 CFR 880.2910
, 21 CFR 870.1875, 21 CFR 870.2340 | 21 CFR 870.2340 | 21 CFR 880.2910 | 21 CFR 870.1875 | 21 CFR 870.1875
Withings BeamO is intended to record, display (when prescribed or used under the care of a physician), store, and transfer single-channel electrocardiogram (ECG) rhythms. It is indicated for use with individuals 22 years and older.
Withings BeamO is a non-sterile, contactless, reusable clinical thermometer intended for the intermittent determination of human body temperature over the temporal artery as the measurement site on people of all ages.
Withings BeamO is also an electronic stethoscope that enables the recording and transmission of auscultation sound data. Withings BeamO is intended to be used by professional users in a clinical environment or by lay users in a non-clinical environment on people of all ages. The electronic stethoscope is for medical diagnostics purposes only. The device is not intended for self-diagnosis.
Withings BeamO, model name SCT02, is a multi-function handheld battery powered device with ECG, stethoscope, temperature capabilities. Withings BeamO can record a 1-lead ECG using two stainless steel electrodes. It analyzes the data collected by the integrated two stainless steel electrodes to generate an one-lead ECG waveform and provides the ECG recording to the user for a given 30 second measurement.
Withings BeamO is also a contactless thermometer that can measure body temperature in adjusted mode.
Withings BeamO is also a digital stethoscope that can be used to auscultate heart and lung sounds. The sensor generates an electric charge when subjected to mechanical vibrations. The charge variations are amplified and digitized by an audio codec. Sound filters are applied to the resulting sound wave in order to listen to the patient's heart and lung sounds with clarity.
Withings BeamO consists of hardware and embedded software. Withings BeamO works in conjunction with a companion software on the Withings App. Withings BeamO communicates with the companion software via Bluetooth Low Energy (BLE). The device measurement results and recordings are synchronized with the companion software using Wi-Fi/Cellular data via the Withings servers.
Withings BeamO does not include ECG analysis or ECG-derived heart rate functionalities.
N/A
Ask a specific question about this device
(425 days)
accessories, HGL
- 21 CFR 870.2300 Cardiac monitor (including cardiotachometer and rate alarm), DRT
- 21 CFR 880.2910
Fetal & Maternal Monitor (Model: F15A, F15A Air) is intended for providing Non-Stress testing or fetal monitoring for pregnant women from the 28th week of gestation. It is intended to be used only by trained and qualified personnel in antepartum examination rooms, labor and delivery rooms.
Fetal & Maternal Monitor (Model: F15A, F15A Air) is intended for real time monitoring of fetal and maternal physiological parameters, including non-invasive monitoring and invasive monitoring:
Non-invasive physiological parameters:
- Maternal heart rates (MHR)
- Maternal ECG (MECG)
- Maternal temperature (TEMP)
- Maternal oxygen saturation (SpO2) and pulse rates (PR)
- Fetal heart rates (FHR)
- Fetal movements (FM)
- FTS-3
Note: SpO2 and PR are not available in F15A Air.
Invasive physiological parameters:
- Uterine activity
- Direct ECG (DECG)
The F15A series fetal and maternal monitor can monitor multiple physiological parameters of the fetus/mother in real time. F15A series can display, store, and print patient information and parameters, provide alarms of fetal and maternal parameters, and transmit patient data and parameters to Central Monitoring System.
F15A series fetal and maternal monitors mainly provide following primary feature:
Non-invasive physiological parameters:
- Maternal heart rates (MHR)
- Maternal ECG (MECG)
- Maternal temperature (TEMP)
- Maternal oxygen saturation (SpO2) and pulse rates (PR)
- Fetal heart rates (FHR)
- Fetal movements (FM)
- FTS-3
Note: SpO2 and PR are not available in F15A Air.
Invasive physiological parameters:
- Uterine activity
- Direct ECG (DECG)
The provided FDA 510(k) clearance letter and summary for the Fetal & Maternal Monitor (F15A, F15A Air) do not contain the detailed information necessary to fully answer all aspects of your request regarding acceptance criteria and the study that proves the device meets them.
The document focuses primarily on demonstrating substantial equivalence to a predicate device (Edan Instruments, Inc., F9 Express Fetal & Maternal Monitor, K173042) through comparison of intended use, technological characteristics, and conformance to various safety and performance standards. It mentions "functional and system level testing to validate the performance of the devices" and "results of the bench testing show that the subject device meets relevant consensus standards," but it does not specify quantitative acceptance criteria for each individual physiological parameter (e.g., FHR accuracy, SpO2 accuracy) nor the specific results of those tests beyond stating that they comply with standards.
Specifically, the document does not include information on:
- A table of acceptance criteria with specific quantitative targets for each parameter and the reported device performance values against those targets. It only states compliance with standards.
- Sample sizes used for a "test set" in the context of clinical performance evaluation (it mentions "bench testing," but this is typically laboratory-based and doesn't involve patient data in a "test set" sense for AI/algorithm performance validation).
- Data provenance for such a test set (e.g., country of origin, retrospective/prospective).
- Number or qualifications of experts used to establish ground truth.
- Adjudication methods.
- Multi-Reader Multi-Case (MRMC) studies or human reader improvement data with AI assistance.
- Standalone (algorithm-only) performance, as this is a monitoring device, not a diagnostic AI algorithm.
- Type of ground truth (beyond "bench testing" which implies engineered signals or controlled environments).
- Sample size for a training set or how ground truth for a training set was established. This device is a traditional medical device, not an AI/ML-driven diagnostic or interpretative algorithm in the way your request implies.
Therefore, based solely on the provided text, I can only address what is present or infer what is missing.
Here's a breakdown based on the available information:
Analysis of Acceptance Criteria and Performance Testing based on Provided Document
The provided 510(k) summary focuses on demonstrating substantial equivalence to a predicate device (F9 Express Fetal & Maternal Monitor, K173042) by showing that the new device (F15A, F15A Air) has the same intended use and fundamentally similar technological characteristics, with any differences not raising new safety or effectiveness concerns.
