AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use
  • HemoSphere Advanced Monitor with HemoSphere Swan-Ganz Module: The HemoSphere advanced monitor when used with the HemoSphere Swan-Ganz module and Edwards Swan-Ganz catheters is indicated for use in adult and pediatric critical care patients requiring monitoring of cardiac output (continuous [CO] and intermittent [iCO]) and derived hemodynamic parameters in a hospital environment. Pulmonary artery blood temperature monitoring is used to compute continuous and intermittent CO with thermodilution technologies. It may also be used for monitoring hemodynamic parameters in conjunction with a perioperative goal directed therapy protocol in a hospital environment. Refer to the Edwards Swan-Ganz catheter and Swan-Ganz Jr catheter indications for use statements for information on target patient population specific to the catheter being used. Refer to the Intended Use statement for a complete list of measured and derived parameters available for each patient population.

  • HemoSphere Advanced Monitor with HemoSphere Oximetry Cable: The HemoSphere Advanced Monitor when used with the HemoSphere Oximetry Cable and Edwards oximetry catheters is indicated for use in adult and pediatric critical care patients requiring monitoring of venous oxygen saturation (SvO2 and ScvO2) and derived hemodynamic parameters in a hospital environment. Refer to the Edwards oximetry catheter indications for use statement for information on target patient population specific to the catheter being used. Refer to the Intended Use statement for a complete list of measured and derived parameters available for each patient population.

  • HemoSphere Advanced Monitor with HemoSphere Pressure Cable: The HemoSphere advanced monitor when used with the HemoSphere pressure cable is indicated for use in adult and pediatric critical care patients in which the balance between cardiac function, fluid status, vascular resistance and pressure needs continuous assessment. It may be used for monitoring of hemodynamic parameters in conjunction with a perioperative goal directed therapy protocol in a hospital environment. Refer to the Edwards FloTrac sensor, FloTrac Jr sensor, Acumen IQ sensor, and TruWave disposable pressure transducer indications for use statements for information on target patient populations specific to the sensor/transducer being used. The Edwards Acumen Hypotension Prediction Index software feature provides the clinician with physiological insight into a patient's likelihood of future hypotensive events and the associated hemodynamics. The Acumen HPI feature is intended for use in surgical or non-surgical patients receiving advanced hemodynamic monitoring. The Acumen HPI feature is considered to be additional quantitative information regarding the patient's physiological condition for reference only and no therapeutic decisions should be made based solely on the Acumen Hypotension Prediction Index (HPI) parameter. Refer to the Intended Use statement for a complete list of measured and derived parameters available for each patient population.

  • HemoSphere Advanced Monitor with Acumen Assisted Fluid Management Feature and Acumen IQ Sensor: The Acumen Assisted Fluid Management (AFM) software feature provides the clinician with physiological insight into a patient's estimated response to fluid therapy and the associated hemodynamics. The Acumen AFM software feature is intended for use in surgical patients >=18 years of age, that require advanced hemodynamic monitoring. The Acumen AFM software feature offers suggestions regarding the patient's physiological condition and estimated response to fluid therapy. Acumen AFM fluid administration suggestions are offered to the clinician; the decision to administer a fluid bolus is made by the clinician, based upon review of the patient's hemodynamics. No therapeutic decisions should be made based solely on the Assisted Fluid Management suggestions. The Acumen Assisted Fluid Management software feature may be used with the Acumen AFM Cable and Acumen IQ fluid meter.

  • HemoSphere Advanced Monitor with HemoSphere Technology Module and ForeSight Oximeter Cable: The non-invasive ForeSight oximeter cable is intended for use as an adjunct monitor of absolute regional hemoglobin oxygen saturation of blood under the sensors in individuals at risk for reduced-flow or no flow ischemic states. The ForeSight Oximeter Cable is also intended to monitor relative changes of total hemoglobin of blood under the sensors. The ForeSight Oximeter Cable is intended to allow for the display of StO2 and relative change in total hemoglobin on the HemoSphere advanced monitor.

    • When used with large sensors, the ForeSight Oximeter Cable is indicated for use on adults and transitional adolescents >=40 kg.
    • When used with medium sensors, the ForeSight Oximeter Cable is indicated for use on pediatric subjects >=3 kg.
    • When used with small sensors, the ForeSight Oximeter Cable is indicated for cerebral use on pediatric subjects <8 kg and non-cerebral use on pediatric subjects <5kg.

    Refer to the Intended Use statement for a complete list of measured and derived parameters available for each patient population.

  • HemoSphere Advanced Monitor with HemoSphere ClearSight Module: The HemoSphere advanced monitor when used with the HemoSphere ClearSight module, pressure controller or Smart Pressure Controller and a compatible Edwards finger cuff are indicated for patients over 18 years of age in which the balance between cardiac function, fluid status and vascular resistance needs continuous assessment. It may be used for monitoring hemodynamic parameters in conjunction with a perioperative goal directed therapy protocol in a hospital environment. In addition, the noninvasive system is indicated for use in patients with comorbidities for which hemodynamic optimization is desired and invasive measurements are difficult. The HemoSphere advanced monitor and compatible Edwards finger cuffs noninvasively measures blood pressure and associated hemodynamic parameters. The Edwards Lifesciences Acumen Hypotension Prediction Index feature provides the clinician with physiological insight into a patient's likelihood of future hypotensive events and the associated hemodynamics. The Acumen HPI feature is intended for use in surgical or non-surgical patients receiving advanced hemodynamic monitoring. The Acumen HPI feature is considered to be additional quantitative information regarding the patient's physiological condition for reference only and no therapeutic decisions should be made based solely on the Hypotension Prediction Index (HPI) parameter.

  • Indication for Acumen IQ Plus and VitaWave Plus finger cuffs: The Acumen IQ Plus and VitaWave Plus finger cuff adult indicated for patients over 18 years of age to continuously blood pressure and associated hemodynamic parameters when used with a compatible Edwards monitoring platform.

  • Smart Pressure Controller: The Smart Pressure Controller is intended for use with an Edwards compatible noninvasive monitoring system - composed of compatible monitor, pressure source (pump), compatible Edwards finger cuff(s) and pressure controller - for continuous noninvasive measurement of blood pressure and associated hemodynamic parameters. Refer to the operator's manual of the compatible Edwards monitor being used for specific information on the intended use environment and patient population.

