K Number
K242170
Date Cleared
2025-04-18

(268 days)

Product Code
Regulation Number
866.5270
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

K-ASSAY® CRP (Ver.2) is intended to be used for the quantitative determination of C-reactive protein (CRP) in human serum and plasma (potassium-EDTA or lithium-heparin) by immunoturbidimetric assay. Measurement of CRP aids in the detection and evaluation of infection, tissue injury, inflammatory disorders and associated diseases. FOR IN VITRO DIAGNOSTIC USE.

Device Description

The K-ASSAY® CRP (Ver.2) assay quantifies C-reactive protein based on immunoturbidimetric assay. The reagent uses latex combined with goat polyclonal antibody specific to human CRP. By adsorbing CRP in the sample to the surface of the latex particles and reacting it with the anti-CRP antibody, specific aggregation corresponding to the CRP concentration occurs. Since the absorbance of the reaction changes in proportion to the amount of aggregation, the concentration of CRP in the sample is determined based on the calibration curve prepared using a standard of known CRP concentrations. The K-ASSAY® CRP (Ver.2) assay can be run using a chemistry analyzer. 6 levels of calibrators from the K-ASSAY® CRP Calibrator (Ver.2) calibrators are used for quantifying the levels of CRP present in the patient's sample.

AI/ML Overview

This document describes the FDA 510(k) clearance for the K-ASSAY CRP (Ver.2) IVD device. This device is an in-vitro diagnostic test system, which means it analyzes biological samples in a lab setting rather than directly interacting with a patient.

Therefore, the concepts of "human readers," "AI assistance," "effect size," "multi reader multi case (MRMC) comparative effectiveness study," and "standalone (i.e. algorithm only without human-in-the loop performance)" are not applicable in this context. These terms are typically used for medical imaging AI/ML devices where human interpretation is involved.

For this in-vitro diagnostic device, the "acceptance criteria" and "study that proves the device meets the acceptance criteria" are related to its analytical performance characteristics when compared to a predicate device, and the reported device performance refers to the results of these analytical studies.

Here's the breakdown of the information provided within the scope of this IVD device:

1. Table of Acceptance Criteria and Reported Device Performance

For an IVD device like the K-ASSAY® CRP (Ver.2), the acceptance criteria are generally established by demonstrating equivalent or superior analytical performance compared to a legally marketed predicate device, and meeting established CLSI guidelines for accuracy, precision, linearity, and interference. The "reported device performance" refers to the actual study results for these characteristics.