1. A table of acceptance criteria and the reported device performance
The document does not provide a specific table with quantitative acceptance criteria for each physiological parameter (e.g., FHR accuracy, SpO2 accuracy) and the corresponding reported performance values obtained in testing. Instead, it states that the device was assessed for conformity with relevant consensus standards. For example, it lists:
- IEC 60601-2-37:2015: Particular requirements for the basic safety and essential performance of ultrasonic medical diagnostic and monitoring equipment (relevant for FHR).
- ISO 80601-2-61:2017+A1:2018: Particular requirements for basic safety and essential performance of pulse oximeter equipment (relevant for SpO2).
- ISO 80601-2-56:2017+A1:2018: Particular requirements for basic safety and essential performance of clinical thermometers for body temperature measurement (relevant for TEMP).
- IEC 60601-2-27:2011: Particular requirements for the basic safety and essential performance of electrocardiographic monitoring equipment (relevant for MECG/DECG).
Acceptance Criteria (Inferred from standards compliance): The acceptance criteria are implicitly the performance requirements specified within these listed consensus standards. These standards set limits for accuracy, precision, response time, and other performance metrics for each type of measurement.
Reported Device Performance: The document states: "The results of the bench testing show that the subject device meets relevant consensus standards." This implies that the measured performance statistics (e.g., accuracy, bias, precision) for each parameter fell within the acceptable limits defined by the respective standards. However, the specific measured values are not provided in this summary.
2. Sample sized used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)
The document mentions "Bench Testing" which implies laboratory-based testing using simulators, controlled signals, or phantoms, rather than a "test set" involving patient data. There is no information provided regarding:
- Sample size (e.g., number of recordings, duration of recordings, number of simulated cases) for the bench tests for each parameter.
- Data provenance (e.g., country of origin, retrospective or prospective) as this is not a study involving patient data collection for performance validation.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)
This is not applicable and not provided. For a traditional physiological monitor, ground truth for bench testing is typically established using:
- Calibrated reference equipment/simulators: e.g., ECG simulators to generate known heart rates, SpO2 simulators to generate known oxygen saturation levels.
- Physical standards/phantoms: e.g., temperature baths at known temperatures.
- Known physical properties: e.g., precise weights for pressure transducers.
Clinical experts are not involved in establishing ground truth for bench performance of these types of physiological measurements.
4. Adjudication method (e.g. 2+1, 3+1, none) for the test set
This is not applicable and not provided. Adjudication methods are relevant for human expert review of complex clinical data (e.g., medical images for AI validation) to establish a consensus ground truth. For bench testing of physiological monitors, ground truth is objectively determined by calibrated instruments or defined physical parameters.
5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
This is not applicable and not provided. An MRMC study is typically performed to evaluate the diagnostic accuracy of AI-assisted human interpretations versus unassisted human interpretations for AI-driven diagnostic devices. The Fetal & Maternal Monitor is a physiological monitoring device, not an AI-assisted diagnostic imaging or interpretation system. It measures and displays physiological parameters; it does not provide AI-driven assistance for human "readers" to interpret complex clinical information.
6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done
The device is a monitor that directly measures physiological parameters. It is not an "algorithm only" device in the sense of an AI model providing a diagnostic output. Its performance (e.g., FHR accuracy) is its standalone performance, as it directly measures these parameters. The document states "functional and system level testing to validate the performance of the devices," which would represent this type of standalone performance for the measurement functionalities. However, specific quantitative results are not given, only compliance with standards.
7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)
As explained in point 3, the ground truth for bench testing of physiological monitors is established using calibrated reference equipment/simulators and physical standards.
8. The sample size for the training set
This is not applicable and not provided. This device is a traditional physiological monitor, not a machine learning model that requires a "training set." Its algorithms for parameter measurement are based on established physiological principles and signal processing techniques, not on statistical learning from large datasets.
9. How the ground truth for the training set was established
This is not applicable and not provided for the same reasons as point 8.
Ask a specific question about this device
(188 days)
Measurement, Blood-Pressure, Non-Invasive | DXN |
| 21 CFR 870.1110 Computer, Blood-Pressure | DSK |
| 21 CFR 880.2910
The monitors are intended to be used for monitoring, storing, recording, and reviewing of, and to generate alarms for, multiple physiological parameters of adults and pediatrics (including neonates). The monitors are intended for use by trained healthcare professionals in hospital environments.
The monitored physiological parameters include: ECG, respiration (RESP), temperature (TEMP), functional oxygen saturation of arterial hemoglobin (SpO₂), pulse rate (PR), non-invasive blood pressure (NIBP), invasive blood pressure (IBP), carbon dioxide (CO2), and cardiac output (C.O.).
The arrhythmia detection and ST Segment analysis are intended for adult patients.
The NIBP monitoring supports iCUFS algorithm and iFAST algorithm. The iCUFS algorithm is intended for adult, pediatric and neonatal patients. The iFAST algorithm is intended for adult and pediatric patients (≥3 years of age). Both measurement algorithms are also intended for use with pregnant women, including pre-eclamptic patients. NIBP MAP is not applicable to pregnant women.
The Spot Temp with T2A module can only measure temperature of adult and pediatric (> 1 year of age) patients.
The monitors are not intended for MRI environments.
The cardiac output (C.O.) is only intended for adult patients.
The CX&UX series Patient Monitor including CX10/CX12/CX15/UX10/UX12/UX15 can perform long-time continuous monitoring of multiple physiological parameters. Also, it is capable of storing, displaying, analyzing and controlling measurements, and it will indicate alarms in case of abnormalities so that doctors and nurses can respond to the patient's situation as appropriate.
Minor differences from the predicate device are limited to some modifications of monitoring parameter specifications. These updates do not change the fundamental scientific technology of the cleared predicate device and thus do not raise any questions about the safety and effectiveness of the subject device.
The provided FDA 510(k) clearance letter details the device's technical specifications and comparisons to predicate devices, along with the non-clinical performance data and adherence to various IEC and ISO standards. However, it explicitly states: "Clinical data: The subject device did not require new clinical studies to support substantial equivalence."
