Search Filters

Search Results

Found 92 results

510(k) Data Aggregation

    K Number
    K242170
    Date Cleared
    2025-04-18

    (268 days)

    Product Code
    Regulation Number
    866.5270
    Reference & Predicate Devices
    Why did this record match?
    510k Summary Text (Full-text Search) :

    98168

    Re: K242170
    Trade/Device Name: K-ASSAY CRP (Ver.2)
    Regulation Number: 21 CFR 866.5270
    Classification Name:** C-reactive protein immunological test system
    Regulation Number: 21 CFR 866.5270
    --------------------------|-------------------------------------------|
    | Classification | Class 2 (866.5270
    ) | Class 2 (866.5270) |
    | Product Code | DCK | DCK |
    | Indications for Use | K-ASSAY® CRP (Ver.2) is

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    K-ASSAY® CRP (Ver.2) is intended to be used for the quantitative determination of C-reactive protein (CRP) in human serum and plasma (potassium-EDTA or lithium-heparin) by immunoturbidimetric assay. Measurement of CRP aids in the detection and evaluation of infection, tissue injury, inflammatory disorders and associated diseases. FOR IN VITRO DIAGNOSTIC USE.

    Device Description

    The K-ASSAY® CRP (Ver.2) assay quantifies C-reactive protein based on immunoturbidimetric assay. The reagent uses latex combined with goat polyclonal antibody specific to human CRP. By adsorbing CRP in the sample to the surface of the latex particles and reacting it with the anti-CRP antibody, specific aggregation corresponding to the CRP concentration occurs. Since the absorbance of the reaction changes in proportion to the amount of aggregation, the concentration of CRP in the sample is determined based on the calibration curve prepared using a standard of known CRP concentrations. The K-ASSAY® CRP (Ver.2) assay can be run using a chemistry analyzer. 6 levels of calibrators from the K-ASSAY® CRP Calibrator (Ver.2) calibrators are used for quantifying the levels of CRP present in the patient's sample.

    AI/ML Overview

    This document describes the FDA 510(k) clearance for the K-ASSAY CRP (Ver.2) IVD device. This device is an in-vitro diagnostic test system, which means it analyzes biological samples in a lab setting rather than directly interacting with a patient.

    Therefore, the concepts of "human readers," "AI assistance," "effect size," "multi reader multi case (MRMC) comparative effectiveness study," and "standalone (i.e. algorithm only without human-in-the loop performance)" are not applicable in this context. These terms are typically used for medical imaging AI/ML devices where human interpretation is involved.

    For this in-vitro diagnostic device, the "acceptance criteria" and "study that proves the device meets the acceptance criteria" are related to its analytical performance characteristics when compared to a predicate device, and the reported device performance refers to the results of these analytical studies.

    Here's the breakdown of the information provided within the scope of this IVD device:

    1. Table of Acceptance Criteria and Reported Device Performance

    For an IVD device like the K-ASSAY® CRP (Ver.2), the acceptance criteria are generally established by demonstrating equivalent or superior analytical performance compared to a legally marketed predicate device, and meeting established CLSI guidelines for accuracy, precision, linearity, and interference. The "reported device performance" refers to the actual study results for these characteristics.

    Acceptance Criteria (Implicit from predicate comparison and CLSI guidelines)Reported Device Performance (K-ASSAY® CRP (Ver.2))
    Intended Use Equivalence: Quantitative determination of CRP in human serum and plasma for detection/evaluation of infection, tissue injury, inflammatory disorders.K-ASSAY® CRP (Ver.2) is intended for the quantitative determination of CRP in human serum and plasma (potassium-EDTA or lithium-heparin) by immunoturbidimetric assay. Measurement of CRP aids in the detection and evaluation of infection, tissue injury, inflammatory disorders and associated diseases.
    Methodology Equivalence: Latex-enhanced immunoturbidimetric assay.Latex-enhanced (immuno)turbidimetric assay.
    Calibration Equivalence/Validation: Appropriate calibrator levels for the intended range.6 levels of calibrators (0.0, 10.0, 50.0, 150.0, 300.0 and 400.0 mg/L).
    Assay Range (Comparable to predicate): Demonstrates clinically relevant measuring range.Claimed Assay Range: 5.0 - 400.0 mg/L. (Predicate: 0.2 – 480 mg/L)
    Precision: Demonstrate acceptable repeatability, within-run, between-run, within-day, between-day, within-laboratory, between-site, and reproducibility CVs as per CLSI EP05-A3 guidelines.Single Site Precision (Combined Data From 3 Lots):
    • Within-Run CV: 0.8% to 2.2%
    • Between-Run CV: 0.0% to 0.6%
    • Within-Lot CV: 0.9% to 1.9%
    • Between-Lot CV: 0.0% to 1.2%
    • Total CV: 0.9% to 2.3%
      Multisite Precision (Combined Data From 3 Analyzers):
    • Repeatability CV: 0.5% to 1.5%
    • Between-Run CV: 0.0% to 1.1%
    • Between-Day CV: 0.6% to 1.8%
    • Between-Site CV: 0.9% to 1.8%
    • Reproducibility CV: 1.1% to 2.5% |
      | Interference: No significant interference from common endogenous and exogenous substances (recovery within 10% of initial value). | Endogenous Substances (up to listed concentrations, no significant interference): Bilirubin C (40 mg/dL), Bilirubin F (40 mg/dL), Cholesterol (300 mg/dL), Hemoglobin (1,000 mg/dL), Intralipid (500 mg/dL), Rheumatoid Factor (1,000 IU/mL), Triglycerides (1,000 mg/dL).
      Exogenous Substances (up to listed concentrations, no significant interference): Acetaminophen (1.5 mM), Amoxicillin (400 µmol/L), Aspirin (3.6 mM), Cephalexin (360 µmol/L), Fluconazole (480 µmol/L), Ibuprofen (2.5 mg/dL), Methotrexate (1,400 µmol/L), Prednisolone (2 µmol/L), Vitamin C (500 mg/L). |
      | Method Comparison: Strong correlation and minimal bias against the predicate device. | Regression Equation: y = 1.005x - 0.002, r = 0.999 (n = 175 clinical native serum samples), where y = K-ASSAY® CRP (Ver.2), x = predicate device. |
      | Linearity: Demonstrates linearity across the claimed assay range. | Regression Equation: y = 0.9709x - 1.095, r = 0.999 (tested range: 4.6 - 441.2 mg/L). |
      | Limit of Quantitation (LoQ): Clinically acceptable LoQ (within-laboratory precision ≤ 20% CV). | Reported LoQ: 1.0 mg/L (highest observed across 3 reagent lots). Claimed LoQ: 5.0 mg/L. |
      | Matrix Comparison: No significant difference in results across different sample matrices (serum vs. plasma). | K2-EDTA Plasma vs Serum: y = 1.007x - 0.141, r = 0.999
      Li-Heparin Plasma vs Serum: y = 0.972x + 0.074, r = 0.999 |
      | Expected Values (Verification): Distribution in normal population consistent with clinical literature. | 168 normal U.S. serum samples tested; 4 out of 168 samples (>5.0 mg/L, 2.4%) consistent with literature (≤ 5 mg/L indicates apparently healthy). |

    2. Sample Sizes Used for the Test Set and Data Provenance

    • Method Comparison: n = 175 clinical native serum samples.
    • Linearity: The number of samples/dilutions used is not explicitly stated, but the tested range was 4.6 - 441.2 mg/L.
    • Limit of Quantitation (LoQ): Not specified in terms of sample size for the LoQ determination itself (typically involves low-concentration samples measured multiple times).
    • Matrix Comparison: 42 donor samples (each collected into 3 different tubes: serum, K2-EDTA plasma, Li-Heparin plasma).
    • Precision (Single-site): For each of the 7 samples/controls, N=240 individual measurements (2 runs/day x 2 replicates/run x 20 days x 3 lots).
    • Precision (Multisite): For each of the 7 samples/controls, N=75 individual measurements (1 run/day x 5 replicates/run x 5 days x 3 analyzers).
    • Interference: The number of unique samples tested for interference is not explicitly stated. Typically, a few samples (e.g., low, medium, high concentration) are spiked with interferents and compared to unspiked controls.
    • Expected Values: 168 normal serum samples.
    • Data Provenance: The document states for "Expected Values" that 168 normal serum samples were "taken from healthy individuals in the U.S." This indicates a U.S. origin for at least this specific study. For other studies (Method Comparison, Precision, Linearity, Interference, Matrix Comparison), the specific country of origin is not mentioned. All studies are retrospectively analyzed in the sense that they were completed performance validation studies submitted to FDA.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

    For an in-vitro diagnostic device like this CRP assay, the "ground truth" is established through highly accurate and precise reference methods or established predicate devices, adhering to rigorous analytical chemistry best practices and guidelines (e.g., CLSI standards). There isn't a panel of human "experts" like radiologists interpreting images. The "ground truth" is quantitative and objective, derived from reference measurements.

