K Number
K190088
Date Cleared
2019-04-17

(90 days)

Product Code
Regulation Number
866.5775
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

QUANTA Flash RF IgM is a chemiluminescent immunoassay for the quantitative determination of IgM rheumatoid factor (RF) antibodies in human serum. The presence of IgM RF antibodies, in conjunction with clinical findings and other laboratory tests, is an aid in the diagnosis of rheumatoid arthritis (RA).

QUANTA Flash RF IgA is a chemiluminescent immunoassay for the semi-quantitative determination of lgA rheumatoid factor (RF) antibodies in human serum. The presence of IgA RF antibodies, in conjunction with clinical findings and other laboratory tests, is an aid in the diagnosis of rheumatoid arthritis (RA).

Device Description

The principle of the assays is chemiluminescent microparticle immunoassay, a variation of solid phase immunoassay. The QUANTA Flash® RF IgM and QUANTA Flash® RF IgA assays are designed to run on the BIO-FLASH® instrument. This platform is a fully automated closed system with continuous load and random access capabilities that automatically processes the samples, runs the assay and reports the results. It includes liquid handling hardware, luminometer and computer with software-user interface. The QUANTA Flash® RF IgM and QUANTA Flash® RF IgA assays utilize a reagent cartridge format, which is compatible with the BIO-FLASH® instrument.

Rabbit polyclonal antibodies are coated onto paramagnetic beads, which are stored in the reagent cartridge under conditions that preserve the antibody in its reactive state. When the assay cartridge is ready to be used for the first time, the entire cartridge is inverted several times to thoroughly mix the reagents. The reagent cartridge is then loaded onto the BIO-FLASH instrument.

A patient serum sample is diluted 1:22.7 by the instrument using system rinse in a disposable plastic cuvette. An aliquot of the diluted patient serum, coupled beads, and assay buffer are combined into a second cuvette, and mixed. This cuvette is incubated at 37°C. The beads are then magnetized and washed several times. Isoluminol conjugated anti-human IgM (QUANTA Flash® RF IgM) or anti-human lgA (QUANTA Flash® RF IgA) antibody is then added to the cuvette, and incubated at 37°C. Again, the beads are magnetized and washed repeatedly. The isoluminol conjugate produces a luminescent reaction when "Trigger" reagents are added to the light produced from this reaction is measured as Relative Light Units (RLU) by the BIO-FLASH optical system. RLU values are proportional to the amount of bound isoluminol conjugate, which in turn is proportional to the amount of RF antibodies bound to the antibodies on the beads.

The QUANTA Flash RF IgM and QUANTA Flash RF IgA assays utilize a predefined lot specific Master Curve that is uploaded into the instrument through the reagent cartridge barcode. Based on the results obtained by running the Calibrators, an instrument specific Working Curve is created, which is used by the software to calculate international units per milliliter (IU/mL) (QUANTA Flash® RF IgM) or chemiluminescent units (CU) (QUANTA Flash® RF IgA) from the RLU value obtained for each sample.

QUANTA Flash RF IgM Calibrators, QUANTA Flash RF IgM Controls, QUANTA Flash RF IgA Calibrators and QUANTA Flash RF IgA Controls are sold separately.

The QUANTA Flash® RF IgM Reagents / QUANTA Flash® RF IgA Reagents kit contains the following materials:

One (1) QUANTA Flash RF IgM / RF IgA Reagent Cartridge

QUANTA Flash RF IgM Reagent Cartridge contains the following reagents for 100 determinations:

  • a. Rabbit pAb coated paramagnetic beads.
  • b. Assay buffer - colored pink, containing protein stabilizers and preservatives.
  • Tracer IgM Isoluminol labeled anti-human IgM antibody, containing buffer, protein C. stabilizers and preservative.

QUANTA Flash RF IgA Reagent Cartridge contains the following reagents for 100 determinations:

  • a. Rabbit pAb coated paramagnetic beads.
  • b. Assay buffer - colored pink, containing protein stabilizers and preservatives.
  • Tracer IgA Isoluminol labeled anti-human IgA antibody, containing buffer, protein C. stabilizers and preservative.
AI/ML Overview

The document describes the analytical and clinical performance characteristics of the QUANTA Flash® RF IgM and QUANTA Flash® RF IgA Reagents, which are chemiluminescent immunoassays for the quantitative or semi-quantitative determination of rheumatoid factor (RF) antibodies in human serum. These assays are intended to aid in the diagnosis of rheumatoid arthritis (RA) in conjunction with clinical findings and other laboratory tests.

Here's a breakdown of the requested information:

1. Table of Acceptance Criteria and Reported Device Performance

Since this document describes two assays (RF IgM and RF IgA), the acceptance criteria and performance are presented for each. The acceptance criteria for analytical performance studies are generally stated in the document (e.g., %CV < 12% for precision), while clinical performance acceptance is implied by the reported sensitivity and specificity, demonstrating substantial equivalence to the predicate device.

QUANTA Flash® RF IgM Reagents

CriterionAcceptance CriteriaReported Device Performance
Precision (Total %CV)< 12% (Within-laboratory precision)Varies by sample concentration; Max observed was 7.6% (Sample 10, 408.8 IU/mL)
Reproducibility (Between-sites %CV)< 12%Varies by sample concentration; Max observed was 10.4% (Sample 8, 412.9 IU/mL)
Reproducibility (Between-lots %CV)< 12%Varies by sample concentration; Max observed was 9.4% (Sample 5, 209.8 IU/mL)
Limit of Detection (LoD)Below Analytical Measuring Range (AMR)0.1 IU/mL
Limit of Quantification (LoQ)Total imprecision CV% <20%0.3 IU/mL (at CV% <20%)
Analytical Measuring Range (AMR)N/A (defined range)0.3 IU/mL - 490.0 IU/mL
Auto-rerun FunctionAuto-dilution provides reportable results up to 9,800.0 IU/mLAchieved up to 9,800.0 IU/mL
High Concentration Hook EffectNo hook effect within the reported rangeNo hook effect up to 1,830.6 IU/mL (theoretical value)
LinearityBest fitting polynomial is linear OR nonlinearity < 15% or ±0.75 IU/mL for negative samplesAll samples showed linearity; Sample 3 (1.5 - 15.3 IU/mL) showed second order polynomial but nonlinearity (-5.2% to 3.8% or -0.1 to 0.6 IU/mL) met acceptance criteria.
Interference (Recovery)85% - 115% recovery for positive samples, or ± 15% of cutoff (±0.75 IU/mL) difference for negative samples, whichever is greater (for specified interferents)Bilirubin (89.0-98.2%), Hemoglobin (94.5-95.0%), Triglycerides (91.2-112.2%), Cholesterol (89.7-91.6%), Methotrexate (96.1-102.8%), Prednisone (101.9-109.5%) met acceptance. (Human IgG and Ascorbic Acid not tested for IgM directly, but for IgA)
Sample Stability (Recovery)85-115% for positive, 80-120% for negativeAll samples met criteria for temperature (RT 48 hrs, 2-8°C 14 days) and freeze/thaw (3 cycles).
Reagent Shelf Life (Recovery)Lower and upper 95% CI of regression line between 80% and 120% at Day 14 (accelerated stability)Initial one-year expiration dating assigned for all components
Reagent In-use StabilityBased on 95% CI of regression line (85-115% recovery) or ≥2% data points <75% or ≥125% recovery80 days
Clinical SensitivitySubstantial equivalence to predicate device implied (not explicitly stated as a numerical criterion)69.6% (95% CI: 64.1 – 74.6%)
Clinical SpecificitySubstantial equivalence to predicate device implied (not explicitly stated as a numerical criterion)88.3% (95% CI: 84.8 - 91.1%)
Method Comparison (Predicate ELISA)PPA, NPA, TPA demonstrating substantial equivalence (not explicitly stated as numerical criterion)NPA: 96.4% (93.6-98.0%), PPA: 81.1% (76.0 - 85.3%), TPA: 89.1% (86.3 - 91.4%); Spearman's rs of 0.85 (0.82 - 0.87)

