K Number
K220178
Date Cleared
2022-03-23

(61 days)

Product Code
Regulation Number
866.5510
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The IgE assay is intended for use in the quantitative determination of Total Immunoglobulin E (lgE) concentration in human serum and plasma (lithium heparin, K2-EDTA, K3-EDTA, K3-EDTA) on Beckman Coulter AU/DxC AU clinical chemistry analyzers. The determination aids in the diagnosis of IgEmediated allergic disorders in conjunction with other clinical findings. For in vitro diagnostic use only.

Device Description

The Total Immunoglobulin E (IgE) reagent kit is in a liquid ready-to-use format designed for optimal performance on Beckman Coulter's AU/DxC AU clinical chemistry analyzers. Each reagent kit contains one buffer reagent (R1), one antibody reagent (R2), and a six-level lot matched calibrator set. The IgE reagent test system utilizes a turbidimetric immunoassay methodology. The AU analyzer measures the change in absorbance at 800 nm to calculate and express the concentration of immunoglobulin E in the test sample based on a stored calibration curve. The IgE assay is traceable to the World Health Organization (WHO) 3rd International Standard 11/234.

AI/ML Overview

The provided document is a 510(k) Premarket Notification for the Beckman Coulter Total Immunoglobulin E (IgE) assay. It describes the device's performance characteristics and how they meet established acceptance criteria, demonstrating substantial equivalence to a predicate device.

Here's an analysis of the acceptance criteria and study data based on the provided document:

1. Table of Acceptance Criteria and Reported Device Performance

The document presents several tables detailing acceptance criteria and study results for different performance parameters.

Table 1: Method Comparison Summary (from Table 9.1.1)

This table compares the Beckman Coulter IgE Assay (candidate device) against the Roche Elecsys IgE II Assay (predicate device).

Acceptance CriteriaParameterReported Device PerformancePass/Fail
Not explicitly stated as a separate column, but implied by the comparison to predicate.Slope0.966 [0.950 - 0.981]Pass (implied, as good correlation is shown)
Intercept (IU/mL)1.0 [-1.0 - 3.0]Pass (implied, as good correlation is shown)
R (Correlation Coefficient)0.996Pass (implied, as good correlation is shown)

Table 2: Precision Performance Summary (from Table 9.2.1)

Test SampleMean (IU/mL)Acceptance Criteria (Repeatability CV%)Reported Repeatability CV (%)Acceptance Criteria (Total Precision CV%)Reported Total Precision CV (%)Pass/Fail
Control 1113.5≤7.0%1.7≤7.5%2.0Pass
Control 2229.4≤7.0%1.0≤7.5%1.6Pass
Pool 170.4≤5.0 IU/mL3.0 (CV%) / 2.1 (IU/mL)≤7.0 IU/mL3.3 (CV%) / 2.3 (IU/mL)Pass
Pool 2167.9≤7.0%1.4≤7.5%2.4Pass
Pool 3413.6≤7.0%0.9≤7.5%1.4Pass

Note: For Pool 1, the document states criteria of "≤5.0 IU/mL" for repeatability and "≤7.0 IU/mL" for total precision. The reported results are 2.1 IU/mL (repeatability) and 2.3 IU/mL (total precision), which meet these criteria.

Table 3: Analytical Range (Linearity) Study Summary (from Table 9.3.1)

Sample TypeAcceptance Criterion (Linear Range)Reported Linear From (IU/mL)Reported Linear To (IU/mL)Pass/Fail
Serum20 – 500 IU/mL (with Allowable Difference of 10.0 IU/mL or 10%*)13.2575.5Pass

Further breakdown of linearity criteria for individual points in Table 9.3.2 shows "Bias Spec" of 10.0 for values ≤100 IU/mL and "%Bias Spec" of 10.0 for values > 100 IU/mL. All individual points "Pass".

