Search Filters

Search Results

Found 47 results

510(k) Data Aggregation

    K Number
    K192727
    Date Cleared
    2020-05-20

    (236 days)

    Product Code
    Regulation Number
    866.5775
    Reference & Predicate Devices
    N/A
    Why did this record match?
    510k Summary Text (Full-text Search) :

    K192727

    Trade/Device Name: K-ASSAY® RF (Ver.2), K-ASSAY® RF Calibrator (Ver.2) Regulation Number: 21 CFR 866.5775

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The K-ASSAY® RF (Ver.2) assay is for the quantitative determination of human IgG rheumatoid factor antibodies in patient serum or plasma (citric acid, EDTA, or lithium heparin) based on immunoturbidimetric assay. The presence of IgG RF antibodies, in conjunction with clinical findings and other laboratory tests, is an aid in the diagnosis of rheumatord arthritis (RA). FOR IN VITRO DIAGNOSTIC USE.

    The K-ASSAY® RF Calibrator (Ver.2) is intended to be used to calibrate the K-ASSAY® RF (Ver.2) immunoturbidimetric assay. FOR IN VITRO DIAGNOSTIC USE.

    Device Description

    Not Found

    AI/ML Overview

    The provided text is a 510(k) premarket notification letter from the FDA for a medical device called "K-ASSAY® RF (Ver.2), K-ASSAY® RF Calibrator (Ver.2)". This device is an in-vitro diagnostic test for quantitative determination of human IgG rheumatoid factor antibodies.

    The letter explicitly states that the device is a "Rheumatoid Factor Immunological Test System" (Regulation Name: 21 CFR 866.5775) and is a "Class II" device.

    Crucially, this document is an FDA clearance letter for a laboratory test (an in-vitro diagnostic device), not an AI/ML-based medical device that would have acceptance criteria based on performance metrics like sensitivity, specificity, or AUC, or involve human readers and their improvement with AI assistance.

    Therefore, most of the requested information regarding acceptance criteria, study design for AI/ML, human readers, ground truth establishment, etc., is not applicable to this type of device and will not be found in this document.

    The "acceptance criteria" for this type of device typically revolve around demonstrating substantial equivalence to a predicate device, and performance measures like precision, accuracy, linearity, measuring range, interference, and agreement studies for clinical performance. These are standard validation tests for IVD products, not AI performance metrics.

    In summary, based on the provided text, I cannot provide the requested information because it pertains to an AI/ML medical device, which the K-ASSAY® RF (Ver.2) is not.

    I can, however, extract the following relevant information from the document as it pertains to a medical device:

    1. Device Name: K-ASSAY® RF (Ver.2), K-ASSAY® RF Calibrator (Ver.2)
    2. Regulation Number: 21 CFR 866.5775
    3. Regulation Name: Rheumatoid Factor Immunological Test System
    4. Regulatory Class: Class II
    5. Indications for Use: The K-ASSAY® RF (Ver.2) assay is for the quantitative determination of human IgG rheumatoid factor antibodies in patient serum or plasma (citric acid, EDTA, or lithium heparin) based on immunoturbidimetric assay. The presence of IgG RF antibodies, in conjunction with clinical findings and other laboratory tests, is an aid in the diagnosis of rheumatoid arthritis (RA). FOR IN VITRO DIAGNOSTIC USE. The K-ASSAY® RF Calibrator (Ver.2) is intended to be used to calibrate the K-ASSAY® RF (Ver.2) immunoturbidimetric assay. FOR IN VITRO DIAGNOSTIC USE.
    6. Type of Use: Prescription Use
    7. Predicate Device: The letter states the device is "substantially equivalent" to legally marketed predicate devices, but the specific predicate device is not named in this letter.

    To reiterate, the questions about acceptance criteria, study design, sample sizes, experts, ground truth, MRMC studies, standalone performance, and training sets are not applicable to this medical device submission as described in the provided FDA 510(k) clearance letter.

    Ask a Question

    Ask a specific question about this device

    K Number
    K191562
    Manufacturer
    Date Cleared
    2020-03-06

    (267 days)

    Product Code
    Regulation Number
    866.5340
    Reference & Predicate Devices
    Why did this record match?
    510k Summary Text (Full-text Search) :

    : | Class II / 510(k) required |
    | Classification Name: | §866.5775

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Yumizen C1200 Ferritin reagent is intended for the quantitative in vitro diagnostic determination of Ferritin in human serum by latex-enhanced immunoturbidimetric assay. Measurements of ferritin aid in the diagnosis of diseases affecting iron metabolism, hemochromatosis (iron overload) and iron deficiency anemia.

    Yumizen C1200 Transferrin reagent is intended for the quantitative in vitro diagnostic determination of transferrin in human serum and lithium heparin plasma by turbidimetry.

    Measurement of transferrin levels ads in the diagnosis of malnutrition, acute inflammation, infection, and iron deficiency anemia.

    Yumizen C1200 Rheumatoid Factor reagent is intended for the quantitative in vitro diagnostic determination of rheumatoid factor in human serum by latex-enhanced immunoturbidimetric assay. Measurement of rheumatoid factor may aid in the diagnosis of rheumatoid arthritis.

    Device Description

    Not Found

    AI/ML Overview

    The document describes the analytical performance characteristics of three devices: Yumizen C1200 Ferritin, Yumizen C1200 Transferrin, and Yumizen C1200 Rheumatoid Factor. Each device is intended for the quantitative in vitro diagnostic determination of specific substances in human serum, and sometimes plasma, using immunoturbidimetric or turbidimetric assays.

    Here's an analysis of the acceptance criteria and study details for each device:


