Search Filters

Search Results

Found 32 results

510(k) Data Aggregation

    K Number
    K240593
    Manufacturer
    Date Cleared
    2024-08-23

    (175 days)

    Product Code
    Regulation Number
    882.1400
    Reference & Predicate Devices
    Why did this record match?
    Product Code :

    OLT

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The VEEGix EEG System, is intended to be used for measuring and recording the electrical activity of a patient's brain, obtained by placing electrodes on the forehead and wirelessly transmitting the electroencephalographic (EEG) signals for storage and display.

    The VEEGix EEG System, is intended for use in the acquisition of EEG signals, displaying them in real time and storing them for later review and analysis.

    The VEEGix EEG System, is intended for use in a hospital Operating Room. Post Anesthesia Care Unit, Intensive Care Unit, Emergency Department, or in other medical facilities such as inpatient and outpatient (ambulatory) surgery settings.

    The VEEGix EEG System is indicated for use on patients 18 years of age or older and is to be used by licensed medical professionals who have been adequately trained in the use and interpretation of EEG data for determination of brain state.

    The VEEGix EEG System does not provide any diagnostic conclusion about the patient's condition.

    The VEEGix EEG System is not to be used as a stand-alone in the evaluation or diagnosis of a disease or other condition.

    The VEEGix EEG System is not intended for use in life support systems.

    Device Description

    The VEEGix™ EEG system (subject device) is a Electrocochleographic (EEG) device to allow healthcare practitioners, nurses, neurologists and qualified technicians working in hospital Operating Room, Post Anesthesia Care Unit, Intensive Care Unit, Emergency Department, or in other medical facilities such as inpatient and outpatient (ambulatory) surgery settings to observe a patient's EEG signals and its derivatives.

    The use of the subject device is a straight-forward procedure of applying the VEEGix Electrode, which is a lightweight band, onto the patient's scalp in which the electrodes acquire the patient's EEG signals. Those signals are transmitted via the VEEGix EEG module) to the VEEGix iOS app (application) which display the patient's EEG measurements in real-time to allow healthcare practitioners, nurses, neurologists and qualified technicians to monitor and record those EEG measurements.

    AI/ML Overview

    Here's a breakdown of the acceptance criteria and study details for the VEEGix EEG System, based on the provided FDA 510(k) summary:

    1. Table of Acceptance Criteria and Reported Device Performance

    The document doesn't explicitly list a table of quantitative acceptance criteria with corresponding performance values for the software component of the device. Instead, the "Technology Comparison" table highlights specific technical characteristics where the VEEGix EEG System aligns with or exceeds its predicates, implying these are the implicit acceptance criteria for technical equivalence. The clinical performance section describes the methodology used to demonstrate comparable EEG signal quality to a conventional clinical EEG system, which serves as the overarching performance goal.

    Implicit Acceptance Criteria and Reported Device Performance (based on "Technology Comparison" and "Clinical Performance Testing" sections):

    Acceptance Criteria (Implied)Reported Device Performance (VEEGix EEG System)Reference/Context
    Technical Equivalence
    Classification RegulationClass II per 882.1400Same as primary predicate
    Product Code(s)OLT, OMC, ORT, GXYSame as primary predicate
    Indications for UseSubset of primary, same as secondary predicateNo new safety/effectiveness questions
    ModalitiesEEGSame as predicates
    Environment of UseHospital OR, PACU, ICU, ED, inpatient/outpatient surgery settingsSame as primary predicate, included in secondary
    Number of EEG Channels1 bipolarSame as secondary predicate, substantially equivalent to primary
    Number of Electrodes3 (Fp1, Fpz, Fp2)Same as secondary predicate, substantially equivalent to primary
    Sensing ElectrodesSilver, disposableSubstantially equivalent
    Power SourceBatterySame as predicates
    System ComponentsElectrode Array, Sensor module, acquisition, Tablet for memory and data viewingSame as primary predicate, similar to secondary
    Screen Display DetailsRaw EEG Waveforms; Signal Quality (Channel Connection, Artifact, Noise, Interference); Spectral Parameters (EEG power spectrum, SEF 95%, DSA, α, Β, δ, θ, γ)Substantially equivalent to primary (missing 10/20 Hz markers, B band division not raising new questions), secondary display capabilities included
    Storage for offline recordingYes, in the tablet displaySame as primary predicate
    Electrode ImpedanceYesSame as primary predicate
    Detection for Leads OffYesSame as predicates
    File output capabilityYesSame as predicates
    Real-time EEG DisplayYes, wireless to tabletSame as predicates
    Processed EEG Bandwidth0.5 Hz to 80 HzSubstantially equivalent as primary predicate (0.5 Hz to 45 Hz)
    Automatic Artifact IdentificationYesSame as primary predicate
    Common Mode Rejection> 100 dBSubstantially equivalent as primary predicate (> 90 dB)
    Amplifier Input Impedance200 GΩSubstantially equivalent as primary predicate (> 500 M Ohm)
    Electrode Impedance TestYesSame as primary predicate
    Patient Contains IsolationReinforced Insulation (2 MOPP), battery poweredSubstantially equivalent as primary predicate
    Event MarkersYes, artifact detectionSame as primary predicate
    Burst Suppression DisplayYes, Suppression Ratio (%)Substantially equivalent as primary predicate (Burst Suppression Probability)
    Display InterfaceTablet display indicates connection and operationSame as primary predicate
    BiocompatibilityTested per ISO 10993 (cytotoxicity, irritation, sensitization)Substantially equivalent
    Clinical Performance
    Comparable EEG signal quality to conventional clinical EEG systemsPearson's correlation coefficient (r=0.84, p
    Ask a Question

    Ask a specific question about this device

    K Number
    K220254
    Manufacturer
    Date Cleared
    2022-11-30

    (303 days)

    Product Code
    Regulation Number
    882.1400
    Reference & Predicate Devices
    Why did this record match?
    Product Code :

    OLT

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Neuron-Spectrum-AM system with Neuron-Spectrum.NET software is intended for use as a digital neurophysiological system for recording, processing, and displaying biopotential signals such as Electroencephalography (EEG) and Polysomnography (PSG) derived from Electroencephalography (EEG) by the means of a dedicated software module and dedicated electrodes.

    The device is portable and can register up to 21 EEG channels, 4 polygraphic channels, and 1 direct current channel.

    The device does not provide alarms, does not provide automated event marking and does not provide to the user any diagnostic conclusion about the patient's condition. They are in patient care institutions, diagnostics centers, neurosurgical hospitals, experimental laboratories. The device can also be used as a home use device under supervision of qualified personnel. The patient group includes all ages and sexes.

    Device Description

    The Neuron-Spectrum-AM with Neuron-Spectrum.NET Software (Subject device) is an ambulatory wireless digital neurophysiological device capable of recording, processing, and displaying electroencephalography (EEG), video EEG, long-term monitoring (LTM), and polysomnography (PSG) biopotential signals.

    As listed in Table 1, The NS-AM device comprises an Electronic Unit, a Battery Adaptor, an External Rechargeable Battery Bank, a Carrying Pouch and the Neuron-Spectrum.NET desktop application software. The Neuron-Spectrum.NET Software application was cleared in the Predicate's 510(k) submission (K133995).