Table 4: Detection Limit Study Summary (LoB, LoD & LoQ) (from Table 9.4.1)

ParameterAcceptance CriteriaReagent Lot 1 ResultReagent Lot 2 ResultPass/Fail
LoB (IU/mL)≤10.05.97.5Pass
LoD (IU/mL)≤15.013.812.5Pass
LoQ (IU/mL)≤20.0 at ≤35% CV19.6 at 11.9% CV17.8 at 7.8% CVPass

Table 5: Anticoagulant Study Results Summary (from Table 9.7.1)

Plasma TypeAcceptance Criteria (Slope)Reported SlopeAcceptance Criteria (Intercept)Reported InterceptAcceptance Criteria (R)Reported RPass/Fail
Na Heparin[0.9 - 1.1]0.989[± 20 IU/mL]0.0[≥ 0.97]0.999Pass
Li Heparin[0.9 - 1.1]0.989[± 20 IU/mL]-0.5[≥ 0.97]0.999Pass
K2 EDTA[0.9 - 1.1]0.986[± 20 IU/mL]-1.7[≥ 0.97]0.997Pass
K3 EDTA[0.9 - 1.1]0.964[± 20 IU/mL]-2.4[≥ 0.97]0.998Pass

Table 6: In-Use Stability Study Summary (from Table 9.8.1)

Stability ParameterClaim (days)Tested to (days)Pass/Fail
Reagent open bottle*2829Pass
Calibration interval†1415Pass
Calibrator open bottle4554Pass

Note: For the stability studies, the acceptance criteria are implicitly met if the "Pass" status is indicated after testing beyond the claimed duration, implying that the mean IgE recovery criteria (within 10 IU/mL or 10% of Day 0 mean) were met.

2. Sample sizes used for the test set and the data provenance

  • Method Comparison: 136 fresh serum samples spanning the analytical measuring range. Provenance not explicitly stated.
  • Precision: Three lots of IgE reagent, tested using two levels of human serum-based quality control material and three patient pools. The experimental design used duplicate sample analysis twice daily over twenty working days (N=80 results for each control/pool). Provenance not explicitly stated.
  • Linearity: 15-level linearity test set of inter-diluted patient pools. Provenance not explicitly stated.
  • Sensitivity (LoB, LoD, LoQ): LoB evaluated four unique lots of Immunoglobulin (Ig)-depleted human serum. LoD and LoQ used native patient pools diluted with Ig-depleted serum. Provenance not explicitly stated.
  • Analytical Specificity (Interference): Test samples containing various interferents (Hemoglobin, Bilirubin, Lipemia, RF, 21 common drugs and concentrations). Specific sample sizes per interferent are not given but are implied by tables (e.g., "Level Tested"). Provenance not explicitly stated.
  • Anticoagulant Studies: Freshly drawn serum and plasma from apparently healthy adult volunteer donors. Five specimen tubes were drawn from each donor: one serum tube and one tube of each type of anticoagulant. Sample sizes listed for each anticoagulant type: Na Heparin (N=55), Li Heparin (N=55), K2 EDTA (N=55), K3 EDTA (N=53). Provenance not explicitly stated.
  • In-Use Stability: Three levels of quality control material were evaluated. Specific N is not detailed, but implied by the "classical design for sampling, storage, and testing." Provenance not explicitly stated.

Generally, the data provenance (e.g., country of origin, retrospective/prospective) is not explicitly mentioned for any of the studies in the provided document. The studies appear to be prospective analytical performance studies designed to evaluate the device's technical specifications.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

This device is an in vitro diagnostic (IVD) quantitative assay for Total IgE. The "ground truth" for such devices is typically established through:

  • Traceability to International Standards: The IgE assay is stated to be traceable to the World Health Organization (WHO) 3rd International Standard 11/234. This is a primary method of establishing "ground truth" for quantitative lab tests, where the standard itself defines the "true" concentration.
  • Reference Methods: The method comparison study uses a legally marketed predicate device (Roche Elecsys IgE II Assay) as the comparative "truth" or reference standard for patient samples. The predicate device itself would have undergone similar rigorous validation, likely traceable to international standards.

