Search Results
Found 105 results
510(k) Data Aggregation
(403 days)
Re: K231151
Trade/Device Name: Kenota 1 Total IgE, Kenota 1 (instrument) Regulation Number: 21 CFR 866.5510
---------------------------------|-----------------|
| DGC | Class II | 21 CFR 866.5510
The Kenota 1 Total IgE is an in vitro test system intended for semi-quantitative measurement of total IgE in human capillary whole blood on the Kenota 1 instrument. It is intended for in vitro diagnostic use as an aid in the clinical diagnosis of IgE-mediated allergic disorders in conjunction with other clinical findings, and is to be used in allergist/immunologist offices.
The Kenota 1 Total IgE test system consists of Kenota 1 Total IgE Cartridge (reagent) and other test components, including Kenota 1 Total IgE External Controls (EC), Kenota 1 (instrument), Kenota 1 Sample Collection Kit (SCK), and Kenota 1 Developing Solution.
Here's an analysis of the provided text to extract acceptance criteria and study details for the Kenota 1 Total IgE device:
Acceptance Criteria and Reported Device Performance
The acceptance criteria for semi-quantitative classification are implied by the "Percent in ATE" (Allowable Test Error) within the method comparison study, where ATE means the result is within one category of the comparator. While explicit thresholds for numeric precision are provided, the primary acceptance for classification appears centered around this categorical agreement.
Performance Metric | Acceptance Criteria (Implied/Directly Stated) | Reported Device Performance |
---|---|---|
Semi-quantitative Agreement (Overall) | Not explicitly stated as a percentage, but shown for each range and overall. The study aims to achieve a high "Percent in ATE" (Allowable Test Error). | 96.4% in ATE (344/357), with a 95% CI of (93.9%; 97.9%) |
Semi-quantitative Agreement (Low Range) | Implied high percentage in ATE | 96.8% in ATE (183/189) |
Semi-quantitative Agreement (Middle Range) | Implied high percentage in ATE | 96.6% in ATE (84/87) |
Semi-quantitative Agreement (High Range) | Implied high percentage in ATE | 95.1% in ATE (77/81) |
Erroneous Results (LER) | Implied low percentage, ideally 0% | 0.0% in LER (0/357), with a 95% CI of (0.0%; 1.1%) |
Within-laboratory Precision (Cartridge lots) - %CV | Not explicitly stated, but generally, lower %CV indicates better precision. | Low: 19.6%; Medium: 14.5%; High: 15.9%; Very High: 9.7% |
Within-laboratory Precision (Instruments) - %CV | Not explicitly stated, but generally, lower %CV indicates better precision. | Low: 13.6%; Medium: 14.0%; High: 15.3%; Very High: 6.8% |
Fingerstick Repeatability - %CV | Not explicitly stated, but generally, lower %CV indicates better repeatability. | 5-34 kU/L: 10.6%; 35-100 kU/L: 8.7%; 101-200 kU/L: 9.8%; 201-540 kU/L: 8.4%; 541-900 kU/L: 7.2%; Entire Range: 8.3% |
Multi-site Reproducibility (Controls) - %CV | Not explicitly stated, but generally, lower %CV indicates better reproducibility. | Low Control: 12.2%; High Control: 11.6% |
Linearity (Categorical Agreement) | 100% agreement expected where possible, with some allowance for boundary cases (e.g., 900 range, 53.3% for ±2 degrees tilt. | |
Lighting (Bias) |
Ask a specific question about this device
(28 days)
Trade/Device Name: N Antisera to Human Immunoglobulins (IgG, IgA, and IgM) Regulation Number: 21 CFR 866.5510
Immunoglobulins (IgG, IgA, and IgM) |
| Regulation Number: | 21 CFR 866.5510
Immunoglobulins (IgG, IgA, and IgM) / K083445 |
| Regulation Number: | 21 CFR 866.5510
br>and IgM) |
| Regulation Number
Regulation Description | 21 CFR 866.5510
In-vitro diagnostic reagents for the quantitative determination of immunoglobulins (IgG. IgA and IgM) in human serum, heparinized and EDTA plasma, and IgG in human urine and cerebrospinal fluid (CSF) by means of immunonephelometry on the BN II and BN ProSpec® System. Measurements of IgG aid in the diagnosis of abnormal protein metabolism and the body's lack of ability to resist infectious agents.
The N Antiserum to Human IgG reagent containing animal serum, produced by immunization of rabbits with highly purified human immunoglobulin (
The provided text describes a special 510(k) premarket notification for a modified device, "N Antisera to Human Immunoglobulins (IgG, IgA, and IgM)". The sole modification is the addition of a High Dose Hook (HDH) effect claim for IgG in cerebrospinal fluid (CSF) samples.
Here's a breakdown of the requested information based on the provided document:
1. A table of acceptance criteria and the reported device performance
Acceptance Criteria | Reported Device Performance |
---|---|
Minimum High Dose Hook limit for CSF samples | High Dose Hook limit shown for all three lots up to the maximum measured concentration of 1130 mg/L |
Adherence to a minimum HDH limit of up to 412 mg/L | Exceeded; the device demonstrated an HDH limit of 1130 mg/L |
2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)
- Sample Size: The study used a "CSF high sample pool." The exact number of individual patient samples contributing to this pool (if multiple were pooled) is not specified. However, the study involved a dilution scheme with twelve (12) individual dilution levels, including the neat sample.
- Data Provenance: Not explicitly stated. The manufacturer is Siemens Healthcare Diagnostics Products GmbH, located in Marburg, Germany, which suggests the study was likely conducted in Germany or a similar geographic region. It is implicitly a prospective study designed to evaluate the HDH effect for the modified device.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)
This section is not applicable as the device is an in-vitro diagnostic reagent for quantitative determination, not an imaging or interpretive device that would typically require expert ground truth establishment in the described manner. The "ground truth" for this type of test is the quantitatively measured concentration of IgG in a sample, established through laboratory methods and comparison to known standards.
4. Adjudication method (e.g. 2+1, 3+1, none) for the test set
This section is not applicable. Adjudication methods like 2+1 or 3+1 are typically used for establishing ground truth in interpretive studies (e.g., radiologists reviewing images). For a quantitative in-vitro diagnostic test, the "ground truth" is determined by the analytical method itself against known standards.
5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
This is not applicable. The device is an in-vitro diagnostic reagent, not an AI-assisted diagnostic tool or an imaging device requiring human reader interpretation. No MRMC study was performed.
6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done
This refers to the performance of the analytical system (N Antisera reagent on BN II System) without human intervention in the measurement process. The HDH study performed is a standalone performance evaluation of the reagent/instrument system. The acceptance criteria and performance data in the table above demonstrate this standalone performance.
7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)
For this in-vitro diagnostic device, the "ground truth" for the High Dose Hook study was established by creating known dilutions of a high-concentration CSF sample, which were then measured by the device. The reported concentrations for these dilutions serve as the reference. The ultimate analytical ground truth for the quantitative measurement itself is tied to international standards like ERM-DA470k/IFCC (as stated in the "Traceability/Standardization" section).
8. The sample size for the training set
This document does not describe the development or training of a machine learning model, so there is no training set in the typical sense. The "training" for such a diagnostic test involves method development, optimization, and validation using various samples and controls, but these are not referred to as a "training set" for an algorithm.
9. How the ground truth for the training set was established
As there is no training set for an algorithm, this question is not applicable.
Ask a specific question about this device
(474 days)
Republic of China
Re: K221114
Trade/Device Name: Immunoglobulin G (IgG) Regulation Number: 21 CFR 866.5510
| CFN |
| Regulation Number: | 21 CFR §866.5510
System reagent for the quantitative determination of IgG immunoglobulins in human serum, plasma and cerebrospinal fluid on Beckman Coulter AU/DxC AU analyzers. The measurement of IgG aids in the diagnosis of abnormal protein metabolism and the body's lack of ability to resist infectious agents.
The device consists of two reagents: R1 buffer (Tris buffer pH 7.2, polyethylene glycol 6000) and R2 (goat anti-IgG antiserum). The reagents contain sodium azide as preservative.
When a sample is mixed with R1 buffer and R2 antiserum solution, human IqG reacts specifically with anti-human IgG antibodies to yield insoluble aggregates. Immune complexes formed in solution scatter light in proportion to their size, shape, and concentration. Turbidimeters then measure the reduction of incidence light due to reflection, absorption, or scatter. The decrease in intensity of light transmitted (increase in absorbance) through particles suspended in solution is a result of complexes formed during the antigen-antibody reaction.
The Beckman Coulter Immunoglobulin G (IgG) reagent for quantitative determination of IgG immunoglobulins in human serum, plasma, and cerebrospinal fluid on Beckman Coulter AU/DxC AU analyzers, device K221114, underwent various non-clinical (bench) studies to demonstrate substantial equivalence to its predicate device (K162208).
