(29 days)
For in vitro diagnostic use in the quantitative determination of total IgE in serum and plasma (EDTA and lithium heparin) using the ADVIA Centaur®, ADVIA Centaur XP, and ADVIA Centaur XPT systems.
The ADVIA Centaur Total IgE (tlgE) assay is a two-site sandwich immunoassay using direct chemiluminometric technology, which uses constant amounts of two antibodies to IqE. Results are determined using a calibration curve that is generated specifically on each instrument by a 2-point calibration and a master curve with the reagent bar code. The ADVIA Centaur Total IgE (tlgE) assay is intended for use on the ADVIA Centaur family of analyzers. The ADVIA Centaur Calibrator 80 is a set of 2 level calibrators for the assay. Siemens Healthcare Diagnostics recommends the use of commercially available quality control materials with at least 2 levels (low and high).
Here's an analysis of the acceptance criteria and the study that proves the device meets them, based on the provided FDA 510(k) summary:
Device: ADVIA Centaur Total IgE (tIgE) Assay
Purpose of Submission: Addition of plasma (EDTA and lithium heparin) sample claim and updating the detection capability claim.
1. Table of Acceptance Criteria and Reported Device Performance
| Acceptance Criteria Category | Specific Acceptance Criteria | Reported Device Performance |
|---|---|---|
| Detection Capability | LoB: 1.5 IU/mL | LoB: 1.5 IU/mL |
| LoD: 2.0 IU/mL | LoD: 2.0 IU/mL | |
| LoQ: 2.5 IU/mL | LoQ: 2.5 IU/mL | |
| Specimen Equivalence | The assay is designed to have a slope of 0.90–1.10 for alternate tube types versus serum. | Dipotassium EDTA Plasma vs. Serum: 0.99 (95% CI: 0.975 - 1.012) |
| Lithium Heparin vs. Serum: 1.00 (95% CI: 0.989 - 1.020) | ||
| Interference | (Implicit, likely a pre-defined acceptable bias percentage for reported interferents) | Dipotassium EDTA (9.0 mg/mL): Bias -1.7% (at 121.51 IU/mL), Bias 1.4% (at 1624.13 IU/mL) |
| Heparin (75 U/mL): Bias -1.7% (at 167.48 IU/mL), Bias -1.1% (at 1450.12 IU/mL) |
2. Sample Size Used for the Test Set and Data Provenance
-
Specimen Equivalence by Method Comparison:
- Sample Size: N = 73 for both Dipotassium EDTA Plasma vs. Serum and Lithium Heparin vs. Serum.
- Data Provenance: Not explicitly stated (e.g., country of origin, retrospective/prospective). However, given it's for an in vitro diagnostic device and involves method comparison, it typically involves collected patient samples.
-
Detection Capability (LoB, LoD, LoQ) and Interferences:
- Sample Size: Not explicitly stated how many individual samples were used to determine LoB, LoD, LoQ, or for interference testing. These typically involve a series of measurements on spiked or known concentration samples rather than a large set of patient samples like method comparison.
- Data Provenance: Not explicitly stated.
3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications
This submission document primarily focuses on analytical performance of an in vitro diagnostic (IVD) assay and does not involve human expert interpretation of results to establish ground truth in the way a medical imaging AI would. Therefore, this section is not applicable in the traditional sense for this type of device. The "ground truth" for this device's performance is established by the known concentrations of analytes in controls, calibrators, and spiked samples, and comparison to a reference method (in the case of method comparison studies).
4. Adjudication Method for the Test Set
Not applicable. As noted above, this is an analytical performance study for an IVD assay, not a study involving human reader interpretation requiring adjudication.
5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done
No. A MRMC study is typically performed for AI-assisted diagnostic imaging devices to evaluate the impact of AI on human reader performance. This submission is for an in vitro diagnostic assay, which does not involve human interpretation of images in this context.
6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done
This device is a standalone device in terms of its operation. It's an automated immunoassay system (ADVIA Centaur family of analyzers) that quantitatively determines total IgE. Its performance characteristics (detection capability, specimen equivalence, interference) are evaluated as the algorithm/system running independently on samples. There isn't a human-in-the-loop interacting with the direct measurement as there would be with an AI assisting image interpretation.
7. The Type of Ground Truth Used
- Detection Capability (LoB, LoD, LoQ): Established using statistical methods defined by CLSI EP17-A2, relying on repeated measurements of blank samples and samples with known low concentrations of the analyte.
