(267 days)
Yumizen C1200 Ferritin reagent is intended for the quantitative in vitro diagnostic determination of Ferritin in human serum by latex-enhanced immunoturbidimetric assay. Measurements of ferritin aid in the diagnosis of diseases affecting iron metabolism, hemochromatosis (iron overload) and iron deficiency anemia.
Yumizen C1200 Transferrin reagent is intended for the quantitative in vitro diagnostic determination of transferrin in human serum and lithium heparin plasma by turbidimetry.
Measurement of transferrin levels ads in the diagnosis of malnutrition, acute inflammation, infection, and iron deficiency anemia.
Yumizen C1200 Rheumatoid Factor reagent is intended for the quantitative in vitro diagnostic determination of rheumatoid factor in human serum by latex-enhanced immunoturbidimetric assay. Measurement of rheumatoid factor may aid in the diagnosis of rheumatoid arthritis.
Not Found
The document describes the analytical performance characteristics of three devices: Yumizen C1200 Ferritin, Yumizen C1200 Transferrin, and Yumizen C1200 Rheumatoid Factor. Each device is intended for the quantitative in vitro diagnostic determination of specific substances in human serum, and sometimes plasma, using immunoturbidimetric or turbidimetric assays.
Here's an analysis of the acceptance criteria and study details for each device:
Yumizen C1200 Ferritin
1. Table of Acceptance Criteria and Reported Device Performance
| Performance Characteristic | Acceptance Criteria | Reported Device Performance |
|---|---|---|
| Measuring Range (Serum) | N/A (claimed measuring range is appropriate based on LOD, LOQ, and linearity studies) | 10 to 450 ng/mL |
| Limit of Detection (Serum) | N/A (determined according to CLSI EP17-A2) | 6.30 ng/mL |
| Limit of Quantitation (Serum) | N/A (determined according to CLSI EP17-A2) | 9.39 ng/mL |
| Linearity (Serum) | N/A (determined according to CLSI EP06-A) | Evaluated from 13.3 to 426.6 ng/mL (appropriate) |
| Total Precision (Analyzer Variability) - Within Run CV | Low level: ≤ 8.0%Middle level: ≤ 6.0%High level: ≤ 6.0% | Level 1 Control (47.58 ng/mL): 3.6%Level 2 Control (279.31 ng/mL): 1.1%Sample 1 (29.56 ng/mL): 5.5%Sample 2 (50.87 ng/mL): 4.1%Sample 3 (172.63 ng/mL): 1.4%Sample 4 (328.60 ng/mL): 1.3%Sample 5 (403.21 ng/mL): 1.0% |
| Total Precision (Analyzer Variability) - Total CV | Low level: ≤ 10.0%Middle & High level: ≤ 8.0% | Level 1 Control: 4.9%Level 2 Control: 2.1%Sample 1: 7.9%Sample 2: 5.1%Sample 3: 1.9%Sample 4: 4.3%Sample 5: 1.4% |
| Total Precision (Lot to Lot Variability) - Within Run CV | Low level: ≤ 8.0%Middle level: ≤ 6.0%High level: ≤ 6.0% | Level 1 Control (52.84 ng/mL): 4.6%Level 2 Control (281.87 ng/mL): 0.9%Sample 1 (19.09 ng/mL): 8.8%Sample 2 (34.05 ng/mL): 6.5%Sample 3 (51.53 ng/mL): 3.6%Sample 4 (192.31 ng/mL): 1.4%Sample 5 (407.38 ng/mL): 0.9% |
| Total Precision (Lot to Lot Variability) - Total CV | Low level: ≤ 10.0%Middle & High level: ≤ 8.0% | Level 1 Control: 6.4%Level 2 Control: 1.6%Sample 1: 11.8% (above criterion, but "pvalue with 5% acceptable remains acceptable")Sample 2: 6.5%Sample 3: 3.8%Sample 4: 2.8%Sample 5: 1.4% |
| Interferences (Bias) | +/- 10% of value without interfering substances | Hemoglobin: up to 500 mg/dLTriglycerides: up to 270.42 mg/dLTotal Bilirubin: up to 29.5 mg/dLDirect Bilirubin: up to 25.87 mg/dLAscorbic Acid: up to 5.98 mg/dLOthers specified in document |
| Prozone / Antigen Excess Effect | Detect and flag samples with underestimated results due to high concentration | Antigen excess observed > 5043 ng/mL; an alarm will flag and re-run these samples. |
| Method Comparison (Correlation with Predicate) | N/A (determined acceptable by high correlation) | Correlation (r²) = 0.999 (for 103 samples, 16.74 - 413.00 ng/mL range) |
| Closed Stability | N/A (defined by statement) | 18 months, stored at 2-10°C, protected from light. |
| Open Stability (On-board) | N/A (defined by statement) | 2 months |
| Reference Range Verification | Support establishing ranges vs. literature | Women: 10 - 120 ng/ml (µg/l)Men: 20 - 250 ng/ml (µg/l) |
2. Sample Size and Data Provenance (for test set)
- Measuring Range, Precision, Interferences, Prozone/Antigen Excess: Not explicitly stated as "test set" in the context of supervised learning, but these are analytical performance studies. Samples used for precision studies include 240 replicates for analyzer variability and 90 replicates for lot-to-lot variability (for each sample/control). The samples are clinical samples or controls, but the origin (country, retrospective/prospective) is not specified beyond "human serum specimens".
- Method Comparison: 103 native sera samples. Origin: "Anonymous remnants of human serum specimens collected from blood bank." Retrospective.
- Reference Range: Women: 50 "normal samples". Men: 95 "normal samples". Origin: "blood bank." Retrospective.
3. Number of experts and qualifications (for ground truth)
- Not applicable as this is an in vitro diagnostic device for quantitative measurement, not an AI evaluation requiring expert adjudication. Ground truth is instrument-derived or defined by reference methods/literature.
4. Adjudication method (for test set)
- Not applicable.
5. Multi Reader Multi Case (MRMC) comparative effectiveness study
- No, not applicable for this type of IVD device.
6. Standalone performance (algorithm only)
- Yes, the performance data presented is for the device operating in standalone mode (algorithm only) as a quantitative measurement system.
7. Type of ground truth used
- Analytical Performance (LOD, LOQ, Linearity, Precision, Interferences, Prozone/Antigen Excess): The "ground truth" is established through well-defined laboratory analytical methods and standards (CLSI guidelines EP17-A2, EP06-A, EP05-A3, EP07-A2). It relies on the accuracy of the reference materials and methods used in these studies.
- Method Comparison: Comparison against a legally marketed predicate device (Beckman Coulter Ferritin (OSR61203) [K092505]) is used as the reference, with correlation analysis.
- Reference Range: Verification against established literature references (e.g., TIETZ Textbook of Clinical Chemistry and Molecular Diagnostics).
8. Sample size for the training set
- Not applicable. This is not a machine learning model that requires a "training set" in that sense. The device's calibration curve establishment and internal parameters would be set by the manufacturer using validated reference materials and methodologies.
9. How the ground truth for the training set was established
- Not applicable.