Here is a summary of the acceptance criteria and reported device performance based on the provided document:
1. Table of Acceptance Criteria and Reported Device Performance
Study Type | Sample Type | Acceptance Criteria | Reported Device Performance | Pass/Fail |
---|---|---|---|---|
Method Comparison | Serum | Slope: Not explicitly stated, but R-value of 0.9981 suggests strong correlation. | Slope: 1.015, Intercept: -25.422, R: 0.9981 | Pass |
CSF | Slope: Not explicitly stated, but R-value of 0.9995 suggests strong correlation. | Slope: 0.998, Intercept: 0.1141, R: 0.9995 | Pass | |
Linearity/Reportable Range | Serum | Linear Range: 75-3,000 mg/dL | ||
Allowable Difference: ±8% between 375-3,000 mg/dL; ±30 mg/dL between 75-375 mg/dL | Linear From: 73.2868 mg/dL | |||
Linear To: 3261.9190 mg/dL | Pass | |||
CSF | Linear Range: 2-50 mg/dL | |||
Allowable Difference: ±10% between 2-50 mg/dL; ±0.5 mg/dL between 2.0-5 mg/dL | Linear From: 1.9 mg/dL | |||
Linear To: 53.0 mg/dL | Pass | |||
Sensitivity (LOQ) | Serum |
Ask a specific question about this device
(421 days)
d'Amunt, Barcelona 08186 Spain
Re: K214068
Trade/Device Name: Quantia IgE Regulation Number: 21 CFR 866.5510
|
| Regulatory Information | Regulation Number
21 CFR 866.5510
Automated latex enhanced immunoassay for the quantitative in vitro determination of total immunoglobulin E (1gE) in human serum or plasma (EDTA, heparin, citrate) using the ARCHITECT c Systems. The measurement of total IgE is useful in the clinical diagnosis of IgE-mediated allergies, if used in conjunction with other clinical studies.
The Quantia IgE reagent is a suspension of polystyrene latex particles of uniform size coated with mouse anti-human IgE. When a sample containing IgE is mixed with the latex reagent and the reaction buffer included in the kit, agglutination occurs. The degree of agglutination is directly proportional to the concentration of IgE in the sample and is determined by measuring the decrease of transmitted light caused by the aggregates. Methodology: Turbidimetric/Immunoturbidimetric.
The provided document outlines the acceptance criteria and study results for the Quantia IgE assay, a device for quantitatively determining total IgE in human serum or plasma. It's important to note that this document is a 510(k) summary, focusing on demonstrating substantial equivalence to a predicate device after a modification, rather than a comprehensive de novo approval study. Therefore, some information typically found in a de novo clinical trial report (e.g., specific details on training set size, number of experts for training ground truth) might not be explicitly detailed.
Here's an analysis of the provided information:
1. Table of Acceptance Criteria and Reported Device Performance
The acceptance criteria are generally implied by the performance characteristics presented and their comparison to the predicate device or established clinical standards (e.g., CLSI guidelines). The document focuses on demonstrating that the modified device performs as well as the predicate and meets relevant analytical performance metrics.
Performance Metric | Predicate Device (K050493) Acceptance Criteria/Performance | Subject Device (Modified Quantia IgE) Reported Performance |
---|---|---|
Linearity (Reportable Range) | 25.0 - 1000.0 IU/mL | 20.0 - 1000.0 IU/mL (Acceptable linearity demonstrated across this range) |
Limit of Blank (LoB) | Not defined | 6.2 IU/mL |
Limit of Detection (LoD) | 12.9 IU/mL | 11.6 IU/mL |
Limit of Quantitation (LoQ) | 25.0 IU/mL | 20.0 IU/mL |
Precision (Total %CV) |
Ask a specific question about this device
(61 days)
California 92821
Re: K220178
Trade/Device Name: Total Immunoglobulin E (IgE) Regulation Number: 21 CFR 866.5510
Immunoglobulin E (IgE)
3.2 Classification Name
Immunoglobulins A, G, M, D, and E immunological test system [866.5510
The IgE assay is intended for use in the quantitative determination of Total Immunoglobulin E (lgE) concentration in human serum and plasma (lithium heparin, K2-EDTA, K3-EDTA, K3-EDTA) on Beckman Coulter AU/DxC AU clinical chemistry analyzers. The determination aids in the diagnosis of IgEmediated allergic disorders in conjunction with other clinical findings. For in vitro diagnostic use only.
The Total Immunoglobulin E (IgE) reagent kit is in a liquid ready-to-use format designed for optimal performance on Beckman Coulter's AU/DxC AU clinical chemistry analyzers. Each reagent kit contains one buffer reagent (R1), one antibody reagent (R2), and a six-level lot matched calibrator set. The IgE reagent test system utilizes a turbidimetric immunoassay methodology. The AU analyzer measures the change in absorbance at 800 nm to calculate and express the concentration of immunoglobulin E in the test sample based on a stored calibration curve. The IgE assay is traceable to the World Health Organization (WHO) 3rd International Standard 11/234.
The provided document is a 510(k) Premarket Notification for the Beckman Coulter Total Immunoglobulin E (IgE) assay. It describes the device's performance characteristics and how they meet established acceptance criteria, demonstrating substantial equivalence to a predicate device.
Here's an analysis of the acceptance criteria and study data based on the provided document:
1. Table of Acceptance Criteria and Reported Device Performance
The document presents several tables detailing acceptance criteria and study results for different performance parameters.
Table 1: Method Comparison Summary (from Table 9.1.1)
This table compares the Beckman Coulter IgE Assay (candidate device) against the Roche Elecsys IgE II Assay (predicate device).
Acceptance Criteria | Parameter | Reported Device Performance | Pass/Fail |
---|---|---|---|
Not explicitly stated as a separate column, but implied by the comparison to predicate. | Slope | 0.966 [0.950 - 0.981] | Pass (implied, as good correlation is shown) |
Intercept (IU/mL) | 1.0 [-1.0 - 3.0] | Pass (implied, as good correlation is shown) | |
R (Correlation Coefficient) | 0.996 | Pass (implied, as good correlation is shown) |
Table 2: Precision Performance Summary (from Table 9.2.1)
Test Sample | Mean (IU/mL) | Acceptance Criteria (Repeatability CV%) | Reported Repeatability CV (%) | Acceptance Criteria (Total Precision CV%) | Reported Total Precision CV (%) | Pass/Fail |
---|---|---|---|---|---|---|
Control 1 | 113.5 | ≤7.0% | 1.7 | ≤7.5% | 2.0 | Pass |
Control 2 | 229.4 | ≤7.0% | 1.0 | ≤7.5% | 1.6 | Pass |
Pool 1 | 70.4 | ≤5.0 IU/mL | 3.0 (CV%) / 2.1 (IU/mL) | ≤7.0 IU/mL | 3.3 (CV%) / 2.3 (IU/mL) | Pass |
Pool 2 | 167.9 | ≤7.0% | 1.4 | ≤7.5% | 2.4 | Pass |
Pool 3 | 413.6 | ≤7.0% | 0.9 | ≤7.5% | 1.4 | Pass |
Note: For Pool 1, the document states criteria of "≤5.0 IU/mL" for repeatability and "≤7.0 IU/mL" for total precision. The reported results are 2.1 IU/mL (repeatability) and 2.3 IU/mL (total precision), which meet these criteria.
Table 3: Analytical Range (Linearity) Study Summary (from Table 9.3.1)
Sample Type | Acceptance Criterion (Linear Range) | Reported Linear From (IU/mL) | Reported Linear To (IU/mL) | Pass/Fail |
---|---|---|---|---|
Serum | 20 – 500 IU/mL (with Allowable Difference of 10.0 IU/mL or 10%*) | 13.2 | 575.5 | Pass |
Further breakdown of linearity criteria for individual points in Table 9.3.2 shows "Bias Spec" of 10.0 for values ≤100 IU/mL and "%Bias Spec" of 10.0 for values > 100 IU/mL. All individual points "Pass".