This means that the submission for this Patient Monitor device (CX10, CX12, CX15, UX10, UX12, UX15) relies on demonstrating substantial equivalence to a legally marketed predicate device (Edan Instruments, Inc., Patient Monitor Model iX10, iX12, iX15, K232962) through non-clinical performance testing and software verification/validation, rather than new clinical trials or studies involving human patients.
Therefore, the requested information regarding acceptance criteria and studies that prove the device meets acceptance criteria through clinical performance (e.g., sample size for test set, expert involvement, MRMC studies, ground truth establishment for test/training sets, effect size of human reader improvement with AI) cannot be extracted from this document, as such clinical studies were explicitly not required for this 510(k) submission.
The document focuses on demonstrating that the new device's technical specifications and performance are similar to the predicate device, and that it complies with relevant safety and performance standards through bench testing.
Here's what can be extracted from the provided text regarding acceptance criteria and the type of study performed, specifically focusing on the non-clinical aspects:
Device: Patient Monitor (CX10, CX12, CX15, UX10, UX12, UX15)
The acceptance criteria for this device are implicitly tied to its performance meeting the standards and accuracy specifications of the predicate device and relevant international standards. Since no new clinical studies were conducted, the "proof" comes from non-clinical bench testing and software validation.
1. Table of Acceptance Criteria and Reported Device Performance (Non-Clinical/Bench Testing)
| Parameter/Acceptance Criteria Type | Details of Acceptance Criteria (Implicit from Standards Compliance & Predicate Equivalence) | Reported Device Performance (as demonstrated by compliance) |
|---|---|---|
| Electrical Safety | Compliance with IEC 60601-1 Edition 3.2 2020-08 | Complies with IEC 60601-1 Edition 3.2 2020-08 |
| Electromagnetic Compatibility (EMC) | Compliance with IEC 60601-1-2:2014 (Fourth Edition) | Complies with IEC 60601-1-2:2014 (Fourth Edition) |
| Alarm Systems | Compliance with IEC 60601-1-8:2020 (General requirements, tests, and guidance for alarm systems) | Complies with IEC 60601-1-8:2020 |
| ECG Monitoring Equipment Performance | Compliance with IEC 60601-2-27:2011 (Basic safety and essential performance of electrocardiographic monitoring equipment) | Complies with IEC 60601-2-27:2011 |
| Invasive Blood Pressure Monitoring Performance | Compliance with IEC 60601-2-34:2011 (Basic safety, including essential performance, of invasive blood pressure monitoring equipment) | Complies with IEC 60601-2-34:2011 |
| Automated Non-Invasive Sphygmomanometers Performance | Compliance with IEC 80601-2-30:2018 (Basic safety and essential performance of automated non-invasive sphygmomanometers) | Complies with IEC 80601-2-30:2018 |
| Multifunction Patient Monitoring Performance | Compliance with IEC 80601-2-49:2018 (Basic safety and essential performance of multifunction patient monitoring equipment) | Complies with IEC 80601-2-49:2018 |
| Respiratory Gas Monitors Performance | Compliance with ISO 80601-2-55:2018 (Basic safety and essential performance of respiratory gas monitors) | Complies with ISO 80601-2-55:2018 |
| Clinical Thermometers Performance | Compliance with ISO 80601-2-56:2017+A1:2018 (Basic safety and essential performance of clinical thermometers) | Complies with ISO 80601-2-56:2017+A1:2018 |
| Pulse Oximeter Equipment Performance | Compliance with ISO 80601-2-61:2017 (Basic safety and essential performance of pulse oximeter equipment) | Complies with ISO 80601-2-61:2017 |
| Wireless Coexistence | Compliance with IEEE ANSI USEMCSC C63.27 (Evaluation of Wireless Coexistence) | Complies with IEEE ANSI USEMCSC C63.27 |
| Software Functionality | Compliance with FDA Guidance "Content of Premarket Submissions for Device Software Functions" | Software verification and validation testing conducted and documentation provided as recommended. |
| Accuracy Specifications (Example: RESP) | 6 rpm to 200 rpm: ±2 rpm | Reported as meeting this accuracy specification. |
| Accuracy Specifications (Example: IBP) | ±2% or ±1 mmHg, whichever is greater (excluding sensor error) | Reported as meeting this accuracy specification. |
2. Sample Size Used for the Test Set and Data Provenance:
- Sample Size: Not applicable in terms of human subjects or patient data test sets, as "new clinical studies" were not required. The "test set" refers to bench testing and functional system-level validation. The specific number of test cycles or a detailed breakdown of test cases for bench testing is not provided in this summary.
- Data Provenance: The data primarily originates from Edan Instruments Inc. (Shenzhen, Guangdong, China) through internal engineering and quality assurance processes for non-clinical bench testing and software validation. It is not patient data, so concepts like "retrospective or prospective" do not apply.
3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications:
- Not applicable for clinical ground truth: Since no clinical studies were performed requiring human interpretation or diagnosis for a test set, no medical experts (e.g., radiologists) were used to establish ground truth in this context.
- Internal experts: Bench testing and software validation would have involved engineers and quality assurance professionals, whose qualifications are implicit in the quality system (21 CFR Part 820) but not specified in detail here.
4. Adjudication Method for the Test Set:
- Not applicable: Adjudication methods (e.g., 2+1, 3+1) are relevant for clinical studies involving multiple readers. This was not a clinical study. Bench testing relies on established technical specifications and standard compliance.
5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was Done:
- No: No MRMC study was performed as no new clinical studies were required or conducted. Therefore, there's no effect size of human readers improving with AI assistance. The device is a patient monitor, not an AI-assisted diagnostic tool.
6. If a Standalone (i.e. algorithm only without human-in-the loop performance) was Done:
- Yes (for the technical components): The "performance testing-Bench" effectively represents a standalone evaluation of the device's functional components (ECG, NIBP, SpO2, etc.) and software against defined technical specifications and standards. The "software verification and validation testing" also represents a standalone evaluation of the algorithm and software functions. The specific algorithms (e.g., iCUFS, iFAST for NIBP, arrhythmia analysis logic) are tested independently for their accuracy against known inputs or reference standards as part of bench testing.