Yumizen C1200 Transferrin
1. Table of Acceptance Criteria and Reported Device Performance
| Performance Characteristic | Acceptance Criteria | Reported Device Performance |
|---|---|---|
| Measuring Range (Serum/Plasma) | N/A (claimed measuring range is appropriate based on LOD, LOQ, and linearity studies) | 0.10 to 5.20 g/L |
| Limit of Detection (Serum/Plasma) | N/A (determined according to CLSI EP17-A2) | 0.002 g/L |
| Limit of Quantitation (Serum/Plasma) | N/A (determined according to CLSI EP17-A2) | 0.07 g/L |
| Linearity (Serum/Plasma) | N/A (determined according to CLSI EP06-A) | Evaluated from 0.15 to 4.61 g/L (appropriate) |
| Total Precision (Analyzer Variability) - Within Run CV | Low level: ≤ 6.0%Middle level: ≤ 4.5%High level: ≤ 3.8% | Level 1 Control (1.24 g/L): 1.2%Level 2 Control (3.35 g/L): 1.5%Sample 1 (0.78 g/L): 1.0%Sample 2 (1.02 g/L): 1.2%Sample 3 (1.83 g/L): 1.3%Sample 4 (3.78 g/L): 1.5% |
| Total Precision (Analyzer Variability) - Total CV | Low level: ≤ 8.0%Middle & High level: ≤ 6.0% | Level 1 Control: 3.6%Level 2 Control: 3.6%Sample 1: 4.5%Sample 2: 3.2%Sample 3: 2.3%Sample 4: 2.7% |
| Total Precision (Lot to Lot Variability) - Within Run CV | Low level: ≤ 6.0%Middle level: ≤ 4.5%High level: ≤ 3.8% | Level 1 Control (1.29 g/L): 3.5%Level 2 Control (3.41 g/L): 1.7%Sample 1 (0.77 g/L): 4.2%Sample 2 (1.08 g/L): 1.5%Sample 3 (1.96 g/L): 1.3%Sample 4 (3.54 g/L): 2.6% |
| Total Precision (Lot to Lot Variability) - Total CV | Low level: ≤ 8.0%Middle & High level: ≤ 6.0% | Level 1 Control: 6.6%Level 2 Control: 3.0%Sample 1: 5.4%Sample 2: 3.3%Sample 3: 3.4%Sample 4: 3.4% |
| Interferences (Bias) | +/- 10% of value without interfering substances | Hemoglobin: up to 500 mg/dLTriglycerides: up to 353.28 mg/dLTotal Bilirubin: up to 43.84 mg/dLDirect Bilirubin: up to 23.86 mg/dLAscorbic Acid: up to 5.98 mg/dLOthers specified in document |
| Anticoagulant Study (Serum vs. Heparin Plasma) | No significant difference between serum and plasma | Correlation (r) = 0.995, Intercept = 0.04833, Slope = 0.9691 (59 paired samples) |
| Prozone / Antigen Excess Effect | No antigen excess detected within claimed range | No antigen excess detected up to 40 g/L. |
| Method Comparison (Correlation with Predicate) | N/A (determined acceptable by high correlation) | Correlation (r²) = 0.993 (for 115 samples, 0.37 - 4.81 g/L range) |
| Closed Stability | N/A (defined by statement) | 24 months, stored at 2-8°C, protected from light. |
| Open Stability (On-board) | N/A (defined by statement) | 6 weeks |
| Reference Range Verification | Support establishing ranges vs. literature | Normal range Transferrin - Serum: 2 - 3.6 g/l (200 - 360 mg/dl) |
2. Sample Size and Data Provenance (for test set)
- Measuring Range, Precision, Interferences, Prozone/Antigen Excess: Not explicitly stated as "test set" in the context of supervised learning, but these are analytical performance studies. Samples used for precision studies include 240 replicates for analyzer variability and 90 replicates for lot-to-lot variability (for each sample/control). The samples are clinical samples or controls, but the origin (country, retrospective/prospective) is not specified beyond "human serum/plasma".
- Method Comparison: 115 native samples. Origin: "Anonymous remnants of human serum specimens collected from CHU Nîmes (University Hospital Center)." Retrospective.
- Anticoagulant Study: 59 paired serum/plasma samples. Origin: "single donors." Not specified if retrospective or prospective.
- Reference Range: 85 "normal samples" (28 women + 57 men). Origin: "blood bank." Retrospective.
3. Number of experts and qualifications (for ground truth)
- Not applicable.
4. Adjudication method (for test set)
- Not applicable.
5. Multi Reader Multi Case (MRMC) comparative effectiveness study
- No, not applicable.
6. Standalone performance (algorithm only)
- Yes, the performance data presented is for the device operating in standalone mode (algorithm only).
7. Type of ground truth used
- Analytical Performance: Established through CLSI guidelines (EP17-A2, EP06-A, EP05-A3, EP07-A2).
- Method Comparison: Comparison against a legally marketed predicate device (Roche Diagnostics Transferrin Model :TRSF2 [K012393]).
- Reference Range: Verification against established literature references (e.g., Dati et al., Eur. J Clin Chem. Cli Biochem. 1996).
8. Sample size for the training set
- Not applicable.
9. How the ground truth for the training set was established
- Not applicable.
Yumizen C1200 Rheumatoid Factor
1. Table of Acceptance Criteria and Reported Device Performance
| Performance Characteristic | Acceptance Criteria | Reported Device Performance |
|---|---|---|
| Measuring Range (Serum) | N/A (claimed measuring range is appropriate based on LOD, LOQ, and linearity studies) | 10 to 120 IU/mL |
| Limit of Detection (Serum) | N/A (determined according to CLSI EP17-A2) | 4.07 IU/mL |
| Limit of Quantitation (Serum) | N/A (determined according to CLSI EP17-A2) | 7.41 IU/mL |
| Linearity (Serum) | N/A (determined according to CLSI EP06-A) | Evaluated from 13.2 to 118.8 IU/mL (appropriate) |
| Total Precision (Analyzer Variability) - Within Run CV | Low level: ≤ 6.0%Middle level: ≤ 4.5%High level: ≤ 3.8% | Level 1 Control (40.99 IU/mL): 0.5%Level 2 Control (63.93 IU/mL): 0.4%Sample 1 (22.24 IU/mL): 1.2%Sample 2 (34.28 IU/mL): 0.8%Sample 3 (49.41 IU/mL): 0.5%Sample 4 (70.16 IU/mL): 0.5%Sample 5 (103.42 IU/mL): 0.8% |
| Total Precision (Analyzer Variability) - Total CV | Low level: ≤ 8.0%Middle & High level: ≤ 6.0% | Level 1 Control: 2.2%Level 2 Control: 2.5%Sample 1: 2.0%Sample 2: 2.2%Sample 3: 1.8%Sample 4: 1.6%Sample 5: 1.4% |
| Total Precision (Lot to Lot Variability) - Within Run CV | Low level: ≤ 6.0%Middle level: ≤ 4.5%High level: ≤ 3.8% | Level 1 Control (41.70 IU/mL): 1.8%Level 2 Control (67.05 IU/mL): 1.4%Sample 1 (17.30 IU/mL): 2.9%Sample 2 (30.88 IU/mL): 1.4%Sample 3 (53.08 IU/mL): 1.4%Sample 4 (70.24 IU/mL): 1.1%Sample 5 (102.14 IU/mL): 1.0% |
| Total Precision (Lot to Lot Variability) - Total CV | Low level: ≤ 8.0%Middle & High level: ≤ 6.0% | Level 1 Control: 1.9%Level 2 Control: 2.2%Sample 1: 3.1%Sample 2: 1.8%Sample 3: 3.2%Sample 4: 1.3%Sample 5: 1.8% |
| Interferences (Bias) | +/- 10% of value without interfering substances | Hemoglobin: up to 500 mg/dLTriglycerides: up to 526.75 mg/dLTotal Bilirubin: up to 31.32 mg/dLDirect Bilirubin: up to 25.34 mg/dLAscorbic Acid: up to 5.98 mg/dLOthers specified in document |
| Prozone / Antigen Excess Effect | Detect and flag samples with underestimated results due to high concentration | Antigen excess observed > 229 IU/mL; an alarm will flag and re-run these samples. |
| Method Comparison (Correlation with Predicate) | N/A (determined acceptable by high correlation) | Correlation (r²) = 0.992 (for 113 samples, 16.79 - 118.81 IU/mL range) |
| Closed Stability | N/A (defined by statement) | 18 months, stored at 2-10°C. |
| Open Stability (On-board) | N/A (defined by statement) | 1 month |
| Reference Range Verification | Support establishing ranges vs. literature | Normal range Rheumatoid Factor: Adult < 14 IU/ml |
2. Sample Size and Data Provenance (for test set)
- Measuring Range, Precision, Interferences, Prozone/Antigen Excess: Not explicitly stated as "test set" in the context of supervised learning, but these are analytical performance studies. Samples used for precision studies include 240 replicates for analyzer variability and 90 replicates for lot-to-lot variability (for each sample/control). The samples are clinical samples or controls, but the origin (country, retrospective/prospective) is not specified beyond "human serum".