  • Intended Use: The HemoSphere advanced monitoring platform is intended to be used by qualified personnel or trained clinicians in a critical care environment in a hospital setting. The Viewfinder remote mobile application can be used for supplemental near real-time remote display of monitored hemodynamic parameter data as well as faults, alerts and notifications generated by the HemoSphere advanced monitoring platform. The HemoSphere advanced monitoring platform is intended for use with compatible Edwards Swan-Ganz and oximetry catheters, Swan-Ganz Jr catheters, FloTrac sensors, FloTrac Jr sensors, Acumen IQ sensors, TruWave disposable pressure transducers, ForeSight/ForeSight Jr sensors, Acumen IQ fluid meter, and ClearSight/ClearSight Jr/Acumen IQ/Acumen IQ Plus/VitaWave/VitaWave Plus finger cuffs

Device Description

The HemoSphere Advanced Monitor was designed to simplify the customer experience by providing one platform with modular solutions for all hemodynamic monitoring needs. The user can choose from available optional sub-system modules or use multiple sub-system modules at the same time. This modular approach provides the customer with the choice of purchasing and/or using specific monitoring applications based on their needs. Users are not required to have all of the modules installed at the same time for the platform to function.

AI/ML Overview

The provided FDA 510(k) clearance letter and summary for the Edwards Lifesciences HemoSphere Advanced Monitor (HEM1) and associated components outlines the device's indications for use and the testing performed to demonstrate substantial equivalence to predicate devices. However, it does not contain the detailed acceptance criteria or the specific study results (performance data) in the format typically required to answer your request fully, especially for acceptance criteria and performance of an AI/algorithm-based feature like the Hypotension Prediction Index (HPI) or Assisted Fluid Management (AFM).

The document states:

  • "Completion of all verification and validation activities demonstrated that the subject devices meet their predetermined design and performance specifications."
  • "Measured and derived parameters were tested using a bench simulation. Additionally, system integration and mechanical testing was successfully conducted to verify the safety and effectiveness of the device. All tests passed."
  • "Software verification testing was conducted, and documentation was provided per FDA's Guidance for Industry and FDA Staff, "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices". All tests passed."

This indicates that internal performance specifications were met, but the specific metrics, thresholds, and study designs for achieving those specifications are not detailed in this public summary.

Therefore, I cannot populate the table with specific numerical performance data against acceptance criteria for the HPI or AFM features, nor can I provide details on sample size, expert ground truth establishment, or MRMC studies, as this information is not present in the provided text.

The text primarily focuses on:

  • Substantial equivalence to predicate devices.
  • Indications for Use for various HemoSphere configurations and modules.
  • Description of software and hardware modifications (e.g., integration of HPI algorithm, new finger cuffs).
  • General categories of testing performed (Usability, System Verification, Electrical Safety/EMC, Software Verification) with a blanket statement that "All tests passed."

Based on the provided document, here's what can and cannot be stated:


1. A table of acceptance criteria and the reported device performance

Cannot be provided with specific numerical data or thresholds from the given text. The document only states that "all verification and validation activities demonstrated that the subject devices meet their predetermined design and performance specifications." No specific acceptance criteria values (e.g., "Accuracy > X%", "Sensitivity > Y%", "Mean Absolute Error < Z") or reported performance values are publicly disclosed in this summary for any parameter, including HPI or AFM. For measured and derived parameters (like CO, MAP, etc.), it states they were tested using bench simulation, and "All tests passed," implying they met internal accuracy specifications for physical measurements, but these are not detailed.

2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

Cannot be provided from the given text. The document mentions "bench simulation" for measured and derived parameters, but does not provide sample sizes for these, or the type/provenance of data for testing the HPI or AFM algorithms.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

Cannot be provided from the given text. The document doesn't describe the process of establishing ground truth for the algorithms, nor does it mention the number or qualifications of experts involved in such a process.

4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

Cannot be provided from the given text. There is no mention of adjudication methods for any test sets.

5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

Cannot be provided from the given text. The document does not describe any MRMC studies or human-in-the-loop performance evaluation regarding the HPI or AFM features. The HPI and AFM features are described as providing "physiological insight" and "suggestions," not as tools requiring reader interpretation in a comparative effectiveness study as typically seen with imaging AI.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

Likely yes, based on the nature of the algorithms, but no specific performance metrics are provided. The HPI and AFM features are stated to provide "quantitative information" and "suggestions." The text indicates "System Verification (Non-Clinical Performance)" and "Software Verification" were performed, suggesting standalone evaluation against internal specifications, but no detailed results are provided. The HPI algorithm itself was "previously cleared in K230057," implying its standalone performance would have been evaluated during that prior clearance, but those details are not in this document.

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

Cannot be definitively stated from the given text. For the HPI feature, which predicts future hypotensive events, ground truth would typically involve actual patient outcomes (e.g., observed hypotensive events). For AFM, which suggests response to fluid therapy, ground truth might involve observed physiological responses to fluid boluses. However, the document does not specify how these ground truths were established for the purpose of testing the algorithms.

8. The sample size for the training set

Cannot be provided from the given text. The document does not mention details about the training data for the algorithms.

9. How the ground truth for the training set was established

Cannot be provided from the given text. The document does not mention details about the training data or its ground truth establishment.


Summary of Device Features Mentioned in Relation to Performance/Testing (General):

  • HemoSphere Advanced Monitor and various modules/accessories: The document primarily describes this as a monitoring platform for various hemodynamic parameters (CO, SvO2, MAP, etc.). Performance for these measured and derived parameters was tested via "bench simulation," and "All tests passed," implying they met internal benchmarks for accuracy and reliability.
  • Acumen Hypotension Prediction Index (HPI) software feature: This feature provides "physiological insight into a patient's likelihood of future hypotensive events." It was integrated from a previously cleared device (K230057). The document states "There are no changes to the Acumen HPI algorithm from what was cleared in K230057." This implies that the acceptance criteria and supporting studies for the HPI algorithm itself would be found in the K230057 clearance documentation, not typically resubmitted in detail for integration into another platform unless the integration process significantly altered its functionality or intended use.
  • Acumen Assisted Fluid Management (AFM) software feature: This feature provides "physiological insight into a patient's estimated response to fluid therapy" and "suggestions." It also mentions "Acumen AFM fluid administration suggestions are offered to the clinician; the decision to administer a fluid bolus is made by the clinician, based upon review of the patient's hemodynamics. No therapeutic decisions should be made based solely on the Assisted Fluid Management suggestions." This language suggests it's a supportive, advisory tool, rather than a diagnostic one requiring strict accuracy metrics in the same way. No performance specifics for AFM are given.
  • Usability Study: Conducted to ensure primary operating functions and critical tasks can be performed without patient or user harm. Determined that "intended users can perform primary operating functions and critical tasks of the system without any usability issues that may lead to patient or user harm." This is an acceptance criterion for human factors, but not for algorithmic performance.
  • Electrical Safety and EMC, Software Verification: All tests passed. These are general product safety and quality criteria, not specific to the performance of the predictive algorithms.