    The Subject device supports up to 21 EEG channels, 4 wide-band polygraphic channels, 1 breath channel, and 1 direct current channel for a total of 27 analog input channels.

    The Electronic Unit acquires, records and transmits biopotential signals such electroencephalography (EEG) and polysomnography (PSG). The built-in 2.4GHz Wi-Fi radio allows real-time transfer of the collected biopotential signals to the computer running the Neuron-Spectrum.NET software. The biopotential signals are also recorded onto a removable SD memory card as back up for later post analysis.

    The Electronic Unit includes a front panel ON/OFF power button, a light sensor and a User Interface (UI). The UI consist of a liquid crystal display and three menu navigation buttons. Figure 4 shows the UI main functions. The Electronic Unit can be powered by replaceable internal AA batteries or by the External Rechargeable Battery Bank.

    The Electronic Unit and the External Rechargeable Battery Bank reside side-by-side inside the Carrying Pouch. The pouch is strapped and worn over the patient clothing while in use.

    The Neuron-Spectrum.NET software is a computer application that receives, records, processes, and displays the biopotential signals collected by the electronic unit on the PC display. The main operations provided by the Neuron-Spectrum.NET software are:

    • EEG Acquisition
    • · EEG Review, Editing, Storing, Exporting.
    • EEG Analysis
    • Creation of Exam Reports
    • · Program Setup

    The electronic unit measures (141 H x 96 W x 36 D) mm and weighs about 270 grams. The External Rechargeable Battery Bank unit measures (165 H x 106 W x 25 D) mm and weighs about 570 grams. The combined weight for the electronic unit, the battery adaptor and the External Rechargeable Battery Bank is approximately 880 grams.

    AI/ML Overview

    The provided FDA 510(k) summary (K220254) for the Neurosoft Ltd. Neuron-Spectrum-AM with Neuron-Spectrum.NET Software describes the device and its claimed substantial equivalence to a predicate device (K133995). Given the information, here's a breakdown of the requested details:

    1. Table of Acceptance Criteria and Reported Device Performance

    This 510(k) summary does not explicitly state specific, quantifiable acceptance criteria (e.g., "sensitivity must be > X%") for the device's diagnostic performance, nor does it report specific performance metrics like sensitivity, specificity, accuracy, etc., for the device's clinical efficacy.

    Instead, the acceptance criteria are implicitly met through demonstrating substantial equivalence to a predicate device by showing that it "is as safe, as effective, and performs as well as or better than the Neuron-Spectrum-4/P with Neuron-Spectrum.NET Predicate device".

    The "reported device performance" is focused on passing various non-clinical engineering and software tests to demonstrate this substantial equivalence.

    Acceptance Criteria Category (Implicit)Reported Device Performance (as demonstrated by non-clinical testing)
    Device Safety & Effectiveness (Overall)"The documentation and test results provided... demonstrate that the Neuron-Spectrum-AM with Neuron-Spectrum.NET Software device is as safe, as effective, and performs as well as or better than the Neuron-Spectrum-4/P with Neuron-Spectrum.NET Predicate device."
    BiocompatibilityAll patient-contacting materials (Housing, External Rechargeable Battery Bank, Carrying Pouch) were found to be Biocompatible.
    Electrical SafetyPassed and complies with applicable standards (e.g., IEC 60601-1, IEC 60601-2-26, IEC 60601-2-40, IEC 60601-1-11).
    Electromagnetic Compatibility (EMC)Passed and complies with applicable standards (e.g., IEC 60601-1-2).
    Wireless Coexistence & RFID ImmunityPassed these tests, ensuring proper functioning in environments with other wireless devices.
    Bench Testing (Signal acquisition quality compared to predicate)The summary states that "bench testing concluded that the Subject device meets and complies with the ... bench testing requirements." While specific metrics are not explicitly stated as acceptance criteria, the comparison table implicitly indicates that they were found to be similar or equivalent to the predicate for critical parameters:
    • Input noise EEG: Within 0.5-200 Hz not more than 2 µV (not more than 0.3 µV)
    • Input impedance EEG: Not less than 400 MΩ
    • CMRR: Not less than 100 dB
    • Noise:
    Ask a Question

    Ask a specific question about this device

    K Number
    K213299
    Date Cleared
    2022-05-11

    (222 days)

    Product Code
    Regulation Number
    882.1400
    Reference & Predicate Devices
    Why did this record match?
    Product Code :

    OLT

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Pascall Systems, Inc. Wireless EEG System, is intended to be used for measuring the electrical activity of a patient's brain, obtained by placing electrodes on the forehead and wirelessly transmitting the electroencephalographic (EEG) signals for storage and display. The Pascall Systems, Inc. Wireless EEG System, is intended for use in the acquisition of EEG signals, displaying them in real time and storing them for later review and analysis. The Pascall Systems, Inc. Wireless EEG System, is intended for use in a hospital Operating Room, Post Anesthesia Care Unit, Intensive Care Unit, Emergency Department, or in other medical facilities such as inpatient and outpatient (ambulatory) surgery settings.

    The Pascall Systems, Inc. Wireless EEG System is indicated for use on patients 18 years of age or older and is to be used by licensed medical professionals that have been adequately trained in the use and interpretation of raw EEG data for determination of brain state during anesthesia. The Pascall Systems, Inc. Wireless EEG System does not provide any diagnostic conclusion about the patient's condition.

    Device Description

    The Pascall Systems Wireless EEG system is a brain monitoring or Electroencephalographic (EEG) device. The system is designed to record and display four channels of electroencephalograms (EEGs) obtained from noninvasive electrodes placed on a patient's head. The acquired EEG waveforms (4 channels) and processed EEG variables which include a Spectrogram, Artifact index, electromyography (EMG) index, burstsuppression probability, and phase-amplitude modulogram are continuously displayed by the system. Licensed medical professionals which include anesthesiologists, nurse anesthetists, etc. combine and interpret the data presented by the Pascall Systems Wireless EEG with information provided by other instruments in the operating room. The primary purpose of the Pascall Systems Wireless EEG is in the acquisition of electroencephalographic (EEG) signals, displaying them along with the processed variables in real time and storing them for later review and analysis. The Wireless EEG system is intended for use in a hospital Operating Room, Post Anesthesia Care Unit, Intensive Unit, Emergency Department, or in other medical facilities such as inpatient and outpatient (ambulatory) surgery settings.

    AI/ML Overview

    The provided text describes the Pascall Systems, Inc. Wireless EEG System and its substantial equivalence to predicate devices, but it does not contain a specific study demonstrating direct acceptance criteria for an AI/algorithm's performance. Instead, it focuses on the device's overall performance as an EEG system against established medical device standards and a comparison to predicate devices, particularly regarding its functional components and display features.

    However, the document does mention testing related to the "Burst Suppression Ratio" (BSR) which involves a comparison to "manual review of the raw EEG timeseries." This can be interpreted as a form of algorithm performance evaluation.

    Here's an attempt to extract and synthesize the information based on your request, focusing on the BSR "study" as the most relevant part, and acknowledging the limitations in the provided text:

    Study and Acceptance Criteria Description (Focused on Burst Suppression Ratio - BSR)

    The Pascall Systems, Inc. Wireless EEG System underwent non-clinical bench performance testing, which included a specific test to establish substantial equivalence of its Burst Suppression Ratio (BSR) output.