Acceptance Criteria (Implicit from predicate comparison and CLSI guidelines)Reported Device Performance (K-ASSAY® CRP (Ver.2))
Intended Use Equivalence: Quantitative determination of CRP in human serum and plasma for detection/evaluation of infection, tissue injury, inflammatory disorders.K-ASSAY® CRP (Ver.2) is intended for the quantitative determination of CRP in human serum and plasma (potassium-EDTA or lithium-heparin) by immunoturbidimetric assay. Measurement of CRP aids in the detection and evaluation of infection, tissue injury, inflammatory disorders and associated diseases.
Methodology Equivalence: Latex-enhanced immunoturbidimetric assay.Latex-enhanced (immuno)turbidimetric assay.
Calibration Equivalence/Validation: Appropriate calibrator levels for the intended range.6 levels of calibrators (0.0, 10.0, 50.0, 150.0, 300.0 and 400.0 mg/L).
Assay Range (Comparable to predicate): Demonstrates clinically relevant measuring range.Claimed Assay Range: 5.0 - 400.0 mg/L. (Predicate: 0.2 – 480 mg/L)
Precision: Demonstrate acceptable repeatability, within-run, between-run, within-day, between-day, within-laboratory, between-site, and reproducibility CVs as per CLSI EP05-A3 guidelines.Single Site Precision (Combined Data From 3 Lots):- Within-Run CV: 0.8% to 2.2%- Between-Run CV: 0.0% to 0.6%- Within-Lot CV: 0.9% to 1.9%- Between-Lot CV: 0.0% to 1.2%- Total CV: 0.9% to 2.3%Multisite Precision (Combined Data From 3 Analyzers):- Repeatability CV: 0.5% to 1.5%- Between-Run CV: 0.0% to 1.1%- Between-Day CV: 0.6% to 1.8%- Between-Site CV: 0.9% to 1.8%- Reproducibility CV: 1.1% to 2.5%
Interference: No significant interference from common endogenous and exogenous substances (recovery within 10% of initial value).Endogenous Substances (up to listed concentrations, no significant interference): Bilirubin C (40 mg/dL), Bilirubin F (40 mg/dL), Cholesterol (300 mg/dL), Hemoglobin (1,000 mg/dL), Intralipid (500 mg/dL), Rheumatoid Factor (1,000 IU/mL), Triglycerides (1,000 mg/dL).Exogenous Substances (up to listed concentrations, no significant interference): Acetaminophen (1.5 mM), Amoxicillin (400 µmol/L), Aspirin (3.6 mM), Cephalexin (360 µmol/L), Fluconazole (480 µmol/L), Ibuprofen (2.5 mg/dL), Methotrexate (1,400 µmol/L), Prednisolone (2 µmol/L), Vitamin C (500 mg/L).
Method Comparison: Strong correlation and minimal bias against the predicate device.Regression Equation: y = 1.005x - 0.002, r = 0.999 (n = 175 clinical native serum samples), where y = K-ASSAY® CRP (Ver.2), x = predicate device.
Linearity: Demonstrates linearity across the claimed assay range.Regression Equation: y = 0.9709x - 1.095, r = 0.999 (tested range: 4.6 - 441.2 mg/L).
Limit of Quantitation (LoQ): Clinically acceptable LoQ (within-laboratory precision ≤ 20% CV).Reported LoQ: 1.0 mg/L (highest observed across 3 reagent lots). Claimed LoQ: 5.0 mg/L.
Matrix Comparison: No significant difference in results across different sample matrices (serum vs. plasma).K2-EDTA Plasma vs Serum: y = 1.007x - 0.141, r = 0.999Li-Heparin Plasma vs Serum: y = 0.972x + 0.074, r = 0.999
Expected Values (Verification): Distribution in normal population consistent with clinical literature.168 normal U.S. serum samples tested; 4 out of 168 samples (>5.0 mg/L, 2.4%) consistent with literature (≤ 5 mg/L indicates apparently healthy).

2. Sample Sizes Used for the Test Set and Data Provenance

  • Method Comparison: n = 175 clinical native serum samples.
  • Linearity: The number of samples/dilutions used is not explicitly stated, but the tested range was 4.6 - 441.2 mg/L.
  • Limit of Quantitation (LoQ): Not specified in terms of sample size for the LoQ determination itself (typically involves low-concentration samples measured multiple times).
  • Matrix Comparison: 42 donor samples (each collected into 3 different tubes: serum, K2-EDTA plasma, Li-Heparin plasma).
  • Precision (Single-site): For each of the 7 samples/controls, N=240 individual measurements (2 runs/day x 2 replicates/run x 20 days x 3 lots).
  • Precision (Multisite): For each of the 7 samples/controls, N=75 individual measurements (1 run/day x 5 replicates/run x 5 days x 3 analyzers).
  • Interference: The number of unique samples tested for interference is not explicitly stated. Typically, a few samples (e.g., low, medium, high concentration) are spiked with interferents and compared to unspiked controls.
  • Expected Values: 168 normal serum samples.
  • Data Provenance: The document states for "Expected Values" that 168 normal serum samples were "taken from healthy individuals in the U.S." This indicates a U.S. origin for at least this specific study. For other studies (Method Comparison, Precision, Linearity, Interference, Matrix Comparison), the specific country of origin is not mentioned. All studies are retrospectively analyzed in the sense that they were completed performance validation studies submitted to FDA.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

For an in-vitro diagnostic device like this CRP assay, the "ground truth" is established through highly accurate and precise reference methods or established predicate devices, adhering to rigorous analytical chemistry best practices and guidelines (e.g., CLSI standards). There isn't a panel of human "experts" like radiologists interpreting images. The "ground truth" is quantitative and objective, derived from reference measurements.