To obtain the detailed performance data, acceptance criteria, sample sizes, and ground truth information for the HPI or AFM algorithms, one would typically need to refer to the original 510(k) submission for the HPI algorithm (K230057) and potentially separate documentation for the AFM feature, which are not included in this general clearance letter for the HemoSphere platform update.

FDA 510(k) Clearance Letter - K243781

Page 1

U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov

Doc ID # 04017.08.00

July 23, 2025

Edwards Lifesciences
Varad Raghuwanshi
Director, Regulatory Affairs
One Edwards Way
Irvine, California 92614

Re: K243781

Trade/Device Name: HemoSphere Advanced Monitor (HEM1); HemoSphere Technology Module (HEMTOM10); HemoSphere ClearSight Module (HEMCSM10); Smart Pressure Controller (PC1Q); Acumen IQ Plus Finger Cuff (AIQCA2); HemoSphere Pressure Cable (HEMPSC100); HemoSphere Vita Monitor (HEMVITA1); HemoSphere Vita Technology Module (HEMVTOM1); HemoSphere VitaWave module (HEMVWM1); VitaWave Plus finger cuff (VWCA2)

Regulation Number: 21 CFR 870.1425
Regulation Name: Programmable Diagnostic Computer
Regulatory Class: Class II
Product Code: DQK, DQE, QAQ, MUD, DXN, DSB, FLL, QMS
Dated: December 7, 2024
Received: December 9, 2024

Dear Varad Raghuwanshi:

We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of

Page 2

K243781 - Varad Raghuwanshi Page 2

Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).

Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.

All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the

Page 3

K243781 - Varad Raghuwanshi Page 3

Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Stephen C. Browning -S

LCDR Stephen Browning
Assistant Director
Division of Cardiac Electrophysiology,
Diagnostics, and Monitoring Devices
Office of Cardiovascular Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health

Enclosure

Page 4

FORM FDA 3881 (8/23) Page 1 of 3

DEPARTMENT OF HEALTH AND HUMAN SERVICES
Food and Drug Administration

Indications for Use

Form Approved: OMB No. 0910-0120
Expiration Date: 07/31/2026
See PRA Statement below.

510(k) Number (if known): K243781

Device Name: HemoSphere Advanced Monitor (HEM1); HemoSphere Technology Module (HEMTOM10); HemoSphere ClearSight Module (HEMCSM10);Smart Pressure Controller (PC1Q); HemoSphere Pressure Cable (HEMPSC100); Acumen IQ Plus Finger Cuff (AIQCA2); HemoSphere Vita Technology Module (HEMVTOM1);HemoSphere VitaWave module (HEMVWM1);VitaWave Plus fin

Indications for Use (Describe)

HemoSphere Advanced Monitor with HemoSphere Swan-Ganz Module

The HemoSphere advanced monitor when used with the HemoSphere Swan-Ganz module and Edwards Swan-Ganz catheters is indicated for use in adult and pediatric critical care patients requiring monitoring of cardiac output (continuous [CO] and intermittent [iCO]) and derived hemodynamic parameters in a hospital environment. Pulmonary artery blood temperature monitoring is used to compute continuous and intermittent CO with thermodilution technologies. It may also be used for monitoring hemodynamic parameters in conjunction with a perioperative goal directed therapy protocol in a hospital environment. Refer to the Edwards Swan-Ganz catheter and Swan-Ganz Jr catheter indications for use statements for information on target patient population specific to the catheter being used. Refer to the Intended Use statement for a complete list of measured and derived parameters available for each patient population.

HemoSphere Advanced Monitor with HemoSphere Oximetry Cable

The HemoSphere Advanced Monitor when used with the HemoSphere Oximetry Cable and Edwards oximetry catheters is indicated for use in adult and pediatric critical care patients requiring monitoring of venous oxygen saturation (SvO2 and ScvO2) and derived hemodynamic parameters in a hospital environment. Refer to the Edwards oximetry catheter indications for use statement for information on target patient population specific to the catheter being used.

Refer to the Intended Use statement for a complete list of measured and derived parameters available for each patient population.

HemoSphere Advanced Monitor with HemoSphere Pressure Cable

The HemoSphere advanced monitor when used with the HemoSphere pressure cable is indicated for use in adult and pediatric critical care patients in which the balance between cardiac function, fluid status, vascular resistance and pressure needs continuous assessment. It may be used for monitoring of hemodynamic parameters in conjunction with a perioperative goal directed therapy protocol in a hospital environment. Refer to the Edwards FloTrac sensor, FloTrac Jr sensor, Acumen IQ sensor, and TruWave disposable pressure transducer indications for use statements for information on target patient populations specific to the sensor/transducer being used.

The Edwards Acumen Hypotension Prediction Index software feature provides the clinician with physiological insight into a patient's likelihood of future hypotensive events and the associated hemodynamics. The Acumen HPI feature is intended for use in surgical or non-surgical patients receiving advanced hemodynamic monitoring. The Acumen HPI feature is considered to be additional quantitative information regarding the patient's physiological condition for reference only and no therapeutic decisions should be made based solely on the Acumen Hypotension Prediction Index (HPI) parameter.

Refer to the Intended Use statement for a complete list of measured and derived parameters available for each patient population.

Page 5

FORM FDA 3881 (8/23) Page 2 of 3

HemoSphere Advanced Monitor with Acumen Assisted Fluid Management Feature and Acumen IQ Sensor:

The Acumen Assisted Fluid Management (AFM) software feature provides the clinician with physiological insight into a patient's estimated response to fluid therapy and the associated hemodynamics. The Acumen AFM software feature is intended for use in surgical patients ≥18 years of age, that require advanced hemodynamic monitoring. The Acumen AFM software feature offers suggestions regarding the patient's physiological condition and estimated response to fluid therapy. Acumen AFM fluid administration suggestions are offered to the clinician; the decision to administer a fluid bolus is made by the clinician, based upon review of the patient's hemodynamics. No therapeutic decisions should be made based solely on the Assisted Fluid Management suggestions.