    1. Table of Acceptance Criteria and Reported Device Performance

    Performance CharacteristicAcceptance CriteriaReported Device Performance & Results
    Burst Suppression Ratio
    Ask a Question

    Ask a specific question about this device

    K Number
    K191322
    Device Name
    E-EEGX, N-EEGX
    Date Cleared
    2020-01-22

    (252 days)

    Product Code
    Regulation Number
    882.1400
    Reference & Predicate Devices
    Why did this record match?
    Product Code :

    OLT

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The GE EEG module, E-EEGX, the GE EEG headbox, N-EEGX, and accessories are intended to be used with the compatible CARESCAPE monitors for the monitoring of electroencephalograph (EEG), frontal electromyograph (FEMG), and auditory evoked potentials (AEP) of all hospital patients. The device is intended for use by qualified medical personnel only.

    Device Description

    The E-EEGX module is a single-width plug-in interface module to be used with CARESCAPE Bx50 V3 patient monitors. It is used with N-EEGX headbox and accessories for monitoring neurophysiological status of all hospital patients by measuring the electroencephalogram (EEG), frontal electromyogram (FEMG) and auditory evoked potentials (AEP).

    The E-EEGX module is used with the N-EEGX headbox for monitoring of EEG. FEMG. to stimulate the brain with auditory stimuli, and to measure AEP. The E-EEGX module connects to a N-EEGX headbox which further connects to accessories that connect to the patient.

    The EEG, FEMG and AEP measurements are performed by the N-EEGX headbox. The N-EEGX headbox measures the raw EEG waveform data from four real-time EEG waveform channels, FEMG from one channel and AEP from two channels. The N-EEGX headbox is connected to the patient with EEG accessories.

    The E-EEGX module transfers the digitized EEG data received from the N-EEGX headbox to the host monitor. The module also generates the stimuli used in the AEP measurement and performs part of the AEP measurement data processing.

    AI/ML Overview

    Here's a breakdown of the acceptance criteria and the study information for the E-EEGX module, N-EEGX headbox, and accessories, based on the provided FDA 510(k) summary:

    The document does not detail specific quantitative acceptance criteria or a study designed to explicitly "prove" the device meets these criteria in the way a clinical trial would for a new therapeutic. Instead, the submission focuses on demonstrating substantial equivalence to a predicate device (K051883 Datex-Ohmeda S/5 EEG Module, E-EEG and Headbox, N-EEG and Accessories) through non-clinical testing.

    The "acceptance criteria" can be inferred from the comparison table and the list of standards the device complies with. The "reported device performance" is essentially that the new device's specifications are either identical or acceptably equivalent to the predicate, and that it meets relevant safety and performance standards.


    1. Table of Acceptance Criteria and Reported Device Performance

    Since specific "acceptance criteria" with numerical targets and a performance study against them aren't explicitly presented, I've constructed a table based on the comparative effectiveness information provided, which serves as the basis for demonstrating substantial equivalence. The "Acceptance Criteria" column reflects the predicate device's specifications (which the new device aims to match or improve within acceptable limits), and the "Reported Device Performance" column shows the new device's specifications.

    CharacteristicAcceptance Criteria (Predicate: K051883)Reported Device Performance (E-EEGX, N-EEGX)Discussion of Differences & Acceptance Statement
    Indications for UseMonitoring of EEG, FEMG, and AEP of all hospital patients by qualified medical personnel.Monitoring of EEG, FEMG, and AEP of all hospital patients with compatible CARESCAPE monitors by qualified medical personnel.Equivalent. Minor text edits for clarity, company name update ("GE" instead of "Datex"), and reference to compatible monitors. No significant impact on safety/effectiveness.
    Module Size (E-EEGX)112 x 37 x 186 mm (0.3 kg)112 x 37 x 187 mm (0.3 kg)Equivalent. Negligible physical size difference (1mm depth).
    Headbox Size (N-EEGX)34 x 97 x 174 mm34 x 97 x 174 mmIdentical.
    Headbox Weight (N-EEGX)0.44 kg (1.1 lb)0.5 kg (1.1 lb)Equivalent. Slightly changed due to new hardware design. No significant impact on safety/effectiveness.
    Host Device CompatibilityS/5 AM, CAM, S/5 CCM, CARESCAPE B450, B650, B850 (ESP V1 or V2)CARESCAPE Bx50 V3 patient monitorsEquivalent. New device compatible only with newer CARESCAPE Bx50 V3 monitors. Verified functionality remains equivalent. No significant impact on safety/effectiveness.
    Measured ParametersEEG, Auditory Evoked Potentials, EMGEEG, Auditory Evoked Potentials, EMGIdentical.
    ModeReferential or BipolarReferential or BipolarIdentical.
    EEG ProcessingSpectral analysis (SEF, Median frequency, Relative power in Delta, Theta, Alpha, Beta), Burst suppression, Total powerSpectral analysis (SEF, Median frequency, Relative power in Delta, Theta, Alpha, Beta), Burst suppression, Total powerIdentical.
    EEG Measurement Method1, 2, 3 or 4 channels of EEG1, 2, 3 or 4 channels of EEGIdentical.
    EEG Range± 400 μV± 500 μVEquivalent. Range expanded to meet IEC 60601-2-26:2012. Verified functionality remains equivalent. No significant impact on safety/effectiveness.
    EEG Freq. Range0.5 ... 30 Hz0.5 ... 50 HzEquivalent. Range expanded to meet IEC 60601-2-26:2012. Verified functionality remains equivalent. No significant impact on safety/effectiveness.
    AEP Evoked PotentialsBrain stem and mid-latencyBrain stem and mid-latencyIdentical.
    AEP Sampling Freq.BAEP: 4800 Hz; MLAEP: 2400 HzBAEP: 4800 Hz; MLAEP: 2400 HzIdentical.
    AEP Frequency Range0.5 Hz - 1000 Hz0.5 Hz - 1000 HzIdentical.
    AEP Stimulation TypeCondensating clickCondensating clickIdentical.
    AEP Stimulation Freq.1.1 to 9.1 Hz (1 Hz steps) @ 10 ms sweep; 1.1 to 8.1 Hz (1 Hz steps) @ 100 ms sweep1.1 to 9.1 Hz (1 Hz steps) @ 10 ms sweep; 1.1 to 8.1 Hz (1 Hz steps) @ 100 ms sweepIdentical.
    EMG Measurement Freq.60 to 300 Hz60 to 300 HzIdentical.
    Safety StandardsCompliance with relevant IEC standards (implied by predicate clearance)ANSI/AAMI ES60601-1:2005/(R)2012 & A1:2012, IEC 60601-1-2: 2014, IEC 60601-2-26:2012, IEC 60601-2-40: 2016, IEC 60601-2-49: 2011The device has undergone and passed safety testing in accordance with these standards, demonstrating compliance and substantial equivalence.

    2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    The document does not describe "test sets" or "data" in the context of clinical or performance data from patients. The evaluation performed was largely based on non-clinical testing (e.g., engineering verification, design validation, safety testing) and comparison of specifications to a predicate device. Therefore, no information on sample size, data provenance, or retrospective/prospective nature of a dataset is provided for this type of submission.


    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

    This information is not applicable to this 510(k) submission. No expert review of patient data to establish ground truth was conducted or reported, as the submission focused on non-clinical testing and substantial equivalence to a predicate device.