Therefore, for this type of device, the concept of "experts" in the sense of clinicians or radiologists establishing a diagnostic "ground truth" for individual cases (like in an imaging study) is not directly applicable. The "ground truth" for the test set is inherent in the certified international standard and the established performance of the predicate device.

4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

Not applicable. Adjudication methods like 2+1 or 3+1 consensus are typically used in studies involving subjective interpretation (e.g., imaging reads) to establish a consensus ground truth. For quantitative in vitro diagnostic assays, the "truth" is determined by reference methods, international standards, and analytical evaluation against defined statistical criteria, not by human consensus or adjudication of individual results.

5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

Not applicable. This is an in vitro diagnostic (IVD) assay, not an AI-powered diagnostic imaging device or an AI-assisted interpretation tool for human readers. Therefore, MRMC studies and the concept of "human readers improving with AI assistance" do not apply.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

Yes, the studies presented are all "standalone" in the sense that they evaluate the performance of the analytical instrument and reagent system itself (the "algorithm only" in a broader sense of the measurement process) without direct human intervention in the result generation or interpretation step for the purpose of the performance evaluation. The device provides a quantitative result (IgE concentration), which is then used by clinicians in conjunction with other clinical findings. The performance data (e.g., precision, linearity, sensitivity, method comparison) directly reflects the device's inherent analytical capabilities.

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

The ground truth for this IgE assay is established primarily through:

  • International Standards: The assay's traceability to the WHO 3rd International Standard 11/234 is explicitly stated and serves as the ultimate reference for IgE concentration.
  • Reference Measurement Procedures/Predicate Device: For method comparison, the Roche Elecsys IgE II Assay serves as the comparative "reference" for patient samples. This predicate device itself is assumed to be traceable to international standards and validated.
  • A priori defined analytical specifications: For precision, linearity, and detection limits, the "ground truth" or acceptance criteria are statistically derived technical specifications (e.g., CV limits, bias limits) that the device must meet to be considered analytically robust.

8. The sample size for the training set

The document does not describe a "training set" in the context of machine learning or AI models because this is not an AI/ML-based device. It is a traditional in vitro diagnostic immunoassay. The performance studies detailed are validation studies for the cleared device.

9. How the ground truth for the training set was established

Not applicable, as there is no "training set" in the AI/ML sense for this traditional IVD product. If "training set" broadly refers to materials used during the development and calibration of the assay, the document indicates that the assay is traceable to the WHO 3rd International Standard 11/234, and a six-level lot-matched calibrator set is included, which would be used to establish the calibration curve for the assay.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, with the letters "FDA" in a blue square. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

March 23, 2022

Beckman Coulter, Inc. Veronica Colinayo Staff Regulatory Affairs 250 S. Kraemer Boulevard, Mail Stop B1.SE.03 Brea, California 92821

Re: K220178

Trade/Device Name: Total Immunoglobulin E (IgE) Regulation Number: 21 CFR 866.5510 Regulation Name: Immunoglobulins A, G, M, D, And E Immunological Test System Regulatory Class: Class II Product Code: DGC Dated: January 20, 2022 Received: January 21, 2022

Dear Veronica Colinayo:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's

{1}------------------------------------------------

requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801 and Part 809); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Ying Mao. Ph.D. Chief Division of Immunology and Hematology Devices OHT7: Office of In Vitro Diagnostics and Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

Submission Number (if known)

K220178

Device Name

Total Immunoglobulin E (IgE)

Indications for Use (Describe)

The IgE assay is intended for use in the quantitative determination of Total Immunoglobulin E (lgE) concentration in human serum and plasma (lithium heparin, K2-EDTA, K3-EDTA, K3-EDTA) on Beckman Coulter AU/DxC AU clinical chemistry analyzers. The determination aids in the diagnosis of IgEmediated allergic disorders in conjunction with other clinical findings. For in vitro diagnostic use only.