In this case:

  • Method Comparison: The predicate device (K-ASSAY® CRP (3), K023828) serves as the comparator for method comparison, which represents the established "ground truth" for clinical samples.
  • Linearity, LoQ, Precision, Interference, Matrix Comparison: The "ground truth" for these analytical performance studies is established by rigorous laboratory protocols, highly calibrated equipment, reference materials, and adherence to CLSI guidelines. The performance is assessed against predefined statistical criteria rather than expert consensus on individual cases.
  • Expected Values: The ground truth comes from established clinical literature and verified normal ranges based on studies of healthy populations.

4. Adjudication Method for the Test Set

Not applicable for an IVD device. Adjudication methods (e.g., 2+1, 3+1) are used in studies where subjective human interpretation (e.g., image reading) requires consensus building for ground truth establishment. For quantitative IVD assays, the results are numerical and objectively measured; therefore, there's no need for adjudication.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done

No. As explained above, this is an in-vitro diagnostic device, not an imaging AI/ML device involving human readers.

6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done

Yes, in essence, all the analytical performance studies (Method Comparison, Linearity, LoQ, Precision, Interference, Matrix Comparison) are "standalone" in nature for an IVD. The device's performance is evaluated based solely on its ability to accurately and precisely measure CRP concentrations in samples, without any human interpretation of the measurement itself or the involvement of an "algorithm" in the AI/ML sense. The device directly processes the sample and outputs a quantitative result.

7. The Type of Ground Truth Used

The ground truth for this IVD device's performance studies is primarily based on:

  • Reference Method/Predicate Device Comparison: For the method comparison study, the predicate device serves as the reference against which the new device's measurements are compared.
  • Reference Materials and Calibrators: For linearity, LoQ, and precision studies, the ground truth for concentration values is established using certified reference materials and meticulously prepared calibrators with known concentrations.
  • Spiked Samples: For interference studies, known amounts of interfering substances are added to samples, and the true CRP concentration (before interference) serves as the baseline ground truth.
  • Clinical Literature/Established Norms: For "Expected Values," the ground truth is derived from widely accepted clinical ranges and population studies cited in the literature.

8. The Sample Size for the Training Set

This document describes a 510(k) submission for a traditional IVD device, not an AI/ML-based device. Therefore, there is no "training set" in the context of machine learning. The device's performance is based on its chemical and physical principles (latex-enhanced immunoturbidimetric assay) and validated through the analytical studies detailed.

9. How the Ground Truth for the Training Set Was Established

Not applicable, as there is no "training set" for this type of IVD device.

FDA 510(k) Clearance Letter and Documentation

Page 1

U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov

Doc ID # 04017.07.05

April 18, 2025

Kamiya Biomedical Company, LLC
Shawn Kaplan
Diagnostics Product Manager
12779 Gateway Dr S
Tukwila, Washington 98168

Re: K242170
Trade/Device Name: K-ASSAY CRP (Ver.2)
Regulation Number: 21 CFR 866.5270
Regulation Name: C-Reactive Protein Immunological Test System
Regulatory Class: Class II
Product Code: DCK
Dated: March 21, 2025
Received: March 21, 2025

Dear Shawn Kaplan:

We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Page 2

K242170 - Shawn Kaplan
Page 2

Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).

Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801 and Part 809); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.