    Yumizen C1200 Ferritin

    1. Table of Acceptance Criteria and Reported Device Performance

    Performance CharacteristicAcceptance CriteriaReported Device Performance
    Measuring Range (Serum)N/A (claimed measuring range is appropriate based on LOD, LOQ, and linearity studies)10 to 450 ng/mL
    Limit of Detection (Serum)N/A (determined according to CLSI EP17-A2)6.30 ng/mL
    Limit of Quantitation (Serum)N/A (determined according to CLSI EP17-A2)9.39 ng/mL
    Linearity (Serum)N/A (determined according to CLSI EP06-A)Evaluated from 13.3 to 426.6 ng/mL (appropriate)
    Total Precision (Analyzer Variability) - Within Run CVLow level: ≤ 8.0%
    Middle level: ≤ 6.0%
    High level: ≤ 6.0%Level 1 Control (47.58 ng/mL): 3.6%
    Level 2 Control (279.31 ng/mL): 1.1%
    Sample 1 (29.56 ng/mL): 5.5%
    Sample 2 (50.87 ng/mL): 4.1%
    Sample 3 (172.63 ng/mL): 1.4%
    Sample 4 (328.60 ng/mL): 1.3%
    Sample 5 (403.21 ng/mL): 1.0%
    Total Precision (Analyzer Variability) - Total CVLow level: ≤ 10.0%
    Middle & High level: ≤ 8.0%Level 1 Control: 4.9%
    Level 2 Control: 2.1%
    Sample 1: 7.9%
    Sample 2: 5.1%
    Sample 3: 1.9%
    Sample 4: 4.3%
    Sample 5: 1.4%
    Total Precision (Lot to Lot Variability) - Within Run CVLow level: ≤ 8.0%
    Middle level: ≤ 6.0%
    High level: ≤ 6.0%Level 1 Control (52.84 ng/mL): 4.6%
    Level 2 Control (281.87 ng/mL): 0.9%
    Sample 1 (19.09 ng/mL): 8.8%
    Sample 2 (34.05 ng/mL): 6.5%
    Sample 3 (51.53 ng/mL): 3.6%
    Sample 4 (192.31 ng/mL): 1.4%
    Sample 5 (407.38 ng/mL): 0.9%
    Total Precision (Lot to Lot Variability) - Total CVLow level: ≤ 10.0%
    Middle & High level: ≤ 8.0%Level 1 Control: 6.4%
    Level 2 Control: 1.6%
    Sample 1: 11.8% (above criterion, but "pvalue with 5% acceptable remains acceptable")
    Sample 2: 6.5%
    Sample 3: 3.8%
    Sample 4: 2.8%
    Sample 5: 1.4%
    Interferences (Bias)+/- 10% of value without interfering substancesHemoglobin: up to 500 mg/dL
    Triglycerides: up to 270.42 mg/dL
    Total Bilirubin: up to 29.5 mg/dL
    Direct Bilirubin: up to 25.87 mg/dL
    Ascorbic Acid: up to 5.98 mg/dL
    Others specified in document
    Prozone / Antigen Excess EffectDetect and flag samples with underestimated results due to high concentrationAntigen excess observed > 5043 ng/mL; an alarm will flag and re-run these samples.
    Method Comparison (Correlation with Predicate)N/A (determined acceptable by high correlation)Correlation (r²) = 0.999 (for 103 samples, 16.74 - 413.00 ng/mL range)
    Closed StabilityN/A (defined by statement)18 months, stored at 2-10°C, protected from light.
    Open Stability (On-board)N/A (defined by statement)2 months
    Reference Range VerificationSupport establishing ranges vs. literatureWomen: 10 - 120 ng/ml (µg/l)
    Men: 20 - 250 ng/ml (µg/l)

    2. Sample Size and Data Provenance (for test set)

    • Measuring Range, Precision, Interferences, Prozone/Antigen Excess: Not explicitly stated as "test set" in the context of supervised learning, but these are analytical performance studies. Samples used for precision studies include 240 replicates for analyzer variability and 90 replicates for lot-to-lot variability (for each sample/control). The samples are clinical samples or controls, but the origin (country, retrospective/prospective) is not specified beyond "human serum specimens".
    • Method Comparison: 103 native sera samples. Origin: "Anonymous remnants of human serum specimens collected from blood bank." Retrospective.
    • Reference Range: Women: 50 "normal samples". Men: 95 "normal samples". Origin: "blood bank." Retrospective.

    3. Number of experts and qualifications (for ground truth)

    • Not applicable as this is an in vitro diagnostic device for quantitative measurement, not an AI evaluation requiring expert adjudication. Ground truth is instrument-derived or defined by reference methods/literature.

    4. Adjudication method (for test set)

    • Not applicable.

    5. Multi Reader Multi Case (MRMC) comparative effectiveness study

    • No, not applicable for this type of IVD device.

    6. Standalone performance (algorithm only)

    • Yes, the performance data presented is for the device operating in standalone mode (algorithm only) as a quantitative measurement system.

    7. Type of ground truth used

    • Analytical Performance (LOD, LOQ, Linearity, Precision, Interferences, Prozone/Antigen Excess): The "ground truth" is established through well-defined laboratory analytical methods and standards (CLSI guidelines EP17-A2, EP06-A, EP05-A3, EP07-A2). It relies on the accuracy of the reference materials and methods used in these studies.
    • Method Comparison: Comparison against a legally marketed predicate device (Beckman Coulter Ferritin (OSR61203) [K092505]) is used as the reference, with correlation analysis.
    • Reference Range: Verification against established literature references (e.g., TIETZ Textbook of Clinical Chemistry and Molecular Diagnostics).

    8. Sample size for the training set

    • Not applicable. This is not a machine learning model that requires a "training set" in that sense. The device's calibration curve establishment and internal parameters would be set by the manufacturer using validated reference materials and methodologies.

    9. How the ground truth for the training set was established

    • Not applicable.

    Yumizen C1200 Transferrin

    1. Table of Acceptance Criteria and Reported Device Performance

    Performance CharacteristicAcceptance CriteriaReported Device Performance
    Measuring Range (Serum/Plasma)N/A (claimed measuring range is appropriate based on LOD, LOQ, and linearity studies)0.10 to 5.20 g/L
    Limit of Detection (Serum/Plasma)N/A (determined according to CLSI EP17-A2)0.002 g/L
    Limit of Quantitation (Serum/Plasma)N/A (determined according to CLSI EP17-A2)0.07 g/L
    Linearity (Serum/Plasma)N/A (determined according to CLSI EP06-A)Evaluated from 0.15 to 4.61 g/L (appropriate)
    Total Precision (Analyzer Variability) - Within Run CVLow level: ≤ 6.0%
    Middle level: ≤ 4.5%
    High level: ≤ 3.8%Level 1 Control (1.24 g/L): 1.2%
    Level 2 Control (3.35 g/L): 1.5%
    Sample 1 (0.78 g/L): 1.0%
    Sample 2 (1.02 g/L): 1.2%
    Sample 3 (1.83 g/L): 1.3%
    Sample 4 (3.78 g/L): 1.5%
    Total Precision (Analyzer Variability) - Total CVLow level: ≤ 8.0%
    Middle & High level: ≤ 6.0%Level 1 Control: 3.6%
    Level 2 Control: 3.6%
    Sample 1: 4.5%
    Sample 2: 3.2%
    Sample 3: 2.3%
    Sample 4: 2.7%
    Total Precision (Lot to Lot Variability) - Within Run CVLow level: ≤ 6.0%
    Middle level: ≤ 4.5%
    High level: ≤ 3.8%Level 1 Control (1.29 g/L): 3.5%
    Level 2 Control (3.41 g/L): 1.7%
    Sample 1 (0.77 g/L): 4.2%
    Sample 2 (1.08 g/L): 1.5%
    Sample 3 (1.96 g/L): 1.3%
    Sample 4 (3.54 g/L): 2.6%
    Total Precision (Lot to Lot Variability) - Total CVLow level: ≤ 8.0%
    Middle & High level: ≤ 6.0%Level 1 Control: 6.6%
    Level 2 Control: 3.0%
    Sample 1: 5.4%
    Sample 2: 3.3%
    Sample 3: 3.4%
    Sample 4: 3.4%
    Interferences (Bias)+/- 10% of value without interfering substancesHemoglobin: up to 500 mg/dL
    Triglycerides: up to 353.28 mg/dL
    Total Bilirubin: up to 43.84 mg/dL
    Direct Bilirubin: up to 23.86 mg/dL
    Ascorbic Acid: up to 5.98 mg/dL
    Others specified in document
    Anticoagulant Study (Serum vs. Heparin Plasma)No significant difference between serum and plasmaCorrelation (r) = 0.995, Intercept = 0.04833, Slope = 0.9691 (59 paired samples)
    Prozone / Antigen Excess EffectNo antigen excess detected within claimed rangeNo antigen excess detected up to 40 g/L.
    Method Comparison (Correlation with Predicate)N/A (determined acceptable by high correlation)Correlation (r²) = 0.993 (for 115 samples, 0.37 - 4.81 g/L range)
    Closed StabilityN/A (defined by statement)24 months, stored at 2-8°C, protected from light.
    Open Stability (On-board)N/A (defined by statement)6 weeks
    Reference Range VerificationSupport establishing ranges vs. literatureNormal range Transferrin - Serum: 2 - 3.6 g/l (200 - 360 mg/dl)