The Acumen Assisted Fluid Management software feature may be used with the Acumen AFM Cable and Acumen IQ fluid meter.

HemoSphere Advanced Monitor with HemoSphere Technology Module and ForeSight Oximeter Cable

The non-invasive ForeSight oximeter cable is intended for use as an adjunct monitor of absolute regional hemoglobin oxygen saturation of blood under the sensors in individuals at risk for reduced-flow or no flow ischemic states. The ForeSight Oximeter Cable is also intended to monitor relative changes of total hemoglobin of blood under the sensors. The ForeSight Oximeter Cable is intended to allow for the display of StO2 and relative change in total hemoglobin on the HemoSphere advanced monitor.

  • When used with large sensors, the ForeSight Oximeter Cable is indicated for use on adults and transitional adolescents ≥40 kg.
  • When used with medium sensors, the ForeSight Oximeter Cable is indicated for use on pediatric subjects ≥3 kg.
  • When used with small sensors, the ForeSight Oximeter Cable is indicated for cerebral use on pediatric subjects <8 kg and non-cerebral use on pediatric subjects <5kg.

Refer to the Intended Use statement for a complete list of measured and derived parameters available for each patient population.

HemoSphere Advanced Monitor with HemoSphere ClearSight Module

The HemoSphere advanced monitor when used with the HemoSphere ClearSight module, pressure controller or Smart Pressure Controller and a compatible Edwards finger cuff are indicated for patients over 18 years of age in which the balance between cardiac function, fluid status and vascular resistance needs continuous assessment. It may be used for monitoring hemodynamic parameters in conjunction with a perioperative goal directed therapy protocol in a hospital environment. In addition, the noninvasive system is indicated for use in patients with comorbidities for which hemodynamic optimization is desired and invasive measurements are difficult. The HemoSphere advanced monitor and compatible Edwards finger cuffs noninvasively measures blood pressure and associated hemodynamic parameters.

The Edwards Lifesciences Acumen Hypotension Prediction Index feature provides the clinician with physiological insight into a patient's likelihood of future hypotensive events and the associated hemodynamics. The Acumen HPI feature is intended for use in surgical or non-surgical patients receiving advanced hemodynamic monitoring. The Acumen HPI feature is considered to be additional quantitative information regarding the patient's physiological condition for reference only and no therapeutic decisions should be made based solely on the Hypotension Prediction Index (HPI) parameter.

Indication for Acumen IQ Plus and VitaWave Plus finger cuffs:

The Acumen IQ Plus and VitaWave Plus finger cuff adult indicated for patients over 18 years of age to continuously blood pressure and associated hemodynamic parameters when used with a compatible Edwards monitoring platform.

Smart Pressure Controller:

Page 6

FORM FDA 3881 (8/23) Page 3 of 3

The Smart Pressure Controller is intended for use with an Edwards compatible noninvasive monitoring system - composed of compatible monitor, pressure source (pump), compatible Edwards finger cuff(s) and pressure controller - for continuous noninvasive measurement of blood pressure and associated hemodynamic parameters.

Refer to the operator's manual of the compatible Edwards monitor being used for specific information on the intended use environment and patient population.

Type of Use (Select one or both, as applicable)

☒ Prescription Use (Part 21 CFR 801 Subpart D) ☐ Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services
Food and Drug Administration
Office of Chief Information Officer
Paperwork Reduction Act (PRA) Staff
PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

Page 7

510(k) Summary – HemoSphere Alta Advanced Monitoring Platform

I. Submitter:

Sponsor: Edwards Lifesciences LLC
One Edwards Way
Irvine, CA 92614

Establishment Registration Number: 2015691

Contact Person: Varad Raghuwanshi
Director, Regulatory Affairs
One Edwards Way
Irvine, CA 92614
Telephone: (213) 479-4688

Date Prepared: July 22, 2025

II. Device Information:

Platform Name (Name of the Device): HemoSphere Advanced Monitoring Platform

Trade Name:

  • HemoSphere Advanced Monitor- HEM1
  • HemoSphere Vita Monitor- HEMVITA1
  • HemoSphere ClearSight Module- HEMCSM10
  • HemoSphere VitaWave Module- HEMVWM1
  • HemoSphere Technology Module- HEMTOM10
  • HemoSphere Vita Technology Module- HEMVTOM1
  • HemoSphere Pressure Cable- HEMPSC100
  • Smart Pressure Controller- PC1Q
  • Acumen IQ Plus Finger cuff- AIQCA2
  • VitaWave Plus Finger cuff- VWCA2

Common Name: Programmable Diagnostic Computer, Fiber Optic Oximeter Catheter, Adjunctive predictive Cardiovascular Indicator, Medium-Term Adjunctive Predictive Cardiovascular Indicator, Adjunctive Open Loop Fluid Therapy Recommender, and Noninvasive blood pressure measurement system

Classification Name: Programmable Diagnostic Computer 21 CFR 870.1425
Fiberoptic Oximeter Catheter 21 CFR 870.1230

Page 8

Adjunctive Predictive Cardiovascular Indicator 21 CFR 870.2210
Oximeter, Tissue Saturation 21 CFR 870.2700
Noninvasive blood pressure measurement system 21 CFR 870.1130
Impedance plethysmograph 21 CFR 870.2770
Thermometer, Electronic, Clinical 21 CFR 880.2910
Adjunctive Open Loop Fluid Therapy Recommender 21 CFR 870.5600

Product Code and Regulatory Class:

  • DQK, Class II
  • DQE, Class II
  • QAQ, Class II
  • MUD, Class II
  • DXN, Class II
  • DSB, Class II
  • FLL, Class II
  • QMS, Class II

III. Predicate Device

Primary Predicate Device:

  • HemoSphere Advanced Monitoring Platform manufactured by Edwards Lifesciences, K223865, cleared June 9th, 2023. It is the base device on which the various hardware and software modifications have been implemented. It has been utilized for substantial equivalence in terms of the graphical user interface (GUI) used, indication for use, intended use, technological characteristics, accessories, basic device functionality, new configuration Pressure controller (PC1Q).