All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-

Page 3

K242170 - Shawn Kaplan
Page 3

assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Ying Mao -S

Ying Mao, Ph.D.
Branch Chief
Division of Immunology and Hematology Devices
OHT7: Office of In Vitro Diagnostics
Office of Product Evaluation and Quality
Center for Devices and Radiological Health

Enclosure

Page 4

FORM FDA 3881 (8/23)
Page 1 of 1
PSC Publishing Services (301) 443-6740 EF

DEPARTMENT OF HEALTH AND HUMAN SERVICES
Food and Drug Administration

Indications for Use

Form Approved: OMB No. 0910-0120
Expiration Date: 07/31/2026
See PRA Statement below.

510(k) Number (if known)
K242170

Device Name
K-ASSAY CRP (Ver.2)

Indications for Use (Describe)
K-ASSAY CRP (Ver.2) is intended to be used for the quantitative determination of C-reactive protein (CRP) in human serum and plasma (potassium-EDTA or lithium-heparin) by immunoturbidimetric assay. Measurement of CRP aids in the detection and evaluation of infection, tissue injury, inflammatory disorders and associated diseases. FOR IN VITRO DIAGNOSTIC USE.

Type of Use (Select one or both, as applicable)
☒ Prescription Use (Part 21 CFR 801 Subpart D)
☐ Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.
DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services
Food and Drug Administration
Office of Chief Information Officer
Paperwork Reduction Act (PRA) Staff
PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

Page 5

KAMIYA BIOMEDICAL COMPANY
12779 Gateway Drive S, Tukwila, WA 98168 USA
Tel: 206.575.8068
Fax: 206.575.8094

KAMIYA BIOMEDICAL COMPANY K-ASSAY® CRP (Ver.2) Assay 510(k) Summary

510(k) Summary

510(k) Number: K242170

1. Submitter

Kamiya Biomedical Company, LLC
12779 Gateway Drive S
Tukwila, WA 98168 U.S.A.

Contact: Shawn Kaplan
Director, Diagnostics
Email: diagnostics@k-assay.com
PH: 206-575-8068
FAX: 206-575-8094

Date Prepared: April 18, 2025

2. Device

Trade Name: K-ASSAY® CRP (Ver.2)
Common Name: C-Reactive Protein, Antigen, Antiserum, And Control
Classification Name: C-reactive protein immunological test system
Regulation Number: 21 CFR 866.5270
Regulatory Class: Class 2
Review Panel: Immunology
Product Code: DCK

3. Predicate Device

Trade Name: K-ASSAY® CRP (3)
510(k) Number: K023828

4. Device Description

The K-ASSAY® CRP (Ver.2) assay quantifies C-reactive protein based on immunoturbidimetric assay. The reagent uses latex combined with goat polyclonal antibody specific to human CRP.

Page 6

By adsorbing CRP in the sample to the surface of the latex particles and reacting it with the anti-CRP antibody, specific aggregation corresponding to the CRP concentration occurs. Since the absorbance of the reaction changes in proportion to the amount of aggregation, the concentration of CRP in the sample is determined based on the calibration curve prepared using a standard of known CRP concentrations.

The K-ASSAY® CRP (Ver.2) assay can be run using a chemistry analyzer. 6 levels of calibrators from the K-ASSAY® CRP Calibrator (Ver.2) calibrators are used for quantifying the levels of CRP present in the patient's sample.

5. Intended Use / Indications for Use

K-ASSAY® CRP (Ver.2) is intended to be used for the quantitative determination of C-reactive protein (CRP) in human serum and plasma (potassium-EDTA or lithium-heparin) by immunoturbidimetric assay. Measurement of CRP aids in the detection and evaluation of infection, tissue injury, inflammatory disorders and associated diseases. FOR IN VITRO DIAGNOSTIC USE.

6. Comparison With Predicate Device

The table below describes the similarities and differences between the candidate and predicate devices. Substantial equivalence was demonstrated by direct method comparison testing as well as comparison of various assay performance characteristics.