    2. Sample Size and Data Provenance (for test set)

    • Measuring Range, Precision, Interferences, Prozone/Antigen Excess: Not explicitly stated as "test set" in the context of supervised learning, but these are analytical performance studies. Samples used for precision studies include 240 replicates for analyzer variability and 90 replicates for lot-to-lot variability (for each sample/control). The samples are clinical samples or controls, but the origin (country, retrospective/prospective) is not specified beyond "human serum/plasma".
    • Method Comparison: 115 native samples. Origin: "Anonymous remnants of human serum specimens collected from CHU Nîmes (University Hospital Center)." Retrospective.
    • Anticoagulant Study: 59 paired serum/plasma samples. Origin: "single donors." Not specified if retrospective or prospective.
    • Reference Range: 85 "normal samples" (28 women + 57 men). Origin: "blood bank." Retrospective.

    3. Number of experts and qualifications (for ground truth)

    • Not applicable.

    4. Adjudication method (for test set)

    • Not applicable.

    5. Multi Reader Multi Case (MRMC) comparative effectiveness study

    • No, not applicable.

    6. Standalone performance (algorithm only)

    • Yes, the performance data presented is for the device operating in standalone mode (algorithm only).

    7. Type of ground truth used

    • Analytical Performance: Established through CLSI guidelines (EP17-A2, EP06-A, EP05-A3, EP07-A2).
    • Method Comparison: Comparison against a legally marketed predicate device (Roche Diagnostics Transferrin Model :TRSF2 [K012393]).
    • Reference Range: Verification against established literature references (e.g., Dati et al., Eur. J Clin Chem. Cli Biochem. 1996).

    8. Sample size for the training set

    • Not applicable.

    9. How the ground truth for the training set was established

    • Not applicable.

    Yumizen C1200 Rheumatoid Factor

    1. Table of Acceptance Criteria and Reported Device Performance

    Performance CharacteristicAcceptance CriteriaReported Device Performance
    Measuring Range (Serum)N/A (claimed measuring range is appropriate based on LOD, LOQ, and linearity studies)10 to 120 IU/mL
    Limit of Detection (Serum)N/A (determined according to CLSI EP17-A2)4.07 IU/mL
    Limit of Quantitation (Serum)N/A (determined according to CLSI EP17-A2)7.41 IU/mL
    Linearity (Serum)N/A (determined according to CLSI EP06-A)Evaluated from 13.2 to 118.8 IU/mL (appropriate)
    Total Precision (Analyzer Variability) - Within Run CVLow level: ≤ 6.0%
    Middle level: ≤ 4.5%
    High level: ≤ 3.8%Level 1 Control (40.99 IU/mL): 0.5%
    Level 2 Control (63.93 IU/mL): 0.4%
    Sample 1 (22.24 IU/mL): 1.2%
    Sample 2 (34.28 IU/mL): 0.8%
    Sample 3 (49.41 IU/mL): 0.5%
    Sample 4 (70.16 IU/mL): 0.5%
    Sample 5 (103.42 IU/mL): 0.8%
    Total Precision (Analyzer Variability) - Total CVLow level: ≤ 8.0%
    Middle & High level: ≤ 6.0%Level 1 Control: 2.2%
    Level 2 Control: 2.5%
    Sample 1: 2.0%
    Sample 2: 2.2%
    Sample 3: 1.8%
    Sample 4: 1.6%
    Sample 5: 1.4%
    Total Precision (Lot to Lot Variability) - Within Run CVLow level: ≤ 6.0%
    Middle level: ≤ 4.5%
    High level: ≤ 3.8%Level 1 Control (41.70 IU/mL): 1.8%
    Level 2 Control (67.05 IU/mL): 1.4%
    Sample 1 (17.30 IU/mL): 2.9%
    Sample 2 (30.88 IU/mL): 1.4%
    Sample 3 (53.08 IU/mL): 1.4%
    Sample 4 (70.24 IU/mL): 1.1%
    Sample 5 (102.14 IU/mL): 1.0%
    Total Precision (Lot to Lot Variability) - Total CVLow level: ≤ 8.0%
    Middle & High level: ≤ 6.0%Level 1 Control: 1.9%
    Level 2 Control: 2.2%
    Sample 1: 3.1%
    Sample 2: 1.8%
    Sample 3: 3.2%
    Sample 4: 1.3%
    Sample 5: 1.8%
    Interferences (Bias)+/- 10% of value without interfering substancesHemoglobin: up to 500 mg/dL
    Triglycerides: up to 526.75 mg/dL
    Total Bilirubin: up to 31.32 mg/dL
    Direct Bilirubin: up to 25.34 mg/dL
    Ascorbic Acid: up to 5.98 mg/dL
    Others specified in document
    Prozone / Antigen Excess EffectDetect and flag samples with underestimated results due to high concentrationAntigen excess observed > 229 IU/mL; an alarm will flag and re-run these samples.
    Method Comparison (Correlation with Predicate)N/A (determined acceptable by high correlation)Correlation (r²) = 0.992 (for 113 samples, 16.79 - 118.81 IU/mL range)
    Closed StabilityN/A (defined by statement)18 months, stored at 2-10°C.
    Open Stability (On-board)N/A (defined by statement)1 month
    Reference Range VerificationSupport establishing ranges vs. literatureNormal range Rheumatoid Factor: Adult
    Ask a Question

    Ask a specific question about this device

    K Number
    K190088
    Date Cleared
    2019-04-17

    (90 days)

    Product Code
    Regulation Number
    866.5775
    Reference & Predicate Devices
    Why did this record match?
    510k Summary Text (Full-text Search) :

    Trade/Device Name: QUANTA Flash RF IgM Reagents, QUANTA Flash RF IgA Reagents Regulation Number: 21 CFR 866.5775
    | |
    | Regulation Number | 866.5775

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    QUANTA Flash RF IgM is a chemiluminescent immunoassay for the quantitative determination of IgM rheumatoid factor (RF) antibodies in human serum. The presence of IgM RF antibodies, in conjunction with clinical findings and other laboratory tests, is an aid in the diagnosis of rheumatoid arthritis (RA).