    In this case:

    • Method Comparison: The predicate device (K-ASSAY® CRP (3), K023828) serves as the comparator for method comparison, which represents the established "ground truth" for clinical samples.
    • Linearity, LoQ, Precision, Interference, Matrix Comparison: The "ground truth" for these analytical performance studies is established by rigorous laboratory protocols, highly calibrated equipment, reference materials, and adherence to CLSI guidelines. The performance is assessed against predefined statistical criteria rather than expert consensus on individual cases.
    • Expected Values: The ground truth comes from established clinical literature and verified normal ranges based on studies of healthy populations.

    4. Adjudication Method for the Test Set

    Not applicable for an IVD device. Adjudication methods (e.g., 2+1, 3+1) are used in studies where subjective human interpretation (e.g., image reading) requires consensus building for ground truth establishment. For quantitative IVD assays, the results are numerical and objectively measured; therefore, there's no need for adjudication.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done

    No. As explained above, this is an in-vitro diagnostic device, not an imaging AI/ML device involving human readers.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done

    Yes, in essence, all the analytical performance studies (Method Comparison, Linearity, LoQ, Precision, Interference, Matrix Comparison) are "standalone" in nature for an IVD. The device's performance is evaluated based solely on its ability to accurately and precisely measure CRP concentrations in samples, without any human interpretation of the measurement itself or the involvement of an "algorithm" in the AI/ML sense. The device directly processes the sample and outputs a quantitative result.

    7. The Type of Ground Truth Used

    The ground truth for this IVD device's performance studies is primarily based on:

    • Reference Method/Predicate Device Comparison: For the method comparison study, the predicate device serves as the reference against which the new device's measurements are compared.
    • Reference Materials and Calibrators: For linearity, LoQ, and precision studies, the ground truth for concentration values is established using certified reference materials and meticulously prepared calibrators with known concentrations.
    • Spiked Samples: For interference studies, known amounts of interfering substances are added to samples, and the true CRP concentration (before interference) serves as the baseline ground truth.
    • Clinical Literature/Established Norms: For "Expected Values," the ground truth is derived from widely accepted clinical ranges and population studies cited in the literature.

    8. The Sample Size for the Training Set

    This document describes a 510(k) submission for a traditional IVD device, not an AI/ML-based device. Therefore, there is no "training set" in the context of machine learning. The device's performance is based on its chemical and physical principles (latex-enhanced immunoturbidimetric assay) and validated through the analytical studies detailed.

    9. How the Ground Truth for the Training Set Was Established

    Not applicable, as there is no "training set" for this type of IVD device.

    Ask a Question

    Ask a specific question about this device

    K Number
    K233242
    Date Cleared
    2024-01-18

    (112 days)

    Product Code
    Regulation Number
    866.5270
    Reference & Predicate Devices
    Why did this record match?
    510k Summary Text (Full-text Search) :

    Trade/Device Name: Atellica® CH High Sensitivity C-Reactive Protein 2 (hCRP2) Regulation Number: 21 CFR 866.5270
    system FDA Classification: Class II Review Panel: Immunology Product Code: NQD Regulation Number: 21 CFR 866.5270

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Atellica® CH High Sensitivity C-Reactive Protein 2 (hCRP2) assay is for in vitro diagnostic use in the quantitative determination of the concentration of C-Reactive Protein (CRP) in human serum and plasma (lithium heparin, sodium heparin or K2 EDTA) on the Atellica® CH Analyzer.

    Measurements from Atellica® CH High Sensitivity C -Reactive Protein 2 (hCRP2) may be used as an aid in identification of individuals at risk for future cardiovascular disease. Measurement of hCRP2, when used in coniunction with traditional clinical laboratory evaluation of acute coronary syndromes, may be useful as an independent marker of prognosis for recurrent events in patients with stable coronary disease or acute coronary syndromes.

    Device Description

    The Atellica CH High Sensitivity C-Reactive Protein 2 (hCRP2) assay is used for the quantitative determination of C-Reactive protein in human serum and plasma using the Atellica CH analyzer. This device is two ready-to-use reagent packs consisting of 23.1mL Phosphate buffer, polidocanol (1.9g/L), and sodium azide (0.1%) in Pack 1 and 12.3mL Mouse anti-CRP monoclonal antibodies (13mg/L), polystyrene particles (1g/L), human albumin (0.05%) and sodium azide (

    AI/ML Overview

    Here's a summary of the acceptance criteria and the study that proves the device meets them, based on the provided FDA 510(k) summary:

    Device: Atellica® CH High Sensitivity C-Reactive Protein 2 (hCRP2) assay

    1. Table of Acceptance Criteria and Reported Device Performance:

    Performance CharacteristicAcceptance Criteria (Design Goal)Reported Device Performance
    Detection Capability
    Limit of Blank (LoB)LoB ≤ Limit of Detection (LoD)0.06 mg/L
    Limit of Detection (LoD)LoD ≤ Limit of Quantitation (LoQ)0.11 mg/L
    Limit of Quantitation (LoQ)≤ 0.16 mg/L (with 9.50 mg/L (which is the upper limit of the measuring interval). This indicates the assay correctly identifies concentrations above the measuring interval and avoids a hook effect within the intended range.

    2. Sample Size Used for the Test Set and Data Provenance:

    • Detection Capability:
      • LoB: 3 reagent lots, 6 blank samples, 5 replicates per sample = 90 determinations.
      • LoD: 495 determinations (270 blank, 225 low-level replicates) using 3 reagent lots.
      • LoQ: 5 native low analyte serum samples, 5 replicates each = 25 determinations.
    • Precision (Repeatability/Within-Lab): 80 replicates per specimen (5 serum samples + 1 QC, total 6 specimens).
    • Reproducibility (Multi-site/Multi-lot): 225 replicates per specimen (5 serum samples + 3 QCs, total 8 specimens).
    • Assay Comparison (Method Comparison): 100 patient samples.
    • Specimen Equivalency: 55 patient samples for each specimen type (Sodium Heparin, Potassium EDTA, Lithium Heparin).
    • Interferences (HIL & Non-interfering Substances): Not explicitly stated, but typically involves a smaller number of samples spiked with interferents at multiple analyte concentrations.

    Data Provenance: Not explicitly stated, but the sample types are human serum and plasma, diluted with Atellica CH diluent (saline) or enriched as needed. This usually implies a mix of commercially sourced and/or in-house collected human samples. There is no information regarding the country of origin or whether the studies were retrospective or prospective.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts:

    This device is an in vitro diagnostic test for C-Reactive Protein (CRP) concentration. The "ground truth" for a quantitative assay like this is typically established by:

    • Reference methods/standards (e.g., ERM-DA474/IFCC for CRP, as mentioned).
    • Comparative analysis against a legally marketed predicate device (BN ProSpec CardioPhase hsCRP).
    • Standard preparation and gravimetric/volumetric assurance for spiked samples or known concentrations.

    Therefore, "experts" in the sense of clinical reviewers or pathologists establishing a diagnostic ground truth is not applicable here. The accuracy of the measurements is compared against established analytical criteria and methodologies.

    4. Adjudication Method for the Test Set:

    Not applicable in the context of this type of IVD performance study, as there is no subjective interpretation requiring adjudication of results from different observers. The output is a quantitative value (mg/L).

    5. If a Multi Reader Multi Case (MRMC) Comparative Effectiveness Study Was Done:

    No, an MRMC comparative effectiveness study was not done. This type of study is relevant for imaging or diagnostic tests where human interpretation plays a significant role and AI assistance might influence reader performance. For a quantitative in vitro diagnostic assay like high-sensitivity CRP, the measurement is automated and objective.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done:

    Yes, the studies described (Detection Capability, Precision, Reproducibility, Assay Comparison, Specimen Equivalency, Interferences, High-Dose Hook Effect) are all standalone performance evaluations of the Atellica CH High Sensitivity C-Reactive Protein 2 (hCRP2) assay as an automated laboratory test on the Atellica CH Analyzer. The device is intended for in vitro diagnostic use, meaning it operates without direct human interpretive input beyond running the test and reading the numerical result.