QUANTA Flash® RF IgA Reagents

CriterionAcceptance CriteriaReported Device Performance
Precision (Total %CV)< 12% (Within-laboratory precision)Varies by sample concentration; Max observed was 6.5% (Sample 8, 721.1 CU)
Reproducibility (Between-sites %CV)< 12%Varies by sample concentration; Max observed was 7.8% (Sample 8, 738.4 CU)
Reproducibility (Between-lots %CV)< 12%Varies by sample concentration; Max observed was 5.2% (Sample 8, 722.7 CU)
Limit of Detection (LoD)Below Analytical Measuring Range (AMR)0.5 CU
Limit of Quantification (LoQ)Total imprecision CV% <20%1.2 CU (at CV% <20%). AMR starts at 1.3 CU.
Analytical Measuring Range (AMR)N/A (defined range)1.3 CU - 900.0 CU
Auto-rerun FunctionAuto-dilution provides reportable results up to 18,000.0 CUAchieved up to 18,000.0 CU
High Concentration Hook EffectNo hook effect within the reported rangeNo hook effect up to 42,710.4 CU (theoretical value)
LinearityBest fitting polynomial is linear OR nonlinearity < 15% or ±3 CU for negative samplesAll samples showed linearity; Sample 1 (104.5 - 1045.0 CU) showed third order polynomial but nonlinearity (-10.8% to 5.6%) met acceptance criteria.
Interference (Recovery)85% - 115% recovery for positive samples, or ± 15% of cutoff (±3 CU) difference for negative samples, whichever is greater (for specified interferents)Bilirubin (92.5-93.1%), Hemoglobin (90.9-94.7%), Triglycerides (97.8-105.4%), Cholesterol (91.0-91.0%), Human IgG (91.6-95.7%), Ascorbic Acid (98.1-102.6%), Methotrexate (92.9-105.8%), Prednisone (93.0-94.9%) met acceptance.
Sample Stability (Recovery)85-115% for positive, 80-120% for negativeAll samples met criteria for temperature (RT 48 hrs, 2-8°C 14 days) and freeze/thaw (3 cycles).
Reagent Shelf Life (Recovery)Lower and upper 95% CI of regression line between 80% and 120% at Day 14 (accelerated stability)Initial one-year expiration dating assigned for all components
Reagent In-use StabilityBased on 95% CI of regression line (85-115% recovery) or ≥2% data points <75% or ≥125% recovery80 days
Clinical SensitivitySubstantial equivalence to predicate device implied (not explicitly stated as a numerical criterion)56.8% (95% CI: 51.1 – 62.3%)
Clinical SpecificitySubstantial equivalence to predicate device implied (not explicitly stated as a numerical criterion)90.5% (95% CI: 87.3 – 93.0%)
Method Comparison (Predicate ELISA)PPA, NPA, TPA demonstrating substantial equivalence (not explicitly stated as numerical criterion)NPA: 97.5% (95.4–98.6%), PPA: 87.6% (82.5 – 91.3%), TPA: 94.0% (91.8 – 95.6%)

2. Sample Size Used for the Test Set and the Data Provenance

  • Test Set (Clinical Validation Study): A total of 706 characterized samples were used.
    • Data Provenance: The document does not explicitly state the country of origin. However, given it's an FDA 510(k) submission, it's generally understood that the studies conform to regulatory requirements for market clearance in the US.
    • Retrospective/Prospective: The document does not explicitly state whether the samples were collected retrospectively or prospectively. It mentions a "cohort of characterized samples, none of which were used for establishing the reference range," suggesting they were pre-existing and evaluated in a retrospective manner.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and the Qualifications of Those Experts

  • This information is not provided in the document. The document describes a "cohort of characterized samples" but does not detail how the characterization (diagnosis of RA vs. control conditions) was established, nor does it mention the number or qualifications of experts involved in this process. This study is not an MRMC study or an AI-based diagnostic study requiring expert ground truth for image interpretation. Instead, it's about the performance of an in-vitro diagnostic (IVD) assay where the "ground truth" for clinical performance is the clinical diagnosis of the patient.

4. Adjudication Method for the Test Set

  • This concept is not applicable to this type of study. Since the device is an IVD assay measuring biomarkers, there is no "adjudication" in the sense of comparing human reads with an AI output or resolving discrepancies among readers. The assay itself provides a quantitative or semi-quantitative result. The "ground truth" for clinical sensitivity and specificity is the medical diagnosis of rheumatoid arthritis based on various clinical findings and laboratory tests.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done, If So, What Was the Effect Size of How Much Human Readers Improve with AI vs Without AI Assistance

  • No, an MRMC comparative effectiveness study was not done. This document describes an in-vitro diagnostic (IVD) device, not an AI-assisted diagnostic tool that would involve human "readers" or image interpretation. Therefore, this type of study is not relevant to the described device.

6. If a Standalone (i.e., Algorithm Only Without Human-in-the-Loop Performance) Was Done

  • Yes, the primary study described is a standalone performance study of the assay. The QUANTA Flash® RF IgM and IgA assays provide a quantitative or semi-quantitative result directly from the sample. Their performance in terms of precision, linearity, interference, stability, and clinical sensitivity/specificity is evaluated as the standalone performance of the assay itself, without a human "interpretation" component that would be integrated into the device's output.

7. The Type of Ground Truth Used

  • Clinical Diagnosis: For the clinical performance characteristics (sensitivity and specificity), the ground truth was the diagnosis of rheumatoid arthritis (RA) for patient samples and the characterization of control patient groups with other conditions or as apparently healthy donors. This "characterization" implies a pre-existing medical diagnosis, likely based on a combination of clinical findings, medical history, and other standard laboratory tests.
  • Reference Values/Spiked Samples: For analytical performance characteristics (e.g., linearity, LoQ, interference), the ground truth involved reference standards, known concentrations, or spiked samples where the target analyte amount was precisely known.