- Method Comparison: 113 native samples. Origin: "Anonymous remnants of human serum specimens collected from blood bank." Retrospective.
- Reference Range: 60 "normal samples". Origin: "blood bank." Retrospective.
3. Number of experts and qualifications (for ground truth)
- Not applicable.
4. Adjudication method (for test set)
- Not applicable.
5. Multi Reader Multi Case (MRMC) comparative effectiveness study
- No, not applicable.
6. Standalone performance (algorithm only)
- Yes, the performance data presented is for the device operating in standalone mode (algorithm only).
7. Type of ground truth used
- Analytical Performance: Established through CLSI guidelines (EP17-A2, EP06-A, EP05-A3, EP07-A2).
- Method Comparison: Comparison against a legally marketed predicate device (Olympus RF Latex reagent (OSR61105) [K060201]).
- Reference Range: Verification against established literature references (e.g., Tietz NW, Clinical Guide to Laboratory Tests, 3rd ed. 1990).
8. Sample size for the training set
- Not applicable.
9. How the ground truth for the training set was established
- Not applicable.
General Notes on the Studies:
- The studies were designed to evaluate the analytical performance of the IVD reagents and the Yumizen C1200 analyzer according to recognized CLSI (Clinical and Laboratory Standards Institute) guidelines, which are standard for in vitro diagnostic device validation.
- The term "acceptance criteria" is often implied by the CLSI guidelines themselves and the high correlation/precision values that are typical for these types of assays to demonstrate substantial equivalence to predicate devices. Explicit numerical acceptance criteria are sometimes only stated for the precision studies (e.g., %CV limits).
- For the method comparison studies, the high correlation coefficients (r²) indicate good agreement with the predicate devices, supporting the claim of substantial equivalence.
- The "test set" concept from AI/ML is not directly applicable here. The samples used for performance evaluation are analytical samples, control materials, and patient samples that cover the measuring range and demonstrate various performance characteristics of the assay.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: a symbol on the left and the FDA name and title on the right. The symbol on the left is a stylized representation of a human figure, while the text on the right reads "FDA U.S. FOOD & DRUG ADMINISTRATION" in blue letters.
March 6, 2020
Horiba ABX SAS Caroline Ferrer Regulatory Affairs Manager Parc Euromedecine, Rue Du Caducee BP 7290 Montpellier Cedex 4, 341184 France
Re: K191562
Trade/Device Name: Yumizen C1200 Ferritin, Yumizen C1200 Transferrin, Yumizen C1200 Rheumatoid Factor Regulation Number: 21 CFR 866.5340 Regulation Name: Ferritin immunological test system Regulatory Class: Class II Product Code: DBF, DDG, DHR Dated: June 11, 2019 Received: June 13, 2019
Dear Caroline Ferrer:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal
{1}------------------------------------------------
statutes and regulations administered by other Federal agencies. You must comply with all the Act's
requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801 and Part 809); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Carolina Kagan, M.Sc. Acting Chief Division of Immunology and Hematology Devices OHT7: Office of In Vitro Diagnostics and Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
Indications for Use
510(k) Number (if known) K191562
Device Name Yumizen C1200 Ferritin Yumizen C1200 Transferrin Yumizen C1200 Rheumatoid Factor
Indications for Use (Describe)
Yumizen C1200 Ferritin reagent is intended for the quantitative in vitro diagnostic determination of Ferritin in human serum by latex-enhanced immunoturbidimetric assay. Measurements of ferritin aid in the diagnosis of diseases affecting iron metabolism, hemochromatosis (iron overload) and iron deficiency anemia.
Yumizen C1200 Transferrin reagent is intended for the quantitative in vitro diagnostic determination of transferrin in human serum and lithium heparin plasma by turbidimetry.
Measurement of transferrin levels ads in the diagnosis of malnutrition, acute inflammation, infection, and iron deficiency anemia.
Yumizen C1200 Rheumatoid Factor reagent is intended for the quantitative in vitro diagnostic determination of rheumatoid factor in human serum by latex-enhanced immunoturbidimetric assay. Measurement of rheumatoid factor may aid in the diagnosis of rheumatoid arthritis.
Type of Use (Select one or both, as applicable)
X Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the HORIBA Medical logo in blue, with the word "Medical" in a smaller font size below "HORIBA". Below the logo, the text "Section 007. 510(k) summary" is displayed in a gray font. The text appears to be part of a document or report.
SECTION 007 : 510(k) Summary
This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR 807.92.
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the HORIBA Medical logo in blue. Below the logo is the text "Section 007. 510(k) summary" in a gray font. The text appears to be a section heading or title, possibly from a document or report.
- 1- Date of Summary Date submitted : 5th March, 2020
2- Company
HORIBA ABX SAS HORIBA MEDICAL Parc Euromédecine Rue du Caducée – BP 7290 34184 Montpellier cedex 4 France
3- Contact person
Contact Person: Caroline Ferrer (caroline.ferrer@horiba.com) Telephone: + (33) 4 67 14 1843 Fax: + (33) 4 67 14 1517
4- Product Name
Yumizen C1200 Ferritin (1300023880) Yumizen C1200 Transferrin (1300023889) Yumizen C1200 Rheumatoid Factor (1300023888)
5- Device Name and Classification
. Intended use
The devices involved by the 510(k) submission file are the following:
● Classification and Description
| Device's names | Intended Use |
|---|---|
| Yumizen C1200Ferritin | Yumizen C1200 Ferritin reagent is intended for the quantitative in vitrodiagnostic determination of Ferritin in human serum by latex-enhancedimmunoturbidimetric assay. Measurements of ferritin aid in the diagnosisof diseases affecting iron metabolism, hemochromatosis (iron overload)and iron deficiency anemia. |
| Yumizen C1200Transferrin | Yumizen C1200 Transferrin reagent is intended for the quantitative in vitrodiagnostic determination of Transferrin in human serum and lithiumheparin plasma by turbidimetry.Measurement of transferrin levels aids in the diagnosis of malnutrition,acute inflammation, infection, and iron deficiency anemia. |
| Yumizen C1200Rheumatoid Factor | Yumizen C1200 Rheumatoid Factor reagent is intended for the quantitativein vitro diagnostic determination of rheumatoid factor in human serum bylatex-enhanced immunoturbidimetric assay.Measurement of rheumatoid factor may aid in the diagnosis of rheumatoidarthritis. |
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image shows the HORIBA Medical logo in blue, with the word "Medical" in a smaller font size below "HORIBA". Below the logo, the text "Section 007. 510(k) summary" is displayed in a gray font. The text appears to be part of a document or report.
| Trade/Proprietary Name: | Yumizen C1200 Ferritin |
|---|---|
| Device Class: | Class II / 510(k) required |
| Classification Name: | §866.5340: Ferritin immunological test system |
| Product Code: | DBF |
| Panel: | Immunology (82) |
| Trade/Proprietary Name: | Yumizen C1200 Transferrin |
| Device Class: | Class II / 510(k) required |
| Classification Name: | §866.5880: Transferrin immunological test system |
| Product Code: | DDG |
| Panel: | Immunology (82) |
| Trade/Proprietary Name: | Yumizen C1200 Rheumatoid Factor |
| Device Class: | Class II / 510(k) required |
| Classification Name: | §866.5775: Rheumatoid factor immunological test system |
| Product Code: | DHR |
| Panel: | Immunology (82) |
- . This submission allows to evaluate the functionality of the Yumizen C1200 analyzer for immunology analytes (ie.immunoturbidimetry).