7. The Type of Ground Truth Used:
- Technical/Reference Standards: For the bench testing, the "ground truth" would be derived from:
- Reference standards/simulators: Calibrated medical equipment, physiological simulators, and test signals (e.g., known ECG waveforms, simulated blood pressure readings, temperature standards) are used to provide the "true" values against which the device's measurements are compared.
- Defined specifications: The device's internal design specifications and the requirements of the referenced IEC/ISO standards serve as the "ground truth" for compliance testing.
- Not clinical ground truth: No expert consensus, pathology, or outcomes data from real patients were used for establishing ground truth for this submission.
8. The Sample Size for the Training Set:
- Not applicable: The device is a patient monitor, not a machine learning/AI algorithm that typically undergoes a distinct "training" phase with a large dataset. Its functionality is based on established physiological measurement principles and programmed algorithms. Any internal calibration or algorithm refinement would be part of the product development process, not a dedicated "training set" in the AI/ML sense.
9. How the Ground Truth for the Training Set Was Established:
- Not applicable: As there was no "training set" in the context of an AI/ML model, the concept of establishing ground truth for it does not apply to this 510(k) submission.
In summary, this 510(k) clearance relies on demonstrating that the new Patient Monitor is substantially equivalent to a previously cleared predicate device, primarily through robust non-clinical bench testing and software validation, proving compliance with established medical device standards and functional specifications. No new clinical studies with patient data were required or conducted for this specific submission.
Ask a specific question about this device
(226 days)
CFR 870.1130
Impedance plethysmograph 21 CFR 870.2770
Thermometer, Electronic, Clinical 21 CFR 880.2910
-
HemoSphere Advanced Monitor with HemoSphere Swan-Ganz Module: The HemoSphere advanced monitor when used with the HemoSphere Swan-Ganz module and Edwards Swan-Ganz catheters is indicated for use in adult and pediatric critical care patients requiring monitoring of cardiac output (continuous [CO] and intermittent [iCO]) and derived hemodynamic parameters in a hospital environment. Pulmonary artery blood temperature monitoring is used to compute continuous and intermittent CO with thermodilution technologies. It may also be used for monitoring hemodynamic parameters in conjunction with a perioperative goal directed therapy protocol in a hospital environment. Refer to the Edwards Swan-Ganz catheter and Swan-Ganz Jr catheter indications for use statements for information on target patient population specific to the catheter being used. Refer to the Intended Use statement for a complete list of measured and derived parameters available for each patient population.
-
HemoSphere Advanced Monitor with HemoSphere Oximetry Cable: The HemoSphere Advanced Monitor when used with the HemoSphere Oximetry Cable and Edwards oximetry catheters is indicated for use in adult and pediatric critical care patients requiring monitoring of venous oxygen saturation (SvO2 and ScvO2) and derived hemodynamic parameters in a hospital environment. Refer to the Edwards oximetry catheter indications for use statement for information on target patient population specific to the catheter being used. Refer to the Intended Use statement for a complete list of measured and derived parameters available for each patient population.
-
HemoSphere Advanced Monitor with HemoSphere Pressure Cable: The HemoSphere advanced monitor when used with the HemoSphere pressure cable is indicated for use in adult and pediatric critical care patients in which the balance between cardiac function, fluid status, vascular resistance and pressure needs continuous assessment. It may be used for monitoring of hemodynamic parameters in conjunction with a perioperative goal directed therapy protocol in a hospital environment. Refer to the Edwards FloTrac sensor, FloTrac Jr sensor, Acumen IQ sensor, and TruWave disposable pressure transducer indications for use statements for information on target patient populations specific to the sensor/transducer being used. The Edwards Acumen Hypotension Prediction Index software feature provides the clinician with physiological insight into a patient's likelihood of future hypotensive events and the associated hemodynamics. The Acumen HPI feature is intended for use in surgical or non-surgical patients receiving advanced hemodynamic monitoring. The Acumen HPI feature is considered to be additional quantitative information regarding the patient's physiological condition for reference only and no therapeutic decisions should be made based solely on the Acumen Hypotension Prediction Index (HPI) parameter. Refer to the Intended Use statement for a complete list of measured and derived parameters available for each patient population.
-
HemoSphere Advanced Monitor with Acumen Assisted Fluid Management Feature and Acumen IQ Sensor: The Acumen Assisted Fluid Management (AFM) software feature provides the clinician with physiological insight into a patient's estimated response to fluid therapy and the associated hemodynamics. The Acumen AFM software feature is intended for use in surgical patients >=18 years of age, that require advanced hemodynamic monitoring. The Acumen AFM software feature offers suggestions regarding the patient's physiological condition and estimated response to fluid therapy. Acumen AFM fluid administration suggestions are offered to the clinician; the decision to administer a fluid bolus is made by the clinician, based upon review of the patient's hemodynamics. No therapeutic decisions should be made based solely on the Assisted Fluid Management suggestions. The Acumen Assisted Fluid Management software feature may be used with the Acumen AFM Cable and Acumen IQ fluid meter.
-
HemoSphere Advanced Monitor with HemoSphere Technology Module and ForeSight Oximeter Cable: The non-invasive ForeSight oximeter cable is intended for use as an adjunct monitor of absolute regional hemoglobin oxygen saturation of blood under the sensors in individuals at risk for reduced-flow or no flow ischemic states. The ForeSight Oximeter Cable is also intended to monitor relative changes of total hemoglobin of blood under the sensors. The ForeSight Oximeter Cable is intended to allow for the display of StO2 and relative change in total hemoglobin on the HemoSphere advanced monitor.
- When used with large sensors, the ForeSight Oximeter Cable is indicated for use on adults and transitional adolescents >=40 kg.
- When used with medium sensors, the ForeSight Oximeter Cable is indicated for use on pediatric subjects >=3 kg.