    QUANTA Flash RF IgA is a chemiluminescent immunoassay for the semi-quantitative determination of lgA rheumatoid factor (RF) antibodies in human serum. The presence of IgA RF antibodies, in conjunction with clinical findings and other laboratory tests, is an aid in the diagnosis of rheumatoid arthritis (RA).

    Device Description

    The principle of the assays is chemiluminescent microparticle immunoassay, a variation of solid phase immunoassay. The QUANTA Flash® RF IgM and QUANTA Flash® RF IgA assays are designed to run on the BIO-FLASH® instrument. This platform is a fully automated closed system with continuous load and random access capabilities that automatically processes the samples, runs the assay and reports the results. It includes liquid handling hardware, luminometer and computer with software-user interface. The QUANTA Flash® RF IgM and QUANTA Flash® RF IgA assays utilize a reagent cartridge format, which is compatible with the BIO-FLASH® instrument.

    Rabbit polyclonal antibodies are coated onto paramagnetic beads, which are stored in the reagent cartridge under conditions that preserve the antibody in its reactive state. When the assay cartridge is ready to be used for the first time, the entire cartridge is inverted several times to thoroughly mix the reagents. The reagent cartridge is then loaded onto the BIO-FLASH instrument.

    A patient serum sample is diluted 1:22.7 by the instrument using system rinse in a disposable plastic cuvette. An aliquot of the diluted patient serum, coupled beads, and assay buffer are combined into a second cuvette, and mixed. This cuvette is incubated at 37°C. The beads are then magnetized and washed several times. Isoluminol conjugated anti-human IgM (QUANTA Flash® RF IgM) or anti-human lgA (QUANTA Flash® RF IgA) antibody is then added to the cuvette, and incubated at 37°C. Again, the beads are magnetized and washed repeatedly. The isoluminol conjugate produces a luminescent reaction when "Trigger" reagents are added to the light produced from this reaction is measured as Relative Light Units (RLU) by the BIO-FLASH optical system. RLU values are proportional to the amount of bound isoluminol conjugate, which in turn is proportional to the amount of RF antibodies bound to the antibodies on the beads.

    The QUANTA Flash RF IgM and QUANTA Flash RF IgA assays utilize a predefined lot specific Master Curve that is uploaded into the instrument through the reagent cartridge barcode. Based on the results obtained by running the Calibrators, an instrument specific Working Curve is created, which is used by the software to calculate international units per milliliter (IU/mL) (QUANTA Flash® RF IgM) or chemiluminescent units (CU) (QUANTA Flash® RF IgA) from the RLU value obtained for each sample.

    QUANTA Flash RF IgM Calibrators, QUANTA Flash RF IgM Controls, QUANTA Flash RF IgA Calibrators and QUANTA Flash RF IgA Controls are sold separately.

    The QUANTA Flash® RF IgM Reagents / QUANTA Flash® RF IgA Reagents kit contains the following materials:

    One (1) QUANTA Flash RF IgM / RF IgA Reagent Cartridge

    QUANTA Flash RF IgM Reagent Cartridge contains the following reagents for 100 determinations:

    • a. Rabbit pAb coated paramagnetic beads.
    • b. Assay buffer - colored pink, containing protein stabilizers and preservatives.
    • Tracer IgM Isoluminol labeled anti-human IgM antibody, containing buffer, protein C. stabilizers and preservative.

    QUANTA Flash RF IgA Reagent Cartridge contains the following reagents for 100 determinations:

    • a. Rabbit pAb coated paramagnetic beads.
    • b. Assay buffer - colored pink, containing protein stabilizers and preservatives.
    • Tracer IgA Isoluminol labeled anti-human IgA antibody, containing buffer, protein C. stabilizers and preservative.
    AI/ML Overview

    The document describes the analytical and clinical performance characteristics of the QUANTA Flash® RF IgM and QUANTA Flash® RF IgA Reagents, which are chemiluminescent immunoassays for the quantitative or semi-quantitative determination of rheumatoid factor (RF) antibodies in human serum. These assays are intended to aid in the diagnosis of rheumatoid arthritis (RA) in conjunction with clinical findings and other laboratory tests.

    Here's a breakdown of the requested information:

    1. Table of Acceptance Criteria and Reported Device Performance

    Since this document describes two assays (RF IgM and RF IgA), the acceptance criteria and performance are presented for each. The acceptance criteria for analytical performance studies are generally stated in the document (e.g., %CV

    Ask a Question

    Ask a specific question about this device

    K Number
    K182747
    Manufacturer
    Date Cleared
    2018-12-18

    (81 days)

    Product Code
    Regulation Number
    866.5775
    Reference & Predicate Devices
    Why did this record match?
    510k Summary Text (Full-text Search) :

    , Michigan 49002

    Re: K182747

    Trade/Device Name: EliA RF IgM Immunoassay Regulation Number: 21 CFR 866.5775
    Regulation section: 21 CFR§866.5775 - Rheumatoid Factor Immunological Test System

    • 2.
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    EliA RF IgM is intended for the in vitro quantitative measurement of IgM class rheumatoid factor antibodies in human serum and plasma (Li-heparin, EDTA) to aid in the diagnosis of rheumatoid arthritis in conjunction with other laboratory and clinical findings. EliA RF IgM uses the EliA IgM method on the instrument Phadia 2500/5000.

    Device Description

    The method-specific reagents are identical with K102673, but are filled in containers specific for the Phadia 2500/5000 instrument. Each device consists of: Test Wells: EliA RF IqM Wells are coated with aggregated rabbit IgG 4 carriers (12 wells each), ready to use; EliA Sample Diluent: EliA Sample Diluent: PBS containing BSA, detergent and 0.095% (w/v) sodium azide - 6 bottles, 48 mL each, ready to use; or 6 bottles, 400 mL each, ready to use; EliA IgM Method Reagents: EliA IgM Conjugate 50 or 200: ß-Galactosidase labeled anti-IgM (mouse monoclonal antibodies) in PBS containing BSA and 0.06% (w/v) sodium azide - 6 wedge shaped bottles, 5 mL each, ready to use; or 6 wedge shaped bottles, 19 mL each, ready to use - -EliA IgM Calibrator Strips: Human IgM (0, 10, 35, 80, 500, 1000 µg/L) in PBS containing BSA, detergent and 0.095% (w/v) sodium azide - 5 strips, 6 singleuse vials per strip, 0.3 mL each, ready to use; - EliA IgM Curve Control Strips: Human IgM (80 µg/L) in PBS containing BSA, detergent and 0.095% (w/v) sodium azide – 5 strips, 6 single-use vials per strip, 0.3 mL each, ready to use; - EliA IgM Calibrator Well: Coated with mouse monoclonal antibodies 4 carriers (12 wells each), ready to use. The Phadia EliA Immunodiagnostic System is an automated system for immunodiagnostic testing. The EliA reagents are available as modular packages, each purchased separately. All packages are required to carry out an EliA RF IgM test.