Additional Predicate Devices:

  • HemoSphere Alta Advanced Monitoring Platform (K242451, Cleared Dec 9th, 2024) is being utilized as a secondary predicate for Fluid Meter Only Mode, modification to provide an option to use the existing pulse rate instead of heart rate input for calculating associated parameters, and additional miscellaneous GUI updates for user convenience such as HPI Yellow display and Cardiac Power Output (CPO) as a key parameter. It has been utilized for substantial equivalence in terms of the intended use, and graphical user interface (GUI) used.

Page 9

  • Acumen IQ finger cuff (AIQCA) by Edwards Lifesciences, K230919, cleared October 24th, 2023, is being utilized for the Acumen IQ Plus finger cuff (AIQCA2). It has been utilized for substantial equivalence in terms of the indication for use, intended use, and technological characteristics

  • Acumen Hypotension Prediction Index (HPI) algorithm by Edwards Lifesciences, K230057, cleared on June 8th, 2023, utilized for the HPI Adjustable MAP (Hypotension threshold) feature.

IV. Device Description

Device Description:

The HemoSphere Advanced Monitor was designed to simplify the customer experience by providing one platform with modular solutions for all hemodynamic monitoring needs. The user can choose from available optional sub-system modules or use multiple sub-system modules at the same time. This modular approach provides the customer with the choice of purchasing and/or using specific monitoring applications based on their needs. Users are not required to have all of the modules installed at the same time for the platform to function.

V. Indications for Use:

HemoSphere Advanced Monitor with HemoSphere Swan-Ganz Module

The HemoSphere advanced monitor when used with the HemoSphere Swan-Ganz module and Edwards Swan-Ganz catheters is indicated for use in adult and pediatric critical care patients requiring monitoring of cardiac output (continuous [CO] and intermittent [iCO]) and derived hemodynamic parameters in a hospital environment. Pulmonary artery blood temperature monitoring is used to compute continuous and intermittent CO with thermodilution technologies. It may also be used for monitoring hemodynamic parameters in conjunction with a perioperative goal directed therapy protocol in a hospital environment. Refer to the Edwards Swan-Ganz catheter and Swan-Ganz Jr catheter indications for use statements for information on target patient population specific to the catheter being used. Refer to the Intended Use statement for a complete list of measured and derived parameters available for each patient population.

Page 10

HemoSphere Advanced Monitor with HemoSphere Oximetry Cable

The HemoSphere Advanced Monitor when used with the HemoSphere Oximetry Cable and Edwards oximetry catheters is indicated for use in adult and pediatric critical care patients requiring monitoring of venous oxygen saturation (SvO2 and ScvO2) and derived hemodynamic parameters in a hospital environment. Refer to the Edwards oximetry catheter indications for use statement for information on target patient population specific to the catheter being used.

Refer to the Intended Use statement for a complete list of measured and derived parameters available for each patient population.

HemoSphere Advanced Monitor with HemoSphere Pressure Cable

The HemoSphere advanced monitor when used with the HemoSphere pressure cable is indicated for use in adult and pediatric critical care patients in which the balance between cardiac function, fluid status, vascular resistance and pressure needs continuous assessment. It may be used for monitoring of hemodynamic parameters in conjunction with a perioperative goal directed therapy protocol in a hospital environment. Refer to the Edwards FloTrac sensor, FloTrac Jr sensor, Acumen IQ sensor, and TruWave disposable pressure transducer indications for use statements for information on target patient populations specific to the sensor/transducer being used.

The Edwards Acumen Hypotension Prediction Index software feature provides the clinician with physiological insight into a patient's likelihood of future hypotensive events and the associated hemodynamics. The Acumen HPI feature is intended for use in surgical or non-surgical patients receiving advanced hemodynamic monitoring. The Acumen HPI feature is considered to be additional quantitative information regarding the patient's physiological condition for reference only and no therapeutic decisions should be made based solely on the Acumen Hypotension Prediction Index (HPI) parameter.

Refer to the Intended Use statement for a complete list of measured and derived parameters available for each patient population.

HemoSphere Advanced Monitor with Acumen Assisted Fluid Management Feature and Acumen IQ Sensor:

The Acumen Assisted Fluid Management (AFM) software feature provides the clinician with physiological insight into a patient's estimated response

Page 11

to fluid therapy and the associated hemodynamics. The Acumen AFM software feature is intended for use in surgical patients ≥18 years of age, that require advanced hemodynamic monitoring. The Acumen AFM software feature offers suggestions regarding the patient's physiological condition and estimated response to fluid therapy. Acumen AFM fluid administration suggestions are offered to the clinician; the decision to administer a fluid bolus is made by the clinician, based upon review of the patient's hemodynamics. No therapeutic decisions should be made based solely on the Assisted Fluid Management suggestions.

The Acumen Assisted Fluid Management software feature may be used with the Acumen AFM Cable and Acumen IQ fluid meter.

HemoSphere Advanced Monitor with HemoSphere Technology Module and ForeSight Oximeter Cable

The non-invasive ForeSight oximeter cable is intended for use as an adjunct monitor of absolute regional hemoglobin oxygen saturation of blood under the sensors in individuals at risk for reduced-flow or no flow ischemic states. The ForeSight Oximeter Cable is also intended to monitor relative changes of total hemoglobin of blood under the sensors. The ForeSight Oximeter Cable is intended to allow for the display of StO2 and relative change in total hemoglobin on the HemoSphere advanced monitor.

  • When used with large sensors, the ForeSight Oximeter Cable is indicated for use on adults and transitional adolescents ≥40 kg.
  • When used with medium sensors, the ForeSight Oximeter Cable is indicated for use on pediatric subjects ≥3 kg.
  • When used with small sensors, the ForeSight Oximeter Cable is indicated for cerebral use on pediatric subjects <8 kg and non-cerebral use on pediatric subjects <5kg.

Refer to the Intended Use statement for a complete list of measured and derived parameters available for each patient population.