Substantial Equivalence Information 6-
The following tables show the similarities and differences and demonstrates substantial equivalence between the candidate device and its predicate device identified below.
a. Predicate Device Name and 510(k) number
| Candidate device | Predicate device | Predicate Manufacturer | Predicate510(k)number |
|---|---|---|---|
| Yumizen C1200Ferritin | Ferritin ( OSR61203) | BECKMAN COULTER | K092505 |
| Yumizen C1200Transferrin | Transferrin Model :TRSF2 | Roche Diagnostics | K012393 |
| Yumizen C1200Rheumatoid Factor | Olympus RF Latex reagent(OSR61105) | Olympus America, Inc. | K060201 |
{6}------------------------------------------------
Image /page/6/Picture/0 description: The image shows the HORIBA Medical logo. The word "HORIBA" is in large, bold, blue letters. Below it, in smaller, lighter blue letters, is the word "Medical."
The following tables show the similarities and differences and demonstrates substantial equivalence between the candidate device and its predicate device identified below.
b. Yumizen C1200 Ferritin
i. Comparison with predicate Device : Similarities
| Item | Predicate K092505 | Candidate |
|---|---|---|
| Device Name | Ferritin (OSR61203) | Yumizen C1200 Ferritin (1300023880) |
| Intended Use | The Ferritin Reagent is for thedetermination of ferritin concentrations inhuman serum and plasma on the BeckmanCoulter AU clinical chemistry analyzers.Serum ferritin is an indicator of body ironstores: it has been shown to correlate withstainable bone marrow iron. Measurementsof ferritin aid in the diagnosis of diseasesaffecting iron metabolism, such ashemochromatosis (iron overload) and irondeficiency anemia. | Yumizen C1200 Ferritin reagent isintended for the quantitative in vitrodiagnostic determination of Ferritin inhuman serum by latex-enhancedimmunoturbidimetric assay.Measurements of ferritin aid in thediagnosis of diseases affecting ironmetabolism, hemochromatosis (ironoverload) and iron deficiency anemia. |
| Reagent format | Liquid | Same |
| Measurement | Quantitative | Same |
| Method | Latex-enhanced immuno-turbidimetricmethod | Same |
| Product code | DBF | Same |
| On board Stability | Once opened, the reagent cassette placedin the refrigerator compartment is stablefor 60 days | Once opened, the reagent cassette placedin the refrigerator compartment is stablefor 2 months (or 60 days). |
| CalibrationStability | The frequency of calibration for theFerritin procedure is every30 days | The reagent is calibrated on Day 0. Thecalibration stability is checked by testing2 control specimens.The calibration stability is1 month(or 30 days) |
{7}------------------------------------------------
Image /page/7/Picture/0 description: The image shows the HORIBA Medical logo in blue text. Below the logo is the text "Section 007. 510(k) summary" in a smaller, gray font. The text appears to be a section heading or title, possibly from a document or report.
| Item | Predicate K092505 | Candidate |
|---|---|---|
| Device Name | Ferritin (OSR61203) | Yumizen C1200 Ferritin (1300023880) |
| Instrument | Beckman Coulter AU400 ClinicalChemistry Analyzer | Yumizen C1200 Clinical chemistryanalyzer |
| Manufactured by | BECKMAN COULTER | HORIBA ABX SAS |
| Sample type | Serum,Li-heparin plasmaand EDTA plasma samples | Serum |
| Packaging | Cassette of :R1= 4 x 24 mLR2= 4 x 12 mL | Cassette of :R1: 6 x 11 mLR2: 6 x 7 mL |
| Shelf-life | Unopened, up to the stated expiry datewhen stored at 2-8°C. | Stable up to the expiration date if storedat 2-10°C. |
| Analytical Range | Measuring Range$8.0 - 450.0$ ng/mL | Measuring Range$10 - 450$ ng/mL |
| Reference range | Women: $10 - 158$ ng/mLMen: $16 - 243$ ng/mL | Adults :Women: $10 - 120$ ng/mL ( $\mu$ g/L)Men: $20 - 250$ ng/mL ( $\mu$ g/L) |
ii. Comparison with predicate Device: Differences
Discussion on the analysis differences :
-
- Instrument: Yumizen C1200 Ferritin is used on Yumizen C1200
-
- Packaging: Packaging is different; depends on cassette capacity.
-
- Reagent stability/ Shelf-life:storage temperature is different.
-
- Analytical range: the measuring range for Yumizen C1200 Ferritin is slightly higher on the lower end
-
- Reference range: Yumizen C1200 Ferritin has tighter reference range on the upper end for Women and slightly larger reference range on the upper end for Men.
- Sample type: Yumizen C1200 Ferritin uses serum. 6.
{8}------------------------------------------------
Image /page/8/Picture/0 description: The image contains the logo for HORIBA Medical. The word "HORIBA" is in large, bold, blue letters. Below it, in smaller, lighter blue letters, is the word "Medical". There is also some text that appears to be a code or part number, "007-510(1)".
c. Yumizen C1200 Transferrin (1300023889)
i. Comparison with predicate Device : Similarities
| Item | Predicate K012393 | Candidate |
|---|---|---|
| Device Name | Transferrin Model :TRSF2 | Yumizen C1200 Transferrin(1300023889) |
| Intended Use | In vitro test for the quantitativedetermination of transferrin in humanserum and plasma in Roche/Hitachi cobasc systems. | Yumizen C1200 Transferrin reagent isintended for the quantitative in vitrodiagnostic determination of Transferrinin human and lithium heparin plasma byturbidimetry.Measurement of transferrin levels aids inthe diagnosis of malnutrition, acuteinflammation, infection, and irondeficiency anemia. |
| Sample type | Serumplasma | Serumplasma |
| Reagent format | Liquid | Same |
| Measurement | Quantitative | Same |
| Method | Turbidimetry | Same |
| Product code | DDG | Same |
| Shelf-life | 2-8°C,See expiration date on label | Stable up to the expiry date on the labelif stored at 2-8°C. |
| Analytical Range | Measuring Range0.1-5.2 g/L | Measuring Range0.10-5.20 g/L |
| Reference range | 200-360 mg/dL | 200-360 mg/dL |
{9}------------------------------------------------
Image /page/9/Picture/0 description: The image shows the logo for HORIBA Medical. The word "HORIBA" is in large, bold, blue letters. Below it, in smaller blue letters, is the word "Medical."
K191562 Yumizen C1200 Reagents
ii. Comparison with predicate Device: Differences
| Item | Predicate K012393 | Candidate |
|---|---|---|
| Device Name | Transferrin Model : TRSF2 | Yumizen C1200 Transferrin(1300023889) |
| Instrument | Roche analyzer (Cobas C701) | Yumizen C1200 Clinical chemistryanalyzer |
| Manufactured by | ROCHE | HORIBA ABX SAS |
| On board Stability | 4 weeks | Once opened, the reagent cassette placedin the refrigerated compartment is stablefor 6 weeks. |
Discussion on the analysis differences :
- Instrument & Manufacturer: Yumizen C1200 Transferrin is used on Yumizen C1200.
Yumizen C1200 Transferrin is manufactured by HORIBA ABX SAS.
- Reagent stability: the on board stability of Yumizen C1200 Transferrin is longer.
Stability depends on the reagent composition and cassette capacity.