    7. The Type of Ground Truth Used:

    The ground truth for this device is established through:

    • Analytical Standards: The assay is traceable to the International Federation of Clinical Chemistry and Laboratory Medicine (IFCC) reference material ERM-DA474/IFCC, which serves as a primary ground truth for CRP concentration.
    • Predicate Device Comparison: The performance is compared against a legally marketed predicate device (BN ProSpec CardioPhase hsCRP assay), where the predicate's measurements serve as a comparative standard.
    • Reference Intervals: Expected values for cardiovascular risk prediction are based on established clinical guidelines (Pearson TA et al., 2003).
    • Spiked Samples: For interference studies, known concentrations of interfering substances are added to samples, and the known concentration of the analyte is the ground truth.

    8. The Sample Size for the Training Set:

    Not explicitly stated in the 510(k) summary. For a device like this, the "training set" would refer to the samples used during the development and optimization phase of the assay (e.g., reagent formulation, calibration curve development), rather than a machine learning training set. The approval document focuses on the validation or test sets.

    9. How the Ground Truth for the Training Set Was Established:

    Similar to point 7, the ground truth for potential "training" (development/optimization) would involve analytical standards (like ERM-DA474/IFCC), purified CRP, and well-characterized human serum/plasma samples, often with known CRP concentrations determined by reference methods or gravimetric/volumetric preparation. The goal would be to develop a robust assay that accurately measures CRP across its analytical range.

    Ask a Question

    Ask a specific question about this device

    K Number
    K232624
    Date Cleared
    2023-11-27

    (90 days)

    Product Code
    Regulation Number
    866.5270
    Reference & Predicate Devices
    Why did this record match?
    510k Summary Text (Full-text Search) :

    Marburg, Hessen 3504 Germany

    Re: K232624

    Trade/Device Name: CardioPhase hsCRP Regulation Number: 21 CFR 866.5270
    Regulation Number:
    ------------
    Regulation Number:
    Regulation Number
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    CardioPhase hsCRP is an in-vitro diagnostic reagent for the quantitative determination of C-reactive protein (CRP) in human serum, and heparin and EDTA plasma by means of particle enhanced immunonephelometry using the BN II and BN ProSpec® System. In acute phase response, increased levels of a number of plasma proteins, including C-reactive protein, is observed. Measurement of CRP is useful for the detection and evaluation of infection, tissue injury, inflammatory disorders and associated diseases. High sensitivity CRP (hsCRP) measurements may be used as an independent risk marker for the identification of individuals at risk for future cardiovascular disease. Measurements of hsCRP, when used in conjunction with traditional clinical laboratory evaluation of acute coronary syndromes, may be useful as an independent marker of prognosis for recurrent events, in patients with stable coronary disease or acute coronary syndromes.

    Device Description

    The CardioPhase hsCRP assay is an in-vitro diagnostic reagent for the quantitative determination of Creactive protein (CRP) in human serum, and heparinized and EDTA plasma by means of particleenhanced immunoassay determination. The assay is traceable to the international standard ERM-DA474/IFCC. N Rheumatology Standard SL (cleared under K964527) is used for the establishment of reference curves for the immunonephelometric determination of C-reactive protein on the BN II and BN ProSpec® Systems. This calibrator consists of a mixture of human sera and elevated concentrations of CRP. The CardioPhase hsCRP reagent is a suspension of polystyrene (Latex) particles to which mouse monoclonal anti-human CRP antibodies (

    AI/ML Overview

    Here's a breakdown of the acceptance criteria and the study that proves the device meets them, based on the provided text:

    1. Table of Acceptance Criteria and Reported Device Performance

    The acceptance criteria for the method comparison study were that "Results from each lot of CardioPhase hsCRP met the predefined acceptance criteria." While the specific numerical acceptance criteria (e.g., maximum allowable bias) are not explicitly detailed in a table format within the provided text, the successful outcome is stated, and the resulting performance is presented as follows:

    Performance MetricLot 1 CardioPhase hsCRPLot 2 CardioPhase hsCRPLot 3 CardioPhase hsCRP
    Sample Size (N)119116113
    Range5.523 – 197.746 mg/L5.378 – 199.150 mg/L5.501 – 199.503 mg/L
    Regression Equation (y = mx + b)y = 0.959x + 0.932 mg/Ly = 0.955x + 0.584 mg/Ly = 1.032x - 0.070 mg/L
    Correlation Coefficient (r)0.9940.9960.994
    Coefficient of Determination (r²)0.9890.9910.989
    Observed Max Predicted Bias (for 10, 100, 150, 200 mg/L)5.2% (relative)Not explicitly stated per lot, but given as overall maximum.Not explicitly stated per lot, but given as overall maximum.
    Overall Max Predicted Bias5.2% (relative)5.2% (relative)5.2% (relative)

    2. Sample Sizes Used for the Test Set and Data Provenance

    • Sample Size:
      • Lot 1: N = 119
      • Lot 2: N = 116
      • Lot 3: N = 113
      • The total number of samples used in the method comparison study is the sum of these, which is 348.
    • Data Provenance: The study was conducted at the "company site in Marburg, Germany." The samples used were "Native serum samples." The text does not explicitly state whether the samples were retrospective or prospective, but the phrasing "Native serum samples were measured" suggests they were existing samples at the time of the study rather than collected specifically for this study.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Their Qualifications

    This information is not provided in the text. The study describes a method comparison between two quantitative laboratory assays (CardioPhase hsCRP and RCRP Flex reagent cartridge). For this type of in-vitro diagnostic device, the "ground truth" is typically defined by the reference method or the predicate device's measurement, not by human expert consensus or adjudication in the way it might be for image-based diagnostic AI.

    4. Adjudication Method for the Test Set

    Not applicable for this type of in-vitro diagnostic device and study design. The comparison is quantitative between two analytical methods, not involving human interpretation requiring adjudication.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done

    No, a Multi-Reader Multi-Case (MRMC) comparative effectiveness study was not done. This type of study is typically performed for image analysis or other diagnostic tools where human interpretation is a key component. The CardioPhase hsCRP is an in-vitro diagnostic reagent for quantitative measurement, which does not involve human readers in the same way.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done

    Yes, the study described is a standalone performance study of the CardioPhase hsCRP assay compared to a predicate device. It evaluates the device's ability to quantitatively determine C-reactive protein concentrations independently. No human-in-the-loop component is mentioned for the performance evaluation itself.

    7. The Type of Ground Truth Used

    The "ground truth" for this method comparison study was established by the predicate device, RCRP Flex® reagent cartridge, which runs on the Dimension clinical chemistry system. Both the proposed device (CardioPhase hsCRP) and the predicate device are traceable to the international standard ERM-DA474/IFCC for C-reactive protein measurements. Therefore, the predicate device's measurements serve as the reference for comparison, and that reference itself is traceable to an international standard.

    8. The Sample Size for the Training Set

    The text does not specify a separate training set or its sample size. The described "method comparison study" is focused on verifying the performance of the device for regulatory submission, using a test set of samples. For in-vitro diagnostic devices, "training sets" are usually relevant for developing the assay itself (e.g., optimizing reagent concentrations, reaction conditions), but this information is not typically detailed in a 510(k) summary with respect to a "training set" of patient data for algorithm development.

    9. How the Ground Truth for the Training Set Was Established

    As no specific "training set" of patient samples is described in the provided text in the context of algorithm development, this information is not applicable. The assay itself relies on a biochemical principle and calibration traceable to an international standard (ERM-DA474/IFCC), and the calibration is established using N Rheumatology Standard SL, which is traceable to Siemens internal Master Calibrator, which is in turn directly traceable to ERM-DA474/IFCC.

    Ask a Question

    Ask a specific question about this device

    K Number
    K220977
    Date Cleared
    2023-07-20

    (472 days)

    Product Code
    Regulation Number
    862.1665
    Reference & Predicate Devices
    Why did this record match?
    510k Summary Text (Full-text Search) :

    Class II Classification Name: C-reactive protein immunological test system Regulation Number: 21 CFR § 866.5270

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Beckman Coulter DxC 500 AU Clinical Chemistry Analyzer is an automated chemistry analyzer that measures analytes in samples, in combination with appropriate reagents, calibrators, quality control (QC) material and other accessories. This system is for in vitro diagnostic use only.

    The Glucose test system is for the quantitative measurement of glucose in human serum, plasma, urine and cerebrospinal fluid on Beckman Coulter AU/DxC AU analyzers. Glucose measurements are used in the diagnosis and treatment of carbohydrate metabolism disorders including diabetes mellitus, neonatal hypoglycemia, and of pancreatic islet cell carcinoma.