    AI/ML Overview

    Here's a breakdown of the acceptance criteria and study information for the EliA RF IgM Immunoassay on the Phadia 2500/5000 instruments, based on the provided FDA 510(k) summary:

    1. Acceptance Criteria and Reported Device Performance

    The core of this submission is to demonstrate that the EliA RF IgM Immunoassay, previously cleared on the Phadia 250 instrument (K102673), performs substantially equivalently when used on the new Phadia 2500/5000 instrument platform. Therefore, the acceptance criteria are primarily focused on method comparison and analytical performance to show this equivalence. No explicit "acceptance criteria" are provided as a single table within this document for each performance characteristic, but they are implied by the statistical analyses and cut-offs used.

    Here's an organized table presenting the relevant acceptance criteria (implied or stated) and the reported device performance:

    Performance CharacteristicAcceptance Criteria (Implied/Stated)Reported Device Performance
    PrecisionNot explicitly stated for each concentration, but typically CV% should be within acceptable clinical limits for diagnostic assays.EliA RF IgM on Phadia 2500/5000:
    Mean (IU/mL)Within-Run SD (%CV)Between-Run SD (%CV)
    1.60.2 (12.7)0.1 (7.1)
    3.90.2 (5.0)0.2 (4.5)
    7.00.3 (4.8)0.2 (3.1)
    73.71.8 (2.5)3.1 (4.2)
    169.67.3 (4.3)7.9 (4.6)
    (Table from Section 10a)
    LinearitySlope for regression lines should be close to 1, intercept close to 0, R² close to 1.EliA RF IgM on Phadia 2500/5000:
    Dilution range (IU/mL)SlopeIntercept
    0.9 - 41.01.020.37
    0.9 - 29.21.020.18
    2.2 - 195.50.96-1.11
    0.8 - 71.51.010.24
    (Table from Section 10b)
    Reportable RangeDefined by LoD and upper limit (200 IU/mL).EliA RF IgM on Phadia 2500/5000:
    Reportable range: 0.6 to 200 IU/mL.
    Measuring range (LoQ to upper limit): 1.0 to 200 IU/mL. (Section 10b)
    Detection Limit (LoD)Determined consistent with CLSI EP17-A; proportions of false positives (α)
    Ask a Question

    Ask a specific question about this device

    K Number
    K162263
    Date Cleared
    2017-05-03

    (265 days)

    Product Code
    Regulation Number
    866.5775
    Reference & Predicate Devices
    Why did this record match?
    510k Summary Text (Full-text Search) :

    B15 1QT UK

    Re: K162263

    Trade/Device Name: Optilite® Rheumatoid Factor Kit Regulation Number: 21 CFR 866.5775
    Regulation section: 21 CFR 866.5775, Rheumatoid factor immunological test system

    • 2.
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Optilite Rheumatoid Factor (RF) Kit is intended for the quantitative in vitro measurement of rheumatoid factor in serum using the Binding Site Optilite analyser. Measurement of rheumatoid factor may aid in the diagnosis of rheumatoid arthritis. This test should be used in conjunction with other laboratory and clinical findings.

    Device Description

    The Optilite Rheumatoid Factor Kit comprises the following reagents: Reaction Buffer, Latex Reagent, RF Controls (supplied at 2 levels, Low and High), and RF Calibrator.

    AI/ML Overview

    Here's an analysis of the provided text, focusing on the acceptance criteria and the study that proves the device meets them, structured according to your requested information:

    1. Table of Acceptance Criteria and the Reported Device Performance

    Performance CharacteristicAcceptance CriteriaReported Device Performance
    Precision/Reproducibility
    Total Precision%CV
    Ask a Question

    Ask a specific question about this device

    K Number
    K160070
    Date Cleared
    2016-12-07

    (329 days)

    Product Code
    Regulation Number
    866.5775
    Reference & Predicate Devices
    N/A
    Why did this record match?
    510k Summary Text (Full-text Search) :

    K160070

    Trade/Device Name: Rheumatoid Factor (RF) Kit For Use On SPAPLUS® Regulation Number: 21 CFR 866.5775

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Rheumatoid Factor (RF) Kit for use on SPAPLUS is intended for the quantitative in vitro measurement of rheumatoid factors in serum using the Binding Site SPAPLUS analyser. Measurement of rheumatoid factor may aid in the diagnosis of rheumatoid arthritis. This test should be used in conjunction with other laboratory and clinical findings.

    Device Description

    Not Found

    AI/ML Overview

    This document is a 510(k) premarket notification decision letter from the FDA for a medical device. It does not contain information about acceptance criteria or specific study details that would allow me to populate the requested table and answer the detailed questions about device performance and study design.

    The document primarily states that the device, "Rheumatoid Factor (RF) Kit For Use On SPAPLUS®," is substantially equivalent to legally marketed predicate devices. It also outlines regulatory requirements and provides contact information for further inquiries.

    Therefore, I cannot fulfill the request based on the provided text.

    Ask a Question

    Ask a specific question about this device

    Why did this record match?
    510k Summary Text (Full-text Search) :

    Quality Controls ADVIA Centaur® Anti-CCP IgG (aCCP) Master Curve Materials Regulation Number: 21 CFR 866.5775
    Trade Name: ADVIA Centaur® anti-CCP IgG (aCCP) assay Common Name: Anti-CCP test Governing Regulation: 866.5775

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The ADVIA Centaur® anti-CCP IgG (aCCP) assay is for in vitro diagnostic use in the semi-quantitative determination of the IgG class of autoantibodies specific to cyclic citrullinated peptide (CCP) in human serum or plasma (K2-EDTA and lithium heparin) using the ADVIA Centaur XP system. Detection of anti-CCP antibodies is used as an aid in the diagnosis of Rheumatoid Arthritis (RA) and should be used in conjunction with other clinical information. Autoantibody levels represent one parameter in a multi-criteria diagnostic process, encompassing both clinical and laboratory-based assessments.

    Quality Control:
    The ADVIA Centaur® Anti-CCP IgG (aCCP) quality control material is for in vitro diagnostic use to monitor the precision and accuracy of the ADVIA Centaur aCCP assay using the ADVIA Centaur systems.

    Master Curve Material (MCM):
    The ADVIA Centaur® Anti-CCP IgG (aCCP) Master Curve Material (MCM) is for in vitro diagnostic use in the verification of calibration and reportable range of the ADVIA Centaur aCCP assay.

    Device Description

    The ADVIA Centaur aCCP assay is a fully automated, two-step immunoassay using chemiluminescent technology. The assay utilizes an acridinium ester-labeled anti-human lgG as the Lite Reagent. The Solid Phase consists of biotinylated CCP coupled to streptavidin which is then coated onto magnetic latex microparticles.