HemoSphere Advanced Monitor with HemoSphere ClearSight Module

The HemoSphere advanced monitor when used with the HemoSphere ClearSight module, pressure controller or Smart Pressure Controller and a compatible Edwards finger cuff are indicated for patients over 18 years of age in which the balance between cardiac function, fluid status and vascular resistance needs continuous assessment. It may be used for monitoring hemodynamic parameters in conjunction with a perioperative goal directed therapy protocol in a hospital environment. In addition, the

Page 12

noninvasive system is indicated for use in patients with comorbidities for which hemodynamic optimization is desired and invasive measurements are difficult. The HemoSphere advanced monitor and compatible Edwards finger cuffs noninvasively measures blood pressure and associated hemodynamic parameters.

The Edwards Lifesciences Acumen Hypotension Prediction Index feature provides the clinician with physiological insight into a patient's likelihood of future hypotensive events and the associated hemodynamics. The Acumen HPI feature is intended for use in surgical or non-surgical patients receiving advanced hemodynamic monitoring. The Acumen HPI feature is considered to be additional quantitative information regarding the patient's physiological condition for reference only and no therapeutic decisions should be made based solely on the Hypotension Prediction Index (HPI) parameter.

Indication for Acumen IQ Plus and VitaWave Plus finger cuffs:

The Acumen IQ Plus and VitaWave Plus finger cuff adult indicated for patients over 18 years of age to continuously blood pressure and associated hemodynamic parameters when used with a compatible Edwards monitoring platform.

Smart Pressure Controller:

The Smart Pressure Controller is intended for use with an Edwards compatible noninvasive monitoring system - composed of compatible monitor, pressure source (pump), compatible Edwards finger cuff(s) and pressure controller - for continuous noninvasive measurement of blood pressure and associated hemodynamic parameters.

Refer to the operator's manual of the compatible Edwards monitor being used for specific information on the intended use environment and patient population.

Page 13

Intended Use:

The HemoSphere advanced monitoring platform is intended to be used by qualified personnel or trained clinicians in a critical care environment in a hospital setting.

The Viewfinder remote mobile application can be used for supplemental near real-time remote display of monitored hemodynamic parameter data as well as faults, alerts and notifications generated by the HemoSphere advanced monitoring platform.

The HemoSphere advanced monitoring platform is intended for use with compatible Edwards Swan-Ganz and oximetry catheters, Swan-Ganz Jr catheters, FloTrac sensors, FloTrac Jr sensors, Acumen IQ sensors, TruWave disposable pressure transducers, ForeSight/ForeSight Jr sensors, Acumen IQ fluid meter, and ClearSight/ClearSight Jr/Acumen IQ/Acumen IQ Plus/VitaWave/VitaWave Plus finger cuffs

A comprehensive list of parameters available while monitoring with the HemoSphere advanced monitor and a connected HemoSphere Swan-Ganz module are listed below. Only iCO, iCI, iSVR, and iSVRI are available to the pediatric patient population.

ParameterDescriptionSub-system technology usedPatient PopulationHospital Environment
COcontinuous cardiac outputHemoSphere Swan-Ganz moduleAdult onlyOperating Room, Intensive Care Unit, Emergency Room
sCOSTAT cardiac output
CPOCardiac Power Output
CIcontinuous cardiac index
sCISTAT cardiac index
EDVright ventricular end diastolic volume
sEDVSTAT right ventricular end diastolic volume
EDVIright ventricular end diastolic volume index
sEDVISTAT right ventricular end diastolic volume index
HRavgaveraged heart rate
LVSWIleft ventricular stroke work index
PVRpulmonary vascular resistance
PVRIpulmonary vascular resistance index
RVEFright ventricular ejection fraction
sRVEFSTAT right ventricular ejection fraction
RVSWIright ventricular stroke work index
SVstroke volume
SVIstroke volume index
SVRsystemic vascular resistance

Page 14

SVRIsystemic vascular resistance indexAdult and Pediatric
iCOintermittent cardiac output
iCIintermittent cardiac index
iSVRintermittent systemic vascular resistance
iSVRIintermittent systemic vascular resistance index

A comprehensive list of parameters available for adult and pediatric patient populations while monitoring with the HemoSphere advanced monitor and a connected HemoSphere oximetry cable are listed below:

ParameterDescriptionSub-system technology usedPatient PopulationHospital Environment
SvO2Mixed Venous Oxygen SaturationHemoSphere oximetry cableAdult and PediatricOperating Room, Intensive Care Unit, Emergency Room
ScvO2Central Venous Oxygen Saturation

A comprehensive list of parameters available for adult and pediatric patient populations while monitoring with the HemoSphere advanced monitor and both a connected HemoSphere Swan-Ganz module and oximetry cable are listed below:

ParameterDescriptionSub-system technology usedPatient PopulationHospital Environment
DO2Oxygen DeliveryHemoSphere Swan-Ganz module and HemoSphere oximetry cableAdult and PediatricOperating Room, Intensive Care Unit, Emergency Room
DO2IOxygen Delivery Indexed
VO2Oxygen Consumption
VO2eEstimated Oxygen Consumption when ScvO2 is being monitored
VO2IOxygen Consumption Index
VO2IeEstimated Oxygen Consumption Index when ScvO2 is being monitored

A comprehensive list of parameters available while monitoring with the HemoSphere advanced monitor and both a connected HemoSphere Swan-Ganz module and pressure cable are listed below:

ParameterDescriptionSub-system technology usedPatient PopulationHospital Environment
CO20s20-second cardiac outputHemoSphere Swan-Ganz module andAdult onlyOperating room, intensive care unit, emergency room
CI20s20-second cardiac index
SV20s20-second stroke volume

Page 15

SVI20s20-second stroke volume indexHemoSphere pressure cable

¹ 20-second flow parameters are only available if the 20s flow parameter feature is enabled. Please Contact your local Edwards representative for more information on enabling this advanced feature.

A comprehensive list of parameters available while monitoring with the HemoSphere advanced monitor and a connected HemoSphere pressure cable are listed below:

ParameterDescriptionSub-system technology usedPatient PopulationHospital Environment
COContinuous Cardiac Output¹HemoSphere pressure cableAdult and Pediatric ≥ 12 years of ageOperating Room, Intensive Care Unit, Emergency Room
CIContinuous Cardiac Index¹
CPOCardiac Power output
CPICardiac Power index
DIAARTSystemic arterial diastolic blood pressure
MAPMean Arterial Pressure
PPVpulse pressure variation¹
PRPulse rate
SVStroke Volume¹
SVIStroke Volume Index¹
SVRSystemic Vascular Resistance¹
SVRISystemic Vascular Resistance¹Index
SVVStroke Volume Variation¹
SYSARTSystolic Blood Pressure
CVPCentral Venous PressureAdult only
DIAPAPpulmonary artery diastolic blood pressure
dP/dtSystolic slope²
EadynDynamic Arterial Elastance²
HPIAcumen Hypotension Prediction Index
MPAPMean Pulmonary Arterial Pressure
SYSPAPSystolic pulmonary artery blood pressure

¹FloTrac parameters are available when using a FloTrac/Acumen IQ sensor and if the FloTrac feature is enabled.
²HPI parameters are available when using an Acumen IQ sensor and if the HPI feature is activated.

