Search Results
Found 2 results
510(k) Data Aggregation
(165 days)
The Osteon Precision Milled Suprastructure is indicated for attachments in the treatment of partially or fully edentulous jaws for the purpose of restoring function. The Osteon Precision Milled Suprastructures are intended for attachment to a minimum of two (2) abutments.
The Osteon Milled Suprastructure is indicated for compatibility with the following abutment systems:
- · Astra Tech Implant System® Multi Base Abutment EV, 4.8mm, max 30°
- · BioHorizons Multi Unit Abutment, 4.8mm, max 30°
- · CONELOG® Implant System
- · Biomet 3 i Multi Unit Abutments, 4.8mm, max 30°
- · TSXTM Implants
- · Tapered Screw-Vent Implant
- · DESS Dental Multi Unit Abutments, 3.4-5.7 mm, 0°
- · 3i OSSEOTITE®
- · Astra Tech OsseoSpeed™
- · Neodent Grand Morse
- NobelReplace® Trilobe
- · NobelReplace® Conical
- · Nobel Brånemark System®
- · Straumann BLX Implants
- · DESS Dental Multi Unit Abutments. Angled. 3.4-6.5 mm, max 30°
- · NobelActive® NobelParallel Conical
- Straumann® Bone Level
- · Zimmer Screw Vent® and Tapered Screw-Vent®
- · Dentium SuperLine® Abutments, 4.5-5.5 mm, max30°
- · Genesis ACTIVE™ Multi-Unit Abutments, 4.8mm, max 30°
- · Implant Direct GPS® Angled Abutment, 5.0mm, max 30°
- KDG Abutments, 4.8mm, max 30°
- · Keystone Multi Unit Abutment, 4.8mm, 0°
- · Medentika Multi Unit Abutments, 4.8mm, max 30°
- · EV Series Dentsply® Implants Astratech Osseospeed®
- · F Series Nobel Biocare NobelActive® NobelReplace® Conical
- · H Series Biomet 3i Certain®
- L Series Straumann Bone Level
- · N Series Straumann Soft tissue Level
- · R Series Zimmer Dental Tapered Screwvent®
- Medentika Multi Unit Abutments, 4.8mm, 0°
- E Series Nobel Biocare Replace™ Select
- · I Series Biomet 3i Osseotite®
- K Series Nobel Biocare™ Branemark
- · S Series Astra Tech OsseoSpeedTM
- · T Series Dentsply Friadent® Frialit/Xive®
- · MegaGen Multi Unit Abutments, 4.8mm, max 30°
- · Xpeed® AnyRidge® Internal Implant System
- AnyOne® Internal Implant System
- AnyRidge® Octa 1 Implant System
- AnyOne® External Implant System
- AnyRidge® Octa 1 Implant System
- AnyOne® Internal Implant System
- Rescue Internal Implant System
- MIS Multi-unit Abutments, 4.8mm
- · C1 Conical Connection Implant System, max 30°
- · V3 Conical Connection Implant System, max 30°
- · Internal Hex Implant System, max 30°
- · Conical Connection, max 30°
- · Neodent GM Mini Conical Abutment, 4.8 mm, max 30°
- · Nobel Biocare™ Brånemark Multi Unit Abutment, 4.8 mm, max 17°
- · Nobel Biocare™ Multi Unit Abutment Plus, 4.8 mm, max 30°
- · Nobel Biocare™ Multi Unit Abutment, 4.8 mm, max 30°
- Nobel Biocare™ Multi Unit Abutments for Straumann and Astra Tech System, 4.8 mm, max 30°
- Nobel Biocare™ Multi Unit Abutments for Astra Tech, Camlog and Ankylos Implant Systems, 4.8 mm, max 30°
- · Nobel Biocare Xeal Abutments, 4.8 mm, max 30°
- · OSSTEM Multi Unit Abutment, 4.8mm, max 30°
- SS SA Fixture Implants
- SA Implant System
- · ET US SSS Prosthetic System
- · Paltop Multi Unit Abutment, 5.0 mm, max 17°
- · Southern Compact Conical Abutments, 4.8 mm
- MAX Implant System, 0°
- · Provata Implant System, max 30°
- · Deep Conical (DC) Implants, 0°
- · Piccolo Implants, 0°
- · External Hex Implants, max 30°
- Straumann® BLX Screw Retained Abutment, 4.6 mm, max 30°
- · Straumann® Screw Retained Abutment, 4.6 mm, max 30°
- · Zimmer Angled Tapered Abutments, 4.5 mm, max 30°
The Osteon Precision Milled Suprastructures (also referred to as superstructures) are metallic dental restorative devices that are intended to be attached by screw retention to dental implant abutments to aid in the treatment of partial and totally edentulous patients for the purpose of restoring chewing function. These suprastructures attach to dental implant abutments using the prosthetic screws from the original equipment manufacturers (OEM) and are used to support the final multi-unit restoration.