8. The Sample Size for the Training Set

  • This document describes the regulatory submission for an IVD reagent kit. The concept of a "training set" is more relevant to machine learning algorithms. For IVD development, the samples used to establish initial parameters like the Master Curve for calibration are distinct from a "training set" in an AI context.
    • Reference Range/Cutoff Establishment:
      • Reference population: 191 subjects (117 apparently healthy donors, plus various infectious disease and autoimmune cohorts).
      • RA patient specimens: 42 diagnosed RA patient specimens.
    • Standards for Master Curve: The Master Curve for each assay (IgM and IgA) consists of 6 different Standards. The document implies these standards are pre-defined and used during manufacturing to create the lot-specific Master Curve. The specific number of runs or samples used to define these initial standards is not detailed, but they are manufactured as high-quality reference materials.

9. How the Ground Truth for the Training Set Was Established

  • As noted above, "training set" doesn't directly apply in the AI sense. However, for the samples used to establish the reference range and cutoff values:
    • The reference population (191 subjects) was classified based on their underlying health status (e.g., "apparently healthy donors," "Infectious Disease Controls," "Rheumatoid Arthritis"). This classification represents the ground truth for establishing the assay's normal range and discriminating cutoff.
    • The cutoff values were established in accordance with CLSI EP28-A3c: Defining, Establishing, and Verifying Reference Intervals in the Clinical Laboratory; Approved Guideline - Third Edition.
    • The non-parametric percentile method was used due to the non-normal distribution of results.
    • For QUANTA Flash RF IgM, the cutoff of 5 IU/mL was set based on the distribution of results in the reference population and the (known) positive RA samples to ensure optimal differentiation.
    • For QUANTA Flash RF IgA, the cutoff was established at 6,000 RLU (assigned 20 CU), which was greater than the 95th percentile of control results (3,767 RLU).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

April 17, 2019

Inova Diagnostics, Inc. Roger Albesa Manager, Research and Development 9900 Old Grove Rd. San Diego, California 92131-1638

Re: K190088

Trade/Device Name: QUANTA Flash RF IgM Reagents, QUANTA Flash RF IgA Reagents Regulation Number: 21 CFR 866.5775 Regulation Name: Rheumatoid factor immunological test system Regulatory Class: Class II Product Code: DHR Dated: January 16, 2019 Received: January 17, 2019

Dear Roger Albesa:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's

{1}------------------------------------------------

requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801 and Part 809); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/CombinationProducts/GuidanceRegulatoryInformation/ucm597488.htm); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely.

Doug Jeffery Acting Deputy Director Division of Immunology and Hematology Devices Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Please wait...

If this message is not eventually replaced by the proper contents of the document, your PDF viewer may not be able to display this type of document.

You can upgrade to the latest version of Adobe Reader for Windows®, Mac, or Linux® by visiting http://www.adobe.com/go/reader_download.

For more assistance with Adobe Reader visit http://www.adobe.com/go/acrreader.

Windows is either a registered trademark of Microsoft Corporation in the United States and/or other countries. Mac is a trademark of Apple Inc., registered in the United States and other countries. Linux is the registered trademark of Linus Torvalds in the U.S. and other countries.

{3}------------------------------------------------

Image /page/3/Picture/0 description: The image shows the logo for Inova Diagnostics, a Werfen Company. The logo consists of two stylized shapes, one red and one blue, that are arranged to resemble a stylized letter "W". To the right of the shapes is the text "Inova Diagnostics" in a gray sans-serif font. Below the company name is the text "A Werfen Company" in a smaller, lighter gray font.

510(k) Summary

QUANTA Flash® RF IgM Reagents QUANTA Flash® RF IgA Reagents

Table of Contents Administrative data........................................................................................................................................................... Predicate device.............................................................................................................................................................. Device description............................................................................................................................................................ Intended use(s) .............................................................................................................................................................. Indications for use…………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………… Substantial equivalence ...................................................................................................................................................... Comparison to predicate device ............................................................................................................................................... Analytical performance characteristics ....................................................................................................................................... Quantitation and units of measure ............................................................................................................................................ Precision .................................................................................................................................................................... Reproducibility Studies ...................................................................................................................................................... Limit of Blank (LoB), Limit of Detection (LoD), and Limit of Quantitation (LoQ)............................................................................................... Analytical Measuring Range (AMR) ............................................................................................................................................. Auto-rerun function and reportable results ................................................................................................................................... High concentration hook effect................................................................................................................................................ Linearity .................................................................................................................................................................... Interference ................................................................................................................................................................. Sample Stability and Handling ................................................................................................................................................ Reagent Stability ............................................................................................................................................................ Cut-off, reference range...................................................................................................................................................... Clinical performance characteristics.......................................................................................................................................... Clinical sensitivity, specificity ............................................................................................................................................ Expected values............................................................................................................................................................... Comparison with predicate device .............................................................................................................................................

This summary of the 510(k) safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR 807.92.

{4}------------------------------------------------

Administrative data

Submitter:Inova Diagnostics, Inc9900 Old Grove Road,San Diego, CA, 92131
Purpose of submission:New device(s)
Devices in the submission:QUANTA Flash® RF IgM ReagentsQUANTA Flash® RF IgA Reagents
Scientific contact:Roger Albesa, Manager, Research and DevelopmentInova Diagnostics, Inc.9900 Old Grove Road, San Diego, CA, 92131Phone: 858-586-9900 x1391Fax: 858-863-0025Email: ralbesa@inovadx.com
Quality Systems contact:Ronda Elliott, VP, Quality Systems and RAInova Diagnostics, Inc9900 Old Grove Road, San Diego, CA, 92131Phone: 858-586-9900 x1381Fax: 858-863-0025Email: relliot@inovadx.com
Device name (assay kit):Proprietary name:QUANTA Flash® RF IgM ReagentsQUANTA Flash® RF IgA Reagents
Common name:Rheumatoid factor IgM, IgA ChemiluminescenceImmunoassay
Classification name:System, Test, Rheumatoid Factor
Regulation DescriptionRheumatoid Factor immunological test system
Regulation Medical SpecialtyImmunology
Review PanelImmunology
Product CodeDHR
Regulation Number866.5775
Device Class2

{5}------------------------------------------------

Predicate device

QUANTA Lite® RF IgM ELISA, 510(k) number: K971614. Date declared: April 30, 1997. QUANTA Lite® RF IgA ELISA, 510(k) number: K983084. Date declared: September 2, 1998.

Device description

The principle of the assays is chemiluminescent microparticle immunoassay, a variation of solid phase immunoassay. The QUANTA Flash® RF IgM and QUANTA Flash® RF IgA assays are designed to run on the BIO-FLASH® instrument. This platform is a fully automated closed system with continuous load and random access capabilities that automatically processes the samples, runs the assay and reports the results. It includes liquid handling hardware, luminometer and computer with software-user interface. The QUANTA Flash® RF IgM and QUANTA Flash® RF IgA assays utilize a reagent cartridge format, which is compatible with the BIO-FLASH® instrument.

Rabbit polyclonal antibodies are coated onto paramagnetic beads, which are stored in the reagent cartridge under conditions that preserve the antibody in its reactive state. When the assay cartridge is ready to be used for the first time, the entire cartridge is inverted several times to thoroughly mix the reagents. The reagent cartridge is then loaded onto the BIO-FLASH instrument.