Type of Use (Select one or both, as applicable)

( Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

*DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW *

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Image /page/3/Picture/0 description: The image contains the logo for Beckman Coulter. The logo consists of a red oval shape with two white curved lines inside, resembling a stylized eye or a wave pattern. To the right of the oval is the company name, "BECKMAN COULTER," written in bold, black capital letters, with "BECKMAN" on the top line and "COULTER" on the bottom line.

1.0 Submitted By

Veronica V. Colinayo Staff Regulatory Affairs Beckman Coulter, Inc. 250 S. Kraemer Blvd. Mail Stop: B1.SE.03 Brea, CA 92821 Telephone: (714) 961-6515 Fax: (714) 961-4234 Email: vvcolinayo@beckman.com

2.0 Date Submitted

January 20, 2022

3.0 Device Name(s)

3.1 Proprietary Names

Total Immunoglobulin E (IgE)

3.2 Classification Name

Immunoglobulins A, G, M, D, and E immunological test system [866.5510, Product Code DGC]

4.0 Predicate Device

Candidate DevicePredicate DeviceManufacturerDocketNumber
Total Immunoglobulin E(IgE)Elecsys IgE IIImmunoassayRoche DiagnosticsK061970

5.0 Device Description

The Total Immunoglobulin E (IgE) reagent kit is in a liquid ready-to-use format designed for optimal performance on Beckman Coulter's AU/DxC AU clinical chemistry analyzers. Each reagent kit contains one buffer reagent (R1), one antibody reagent (R2), and a six-level lot matched calibrator set. The IgE reagent test system utilizes a turbidimetric immunoassay methodology. The AU analyzer measures the change in absorbance at 800 nm to calculate and express the concentration of immunoglobulin E in the test sample based on a stored calibration curve. The IgE assay is traceable to the World Health Organization (WHO) 3rd International Standard 11/234.

{4}------------------------------------------------

Image /page/4/Picture/0 description: The image shows the logo for Beckman Coulter. The logo consists of a red circle with two white curved lines inside, resembling a stylized wave or heartbeat. To the right of the circle, the words "BECKMAN" and "COULTER" are stacked on top of each other in a bold, sans-serif font.

6.0 Intended Use

The IgE assay is intended for use in the quantitative determination of Total Immunoglobulin E (IgE) concentration in human serum and plasma (lithium heparin, sodium heparin, K2 EDTA, K3 EDTA) on Beckman Coulter AU/DxC AU clinical chemistry analyzers. The determination aids in the diagnosis of IgE-mediated allergic disorders in conjunction with other clinical findings. For in vitro diagnostic use only.

7.0 Comparison to the Predicate

The following table describes the similarities and differences between the candidate device and the predicate device identified in Section 4.0 of this summary:

DeviceSimilarities
Intended Use: quantitative determination of total IgE for IVD useSample types: serum, plasma (heparin & EDTA)Format: liquid, ready-to-useAntibody: monoclonal, mouseStorage conditions: 2-8°CSame as Roche'sIgE II assay on the Cobas Immunoassay System
Differences
BeckmanCoulter IgEassay forAU/DxC AUClinicalChemistryAnalyzersOperating Principle
Assay Format
Calibrator scheme
Calibration StabilityIgE: 14 daysIgE II: 7 days
TraceabilityIgE: WHO 3rd IRP 11/234IgE II: WHO 2nd IRP 75/502
Analytical Measuring RangeIgE: 20 - 500 IU/mLIgE II: 0.100 - 2500 IU/mL
Extended Measuring Range(manual or auto-dilution)IgE: 500 - 1,000 IU/mLIgE II: >2,500 up to 50,000 IU/mL
Limit of DetectionIgE: 15 IU/mLIgE II: 0.100 IU/mL
Limit of QuantitationIgE: 20 IU/mL with ≤ 35% CVIgE II: 0.500 IU/mL with < 20% CV

{5}------------------------------------------------

Image /page/5/Picture/0 description: The image shows the logo for Beckman Coulter. The logo consists of a red circular shape on the left, with two curved white lines inside. To the right of the circle, the words "BECKMAN" and "COULTER" are written in bold, black letters, with "BECKMAN" on top of "COULTER".