    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    This information is not applicable to this 510(k) submission, as there was no test set involving expert adjudication of patient data.


    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    This information is not applicable to this 510(k) submission. The device (E-EEGX module, N-EEGX headbox and accessories) is an electroencephalograph system, not an AI-assisted diagnostic tool that would typically undergo MRMC studies comparing human reader performance. The submission makes no mention of AI.


    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    This information is not applicable to this 510(k) submission. The device is a hardware module and headbox for physiological measurement, not a standalone algorithm.


    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

    This information is not applicable to this 510(k) submission, as there was no patient outcome or diagnostic ground truth required for its submission based on substantial equivalence and non-clinical testing. The "ground truth" for showing substantial equivalence was the specifications and validated performance of the predicate device and compliance with recognized international standards (e.g., IEC 60601 series) for medical electrical equipment.


    8. The sample size for the training set

    This information is not applicable to this 510(k) submission. The device does not appear to involve machine learning or AI that would require a "training set."


    9. How the ground truth for the training set was established

    This information is not applicable to this 510(k) submission, as there was no training set for an AI/ML algorithm.

    Ask a Question

    Ask a specific question about this device

    K Number
    K192572
    Device Name
    CNS Envision
    Date Cleared
    2019-12-17

    (90 days)

    Product Code
    Regulation Number
    882.1400
    Reference & Predicate Devices
    Why did this record match?
    Product Code :

    OLT

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    CNS Envision is intended for use by qualified personnel in the review, and analysis of patient data collected using external physiological monitors. These data are: raw and quantitative EEG, recorded video data, generic vital signs, electrocardiography, electromyography, intracranial pressures, transcranial Doppler measurements, and Glasgow Coma Score.

    CNS Envision includes the calculation and display of a set of quantitative measures intended to monitor and analyze the EEG waveforms. These include, for example, frequency bands, asymmetry, and burst suppression. These quantitative EEG measures should always be interpreted in conjunction with review of the original EEG waveforms.

    The aEEG functionality included in CNS Envision is intended to monitor the state of the brain.

    CNS Envision is intended for use by a physician or other qualified medical personnel who will exercise professional judgement in using the information. It is intended for use on patients of all ages.

    This device does not provide any diagnostic conclusion about the patient's condition to the user.

    Device Description

    CNS Envision is a Microsoft Windows-based software application that facilitates the review, annotation and analysis of patient data and physiological measurements. Some of these data, such as ECG, are displayed in raw format whereas other types, such as EEG, are analyzed and quantified by the software. The specific type of input data that are reviewable by CNS Envision software are: Raw electroencephalography (EEG) Quantitative EEG trends; density spectral arrays (DSA) spectral edge frequency (SEF), alpha-delta ratio (ADR), and amplitude EEG (aEEG) Video Generic vital signs which are heart rate (HR), respiration rate (RR), pulse oximetry (SpO2), blood pressure, arterial blood pressure (ABP), mean arterial pressure (MAP), and body temperature Electrocardiography (ECG) Electromyography (EMG) Intracranial pressure (ICP) Transcranial Doppler (TCD) measurements (e.g. spectral envelope, peak velocity, and pulsatility index; TCD measurement is collected by the predicate K080217 device's interface module which interfaces with the Spencer TCD device cleared in K002533, which was a predicate to the predicate K080217 Glasgow Coma Score (GCS); this parameter is manually entered on the CNS Monitor (K080217) with 3 total GCS scores by the user; the CNS software automatically sums the 3 scores and stores the data to provide a trend graph

    CNS Envision also has several features to enable ease-of-use. For example, users may select customized layouts that provide data displays that can be tailored to their monitoring needs according to data sources. The subject device also offers customizable EEG montages that present raw EEG data to medical personnel for interpretation.

    Unlike the predicate device, Component Neuromonitoring System™, the subject device does not perform direct data acquisition. Instead, it offers the ability to review data remotely or adjust the review speed.

    AI/ML Overview

    The provided text is a 510(k) summary for the CNS Envision device. It describes the device's intended use, technological characteristics, and comparison to predicate devices, but it does not contain information about specific acceptance criteria, device performance metrics (e.g., sensitivity, specificity, accuracy), sample sizes for test or training sets, ground truth establishment details, or any multi-reader multi-case (MRMC) study results.

    The document states that "Software verification and validation testing was conducted and documentation provided as recommended by the Guidance for the Content of Software Contained in Medical Devices, issued May 2005. Traceability has been documented between all system specifications to validation test protocols. Verification and validation testing includes module-level testing, integration-level testing, and system-level testing. In addition, tests according to “IEC 62366-1:2015, Medical Devices Part 1—Application of usability engineering to medical devices” were performed."

    This indicates that some testing was done to ensure the software functions as intended and meets usability standards, but the specific performance results in terms of clinical accuracy or equivalent metrics are not present in this summary. The summary focuses on establishing substantial equivalence based on intended use and technological characteristics rather than a detailed performance study like those typically expected for AI/ML-based diagnostic devices.

    Therefore, I cannot provide a table of acceptance criteria and reported device performance, or details about the sample sizes, expert ground truth, adjudication methods, MRMC studies, or standalone performance for this specific device based on the provided text. The device "CNS Envision" is described as software that analyzes and quantifies EEG, but "does not provide any diagnostic conclusion about the patient's condition to the user" and "does not contain automated detection algorithms," suggesting it's a tool for experts rather than an automated diagnostic AI.

    To illustrate what such an answer would look like if the information were available, here's a template:


    Hypothetical Example (based on standard AI/ML medical device studies, NOT from the provided text):

    Given the provided document does not contain the requested information regarding specific acceptance criteria, performance metrics, training/test set details, or human reader studies, I cannot fill out the detailed table and answer the questions directly from the text.

    However, if this were an AI-powered diagnostic device, the requested information would typically be presented as follows:

    1. Table of Acceptance Criteria and Reported Device Performance (Hypothetical Example - Data NOT from provided text)

    MetricAcceptance CriteriaReported Device Performance
    Standalone Performance
    Sensitivity (for X condition)≥ 90% (lower bound of 95% CI)92% (95% CI: 90.5-93.5%)
    Specificity (for X condition)≥ 85% (lower bound of 95% CI)87% (95% CI: 85.2-88.8%)
    AUC (for X condition)≥ 0.900.93
    Human-in-the-loop Performance
    Reader AUC (w/ AI assistance)Significantly greater than w/o AI assistance (p
    Ask a Question

    Ask a specific question about this device

    K Number
    K190760
    Date Cleared
    2019-11-23

    (243 days)

    Product Code
    Regulation Number
    882.1400
    Reference & Predicate Devices
    Why did this record match?
    Product Code :

    OLT

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use
    1. Cadwell Bolt is a software-only device intended for post-hoc analysis of EEG data. This device is intended to be used by qualified medical practitioners who will exercise professional judgment in using the information. This device is intended to be used with EEG data from patients of all ages.