    System reagent for the quantitative determination of C-Reactive Protein in human serum and plasma on Beckman Coulter AU/DxC AU Analyzers. Measurement of CRP is useful for the detection and evaluation of infection, tissue injury, inflammatory disorders and associated diseases. Measurements may also be useful as an aid in the identification of individuals at risk for future cardiovascular disease. High sensitivity CRP (hsCRP) measurements, when used in conjunction with traditional clinical laboratory evaluation of acute coronary syndromes, may be useful as an independent marker of prognosis for recurrent events, in patients with stable coronary disease or acute coronary syndromes. Reagents for the quantitative determination of Sodium, Potassium and Chloride concentrations in human serum, plasma and urine on the Beckman Coulter ISE modules.

    The sodium test system is intended for the quantitative measurement of sodium in serum, plasma, and urine.

    Measurements obtained by this device are used in the diagnosis and treatment of aldosteronism (excessive secretion of the hormone aldosterone), diabetes insipidus (chronic excretion of dilute urine, accompanied by extreme thirst), adrenal hypertension, Addison's disease (caused by destruction of the adrenal glands), dehydration, inappropriate antidiuretic hormone secretion, or other diseases involving electrolyte imbalance.

    The potassium test system is intended for the quantitative measurement of potassium in serum, plasma, and urine. Measurements obtained by this device are used to monitor electrolyte balance in the diagnosis and treatment of disease conditions characterized by low or high blood potassium levels.

    The chloride test system is intended for the quantitative measurement of the level of chloride in plasma, serum, and urine. Chloride measurements are used in the diagnosis and treatment of electrolyte and metabolic disorders such as cystic fibrosis and diabetic acidosis.

    Device Description

    The Beckman Coulter DxC 500 AU Clinical Chemistry Analyzer carries out automated analysis of serum, plasma, urine samples and other body fluids and automatically generates results. The device is an automated photometric clinical analyzer that measures analytes in samples, in combination with appropriate reagents, calibrators, quality control (QC) material and other accessories. This system is for in vitro diagnostic use only. Electrolyte measurement is performed using a single cell lon Selective Electrode (ISE) which is also common among the other members of the AU family.

    The ISE module for Na+, K+, and Cl- employs crown ether membrane electrodes for sodium and potassium and a molecular oriented PVC membrane for chloride that are specific for each ion of interest in the sample. An electrical potential is developed according to the Nernst Equation for a specific ion. When compared to the Internal Reference Solution, this electrical potential is translated into voltage and then into the ion concentration of the sample.

    In this Beckman Coulter procedure, glucose is phosphorylated by hexokinase (HK) in the presence of adenosine triphosphate (ATP) and magnesium ions to produce glucose-6-phosphate (G-6-P) and adenosine diphosphate (ADP). Glucose-6phosphate dehydrogenase (G6P-DH) specifically oxidizes G-6-P to 6phosphogluconate with the concurrent reduction of nicotinamide adenine dinucleotide (NAD+) to nicotinamide adenine dinucleotide, reduced (NADH). The change in absorbance at 340/660 nm is proportional to the amount of glucose present in the sample.

    The CRP Latex reagent is an in vitro diagnostic device that consists of ready to use buffer and latex particles coated with rabbit anti-CRP antibodies. In this procedure, the measurement of the rate of decrease in light intensity transmitted through particles suspended in solution is the result of complexes formed during the immunological reaction between the CRP of the patient serum and rabbit anti-CRPantibodies coated on latex particles. Two measuring range settings are available: Normal application (CRP Concentrations ranging between 5.0-480 mg/L) and Highly Sensitive (Cardiac) Application- (CRP concentrations ranging between 0.2-80mg/L).

    AI/ML Overview

    This document describes the acceptance criteria and supporting study for the Beckman Coulter DxC 500 AU Clinical Chemistry Analyzer and its associated reagents (Glucose, CRP Latex, ISE Reagents for Sodium, Potassium, and Chloride).

    1. Table of Acceptance Criteria and Reported Device Performance

    The device performance was evaluated across several metrics. The table below summarizes the acceptance criteria (often implied by the "Pass" result and the specific targets within the CLSI guidelines references) and the reported performance for key tests:

    Reagent/ISE & Sample TypeMetricAcceptance Criteria (Implied)Reported PerformanceResult
    hsCRP (Cardiac) (Serum)Method ComparisonSlope: ~1.0; Bias: Low; R: ~1.0Slope: 0.990; Bias: 0.4% at 3mg/L; R: 0.9997Pass
    Linearity
    Ask a Question

    Ask a specific question about this device

    K Number
    K221119
    Date Cleared
    2023-03-17

    (333 days)

    Product Code
    Regulation Number
    866.5270
    Reference & Predicate Devices
    Why did this record match?
    510k Summary Text (Full-text Search) :

    New York 10591

    Re: K221119

    Trade/Device Name: RCRP Flex reagent cartridge Regulation Number: 21 CFR 866.5270
    Product Code: | DCN |
    | Regulation Number: | 21 CFR 866.5270

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The C-Reactive Protein Extended Range (RCRP) method used on the Dimension® clinical chemistry system is an in vitro diagnostic test intended for the quantitative determination of CRP in human serum and plasma (lithium heparin). Measurement of C-Reactive Protein is useful for the detection and evaluation of infection, tissue injury, inflammatory disorders and associated diseases.

    Device Description

    The RCRP method is based on a particle enhanced turbidimetric immunoassay (PETIA) technique. Synthetic particles coated with antibody to C-Reactive Protein (AbPR) aggregate in the presence of C-Reactive Protein in the sample. The increase in turbidity which accompanies aggregation is proportional to the C-Reactive Protein concentration.

    AI/ML Overview

    This document describes the RCRP Flex® reagent cartridge, an in vitro diagnostic test for the quantitative determination of C-Reactive Protein (CRP) in human serum and plasma. The submission is a special 510(k) for a modified device, primarily due to an update in traceability from IFCC CRM 470 to ERM-DA474/IFCC reference material and a change in the analytical measurement range (AMR).

    Here's an analysis of the acceptance criteria and study data based on the provided text:

    1. Table of Acceptance Criteria and Reported Device Performance

    The acceptance criteria and observed performance are provided for the method comparison studies.

    AttributeAcceptance CriteriaReported Device Performance (Modified RCRP vs. Predicate RCRP)Pass/FailReported Device Performance (RCRP Dimension RXL vs. N High Sensitivity CRP)Pass/Fail
    Slope$1.00 \pm 0.1$0.99Pass0.95Pass
    y-intercept$0.0 \pm 2.0$ mg/L-0.5 mg/LPass-1.6 mg/LPass
    Correlation Coefficient (r)$\geq 0.9600$1.000Pass0.997Pass

    Detection Capability (LoB, LoD, LoQ)

    Specimen TypeDetection CapabilityAcceptance CriteriaResult (mg/L / mg/dL)
    Serum and Lithium Heparin PlasmaLoB≤ LoD0.6 mg/L (0.06 mg/dL)
    LoD≤ LoQ1.0 mg/L (0.10 mg/dL)
    LoQ≤ 5.0 mg/L5.0 mg/L (0.50 mg/dL)

    Interference

    Endogenous Substance TestedEndogenous Substance ConcentrationAnalyte ConcentrationAcceptance Criteria (Bias)Bias (%)
    Hemoglobin (hemolysate)[500 mg/dL] 5.0 g/L[11.6 mg/L] 1.16 mg/dLBias exceeding 10% is interference0%
    Bilirubin (Unconjugated)[40 mg/dL] 684 µmol/L[11.7 mg/L] 1.17 mg/dLBias exceeding 10% is interference2%
    Lipemia (Intralipid)[250 mg/dL] 2.5 g/L[11.8 mg/L] 1.18 mg/dLBias exceeding 10% is interference-9%
    Lipemia (Triglyceride Fraction)[750 mg/dL] 7.5 g/L[11.1 mg/L] 1.11 mg/dLBias exceeding 10% is interference-7%

    2. Sample Sample Size Used for the Test Set and Data Provenance

    • Method Comparison – Modified RCRP assay vs Predicate RCRP assay:

      • Sample Size: 132 individual human native serum samples.
      • Data Provenance: Samples were obtained from "specimen vendors". The country of origin is not specified, nor is whether the data is retrospective or prospective.
    • Method Comparison - RCRP assay on Dimension RXL system vs N High Sensitivity CRP on the BN™ System:

      • Sample Size: 171 for the overall comparison (5.3 to 241.3 mg/L) and 39 for the narrower range (5.3 to 20.2 mg/L).
      • Data Provenance: This study involved "re-analyzed historical IFU data." The original provenance (country, retrospective/prospective) of this historical data is not specified.
    • Linearity Testing: Not specified, but generally involves a set of diluted samples or spiked matrix covering the analytical measurement range.