A list of Acumen Assisted Fluid Management (AFM) outputs available for surgical patients ≥18 years of age while monitoring with the HemoSphere advanced monitor and a connected HemoSphere pressure cable are listed below:

AFM outputSub-system technology usedPatient PopulationHospital Environment
Fluid Bolus Suggested

Page 16

Test Bolus SuggestedHemoSphere pressure cable≥18 years of age onlyOperating room
Fluid Not Suggested
Suggestions Suspended
Bolus In Progress...
Bolus Complete
Bolus Complete; Analyzing Hemodynamic Response
Tracked Case Vol.
Flow Rate
Bolus Volume

AFM outputs are available when using an Acumen IQ sensor and if the AFM feature is activated. Flow rate and Bolus Volume are visible when using automatic fluid tracking mode.

A comprehensive list of parameters available for adult and pediatric patient populations while monitoring with the HemoSphere advanced monitor and both a connected HemoSphere pressure cable and oximetry cable are listed below:

ParameterDescriptionSub-system technology usedPatient PopulationHospital Environment
DO2Oxygen DeliveryHemoSphere pressure cable and HemoSphere oximetry cableAdult onlyOperating Room, Intensive Care Unit, Emergency Room
DO2IOxygen Delivery Indexed
VO2Oxygen Consumption
VO2eEstimated Oxygen Consumption when ScvO2 is being monitored
VO2IOxygen Consumption Index
VO2IeEstimated Oxygen Consumption Index when ScvO2 is being monitored

Tissue oxygen saturation, StO2, can be monitored with the HemoSphere advanced monitor, a connected HemoSphere technology module, and the ForeSight oximeter cable as listed below.

ParameterDescriptionSub-system technology usedPatient PopulationHospital Environment
StO2Tissue oxygen saturationForeSight oximeter cable and HemoSphere technology moduleAdult and PediatricOperating Room, Intensive Care Unit, Emergency Room
ΔctHbrelative change in total hemoglobin

A comprehensive list of parameters available while monitoring with the HemoSphere advanced monitor and a connected HemoSphere ClearSight module are listed below:

ParameterDescriptionSub-system technology usedPatient PopulationHospital Environment
COContinuous Cardiac OutputHemoSphereAdult andOperating

Page 17

CPOCardiac Power OutputClearSight modulepediatric ≥ 12Room, Intensive Care Unit, Emergency Room
CIContinuous Cardiac Index
DIAARTarterial diastolic blood pressure
SYSARTSystolic Blood Pressure
MAPMean Arterial Pressure
PPVpulse pressure variation
PRPulse rate
SVStroke Volume
SVIStroke Volume Index
SVRSystemic Vascular Resistance
SVRISystemic Vascular Resistance Index
SVVStroke Volume Variation
dP/dtsystolic slopeAdult only
EadynDynamic Arterial Elastance¹
HPIAcumen Hypotension PredictionIndex¹

¹HPI parameters are available when using an Acumen IQ or Acumen IQ Plus finger cuff and heart reference sensor (HRS), if applicable, and if the HPI feature is activated. Activation is only available in certain areas. Please contact your local Edwards representative for more information on enabling this advanced feature.

NOTE: CO/CI and SV/SVI are measured using a reconstructed brachial arterial waveform. All other monitored parameters use a reconstructed radial arterial waveform. SVR/SVRI are derived from CO/CI and MAP along with an entered or monitored CVP value.

A comprehensive list of parameters available for adult patient populations while monitoring with the HemoSphere advanced monitor and both a connected HemoSphere ClearSight module and oximetry cable are listed below:

ParameterDescriptionSub-System Module UsedPatient PopulationHospital Environment
DO2Oxygen DeliveryHemoSphere ClearSight Module and HemoSphere Oximetry CableAdult onlyOperating Room, Intensive Care Unit
DO2IOxygen Delivery Indexed
VO2Oxygen Consumption
VO2eEstimated Oxygen Consumption when ScvO2 is being monitored
VO2IOxygen Consumption Index
VO2IeEstimated Oxygen Consumption Index when ScvO2 is being monitored

Page 18

CPOCardiac Power OutputClearSight modulepediatric ≥ 12Room, Intensive Care Unit, Emergency Room
CIContinuous Cardiac Index
DIAARTarterial diastolic blood pressure
SYSARTSystolic Blood Pressure
MAPMean Arterial Pressure
PPVpulse pressure variation
PRPulse rate
SVStroke Volume
SVIStroke Volume Index
SVRSystemic Vascular Resistance
SVRISystemic Vascular Resistance Index
SVVStroke Volume Variation
dP/dtsystolic slopeAdult only
EadynDynamic Arterial Elastance¹
HPIAcumen Hypotension PredictionIndex¹

¹HPI parameters are available when using an Acumen IQ or Acumen IQ Plus finger cuff and heart reference sensor (HRS), if applicable, and if the HPI feature is activated. Activation is only available in certain areas. Please contact your local Edwards representative for more information on enabling this advanced feature.

NOTE: CO/CI and SV/SVI are measured using a reconstructed brachial arterial waveform. All other monitored parameters use a reconstructed radial arterial waveform. SVR/SVRI are derived from CO/CI and MAP along with an entered or monitored CVP value.