ItemK-ASSAY® CRP (Ver.2) assay (candidate device)K-ASSAY® CRP (3) assay (predicate device)
ClassificationClass 2 (866.5270)Class 2 (866.5270)
Product CodeDCKDCK
Indications for UseK-ASSAY® CRP (Ver.2) is intended to be used for the quantitative determination of C-reactive protein (CRP) in human serum and plasma (potassium-EDTA or lithium-heparin) by immunoturbidimetric assay. Measurement of CRP aids in the detection and evaluation of infection, tissue injury, inflammatory disorders and associated diseases. FOR IN VITRO DIAGNOSTIC USE.K-ASSAY® CRP (3) is intended to be used as a high-sensitive assay for the quantitative determination of CRP in serum and plasma by immunoturbidimetric assay. Measurement of C-Reactive Protein aids in the detection and evaluation of tissue injury, inflammatory disorders, and related diseases.
AssayLatex-enhanced (immuno)turbidimetric assayLatex-enhanced (immuno)turbidimetric assay
Calibration:6 levels (0.0, 10.0, 50.0, 150.0, 300.0 and 400.0 mg/L)Multi-point calibrators E, D and F; 5 levels for each calibrator set
Range:5.0 - 400.0 mg/L0.2 – 480 mg/L

Page 7

PrecisionSingle SiteRepeatability CV = 0.4 to 1.8 %Between-Run CV = 0.0 to 1.4 %Within-Day CV = 0.7 to 1.8 %Between-Day CV = 0.0 to 1.7 %Within-Laboratory CV = 0.8 to 2.4 %Multi-SiteRepeatability CV = 0.5 to 1.5 %Between-Run CV = 0.0 to 1.1 %Between-Day CV = 0.6 to 1.8 %Between-Site CV = 0.9 to 1.8 %Reproducibility CV = 1.1 to 2.5 %Within Run CV = 0.53 to 0.81 %Between Runs CV = 0.39 to 1.88%
InterferencesThe following substances were tested up to the listed concentrations, and no significant interference was observed:Endogenous SubstancesBilirubin C, (Conjugated) 40 mg/dLBilirubin F (Unconjugated) 40 mg/dLCholesterol 300 mg/dLHemoglobin 1,000 mg/dLIntralipid 500 mg/dLRheumatoid Factor 1,000 IU/mLTriglycerides 1,000 mg/dLExogenous SubstancesAcetaminophen 1.5 mMAmoxicillin 400 µmol/LAspirin (Acetylsalicylic Acid) 3.6 mMCephalexin 360 µmol/LFluconazole 480 µmol/LIbuprofen 2.5 mg/dLMethotrexate 1,400 µmol/LPrednisolone 2 µmol/LVitamin C (Ascorbic Acid) 500 mg/LBilirubin C No interference up to 30 mg/dLBilirubin F No interference up to 30 mg/dLHemoglobin No interference up to 500 mg/dLLipid No interference up to 5% IntrafatRheumatoid Factor No interference up to 560 IU/mL
Method Comparisony = K-ASSAY® CRP (Ver.2), x = predicate devicey = 1.005x - 0.002 r = 0.999 n = 175 clinical native serum samples

Page 8

7. Standards and Guidance Documents

The following standards and guidance documents have been used as a basis for the procedures described in this submission:

CLSI Guidelines:

  • CLSI EP05-A3: Evaluation of Precision of Quantitative Measurement Procedures; Approved Guideline – Third Edition. October 2014.
  • CLSI EP06-Ed2: Evaluation of Linearity of Quantitative Measurement Procedures – Second Edition. November 2020.
  • CLSI EP07: Interference Testing in Clinical Chemistry – Third Edition. April 2018.
  • CLSI EP09c: Measurement Procedure Comparison and Bias Estimation Using Patient Samples – Third Edition. June 2018.
  • CLSI EP17-A2: Evaluation of Detection Capability for Clinical Laboratory Measurement Procedures; Approved Guideline – Second Edition. June 2012.
  • CLSI EP25-Ed2: Evaluation of Stability of In Vitro Medical Laboratory Test Reagents - Second Edition. April 2023.
  • CLSI EP28-A3c: Defining, Establishing, and Verifying Reference Intervals in the Clinical Laboratory; Approved Guideline - Third Edition. October 2010.