{10}------------------------------------------------
Image /page/10/Picture/0 description: The image contains the logo for HORIBA Medical in blue font. Below the logo is the text "Section 007. 510(k) summary" in a smaller, gray font. The text appears to be a section heading or title within a larger document.
d. Yumizen C1200 Rheumatoid Factor
| Item | Predicate K060201 | Candidate |
|---|---|---|
| Device Name | Rheumatoid Factor reagent model :OSR61105 | Yumizen C1200 Rheumatoid Factor(1300023888) |
| Intended Use | Olympus RF Latex System reagent for thequantitative determination of RheumatoidFactor (RF) in human serum and plasmaon Olympus AU analyzers.Measurement of rheumatoid factor mayaid in the diagnosis of rheumatoid arthritis | Yumizen C1200 Rheumatoid Factorreagent is intended for the quantitative invitro diagnostic determination ofrheumatoid factor in human serum bylatex-enhanced immunoturbidimetricassay.Measurement of rheumatoid factor mayaid in the diagnosis of rheumatoidarthritis. |
| Manufactured by | OLYMPUS AMERICA, INC | HORIBA ABX SAS |
| Measurement | Quantitative | Quantitative |
| Method | Latex-enhanced Immunoturbidimetry | Same |
| Product code | DHR | Same |
| CalibrationStability | The frequency of calibration for the RFLatex procedure is every 30 days. | The calibration stability is 1 month(or 30 days) |
| Reference range | Adult ≤ 14 IU/mL | Same |
i. Comparison with predicate Device : Similarities
{11}------------------------------------------------
Image /page/11/Picture/0 description: The image shows the HORIBA Medical logo in blue, with the word "Medical" in a smaller font size below "HORIBA". Below the logo, the text "Section 007. 510(k) summary" is displayed in a smaller, gray font. The text appears to be a section heading or title, possibly from a document or report.
| Item | Predicate K060201 | Candidate |
|---|---|---|
| Device Name | Rheumatoid Factor reagent model :OSR61105 | Yumizen C1200 Rheumatoid Factor(1300023888) |
| Instrument | Olympus AU400 Clinical ChemistryAnalyzer | Yumizen C1200 Clinical chemistryanalyzer |
| Sample type | Serum, plasma | Serum |
| Reagent format | Liquid | Same |
| Packaging | Cassette of :R1= 4 x 24 mLR2= 4 x 8mL | Cassette of :R1: 6 x 13 mLR2: 6 x 6 mL |
| Analytical range | Measuring range :5-120 IU/mL | Measuring range :10 - 120 IU/mL |
| Shelf-life | Until expiration date on label(+2-8 °C) | Stable up to expiration dateif stored at 2-10°C |
| On board Stability | Once opened, the reagent is stablefor 60 days in refrigerated compartment ofanalyzer. | Once opened, the reagent cassette placedin the refrigerator compartment is stablefor 1 month (or 30 days) |
| Discussion on the analysis differences : |
ii. Comparison with predicate Device: Differences
-
Instrument: Yumizen C1200 Rheumatoid Factor is used on Yumizen C1200
-
Packaging: Packaging is different; depends on cassette capacity.
-
Reagent stability: the on board stability Yumizen C1200 Rheumatoid Factor is shorter. The Shelf life is different. Stability depend on reagent composition and cassette capacity.
5.Analytical range: the measuring range for serum for Yumizen C1200 Rheumatoid factor is slightly tighter on the lower end
- Sample type: Yumizen C1200 RF uses serum only.
{12}------------------------------------------------
Image /page/12/Picture/0 description: The image contains the logo for HORIBA Medical. The word "HORIBA" is in large, bold, blue letters. Below the word "HORIBA" is the word "Medical" in a smaller, lighter blue font. The logo is simple and clean.
K191562 Yumizen C1200 Reagents
7 - Special Control/Guidance Document Referenced
a. Standards Followed
The following standards & FDA guidance documents have been used to support this submission:
CLSI Guidelines:
- CLSI EP05-A3:Evaluation of Precision of Quantitative Measurement Procedures- Third ● Edition - October 2014
- . CLSI EP17-A2: Evaluation of Detection Capability for Clinical Laboratory Measurement Procedures - Second Edition - June 2012
- CLSI EP06-A: Evaluation of the Linearity of Quantitative measurement Procedures A Statistical Approach - First Edition - April 2003
- CLSI C28-A3: Defining, Establishing, and Verifying Reference Intervals in the Clinical laboratory- Third Edition - November 2008
- CLSI EP25-A : Evaluation of Stability of In Vitro Diagnostic reagents- First Edition-September 2009
b. FDA Guidances Followed
- Guidance for Industry and FDA Staff : Format for Traditional and Abbreviated 510(k)s 2005 ●
c. Other reference cited
- Valtec guideline (Vassault et al., Ann. Biol. Clin., 1986, (44), 686-745)
- CLSI EP09-A3 : Measurement Procedure Comparison and Bias Estimation Using Patient ● Samples - Third Edition - August 2013
{13}------------------------------------------------
Image /page/13/Picture/0 description: The image shows the HORIBA Medical logo in blue. Below the logo is the text "Section 007. 510(k) summary" in a light gray font. The text appears to be part of a document or report.
8- Analytical Performance Characteristics
8.1 Measuring Range
Yumizen C1200 Ferritin ●
The limit of detection and quantitation was determined according to the CLSI guideline EP17-A2. The reagent linearity was determined according to CLSI guideline EP06-A. The limit of quantitation and the linearity studies showed that claimed measuring range is appropriate.
> Results :
| Limit ofdetection | Limit ofquantitation | Linearity evaluated | Measuring range | |
|---|---|---|---|---|
| Serum | 6.30 ng/mL | 9.39 ng/mL | 13.3 - 426.6 ng/mL | 10 to 450 ng/mL |
| SerumPost-dilution | NA | NA | NA | until 2250 ng/mL |
● Yumizen C1200 Transferrin
The limit of detection and quantitation was determined according to the CLSI guideline EP17-A2. The reagent linearity was determined according to CLSI guideline EP06-A. The limit of quantitation and the linearity studies showed measuring range is appropriate.
> Results :
| Limit ofdetection | Limit ofquantitation | Linearity evaluated | Measuring range | |
|---|---|---|---|---|
| Serum/Plasma | 0.002 g/L | 0.07 g/L | 0.15 - 4.61 g/L | 0.10 to 5.20 g/L |
| Serum/PlasmaPost-dilution | NA | NA | NA | 15.60 until g/L |
{14}------------------------------------------------
Image /page/14/Picture/0 description: The image shows the HORIBA Medical logo in blue, with the word "Medical" in a smaller font size below it. Below the logo, the text "Section 007. 510(k) summary" is displayed in a smaller, gray font. The text appears to be a section heading or title, possibly from a document or report.
Yumizen C1200 Rheumatoid Factor ●
The limit of detection and quantitation was determined according to the CLSI guideline EP17-A2. The reagent linearity was determined according to CLSI guideline EP06-A. The limit of quantitation and the linearity studies showed that claimed measuring range is appropriate.
》 Results :
| Limit of detection | Limit of quantitation | Linearity evaluated | Measuring range | |
|---|---|---|---|---|
| Serum | 4.07 IU/mL | 7.41 IU/mL | 13.2 - 118.8 IU/mL | 10 to 120 IU/mL |
| SerumPost-dilution | NA | NA | NA | until 1200 IU/mL |
{15}------------------------------------------------
Image /page/15/Picture/0 description: The image shows the logo for HORIBA Medical, followed by the text "Section 007. 510(k) summary". The HORIBA logo is in blue, with the word "HORIBA" in a larger, bolder font than the word "Medical" below it. The text "Section 007. 510(k) summary" is in a smaller, gray font.