- When used with small sensors, the ForeSight Oximeter Cable is indicated for cerebral use on pediatric subjects <8 kg and non-cerebral use on pediatric subjects <5kg.
Refer to the Intended Use statement for a complete list of measured and derived parameters available for each patient population.
-
HemoSphere Advanced Monitor with HemoSphere ClearSight Module: The HemoSphere advanced monitor when used with the HemoSphere ClearSight module, pressure controller or Smart Pressure Controller and a compatible Edwards finger cuff are indicated for patients over 18 years of age in which the balance between cardiac function, fluid status and vascular resistance needs continuous assessment. It may be used for monitoring hemodynamic parameters in conjunction with a perioperative goal directed therapy protocol in a hospital environment. In addition, the noninvasive system is indicated for use in patients with comorbidities for which hemodynamic optimization is desired and invasive measurements are difficult. The HemoSphere advanced monitor and compatible Edwards finger cuffs noninvasively measures blood pressure and associated hemodynamic parameters. The Edwards Lifesciences Acumen Hypotension Prediction Index feature provides the clinician with physiological insight into a patient's likelihood of future hypotensive events and the associated hemodynamics. The Acumen HPI feature is intended for use in surgical or non-surgical patients receiving advanced hemodynamic monitoring. The Acumen HPI feature is considered to be additional quantitative information regarding the patient's physiological condition for reference only and no therapeutic decisions should be made based solely on the Hypotension Prediction Index (HPI) parameter.
-
Indication for Acumen IQ Plus and VitaWave Plus finger cuffs: The Acumen IQ Plus and VitaWave Plus finger cuff adult indicated for patients over 18 years of age to continuously blood pressure and associated hemodynamic parameters when used with a compatible Edwards monitoring platform.
-
Smart Pressure Controller: The Smart Pressure Controller is intended for use with an Edwards compatible noninvasive monitoring system - composed of compatible monitor, pressure source (pump), compatible Edwards finger cuff(s) and pressure controller - for continuous noninvasive measurement of blood pressure and associated hemodynamic parameters. Refer to the operator's manual of the compatible Edwards monitor being used for specific information on the intended use environment and patient population.
-
Intended Use: The HemoSphere advanced monitoring platform is intended to be used by qualified personnel or trained clinicians in a critical care environment in a hospital setting. The Viewfinder remote mobile application can be used for supplemental near real-time remote display of monitored hemodynamic parameter data as well as faults, alerts and notifications generated by the HemoSphere advanced monitoring platform. The HemoSphere advanced monitoring platform is intended for use with compatible Edwards Swan-Ganz and oximetry catheters, Swan-Ganz Jr catheters, FloTrac sensors, FloTrac Jr sensors, Acumen IQ sensors, TruWave disposable pressure transducers, ForeSight/ForeSight Jr sensors, Acumen IQ fluid meter, and ClearSight/ClearSight Jr/Acumen IQ/Acumen IQ Plus/VitaWave/VitaWave Plus finger cuffs
The HemoSphere Advanced Monitor was designed to simplify the customer experience by providing one platform with modular solutions for all hemodynamic monitoring needs. The user can choose from available optional sub-system modules or use multiple sub-system modules at the same time. This modular approach provides the customer with the choice of purchasing and/or using specific monitoring applications based on their needs. Users are not required to have all of the modules installed at the same time for the platform to function.
The provided FDA 510(k) clearance letter and summary for the Edwards Lifesciences HemoSphere Advanced Monitor (HEM1) and associated components outlines the device's indications for use and the testing performed to demonstrate substantial equivalence to predicate devices. However, it does not contain the detailed acceptance criteria or the specific study results (performance data) in the format typically required to answer your request fully, especially for acceptance criteria and performance of an AI/algorithm-based feature like the Hypotension Prediction Index (HPI) or Assisted Fluid Management (AFM).
The document states:
- "Completion of all verification and validation activities demonstrated that the subject devices meet their predetermined design and performance specifications."
- "Measured and derived parameters were tested using a bench simulation. Additionally, system integration and mechanical testing was successfully conducted to verify the safety and effectiveness of the device. All tests passed."
- "Software verification testing was conducted, and documentation was provided per FDA's Guidance for Industry and FDA Staff, "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices". All tests passed."
This indicates that internal performance specifications were met, but the specific metrics, thresholds, and study designs for achieving those specifications are not detailed in this public summary.
Therefore, I cannot populate the table with specific numerical performance data against acceptance criteria for the HPI or AFM features, nor can I provide details on sample size, expert ground truth establishment, or MRMC studies, as this information is not present in the provided text.
The text primarily focuses on:
- Substantial equivalence to predicate devices.
- Indications for Use for various HemoSphere configurations and modules.
- Description of software and hardware modifications (e.g., integration of HPI algorithm, new finger cuffs).
- General categories of testing performed (Usability, System Verification, Electrical Safety/EMC, Software Verification) with a blanket statement that "All tests passed."
Based on the provided document, here's what can and cannot be stated:
1. A table of acceptance criteria and the reported device performance
Cannot be provided with specific numerical data or thresholds from the given text. The document only states that "all verification and validation activities demonstrated that the subject devices meet their predetermined design and performance specifications." No specific acceptance criteria values (e.g., "Accuracy > X%", "Sensitivity > Y%", "Mean Absolute Error < Z") or reported performance values are publicly disclosed in this summary for any parameter, including HPI or AFM. For measured and derived parameters (like CO, MAP, etc.), it states they were tested using bench simulation, and "All tests passed," implying they met internal accuracy specifications for physical measurements, but these are not detailed.
2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)
Cannot be provided from the given text. The document mentions "bench simulation" for measured and derived parameters, but does not provide sample sizes for these, or the type/provenance of data for testing the HPI or AFM algorithms.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)
Cannot be provided from the given text. The document doesn't describe the process of establishing ground truth for the algorithms, nor does it mention the number or qualifications of experts involved in such a process.
4. Adjudication method (e.g. 2+1, 3+1, none) for the test set
Cannot be provided from the given text. There is no mention of adjudication methods for any test sets.