The Osteon Precision Milled Suprastructure is designed for an individual patient from scans of the patient's dental impression. The suprastructure is manufactured with the aid of Computer Aided Design (CAD) and Computer Aided Manufacturing (CAM) technology. All CAD/CAM fabrication is performed by Osteon Medical.
Osteon Precision Milled Suprastructures facilitate the attachment of both removable and fixed dental prosthesis and hence are categorized as Type A and Type B.
- Type A: Intended to act as a supporting structure to facilitate the attachment of a removable dental prosthesis and include Primary Bar and Nexus Removable Bar.
- . Type B: Intended to act as a supporting structure to facilitate the attachment of a fixed dental prosthesis and include Melbourne Bar and Nexus Fixed Bar (including Nexus Hybrid, Nexus Bridge, Micro Nexus, and Nexus Wraparound).
The provided FDA 510(k) summary (K233083) describes the acceptance criteria and study for the Osteon Precision Milled Suprastructure.
1. Table of Acceptance Criteria and Reported Device Performance
This submission is a 510(k) for an expansion of an existing device (K221019) to add compatibility with new OEM abutment systems and introduce three new Type B Nexus Fixed Bars. The acceptance criteria are primarily based on demonstrating substantial equivalence to the predicate devices in terms of intended use, technological characteristics, and material composition.
The tables below synthesize the design specifications (acceptance criteria) and the device's performance (reported as meeting these specifications) by demonstrating substantial equivalence to predicate devices.
Table 1: Comparison of General Characteristics and Materials (summarized for clarity)
Characteristic | Acceptance Criteria (Predicate) | Reported Device Performance (Subject Device) |
---|---|---|
Intended Use | Dental implant abutment for restoring chewing function in partially or fully edentulous jaws, attached to a minimum of two abutments. Specific compatible OEM abutment systems listed. | Substantially Equivalent. The subject device shares the same intended use. The primary difference is an expanded list of compatible OEM abutment systems. |
Device Material | Titanium alloy Ti-6Al-4V (ASTM F136) | Meets/Substantially Equivalent. Manufactured from titanium alloy conforming to ASTM F136. |
Design/Technology | CAD/CAM milling from single milling blanks | Meets/Substantially Equivalent. CAD/CAM milling from single milling blanks. |
Design/Construction | Patient specific/machined | Meets/Substantially Equivalent. Patient specific/machined. |
Sterility | Supplied Nonsterile | Meets/Substantially Equivalent. Supplied Nonsterile (to be sterilized by end-user). |
Prescription/OTC | Prescription only | Meets/Substantially Equivalent. Prescription only. |
Recommended Cleaning & Maintenance | Proper oral hygiene | Meets/Substantially Equivalent. Proper oral hygiene. |
Table 2: Design Specifications (Acceptance Criteria vs. Reported Performance)
Design Parameters | Type A (Removable Prosthesis) Acceptance Criteria (Predicate) | Type B (Fixed Prosthesis) Acceptance Criteria (Predicate) | Type A (Removable Prosthesis) Reported Performance (Subject Device) | Type B (Fixed Prosthesis) Reported Performance (Subject Device) |
---|---|---|---|---|
Total Cylinders | 2-10 | 2-10 | 2-10 | 2-10 |
Suprastructure Span Between Cylinders (mm) | 1-30 mm | 1-30 mm | 1-30 mm | 1-30 mm |
Suprastructure Height (mm) | 3-12 mm | 3*-22 mm | 3-12 mm | 3*-22 mm |
Suprastructure Width (mm) | 3.4-12 mm | 3.4**-12 mm | 3.4-12 mm | 3.4**-12 mm |
Distal Cantilever Section (mm) | 0-15 mm | 0-15 mm | 0-15 mm | 0-15 mm |
Cylinder Height (mm) | 0-4.6 mm | 0-4.6 mm | 0-4.6 mm | 0-4.6 mm |
Cylinder Diameter (mm) | 4.5-8 mm | 4.5-8 mm | 4.5-8 mm | 4.5-8 mm |
*Note for Suprastructure Height: The Minimum Suprastructure Height for the Micro Nexus Bar (a new Type B bar in the subject device) is 2.5 mm, which is within the stated range of 3*-22 mm, where * indicates variability for this new bar.
**Note for Suprastructure Width: The Minimum Suprastructure Width for the Micro Nexus Bar (a new Type B bar in the subject device) is 2.5 mm, which is within the stated range of 3.4**-12 mm, where ** indicates variability for this new bar.