A patient serum sample is diluted 1:22.7 by the instrument using system rinse in a disposable plastic cuvette. An aliquot of the diluted patient serum, coupled beads, and assay buffer are combined into a second cuvette, and mixed. This cuvette is incubated at 37°C. The beads are then magnetized and washed several times. Isoluminol conjugated anti-human IgM (QUANTA Flash® RF IgM) or anti-human lgA (QUANTA Flash® RF IgA) antibody is then added to the cuvette, and incubated at 37°C. Again, the beads are magnetized and washed repeatedly. The isoluminol conjugate produces a luminescent reaction when "Trigger" reagents are added to the light produced from this reaction is measured as Relative Light Units (RLU) by the BIO-FLASH optical system. RLU values are proportional to the amount of bound isoluminol conjugate, which in turn is proportional to the amount of RF antibodies bound to the antibodies on the beads.

The QUANTA Flash RF IgM and QUANTA Flash RF IgA assays utilize a predefined lot specific Master Curve that is uploaded into the instrument through the reagent cartridge barcode. Based on the results obtained by running the Calibrators, an instrument specific Working Curve is created, which is used by the software to calculate international units per milliliter (IU/mL) (QUANTA Flash® RF IgM) or chemiluminescent units (CU) (QUANTA Flash® RF IgA) from the RLU value obtained for each sample.

QUANTA Flash RF IgM Calibrators, QUANTA Flash RF IgM Controls, QUANTA Flash RF IgA Calibrators and QUANTA Flash RF IgA Controls are sold separately.

The QUANTA Flash® RF IgM Reagents / QUANTA Flash® RF IgA Reagents kit contains the following materials:

One (1) QUANTA Flash RF IgM / RF IgA Reagent Cartridge

{6}------------------------------------------------

QUANTA Flash RF IgM Reagent Cartridge contains the following reagents for 100 determinations:

  • a. Rabbit pAb coated paramagnetic beads.
  • b. Assay buffer - colored pink, containing protein stabilizers and preservatives.
  • Tracer IgM Isoluminol labeled anti-human IgM antibody, containing buffer, protein C. stabilizers and preservative.

QUANTA Flash RF IgA Reagent Cartridge contains the following reagents for 100 determinations:

  • a. Rabbit pAb coated paramagnetic beads.
  • b. Assay buffer - colored pink, containing protein stabilizers and preservatives.
  • Tracer IgA Isoluminol labeled anti-human IgA antibody, containing buffer, protein C. stabilizers and preservative.

Intended use(s)

QUANTA Flash RF IgM is a chemiluminescent immunoassay for the quantitative determination of IgM rheumatoid factor (RF) antibodies in human serum. The presence of IgM RF antibodies, in conjunction with clinical findings and other laboratory tests, is an aid in the diagnosis of rheumatoid arthritis (RA).

QUANTA Flash RF IgA is a chemiluminescent immunoassay for the semi-quantitative determination of lgA rheumatoid factor (RF) antibodies in human serum. The presence of IgA RF antibodies, in conjunction with clinical findings and other laboratory tests, is an aid in the diagnosis of rheumatoid arthritis (RA).

Indications for use

Same as Intended use.

Substantial equivalence

The QUANTA Flash RF IgM Reagents and the QUANTA Flash RF IgA Reagents have the same intended use and assay principle as the predicate device.

{7}------------------------------------------------

Comparison to predicate device

QUANTA Flash RF IgM Reagents

Similarities
ItemQUANTA Flash RF IgM ReagentsQUANTA Lite RF IgM ELISA
Intended useQUANTA Flash RF IgM is achemiluminescent immunoassay for thequantitative determination of IgMrheumatoid factor (RF) antibodies inhuman serum. The presence of IgM RFantibodies, in conjunction with clinicalfindings and other laboratory tests, is anaid in the diagnosis of rheumatoidarthritis (RA).QUANTA Lite RF IgM is an enzyme-linked immunosorbent assay (ELISA) forthe semi-quantitative detection of IgMrheumatoid factor (RF) antibodies inpatient sera. The presence of theseantibodies, when considered inconjunction with other laboratory andclinical findings, is an aid in thediagnosis of rheumatoid arthritis (RA).
Assay methodologySolid phase (heterogeneous)immunoassaySolid phase (heterogeneous)immunoassay
AntigenRabbit polyclonal antibodyRabbit polyclonal antibody
Sample typeHuman serumHuman serum
Differences
ItemQUANTA Flash RF IgM ReagentsQUANTA Lite RF IgM ELISA
Detection/Operating principleChemiluminescent immunoassayEnzyme-linked immunosorbent assay
Solid phaseParamagnetic microparticles (beads)96-well polystyrene plate
ConjugateIsoluminol conjugated monoclonal anti-human IgM antibodyHRP conjugated monoclonal anti-human IgM antibody
UnitsInternational Units per milliliter (IU/mL)Units (U)
Cut-off5.0 IU/mL6.0 Units
AnalyticalMeasuring Range0.3 - 490.0 IU/mL0.0 – 100.0 Units
CalibrationLot specific Master Curve + Twocalibrators (sold separately)Five lot specific calibrators

{8}------------------------------------------------

Similarities
ItemQUANTA Flash RF IgA ReagentsQUANTA Lite RF IgA ELISA
Intended useQUANTA Flash RF IgA is achemiluminescent immunoassay for thesemi-quantitative determination of IgArheumatoid factor (RF) antibodies inhuman serum. The presence of IgA RFantibodies, in conjunction with clinicalfindings and other laboratory tests, is anaid in the diagnosis of rheumatoidarthritis (RA).QUANTA Lite RF IgA is an enzyme-linked immunosorbent assay (ELISA) forthe semi-quantitative detection of IgArheumatoid factor (RF) antibodies inpatient sera. The presence of theseantibodies, when considered inconjunction with other laboratory andclinical findings, is an aid in thediagnosis of rheumatoid arthritis (RA).
Assay methodologySolid phase (heterogeneous)immunoassaySolid phase (heterogeneous)immunoassay
AntigenRabbit polyclonal antibodyRabbit polyclonal antibody
Sample typeHuman serumHuman serum

QUANTA Flash RF IgA Reagents

Differences
ItemQUANTA Flash RF IgA ReagentsQUANTA Lite RF IgA ELISA
Detection/Operating principleChemiluminescent immunoassayEnzyme-linked immunosorbent assay
Solid phaseParamagnetic microparticles (beads)96-well polystyrene plate
ConjugateIsoluminol conjugated monoclonal anti-human IgA antibodyHRP conjugated monoclonal anti-human IgA antibody
UnitsChemiluminescent Units (CU)Units (U)
Cut-off20.0 CU6.0 Units
AnalyticalMeasuring Range1.3 – 900.0 CU0.0 – 100.0 Units
CalibrationLot specific Master Curve + Twocalibrators (sold separately)Five lot specific calibrators

{9}------------------------------------------------

Analytical performance characteristics

Quantitation and units of measure

For quantitation, the QUANTA Flash RF IgM and QUANTA Flash RF IgA assays utilize a lot specific Master Curve that is uploaded onto the instrument through the reagent cartridge barcode. The Master Curve for QUANTA Flash RF IgM and QUANTA Flash RF IgA consist of 6 different Standards each. These Master Curve Standards are used to create the lot specific Master Curve during the manufacturing procedure.