FDA Guidance:

  • Guidance for Industry and FDA Staff: Review Criteria for Assessment of C-Reactive Protein (CRP), High Sensitivity C-Reactive Protein (hsCRP) and Cardiac C-Reactive Protein (cCRP) Assays. September 2005.

Page 9

8. Performance Data

Testing was performed to confirm assay performance. A summary of results follows.

Method Comparison

Testing was performed following CLSI guideline, EP09c 3rd edition. Samples ranging from 5.4 - 409.5 mg/L were tested using the K-ASSAY® CRP (Ver.2) assay on an Abbott Architect c8000 analyzer and compared to results from a predicate assay also using the same analyzer. Analysis yielded the following results:

y = K-ASSAY® CRP (Ver.2), x = predicate device
y = 1.005x – 0.002 r = 0.999 n = 175

Linearity

Testing was performed on an Abbott Architect c8000 analyzer following CLSI guideline EP06, 2nd edition. Testing the linear range of 4.6 - 441.2 mg/L yielded the following regression equation:
y = 0.9709x - 1.095, r = 0.999

Limit of Quantitation (LoQ)

The detection limits of the assay were evaluated based on CLSI guideline EP17-A2. The LoQ was defined as the concentration achieving within-laboratory precision of ≤ 20% CV across all 3 reagent lots, with the highest observed LoQ being 1.0 mg/L. However, the claimed LoQ for the K-ASSAY® CRP (Ver.2) is 5.0 mg/L.

Assay Range

As confirmed by linearity and LoQ testing, the assay range has been set as: 5 - 400 mg/L.

Page 10

Matrix Comparison

42 donor samples were each collected into 3 collection tubes to assess the effects of anticoagulants. Values of serum samples ranging from 5.1 - 399.1 mg/L were compared with values for plasma samples collected using Potassium-EDTA and Lithium-Heparin, yielding the following results by regression analysis:

K2-EDTA Plasma (y) vs Serum (x):
y = 1.007x - 0.141, r = 0.999

Li-Heparin Plasma (y) vs Serum (x):
y = 0.972x + 0.074, r = 0.999

Precision

Estimates of imprecision, based on CLSI recommendations, are consistent with typical performance.

Single-site precision was evaluated using 3 unique lots of K-ASSAY® CRP (Ver.2) on an Abbott Architect c8000 analyzer. 2 commercial controls and 5 human serum samples were analyzed in 2 runs per day, with 2 replicates per run, for 20 days.

Multisite precision was evaluated using 1 lot of K-ASSAY® CRP (Ver.2) on 3 Abbott Architect c8000 analyzers. 2 commercial controls and 5 human serum samples were analyzed in 1 run per day, with 5 replicates per run, for 5 days.

Data was evaluated using the CLSI approved guideline, EP05-A3 3rd edition, and representative comparative performance data appears in the tables below.

Single-Site Precision (Combined Data From 3 Lots)

SampleMean (mg/L)NWithin-RunBetween-RunWithin-LotBetween-LotTotal
SDCV %SDCV %SDCV %SDCV %SDCV %
Control Level 19.82400.21.7%0.00.4%0.21.7%0.00.1%0.21.7%
Control Level 231.72400.31.1%0.20.5%0.41.1%0.00.0%0.41.1%
Sample 15.42400.12.2%0.00.0%0.11.9%0.11.2%0.12.3%
Sample 211.12400.11.2%0.10.5%0.11.3%0.00.3%0.11.3%
Sample 345.02400.40.8%0.30.6%0.40.9%0.10.1%0.40.9%
Sample 4196.52401.80.9%0.80.4%1.80.9%0.80.4%1.91.0%
Sample 5330.72404.61.4%1.30.4%4.71.4%0.00.0%4.71.4%

Page 11

Multisite Precision (Combined Data From 3 Analyzers)