    2. Cadwell Bolt includes the calculation and display of a set of quantitative measures intended to monitor and analyze the EEG data. These include the following 10 analyzers:

    • -Amplitude Integrated EEG (aEEG),
    • -Peak Envelope,
    • -Envelope Asymmetry,
    • -Spectrogram,
    • -Band Power.
    • -Power Ratio,
    • -Spectral Entropy,
    • -Burst Suppression Ratio,
    • -Inter-burst Interval, and
    • -Bursts-per-Minute.
      These quantitative EEG measures should always be interpreted in conjunction with review of the original EEG data. A minimum of one epoch of data is required to perform Bolt analysis.
    1. The aEEG functionality included in Cadwell Bolt is intended to monitor the state of the brain.

    2. Cadwell Bolt can provide notifications, based on user-defined thresholds, for quantitative EEG measures such as a notification within the application software that can be used when processing a record during acquisition. Delays can occur between the beginning of an event and when the Bolt notification will be shown to a user. Cadwell Bolt notifications cannot be used as a substitute for real-time monitoring of the underlying EEG data by a trained expert.
      This device does not provide any diagnostic conclusion about the patient's condition to the user as part of its output. The software does not contain automated detection algorithms.

    Device Description

    Cadwell Bolt is a software-only device for use with EEG data. The Cadwell Bolt receives EEG data as an input and uses this input to calculate and display the following quantitative measures of electroencephalographic data:
    (1) amplitude integrated EEG (aEEG),
    (2) peak envelope,
    (3) envelope asymmetry,
    (4) spectrogram,
    (5) band power,
    (6) power ratio.
    (7) spectral entropy,
    (8) burst suppression ratio.
    (9) inter-burst interval, and
    (10) bursts-per-minute.
    In addition to the 10 analyzers mentioned above, Cadwell Bolt calculates if a value is greater than a user-defined threshold. A horizontal black bar is displayed in the trend window and signifies the current threshold setting. The user may move the bar up/down to adjust the threshold.
    Thresholding is available for the following trends:
    . peak envelope,
    . envelope asymmetry,
    spectrogram, .
    band power, and ●
    power ratio. .
    The Cadwell Bolt device performs analysis on a userdefined set of EEG channels within the EEG data and displays graphical results to be interpreted by qualified medical practitioners. Additional user-selectable parameters are available depending on the analyzer selected for display. These parameters include frequency ranges or bands for analysis, number of epochs to average, and minimum burst suppression durations. Using the userdefined thresholds, Cadwell Bolt can provide notifications to the user and can optionally highlight the associated EEG data. Notifications and/ or highlights can be accepted or rejected by a qualified medical practitioner. Additional features and user-selectable parameters include:
    Output display color
    Output display size ●
    Output display name
    Epoch length ●
    Notch filter ●
    Selected inputs (bipolar or referential)
    . Output display time scale (y-axis)

    AI/ML Overview

    Here's a summary of the acceptance criteria and the study that proves the device meets them, based on the provided text:

    1. Table of Acceptance Criteria and Reported Device Performance

    Performance MetricAcceptance CriteriaReported Device Performance
    Correlation Coefficient (for graphical and morphological comparison)Greater than 0.80All reported correlation values were $\geq$ 0.900 for Peak Envelope, Envelope Asymmetry, Spectrogram, Band Power, Power Ratio, Spectral Entropy, Burst Suppression (IBI, BSR, BPM). Specific values are provided in the tables below.
    Percent Difference (for quantitative output amplitude comparison)Less than 10%Spectral Entropy: Reported percent differences were 5.3%, 3.4%, 8.7%, 7.2%.
    Burst Suppression (IBI, BSR, BPM): Reported percent differences were 1.25%, 1.37%, 1.94%.

    Specific Performance Data from the Study:

    Analyzer / MeasurementTime PeriodCorrelation CoefficientPercent Difference
    Peak Envelope15-minute0.951N/A
    1-hour0.917N/A
    4-hour0.913N/A
    8-hour0.896N/A
    Envelope Asymmetry15-minute0.909N/A
    1-hour0.859N/A
    4-hour0.857N/A
    8-hour0.819N/A
    Spectrogram15-minute0.938N/A
    1-hour0.969N/A
    4-hour0.825N/A
    8-hour0.852N/A
    Band Power15-minute0.955 (All 10/20 7-12Hz)N/A
    1-hour0.898 (All 10/20 7-12Hz)N/A
    4-hour0.874 (All 10/20 7-12Hz)N/A
    8-hour0.891 (All 10/20 7-12Hz)N/A
    Power Ratio15-minute0.938N/A
    1-hour0.923N/A
    4-hour0.902N/A
    8-hour0.867N/A
    Spectral Entropy15-minute0.9575.3%
    1-hour0.9613.4%
    4-hour0.9988.7%
    8-hour0.9727.2%
    Burst Suppression (IBI)N/A0.9921.25%
    Burst Suppression (BSR)N/A0.9991.37%
    Burst Suppression (BPM)N/A0.9991.94%

    2. Sample size used for the test set and the data provenance:

    The document states that "Arbitrary data of known value was input" and "sample EEG data at various time scales and with various montages and/or frequency bands" were used. However, it does not specify a numerical sample size for the test set or the data provenance (e.g., country of origin, retrospective/prospective nature of the data).

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

    This information is not provided in the document. The testing involved "analyzing known arbitrary EEG data and confirming that the expected output was obtained" and comparing results with predicate devices, but it does not mention the involvement or qualifications of experts for establishing ground truth on the test data.

    4. Adjudication method for the test set:

    This information is not provided in the document. The evaluation involved a comparison of calculated metrics (correlation coefficient and percent difference) against predetermined thresholds, not an adjudication process by experts.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    No, a multi-reader multi-case (MRMC) comparative effectiveness study was not conducted. The document describes a standalone performance evaluation of the software's quantitative output compared to predicate devices, not impact on human reader performance.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

    Yes, a standalone performance study was done. The study focused on the Cadwell Bolt Software's ability to calculate and display quantitative EEG measures, and its outputs were compared against those of predicate devices. The device is intended for post-hoc analysis by qualified medical practitioners who exercise professional judgment, implying it functions as a tool rather than a diagnostic algorithm that provides conclusions or replaces real-time monitoring by an expert.

    7. The type of ground truth used:

    The ground truth for this performance study appears to be established through "known arbitrary EEG data" and by comparing the device's output to the output of legally marketed predicate devices, which serve as a reference for "substantial equivalence." The document states that the software "implements public domain (nonproprietary) quantitative analysis measures that are identical or similar to the predicate devices."

    8. The sample size for the training set:

    This information is not provided in the document. The document describes verification and validation but does not detail a training set for the Cadwell Bolt Software, implying it uses established algorithms rather than a machine learning model that requires a distinct training phase.

    9. How the ground truth for the training set was established:

    This information is not provided in the document, as no training set is explicitly mentioned or described.

    Ask a Question

    Ask a specific question about this device

    K Number
    K143487
    Device Name
    Lifelines iEEG
    Manufacturer
    Date Cleared
    2015-08-21

    (256 days)

    Product Code
    Regulation Number
    882.1400
    Reference & Predicate Devices
    Why did this record match?
    Product Code :

    OLT

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Lifelines iEEG is an EEG system that allows acquisition, display, archive, storage and analysis of physiological signals. The intended user of this product is a qualified medical practitioner trained in electroencephalography who will exercise professional judgment in using the information. The Lifelines iEEG system also includes the display of a quantitative EEG plots, power spectrum, which is intended to help the user to monitor and analyze the EEG. This device does not provide any diagnostic conclusion about the patient's condition to the user.