Table 4: Detection Limit Study Summary (LoB, LoD & LoQ) (from Table 9.4.1)
Parameter | Acceptance Criteria | Reagent Lot 1 Result | Reagent Lot 2 Result | Pass/Fail |
---|---|---|---|---|
LoB (IU/mL) | ≤10.0 | 5.9 | 7.5 | Pass |
LoD (IU/mL) | ≤15.0 | 13.8 | 12.5 | Pass |
LoQ (IU/mL) | ≤20.0 at ≤35% CV | 19.6 at 11.9% CV | 17.8 at 7.8% CV | Pass |
Table 5: Anticoagulant Study Results Summary (from Table 9.7.1)
Plasma Type | Acceptance Criteria (Slope) | Reported Slope | Acceptance Criteria (Intercept) | Reported Intercept | Acceptance Criteria (R) | Reported R | Pass/Fail |
---|---|---|---|---|---|---|---|
Na Heparin | [0.9 - 1.1] | 0.989 | [± 20 IU/mL] | 0.0 | [≥ 0.97] | 0.999 | Pass |
Li Heparin | [0.9 - 1.1] | 0.989 | [± 20 IU/mL] | -0.5 | [≥ 0.97] | 0.999 | Pass |
K2 EDTA | [0.9 - 1.1] | 0.986 | [± 20 IU/mL] | -1.7 | [≥ 0.97] | 0.997 | Pass |
K3 EDTA | [0.9 - 1.1] | 0.964 | [± 20 IU/mL] | -2.4 | [≥ 0.97] | 0.998 | Pass |
Table 6: In-Use Stability Study Summary (from Table 9.8.1)
Stability Parameter | Claim (days) | Tested to (days) | Pass/Fail |
---|---|---|---|
Reagent open bottle* | 28 | 29 | Pass |
Calibration interval† | 14 | 15 | Pass |
Calibrator open bottle | 45 | 54 | Pass |
Note: For the stability studies, the acceptance criteria are implicitly met if the "Pass" status is indicated after testing beyond the claimed duration, implying that the mean IgE recovery criteria (within 10 IU/mL or 10% of Day 0 mean) were met.
2. Sample sizes used for the test set and the data provenance
- Method Comparison: 136 fresh serum samples spanning the analytical measuring range. Provenance not explicitly stated.
- Precision: Three lots of IgE reagent, tested using two levels of human serum-based quality control material and three patient pools. The experimental design used duplicate sample analysis twice daily over twenty working days (N=80 results for each control/pool). Provenance not explicitly stated.
- Linearity: 15-level linearity test set of inter-diluted patient pools. Provenance not explicitly stated.
- Sensitivity (LoB, LoD, LoQ): LoB evaluated four unique lots of Immunoglobulin (Ig)-depleted human serum. LoD and LoQ used native patient pools diluted with Ig-depleted serum. Provenance not explicitly stated.
- Analytical Specificity (Interference): Test samples containing various interferents (Hemoglobin, Bilirubin, Lipemia, RF, 21 common drugs and concentrations). Specific sample sizes per interferent are not given but are implied by tables (e.g., "Level Tested"). Provenance not explicitly stated.
- Anticoagulant Studies: Freshly drawn serum and plasma from apparently healthy adult volunteer donors. Five specimen tubes were drawn from each donor: one serum tube and one tube of each type of anticoagulant. Sample sizes listed for each anticoagulant type: Na Heparin (N=55), Li Heparin (N=55), K2 EDTA (N=55), K3 EDTA (N=53). Provenance not explicitly stated.
- In-Use Stability: Three levels of quality control material were evaluated. Specific N is not detailed, but implied by the "classical design for sampling, storage, and testing." Provenance not explicitly stated.
Generally, the data provenance (e.g., country of origin, retrospective/prospective) is not explicitly mentioned for any of the studies in the provided document. The studies appear to be prospective analytical performance studies designed to evaluate the device's technical specifications.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts
This device is an in vitro diagnostic (IVD) quantitative assay for Total IgE. The "ground truth" for such devices is typically established through:
- Traceability to International Standards: The IgE assay is stated to be traceable to the World Health Organization (WHO) 3rd International Standard 11/234. This is a primary method of establishing "ground truth" for quantitative lab tests, where the standard itself defines the "true" concentration.
- Reference Methods: The method comparison study uses a legally marketed predicate device (Roche Elecsys IgE II Assay) as the comparative "truth" or reference standard for patient samples. The predicate device itself would have undergone similar rigorous validation, likely traceable to international standards.
Therefore, for this type of device, the concept of "experts" in the sense of clinicians or radiologists establishing a diagnostic "ground truth" for individual cases (like in an imaging study) is not directly applicable. The "ground truth" for the test set is inherent in the certified international standard and the established performance of the predicate device.
4. Adjudication method (e.g. 2+1, 3+1, none) for the test set
Not applicable. Adjudication methods like 2+1 or 3+1 consensus are typically used in studies involving subjective interpretation (e.g., imaging reads) to establish a consensus ground truth. For quantitative in vitro diagnostic assays, the "truth" is determined by reference methods, international standards, and analytical evaluation against defined statistical criteria, not by human consensus or adjudication of individual results.
5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
Not applicable. This is an in vitro diagnostic (IVD) assay, not an AI-powered diagnostic imaging device or an AI-assisted interpretation tool for human readers. Therefore, MRMC studies and the concept of "human readers improving with AI assistance" do not apply.
6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done
Yes, the studies presented are all "standalone" in the sense that they evaluate the performance of the analytical instrument and reagent system itself (the "algorithm only" in a broader sense of the measurement process) without direct human intervention in the result generation or interpretation step for the purpose of the performance evaluation. The device provides a quantitative result (IgE concentration), which is then used by clinicians in conjunction with other clinical findings. The performance data (e.g., precision, linearity, sensitivity, method comparison) directly reflects the device's inherent analytical capabilities.
7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)
The ground truth for this IgE assay is established primarily through:
- International Standards: The assay's traceability to the WHO 3rd International Standard 11/234 is explicitly stated and serves as the ultimate reference for IgE concentration.
- Reference Measurement Procedures/Predicate Device: For method comparison, the Roche Elecsys IgE II Assay serves as the comparative "reference" for patient samples. This predicate device itself is assumed to be traceable to international standards and validated.
- A priori defined analytical specifications: For precision, linearity, and detection limits, the "ground truth" or acceptance criteria are statistically derived technical specifications (e.g., CV limits, bias limits) that the device must meet to be considered analytically robust.
8. The sample size for the training set
The document does not describe a "training set" in the context of machine learning or AI models because this is not an AI/ML-based device. It is a traditional in vitro diagnostic immunoassay. The performance studies detailed are validation studies for the cleared device.
9. How the ground truth for the training set was established
Not applicable, as there is no "training set" in the AI/ML sense for this traditional IVD product. If "training set" broadly refers to materials used during the development and calibration of the assay, the document indicates that the assay is traceable to the WHO 3rd International Standard 11/234, and a six-level lot-matched calibrator set is included, which would be used to establish the calibration curve for the assay.
Ask a specific question about this device
(381 days)
The HYDRASHIFT 2/4 isatuximab kit is intended for the qualitative detection of monoclonal proteins in human serum by immunofixation electrophoresis. The HYDRASHIFT 2/4 isatuximab kit is to be used in conjunction with the HYDRAGEL IF kit and the semi-automated HYDRASYS 2 electrophoresis apparatus. The electropherograms are evaluated visually for the presence of specific reactions with the suspect monoclonal proteins. The HYDRASHIFT 2/4 isatuximab kit removes isatuximab IgG kappa interference and enables the visual evaluation of the presence of monoclonal proteins on HYDRAGEL IF kits in patients who have received isatuximab therapy.
For In Vitro Diagnostic use. For Prescription Use Only.
HYDRASYS 2 is a semi-automated multi-parameter system for start-to finish agarose gel electrophoresis: application of samples, migration, drying, staining, destaining and final-stage drying.
Abnormal bands in serum protein electrophoregrams, primarily those in the beta globulin and gamma globulin zones, are always suspected to be monoclonal proteins (M-proteins, paraproteins, monoclonal immunoqlobulins) and therefore, an indication of performing an Immunofixation technique to type and confirm the monoclonal gammopathies.
lsatuximab is a human therapeutic IgG kappa monoclonal antibody and as such, during the clinical monitoring of patients treated with isatuximab, this antibody simulates a band detected by serum protein electrophoresis and immunofixation in the gamma region. It can simulate an endogenous IgG kappa paraprotein.
The HYDRASHIFT isatuximab immunofixation procedure, performed on the HYDRAGEL IF 2/4 gel, is based on the creation of an isatuximab / anti-isatuximab antibody complex and shifting it outside the gammaglobulins zone. With the HYDRASHIFT isatuximab procedure, the isatuximab / anti-isatuximab antibody complex is visualized in alpha-1 zone on IgG and Kappa immunofixation tracks and then the interference is removed from the gamma zone.
The provided text describes the performance data for the HYDRASHIFT 2/4 isatuximab kit, which is intended for the qualitative detection of monoclonal proteins in human serum by immunofixation electrophoresis, specifically by removing isatuximab IgG kappa interference.
Here's an analysis of the acceptance criteria and study proving the device meets them, based on the provided document:
Acceptance Criteria and Reported Device Performance
The core acceptance criterion for this device, based on the performance data presented, is 100% concordance in visual evaluation of the presence or absence of monoclonal proteins, particularly after the removal of isatuximab interference.