SampleMean (mg/L)NRepeatabilityBetween-RunBetween-DayBetween-SiteReproducibility
SDCV %SDCV %SDCV %SDCV %SDCV %
Control Level 110.0750.11.1%0.00.4%0.11.1%0.11.3%0.21.7%
Control Level 231.4750.20.7%0.31.0%0.41.2%0.31.1%0.51.6%
Sample 15.4750.11.2%0.00.0%0.11.2%0.11.1%0.11.6%
Sample 211.1750.11.0%0.00.0%0.11.0%0.21.8%0.22.0%
Sample 345.0750.20.5%0.20.4%0.30.6%0.40.9%0.51.1%
Sample 4200.6751.30.7%1.70.8%2.11.1%1.91.0%2.91.4%
Sample 5344.5755.11.5%3.71.1%6.31.8%5.81.7%8.62.5%

Interference

Testing was performed on an Abbott Architect c8000 analyzer according to the CLSI EP07-3rd edition guideline with the following results.

Endogenous and Exogenous Substances

Results of studies conducted to evaluate the susceptibility of the method to interference from common or known endogenous and exogenous substances that could interfere with the assay are listed below. The criterion for no significant interference is recovery within 10% of the initial value (sample containing no interferent).

The following substances were tested up to the listed concentrations, and no significant interference was observed:

Endogenous Substances

  • Bilirubin C, (Conjugated) 40 mg/dL
  • Bilirubin F (Unconjugated) 40 mg/dL
  • Cholesterol 300 mg/dL
  • Hemoglobin 1,000 mg/dL
  • Intralipid 500 mg/dL
  • Rheumatoid Factor 1,000 IU/mL
  • Triglycerides 1,000 mg/dL

Exogenous Substances

  • Acetaminophen 1.5 mM
  • Amoxicillin 400 µmol/L
  • Aspirin (Acetylsalicylic Acid) 3.6 mM
  • Cephalexin 360 µmol/L
  • Fluconazole 480 µmol/L
  • Ibuprofen 2.5 mg/dL

Page 12

  • Methotrexate 1,400 µmol/L
  • Prednisolone 2 µmol/L
  • Vitamin C (Ascorbic Acid) 500 mg/L

Expected Values

For evaluation of infection, tissue injury, and inflammation disorders, the following suggested ranges from widely accepted literature references are used.¹,²,³

  • ≤ 5 mg/L indicates apparently healthy individuals
  • ≥ 10 mg/L is clinically significant

As a verification, 168 normal serum samples taken from healthy individuals in the U.S. were tested with the K-ASSAY® CRP (Ver.2) assay based on CLSI C28-A3. All samples were run on an Abbott Architect c8000 analyzer. The results showed that 4 out of the 168 samples were >5.0 mg/L (2.4%).

  1. U.S. FDA, Review Criteria for Assessment of C-Reactive Protein (CRP), High Sensitivity C-Reaction Protein (hsCRP) and Cardiac C-Reaction Protein (cCRP) Assays, rev. Sep 22, 2005.

  2. Dati F., et al. Consensus of a group of professional societies and diagnostic companies on guidelines for interim reference ranges for 14 proteins in serum based on the standardization against the IFCC/ BCR/ CAP reference material (CRM 470). European Journal of Clinical Chemistry and Clinical Biochemistry. 34(6):517-520, 1996.

  3. Rifai, N. and Ridker, P.M. Population distributions of C-reactive protein in apparently healthy men and women in the United States: Implication for clinical interpretation. Clin Chem, 49(4):666-669, 2003.

9. Conclusion

The performance of the K-ASSAY® CRP (Ver.2) assay is substantially equivalent to other similar assays on the market.

§ 866.5270 C-reactive protein immunological test system.

(a)
Identification. A C-reactive protein immunological test system is a device that consists of the reagents used to measure by immunochemical techniques the C-reactive protein in serum and other body fluids. Measurement of C-reactive protein aids in evaluation of the amount of injury to body tissues.(b)
Classification. Class II (performance standards).