    AI/ML Overview

    Here's an analysis of the acceptance criteria and the study that proves the device meets them, based on the provided text:

    Device Name: ADVIA Centaur® Anti-CCP IgG (aCCP) Assay

    1. Table of Acceptance Criteria and Reported Device Performance:

    The document doesn't explicitly state "acceptance criteria" as a separate, quantitative table derived from a pre-defined standard (like a predicate device's performance). Instead, it compares the performance of the new device (ADVIA Centaur aCCP assay) to a predicate device (ARCHITECT Anti-CCP Assay) for several characteristics and implies that similar performance constitutes meeting the criteria for substantial equivalence. I will reconstruct a table showing the implied acceptance criteria (based on the predicate's performance or general expectations for such an assay) and the reported performance.

    Performance MetricImplied Acceptance Criteria (Based on Predicate or General Assay Expectations)Reported Device Performance (ADVIA Centaur aCCP Assay)
    Intended UseSemi-quantitative determination of IgG class anti-CCP antibodies, aid in RA diagnosis, used with other clinical info.Matches: Semi-quantitative determination of IgG class anti-CCP antibodies, aid in RA diagnosis, used with other clinical info.
    Assay TechnologyAutomated, Chemiluminescent Microparticle Immunoassay (CMIA)Matches: Automated, Chemiluminescent Microparticle Immunoassay (CMIA)
    Specimen TypeHuman serum, serum separator tubes, human plasma (lithium heparin, potassium EDTA)Matches: Human serum, serum separator tubes, human plasma (lithium heparin, potassium EDTA)
    Capture AntibodyCyclic citrullinated peptide (CCP), second generationMatches: Cyclic citrullinated peptide (CCP), second generation
    Conjugate AntibodyMouse anti-human IgG: acridinium-labeledMatches: Mouse anti-human IgG: acridinium-labeled
    Storage ConditionsIntended Storage of 2-8 °CMatches: Intended Storage of 2-8 °C
    Calibrator Range0.0-200.0 U/mLMatches: 0.0-200.0 U/mL
    Suggested Cut-Off5.0 U/mLMatches: 5.0 U/mL
    Interference (Total Protein)No interference from Total Protein (12 g/dL)Matches: No interference from Total Protein (12 g/dL)
    Interference (Rheumatoid Factor)No interference from Rheumatoid Factor (200 IU/mL)Matches: No interference from Rheumatoid Factor (200 IU/mL)
    Cross-Reactivity (General)No significant cross-reactivity with specified autoantibodies (e.g., SSA, SSB, Sm, RNP, Scl-70, TPO, Jo-1, ds-DNA, Ribosomal P)Matches: No significant cross-reactivity with specified autoantibodies (SSA, SSB, Sm, RNP, Scl-70, TPO, Jo-1, ds-DNA, Ribosomal P). Also M2, Chromatin.
    Sample StabilitySpecimens stable for up to 7 days at 2-8ºC or 22 hours at 30ºC, avoid > 3 freeze/thaw cycles.Separated specimens stable for up to 22 hours at room temp or up to 7 days at 2-8 ºC. Avoid > 2 freeze-thaw cycles. (Slight difference in freeze/thaw cycle recommendation, but generally comparable).
    Imprecision (Within-Lab %CV)Predicate: Within-run CV of 2.0% to 4.7% and total CV of 2.8 to 7.7% (2.7 to 195.3 U/mL).
    New device design spec: 50 U/mL.Meets Design Spec: Ranged from 3.0 to 4.3% (2.37 to 111.53 U/mL).
    Sensitivity (Limit of Detection)≤ 0.5 U/mLBetter: 0.40 U/mL (design goal was ≤ 1.50 U/mL)
    Measurable Range0.5 - 200.0 U/mLComparable/Slightly Wider: 0.40 – 200.0 U/mL
    On-board Reagent StabilityMax 30 daysBetter: Max 60 days
    High Dose HookNot explicitly stated but assumed desirable to avoid.Patient samples up to 3000.00 U/mL will assay > 200.00 U/mL (no high-dose hook effect within this range).
    Interference (General)No significant effects from hemolyzed, icteric, lipemic samples within specified limits.Hemolyzed (1000 mg/dL Hb), Icteric (40 mg/dL unconj/conj bilirubin), Lipemic (1500 mg/dL Intralipid, 2450 mg/dL triglycerides). Biotin (500 ng/dL), Caprine IgG (6 g/dL).
    Dilution Linearity (Recovery)Not explicitly stated but assumed desirable to be within an acceptable range (e.g., 80-120%).Sample 1 (140.63 U/mL) diluted 1:10 showed 107.87% recovery. Sample 2 (180.08 U/mL) diluted 1:10 showed 105.62% recovery.
    Clinical Concordance (Overall % Agreement)"Substantially equivalent performance" to predicate. Implicitly, high agreement.96.84% (CI 93.89 - 98.39%) with ARCHITECT Anti-CCP.
    Clinical Sensitivity (RA diagnosis)"Substantially equivalent performance." Implicitly, acceptable sensitivity for RA.68.08% (CI 62.5-73.3%)
    Clinical Specificity (Non-RA population)"Substantially equivalent performance." Implicitly, acceptable specificity for RA.97.17% (CI 95.22-98.49%)

    2. Sample Sizes and Data Provenance:

    • Test Set (Clinical Performance):
      • Method Comparison Study: 253 samples (143 confirmed positive for RA, and 110 samples where other auto-antibodies may be present). Data provenance not explicitly stated (e.g., country of origin, retrospective/prospective), but it is a clinical study.
      • Clinical Sensitivity and Specificity Study: 767 patient samples.
        • 307 confirmed-positive RA subjects.
        • 460 non-RA subjects with potentially cross-reacting conditions (22 subgroups).
        • Data provenance not explicitly stated (e.g., country of origin, retrospective/prospective).
    • Training Set: Not explicitly mentioned for this type of in vitro diagnostic assay. Immunoassays are generally calibrated and validated, not "trained" in the machine learning sense. The "training" in this context would likely refer to the development and optimization of the assay reagents and parameters.

    3. Number of Experts and Qualifications for Ground Truth (Test Set):

    • For RA Confirmation: "307 confirmed-positive RA subjects that were classified according to the American College of Rheumatology criteria." This implies expert clinical diagnosis based on established criteria, but the specific number and qualifications of individuals making these diagnoses are not provided.
    • For Non-RA Subjects: "22 subgroups of non-RA subjects (n=460) with potentially cross-reacting conditions." This also implies clinical diagnosis by experts, but details are not provided.
    • For Autoantibody Presence (Method Comparison): "110 samples where other auto-antibodies may be present." This suggests expert determination of autoantibody status, but details are not given.

    4. Adjudication Method for the Test Set:

    Not applicable or explicitly stated as this is an immunoassay, not an imaging device requiring expert adjudication of reader interpretations. The ground truth for clinical sensitivity and specificity is based on clinical diagnosis (American College of Rheumatology criteria for RA).

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study:

    No, an MRMC comparative effectiveness study was not done. This type of study is typically performed for imaging devices or AI tools that assist human readers in interpretation. The ADVIA Centaur anti-CCP IgG assay is a standalone diagnostic laboratory test.