Acceptance Criteria (Inferred from Performance Goals) | Reported Device Performance |
---|---|
Repeatability: Consistent visual evaluation of monoclonal proteins within the same gel and consistent removal of isatuximab interference. | 100% concordant results for all tested samples across 4 runs within the same gel. |
Reproducibility (between gels, lots, instruments): Consistent visual evaluation of monoclonal proteins and consistent removal of isatuximab interference across different gels, kits from different lots, and different instruments over multiple days. Consistency in characterization (normal or abnormal with monoclonal components) should be maintained. | 100% concordant results for all tested samples across 9 runs (3 gels/day x 3 days) on 3 different instruments and with 3 different kit lots. |
Comparative Studies (Internal & External): The device must effectively remove isatuximab interference while accurately characterizing normal and pathological serum samples, demonstrating 100% agreement between the standard procedure and the HYDRASHIFT procedure for the characterization of monoclonal proteins in both spiked and native samples, and in samples where isatuximab interference is present or absent. The device should allow for clear visualization and characterization of monoclonal proteins after interference removal. | Internal Study (53 samples): 100% complete agreement between native and isatuximab-spiked samples for 26 normal and 27 pathological serum samples. Monoclonal proteins were detected and characterized with 100% concordance. |
External Study No. 1 (204 samples): 100% concordant results for 69 normal serum samples and 135 pathological serum samples between the HYDRAGEL 4 IF Acid Violet Dynamic Mask kit and the HYDRASHIFT 2/4 isatuximab procedure. The kit successfully shifted isatuximab in 90 samples where it was visualized. | |
External Study No. 2 (203 samples): 100% concordant results for 68 normal serum samples and 135 pathological serum samples between the HYDRAGEL 4 IF Acid Violet Standard Mask kit and the HYDRASHIFT 2/4 isatuximab procedure. The kit successfully shifted isatuximab in 90 samples where it was visualized. | |
Sensitivity (Detection Limit of Isatuximab Interference): The device should effectively handle isatuximab interference at relevant clinical concentrations. | The detection limit of isatuximab and/or isatuximab/anti-isatuximab antibody complex visualized is 0.3 g/L. |
Interference: The device results should not be affected by common endogenous interfering factors or specific drugs relevant to the patient population. | No interference detected due to specified concentrations of unconjugated bilirubin, conjugated bilirubin, triglycerides, hemoglobin, rheumatoid factor, Human Anti-Mouse Antibody (HAMA), and various chemotherapy drugs (Pomalidomide, Carfilzomib, Dexamethasone, Ixazomib, Cyclophosphamide, Melphalan, Prednisone, Lenalidomide, Bortezomib). |
Study Details Proving Device Meets Acceptance Criteria
1. A table of acceptance criteria and the reported device performance:
(See table above)
2. Sample sizes used for the test set and the data provenance:
- Repeatability Study: 10 different serum samples (2 controls, 8 spiked with monoclonal components). 4 runs per sample within the same gel.
- Reproducibility Study: 8 different serum samples with monoclonal components + Normal Control Serum + Isatuximab Control. Each sample analyzed on 9 runs (1 analysis per gel, over 3 working days) on 3 HYDRASYS 2 instruments with 3 lots of HYDRASHIFT 2/4 isatuximab kit.
- Comparative Studies:
- Internal Study: 53 serum samples (26 normal, 27 pathological).
- External Study No. 1: 204 serum samples.
- External Study No. 2: 203 serum samples. (Note: "The same serum samples were analyzed in both external studies with exception of one sample included in external study 1." suggesting ~204 unique samples across external studies.)
Data Provenance:
- Internal Study: Conducted by Sebia (manufacturer).
- External Studies: Performed in the USA.
- Retrospective/Prospective: Not explicitly stated, but the description of "serum samples" and "analyzed with" suggests they were existing or collected samples, making it likely retrospective for the comparative studies. The repeatability and reproducibility studies appear to be prospective experimental designs.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:
The document does not specify the number or qualifications of experts used to establish the ground truth. The results are described as "100% concordant" based on detection and characterization of monoclonal proteins, which implies a pre-established or expert-verified classification for each sample. However, the exact process or personnel involved in this ground truth establishment are not detailed. Given it's an in vitro diagnostic (IVD) device for lab use, the "ground truth" would likely be the result of a reference method interpretation by qualified lab personnel.
4. Adjudication method (e.g., 2+1, 3+1, none) for the test set:
The document does not specify any adjudication method. The outcome measures are presented as "100% concordant results," which suggests a clear, unambiguous read for each test, or that any discrepancies were resolved, though the process for resolution is not described.
5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:
No MRMC study was performed or described. This device is an in vitro diagnostic (IVD) kit for immunofixation electrophoresis, not an AI-assisted diagnostic imaging device for human readers. It automates a lab procedure and provides an electropherogram for visual evaluation. The "visual evaluation" mentioned refers to lab personnel interpreting the results of the gel electrophoresis.
6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:
Yes, effectively. This device is an IVD kit. Its performance is evaluated on its ability to process serum samples and produce an electropherogram that then is visually evaluated. The "performance data" sections (repeatability, reproducibility, comparative studies) essentially describe the standalone performance of the kit itself in consistently producing the expected analytical result (i.e., shifting the isatuximab band and allowing for accurate detection/characterization of monoclonal proteins). While the final "reading" is visual, the kit's function is mechanistic/chemical, not algorithmic interpretation requiring human oversight in the AI sense.
7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):
The ground truth appears to be based on the presence and characterization of monoclonal proteins in serum samples, determined by standard laboratory methods and potentially confirmed by expert consensus in a clinical laboratory setting. For the comparative studies, samples were characterized as "normal" or "abnormal with monoclonal components," implying a reference standard or pre-established truth for each sample type. The "native" samples and "spiked" samples with isatuximab served as a form of ground truth for evaluating the device's ability to differentiate between intrinsic monoclonal proteins and isatuximab interference.
8. The sample size for the training set:
Not applicable/Not specified. This is an in vitro diagnostic (IVD) kit, not a machine learning or AI-based device that typically requires a separate "training set" for model development. The development process for such a kit involves chemical and biochemical optimization rather than data-driven learning.
9. How the ground truth for the training set was established:
As above, not applicable. The device is a wet-lab kit, not an AI algorithm.
Ask a specific question about this device
(190 days)
Immunoglobulin A, Yumizen C1200 Immunoglobulin G, Yumizen C1200 Immunoglobulin M Regulation Number: 21 CFR 866.5510
Class II / 510(k) required |
| Classification Name: | §866.5510
Class II / 510(k) required |
| Classification Name: | §866.5510
Class II / 510(k) required |
| Classification Name: | §866.5510
Yumizen C1200 Immunoglobulin A reagent is intended for the quantitative in vitro diagnostic determination of Immunoglobulin A (IgA) in serum and lithium heparin plasma by immunoturbidimetry on Yumizen analyzers.Measurement of this immunoglobulin aids in the diagnosis of abnormal protein metabolism and the body's lack of ability to resist infectious agents. This test should be used in conjunction with other findings,
Yumizen C1200 Immunoglobulin G reagent is intended for the quantitative in vitro diagnostic determination of Immunoglobulin G (IgG) in serum and lithium heparin plasma by immunoturbidimetry on Yumizen analyzers.Measurement of this immunoglobulin aids in the diagnosis of abnormal protein metabolism and the body's lack of ability to resist infectious agents. This test should be used in conjunction with other findings.
Yumizen C1200 Immunoglobulin M reagent is intended for the quantitative in vitro diagnostic determination of Immunoglobulin M (IgM) in serum and lithium heparin plasma by immunoturbidimetry on Yumizen analyzers. Measurement of this immunoglobulin aids in the diagnosis of abnormal protein metabolism and the body's lack of ability to resist infectious agents.
This test should be used in conjunction with other laboratory and clinical findings.
This submission consists in the Yumizen C1200 Immunoglobulin A (1300023881), Yumizen C1200 Immunoglobulin G (1300023883) and Yumizen C1200 Immunoglobulin M (1300023884) reagent for serum and plasma testing for Yumizen C1200 reagent.
The Yumizen C1200 Level 1 Protein Control (1300023944) and Yumizen C1200 Level 2 Protein Control (1300023945) for use on Yumizen C1200 Analyzer and the Yumizen C1200 Protein Cal (1300023893) for use on Yumizen C1200 Analyzer are sold separately.
This document describes the analytical performance characteristics of the Yumizen C1200 Immunoglobulin A, G, and M reagents, intended for quantitative in vitro diagnostic determination of immunoglobulins in serum and lithium heparin plasma. The study aims to demonstrate that the device meets acceptance criteria for various analytical parameters, ensuring its safety and effectiveness.
