510(k) Data Aggregation

    K Number
    K232624
    Date Cleared
    2023-11-27

    (90 days)

    Product Code
    Regulation Number
    866.5270
    Reference & Predicate Devices
    Why did this record match?
    Reference Devices :

    K964527, K943997, K001647, K212559

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    CardioPhase hsCRP is an in-vitro diagnostic reagent for the quantitative determination of C-reactive protein (CRP) in human serum, and heparin and EDTA plasma by means of particle enhanced immunonephelometry using the BN II and BN ProSpec® System. In acute phase response, increased levels of a number of plasma proteins, including C-reactive protein, is observed. Measurement of CRP is useful for the detection and evaluation of infection, tissue injury, inflammatory disorders and associated diseases. High sensitivity CRP (hsCRP) measurements may be used as an independent risk marker for the identification of individuals at risk for future cardiovascular disease. Measurements of hsCRP, when used in conjunction with traditional clinical laboratory evaluation of acute coronary syndromes, may be useful as an independent marker of prognosis for recurrent events, in patients with stable coronary disease or acute coronary syndromes.

    Device Description

    The CardioPhase hsCRP assay is an in-vitro diagnostic reagent for the quantitative determination of Creactive protein (CRP) in human serum, and heparinized and EDTA plasma by means of particleenhanced immunoassay determination. The assay is traceable to the international standard ERM-DA474/IFCC. N Rheumatology Standard SL (cleared under K964527) is used for the establishment of reference curves for the immunonephelometric determination of C-reactive protein on the BN II and BN ProSpec® Systems. This calibrator consists of a mixture of human sera and elevated concentrations of CRP. The CardioPhase hsCRP reagent is a suspension of polystyrene (Latex) particles to which mouse monoclonal anti-human CRP antibodies (

    AI/ML Overview

    Here's a breakdown of the acceptance criteria and the study that proves the device meets them, based on the provided text:

    1. Table of Acceptance Criteria and Reported Device Performance

    The acceptance criteria for the method comparison study were that "Results from each lot of CardioPhase hsCRP met the predefined acceptance criteria." While the specific numerical acceptance criteria (e.g., maximum allowable bias) are not explicitly detailed in a table format within the provided text, the successful outcome is stated, and the resulting performance is presented as follows:

    Performance MetricLot 1 CardioPhase hsCRPLot 2 CardioPhase hsCRPLot 3 CardioPhase hsCRP
    Sample Size (N)119116113
    Range5.523 – 197.746 mg/L5.378 – 199.150 mg/L5.501 – 199.503 mg/L
    Regression Equation (y = mx + b)y = 0.959x + 0.932 mg/Ly = 0.955x + 0.584 mg/Ly = 1.032x - 0.070 mg/L
    Correlation Coefficient (r)0.9940.9960.994
    Coefficient of Determination (r²)0.9890.9910.989
    Observed Max Predicted Bias (for 10, 100, 150, 200 mg/L)5.2% (relative)Not explicitly stated per lot, but given as overall maximum.Not explicitly stated per lot, but given as overall maximum.
    Overall Max Predicted Bias5.2% (relative)5.2% (relative)5.2% (relative)

    2. Sample Sizes Used for the Test Set and Data Provenance

    • Sample Size:
      • Lot 1: N = 119
      • Lot 2: N = 116
      • Lot 3: N = 113
      • The total number of samples used in the method comparison study is the sum of these, which is 348.
    • Data Provenance: The study was conducted at the "company site in Marburg, Germany." The samples used were "Native serum samples." The text does not explicitly state whether the samples were retrospective or prospective, but the phrasing "Native serum samples were measured" suggests they were existing samples at the time of the study rather than collected specifically for this study.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Their Qualifications

    This information is not provided in the text. The study describes a method comparison between two quantitative laboratory assays (CardioPhase hsCRP and RCRP Flex reagent cartridge). For this type of in-vitro diagnostic device, the "ground truth" is typically defined by the reference method or the predicate device's measurement, not by human expert consensus or adjudication in the way it might be for image-based diagnostic AI.

    4. Adjudication Method for the Test Set

    Not applicable for this type of in-vitro diagnostic device and study design. The comparison is quantitative between two analytical methods, not involving human interpretation requiring adjudication.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done

    No, a Multi-Reader Multi-Case (MRMC) comparative effectiveness study was not done. This type of study is typically performed for image analysis or other diagnostic tools where human interpretation is a key component. The CardioPhase hsCRP is an in-vitro diagnostic reagent for quantitative measurement, which does not involve human readers in the same way.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done

    Yes, the study described is a standalone performance study of the CardioPhase hsCRP assay compared to a predicate device. It evaluates the device's ability to quantitatively determine C-reactive protein concentrations independently. No human-in-the-loop component is mentioned for the performance evaluation itself.

    7. The Type of Ground Truth Used

    The "ground truth" for this method comparison study was established by the predicate device, RCRP Flex® reagent cartridge, which runs on the Dimension clinical chemistry system. Both the proposed device (CardioPhase hsCRP) and the predicate device are traceable to the international standard ERM-DA474/IFCC for C-reactive protein measurements. Therefore, the predicate device's measurements serve as the reference for comparison, and that reference itself is traceable to an international standard.

    8. The Sample Size for the Training Set

    The text does not specify a separate training set or its sample size. The described "method comparison study" is focused on verifying the performance of the device for regulatory submission, using a test set of samples. For in-vitro diagnostic devices, "training sets" are usually relevant for developing the assay itself (e.g., optimizing reagent concentrations, reaction conditions), but this information is not typically detailed in a 510(k) summary with respect to a "training set" of patient data for algorithm development.

    9. How the Ground Truth for the Training Set Was Established

    As no specific "training set" of patient samples is described in the provided text in the context of algorithm development, this information is not applicable. The assay itself relies on a biochemical principle and calibration traceable to an international standard (ERM-DA474/IFCC), and the calibration is established using N Rheumatology Standard SL, which is traceable to Siemens internal Master Calibrator, which is in turn directly traceable to ERM-DA474/IFCC.