5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
Cannot be provided from the given text. The document does not describe any MRMC studies or human-in-the-loop performance evaluation regarding the HPI or AFM features. The HPI and AFM features are described as providing "physiological insight" and "suggestions," not as tools requiring reader interpretation in a comparative effectiveness study as typically seen with imaging AI.
6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done
Likely yes, based on the nature of the algorithms, but no specific performance metrics are provided. The HPI and AFM features are stated to provide "quantitative information" and "suggestions." The text indicates "System Verification (Non-Clinical Performance)" and "Software Verification" were performed, suggesting standalone evaluation against internal specifications, but no detailed results are provided. The HPI algorithm itself was "previously cleared in K230057," implying its standalone performance would have been evaluated during that prior clearance, but those details are not in this document.
7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)
Cannot be definitively stated from the given text. For the HPI feature, which predicts future hypotensive events, ground truth would typically involve actual patient outcomes (e.g., observed hypotensive events). For AFM, which suggests response to fluid therapy, ground truth might involve observed physiological responses to fluid boluses. However, the document does not specify how these ground truths were established for the purpose of testing the algorithms.
8. The sample size for the training set
Cannot be provided from the given text. The document does not mention details about the training data for the algorithms.
9. How the ground truth for the training set was established
Cannot be provided from the given text. The document does not mention details about the training data or its ground truth establishment.
Summary of Device Features Mentioned in Relation to Performance/Testing (General):
- HemoSphere Advanced Monitor and various modules/accessories: The document primarily describes this as a monitoring platform for various hemodynamic parameters (CO, SvO2, MAP, etc.). Performance for these measured and derived parameters was tested via "bench simulation," and "All tests passed," implying they met internal benchmarks for accuracy and reliability.
- Acumen Hypotension Prediction Index (HPI) software feature: This feature provides "physiological insight into a patient's likelihood of future hypotensive events." It was integrated from a previously cleared device (K230057). The document states "There are no changes to the Acumen HPI algorithm from what was cleared in K230057." This implies that the acceptance criteria and supporting studies for the HPI algorithm itself would be found in the K230057 clearance documentation, not typically resubmitted in detail for integration into another platform unless the integration process significantly altered its functionality or intended use.
- Acumen Assisted Fluid Management (AFM) software feature: This feature provides "physiological insight into a patient's estimated response to fluid therapy" and "suggestions." It also mentions "Acumen AFM fluid administration suggestions are offered to the clinician; the decision to administer a fluid bolus is made by the clinician, based upon review of the patient's hemodynamics. No therapeutic decisions should be made based solely on the Assisted Fluid Management suggestions." This language suggests it's a supportive, advisory tool, rather than a diagnostic one requiring strict accuracy metrics in the same way. No performance specifics for AFM are given.
- Usability Study: Conducted to ensure primary operating functions and critical tasks can be performed without patient or user harm. Determined that "intended users can perform primary operating functions and critical tasks of the system without any usability issues that may lead to patient or user harm." This is an acceptance criterion for human factors, but not for algorithmic performance.
- Electrical Safety and EMC, Software Verification: All tests passed. These are general product safety and quality criteria, not specific to the performance of the predictive algorithms.
To obtain the detailed performance data, acceptance criteria, sample sizes, and ground truth information for the HPI or AFM algorithms, one would typically need to refer to the original 510(k) submission for the HPI algorithm (K230057) and potentially separate documentation for the AFM feature, which are not included in this general clearance letter for the HemoSphere platform update.
Ask a specific question about this device
(315 days)
. §880.2910 Clinical electronic thermometer.
21 C.F.R. §882.1400 Electroencephalograph
21 C.F.R.
The monitor B105M, B125M, B155M, B105P and B125P are portable multi-parameter patient monitors intended to be used for monitoring, recording, and to generate alarms for multiple physiological parameters of adult, pediatric, and neonatal patients in a hospital environment and during intra-hospital transport.
The monitor B105M, B125M, B155M, B105P and B125P are intended for use under the direct supervision of a licensed health care practitioner.
The monitor B105M, B125M, B155M, B105P and B125P are not Apnea monitors (i.e., do not rely on the device for detection or alarm for the cessation of breathing). These devices should not be used for life sustaining/supporting purposes.
The monitor B105M, B125M, B155M, B105P and B125P are not intended for use during MRI.
The monitor B105M, B125M, B155M, B105P and B125P can be stand-alone monitors or interfaced to other devices via network.
The monitor B105M, B125M, B155M, B105P and B125P monitor and display: ECG (including ST segment, arrhythmia detection, ECG diagnostic analysis and measurement), invasive blood pressure, heart/pulse rate, oscillometric non-invasive blood pressure (systolic, diastolic and mean arterial pressure), functional oxygen saturation (SpO2) and pulse rate via continuous monitoring (including monitoring during conditions of clinical patient motion or low perfusion), temperature with a reusable or disposable electronic thermometer for continual monitoring Esophageal/Nasopharyngeal/Tympanic/Rectal/Bladder/Axillary/Skin/Airway/Room/Myocardial/Core/Surface temperature, impedance respiration, respiration rate, airway gases (CO2, O2, N2O, anesthetic agents, anesthetic agent identification and respiratory rate), Cardiac Output (C.O.), Entropy, neuromuscular transmission (NMT) and Bispectral Index (BIS).
The monitor B105M, B125M, B155M, B105P and B125P are able to detect and generate alarms for ECG arrhythmias: Asystole, Ventricular tachycardia, VT>2, Ventricular Bradycardia, Accelerated Ventricular Rhythm, Ventricular Couplet, Bigeminy, Trigeminy, "R on T", Tachycardia, Bradycardia, Pause, Atrial Fibrillation, Irregular, Multifocal PVCs, Missing Beat, SV Tachy, Premature Ventricular Contraction (PVC), Supra Ventricular Contraction (SVC) and Ventricular fibrillation.
The proposed monitors B105M, B125M, B155M, B105P and B125P are new version of multi-parameter patient monitors developed based on the predicate monitors B105M, B125M, B155M, B105P and B125P (K213490) to provide additional monitored parameter Bispectral Index (BIS) by supporting the additional optional E-BIS module (K052145) which used in conjunction with Covidien BISx module (K072286).