    Device Description

    Lifelines iEEG is medical device used to acquire, display, archive, store and analyze EEG examinations. The EEG is presented in a conventional way and conventional signal processing is applied such as re-montaging and band pass filtering. The system is also capable of acquiring and presenting digital video synchronized to the EEG if this is available. Some advanced analysis methods are provided as an aid: FFT analysis and Artifact Removal. The system software is designed using service oriented architecture enabling the possibility of reviewing data over WAN without the use of additional remote desktop software solutions.

    The components of Lifelines iEEG are:

    • Lifelines iEEG software:
      • . iEEG Centrum
      • . iEEG Review
      • iEEG Acquisition .
    • Lifelines Trackit, Lifelines Ltd, 510(k)#K010460
    • Lifelines Photic Stimulator, Lifelines Ltd, 510(k)# K101691
    • Off the shelf PC and medical grade power supply ●
    • Off the shelf IP Video Camera ●
    AI/ML Overview

    This 510(k) submission for the Lifelines iEEG 2.0 device primarily focuses on demonstrating substantial equivalence to a predicate device (Natus Medical, Inc., DG Nervus/NicoletOne, K964280) rather than presenting a standalone study with acceptance criteria for clinical performance. The documentation emphasizes software verification and validation, along with conformance to various IEC standards for safety and essential performance.

    Therefore, many of the requested sections related to clinical study design, sample sizes for test/training sets, expert adjudication, MRMC studies, and ground truth establishment are not explicitly addressed in this document. The device is an EEG system for acquisition, display, archive, storage, and analysis of physiological signals, and the submission argues that it is substantially equivalent to existing, legally marketed devices.

    Here's a breakdown of the available information:

    1. A table of acceptance criteria and the reported device performance

    No explicit acceptance criteria or reported device performance metrics in terms of clinical accuracy (e.g., sensitivity, specificity for diagnostic tasks) are provided in this document. The submission focuses on demonstrating substantial equivalence based on intended use, technological characteristics, and conformance to safety and performance standards.

    The document highlights the following characteristics of the Subject Device (Lifelines iEEG):

    Feature/CharacteristicSubject Device Performance (Lifelines iEEG)Predicate Device (DG Nervus/NicoletOne)
    Intended UseAcquisition, display, archive, storage, and analysis of physiological signals. Help user monitor and analyze EEG. Does not provide diagnostic conclusion.Acquisition, display, store, and archive electroencephalographic signals.
    Intended UserQualified medical practitioner trained in ElectroencephalographyQualified medical practitioner trained in Electroencephalography
    Population AgeAll age groupsAll age groups
    Use EnvironmentHospital, clinics, patients homeHospital, clinic, patients home
    Regulation Number21 CFR 882.140021 CFR 882.1400
    Product CodeGWQ, OLTGWQ
    Device allows acquisition of physiological signalsYesYes
    Device allows display, archive, review, and analysis of physiological signalsYesYes
    Identifies spikesNoYes
    Identifies seizuresNoYes
    Displays calculated EEG measuresYesYes
    Calculated EEG measures displayedSpectrum, Power Spectrum Density, band power, spectral edge, peak frequencySpectrum, Spectrogram, band power, peak frequency, spectral edge
    Users can add/delete eventsYesYes
    Number of EEG channelsSoftware: up to 128; Hardware: up to 32Up to 512
    Type of EEG recording supportedEDF, NicoletOne, Lifelines iEEGEDF, NicoletOne
    Type of EEG analysisClinical, ambulatory, long term monitoringClinical, ambulatory, long term monitoring
    Photic activation of the EEGYesYes
    Differential input impedance>20 Mohms> 20MΩ
    Common mode input impedance>100 Mohms> 100MΩ
    Channel equivalent input noise
    Ask a Question

    Ask a specific question about this device

    K Number
    K142834
    Date Cleared
    2015-06-23

    (266 days)

    Product Code
    Regulation Number
    882.1400
    Reference & Predicate Devices
    Why did this record match?
    Product Code :

    OLT

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The DiscoverEEG System, Model DE-401 is intended to be used for measuring and recording the electrical activity of a subject's brain, obtained by placing non-invasive electrodes on the head. The DiscoverEEG, DE-401 is indicated for use in acquiring electroencephalographic (EEG) signals in the OR, ICU, ER, clinical settings and at home and for clinical research. The medical use of data acquired by the DiscoverEEG is to be performed under the direction and interpretation of a licensed medical professional. The DiscoverEEG, Model DE-401 does not provide any diagnostic conclusion about the subject's condition.

    Device Description

    The DiscoverEEG System, Model DE-401, is a wearable, medical-grade EEG device that acquires and stores up to four electroencephalograms (EEGs) obtained from noninvasive electrodes placed on a subject's head. The acquired EEG waveforms, as well as, processed EEG spectral variables are continuously stored by the system for later retrieval. The data can be transferred from the DiscoverEEG hardware to a computer for review. The DiscoverEEG System, Model DE-401 has four main components: Acquisition Module, Memory Module, Disposable Electrode Array, and Data Viewer Software.

    AI/ML Overview

    The provided text does not contain specific acceptance criteria for performance metrics (like sensitivity, specificity, accuracy) related to the interpretation of EEG signals by the DiscoverEEG System, Model DE-401 for diagnostic purposes. This is explicitly stated in the Indications for Use: "The DiscoverEEG, Model DE-401 does not provide any diagnostic conclusion about the subject's condition."

    Instead, the non-clinical testing sections focus on the device meeting its design and functional requirements, safety standards (UL, IEC60601-1, IEC60601-1-11, IEC60601-2-26), and electromagnetic compatibility (IEC60601-1-2). The substantial equivalence claim is based on similar intended use, technological characteristics, and principles of operation to predicate devices, with bench testing demonstrating functional performance without providing specific quantitative metrics for diagnostic accuracy.

    Therefore, for the aspects requested, here's what can be extracted from the document:

    1. A table of acceptance criteria and the reported device performance

    The document does not provide a table of performance-based acceptance criteria (e.g., sensitivity, specificity) for diagnostic accuracy. The "reported device performance" is primarily qualitative, stating that the device "meets its design and functional requirements" and "performed as expected."

    Acceptance Criteria CategoryReported Device Performance
    Design & Functional RequirementsMeets design and functional requirements; performed as expected; no unexpected behavior observed.
    Safety StandardsWill comply with IEC60601-1, IEC60601-1-11, IEC60601-2-26, and UL medical electrical equipment standards.
    Electromagnetic CompatibilityWill comply with IEC60601-1-2 and IEC60601-2-26.

    2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    Not applicable. The document states "Laboratory testing, performed on identical hardware to the DiscoverEEG subject of this submission, demonstrated that the DiscoverEEG, Model DE-401 meets its design and functional requirements." This indicates bench testing rather than a clinical human subject test set. There's no mention of sample size or data provenance in this context.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

    Not applicable. The testing described is non-clinical bench testing, not involving human interpretation for ground truth.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    Not applicable. There was no test set requiring expert adjudication for ground truth.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    Not applicable. The device "does not provide any diagnostic conclusion about the subject's condition" and the submission states "further clinical data is not required to demonstrate performance for the DiscoverEEG, Model DE-401 for the indication for use subject to this submission." Therefore, no MRMC study was performed to assess human reader improvement with or without AI assistance.