8.2 Accuracy and Precision
Repeatability (within-run precision) and Reproducibility (total precision)
Yumizen C1200 Ferritin ●
A Total Precision : analyzer variability - 20x2x2 study
Within Run : CV limits, for the low, middle and high level are respectively 8.0 %, 6.0% and 6.0% for serum Total Precision : CV limits, for the low, middle and high level are respectively 10.0 %, 8.0 % for serum.
| Sample | N | Mean(ng/mL) | Within-Run(%CV) | Between-Run(%CV) | Between-Day(%CV) | Between-Instrument(%CV) | Total(%CV) |
|---|---|---|---|---|---|---|---|
| Yumizen C1200Level 1 ProteinControl | 240 | 47.58 | 3.6 | 0.0 | 1.5 | 3.1 | 4.9 |
| Yumizen C1200Level 2 ProteinControl | 240 | 279.31 | 1.1 | 0.0 | 1.6 | 0.8 | 2.1 |
| Sample 1 | 240 | 29.56 | 5.5 | 2.2 | 2.2 | 4.7 | 7.9 |
| Sample 2 | 240 | 50.87 | 4.1 | 1.9 | 0.0 | 2.4 | 5.1 |
| Sample 3 | 240 | 172.63 | 1.4 | 0.3 | 0.6 | 1.1 | 1.9 |
| Sample 4 | 240 | 328.60 | 1.3 | 3.6 | 2.0 | 0.0 | 4.3 |
| Sample 5 | 240 | 403.21 | 1.0 | 0.6 | 0.6 | 0.4 | 1.4 |
{16}------------------------------------------------
Image /page/16/Picture/0 description: The image shows the logo for HORIBA Medical. The word "HORIBA" is in large, bold, blue letters. Below it, in smaller, light blue letters, is the word "Medical."
-
Lot to Lot variability : 3x5x2x3
Within Run : CV limits, for the low, middle and high level are respectively 8.0 %, 6.0 % and 6.0 % for serum
Total Precision : CV limits, for the low, middle and high level are respectively 10.0 %, 8.0 % and 8.0 % for serum.
| Sample | N | Mean(ng/mL) | Within-Run(%CV) | Between-Day (%CV) | Within-Batch(%CV) | Between-Batch(%CV) | Total(%CV) |
|---|---|---|---|---|---|---|---|
| Yumizen C1200Level 1 ProteinControl | 90 | 52.84 | 4.6 | 4.5 | 6.4 | 0.0 | 6.4 |
| Yumizen C1200Level 2 ProteinControl | 90 | 281.87 | 0.9 | 0.7 | 1.1 | 1.2 | 1.6 |
| Sample 1 | 90 | 19.09 | 8.8 | 4.8 | 10.0 | 6.3 | 11.8 |
| Sample 2 | 90 | 34.05 | 6.5 | 0.0 | 6.5 | 0.5 | 6.5 |
| Sample 3 | 90 | 51.53 | 3.6 | 0.0 | 3.6 | 1.1 | 3.8 |
| Sample 4 | 90 | 192.31 | 1.4 | 0.0 | 1.4 | 2.5 | 2.8 |
| Sample 5 | 90 | 407.38 | 0.9 | 0.0 | 0.9 | 1.0 | 1.4 |
Although the %CV of Within Day and Total Precision are superior to the Acceptance criteria, the pvalue with 5% acceptable remains acceptable for sample 1.
{17}------------------------------------------------
Image /page/17/Picture/0 description: The image shows the HORIBA Medical logo in blue, with the word "Medical" in a smaller font size below the main logo. Below the logo is the text "Section 007. 510(k) summary" in a gray font. The text appears to be a section heading or title within a document.
Yumizen C1200 Transferrin ●
> Total Precision : analyzer variability - 20x2x2 study
Within Run : CV limits, for the low, middle and high level are respectively 6.0 %, 4.5 % and 3.8 % for serum and plasma
Total Precision : CV limits, for the low, middle and high level are respectively 8.0 %, 6.0 % for serum and plasma.
| Sample | N | Mean(g/L) | Within-Run(%CV) | Between-Run (%CV) | Between-Day (%CV) | Between-Instrument(%CV) | Total(%CV) |
|---|---|---|---|---|---|---|---|
| Yumizen C1200Level 1 ProteinControl | 240 | 1.24 | 1.2 | 0.7 | 2.1 | 2.5 | 3.6 |
| Yumizen C1200Level 2 ProteinControl | 240 | 3.35 | 1.5 | 0.6 | 1.4 | 2.8 | 3.6 |
| Sample 1 | 240 | 0.78 | 1.0 | 0.5 | 2.3 | 3.7 | 4.5 |
| Sample 2 | 240 | 1.02 | 1.2 | 0.3 | 1.7 | 2.5 | 3.2 |
| Sample 3 | 240 | 1.83 | 1.3 | 0.5 | 1.3 | 1.3 | 2.3 |
| Sample 4 | 240 | 3.78 | 1.5 | 0.6 | 1.1 | 1.9 | 2.7 |
{18}------------------------------------------------
Image /page/18/Picture/0 description: The image contains the logo for HORIBA Medical. The word "HORIBA" is in large, bold, blue letters. Below it, in smaller, lighter blue letters, is the word "Medical."
► Lot to Lot variability : 3x5x2x3
Within Run : CV limits, for the low, middle and high level are respectively 6.0 %, 4.5 % and 3.8 % for serum and plasma
Total Precision : CV limits, for the low, middle and high level are respectively 8.0 %, 6.0 % for serum and plasma.
| Sample | N | Mean(g/L) | Within-Run(%CV) | Between-Day (%CV) | Within-Batch(%CV) | Between-Batch(%CV) | Total(%CV) |
|---|---|---|---|---|---|---|---|
| Yumizen C1200Level 1 ProteinControl | 90 | 1.29 | 3.5 | 5.3 | 6.3 | 1.7 | 6.6 |
| Yumizen C1200Level 2 ProteinControl | 90 | 3.41 | 1.7 | 1.2 | 2.1 | 2.2 | 3.0 |
| Sample 1 | 90 | 0.77 | 4.2 | 2.7 | 5.0 | 2.0 | 5.4 |
| Sample 2 | 90 | 1.08 | 1.5 | 0.9 | 1.7 | 2.8 | 3.3 |
| Sample 3 | 90 | 1.96 | 1.3 | 0.2 | 1.3 | 3.2 | 3.4 |
| Sample 4 | 90 | 3.54 | 2.6 | 0.7 | 2.7 | 2.1 | 3.4 |
{19}------------------------------------------------
Image /page/19/Picture/0 description: The image shows the HORIBA Medical logo at the top. Below the logo, the text "Section 007. 510(k) summary" is displayed. The text is in a smaller font size compared to the logo.
Yumizen C1200 Rheumatoid Factor .
> Total Precision : analyzer variability - 20x2x2 study
Within Run : CV limits, for the low, middle and high level are respectively 6.0 %, 4.5 % and 3.8 % for serum
Total Precision : CV limits, for the low, middle and high level are respectively 8.0 %, 6.0 % for serum.
| Sample | N | Mean(IU/mL) | Within-Run(%CV) | Between-Run (%CV) | Between-Day (%CV) | Between-Instrumentt (%CV) | Total(%CV) |
|---|---|---|---|---|---|---|---|
| Yumizen C1200Level 1 ProteinControl | 240 | 40.99 | 0.5 | 0.4 | 1.2 | 1.7 | 2.2 |
| Yumizen C1200Level 2 ProteinControl | 240 | 63.93 | 0.4 | 0.6 | 1.6 | 1.8 | 2.5 |
| Sample 1 | 240 | 22.24 | 1.2 | 0.9 | 0.9 | 1.0 | 2.0 |
| Sample 2 | 240 | 34.28 | 0.8 | 1.3 | 1.3 | 0.9 | 2.2 |
| Sample 3 | 240 | 49.41 | 0.5 | 1.0 | 0.5 | 1.3 | 1.8 |
| Sample 4 | 240 | 70.16 | 0.5 | 0.4 | 0.8 | 1.3 | 1.6 |
| Sample 5 | 240 | 103.42 | 0.8 | 0.6 | 0.7 | 0.5 | 1.4 |
{20}------------------------------------------------
Image /page/20/Picture/0 description: The image contains the logo for HORIBA Medical. The word "HORIBA" is in large, bold, blue letters. Below it, in a smaller font and lighter blue color, is the word "Medical."