1. Table of Acceptance Criteria and Reported Device Performance
Feature/Metric | Acceptance Criteria | Yumizen C1200 Immunoglobulin A Performance | Yumizen C1200 Immunoglobulin G Performance | Yumizen C1200 Immunoglobulin M Performance |
---|---|---|---|---|
Measuring Range | Appropriateness supported by LOD, LOQ, and linearity studies. | Serum: 0.10 to 7.00 g/L | ||
(Post-dilution): 7.00 to 21.00 g/L | ||||
(Linearity Range): 0.21 - 6.60 g/L (Slope 1.027, R^2 0.9975) | Serum: 0.75 to 30.00 g/L | |||
(Post-dilution): 30.00 to 90.00 g/L | ||||
(Linearity Range): 0.82 – 29.42 g/L (Slope 0.9965, R^2 0.9986) | Serum: 0.20 to 5.00 g/L | |||
(Post-dilution): 5.00 to 15.00 g/L | ||||
(Linearity Range): 0.26 - 4.16 g/L (Slope 1.013, R^2 0.9994) | ||||
Precision | Within-Run CV: Low (≤4.5%), Middle (≤3.8%), High (≤3%) | |||
Total Precision CV: Low (≤6.0%), Middle (≤5.0%), High (≤4.0%) | Total Precision (Analyzer Variability): | |||
Control Level 1: 3.7% | ||||
Control Level 2: 3.1% | ||||
Samples: 1.7%-3.8% | ||||
Lot-to-Lot Variability: | ||||
Control Level 1: 1.2% | ||||
Control Level 2: 1.1% | ||||
Samples: 1.2%-4.1% | Total Precision (Analyzer Variability): | |||
Control Level 1: 2.9% | ||||
Control Level 2: 3.3% | ||||
Samples: 1.8%-3.0% | ||||
Lot-to-Lot Variability: | ||||
Control Level 1: 1.7% | ||||
Control Level 2: 1.9% | ||||
Samples: 1.4%-2.3% | Total Precision (Analyzer Variability): | |||
Control Level 1: 2.1% | ||||
Control Level 2: 1.8% | ||||
Samples: 1.3%-2.4% | ||||
Lot-to-Lot Variability: | ||||
Control Level 1: 1.6% | ||||
Control Level 2: 1.7% | ||||
Samples: 1.0%-2.7% | ||||
Interferences | Acceptable bias: +/-10% of value without interfering substances | Reported highest values for which no interferences >10% were observed for various substances (Hemoglobin, Triglycerides, Bilirubin, Ascorbic Acid, Acetylsalicylic Acid, Ibuprofen, Acetaminophen). | Reported highest values for which no interferences >10% were observed for various substances (Hemoglobin, Triglycerides, Bilirubin, Ascorbic Acid, Acetylsalicylic Acid, Ibuprofen, Acetaminophen). | Reported highest values for which no interferences >10% were observed for various substances (Hemoglobin, Triglycerides, Bilirubin, Ascorbic Acid, Acetylsalicylic Acid, Ibuprofen, Acetaminophen). |
Matrix Comparison | No significant difference between serum and heparinized plasma specimens (implied by correlation and slope close to 1). | IgA: | ||
N=62 (paired serum/heparin plasma) | ||||
Slope: 1.000 | ||||
Correlation: 0.999 | IgG: | |||
N=43 (paired serum/heparin plasma) | ||||
Slope: 0.9929 | ||||
Correlation: 0.988 | IgM: | |||
N=43 (paired serum/heparin plasma) | ||||
Slope: 1.000 | ||||
Correlation: 0.999 | ||||
Method Comparison | Demonstrated substantial equivalence through correlation with predicate device. | IgA: | ||
N=190 (native serum samples) | ||||
Slope: 0.9941 | ||||
Correlation: 0.993 | IgG: | |||
N=214 (native serum samples) | ||||
Slope: 1.016 | ||||
Correlation: 0.993 | IgM: | |||
N=153 (native serum samples) | ||||
Slope: 1.005 | ||||
Correlation: 0.993 | ||||
Reagent Stability | Shelf life and on-board stability for opened reagents. | Closed: 24 months at 2-8°C | ||
On-Board (Opened): 6 weeks | Closed: 24 months at 2-8°C | |||
On-Board (Opened): 6 weeks | Closed: 24 months at 2-8°C | |||
On-Board (Opened): 6 weeks | ||||
Reference Range | Verification studies support established ranges through literature. | 0.70 - 4.00 g/L (70 - 400 mg/dL) | 7 – 16 g/L (700 - 1600 mg/dL) | 0.40 - 2.30 g/L (40 - 230 mg/dL) |
2. Sample Sizes Used for the Test Set and Data Provenance
- Measuring Range (Linearity):
- IgA, IgG, IgM: Samples were "spiked" to create different concentrations, then serially diluted. The exact number of initial samples for spiking is not specified, but dilutions were "assayed in quadruplicate within a single run."
- Precision (Total Precision: analyzer variability - 20x2x2 study):
- IgA, IgG, IgM: 5 human sera samples and 2 levels of Yumizen C1200 Protein Control. Tested with "two replicates per run, two runs per day for 20 days on each of three analyzers" (n=240 per sample).
- Precision (Lot to Lot variability study: 3x5x2x3):
- IgA, IgG, IgM: 5 human sera samples and 2 levels of Yumizen C1200 Protein Control. Tested in "triplicates per run, two runs per day for five days on each of three lots" (n=90 per sample).
- Interferences: The exact sample size is not stated, but the study implies testing samples with varying concentrations of interfering substances to determine acceptable bias.
- Matrix Comparison:
- IgA: 62 paired samples (serum and heparinized plasma) from single donors.
- IgG: 45 paired samples (serum and heparinized plasma) from single donors. Of these, 43 were used for correlation analysis.
- IgM: 43 paired samples (serum and heparinized plasma) from single donors.
- Method Comparison:
- IgA: 190 "native samples" from human serum.
- IgG: 214 "native samples" from human serum.
- IgM: 153 "native samples" from human serum.
Data Provenance: The human serum samples used for precision, matrix comparison, and method comparison studies were "anonymous remnants of human serum specimens collected from blood bank." "Spiked" samples were used for linearity studies, and "normal samples" from a blood bank were used for reference range verification. The document does not explicitly state the country of origin, but the manufacturer is based in France. The studies appear to be prospective analytical performance evaluations.
3. Number of Experts Used to Establish the Ground Truth for the Test Set and Their Qualifications
No external experts are mentioned for establishing ground truth in these analytical performance studies. The studies rely on established CLSI guidelines for evaluation and comparison with a legally marketed predicate device. The ground truth for quantitative measurements is the direct measurement by the devices themselves and comparison against predicate devices or known spiked concentrations.
4. Adjudication Method for the Test Set
Not applicable. These are analytical performance studies for an in vitro diagnostic device, not studies involving human interpretation or adjudication.
5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study
No. This document describes the analytical performance of an in vitro diagnostic reagent and analyzer system, not a device requiring human interpretation for diagnostic purposes where MRMC studies would typically be conducted.
6. Standalone (Algorithm Only Without Human-in-the-Loop Performance) Study
Yes, this can be considered a standalone performance study. The Yumizen C1200 system (reagents and analyzer) performs the quantitative determination of immunoglobulins. The studies evaluate the analytical capabilities of the system itself, such as accuracy (via method comparison, linearity), precision, interference, and stability, without direct human intervention in the result generation or diagnostic interpretation loop. The intent is to demonstrate the device's ability to accurately measure predefined analytes.
7. Type of Ground Truth Used
The ground truth for these analytical performance studies is established through:
- Known concentrations: For linearity studies, spiked samples with known concentrations were used.
- Predicate device measurements: For method comparison, results from the candidate device were compared against measurements obtained from legally marketed predicate devices (Beckman Coulter's Olympus IgA, IgG, IgM reagents on AU analyzers).
- CLSI guidelines and established methodologies: The studies adhere to CLSI guidelines (e.g., EP05-A3 for precision, EP17-A2 for detection capability, EP06-A for linearity, EP25-A for stability, C28-A3 for reference intervals, EP-9A3 for method comparison) which define accepted methods for evaluating analytical performance and establishing performance characteristics.
- Literature-established reference ranges: For reference range verification, the device's measurements on "normal" samples were compared against ranges cited in scientific literature (e.g., Dati et al., 1996).
8. Sample Size for the Training Set
The document does not explicitly delineate a "training set" in the context of machine learning or AI models. This device is an in vitro diagnostic reagent and analyzer system, not an AI/ML-based diagnostic algorithm that typically undergoes a distinct training phase with a dedicated dataset. The performance studies described here are for analytical validation rather than algorithm training.
9. How the Ground Truth for the Training Set Was Established
Not applicable, as there is no mention of a "training set" in the context of an AI/ML model for this in vitro diagnostic device. The analytical evaluations described involve testing the reagent and instrument system, not training a learning algorithm.