A comprehensive list of parameters available for adult patient populations while monitoring with the HemoSphere advanced monitor and both a connected HemoSphere ClearSight module and oximetry cable are listed below:

ParameterDescriptionSub-System Module UsedPatient PopulationHospital Environment
DO2Oxygen DeliveryHemoSphere ClearSight Module and HemoSphere Oximetry CableAdult onlyOperating Room, Intensive Care Unit
DO2IOxygen Delivery Indexed
VO2Oxygen Consumption
VO2eEstimated Oxygen Consumption when ScvO2 is being monitored
VO2IOxygen Consumption Index
VO2IeEstimated Oxygen Consumption Index when ScvO2 is being monitored

Page 19

Software modifications to the subject HemoSphere Advanced Monitoring Platform (K223865, Cleared June 9th, 2023):

  • Software modification to the HemoSphere Advanced Monitor (HEM1): The existing software components of HemoSphere Advanced Monitor (HEM1-SHM and DBM) have been updated to support graphical user interface (GUI) associated with the new features introduced to the HemoSphere Advanced Monitoring Platform.

The GUI updates include the following features:

  • Fluid meter only Mode: For user convenience, the automatic bolus tracking feature can now also be used independently in "Fluid Meter Only" mode to track fluid administration. In this mode, users can track fluid delivery via an existing Acumen fluid meter instead of manually tracking and logging. During this mode, the AFM algorithm will be disabled, and the user will not get any bolus suggestions.

  • Modification to provide an option to use the existing pulse rate instead of heart rate input for calculating associated parameters: The HemoSphere Advanced Monitoring Platform currently requires a heart rate (HR) signal to be computed to calculate parameters such as EDV, EDVI, RVEF, sEDV, sEDVI, SV and SVI parameters. Modifications have been made to allow users to have the option to use HR or alternatively utilize existing pulse rate (PR)for the parameter calculation.

  • Miscellaneous Graphical User Interface (GUI) updates: The HemoSphere Advanced Monitoring Platform GUI has been updated to support the features listed below and to provide user convenience.

  • Integration of the existing Acumen Hypotension Prediction Index (HPI) Algorithm: The previously cleared HPI Adjustable MAP target algorithm that was cleared in K230057 on June 8, 2023, is being integrated to the HemoSphere Advanced Monitoring Platform. There are no changes to the Acumen HPI algorithm from what was cleared in K230057. The same feature is now integrated

Page 20

under the name "Hypotension threshold" into the HemoSphere Advanced Monitoring Platform.

The existing HemoSphere Pressure Cable (HEMPSC100) and HemoSphere ClearSight Module (HEMCSM10) software have been updated to support integration of the HPI Adjustable MAP target algorithm.

  • Software modification to the HemoSphere ClearSight/VitaWave Module (HEMCSM10/HEMVWM1): The existing HemoSphere ClearSight Module software has been updated to support a new configuration of the Pressure Controller.

Hardware modifications to the subject HemoSphere Advanced Monitoring Platform (original K223865, Cleared June 9th, 2023):

  • Smart Pressure Controller: A new hardware configuration of the existing accessory, Pressure controller (PC2/HEMPC2) most recently cleared in K223865, is being introduced to provide correction for the hydrostatic height difference between the finger and heart under the tradename "Smart Pressure Controller" (Model- PC1Q).

Hardware modifications to the subject Acumen IQ finger cuff, Model: AIQCA (K230919, Cleared October 24th, 2023):

  • Acumen IQ Plus finger cuff (AIQCA2): A new configuration of the Acumen IQ finger cuff is being introduced to provide continuous noninvasive measurement of blood pressure and hemodynamic monitoring under the tradename of Acumen IQ Plus finger cuff (AIQCA2) and VitaWave Plus finger cuff (VWCA2).

Labelling Modifications to the HemoSphere Advanced Monitoring Platform and Acumen IQ finger Cuff to address the device modifications

New labelling is being created for the updated Intended use and Indications for use for Acumen IQ/VitaWave Plus finger cuff.

Page 21

  • Indication update to include Acumen Hypotension Prediction Index (HPI) modification cleared in K230057 on June 8, 2023: The Acumen Hypotension Prediction Index (HPI) feature indication has been updated to include the Adjustable MAP targets and patient population to include non-surgical patients during non-invasive monitoring cleared in K230057 on June 8, 2023.

Performance Data:

The following verification activities were performed in support of a substantial equivalence determination.

Usability Study

Usability study was conducted per FDA's guidance document "Applying Human Factors and Usability Engineering to Medical Devices" to investigate primary operating functions and critical tasks of the system for any usability issues regarding the HemoSphere Advanced Monitoring Platform that may lead to patient or user harm.

The usability study demonstrated that the intended users can perform primary operating functions and critical tasks of the system without any usability issues that may lead to patient or user harm.

System Verification (Non-Clinical Performance):

Completion of all verification and validation activities demonstrated that the subject devices meet their predetermined design and performance specifications. Verification activities performed confirmed that the differences in the design and materials used did not adversely affect the safety and effectiveness of the subject device.

Measured and derived parameters were tested using a bench simulation. Additionally, system integration and mechanical testing was successfully conducted to verify the safety and effectiveness of the device. All tests passed.

Electrical Safety and Electromagnetic Compatibility (EMC)

Electrical safety and EMC testing were conducted on the subject HemoSphere Advanced Monitoring Platform, consisting of the HemoSphere Monitor, HemoSphere Vita Monitor, HemoSphere Swan-Ganz Module, HemoSphere Oximetry Cable, HemoSphere Pressure Cable, HemoSphere Technology Module, ForeSight Oximeter Cable, ClearSight Module, VitaWave Module, Pressure Controller, Heart Reference Sensor, Acumen AFM Cable, and Smart Pressure Controller. The system complies with the IEC 60601-1, IEC 60601-1-2, IEC 60601-1-6, IEC 60601-1-8,

Page 22

IEC 62304, IEC 62366-1, IEC 60601-2-34, IEC 60601-2-57, IEC 60601-2-49 and IEC 80601-2-49. All tests passed.

Software Verification

Software verification testing was conducted, and documentation was provided per FDA's Guidance for Industry and FDA Staff, "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices". All tests passed.

Conclusion

The technological characteristics of the subject and predicate devices are identical. The HemoSphere Advanced Monitoring Platform has successfully passed functional and performance testing, including software verification and validation, system integration, mechanical, electrical, human factor usability, and safety testing. The testing performed demonstrates that the HemoSphere Advanced Monitoring Platform is substantially equivalent to the legally marketed predicates.

§ 870.1425 Programmable diagnostic computer.

(a)
Identification. A programmable diagnostic computer is a device that can be programmed to compute various physiologic or blood flow parameters based on the output from one or more electrodes, transducers, or measuring devices; this device includes any associated commercially supplied programs.(b)
Classification. Class II (performance standards).