    6. Standalone Performance:

    Yes, standalone performance was done. The entire document describes the standalone performance of the ADVIA Centaur aCCP assay in various analytical and clinical studies (linearity, dilution linearity, detection capability, high dose hook, cross-reactivity, interference, precision, specimen collection comparison, clinical concordance with a predicate, and clinical sensitivity/specificity against clinical diagnosis). The results presented are solely the performance of the algorithm/assay without human intervention in the result determination.

    7. Type of Ground Truth Used:

    • Analytical Performance: Based on reference materials, spiked samples with known concentrations, or established analytical methods.
    • Clinical Performance (Sensitivity/Specificity):
      • RA Diagnosis: American College of Rheumatology criteria (clinical diagnosis) for RA.
      • Non-RA Status: Clinical diagnosis of various conditions (22 subgroups).
    • Method Comparison: The predicate device (ARCHITECT Anti-CCP Assay) results were used as a reference point for agreement.

    8. Sample Size for the Training Set:

    Not applicable in the machine learning sense. For an immunoassay, "training" is typically assay development, which involves optimizing reagents and parameters. The document doesn't specify sample sizes used during the development phase.

    9. How the Ground Truth for the Training Set Was Established:

    Not applicable. For an immunoassay, ground truth during development would involve well-characterized samples (e.g., confirmed positive/negative for CCP antibodies, known concentrations) to optimize the assay's performance characteristics. Specific details on this are not provided in the summary.

    Ask a Question

    Ask a specific question about this device

    Why did this record match?
    510k Summary Text (Full-text Search) :

    Enhanced RF IgM Antibody ELISA ImmuLisa Enhanced RF Antibody Screen ELISA Regulation Number: 21 CFR 866.5775
    ™ RF IgM Antibody ELISA ImmuLisa Enhanced™ RF IgA/IgG/IgM Antibody ELISA Regulation Number: 21 CFR §866.5775

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Enzyme linked immunoassay (ELISA) for the qualitative or semi-quantitative detection of Rheumatoid Factor IgA antibodies in human serum to aid in the diagnosis of theumatoid arthritis (RA) in conjunction with other laboratory tests and clinical findings.
    Enzyme linked immunoassay (ELISA) for the qualitative or semi-quantitative detection of Rheumatoid Factor IgG antibodies in human serum to aid in the diagnosis of theumatoid arthritis (RA) in conjunction with other laboratory tests and clinical findings.
    Enzyme linked immunoassay (ELISA) for the qualitative or semi-quantitative detection of Rheumatoid Factor IgM antibodies in human serum to aid in the diagnosis of theumatoid arthritis (RA) in conjunction with other laboratory tests and clinical findings.
    Enzyme linked immunoassay (ELISA) for the qualitative detection of Rheumatoid Factor IgA, IgG and IgM antibodies in human serum to aid in the diagnosis of theumatoid arthritis (RA) in conjunction with other laboratory tests and clinical findings.

    Device Description

    Not Found

    AI/ML Overview

    This document is a 510(k) Premarket Notification from the FDA regarding the ImmuLisa Enhanced™ RF IgA Antibody ELISA, ImmuLisa Enhanced™ RF IgG Antibody ELISA, ImmuLisa Enhanced™ RF IgM Antibody ELISA, and ImmuLisa Enhanced™ RF IgA/IgG/IgM Antibody ELISA. It doesn't contain information about acceptance criteria or a study that proves the device meets specific performance metrics.

    The document primarily states that the FDA has reviewed the submission and determined that the device is substantially equivalent to legally marketed predicate devices. It outlines regulatory information such as:

    • Trade/Device Name: ImmuLisa Enhanced™ RF IgA Antibody ELISA, ImmuLisa Enhanced™ RF IgG Antibody ELISA, ImmuLisa Enhanced™ RF IgM Antibody ELISA, ImmuLisa Enhanced™ RF IgA/IgG/IgM Antibody ELISA
    • Regulation Number: 21 CFR 866.5775
    • Regulation Name: Rheumatoid factor immunological test system
    • Regulatory Class: Class II
    • Product Code: DHR
    • Indications for Use: Enzyme linked immunoassay (ELISA) for the qualitative or semi-quantitative detection of Rheumatoid Factor IgA, IgG, and/or IgM antibodies in human serum to aid in the diagnosis of rheumatoid arthritis (RA) in conjunction with other laboratory tests and clinical findings.
    • Type of Use: Prescription Use

    Therefore, I cannot provide the requested information about acceptance criteria, device performance, study details (sample sizes, data provenance, ground truth establishment, expert qualifications, adjudication methods), MRMC studies, or standalone algorithm performance, as these details are not present in the provided text.

    Ask a Question

    Ask a specific question about this device

    K Number
    K143754
    Date Cleared
    2015-09-21

    (264 days)

    Product Code
    Regulation Number
    866.5775
    Reference & Predicate Devices
    Why did this record match?
    510k Summary Text (Full-text Search) :

    Quanta Flash® CCP3 Quanta Flash® CCP3 Calibrators Quanta Flash® CCP3 Controls Regulation Number: 21 CFR 866.5775
    |

    5

    | Regulation Number | 866.5775

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    QUANTA Flash CCP3 is a chemiluminescent immunoassay for the semi-quantitative determination of IgG anti-CCP3 antibodies in human serum. The presence of anti-CCP3 antibodies, in conjunction with clinical findings and other laboratory tests, is an aid in the diagnosis of rheumatoid arthritis.

    QUANTA Flash CCP3 Calibrators are intended for use with the QUANTA Flash CCP3 chemiluminescent immunoassay for the determination of IgG anti-CCP3 antibodies in human serum. Each calibrator establishes a point of reference for the working curve that is used to calculate unit values.

    QUANTA Flash CCP3 Controls are intended for use with the QUANTA Flash CCP3 chemiluminescent immunoassay for quality control in the determination of IgG anti-CCP3 antibodies in human serum.

    Device Description

    The QUANTA Flash CCP3 assay is designed to run on the BIO-FLASH® instrument. This platform is a fully automated closed system with continuous load and random access capabilities that automatically processes the samples, runs the assay and reports the results. It includes liquid handling hardware, luminometer and computer with software-user interface. The QUANTA Flash CCP3 assay utilizes a reagent cartridge format, which is compatible with the BIO-FLASH instrument.