8.0 Comparison testing

Comparative studies were conducted for the candidate IgE reagent test system on the DxC 700 AU Clinical Chemistry Analyzer. Equivalence was demonstrated through a method comparison study. Additional performance studies verified that the technological differences between the candidate and predicate devices did not adversely affect safety and effectiveness. Discussions of the following performance parameters are presented in this 510(k) Summary:

Method Comparison Imprecision Linearity Sensitivity Reference Interval Specificity Anticoagulants In-use (Reagent Onboard & Calibration) Stability

9.0 Summary of Performance Data

The data in the Premarket Notification supports a finding of substantial equivalence to measurand test systems already in commercial distribution.

9.1 Method Comparison Summary

Method comparison and bias estimation experiments were designed in accordance with the CLSI Guideline EP09c "Measurement Procedure Comparison and Bias Estimation Using Patient Samples - Third Edition". The study evaluated 136 fresh serum samples spanning the analytical measuring range of the candidate IgE assay, where test sample concentrations ranged from approximately 25 to 499 IU/mL IgE as measured by the predicate assay. The study results are summarized in Table 9.1.1 based on Weighted Deming regression analysis.

Method YSlope[95% C.I.]Intercept(IU/mL)[95% C.I.]RNMethod X
BeckmanCoulterIgE Assay0.966[0.950 - 0.981]1.0[-1.0 - 3.0]0.996136Roche ElecsysIgE II Assay

{6}------------------------------------------------

Image /page/6/Picture/0 description: The image shows the logo for Beckman Coulter. The logo consists of a red circle with two white curved lines inside, resembling a stylized wave or flow. To the right of the circle is the company name, "BECKMAN" in bold, followed by "COULTER" in a smaller, bold font, stacked below the first word.

9.2 Precision

Repeatability (within-run) and within-laboratory (total) precision studies were designed from CLSI Guideline EP05-A3 "Evaluation of Quantitative Measurement Procedures; Approved Guideline - Third Edition". Precision studies evaluated three lots of IgE reagent on a single DxC 700 AU analyzer, and one IgE reagent lot on three DxC 700 AU analyzers. Test samples included two levels of human serum-based quality control material and three patient pools. The experimental design used duplicate sample analysis twice daily over twenty working days (N=80) in random order. The performance summary for repeatability (within-run) and total (withinlaboratory) imprecision for the candidate IgE assay is provided in Table 9.2.1. The total imprecision estimates include the between-run, between-day, between-lot, and between-instrument components of variance.

TestSampleMean(IU/mL)RepeatabilityResultsCriteria*Total PrecisionResultCriteria*Pass/Fail
Control 1113.51.91.7≤7.0%CV2.32.0≤7.5%CVPass
Control 2229.42.31.0≤7.0%CV3.71.6≤7.5%CVPass
Pool 170.42.13.0≤5.0IU/mL2.33.3≤7.0IU/mLPass
Pool 2167.92.41.4≤7.0%CV4.02.4≤7.5%CVPass
Pool 3413.63.90.9≤7.0%CV5.71.4≤7.5%CVPass

Table 9.2.1 IgE Reagent Imprecision Performance Summary

*Repeatability criteria is 5.0 IU/mL for mean recovery values ≤ 71.4 IU/mL, and 7.0% CV for mean recovery values > 71.4 IU/mL; total precision criteria is 7.0 IU/mL for mean recovery values ≤ 93.3 IU/mL, and 7.5% CV for mean recovery values > 93.3 IU/mL.

{7}------------------------------------------------

Image /page/7/Picture/0 description: The image shows the logo for Beckman Coulter. The logo consists of a red circle with two white curved lines inside, resembling a stylized eye. To the right of the circle, the words "BECKMAN" and "COULTER" are written in bold, black letters, with "BECKMAN" on top of "COULTER".