List of QUANTA Flash RF IgM Standards:

MaterialAssigned Value
RF IgM Master Curve Standard 10.0 IU/mL
RF IgM Master Curve Standard 21.9 IU/mL
RF IgM Master Curve Standard 37.8 IU/mL
RF IgM Master Curve Standard 431.0 IU/mL
RF IgM Master Curve Standard 5124.1 IU/mL
RF IgM Master Curve Standard 6496.5 IU/mL

List of QUANTA Flash RF IgA Standards:

MaterialAssigned Value
RF IgA Master Curve Standard 10.0 CU
RF IgA Master Curve Standard 27.2 CU
RF IgA Master Curve Standard 314.4 CU
RF IgA Master Curve Standard 457.5 CU
RF IgA Master Curve Standard 5230.0 CU
RF IgA Master Curve Standard 6919.9 CU

Precision

The precision of the QUANTA Flash RF IgM and QUANTA Flash RF IgA assays was evaluated on 10 samples for the IgM subtype and 9 samples for the IgA one, containing various concentrations of RF antibodies in accordance with CLSI EP05-A3, Evaluation of Quantitative Measurement Procedures; Approved Guideline. Samples were run in duplicates, twice a day, for 20 days.

Data were analyzed with the Analyse-it for Excel method evaluation software, and repeatability (withinrun), between run, between day and within-laboratory precision) were calculated. Acceptance criteria: Total %CV: < 12%

{10}------------------------------------------------

QUANTA Flash RF IgMRepeatabilityBetween-RunBetween-DayWithin-Laboratory Precision
Sample IDNMean (IU/mL)SD (IU/mL)CV (%)SD (IU/mL)CV (%)SD (IU/mL)CV (%)SD (IU/mL)CV (%)
1801.20.075.4%0.043.1%0.043.3%0.087.0%
2802.30.094.1%0.125.2%0.041.6%0.156.8%
3802.70.103.8%0.103.6%0.103.5%0.176.3%
4803.70.174.6%0.123.2%0.082.2%0.226.0%
5805.40.173.2%0.173.2%0.152.9%0.295.4%
68020.60.824.0%0.381.8%0.582.8%1.075.2%
78083.92.883.4%2.713.2%2.112.5%4.495.3%
880181.38.954.9%8.624.8%0.000.0%12.436.9%
980379.316.934.5%22.475.9%4.241.1%28.457.5%
1080408.819.254.7%21.635.3%10.832.6%30.917.6%

Results are summarized in the tables below.

QUANTA Flash RF IgARepeatabilityBetween-RunBetween-DayWithin-LaboratoryPrecision
SampleIDNMean(CU)SD(CU)CV(%)SD(CU)CV(%)SD(CU)CV(%)SD(CU)CV(%)
18010.60.373.5%0.302.8%0.403.7%0.625.8%
28011.20.595.3%0.040.4%0.221.9%0.635.6%
38020.90.844.0%0.000.0%0.673.2%1.085.2%
48021.60.723.3%0.271.2%0.542.5%0.944.3%
58048.72.084.3%0.691.4%0.551.1%2.264.6%
680107.02.532.4%2.852.7%2.252.1%4.424.1%
780415.811.172.7%5.801.4%8.782.1%15.353.7%
880721.131.704.4%15.992.2%30.784.3%46.996.5%
980749.325.303.4%20.222.7%21.922.9%39.115.2%

Reproducibility Studies

Reproducibility between sites (instruments)

Eight samples were tested according to CLSI EP05-A3 Evaluation of Quantitative Measurement Procedures; Approved Guideline, at three different sites. Samples were run in replicates of 5, once a day, for 5 days, to generate 25 data points per site. Data were analyzed with the Analyse-it for Excel method evaluation software to calculate between site precision.

Acceptance criteria: Reproducibility %CV: < 12%

{11}------------------------------------------------

QUANTA Flash RF IgMRepeatabilityBetween-DayWithin-SiteBetween-SiteReproducibility
SampleNMeanSDCVSDCVSDCVSDCVSDCV
ID(IU/mL)(IU/mL)(%)(IU/mL)(%)(IU/mL)(%)(IU/mL)(%)(IU/mL)(%)
1751.70.16.4%0.14.4%0.17.8%0.12.7%0.18.2%
2753.00.14.0%0.11.7%0.14.4%0.27.5%0.38.7%
3754.60.23.3%0.24.0%0.25.2%0.12.2%0.35.6%
4756.30.22.9%0.34.1%0.35.0%0.57.8%0.69.3%
57523.20.93.9%0.72.9%1.14.9%0.21.1%1.25.0%
67592.82.42.6%2.93.1%3.74.0%5.15.5%6.36.8%
775200.39.44.7%10.75.4%14.27.1%4.92.4%15.17.5%
875412.915.53.8%17.34.2%23.35.6%36.28.8%43.010.4%

Results are summarized in the tables below.

QUANTA Flash RF IgARepeatabilityBetween-DayWithin-SiteBetween-SiteReproducibility
SampleNMean(CU)SD(CU)CV(%)SD(CU)CV(%)SD(CU)CV(%)SD(CU)CV(%)SD(CU)CV(%)
17511.20.54.2%0.43.3%0.65.3%0.43.1%0.76.2%
27511.30.54.0%0.22.1%0.54.5%0.32.5%0.65.2%
37521.90.73.4%0.10.4%0.73.4%0.62.9%1.04.5%
47523.20.93.9%0.21.0%0.94.0%0.83.6%1.35.4%
57551.71.32.5%0.71.2%1.52.8%0.50.9%1.52.9%
675112.63.22.9%2.32.1%4.03.5%0.00.0%4.03.5%
775462.413.22.9%10.92.4%17.13.7%19.84.3%26.15.6%
875738.427.33.7%2.53.0%35.34.8%45.36.1%57.47.8%

Reproducibility between lots

Eight samples were tested according to CLSI EP05-A3 Evaluation of Quantitative Measurement Procedures; Approved Guideline, using three different lots. Samples were run in replicates of 5, once a day, for 5 days, to generate 25 data points per lot, 75 data points total for each sample. Data were analyzed with the Analyse-it for Excel method evaluation software to calculate between lot precision.

Acceptance criteria: Reproducibility %CV: < 12%

{12}------------------------------------------------

QUANTA Flash RF IgMRepeatabilityBetween-DayWithin-LotBetween-LotReproducibility
SampleNMean(IU/mL)SD(IU/mL)CV(%)SD(IU/mL)CV(%)SD(IU/mL)CV(%)SD(IU/mL)CV(%)SD(IU/mL)CV(%)
1752.90.13.8%0.01.6%0.14.1%0.01.0%0.14.2%
2753.30.13.1%0.12.2%0.13.8%0.13.4%0.25.1%
3754.60.23.6%0.13.0%0.24.7%0.00.0%0.24.7%
47523.20.83.4%0.73.1%1.14.6%0.41.6%1.14.8%
575209.89.84.7%14.06.7%17.18.2%9.84.7%19.79.4%
6751.60.15.2%0.15.5%0.17.6%0.14.1%0.18.7%
77592.93.84.1%2.42.6%4.54.9%4.85.2%6.67.1%
8756.00.23.8%0.22.9%0.34.8%0.11.4%0.35.0%

Results are summarized in the tables below.