Ask a specific question about this device
(29 days)
York 10591
Re: K193493
Trade/Device Name: ADVIA Centaur Total IgE (tIgE) Regulation Number: 21 CFR 866.5510
| DGC |
| Regulation Number | 21 CFR 866.5510
For in vitro diagnostic use in the quantitative determination of total IgE in serum and plasma (EDTA and lithium heparin) using the ADVIA Centaur®, ADVIA Centaur XP, and ADVIA Centaur XPT systems.
The ADVIA Centaur Total IgE (tlgE) assay is a two-site sandwich immunoassay using direct chemiluminometric technology, which uses constant amounts of two antibodies to IqE. Results are determined using a calibration curve that is generated specifically on each instrument by a 2-point calibration and a master curve with the reagent bar code. The ADVIA Centaur Total IgE (tlgE) assay is intended for use on the ADVIA Centaur family of analyzers. The ADVIA Centaur Calibrator 80 is a set of 2 level calibrators for the assay. Siemens Healthcare Diagnostics recommends the use of commercially available quality control materials with at least 2 levels (low and high).
Here's an analysis of the acceptance criteria and the study that proves the device meets them, based on the provided FDA 510(k) summary:
Device: ADVIA Centaur Total IgE (tIgE) Assay
Purpose of Submission: Addition of plasma (EDTA and lithium heparin) sample claim and updating the detection capability claim.
1. Table of Acceptance Criteria and Reported Device Performance
Acceptance Criteria Category | Specific Acceptance Criteria | Reported Device Performance |
---|---|---|
Detection Capability | LoB: 1.5 IU/mL | LoB: 1.5 IU/mL |
LoD: 2.0 IU/mL | LoD: 2.0 IU/mL | |
LoQ: 2.5 IU/mL | LoQ: 2.5 IU/mL | |
Specimen Equivalence | The assay is designed to have a slope of 0.90–1.10 for alternate tube types versus serum. | Dipotassium EDTA Plasma vs. Serum: 0.99 (95% CI: 0.975 - 1.012) |
Lithium Heparin vs. Serum: 1.00 (95% CI: 0.989 - 1.020) | ||
Interference | (Implicit, likely a pre-defined acceptable bias percentage for reported interferents) | Dipotassium EDTA (9.0 mg/mL): Bias -1.7% (at 121.51 IU/mL), Bias 1.4% (at 1624.13 IU/mL) |
Heparin (75 U/mL): Bias -1.7% (at 167.48 IU/mL), Bias -1.1% (at 1450.12 IU/mL) |
2. Sample Size Used for the Test Set and Data Provenance
-
Specimen Equivalence by Method Comparison:
- Sample Size: N = 73 for both Dipotassium EDTA Plasma vs. Serum and Lithium Heparin vs. Serum.
- Data Provenance: Not explicitly stated (e.g., country of origin, retrospective/prospective). However, given it's for an in vitro diagnostic device and involves method comparison, it typically involves collected patient samples.
-
Detection Capability (LoB, LoD, LoQ) and Interferences:
- Sample Size: Not explicitly stated how many individual samples were used to determine LoB, LoD, LoQ, or for interference testing. These typically involve a series of measurements on spiked or known concentration samples rather than a large set of patient samples like method comparison.
- Data Provenance: Not explicitly stated.
3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications
This submission document primarily focuses on analytical performance of an in vitro diagnostic (IVD) assay and does not involve human expert interpretation of results to establish ground truth in the way a medical imaging AI would. Therefore, this section is not applicable in the traditional sense for this type of device. The "ground truth" for this device's performance is established by the known concentrations of analytes in controls, calibrators, and spiked samples, and comparison to a reference method (in the case of method comparison studies).
4. Adjudication Method for the Test Set
Not applicable. As noted above, this is an analytical performance study for an IVD assay, not a study involving human reader interpretation requiring adjudication.
5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done
No. A MRMC study is typically performed for AI-assisted diagnostic imaging devices to evaluate the impact of AI on human reader performance. This submission is for an in vitro diagnostic assay, which does not involve human interpretation of images in this context.
6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done
This device is a standalone device in terms of its operation. It's an automated immunoassay system (ADVIA Centaur family of analyzers) that quantitatively determines total IgE. Its performance characteristics (detection capability, specimen equivalence, interference) are evaluated as the algorithm/system running independently on samples. There isn't a human-in-the-loop interacting with the direct measurement as there would be with an AI assisting image interpretation.
7. The Type of Ground Truth Used
- Detection Capability (LoB, LoD, LoQ): Established using statistical methods defined by CLSI EP17-A2, relying on repeated measurements of blank samples and samples with known low concentrations of the analyte.
- Specimen Equivalence by Method Comparison: The "ground truth" for the comparison (serum) is considered the established method against which the alternative sample types (EDTA plasma, lithium heparin plasma) are being evaluated. This relies on the accuracy of the serum measurement itself.
- Interferences: Established by spiking known interfering substances at specified concentrations into samples with known IgE concentrations and measuring the resulting bias.
8. The Sample Size for the Training Set
Not applicable. This is an analytical performance study for a chemical assay. There is no "training set" in the machine learning sense for this device. The assay's parameters and calibration are established through laboratory procedures, calibrators, and master curves, not through a data-driven training process in the way an AI algorithm is trained.
9. How the Ground Truth for the Training Set Was Established
Not applicable. As there is no "training set" in the AI sense for this device, there is no ground truth established for it. The assay relies on known chemical reactions, calibrated reagents, and master curves set during the assay's development and manufacturing.
Ask a specific question about this device
(88 days)
30093
Re: K192095
Trade/Device Name: CAPI 3 Immunotyping, Capillarys 3 Tera Regulation Number: 21 CFR 866.5510
| 21 CFR Part 866.5510, 866.5550, 866.1630,
The CAPI 3 IMMUNOTYPING kit is designed for the qualitative detection and the characterization of monoclonal proteins (immunotyping) in human urine and serum with the CAPILLARYS 3 TERA instrument, SEBIA, for capillary electrophoresis. It is used in conjunction with the CAPI 3 PROTEIN(E) 6 kit, SEBIA, designed for proteins separation into 6 major fractions in alkaline buffer (pH 9.9).
The CAPILLARYS 3 TERA instrument performs all procedural sequences automatically to obtain a profile for qualitative analysis. Each urine or serum sample is mixed with individual antisera that are specific against gamma (Ig G). alpha (Ig A) and mu (Ig M) heavy chains, and kappa (free and bound) light chains and lambda (free and bound) light chains, respectively.
The proteins, separated in silica capillaries, are directly detected by their absorbance at 200 nm.
The electrophoregrams are evaluated visually to detect the presence of specific reactions with the suspect monoclonal proteins.
For In Vitro Diagnostic Use.
The capillary electrophoresis provides complete automation with fast separation and good resolution. This electrokinetic separation technique is carried out in a silica glass tube (i.e., capillary) with internal diameter lower than 100 um filled with a buffer composed of electrolytes.
The CAPILLARYS 3 TERA instrument uses the principle of capillary electrophoresis in free solution. With this technique, charged molecules are separated by their electrophoretic mobility in an alkaline buffer with a specific pH. Separation occurs according to the electrolyte pH and electroosmotic flow. The CAPILLARYS 3 TERA instrument has silica capillaries functioning in parallel allowing 12 simultaneous analyses.
In capillary electrophoresis, abnormal fractions detected in serum or urine protein electrophoregrams, primarily those in the beta globulin and gamma globulin zones, are always suspect of being monoclonal proteins (M-proteins, paraproteins, monoclonal immunoglobulins). With the CAPI 3 IMMUNOTYPING procedure, the immunotyping procedure uses specific antibodies to identify these abnormal fractions.
In immunotyping a sample dilution is prepared and injected at the anodic end of six capillaries. The reference pattern (ELP pattern), which is a complete electrophoretic pattern of the sample's proteins, is obtained by mixing the sample with the ELP solution and injection into the 1st capillary. The antisera patterns are obtained by sample aspiration into the 5 subsequent capillaries. Previously diluted samples are mixed with specific antisera against gamma (Ig G), alpha (lg A), mu (lg M) heavy chains, and free and bound Kappa and Lambda light chains. Protein separation is performed in a high voltage electrical field. The separated proteins are detected using absorbance at 200 nm at the cathodic end of the capillary. After the analysis, the capillaries are immediately washed with a wash solution and filled with buffer which prepares the capillaries for the next analysis.
The immunotyping is performed in four automated steps:
- Dilution of serum or urine samples with a specific diluent in the pre-dilution well of the reagent cup. This dilution is made according to the sample's immunoglobulin concentration.
- Mixing diluted serum sample with specific antisera. The antigen - antibody complex is formed rapidly in liquid medium without the need for extra incubation step or removal of the immune complexes.
- Injection of the prepared samples with simultaneous aspiration into 6 capillaries at the anodic end. Protein separation occurs when a high voltage field is applied to the alkaline buffer. The separated proteins are detected using absorbance at 200 nm at the cathodic end of the capillary.