9.3 Analytical Range (Linearity)

Analytical range (linearity) studies were designed in accordance with the CLSI guideline EP06-A "Evaluation of the Linearity of Quantitative Measurement Procedures: A Statistical Approach; Approved Guideline". The study used a 15-level linearity test set of inter-diluted patient pools that spanned the claimed analytical measuring range of the candidate IgE assay. The initial assessment of the study data in accordance with the EP06-A document found the modelled non-linearity to be not statistically significant, and the assay was deemed linear; as such, no further evaluation of non-linear models against a bias criterion was required. Table 9.3.1 provides the linearity study summary that supports the analytical measuring range claim for the candidate IgE assay, and Table 9.3.2 provides the linearity test results.

SampleTypeAcceptance CriterionResultsPass/Fail
Linear Range(IU/mL)Allowable DifferenceSpecificationLinear FromLinear To
Serum20 – 50010.0 IU/mL or 10.0%*13.2575.5Pass
Weighted Linear Regressiony = 1.01273x + 0.630R2 = 0.99990

Table 9.3.1 IgE Linearity Study Summary

*Values ≤100 IU/mL use the unit specification, and > 100 IU/mL use the percent specification.

SampleTargetAveragePredictedOrder 1PredictedNonlinearCOVBiasBiasSpec%Bias%BiasSpecPass/Fail
LIN_0113.213.214.0N/A*100.0N/A*10.0N/A*-Pass
LIN_0215.116.415.9N/A*100.0N/A*10.0N/A*-Pass
LIN_0324.525.225.4N/A*100.0N/A*10.0N/A*-Pass
LIN_0450.752.152.0N/A*100.0N/A*10.0N/A*-Pass
LIN_0588.290.990.0N/A*100.0N/A*10.0N/A*-Pass
LIN_06155.7157.6158.3N/A*100.0N/A*-N/A*10.0Pass
LIN_07215.6220.0219.0N/A*100.0N/A*-N/A*10.0Pass
LIN_08266.2269.2270.2N/A*100.0N/A*-N/A*10.0Pass
LIN_09331.8335.3336.7N/A*100.0N/A*-N/A*10.0Pass
LIN_10388.1392.2393.7N/A*100.0N/A*-N/A*10.0Pass
LIN_11425.6428.7431.6N/A*100.0N/A*-N/A*10.0Pass
LIN_12463473.8469.5N/A*100.0N/A*-N/A*10.0Pass
LIN_13511.8520.9518.9N/A*100.0N/A*-N/A*10.0Pass
LIN_14530.5542.6537.9N/A*100.0N/A*-N/A*10.0Pass
LIN_15575.5575.5583.5N/A*100.0N/A*-N/A*10.0Pass

Table 9.3.2 IgE Linearity Study Degree of Non-Linearity Test Results

*Not Applicable: The non-linearity was deemed not statistically significant, therefore no additional bias assessment between the non-linear and linear models was required.

{8}------------------------------------------------

Image /page/8/Picture/0 description: The image shows the logo for Beckman Coulter. The logo consists of a red circle with two white curved lines inside, and the words "BECKMAN COULTER" in black, with "BECKMAN" on top of "COULTER". The logo is simple and modern, and the colors are eye-catching.

9.4 Sensitivity (Detection Capability)

Limit of Blank (LoB), Limit of Detection (LoD), and Limit of Quantitation (LoQ) studies were designed from the CLSI guideline EP17-A2 "Evaluation of Detection Capability for Clinical Laboratory Measurement Procedures Approved Guideline - Second Edition" using the Classical Approach. The EP17 study evaluated two lots of candidate lgE reagent on one DxC 700 AU analyzer. The LoB evaluation assessed four unique lots of Immunoglobulin (Ig)-depleted human serum as the individual blank samples, and the LoD and LoQ evaluations used native patient pools diluted with Ig-depleted serum to achieve the low analyte levels. The results of the Detection Capability Study are shown in Table 9.4.1.