QUANTA Flash RF IgARepeatabilityBetween-DayWithin-LotBetween-LotReproducibility
SampleIDNMean(CU)SD(CU)CV(%)SD(CU)CV(%)SD(CU)CV(%)SD(CU)CV(%)SD(CU)CV(%)
17511.70.54.3%0.43.8%0.75.7%0.00.0%0.75.7%
27522.60.73.0%0.62.7%0.94.0%0.20.9%0.94.1%
37522.50.62.9%0.41.8%0.83.4%0.62.6%1.04.3%
47552.91.32.5%0.81.6%1.63.0%0.40.8%1.63.1%
575487.013.62.8%11.52.4%17.83.7%7.21.5%19.23.9%
67511.40.43.4%0.22.0%0.54.0%0.00.3%0.54.0%
775113.23.32.9%2.32.0%4.03.5%0.00.0%4.03.5%
875722.721.93.0%23.73.3%32.34.5%18.72.6%37.35.2%

Limit of Blank (LoB), Limit of Detection (LoD), and Limit of Quantitation (LoQ)

The LoD of the QUANTA Flash RF IgM and QUANTA Flash RF IgA assays are 0.1 IU/mL and 0.5 CU respectively, which are below the analytical measuring range of the respective assays. The LoD of each assay was determined by using two reagent lots, consistent with CLSI EP17-A2 guideline with proportions of false positives (alpha) less than 5% and false negatives (beta) less than 5%; based on 240 determinations, with 60 measurements on blank samples and 60 measurements of low level samples, per reagent lot. The LoB of the QUANTA Flash RF IgM is 0.0 IU/mL (450 RLU) while the LoB of the QUANTA Flash RF IgA is 0.3 CU (420 RLU).

Four low level samples were tested in replicates of five on two reagent lots, once per day, for 3 days, obtaining 30 data points per sample to generate data used to calculate the LoQ for the QUANTA Flash RF IgM and QUANTA Flash RF IgA assays. The LoQ was determined in each case by calculating the total imprecision of each sample.

Acceptance criteria: Total imprecision CV% <20%.

{13}------------------------------------------------

510(k) Summary QUANTA Flash® RF IgA Reagents

The LoQ for the QUANTA Flash RF IgM assay has been found to be at 0.3 IU/mL. The LoQ for the QUANTA Flash RF IgA assay has been found to be at 1.2 CU. Even though the LoQ for the QUANTA Flash RF IgA has been found to be at 1.2 CU, the AMR of the QUANTA Flash RF IgA will start at 1.3 CU.

Analytical Measuring Range (AMR)

QUANTA Flash RF IgM:0.3 IU/mL - 490.0 IU/mL
QUANTA Flash RF IgA:1.3 CU - 900.0 CU

Auto-rerun function and reportable results

The BIO-FLASH software has an auto-rerun option available. If this option is selected, the instrument will automatically rerun any sample that has a result of >490.0 IU/mL, for the QUANTA Flash RF IgM, or >900.0 CU, for the QUANTA Flash RF IgA, after further diluting it by 20 fold, thereby bringing the measured value within the AMR. The final result will be calculated by the software by taking into account the additional dilution factor.

For the QUANTA Flash RF IgM, the highest value that can be directly measured is 490.0 IU/mL, so the highest value that can be reported is 9,800.0 IU/mL.

For the QUANTA Flash RF IgA, the highest value that can be directly measured is 900.0 CU, so the highest value that can be reported is 18,000.0 CU.

High concentration hook effect

To assess hook effect, measurement signal in relative light units (RLU) was examined by performing serial dilutions of two high positive samples (with results above the AMR of each assay when tested as neat samples). RLU values showed increasing antibody concentrations above the AMR for both assays, thereby confirming that high positive specimens above the AMR do not show hook effect up to 1,830.6 IU/mL for the QUANTA Flash RF IgM assay and 42,710.4 CU for the QUANTA Flash RF IgA assay (theoretical value calculated using the highest value in the AMR and its dilution factor).

Linearity

The linearity of the AMR of the QUANTA Flash RF IgM and QUANTA Flash RF IgA was evaluated by a study according to CLSI EP06-A, Evaluation of the Linearity of Quantitative Measurement Procedures: A Statistical Approach; Approved Guideline. The linearity was evaluated using four human serum samples for the QUANTA Flash RF IgM and three human serum samples for the QUANTA Flash RF IgA with various RF antibody concentrations which were combined with another human serum sample containing low levels of RF antibodies in 10% increments (from 0% to 90% of low sample) to obtain values that cover the entire AMR of the assays. The dilutions were assayed in duplicates. Results were analyzed according to the guideline performing regression analysis and identifying the best fitting polynomial.

{14}------------------------------------------------

Acceptance criteria:

  • Best fitting polynomial is a linear one, otherwise, the difference between the best-fitting nonlinear and linear polynomial is less than 15% or ±0.75 IU/mL / ±3 CU for negative samples (allowable nonlinearity).

For the QUANTA Flash RF IgM, all samples have been found that the best fitting polynomial is a linear one except Sample 3, where the best fitting polynomial found was a second order polynomial. The nonlinearity for Sample 3 ranged from -5.2% to 3.8% and from -0.1 IU/mL to 0.6 IU/mL, fulfilling the acceptance criteria.

SerumSamplesTest Range(IU/mL)Slope(95% CI)Y-Intercept(95% CI)Average %Recovery
160.5 – 544.10.92(0.88 – 0.96)25.9(11.9 – 39.9)0.99104.0%
210.8 – 108.31.04(0.99 - 1.10)0.6(-2.8 - 4.0)0.99105.6%
31.5 - 15.31.00(0.95 - 1.05)-0.6(-1.1 – -0.1)0.9991.2%
40.2 – 1.51.07(0.99 - 1.14)-0.1(-0.1 - 0.0)0.9896.7%
Combined0.2 – 544.10.98(0.97 – 0.99)2.7(0.5 - 4.9)1.0099.3%

All four samples showed dilution linearity individually and in combination.

For the QUANTA Flash RF IgA, all samples have been found that the best fitting polynomial is a linear one except Sample 1, where the best fitting polynomial found was a third order polynomial. The nonlinearity for Sample 1 ranged from -10.8% to 5.6%, fulfilling the acceptance criteria.

SerumSamplesTest Range(CU)Slope(95% CI)Y-Intercept(95% CI)R 2Average %Recovery
1104.5 - 1045.01.00(0.96 - 1.04)12.8(-12.4 - 38.1)0.99102.9%
210.8 - 107.70.94(0.90 - 0.98)5.1(2.7 - 7.6)0.99106.6%
31.1 - 11.10.97(0.92 - 1.02)0.3(-0.1 - 0.6)0.99104.9%
Combined1.1 - 1045.01.01(1.00 - 1.02)1.7(-2.8 - 6.2)1.00104.8%

All three samples showed dilution linearity individually and in combination.