- Overlay of the ELP pattern on the antisera patterns (Ig G, Ig A, Ig M, Kappa and Lambda) allows characterization of suspected monoclonal component.
Here's an analysis of the acceptance criteria and study data for the CAPI 3 IMMUNOTYPING device, based on the provided FDA 510(k) summary:
1. Table of Acceptance Criteria and Reported Device Performance
The FDA 510(k) summary does not explicitly list predefined quantitative acceptance criteria with specific numerical thresholds for all performance metrics (e.g., repeatability, reproducibility, accuracy). Instead, for qualitative assessments, the acceptance criterion appears to be "concordant results" or "100% agreement." For sensitivity, the acceptance criterion is implicitly shown by the reported detection limits.
Performance Metric | Acceptance Criteria (Implicit from Document) | Reported Device Performance |
---|---|---|
Repeatability (Within-run and Between-capillaries) | All samples should give concordant results within run and between capillaries. | For each tested reagent (ELP, Anti-IgG, Anti-IgA, Anti-IgM, Anti-Kappa, Anti-Lambda), all 4 urine samples (including Bence Jones proteins) gave concordant results within run and between capillaries. All dilution programs were tested. |
Reproducibility (Between lots and instruments) | All samples should give concordant results across different runs, instruments, and reagent lots. | All 3 urine samples (including Bence Jones proteins) gave concordant results for all 18 runs (over 5 working days, 2 times/day) on the 3 CAPILLARYS 3 TERA instruments and with the 3 lots of CAPI 3 IMMUNOTYPING kit. All dilution programs were tested. |
Sensitivity (Detection Limit) | Monoclonal proteins in urine should be detectable at clinically relevant low concentrations. (No explicit numerical acceptance criterion is stated, but the ability to detect at specified g/L or mg/dL is the measure). | Detection Limits Reported: |
- Lambda free: 0.010 g/L (1.0 mg/dL)
- Kappa free: 0.030 g/L (3.0 mg/dL)
- IgG Lambda: 0.004 g/L (0.4 mg/dL)
- Lambda (part of IgG Lambda): 0.004 g/L (0.4 mg/dL) |
| Sample Stability (2-8 °C) | Results after 1 week storage at 2-8 °C should comply with "acceptance criteria defined by SEBIA." (Specific criteria not detailed in document). | The results obtained complied with the acceptance criteria defined by SEBIA. Conclusion: Urine samples can be stored for 1 week between 2 and 8 °C. |
| Sample Stability (-70/-80 °C) | Results after 1 month storage at -70/-80 °C should comply with "acceptance criteria defined by SEBIA." (Specific criteria not detailed in document). | The results obtained complied with the acceptance criteria defined by SEBIA. Conclusion: Urine samples can be stored for 1 month between -70 and -80 ℃. |
| Kit Stability (CAPI 3 IMMUNOTYPING sample diluent) | Stable for 2 years at 2 - 30 °C. (The acceptance criterion is the claimed stability period and conditions). | 2 years at 2 - 30 °C (Reported as claimed stability). |
| Kit Stability (CAPI 3 IMMUNOTYPING ELP solution) | Stable for 2 years at 2 - 8 °C. (The acceptance criterion is the claimed stability period and conditions). | 2 years at 2 - 8 °C (Reported as claimed stability). |
| Kit Stability (CAPI 3 IMMUNOTYPING antisera) | Stable for 2 years at 2 - 8 °C. (The acceptance criterion is the claimed stability period and conditions). | 2 years at 2 - 8 °C (Reported as claimed stability). |
| On-Board Stability (Sample Diluent, ELP, Antisera) | Stable for 2 months on-board the instrument. (The acceptance criterion is the claimed stability period). | 2 months (Reported as claimed stability). |
| Accuracy/Concordance | 100% agreement between the test technique (CAPI 3 IMMUNOTYPING on CAPILLARYS 3 TERA) and the reference technique (CAPILLARYS IMMUNOTYPING on CAPILLARYS 2) for qualitative results, including presence/absence and type of monoclonal component. | This study demonstrated 100% agreement between the two techniques for all 52 urine samples (8 without, 44 with monoclonal components of various types). This 100% agreement was confirmed for specific monoclonal protein types (IgG Lambda, IgG Kappa, IgG Lambda with Lambda free, IgG Kappa with Kappa free, Lambda free, Kappa free, IgM Kappa with Kappa free, IgA Lambda, IgA Lambda with Lambda free, and without Monoclonal). |
2. Sample Size Used for the Test Set and Data Provenance
- Repeatability: 4 different urine samples with monoclonal proteins (including Bence Jones proteins).
- Reproducibility: 3 different urine samples with monoclonal proteins (including Bence Jones proteins).
- Sensitivity: 3 pathological urine samples, serially diluted in normal urine.
- Sample Stability (2-8 °C): 10 urine samples (normal and pathological).
- Sample Stability (-70/-80 °C): 10 urine samples (normal and pathological).
- Accuracy/Concordance: 52 urine samples (8 without monoclonal component, 44 with monoclonal component including Bence Jones proteins).
Data Provenance: The document does not explicitly state the country of origin of the data or whether it was retrospective or prospective. Given that it's a 510(k) submission from Sebia (manufacturing in France, submitter in USA), and refers to "urine samples," these are likely clinical samples, but details on their origin are not provided. The phrase "analyzed at the beginning of the study (reference) and after X storage (test)" suggests a prospective element for the stability studies on stored samples.
3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications
The document states: "All electrophoregrams were evaluated visually for the qualitative results." However, it does not specify the number of experts used for ground truth establishment or their qualifications (e.g., "radiologist with 10 years of experience"). This is a significant omission from the perspective of external validation of the ground truth. It is implied that visual evaluation by trained personnel (likely laboratory professionals or experts in electrophoresis interpretation) was the method, but no further details are provided.
4. Adjudication Method for the Test Set
The document does not detail an adjudication method (such as 2+1 or 3+1 consensus) for the visual evaluation of electrophoregrams. It simply states they were "evaluated visually." This implies a single evaluator, or an internal process where disagreement (if any) was resolved, but the process is not described.
5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study
No multi-reader multi-case (MRMC) comparative effectiveness study was described. The accuracy/concordance study compares the new device's visual interpretation results to an existing, cleared device's visual interpretation results, which is a comparison of two methods, not an assessment of human readers with and without AI assistance. The device itself is a qualitative detection and characterization system, and the evaluation stated is "electrophoregrams were evaluated visually," indicating human interpretation is still involved in the final result.
6. Standalone Performance Study
The studies described (repeatability, reproducibility, sensitivity, stability, accuracy/concordance) evaluate the performance of the integrated system (CAPI 3 IMMUNOTYPING kit on CAPILLARYS 3 TERA instrument). The "electrophoregrams are evaluated visually" suggests human-in-the-loop performance, rather than a purely standalone algorithm. The device produces a "protein profile for qualitative analysis," which is then visually interpreted. Therefore, a standalone (algorithm only) performance study, in the sense of an AI algorithm making a definitive diagnosis without human oversight, does not appear to have been performed or is not directly applicable to the described use case which involves visual evaluation.
7. Type of Ground Truth Used
The ground truth for the test sets (e.g., for accuracy, repeatability, reproducibility studies) appears to be established by the visual evaluation of electrophoregrams performed by an expert or experts using a "reference technique" or the device itself.
- For the accuracy/concordance study, the "reference technique" (CAPILLARYS IMMUNOTYPING URINE procedure performed with the CAPILLARYS 2 instrument) served as the comparator for establishing ground truth, assuming the predicate device's results are considered the established truth.
- For other studies like repeatability and reproducibility, the "ground truth" seems to implicitly be the correct identification of presence/absence and type of monoclonal protein based on the expected outcome for the known pathological samples, assessed by visual evaluation.
- No mention of pathology, outcomes data, or independent gold standard beyond another electrophoretic method is provided.
8. Sample Size for the Training Set
The document does not mention a "training set" or explicitly describe the development of an algorithm that learns from data. This device is an immunodiagnostic test system based on capillary electrophoresis and specific antisera, producing profiles that are then visually interpreted. It is not presented as an AI/ML-driven device requiring a training set in the conventional sense. The "training" for such a system would typically involve its chemical and mechanical design, calibration, and validation against a standard.
9. How the Ground Truth for the Training Set Was Established
As no training set is described for an AI/ML algorithm, this question is not applicable. The assay's "truth" is inherent in its biochemical principle (antigen-antibody reaction, electrophoretic separation) and the expertise of interpreting the resulting electrophoregrams.