Reagent LotLoB ResultLoB ClaimLoD ResultLoD ClaimLoQ ResultLoQ ClaimPass/Fail
(IU/mL)
15.9≤10.013.8≤15.019.6 at 11.9% CV≤20.0 at ≤35% CVPass
27.512.517.8 at 7.8% CVPass
Table 9.4.1 IqE Detection Limit Study Summary (LoB, LoD & LoQ)

{9}------------------------------------------------

Image /page/9/Picture/0 description: The image shows the logo for Beckman Coulter. The logo consists of a red circle with two white curved lines inside, resembling a stylized wave or flow. To the right of the circle, the words "BECKMAN" and "COULTER" are stacked on top of each other in a bold, sans-serif font.

9.5 Reference Interval (Expected values)

IgE reference intervals are significantly influenced by age, sex, geographic location, microflora of the gastrointestinal tract, diet of the population, as well as environmental factors such as climate change.1 As such, Beckman Coulter recommends that each clinical laboratory establish a range of expected values for its own local population, as dictated by good laboratory practices. The reference intervals presented in Table 7.5.1 were taken from literature.2

AnalyteSample TypeCondition (age)Levels (kIU/L)*
IgESerum0 - <7 years<25 - 440
IgESerum7 - <19 years<25 - 450
IgESerumAdult (20-60 years)0 - 160

Table 9.5.1 IgE Expected Values2

*Equivalent to IU/mL

  • 1 CLSI I/LA20, Analytical Performance Characteristics, Quality Assurance, and Clinical Utility of Immunological Assays for Human Immunoglobulin E Antibodies of Defined Allergen Specificities, 3rd Edition, October 2016, pp. 23; 27.
  • 2 Rifai N, Tietz Handbook of Clinical Chemistry and Molecular Diagnostics, 6th Ed., Elsevier (2018).

{10}------------------------------------------------

Image /page/10/Picture/0 description: The image contains the logo for Beckman Coulter. The logo consists of a red oval shape with two white curved lines inside, placed to the left of the company name. The text "BECKMAN" is written in bold, uppercase letters above the text "COULTER", which is also in bold, uppercase letters.

9.6 Analytical Specificity

Interference studies were designed in accordance with the CLSI Guidelines EP07 "Interference Testing in Clinical Chemistry - Third Edition" and EP37 "Supplemental Tables for Interference Testing in Clinical Chemistry" to identify and evaluate substances that could potentially interfere with the candidate InE assay. The criteria for no significant interference (NSI) required the test sample (containing interferent) recovery mean to be within ± 10 IU/mL for recovered values ≤ 100 IU/mL, or 10% of the control recovery mean values > 100 IU/mL (sample containing no interferent).

The interference study results are presented in Tables 9.6.1 and 9.6.2. The drug omalizumab is a known interferent with the IgE assay methodology1 and was not tested; falsely decreased results may occur in patients being treated with the drug. Antibody specificity studies were performed externally by the supplier of the monoclonal antibody component and showed no cross-reactivity with human IgA, IqD, IgG, or IgM.

SubstanceSample TypeSourceLevel TestedObserved Effect
HemoglobinSerumHemolysate (human)1,000 mg/dLNSI
Unconjugated BilirubinSerumPorcine60 mg/dLNSI
LipemiaSerumIntralipid1,000 mg/dLNSI
RFSerumHuman250 IU/mLNSI

Table 9.6.1 IgE Endogenous Interferents Study Summary

NSI = No Significant Interference within ± 10.0 IU/mL or 10%

SubstancesSample TypeAnalyte Level(IU/mL)Range ofObservedMean % BiasObservedEffect
21 commondrugs andconcentrationsSerum~160-1.6% to 0.9%NSI

NSI = No Significant Interference within ± 10.0 IU/mL or 10%

1 Hamilton RG. Accuracy of Food and Drug Administration-cleared IgE antibody assays in the presence of anti-IgE (omalizumab). J Allergy Clin Immunol. 2006;117(4):759-766

{11}------------------------------------------------

Image /page/11/Picture/0 description: The image shows the logo for Beckman Coulter. The logo consists of a red circle with a white, wave-like design inside. To the right of the circle, the words "BECKMAN" and "COULTER" are written in bold, black letters, with "BECKMAN" on top of "COULTER".