    • Detection Capability (LoB, LoD, LoQ): Not specified.

    • Precision:

      • Sample Size: 6 serum samples, analyzed with N=10 replicates each day for 5 days (total of 50 replicates per sample level).
    • Specimen Equivalency:

      • Sample Size: 73 samples.
      • Data Provenance: Not specified, but likely from specimen vendors similar to the method comparison.
    • Interference:

      • Sample Size: Not explicitly stated, but typically involves a control sample and test samples (with interferent) for assessment of bias.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

    This document describes an in vitro diagnostic (IVD) test, specifically an immunoassay for C-Reactive Protein. The "ground truth" for such devices is typically established through a reference method or known concentrations of certified reference materials, not through expert consensus or interpretation in the same way an imaging AI might.

    • No human experts (e.g., radiologists) were used to establish ground truth for this type of device. The assessment is based on measured concentrations against established reference standards.

    4. Adjudication Method for the Test Set

    Not applicable. As described above, the ground truth for an IVD device like this is based on quantitative measurements and reference materials, not subjective interpretations requiring adjudication.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    Not applicable. This is an in vitro diagnostic assay, not an AI-powered image analysis or diagnostic assist device that would involve human readers.

    6. If a Standalone (i.e. algorithm only without human-in-the loop performance) was done

    This device is a standalone algorithm/reagent system designed for automated quantitative measurement on a clinical chemistry system. Its performance is evaluated intrinsically through various analytical studies (method comparison, linearity, precision, detection capability, interference) without human-in-the-loop performance influencing the measurement itself. The results are then interpreted by clinicians.

    7. The Type of Ground Truth Used

    The ground truth for this device is based on:

    • Reference Materials: For standardization, the device is traceable to ERM-DA474/IFCC reference material (and previously IFCC CRM 470). These are internationally recognized certified reference materials for CRP.
    • Comparative Methods: The performance is benchmarked against a predicate RCRP assay and the N High Sensitivity CRP on the BN™ System. These are established laboratory methods.
    • Clinical Laboratory Standards (CLSI): The studies follow guidelines from CLSI, which define how to robustly evaluate analytical performance parameters like precision, linearity, and detection limits.

    8. The Sample Size for the Training Set

    This document does not describe a machine learning or AI model that requires a distinct "training set" in the conventional sense. The device is a chemical reagent and assay system. Its "training" or development would involve extensive experimentation and optimization during the R&D phase to ensure reagent stability, reaction kinetics, and signal transduction are robust and accurate. This is not typically quantified as a "training set size" like in AI/ML contexts.

    9. How the Ground Truth for the Training Set Was Established

    Not applicable, as this is not an AI/ML device with a defined training set and ground truth in that context. The "ground truth" for the development of such an assay would be through rigorous chemical and biological characterization, using known concentrations of analytes, reference materials, and established analytical chemistry principles.

    Ask a Question

    Ask a specific question about this device

    K Number
    K212559
    Date Cleared
    2022-12-16

    (490 days)

    Product Code
    Regulation Number
    866.5270
    Reference & Predicate Devices
    Why did this record match?
    510k Summary Text (Full-text Search) :

    Marburg, 35041 Germany

    Re: K212559

    Trade/Device Name: CardioPhase® hsCRP Regulation Number: 21 CFR 866.5270
    |
    | Classification Name: | system, test, C-Reactive Protein per 21CFR
    866.5270
    Phase
    hsCRP | C-reactive protein
    immunological test
    system | Class II per
    21CFR
    866.5270

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    CardioPhase® hsCRP is an in-vitro diagnostic reagent for the quantitative determination of C-reactive protein (CRP) in human serum, and heparin and EDTA plasma by means of particle enhanced immunonephelometry using the BN II and BN ProSpec® System. In acute phase response, increased levels of plasma proteins, including C-reactive protein, is observed. Measurement of CRP is useful for the detection and evaluation of infection, tissue injury, inflammatory disorders and associated diseases. High sensitivity CRP (hsCRP) measurements may be used as an independent risk marker for the identification of individuals at risk for future cardiovascular disease. Measurements of hsCRP, when used in conjunction with traditional clinical laboratory evaluation of acute coronary syndromes, may be useful as an independent marker of prognosis for recurrent events, in patients with stable coronary disease or acute coronary syndromes.

    Device Description

    The CardioPhase hsCRP assay is an in vitro diagnostic reagent for the quantitative determination C-reactive protein, in human serum, and heparinized and EDTA plasma by means of particle-enhanced immunoassay determination. Polystyrene particles coated with monoclonal antibodies specific to human CRP are aggregated when mixed with samples containing CRP. These aggregates scatter a beam of light passed through the sample. The intensity of the scattered light is proportional to the concentration of the relevant protein in the sample. The result is evaluated by comparison with a standard of known concentration.

    AI/ML Overview

    This document describes the CardioPhase® hsCRP device, a C-reactive protein immunological test system, and a Special 510(k) submission for a change in its reference standard material from ERM-DA470 to ERM-DA474/IFCC.

    Here's an analysis of the acceptance criteria and the study that proves the device meets them:

    1. Table of Acceptance Criteria and Reported Device Performance

    The document doesn't explicitly state numerical acceptance criteria for all performance characteristics, but rather describes the studies performed and their satisfactory outcomes which imply meeting internal acceptance limits. For instance, for linearity, the stated ranges confirm the measuring range, implying that the observed linearity falls within acceptable deviations. For matrix comparison, high correlation coefficients and slopes close to 1 with intercepts close to 0 indicate acceptable equivalence.

    Performance CharacteristicAcceptance Criteria (Implicit/Explicit)Reported Device Performance
    Detection CapabilitiesLimit of Blank (LoB): Values below the respective Limit of Quantitation (LoQ).
    Limit of Detection (LoD): Greater than LoB and equal to or below LoQ.
    Limit of Quantitation (LoQ): An imprecision goal of less than 20% CV. (Specific numerical LoB/LoD not provided as acceptance criteria, but derived from the study.)LoB: All results measured on blank samples yielded results below the respective LoQ.
    LoD: Calculated parametrically, greater than LoB and equal to or below LoQ.
    LoQ: Set to 0.094 mg/L based on the sample/instrument/reagent lot combination with the highest imprecision observed in the study (
    Ask a Question

    Ask a specific question about this device

    K Number
    K201256
    Date Cleared
    2022-11-04

    (907 days)

    Product Code
    Regulation Number
    866.5270
    Reference & Predicate Devices
    Why did this record match?
    510k Summary Text (Full-text Search) :

    Procise CRP Assay Kit, ProciseDx Instrument, ProciseDx Calibration Cartridge Regulation Number: 21 CFR 866.5270

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Procise CRP assay is a time-resolved fluorescence energy transfer immunoassay for the quantitative determination of C-Reactive Protein (CRP) levels in human serum. The test is carried out by means of the ProciseDx Analyzer.

    Measurement of CRP aids in evaluation of injury to body tissues, inflammatory disorders. The instrument and assay are for use by trained professionals in the clinical laboratory. For in vitro diagnostic use only. Not for point of care use.

    Device Description

    The Procise CRP assay is a homogeneous sandwich immunoassay assay that uses a fluorescence resonance energy transfer (FRET) signal to detect and quantify CRP. FRET is a process in which a donor molecule, in an excited state, transfers excitation energy to an acceptor fluorophore when the two are brought into close proximity. Upon excitation at a characteristic wavelength the energy absorbed by the donor is transferred to the acceptor, which in turn emits light energy. The level of light emitted from the acceptor fluorophore is directly proportional to the degree of donor/acceptor complex formation.

    The Procise CRP assay format is designed as a competitive format. A monoclonal anti-CRP antibody and exogenous CRP antigen are labeled with donor and acceptor fluorophores, respectively. The monoclonal antibody specific for CRP is labelled with the donor fluorophores and the CRP antigen is labelled with the acceptor fluorophore. Similar to other competitive assay formats, as the concentration of CRP increases a proportional decrease in the signal is observed.

    AI/ML Overview

    The Procise CRP Assay Kit, ProciseDx Instrument, and ProciseDx Calibration Cartridge (K201256) is a device for the quantitative determination of C-Reactive Protein (CRP) levels in human serum. It is intended to aid in the evaluation of injury to body tissues, infection, and inflammatory disorders.