- Specimen Equivalence by Method Comparison: The "ground truth" for the comparison (serum) is considered the established method against which the alternative sample types (EDTA plasma, lithium heparin plasma) are being evaluated. This relies on the accuracy of the serum measurement itself.
- Interferences: Established by spiking known interfering substances at specified concentrations into samples with known IgE concentrations and measuring the resulting bias.
8. The Sample Size for the Training Set
Not applicable. This is an analytical performance study for a chemical assay. There is no "training set" in the machine learning sense for this device. The assay's parameters and calibration are established through laboratory procedures, calibrators, and master curves, not through a data-driven training process in the way an AI algorithm is trained.
9. How the Ground Truth for the Training Set Was Established
Not applicable. As there is no "training set" in the AI sense for this device, there is no ground truth established for it. The assay relies on known chemical reactions, calibrated reagents, and master curves set during the assay's development and manufacturing.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: a symbol on the left and the FDA name on the right. The symbol on the left is a stylized image of a human figure, while the FDA name on the right is written in blue letters. The words "U.S. FOOD & DRUG ADMINISTRATION" are written in a clear, sans-serif font.
January 15, 2020
Siemens Healthcare Diagnostics Inc. Ian Thompson Regulatory Clinical Affairs Specialist 511 Benedict Avenue Tarrytown, New York 10591
Re: K193493
Trade/Device Name: ADVIA Centaur Total IgE (tIgE) Regulation Number: 21 CFR 866.5510 Regulation Name: Immunoglobulins A, G, M, D, and E immunological test system Regulatory Class: Class II Product Code: DGC Dated: December 16, 2019 Received: December 17, 2019
Dear Ian Thompson:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's
{1}------------------------------------------------
requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801 and Part 809); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Carolina Kagan Acting Chief Division of Immunology and Hematology Devices OHT7: Office of In Vitro Diagnostics and Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
Indications for Use
510(k) Number (if known) K193493
Device Name ADVIA Centaur® Total IgE (tIgE)
Indications for Use (Describe)
For in vitro diagnostic use in the quantitative determination of total IgE in serum and plasma (EDTA and lithium heparin) using the ADVIA Centaur®, ADVIA Centaur XP, and ADVIA Centaur XPT systems.
Type of Use (Select one or both, as applicable)
X Prescription Use (Part 21 CFR 801 Subpart D)
| Over-The-Counter Use (21 CFR 801 Subpart C)
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
This 510(k) Summary of Safety and Effectiveness is being submitted in accordance with the requirements of 21 CFR 807.92 and the Safe Medical Device Act of 1990.
The assigned 510(k) Number is: K193493
1. Date Prepared
December 13, 2019
2. Applicant Information
| Contact: | Ian ThompsonRegulatory Clinical Affairs Specialist |
|---|---|
| Address: | Siemens Healthcare Diagnostics Inc.511 Benedict AvenueTarrytown, NY 10591 |
| Phone: | 914-524-2296 |
| Fax: | 914-524-2101 |
| Email: | ian_thompson@siemens-healthineers.com |
Regulatory Information 3.
Table 1. Regulatory Information for ADVIA Centaur® Total IgE (tlgE)
| Trade Name | ADVIA Centaur Total IgE (tIgE) |
|---|---|
| Device | IgE, Antigen, Antiserum, Control |
| Regulation Description | Immunoglobulins A, G, M, D, and E immunological test system |
| FDA Classification | Class II |
| Review Panel | Immunology |
| Product Code | DGC |
| Regulation Number | 21 CFR 866.5510 |
4. Predicate Device Information
Predicate Device Name: ADVIA Centaur® Total IgE (tIgE)
510(k) Number: K920372
The ADVIA Centaur Total IgE (tlgE) assay with the addition of the plasma (EDTA and lithium heparin) sample and new detection capability [Limit of Blank (LoB), Limit of Detection (LoD), and Limit of Quantitation (LoQ)] claims in the Instructions for Use (Package Inserts) are substantially equivalent to the ADVIA Centaur Total IgE (tlgE) assay that was cleared under 510(k) K920372, as shown below in the Substantial Equivalence Information section.
5. Intended Use
For in vitro diagnostic use in the quantitative determination of total IgE in serum and plasma (EDTA and lithium heparin) using the ADVIA Centaur®, ADVIA Centaur XP, and ADVIA Centaur XPT systems.