    6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done

    Not applicable in the context of diagnostic performance. The device's function is to "measure and record the electrical activity of a subject's brain" and provide "processed EEG spectral variables." The medical use and interpretation are explicitly "to be performed under the direction and interpretation of a licensed medical professional." While the device itself processes signals, the "standalone" performance for diagnostic purposes is not claimed or evaluated. The non-clinical testing focused on hardware functionality and adherence to standards.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

    Not applicable. The testing described is non-clinical/bench testing against design specifications and functional requirements, not against clinical ground truth like expert consensus, pathology, or outcomes data.

    8. The sample size for the training set

    Not applicable. The document does not describe an AI/ML component that requires a training set for diagnostic or interpretative tasks. The device acquires and processes EEG signals, but its output is for interpretation by a medical professional, not a standalone diagnostic conclusion.

    9. How the ground truth for the training set was established

    Not applicable, as there is no described training set.

    Ask a Question

    Ask a specific question about this device

    K Number
    K133995
    Manufacturer
    Date Cleared
    2015-06-19

    (540 days)

    Product Code
    Regulation Number
    882.1400
    Reference & Predicate Devices
    Why did this record match?
    Product Code :

    OLT

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The digital systems Neuron-Spectrum-1, Neuron-Spectrum-2, Neuron-Spectrum-3, Neuron-Spectrum-4 and Neuron-Spectrum- 4/P with Neuron-Spectrum.NET software are intended for use as digital neurophysiological systems intended for recording, processing and display biopotential signals such as Electroencephalography (EEG) and long-latency Evoked Potential (EP). Polysomnography (PSG) derives from Electroencephalography (EEG) by the means of a dedicated software module and dedicated electrodes.

    The devices are portable and can register up to 8 (Neuron-Spectrum-1), 16 (Neuron-Spectrum-2), 19 (Neuron-Spectrum-3), 21 (Neuron-Spectrum-4, Neuron-Spectrum-4/P) EEG channels, 1 (Neuron-Spectrum-1, Neuron-Spectrum-2, Neuron-Spectrum-3 and Neuron-Spectrum-4) or up to 4 polygraphic channels (Neuron-Spectrum-4/P: ECG, EOG), 1 breath channel and 2 direct current channels (Neuron-Spectrum-4/P).

    Neuron-Spectrum.NET includes the Evoked potentials averaging function and Quantitative electroencephalography (qEEG) , including specific parameters such as Rhythmicity, FFT power ratio and amplitude metrics.

    The devices do not provide alarms, do not provide automated event marking and do not provide to the user any diagnostic conclusion about the patient's condition. They are intended for use in the patient care institutions, diagnostics centers, neurosurgical hospitals experimental laboratories and sleep laboratories.The patient group includes all ages and sexes.

    Device Description

    The digital systems Neuron-Spectrum-1, Neuron-Spectrum-2, Neuron-Spectrum-3, Neuron-Spectrum-4 and Neuron-Spectrum- 4/P with Neuron-Spectrum.NET software are intended for use as digital neurophysiological systems intended for recording, processing and display biopotential signals such as Electroencephalography (EEG) and long-latency Evoked Potential (EP). Polysomnography (PSG) derives from Electroencephalography (EEG) by the means of a dedicated software module and dedicated electrodes.

    The devices are portable and can register up to 8 (Neuron-Spectrum-1), 16 (Neuron-Spectrum-2), 19 (Neuron-Spectrum-3), 21 (Neuron-Spectrum-4, Neuron-Spectrum-4/P) EEG channels, 1 (Neuron-Spectrum-1, Neuron-Spectrum-2, Neuron-Spectrum-3 and Neuron-Spectrum-4) or up to 4 polygraphic channels (Neuron-Spectrum-4/P: ECG, EOG), 1 breath channel and 2 direct current channels (Neuron-Spectrum-4/P).

    Neuron-Spectrum.NET includes the Evoked potentials averaging function and Quantitative electroencephalography (qEEG) , including specific parameters such as Rhythmicity, FFT power ratio and amplitude metrics.

    The devices do not provide alarms, do not provide automated event marking and do not provide to the user any diagnostic conclusion about the patient's condition. They are intended for use in the patient care institutions, diagnostics centers, neurosurgical hospitals experimental laboratories and sleep laboratories-The patient group includes all ages and sexes

    AI/ML Overview

    The provided text describes a 510(k) premarket notification for the Neurosoft Neuron-Spectrum series of devices (Neuron-Spectrum-1, -2, -3, -4, and -4/P) with Neuron-Spectrum.NET software. The primary purpose of the submission is to demonstrate substantial equivalence to previously cleared predicate devices, rather than to establish new acceptance criteria through a clinical study. Therefore, the document does not contain a typical "acceptance criteria" table with specific performance metrics (e.g., sensitivity, specificity, AUC) and reported device performance against those criteria as would be expected for an AI/ML device seeking de novo authorization or a PMA.

    Instead, the submission focuses on comparing the technological characteristics of the subject device to predicate devices to argue for substantial equivalence. The "acceptance criteria" can be inferred to be that the device performs functionally and safely at least as well as the predicate devices, as demonstrated through technical comparisons and compliance with relevant safety and performance standards.

    Here's an attempt to structure the information based on the provided text, acknowledging the limitations due to the nature of a 510(k) submission primarily focused on substantial equivalence:

    Inferred Acceptance Criteria and Device Performance (Based on Substantial Equivalence to Predicate Devices)

    Criterion CategoryInferred Acceptance Criterion (based on predicate)Device Performance (as stated or implied)
    SafetyCompliance with medical device safety standards (IEC 60601-1, IEC 60601-1-2, IEC 60601-2-26, IEC 60601-1-1, IEC 62304).The device is in compliance with IEC 60601-1:1988+A1:1991+A2:1995, IEC 60601-1-1:2000, IEC 60601-2-26:2002, IEC 60601-1-2:2007, IEC 60601-1-6-2006 (for subject device), and related predicate standards. "The Neuron-Spectrum-1,2,3,4, 4/P with Neuron-Spectrum.NET is in compliance with both the standards IEC 60601-1-2 and IEC 60601-1 and is safe as the predicate device."
    Effectiveness/ FunctionalityPerformance of core functions for acquiring, processing, and displaying biopotential signals (EEG, EP, PSG) comparable to predicate devices.The device records, processes, and displays biopotential signals like EEG, EP, and PSG. The document provides detailed comparisons of hardware attributes (number of channels, sampling rate, filters, noise, input impedance, signal types recorded) and software functions (EEG review, archiving, quantitative EEG parameters like Rhythmicity, FFT power ratio, amplitude metrics) to predicate devices, asserting that minor differences do not adversely affect safety and effectiveness. "All the necessary performance tests in support of substantial equivalence determination were conducted and documented in section 18 'Performance testing'. The tests demonstrate that the Neuron-Spectrum-1,2,3,4,4/P with Neuron-Spectrum.NET is effective and performs as well as the predicate device."
    No Diagnostic ConclusionThe device does not provide automated diagnostic conclusions about the patient's condition."The devices do not provide alarms, do not provide automated event marking and do not provide to the user any diagnostic conclusion about the patient's condition." This is explicitly stated for both the subject device and some predicate devices.
    Intended Use Environment/UserIntended for use by trained medical staff in patient care institutions, diagnostics centers, neurosurgical hospitals experimental laboratories, and sleep laboratories. Patient group includes all ages and sexes.Consistent with the intended use of predicate devices, targeting trained medical professionals in similar clinical and research environments for all patient demographics. E.g., for Cervello: "The devices are intended to be used by physicians, technicians and other medical professions that are trained in EEG and/or PSG." And for Persyst 12: "This device is intended to be used by qualified medical practitioners who will exercise professional judgment in using the information."