2. Sample size used for the test set and the data provenance
The submission is a 510(k) for an expansion of an existing device (K221019). The "study" proving the device meets acceptance criteria primarily relies on non-clinical data and demonstration of substantial equivalence to predicate devices.
- Test set/Sample size: The document does not describe a "test set" in the traditional sense of a clinical or imaging study with a defined sample size for statistical analysis. Instead, it relies on:
- Reverse engineering analysis of OEM abutments and OEM abutment screws: This was done to confirm compatibility. The sample size for this analysis is not explicitly stated but would involve the specific OEM abutments listed for compatibility with the subject device.
- Fatigue testing: The document states, "Fatigue testing was not performed since the Subject devices are abutment-borne and are not intended to compensate for angulation in excess of the maximum angulation of OEM angled abutments..." This indicates that engineering analysis and design specifications, rather than physical fatigue testing, were used to demonstrate performance in this regard.
- Data provenance: Not directly applicable as there isn't a "test set" from patients. The data relies on engineering analysis, existing predicate device data, and compliance with standards.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts
Not applicable. This is not a study requiring expert readers to establish ground truth. Substantial equivalence claims are based on engineering, material, and design comparisons to legally marketed devices.
4. Adjudication method (e.g. 2+1, 3+1, none) for the test set
Not applicable. There is no "test set" or diagnostic performance study that would require an adjudication method.
5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
Not applicable. This device is a physical dental suprastructure, not an AI-powered diagnostic tool.
6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done
Not applicable. This device is a physical dental suprastructure, not an algorithm.
7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)
The "ground truth" equivalent for this submission is the established safety and effectiveness of the primary predicate device (K221019) and various reference devices, supported by:
- Compliance with recognized standards: Biocompatibility (ISO 10993-5, ISO 10993-12), sterilization validation (ANSI/AAMI/ISO 17665-1, ANSI/AAMI/ISO 17665-2, ANSI/AAMI/ISO 14937).
- Material specifications: Conformance to ASTM F136 for titanium alloy.
- Engineering analysis and design specifications: Comparison of design features to predicate devices and confirmation of compatibility through reverse engineering.
8. The sample size for the training set
Not applicable. This is not a machine learning or AI device that requires a training set.
9. How the ground truth for the training set was established
Not applicable. As above, there is no training set for this device.
Ask a specific question about this device
(145 days)
The OSSTEM Abutment system is intended for use with a dental implant to provide support for prosthetic restorations such as crowns, bridges, or overdentures.
TS Abutment System is made of titanium alloy. TS Abutment System is intended for use with a dental implant to provide support for prosthetic restorations such as crowns, bridges, or overdentures. TS Abutment System is similar to other commercially available products based on the intended use, technology used, material composition employed and performance characteristics.
The provided document is a 510(k) summary for the Osstem Implant Co., Ltd. TS Abutment System. It describes the device and claims substantial equivalence to previously cleared predicate devices. The document does not describe a study that uses acceptance criteria in the context of device performance metrics like accuracy, sensitivity, or specificity, nor does it involve expert reviews or clinical trials for such metrics.
Instead, the submission focuses on demonstrating substantial equivalence through comparisons of technical characteristics and performance testing (bench testing, biocompatibility, sterilization validation, etc.) against existing predicate devices and established standards.
Therefore, most of the requested information regarding acceptance criteria for device performance (accuracy, sensitivity), sample sizes for test sets, data provenance, number and qualifications of experts, adjudication methods, MRMC studies, standalone performance, and ground truth establishment would not be applicable to this type of regulatory submission. This submission does not pertain to an AI/ML powered device, but rather a physical dental implant component.
However, I can extract information related to the technical comparisons that demonstrate substantial equivalence.
1. A table of acceptance criteria and the reported device performance
The document does not present explicit acceptance criteria in terms of numerical performance metrics (e.g., specific thresholds for accuracy, sensitivity). Instead, the acceptance is based on demonstrating that the subject device's characteristics and performance are "substantially equivalent" to predicate devices, often by adhering to established standards or by justifying that any differences do not raise new questions of safety or effectiveness. The reported "performance" for most components is implicitly that they function equivalently to their predicates and meet relevant mechanical and biological standards.
Here's a summary of the non-clinical performance and "acceptance criteria" through comparative analysis:
Aspect | Acceptance Criteria (Implicit) | Reported Device Performance |
---|---|---|
Mechanical Properties (Fatigue) | The fatigue performance of the proposed device is considered to be equal to or greater than that of the reference/primary predicate device. This is often justified by: |
- Same design, function, connection, and platform as predicate.