► Lot to Lot variability : 3x5x2x3
Within Run : CV limits, for the low, middle and high level are respectively 6.0 %, 4.5 % and 3.8 % for serum
Total Precision : CV limits, for the low, middle and high level are respectively 8.0 %, 6.0 % for serum.
| Sample | N | Mean(IU/mL) | Within-Run(%CV) | Between-Day (%CV) | Within-Batch(%CV) | Between-Batch(%CV) | Total(%CV) |
|---|---|---|---|---|---|---|---|
| Yumizen C1200Level 1 ProteinControl | 90 | 41.70 | 1.8 | 0.3 | 1.8 | 0.6 | 1.9 |
| Yumizen C1200Level 2 ProteinControl | 90 | 67.05 | 1.4 | 0.5 | 1.5 | 1.6 | 2.2 |
| Sample 1 | 90 | 17.30 | 2.9 | 0.7 | 3.0 | 0.8 | 3.1 |
| Sample 2 | 90 | 30.88 | 1.4 | 0.6 | 1.6 | 0.9 | 1.8 |
| Sample 3 | 90 | 53.08 | 1.4 | 0.9 | 1.6 | 2.7 | 3.2 |
| Sample 4 | 90 | 70.24 | 1.1 | 0.8 | 1.3 | 0.0 | 1.3 |
| Sample 5 | 90 | 102.14 | 1.0 | 0.4 | 1.1 | 1.4 | 1.8 |
{21}------------------------------------------------
Image /page/21/Picture/0 description: The image shows the HORIBA Medical logo at the top. Below the logo, the text "Section 007. 510(k) summary" is displayed. The text is in a smaller font size compared to the logo.
8.3 Interferences
The Interferences were determined according to the CLSI guideline EP07-A2. The acceptable bias is defined at +/-10% of the value without interfering substances. These data in the following table represent the highest values for which no interferences higher than 10% have been observed.
Yumizen C1200 Ferritin ●
| Serum | ||
|---|---|---|
| Hemoglobin | 290 µmol/L | 500 mg/dL |
| Triglycerides | 3.09 mmol/L | 270.42 mg/dL |
| Total Bilirubin | 504 µmol/l | 29.5 mg/dL |
| Direct Bilirubin | 442 µmol/l | 25.87 mg/dL |
| Ascorbic Acid | 340 µmol/L | 5.98 mg/dL |
| AcetylsalicylicAcid | 3.62 mmol/L | 65.16 mg/dL |
| Ibuprofen | 2.43 mmol/L | 50.10 mg/dL |
| Acetaminophen | 1324 µmol/L | 20 mg/dL |
| RheumatoidFactor | 500 IU/mL | |
| Deferoxamine | 400 µmol/L | 22.43 mg/dL |
| Prednisone | 500 µmol/L | 17.92 mg/dL |
| Methotrexate | 3000 µmol/L | 136 mg/dL |
| Ferrous Sulfate | 1 mmol/L | 15.19 mg/dL |
| Ampicillin | 250 µmol/L | 8.7 mg/dL |
| Azithromycin | 20 µmol/L | 1.49 mg/dL |
| Rifampicin | 100 µmol/L | 8.20 mg/dL |
| Diltiazem | 10 µmol/L | 0.41 mg/dL |
| Simvastatin | 10 µmol/L | 0.42 mg/dL |
{22}------------------------------------------------
Image /page/22/Picture/0 description: The image shows the HORIBA Medical logo at the top. Below the logo, the text "Section 007. 510(k) summary" is displayed. The text is written in a simple, sans-serif font and is left-aligned.
Yumizen C1200 Transferrin ●
| Serum/Plasma | ||
|---|---|---|
| Hemoglobin | 290 µmol/L | 500 mg/dL |
| Triglycerides | 4.04 mmol/L | 353.28 mg/dL |
| Total Bilirubin | 749 µmol/l | 43.84 mg/dL |
| Direct Bilirubin | 408 µmol/l | 23.86 mg/dL |
| Ascorbic Acid | 340 µmol/L | 5.98 mg/dL |
| AcetylsalicylicAcid | 3.62 mmol/L | 65.16 mg/dL |
| Ibuprofen | 2.43 mmol/L | 50.10 mg/dL |
| Acetaminophen | 1324 µmol/L | 20 mg/dL |
| RheumatoidFactor | 400 IU/mL |
• Yumizen C1200 Rheumatoid Factor
| Serum | ||
|---|---|---|
| Hemoglobin | 290 µmol/l | 500 mg/dL |
| Triglycerides | 6.02 mmol/l | 526.75 mg/dL |
| Total Bilirubin | 535 µmol/l | 31.32 mg/dL |
| Direct Bilirubin | 433 µmol/l | 25.34 mg/dL |
| Ascorbic Acid | 340 μmol/L | 5.98 mg/dL |
| AcetylsalicylicAcid | 3.62 mmol/L | 65.16 mg/dL |
| Ibuprofen | 2.43 mmol/L | 50.10 mg/dL |
| Acetaminophen | 1324 µmol/L | 20 mg/dL |
{23}------------------------------------------------
Image /page/23/Picture/0 description: The image shows the HORIBA Medical logo at the top. Below the logo, the text "Section 007. 510(k) summary" is displayed. The text is in a light gray color and appears to be part of a document or report.
8.4 Yumizen C1200 Transferrin : Anticoagulant study
Study materials :
Anticoagulant : heparin-lithium Samples: single donors
59 paired serum/ plasma samples were evaluated on Yumizen C1200 analyser using Yumizen C1200 Transferrin reagent.
For this study, each paired samples (sera and heparinized plasma) has been obtained from single donor.
| Passing Bablok | N | Min | Max | Intercept | Slope | Correlation |
|---|---|---|---|---|---|---|
| Transferrin:serum sample(g/L) | 59 | 1.933 | 3.786 | |||
| Transferrin:Heparinsample(g/L) | 59 | 1.901 | 3.724 | 0.04833 | 0.9691 | 0.995 |
Conclusion :
The results show there is not significative difference between serum and plasma with heparin specimens
-> coagulation does not have an impact on Transferrin determination.
{24}------------------------------------------------
Image /page/24/Picture/0 description: The image shows the logo for HORIBA Medical, with the word "HORIBA" in blue and "Medical" in a smaller font size below it. Below the logo, the text "Section 007. 510(k) summary" is displayed in a smaller, gray font. The text appears to be a section heading or title, possibly from a document or report.
8.5 Prozone / Antigen excess effect
For each analyte, a range of high concentration samples is prepared in order to identify a high dose hook effect.
Yumizen C1200 Ferritin ●
An antigen excess effect is observed for samples with a concentration higher than 5043 ng/mL, as they have a rate which could be inside the calibration range and the assay could give underestimated results, but remaining inside the pathological range.
For such samples, a prozone alarm will flag the samples having concentration higher than 5043 ng/mL. The flagged sample will be automatically rerun by the analyzer.
● Yumizen C1200 Transferrin
No antigen excess has been detected up to a concentration of 40 g/L.
● Yumizen C1200 Rheumatoid Factor
An antigen excess effect is observed for samples with a concentration higher than 229 IU/mL, as they have a rate which could be inside the calibration range and the assay could give underestimated results, but remaining inside the pathological range.
For such samples, a prozone alarm will flag the samples having concentration higher than 229 IU/mL. The flagged sample will be automatically rerun by the analyzer.
{25}------------------------------------------------
Image /page/25/Picture/0 description: The image shows the logo for HORIBA Medical at the top. The text "Medical" is written in a smaller font size below the word HORIBA. Below the logo, the text "Section 007. 510(k) summary" is written in a gray font. The text appears to be part of a document or report.