    Here's an analysis of the acceptance criteria and the study proving the device meets them:

    1. Table of Acceptance Criteria and Reported Device Performance

    The FDA clearance document does not explicitly state "acceptance criteria" in a quantified table for all performance characteristics. Instead, it presents the results of individual studies and implicitly expects them to demonstrate substantial equivalence to the predicate device and meet typical analytical performance standards for in vitro diagnostic devices. Below is a summary of the performance characteristics obtained from the studies. For qualitative criteria like specificity, "No significant interference" serves as the reported performance, implying it met an internal acceptance threshold. For quantitative metrics like precision, the reported values are the performance.

    Performance CharacteristicAcceptance Criteria (Implicit/Typical for IVD)Reported Device Performance (Procise CRP)
    Precision (Within-Laboratory)Typically low Coefficient of Variation (CV%) for various CRP concentrations. Expected to be within acceptable limits for clinical diagnostic use.Sample 1 (5.9 mg/L): Total SD=0.4, %CV=7.4
    Sample 2 (9.2 mg/L): Total SD=0.8, %CV=8.6
    Sample 3 (35.5 mg/L): Total SD=1.6, %CV=4.4
    Sample 4 (75.1 mg/L): Total SD=4.6, %CV=6.1
    Sample 5 (93.5 mg/L): Total SD=6.4, %CV=6.8
    Sample 6 (125.6 mg/L): Total SD=10.0, %CV=8.0
    Precision (Between-site)Typically low Coefficient of Variation (CV%) across different sites. Expected to be within acceptable limits for clinical diagnostic use.Sample 1 (7.27 mg/L): Total SD=0.6, %CV=7.6
    Sample 2 (25.3 mg/L): Total SD=2.0, %CV=7.8
    Sample 3 (72.7 mg/L): Total SD=6.5, %CV=9.0
    QC1 (5.53 mg/L): Total SD=0.6, %CV=10.0
    QC2 (48.7 mg/L): Total SD=4.5, %CV=9.3
    LinearityLinear regression R² value close to 1, slope close to 1, and intercept close to 0 within the measuring range.Range (3.6 - 161.0 mg/L): Slope = 0.99 (95% CI: 0.97 – 1.00), Intercept = 0.05 (95% CI: -0.94 – 1.05), R² = 1.00
    Analytical Specificity/InterferenceMean bias
    Ask a Question

    Ask a specific question about this device

    K Number
    K192028
    Manufacturer
    Date Cleared
    2020-06-25

    (335 days)

    Product Code
    Regulation Number
    866.5270
    Reference & Predicate Devices
    Why did this record match?
    510k Summary Text (Full-text Search) :

    Montpellier Cedex 4, 34184France

    Re: K192028

    Trade/Device Name: Yumizen C1200 CRP Regulation Number: 21 CFR 866.5270
    | Class II / 510(k) required |
    | Classification Name: | §866.5270

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Yumizen C1200 CRP reagent is intended for the quantitative in vitro diagnostic determination of the C-reactive protein in human serum and lithium heparin plased on an immunoturbidimetric assay. Measurement of C-reactive protein aids in evaluation of the amount of injury to body tissues and for evaluation of infections, tissue injury and inflammatory disorders. This test should be used in conjunction with other laboratory and clinical findings.

    Device Description

    Yumizen C1200 CRP (Licensed for USP6, 248, 597/ USP6, 828, 158 and equivalent patents in other countries) is a latex-enhanced immunoturbidimetric assay developed to accurately measure CRP levels in serum and plasma samples for conventional CRP ranges.

    When an antigen-antibody reaction occurs between CRP in a sample and anti-CRP antibody which has been sensitized to latex particles, agglutination results. This agglutination is detected as an absorbance change, with the magnitude of the change being proportional to the quantity of CRP in the sample. The actual concentration is then determined by interpolation from a calibration curve prepared from calibrators of known concentration.

    Reagents Yumizen C1200 CRP is ready-to-use.

    Reagent 1: Buffer solution: Glycine buffer solution Reagent 2: Latex suspension: 0.20% w/v suspension of latex particles sensitized with anti-CRP antibodies (rabbit)

    After measurements are taken, reagent cassettes should remain in the refrigerated tray.

    Care should be taken not to interchange the caps with others cassettes.

    Reagents with different lot numbers should not be interchanged or mixed.

    This submission consists of the Yumizen C1200 CRP (1300023877) reagent for serum and plasma testing for Yumizen C1200 reagent CRP, the submission includes the controls Yumizen C1200 Level 1 Protein Control (1300023944) and Yumizen C1200 Level 2 Protein Control (1300023945) for use on Yumizen C1200 Analyzer. The submission for Yumizen C1200 reagent CRP also includes the corresponding calibrator Yumizen C1200 CRP Cal (1300023899) for use on Yumizen C1200 Analyzer.

    AI/ML Overview

    The acceptance criteria and study proving the device meets them are detailed below for the Yumizen C1200 CRP reagent.

    1. A table of acceptance criteria and the reported device performance

    Performance CharacteristicAcceptance CriteriaReported Device Performance
    Measuring RangeLimit of Quantitation (LOQ): Not explicitly stated as an acceptance criterion, but determined according to CLSI EP17-A2.LOQ: 5 mg/L (Serum)
    Linearity: Not explicitly stated as an acceptance criterion, but evaluated according to CLSI EP06-A. Range should cover desirable range and extend to lowest and highest ends.Linearity Range: 9.42 to 150.78 mg/L (Serum)
    Measuring Range: Not explicitly stated as an acceptance criterion, but based on LOQ and linearity studies.Measuring Range: 5.0 to 160 mg/L (until 800 mg/L with post-dilution)
    Accuracy and PrecisionWithin-run CV limits: Low level: 9.0%, Middle level: 4.5%, High level: 3.8%Within-run CV: All reported values for samples and controls are well below the limits (e.g., 0.8% - 1.8%)
    Total precision CV limits: Low level: 12.0%, Middle level: 6.0%, High level: 5.0%Total Precision CV: All reported values for samples and controls are well below the limits (e.g., 1.5% - 2.9%)
    InterferencesAcceptable bias is defined at +/-10% of the value without interfering substances.Highest reported values for various interferents show no interference higher than 10%.
    Matrix ComparisonNot explicitly stated as an acceptance criterion, but results should show no significant difference between serum and plasma with heparin specimens.Correlation (R) of 0.996 and slope (0.8973 – 1.007) and intercept (-0.1611 – +0.6459) for 38 samples; concluded "no significative difference."
    Method ComparisonNot explicitly stated as an acceptance criterion, but evaluated using NCCLS (CLSI) EP-9A3.Correlation (R2) of 0.998 and slope (0.9680 – 0.9976) and intercept (-0.1357 – +0.6287) for 102 samples.
    Reagent StabilityClosed stability: Stable up to the expiry date on the label if stored at 2-10°C.Shelf Life: 24 months.
    Open stability (on-board): Not explicitly stated as an acceptance criterion, but assessed.On-board stability: 8 weeks.
    Reference RangeVerification studies should support established reference ranges in literature for adults: 20-60 years
    Ask a Question

    Ask a specific question about this device

    K Number
    K192072
    Date Cleared
    2020-02-21

    (203 days)

    Product Code
    Regulation Number
    866.5270
    Reference & Predicate Devices
    Why did this record match?
    510k Summary Text (Full-text Search) :

    Indiana 46250

    Re: K192072

    Trade/Device Name: Tina-quant C-Reactive Protein IV Regulation Number: 21 CFR 866.5270
    |
    | Product Codes,
    Regulation Numbers | DCN, 21 CFR § 866.5270

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Tina-quant® C-Reactive Protein IV is an immunoturbidimetric assay for the in vitro quantitative determination of CRP in human serum and plasma on cobas c systems.

    A C-reactive protein immunological test system is a device that consists of the reagents used to measure by immunochemical techniques the C-reactive protein in serum and plasma. Measurement of C-reactive protein aids in evaluation of the amount of injury to body tissues.

    Device Description

    The Tina-quant® C-Reactive Protein IV reagent will be a liquid ready to use 2 component particle enhanced immunoturbidimetric assay.

    Reagents - working solutions

    R1: TRIS* buffer with bovine serum albumin; preservatives

    R2 Latex particles coated with anti-CRP (mouse) in glycine buffer; immunoglobulins (mouse); preservative

    • TRIS= Tris(hydroxymethyl)-aminomethane

    The Tina-quant® C-Reactive Protein IV assay will be based on the DUREL technology (dual radius enhanced latex - technology) which is also used in C-Reactive Protein Gen.3 predicate method. Human CRP agglutinates with latex particles coated with monoclonal anti-CRP antibodies. The aggregates are determined turbidimetrically.