In addition to the added parameter, the proposed monitors also offer below several enhancements:
- Provided data connection with GE HealthCare anesthesia devices to display the parameters measured from anesthesia devices (Applicable for B105M, B125M and B155M).
- Modified Early Warning Score calculation provided.
- Separated low priority alarms user configurable settings from the combined High/Medium/Low priority options.
- Provided additional customized notification tool to allow clinician to configure the specific notification condition of one or more physiological parameters measured by the monitor. (Applicable for B105M, B125M and B155M).
- Enhanced User Interface in Neuromuscular Transmission (NMT), Respiration Rate and alarm overview.
- Provided Venous Stasis to assist venous catheterization with NIBP cuff inflation.
- Supported alarm light brightness adjustment.
- Supported alarm audio pause by gesture (Not applicable for B105M and B105P).
- Supported automatic screen brightness adjustment.
- Supported network laser printing.
- Continuous improvements in cybersecurity
The proposed monitors B105M, B125M, B155M, B105P and B125P retain equivalent hardware design based on the predicate monitors and removal of the device Trim-knob to better support cleaning and disinfecting while maintaining the same primary function and operation.
Same as the predicate device, the five models (B105M, B125M, B155M, B105P and B125P) share the same hardware platform and software platform to support the data acquisition and algorithm modules. The differences between them are the LCD screen size and configuration options. There is no change from the predicate in the display size.
As with the predicate monitors B105M, B125M, B155M, B105P and B125P (K213490), the proposed monitors B105M, B125M, B155M, B105P and B125P are multi-parameter patient monitors, utilizing an LCD display and pre-configuration basic parameters: ECG, RESP, NIBP, IBP, TEMP, SpO2, and optional parameters which include CO2 and Gas parameters provided by the E-MiniC module (K052582), CARESCAPE Respiratory modules E-sCO and E-sCAiO (K171028), Airway Gas Option module N-CAiO (K151063), Entropy parameter provided by the E-Entropy module (K150298), Cardiac Output parameter provided by the E-COP module (K052976), Neuromuscular Transmission (NMT) parameter provided by E-NMT module (K051635) and thermal recorder B1X5-REC.
The proposed monitors B105M, B125M, B155M, B105P and B125P are not Apnea monitors (i.e., do not rely on the device for detection or alarm for the cessation of breathing). These devices should not be used for life sustaining/supporting purposes. Do not attempt to use these devices to detect sleep apnea.
As with the predicate monitors B105M, B125M, B155M, B105P and B125P (K213490), the proposed monitors B105M, B125M, B155M, B105P and B125P also can interface with a variety of existing central station systems via a cabled or wireless network which implemented with identical integrated WiFi module. (WiFi feature is disabled in B125P/B105P).
Moreover, same as the predicate monitors B105M, B125M, B155M, B105P and B125P (K213490), the proposed monitors B105M, B125M, B155M, B105P and B125P include features and subsystems that are optional or configurable, and it can be mounted in a variety of ways (e.g., shelf, countertop, table, wall, pole, or head/foot board) using existing mounting accessories.
The provided FDA 510(k) clearance letter and summary for K242562 (Monitor B105M, Monitor B125M, Monitor B155M, Monitor B105P, Monitor B125P) do not contain information about specific acceptance criteria, reported device performance metrics, or details of a study meeting those criteria for any of the listed physiological parameters or functionalities (e.g., ECG or arrhythmia detection).
Instead, the documentation primarily focuses on demonstrating substantial equivalence to a predicate device (K213490) by comparing features, technology, and compliance with various recognized standards and guidance documents for safety, EMC, software, human factors, and cybersecurity.
The summary explicitly states: "The subject of this premarket submission, the proposed monitors B105M/B125M/B155M/B105P/B125P did not require clinical studies to support substantial equivalence." This implies that the changes introduced in the new device versions were not considered significant enough to warrant new clinical performance studies or specific quantitative efficacy/accuracy acceptance criteria beyond what is covered by the referenced consensus standards.
Therefore, I cannot provide the requested information from the given text:
- A table of acceptance criteria and the reported device performance: This information is not present. The document lists numerous standards and tests performed, but not specific performance metrics or acceptance thresholds.
- Sample size used for the test set and the data provenance: Not explicitly stated for performance evaluation, as clinical studies were not required. The usability testing mentioned a sample size of 16 US clinical users, but this is for human factors, not device performance.
- Number of experts used to establish the ground truth for the test set and the qualifications of those experts: Not applicable, as detailed performance studies requiring expert ground truth are not described.
- Adjudication method (e.g. 2+1, 3+1, none) for the test set: Not applicable.
- If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance: Not applicable. This device is a patient monitor, not an AI-assisted diagnostic tool that would typically involve human readers.
- If a standalone (i.e. algorithm only without human-in-the loop performance) was done: The document describes "Bench testing related to software, hardware and performance including applicable consensus standards," which implies standalone testing against known specifications or simulated data. However, specific results or detailed methodologies for this type of testing are not provided beyond the list of standards.
- The type of ground truth used (expert consensus, pathology, outcomes data, etc.): Not explicitly stated for performance assessment. For the various parameters (ECG, NIBP, SpO2, etc.), it would typically involve reference equipment or validated methods as per the relevant IEC/ISO standards mentioned.
- The sample size for the training set: Not applicable, as this is not an AI/ML device that would require explicit training data in the context of this submission.
- How the ground truth for the training set was established: Not applicable.
In summary, the provided document focuses on demonstrating that the new monitors are substantially equivalent to their predicate through feature comparison, adherence to recognized standards, and various non-clinical bench tests (e.g., hardware, alarms, EMC, environmental, reprocessing, human factors, software, cybersecurity). It does not contain the detailed performance study results and acceptance criteria typically found for novel diagnostic algorithms or AI-driven devices.