    Synthetic cyclic citrullinated peptide is coated onto paramagnetic beads. The bead suspension is lyophilized and stored in the bead tube. Prior to use in the BIO-FLASH system, the sealed reagent tubes are pierced with the reagent cartridge lid and the beads are rehydrated and resuspended using resuspension buffer by pipetting up and down with a transfer pipette. The reagent cartridge is then loaded onto the BIO-FLASH instrument. Samples are also loaded onto the instrument in sample racks. Serum samples are diluted by the BIO-FLASH with system rinse in a small disposable plastic cuvette. Small amounts of the diluted patient serum, the beads, and assay buffer are combined into a second cuvette, and mixed. This cuvette is then incubated at 37°C. The beads are magnetized and washed several times. Isoluminol conjugated anti-human IgG antibodies are then added to the cuvette, and again incubated at 37°C. The beads are magnetized and washed repeatedly. The isoluminol conjugate is oxidized when Trigger 1 (Fe(II)coproporphyrin in sodium hydroxide solution) and Trigger 2 (ureahydrogen peroxide in sodium chloride solution) are added to the flash of light produced from this reaction is measured as Relative Light Units (RLU) by the BIO-FLASH optical system. The RLU are proportional to the amount of isoluminol conjugate that is bound to the human IgG, which is in turn proportional to the amount of anti-CCP3 antibodies bound to the corresponding beads.

    For quantitation, the QUANTA Flash CCP3 assay utilizes a predefined lot specific Master Curve that is uploaded onto the instrument through the reagent cartridge barcode. Every new lot number of reagent cartridge must be calibrated before first use, with the QUANTA Flash CCP3 Calibrators. Based on the results obtained with the two Calibrators included in the Calibrator Set (sold separately), an instrument specific Working Curve is created, which is used to calculate chemiluminescent units (CU) from the instrument signal (RLU) obtained for each sample.

    The QUANTA Flash CCP3 kit contains the following materials:

    One (1) QUANTA Flash CCP3 Reagent Cartridge One (1) vial of Resuspension buffer One (1) Transfer pipette

    The QUANTA Flash CCP3 reagent cartridge contains the following reagents for 100 determinations:

    • a. CCP3 coated paramagnetic beads, lyophilized.
    • b. Assay buffer - colored pink, containing Tris-buffered saline, Tween 20, protein stabilizers and preservatives.
    • C. Tracer IgG - Isoluminol labeled anti-human IgG antibodies in buffer, containing protein stabilizers and preservative.

    The QUANTA Flash CCP3 Calibrators kit contains two vials of Calibrator 1 and two vials of Calibrator 2:

    QUANTA Flash CCP3 Calibrators:

    • QUANTA Flash CCP3 Calibrator 1: Two (2) barcode labeled tubes containing 0.7 mL prediluted, ready to use reagent. Calibrators contain human antibodies to CCP3 in stabilizer and preservative.

    • QUANTA Flash CCP3 Calibrator 2: Two (2) barcode labeled tubes containing 0.7 mL prediluted, ready to use reagent. Calibrators contain human antibodies to CCP3 in stabilizer and preservative.
      The QUANTA Flash CCP3 Controls kit contains two vials of Negative Control and two vials of Positive Control:

    QUANTA Flash CCP3 Controls:

    • QUANTA Flash CCP3 Negative Control: Two (2) barcode labeled tubes containing 0.5 mL, ready to use reagent. Controls contain human antibodies to CCP3 in stabilizer and preservative.
    • QUANTA Flash CCP3 Positive Control: Two (2) barcode labeled tubes containing 0.5 mL, ready to use reagent. Controls contain human antibodies to CCP3 in stabilizer and preservative.
    AI/ML Overview

    The provided text describes the QUANTA Flash® CCP3 chemiluminescent immunoassay and its associated calibrators and controls. The study focuses on establishing the substantial equivalence of the new device to a predicate device (QUANTA Lite® CCP3 IgG ELISA) and demonstrating its analytical and clinical performance characteristics.

    Here's an analysis of the acceptance criteria and study as requested:

    1. Table of Acceptance Criteria and Reported Device Performance

    The document details numerous acceptance criteria for various analytical and clinical performance characteristics. Below is a summary of some key areas:

    Performance CharacteristicAcceptance CriteriaReported Device Performance
    PrecisionTotal %CV:
    Ask a Question

    Ask a specific question about this device

    K Number
    K121576
    Date Cleared
    2013-04-04

    (309 days)

    Product Code
    Regulation Number
    866.5775
    Reference & Predicate Devices
    Why did this record match?
    510k Summary Text (Full-text Search) :

    Regulation 21 CFR 866.5775 1

    section:

    1. Classification: Class II

    IgG Assay and IMMULITE 2000 Anti-CCP IgG Calibration Verification Material Regulation Number: 21 §CFR 866.5775

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The IMMULITE 2000 Anti-CCP IgG assay is an in vitro diagnostic immunoassay for the semi-quantitative determination of the IgG class of autoantibodies specific to cyclic citrullinated peptide (CCP) in human serum (including Serum Separator Tubes) or plasma (EDTA or lithium heparin) on the IMMULITE 2000 system. Detection of anti-CCP antibodies is used as an aid in the diagnosis of Rheumatoid Arthritis (RA) and should be used in conjunction with other clinical information. Autoantibody levels represent one parameter in a multi-criteria diagnostic process, encompassing both clinical and laboratory-based assessments.

    The IMMULITE 2000 Anti-CCP IgG Calibration Verification Material (CVM) is for in vitro diagnostic use, as a control for calibration verification of the IMMULITE 2000 Anti-CCP IgG assay on the IMMULITE 2000 system.

    Device Description

    The IMMULITE 2000 Anti-CCP IgG assay consists of the following components:

    • Anti-CCP IgG bead pack coated with cyclic citrullinated peptide . (CCP) antigen
    • Anti-CCP IgG reagent wedge containing bovine calf intestine . conjugated to a monoclonal murine anti-human IgG antibody
    • Anti-CCP IgG adjustors, low and high, containing lyophilized . human serum with IgG reactive to CCP
    • Anti-CCP IgG controls, negative and positive, containing human . serum
    • . Autoantibody sample diluent containing protein/buffer matrix

    The IMMULITE Anti-CCP IgG Calibration Verification Material consists of one set of four vials, containing low, intermediate and high levels of lyophilized human serum with IgG reactive to cyclic citrullinated peptide (CCP), in buffer with preservative, plus an anti-CCP-free sample.

    AI/ML Overview

    The IMMULITE® 2000 Anti-CCP IgG Assay is an in vitro diagnostic immunoassay for the semi-quantitative determination of the IgG class of autoantibodies specific to cyclic citrullinated peptide (CCP). It is used as an aid in the diagnosis of Rheumatoid Arthritis (RA) and should be used in conjunction with other clinical information.

    Here's an analysis of the acceptance criteria and the study that proves the device meets those criteria:


    1. Table of Acceptance Criteria and Reported Device Performance

    The provided document does not explicitly state acceptance criteria in a single, consolidated table. However, performance characteristics are reported, which implicitly define what the manufacturer considers acceptable for their device to perform as intended and to demonstrate substantial equivalence to the predicate device. Based on the performance characteristics presented, the following can be inferred as the "acceptance criteria" through the demonstrated results.

    Performance CharacteristicAcceptance Criteria (Implicit from Results)Reported Device Performance
    Precision (Total CV%)Generally, Coefficients of Variation (CV%) for diagnostic assays should be within acceptable limits (e.g.,
    Ask a Question

    Ask a specific question about this device

    Page 1 of 5