    AI/ML Overview

    The document describes the non-clinical performance evaluation of the Tina-quant® C-Reactive Protein IV device. Here's a summary of the acceptance criteria and the study that proves the device meets them:

    1. A table of acceptance criteria and the reported device performance

    Performance CharacteristicAcceptance Criteria or Expected Outcome (or Predicate Performance)Reported Device Performance (Tina-quant® C-Reactive Protein IV)
    PrecisionRepeatability (CV%)Repeatability (CV%)
    Precinorm ProteinPredicate: 1.2%1.3%
    Precipath ProteinPredicate: 1.3%1.6%
    Human Serum SamplesPredicate: 3.6%, 1.6%, 1.2% (for different samples)1.5% (Serum 2), 1.6% (Serum 3), 2.2% (Serum 4), 2.0% (Serum 5), 1.3% (Serum 6)
    Intermediate Precision (CV%)Intermediate Precision (CV%)Intermediate Precision (CV%)
    Precinorm ProteinPredicate: 2.9%1.5%
    Precipath ProteinPredicate: 1.9%1.9%
    Human Serum SamplesPredicate: 11.1%, 3.9%, 1.7% (for different samples)1.6% (Serum 2), 1.9% (Serum 3), 2.4% (Serum 4), 2.4% (Serum 5), 1.8% (Serum 6)
    Analytical Sensitivity
    Limit of Blank (LoB)Claimed LoB: 0.2 mg/LObserved LoB: 0.0700 mg/L (Meets criterion: Observed is less than claimed)
    Limit of Detection (LoD)Claimed LoD: 0.3 mg/LObserved LoD: 0.137 mg/L (Meets criterion: Observed is less than claimed)
    Limit of Quantitation (LoQ)Claimed LoQ: 3 mg/L (Predicate was 0.6 mg/L, but 3 mg/L is claimed for the candidate device. The study shows the device is capable of quantifying lower.)Observed LoQ: 0.313 mg/L (Exceeds criterion as it is much lower than the claimed 3 mg/L, indicating better performance. The claimed measuring range for Tina-quant® C-Reactive Protein IV starts at 3 mg/L, so the LoQ of 0.313 mg/L is well below the lowest reportable value, which is acceptable.)
    Linearity/Assay Reportable RangeConsistent with predicate, good correlation (R > 0.999), and linearity across the intended measuring range.Linear Regression: y=1.002x-0.0169; Pearson correlation coefficient (R)=0.9994. Claimed Measuring Range: 3 to 350 mg/L. (Meets criterion)
    Endogenous InterferenceNo interference up to specified levels for lipemia, hemolysis, bilirubin, ditauro bilirubin, albumin, IgG, and RF Factor.No interference up to: Lipemia: 1000 L Index; Hemolysis: 1000 H Index; Bilirubin: 60 I Index; Ditauro Bilirubin: 60 I Index; Albumin: 60 g/L; IgG: 50 g/L; RF Factor: 1200 IU/mL. (Meets criterion)
    Exogenous Interference (Drugs)No interference up to specified drug concentrations.No interference up to: N-Acetylcysteine: 1660 mg/L; Ampicillin-Na: 1000 mg/L; Ascorbic acid: 300 mg/L; Cefoxitin: 6600 mg/L; Heparin: 5000 IU/L; Levodopa: 20 mg/L; Methyldopa: 22.5 mg/L; Metronidazole: 200 mg/L; Doxycyclin: 50 mg/L; Acetylsalicylic acid: 1000 mg/L; Rifampicin: 60 mg/L; Ticarcillin: 225 mg/L; Penicillamin: 24 mg/L; Phenylbutazone: 400 mg/L; Cyclosporine: 5 mg/L; Acetaminophen: 200 mg/L; Ibuprofen: 500 mg/L; Theophylline: 100 mg/L. (Meets criterion)
    Sample Matrix ComparisonGood agreement between serum and plasma types (Li-Heparin, K2-EDTA, K3-EDTA) with correlation (r) close to 1 and slope near 1, intercept near 0.Linear Regression: Serum vs. Li-Heparin: y = 1.029x - 0.192, r = 0.999; Serum vs. K2-EDTA: y = 1.024x - 0.201, r = 0.999; Serum vs. K3-EDTA: y = 1.024x - 0.258, r = 0.999. (Meets criterion)
    Method Comparison to PredicateStrong correlation with the predicate device (R > 0.999) and slope near 1, intercept near 0.Passing/Bablok Regression: Slope = 0.985, Intercept = +0.278, Correlation (Pearson) = 0.999. Weighted Deming Regression: Slope = 0.979, Intercept = +0.296, Correlation = 0.999. (Meets criterion for substantial equivalence)
    StabilityMeet Roche Diagnostic's claims on package labeling.Data supports Roche Diagnostic's claims. (Meets criterion)

    Study Details:

    The performance studies were designed in accordance with CLSI guidelines (EP5-A3, EP17-A2, EP6-A) and FDA guidance for CRP assays.

    2. Sample sized used for the test set and the data provenance:

    • Precision (Repeatability & Intermediate Precision):
      • Controls: Precinorm Protein and Precipath Protein.
      • Human Serum Samples: 4 human serum samples were used to evaluate repeatability and intermediate precision, plus 6 human serum samples for the specific results documented in the table. Each sample/control was analyzed with multiple aliquots (e.g., 2 aliquots for 21 days for precision).
      • Data Provenance: Not explicitly stated (e.g., country of origin), but it's noted as "human serum samples." The studies are prospective for the device evaluation.
    • Analytical Sensitivity (LoB, LoD, LoQ):
      • LoB: Ten aliquots of analyte-free saline (N=60 determinations per lot).
      • LoD: Five samples of low analyte level human serum (N=60 determinations per lot).
      • LoQ: Nine low concentration human serum samples (N=25 determinations per sample per lot).
      • Data Provenance: Not explicitly stated for specific origin, but uses "analyte free saline" and "human serum."
    • Linearity/Assay Reportable Range:
      • Samples: Dilution series prepared from native unmodified human serum sample pools, resulting in 15 levels. Each level measured in triplicate (n ≥ 3).
      • Data Provenance: "native unmodified human serum sample pools."
    • Endogenous Interference:
      • Samples: Effects determined at 2 CRP levels (5-10 mg/L and 35-100 mg/L) utilizing 11-level serial dilution sets of interfering substances. Each level measured in triplicate.
      • Data Provenance: Not explicitly stated, likely lab-prepared samples with added interferents.
    • Exogenous Interference (Drugs):
      • Samples: Effects determined at 2 CRP levels (5-10 mg/L and 35-100 mg/L) using pooled samples spiked with drugs. N=5 results per drug.
      • Data Provenance: Lab-prepared samples with added drugs.
    • Sample Matrix Comparison:
      • Samples: Native samples from single donors drawn into serum, Li-Heparin, K2-EDTA, and K3-EDTA plasma primary tubes.
      • Range Tested: 3.34 to 344 mg/L.
      • Data Provenance: "native samples (single donors)."
    • Method Comparison to Predicate:
      • Samples: One hundred ten (N=110) native, unaltered serum samples.
      • Data Provenance: "native, unaltered serum samples."

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g., radiologist with 10 years of experience)

    This device is an in-vitro diagnostic (IVD) assay for quantitative determination of C-Reactive Protein (CRP). The "ground truth" for such devices is established by reference methods, certified reference materials, and accepted analytical techniques, not by expert interpretation in the same way imaging studies might use radiologists. The measurements for the test set samples (e.g., CRP concentrations) are determined by the analytical methods themselves, often verified against established standards or predicate devices. There is no mention of human experts establishing ground truth for the test set in the context of interpretation.

    4. Adjudication method (e.g., 2+1, 3+1, none) for the test set

    For IVD assays, ground truth is typically analytical, not based on expert adjudication of observations. The "truth" is the measured concentration, often confirmed by multiple replicates or comparative methods, as opposed to a consensus reading process found in clinical imaging studies. Therefore, no adjudication method (like 2+1 or 3+1) is applicable or described here.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    No, an MRMC comparative effectiveness study was not performed. This device is an in-vitro diagnostic instrument for measuring a biomarker (CRP). It does not involve human readers interpreting images or results that would be "assisted" by AI. The device directly produces quantitative results.