- Adherence to ISO 14801 standard (maintaining 11mm distance between embedding plane and hemispherical loading member).
- Differences (e.g., added lengths or diameters) not creating a new worst-case scenario. | Rigid Abutment: No additional fatigue testing conducted, as the proposed device has the same design, function, connection, platform, and moment arm as the reference/predicate, and added lengths are within acceptable ranges or previously cleared for similar components.
Transfer Abutment: No additional fatigue testing conducted for similar reasons, justifying that changes in connection features were evaluated and not a new worst case, and that the device has the same moment arm.
Angled Abutment: No additional fatigue testing conducted, as differences in feature (guide for connecting with implant body) do not change the fundamental design or platform and the proposed device's fatigue is considered equal to or greater than the predicate.
FreeForm ST Abutment: No additional fatigue testing conducted, as added dimensions are within the range of the reference device and it maintains the same moment arm.
Temporary Abutment: No additional performance test needed as it's for temporary use, and the proposed gingiva height has primary predicate clearance.
Multi Abutment: No additional fatigue testing conducted, as added dimensions (gingival height) are covered by primary predicates and it maintains the same moment arm.
Ti Cylinder Screw: No fatigue testing considered necessary as it's a component using with cylinder, and material biocompatibility is established.
Convertible Abutment: No additional performance tests as difference is sterilization process. Its mechanical performance is expected to be equivalent to predicate. |
| Biocompatibility | The device materials must have acceptable biocompatibility, typically demonstrated by using materials already cleared in predicate devices or by following ISO 10993-1. | The TS Abutment System uses the same materials (e.g., Ti-6Al-4V (ASTM F136), Titanium Gr.3 (ASTM F67), Titanium Gr.4 (ASTM F67)), manufacturer, manufacturing process, and surface treatment as primary predicate and reference devices. Therefore, no additional biocompatibility testing was conducted. |
| Sterilization Validation | The sterilization process must be validated to ensure sterility, or justified if changes from predicates do not create a new worst-case scenario. For sterile devices, shelf-life must be validated. | For those components provided sterile, the validation of the gamma irradiation process was previously conducted for the predicate device. Changes in dimensions for the subject device do not create a new worst-case scenario for sterilization, thus no additional validation was required. Shelf-life for sterile components (e.g., Transfer Abutment, Angled Abutment, Multi Abutment, Convertible Abutment, Port Abutment, Stud Abutment, Healing Abutment) is established as 8 years. For non-sterile devices made of titanium, no specific shelf-life is considered for the material itself due to its known stability. |
| Pyrogen Test | Absence of pyrogens to meet established standards (e.g., ISO 10993-11:2006 and USP). | Bacterial Endotoxin Test Report on implants was referenced from K161604, indicating previous compliance. |
| MR Compatibility | Evaluation of magnetically induced displacement force and torque based on FDA guidance "Testing and Labeling Medical Devices for Safety in the Magnetic Resonance (MR) Environment." | Non-clinical worst-case MRI review was performed using scientific rationale and published literature (e.g., Woods et al., 2019), addressing parameters for magnetically induced displacement force and torque for all compatible implant bodies, abutments, and fixation screws and material composition. |
2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)
- Sample Size for Test Set: Not applicable. This submission relies on comparisons to predicate devices and existing standards, not a new test set for performance evaluation in the clinical sense (e.g., diagnostic accuracy).
- Data Provenance: The document references previously cleared 510(k) submissions (e.g., K182091, K161689, K221684, K163634, K161604, K120847) from Osstem Implant Co., Ltd. and one from Southern Implants (Pty) Ltd. The specific country of origin for the data from these predicate clearances is not detailed in this summary, but the manufacturer of the subject device is based in the Republic of Korea. The data is retrospective, as it refers to performance data and justifications from prior clearances.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)
Not applicable. No "ground truth" as typically understood in AI/ML diagnostic performance studies was established or used for this device's submission. The evaluation is based on engineering principles, materials science, and conformity to recognized standards.
4. Adjudication method (e.g. 2+1, 3+1, none) for the test set
Not applicable. No adjudication method for a test set was used.
5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
Not applicable. This is not an AI-powered device.
6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done
Not applicable. This is not an AI-powered device.
7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)
Not applicable. No "ground truth" of this nature was used. The ground truth, in a regulatory sense for this type of device, is conformity to established material specifications, manufacturing processes, and recognized performance standards (like ISO 14801 for fatigue testing).
8. The sample size for the training set
Not applicable. This is not an AI-powered device.
9. How the ground truth for the training set was established
Not applicable. This is not an AI-powered device.
Ask a specific question about this device
Page 1 of 1