These data demonstrate the linearity of the analytical measuring range (0.3 IU/mL – 490.0 IU/mL) of the QUANTA Flash RF IgM assay and the analytical measuring range (1.3 CU – 900.0 CU) of the QUANTA Flash RF IgA assay.

{15}------------------------------------------------

Interference

The interference study was performed according to CLSI EPO7-A2, Interference Testing in Clinical Chemistry; Approved Guideline - Second Edition. A set of four human serum specimens, one high positive, one low positive, one near the cutoff and one negative sample were tested using the following interfering substances (bilirubin, hemoglobin, triglycerides, cholesterol, human lgG, ascorbic acid, methotrexate and prednisone). All interferents were spiked into every specimen in 10% of total specimen volume, and the resulting samples were assessed in triplicates with the QUANTA Flash RF IgM and the QUANTA Flash RF IgA assays. Recovery of the unit values was calculated compared to control samples spiked with the same volume of diluents (10% of total sample volume). Acceptance criteria for the interference studies were 85% - 115% recovery, or ± 15% of the cut-off (±0.75 IU/mL or ±3 CU) difference, whichever is greater.

No interference was detected with any of the assays with bilirubin up to 1 mg/mL (recovery: from 89.0% to 98.2% and from 92.5 to 93.1 for the IgM and IgA subtypes respectively), hemoglobin up to 2 mg/mL (recovery: from 94.5% to 95.0% and from 90.9 to 94.7 for the IgM and IgA subtypes respectively), triglycerides up to 1000 mg/dL (recovery: from 91.2% to 112.2% and from 97.8 to 105.4 for the IgM and lgA subtypes respectively), cholesterol up to 332.5 mg/dL (recovery: from 89.7% to 91.6% and from 91.0 to 91.0 for the IgM and IgA subtypes respectively), human IgG up to 70 mg/mL (recovery: from 86.8% to 89.3% and from 91.6 to 95.7 for the IgA subtypes respectively), ascorbic acid up to 60.1 mg/L (recovery: from 95.6% to 101.7% and from 98.1 to 102.6 for the IgA subtypes respectively), methotrexate up to 9.1 mg/mL (recovery: from 96.1% to 102.8% and from 92.9 to 105.8 for the IgM and lgA subtypes respectively), and prednisone up to 0.3 mg/L (recovery: from 101.9% to 109.5% and from 93.0 to 94.9 for the IgM and IgA subtypes respectively).

Sample Stability and Handling

For the QUANTA Flash RF IgM, five human serum samples, encompassing negative (n=1), around the cut-off (n=1), and positive samples (n=3), and for the QUANTA Flash RF IgA four human serum samples, encompassing negative (n=1), around the cut-off (n=1), and positive samples (n=2), were tested in duplicates for up to 21 days while stored at 2-8°C, up to 48 hours while stored at room temperature, and after repeated freeze/thaw cycles up to 3 cycles.

Results were compared to those obtained on control samples (time zero / zero cycles).

Acceptance criteria: recovery is between 85-115% for positive samples, and between 80-120% for negative samples (<5 IU/mL or <20 CU).

All samples fulfilled the acceptance criteria at each time point for each condition. Based on these results, we recommend that samples are stored up to 48 hours at room temperature, up to 14 days at 2-8°C, and can be subjected to up to 3 freeze/thaw cycles (when samples are stored at or below -20°C).

{16}------------------------------------------------

Reagent Stability

Shelf life

To establish the initial claim for shelf life, accelerated stability studies were performed for 3 weeks at 37 °C, where one week is equal to six months at 5 ± 3°C.

Accelerated stability testing was performed on each of the following sealed components to establish initial stability claim:

• RF IgM beads(3 Lots)
• Anti-IgM tracer(3 Lots)
• RF IgA beads(3 Lots)

Each week a new sealed component was placed in the incubator, and all components were tested at the end of the experiment together with the one that was stored at 5 ± 3°C. The recovery of the measured values was calculated for each time point (compared to those obtained with 5 ± 3°C stored reagent). All calculations were performed by comparing results of sealed components stored at 5 ± 3℃ (control) to those stored at 37 ± 3℃ (test) for 1, 2 and 3 weeks, where one week is equal to six months at 5 ± 3℃. Linear regression analysis was performed between recovery values and the number of days.

Acceptance criteria for one year preliminary expiration dating:

With regression analysis, the lower and upper 95% Cl interval of the regression line is between 80% and 120% recovery at day 14.

All components tested fulfilled the acceptance criteria above, so one year expiration dating was assigned to each component

In-use (onboard) stability

Reagent Cartridge

To establish the in-use stability of the QUANTA Flash RF IgM and QUANTA Flash RF IgA reagent cartridges, two lots of reagent cartridge were tested using up to 10 human serum samples (with different reactivity levels). The specimens were tested periodically for of 91 days. Percent recoveries were calculated compared to the day zero average values, and linear regression analysis was performed by plotting percent recovery against the number of days. The claim was established using the following criteria (using the one that is fulfilled first):

  • The stability claim is established at the actual measurement day proceeding the 95% confidence interval of the regression line reaches 85% or 115% recovery, or

  • At the actual measurement day preceding the day when ≥2% of the recovery data, (3 data points) is <75% or ≥125% recovery.

{17}------------------------------------------------

The onboard stability results for the QUANTA Flash RF IgM are as follows: Lot RP0005: 83 days Lot RP0006: 90 days Using these criteria, the in-use (onboard) stability of the QUANTA Flash RF IgM reagent cartridge was set at 80 days.

The onboard stability results for the QUANTA Flash RF IgA are as follows: Lot RP0003: 90 days

Lot RP0004: 84 days

Using these criteria, the in-use (onboard) stability of the QUANTA Flash RF IgA reagent cartridge was set at 80 days.

Real time stability

Real time stability testing has been scheduled to be performed every three or six months on the QUANTA Flash RF IgM and QUANTA Flash RF IgA Reagents kits, to verify the one year expiration that was assigned based on accelerated stability studies. At the time of the submission, results were available up to 13 months for QUANTA Flash RF IgM Reagent kits and for QUANTA Flash RF IgA Reagent kits.

A negative sample (Negative Control), a low positive sample (Positive Control), and a high positive sample were tested in replicates of 6 (replicates of 9 at time zero) at each time point.

  • Acceptance criteria: results should fall within their respective ranges.

All results to date were within the acceptance limits.