6. Indications for Use
Same as Intended use.
{4}------------------------------------------------
Special Conditions for Use Statement(s): For prescription use only
Device Description 7.
The ADVIA Centaur Total IgE (tlgE) assay is a two-site sandwich immunoassay using direct chemiluminometric technology, which uses constant amounts of two antibodies to IqE. Results are determined using a calibration curve that is generated specifically on each instrument by a 2-point calibration and a master curve with the reagent bar code. The ADVIA Centaur Total IgE (tlgE) assay is intended for use on the ADVIA Centaur family of analyzers. The ADVIA Centaur Calibrator 80 is a set of 2 level calibrators for the assay. Siemens Healthcare Diagnostics recommends the use of commercially available quality control materials with at least 2 levels (low and high).
The ADVIA Centaur Total IgE (tlgE) reagent kit contains the following:
-
ADVIA Centaur Total IgE ReadyPack primary reagent pack contains Lite Reagent and Solid . Phase Reagent.
Materials Required but Not provided: -
ADVIA Centaur Calibrator 80: consists of 2 levels (low and high) of human IgE calibrators in ● equine serum and preservatives; lyophilized.
Optional Reagents: -
ADVIA Centaur IgE Diluent: consists of IgE-free human plasma with sodium azide (0.1%).
-
ADVIA Centaur IgE Master Curve Material: consists of MCM 1 that is Iyophilized human plasma ● with sodium azide (0.1% after reconstitution) and preservatives and MCM 2-7 that are various levels of IgE in Iyophilized human plasma with sodium azide (0.1% after reconstitution) and preservatives.
8. Purpose of the Submission
The purpose of this submission is for the addition of the plasma (EDTA and lithium heparin) sample claim and updating the detection capability claim for the ADVIA Centaur Total IgE (tlgE) assay.
Substantial Equivalence Information - Comparison of Candidate 9. Device and Predicate Device
The following table demonstrates substantial equivalence between the ADVIA Centaur Total IgE (tlgE) assay (Candidate Device) that has modified Instructions for Use (Package Inserts) with the addition of plasma (EDTA and lithium heparin) sample and new detection capability (LoB, LoD, and LoQ) claims and the currently marketed ADVIA Centaur Total IgE (tlgE) assay (Predicate Device) that was cleared under 510(k) K920372.
{5}------------------------------------------------
| Trade Name | Candidate Device | Predicate Device |
|---|---|---|
| ADVIA Centaur Total IgE (tIgE)(Modified Labeling) | ADVIA Centaur Total IgE (tIgE)(Unmodified Labeling) | |
| Intended Use /Indications for Use | For in vitro diagnostic use in thequantitative determination of total IgE inserum and plasma (EDTA and lithiumheparin) using the ADVIA Centaur®, ADVIACentaur XP, and ADVIA CentaurXPT systems. | For in vitro diagnostic use in thequantitative determination of total IgE inserum using the ADVIA Centaur®,ADVIA Centaur XP, and ADVIA CentaurXPT systems. |
| Measurement | Quantitative | Same |
| DetectionCapability | LoB: 1.5 IU/mLLoD: 2.0 IU/mLLoQ: 2.5 IU/mL | Analytical Sensitivity: 1.5 IU/mLNA*NA |
| Assay Range | Serum and plasma: 2.5-3000 IU/mL | Serum: 1.5-3000 IU/mL |
| Operating Principle | two-site sandwich immunoassay | Same |
| Technology | Direct chemiluminescent | Same |
| Sample Type | Serum, plasma (EDTA and lithium heparin) | Serum |
| Sample Volume | 30 µL (serum and plasma) | 30 µL (serum) |
| Traceability/Standardization | World Health Organization (WHO)75/502 | Same |
| Calibration | 2-point | Same |
| Calibrator/Levels | Calibrator 80/2 levels | Same |
| Controls/Levels | Commercial Controls/2 levels | Same |
| Master CurveMaterials | Seven levels (MCM1-7) | Same |
| Detection Antibody | Goat anti-human IgE antibody labeled withacridinium ester | Same |
| Capture Antibody | Mouse anti-human IgE antibody covalentlycoupled to paramagnetic particles | Same |
- NA = Not applicable.
Standard/Guidance Document References 10.
The following recognized standards from Clinical Laboratory Standards Institute (CLSI) were used as a basis of the study procedures described in this submission:
- Interference Testing in Clinical Chemistry; Approved Guideline-Third Edition. (CLSI EP07-● ed3).