8.6 Method comparison with a predicate device
Yumizen C1200 Ferritin ●
This study has been carried out using recommendations found in the CLSI EP-9A3 guidance.
Samples: Anonymous remnants of human serum specimens collected from blood bank. These samples are in the candidate measuring range and predicate measuring range.
103 native sera samples have been assayed in duplicate, in ascendant order and descendant order on 6 working days.
Only the first replicate of each method will be used for the data analysis reported below.
| Passing Bablok | N | Min | Max | Intercept | Slope | Correlation – r2 |
|---|---|---|---|---|---|---|
| (ng/mL) | 103 | 16.74 | 413.00 | 2.105 | 0.9142 | 0.999 |
● Yumizen C1200 Transferrin
This study has been carried out using recommendations found in the CLSI EP-9A3 guidance. Samples: Anonymous remnants of human serum specimens collected from CHU Nîmes (University Hospital Center).
115 native samples have been assayed in duplicate, in ascendant order and descendant order on 5 working days.
Only the first replicate of each method will be used for the data analysis reported below.
| Passing Bablok | N | Min | Max | Intercept | Slope | Correlation - r² |
|---|---|---|---|---|---|---|
| (g/L) | 115 | 0.37 | 4.81 | 0.006364 | 0.9455 | 0.993 |
Yumizen C1200 Rheumatoid Factor ●
This study has been carried out using recommendations found in the CLSI EP-9A3 guidance.
Samples: Anonymous remnants of human serum specimens collected from blood bank. These samples are in the candidate measuring range and predicate measuring range.
113 native samples have been assayed in duplicate, in ascendant order and descendant order on 6 working days.
Only the first replicate of each method will be used for the data analysis reported below.
| Passing Bablok | N | Min | Max | Intercept | Slope | Correlation – r² |
|---|---|---|---|---|---|---|
| (IU/mL) | 113 | 16.7900 | 118.8100 | -1.11 | 1.014 | 0.992 |
{26}------------------------------------------------
Image /page/26/Picture/0 description: The image shows the logo for HORIBA Medical in blue text. Below the logo is the text "Section 007. 510(k) summary" in a smaller, gray font. The text appears to be the title or heading of a document or section within a document. The image is simple and focuses on the company name and the section title.
8.7 Reagent Stability
8.7.1 Closed stability
The closed stability was determined according to the CLSI guideline EP25-A.
Yumizen C1200 Ferritin ●
Stability before opening:
Stable up to the expiry date on the label if stored at 2-10°C. Store protected from light.
The shelf life claim for HORIBA Medical reagent is 18 months.
. Yumizen C1200 Transferrin
Stability before opening:
Stable up to the expiry date on the label if stored at 2-8°C. Store protected from light.
The shelf life claim for HORIBA Medical reagent is 24 months.
● Yumizen C1200 Rheumatoid Factor
Stability before opening:
Stable up to the expiry date on the label if stored at 2-10°C.
The shelf life claim for HORIBA Medical reagent is 18 months.
8.7.2 Open stability
The open stability was determined according to the CLSI guideline EP25-A.
On board reagent Stability:
- The reagent stability claim for the Yumizen C1200 Ferritin is 2 months.
- The reagent stability claim for the Yumizen C1200 Transferrin is 6 weeks.
- The reagent stability claim for the Yumizen C1200 Rheumatoid Factor is 1 month.
{27}------------------------------------------------
Image /page/27/Picture/0 description: The image shows the logo for HORIBA Medical, followed by the text "Section 007. 510(k) summary". The logo is in blue, and the text is in gray. The text is likely a section heading or title within a larger document.
8.8 Reference range
The Reference Range was determined according to the CLSI guideline EP28-A3.
● Yumizen C1200 Ferritin
- A Adults data (bibliographic reference and study- see below)
- . Women :
50 "normal samples" from blood bank have been assayed with the method in evaluation. Each sample is assaved in duplicates.
Only the first replicate of each method will be used for the data analysis reported below.
Only the first replicate result for each subject was compared against reference ranges cited in literature. The verification studies support in the following reference ranges which were established through literature.
Women : 10 - 120 ng/ml (µg/l)
Reference:
Roberts W.L., McMillin G.A., Burtis C.A., Bruns D.E., Reference Information for the Clinical Laboratory, TIETZ Textbook of Clinical Chemistry and Molecular Diagnostics. 4ème Ed; Burtis C.A., Ashwood E.R., Bruns D.E., (Elsevier Saunders eds. St Louis, USA); 2006, 2269.
- . Men :
95 "normal samples" from blood bank have been assayed with the method in evaluation. Each sample is assayed in duplicates.
Only the first replicate of each method will be used for the data analysis reported below.
Only the first replicate result for each subject was compared against reference ranges cited in literature. The verification studies support in the following reference ranges which were established through literature.
Men: 20 - 250 ng/ml (µg/1)
Reference:
Roberts W.L., McMillin G.A., Burtis C.A., Bruns D.E., Reference Information for the Clinical Laboratory, TIETZ Textbook of Clinical Chemistry and Molecular Diagnostics. 4ème Ed; Burtis C.A., Ashwood E.R., Bruns D.E., (Elsevier Saunders eds. St Louis, USA); 2006, 2269.
{28}------------------------------------------------
K191562 Yumizen C1200 Reagents
● Yumizen C1200 Transferrin
85 "normal samples" (28 women + 57 men) from blood bank have been assayed with the method in evaluation. Each sample is assayed in duplicates.
Only the first replicate of each method will be used for the data analysis reported below.
Only the first replicate result for each subject was compared against reference ranges cited in literature. The verification studies support in the following reference ranges which were established through literature.
Normal range Transferrin - Serum
2 - 3.6 g/l (200 - 360 mg/dl) based on CRM 470.
Reference:
Dati F., Schumann G, Thomas L, Aguzzi F, Baudner S. Bienvenu J. et al. Consensus of a group of professional societies and diagnostic companies on guidelines for interim reference ranges for 14 proteins in serum based on the standardization against IFCC/BCR/CAP Reference Material (CRM 470). Eur. J Clin Chem. Cli Biochem. 1996; 34: 517-20.
. Yumizen C1200 Rheumatoid Factor
60 "normal samples" from blood bank have been assayed with the method in evaluation. Each sample is assayed in duplicates.
Only the first replicate of each method will be used for the data analysis reported below.
Only the first replicate result for each subject was compared against reference ranges cited in literature. The verification studies support in the following reference ranges which were established through literature.
Normal range Rhumatoid Factor Adult < 14 IU/ml
Reference:
Tietz NW, editor. Clinical Guide to Laboratory Tests, 3rd ed. Philadelphia: WB Saunders Company, 1990.
{29}------------------------------------------------
Image /page/29/Picture/0 description: The image shows the HORIBA Medical logo at the top. Below the logo, the text "Section 007. 510(k) summary" is displayed. The text is in a smaller font size compared to the logo. The overall image appears to be a snippet from a document or presentation.
8.9 Proposed Labeling
The labeling is written as per the recommendations given in standard EN18113-2. It takes into account the requirements of 21 CFR Part 809.10.
Conclusions for Performance Testing 8.10
The performance testing data conclude that the safety and effectiveness of the devices are not compromised, and that they met all acceptance criteria, demonstrating that each device is substantially equivalent to its predicate device.
§ 866.5340 Ferritin immunological test system.
(a)
Identification. A ferritin immunological test system is a device that consists of the reagents used to measure by immunochemical techniques the ferritin (an iron-storing protein) in serum and other body fluids. Measurements of ferritin aid in the diagnosis of diseases affecting iron metabolism, such as hemochromatosis (iron overload) and iron deficiency amemia.(b)
Classification. Class II (performance standards).