9.7 Anticoagulant Studies (Serum vs. Plasma)

Methods comparison studies were used to evaluate plasma as an equivalent sample type. The methods comparison and bias estimation experiments were designed using the CLSI Guideline EP09c "Measurement Procedure Comparison and Bias Estimation Using Patient Samples; Third Edition". The study utilized freshly drawn serum and plasma from apparently healthy adult volunteer donors, where five specimen tubes were drawn from each donor: one serum tube and one tube of each type of anticoagulant. The acceptance criteria and study results are summarized in Table 9.7.1 based on Weighted Deming regression analysis.

PlasmaTypeLevelTestedNSlope[0.9 - 1.1]Intercept[± 20 IU/mL]R[≥ 0.97]Pass/Fail
Na Heparin17 Units/mL550.9890.00.999Pass
Li Heparin17 Units/mL550.989-0.50.999Pass
K2 EDTA1.8 mg/mL550.986-1.70.997Pass
K3 EDTA1.8 mg/mL530.964-2.40.998Pass

Table 9.7.1 IgE Anticoagulant Study Results Summary

{12}------------------------------------------------

Image /page/12/Picture/0 description: The image shows the logo for Beckman Coulter. The logo consists of a red oval shape on the left side, with two white curved lines inside. To the right of the oval is the company name, "BECKMAN COULTER", in bold, black letters.

9.8 In-Use Stability

In-Use Stability studies were designed to verify the open-bottle stability claims for the candidate IqE kit components and the IqE assay calibration interval. Testing was designed in accordance with the CLSI EP25-A guideline "Evaluation of Stability of In Vitro Diagnostic Reagents; Approved Guideline" using a classical design for sampling, storage, and testing. The study design evaluated in-use metrics at both the start and end of the shelf stability period for reagent kits stored under normal conditions (2-8°C) and including reagent kits that were exposed to post-shipping stress conditions. Three levels of quality control material were evaluated through the duration of the test period, where the mean lgE recovery at the last time point must be within 10 IU/mL or 10% of the mean Day 0 recovery. Table 9.8.1 summarizes the study results for the prestressed reagent kits beyond the claimed shelf life of 24 months.

Pre-stressConditionShelf Life(months)Stability Parameter(2-8°C storage)Claim(days)Tested to(days)Pass/Fail
WinterSummer29Reagent open bottle*2829Pass
WinterSummer29Calibration interval†1415Pass
WinterSummer29Calibrator open bottle4554Pass

Table 9.8.1: IqE In-Use Stability Study Summary

*Reagent stored onboard (on instrument) for study duration

*For Calibration Cycle 1, the Day 15 test point uses the Day 0 calibration curve; for Calibration Cycle 2, the Day 15 test point uses a new calibration curve based on a Day 15 recalibration.

{13}------------------------------------------------

Image /page/13/Picture/0 description: The image shows the logo for Beckman Coulter. The logo consists of a red oval shape on the left, with two white curved lines inside. To the right of the oval is the company name, "BECKMAN COULTER," written in a bold, sans-serif font. The word "BECKMAN" is stacked on top of the word "COULTER."

10.0 Conclusion

Beckman Coulter's Total Immunoglobulin E (IgE) Reagent is substantially equivalent to the Roche Elecsys IgE II reagent test system (K061970) as demonstrated through methods comparison and precision studies. The performance testing presented in this submission provides evidence that the device is safe and effective in its intended use.

This 510(k) Summary is being submitted in accordance with the requirements of the Safe Medical Device Act of 1990 and the implementing regulation 21 CFR 807.92.

§ 866.5510 Immunoglobulins A, G, M, D, and E immunological test system.

(a)
Identification. An immunoglobulins A, G, M, D, and E immunological test system is a device that consists of the reagents used to measure by immunochemical techniques the immunoglobulins A, G, M, D, an E (serum antibodies) in serum. Measurement of these immunoglobulins aids in the diagnosis of abnormal protein metabolism and the body's lack of ability to resist infectious agents.(b)
Classification. Class II (performance standards).