- Evaluation of Detection Capability for Clinical Laboratory Measurement Procedures; Approved ● Guideline-Second Edition (CLSI EP17-A2).
Test Principle 11.
The ADVIA Centaur Total IgE (tlgE) assay is a two-site sandwich immunoassay using direct chemiluminometric technology, which uses constant amounts of two antibodies to lgE. The first antibody, in the Lite Reagent, is a goat anti-human IqE antibody labeled with acridinium ester. The second antibody, in the Solid Phase, is a mouse anti-human IgE antibody, which is covalently coupled to paramagnetic particles.
{6}------------------------------------------------
12. Performance Characteristics
The addition of the plasma (EDTA and lithium heparin) sample and detection capability (LoB, LoD, and LoQ) claims in the Instructions for Use (Package Inserts) for the ADVIA Centaur Total IgE (tlgE) assay was demonstrated by testing the performance characteristics with the following studies:
- . Specimen Equivalence by Method Comparison
- . Detection Capability (LoB, LoD, LoQ)
- . Interferences: EDTA and Heparin
The plasma (EDTA and lithium heparin) sample and detection capability (LoB, LoD, and LoQ) claims for the ADVIA Centaur Total IgE (tlgE) assay do not require the collection of additional analytical performance data. Therefore, all analytical performance data previously reviewed for the ADVIA Centaur Total IgE (tlqE) assay continues to apply to this assay, because the assay was not modified.
Detection Limit 12.1
Detection capability was determined in accordance with CLSI Document EP17-A2.
| Limit of Blank (LoB) | 1.5 IU/mL |
|---|---|
| Limit of Detection (LoD) | 2.0 IU/mL |
| Limit of Quantitation (LoQ) | 2.5 IU/mL |
The LoB corresponds to the highest measurement likely to be observed for a blank sample with a probability of 95%.
The LoD corresponds to the lowest concentration of total IgE that can be detected with a probability of 95%.
The LoQ corresponds to the lowest amount of total IgE in a sample at which the within laboratory CV is ≤ 20%.
12.2 Specimen Equivalence by Method Comparison
Specimen equivalency was determined with the weighted Deming linear regression model in accordance with CLSI Document EP09-A3. The following results were obtained:
| Comparison | N* | Sample Interval | Slope(95% CI) | Intercept(95% CI) | CorrelationCoefficient (r) |
|---|---|---|---|---|---|
| Dipotassium EDTA Plasmavs. Serum | 73 | 2.80 – 2748.84 IU/mL | 0.99(0.975 - 1.012) | 0.28 IU/mL(0.086 - 0.475) | 1.00 |
| Lithium Heparin vs. Serum | 73 | 2.80 - 2748.84 IU/mL | 1.00(0.989 - 1.020) | 0.18 IU/mL(-0.102 - 0.458) | 1.00 |
- N = Number of samples tested.
The assay is designed to have a slope of 0.90–1.10 for alternate tube types versus serum.
12.3 Interferences: EDTA and Heparin
Interference testing was performed in accordance with CLSI Document EP07-ed3. The following results were obtained:
{7}------------------------------------------------
| Interferent | Interferent Concentration | Analyte Concentration (IU/mL) | Bias (%) |
|---|---|---|---|
| Dipotassium EDTA | 9.0 mg/mL | 121.51 | -1.7 |
| 1624.13 | 1.4 | ||
| Heparin | 75 U/mL | 167.48 | -1.7 |
| 1450.12 | -1.1 |
Clinical Studies 12.4
Not applicable.
12.5 Clinical Cut-off
Not applicable.
13. Conclusion
The ADVIA Centaur Total IgE (tlgE) assay with the addition of the plasma (EDTA and lithium heparin) sample and new detection capability (LoB, LoD, and LoQ) claims in the Instructions for Use (package insert) is substantially equivalent to the currently marketed ADVIA Centaur Total IgE (tlgE) assay (K920372).
§ 866.5510 Immunoglobulins A, G, M, D, and E immunological test system.
(a)
Identification. An immunoglobulins A, G, M, D, and E immunological test system is a device that consists of the reagents used to measure by immunochemical techniques the immunoglobulins A, G, M, D, an E (serum antibodies) in serum. Measurement of these immunoglobulins aids in the diagnosis of abnormal protein metabolism and the body's lack of ability to resist infectious agents.(b)
Classification. Class II (performance standards).