Ask a specific question about this device
(29 days)
Trade/Device Name: Human IgA liquid reagent kit for use on SPAPLUS Regulation Number: 21 CFR 866.5510
Regulation section:
21 CFR 866.5510, Immunoglobulins A, G, M, D, and E immunological test system
- 2
This kit is intended for the quantitative in vitro determination of human IgA in serum, lithium heparin or EDTA plasma, using the Binding Site SPAPLUS turbidimetric analyser. Measurement of IgA aids in the diagnosis of abnormal protein metabolism and the body's lack of ability to resist infectious agents. The test results are to be used in conjunction with other clinical and laboratory findings.
Human IqA liquid reagent kit for use on SPAPLUS® comprises the following reagents:
Antiserum: Monospecific goat anti IgA supplied in stabilised liquid form. lt contains 0.099% sodium azide, 0.1% E-amino-n-caproic acid (EACA), 0.5% BSA and 0.01% benzamidine as preservatives.
Calibrator and Controls: These consist of pooled human serum and are supplied in stabilised liquid form. The concentration of IgA given on the quality control certificate has been obtained by comparison with European Reference Material ERM-DA470k. They contain 0.099% sodium azide, 0.1% EACA and 0.01% benzamidine as preservatives.
Reaction Buffer: Containing 0.099% sodium azide as a preservative.
The provided document describes the "Human IgA liquid reagent kit for use on SPAPLUS" by The Binding Site Group Ltd. This device is intended for the quantitative in vitro determination of human IgA in serum, lithium heparin, or EDTA plasma using the Binding Site SPAPLUS turbidimetric analyzer. The submission is a special 510(k) for a modification to an existing device (K103824), specifically a change in the source of the detection antibody from sheep to goat.
Here's an analysis of the acceptance criteria and the study that proves the device meets them:
1. A table of acceptance criteria and the reported device performance
The document does not explicitly present acceptance criteria in a dedicated table format with corresponding reported device performance for all aspects. Instead, acceptance criteria are described within the context of each study and then a conclusion about meeting those criteria is stated. I will extract and present this information in a table format where possible.
Test Parameter / Characteristic | Acceptance Criteria (Implicit/Explicit) | Reported Device Performance | Conclusion on Meeting Criteria |
---|---|---|---|
Precision (Repeatability & Within Lab) | No explicit numerical acceptance criteria given, but the expectation is "no change in performance compared to the device cleared in K103824". | CV% ranges from 1.7% to 7.1%. | Do not indicate any change in performance compared to K103824. |
Precision (Between Instrument) | No explicit numerical acceptance criteria given, but the expectation is "no change in performance compared to the device cleared in K103824". | CV% ranges from 0.0% to 1.8%. | Do not indicate any change in performance compared to K103824. |
Precision (Between Lot) | No explicit numerical acceptance criteria given, but the expectation is "no change in performance compared to the device cleared in K103824". | CV% ranges from 0.0% to 3.2%. | Do not indicate any change in performance compared to K103824. |
Linearity/Assay Reportable Range | No explicit numerical acceptance criteria given, but the expectation is "comparable to those currently presented in the product insert" and "no change in performance compared to K103824". | Linear regression equation: y=0.993x - 0.230 g/L with an R value of 0.996. | Comparable to product insert, do not indicate any change in performance compared to K103824. |
Accelerated Kit Stability | Maximum allowable difference of ±15% to verify the stability claim of 18 months. | All controls, internal reference, and samples for two batches (455406 and 455407) passed, achieving an equivalent stability of 561 days at 4ºC, which exceeds the required 395 days (18 months). | Pass for all tested parameters and batches. |
On-Board Stability | No explicit numerical acceptance criteria given, but the expectation is "no difference in the cleared on-board stability claim". | Studies showed no difference in the cleared on-board stability claim. | No difference observed. |
Limit of Quantitation (LoQ) | Allowable CV of 8%. | LoQ claim was validated by all samples reporting within the acceptance criteria. | LoQ claim validated. |
Limit of Detection (LoD) & Limit of Blank (LoB) | No explicit numerical acceptance criteria given, but the expectation is "no change in performance compared to the device cleared in K103824". | LoD estimated as 0.003 g/L, LoB as 0.001 g/L. No change in performance observed after antisera change. | Do not indicate any change in performance compared to K103824. |
Method Comparison with Predicate Device (Bland Altman) | No explicit numerical acceptance criteria given, but the expectation is "no change in performance compared to K103824". | Mean Bias: -2.18%, 95% Limits of Agreement: 0.55% to 3.17%. | Do not indicate any change in performance compared to K103824. |
Method Comparison with Predicate Device (Passing Bablok) | No explicit numerical acceptance criteria given, but the expectation is "no change in performance compared to K103824". | Equation: y=1.017x + 0.002, Slope 95% CI: 1.004 to 1.029, Intercept 95% CI: -0.029 to 0.026. Correlation coefficient: 0.998. | Do not indicate any change in performance compared to K103824. |
Expected Values/Reference Range Transfer | ≤2 samples falling outside of the limits of the reference interval to be transferred. | All 20 samples tested gave results within the reference interval (1.553 to 4.840 g/L). | Meets acceptance criteria, reference interval can be transferred. |
2. Sample sized used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)
- Precision Studies:
- Repeatability and Within Laboratory: 4 different samples, 80 data points for each level (total 320 data points).
- Between Instrument: 4 different samples, 24 data points for each level (total 96 data points).
- Between Lot: 4 different samples, 24 data points for each level (total 96 data points).
- Linearity Study: A high pool (8.19g/L) and a low pool (0.12 g/L) were used to create a dilution series. Each diluted sample was tested in 3 replicates.
- Accelerated Stability: 6 replicates of controls, internal reference, and 3 different samples were tested.
- Limit of Quantitation (LoQ) Validation: 4 samples were tested using two reagent lots.
- Method Comparison with Predicate Device: 102 serum samples and 42 plasma samples (total 144 samples).
- Expected Values/Reference Range Transfer: 20 samples from "apparently healthy US donors".
Data Provenance:
- For the Expected Values/Reference Range Transfer study, samples were from "apparently healthy US donors".
- For other studies, the country of origin or whether the data was retrospective or prospective is not explicitly stated. The studies were carried out by The Binding Site Group Ltd., which is based in the UK, so it's likely testing was conducted there unless otherwise specified.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)
This product is an in vitro diagnostic (IVD) reagent kit for quantitative Immunoglobulin A (IgA) determination. The "ground truth" in this context refers to the accurate measurement of IgA concentration. The document mentions traceability to European Reference Material ERM-DA470k/IFCC for calibration. This implies that the standard itself serves as the "ground truth" reference, rather than independent expert consensus on clinical interpretation. There are no mentions of experts establishing ground truth in the sense of clinical diagnosis or image interpretation.
4. Adjudication method (e.g. 2+1, 3+1, none) for the test set
Adjudication methods like 2+1 or 3+1 are typically used in studies involving subjective interpretation by multiple human readers (e.g., in radiology studies) to resolve discrepancies and establish a consensus "ground truth." Since this device is a quantitative IVD assay and its performance is validated against analytical standards, reference materials, and comparative measurements, such adjudication methods are not applicable and therefore not used.
5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
No, a multi-reader multi-case (MRMC) comparative effectiveness study was not done. This type of study is relevant for AI-powered diagnostic tools where human readers (e.g., radiologists) use AI assistance to improve their diagnostic accuracy. This device is a standalone quantitative laboratory test kit, not an AI or imaging diagnostic aid, so an MRMC study is not applicable.
6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done
Yes, the performance studies described are for the standalone device (reagent kit on the SPAPLUS analyzer) without human-in-the-loop performance influencing the measurement part of the device's function. The analytical performance, stability, and comparison studies evaluate the kit's ability to accurately measure IgA concentrations. The device's output is a quantitative value, not an interpretation of data that requires human feedback for performance assessment.
7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)
The ground truth for this quantitative assay is primarily based on:
- Reference Materials: Calibration traceability to ERM-DA470k/IFCC. This is the international standard for immunoglobulin measurements.
- Reference Methods: The predicate device itself (K103824) serves as a comparative "ground truth" for method comparison studies, demonstrating substantial equivalence.
- Defined Concentrations: For linearity and precision studies, samples with known or precisely diluted concentrations are used.
- Clinical Reference Intervals: The ability to transfer established reference intervals (based on a healthy population) is also part of the "ground truth" for clinical usability.
8. The sample size for the training set
This document describes a special 510(k) submission for a modification to an existing device, which mostly involves re-validation studies to ensure the modification (change in antibody source from sheep to goat) does not alter performance compared to the cleared predicate. It does not describe the development of a novel algorithm or AI model that requires a distinct "training set." The studies performed are validation studies, not training. Therefore, a "training set" in the context of machine learning is not applicable here.
9. How the ground truth for the training set was established
As there is no "training set" in the machine learning sense for this device submission, this question is not applicable. The device's underlying principles (immunoturbidimetry) are well-established, and its performance is validated against analytical standards rather than learned from a dataset.
Ask a specific question about this device
Page 1 of 11