Ask a specific question about this device
(125 days)
(EAR-E101); Ear Thermometer (EAR-E102); Ear Thermometer (EAR-E103)
Regulation Number: 21 CFR 880.2910
(EAR-E101); Ear Thermometer (EAR-E102); Ear Thermometer (EAR-E103)
Regulation Number: 21 CFR 880.2910
Thermometer |
| Model | models EAR-E101, EAR-E102, EAR-E103 |
| Regulation Number | 21 CFR 880.2910
Equipment & Supply Co., Ltd | -- |
| Product Code | FLL | FLL | Same |
| Regulation No. | 21 CFR 880.2910
| 21 CFR 880.2910 | Same |
| Classification | II | II | Same |
| **Indications for Use / Intended
The Ear Thermometer is intended to measure human body temperature of people over three months from surface of eardrum. It applies to both professional use and home use.
The ear thermometer is a handheld device that displays the temperature of the measured patient by measuring the thermal radiation of the eardrum. Measurement unit: °C or °F. The results can be displayed on LCD. The thermometers are powered by 1.5V×2 (AAA or AA) alkaline batteries, which can be used for people over three months. A thermopile sensor is employed to detect or monitor the infrared thermal energy emitted from the eardrum, which is converted into temperature measurement with the unit of °C or °F. All the models share the same critical components, intended use, working principle and similar product design, and compose of a sensor, PCB, buttons, LCD display and housing. Functions include temperature measurement, memory reading recall, unit switch, low battery detection and high temperature indicator.
The FDA 510(k) clearance letter and accompanying 510(k) Summary for the Ear Thermometers (EAR-E101, EAR-E102, EAR-E103) provide information on acceptance criteria and supporting studies.
1. Table of Acceptance Criteria and Reported Device Performance:
The document primarily focuses on demonstrating substantial equivalence to a predicate device (Jiangsu Yuyue Medical Equipment & Supply Co., Ltd, Infrared Ear Thermometers YHT101 and YHT200, K203583). Thus, the "acceptance criteria" are largely framed as demonstrating equivalence or adherence to relevant standards. The performance data is presented in comparison to the predicate and overall compliance with standards.
| Acceptance Criteria (Stated or Implied by Comparison) | Reported Device Performance (Subject Device) |
|---|---|
| Accuracy (Clinical) | Measurement Range: 32.0℃~42.9℃ (89.6°F-109.2°F) Accuracy: ±0.2℃/±0.4°F (for 32.0℃ to 42.9℃) |
| Biocompatibility | Passed Cytotoxicity, Sensitization, and Irritation tests per ISO 10993-1, ISO 10993-5, ISO 10993-10, ISO 10993-23 |
| Electrical Safety | Compliant with IEC 60601-1, IEC 60601-1-11, ISO 80601-2-56 |
| Electromagnetic Compatibility (EMC) | Compliant with IEC 60601-1-2 |
| Operating Environments | Temperature: 15°C~ 40°C Humidity: ≤ 85% RH, non-condensing Atmospheric pressure: 70–106 kPa |
| Transport and Storage Environments | Ambient Temperature: -20°C to 55°C Relative Humidity: ≤ 93% RH, non-condensing Atmospheric pressure: 50 kPa to 106 kPa |
| Regulatory Compliance | Compliant with 21 CFR 880.2910 (Clinical Electronic Thermometer) |
| Indications for Use (Patient Population) | Measures human body temperature of people over three months from surface of eardrum. (Narrower than predicate, excluding preterm and newborns) |
2. Sample size used for the test set and the data provenance:
- Non-clinical Data (Bench Testing): The document does not specify a "sample size" in terms of number of devices for bench testing. It states that "The device has been tested according to the following standards," implying tests were performed on representative samples to ensure compliance.
- Clinical Data: The summary states, "The clinical testing has been conducted per ISO 80601-2-56 Medical electrical equipment—Part 2-56: Particular requirements for basic safety and essential performance of clinical thermometers for body temperature measurement." However, the specific sample size (number of patients/measurements) and data provenance (e.g., country of origin, retrospective/prospective) for this clinical study are NOT provided in the given document.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:
- Clinical Data: The document does not specify the number of experts or their qualifications for establishing ground truth in the clinical study. As it's an ear thermometer, the "ground truth" would typically come from a reference temperature measurement method (e.g., rectal thermometry) rather than expert interpretation of an image or signal.
4. Adjudication method for the test set:
- Clinical Data: Given that the device measures a quantitative value (temperature), an adjudication method in the context of expert consensus (like multiple readers for an image) is generally not applicable. The comparison would be between the device's reading and the reference method's reading. The document does not describe any specific adjudication method for the clinical test set.
5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:
- No, an MRMC comparative effectiveness study was NOT done. This type of study (MRMC) is relevant for diagnostic imaging devices where human interpretation plays a significant role, sometimes aided by AI. This document is for an ear thermometer, a direct measurement device, not an AI-assisted diagnostic tool.
6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:
- Yes, a standalone performance was done for the ear thermometer. The device itself is designed to provide temperature readings directly, without requiring human interpretation or input to calculate the core measurement. The bench and clinical testing described aim to quantify this standalone performance in terms of accuracy and adherence to standards.
7. The type of ground truth used:
- Clinical Data: While not explicitly stated, for a clinical thermometer, the ground truth for performance testing is typically established by comparison against a standardized, highly accurate reference thermometer (e.g., a rectal thermometer or an oral thermometer with known accuracy) or a specialized blackbody calibrator in a controlled environment as specified by standards like ISO 80601-2-56. The document mentions "clinical accuracy test" and "performance test," strongly implying such a comparative methodology.
8. The sample size for the training set:
- Not applicable / Not provided. The device is an ear thermometer, which is a sensor-based measurement tool, not an AI/machine learning algorithm that requires a "training set" in the conventional sense. The device's performance is determined by its design, calibration, and adherence to physical principles and engineering specifications, which are validated through bench and clinical testing.
9. How the ground truth for the training set was established:
- Not applicable. As noted above, the concept of a "training set" and its associated ground truth establishment is not relevant for this type of medical device submission.
Ask a specific question about this device
Page 1 of 77