{18}------------------------------------------------

Cut-off, reference range

QUANTA Flash RF IgM:
Negative<5 IU/mL
Positive≥5 IU/mL
QUANTA Flash RF IgA:
Negative<20 CU
Positive≥20 CU

The reference population for establishing the reference interval for the QUANTA Flash RF IgM and QUANTA Flash RF IgA assays consisted of 191 subjects:

Sample GroupN
Apparently healthy donors117
Infectious Disease Controls (HBV, HCV, HIV, Syphilis)27
False Positive Cohort (high reactivity samples)13
Antiphospholipid syndrome (APS)12
Systemic Lupus Erythematosus (SLE)9
Celiac Disease8
Vasculitis5

All specimens were the same matrix (human serum) as specified in the Intended Use. All specimens were unaltered. The cut-off values were established in accordance to CLSI EP28-A3c: Defining, Establishing, and Verifying Reference Intervals in the Clinical Laboratory; Approved Guideline - Third Edition. The Analyse-it for Excel software was used to make the calculations. The distribution of the results was non-normal (Shapiro-Wilk p<0.0001), so the non-parametric percentile method was used.

Additionally, 42 diagnosed rheumatoid arthritis (RA) patient specimens were assayed to aid in the determination of the cutoff values.

For the QUANTA Flash RF IgM, based on the distribution of result values and in these (known) positive samples, a cutoff of 5 IU/mL (16,300 RLU) has been set to ensure optimal differentiation between negatives and positives samples.

For the QUANTA Flash RF IgA, based on the distribution of result values and in these (known) positive samples, the cutoff was established at 6,000 RLU to ensure optimal differentiation between negatives and positives samples. The cut-off was established at a value greater than the 95th percentile of the control results (3,767 RLU) and it was assigned a value of 20 CU.

{19}------------------------------------------------

Clinical performance characteristics

Clinical sensitivity, specificity

A cohort of characterized samples, none of which were used for establishing the reference range, was used to validate the clinical performance of the QUANTA Flash RF IgM and the QUANTA Flash RF IgA assays. A total of 706 characterized samples were included in this Validation Set. All samples were run on the QUANTA Flash RF IgM and the QUANTA Flash RF IgA assays. The distribution of the cohort and the RF positivity rate is in the Table below:

Patient GroupNRF IgMN PositiveRF IgM% PositiveRF IgAN PositiveRF IgA% Positive
Ulcerative Colitis50816.0%1020.0%
Crohn's Disease30310.0%26.7%
Osteoarthritis30310.0%00.0%
Autoimmune Hepatitis3000.0%13.3%
Sjögren's Syndrome30723.3%620.0%
Systemic Sclerosis30930.0%1033.3%
Celiac Disease2900.0%00.0%
Systemic Lupus Erythematosus28414.3%414.3%
Hepatitis B Virus2015.0%00.0%
Hepatitis C Virus1900.0%15.3%
Polymyalgia Rheumatica15213.3%00.0%
Psoriatic Arthritis1200.0%00.0%
Syphilis1218.3%00.0%
Parvo virus infection10220.0%220.0%
Lyme disease1000.0%00.0%
Fibromyalgia10110.0%00.0%
Ankylosing Spondylitis800.0%00.0%
Mixed connective tissue disease8112.5%112.5%
Hashimoto's disease800.0%00.0%
Grave's disease6116.7%00.0%
Polymyositis10330.0%220.0%
Dermatomyositis5240.0%00.0%
Total Controls4104811.7%399.5%
Rheumatoid Arthritis29620669.6%16856.8%
Total706----

{20}------------------------------------------------

510(k) Summary QUANTA Flash® RF IgM Reagents

510(k) Summary QUANTA Flash® RF IgA Reagents
Clinical Analysis (N=706)QUANTA Flash RF IgMAnalysis(95% confidence)
DiagnosisPositiveNegativeTotal
RA20690296Sensitivity: 69.6% (64.1 – 74.6%)
Controls48362410Specificity: 88.3% (84.8 - 91.1%)
Total254452706

Clinical sensitivity and specificity of the QUANTA Flash RF IgM were analyzed in the table below:

Clinical sensitivity and specificity of the QUANTA Flash RF IgA were analyzed in the table below:

Clinical Analysis (N=706)QUANTA Flash RF IgAAnalysis(95% confidence)
PositiveNegativeTotal
DiagnosisRA168128296Sensitivity: 56.8% (51.1 – 62.3%)
Controls39371410Specificity: 90.5% (87.3 – 93.0%)
Total207499706

Expected values

QUANTA Flash RF IgM

The expected value in the normal population is "negative". RF IgM antibody levels were analyzed using the QUANTA Flash RF IgM assay on a panel of 100 apparently healthy blood donors (50 females/50 males, ages 24 to 70 years, with an average and median age of 50 and 51 years respectively). With a cutoff of 5 IU/mL, six samples (6.0%) were positive on the QUANTA Flash RF IgM. The mean and median concentration were 3.6 and 0.5 IU/mL respectively, and the values ranged from <0.3 to 219.1 IU/mL.

QUANTA Flash RF IgA

The expected value in the normal population is "negative". RF IgA antibody levels were analyzed using the QUANTA Flash RF IgA assay on a panel of 100 apparently healthy blood donors (50 females/50 males, ages 24 to 70 years, with an average and median age of 50 and 51 years respectively). With a cutoff of 20 CU, six samples (6.0%) were positive on the QUANTA Flash RF IgA. The mean and median concentration were 8.2 and 2.9 CU respectively, and the values ranged from <1.3 to 324.9 CU.

{21}------------------------------------------------

Comparison with predicate device

Samples for method comparison analysis included all samples from the clinical validation study. These samples were tested on both the QUANTA Flash RF assays and on their predicate ELISA.

Of the 706 samples tested, 577 samples fell within the AMR of the QUANTA Flash RF IgM assay, while 612 fell within the AMR of the QUANTA Flash RF IgA assay.

Method ComparisonQUANTA Flash RF IgMPercent Agreement
Within AMR (N=577)NegativePositiveTotal(95% Confidence)
PredicateELISANegative29111302NPA: 96.4 (93.6-98.0)
Positive52223275PPA: 81.1 (76.0 - 85.3)
Total343234577TPA: 89.1 (86.3 - 91.4)

Method comparison of the QUANTA Flash RF IgM with the predicate device using samples within the AMR.

PPA= Positive Percent Agreement; NPA= Negative Percent; TPA= Total Percent Agreement

Additionally, a quantitative comparison has been performed on the 577 samples in the AMR of the QUANTA Flash RF IgM assay utilizing a Spearman correlation. The results revealed a Spearman's rs of 0.85 (95% Cl, 0.82 - 0.87).

Method comparison of the QUANTA Flash RF IgA with the predicate device using samples within the AMR.

Method ComparisonWithin AMR (N=612)QUANTA Flash RF IgAPercent Agreement(95% Confidence)
NegativePositiveTotal
PredicateELISANegative38510395NPA: 97.5 (95.4–98.6)
Positive27190217PPA: 87.6 (82.5 – 91.3)
Total412200612TPA: 94.0 (91.8 – 95.6)

PPA= Positive Percent Agreement; NPA= Negative Percent Agreement; TPA= Total Percent Agreement

§ 866.5775 Rheumatoid factor immunological test system.

(a)
Identification. A rheumatoid factor immunological test system is a device that consists of the reagents used to measure by immunochemical techniques the rheumatoid factor (antibodies to immunoglobulins) in serum, other body fluids, and tissues. Measurement of rheumatoid factor may aid in the diagnosis of rheumatoid arthritis.(b)
Classification. Class II (performance standards).