    6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done

    Yes, the performance studies described are for the device (Tina-quant® C-Reactive Protein IV on a cobas c 501 analyzer) operating in a standalone capacity. It measures CRP in samples without human intervention in the interpretive output. The studies evaluate the analytical performance of the instrument and reagents themselves.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

    The ground truth for this IVD device is based on:

    • Analytical Standards: For LoB, LoD, LoQ, and Linearity, ground truth is established through statistical derivation from measurements of samples with known low or varying concentrations, or through the use of analyte-free samples or reference materials.
    • Predicate Device Comparison: The "Method Comparison to Predicate" establishes the ground truth by comparing the device's results against a legally marketed and established predicate device (Roche Diagnostics C-Reactive Protein Gen.3), assuming the predicate provides the accepted reference values.
    • Spiked Samples/Known Concentrations: For interference studies (endogenous and exogenous), ground truth is established by adding known amounts of interfering substances or drugs to samples with known CRP concentrations and observing the recovery of the expected CRP value.
    • Certified Reference Materials (CRM): The "Traceability/Standardization" indicates that the candidate device is standardized against the certified reference material of IRMM (Institute for Reference Materials and Measurements) ERM-DA474/IFCC. This is a primary source of analytical ground truth.

    8. The sample size for the training set

    This document describes a premarket notification (510(k)) submission for a medical device. The "training set" concept is typically associated with machine learning or AI models. This device is an immunoturbidimetric assay, which is a traditional analytical chemistry method, not an AI/ML system. Therefore, there is no "training set" in the context of machine learning. The studies described are performance evaluations to demonstrate that the device meets its analytical specifications.

    9. How the ground truth for the training set was established

    As there is no "training set" for an AI/ML model for this traditional IVD assay, this question is not applicable. The device's performance is validated against analytical performance criteria as described above in point 7.

    Ask a Question

    Ask a specific question about this device

    K Number
    K192118
    Device Name
    CRP Vario
    Manufacturer
    Date Cleared
    2019-11-08

    (94 days)

    Product Code
    Regulation Number
    866.5270
    Reference & Predicate Devices
    Why did this record match?
    510k Summary Text (Full-text Search) :

    Robert Koch, 2 Milano, 20152 It

    Re: K192118

    Trade/Device Name: CRP Vario Regulation Number: 21 CFR 866.5270
    Classification Name | C-Reactive Protein, Antigen, Antiserum, and Control
    Device, (21CFR 866.5270

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The CRP Vario assay [CRPVa] is for in vitro diagnostic use in the quantitative immunoturbidimetric determination of C-reactive protein in human serum and plasma (sodium and lithium heparin) using the ARCHITECT c Systems. Measurement of C-reactive protein is useful in the detection and evaluation of infection, tissue injury and inflammatory disorders.

    CRP Calibrators (including CRP Calibrator Set, CRP Calibrator HS and CRP Calibrator WR) are intended to be used for the calibration of the CRP Vario for the quantitative determination of C-reactive protein in human serum or plasma samples.

    Device Description

    The CRP Vario assay is intended for the quantitative immunoturbidimetric determination of C-reactive protein in human serum and plasma. It is supplied as a two-reagent kit. The kit contains Reagent 1 (Glycine buffer) and Reagent 2 (Anti-CRP polyclonal antibodies adsorbed on latex particles). Two different sizes of the product are available. The submission also describes CRP Calibrators (CRP Calibrator Set, CRP Calibrator HS, and CRP Calibrator WR) which are prepared by diluting CRP with human serum and stabilized by adding sodium azide.

    AI/ML Overview

    Here's an analysis of the acceptance criteria and study as described in the provided document:

    Acceptance Criteria and Device Performance for CRP Vario

    1. Table of Acceptance Criteria and Reported Device Performance

    Note: The document explicitly refers to an "Acceptance Criteria" column for Method Comparison in Table 15-3. For other performance characteristics, the "acceptance criteria" can be inferred by comparing the candidate device's stated performance to that of the predicate device, or by general statements about meeting recommendations or consistency with original submissions.

    FeatureAcceptance CriteriaReported Device Performance (Candidate Device CRP Vario)Met?
    Method Comparison
    High Sensitivity Method (Slope)0.95 – 1.050.969 (0.964 – 0.974)Yes
    High Sensitivity Method (R)≥ 0.9751.000Yes
    High Sensitivity Method (n)≥ 100111Yes
    Standard Method (Slope)0.95 – 1.050.956 (0.925 – 0.997)Yes
    Standard Method (R)≥ 0.9751.000Yes
    Standard Method (n)≥ 100119Yes
    Wide Range Method (Slope)0.95 – 1.050.976 (0.944 – 0.993)Yes
    Wide Range Method (R)≥ 0.9751.000Yes
    Wide Range Method (n)≥ 100119Yes
    Limit of Quantitation (LoQ)
    LOQ for hsCRP≤ 0.03 mg/dL"The LOQ results are consistent with the original submission." (Original submission stated 0.01 mg/dL for High Sensitivity Method, which is the hsCRP method).Yes
    LOQ for Standard and Wide Range Methods≤ 0.05 mg/dL"The LOQ results are consistent with the original submission." (Original submission stated 0.02 mg/dL for Standard and Wide Range Methods).Yes
    Linearity
    High Sensitivity MethodMeets pre-specified allowable error criteria against 2nd and 3rd order polynomial fits.Found to extend up to 16.00 mg/dL.Yes
    Standard MethodMeets pre-specified allowable error criteria against 2nd and 3rd order polynomial fits.Found to extend up to 32.00 mg/dL.Yes
    Wide Range MethodMeets pre-specified allowable error criteria against 2nd and 3rd order polynomial fits.Found to extend up to 48.00 mg/dL.Yes

    2. Sample Size Used for the Test Set and Data Provenance

    • Method Comparison Test Set:

      • High Sensitivity Method: N = 111 human serum samples
      • Standard Method: N = 119 human serum samples
      • Wide Range Method: N = 119 human serum samples
      • Data Provenance: The document states "Human serum samples, spanning the measuring range, were tested." No information is provided regarding the country of origin or whether the data was retrospective or prospective.
    • Limit of Quantitation (LoQ) Test Set:

      • The study used serum pools diluted with SeraSub (a synthetic serum) to create 8 low-level analyte samples (0.08, 0.06, 0.05, 0.04, 0.03, 0.02, 0.01, and 0.005 mg/dL).
      • Each sample was tested in a minimum of 10 replicates per run, with 2 runs per day over 3 days, resulting in a total of 60 replicates per sample.
    • Linearity Test Set:

      • A high CRP serum pool was mixed with a low serum pool to generate 12 concentrations.
      • Each concentration level was tested in triplicate determinations.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Their Qualifications

    This section is not applicable to this type of device and study. The CRP Vario is an in vitro diagnostic device that quantifies C-reactive protein. Its performance is evaluated against reference methods (e.g., predicate devices, established analytical protocols) and CLSI guidelines, not against expert human interpretations of images or clinical reports. There are no "experts" establishing a subjective ground truth in the way described for AI in image analysis.

    4. Adjudication Method for the Test Set

    This section is not applicable. As an in vitro diagnostic device, the performance is determined by quantitative measurements and statistical comparisons with reference methods, not by adjudication of interpretations by multiple human readers.

    5. If a Multi Reader Multi Case (MRMC) Comparative Effectiveness Study Was Done

    No, an MRMC comparative effectiveness study was not done. This type of study evaluates human reader performance often in the context of imaging diagnostics. The CRP Vario is an in vitro diagnostic for laboratory analysis, not a device requiring human interpretation in this manner.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done

    Yes, the studies reported here (Method Comparison, Linearity, Limit of Quantitation) are standalone performance evaluations of the CRP Vario assay. The device itself is an automated immunoturbidimetric system (ARCHITECT c8000 system) for quantitative measurement, operating without human interpretation of the final result, beyond standard laboratory quality control and result reporting.

    7. The Type of Ground Truth Used

    The "ground truth" for the performance evaluation of the CRP Vario was established by:

    • Reference Methods / Predicate Device: For the Method Comparison study, the candidate CRP Vario (traceable to ERM-DA472/IFCC) was compared against a Beckman Coulter CRP Latex REF. OSR6199 (traceable to CRM 470) on an AU5800 platform.
    • CLSI Guidelines: Formal CLSI (Clinical and Laboratory Standards Institute) protocols (EP09-A3 for Method Comparison, EP06-A for Linearity, EP17-A2 for LoQ) were followed, indicating that the acceptable statistical and analytical measures defined by these standards served as the basis for evaluating performance.
    • Dilution Series / Known Concentrations: For Linearity and LoQ, the "ground truth" was derived from precisely prepared serum pools and their serial dilutions to known or targeted concentrations.

    8. The Sample Size for the Training Set

    This section is not applicable. The CRP Vario is a re-submission for a modification of an existing in vitro diagnostic device, not an AI/ML device that requires a distinct "training set" in the computational sense. Its performance is characterized through analytical verification and validation studies.

    9. How the Ground Truth for the Training Set Was Established

    This section is not applicable for the reasons stated in point 8.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 10