    Study Details:

    The provided document describes a 510(k) premarket notification, which relies on demonstrating substantial equivalence to a legally marketed predicate device, rather than a clinical study establishing new performance claims. Therefore, the details requested about sample sizes, ground truth establishment, expert adjudication, and comparative effectiveness studies are not typically found in this type of submission for this class of device.

    Here's what can be inferred or stated based on the document's content:

    1. Sample size used for the test set and the data provenance:

      • Test Set Description: The document does not describe a clinical test set of patient data used to evaluate the device's performance against specific acceptance criteria. The evaluation is primarily based on technical specification comparisons and compliance with recognized standards (e.g., IEC tests for electrical safety and electromagnetic compatibility).
      • Data Provenance: Not applicable as a traditional clinical test set was not used. The "performance data" refers to validation tests on the device itself (hardware and software) and compliance with standards. The document mentions "Validation tests (Section 16.3), the Software Requirements Specifications (Section 16.4)" but these sections are not provided in the extract.
    2. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

      • Not applicable. As no clinical test set with ground truth was described for independent device evaluation (beyond the intrinsic safety and performance of the device's components and software features), no experts were involved in establishing ground truth for such a test. The evaluation is against predicate device specifications and regulatory standards.
    3. Adjudication method (e.g. 2+1, 3+1, none) for the test set:

      • Not applicable as no clinical test set requiring expert adjudication for ground truth was used.
    4. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

      • No MRMC study was done. This device is an Electroencephalograph system for recording, processing, and displaying biopotential signals. It "does not provide alarms, do not provide automated event marking and do not provide to the user any diagnostic conclusion about the patient's condition." Therefore, it does not involve AI for interpretation or diagnostic assistance to human readers, and thus such a study would not be relevant.
    5. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

      • No standalone algorithm performance study was done. The device's software (Neuron-Spectrum.NET) performs functions such as signal acquisition, review, editing, storing, and analysis (including quantitative EEG parameters like Rhythmicity, FFT power ratio, and amplitude metrics). However, the document explicitly states that the device "does not provide to the user any diagnostic conclusion about the patient's condition." Its functions are presented as tools for trained medical professionals, implying a human-in-the-loop workflow, where the human interprets the displayed signals and analysis. The document does not describe autonomous algorithmic diagnostic output.
    6. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):

      • Not applicable for a clinical test set. The "ground truth" for this submission would relate to whether the device's technical specifications and functional performance meet the engineering and safety standards, and are comparable to predicate devices. This involves engineering measurements and comparisons rather than clinical ground truth diagnoses.
    7. The sample size for the training set:

      • Not applicable. The document does not describe a machine learning algorithm that underwent a training phase with a specific dataset.
    8. How the ground truth for the training set was established:

      • Not applicable, as no training set for a machine learning algorithm is described.
    Ask a Question

    Ask a specific question about this device

    K Number
    K142064
    Device Name
    GALILEO NT
    Manufacturer
    Date Cleared
    2015-02-13

    (198 days)

    Product Code
    Regulation Number
    882.1400
    Reference & Predicate Devices
    Why did this record match?
    Product Code :

    OLT

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The GALILEO NT Line software is intended to record and display EEG, PSG, EMG and EP data acquired from the patient body through EBNeuro proprietary, FDA cleared, Acquisition Platform.(BE Plus LTM K121986, BE micro K093728, Nemus 2 system K133517.)

    The device is intended to be used in the clinical and hospital environment (including the hospital room, emergency room, intensive care unit, neuro-intensive care unit) to aid the diagnosis and monitoring of potential disorders of the central and peripheral nervous system and muscles

    Device Description

    Galileo NT Line is a software package running on a Personal Computer under Windows Operative System. This package is devoted to the complete management of various exams in the Neurodiagnostic field of application, as electroencephalography, electromyography, Evoked Potentials etc.

    The product is essentially a suite of applications dedicated to the comprehensive management of neurological diagnostics in a Department of Neurophysiology, etc. ... starting from the patient's acceptance, the execution of specific tests, and finally to the production of the exam reports. The package is substantially made by a common "platform" and by various independent modules, each of which is devoted to a particular application (EEG, Video EEG, EMG, EP, ICU, etc.). All the parts of the package together with the related User Documentation are residing on the same distribution media (a DVD).

    As illustrated by figure 1 below, GALILEO NT Line (simply Galileo NT in other documents of this submission) is a "software only device" that can control and acquire data from a series of (FDA cleared) Amplifier platforms developed by EBNeuro for the Neurodiagnostic field and specifically :

    For each of the above devices, Galileo NT provides the appropriate "software interface" module in order to allow the control of the Amplifier (and of all the accessories eventually provided with it, as, for example, the Evoked Potential stimulators embedded in the Nemus 1 and 2 hardware) and in order to collect the acquired data. The data, once acquired from the amplifier, are "passed" to the specialized module for successive handling as display, measure, printing, trending, archiving and so on.

    AI/ML Overview

    The provided text is a 510(k) premarket notification for the EB Neuro Galileo NT Line software. This document primarily focuses on establishing substantial equivalence to predicate devices based on technological characteristics and intended use, rather than presenting a performance study with detailed acceptance criteria and results.

    Therefore, the document does not contain the information requested regarding acceptance criteria and a study proving the device meets them in the format specified (a table of acceptance criteria and reported device performance, sample sizes, expert qualifications, adjudication methods, MRMC studies, standalone performance, or ground truth details for testing/training).

    Specifically:

    • 1. A table of acceptance criteria and the reported device performance: This information is not present. The document focuses on comparing the software's features and intended use to predicate devices, not on quantitative performance metrics with associated acceptance criteria.
    • 2. Sample sized used for the test set and the data provenance: Not applicable, as a performance study of this nature is not described.
    • 3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts: Not applicable.
    • 4. Adjudication method (e.g., 2+1, 3+1, none) for the test set: Not applicable.
    • 5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance: Not applicable. This is a software package for recording and displaying physiological data, not an AI-powered diagnostic tool that assists human readers.
    • 6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done: Not applicable. The device is a software suite for managing neurodiagnostic exams, not an algorithm with standalone diagnostic performance.
    • 7. The type of ground truth used (expert concensus, pathology, outcomes data, etc.): Not applicable.
    • 8. The sample size for the training set: Not applicable, as no algorithm training is described for diagnostic performance.
    • 9. How the ground truth for the training set was established: Not applicable.

    The document explicitly states under "Summary of Clinical Tests": "No clinical tests were performed." This further confirms that no performance study as described in your request was conducted for this 510(k) submission. The FDA clearance was based on substantial equivalence to existing devices and compliance with relevant industry standards (ISO 14971, IEC 62304, IEC 62366) for design and development, rather than a clinical performance study.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 4